Professional Documents
Culture Documents
Pacing
Pacing
Log in
Access keysNCBI HomepageMyNCBI HomepageMain ContentMain Navigation
Bookshelf
Search databaseBooksAll
DatabasesAssemblyBiocollectionsBioProjectBioSampleBooksClinVarConserved
DomainsdbGaPdbVarGeneGenomeGEO DataSetsGEO ProfilesGTRHomoloGeneIdentical Protein
GroupsMedGenMeSHNLM CatalogNucleotideOMIMPMCPopSetProteinProtein ClustersProtein Family
ModelsPubChem BioAssayPubChem CompoundPubChem
SubstancePubMedSNPSRAStructureTaxonomyToolKitToolKitAllToolKitBookgh
Search term
Search
Browse Titles
Advanced
Help
StatPearls [Internet].
Show details
Search term
Overdrive Pacing
Michael Self; Christopher R. Tainter.
Author Information
Go to:
Introduction
Temporary cardiac pacing (TCP) is a type of exogenous cardiac pacing in which an external
energy source delivers electrical impulses to stimulate the heart to contract faster than its
native rate. This intervention can be used to over-ride a malignant tachydysrhythmia or
compensate for symptomatic bradycardia. TCP is typically used for dysrhythmias refractory
to pharmacological therapies or cardioversion. Temporary cardiac pacing is not a new
intervention; it was first attempted in 1952 when Paul Zoll used hypodermic needles in the
chest wall to deliver a pulsating external current for two patients with asystole.[1] Today,
TCP is available in a broad spectrum of critical care settings, from pre-hospital to the
intensive care unit, delivered via a variety of modalities, including transcutaneous,
transvenous, epicardial, and transesophageal.[2] This activity will focus on temporary
cardiac pacing in critical care settings.
Go to:
Contraindications
There are no absolute contraindications to temporary cardiac pacing. However, TCP should
be avoided or used with caution in certain scenarios. The most common reason to avoid any
kind of temporary cardiac pacing is hemodynamically stable bradydysrhythmias with rare or
tolerable symptoms. Clinicians should avoid cardiac pacing in hypothermia (unnecessary)
and prolonged bradyasystolic arrest (futile). Clinicians should also avoid transvenous pacing
in patients with a prosthetic tricuspid valve, as catheter placement may damage the valve or
the catheter may become stuck in the valve, and patients with an excessive risk of bleeding,
including those with acute myocardial infarction receiving thrombolytics, anticoagulants, and
antiplatelet agents.[1][2][3]
Go to:
Equipment
Temporary transcutaneous pacing requires a pulse generator and monitoring unit with
standard defibrillation pads. These are ubiquitous in most medical settings. The pulse
generator must have a pacing function. ECG electrodes, non-invasive or invasive blood
pressure monitoring and pulse oximetry are strong recommendations.[1][6]
Temporary transvenous pacing requires central venous access, typically with a 6 French (Fr)
venous percutaneous introducer sheath, a transvenous pacing catheter, and an external pulse
generator. Ultrasound guidance for central venous access is also a strong recommendation. A
12 lead-capable ECG machine, cardiac monitor, non-invasive or invasive blood pressure
monitoring, and pulse oximetry should be available. This procedure is sterile, and standard
sterile technique is necessary.[1][2][3]
The transvenous pacing catheter is mostly bipolar, 3 Fr to 5 Fr in diameter and 100 cm long.
Lines are typically present at 10 cm intervals to estimate catheter depth. Catheters may be
flexible, semi-floating, or rigid. For most emergent indications, when temporary pacing
would be in order, a semi-floating catheter with a balloon is used. The balloon holds 1.5 ccs
of air and requires testing before insertion. The leading end of the catheter has two
electrodes, of which the negative is most distal. Adapters allow the electrodes to be attached
to the pulse generator.[2][3]
The external pulse generator delivers an electrical current through the pacing catheter,
measured in milliamperes (mA). Generators share the same basic features, including
electrical output and cardiac sensing components. These are present as dials on the
generator’s face. Output control regulates the current delivered, functionally controlling the
ability to obtain electrical capture. Rate control selects the pacing rate. Sensitivity control
establishes the threshold at which a sensed intrinsically generated current inhibits the
pacemaker from firing. This setting is the choice for demand (synchronous) pacing. For most
emergent indications, the sensitivity control is turned to the lowest setting, providing
asynchronous (fixed rate) pacing.[2][3]
Epicardial pacing typically requires an external pulse generator and surgically placed
epicardial pacing electrodes. The electrode location requires confirmation with the surgical
team.[2][7]
Go to:
Technique
Temporary Transcutaneous Pacing (TTCP)
Depending on the clinical scenario, consider analgesia and/or sedation before or
immediately after, initiating transcutaneous pacing as the electrical current needed
for capture is painful.
Place the pacing pads in an anterior-posterior (preferred) or anterior-lateral
configuration.
Attach the pads to the pulse generator, most often a defibrillator with pacing
capabilities. Some pulse generators require the attachment of cardiac monitoring
leads for proper functioning.
Select the pacing function on the defibrillator.
Select the desired rate per indication, typically 60 to 80 BPM or 10 to 30 BPM
greater than the intrinsic rate.
Starting at 70 milliamps (mA), increase the output by 5 to 10 mA until the
initiation of capture, indicated by a wide-complex QRS following every pacer
spike (electrical capture) and signs of improved perfusion (mechanical capture).
It is critically important to confirm ventricular contraction (mechanical capture), as
it is not difficult to be misled by electrical activity representing a QRS complex,
which is possible by confirming a pulse that matches the set pacemaker rate by
palpation echocardiogram, pulse oximetry, or arterial waveform.
The current at which capture is obtained is called the threshold current.
Once capture is confirmed, set the current at 5 to 10 mA higher than the threshold
current to prevent loss of capture.
If the pacing rate does not get captured at greater than 120 mA, reposition the pads
and repeat the above steps.[1][6]
Temporary Transvenous Pacing (TTVP)
Ensure that the pulse generator is in good working condition and all equipment is
at the bedside.
Obtain central venous access using an appropriately sized percutaneous venous
introducer sheath, typically 6 Fr. Inappropriately sized sheaths may be unable to
pass the pacing catheter or may leak around it. The right internal jugular vein and
left subclavian veins are the preferred vessels because of a more direct path to the
right ventricle (RV). Ultrasound guidance for placement is highly recommended.
Detailed instruction on obtaining central venous access is beyond the scope of this
article.
Placement without ECG Guidance
Connect the pacing electrodes to the pulse generator.
Insert the catheter into the introducer sheath so that the balloon and electrodes are
past the distal end of the introducer sheath (approximately at the 20 cm mark on
the catheter). Note: do not inflate the balloon while it is within the introducer
sheath.
Turn on the pulse generator. Set the pacing generator to the desired rate depending
on the indication, typically 60 to 80 BMP or at least 10 BMP greater than the
native rate. The initial output requires setting to 2 to 5 mA. Decrease the
sensitivity to the lowest level.
Inflate the balloon and advance the catheter slowly. The cardiac monitor will
typically show pacer spikes. When the catheter passes the right ventricle and
contacts the endocardium, a wide QRS complex with an LBBB pattern will follow
every pacer spike (electrical capture). It is also reasonable to intermittently deflate
the balloon to check for capture.
Deflate the balloon, secure the catheter in place, and make a note of catheter depth.
Ensure mechanical capture by evaluating: signs of perfusion, peripheral/central
pulse rate equals pulse generator set rate, pulse oximetry waveform, arterial line
waveform.
Decrease the output slowly until capture is lost. Increase the output to regain
capture; this is the threshold current, typically less than 1 mA. Increase the output
to approximately 2.5 times the threshold current, typically 2 to 3 mA.
If demand pacing is the goal, adjust the sensitivity such that native cardiac
electrical impulses inhibit the pacemaker. See the section on epicardial pacing
below for a more in-depth discussion of demand pacing.
Placement with ECG Guidance
Connect the negative (distal) electrode to ECG lead V1 using an alligator clip.
Insert the catheter into the introducer sheath so that the balloon and electrodes are
past the distal end of the introducer sheath (approximately at the 20 cm mark on
the catheter). Note: do not inflate the balloon while it is within the introducer
sheath.
Inflate the balloon and slowly advance the catheter. Closely monitor the ECG.
ECG morphology should change predictably with electrode location.
o In the high right atrium (RA), there will be a large negative p-wave,
typically greater than the QRS complex, followed by a negative QRS
complex.
o As the catheter passes through the RA, the p-wave becomes biphasic and
then positive.
o When the catheter passes into the RV, the p-wave becomes smaller and
negative, followed by a deeply negative QRS complex.
o Note: the duration of the QRS complex will depend on the native cardiac
activity. For example, an AV nodal escape rhythm will have a narrow QRS
complex, while a ventricular escape rhythm will have a wide QRS
complex.
o When the catheter makes contact with the right ventricular endocardium, an
injury pattern will result, with a deep, negative QRS complex followed by
marked ST elevation.
Advance the catheter until observing the RV pattern or RV endocardial pattern.
Deflate the balloon, secure the catheter in place, and make a note of catheter depth.
Ensure mechanical capture by evaluating: signs of perfusion, peripheral/central
pulse rate equals pulse generator set rate, pulse oximetry waveform, arterial line
waveform.
Decrease the output slowly until capture is lost. Increase the output to regain
capture; this is the threshold current, typically less than 1 mA. Increase the output
to approximately 2.5 times the threshold current, typically 2 to 3 mA.
If demand pacing is the desired goal, adjust the sensitivity such that native cardiac
electrical impulses inhibit the pacemaker. See the section on epicardial pacing for
a more in-depth discussion of demand pacing. [1][2][3]
Epicardial Pacing
Select the desired parameters on the pulse generator, depending on the clinical
scenario. For the prevention or treatment of tachyarrhythmias, this may be
asynchronous. For the prevention or treatment of bradydysrhythmias, this may be
asynchronous or synchronous (demand) pacing.
If demand pacing is the objective, adjust the sensitivity such that native cardiac
electrical impulses inhibit the pacemaker. Then set the pacemaker rate. In demand
pacing, this represents the backup rate, and the pacemaker will deliver an impulse
if it does not sense a native electrical impulse at a rate greater than the backup
rate.
Adjust the pacemaker output and evaluate for signs of mechanical capture as
described above.[2][7]
Go to:
Complications
Complications of TTCP include pain, failure to obtain capture, loss of capture, and rarely
cutaneous burns. Complications of TTVP are more numerous, and many are related to
central venous access, including infection, bleeding, damage to nearby structures, vein
thrombosis, air embolism, pneumothorax, and others. Additionally, TTVP can sometimes
induce ventricular tachycardia or ventricular fibrillation.[1][2][3][8][9]
Go to:
Clinical Significance
Select clinical pearls regarding TCP include:
Careful patient selection is paramount as the risks associated with certain types
of TCP outweigh the benefits.
TCP should be used when maximum medical therapy, including cardioversion
when appropriate, has failed.
Assuming that electrical capture is equivalent to ventricular or mechanical capture
is one of the most significant pitfalls in TCP. Electrical activity in the form of a
QRS complex is sometimes present without subsequent ventricular contractions.
TTCP is especially prone to this due to the chest wall impedance and greater depth
of the ventricles from the pacing pads. Ensure mechanical capture is present by
confirming a pulse that matches the set pacemaker rate by palpation,
echocardiogram, pulse oximetry, or arterial waveform. Check for signs of
mechanical capture as mentioned above and re-evaluate them frequently,
especially after transfers, procedures, or repositioning.[6]
Another pitfall related to TTCP is basing capture on the ECG waveform of a
separate cardiac monitor. Many defibrillators/pulse generators require the
placement of their own ECG electrodes, and any decision about electrical capture
should have their basis in ECG waveforms on the defibrillator/pulse generator
itself, rather than a separate cardiac monitor.
TTVP catheter placement should be confirmed by a chest X-ray. Continuous
echocardiography may also be useful to confirm lead placement.[10][11]
Go to:
Review Questions
Access free multiple choice questions on this topic.
Comment on this article.
Go to:
References
1.
Gammage MD. Temporary cardiac pacing. Heart. 2000 Jun;83(6):715-20. [PMC
free article] [PubMed]
2.
Sullivan BL, Bartels K, Hamilton N. Insertion and Management of Temporary
Pacemakers. Semin Cardiothorac Vasc Anesth. 2016 Mar;20(1):52-62. [PubMed]
3.
Harrigan RA, Chan TC, Moonblatt S, Vilke GM, Ufberg JW. Temporary
transvenous pacemaker placement in the Emergency Department. J Emerg
Med. 2007 Jan;32(1):105-11. [PubMed]
4.
Kowey PR, Engel TR. Overdrive pacing for ventricular tachyarrhythmias: a
reassessment. Ann Intern Med. 1983 Nov;99(5):651-6. [PubMed]
5.
Blommaert D, Gonzalez M, Mucumbitsi J, Gurné O, Evrard P, Buche M, Louagie
Y, Eucher P, Jamart J, Installé E, De Roy L. Effective prevention of atrial
fibrillation by continuous atrial overdrive pacing after coronary artery bypass
surgery. J Am Coll Cardiol. 2000 May;35(6):1411-5. [PubMed]
6.
Bektas F, Soyuncu S. The efficacy of transcutaneous cardiac pacing in ED. Am J
Emerg Med. 2016 Nov;34(11):2090-2093. [PubMed]
7.
Reade MC. Temporary epicardial pacing after cardiac surgery: a practical review:
part 1: general considerations in the management of epicardial
pacing. Anaesthesia. 2007 Mar;62(3):264-71. [PubMed]
8.
Carrizales-Sepúlveda EF, González-Sariñana LI, Ordaz-Farías A, Vera-Pineda R,
Flores-Ramírez R. Thermal burn resulting from prolonged transcutaneous pacing
in a patient with complete heart block. Am J Emerg Med. 2018
Aug;36(8):1523.e5-1523.e6. [PubMed]
9.
Hill PE. Complications of permanent transvenous cardiac pacing: a 14-year review
of all transvenous pacemakers inserted at one community hospital. Pacing Clin
Electrophysiol. 1987 May;10(3 Pt 1):564-70. [PubMed]
10.
Pinneri F, Frea S, Najd K, Panella S, Franco E, Conti V, Corgnati G.
Echocardiography-guided versus fluoroscopy-guided temporary pacing in the
emergency setting: an observational study. J Cardiovasc Med (Hagerstown). 2013
Mar;14(3):242-6. [PubMed]
11.
Ferri LA, Farina A, Lenatti L, Ruffa F, Tiberti G, Piatti L, Savonitto S. Emergent
transvenous cardiac pacing using ultrasound guidance: a prospective study versus
the standard fluoroscopy-guided procedure. Eur Heart J Acute Cardiovasc
Care. 2016 Apr;5(2):125-9. [PubMed]
12.
Zingg W, Cartier V, Inan C, Touveneau S, Theriault M, Gayet-Ageron A, Clergue
F, Pittet D, Walder B. Hospital-wide multidisciplinary, multimodal intervention
programme to reduce central venous catheter-associated bloodstream
infection. PLoS One. 2014;9(4):e93898. [PMC free article] [PubMed]
13.
Kumar K, Zarychanski R, Bell DD, Manji R, Zivot J, Menkis AH, Arora RC.,
Cardiovascular Health Research in Manitoba Investigator Group. Impact of 24-
hour in-house intensivists on a dedicated cardiac surgery intensive care unit. Ann
Thorac Surg. 2009 Oct;88(4):1153-61. [PubMed]
14.
Stephens RS, Whitman GJ. Postoperative Critical Care of the Adult Cardiac
Surgical Patient. Part I: Routine Postoperative Care. Crit Care Med. 2015
Jul;43(7):1477-97. [PubMed]
Copyright © 2022, StatPearls Publishing LLC.
This book is distributed under the terms of the Creative Commons Attribution 4.0 International License
any medium or format, as long as you give appropriate credit to the original author(s) and the source, a link is provided to
the Creative Commons license, and any changes made are indicated.
Share on Facebook
Share on Twitter
Views
PubReader
Print View
Cite this Page
In this Page
Continuing Education Activity
Introduction
Anatomy and Physiology
Indications
Contraindications
Equipment
Technique
Complications
Clinical Significance
Enhancing Healthcare Team Outcomes
Nursing, Allied Health, and Interprofessional Team Interventions
Review Questions
References
Bulk Download
Bulk download StatPearls data from FTP
Related information
PMC
PubMed
National Library of Medicine
8600 Rockville Pike
Bethesda, MD 20894
Web Policies
FOIA
HHS Vulnerability Disclosure
Help
Accessibility
Careers
NLM
NIH
HHS
USA.gov