You are on page 1of 7

European Journal of Orthodontics 36 (2014) 550–556 © The Author 2013.

r 2013. Published by Oxford University Press on behalf of the European Orthodontic Society.
doi:10.1093/ejo/cjt023 All rights reserved. For permissions, please email: journals.permissions@oup.com
Advance Access publication 18 April 2013

Finite element analysis of mono- and bicortical mini-implant


stability
Christof Holberg, Philipp Winterhalder, Ingrid Rudzki-Janson and Andrea Wichelhaus

Downloaded from https://academic.oup.com/ejo/article/36/5/550/405037 by Universitatea Tehnica ''Gheorghe Asachi'', Iasi user on 17 May 2022
Department of Orthodontics, University of Munich, Bavaria, Germany

Correspondence to: Christof Holberg, Department of Orthodontics, University of Munich, Goethestrasse 70, 80336
Munich, Germany. E-mail: cholberg@med.lmu.de

SUMMARY
Objectives  Loosening and loss rates of monocortical mini-implants are relatively high, therefore the fol-
lowing null hypothesis was tested: ‘The local bone stress in mono and bicortically-anchored mini-implants
is identical’.
Material and Methods Anisotropic Finite Element Method (FEM) models of the mandibular bone,
including teeth, periodontal ligaments, orthodontic braces, and mini-implants of varying length, were
created. The morphology was based on the Computed Tomography data of an anatomical preparation.
All mini-implants with varying insertion depths (monocortical short, monocortical long, bicortical) were
typically loaded, and the induced effective stress was calculated in the cervical area of the cortical bone.
The obtained values were subsequently analysed descriptively and exploratively using the SPSS 19.0
software.
Results  The null hypothesis was rejected, since the stress values of each anchorage type differed sig-
nificantly (Kruskal–Wallis Test, P < 0.001). Therefore, the lowest effective stress values were induced in
bicortical anchorage (mean  =  0.65 MPa, SD  =  0.06 MPa) and the highest were induced in monocortical
(short) anchorage of the mini-implants (mean = 1.79 MPa, SD = 0.29 MPa). The Spearman rank correlation
was 0.821 (P < 0.001).
Conclusions  The deeper the mini-implant was anchored, the lower were the effective stress values in
the cervical region of the cortical bone. Bicortical implant anchorage is biomechanically more favourable
than monocortical anchorage; therefore, bicortical anchorage should be especially considered in chal-
lenging clinical situations requiring heavy anchorage.

Introduction
and the magnitude of the applied force, the thread design
To avoid undesired effects on the adjacent teeth, sufficient (Wiechmann et  al., 2007), the thread diameter (Morarend
anchorage should be ensured during orthodontic tooth et al., 2009), and above all, the length of the mini-implant
movement (Angle, 1899). In addition to dental anchorage, (Mortensen et  al., 2009). The chosen screw length of the
where multiple teeth can be assembled to anchor blocks, mini-implant determines the kind of anchorage (mono- or
skeletal anchorage is a proven method, particularly in bicortical). Thus monocortical anchorage obviously reduces
challenging clinical situations (Feldmann and Bondemark, the risk of injury to the adjacent anatomic structures such as
2006). For example, the anchorage system is strongly periodontal ligaments (PDLs) (Lemieux et al., 2011).Longer
loaded during mandibular molar mesialization, as in such mini-implants, however, could probably enhance primary
cases either an extraoral (Benauwt, 1974) or a skeletal stability, leading to a reduced risk of loosening or loss (Wu
anchorage system is strongly recommended (Freudenthaler et al., 2007), even if the injury risk to adjacent anatomical
et al., 2001). For skeletal anchorage, dental implants (Gallas structures may be increased (Lemieux et  al., 2011). Thus
et al., 2005), mini-plates (Sato et al., 2007; Cornelis et al., the factors of required anchorage and injury risk should be
2008), palatal implants (Gedrange et  al., 2005; Wehrbein considered when choosing a specific mini-implant length
and Gollner, 2007), and mini-implants (Kanomi, 1997) (Sung et al, 2010; Çifter and Saraç, 2011). In challenging
are used. Mini-implants have been particularly favoured clinical situations requiring heavy anchorage, the insertion
for skeletal anchorage in recent years (Reynders et  al., of bicortically anchored mini-implants comes into
2009). However, the loss and loosening rates of these consideration. Since a detailed biomechanical comparison
usually monocortically inserted mini-implants is relatively between mono- and bicortical anchorage systems is not
high at 10–30% (Crismani et  al., 2010). Many factors available in literature as yet; this will be the subject of the
impact on the success rate (Stahl et al., 2009; Chang et al., present study. In particular, the following questions will be
2012). These include the direction (Pickard et  al., 2010) answered. Are there biomechanical differences between a
MONO- OR BICORTICAL MINI-IMPLANTS 551

monocortical short, a monocortical long, and a bicortical


insertion of the mini-implants? Are the biomechanical
differences of importance? Do the differences affect the
quality of the primary stability? Thus, the null hypothesis
was, ‘The load of the peri-implant bone in mono or bicortical
anchorage is identical’.

Downloaded from https://academic.oup.com/ejo/article/36/5/550/405037 by Universitatea Tehnica ''Gheorghe Asachi'', Iasi user on 17 May 2022
Material and methods
An anatomical preparation of a mandibular segment
(18  year old man) with tooth 46 missing served as the
morphological basis for the Finite Element Method (FEM)
models (Figure 1). This anatomical specimen was scanned
using Computed Tomography (CT), and the morphological
data were acquired by manual segmentation of the axial CT
layers using the Amira™ software (Visage Imaging GmbH,
Berlin, Germany). Segment models of the compact and can-
cellous mandible, complete models of tooth 44, 45, and 47,
and models of the corresponding PDLs resulted after cross-
linking the point cloud (Delauney Triangulation) to three-
dimensional polygon meshes. The surface structure of the
polygon meshes was subsequently post-processed and opti-
mized using the Rapidform™ software (INUS Technology
Inc., Seoul, Korea).

Pre-processing
Using Polytrans™ software (Okino Inc., Ontario, Figure 2  Cross-section of the CAD models showing the variable length
Canada), the polygon meshes were converted into of the mini-implant that was anchored mono- or bicortically.
analytical solid models (IGES format). All constructible
elements (brackets, powerarm and mini-implant)
and subtraction in the Mechanical Desktop™ software
were designed virtually using established Computer
(Autodesk GmbH, Munich, Germany). The first CAD
Aided Design (CAD) tools in the Inventor™ software
model represented short monocortical anchorage by using
(Autodesk GmbH, Munich, Deutschland). The resulting
a mini-implant with a diameter of 1.6 mm and a length
solid models could be combined to three compound
of 5 mm. The mini-implant in the second model (7 mm
models (Figure  2) using Boolean operations of addition
length) was also inserted monocortically but reached the
region of the cancellous bone. Finally, in the third CAD
model bicortical anchorage was performed by using
a mini-implant with a diameter of 1.6 mm and a length
of 10 mm. The resulting CAD models could now be
cross-linked three-dimensionally to FEM models using
the ANSYS™ 11.0 simulation software (Ansys Inc.,
Canonsburg, Pennsylvania, USA). The finite elements
used were parabolic tetrahedrons with one node at
each corner and one additional node in the middle of
each edge. Accordingly, this tetrahedral element was a
nonlinear, 10-noded three-dimensional finite element.
Loading and boundary conditions (Tables 1 and 2) were
defined corresponding to the standard values found in the
literature. A force and a counterforce each of 1.5 N was
applied to all simulation models. The orientation of the
force vector corresponded to the closed coil depicted in
Figure 1. As a boundary condition, multiple nodes could
Figure 1  CAD model consisting of various materials (cortical bone, can-
cellous bone, teeth with enamel, dentin and periodontal ligament, brackets, be fixed in the simulation model as far away as possible
powerarm, and a mini-implant with a variable length). from the region of interest (ROI). The contact conditions
552 C. HOLBERG ET AL.

Table 1  Material properties of the mini-implant and other materials used in the present study; Young´s modulus specifies the elasticity
and Poisson´s ratio the transverse contraction characteristics of the material. 

Young´s modulus Poisson´s ratio References

Mini-implant 110 000 MPa 0.35 (Motoyoshi et al., 2009)


Brackets 210.0 MPa 0.30 (Chen et al., 2008)

Downloaded from https://academic.oup.com/ejo/article/36/5/550/405037 by Universitatea Tehnica ''Gheorghe Asachi'', Iasi user on 17 May 2022
Adhesive 8.823 MPa 0.25 (Lin et al., 2011)
Powerarm (Stainless Steel) 193.0 MPa 0.30 (Chen et al., 2008)
Enamel 80 000 MPa 0.25 (He et al., 2006)
Dentin 24 400 MPa 0.43 (Xu et al., 1998; Kinney et al., 2004)
PDL 50 MPa 0.49 (Rees and Jacobsen, 1997; Fill et al., 2011)
Nerve tissue 0.58 MPa 0.42 (Ichihara et al., 2001; Borschel et al., 2003)

between the structures were defined as fixed—only the


Table 2  Anisotropic material properties of cortical and contact between the bone and the mini-implant was
cancellous bone used in the present study (Dechow et al., 1993;
O’Mahony et al., 2000; O’Mahony et al., 2001). 
friction dependent with non-linear lifting (coefficient
µ = 0.3).

Compact Bone Cancellous Bone Stress calculation


Ex 12 600 MPa 1148 MPa An ROI was defined, where the effective stress (in MPa)
Ey 19 400 MPa 1148 MPa should be calculated. This ROI was formed by the buccal
Ez 12 600 MPa 210 MPa
part of the cortical bone surrounding the head of the mini-
Gxy 5700 MPa 434 MPa
Gyz 5700 MPa 68 MPa implant by a diameter of 8 mm (Figure  3). In this region
Gxz 4850 MPa 68 MPa the effective stress (in MPa) was calculated at all nodes.
vxy 0.253 0.322 Following this, the values were sorted by their size, and the
vyz 0.390 0.055
vxz 0.300 0.055 top 1000 highest stress values were considered for further
statistical analysis, as these peak loads could be responsible

Figure 3  Comparison of induced stress values (in MPa) in the region of interest (cortical bone around the head of the mini-implant). The implant body
itself was removed for a better view. Cross-section (left row) showing the distribution of the induced effective stresses (in MPa).
MONO- OR BICORTICAL MINI-IMPLANTS 553

for possible mini-implant complications such as loss of the variant of the monocortical mini-implant showed somewhat
primary stability, loosening, or even loss of the mini-implant. lower stress values. The mean effective stress was 1.55 MPa
(SD 0.28 MPa). The median was 1.48 MPa, the maximum
Statistical analysis 2.43 MPa, and the minimum 1.22 MPa (Table 3).
The obtained values were analysed descriptively and explora-
Bicortical anchorage
tively using the SPSS™ 15.0 software. The Kolmogorov–

Downloaded from https://academic.oup.com/ejo/article/36/5/550/405037 by Universitatea Tehnica ''Gheorghe Asachi'', Iasi user on 17 May 2022
Smirnov test was performed to check the variables for normal After applying a force of 1.5 N perpendicular to the axis
distribution. The Kruskal–Wallis test was used to find sig- of the monocortical short-inserted mini-implant, the mean
nificant differences between the samples; following this, the effective stress in the cortical bone was 0.65 MPa (SD
Mann–Whitney U-test was used in pairs. The correlation 0.09 MPa). The median was 0.64 MPa, the maximum value
coefficient according to Spearman was calculated to quantify 0.90 MPa, and the minimum 0.57 MPa (Table 3). The box
the correlation between the length of the mini-implant and plots in Figure 5 illustrate the statistical differences between
the induced effective stress in the peri-implant bone. the three mini-implant anchorage variants.

Statistical analysis
Results
The variable of effective stress was not normally distributed
The samples for the anchorage variants ‘monocortical
in all types of anchorage. The hypothesis of a normal dis-
short’, ‘monocortical long’, and ‘bicortical’ differed with
tribution could be rejected in the Kolmogorov–Smirnov test
regard to their descriptive statistics. The mean effective
stress (in MPa) was clearly higher in mono than in bicortical
anchorage (Figure 3). The peri-implant stress values in the Table 3  Descriptive statistics of effective stress values (n=1000)
cortical bone for the monocortical variant were more than calculated in the cervical region of the cortical bone close to the
twice those for the bicortical variant. The individual values, mini-implant. 
which were sorted in a descending order, are depicted in
Figure 4 to give a comparison between the anchorage types. Parameter Unit Monocortical Monocortical Bicortical
(short) (long)
Monocortical anchorage
n — 1000 1000 1000
After applying a force of 1.5 N perpendicular to the axis Mean MPa 1.79 1.55 0.65
of the monocortical short-inserted mini-implant, the mean Median MPa 1.71 1.48 0.64
effective stress in the cortical bone was 1.79 MPa (SD Std.-Dev. MPa 0.29 0.28 0.06
Minimum MPa 1.43 1.22 0.57
0.29 MPa). The median was 1.71 MPa, the maximum value Maximum MPa 2.76 2.43 0.90
2.76 MPa, and the minimum value 1.43 MPa. The long

Figure 4  Comparison of the peak stress values with different types of


anchorage calculated at the nodes of the region of interest (cortical bone Figure 5  Boxplot visualization comparing the stress values in the region
around the head of the mini-implant). In the chart all presented nodes were of interest (cortical bone around the head of the mini-implant) with three
rearranged by their value size. different types of mini-implant anchorage.
554 C. HOLBERG ET AL.

with a very high significance (P < 0.001). Due to the lack of 2010; Lombardo et al., 2010). This virtual design of organic
normal distribution, non-parametric tests were used. In the structures increases the methodical error. To keep this as
Kruskal–Wallis test the mean values of the samples differed low as possible, the FEM models of the present study were
with high significance (P < 0.001). A  pairwise testing of not constructed but rather generated from the CT data of
the samples using the Mann–Whitney U-test yielded highly an anatomical preparation. To further reduce the systematic
significant differences for each variant (P < 0.001). The error, no absolute values were considered to draw the
rank correlation according to Spearman was in the negative conclusion, only the differences between the simulations.

Downloaded from https://academic.oup.com/ejo/article/36/5/550/405037 by Universitatea Tehnica ''Gheorghe Asachi'', Iasi user on 17 May 2022
range at −0.821 (P < 0.001), which corresponds to a high Since all simulations were affected by the simplification
correlation. When only including the 10 highest stress val- effects to the same extent, the analysis of the differences
ues in the samples, the correlation coefficient according to resulted in an additional increase of validity.
Spearman was −0.943 (P < 0.001), which corresponds to a
very high correlation. The negative correlation coefficients Cortical stress
indicated that the peri-implant stress values were lower
The null hypothesis (‘The local bone stress in mono and
when the mini-implant was longer and inserted deeper into
bicortically-anchored mini-implants is identical’) could
the alveolar bone.
be discarded since highly significant differences were
determined between the different anchorage types. Since
Discussion in bicortical anchorage the peak stresses induced in the
cortical bone were the lowest, this type of anchorage
The FEM is a proven method to analyse mini-implant-
seems to be superior according to a long-lasting primary
based anchorage systems biomechanically (Motoyoshi
stability. Respectively, the risk of mini-implant loss or
et  al., 2005; Stahl et  al., 2009; Chatzigianni et  al., 2011;
loosening should be reduced in bicortical anchorage. The
Ammar et  al., 2011; Suzuki et  al., 2011; Woodall et  al.,
present results are supported by the study of Brettin et al.
2011; Liu et al., 2012; Singh et al., 2012). In the present
(2008) in which the deflection forces of mono or bicor-
study, three anchorage variants of mini-implants were
tically anchored mini-implants were calculated using a
compared biomechanically using the FEM. As in other
simplified FEM model and an anatomical preparation.
FEM studies, the FEM models represented an idealization
Here the bicortically anchored mini-implants underwent
and simplification of the reality (Cattaneo et  al., 2005;
less loosening than the monocortical ones. Additionally,
Stahl et al., 2009; Pickard et al. 2010; Chatzigianni et al.,
biomechanical differences between variants of monocor-
2011). For example, the mechanical properties of the PDL
tical anchorage could be analysed in the present study.
are very complex in reality (Poppe et al., 2002). Different
Since the short monocortical anchorage is limited to the
approaches are taken to consider this complexity in FEM
outer cortex, the stresses were higher than with longer
studies; for instance, sometimes the material properties of
mini-implants, which are extended into the cancellous
the PDL are defined as linear (Rees and Jacobsen, 1997;
bone. The stress differences between these monocortical
Poiate et  al., 2009; Fill et  al., 2011), but this approach
variants were highly significant, but the extent of the dif-
often seems to be an oversimplification. Often the material
ferences was much lower than between monocortical and
properties are defined as ‘bilinear’, leading to an ‘ultimate
bicortical anchorage.
strain’ value between the linear ranges (Ziegler et al., 2005;
Cattaneo et al., 2009; Dong-Xu et al., 2011). This approach
Monocortical anchorage
is more accurate because both ‘curled’ and ‘uncurled’
PDL modes are taken into account (Ziegler et  al., 2005; For a sustainable primary stability of orthodontic mini-
Dong-Xu et al., 2011). Another approach is the description implants it is important to prevent loosening of the mini-
of the PDL by the hyper-elastic Ogden constitutive law implant by overloading the peri-implant bone. In particular,
(Shibata et  al., 2006). In the present study, the material the stress induced in the cervical region of the peri-implant
properties of the PDL were defined as linear. This bone is important here (Singh et al., 2012). The lower the
simplification was acceptable, as the PDL was far away induced stress is here, the lower the risk of loosening of the
from the ROI and had only a small effect on the results. mini-implant (Florvaag et al., 2010). With short monocorti-
In the present study, the morphology of the bone was cal anchorage, however, relatively high stresses are induced
differentiated into a cortical and cancellous segment, and in the region of the cervical peri-implant bone. These were
the corresponding anisotropic material properties could somewhat higher than in long monocortical anchorage but
be assigned. All the material properties used were taken clearly higher than in bicortical anchorage. Not only was the
from the literature. In comparison to other FEM studies, peri-implant bone in the distal region of the mini-implant
the morphological accuracy of the present FEM model affected by stresses (where the tensile force was applied by
was relatively high, because the alveolar bone segments the closed coil) so too was the mesial region. Obviously,
were sometimes virtually constructed by the authors and the head of the short mini-implant was pulled in a distal
not transferred from individual patient data (Pickard et al., direction, causing stresses in the mesial region by a leverage
MONO- OR BICORTICAL MINI-IMPLANTS 555

effect of the implant body. This interesting effect was not Chen C S, Hsu M L, Chang K D, Kuang S H, Chen P T, Gung Y W 2008
Failure analysis: enamel fracture after debonding orthodontic brackets.
found in long monocortical anchorage, including anchor- The Angle Orthodontist 78: 1071–1077
age in the cancellous bone. Apparently peri-implant stresses Çifter M, Saraç M 2011 Maxillary posterior intrusion mechanics with mini-
caused by this leverage effect can be prevented by choosing implant anchorage evaluated with the finite element method. American
longer mini-implants. Journal of Orthodontics and Dentofacial Orthopedics 140: e233–e241
Cornelis M A et  al. 2008 Orthodontic loading of titanium miniplates in
dogs: microradiographic and histological evaluation. Clinical Oral

Downloaded from https://academic.oup.com/ejo/article/36/5/550/405037 by Universitatea Tehnica ''Gheorghe Asachi'', Iasi user on 17 May 2022
Bicortical anchorage Implants Research 19: 1054–1062
The cervical bone stress was clearly lower in the bicortical Crismani A G, Bertl M H, Celar A G, Bantleon H P, Burstone C J 2010
Miniscrews in orthodontic treatment: review and analysis of published
variant than in all variants of monocortical anchorage. The clinical trials. American Journal of Orthodontics and Dentofacial
reduction factor was 2–3, leading to low stresses of the cer- Orthopedics 137: 108–113
vical bone in the direction of the applied force. In contrast Dechow P C, Nail G A, Schwartz-Dabney C L, Ashman R B 1993 Elastic
to monocortical anchorage, the mesial region of the peri- properties of human supraorbital and mandibular bone. American
Journal of Physical Anthropology 90: 291–306
implant bone did not show relevant stresses. Obviously, in
Dong-Xu L, Hong-Ning W, Chun-Ling W, Hong L, Ping S, Xiao Y
bicortical anchorage the mini-implant is completely fixed 2011 Modulus of elasticity of human periodontal ligament by optical
by the opposite cortical bone, effectively preventing a lever- measurement and numerical simulation. The Angle Orthodontist 81:
age effect of the implant body. The overall stresses of the 229–236
cervical bone are very low in bicortical anchorage; thus the Feldmann I, Bondemark L 2006 Orthodontic anchorage: a systematic
review. The Angle Orthodontist 76: 493–501
risk of loosening or loss of the mini-implant is reduced in
Fill T S, Carey J P, Toogood R W, Major P W 2011 Experimentally
comparison to monocortical anchorage. determined mechanical properties of, and models for, the periodon-
tal ligament: critical review of current literature. Journal of Dental
Biomechanics 2011: 312980
Conclusion Florvaag B et  al. 2010 Biomechanical properties of orthodontic minis-
crews. An in-vitro study. Journal of Orofacial Orthopedics = Fortschritte
From a biomechanical point of view bicortical mini-implant der Kieferorthopädie 71: 53–67
anchorage is more favourable than monocortical anchorage, Freudenthaler J W, Haas R, Bantleon H P 2001 Bicortical titanium screws
because the induced stress in the peri-implant bone is less. for critical orthodontic anchorage in the mandible: a preliminary report
on clinical applications. Clinical Oral Implants Research 12: 358–363
Therefore, bicortical anchorage should be especially con-
Gallas M M, Abeleira M T, Fernández J R, Burguera M 2005 Three-
sidered in challenging clinical situations requiring heavy dimensional numerical simulation of dental implants as orthodontic
anchorage. This implies a beneficial morphology, allowing anchorage. European Journal of Orthodontics 27: 12–16
a bicortical insertion without increased risk of injury to the Gedrange T, Hietschold V, Mai R, Wolf P, Nicklisch M, Harzer W 2005
adjacent structures. An evaluation of resonance frequency analysis for the determination of
the primary stability of orthodontic palatal implants. A study in human
cadavers. Clinical Oral Implants Research 16: 425–431
References He L H, Fujisawa N, Swain M V 2006 Elastic modulus and stress-strain
response of human enamel by nano-indentation. Biomaterials 27:
Ammar H H, Ngan P, Crout R J, Mucino V H, Mukdadi O M 2011 Three- 4388–4398
dimensional modeling and finite element analysis in treatment planning Ichihara K, Taguchi T, Shimada Y, Sakuramoto I, Kawano S, Kawai S 2001
for orthodontic tooth movement. American Journal of Orthodontics and Gray matter of the bovine cervical spinal cord is mechanically more rigid
Dentofacial Orthopedics 139: e59–e71 and fragile than the white matter. Journal of Neurotrauma 18: 361–367
Angle E H 1899 Classification of malocclusion. Dental Cosmos 41: Kanomi R 1997 Mini-implant for orthodontic anchorage. Journal of
350–357 Clinical Orthodontics 31: 763–767
Benauwt A 1974 [Use of the Delaire facial mask for the mesial movement Kinney J H, Gladden J R, Marshall G W, Marshall S J, So J H, Maynard J
of the mandibular arch]. L’ Orthodontie française 45: 252–256 D 2004 Resonant ultrasound spectroscopy measurements of the elastic
Borschel G H, Kia K F, Kuzon W M Jr, Dennis R G 2003 Mechanical prop- constants of human dentin. Journal of Biomechanics 37: 437–441
erties of acellular peripheral nerve. The Journal of Surgical Research Lemieux G et  al. 2011 Computed tomographic characterization of
114: 133–139 mini-implant placement pattern and maximum anchorage force in
Brettin B T et  al. 2008 Bicortical vs monocortical orthodontic skel- human cadavers. American Journal of Orthodontics and Dentofacial
etal anchorage. American Journal of Orthodontics and Dentofacial Orthopedics 140: 356–365
Orthopedics 134: 625–635 Lin C L, Huang S F, Tsai H C, Chang W J 2011 Finite element sub-mode-
Cattaneo P M, Dalstra M, Melsen B 2005 The finite element method: a ling analyses of damage to enamel at the incisor enamel/adhesive inter-
tool to study orthodontic tooth movement. Journal of Dental Research face upon de-bonding for different orthodontic bracket bases. Journal of
84: 428–433 Biomechanics 44: 134–142
Cattaneo P M, Dalstra M, Melsen B 2009 Strains in periodontal ligament Liu T C, Chang C H, Wong T Y, Liu J K 2012 Finite element analysis of
and alveolar bone associated with orthodontic tooth movement analyzed miniscrew implants used for orthodontic anchorage. American Journal
by finite element. Orthodontics & Craniofacial Research 12: 120–128 of Orthodontics and Dentofacial Orthopedics 141: 468–476
Chang J Z et al. 2012 Effects of thread depth, taper shape, and taper length Lombardo L, Gracco A, Zampini F, Stefanoni F, Mollica F 2010 Optimal
on the mechanical properties of mini-implants. American Journal of palatal configuration for miniscrew applications. The Angle Orthodontist
Orthodontics and Dentofacial Orthopedics 141: 279–288 80: 145–152
Chatzigianni A, Keilig L, Duschner H, Götz H, Eliades T, Bourauel C 2011 Morarend C et  al. 2009 Effect of screw diameter on orthodontic skel-
Comparative analysis of numerical and experimental data of orthodontic etal anchorage. American Journal of Orthodontics and Dentofacial
mini-implants. European Journal of Orthodontics 33: 468–475 Orthopedics 136: 224–229
556 C. HOLBERG ET AL.

Mortensen M G, Buschang P H, Oliver D R, Kyung H M, Behrents R G Shibata T, Botsis J, Bergomi M, Mellal A, Komatsu K 2006 Mechanical
2009 Stability of immediately loaded 3- and 6-mm miniscrew implants behavior of bovine periodontal ligament under tension-compression
in beagle dogs–a pilot study. American Journal of Orthodontics and cyclic displacements. European Journal of Oral Sciences 114: 74–82
Dentofacial Orthopedics 136: 251–259 Singh S, Mogra S, Shetty V S, Shetty S, Philip P 2012 Three-dimensional
Motoyoshi M, Inaba M, Ueno S, Shimizu N 2009 Mechanical anisot- finite element analysis of strength, stability, and stress distribution in
ropy of orthodontic mini-implants. International Journal of Oral and orthodontic anchorage: a conical, self-drilling miniscrew implant sys-
Maxillofacial Surgery 38: 972–977 tem. American Journal of Orthodontics and Dentofacial Orthopedics
Motoyoshi M, Yano S, Tsuruoka T, Shimizu N 2005 Biomechanical effect 141: 327–336

Downloaded from https://academic.oup.com/ejo/article/36/5/550/405037 by Universitatea Tehnica ''Gheorghe Asachi'', Iasi user on 17 May 2022
of abutment on stability of orthodontic mini-implant. A finite element Stahl E, Keilig L, Abdelgader I, Jäger A, Bourauel C 2009 Numerical
analysis. Clinical Oral Implants Research 16: 480–485 analyses of biomechanical behavior of various orthodontic anchorage
O’Mahony A M, Williams J L, Katz J O, Spencer P 2000 Anisotropic elas- implants. Journal of Orofacial Orthopedics  =  Fortschritte der
tic properties of cancellous bone from a human edentulous mandible. Kieferorthopädie 70: 115–127
Clinical Oral Implants Research 11: 415–421 Sung S J, Jang G W, Chun Y S, Moon Y S 2010 Effective en-masse
O’Mahony A M, Williams J L, Spencer P 2001 Anisotropic elasticity of retraction design with orthodontic mini-implant anchorage: a finite
cortical and cancellous bone in the posterior mandible increases peri- element analysis. American Journal of Orthodontics and Dentofacial
implant stress and strain under oblique loading. Clinical Oral Implants Orthopedics 137: 648–657
Research 12: 648–657 Suzuki A, Masuda T, Takahashi I, Deguchi T, Suzuki O, Takano-Yamamoto
Pickard M B, Dechow P, Rossouw P E, Buschang P H 2010 Effects of T 2011 Changes in stress distribution of orthodontic miniscrews and sur-
miniscrew orientation on implant stability and resistance to failure. rounding bone evaluated by 3-dimensional finite element analysis. American
American Journal of Orthodontics and Dentofacial Orthopedics 137: Journal of Orthodontics and Dentofacial Orthopedics 140: e273–e280
91–99 Wehrbein H, Gollner P 2007 Skeletal anchorage in orthodontics – basics
Poiate I A, de Vasconcellos A B, de Santana R B, Poiate E 2009 Three- and clinical application. Journal of Orofacial Orthopedics 68: 443–461
dimensional stress distribution in the human periodontal ligament in Wiechmann D, Meyer U, Büchter A 2007 Success rate of mini- and micro-
masticatory, parafunctional, and trauma loads: finite element analysis. implants used for orthodontic anchorage: a prospective clinical study.
Journal of Periodontology 80: 1859–1867 Clinical Oral Implants Research 18: 263–267
Poppe M, Bourauel C, Jäger A 2002 Determination of the elasticity Woodall N, Tadepalli S C, Qian F, Grosland N M, Marshall S D, Southard
parameters of the human periodontal ligament and the location of the T E 2011 Effect of miniscrew angulation on anchorage resistance.
center of resistance of single-rooted teeth a study of autopsy specimens American Journal of Orthodontics and Dentofacial Orthopedics 139:
and their conversion into finite element models. Journal of Orofacial e147–e152
Orthopedics = Fortschritte der Kieferorthopädie 63: 358–370 Wu J C, Huang J N, Zhao S F 2007 Bicortical microimplant with 2 anchor-
Rees J S, Jacobsen P H 1997 Elastic modulus of the periodontal ligament. age heads for mesial movement of posterior tooth in the beagle dog.
Biomaterials 18: 995–999 American Journal of Orthodontics and Dentofacial Orthopedics 132:
Reynders R, Ronchi L, Bipat S 2009 Mini-implants in orthodontics: a sys- 353–359
tematic review of the literature. American Journal of Orthodontics and Xu H H et  al. 1998 Indentation damage and mechanical properties of
Dentofacial Orthopedics 135: 564.e1–19 human enamel and dentin. Journal of Dental Research 77: 472–480
Sato R, Sato T, Takahashi I, Sugawara J, Takahashi N 2007 Profiling of Ziegler A, Keilig L, Kawarizadeh A, Jäger A, Bourauel C 2005 Numerical
bacterial flora in crevices around titanium orthodontic anchor plates. simulation of the biomechanical behaviour of multi-rooted teeth.
Clinical Oral Implants Research 18: 21–26 European Journal of Orthodontics 27: 333–339

You might also like