3590ET / 3590EGT

Touch screen indicator

CONNECTION DIAGRAMS ENGLISH

www.diniargeo.com
Index
Introduction  4

Electrical warnings  5

Electrical precautions  5

Cable classification  5

Recommended distance between cables  6

Maximum permitted length of cables  6

Earthing system  7

Example of weighbridge earthing  8

Example of silos earthing  9

Motherboard  10

I/O expansion board  12

Optional internal board for rapid connection of multiple scales  13

RJ45 RS232 connector board  13

RS232 / RS485 to Ethernet / WIFI conversion board  14

USB support data saving board  16

RADIO communication board  17

PROFIBUS communication board  17

BLUETOOTH communication board  18

RADIO REMOTE CONTROL board  18

Circuit breaker against electrostatic loads  18

Analog output  19

DAC16O-X  19

DAC16OSER-X  20

Traffic light  21

Chip 485  21

RS485 opto-isolated board  22

Display board  22

3590ET / 3590EGT 3
TECH_MAN_ENG_SCHEMES_3590ET_3590EGT
Introduction

Dear Customer,
Thank you for purchasing a DINI ARGEO product.

This manual contains all instructions for proper installation of the weight indicator with touch screen series 3590.
We thank you for purchasing this scale and we kindly ask you to take note of certain aspects of this manual:

This booklet provides useful information for proper operation and maintenance of the scale referred to herein. It is important to pay the
utmost attention to all the sections that illustrate the simplest and safest way to operate the device.

This publication, or portions thereof, may not be duplicated without written permission from the Manufacturer.

All information herein is based on the data available at the time of publication. The Manufacturer reserves the right to make changes
to its products at any time without notice and without incurring any penalty. We therefore recommend that you always check for any
updates.

N.B.: The individual in charge of operating the scale must ensure that all safety regulations in force in the country of use are applied,
ensuring that the appliance is used in accordance with the purpose it is intended for and to avoid any danger for the user.
The manufacturer declines any liability arising from any weighing operation errors.

We recommend carefully following the instructions when programming the weight indicator, as to do otherwise could jeopardise proper
scale operation.

This manual was written with the utmost care but we always welcome feedback on any inaccuracies you may find.

The instrument is covered by warranty and MUST NOT BE TAMPERED WITH BY THE USER for any reason.
Any attempt to repair or modify the device could expose the user to the risk of electrical shock and will render all warranty conditions
null, thus releasing the Manufacturer from all liability.

All problems with the unit or system must be communicated to the manufacturer or the dealer from which it was purchased. In any
case, DISCONNECT POWER before any operations.

3590ET / 3590EGT 4
TECH_MAN_ENG_SCHEMES_3590ET_3590EGT
Electrical warnings

Electrical precautions
• Mains power supply adjusted within ± 10% of the nominal voltage
• Electrical protections (fuses, etc.) are to be provided by the installer.
• Comply with the minimum recommended distances between different categories of cables.
• The load cell or signal amplifier extension cables, which are used for serial port and analog output connection, must com-
ply with the maximum permitted lengths.
• The load cell or signal amplifier extension cables, which are used for serial port and analog output connection, must be
shielded and must also be inserted alone into the conduit or metal pipe.
• Cell or amplifier cable input into the electrical panel must be autonomous. If possible, they must be connected directly to
the indicator’s terminal board without passing through the conduit with other cables.
• Install an “RC” filter on coil contactors, solenoid valves and all devices that generate electrical interferences.
• If condensation can develop inside the scale’s transmitter, it is recommended to keep the equipment running.
• With regard to all shielded and non-shielded cables (cell cable, PC cable, power supply cable, etc.) that are connected to
the indicator, you must keep the cable as short as possible and make a minimum amount of the cables exit from the shield
in order to be connected to the terminal board;
• If the indicator is situated inside an electrical panel, you must also use a shielded cable for power supply and must keep
the cable as short as possible and far from the cables supplying coils, inverter, electromotive force, etc., and apply a
de-coupling transformer to supply the indicator only.

Cable classification
The various cables are classified according to the signals they transmit:

Category I
• Fieldbus , LAN network ( PROFIBUS, Ethernet, Devicenet...)
• Data shielded cables (RS232 ...)
• Shielded cables for analog digital signals < 25V (sensors, load cells.)
• Low-voltage power supply cables (<60V)
• Coaxial cables

Category II
• DC power supply cables with >60V and <400V
• AC power supply cables with >25V and <400

Category III
• Power supply cables with >400V
• Telephone cable

Category IV
• Any cables subject to lightning hazards

3590ET / 3590EGT 5
TECH_MAN_ENG_SCHEMES_3590ET_3590EGT
Electrical warnings

Recommended distance between cables


• When cables are laid parallel, they must be at the distances specified in the figure below
• These distances are intended in the air and are reduced if the housings are separated by metal boards connected to the
ground
• Cables from different categories can cross (90°)

Category I Category II Category III Category IV

≥ 100 mm
≥ 200 mm
≥ 500 mm
≥ 100 mm
≥ 500 mm
≥ 500 mm

Maximum permitted length of cables


LOAD CELL CONNECTION CABLE
The relative cable can reach the following maximum length from the line in order to connect the load cells:
• 50 metres with a 6 x 0.25 mm2 cable
• 100 metres with a 6 x 0.5 mm2 cable

RS232 CONNECTION CABLE


The relative cable can reach a maximum length of approximately 15 metres from the line for RS 232 connections, with a baud
rate of up to 19200.

RS485 CONNECTION CABLE


The relative cable can reach a maximum length of approximately 1200 metres from the line with a baud rate of up to 9600.

ANALOGUE OUTPUT CONNECTION CABLE


The maximum length for current analog output is:
• 100 metres with a 2 x 0.25 mm2 cable
• 150 metres with a 2 x 0.5 mm2 cable
• 300 metres with a 2 x 1 mm2 cable

The maximum length for voltage analog output is:


• 50 metres with a 2 x 0.25 mm2 cable
• 75 metres with a 2 x 0.5 mm2 cable
• 150 metres with a 2 x 1 mm2 cable

3590ET / 3590EGT 6
TECH_MAN_ENG_SCHEMES_3590ET_3590EGT
Electrical warnings

Earthing system
For correct earthing and perfect system operation, you must connect the indicator, load cells, any junction boxes and weighing
structure to earth.

INDICATOR
Connect the container’s external earth terminal to earth with copper cables having a minimum dimension of 16 mm2.

LOAD CELLS AND JUNCTION BOX


Earthing must be carried out by connecting the earthing cables to the earth bar; the cables must have a minimum dimension of
16 mm2. Also, connect the earth bar to the earth pole with a minimum cable of 50 mm2.

• If the load cells are connected to the indicator via a junction box, you must connect the cable shield from the indicator and
the cell cable shields to the earth terminal of the junction box (refer to the junction box manual) and connect the latter to
earth by means of a copper cable having a minimum dimension of 16 mm2.
• If the load cells are connected directly to the indicator (without using a junction box), you must connect the cell cable
shields to the earthing point (or earth bar) inside the container.
• If the weighing system is for a large and/or outdoor structure, such as a weighbridge, and the junction box is connected to
the indicator with a cable longer than 10 m, or if there are interferences, connect the cable’s braiding from the indicator to
the earth terminal in the junction box and indicator and connect the two earth systems with an earth cable having a mini-
mum dimension of 16 mm2.

WEIGHING STRUCTURE
Connect the weighing structure and any other structures that are not connected (for example silos that leave material on the
weighing structure) to earth with cables having a minimum dimension of 16 mm2.
You must also connect the upper part of each cell with the lower part of the cell with a copper braid having a minimum dimen-
sion of 16 mm2. The upper part must be short circuited with the surface of the weighing structure and the lower part must be
connected to earth with a copper braid having a minimum dimension of 16 mm2.

SERIAL CABLES AND INSTRUMENTS CONNECTED


Connect the serial cable shield to the earthing point (or earth bar) inside the container (on the end part of the cable towards
indicator) and to the earth terminal of the connected instrument (on the end part of the cable towards the instrument connected),
and connect the earth terminal of the connected instrument to earth with copper cables having a minimum dimension of 16 mm2.
To avoid undesired effects, the earthing references of the connection cable and power supply of the instrument connection and
the indicator must have the same power.

i NOTES:

• ll earth cables must have a suitable length in order to obtain an overall resistance of the earthing system
below 1Ω.

• With regard to large and/or outdoor weighing system structures, such as a weighbridge:
- connection to earth must be carried out by connecting the earth cables to an earth bar, and the earth
bar to an earth pole with a minimum cable of 50 mm2.

- The cables must be larger (e.g. 50 mm2 instead of 16 mm2 and 100 mm2 instead of 50 mm2) since the
voltage involved is higher (e.g. lightning);

- the earth pole must be situated at a minimum distance of 10 m from the weighbridge structure;

• You must open SENSE-REFERENCE inside the indicator in order to compensate for the shift due to an increase in
temperature.

• You must check and, if necessary, remove the connection between electrical installation earth and neutral.

3590ET / 3590EGT 7
TECH_MAN_ENG_SCHEMES_3590ET_3590EGT
Electrical warnings

Example of weighbridge earthing

Weight indicator

Junction box situated on


the coast wall of the ditch

Load cell

Weighbridge

Earth cables
Ø 8-sec 50 mm2
Copper perforated plate Ø 11,3 - sec. 100 mm2
placed on the coast wall

Weighbridge

Earth pole placed


under or next to
the weighbridge

Load cell
bypass U-bolt

3590ET / 3590EGT 8
TECH_MAN_ENG_SCHEMES_3590ET_3590EGT
Electrical warnings

Example of silos earthing

Load cell
protection earth
bypass

Load cell

3590ET / 3590EGT 9
TECH_MAN_ENG_SCHEMES_3590ET_3590EGT
Motherboard

DISPLAY

AUTO RS485
SWITCH-ON RESISTORS

If “ON”, the instrument If “ON”, it enables the 120 Ω terminal resistors


turns on automatically. between A(+) and B(-) connectors (22-23)

POWER
DISPLAY J2 J3

RX / CTS
OFF
ON

GND
SEN + / REF + EXC+ SEN + / REF - EXC-

+5V
TX
SENSOR

-
-
J1
IN 1-2
CAL OUT 1-2 OUT 3-4 SLOT RS232
RX / CTS

SLOT SLOT COM3 SIG-


GND

GND
SIG+
+5V

+5V
CTS J3
RX
TX

TX

RS485 EXC-
-
-

-
J1 EXC+
Legal ON J2
J1 RS232 RS232 CH4 CH3 CH2 CH1
Internal use COM1 COM2
OUT3
OUT1

OUT4
OUT2
COM

COM

OFF

SEN + / REF +
IN1

IN2

SEN + / REF -
RX1 / CTS

EARTH
+Vaux

EXC+
+BAT

+Vdc
GND

SIG+
GND
GND

GND

EXC-
A(+)

SIG-
RX2
TX1

TX2
CTS

B(-)
Vin

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30

POWER IN: COM 1 COM 2 COM 3 4 / 6 wires load cell


12 ~ 24 Vdc connection
Max. 48 Vac / 60 Vdc
Max. 500 mA
Max. 10 Ω

AUXILIARY VOLTAGE OUT: Max. 12 ~ 24 Vdc Load cell shield


5,3 ~ 8 Vdc 400 mA max. Max. 20 mA connection

ORIGINAL BATTERY ONLY

VOLTAGE OUT: 12Vdc


400 mA max.
Only if power supply is
connected.

INPUT AND OUTPUT CONNECTION

OUTPUT INPUT
OUT3
OUT1

OUT4
OUT2
COM

COM

IN1
IN2

RX1 / CTS
+Vaux
+BAT
GND

GND
Vin

GND

RX2
TX1

TX2
CTS

1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19

24Vdc

24Vdc

3590ET / 3590EGT 10
TECH_MAN_ENG_SCHEMES_3590ET_3590EGT
INTEGRATED CIRCUITS INSTALLATION

! OK! NO !

3590ET / 3590EGT 11
TECH_MAN_ENG_SCHEMES_3590ET_3590EGT
I/O expansion board

installation integrated inside the scale. 12 OUT and 8 IN; enables expansion of scale power supply up to 8-40Vdc.

SLOT1 SLOT2

SLOT 1 SLOT 2

I + (up to 20 mA)

I + (up to 20 mA)
COM - (0 mA / V)

COM - (0 mA / V)
V + (up to 10 V)

V + (up to 10 V)
OUT10
OUT11
OUT12
OUT13
OUT14
OUT15
OUT16
OUT5
OUT6
OUT7
OUT8
OUT9
COM

COM

IN5
IN6
IN7
IN8
IN3
IN4

GND
31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 GPI0
51 52 53 54 55 56 57 58

Max. 350 Ω Min. 10 KΩ

Max. 48 Vac / 60 Vdc Max. 12 ~ 24 Vdc


Max. 500 mA Max. 20 mA Min. 10 KΩ Max. 350 Ω
Max. 10 Ω

Connection:
DISPLAY

CPU

2
EXPANSION

INPUT AND OUTPUT CONNECTION

OUTPUT INPUT
OUT7
OUT5

OUT8
OUT6
COM

COM

IN3
IN4

IN5

IN6
IN7
IN8

31 32 33 34 35
44 45 46 47 48 49 50

24Vdc

24Vdc

3590ET / 3590EGT 12
TECH_MAN_ENG_SCHEMES_3590ET_3590EGT
Optional internal board for rapid connection of multiple scales

Multi-scale fitting board Installation:

SIG- SIG-
SIG+ SIG+ AUTO
CELL3
CELL2

SWITCH-ON

CELL3
CELL2
EXC- EXC- Se “ON”, lo strumento
si accende automaticamente.
EXC+ EXC+

SIG- POWER
SIG-
DISPLAY
SIG+ SIG+

CELL4
CELL1
CELL4
CELL1

OFF
ON
EXC- J1 EXC- SENSOR
INPUT
EXC+ EXC+ 1-2 SLOT
CAL OUTPUT OUTPUT RS232

RX / CTS
1-2 SLOT 3-4 SLOT COM3

GND

GND
+5V

+5V
CTS
J3

RX
TX

TX
RS485

-
-

-
J1
Legal ON J2
J1 RS232 RS232 CH4 CH3 CH2 CH1
Internal use COM1 COM2
OUT3
OUT1

OUT4
OUT2
COM

COM

OFF

SEN + / REF +
IN1

IN2

SEN + / REF -
EARTH
+Vaux

EXC+
+BAT

+Vdc
CTS2
GND

SIG+
GND
GND

GND

EXC-
A(+)

SIG-
RX1
TX1

TX2

B(-)
Vin

RX
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30
Motherboard
OFF
ON

RS232

ON
RS232 RS232

OFF

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30

RJ45 RS232 connector board

to TX R/-
to motherboard to RX T/+ TXD
GND GND GND
RXD RXD
TXD

i
The AMP connector enables you to rapidly connect and supply the CPU board of the weighing indicator.

3590ET / 3590EGT 13
TECH_MAN_ENG_SCHEMES_3590ET_3590EGT
RS232 / RS485 to Ethernet / WIFI conversion board

Installation integrated inside the scale.

A
GND to GND
-
+VDC to 5÷12VDC
- to motherboard
to A+ SETTINGS TX to RX
A+

OFF ON
RX to TX
to B-
B-

RS232
RS485

DFLT
WIFI

ETHERNET
ANTENNA TX to RX i
RX to TX The AMP connector enables
GND to GND
B you to rapidly connect and
RS232
supply the weighing indicator
CPU board.
WIFI module

RESET Jumper
Closing the DFLT jumper resets default settings of the operating
methods set via the dip switch:
WIFI • IP address: 192.168.16.205
ANTENNA • Subnet Mask: 255.255.255.0
• Baud rate and bit according to the type of combination of dip
switches
B
to GND

Version for installation on a DIN bar


to RX
to TX

RS232
5 6 7
TX RX GND

SETHDIN

POWER RS485
1 2 3 4 ETHERNET
+Vdc GND A(+) B(-)
to 12÷24V
to GND
to A(+)
to B(-)

3590ET / 3590EGT 14
TECH_MAN_ENG_SCHEMES_3590ET_3590EGT
SETTINGS:

OFF ON
ETHERNET/WIFI conversion in «232/485» and vice-versa
1
2 WARNINGS:
3 RS232 and RS485 communication ports cannot be used simultaneously.
4 Set the serial ports to 9600 n-8-1.

OFF ON Network between instruments on RS232


1
Port 485 available for simultaneous communication to other devices.
2
Set the serial ports to 115200 n-8-1.
3
4

OFF ON Modbus RTU converter in Modbus TCP


1
Convert Modbus RTU protocol to Modbus TCP.
2
RS232 and RS485 communication ports cannot be used simultaneously.
3
Set the serial ports to 9600 n-8-2.
4
i
It enables you to rapidly connect and supply the weighing indicator CPU board.

OFF ON RS232 converter in «RS485» and vice-versa


1
ETHERNET/WIFI port disabled.
2
Set the serial ports to 9600 n-8-1.
3
4

OFF ON Enable the option to configure the ethernet/wifi board via browser
1
Set the serial ports to 9600 n-8-1
2
3
4

i
Attention:
every time you change the operating mode, the board resets the standard default settings.

3590ET / 3590EGT 15
TECH_MAN_ENG_SCHEMES_3590ET_3590EGT
USB support data saving board

installation integrated inside the scale. Version for installation on a DIN bar

to GND
to CTS
to RX
to TX
to printer
to GND
to CTS
to Vin
to RX
to TX

RS232
3 4 5 6
TXD RXD RTS GND

OFF
1
ON
USBDIN
RS232 (PRINTER) 2
3
USB CABLE
TX
RX
CTS
Vin
GND

SETTINGS
RS232 (INDICATOR)
GND

5÷12Vdc
CTS
TX
RX

USB UART POWER


1 2
+Vdc GND

to motherboard
to GND
to 5÷12V

The AMP connector enables you to rapidly connect


and supply the CPU board of the weighing indica-
tor

SETTINGS:

OFF ON Configure the baud rate, parity, bit number and stop bit:
1
• ON = 38400-E-8-1
2
• OFF = 9600-N-8-1
3
4

OFF ON Enable CTS for a connected printer:


1
• ON = CTS managed
2
• OFF = CTS not managed
3
4

i
LED description:

RED LED:
ON = Print in progress
Flashing = USB not detected

GREEN LED:
Slow flashing = Ready
Quick flashing = USB or printer not detected

3590ET / 3590EGT 16
TECH_MAN_ENG_SCHEMES_3590ET_3590EGT
RADIO communication board

installation integrated inside the scale TTL version

RADIO
RADIO MODULE
MODULE TTL

RX to TX
TX to RX to 5V 5V
to motherboard
to GND GND
5÷12V to 5÷12V to TX RX
to RX TX
GND to GND

i
The AMP connector enables you to rapidly connect and supply the CPU board of the weighing indicator .

PROFIBUS communication board

Version for installation on a DIN bar

i
It is possible to connect the
indicator to RS232 or RS485

PROFIBUS

GND B-LINE
54321

9876
A-LINE +5VBUS

PROFI232

POWER RS232 RS485


1 2 3 4 5 6 7 8 9
+Vdc GND TX RX GND B(-) A(+) SHD EARTH
to 12÷24V

to GND
to RX

to TX

to GND
to B(-)

to A(+)

to SHD
to EARTH

3590ET / 3590EGT 17
TECH_MAN_ENG_SCHEMES_3590ET_3590EGT
BLUETOOTH communication board

Installation integrated inside the scale.

i
GND to GND The AMP connector enables you to rapidly connect and supply the
5V to 5V CPU board of the weighing indicator.
RX to TX
TX to RX

RADIO REMOTE CONTROL board

installation integrated inside the scale.

i
TLR -1 GND
Connect the module to the indicator’s COM2 via a strip connector
- situated on the board with a wire or directly to the CPU board with a
5V tulip connector situated under the board.
-
ANTENNA TX
RX

Circuit breaker against electrostatic loads

Installation integrated inside the scale.

ANALOG CELL DIGITAL CELL


EXC-
EXC+
SEN+/REF-
SEN+/REF+
SIG-
SIG+

GND
+Vdc
B(-)
A(+)

SHIELD
INDICATOR
SIDE
SEN+/ REF+
SEN+/ REF-

SHIELD
SHIELD

CELL
+Vdc
EXC+
SIG+

GND
EXC-

A(+)
SIG-

B(-)

SIDE

3590ET / 3590EGT 18
TECH_MAN_ENG_SCHEMES_3590ET_3590EGT
Analog output

DAC16O-X
ANALOG
Installation integrated inside the scale. 0-4mA/0-10Vdc. OUTPUT
Assemble the board facing towards the indicator’s CPU.

1
DISPLAY

CPU

EXPANSION

SLOT1 SLOT2
1

ANALOG
OUTPUT
31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58

2
DISPLAY

CPU

EXPANSION

SLOT1 SLOT2
ANALOG
OUTPUT
1

31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58

3590ET / 3590EGT 19
TECH_MAN_ENG_SCHEMES_3590ET_3590EGT
DAC16OSER-X

to GND
to RX
to TX

to A+
to B-
JP2

GND
A+
RX
TX

B-
SW1 GND
V+
to PLC
GND
I+

to +Vdc +
to motherboard 5 ÷ 12 Vdc
to GND -

OPERATING RANGE (only for tension analog output):

OFF ON
1
0 ÷ 10 V
2

OFF ON
1
-10 ÷ 10 V
2

OFF ON
1
0÷5V
2

OFF ON
1
-5 ÷ 5 V
2

3590ET / 3590EGT 20
TECH_MAN_ENG_SCHEMES_3590ET_3590EGT
Traffic light

Integrated and indicator

to GND 5.5V to 7.5V (vaux)

BUZZER to OUT1/OUT3
YELLOW LED to OUT1
T/+
R/-
-
-
+
GREEN LED to motherboard
to OUT2
RED LED to OUT3

BUZZER

ZONE1 ZONE2 ZONE3

Chip 485

OK! NO

3590ET / 3590EGT 21
TECH_MAN_ENG_SCHEMES_3590ET_3590EGT
RS485 opto-isolated board

RS485 opto-isolated board

Display board

CPU
connector

Buzzer

ETHERNET connector
(only for 8” models)

Indicator’s keyboard
connectors

Power expansion
board PC keyboard/ barcode
reader connector
USB port

Power display
board
SD card:
upload of images
inside the indicator Port for USB
connector board

3590ET / 3590EGT 22
TECH_MAN_ENG_SCHEMES_3590ET_3590EGT
This publication, or portions thereof, may not be duplicated without written permission from the Manufacturer. All information contained in this manual is based
on the data available at the time of its publication; the Manufacturer reserves the right to make changes to its products at any time without notice and without
incurring any penalty. We therefore recommend that you always check for any updates.
The individual in charge of the scale operation must ensure that all safety regulations in force in the country of use are applied, ensuring that the appliance is
used in accordance with the purpose it is intended for and to avoid any danger for the user.
The Manufacturer declines any liability arising from any weighing operation errors.
HEAD OFFICE
Via Della Fisica, 20

www.diniargeo.com
Via Dell’Elettronica, 15
SERVICE ASSISTANCE

41042 Spezzano di Fiorano, Modena - Italy


41042 Spezzano di Fiorano, Modena - Italy
Tel. +39 0536 843418 - Fax +39 0536 843521

Tel. +39 0536 921784 - Fax +39 0536 926654


Authorised service centre stamp
The information in this document is approximate and can be subject to variations without prior notice by Dini Argeo, with respect of the norms in force. The official
TECH_MAN_ENG_SCHEMES_3590ET_3590EGT
technical data is available in the updated version on the www.diniargeo.com website or by contacting the Dini Argeo Customer Service.
Rev. 22.11.2021