ELECTRONIC ENTERPRISES (INDIA) PVT. LTD.
PARA ELECTRONICS ~ MANUFACTURING DIVISION
Unit 306, Nimesh Industrial Estate, 90 Feet Cross Road,
Mulund (East) , Mumbai, - 400 081. India
Phone : 91-022-2563 5600) 91-022-2563 2741
Fax : 91-022-2563 7835
Email; pe.servicing@ecipLin Website: www ceipl.in
Certificate No. SER/14-15/055, Date: 29.11.2014
CALIBRATION CERTIFICATE
This is to certify that, Plating Thickness Tester, Model SIDET Sr. No. 2004317
including Pump Unit and with Rubber Gasket with etching hole Average
diameter 2.40 is calibrated against Kocour USA make Plating Thickness Standard,
Nickel on Steel; Sr. No. Ni-52695. This Nickel Plating Standard is manufactured
by Kocour Company Inc. Chicago, USA and is traceable to National Institute of
Standard and Technology, USA.
Following are the calibration results:
Standard Thickness Measured Thickness
13.3 Microns + 5% 13.23 Mierons
Calibration Results are well within specifications of SIDET and are sat
‘Next Calibration is due on 29.11.2015
For ELECTRONIC ENTERPRISES (1) PVT. LTD.
PARA ELECTRONICS ~ MANUFACTURING DIVN.
Authorized Signatory
Registered Office
Electronic Enterprises (I) Prt. Lid,
215-216, Regal Indusrial Estate, Acharya Donde Marg, Sewri, Mumbai, 400 015
Phone : 91-22-241370%6 /91-022-24147970 Fax : 91-022-24133341 / 91-022-24153894
Email : spihale@eeipl.inCOMPANY: Electronic Enterprises (I) Pvt 14 August 2014
KOCOUR ORDER #: 101774
CERTIFICATION
Serial #: Ni-52695
COATING # 1: Nickel ‘THICKNESS: 0.52 MIL: 13,3 um
OVER
COATING # 2: ‘THICKNESS:
OVER
COATING #3:_ THICKNESS:
OVER
SUBSTRATE: Stee!
‘This certifies that the above Kocour thickness standard was prepared at ambient
temperature by the weight/area method and was weighed on a Mettler Semi-Micro Analytical
Balance, Model AE-163, Serial Number E-13040, having weights within Class "E2" tolerances
‘of the National Institute of Standards and Technology (NIST). It is further certified that this
balance was calibrated on 08 January 2014, and is due for re-calibration in January, 2015 by the
Mettler Instrument Corporation. This service includes accuracy test with Mettler test weights #
43 calibrated on 13 March 2013 traceable to the National Institute of Standards and Technology
(NIST) # MTSO61 and if required, adjustments made to original specifications.
This standard isa consumable item and cannot be re-calibrated or re-certified. The
calibration and re-calibration dates listed above reference the certifying balance only.
Kocour Company
Brrer 8 PRA,
Bruce L. Polak
Labertory Mesagic
KOCOUR COMPAKY INC, 4800 SOUTH ST. LOUIS AVENUE - CHICAGO, E0532 USA
TEL (79) 7-111 AX (73) 847080