Professional Documents
Culture Documents
TECHNICAL MANUAL
JOLLY Plus DR Series TECHNICAL MANUAL
SUMMARY
1.6 DIMENSION............................................................................................................................. 20
1.9 SAFETY...................................................................................................................................... 21
1.9.1 INTRODUCTION ....................................................................................................................... 21
1.9.2 MECHANICAL SAFETY DEVICES ............................................................................................ 23
1.9.3 ELECTRICAL SAFETY DEVICES .............................................................................................. 23
1.9.4 PROTECTION FROM RADIATION .......................................................................................... 24
1.9.5 RESIDUAL RISKS ..................................................................................................................... 24
The Technical manual should be read before starting using the equipment.
It is recommended to follow strictly the instructions of the present in order to
avoid any illegitimate use of the unit. The unit is supposed to be used by
qualified and authorized people only.
The installation and maintenance must be carried out by people authorized by
the manufacturer.
The purchaser of the equipment does not buy the system software but only the
right to use the software licence. this is not an exclusive right and it is not
transferable.
0051
Manufacturer:
REVISIONS CHRONOLOGY
CHAPTER 1
EQUIPMENT DESCRIPTION
The following symbols are available in the technical manual, in the user manual and on the labels
applied on the equipment:
Symbol Description
CE compliance mark
Operating Instruction
Serial no.
Catalogue no.
Date of manufacture
Manufacturer
This way up
Fragile
Alternating current
Protective earth
Earth
'ON' (power)
'OFF' (power)
X-ray tube
The mobile equipment is suitable for many medical applications and is supposed to be used by
qualified people only, duly trained and informed about the risks deriving from Ionogenic radiations.
This system is not included in the list of devices requiring continuity of operation. This system does
not come into contact with the patient, only some parts of the device may accidentally touch the
patient. The contact with the operator is contemplated during the equipment operation. The contact
with the patient is non-invasive. The contact with the operator is required for reasons depending on
the equipment utilization.
The equipment has been specifically designed for radiological examinations and diagnostic
investigations carried out in Hospitals and Medical Centres. It is suitable for operation in:
- Respiratory System investigations
- Skeleton investigations
- Sport Medicine
- Paediatrics Orthopaedics
- First-Aid Stations
- Plaster Room
- Operating Theatre
- Traumatology
The typical range of using estimate is from 20 exposure per day to 150 exposure per day (1 day
= 24 hours period).
For paediatric applications are provided specific additional filters, to obtain a filtering
not less than 0.1 mm Cu or 3.5 mm Al
The average life of this equipment depends on the life of the X-ray tube installed in the
monobloc. The compliance with the instructions of this Manual will make it possible to get the
maximum life duration of the X-ray tube. It is not always possible to know in advance when the X-ray
tube’s operation life is coming to an end. This component could even stop operating during the
examination of a patient. For this reason, it is recommended to check the focal spot yearly, in order to
find if there is any degradation of the anode, indicated by the increase of the focal spot dimension and
consequent worse image resolution.
The main causes of early wear and damages to the X-ray tube are:
- Filament prolonged heating
- Discharges in the tube
- Tube over current
- Cold anode overcharge, caused by excessive loads after a long period of inactivity, with
consequent cratering of the track.
Here below how to limit untimely wear and damages to the monobloc:
1. The X-ray tube preheating cycle avoids the overheating since the tube reaches the
operating temperature slowly.
IF A HIGH VALUES EXPOSURE IS STARTED WHEN THE X-RAY TUBE IS STILL COLD, THE
ANODE COULD BE IRREPARABLY DAMAGED.
2. If a high values exposure is started when the device has not been used for a long time,
the X-ray tube could be irreparably damaged.
3. The X-ray tube overcharge is avoided by the software control on the preset kVp and mAs
values. The system prevents the exposure if these values exceed the maximum values
provided for the X-ray tube.
4. Filament temperature preparation. It is recommended to effect the preparation only
when the patient is positioned for the examination.
Description Value
X-ray exposure control • Remote control by wire
- Hand switch with spiral cable extendable up to 3.8 meter length
Pre-selected technique
• 2-point technique (kV, mA)
radiography (APR)
Free technique radiography • 2-point technique (kV, mA)
• 3-point technique (kV, mA, ms)
Safety and protections • Capacitors voltage
• Inverter over current
• mA min and max safety device
• Monobloc working temperature
• X-ray filament safety device
• X-ray over load safety device
• X-ray tube anode rotation
• X-ray tube HU (thermal unit)
Description Values
Model 4 kW 4 kW (40 mA - 100 kV @ 100 ms)
Max power Model 20 kW 20 kW (200 mA - 100 kV @ 100 ms)
Model 30 kW 30 kW (300 mA - 100 kV @ 100 ms)
H.T. control H. F. Inverter – Frequency 40 kHz
Max ripple ≤ 2% @ maximum power
kV value increasing time ≤ 3 ms
1 ms ÷ 6300 ms (35 steps)
Exposure time (1 • 2 • 3 • 6 • 8 • 10 • 12 • 15 • 20 • 25 • 30 • 40 • 50 • 63 • 80 • 100 • 125 • 150
(for all the model) • 200 • 250 • 300 • 350 • 400 • 500 • 630 • 800 • 1000 • 1250 • 1500 • 2000 • 2500
• 3000 • 4000 • 5000 • 6300)
Parameter Accuracy
kV ± 5% of nominal value
mA ± 10% of nominal value
mAs ± 10% ±0.2 of nominal value
Time ± 5% of nominal value
Linearity e Reproducibility
Standard deviation of the average, (average 10 exposures) < 10%
Performance
Feature Value
Width 680 mm
Length 1095 mm
Height with locked arm 1580 mm
Focus-floor distance - variable from 450 to 2000 mm
Monobloc rotation in the yoke -45° / +167°
Monobloc yoke support rotation ± 90°
Collimator rotation ± 90°
Antistatic pivoting wheels
80 mm
diameter
Rear wheels diameter 250 mm
Weight - net ≅ 185 Kg
Unit displacement Manual
Radiographic cassettes holder Nr. 1 Flat panel detector format 35x43 cm (14x17 inches)
1.3.6 COLLIMATOR
ADDITIONAL INFORMATION CAN BE FOUND IN THE COLLIMATOR MANUAL, SUPPLIED WITH THE
EQUIPMENT.
1.3.7 MONOBLOC
11
1
16
2
9
3
10
4
8
18
17
12
13
14
15
Legend
1. Medical Station 2. Transport handles and parking brakes control
3. Supply cable support 4. Flat Panel holder
5. Bracket to lean the unit for overcoming little 6. Rear wheels diameter 250 mm
obstacles
7. Pivoting front wheels 8. Column
9. Arm 10. Arm retainer
11. Arm handles 12. Monobloc’s covers
13. Goniometer 14. X-Ray collimator
15. Handgrip 16. Hand switch for radiographic exposure
17. Automatic circuit breaker (Power ON/OFF) 18. Flat Panel cable’s support
In addition the operator can interface the Medical Station through two external sockets,
respectively type USB and Ethernet, placed on the Medical Station’s support.
1.5 LABEL
The following figure illustrates the position of the equipment’s adhesive label.
Here below you will find the labels as an example.
The monobloc label is placed inside the covers, as illustrated in the following picture.
or
In the picture below are shown the positions of collimator and DAP labels.
1.6 DIMENSION
1.8 CLASSIFICATION
1.9 SAFETY
1.9.1 INTRODUCTION
As a consequence, these instructions are to be carefully read and understood before putting the
equipment at work.
The X-ray equipment must only be used by skilled, qualified personnel, suitably trained and
informed with regard to protection against radiation and the use of X-ray equipment.
The manufacturer is ready to provide the after-sale servicing and, if required, to take care of the
installation.
The end user is responsible for ensuring compliance with applicable standards.
Safety devices and circuits must not be moved, modified or excluded. The plug on the
equipment's power supply cable must only be inserted by authorised personnel and only when the
equipment is to be used.
The equipment must also not be used if any of the warning or alarm devices malfunctions. If the
equipment is used along with other apparatus, components or modules which could be incompatible,
it must be ensured that this poses no risk to patients or operators. In this event consult BMI
Biomedical International s.r.l.
As with any technical apparatus, this X-ray equipment must be used appropriately and be
checked and serviced regularly in accordance with instructions in the Maintenance chapter in this
manual.
BMI BIOMEDICAL INTERNATIONAL s.r.l. can only be liable for the safety of its products if
maintenance, repairs and/or modifications are carried out by the manufacturer or third parties
authorised by BMI Biomedical International in writing.
THE JURIDICAL BINDING VERSION OF THIS DOCUMENT IS THE ITALIAN LANGUAGE ONE.
ANY TRANSLATION, EVEN IF VERIFIED AND APPROVED BY THE MANUFACTURER, IS ONLY
SUPPLIED TO FACILITATE THE USER.
The equipment designed and manufactured by BMI Biomedical International are equipped with
advanced radiological systems to safeguard safety and environment.
The equipment are not dangerous to people and environment if used in the way described in this
manual.
The equipment is sold with the following clause: the manufacturer, its agents and representatives
decline all responsibility for any damage or injury deriving from the wrong utilization of this equipment
or due to lack of maintenance or to the utilization of not original spare parts.
Regarding the disposal of substances like the dielectric oil contained in the monobloc, the anti-X
protective lead, the electronic boards and the electronic components, the provisions of the law in force
both on the national territory and on the territory where the equipment is installed have to be
complied with.
BMI is a leading company in its field and constantly upgrades the equipment according to the
progressive technological knowledge.
Since this equipment emits X-rays, it must be exclusively used complying with the safety
instructions included in this manual and must not be used for other purposes than the diagnostic
ones.
The incorrect utilization of the equipment can cause temporary or permanent injuries.
AS THIS EQUIPMENT PRODUCES X-RAYS, IT MUST ONLY BE USED IN ACCORDANCE WITH THE
SAFETY INSTRUCTIONS CONTAINED IN THIS MANUAL AND IN COMPLIANCE WITH EXISTING
LEGISLATION IN THE COUNTRY WHERE IT IS INSTALLED.
THIS X-RAY DIAGNOSTIC SYSTEM MUST NOT BE USED FOR PURPOSES OTHER THAN THOSE
OUTLINED IN THIS MANUAL.
AFTER THE EQUIPMENT IS INSTALLED, COMPLIANCE WITH ESSENTIAL REQUIREMENTS
STIPULATED BY DIRECTIVE 93/42/EEC MUST BE VERIFIED, AND RESPONSIBILITY ACCEPTED
FOR COMPLIANCE WITH APPLICABLE STANDARDS BOTH IN TERMS OF INITIAL START-UP AND
SUBSEQUENT USE OF THE EQUIPMENT.
The buyer is supposed to use only qualified people (specialized engineers of the manufacturer or
equivalent personnel of a servicing company) to carry out any maintenance intervention on the
equipment. The connections among the different parts of the device, the calibration and the testing
must be carried out by qualified people, authorized by the manufacturer.
Only duly trained and qualified people have access to the internal parts of this device or to the
disassembled parts.
WARNING: TO AVOID THE RISK OF ELECTRIC SHOCK, THIS EQUIPMENT MUST ONLY BE
CONNECTED TO A SUPPLY MAINS WITH PROTECTIVE EARTH.
DANGER OF ELECTRIC SHOCK. ALWAYS CHECK THE CONDITION OF THE DIFFERENTIAL SWITCH
AND OF THE EQUIPMENT SUPPLY CABLE.
THE SYSTEM MUST NOT BE USED IN THE PRESENCE OF ANY ELECTRICAL, MECHANICAL OR
RADIOLOGICAL FAULTS.
THE EQUIPMENT MUST ALSO NOT BE USED IF ANY OF THE WARNING OR ALARM DEVICES
MALFUNCTIONS.
THIS DEVICE IS CONNECTED WITH A PROTECTIVE GROUND DURING OPERATION. FOR THE
PATIENT’S AND THE OPERATOR’S SAFETY, ONLY DEVICES COMPLYING WITH THE SAFETY
STANDARDS, SUITABLE FOR DEVICES IN PERFECT OPERATING CONDITIONS, MUST BE USED.
SINCE THE EQUIPMENT EMITS X-RAYS, IT CAN ENDANGER THE OPERATOR AND THE PATIENT,
IF THE EXPOSURE PARAMETERS ARE NOT RESPECTED.
BMI produce advanced radiographic systems as far as safety and protection of the environment
are concerned. Therefore, if the unit is used correctly, there is no risk to people and environment.
Concerning the risks due to incorrect use of the unit, please refer to the directions and the
recommendations given above. We point out that the unit has been designed to prevent its
overturning up to 10° tilting with respect to the horizontal in transport position.
The following is therefore recommended:
• Do not stop, do not move and do not park the unit on planes having an inclination
greater than 10°.
• Do not try to move the unit without operating on the special “brakes release” handle
• When displacing the unit is careful if there is any obstacle on the floor (cables, steps,
little drops or rises) concerning the residual risks due to any defect of the adopted
protective measures, we remind you that the protection against electric shocks is realized
through an earth connection of the unit metallic parts.
• It is therefore necessary to check periodically the correct operation of the whole earth
circuit, according to the ordinal maintenance schedule described in this manual.
TO IGNORE AND NOT TO COMPLY WITH THE UNIT ALARMS COULD CAUSE AN OVERHEATING OF
THE X-RAY MONOBLOC. THIS CONDITION COULD RESULT IN THE LOSS OF THE MONOBLOC
INSULATION AND IN DAMAGE TO THE X-RAY TUBE. WHEN MOVING THE UNIT, BE CAREFUL
NOT TO HIT THE PATIENT OR THE OPERATOR. DO NOT MOVE THE UNIT TOO QUICKLY; THE
ACCUMULATED KINETIC ENERGY COULD HAZARD THE PERSONNEL NEXT TO THE EQUIPMENT.
This electronic equipment complies with EMC 2014/30/EU Directive, defining the maximum
emission levels allowed for the electronic equipment and the required immunity against
electromagnetic field.
This, however, does not exclude that signals coming from transmitters, like mobile phones or
similar mobile radio transmitters, even if complying with EMC standards, can influence the correct
operation of the Electro medical equipment if used nearby and with a quite high transmitting power. It
is therefore necessary to avoid using these radio transmitters near electronically controlled medical
systems, to exclude any interference.
The electronic unit complying with EMC standards has been designed to avoid any risk of
malfunctioning caused by electromagnetic interferences, when working in standard conditions.
However, the radio signals coming from high frequency transmitters at a quite high transmitting
power, if used near the electronic equipment, may cause electromagnetic incompatibility. Any
transmission with mobile radio transmitter must be avoided.
The mobile phones must be off in the area near the unit. These rules must be complied with
when the unit is working.
In the following pages, please find the EMC Certificate issued by the Italian notified body IMQ.
following symbol:
Note 1 At 80 MHz the upper frequency range applies.
Note 2 These guidelines may not be applicable in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a
The intensities of fixed transmitters, as well as base stations for telephones (mobiles/cordless), land mobile radios,
amateur radios, AM/FM radio transmissions and TV transmissions cannot in theory be forecast with accuracy. To assess the
electromagnetic environment generated by fixed RF transmitters, an electromagnetic analysis of the site must be considered. If
the intensity of the field measured in the location where the Device is used exceeds the aforementioned applicable level of RF
compliance, the Device should be monitored to check for normal operation. If irregularities in performance transpire, additional
measures may be required, such as moving or changing the position of the Device.
b
In the 150 kHz to 80 MHz frequency range the field intensity must be less than 3 V/m.
100 12 m 12 m 23 m
For transmitters specified for a maximum power output not included above, the recommended
distance d in meters (m) can be calculated using the equation for the transmitter frequency, where P
is the maximum rated power output of the transmitter in Watts (W) according to the transmitter
manufacturer.
CHAPTER 2
UTILIZATION
e
a
f
b
LEGEND
Rotation direction of the transportation
a
handle to release the brakes
Transport handle for parking brakes (dead-
b
man brakes) control
c Supply cable support
Bracket to lean the unit for overcoming little
d
obstacles
e Arm
f Arm retainer
ATTENTION! DO NOT PUSH THE EQUIPMENT FRONT SIDE AGAINST DOORS TO OPEN THEM.
REMEMBER TO PRE-SET THE LARGE FOCUS FOR ALL RADIOGRAPHIC EXPOSURES AND TO
CLOSE COMPLETELY THE X-RAY COLLIMATOR LEAD BLADES, IN ORDER TO INCREASE THE
SAFETY LEVEL AGAINST IONOGENIC RADIATIONS.
Assumed that this procedure will reduce the possibility of damaging the X-ray tube anode and the
high voltage components in the monobloc, BMI suggest using this procedure when the equipment is
first turned ON for the day, or when it has not been used for more than 24 hours. No test setup is
required.
This procedure is required only when the Device has not been used continuously for 1 month
after the installation date or if the equipment has not been used continuously for more than 1 month
in the Hospital department.
Forming the monobloc’s X-ray tube prevents it from damages. The X-ray tube formation
procedure consists in making radiographic exposures, according to below reported chart.
For each Device model all the radiographic exposures should be taken using only the X-ray tube
large focus. BMI suggest taking all the radiographic exposures closing the entire X-ray collimator
leaded shutter. The X-ray tube formation procedure will take approximately one hour.
REMEMBER TO SHUT THE X-RAY COLLIMATOR LEAD BLADES TO INCREASE THE SAFETY LEVEL
AGAINST IONOGENIC RADIATIONS.
Perform some exposures of 500 ms and 25 mA (with a break of 1 minute in between) starting
from 70 kV with increase of 5 kV up to 110 kV.
If during an exposure instability occurs, decrease of 5kV the voltage value and perform two
exposures; then he procedure keeps on.
Wait at least 30 minutes (refer to heating & cooling curves) before performing other exposures.
If no anomalies are found, Device is now ready to be used.
ON (I)
OFF (0)
Arm retainer
Handles
Release the arm, standing laterally Lift the arm using the handles
When the handle is released, the arm will remain in the acquired position thanks to a clutch
device integrated in the equipment.
The monobloc with the X-ray collimator assembly can rotate as described below:
• Monobloc yoke support rotation: +90° / -90°
• Monobloc rotation in the yoke: -45° / +167°
Goniometer
Handgrip
2.7 COLLIMATOR
2
3
4 4
5
6
7
LEGEND
1. X-ray field light beam pushbutton 6. Retractable tape measure for SID measuring
2. Longitudinal diaphragm adjustment knob 7. Handgrip for ± 90° rotation on its vertical axis
3. Transversal diaphragm adjustment knob 8. Goniometer
4. Rails guides for accessories 9. Monobloc focus external reference point
5. DAP (optional)
• Press the X-ray field light beam lamp push-button (1) to switch ON the lamp. The lamp
automatically switches OFF after 30 seconds. To avoid the lamp overheating read the
collimator manual.
• Position the collimator/monobloc assembly above the patient.
• Check if the X-ray light beam crosses is centered on the patient’s area to be examined.
• Adjust the X-ray field size using the two knobs available on the front part of the X-ray
collimator (2 and 3). The right knob adjusts the longitudinal shutters; the left one
adjusts the transversal shutters. The knobs make it possible to the operator to set the X-
ray field according to the radiographic cassette size. The collimator field size can vary
from 0×0 cm minimum to 43×43 cm (17”x17”) maximum, at a focus film distance of 100
cm (40”).
• The scale under each knob indicates the longitudinal and transversal field size at 100 cm
(40”) distance.
• Check the SID using the retractable tape measure.
The equipment microprocessor’s software verifies if the values inserted are compatible
with the equipment monobloc model performances, to avoid any failure.
Normally the dummy phantom overall dimensions are 45 cm x 45 cm (PMMA and aluminium
plate).
N. Phantom Structure
1 Abdomen 18 cm PMMA
2 Lumbar spine 18 cm PMMA + 5 mm Al 7x45 at the top and centered
3 Skull 2 cm PMMA + 2 mm Al + 10 cm PMMA + 1 mm Al + 3 cm PMMA
4 Extremity 2 cm PMMA + 2 mm Al + 4 cm PMMA
5 Chest 2 cm PMMA + 2 mm Al + 3 cm PMMA + 5 cm air + 2 cm PMMA + 1 mm
Al + 3 cm PMMA
Collim.
SID
SID
Collim.
2.8.3 MEASURE
• Place the Device collimator on the patient’s area to be exanimate, placing the device
respecting the SID.
• Place the radiographic cassette below the patient’s area to be delaminated.
• Set the collimator X-ray field area (additional information are reported on the device
manual) using its light beam, that allows the operator to obtain on the patient body a
field marker. To switch ON the light beam depress the ON/OFF button, which is located
on the control desk panel and on the collimator itself.
• Adjust the size of the field as described on § 2.7.
• Set the exposure parameters to take the radiography exposure, as described on Digital
Acquisition User Manual.
• Press halfway the preparation button (1st time) in order to spin the X-ray tube rotor.
• Press completely the button (2nd time) in order to take the radiographic exposure.
There will be a warning sound (buzzer sound) to advice the operator that the
radiographic exposure has been taken.
BEFORE TAKING ANY RADIOGRAPHIC EXPOSURE MAKES SURE THAT ALL NECESSARY
PRECAUTIONS AGAINST IONOGENIC RADIATIONS HAVE BEEN TAKEN. AFTER A LONG PERIOD OF
INACTIVITY OF THE EQUIPMENT, IT IS VERY IMPORTANT TO MAKE THE FORMATION OF THE X-
RAY TUBE, TO AVOID ANY H.T. DISCHARGE THAT COULD DAMAGE IT.
THE X-RAY TUBE FORMATION PROCEDURE IS DESCRIBED IN § 2.3.
WHEN THE X-RAY EMISSION IS OVER, THE RELEVANT WARNING LIGHT ON THE CONTROL
CONSOLE DISPLAY GOES OUT AND AT THE SAME TIME THE SOUND SIGNAL STOPS.
AT THIS POINT THE TWO-STEP X-RAY CONTROL PUSH-BUTTON MUST BE RELEASED OR, IF THE
EXPOSURE HAS BEEN CONTROLLED BY THE CONSOLE PUSH-BUTTONS, BOTH OF THEM MUST BE
RELEASED; THE DEVICE IS READY FOR NEXT EXPOSURE
BEFORE TAKING AN X-RAY EXPOSURE PROTECT THE OPERATOR AND THE PATIENT FROM X-RAYS
USING THE SUITABLE LEAD PROTECTIONS.
In the following table are listed all the errors code that the generator can send to the console.
For each error listed, it is explained its meaning, all the possible situations that can lead to the specific
code and the action that can be done for the resolution of the problem.
ERROR ERROR
PROBLEM (could indicate) ACTION
CODE MESSAGE
Before switching to inactive status (XR0) Reset the error and if
the generator sends the message ‘PR0’ the message persists
18 PR0 Timeout
and does not receive an answer from contact Technical
the console. Support.
Selected parameters caused X-ray tube Wait for the tube to cool
to exceed its programmed anode heat down.
19 Anode Heat Limit
limit. The estimated heating unit is
below the limit.
The received message is not supported Check the wiring
Communication or does not exist. harness of the console.
45 Message Not The console could have sent a wrong If the message persists
Supported message or there could be problems on contact Technical
communication line. Support.
The generator is set to an internal state Verify the protocol rules
in which the received command from (if the console software
Communication console is not allowed. This occurs when is not made from BMI).
46 Message Not the protocol rules are not respected. Change only one
Allowed Simultaneous commands are not command at a time.
allowed i.e. changing kV and prep at the
same time.
The preparation button is pressed Wait till the Initialization
during the Initialization phase. phase is finished.
Check if the led DL3 of
control board is lighted
OFF, after 60 seconds
from unit Power ON. If
Capacitors under
98 the led DL3 does not
charge
light OFF check fuse F1
and F2 of the capacitor
box.
If the message persists
contact Technical
Support.
The temperature surway is opened. Wait for the monobloc
The temperature inside the monobloc is to cool down.
Tube Head Over high. Check the wiring
99
Temperature The cable of the temperature surway is harness.
disconnected. Check the integrity of
the temperature surway.
The OVL alarm signal is still active after Contact Technical
60 second from the by-pass of the line Support.
100 OVL Timeout relays.
The capacitor charging phase has
problem.
The preparation switch has been Repeat the exposure if
1st time PB released too soon in one of the following it’s necessary.
101
released situations: Check switch.
• During the preparation phase
persist.
The generator internal status is set to Contact Technical
Error status (due to a previous latching Support.
error still not confirmed from console)
116 Wrong context and meanwhile it receives calibration
related message.
This error could be triggered only when
the generator is in setup mode.
The generator tried to go to an Contact Technical
unexpected internal status. This error Support.
118 Out of range shouldn’t occur and could indicate a
corruption of the flash memory or a
software bug.
The generator received a message from Reset the error and if
console with an invalid value. the message persists
119 Invalid value This could be caused by console which contact Technical
doesn’t comply with the supported Support.
version of the protocol.
The correlation of the current set of Reset the error and if
radiologic parameters fails. the message persists
120 Correlation fail
contact Technical
Support.
When potter is enabled and inserted, Check the wiring
the generator waits for the signal harness of the Potter
OK_POTTER while entering or during Bucky.
the Ready status (PR2). Check that the contact
Potter contact not
121 Error is generated if the signal doesn’t of Potter Bucky is
available
become active within 2 seconds from working.
the previous conditions.
The Potter Bucky is not operating
correctly.
The generator cannot operate with the The DAP device is an
DAP. optional contact
This error can be triggered in the technical support for
following situations: enable it.
• The generator receives command to set Check if it is installed in
DAP status and this last has not been the equipment.
122 DAP Not Present activated from setup. Check the wiring
• Timeout on DAP receiver line: the harness.
generator tries to communicate with the
DAP and receives no answer.
• This can occurs if there are
communication problems on DAP serial
line or the DAP is off (not connected).
Should not occurs: if DAP is not enabled Contact Technical
123 DAP Disabled from X-ray mode then the generator will Support.
automatically ignore it.
The received response from DAP is not Contact Technical
valid or unexpected. This kind of error Support.
could be caused by problems on DAP
124 DAP Error
serial line or from an unsupported DAP
version.
The DAP device is not functional.
This error could be triggered when Verify that the Digital
entering and during the Ready status System with external
DR EXP_OK (PR2) if all of the following conditions synchronization work
126
missing are met: properly.
• Exposure button pressed.
• Function DR enabled.
In general, each timeout error occurs when the generator does not receive the answer
to some specific commands when the connected peripheral (e.g. printer, dap) is not
responding. The timeout period is fixed to 2 seconds. This category of errors can be
caused by problems on communication line, noise or problems on the
console/peripheral side (application crash, out of sync, hardware problems).
2.10 LANGUAGE
The Device are equipped with a digital X-ray imaging system developed by Digitec.
Digitec’s platform is specifically designed for routine radiographic exposures where the image is
captured on a flat panel detector and stored as an electronic patient file in the computer. This
software is a high resolution digital imaging system intended to replace conventional film techniques.
The digital X-ray imaging system enables an operator to acquire, display, process, export images
to portable media, send images over a network for long term storage and distribute hardcopy images
with a laser printer.
Image processing algorithms enable the operator to bring out diagnostic details difficult to see
using conventional imaging techniques. Acquired images are instantly viewable or can be stored
locally for temporary storage.
DO NOT SWITCH OFF THE EQUIPMENT BY REMOVING THE SUPPLY CABLE FROM THE SOCKET
BEFORE DISCONNECTING THE AUTOMATIC CIRCUIT BREAKER.
AFTER SWITCHING OFF, WAIT AT LEAST 30 SECONDS BEFORE SWITCHING ON THE EQUIPMENT.
CHAPTER 3
ORDINARY MAINTENANCE
The purpose of this Periodic Maintenance is to guaranty a continue safe performance of the Unit,
to increase serviceability, to reduce costs (down time, repairs, etc.) and to assure the safety (personal
risk).
The equipment requires periodical checking and maintenance to assure correct operation and
safety standards.
Following information and instructions are therefore to be strictly complied with:
• A prolonged utilization of the equipment may affect the system safety due to the quick wear of
the mechanical and electronic components.
• The correct set up of the electronic and mechanical components and of the software can
prevent failures that could affect the safety and the irradiation exposure level to which the
doctors, the paramedics and the patients are exposed.
• The owner of the equipment is responsible for the maintenance correct execution and is
expected to have preventive checking and technical interventions carried out by expressly
authorized technical personnel.
The following checks and maintenance procedures, at the suggested intervals, are the
manufacturer’s recommendation for the most effective Periodic Maintenance schedule for the Unit.
Service tasks here described must be performed exclusively by service personnel specifically
trained on medical X-ray Generators.
The below reported chart illustrates the scheduled inspections to be carried out:
PERIOD: CHECKING:
NO SPECIFIC • The Flat Panel calibration is usually required in case the user notice some
CYCLE defect or artefact in the x-ray image or change configuration of x-ray system.
It also depends on how frequently the user make exposure with this system,
since x-ray strength could be reduced by usage. Refer to the Digital Acquisition
Service Manual
DAILY • Check the correct functionally of the hand switch remote control.
CHECKING • Check the correct functionally of the collimator.
• Check the wearing status of the supply cable and its plug.
• Check the parking brakes functionality.
• Perform the daily X-ray tube warm-up procedure (see § 2.2).
• Check the wearing status of the flat panel cable and its circular connector.
Do not use any corrosive or abrasive product or solvent to clean the unit surfaces. To disinfect
the system always comply with the rules in force regarding disinfection procedures and protection
against explosion. For cleaning and disinfection operations, following precautions have to be complied
with:
• Switch OFF the equipment and disconnect the supply cable plug from the socket.
• Be sure that no liquid is poured inside the equipment in order to prevent short circuits and the
corrosion of the electric and electromechanical components.
• Never spray or squirt the liquids directly onto the displays.
• Be cautious of all kinds of cleaning solvents or chemicals when you use it for the sake of
cleaning.
• Some person may be allergic to the ingredients used to prepare the cleaning products.
• Don’t try to put any food, drink or cigarette around the equipment, especially on the monitors.
Keep in mind that some components used to clean the monitors may only be able to be cleaned
using a product designed for cleaning that component, if this is the case it will be mentioned in the
cleaning tips:
• Cloth - A piece of cloth is the best tool to use when rubbing up a component. Although paper
towels or tissues can be used on most hardware as well, we still recommend you to rub it with
a piece of cloth.
• Water or rubbing alcohol – You may moisten a piece of cloth a bit with some water or rubbing
alcohol and rub it on the computer. Unknown solvents may be harmful to the plastics parts.
• Vacuum cleaner - Absorb the dust, dirt, hair, cigarette particles, and other particles out of the
equipment and monitors can be one of the best methods of cleaning a computer. Over time
these items can restrict the airflow in a computer and cause circuitry to corrode.
• Compressed air – not to be used. First, compressed air is extremely forceful, depending also
on its pressure, because compressed air can dislodge particles. These particles are a danger
since they can enter your eyes or abrade skin. The possible damage would depend on the size,
weight, shape, composition, and speed of the particles. Second, compressed air itself is also a
serious hazard. In addition, using air to clean forces the dirt, dust particles and moisture into
the air, making these contaminants.
NEVER ATTEMPT TO CLEAN INTERNALLY THE UNIT. THE UNIT MUST BE ALWAYS KEPT SHUT
UNLESS A SPECIFIC RENEWAL PART PROCEDURE IS TO BE UNDERTAKEN BY HIGHLY
QUALIFIED MAINTENANCE PERSONNEL.
3.4 DISPOSAL
This device is compliant with RoHS2 Directive Recast (Directive 2011/65/EU) related to reduction
of the use of dangerous material on electrical and electronic devices.
Disposal of the device is subject to WEEE Directive Recast (Directive 2012/19/EU), related to
waste generated by electrical and electronic equipment.
As a result, it cannot be treated as domestic waste in the European Union and must be dealt with
by specialist waste disposal companies.
Graphic symbol:
CHAPTER 4
ELECTRICAL AND ELECTRONIC
COMPONENTS
4.1 INTRODUCTION
The below reported picture shows the area where are located inside the equipment the main
electrical and electronic components.
Monobloc
Medical Station
Electronic box
Capacitors Inverter
4.2 PANEL PC
Here below are indicated the operator console external and internal view.
Panel PC
Power Supply
Panel PC
Power Supply
Flat Panel
4
10
11
4.4 INVERTER
Quantity
Nr. Description
Model 4 kW Model 20/30 kW
1 Discharge capacitors power charge push button 1 1
2a Fuse 690V 80A 1 0
2b Fuse 140A 660V 0 1
3 Electrolytic capacitor 6800 µF 2 16
4 Rectifier bridge module 1 1
5 Resistance 50W 470R 2 6
6 Toroidal transformer single phase 1500 VA 0 1
7 Panel fuse holder 10x38 0 1
8 Fuse 10x38F 32A 400V 0 1
9 Solid state relay 0 1
2
5 b
9
2
3
3
CHAPTER 5
INSTALLATION
5.1 INTRODUCTION
5.2 UNPACKING
The equipment is wrapped in cellophane and then put inside a wooden or plywood case.
The equipment is secured by means of wooden crosspieces to avoid any movement during the
transport.
On the external side you can find the handling and storing instructions, the shipping address, the
equipment model and serial number.
Any particular request regarding the packing and the labelling must be specified with the order.
The final customer is suggested to ask for special shock watch labels to be applied on the
external side of the packing case. When the customer receives the equipment, if the shock watch
labels have become red, this will mean that the equipment has been incorrectly handled during the
transport or the storing, and has probably been damaged.
For these reasons the customer must accept the equipment on condition, writing “…acceptance
withheld for damaged packing (or similar words)…” on all copies of the transport documents
accompanying the goods. Later the customer will open the case to check the equipment integrity and
will inform the manufacturer about what happened and about the equipment condition.
The customer must inform of:
- Number of shipping document
- Equipment model and serial number.
- He will supply a detailed report with photographs illustrating the packing situation and
the damages and/or malfunctioning of the mechanical, electrical and electronic parts of
the equipment, if any.
- This communication must be sent within 2 days from the date of receipt of the
equipment. After this delay the supplier will not accept any responsibility for damages or
defects caused by the transport.
- Any request of returning the product back will have to be agreed up on with the
manufacturer who must authorize it in writing.
EXTERNALLY TO THE DEVICE CRATE BMI STICK ON THE DETAILED PROCEDURE TO CORRECTLY
UNPACK THE EQUIPMENT.
Pay attention to the symbols printed outside the cardboard of the package:
Check visually each cable and each connector installed in the equipment to make sure that there
is no damage that could jeopardize the correct operation of the equipment and the operator’s safety.
Following components must be inspected:
• Equipment supply cable and plug.
• X-ray control cable.
• Equipment/monobloc connecting cable.
• X-ray collimator supply cable.
• DAP interface cable (optional).
Check whether all the main parts of the equipment work properly, to avoid any defect that could
cause malfunctioning or endanger the user or the patient.
The mechanical inspections include the following operations:
• Displacement of the equipment mobile parts in order to verify if the movements occur
smoothly and noiseless.
• Parking brakes testing to check if the equipment braking is quick and efficient.
• Check if there is any break and/or dents and if any screw has become loose thus
jeopardizing the equipment operating and safety conditions.
The main controls are:
• Rolling of the wheels to verify if there is any wrap or damage.
• X-ray equipment parking brake to check if the braking occurs properly.
• Check the integrity of the control console and verify if the touch screen display has no
dents or scratches.
Integrity and fastening of the guards to verify if all parts screwed to the equipment structure are
firmly fixed. Check the arm balancing to verify that it keeps the required position when lifted (for
example).
CHAPTER 6
ADJUSTMENTS
& SETUP
6.2 ADJUSTMENTS
• Place the non-invasive kilovolmeter instrument on a flat and horizontal surface under
the X-ray collimator.
• Position a Non-Invasive kVp meter aligned to the X-ray beam at SID 50 cm. Open the
light field to center it with the kVp meter.
• Make an exposure (see the next table for the parameters), read the kVp and take
note of the value.
Parameters
Model
kV mA ms
Model 4 kW 100 40 100
Model 20 kW 100 200 100
Model 30 kW 100 200 100
• Verify that the kV value measured by corresponds to the reference value. The
admitted kV measuring allowance is ± 5%.
To eventually adjust the kV value measured, the technician can adjust on the inverter electronic
board «control logic board» the trimmer R117 (for the position see the layout at page 118).
Check the performance of the x-ray tube in compliance with 10.4.4 of Mexican Standard
considering the effective value of HVL dependent of total filtration of the device.
Take radiographic exposures pre-setting the parameters, as written in the below reported charts:
• Connect the Oscilloscope CH1 between TP18 (mA) and TP1 (GND) on Control logic
board (see page 118).
• Configure CH1 in 1V/div. Time base 20ms/div.
• V and mA relation, measured in the Oscilloscope, is as follows:
mA: 1V = 100 mA (for Model 20/30 kW)
mA: 1V = 20 mA (for Model 4 kW)
• Switch ON the equipment, entering in “operator mode”.
• Take 6 radiographic exposures pre-setting the parameters, according to the Device to
be adjusted, as written in the below reported charts:
ONLY WHEN THE MONOBLOC HAS BEEN WARMED UP, THE mA VALUES MEASURED SHOULD BE
CONSIDERABLE RELIABLE.
Configuration provides the initial settings that must be carried out before starting to work with
the Device. Functionally characteristics of the Device are defined during Configuration.
Calibration data are stored in a memory chip located on the main board.
For enter on the Configuration and Calibration mode the technician must run the BMI-Console
software. This software is installed on the Medical Station on the Administrator profile (refer to the
Digital Acquisition User Manual for understand how to enter in this profile).
The Configuration and Calibration mode is composed by the following pages:
The following pages can be different, depending from the software revision, but the
function is identical.
Password
Counters
mA Calibration
NOT ALL THE MASKS AND ALL THE PARAMETERS ARE AVAILABLE. THE PASSWORD’S LEVEL
DEFINE WHAT THE TECHNICIAN CAN MODIFY.
• To navigate between the setup pages press the two arrows icons
Navigator arrows
• Left arrow icon is used to go back in the masks meanwhile the right arrow icon is
used to go forward in the masks.
ONLY TECHNICIAN AUTHORIZED BY BMI ARE ALLOWED TO ENTER IN THE EQUIPMENT SET UP MENU.
• Type the password (TO BE REQUESTED TO BMI) on the Password mask and then
“BMI-Console” has a serial communication with the main board. Be sure that the correct COM
port is selected.
This mask allows to enable/disable an eventual DAP and set the time during of the collimator
lamp.
• Select the time during of the collimator lamp, using the dedicated arrows.
• For enable/disable the DAP option, push on the dedicated button:
A light gray icon means that is not possible to enable/disable the option.
• The buttons are a toggle buttons.
• DAP options are:
Button Description
DAP Enable the function of the optional DAP
Enable Printer Enable the function of the DAP Printer
mGycm2 Enable the function of the unit of measure when the DAP is enable
dGycm2 Enable the function of the unit of measure when the DAP is enable
• The DAP gain is set by the manufacturer when the Device is in under testing.
• To confirm the selection and to exit from Configuration Mode press the icon
• To confirm the selection and to exit from Configuration Mode press the icon
This section only applies after a monobloc replacement with a different X-RAY tube.
THE DEVICE AUTOMATICALLY RESETS THE CALIBRATION PARAMETERS WHEN THIS PROCEDURE
IS STARTED. EVEN IS SET AGAIN THE SAME X-RAY TUBE.
• Select the desiderate X-RAY tube using the scrolling icon (A).
• Proceed with the tube calibration.
A light gray icon means that is not possible to enable/disable the option.
• For enable/disable the desired option push on the respective button. The button is a
toggle button.
• To confirm the selection and to exit from Configuration Mode press the icon
Button Description
D.R. Enable the DR interface
Potter Enable the potter bucky interface
A.E.C. Not Evailable
Enable the X-ray tube anode rotation.
Starter THIS PARAMETER MUST BE ENABLED FOR GENERATOR WITH
ROTATING ANODE TUBE!!!
Real mAs Enable the visualization of the real mAs on the message pop-up
Exposure Out Enable a digital output for indicate the x-ray exposure
Err Auto Ack Enable the automatic alarm acknowledgment
3P Enable/disable the operative mode of 3 points
mAs Visible
When the 3P is enabled, show on the field of mAs the calculation of
this parameter
Enable Scr Enable the function of the screen for the APR
Lamp 1st time
Enable the power ON of the collimator when it’s pressed the
preparation button.
Enable OP Mod Enable/disable the Operator Mode
Enable APR Enable/disable the APR Mode
Enable Login Enable/disable the Login password, when the Device is switched ON
6.3.7 COUNTER
This mask visualizes the total and partial numbers of radiographic exposures done and the
heating unit thermal capacity generated by the monobloc. It is possible to reset the partial counters in
case of monobloc replacement.
• Enter in Configuration mode (see § 6.3.1).
• Select Counters mask.
• To confirm the selection and to exit from Configuration Mode press the icon
.
Pressing on it
• Adjust the value displayed in the field (with the two arrows under the field), in order
to measure, with the digital multimeter, 2.05 Vdc
• Select the field
• Adjust the value displayed in the field (with the two arrows under the field), in order
to measure, with the digital multimeter, 5.28 Vdc
• To confirm the selection and to exit from Configuration Mode press the icon
DO NOT CARRY OUT THE mA VALUES CALIBRATION WHEN THE MONOBLOC IS COLD.
REMEMBER TO EFFECT A PREHEATING CYCLE MAKING AT LEAST TWO OR BETTER FIVE
EXPOSURES, PRE-SETTING ON OPERATOR MODE THE MINIMUM mA AND KV EXPOSURE
VALUES.
• To change the measure number uses the two arrow keys (for decrease the
IN THE mA CALIBRATION MASK THE INITIALS SF AND LF MEANS RESPECTIVELY THE SMALL
AND THE LARGE FOCUS OF THE X-RAY TUBE.
mA AND KVP RANGES DEPEND ON THE DEVICE MODEL. SEE THE FOLLOWING TABLES FOR
KNOW THEM.
Model 4 kW
kV mA
Measure 1 40 25
LARGE FOCUS
Measure 2 40 40
Measure 3 40 63
Measure 4 40 80
Measure 5 40 100
Measure 6 100 25
Measure 7 100 40
Model 20 kW Model 20 kW
kV mA kV mA
Measure 1 60 50 Measure 8 60 25
SMALL FOCUS
LARGE FOCUS
Model 30 kW Model 30 kW
kV mA kV mA
Measure 1 60 50 Measure 9 60 25
SMALL FOCUS
• Make an exposure again to verify that mA value in the Oscilloscope is the one
selected; if it is not, repeat the procedure to increase or decrease the value.
For example following there is the 7th measure of the Model 4 kW.
CH2: 40 mA
CH1: 100 kV
• Once the correct mA value is reached (take at least two radiographic exposures in
order to verify if the mA value displayed is correct), go on the next step of calibration
• To confirm the selection and to exit from Configuration Mode press the icon
Protection guard
Hided screws
Control logic
KEEP CONTINUOUSLY PRESSED THE DISCHARGE PUSH-BUTTON FOR AT LEAST 100 SECONDS
IN ORDER TO AVOID ELECTRICAL SPARKS IN THE PUSH-BUTTON CONTACT TERMINAL BLOCK.
CHAPTER 7
EXTRAORDINARY
MAINTENANCE
Check whether the hooking system keeps the arm firmly locked and in sequence hook an
unfasten the arm verifying the correct insertion of the hook in its seat. Move the equipment with the
arm locked to verify if it keeps its position. In the negative, replace the “arm retainer assembly”.
“X-ray window”
• Check if there are any dielectric oil spots on the “X-ray window” and on the collimator
plane.
• Reassemble the item previously detached or unwired.
• Make the X-ray field centring (refer to collimator manual).
• If the result of the test is negative, ship to the manufacturer the monobloc to be
repaired or replaced.
• Switch OFF the equipment and disconnect the equipment power supply cord from the
socket.
• Place the equipment on a flat floor and lock its arm.
• Place a dynamometer in the middle of the column (point of application A).
Point A
• Verify that the equipment with the brakes locked does not move applying a pushing
force of ~220 N [± 5 N].
• Verify that the equipment starts to move with the brakes released applying a pushing
force of ~20 N [± 5 N].
• If the test result is negative, adjust the brakes blocks as indicated in § 7.7.1.
Point B
• Push with the dynamometer the arm verifying that to lift up is it necessary to apply a
force of ~50 N [± 10 N].
• Place the dynamometer on the arm, near the handles, before the fork rotating pin
(application point C).
Point C
• Push with the dynamometer the arm verifying that to lift down, is it necessary to
apply a force of ~50 N [± 10 N].
• If the test result is negative, adjust the spring as described in § 7.7.3.
Point D
• If the test result is negative, adjust the fork joke rotation clutch as described in §
7.7.5.
• Check using a dynamometer if the force to be applied on the collimator’s handle to tilt
the monobloc (-45°/+167°) is ~40 N [± 5 N] (application point E).
Point E
• If the test result is negative, adjust the monobloc rotation clutch as described in §
7.7.6.
Upper housing
Lower housing
• Verify on both two housings, using a torch/lamp, that the pulleys, ropes and
mechanical equaliser does not damage them, prejudicing the equipment safety.
During the inspection move continuously up and down the arm in order to better
evaluates the wearing status of all the mechanical components.
• Reassemble the column front guard.
• Close the arm.
Protection guard
Hided screws
KEEP CONTINUOUSLY PRESSED THE DISCHARGE PUSH-BUTTON FOR AT LEAST 100 SECONDS
IN ORDER TO AVOID ELECTRICAL SPARKS IN THE PUSH-BUTTON CONTACT TERMINAL BLOCK.
• Adjust at the same time the brakes stretch acting, half turn per time, on the self-
locking hexagonal nuts (A).
A A
Self locking hexagonal nut:
• To approach the brake to the wheel: screw
both nuts half turn per time.
• To move away the brake from the wheel:
unscrew both nuts half turn per time.
• Grip the transportation handle and release two times the stationary brakes in order to
evaluate if the brakes adjusting has been done correctly.
• Without releasing the stationary brakes, push manually the equipment verifying that it
does not move.
• Verify that each brake sliding block is placed 1 mm away from the wheel.
Brake block
1 mm distance
Black tap
Pulling direction to
remove the guard
Ring nuts
• Pull out the column’s upper cover guard as indicated by an arrow in picture
THE ADJUSTMENT OF THE SPRING MUST BE CARRIED OUT ONLY BY TECHNICIANS EXPRESSLY
AUTHORIZED BY THE MANUFACTURER AND DULY TRAINED.
THE GOOD OPERATION OF THE SPRING IS VERY IMPORTANT FOR THE OPERATOR’S AND THE
PATIENT’S SAFETY.
Clutch sector
• Act on the spring adjusting housing, unloosing the two hexagonal nuts (1 and 2).
2
1
• Shifting both the hexagonal nuts to the midcourse of the threaded bar.
Midcourse
Threaded bar
Nuts
• Adjust the collimator height from the floor at 200 cm, using the collimator retractable
meter, as shown on picture.
Collimator
200 cm
• When the technician has verified that the F.F.D. is 200 cm, screw both the hexagonal
nuts attached to the “Bumper”.
Nuts
“Bumper”
• Grip both hexagonal nuts to the “Bumper”, keeping hold the lower hexagonal nut and
screw the upper hexagonal nut.
Upper nut
Lower nut
“Bumper”
• This procedure does not allow at the arm losing the maximum’s arm SID position.
• Pull down the arm till reaching 90°. In this position the arm should keep the position
due to a correct spring balancing. In case the arm drops down, increase the spring
traction force, screwing both hexagonal nuts, vice versa reduce the spring traction
force, unscrewing both hexagonal nuts. On the following pictures has been shown the
lower hexagonal nut to be adjusted.
Nuts
• Reassemble the column front guard and the column’s upper cover guard
Clutch sector
• Detach the cover disk of the rotation pin, unscrewing the central screw, to be set
aside.
Rotation pin
Fork
Cover disk
Fixing screw
M12 nut
• Rotate the monobloc ± 90° around the pin, checking that it keeps in place on both
positions. If it moves, grip the M12 nut.
DO NOT UNSCREW M12 NUT: THE FORK WILL COME OFF TO THE ARM!
Position of the
lateral clutch
Ring nut
Nut retainer
The pair of steel ropes and the spring equalizing rocker arm have to be replaced every 5 years.
The customer must inform BMI in advance and BMI will authorize the replacement and supply the
spare parts together with the instructions for the replacement.
• Turn off the mobile unit, using circuit breaker, and remove the plug from the socket.
• Remove the Medical Station by unscrewing the six fixing screws.
• Remove the two profiles and the right carter, by unscrewing the lateral screws.
• Remove the protection guard by unscrewing its lateral screws.
Please note that some of them are placed (hided) below the cassette tray.
Medical Station
Screws
Profiles
Protection guard
KEEP CONTINUOUSLY PRESSED THE DISCHARGE PUSH-BUTTON FOR AT LEAST 100 SECONDS
IN ORDER TO AVOID ELECTRICAL SPARKS IN THE PUSH-BUTTON CONTACT TERMINAL BLOCK.
• Detach the circuit breaker to facilitate the disconnection of the power supply cable
wires from it.
• Disconnect the blue (N) and the brown (L) wire from the circuit breaker, as shown in
the picture below.
• Unscrew the cable retainer and remove the power supply cable from the electronic
box.
• Insert a new power supply cable and connect the blue (N) and the brown (L) wire on
the circuit breaker. Mount it on the electronic box in a correct position.
• Secure the power supply cable on electronic box with cable retainer.
• Connect the ground wire on the copper bar by using the screw.
• Connect the other ground wire on the lateral carter, screwing the M5 nut.
• Mount the right carter, the two green profiles and the control console on the unit.
CHAPTER 8
TROUBLESHOOTING
AND SPARE PARTS
8.1 TROUBLESHOOTING
The Device contains many self-diagnostic routines which greatly facilitates troubleshooting.
As a general rule, the first step in any troubleshooting procedure is to verify correct power supply
voltage.
The Medical Station will display status messages on the touch screen display during normal and
abnormal operation of the equipment. This paragraph § 2.9 contains tables of those messages and
suggests actions to be taken by service personnel to correct any malfunctions that may occur.
WHENEVER THE DEVICE IS TURNED OFF, WAIT 5 MINUTES BEFORE TURNING IT ON.
The following table describes the normal states and functions of the status and diagnostic LEDs
on the circuit boards in the unit. A change in the state of an LED may be for a brief duration,
especially during a fault or an exposure.
‘Filament board’
LED NORMAL STATE FUNCTION
DL1 OFF If lit, indicates that excessive current has been.
DL2 ON Indicate presence of the power supply.
‘Control board’
LED SILK SCREEN NORMAL STATE FUNCTION
DL1 OFF Over current
DL2 OFF kV Unbalanced
DL3 OFF Over-load
DL4 NOT RAD/FLUORO
DL5 LP/NOT HP (LOW POWER IS HIGH SIGNAL)
DL6 LP/NOT HP (LOW POWER IS HIGH SIGNAL)
‘Main board’
LED SILK SCREEN NORMAL STATE FUNCTION
Indicate presence of
D27 5V ON (GREEN)
the +5Vdc supply
DL9 PREP OFF Lit when in PREP state
DL3 XRAY OFF Lit when in XRAY state
DL5 XRAY OFF Lit when in XRAY state
DL6 XRAY OFF Lit when in XRAY state
Should some part of the equipment be replaced BMI suggest using only the original spare parts
in order to avoid any malfunctioning of the equipment and the consequent loss of warranty.
The manufacturer does not assume any responsibility for damages to things or people or to the
equipment caused by the utilization of spare parts that have not been previously approved or
purchased by BMI.
SHOULD YOU NEED ANY SPARE PART, GET IN TOUCH WITH BMI SERVICE DEPARTMENT.
IT IS RECOMMENDED TO COMMUNICATE THE EQUIPMENT SERIAL NUMBER AND MODEL TOGETHER
WITH THE SPARE PART CODE.
MEDICAL STATION
Model
Code Description Model 20 Model 30
Model 4 kW
kW kW
Exposure two-step hand switch
BMHS
complete of extensible cable
ELECTRONIC BOX
Model
Code Description Model 20 Model 30
Model 4 kW
kW kW
CAPACITORS BOX
Model
Code Description Model 4 Model 20 Model 30
kW kW kW
CHAPTER 9
SCHEMATICS
CODE: E01S028
DESCRIPTION: Inverter board (mounted inside the inverter – not available as spare part)
CODE: E01S025
DESCRIPTION: Control logic board
TP 1 (Gnd)
TP 18
R117
TP 19
TP 28
CODE: E01S027
DESCRIPTION: Main board
CODE: E01S004
DESCRIPTION: Filament board
CODE: E01S003
DESCRIPTION: Starter board
9.2 SCHEMATICS