Professional Documents
Culture Documents
Service Manual
NOTES
66 76 980_02
CONTENTS
1. Important
2. Introduction
3. Description of functions
5. Service procedures
6. Troubleshooting
7. Preventive maintenance
8. Index
10. Diagrams
66 76 980_02 0–1
NOTES
66 76 980_02
1. IMPORTANT
General
Service documentation for the MAQUET FLOW-i ® Anesthesia System consists of:
User's Manual. The User's Manual is an indispensable complement to the Service Manual
for proper servicing.
Service Manual
Installation Instructions
Spare Parts information
Documentation for all optional equipment included in the FLOW-i system.
The following convention is used throughout this Service Manual:
FLOW-i represents FLOW-i Anesthesia System.
The information in this Service Manual is based on:
FLOW-i System version 2.1.
Serial number of the unit is found on a label attached by the mains power inlet.
System software version number can be found in the Status window on the Control panel.
Make sure that the version of the User's Manual corresponds to the System software version.
In addition to the Important information given here and in the related documents (e. g. in the
User's Manual), always pay attention to applicable local and national regulations.
Responsibility for the safe functioning of the equipment reverts to the owner or user in all
cases in which service or repair has been done by a non-professional or by persons who are
not employed by or authorized by MAQUET, and when the equipment is used for other than
its intended purpose.
Special waste. This product contains electronic and electrical components. Discard
disposable, replaced and left-over parts in accordance with appropriate industrial
and environmental standards.
66 76 980_02 1–1
1. IMPORTANT
Hazard notices
Before disassembling or assembling, make sure that:
All gas conveying parts are cleaned according to instructions in the FLOW-i – User's
Manual.
Gas supply is disconnected (central gas supply and backup gas supply).
Mains power cable is disconnected.
Power button is switched Off. If the Power button is switched On, the power backup battery
will supply power all electrical components in the unit.
Note that PC 1900 Main back-plane, PC 1903 External connectors, PC 1923 Power control,
the Emergency ventilation switch and the Power button are energized by the power backup
battery also when the Power button is switched Off.
The power backup battery can be disconnected either by disconnecting the battery cables at
the battery terminals or by disconnecting the battery cables (Battery and Battery sense) at PC
1903 External connectors.
WARNING! With gas supply connected to the FLOW-i, there are pressurized
components inside the unit. All personnel must exercise extreme caution if fault
tracing or adjustments are performed with gas supply connected.
WARNING! With power supply connected to the FLOW-i, there are energized
electrical components inside the unit. All personnel must exercise extreme caution
if fault tracing or adjustments are performed with power supply connected and with
covers removed.
Installation
Only personnel trained and authorized by MAQUET shall be permitted to install the
FLOW-i. The installation and handing-over procedures are described in the
'Installation Instructions'.
System checkout
After any installation, maintenance or service intervention in the FLOW-i, perform a 'System
checkout' and a ‘Manual check of Emergency ventilation system’ according to instructions in
the FLOW-i – User's Manual.
66 76 980_02 1–2
1. IMPORTANT
Service
The FLOW-i must be serviced at regular intervals by personnel trained and
authorized by MAQUET. Any maintenance or service must be noted in a log book.
It is recommended that maintenance and service is done as a part of a service contract with
MAQUET.
For functionality enhancement, the latest released System software version is always
recommended.
Preventive maintenance of the FLOW-i anesthesia system must be performed by personnel
trained and authorized by MAQUET at least once a year, or every 5000 hours of operation,
whichever comes first.
The power backup battery shall be replaced every three years.
The memory backup batteries on PC 1920, PC 1921 and PC 1922 shall be replaced every
five years.
Special waste. This product contains electronic and electrical components. Discard
disposable, replaced and left-over parts in accordance with appropriate industrial
and environmental standards.
66 76 980_02 1–3
1. IMPORTANT
Environmental declaration
Purpose
This environmental declaration applies to FLOW-i basic unit.
66 76 980_02 1–4
1. IMPORTANT
Other electronics
Gas module, containing multiple PC boards, 4 each
Electromechanical valves (EMV), 14 each
Multigas analyzer (AION), 2 each
O2 sensor (Servomex), 2 each
Fresh gas safety valve pull magnet
Expiratory flow transducer (in each patient cassette)
APL/PEEP valve coil
LED lamp
AC/DC Converter, containing PC boards
Power backup battery (lead)
Memory backup battery (lithium), 3 each
Fan
TFT panel assembly including backlight
Touch screen (glass)
Power button
APL knob
MAN/AUTO switch
O2 Flush valve
Emergency ventilation switch
Lift (C30)
Auxiliary power outlets (option)
Isolation transformer (option)
Construction materials
The construction materials used in FLOW-i in % of the total weight.
66 76 980_02 1–5
1. IMPORTANT
PE (Polyethylene)
PES (Polyethersulfone)
PET (Polyethylene terephtalate)
POM (Polyoxymethylene)
PP (Polypropylene)
PPSU (Polyphenylsulfon)
PS (Polystyrene)
PTFE (Polytetrafluoroethylene)
PUR (Polyurethane)
PVC (Polyvinyl chloride)
PVDF (Polyvinylidene fluoride)
SBC (styrene-butadiene copolymer)
SEBS (Styrene Ethylene Butylene Styrene)
SI (Silicone)
TPE (Thermoplastic elastomer)
TPU (Thermoplastic polyurethane)
Articles of consumption
66 76 980_02 1–6
1. IMPORTANT
66 76 980_02 1–7
1. IMPORTANT
Power consumption
The power consumption depends on the operating mode and system configuration. The values
listed below are calculated values for a 230 V FLOW-i system.
Noise level
Operational mode with average settings: Approximately 45 dB (A).
66 76 980_02 1–8
2. INTRODUCTION
66 76 980_02 2-1
2. INTRODUCTION
General
The anesthesia system is available in different models:
C20
C30
C40
The following table shows the standard equipment for these different models (main
components):
66 76 980_02 2-2
2. INTRODUCTION
Functional units
The system comprises the following basic parts:
1. Control panel
2. Patient monitor (option)
3. Breathing system
4. Vaporizer unit
5. Emergency ventilation system
6. External connections
66 76 980_02 2-3
2. INTRODUCTION
In this technical description, the system is divided into the following functional units (also refer to
the System Overview in chapter Diagrams):
1. Gas Distribution Section
1. Gas inlets and outlets (outlets plugged).
Contains gas inlet filters and one-way
valves OV1-OV6.
2. EVAC outlet connector.
3. PC 1906 Gas block (behind cover).
4. Reflector gas selector.
5. Gas module docking.
66 76 980_02 2-4
2. INTRODUCTION
66 76 980_02 2-5
2. INTRODUCTION
6. Vaporizer
1. Vaporizer docked in slot 1. Available
FLOW-i vaporizers are Isoflurane,
Sevoflurane and Desflurane.
2. Vaporizer inlet valve docking, slot 2.
3. Vaporizer outlet valve docking, slot 2.
66 76 980_02 2-6
2. INTRODUCTION
8. Patient cassette
1. PC 1917 Exp. channel cassette.
2. Insp. valve cover.
3. Insp. sampling connectors.
4. Insp. patient tube connector.
5. Exp. valve cover.
6. Exp. sampling connectors.
7. Exp. patient tube connector.
8. Manual ventilation tube/bag connector.
66 76 980_02 2-7
2. INTRODUCTION
66 76 980_02 2-8
2. INTRODUCTION
66 76 980_02 2-9
2. INTRODUCTION
66 76 980_02 2 - 10
2. INTRODUCTION
17. EVAC
1. EVAC Indicator.
2. EVAC Reservoir.
3. EVAC connector to Expiratory outlet on
patient cassette and Fresh gas safety valve
SV1. On units with S/N below 1500; also
connected to the System checkout valve
PV6.
4. Connector to EVAC outlet.
5. Connector to Drive gas safety valve SV3.
66 76 980_02 2 - 11
2. INTRODUCTION
66 76 980_02 2 - 12
2. INTRODUCTION
66 76 980_02 2 - 13
2. INTRODUCTION
66 76 980_02 2 - 14
2. INTRODUCTION
66 76 980_02 2 - 15
2. INTRODUCTION
66 76 980_02 2 - 16
2. INTRODUCTION
Optional equipment
A number of equipment options, not described above, can be added to the FLOW-i anesthesia
system:
Pendant mounting kits (C40 only)
Adapter kit (mounting for Backup gas holder and Additional arm)
Adjustable arm for patient monitor parameter modules
Top shelf
Universal bracket for C20
Universal bracket for C30
Side rails for hanging accessories
Additional arm
Additional table
Manual breathing bag support arm
Cable support arm
Rear handle for carrier.
66 76 980_02 2 - 17
2. INTRODUCTION
Electronic structure
General
The FLOW-i system is divided into several sub-systems, where different software is executing
on different computer nodes (PC boards). The main functional sub-systems are:
Control / PC 1920 Control
Expiratory Flowmeter / PC 1919 Exp. Channel
Monitoring / PC 1921 Monitoring
Panel / PC 1922 Panel
The communication link between these nodes is a CAN-bus.
The system software can be installed from a USB memory. Separation of the software sub-
systems is handled by the installation program.
CAN bus
Internal communication via CAN bus protocol (CAN = Controller Area Network). The CAN bus is
a simple two-wire differential serial bus system. In the system configuration, the bus bandwidth
is 500 kbit/sec.
66 76 980_02 2 - 18
2. INTRODUCTION
Electronic subsystems
Expiratory Flowmeter
The Expiratory Flowmeter sub-system’s main responsibility is the flow measurement and Patient
Cassette handling.
Exp. Flow System: Responsibility for the flow measurement performed by the Expiratory Flow
Transducer in the Patient Cassette.
Patient Cassette: Responsibility for Patient Cassette handling.
Control
The Control sub-system’s main responsibility is the patient treatment, i.e. how to regulate the
inspiratory and expiratory gas flow.
Breathing: Responsible for all modes of ventilation and delivery of anesthetic agents and gas
mixtures.
Ventilator Settings: Responsible for all ventilator settings.
Persistent Storage: Maintain a persistent data storage area used to save parameter settings.
Metrics: Contains several metrics for various types of information.
System Settings: Responsible for some of the system settings, e.g. storing default patient
category and default mode of ventilation. The system settings will be stored in persistent
memory.
Future Enhancements: Platform for future enhancements and options regarding patient
treatment.
Monitoring
The Monitoring sub-system’s main responsibility is the alarm and monitoring functions.
Alarms: Responsible for detecting and handling alarm situations.
Metrics: Responsible for monitoring the system behavior.
Persistent Storage: Maintain a persistent data storage area to save Alarm settings.
Internal Network Supervision: CAN bus-master responsible for connection of all nodes in the
system and supervision of the CAN network.
System Settings: Responsible for some of the system settings, e.g. default alarm limits.
Supervision of anesthesia delivery: Responsible for supervision of the anesthetic agent
concentration.
Optional functions: Responsible for optional functionality in the system. An option is marked in
the backplane EEPROM and can enable functions such as a ventilation mode or other features.
Future Enhancements: Platform for future enhancements and options regarding monitoring
and alarm functions.
66 76 980_02 2 - 19
2. INTRODUCTION
Panel
The Panel sub-system’s main responsibility is to give the user the possibility to interact with the
system.
Presentation and Modification of Ventilator Settings: Responsible for implementing
mechanisms to let the users modify the ventilator and anesthesia settings, as well as displaying
the ventilator and anesthesia settings to the user.
Presentation of Real-Time Metrics: Responsible for presentation of real-time metrics
generated from measurements in the Monitoring sub-system.
Presentation of Alarms: Responsible for presentation of alarms generated by the Monitoring
sub-system.
Drawing Primitives and Fonts: Will use its own graphics library that includes functions such as
draw line, draw text and draw circle.
Persistent Storage: Maintain a persistent data storage area to save Log data.
Event Logging: Responsible to maintain an Event log. Some of the information in the log will
be generated by the other sub-systems. Date and time will always be stored together with the
events. Typical events saved in the log are:
Alarm activations
Alarm limit changes
Apnea periods
Activation of System Checkout
Changes in ventilator settings
Activation of immediate functions, e.g. O2 Flush
System Diagnostics Logging: Responsible to maintain a system Diagnostics log. The
Diagnostic log is divided into four different logs and generated from all sub-systems. Date and
time will always be stored together with the events as well as some identifier of the sub-system
generating the event. Typical events saved in the logs are:
1. Technical Log:
Details about technical alarms.
2. Test Result Log:
System Checkout results.
Individual test results.
Calibration results.
3. Service Log:
Service notes created by the service technician.
4. Installation Log:
Installation version history.
FCI protocol: Responsible for the external communication protocol FCI (FLOW-i
Communication Interface).
Future Enhancements: Platform for future enhancements and options regarding user data
input and output.
66 76 980_02 2 - 20
2. INTRODUCTION
66 76 980_02 2 - 21
NOTES
66 76 980_02
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3-1
3. DESCRIPTION OF FUNCTIONS
0. General
About this chapter
This text refers to the System Overview diagram in chapter 'Diagrams'.
66 76 980_02 3-2
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3-3
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3-4
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3-5
3. DESCRIPTION OF FUNCTIONS
PC 1906 Gas Block is equipped with a temperature sensor (NTC resistor) to measure the
temperature at the Gas Distribution Section. The output signal is used for regulation of Fan 1.
Information from the ID/Status PROM is read by the FLOW-i system.
AGS
Outlet connector for an AGSS (Anesthetic Gas Scavenging System). The internal EVAC
system, connected to the AGS outlet, collects gases from the:
Expiratory Outlet (exhaust gas)
Fresh Gas Safety Valve SV1
Drive Gas Safety Valve SV3
Pressure Supply Pilot Valves EMV2–8
Vaporizer Scavenging Valves EMV10/EMV12.
66 76 980_02 3-6
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3-7
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3-8
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3-9
3. DESCRIPTION OF FUNCTIONS
1. Pressure Supply Pilot Valves 1. Pressure Supply Pilot Valve output connectors
2. PC 1907 Valve Drivers
3. PC 1781 Exp Pressure Transducer
66 76 980_02 3 - 10
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 11
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 12
3. DESCRIPTION OF FUNCTIONS
APL potentiometer
The APL potentiometer is electrically controlling the APL (Adjustable Pressure Limit) regulation
of the APL/PEEP Valve. APL can be set between spontaneous breathing (fully open) and 80
cmH2O. When fully open, the internal resistance of the system is equal to approx. 2 cmH2O.
The APL potentiometer is enabled only during Manual Ventilation.
A calibration of the APL potentiometer is required e.g. after replacement. Refer to chapter
Service Procedures.
66 76 980_02 3 - 13
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 14
3. DESCRIPTION OF FUNCTIONS
1. Gas inlet
2. Filter
3. Temperature sensor
4. Supply pressure transducer
5. Flow transducer (Delta pressure transducer and net)
6. Nozzle unit with valve diaphragm
7. Solenoid
CAUTION: The Gas Modules are factory calibrated. The Gas Modules must not be
disassembled further than described in chapter 'Preventive Maintenance'.
Gas inlet
When the Gas Modules are mounted in the unit, the gas inlet channel in the filter cover
connects to the Gas Module Docking on the Gas Distribution Section.
The filter housing and the filter cover are provided with matching guide pins. These guide pins
prevent mounting of the filter cover (with gas inlet) on the wrong module.
Filter
The Filter protects the system from particles in the gas delivered to the Gas Modules. The filter
must be replaced during the 'Preventive Maintenance'.
66 76 980_02 3 - 15
3. DESCRIPTION OF FUNCTIONS
Temperature sensor
The temperature of the supplied gas is measured by the Temperature Sensor. This sensor is
situated in the gas flow.
The output signal from this sensor is used to compensate for the gas density variations due to
temperature.
Flow transducer
The gas flows through a net (resistance) which causes a pressure drop. The pressure is
measured on both sides of this net and the differential pressure value is then amplified.
Nozzle unit
The plastic Nozzle Unit contains a valve diaphragm. The valve diaphragm, controlled by the
Solenoid, regulates the gas flow through the Gas Module.
The complete plastic nozzle unit must be replaced during the 'Preventive Maintenance'. After
replacement, allow the diaphragm to settle during approx. 10 minutes before gas pressure is
connected to the Gas Module.
Solenoid
The gas flow through the Gas Module is regulated by the Solenoid via the Nozzle Unit. The
current supplied to the solenoid is regulated so that the Gas Module will deliver a gas flow
according to the settings on the Control Panel.
ID/Status PROM
Information from the ID/Status PROM is read by the FLOW-i.
O2 Flush Valve
The O2 Flush Valve (push-button marked O2+) is used to manually supply the fresh gas with an
additional O2 gas supply of minimum 40 l/min. This gas supply will not go through the vaporizer.
The O2 Flush gas supply will continue as long as the button is depressed.
The O2 Flush Valve is electrically connected to PC 1927 O2 Flush and the system detects if the
O2+ push-button is pressed. If so, ventilation continues without changes in the fresh gas
delivery; the anesthetic flow is not recalculated to adjust to the new higher flow and the
vaporizer continues to deliver the same anesthetic concentration.
The O2 delivery from the O2 Flush Valve is however not dependent on any electric function and
additional O2 can be supplied also in case of a power failure.
O2 with pressure regulated to 200 kPa by REG4 is delivered from the Drive Gas Supply Section
to the O2 Flush Valve.
66 76 980_02 3 - 16
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 17
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 18
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 19
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 20
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 21
3. DESCRIPTION OF FUNCTIONS
6. Vaporizer
There are vaporizers available for three different anesthetic agents; Isoflurane, Sevoflurane and
Desflurane.
The vaporizers are equipped with different filling devices adapted to anesthetic agent
manufacturer’s specifications.
A Liquid Container inside the vaporizer contains the anesthetic agent. When the vaporizer is
activated (a concentration is set and accepted), the Liquid Container will be pressurized to 120
kPa by the Vaporizer Drive Gas.
The liquid is lead via channels in the base plate, through the Safety Valve, to the Vaporizer
Injector. The injectors opening is pulse controlled in order to achieve the desired dose of liquid.
A nozzle plate in front of the injector forms the liquid into a spray in the Vaporizing Chamber.
Fresh Gas to Vaporizer passes through the Vaporizing Chamber and the anesthetic agent is
mixed with the gas. This gas mixture, Fresh Gas from Vaporizer, is then routed to the Patient
Cassette (or AFGO).
The Vaporizing Chamber is heated which promotes conversion of spray into gas. The
temperature is regulated to 47°C for Isoflurane/Sevoflurane and 37˚C for Desflurane.
The Vaporizer Pressure Transducer measures the liquid pressure. The Safety Valve will cut off
the supply of liquid as a safety precaution. Presence of spray is monitored with the optical
Vaporizer Injection Detection.
Sintered metal filters protect the system from particles entering the liquid system. An optical
Liquid Level Sensor measures the liquid level in the Liquid Indicator glass tube. The agent is
filled into the container via a key filling valve. The mechanism is protected by a lid with a Lid
Sensor.
66 76 980_02 3 - 22
3. DESCRIPTION OF FUNCTIONS
Handle the vaporizer with care and never turn a vaporizer containing anesthetic agent upside
down or lay it sideways.
The Vaporizer is factory calibrated and leakage tested. The Vaporizer must not be
disassembled.
Repair or service in case of malfunction must be carried out by MAQUET Critical Care only.
The RMA procedure must be used when returning vaporizers to MAQUET Critical Care.
If the vaporizer is dropped, it must be serviced by MAQUET Critical Care to ensure proper
functionality.
The vaporizer must be emptied before it is returned to MAQUET Critical Care. Refer to
emptying instructions in the User’s Manual.
EMC Gasket
The EMC Gasket is a metal blade located in the Vaporizer Drive Gas inlet. This metal blade will
ensure ground connection between the docking pipe and the vaporizer when docking the
vaporizer.
Docking Valve
The Docking Valve OV9 is a one-way valve located in the Vaporizer Drive Gas inlet. The
Vaporizer Drive Gas docking pipe on the Vaporizer Valve Section will mechanically open OV9
when the vaporizer is docked.
When the vaporizer is disconnected, OV9 will close and thus close the vaporizer pressure
channel. This will keep the Vaporizer Liquid Container closed.
66 76 980_02 3 - 23
3. DESCRIPTION OF FUNCTIONS
Vaporizer Lid
The Vaporizer Lid covers the Safety Fill Valve. The lids position is monitored by a reflective
object sensor (opto switch) on PC 1931 Vaporizer Lid Sensor.
The lid must be closed, or near closed position, in order to activate the vaporizer. If the lid is
detected as open, the vaporizer will be de-activated and thus de-pressurized.
Level Indicator
The anesthetic agent level in the liquid container is indicated in two different ways:
A float in the level meter tube shows the anesthetic agent level. The float is visible in the level
meter tube through the illuminated vaporizer window. The four white LEDs that light up the
window are located on PC 1912 Vaporizer Liquid Detector.
The anesthetic agent level is also monitored by LEDs on PC 1912 Vaporizer Liquid Detector
(sending) and PC 1913 Vaporizer Liquid Level Detector (receiving). These LEDs are mounted
on six different levels corresponding to 5%, 10%, 25%, 50%, 75% and 100%. With this
design, the following anesthetic agent levels will be shown on the Control Panel:
Level < 5%
Level 5–10%
Level 10–25%
Level 25–50%
Level 50–75%
Level 75–100%
Level = 100%
This monitoring function is controlled by PC 1910 Vaporizer Controller Board.
The vaporizer liquid level alarms are:
Low priority clinical alarm when the liquid level is below 10%
Medium priority clinical alarm when the liquid level is below 5%.
The vaporizer will not switch off when empty, but it is not recommended to run the vaporizer dry.
Note: There may be small amounts of anesthetic agent remaining in the vaporizer, although the
indicator shows that the vaporizer is empty.
66 76 980_02 3 - 24
3. DESCRIPTION OF FUNCTIONS
Drain Plug
The vaporizer is equipped with a Drain Plug. This Drain Plug is intended for MAQUET Critical
Care service only and must not be removed during field service. Use a bottle with filling adapter
to empty the vaporizer. Refer to emptying instructions in the User’s Manual.
Vaporizer Injector
The Vaporizer Injector VI supplies anesthetic agent into the fresh gas. Depending on the set
and the measured concentration, it injects the agent in very short pulses. Min. injector pulse
time is 2 ms and max. time is 10 ms. The injected agent speed is approx. 10 m/s. A built-in filter
in the injector separates particles down to 32 µm in the anesthetic agent.
As the injector technology creates a high and disturbing sound, the vaporizer is designed to
reduce this sound. The injector itself has a flexible mounting inside the vaporizer. The injector is
also equipped with a free weight to reduce the injector’s oscillation.
PC 1910 Vaporizer Controller Board contains drivers for the Vaporizer Injector.
66 76 980_02 3 - 25
3. DESCRIPTION OF FUNCTIONS
Vaporizing Chamber
The Vaporizer Injector VI supplies anesthetic agent into the Vaporizing Chamber. Fresh Gas to
Vaporizer, supplied by the Fresh Gas Modules, is routed into the Vaporizer via the Vaporizer
Inlet Valves PV1 or PV3. The fresh gas enters the Vaporizing Chamber where it is mixed with
the vaporized anesthetic agent.
The mixed Fresh Gas from Vaporizer exits the Vaporizer via the Vaporizer Outlet Valves PV2 or
PV4 and is routed to the Patient Cassette (or AFGO).
66 76 980_02 3 - 26
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 27
3. DESCRIPTION OF FUNCTIONS
The following gas connections are integrated in the Fresh Gas Connector Block:
Fresh gas from vaporizer.
O2 from the O2 Flush Valve. Supplies additional O2 to the fresh gas when O2 Flush Valve
button is pressed.
Control Gas Analyzer’s sampling gas return (R2). Return point for CGA gas sample.
Pilot Valve Drive Gas supply (PV7). Plugged connector to hold the drive gas tube intended for
the AFGO Pilot Valve (when the optional AFGO Valve is installed)
Fresh gas to patient cassette via the safety valve.
66 76 980_02 3 - 28
3. DESCRIPTION OF FUNCTIONS
Fresh Gas Safety Valve SV1 is a spring-loaded solenoid valve. The valve is connected to the
Fresh Gas from Vaporizer channel close to the Patient Cassette. The valve is easy to remove
for inspection and cleaning.
A gas channel for the Reflector Gas is integrated in the safety valve unit, but is not connected to
the safety valve function.
The safety valve solenoid is electrically controlled from PC 1919 Exp. channel:
When the solenoid is not activated, the safety valve is kept closed by the spring inside the
valve housing.
When the solenoid is electrically activated, the safety valve will open and the fresh gas is let
out from the fresh gas channel to the EVAC Reservoir thus enabling a decrease in the
inspiratory pressure.
The opening conditions for the safety valve are:
The pressure inside the inspiratory channel in the Patient Cassette is 5 cmH2O above the
preset upper pressure limit. This condition is controlled by the Monitoring subsystem. AUTO
mode only.
The pressure inside the inspiratory channel in the Patient Cassette is 7 cmH2O above the
preset upper pressure limit. This condition is controlled by the Control subsystem. AUTO
mode only.
The pressure inside the inspiratory channel in the Patient Cassette is above:
20 cmH2O if APL is set between SP and 12 cmH2O. Redundant opening condition at 25
cmH2O. MAN mode only.
Set APL +7 cmH2O if APL i set between 13 and 49 cmH2O. Redundant opening condition
at 12 cmH2O. MAN mode only.
Set APL +10 cmH2O if APL i set between 50 and 80 cmH2O. Redundant opening condition
at 15 cmH2O. MAN mode only.
The safety valve will also be opened by some other alarms, e. g. the High continuous
pressure-alarm (PEEP level + 15 cmH2O for more than 15 seconds). AUTO mode only.
The pressure inside the inspiratory channel in the Patient Cassette is above 117 ±7 cmH2O.
This condition is controlled by calibrated electronics and calibration to 117 ±3 cmH2O is
performed during each System checkout. This condition will also open the APL/PEEP valve
and is an extra safety function and the situation will normally not occur. AUTO and MAN
mode.
66 76 980_02 3 - 29
3. DESCRIPTION OF FUNCTIONS
8. Patient Cassette
The Patient Cassette is the main interface for the gases distributed to and from the patient. The
cassette is a removable and easy-to-clean unit.
The gas modules and the unidirectional valves in the cassette control the gas flow through the
cassette. The gas flow is controlled due to:
Selected breathing system.
Selected ventilation mode.
Breath cycle phase.
The Patient Cassette connects to:
Fresh Gas from Vaporizer (via Safety Valve SV1).
Reflector Gas from the Reflector Gas Module.
Volume Reflector, two channels.
Expiratory Outlet.
Manual Ventilation Bag.
CO2 Absorber, inlet and outlet. The absorber connections are equipped with spring-loaded
cut-off valves that will automatically close the connections when the CO2 Absorber is lowered.
Patient tubing, Insp. and Exp.
The Patient Cassette also contains:
Gas sampling and gas return channels for the Gas analyzers.
Pressure measuring points for Insp. and Exp. pressure.
Other integrated cassette main functions are:
The Expiratory Flow Transducer that measures the expiratory gas flow through the cassette.
The APL/PEEP Valve that limits the airway pressure during ventilation.
The Manual Ventilation Valve that connects the Manual Ventilation Bag.
The CO2 Absorber valves and a number of silicone parts in the cassette must be replaced
during Preventive Maintenance.
66 76 980_02 3 - 30
3. DESCRIPTION OF FUNCTIONS
Patient Cassette
66 76 980_02 3 - 31
3. DESCRIPTION OF FUNCTIONS
To reduce water condensation in conjunction with the transducers, the flow transducer is
heated. A heating foil is mounted close to each transducer.
The expiratory gas temperature will differ within the breathing system; the gas will e.g. be
cooled down in the Volume Reflector. These temperature variations will affect the gas flow
calculations and thus also the volume measuring. One temperature sensor, located close to the
expiratory inlet, measures the temperature of the gas flow from the patient. The other
temperature sensor, located close to the flow transducer, measures the gas temperature in
conjunction with the flow transducer. With this design, the temperature from both the reflector
and the patient can be estimated.
APL/PEEP Valve
The APL/PEEP Valve limits the airway pressure to the level set by the operator.
The APL/PEEP Valve consists of a membrane in the cassette that is operated by the axis of the
APL/PEEP Valve Coil. The axis can be activated (pushed up) into desired position in two
different ways:
During automatic and manual ventilation, power supply to the coil is regulated so that the
airway pressure in the patient system is kept on the APL/PEEP level set by the operator.
During emergency manual ventilation, the axis is pushed up into desired position by the
mushroom valve inside the Emergency Valve Actuator mounted underneath the APL/PEEP
Valve Coil. Gas pressure (supplied by REG5) to inflate the mushroom valve is regulated to
correspond to the Mechanical APL level set by the operator.
The APL/PEEP Valve is fully open as long as the axis is not activated.
The APL/PEEP Valve membrane must be replaced during Preventive Maintenance.
There is also a counter for the breathing cycles. Operating capacity for the membrane is
estimated to 10.000.000 breathing cycles. If this limit is passed, or if the membrane for some
reason has become defective, it must be replaced. Remaining capacity (in %) for the membrane
can be shown in the Service & Setting’s Status window. The remaining capacity meter must be
reset after replacement of the membrane.
66 76 980_02 3 - 32
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 33
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 34
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 35
3. DESCRIPTION OF FUNCTIONS
Note
A Service Software for the gas analyzers is required to:
Adjust the potentiometers on the O2 sensor. If the Service Software is not used, the
measurement accuracy of the gas analyzer will be outdated.
Perform an O2 sensor calibration if the O2 sensor and/or the AION have been replaced.
Update the O2 sensor serial number stored in AION EEPROM if the O2 sensor has been
replaced.
66 76 980_02 3 - 36
3. DESCRIPTION OF FUNCTIONS
Update the configuration stored in the AION EEPROM if the AION unit has been replaced.
This is done to secure traceability and to allow correct revision number to be shown on the
Control Panel.
The AION™ is the gas measuring unit. It is a complete unit and must not be disassembled.
66 76 980_02 3 - 37
3. DESCRIPTION OF FUNCTIONS
Pneumatic Module
The pneumatic module interconnects the pump, the gas measurement bench and the O2
sensor. Includes two solenoid valves; the zero calibration valve (ZCV) and the purge flow valve
(PFV)
Zero Calibration Valve (ZCV): To establish a gas zero concentration reference, ambient air is
routed through the Gas Measurement Bench via the Zero Calibration Valve. The ambient gas
intake is located behind the Mechanical APL knob in the Emergency Ventilation section. A filter
is connected to the tubing.
The automatic zero calibration of the AION will be performed:
During warm-up. In CGA, a “sleep mode” will keep the IR light source heated and the pump
disabled to avoid zero calibration and enable a fast start (<10 s).
At a temperature change >1 °C (in the GMB)
Every 4 hours. In CGA, this time controlled calibration can be disabled by the System SW.
At gas span calibration.
Purge Flow Valve (PFV): The purge flow function is used only when a water trap is connected
to the sampling line. It is not required, and disabled, in the CGA.
66 76 980_02 3 - 38
3. DESCRIPTION OF FUNCTIONS
The GMB consists of an IR light source, a rotating filter wheel with optical filters connected to a
DC motor, the gas measurement chamber and an IR light detector.
The measurement of CO2, N2O and the anesthetic agents in the sampling gas is based on the
fact that the different gas components absorb infrared light at specific wavelengths. The filter
wheel has eight optical filters to allow an accurate analysis of any mixture of these gases.
As O2 does not absorb infrared light to the same extent as other breathing gases, O2 is
measured by the separate O2 sensor. However, the presence of O2 causes some interference
and information from the O2 sensor is used to compensate for that interference.
The GMB also includes a PCB with an EEPROM and components for pressure and temperature
measurements. Factory calibration data are stored in the EEPROM and the gas calibration data
is updated during the gas span calibration.
Pump Module
The Pump Module (PM) with a double-action membrane pump has a built-in flow control with its
own flow sensor. The delivered flow is very stable with minimal flow variations. The design of
the pump makes it very reliable with no maintenance necessary. The default sampling flow rate
is 200 ml/min.
Base Module
The Base Module (BM) includes a PC board with CPU and memory functions. This PC board
controls the gas sampling system and performs calculations from the measured parameters.
The analog signals from the GMB are converted to digital signals in the AION CPU.
The AION-SW is stored in a flash memory. As from System version 2.0, installation of the AION-
SW is included in the semi-automatic System software installation procedure. Refer to chapter
‘Service procedures’.
The AION PC board has a 5 V DC/DC converter for its internal power supply.
66 76 980_02 3 - 39
3. DESCRIPTION OF FUNCTIONS
Control O2 Sensor
The O2 sensor is a paramagnetic sensor that measures the O2 content in the sampled gas
mixture. The sensor output is a serial, full duplex UART (RS-232 compatible) signal with 10 ms
update rate.
O2 is distinguished from most other common gases by its paramagnetic properties. This fact is
used by the paramagnetic O2 sensor, which has two nitrogen filled glass spheres mounted on a
strong rare metal taut-band suspension. This assembly is suspended in a symmetrical non-
uniform magnetic field. In the presence of paramagnetic O2, the glass spheres are pushed
further away from the strongest part of the magnetic field. The strength of the torque acting on
the suspension is proportional to the O2 concentration.
A feedback system creates a current that balances the torque acting on the suspension
assembly. The feedback current is directly proportional to the partial pressure of O2 in the
sample. A voltage output is derived that is proportional to the current, which in turn is
proportional to the O2 concentration.
The O2 sensor has no consumable parts and has a consistent performance over time. It
requires no reference gas for zero calibration and has no interference from other respiratory
gases.
The O2 sensor output is delivered to the system, together with gas measurement information
from AION.
AION uses the O2 concentration data to compensate for the O2 interference on the gas
measurements performed by AION.
The O2 sensor is powered with +5 VDC.
66 76 980_02 3 - 40
3. DESCRIPTION OF FUNCTIONS
Sampling valves
The sampling valves Control Gas Analyzer Inlet Valve EMV14 and Control Gas Analyzer Outlet
Valve EMV16 controls the sampling gas flow to and from the Control Gas Analyzer. The
sampling valves are controlled by the software.
The inlet valve EMV14 can switch between the following gas sampling points:
Inspiratory channel (S1). Gas will be sampled from the inspiratory channel in the cassette, i.e.
from the gas delivered to the patient. The gas sample is routed via a heated bacteria filter and
a Nafion™ tube. The Nafion tube equalizes the moisture content and the temperature in the
sampled gas to that of the ambient air. This is to reduce the influence on the measurements
from humidity and to prevent from water condensation inside the sampling system.
Fresh gas channel (S2). Gas will be sampled simultaneously from two different sampling
points in the fresh gas channel; 50% of the gas before the vaporizer and 50% after the
vaporizer. The purpose of this 50/50 gas mix is to get a lower and more stable agent
concentration in the gas sample. If all gas is sampled after the vaporizer, the agent
concentration may temporarily exceed AION’s measuring range (due to the vaporizer’s
injection technique). The software will re-calculate the measured value to the actual agent
concentration. With this design, CGA will be able to measure a much higher agent
concentration compared to the PCA.
The outlet valve EMV16 can switch between the following gas return points:
Inspiratory channel (R1). The gas sample will be returned to the inspiratory channel in the
cassette, close to the sampling point. Used when gas is sampled from the inspiratory channel.
Fresh gas (R2). The gas sample will be returned to the fresh gas channel. The gas return
point is located inside the Fresh Gas Connector Block (or AFGO Valve if mounted). Used
when gas is sampled from the fresh gas channel.
Default sampling setting is fresh gas (S2 > R2).
66 76 980_02 3 - 41
3. DESCRIPTION OF FUNCTIONS
The AION™ is the gas measuring unit. It is a complete unit and must not be disassembled.
66 76 980_02 3 - 42
3. DESCRIPTION OF FUNCTIONS
Pneumatic Module
The pneumatic module interconnects the pump, the gas measurement bench and the O2
sensor. Includes two solenoid valves; the zero calibration valve (ZCV) and the purge flow valve
(PFV)
Zero Calibration Valve (ZCV): To establish a gas zero concentration reference, ambient air is
routed through the Gas Measurement Bench via the Zero Calibration Valve. The ambient gas
intake is located behind the Mechanical APL knob in the Emergency Ventilation section. A filter
is connected to the tubing.
The automatic zero calibration of the AION will be performed:
During warm-up
At a temperature change >1 °C (in the GMB)
Every 4 hours
At gas span calibration.
66 76 980_02 3 - 43
3. DESCRIPTION OF FUNCTIONS
flow forces water and secretions in the sampling line into the water trap container. The forced
purge flow is routed from the water trap directly to the pump and out through the evacuation
outlet. It will not pass the Gas Measurement Bench or the O2 sensor.
If the total flow remains below the limit value, the forced purge flow-function will be repeated up
to four times until further forced purge cycles are inhibited. The purge function is automatic and
is not possible to start manually.
Pump Module
The Pump Module (PM) with a double-action membrane pump has a built-in flow control with its
own flow sensor. The delivered flow is very stable with minimal flow variations. The design of
the pump makes it very reliable with no maintenance necessary. The water trap to be used sets
the default sampling flow rate to 200 ml/min.
During the forced purge flow, the pump capacity will be increased to create the forced purge
flow. This will not change the sampling flow rate.
66 76 980_02 3 - 44
3. DESCRIPTION OF FUNCTIONS
Base Module
The Base Module (BM) includes AION module has a PC board with CPU and memory
functions. This PC board controls the gas sampling system and performs calculations from the
measured parameters. The analog signals from the GMB and from the O2 sensor are converted
to digital signals in the AION CPU.
The AION-SW is stored in a flash memory. As from System version 2.0, installation of the AION-
SW is included in the semi-automatic System software installation procedure. Refer to chapter
‘Service procedures’.
The AION PC board has a 5 V DC/DC converter for its internal power supply.
Patient O2 Sensor
The O2 sensor is a paramagnetic sensor that measures the O2 content in the sampled gas
mixture. The sensor output is an analogue signal. One volt corresponds to a concentration of
100% oxygen.
O2 is distinguished from most other common gases by its paramagnetic properties. This fact is
used by the paramagnetic O2 sensor, which has two nitrogen filled glass spheres mounted on a
strong rare metal taut-band suspension. This assembly is suspended in a symmetrical non-
uniform magnetic field. In the presence of paramagnetic O2, the glass spheres are pushed
further away from the strongest part of the magnetic field. The strength of the torque acting on
the suspension is proportional to the O2 concentration.
A feedback system creates a current that balances the torque acting on the suspension
assembly. The feedback current is directly proportional to the partial pressure of O2 in the
sample. A voltage output is derived that is proportional to the current, which in turn is
proportional to the O2 concentration.
66 76 980_02 3 - 45
3. DESCRIPTION OF FUNCTIONS
The O2 sensor has no consumable parts and has a consistent performance over time. It
requires no reference gas for zero calibration and has no interference from other respiratory
gases.
The O2 sensor analogue output is converted to digital and delivered to the system, together
with gas measurement information from AION.
AION uses the O2 concentration data to compensate for the O2 interference on the gas
measurements performed by AION.
The O2 sensor is powered with +5 VDC.
Water Trap
The DRYLINE™ water trap protects the gas analyzer system from condensed water, secretions,
bacterial contamination and dust. It consists of two parts, a filter housing with hydrophobic anti-
bacterial filter and a container for separated waste.
The filter housing has one gas sample inlet (W1) to be connected to the sampling tube.
In the water trap, the sampling flow is divided in two parts:
Sampling gas flow (W2), approx. 90%.
Purge gas flow (W3), approx. 10%. This flow will create a negative pressure inside the water
trap in order to collect the water droplets.
There are two outlets in the filter housing, one for each gas flow.
66 76 980_02 3 - 46
3. DESCRIPTION OF FUNCTIONS
The Water Trap is connected to the Water Trap Receptacle. There are two gas connections on
the receptacle; one for the sampling gas flow and one for the purge gas flow.
The gas sample is routed via a bacteria filter and a Nafion™ tube. The Nafion tube equalizes
the moisture content and the temperature in the sampled gas to that of the ambient air. This is
to reduce the influence on the measurements from humidity and to prevent from water
condensation inside the sampling system.
Two electrical switches on the water trap receptacle detect if a water trap is present. If no water
trap is connected or if a water trap without container is connected, both switches are inactive.
The pump, and thus the gas flow through the Patient Gas Analyzer, is automatically stopped to
prevent contamination.
The control signal cable from the water trap receptacle PC board is connected to the Main back-
plane.
66 76 980_02 3 - 47
3. DESCRIPTION OF FUNCTIONS
Safety Aspects
The complete CO2 Absorber must be discarded as hazardous waste when
consumed.
Always use MAQUET approved CO2 Absorber.
Never open the CO2 Absorber to replace soda lime.
Further important information regarding replacement, safety aspects and storage of
the CO2 Absorber can be found in the User’s Manual.
CO2 Absorber
Expiratory gas from the Volume Reflector is routed through the CO2 Absorber and thus purified
before it is mixed with Fresh Gas from Vaporizer and delivered to the patient.
The CO2 Absorber will also handle water that may condensate in the absorber.
Furthermore, the color of the absorbent will change when consumed (from white to blue/violet).
The pre-packed CO2 Absorber contains 700 ml of absorbent.
The absorber is placed in the absorber compartment and lifted into position (Bypassed or
Locked) by a spring-loaded piston that is manually actuated by the absorber locking switch.
66 76 980_02 3 - 48
3. DESCRIPTION OF FUNCTIONS
Absorber Sensor
The Absorber Sensor is a micro switch that is actuated by the piston in the absorber lift. The
sensor indicates when:
The absorber is not connected (locking switch position Open or Bypassed)
The absorber is connected (locking switch position Locked).
The Absorber Sensor is electrically connected to PC 1907 Valve Drivers.
66 76 980_02 3 - 49
3. DESCRIPTION OF FUNCTIONS
Volume Reflector
The purpose of the Volume Reflector is to store the patients expired gas so that the gas can be
purified in the absorber for partial re-use during next inspiration phase.
The inside of the reflector is designed as a long gas channel with a total volume of 1.3 liter.
During the expiration phase, expiratory gas from the patient will be routed into the reflector gas
channel.
During the inspiration phase, reflector gas from the Reflector Gas Module will be routed into the
other end of the reflector gas channel. The reflector gas will, like a piston, push the expiratory
gas out from the reflector and through the CO2 Absorber. Fresh Gas according the Control
Panel setting will then be mixed with the purified gas to form the inspiratory gas to the patient.
CAUTION: As the Volume Reflector is exposed to the patients expired gas it must
be included in the routine cleaning procedure, refer to the User’s Manual.
Reflector Socket
The Reflector Socket (volume reflector adaptor) connects the Patient Cassette to the Volume
Reflector. The socket will also secure the reflector in position when mounted.
As the Reflector Socket is exposed to the patients expired gas it must be included in the routine
cleaning procedure, refer to the User’s Manual.
The Reflector Socket seals must be replaced during Preventive Maintenance.
66 76 980_02 3 - 50
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 51
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 52
3. DESCRIPTION OF FUNCTIONS
1. Gas inlet
2. Filter
3. Temperature sensor
4. Supply pressure transducer
5. Flow transducer (Delta pressure transducer and net)
6. Nozzle unit with valve diaphragm
7. Solenoid
CAUTION: The Gas Modules are factory calibrated. The Gas Modules must not be
disassembled further than described in chapter 'Preventive Maintenance'.
Gas inlet
When the Gas Module is mounted in the unit, the gas inlet channel in the filter cover connects to
the Gas Module Docking on the Gas Distribution Section.
The filter housing and the filter cover are provided with matching guide pins. These guide pins
prevent mounting of the filter cover (with gas inlet) on the wrong module.
Filter
The Filter protects the system from particles in the gas delivered to the Gas Module. The filter
must be replaced during the 'Preventive Maintenance'.
66 76 980_02 3 - 53
3. DESCRIPTION OF FUNCTIONS
Temperature sensor
The temperature of the supplied gas is measured by the Temperature Sensor. This sensor is
situated in the gas flow.
The output signal from this sensor is used to compensate for the gas density variations due to
temperature.
Flow transducer
The gas flows through a net (resistance) which causes a pressure drop. The pressure is
measured on both sides of this net and the differential pressure value is then amplified.
Nozzle unit
The plastic Nozzle Unit contains a valve diaphragm. The valve diaphragm, controlled by the
Solenoid, regulates the gas flow through the Gas Module.
The complete plastic nozzle unit must be replaced during the 'Preventive Maintenance'. After
replacement, allow the diaphragm to settle during approx. 10 minutes before gas pressure is
connected to the Gas Module.
Solenoid
The gas flow through the Gas Module is regulated by the Solenoid via the Nozzle Unit. The
current supplied to the solenoid is regulated so that the Gas Module will deliver a gas flow
according to the settings on the Control Panel.
ID/Status PROM
Information from the ID/Status PROM is read by the FLOW-i.
66 76 980_02 3 - 54
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 55
3. DESCRIPTION OF FUNCTIONS
In case of a total power failure (i.e. mains power and power backup battery) or system failure,
this system will allow the patient to be manually ventilated. The emergency ventilation system
comprises:
Emergency Ventilation control panel with:
Emergency Ventilation switch
Oxygen Flow; flowmeter for oxygen gas supply during emergency manual ventilation
Emergency APL Pressure Regulator (Mechanical APL); for APL setting during emergency
manual ventilation
Emergency APL actuator to control the APL valve inside the cassette during emergency
manual ventilation
APL/PEEP Valve Coil that is mechanically controlled by the Emergency APL actuator during
emergency manual ventilation and electrically controlled during normal AUT and MAN
ventilation
Gas analyzer zero calibration gas intake
Oxygen Connector Block.
Note: When Emergency Ventilation is switched Off, the Mechanical APL must be set to
minimum (for units of Version A).
Version A
66 76 980_02 3 - 56
3. DESCRIPTION OF FUNCTIONS
Version B
66 76 980_02 3 - 57
3. DESCRIPTION OF FUNCTIONS
Oxygen Flow
The oxygen flowmeter, graded up to 10 l/min, is used to supply oxygen to the breathing system
during emergency manual ventilation. Max. flow is 16–20 l/min. The gas is connected to the O2
supply to Fresh Gas tubing via the Oxygen Connector Block.
66 76 980_02 3 - 58
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 59
3. DESCRIPTION OF FUNCTIONS
Bleed-flow devices
There are two Bleed-flow devices connected to the Oxygen Connector Block – Version B:
High-pressure bleed-flow device. A possible leakage from the Emergency Ventilation Switch
will be evacuated through this device.
Low-pressure bleed-flow device. Will evacuate pressure from the Emergency APL Actuator.
Without this bleed-flow, a possible leakage from the Emergency APL Pressure Regulator may
inflate the Emergency APL Actuator and interfere with the APL/PEEP setting. Such leakage
will increase the pressure in the patient circuit and will be detected by the System Checkout if
it is outside the SCO limits.
66 76 980_02 3 - 60
3. DESCRIPTION OF FUNCTIONS
Version A Version B
Test plug
During System Checkout (SCO), the patient tubing’s Y-piece must be connected to the Test
plug. With a plugged Y-piece, the SCO leakage checks can be performed.
66 76 980_02 3 - 61
3. DESCRIPTION OF FUNCTIONS
17. EVAC
EVAC is a passive gas scavenging system that evacuates all excess gas, breathing gas as well
as drive gas from the FLOW-i system. The AGS outlet is connected to the hospitals anesthesia
gas scavenging system that will supply the negative pressure required for gas evacuation.
1. EVAC Reservoir
2. Exhaust gas from Patient Cassette
3. Tubing between flow indicator and AGS outlet
4. Evacuation from Drive Gas Safety Valve
5. AGS Flow Indicator
6. Tubing between reservoir and flow indicator
7. Evacuation from Pilot valves and Vaporizer valves
66 76 980_02 3 - 62
3. DESCRIPTION OF FUNCTIONS
EVAC Reservoir
Excess gas is collected and evacuated from the system by means of a negative pressure
created by the hospitals anesthesia gas scavenging system. To handle larger gas volumes,
FLOW-i is equipped with an EVAC Reservoir. This plastic container has an air intake towards
ambient air to avoid excessive high negative pressures in the system.
The EVAC Reservoir, the AGS Flow Indicator and the EVAC Hytrel tubing’s must be
disassembled and cleaned during Preventive Maintenance.
CAUTION: Ensure that the EVAC system is correctly connected after any service
intervention. Running the system with a poor connection may result in anesthetic
agent being emitted into the operation environment. Leakages in the EVAC system
will not be detected by the System Checkout.
66 76 980_02 3 - 63
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 64
3. DESCRIPTION OF FUNCTIONS
Display/Touch Panel
The Display is a 15’’ TFT (Thin Film Transistor) screen for color display of picture- and
alphanumeric data. The screen brightness can be manually adjusted using the membrane
button Screen layout. Selectable brightness levels are 100%, 75% or 50%.
The Touch Panel implies the touch function of the screen and is interactive with information
shown on the Display.
The Backlight lamps inside the Display/Touch Panel are driven from the Backlight Inverter.
Estimated lifetime (with acceptable brightness level) for the lamps is 30.000 hours. Using
Service & Settings, a time meter for the lamps can be shown. The time meter must be reset
after replacement of the lamps.
Except for the backlight lamps, the Display/Touch Panel is handled as one complete part
Backlight Inverter
PC board with driving stage for backlight (lamps) mounted inside the Display/Touch Panel. The
supply voltage delivered by the Backlight Inverter is 660 V.
WARNING! The PC board generates 660 V. All personnel must exercise extreme
caution if fault tracing or adjustments are performed with power supply connected
and with covers removed.
66 76 980_02 3 - 65
3. DESCRIPTION OF FUNCTIONS
Rotary knob
Rotary knob is an optical rotary encoder with switch that is connected to and managed by PC
1925.
Loudspeaker
For generation of sound, e.g. alarm. The loudspeaker generates different tones with individual
sound volumes. At startup and during System Checkout the function of the loudspeaker is
monitored by the microphone on PC 1924. During operation, it is continuously monitored
through current sensing.
66 76 980_02 3 - 66
3. DESCRIPTION OF FUNCTIONS
1. Battery connector
2. Battery sense connector
3. Gas cylinder pressure connector
4. PC 1903 External connectors (behind panel)
5. Lamp connector
6. AC/DC Power inlet
7. AC/DC Converter
8. PC 1923 Power control
AC/DC Converter
The AC/DC Converter is a complete unit adapted for the FLOW-i application.
The AC/DC Converter supplies 12 V_AC_DC to PC 1923 Power Control. The output from the
AC/DC Converter is 11-16 V DC and the output level is controlled by an analog 0-5 V signal.
Mains power is supplied to the AC/DC Converter via the AC/DC Power inlet, which is equipped
with two 6.3 A fuses.
66 76 980_02 3 - 67
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 68
3. DESCRIPTION OF FUNCTIONS
Fan 1
Fan 1 is located behind the fan filter below the CO2 Absorber. The fan forces cooling air into the
unit.
A tacho sensor integrated into the fan provides the fan rotation signal. The system can be
switched On only if fan rotation is detected. Refer to ‘PC 1923 Power Control’ above for further
information.
66 76 980_02 3 - 69
3. DESCRIPTION OF FUNCTIONS
LED Lamp
A LED lamp is a reading lamp integrated into the Control Panel swivel arm. The brightness can
be adjusted with the DIM knob on the swivel arm.
66 76 980_02 3 - 70
3. DESCRIPTION OF FUNCTIONS
1. Fuse F1 Lift fuse (C30 only) 5. Fuse F5 Mains power inlet fuse
2. Fuse F2 Lift fuse (C30 only) 6. Fuse F6 Mains power inlet fuse
3. Mains power inlet 7. PC 1938 Mains connection
8. Isolation transformer fuse
4. Equipotential terminal (earth)
66 76 980_02 3 - 71
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 72
3. DESCRIPTION OF FUNCTIONS
1. PC 1927 O2 Flush
2. O2 Flush button
3. PC 1928 Power On/Off
4. Power button
PC 1927 O2 Flush
PC 1927 O2 Flush comprises electronics for handling of:
O2 Flush. Detects if the O2 Flush valve is open. For redundancy, a 2-pole switch is integrated
in the O2 Flush button. Both poles must be closed to detect the O2 Flush valve as open.
Vaporizer status LED lamps. Show vaporizer status with the Vaporizer LEDs, visible just
above the vaporizer slots. The LEDs are mounted on this board. There are two sets of LEDs,
66 76 980_02 3 - 73
3. DESCRIPTION OF FUNCTIONS
one for each vaporizer slot. Each set of LED contains three LEDs; green, yellow and red. The
LEDs indicate:
Green: The selected vaporizer has passed the System checkout or Vaporizer check, and is
ready to deliver agent.
Yellow:
The vaporizer slot is selected but no vaporizer is present.
The selected vaporizer slot holds an unchecked vaporizer.
The selected slot's vaporizer is being filled.
The selected slot's vaporizer has triggered an alarm, i.e. vaporizer nearly empty.
Red: The vaporizer is disabled. The power to the vaporizer is shut off (the condition is
applicable to both slots, i.e. not only the selected slot).
Not lit/Off: The vaporizer slot is not selected.
Includes an ID PROM. The ID information can be read by the FLOW-i system.
Power button
The Power button is used to start-up the system. A white LED inside the button is lit when the
system is switched On. A glass cover protects the button.
For redundancy, the Power button contains a 2-pole switch. Both poles must be closed to
enable system start-up.
66 76 980_02 3 - 74
3. DESCRIPTION OF FUNCTIONS
PC 1919 is connected to the Patient cassette via PC 1918 Exp. cannel connector.
PC 1919 contains electronics including microprocessor for handling of:
Expiratory flow measurement performed by the Expiratory Flow Transducer inside the
cassette. The output signal Exp. Flow is an analogue signal used in sub-systems Control and
Monitoring.
All electronic connections to and from the Patient Cassette.
Measurement of Fresh gas pressure and Inspiratory pressure via PC 1781 (F) and PC 1781
(I). PC 1919 contains holders and electrical connectors for these pressure transducer boards.
Control of the APL/PEEP Valve and the Fresh gas safety valve functions.
A DC/DC converter on PC 1919 generates 200 VDC to the Expiratory Flow Transducer drivers.
A discharge circuitry makes sure that the capacitor that carries 200 V is discharged when the
cassette is disconnected.
WARNING! The PC board generates 200 V. All personnel must exercise extreme
caution if fault tracing or adjustments are performed with power supply connected
and with covers removed.
The CAN bus is responsible for all communication excluding the analogue flow signal.
A thermistor on PC 1919 monitors the temperature. An alarm is activated if the temperature is
77 ±5 °C or higher or if the temperature sensor is not connected (e.g. due to poor connection).
Includes an ID PROM. The ID information can be read by the FLOW-i system.
Green and red LEDs indicate activities related to PC 1919. The LED’s are described in chapter
Troubleshooting.
Note: The System SW must be re-installed if PC 1919 is replaced.
66 76 980_02 3 - 75
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 76
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 77
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 78
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 79
3. DESCRIPTION OF FUNCTIONS
Cables listed below have passed EMC requirements specified in IEC 60601-1-2 ed. 3:
RS232 cable (standard type, unshielded, length max. 3 m)
RS232 cable (standard type, shielded, length max. 3 m)
VGA cable (standard type, unshielded, length max. 3 m)
LAN cable (RJ-45 connector, unshielded, length max. 3 m)
66 76 980_02 3 - 80
3. DESCRIPTION OF FUNCTIONS
Includes an ID PROM. The ID information can be read by the FLOW-i system. The units System
ID (Serial No.), configuration, operating time, options, etc, is stored in an EEPROM on PC 1900.
Thus, when replacing PC 1900, a spare part that is factory programmed for the concerned unit
must be used. Other information, also required by the system, is stored on PC 1900. Refer to
section ‘Information stored on PC boards’ in chapter ‘Disassembling and assembling’.
66 76 980_02 3 - 81
3. DESCRIPTION OF FUNCTIONS
27. Lift
The Lift is available on C30 model only. With the mains power connected, the lift function is
operated by means of the Up and Down buttons located to the right of the ON/OFF button.
The Lift is a complete unit that contains an AC/DC Converter and a Lift actuator (DC motor).
AC/DC Converter
The lift is supplied with fused (F1 and F2) mains power from PC 1938 Mains connection. The
AC/DC Converter transforms and rectifies the power supply to approx. 24 VDC.
The ring core transformer is protected by a thermo fuse that will trip at approx. 100˚C.
During mains power failure, the backup battery in the anesthesia system will not power the lift.
Lift actuator
The lift actuator used is a 24 VDC permanent magnet motor with 200 stroke length. Lift speed is
>15 mm/s. The actuator is equipped with a brake which increases the self-locking ability.
The lift actuator has a 10% duty cycle; it may be used continuously for a 2 minute period, after
which it must cool down for 18 minutes. If the lift is used before it has cooled down completely,
the time of continuous use will be less than 2 minutes.
The maximum capacity is 150 kg, equivalent to approx. 30 kg of auxiliary equipment. For a
detailed description of equipment weight restrictions, refer to User’s Manual.
66 76 980_02 3 - 82
3. DESCRIPTION OF FUNCTIONS
28. AFGO
Additional Fresh Gas Outlet (AFGO) is an option to the anesthesia system that allows use of an
external partial rebreathing system, such as Bains, Jackson Rees or Mapleson D.
The AFGO Valve PV7 can be switched to supply Fresh Gas from Vaporizer either to the Patient
Cassette or to the Additional Fresh Gas Outlet. With AFGO enabled, fresh gas will be supplied
to the Additional Fresh Gas Outlet.
When using AFGO the circle system is by-passed and the following can be set:
Gas mix
O2 concentration
Fresh gas flow
Agent concentration
Fresh gas flow ranges are from 1.0 to 20 l/min when using the AFGO and the default settings
are the same as in Manual ventilation.
AFGO Valve
With the AFGO Valve PV7, the Fresh Gas from Vaporizer is supplied either to the Patient
Cassette or to the Additional Fresh Gas Outlet. The Additional Fresh Gas Outlet is a standard
22 mm cone and the Fresh Gas flow supplied is max. 20 l/min.
66 76 980_02 3 - 83
3. DESCRIPTION OF FUNCTIONS
On units with S/N below 1500, the AFGO Valve is factory mounted as default. Installation of the
AFGO Valve delivered with the AFGO option is not required.
On units with S/N above 1500, the AFGO Valve must be mounted to replace the Fresh Gas
Connector Block.
66 76 980_02 3 - 84
3. DESCRIPTION OF FUNCTIONS
The factory mounted option Auxiliary Power Outlets also includes the Isolation Transformer. The
transformer supplies power to the three extra mains power outlets and to the integrated patient
monitor power cable.
Max permissible load is:
230 VAC units: 2 A (extra outlets) + 1 A (patient monitor cable)
110 VAC units: 4 A (extra outlets) + 2 A (patient monitor cable)
The Auxiliary power outlet contains two fuses (circuit breakers):
Patient Monitor fuse rated 1 A (230 VAC) or 2 A (110 VAC)
Power Outlets fuse rated 2 A (230 VAC) or 4 A (110 VAC)
During mains power failure, the backup battery will not power the Auxiliary power outlets.
66 76 980_02 3 - 85
3. DESCRIPTION OF FUNCTIONS
1. Isolation transformer
2. Power In
3. Power Out
The Isolation Transformer is a complete unit that contains a transformer switching relay and a
ring core transformer. There are two different Isolation Transformers available, one for 110 VAC
and one for 230 VAC.
The transformer switching relay is a current limitation that connects power to the transformer in
a controlled and safe way. This will prevent fuses to trip due to high start current in the
transformer.
Power supply to the transformer is protected by a fuse (circuit breaker). The fuse is rated 6 A
(230 VAC) or 12 A (110 VAC).
Max permissible load, as specified for the power outlets, is:
230 VAC units: 2 A (extra outlets) + 1 A (patient monitor cable)
110 VAC units: 4 A (extra outlets) + 2 A (patient monitor cable)
The transformer is protected by a thermo fuse that will trip at approx. 110˚C.
66 76 980_02 3 - 86
3. DESCRIPTION OF FUNCTIONS
1. Air inlet
2. O2 inlet
3. Flowmeter outlet (with tube adapter
attached)
4. Flowmeter tube
5. Flowmeter unit regulatory valve
6. On/Off switch for the suction unit
7. Vacuum gauge for the suction unit
8. Suction unit regulatory valve
9. Mount for hydrophobic bacterial and viral
filter
66 76 980_02 3 - 87
3. DESCRIPTION OF FUNCTIONS
Flowmeter unit
The O2 gas supply hose is connected to the O2 inlet (2). O2 supply pressure must be 300-650
kPa. An O2 filter is mounted inside the module.
Outlet flow is regulated by means of the flowmeter regulating valve (5). Current flow is shown by
the flowmeter tube (4) where the center of the ball aligned to the printed scale corresponds to
the actual flow. Max. flow is 10 l/min.
With the tube adapter (3) attached, the flowmeter outlet can be connected to 4, 6 and 8 mm
tubing.
Suction unit
The Air gas supply hose is connected to the Air inlet (1). Air supply pressure must be 300-650
kPa.
The suction pressure is created by an ejector inside the Suction unit. The ejector outlet, inside
the unit, is equipped with a silencer to reduce the sound level from the ejector.
Suction pressure is regulated by means of the On/Off switch (6) and the suction unit regulatory
valve (8). The vacuum gauge (7) shows the current suction pressure.
Max. suction pressure is -60 kPa down to -90 kPa (depending on the Air supply pressure.
The suction jar used in conjunction with the unit must have a volume of at least 0.5 liters. The
tube connecting the suction jar with the suction unit must have an inner diameter of at least 6
mm and not be more than 1 m in length. The tube material should be rigid enough to prevent
occlusion.
A hydrophobic bacterial/viral filter must be connected to the suction unit. It protects the system
against contagious particles and acts as a second overflow protection device in addition to the
device located in the suction jar. The suction unit cannot be operated without a correctly placed
bacterial/viral filter.
A visual inspection and a function check of the Auxiliary O2 and Suction Module must be
performed during the Preventive Maintenance.
66 76 980_02 3 - 88
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 89
3. DESCRIPTION OF FUNCTIONS
The Backup Gas Trolley is equipped with PIN-index connections (yokes) for 2 gas cylinders:
O2 gas cylinder. Max cylinder gas pressure is 20000 kPa (200 bar) before the pressure
regulator.
AIR gas cylinder. Max cylinder gas pressure is 20000 kPa (200 bar) before the pressure
regulator.
N2O gas cylinder. Max cylinder gas pressure is 8000 kPa (80 bar) before the pressure
regulator.
Max. allowed gas cylinder size is: length 655 mm (valve excluded), diameter 140 mm.
The gas inlet channels, inside the yokes, contain inlet filters and one-way valves (OV21 and
OV22). The filters must be replaced during ”Preventive maintenance”.
The cylinder gas pressure is reduced to 350 kPa by the pressure regulators (REG21 and
REG22). The gas is then connected to the Gas distribution section via tubing. The pressure
regulators are preset in factory and cannot be adjusted during field service.
The pressure transducers on PC 1916 Cylinder pressure connection are connected to PC 1903
External connectors, thus the gas pressure in the cylinders will be displayed on the Control
panel. Alarm is activated if the gas pressure is outside the preset alarm limits.
The cable connector on PC 1916 is protected by a cover. When this cover is open (e.g. when
the signal cable is connected), the trolleys locking mechanism is blocked and the trolley cannot
be removed from FLOW-i.
There are two safety valves (SV21 and SV22) on the trolley; one for each gas. The safety
valves are preset to open at a pressure of 650 kPa (6.5 bar).
Filters and seals in the yokes as well as cylinder gas supply filters and O-rings in the Gas
Distribution Section must be replaced during the Preventive Maintenance. The gas tube
connections must also be checked during Preventive Maintenance. The gas tube connections
must be tightened firmly with a wrench. It must not be possible to disconnect the gas tubes by
hand (without wrench).
66 76 980_02 3 - 90
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 91
3. DESCRIPTION OF FUNCTIONS
The Backup Gas Holder is equipped with PIN-index connection (yoke) for 1 gas cylinder:
N2O gas cylinder. Max cylinder gas pressure is 8000 kPa (80 bar) before the pressure
regulator.
Max. allowed gas cylinder size is:
For C20: length 655 mm (valve excluded), diameter 140 mm.
For C30: length 595 mm (valve excluded), diameter 140 mm. A bottom plate mounted on C30
units prevents cylinders with lengths > 595 mm to be connected.
The gas inlet channel, inside the yoke, contains an inlet filter and a one-way valve (OV23). The
filter must be replaced during ”Preventive maintenance”.
The cylinder gas pressure is reduced to 350 kPa by the pressure regulator (REG23). The gas is
then connected to the Gas distribution section via tubing. The pressure regulator is preset in
factory and cannot be adjusted during field service.
The pressure transducer is connected to PC 1903 External connectors, thus the gas pressure in
the cylinder will be displayed on the Control panel. Alarm is activated if the gas pressure is
outside the preset alarm limits.
There is a safety valve (SV23) on the holder. The safety valve is preset to open at a pressure of
650 kPa (6.5 bar).
Filter and seal in the yoke as well as cylinder gas supply filter and O-ring in the Gas Distribution
Section must be replaced during the Preventive Maintenance. The gas tube connections must
also be checked during Preventive Maintenance. The gas tube connections must be tightened
firmly with a wrench. It must not be possible to disconnect the gas tube by hand (without
wrench).
66 76 980_02 3 - 92
3. DESCRIPTION OF FUNCTIONS
66 76 980_02 3 - 93
NOTES
66 76 980_02
4. DISASSEMBLING AND ASSEMBLING
66 76 980_02 4-1
4. DISASSEMBLING AND ASSEMBLING
Preparations
Before disassembling or assembling, make sure that:
All gas conveying parts are cleaned according to instructions in the FLOW-i – User's Manual.
Gas supply is disconnected.
Mains power cable is disconnected.
Power button is switched Off. If the Power button is switched On, the power backup battery
will supply power all electrical components in the unit.
Note that PC 1900 Main back-plane, PC 1903 External connectors, PC 1923 Power control,
the Emergency ventilation switch and the Power button are energized by the power backup
battery also when the Power button is switched Off.
The power backup battery can be disconnected either by disconnecting the battery cables at
the battery terminals or by disconnecting the battery cables (Battery and Battery sense) at
PC 1903 External connectors.
WARNING! With gas supply connected to the FLOW-i, there are pressurized
components inside the unit. All personnel must exercise extreme caution if fault
tracing or adjustments are performed with gas supply connected and with covers
removed.
WARNING! With power supply connected to the FLOW-i, there are energized
electrical components inside the unit. All personnel must exercise extreme caution if
fault tracing or adjustments are performed with power supply connected and with
covers removed.
Handling vaporizers
Handle the vaporizer with care. Avoid turning a vaporizer containing anesthetic agent upside
down or laying it sideways.
The vaporizers are factory calibrated and leakage tested.
The vaporizer must not be disassembled.
If the vaporizer is dropped, it must be serviced by MAQUET to ensure proper functionality.
Repair or service in case of malfunction must be carried out by MAQUET only.
The RMA procedure must be used when returning vaporizers to MAQUET.
The vaporizer must be emptied before it is returned to MAQUET. Use a bottle with filling adapter
to empty the vaporizer. Refer to emptying instructions in the User’s Manual.
Note: There may be small amounts of anesthetic agent remaining in the vaporizer, although the
indicator shows that the vaporizer is empty.
66 76 980_02 4-2
4. DISASSEMBLING AND ASSEMBLING
Handling PC boards
The PC boards contain components that are highly sensitive to static electricity.
Those who come into contact with circuit boards containing sensitive components must take
certain precautions to avoid damaging the components (ESD protection).
When working with ESD sensitive components, always use a grounded wrist band and
grounded work surface. Adequate service tools must also be used.
PC boards (spare parts) must always be kept in protective packaging for sensitive electronic
device.
PC boards must not be inserted or removed while the mains power or battery power is applied
to the PC boards.
Remove and insert the PC boards very carefully to avoid damage to the connectors.
Replacing PC boards
The FLOW-i software is distributed to the following PC boards:
PC 1910 Vaporizer Controller Board (included in vaporizer)
PC 1919 Expiratory Channel
PC 1920 Control
PC 1921 Monitoring
PC 1922 Panel
PC 1923 Power Control
PC 1925 Display CPU.
When delivered as spare parts, these PC boards are equipped with a 'System SW version' that
may differ from the version on the unit to be repaired.
Note: PC 1910 Vaporizer Controller Board is included in vaporizer and not available as spare
part. Software installation on PC 1910 will be performed with the vaporizer connected during
System software installation.
To keep the 'System SW version' used prior to the PC board replacement, the applicable
'System SW version' must be available for reinstallation purposes.
For functionality enhancement, the latest released 'System SW version' is always
recommended. Before installing a new 'System SW version' on a unit, ensure that the software
is fully compatible with all HW-, SW- and mechanical components in the unit. If any compatibility
conflicts are apparent this will be noted on the 'MCC SW download' website.
66 76 980_02 4-3
4. DISASSEMBLING AND ASSEMBLING
66 76 980_02 4-4
4. DISASSEMBLING AND ASSEMBLING
Assembling guidelines
All parts of the FLOW-i system assembled with screws and nuts are tightened with a specified
torque. Thread locking adhesives are used if required.
In order to maintain these specifications over time, it must be ensured that after any service
intervention removed parts are re-assembled and secured according to instructions. Make sure
to follow the guidelines stated below.
Tightening torque
Thread size M3: 0.95 Nm ±15%
Thread size M4: 3.1 Nm ±15%
Thread size M5: 5.0 Nm ±15%
Thread size M6: 7.0 Nm ±15%
Thread size M20: 7.0 Nm ±15% (nut on release button for cassette lid)
Special grease
Lubricants should normally not be used when servicing the unit. If lubricants must be used, use
only very small amounts of grease with P/N 95 73 700. This special grease is O2 compatible.
WARNING! Due to the increased O2 concentration in valves and gas channels, only
special grease intended for use in high O2 concentrations, are allowed.
Internal tubing
During disassembling of the unit, it may be required to disconnect internal tubing. Disconnect
carefully to avoid damage to the tubing or connectors.
CAUTION: Tubing included in the gas sampling system must not be cut before reconnected.
The tubing length is calculated and adapted to the sampling functionality. Proper operation of
the sampling system cannot be ensured if the tubing length is incorrect. Cut tubing must be
replaced.
During assembling of the unit, it is important that all tubing is properly re-connected. Faulty or
poor connections will result in system malfunction.
WARNING! Ensure that the EVAC system, including tubing from scavenging valves,
is correctly connected. Running the system with a poor connection may result in
anesthetic agent being emitted into the operation environment. Leakages in the
EVAC system will not be detected by the System Checkout.
66 76 980_02 4-5
4. DISASSEMBLING AND ASSEMBLING
To connect tubing with collar nut: 2. Tighten the collar nut firmly by hand.
1. Push the tubing onto the nipple until stop.
To disconnect sampling 2. Put the blade under the To connect sampling tubing:
tubing: tubing end and carefully Carefully push the tubing
1. Carefully lift up the locking pry loose the tube. onto the nipple.
sleeve using a flat-blade Push the locking sleeve
screwdriver. over the nipple by hand
(with two fingers).
66 76 980_02 4-6
4. DISASSEMBLING AND ASSEMBLING
Cable clamps
The mains power cable must be secured
with a cable clamp (1)
66 76 980_02 4-7
4. DISASSEMBLING AND ASSEMBLING
Control panel
WARNING! With power supply connected to the FLOW-i, there are energized
electrical components inside the unit, e. g. the backlight lamps that are supplied with
660 V by the Backlight Inverter. All personnel must exercise extreme caution if fault
tracing or adjustments are performed with power supply connected and with the user
interface rear cover removed.
66 76 980_02 4-8
4. DISASSEMBLING AND ASSEMBLING
Covers
Front door
To open the front door:
Remove screws (1) and lift off the vaporizer
slot housing (2).
66 76 980_02 4-9
4. DISASSEMBLING AND ASSEMBLING
Rear cover
Disconnect the Control panel cable (1).
Loosen or remove rail (2).
Carefully lift off the rear cover (3).
Side cover
FLOW-i C20 shown in illustration.
Disconnect the mains power cable (1).
Loosen or remove rail (2).
Remove all screws (3).
Carefully lift off the side cover (4).
66 76 980_02 4 - 10
4. DISASSEMBLING AND ASSEMBLING
Gas modules
The four Gas modules are:
1. Reflector gas module
2. Fresh gas module O2
3. Fresh gas module AIR
4. Fresh gas module N2O.
66 76 980_02 4 - 11
4. DISASSEMBLING AND ASSEMBLING
PC board rack
To open the PC board rack:
Remove the two screws (1).
Lift off the PC board rack cover plate (2).
66 76 980_02 4 - 12
5. SERVICE PROCEDURES
66 76 980_02 5-1
5. SERVICE PROCEDURES
Service key
The Service key is a USB memory stick provided to personnel trained and authorized by
MAQUET to perform installation, service or maintenance of the FLOW-i.
The Service key and its associated access code are personal and must not be handed-over to
anyone else. The access code is unique for each Service key.
With the Service key it is possible to use different access levels of the built-in service software’s.
These service software’s are:
Tools
Service & Settings
See descriptions below.
Note 1: When saving information to a USB memory (logs, screenshots, etc), allow this
procedure to complete before removing the memory stick. Information will be shown on the
control panel when completed.
Note 2: Make sure to log out (select Close) from Service & Settings when completed. If the
Service key is removed with Service & Settings open, a limited access to Service & Settings
remains.
Tools
With the membrane button Menu, a number of functions will be available for all users. These
functions are described in the User’s Manual.
When the Service key is connected, the Menu will be extended with the additional Tools button.
The Tools button will provide two service tools:
Gas analyzer tool. Press this button to display metrics from the Control Gas Analyzer (CGA)
and the Patient Gas Analyzer PGA). Can be used during verification of the gas analyzers.
Vaporizer tool. Press this button to display the active vaporizers status.
Note that CO2 concentration is always displayed in % independent of the unit chosen in Startup
configuration.
66 76 980_02 5-2
5. SERVICE PROCEDURES
The table in chapter Diagrams shows the ‘Authorized biomed’ and the ‘SSU/SSP’ access level
of Service & Settings.
Service & Settings functions are further described in this chapter. See sections Logs, Tests,
Calibrations, Service functions, Service report, Startup configuration and Settings.
66 76 980_02 5-3
5. SERVICE PROCEDURES
CAUTION: The unit must not be switched off during the Installer update process.
Such interrupt of the Installer update will make the concerned PC board defective
and the PC board must be replaced.
As from System version 2.0, the AION gas analyzers (CGA and PGA) are included in the
System software installation. The AION software will automatically be installed during the
System software installation. This will add approx.10 minutes to the installation procedure.
Note: The version of the already installed AION software cannot be checked during the
installation phase. The AION software will thus always be installed when performing a System
software installation.
APL calibration values will be lost during System software installation. Calibrate the APL knob
after a System software installation. The calibration procedure is implemented and described i
Service & Settings > Calibration > APL knob. Service key is required to access Calibration in
S&S.
The System software is distributed to a number of PC boards, see chapter ‘Disassembling
and assembling’, section ‘Replacing PC boards’. If such PC board is replaced, a System
software installation must be performed to assure that correct software version is installed on
all PC boards. A compatibility check of the PC boards is performed during each FLOW-i
startup.
After any installation, maintenance or service intervention in the FLOW-i, perform a 'System
checkout' and a ‘Manual check of Emergency ventilation system’ according to instructions in
the FLOW-i - User's Manual.
66 76 980_02 5-4
5. SERVICE PROCEDURES
Required equipment
USB memory with the System software version to be installed. Only MAQUET approved USB
memory sticks must be used on FLOW-i.
Tool to press the Boot enable switch, e.g. a 2 mm hexagonal wrench.
Service key to access Calibration in Service & Settings.
Installation procedure
1. Copy the Startup configuration to a USB
memory.
2. Switch off the system using the Power
button.
3. Connect the USB memory, with the
System software version to be installed, to
the USB port.
4. Make sure that the vaporizers to be
covered by this software installation are
connected.
5. Open the Emergency ventilation cover to
access the Boot enable switch.
6. Keep the Boot enable switch pressed (e.g.
Pressing the Boot enable switch with the hexagonal wrench as shown in
the illustration) and at the same time, start
the system using the Power button.
7. Release the Boot enable switch.
8. The system will start up and the FLOW-i Installer screen will open. The installer screen
contains a number of installation progress bars. The screenshot below shows a completed
installation.
9. The installation procedure will now start automatically and the installation progress will be
shown on the installation bars. See Color code information below. The total installation time
will be approx. 20 minutes plus 3 minutes for each connected vaporizer.
Note: If an Installer update is required, this will be performed during the first phase of the
System software installation. When the Installer upgrade is completed, the System software
installation must be restarted to finalize the installation.
66 76 980_02 5-5
5. SERVICE PROCEDURES
10. The installation of the different sub-systems is shown on the installer screen as follows:
11. When the installation is successfully completed, switch off the system using the Power
button.
12. If further vaporizers must be updated, connect these vaporizers and repeat the installation
procedure from work step 4. The Vaporizer software installation time will be approx. 3
minutes for each vaporizer.
13. Remove the USB memory with the System software.
14. Start the system using the Power button. A second restart may be required due to changes
in the configuration and/or Technical alarms.
15. In the Status window, check that correct System software version is installed.
16. Calibrate the APL knob. The calibration procedure is implemented and described in Service
& Settings > Calibration > APL knob. Service key is required to access Calibration in S&S.
17. Connect the USB memory containing the Startup configuration and copy the configuration
from the USB memory. Remove the USB memory when completed.
18. Perform a System Checkout and a ‘Manual check of Emergency ventilation system’
according to instructions in the FLOW-i - User's Manual.
Color Description
Black Waiting – Reading on the USB memory not started.
Blue Skipped – Same software version already installed.
Purple Transferring – Software transfer to subsystem in progress.
Yellow Writing – Software transfer to flash memory in progress.
Red Failure – Software installation error (see below).
Gray File Not Found – Software not found on the USB memory.
Green Done / Successful – Software installation completed
66 76 980_02 5-6
5. SERVICE PROCEDURES
Failure
If the System software installation fails, restart the complete installation procedure.
Possible software installation errors:
Files on the USB memory corrupt or not recognized.
Software transfer to flash memory failed.
Checksum error.
66 76 980_02 5-7
5. SERVICE PROCEDURES
Required equipment
USB memory with the Software option to be installed. Only MAQUET approved USB memory
sticks must be used on FLOW-i.
Installation procedure
1. Switch off the system using the Power
button.
2. Connect the USB memory, with the
Software option to be installed, to the USB
port.
66 76 980_02 5-8
5. SERVICE PROCEDURES
66 76 980_02 5-9
5. SERVICE PROCEDURES
Logs
General
A number of logs are accessible in Service & Settings. The logs can be viewed and used for
troubleshooting of the unit.
If further support is required, the logs can be downloaded and sent to HSC via MSupport.
When downloading logs for support purpose, it is important that all required logs are included.
See instructions below.
Required equipment
USB memory for the downloaded logs. Only MAQUET approved USB memory sticks must be
used on FLOW-i.
66 76 980_02 5 - 10
5. SERVICE PROCEDURES
66 76 980_02 5 - 11
5. SERVICE PROCEDURES
Tests
General
A number of tests are accessible in Service & Settings. The tests can be used for
troubleshooting of the unit.
Required equipment
Service key with access to Tests in Service & Settings.
A 14-18-20 mm rubber plug from the service kit (required for the AFGO leakage check only).
66 76 980_02 5 - 12
5. SERVICE PROCEDURES
Leakage checks
Internal leakage
66 76 980_02 5 - 13
5. SERVICE PROCEDURES
Test procedure
1. Connect gas supply (AIR and O2).
2. Plug the AGFO connector. A 14-18-20 mm
rubber plug from the service kit should be
used.
3. Select Tests.
4. Select Leakage check.
5. Select AFGO leakage.
6. Press Start.
66 76 980_02 5 - 14
5. SERVICE PROCEDURES
Calibrations
General
A number of calibrations are accessible in Service & Settings. The calibrations can be used for
verification, maintenance, etc, of the unit.
Required equipment
Service key with access to Calibration in Service & Settings.
Further equipment required is listed for each calibration procedure.
Calibration menu
66 76 980_02 5 - 15
5. SERVICE PROCEDURES
APL knob
The APL potentiometer must be calibrated:
After replacement of the APL potentiometer.
When the set value differs from the value shown on the screen.
After System software installation. APL calibration values stored on PC 1921 Monitoring will
be lost.
After replacement of PC 1921 Monitoring or in case of malfunction in the memory backup
battery on PC 1921.
After any installation, maintenance or service intervention in the FLOW-i, perform a 'System
checkout' and a ‘Manual check of Emergency ventilation system’ according to instructions in the
FLOW-i - User's Manual.
Calibration procedure
1. Select Calibration.
2. Select APL knob.
3. Follow instructions in the Calibration APL
knob menu.
66 76 980_02 5 - 16
5. SERVICE PROCEDURES
Barometer
The Barometer must be calibrated:
If the Barometer test in System Checkout fails.
If a Technical error code indicates Barometer error.
If the Barometric pressure check performed during installation or Preventive Maintenance
shows that calibration is required, i.e. if the ‘Current barometer pressure’ differ more than ±5%
from the actual barometric pressure.
After replacement of PC 1921 Monitoring or in case of malfunction in the memory backup
battery on PC 1921.
After any installation, maintenance or service intervention in the FLOW-i, perform a 'System
checkout' and a ‘Manual check of Emergency ventilation system’ according to instructions in the
FLOW-i - User's Manual.
Required equipment
Barometer or information about the barometric pressure at the installation site.
Calibration procedure
1. Select Calibration.
2. Select Barometer.
3. Enter a ‘New barometer pressure’ and
press Set.
4. Check that the new ‘Current barometer
pressure’ corresponds to the set value.
Note that the new ‘Current barometer
pressure’ value may differ 1-2 cmH2O
from the set value.
66 76 980_02 5 - 17
5. SERVICE PROCEDURES
Gas analyzer
Gas analyzer calibration covers:
Patient gas analyzer. Calibration using Service & Settings, described below.
Control gas analyzer. Calibration using Service & Settings, described below.
Control O2 sensor. Calibration using Service & Settings, described below.
Patient O2 sensor. Calibration using KMG 2000 Service Software, described below.
Before starting any check or calibration of AION or O2 sensors, read the complete ‘Gas
analyzer’ chapter below to be familiar with the procedures.
Leakage in the system will result in a faulty calibration. Always perform a System Checkout
before starting the calibration, refer to the FLOW-i – User’s Manual.
After any installation, maintenance or service intervention in the FLOW-i, perform a 'System
checkout' and a ‘Manual check of Emergency ventilation system’ according to instructions in the
FLOW-i - User's Manual.
Required equipment
Calibration gas kit including gas bottle, regulator and tubing. P/N 52 07 415.
During calibration of the gas analyzers, a sampling flow of 200 ml/min is used. As the flow
regulator on the calibration gas bottle is set to approx. 150 ml/min, the collector bag must be
used as a gas reservoir to ensure sufficient calibration gas supply.
Calibration adapter CGA, P/N 66 86 915.
Gas evacuation system for the calibration gas.
For calibration of the Patient O2 sensor:
PC or laptop with KMG 2000 Service Software.
Service cable, P/N 65 23 570.
66 76 980_02 5 - 18
5. SERVICE PROCEDURES
66 76 980_02 5 - 19
5. SERVICE PROCEDURES
66 76 980_02 5 - 20
5. SERVICE PROCEDURES
66 76 980_02 5 - 21
5. SERVICE PROCEDURES
A completed calibration sequence will display the last page in the calibration procedure.
A dialogue on this page will display the result of the calibration:
If the calibration check resulted in measured calibration values already within valid
intervals, the text ‘Calibration not necessary’ will appear.
If calibration was performed and accepted, the text ‘Calibration accepted by user’ will
appear.
A cancelled calibration will also end at the last page if the cancellation was submitted while
showing a page displaying calibration values. In this case, the text ‘Calibration cancelled by
user. Ensure that the calibration gas bottle valve is closed’ will appear.
The result of the calibration procedure is saved in the Service report log in Service & Settings.
See Logs > Service report.
The result of the calibration procedure (e.g. Last calibration and Last calibration check) is also
saved in Gas analyzer status in Service & Settings. See Status > Gas analyzer.
66 76 980_02 5 - 22
5. SERVICE PROCEDURES
O2 sensor calibration
Note: As the O2 sensor is flow-sensitive, the AION and the O2 sensor must be calibrated
together. The O2 sensor calibration must be carried out when either of the two units are
replaces.
66 76 980_02 5 - 23
5. SERVICE PROCEDURES
Patient O2 sensor
For Patient O2 sensor, calibration is performed with the KMG 2000 Service Software.
The Service Software is required to:
Perform a Patient O2 sensor calibration
Update the Patient O2 sensor serial number stored in AION EEPROM if the O2 sensor has
been replaced.
The Service Software, run on a PC connected to the FLOW-i, will enable a number of service
and troubleshooting options.
The Service Software will e.g. display:
Gas concentrations
Status for the different sub-units
Error logs
Error log frequency
Messages from the gas analyzers.
66 76 980_02 5 - 24
5. SERVICE PROCEDURES
66 76 980_02 5 - 25
5. SERVICE PROCEDURES
8. Follow the instructions in the Service Software on how to perform the calibration.
9. After completed calibration:
Close the calibration gas bottle valve.
Allow the FLOW-i to purge the gas in the calibration setup, including the collection bag,
into the gas evacuation system.
Set the 3-way-valve to ‘room air’.
Remove the calibration gas kit.
10. In Service & Settings, select Disconnect.
11. Exit the Service Software. Switch off the PC and disconnect the service cable.
12. Restart FLOW-i.
13. In Service & Settings:
Select Status
Select Gas analyzer
Check that O2 sensor Last calibration date is updated.
14. This will complete the calibration in the Patient O2 sensor.
66 76 980_02 5 - 26
5. SERVICE PROCEDURES
Touch screen
The touch screen in FLOW-i uses a nine point calibration algorithm. The touch screen must be
calibrated:
If an inaccuracy is seen in the touch screen function.
After replacement of PC 1922 Panel or in case of malfunction in the memory backup battery
on PC 1922.
After any installation, maintenance or service intervention in the FLOW-i, perform a 'System
checkout' and a ‘Manual check of Emergency ventilation system’ according to instructions in the
FLOW-i - User's Manual.
Calibration procedure
1. Select Calibration.
2. Select Touch screen.
The calibration procedure can be restarted by pressing the Delete all marks button. This action
deletes all sample points accepted so far and turns the cross mark color back to black.
The calibration procedure can be cancelled by pressing the Cancel button. This action deletes
all sample points and closes the calibration window. Old calibration values are retained.
66 76 980_02 5 - 27
5. SERVICE PROCEDURES
Service functions
General
Service functions accessible in Service & Settings are:
Valve tests
Gas analyzer.
They can be used for service and troubleshooting of the unit.
Required equipment
Service key with access to Service functions in Service & Settings.
66 76 980_02 5 - 28
5. SERVICE PROCEDURES
Service report
General
Service report in Service & Settings can be used by Field Service Engineers to create and save
service and maintenance information. The selected service tasks will be logged in the Service
report log.
Required equipment
Service key with access to Service report in Service & Settings.
66 76 980_02 5 - 29
5. SERVICE PROCEDURES
Startup configuration
General
Startup configuration in Service & Settings can be used to create and save the units Startup
configuration.
In this menu, it is also possible to save the Startup configuration to a USB memory or copy the
Startup configuration from a USB memory.
Note: Some parts of the Startup configuration may be changed during System software
installation. Copy the Startup configuration to a USB memory prior to a System software
installation and reinstall the configuration afterwards.
Required equipment
Service key with access to Startup configuration in Service & Settings.
In this menu, it is also possible to save the Startup configuration to a USB memory or copy the
Startup configuration from a USB memory. All selectable parameters are shown in the Service &
Settings table in chapter Diagrams.
In the Biomed access code menu it is possible to change the default Biomed access code
(1973) to different four-digit code.
66 76 980_02 5 - 30
5. SERVICE PROCEDURES
Settings
General
Settings in Service & Settings can be used to set date and time and to calculate total anesthetic
agent usage.
Required equipment
Service key with access to Settings in Service & Settings.
Settings menu
66 76 980_02 5 - 31
5. SERVICE PROCEDURES
Leakage detection
General
The leakage test in the System Checkout (SCO), or the in the separate Leakage check, are
performed using a pressure of 50 cmH2O for AUTO ventilation and 30 cmH2O for MAN
ventilation.
The measured leakage for AUTO and MAN ventilation will be displayed separately in ml/min. A
maximum leakage of up to 150 ml/min is allowed for each of the mode (AUTO and MAN).
When a leakage is detected, the leakage check will stop and display 'AUTO ventilation leakage'
or 'MAN ventilation leakage' on the screen.
To locate the leakage, the procedure below describes a method to step-by-step check different
parts of the system.
Test cases 1–5 do not require any tools. As from Test case 6, the following leakage detection
tools are required:
Leakage detector, P/N 65 03 721
Plug kit, P/N 65 03 739
Additional plugs and fittings for FLOW-i, P/N 66 84 720
Calibration manometer, P/N 61 03 527
Note: Before proceeding with the Test cases below, make sure that the patient tubes, sample
filters and water trap with sampling line works properly.
66 76 980_02 5 - 32
5. SERVICE PROCEDURES
Note: The compressible volume in the system Possible cause for leakage in CO2 Absorber:
is calculated during this AUTO ventilation
1. Cracks or damages in plastics of the
leakage test. As the absorber is removed, the
absorber canister.
calculated value will be incorrect.
2. Damaged seals in the absorber valves (in
If a dialog ‘Compressible volume out of range.
the cassette).
Check the breathing system.’ appears during
this test, select ‘Bypass test’.
A complete ‘System Checkout’, with absorber
connected, must be performed after any
service intervention in the FLOW-i.
66 76 980_02 5 - 33
5. SERVICE PROCEDURES
66 76 980_02 5 - 34
5. SERVICE PROCEDURES
Note: As from Test case 6, leakage detection tools are required. Refer to General above.
WARNING! With power supply connected to the FLOW-i, there are energized
electrical components inside the unit. All personnel must exercise extreme caution
if fault tracing or adjustments are performed with power supply connected and with
covers removed.
Test Case 7 – APL/PEEP valve coil, Manual ventilation valve actuator and Fresh gas
safety valve
66 76 980_02 5 - 35
5. SERVICE PROCEDURES
Closed valve
Open valve
66 76 980_02 5 - 36
5. SERVICE PROCEDURES
66 76 980_02 5 - 37
5. SERVICE PROCEDURES
66 76 980_02 5 - 38
5. SERVICE PROCEDURES
66 76 980_02 5 - 39
5. SERVICE PROCEDURES
66 76 980_02 5 - 40
5. SERVICE PROCEDURES
66 76 980_02 5 - 41
5. SERVICE PROCEDURES
66 76 980_02 5 - 42
5. SERVICE PROCEDURES
WARNING! With gas supply connected to the FLOW-i, there are pressurized
components inside the unit. All personnel must exercise extreme caution if fault
tracing or adjustments are performed with gas supply connected and with covers
removed.
WARNING! With power supply connected to the FLOW-i, there are energized
electrical components inside the unit. All personnel must exercise extreme caution if
fault tracing or adjustments are performed with power supply connected and with
covers removed.
Required equipment
Standard service tools.
Connector kit for pressure tester P/N 66 85 355.
Digital handheld pressure gauge, local purchase. Recommended measuring range 0 – 700
kPa, accuracy ±0.2%.
Procedure
1. Open the front door. Refer to chapter
‘Disassembling and assembling’.
2. Connect gas supply (AIR and O2).
3. Connect Mains power.
4. Run the FLOW-i in Standby mode.
5. Connect the Connector kit for pressure
tester (1) to the concerned regulator, see
information below.
6. Check the regulator output pressure on
the pressure gauge (2). See pressure
values below.
66 76 980_02 5 - 43
5. SERVICE PROCEDURES
66 76 980_02 5 - 44
5. SERVICE PROCEDURES
Required equipment
Standard service tools
Patient tubing and Y-piece
Plug for the manual ventilation bag connector on the patient cassette.
Calibration manometer, P/N 61 03 527
Check procedure
1. Connect O2 to FLOW-i.
2. Connect patient tubing and Y-piece.
3. Connect the calibration manometer to the Y-piece. Make sure that the manometer is
plugged with its cap.
4. Plug the manual ventilation bag connector on the patient cassette.
5. Switch on Emergency ventilation.
6. Set the O2 flowmeter to 10 l/min.
7. Turn the APL knob counter-clockwise to stop.
8. Check that the manometer shows 6 cmH2O ±4 cmH2O.
9. Turn the APL knob clockwise to the first tactile resistance can be felt (and a slight 'click').
10. Check that the manometer shows 34 cmH2O ±10 cmH2O.
11. Turn the APL knob clockwise to stop.
12. Check that the manometer shows 80 cmH2O ±12 cmH2O.
66 76 980_02 5 - 45
5. SERVICE PROCEDURES
Calibration procedure
1. Connect O2 to FLOW-i.
2. Pull off the Mechanical APL knob.
3. Connect patient tubing and Y-piece.
4. Connect the calibration manometer to the
Y-piece. Make sure that the manometer
is plugged with its cap.
5. Plug the manual ventilation bag
connector on the patient cassette.
6. Switch on Emergency ventilation.
7. Set the O2 flowmeter to 10 l/min.
66 76 980_02 5 - 46
5. SERVICE PROCEDURES
66 76 980_02 5 - 47
5. SERVICE PROCEDURES
Tilt adjustment
Remove the plastic cover (1) to make the tilt
screw accessible.
Tighten or loosen the screw (2) for tilt
adjustment. Use a 5 mm hexagonal wrench.
Mount the plastic cover (1).
66 76 980_02 5 - 48
5. SERVICE PROCEDURES
Preparations
Switch off the system using the Power button.
Disconnect the mains power cable.
Disconnect the gas supplies (wall and/or cylinder).
Remove patient tubing.
Important: The power backup batteries replacement date must be reset after replacement. The
replacement date is set using the Service report menu (Replaced internal battery).
66 76 980_02 5 - 49
5. SERVICE PROCEDURES
ESD sensitive components. The lithium batteries are mounted on ESD sensitive
PC boards. Refer to chapter 'Disassembling and assembling', section 'Handling PC
boards' for further information regarding ESD sensitive components
Preparations
Switch off the system using the Power button.
Disconnect the mains power cable.
Disconnect the gas supplies (wall and/or cylinder).
Remove patient tubing.
66 76 980_02 5 - 50
5. SERVICE PROCEDURES
Important: The memory backup batteries replacement date must be reset after replacement.
The replacement date is set using the Service report menu (Replaced memory backup
batteries).
66 76 980_02 5 - 51
5. SERVICE PROCEDURES
1
To maintain the lift function, the combined load on the extra equipment and accessories should not
exceed 30 kg
2
Maximum configurations is dependant on the lift capacity (minimum requirement 200 kg) of ceiling
pendants approved for C40 and the maximum capacity of the C40-pendant interface (250 kg). This gives a
range of approx. 50-100 kg for accessories and extra equipment, depending on the ceiling pendant.
3
C30 can only be equipped with the additional arm or the extra backup gas holder.
4
The system can be equipped with either Top shelf or Patient monitor.
66 76 980_02 5 - 52
5. SERVICE PROCEDURES
Procedure
Remove the plastic cable holder (1) and
disconnect the Control panel cable (2).
Disconnect the mains power supply cable
(3) from the Patient monitor (Philips monitor
shown in illustration).
Unscrew the mounting knobs (4) and
remove the Control panel and the Patient
monitor from the current mountings.
66 76 980_02 5 - 53
5. SERVICE PROCEDURES
66 76 980_02 5 - 54
5. SERVICE PROCEDURES
66 76 980_02 5 - 55
5. SERVICE PROCEDURES
66 76 980_02 5 - 56
5. SERVICE PROCEDURES
CAUTION: Disconnect all electrical cables and gas connections before undocking
the system from the ceiling pendant.
CAUTION: Make sure that the swivel arm, extra equipment and accessories, e.g.
support arm and additional table, are positioned as shown in the FLOW-i – User’s
Manual, chapter Storage and transportation.
66 76 980_02 5 - 57
5. SERVICE PROCEDURES
66 76 980_02 5 - 58
5. SERVICE PROCEDURES
Note 1: The Functional check must be performed on all connected gas cylinders, i.e. gas supply
from the Backup gas trolley and the Backup gas holder.
Note 2: If Technical alarms TE 516 or TE 517 appears, this may indicate incorrect connection of
backup gas supply (gas concentration delivery error).
Functional check
Perform a System Checkout with FLOW-i
If an AIR backup gas cylinder is
connected:
connected to central gas supply to ensure
Set Auto/Man switch to MAN.
that the system is fully functional and has
Set APL = SP.
no leakage.
Set Fresh gas flow = 6.0 l/min.
Connect the Y-piece to the System
Set O2 conc. = 40%.
Checkout test plug.
Press Start case.
If an O2 backup gas cylinder is Disconnect the central gas supply for AIR.
connected: A dialog which prompts the user to open
Set Auto/Man switch to MAN. the AIR backup gas supply appears on the
Set APL = SP. control panel.
Set Fresh gas flow = 6.0 l/min. Confirm the dialog.
Set O2 conc. = 40%. Open the AIR gas cylinder and let the
Press Start case. system work for 90 seconds.
Disconnect the central gas supply for O2. Check that no technical alarms appear.
A dialog which prompts the user to open End case.
the O2 backup gas supply appears on the Restore central gas supply.
control panel. Close the AIR gas cylinder.
Confirm the dialog.
Open the O2 gas cylinder and let the
system work for 90 seconds.
Check that no technical alarms appear.
End case.
Restore central gas supply.
Close the O2 gas cylinder.
66 76 980_02 5 - 59
5. SERVICE PROCEDURES
66 76 980_02 5 - 60
6. TROUBLESHOOTING
66 76 980_02 6-1
6. TROUBLESHOOTING
General
Before starting troubleshooting, try to eliminate all possibilities of operational errors. If the
malfunction remains, use the troubleshooting guides below as well as the information in chapter
'Description of functions' to locate the faulty part.
There should be a logical trend to your troubleshooting. First try to trace the faulty function.
Then the most suitable procedure is to track down the exact fault by systematically replacing
individual spare parts, one at a time. One purpose of the Description of functions and the
Diagrams in this Service Manual is to make it easier to trace faults.
When the fault is corrected, carry out a complete 'System Checkout' as described in the
'FLOW-i Anesthesia System – User's Manual'.
The troubleshooting guides below are focused only on technical problems. Information about
clinical related problems can be found in the 'FLOW-i Anesthesia System – User's Manual'.
For functionality enhancement, the latest released System SW version is always recommended.
The information in this chapter applies to System version 2.1 (or higher).
WARNING! With gas supply connected to the FLOW-i, there are pressurized
components inside the unit. All personnel must exercise extreme caution if fault
tracing or adjustments are performed with gas supply connected.
WARNING! With power supply connected to the FLOW-i, there are energized
electrical components inside the unit. All personnel must exercise extreme caution
if fault tracing or adjustments are performed with power supply connected and with
covers removed.
Possible causes to malfunction not mentioned in the following troubleshooting guides are:
The system has not been correctly assembled after cleaning, maintenance or service.
Disconnection or bad connection in cable connectors, PC board connectors, and
interconnection boards (PC 1900, PC 1909, etc).
Pinched cables. Defective cables may short-circuit power supply, CAN-bus signals, etc.
Disconnected or defective gas tubes, breathing gas as well as drive gas.
Note: Repeated disconnection/connection of tubes in quick couplings may cause a leakage.
These possible causes to malfunction must always be considered during troubleshooting.
66 76 980_02 6-2
6. TROUBLESHOOTING
System Checkout
The FLOW-i demands the user to start the automatic System Checkout at every startup of the
unit. The 'FLOW-i Anesthesia System – User's Manual' describes how to perform this System
Checkout.
The System Checkout-description on the following pages gives more detailed information about
the System Checkout. This information can be used e.g. during troubleshooting of the unit.
Some of the recommended actions described below refer to Service & Settings (S&S). The
Service key is required to access S&S. Troubleshooting can of course be performed without
access to S&S, but for some of the recommended actions, S&S will make troubleshooting faster
and easier.
Check if the fault remains after each performed service action. Re-run the full System Checkout
or run the concerned test using S&S.
It is also possible to select System Checkout via the Standby Menu; a Full check as well as
separate Leakage check and Vaporizer check. The Leakage check also measures circuit
compliance. These separate checks do not replace the complete System Checkout.
66 76 980_02 6-3
6. TROUBLESHOOTING
1. O2 flush test
Checks the O2 flush mechanism. The test is divided in three parts: Prior to pressing (1),
pressing (2) and releasing (3) the button. Refer to illustration above.
66 76 980_02 6-4
6. TROUBLESHOOTING
1. When the test is started, it is checked whether O2 supply pressure is present and that the
Patient Cassette is mounted. The O2 flow which is measured by the expiratory flow
transducer must not exceed 1 l/min prior to pressing the O2 Flush button. The test will fail if
the button is not pressed within 20 seconds.
2. The user is requested to press the O2 flush button for approx. 3 seconds. The test waits 1
second before reading the O2 flow. The flow must be within: 35 – 65 l/min. The test will fail
if the button is not released within 22 seconds.
3. When the user releases the O2 flush button after at least 2 seconds, the test again waits 1
second before reading the O2 flow. It must not exceed: 1 l/min.
Checks that the Insp. and Exp. valve discs are in place and are able to move.
The user is requested to open the cassette lid and press ‘Start check’.
The user is requested to observe and ensure the ceramic valve discs are moving.
The test is passed when the user presses ‘Yes’.
If they are not moving properly the user shall press ‘No’ and the test fails.
The user is requested to close the cassette lid and press ‘Continue’.
If the system detects that the cassette lid is open after the user pressed ‘Yes’, a dialog will
appear. The user gets the choice to close the lid, or to press ‘Bypass’ in case the system have
problems detecting the lid position.
Pressing ‘Bypass’ will not affect the test result, but generate an event in the log.
At start of this test, it is verified that the air inlet supply pressure is within range. The APL/PEEP
valve is open. The test is started by opening the Fresh Gas Module AIR. The Vaporizer bypass
valve is closed in order to build-up a pressure of approx. 110 cmH2O (measured by the Fresh
Gas Pressure Transducer). When the target pressure is reached, the bypass valve is opened.
66 76 980_02 6-5
6. TROUBLESHOOTING
This creates a pressure wave which will cause the inspiratory and expiratory valve to move.
Actual generated pressure will be shown in the test log. If the pressure reaches 140 cmH2O or
fails to reach 50 cmH2O, the test will be aborted.
The pressure wave is repeated for 90 s or until the test is manually deactivated.
3. Internal tests
3.1 Absorber switch test
Checks that the Absorber switch is in correct position, i.e. the absorber is docked.
66 76 980_02 6-6
6. TROUBLESHOOTING
4. Barometer test
Checks that the barometric pressure, measured by the internal barometer on the Monitoring
board, is within limits (670 - 1080 hPa):
Monitoring: Checks that the barometric pressure, read by Monitoring, is within limits.
Control: Checks that the barometric pressure, read by Control, is within limits.
The difference between the pressures read by Monitoring and Control must not be greater
than 26 hPa.
66 76 980_02 6-7
6. TROUBLESHOOTING
66 76 980_02 6-8
6. TROUBLESHOOTING
66 76 980_02 6-9
6. TROUBLESHOOTING
66 76 980_02 6 - 10
6. TROUBLESHOOTING
66 76 980_02 6 - 11
6. TROUBLESHOOTING
Flow test
If the N2O module is supplied with gas, the tests of the Fresh Gas Modules are tested two by
two. The Reflector Gas Module is tested individually.
Order: Air + N2O, O2 + N2O, Air + O2 and Reflector.
If the N2O module is not supplied with gas, the tests of the Fresh Gas Modules are first tested
individually and then two by two. The Reflector Gas module is tested individually.
Order: Air, O2, Air + O2 and Reflector.
The modules to be tested are set to deliver 30 l/min.
The Control and Monitoring system receives the measured values from all Gas Modules and
the Patient Cassette. The measured value from the modules must be within the following
limits in order for the test to pass:
Active modules: 29.4 – 30.6 l/min
Inactive modules: -0.096 – 0.096 l/min
Patient Cassette: 90 – 110 % of the sum of all Gas Modules flow as measured by the
same subsystem (CON and MON).
Note: When the Fresh Gas Module O2 is active, the value from Reflector Gas Module will be
ignored and when the Reflector Gas Module is active, the value from Fresh Gas Module O2 will
be ignored. This is due to that the modules are using the same gas supply.
Noise measurement
The amplitude of the sound pulses generated by the ultrasonic transducers/receivers in the
Expiratory Flow Transducer is measured at zero gas flow.
The amplitude of the received sound pulse is measured in an earlier timeframe than during
ordinary operation. The measured amplitude gives an indication of how much of the transmitted
sound pulse that travels through the solid cassette material instead of through the gas mixture in
the Expiratory Flow Transducer.
The values from both upstream and downstream directions are logged.
k factor
First the sum of the flow values from all of Gas Modules in the flow tests is calculated.
Secondly the sum of all the flow values from the Patient Cassette is calculated, where the flow
value from the gas reflector tests is absolute.
The k value is produced by dividing the Gas Module flow sum with the Patient Cassette flow
sum. The k value must be within 0.92 – 1.08.
The k value is tested by multiplying each Patient Cassette flow value measured by the
Monitoring system with the k value. If the sum of the flow values from the Gas Modules differ
more than 8% from the new Patient Cassette flow value in a subtest, the test will fail.
66 76 980_02 6 - 12
6. TROUBLESHOOTING
66 76 980_02 6 - 13
6. TROUBLESHOOTING
In case PGA reports occlusion, a dialog is displayed asking the user to check sampling line
and water trap. The dialog offers the possibility to redo the test.
4. O2 analyzers offset
Checks the Control O2 sensor @ 21%.
Measures room air by opening the zero valve (inside AION).
O2 offset is set to zero.
To stabilize the flow, room air is flushed for 10 s and then sampled 10 times over 2.5 s.
An O2 mean value is calculated. The mean value may not differ more than ±3% (abs) @ 21%
and the noise level must not be more than 6% (abs).
66 76 980_02 6 - 14
6. TROUBLESHOOTING
The CGA outlet valve is then switched back to return the sampled gas into the fresh gas
channel. The average flow must be at least 150 ml/min.
The flow in the inspiratory return channel should be less than in the fresh gas return channel
due to the higher resistance in the inspiratory return channel.
The PGA average flow is measured at the same time and must be at least 80 ml/min.
66 76 980_02 6 - 15
6. TROUBLESHOOTING
Battery power path test. Checks that the power supply switches to battery, when the input
voltage is decreased. The test fails if no battery current is detected.
Internal resistance test. Checks the condition of the battery by measuring that the internal
resistance. The test fails if the resistance is too high.
66 76 980_02 6 - 16
6. TROUBLESHOOTING
Checks that the vaporizer reaches a pressure of 1000 - 1400 hPa within 10 s.
Delivers a fresh gas flow (air) of 12 l/min and reflector flow of 13 l/min.
Opens the Vaporizer Safety Valve.
Requests an agent concentration of 4%.
Checks that a necessary temperature is reached and the measured concentration is 3 - 5%
using the CGA.
Stops the agent delivery and evacuates the vaporizer.
Checks that the vaporizer reaches a pressure below 200 hPa (700 hPa for Desflurane) within
10 s.
Closes the vaporizer Safety valve and pressurizes the vaporizer.
Checks that the vaporizer reaches a pressure of 1000 - 1400 hPa within 10 s.
Request an agent concentration of 4% (12% for Desflurane).
Check that the measured concentration is below 0.5% (2% for Desflurane) using the CGA.
Opens the Vaporizer Safety valve and requests an agent concentration of 4%.
Checks that the measured concentration is 3 - 5% using the CGA.
The agent delivery is stopped and a gas flow is delivered to flush the system.
The CGA pump is stopped and the vaporizer is unlocked.
If the test is passed, time and date for the test will be stored in the Vaporizer.
66 76 980_02 6 - 17
6. TROUBLESHOOTING
The recommended actions for the Control pressure transducer test that failed with CFI 3 are:
With Monitoring pressure transducer test (CFI 15):
1. Replace Inspiratory Pressure Transducer.
Without Monitoring pressure transducer test (CFI 15):
66 76 980_02 6 - 18
6. TROUBLESHOOTING
The recommended actions for the Monitoring pressure transducer test that failed with CFI 15
are:
With Control pressure transducer test (CFI 3):
1. Replace Inspiratory Pressure Transducer.
Without Control pressure transducer test (CFI 3):
1. Replace PC 1921 Monitoring.
The recommended action for this combination of Check Failures is to replace the Inspiratory
Pressure Transducer.
2011-09-28 15:26:14 SFIN Pressure transducer Result Failed
2011-09-28 15:26:14 FIN APL/PEEP valve test Result Passed
2011-09-28 15:25:57 STA APL/PEEP valve test
2011-09-28 15:25:56 FIN Monitoring pressure transducer test Result Failed (15)
2011-09-28 15:25:54 CHK (15) Inspiratory Pressure Transducer offset out of range.
2011-09-28 15:25:51 STA Monitoring pressure transducer test
2011-09-28 15:25:50 FIN Control pressure transducer test Result Failed (3)
2011-09-28 15:25:50 CHK (3) CON Inspiratory Pressure Transducer offset out of range.
2011-09-28 15:25:48 STA Control pressure transducer test
2011-09-28 15:25:46 SSTA Pressure transducer
The recommended actions for the Monitoring flow transducer test that failed with CFI 62 are:
With CFI 63:
1. CFI indicating patient cassette error may be due to other failures in the system, e.g. gas
delivery from the gas modules. In case of multiple CFI, where patient cassette error is indicated,
start troubleshooting the other indicated parts.
2. Replace Patient Cassette.
Without CFI 63:
1. Replace PC 1920 Control.
The recommended actions for the Monitoring flow transducer test that failed with CFI 63 are:
With CFI 62:
1. CFI indicating patient cassette error may be due to other failures in the system, e.g. gas
delivery from the gas modules. In case of multiple CFI, where patient cassette error is indicated,
start troubleshooting the other indicated parts.
2. Replace Patient Cassette.
Without CFI 62:
1. Replace PC 1921 Monitoring.
66 76 980_02 6 - 19
6. TROUBLESHOOTING
The recommended action for this combination of Check Failures is to replace the Patient
Cassette.
2011-09-21 12:55:33 SFIN Flow transducer Result Failed
2011-09-21 12:55:33 FIN Monitoring flow transducer test Result Failed (63, 62)
2011-09-21 12:55:33 DLG "Flow out of range. Check that the breathing circuit is correctly
mounted and connected to the test plug." [Bypass test, retry 1]
2011-09-21 12:54:35 CHK (62) CON Expiratory Flow Transducer flow out of range during O2
stage.
2011-09-21 12:54:35 CHK (63) MON Expiratory Flow Transducer flow out of range during O2
stage.
2011-09-21 12:54:14 DLG "Flow out of range. Check that the breathing circuit is correctly
mounted and connected to the test plug." [Redo test, retry 1]
2011-09-21 12:53:38 CHK (62) CON Expiratory Flow Transducer flow out of range during O2
stage.
2011-09-21 12:53:38 CHK (63) MON Expiratory Flow Transducer flow out of range during O2
stage.
2011-09-21 12:53:17 STA Monitoring flow transducer test
2011-09-21 12:53:16 FIN Control flow transducer test Result Passed
2011-09-21 12:53:02 STA Control flow transducer test
2011-09-21 12:53:01 SSTA Flow transducer
66 76 980_02 6 - 20
6. TROUBLESHOOTING
66 76 980_02
1. O2 flush test
2 O2 Flush Button before depression flow outside interval 0 to 1 l/min without 1. Check/replace O2 Flush Valve.
button depression indication from electrical switch.
2. Check for leakage into the system, e.g. fresh gas modules, emergency
ventilation.
3 O2 Flush Button during depression flow < 35 l/min. 1. Check for leakage into the system, e.g. fresh gas modules, emergency
ventilation.
2. Check internal Instant O2 tubing.
3. Instant O2 supply pressure from REG4 is too low. Check/replace REG4.
4 O2 Flush Button during depression flow > 65 l/min. 1. If the O2 supply pressure is within range, check/replace REG4.
5 Central Gas Supply O2 pressure outside interval of 2500 hPa to 6500 hPa. 1. Check central O2 gas supply.
2. Replace PC 1906 Gas Block.
6 Button depression detected too early, which makes it impossible to properly 1. Check/replace O2 Flush Valve.
measure and check O2 Flush Button before depression flow.
7 Additional button depression detected after release was indicated, which 1. Redo test and press the O2 Flush Button firmly once for at least 2 seconds.
makes it impossible to properly measure and check O2 Flush Button after
release flow.
8 O2 Flush Button after release flow outside interval -1 to 1 l/min. 1. Check/replace O2 Flush Valve.
6 - 21
6. TROUBLESHOOTING
66 76 980_02
10 No docked Patient Cassette detected. The lid (Patient Cassette Cover) may be 1. Check/replace the Patient Cassette.
open.
1 Pressurization cycle peak pressure > 140 cmH2O. 1. Check/replace the Vaporizer Bypass Valve, PV5.
2. Check/replace the Fresh Gas Pressure Transducer.
3. Check/replace the Fresh Gas Module Air.
2 Pressurization cycle peak pressure < 50 cmH2O. 1. Check/replace the Vaporizer Bypass Valve, PV5.
2. Check/replace the Fresh Gas Pressure Transducer.
3. Check/replace the Fresh Gas Module Air.
3 The user has pressed "No" in dialog asking if inspiratory and expiratory valves 1. Check the Insp. Valve and Exp. Valve in the Patient Cassette. Clean /
are moving properly. replace if necessary.
4 Central Gas Supply AIR pressure outside interval of 2500 hPa to 6500 hPa. 1. Check central AIR gas supply.
2. Replace PC 1906 Gas Block.
6 - 22
6. TROUBLESHOOTING
66 76 980_02
3. Internal tests
1 The Absorber Switch indicates that no CO2 Absorber is docked. 1. Check that the Absorber is mounted correctly and the switch is in upright
position.
2. Check/replace the Absorber Sensor.
1 No Water Trap is mounted according to electrical switches in Water Trap 1. Insert Water Trap.
Receptacle.
2. Check cable to the Water Trap Receptacle.
3. Replace Water Trap Receptacle including cable.
2 A Neonate Water Trap is mounted according to electrical switches in Water 1. It is recommended to replace the Infant Water Trap with an Adult Water
Trap Receptacle. Trap.
1 A vaporizer is docked in slot 1 and Disable VAP 1 feedback signal is active at 1. Check for information in the Technical alarms log. For recommended
the beginning of the test. actions, refer to Troubleshooting, section Technical error codes and messages.
2. Move Vaporizer 1 to slot 2. If the corresponding failure occurs also in slot 2;
Vaporizer service.
3. Replace PC 1920 Control.
4. Replace PC 1921 Monitoring.
2 A vaporizer is docked in slot 2 and Disable VAP 2 feedback signal is active at 1. Check for information in the Technical alarms log. For recommended
the beginning of the test. actions, refer to Troubleshooting, section Technical error codes and messages.
2. Move Vaporizer 2 to slot 1. If the corresponding failure occurs also in slot 1;
Vaporizer service.
3. Replace PC 1920 Control.
4. Replace PC 1921 Monitoring.
3 No vaporizer is docked in slot 1 and Disable VAP 1 feedback signal is inactive 1. Replace PC 1920 Control.
at the beginning of the test.
4 No vaporizer is docked in slot 2 and Disable VAP 2 feedback signal is inactive 1. Replace PC 1920 Control.
at the beginning of the test.
5 VAP 1 power check #1 voltage out of range. 1. If several Check failures appear in this test; perform recommended actions
for these failures.
2. Replace PC 1923 Power Control.
6 VAP 2 power check #1 voltage out of range. 1. If several Check failures appear in this test; perform recommended actions
for these failures.
2. Replace PC 1923 Power Control.
6 - 24
6. TROUBLESHOOTING
66 76 980_02
8 A vaporizer is docked in slot 2, which is disabled and Disable VAP2 feedback 1. Replace PC 1920 Control.
signal is inactive.
9 No vaporizer is docked in slot 1, which is enabled and Disable VAP 1 feedback 1. Replace PC 1920 Control.
is active.
2. Replace PC 1921 Monitoring.
10 No vaporizer is docked in slot 2, which is enabled and Disable VAP 2 feedback 1. Replace PC 1920 Control.
is active.
2. Replace PC 1921 Monitoring.
11 VAP 1 power check #2 voltage out of range. 1. If several Check failures appear in this test; perform recommended actions
for these failures.
2. Replace PC 1923 Power Control.
12 VAP 2 power check #2 voltage out of range. 1. If several Check failures appear in this test; perform recommended actions
for these failures.
2. Replace PC 1923 Power Control.
13 No vaporizer is docked in slot 1, which is disabled and Disable VAP 1 feedback 1. Replace PC 1920 Control.
signal is inactive.
14 No vaporizer is docked in slot 2, which is disabled and Disable VAP 2 feedback 1. Replace PC 1920 Control.
signal is inactive.
15 VAP 1 power check #3 voltage out of range. 1. If several Check failures appear in this test; perform recommended actions
for these failures.
6 - 25
6. TROUBLESHOOTING
66 76 980_02
16 VAP 2 power check #3 voltage out of range. 1. If several Check failures appear in this test; perform recommended actions
for these failures.
2. Replace PC 1923 Power Control.
17 VAP 1 disabled at test start. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
2. Move Vaporizer 1 to slot 2. If the corresponding failure occurs also in slot 2;
Vaporizer service.
18 VAP 2 disabled at test start. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
2. Move Vaporizer 2 to slot 1. If the corresponding failure occurs also in slot 1;
Vaporizer service.
1 A vaporizer is docked in slot 1 and Disable VAP 1 feedback signal is active at 1. Check for information in the Technical alarms log. For recommended
the beginning of the test. actions, refer to Troubleshooting, section Technical error codes and messages.
2. Move Vaporizer 1 to slot 2. If the corresponding failure occurs also in slot 2;
Vaporizer service.
3. Replace PC 1921 Monitoring.
4. Replace PC 1920 Control.
2 No vaporizer is docked in slot 1 and Disable VAP 1 feedback signal is inactive 1. Replace PC 1921 Monitoring.
at the beginning of the test.
3 A vaporizer is docked in slot 2 and Disable VAP 2 feedback signal is active at 1. Check for information in the Technical alarms log. For recommended
the beginning of the test. actions, refer to Troubleshooting, section Technical error codes and messages.
6 - 26
6. TROUBLESHOOTING
66 76 980_02
4 No vaporizer is docked in slot 2 and Disable VAP 2 feedback signal is inactive 1. Replace PC 1921 Monitoring.
at the beginning of the test.
5 Disable VAP 1 feedback signal inactive after activation of signal Disable VAP 1. Replace PC 1921 Monitoring.
1.
6 Disable VAP 2 feedback signal inactive after activation of signal Disable VAP 1. Replace PC 1921 Monitoring.
2.
7 Disable VAP 1 feedback signal active after deactivation of signal Disable VAP 1. Replace PC 1921 Monitoring.
1 (VAP 1 docked).
2. Replace PC 1920 Control.
8 Disable VAP 1 feedback signal inactive after activation of signal Disable VAP 1 1. Replace PC 1921 Monitoring.
(Slot 1 empty).
9 Disable VAP 2 feedback signal active after deactivation of signal Disable VAP 1. Replace PC 1921 Monitoring.
2 (VAP 2 docked).
2. Replace PC 1920 Control.
10 Disable VAP 2 feedback signal inactive after activation of signal Disable VAP 2 1. Replace PC 1921 Monitoring.
(Slot 2 empty).
6 - 27
6. TROUBLESHOOTING
66 76 980_02
1 Average sound level < 800, i.e. estimated sound level at a distance of 1 meter 1. Make sure that the holes for the loudspeaker are not covered on the Control
is below 60 dB. panel.
2. Check/replace loudspeaker.
3. Replace PC 1924 Display connections.
2 Audio test feedback signal inactive after buzzer activation. 1. Replace PC 1921 Monitoring
3 Audio test feedback signal active after buzzer deactivation. 1. Replace PC 1921 Monitoring
4 Lower limit activation failed for 5.0 V window. 1. Replace PC 1921 Monitoring
5 Lower limit deactivation failed for 5.0 V window. 1. Replace PC 1921 Monitoring
6 Upper limit activation failed for 5.0 V window. 1. Replace PC 1921 Monitoring
7 Upper limit deactivation failed for 5.0 V window. 1. Replace PC 1921 Monitoring
8 Lower limit activation failed for 3.3 V window. 1. Replace PC 1921 Monitoring
9 Lower limit deactivation failed for 3.3 V window. 1. Replace PC 1921 Monitoring
10 Upper limit activation failed for 3.3 V window. 1. Replace PC 1921 Monitoring
6 - 28
6. TROUBLESHOOTING
66 76 980_02
11 Upper limit deactivation failed for 3.3 V window. 1. Replace PC 1921 Monitoring
12 Lower limit activation failed for 2.5 V window. 1. Replace PC 1921 Monitoring
13 Lower limit deactivation failed for 2.5 V window. 1. Replace PC 1921 Monitoring
14 Upper limit activation failed for 2.5 V window. 1. Replace PC 1921 Monitoring
15 Upper limit deactivation failed for 2.5 V window. 1. Replace PC 1921 Monitoring
16 Lower limit activation failed for 1.5 V window. 1. Replace PC 1921 Monitoring
17 Lower limit deactivation failed for 1.5 V window. 1. Replace PC 1921 Monitoring
18 Upper limit activation failed for 1.5 V window. 1. Replace PC 1921 Monitoring
19 Upper limit deactivation failed for 1.5 V window. 1. Replace PC 1921 Monitoring
3 Removal of docked VAP detected during Initial stage of Vaporizer initial test. 1. Handling error.
4 Removal of docked VAP detected during Evacuation stage of Vaporizer initial 1. Handling error.
test.
5 Removal of docked VAP detected during Internal VAP test stage of Vaporizer 1. Handling error.
initial test.
6 Removal of docked VAP detected during Pressurization stage of Vaporizer 1. Handling error.
initial test.
7 VAP disabled during Initial stage of Vaporizer initial test. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
8 VAP disabled during Evacuation stage of Vaporizer initial test. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
9 VAP disabled during Internal VAP test stage of Vaporizer initial test. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
6 - 30
6. TROUBLESHOOTING
66 76 980_02
11 Open VAP Lid detected during Initial stage of Vaporizer initial test. 1. Handling error.
Notice: This check failure results in test result "Not performed".
2. Clean surface that reflects the PC 1931 Vaporizer Lid Sensor.
3. Vaporizer service.
12 Open VAP Lid detected during Initial stage of Vaporizer initial test. 1. Handling error.
2. Clean surface that reflects the PC 1931 Vaporizer Lid Sensor.
3. Vaporizer service.
13 Open VAP Lid detected during Evacuation stage of Vaporizer initial test. 1. Handling error.
Notice: This check failure results in test result "Not performed".
2. Clean surface that reflects the PC 1931 Vaporizer Lid Sensor.
3. Vaporizer service.
14 Open VAP Lid detected during Internal VAP test stage of Vaporizer initial test. 1. Handling error.
Notice: This check failure results in test result "Not performed".
2. Clean surface that reflects the PC 1931 Vaporizer Lid Sensor.
3. Vaporizer service.
.
15 Open VAP Lid detected during Pressurization stage of Vaporizer initial test 1. Handling error.
Notice: This check failure results in test result "Not performed".
2. Clean surface that reflects the PC 1931 Vaporizer Lid Sensor.
3. Vaporizer service.
16 Notice: This check failure results in test result "Not performed". 1. Handling error.
6 - 31
6. TROUBLESHOOTING
66 76 980_02
17 Notice: This check failure results in test result "Not performed". 1. Handling error.
2. Vaporizer service.
18 Vaporizer not initialized after 15 s wait. 1. Move Vaporizer 1 to slot 2. If the corresponding failure occurs also in slot 2;
Vaporizer service.
2. Check/replace PC 1909 Vaporizer Adapter.
3. Replace PC 1920 Control.
19 VAP pressure not consecutively below 700 hPa (Desflurane) or 200 hPa (other 1. Move Vaporizer 1 to slot 2. If the corresponding failure occurs also in slot 2;
anesthetic agents) during 5 s within a total timeframe of 10 s from evacuation Vaporizer service.
start.
2. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 2; replace Scavenging valve.
3. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 2; replace Pressure valve.
4. Replace PC 1907 Valve Drivers.
20 VAP pressure above 700 hPa detected during Internal VAP test stage. 1. Move Vaporizer 1 to slot 2. If the corresponding failure occurs also in slot 2;
Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 2; replace Pressure valve.
3. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 2; replace Scavenging valve.
4. Replace PC 1907 Valve Drivers.
21 Internal VAP test not finished within 5 s. 1. Move Vaporizer 1 to slot 2. If the corresponding failure occurs also in slot 2;
Vaporizer service.
2. Check/replace PC 1909 Vaporizer Adapter.
6 - 32
6. TROUBLESHOOTING
66 76 980_02
22 Internal VAP test failed during Vaporizer initial test. 1. Vaporizer service.
23 VAP pressure not consecutively within interval 1000 hPa to 1400 hPa during 1. Move Vaporizer 1 to slot 2. If the corresponding failure occurs also in slot 2;
5s within a total timeframe of 15 s from pressurization start. Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 2; replace Scavenging valve.
3. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 2; replace Pressure valve.
4. Replace PC 1907 Valve Drivers.
3 Removal of docked VAP detected during Initial stage of Vaporizer initial test. 1. Handling error.
4 Removal of docked VAP detected during Evacuation stage of Vaporizer initial 1. Handling error.
test.
5 Removal of docked VAP detected during Internal VAP test stage of Vaporizer 1. Handling error.
initial test.
6 - 33
6. TROUBLESHOOTING
66 76 980_02
7 VAP disabled during Initial stage of Vaporizer initial test. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
8 VAP disabled during Evacuation stage of Vaporizer initial test. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
9 VAP disabled during Internal VAP test stage of Vaporizer initial test. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
10 VAP disabled during Pressurization stage of Vaporizer initial test. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
11 Open VAP Lid detected during Initial stage of Vaporizer initial test. 1. Handling error.
Notice: This check failure results in test result "Not performed".
2. Clean surface that reflects the PC 1931 Vaporizer Lid Sensor.
3. Vaporizer service.
12 Open VAP Lid detected during Initial stage of Vaporizer initial test. 1. Handling error.
2. Clean surface that reflects the PC 1931 Vaporizer Lid Sensor.
3. Vaporizer service.
13 Open VAP Lid detected during Evacuation stage of Vaporizer initial test. 1. Handling error.
Notice: This check failure results in test result "Not performed".
2. Clean surface that reflects the PC 1931 Vaporizer Lid Sensor.
3. Vaporizer service.
6 - 34
6. TROUBLESHOOTING
66 76 980_02
15 Open VAP Lid detected during Pressurization stage of Vaporizer initial test. 1. Handling error.
Notice: This check failure results in test result "Not performed".
2. Clean surface that reflects the PC 1931 Vaporizer Lid Sensor.
3. Vaporizer service.
16 Notice: This check failure results in test result "Not performed". 1. Handling error.
2. Vaporizer service.
17 Notice: This check failure results in test result "Not performed". 1. Handling error.
2. Vaporizer service.
18 Vaporizer not initialized after 15 s wait. 1. Move Vaporizer 2 to slot 1. If the corresponding failure occurs also in slot 1;
Vaporizer service.
2. Check/replace PC 1909 Vaporizer Adapter.
3. Replace PC 1920 Control.
19 VAP pressure not consecutively below 700 hPa (Desflurane) or 200 hPa (other 1. Move Vaporizer 2 to slot 1. If the corresponding failure occurs also in slot 1;
anesthetic agents) during 5 s within a total timeframe of 10 s from evacuation Vaporizer service.
start.
2. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 1; replace Scavenging valve.
3. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 1; replace Pressure valve.
4. Replace PC 1907 Valve Drivers.
6 - 35
6. TROUBLESHOOTING
66 76 980_02
20 VAP pressure above 700 hPa detected during Internal VAP test stage. 1. Move Vaporizer 2 to slot 1. If the corresponding failure occurs also in slot 1;
Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 1; replace Pressure valve.
3. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 1; replace Scavenging valve.
4. Replace PC 1907 Valve Drivers.
21 Internal VAP test not finished within 5 s. 1. Move Vaporizer 2 to slot 1. If the corresponding failure occurs also in slot 1;
Vaporizer service.
2. Check/replace PC 1909 Vaporizer Adapter.
3. Replace PC 1920 Control.
22 Internal VAP test failed during Vaporizer initial test. 1. Vaporizer service.
23 VAP pressure not consecutively within interval 1000 hPa to 1400 hPa during 1. Move Vaporizer 2 to slot 1. If the corresponding failure occurs also in slot 1;
5s within a total timeframe of 15 s from pressurization start. Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 1; replace Scavenging valve.
3. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 1; replace Pressure valve.
4. Replace PC 1907 Valve Drivers.
1 Central Gas Supply N20 Pressure ≤ 1000 hPa AND N20 enabled in N2O is connected:
configuration.
1. Check central N2O gas supply.
2. Replace PC 1906 Gas Block.
6 - 36
6. TROUBLESHOOTING
66 76 980_02
2 Central Gas Supply N20 Pressure > 1000 hPa AND N20 disabled in N2O is connected:
configuration.
1. Enable N2O in configuration
N2O is not connected:
2. Replace PC 1906 Gas Block.
1 Central Gas Supply AIR pressure outside the interval 2500 hPa to 6500 hPa. 1. Check central gas supply.
2. Replace PC 1906 Gas Block.
3. Replace PC 1921 Monitoring.
2 Patient Cassette position stage #1, Expiratory flow outside interval 500 to 2500 1. Check gas supply.
ml/min
2. Check/replace Patient Cassette.
3. Check/replace connections to AFGO valve.
4. Check/replace AFGO valve, PV7 incl. EMV18.
5. Check/replace PC 1907 Valve drivers.
6. Check/replace PC 1920 Control.
3 AFGO position stage, Expiratory flow ≥ 500 ml/min. 1. Check/replace AFGO Valve, PV7 incl. EMV18.
2. Check/replace PC 1907 Valve drivers.
3. Check/replace PC 1920 Control.
6 - 37
6. TROUBLESHOOTING
66 76 980_02
5 Patient cassette position stage #1, Fresh Gas pressure > 30 cmH2O. 1. Check that air can flow thru the AFGO plug or the AFGO connection.
2. Check Vaporizer bypass valve, PV5.
3. Check/replace PC 1781 Fresh gas pressure transducer.
4. Check/replace PC 1919 Exp. Channel.
5. Check/replace PC 1920 Control.
6 AFGO position stage, Fresh Gas pressure > 30 cmH2O. 1. Check that air can flow thru the AFGO plug or the AFGO connection.
2. Check Vaporizer bypass valve, PV5.
3. Check/replace PC 1781 Fresh gas pressure transducer.
4. Check/replace PC 1919 Exp. Channel.
5. Check/replace PC 1920 Control.
7 Patient Cassette position stage #2, Fresh Gas pressure > 30 cmH2O. 1. Check that air can flow thru the AFGO plug or the AFGO connection.
2. Check Vaporizer bypass valve, PV5.
3. Check/replace PC 1781 Fresh gas pressure transducer.
4. Check/replace PC 1919 Exp. Channel.
5. Check/replace PC 1920 Control.
8 Option for AFGO not installed. When the option is not installed, the test N/A
finishes with Passed result.
6 - 38
6. TROUBLESHOOTING
66 76 980_02
4. Barometer test
1 Central Gas Supply AIR pressure < 2500 hPa. Central gas supply error. Check central gas AIR pressure with a manometer. If
the manometer pressure is 2500-6500 hPa, but the gas supply pressure
reading on FLOW-i is outside this interval:
1. Replace PC 1906 Gas Block.
2. Replace PC 1921 Monitoring.
2 Central Gas Supply AIR pressure > 6500 hPa. Central gas supply error. Check central gas AIR pressure with a manometer. If
the manometer pressure is 2500-6500 hPa, but the gas supply pressure
reading on FLOW-i is outside this interval:
1. Replace PC 1906 Gas Block.
2. Replace PC 1921 Monitoring.
3 Central Gas Supply O2 pressure < 2500 hPa. Central gas supply error. Check central gas O2 pressure with a manometer. If
the manometer pressure is 2500-6500 hPa, but the gas supply pressure
reading on FLOW-i is outside this interval:
1. Replace PC 1906 Gas Block.
2. Replace PC 1921 Monitoring.
4 Central Gas Supply O2 pressure > 6500 hPa. Central gas supply error. Check central gas O2 pressure with a manometer. If
the manometer pressure is 2500-6500 hPa, but the gas supply pressure
reading on FLOW-i is outside this interval:
1. Replace PC 1906 Gas Block.
2. Replace PC 1921 Monitoring.
5 Central Gas Supply N2O pressure < 2500 hPa. Central gas supply error. Check central gas N2O pressure with a manometer.
If the manometer pressure is 2500-6500 hPa, but the gas supply pressure
6 - 40
6 Central Gas Supply N2O pressure > 4500 hPa. Central gas supply error. Check central gas N2O pressure with a manometer.
If the manometer pressure is 2500-6500 hPa, but the gas supply pressure
reading on FLOW-i is outside this interval:
1. Replace PC 1906 Gas Block.
2. Replace PC 1921.
3. Replace REG6.
7 Absolute difference between Central Gas Supply AIR pressure and MON 1. Unstable central gas AIR supply may pressure may trigger this error. Check
Fresh Gas Module AIR pressure > 10% * MIN (Central Gas Supply AIR central gas AIR supply with a manometer, make sure that the pressure is
pressure, MON Fresh Gas Module AIR pressure). stable and re-run the test.
Notice: Check is only performed when (AIR Cylinder Gas pressure < 1000 hPa
AND Central Gas Supply AIR pressure > 1000 hPa) OR Central Gas Supply
AIR pressure > 3800 hPa.
With CFI 8:
2. Replace PC 1906 Gas Block.
3. Replace Fresh Gas Module AIR.
Without CFI 8:
2. Replace PC 1921 Monitoring.
3. Replace PC 1920 Control.
8 Absolute difference between Central Gas Supply AIR pressure and CON Fresh 1. Unstable central gas AIR supply may pressure may trigger this error. Check
Gas Module AIR pressure > 10% * MIN (Central Gas Supply AIR pressure, central gas AIR supply with a manometer, make sure that the pressure is
MON Fresh Gas Module AIR pressure). stable and re-run the test.
Notice: Check is only performed when (AIR Cylinder Gas pressure < 1000 hPa
AND Central Gas Supply AIR pressure > 1000 hPa) OR Central Gas Supply
AIR pressure > 3800 hPa.
6 - 41
6. TROUBLESHOOTING
66 76 980_02
9 Absolute difference between Central Gas Supply O2 pressure and MON Fresh 1. Unstable central gas O2 supply may pressure may trigger this error. Check
Gas Module O2 pressure > 10% * MIN (Central Gas Supply O2 pressure, central gas O2 supply with a manometer, make sure that the pressure is stable
MON Fresh Gas Module O2 pressure). and re-run the test.
Notice: Check is only performed when (O2 Cylinder Gas pressure < 1000 hPa
AND Central Gas Supply O2 pressure > 1000 hPa) OR Central Gas Supply O2
pressure > 3800 hPa.
With CFI 10, CFI 13 and CFI 14:
2. Replace PC 1906 Gas Block.
With CFI 10 and without CFI 13 and CFI 14:
2. Replace Fresh Gas Module O2.
Without CFI 10:
2. Replace PC 1921 Monitoring.
3. Replace PC 1920 Control.
10 Absolute difference between Central Gas Supply O2 pressure and CON Fresh 1. Unstable central gas O2 supply may pressure may trigger this error. Check
Gas Module O2 pressure > 10% * MIN (Central Gas Supply O2 pressure, central gas O2 supply with a manometer, make sure that the pressure is stable
MON Fresh Gas Module O2 pressure). and re-run the test.
Notice: Check is only performed when (O2 Cylinder Gas pressure < 1000 hPa
AND Central Gas Supply O2 pressure > 1000 hPa) OR Central Gas Supply O2
pressure > 3800 hPa.
With CFI 9, CFI 13 and CFI 14:
6 - 42
6. TROUBLESHOOTING
66 76 980_02
11 Absolute difference between Central Gas Supply N2O pressure and MON 1. Unstable central gas N2O supply may pressure may trigger this error. Check
Fresh Gas Module N2O pressure > 10% * MIN (Central Gas Supply N2O central gas N2O supply with a manometer, make sure that the pressure is
pressure, MON Fresh Gas Module N2O pressure). stable and re-run the test.
Notice: Check is only performed when (N2O Cylinder Gas pressure < 1000
hPa AND Central Gas Supply N2O pressure > 1000 hPa) OR Central Gas
Supply N2O pressure > 3800 hPa.
With CFI 12:
2. Replace PC 1906 Gas Block.
3. Replace Fresh Gas Module N2O.
Without CFI 12:
2. Replace PC 1921 Monitoring.
3. Replace PC 1920 Control.
12 Absolute difference between Central Gas Supply N2O pressure and CON 1. Unstable central gas N2O supply may pressure may trigger this error. Check
Fresh Gas Module N2O pressure > 10% * MIN (Central Gas Supply N2O central gas N2O supply with a manometer, make sure that the pressure is
pressure, MON Fresh Gas Module N2O pressure). stable and re-run the test.
Notice: Check is only performed when (N2O Cylinder Gas pressure < 1000
hPa AND Central Gas Supply N2O pressure > 1000 hPa) OR Central Gas
Supply N2O pressure > 3800 hPa.
With CFI 11:
2. Replace PC 1906 Gas Block.
3. Replace Fresh Gas Module N2O.
Without CFI 11:
6 - 43
6. TROUBLESHOOTING
66 76 980_02
13 Absolute difference between Central Gas Supply O2 pressure and MON 1. Unstable central gas O2 supply may pressure may trigger this error. Check
Reflector Gas Module O2 pressure > 10% * MIN (Central Gas Supply O2 central gas O2 supply with a manometer, make sure that the pressure is stable
pressure, MON Reflector Gas Module O2 pressure). and re-run the test.
Notice: Check is only performed when (O2 Cylinder Gas pressure < 1000 hPa
AND Central Gas Supply O2 pressure > 1000 hPa) OR Central Gas Supply O2
pressure > 3800 hPa.
With CFI 9, CFI 10 and CFI 14:
2. Replace PC 1906 Gas Block.
Without CFI 9 and CFI 10 and with CFI 14:
2. Replace Reflector Gas Module O2.
Without CFI 14:
2. Replace PC 1921 Monitoring.
3. Replace PC 1920 Control.
14 Absolute difference between Central Gas Supply O2 pressure and CON 1. Unstable central gas O2 supply may pressure may trigger this error. Check
Reflector Gas Module O2 pressure > 10% * MIN (Central Gas Supply O2 central gas O2 supply with a manometer, make sure that the pressure is stable
pressure, MON Reflector Gas Module O2 pressure). and re-run the test.
Notice: Check is only performed when (O2 Cylinder Gas pressure < 1000 hPa
AND Central Gas Supply O2 pressure > 1000 hPa) OR Central Gas Supply O2
pressure > 3800 hPa.
With CFI 9, CFI 10 and CFI 13:
2. Replace PC 1906 Gas Block.
Without CFI 9 and CFI 10 and with CFI 13:
2. Replace Reflector Gas Module O2.
Without CFI 13:
2. Replace PC 1920 Control.
3. Replace PC 1921 Monitoring.
6 - 44
6. TROUBLESHOOTING
66 76 980_02
15 CON/MON Fresh Gas Module O2 pressure difference > 150 hPa. 1. Replace PC 1921 Monitoring.
2. Replace PC 1920 Control.
16 CON/MON Fresh Gas Module AIR pressure difference > 150 hPa. 1. Replace PC 1921 Monitoring.
2. Replace PC 1920 Control.
17 CON/MON Fresh Gas Module N2O pressure difference > 150 hPa. 1. Replace PC 1921 Monitoring.
2. Replace PC 1920 Control.
18 CON/MON Reflector Gas Module pressure difference > 150 hPa. 1. Replace PC 1921 Monitoring.
2. Replace PC 1920 Control.
19 AIR Cylinder Gas pressure < 40000 hPa. Check cylinder gas supply AIR pressure with a manometer. If the manometer
Notice: This check failure does not affect the test result. pressure is 40000–200000 hPa, but the gas supply AIR pressure reading on
the FLOW-i is outside this interval:
1. Replace PC 1916 Cylinder Pressure Connection.
2. Replace PC 1921 Monitoring.
20 AIR Cylinder Gas pressure > 200000 hPa. Check cylinder gas supply AIR pressure with a manometer. If the manometer
Notice: This check failure does not affect the test result. pressure is 40000–200000 hPa, but the gas supply AIR pressure reading on
the FLOW-i is outside this interval:
21 O2 Cylinder Gas pressure < 40000 hPa. Check cylinder gas supply O2 pressure with a manometer. If the manometer
Notice: This check failure does not affect the test result. pressure is 40000–200000 hPa, but the gas supply O2 pressure reading on
the FLOW-i is outside this interval:
6 - 45
6. TROUBLESHOOTING
66 76 980_02
22 O2 Cylinder Gas pressure > 200000 hPa. Check cylinder gas supply O2 pressure with a manometer. If the manometer
Notice: This check failure does not affect the test result. pressure is 40000-200000hPa, but the gas supply O2 pressure reading on the
FLOW-i is outside this interval:
1. Replace PC 1916 Cylinder Pressure Connection.
2. Replace PC 1921 Monitoring.
23 N2O Cylinder Gas pressure < 40000 hPa. Check cylinder gas supply N2O pressure with a manometer. If the manometer
Notice: This check failure does not affect the test result. pressure is 40000-200000hPa, but the gas supply N2O pressure reading on
the FLOW-i is outside this interval:
1. Replace PC 1916 Cylinder Pressure Connection.
2. Replace PC 1921 Monitoring.
24 N2O Cylinder Gas pressure > 200000 hPa. Check cylinder gas supply N2O pressure with a manometer. If the manometer
Notice: This check failure does not affect the test result. pressure is 40000-200000hPa, but the gas supply N2O pressure reading on
the FLOW-i is outside this interval:
1. Replace PC 1916 Cylinder Pressure Connection.
2. Replace PC 1921 Monitoring.
25 AIR Backup Gas Supply not detected Disconnect the Backup Gas Trolley cable.
AND
AIR Cylinder Gas pressure > 1000 hPa.
1. If the fault condition does not recur, replace PC 1916 Cylinder Pressure
Connection.
2. If the fault condition remains, replace PC 1903 External Connectors.
6 - 46
6. TROUBLESHOOTING
66 76 980_02
27 N2O Backup Gas Supply not detected Disconnect the Backup Gas Trolley cable.
AND
N2O Cylinder Gas pressure > 1000 hPa.
1. If the fault condition does not recur, replace PC 1916 Cylinder Pressure
Connection.
2. If the fault condition remains, replace PC 1903 External Connectors.
28 System shall not be used with N2O according to N2O gas supply test N2O is connected, but the use of N2O is disabled in the Startup configuration.
AND
Central Gas Supply N2O pressure > 1000 hPa.
1. Check Startup configuration.
29 System shall not be used with N2O according to N2O gas supply test N2O is connected, but the use of N2O is disabled in the Startup configuration.
AND
N2O Cylinder Gas pressure > 1000 hPa.
1. Check Startup configuration.
6 - 47
6. TROUBLESHOOTING
66 76 980_02
1 CON Fresh Gas Pressure Transducer offset outside interval Factory default With Monitoring pressure transducer test (CFI 4):
Fresh Gas Pressure Transducer offset +/- 300 mV.
1. Replace Fresh Gas Pressure Transducer.
Without Monitoring pressure transducer test (CFI 4):
1. Replace PC 1920 Control.
2 CON Reflector Pressure Transducer offset outside interval Reflector Pressure With Monitoring pressure transducer test (CFI 9):
Transducer offset +/- 300 mV.
1. Replace Reflector Pressure Transducer.
Without Monitoring pressure transducer test (CFI 9):
1. Replace PC 1920 Control.
3 CON Inspiratory Pressure Transducer offset outside interval Inspiratory With Monitoring pressure transducer test (CFI 15):
Pressure Transducer offset +/- 300 mV.
1. Replace Inspiratory Pressure Transducer.
Without Monitoring pressure transducer test (CFI 15):
1. Replace PC 1920 Control.
4 CON Expiratory Pressure Transducer offset outside interval Expiratory With Monitoring pressure transducer test (CFI 21):
Pressure Transducer offset +/- 300 mV.
1. Replace Expiratory Pressure Transducer.
Without Monitoring pressure transducer test (CFI 21):
6 - 48
6. TROUBLESHOOTING
66 76 980_02
3 Read only memory area invalid. 1. Replace Fresh Gas Pressure Transducer.
4 MON Fresh Gas Pressure Transducer offset outside interval Factory default With Control pressure transducer test (CFI 1):
Fresh Gas Pressure Transducer offset +/- 300 mV.
1. Replace Fresh Gas Pressure Transducer.
Without Control pressure transducer test (CFI 1):
1. Replace PC 1921 Monitoring.
5 MON Fresh Gas pressure outside interval 55 to 65 cmH2O. With CFI 24:
1. Replace PC 1921 Monitoring.
Without CFI 24:
1. Replace Fresh Gas Pressure Transducer.
2. Replace Fresh Gas Module AIR.
11 Reflector Pressure Transducer gain correction factor outside interval Factory 1. Check for leakage. Follow the step by step instruction in chapter 5 "Service
default Reflector Pressure Transducer gain correction factor +/- 0.05. procedures", section “Leakage detection”.
With CFI 17 and CFI 23:
2. Replace Fresh Gas Pressure Transducer.
Without CFI 17 and CFI 23:
2. Replace Reflector Pressure Transducer.
3. Replace Fresh Gas Pressure Transducer.
15 MON Inspiratory Pressure Transducer offset outside interval Factory default With Control pressure transducer test (CFI 3):
Inspiratory Pressure Transducer offset +/- 300 mV.
6 - 50
6. TROUBLESHOOTING
66 76 980_02
17 Inspiratory Pressure Transducer gain correction factor outside interval Factory 1. Check for leakage. Follow the step by step instruction in chapter 5 "Service
default Inspiratory Pressure Transducer gain correction factor +/- 0.05. procedures", section “Leakage detection”.
With CFI 11 and CFI 23:
2. Replace Fresh Gas Pressure Transducer.
Without CFI 11 and CFI 23:
2. Replace Inspiratory Pressure Transducer.
3. Replace Fresh Gas Pressure Transducer.
21 MON Expiratory Pressure Transducer offset outside interval Factory default With Control pressure transducer test (CFI 4):
Expiratory Pressure Transducer offset +/- 300 mV.
1. Replace Expiratory Pressure Transducer.
Without Control pressure transducer test (CFI 4):
6 - 51
6. TROUBLESHOOTING
66 76 980_02
23 Expiratory Pressure Transducer gain correction factor outside interval Factory 1. Check for leakage. Follow the step by step instruction in chapter 5 "Service
default Expiratory Pressure Transducer gain correction factor +/- 0.05. procedures", section “Leakage detection”.
With CFI 17 and CFI 23:
2. Replace Fresh Gas Pressure Transducer.
Without CFI 17 and CFI 23:
2. Replace Expiratory Pressure Transducer.
3. Replace Fresh Gas Pressure Transducer.
24 Absolute difference of CON Fresh Gas pressure and MON Fresh Gas pressure 1. Replace PC 1921 Monitoring.
> 1 cmH2O
OR
Absolute difference of CON Reflector pressure and MON Reflector pressure >
1 cmH2O
OR
Absolute difference of CON Inspiratory pressure and MON Inspiratory pressure
> 1 cmH2O
OR
Absolute difference of CON Expiratory pressure and MON Expiratory pressure
> 1 cmH2O.
2. Replace PC 1920 Control.
6 - 52
25 Central Gas Supply AIR pressure outside interval 2500 hPa to 6500 hPa. 1. Check central gas supply.
6. TROUBLESHOOTING
66 76 980_02
1 Central Gas Supply AIR pressure outside interval 2500 hPa to 6500 hPa 1. Check central AIR and O2 gas supply.
OR
Central Gas Supply O2 pressure outside interval 2500 hPa to 6500 hPa.
2. Replace PC 1906 Gas Block.
2 Pressure too low during APL performance check #1. 1. Check that the APL/PEEP Valve membrane is clean and correctly seated in
the Patient Cassette.
2. Check for leakage. Follow the step by step instruction in chapter 5 "Service
procedures", section “Leakage detection”.
3. Replace APL/PEEP Valve Coil.
3 APL/PEEP Valve Coil too strong. 1. Make sure the Emergency APL Pressure regulator is set to minimum in the
Emergency Ventilation Section.
2. Check that the APL/PEEP Valve membrane is clean and correctly seated in
the Patient Cassette.
3. Replace Patient Cassette.
4. Check if the Emergency Ventilation is causing the problem by removing the
Mushroom Valve and it's holder from the APL/PEEP Valve Coil.
5. Replace APL/PEEP Valve Coil.
4 APL/PEEP Valve Coil too weak. 1. Make sure the Emergency APL Pressure regulator is set to minimum in the
Emergency Ventilation Section.
2. Check that the APL/PEEP Valve membrane is clean and correctly seated in
the Patient Cassette.
3. Replace Patient Cassette.
6 - 53
6. TROUBLESHOOTING
66 76 980_02
5 APL/PEEP Valve Coil too weak. 1. Make sure the Emergency APL Pressure regulator is set to minimum in the
Emergency Ventilation Section.
2. Check that the APL/PEEP Valve membrane is clean and correctly seated in
the Patient Cassette.
3. Replace Patient Cassette.
4. Check if the Emergency Ventilation is causing the problem by removing the
Mushroom Valve and it's holder from the APL/PEEP Valve Coil.
5. Replace APL/PEEP Valve Coil.
6 New APL/PEEP Valve Coil gain outside interval 7.3 mA/cmH2O to 11.3 1. Make sure the Emergency APL Pressure regulator is set to minimum in the
mA/cmH2O. Emergency Ventilation Section.
2. Check that the APL/PEEP Valve membrane is clean and correctly seated in
the Patient Cassette.
3. Check for leakage. Follow the step by step instruction in chapter 5 "Service
procedures", section “Leakage detection”.
4. Check if the Emergency Ventilation is causing the problem by removing the
Mushroom Valve and it's holder from the APL/PEEP Valve Coil.
5. Replace APL/PEEP Valve Coil.
7 New APL/PEEP Valve Coil offset outside interval 67.68 mA to 156.84 mA. 1. Make sure the Emergency APL Pressure regulator is set to minimum in the
Emergency Ventilation Section.
2. Check that the APL/PEEP Valve membrane is clean and correctly seated in
the Patient Cassette.
3. Check for leakage. Follow the step by step instruction in chapter 5 "Service
procedures", section “Leakage detection”.
4. Check if the Emergency Ventilation is causing the problem by removing the
Mushroom Valve and it's holder from the APL/PEEP Valve Coil.
6 - 54
6. TROUBLESHOOTING
66 76 980_02
8 Arithmetic check for new APL/PEEP Valve Coil offset/gain failed. 1. Make sure the Emergency APL Pressure regulator is set to minimum in the
Emergency Ventilation Section.
2. Check that the APL/PEEP Valve membrane is clean and correctly seated in
the Patient Cassette.
3. Check for leakage. Follow the step by step instruction in chapter 5 "Service
procedures", section “Leakage detection”.
4. Check if the Emergency Ventilation is causing the problem by removing the
Mushroom Valve and it's holder from the APL/PEEP Valve Coil.
5. Replace APL/PEEP Valve Coil.
9 APL performance check #2 (APL = 50.00) pressure outside interval 45 cmH2O 1. Make sure the Emergency APL Pressure regulator is set to minimum in the
to 55 cmH2O Emergency Ventilation Section.
OR
APL performance check #2 (APL = 2.00) pressure outside interval 1 cmH2O to
3.5 cmH2O.
2. Check that the APL/PEEP Valve membrane is clean and correctly seated in
the Patient Cassette.
3. Check for leakage. Follow the step by step instruction in chapter 5 "Service
procedures", section “Leakage detection”.
4. Check if the Emergency Ventilation is causing the problem by removing the
Mushroom Valve and it's holder from the APL/PEEP Valve Coil.
5. Replace APL/PEEP Valve Coil.
6 - 55
6. TROUBLESHOOTING
66 76 980_02
1 Fresh Gas Safety Valve indicated to be open at test start. 1. Replace PC 1919 Exp. Channel.
2. Replace PC 1921 Monitoring.
2 Disable valves feedback signal active while Fresh Gas Safety Valve indicated 1. Replace PC 1919 Exp. Channel.
to be closed.
2. Replace PC 1921 Monitoring.
3 Fresh Gas Safety Valve indicated to be open 5 s after deactivation of disable 1. Replace PC 1919 Exp. Channel.
valves signal.
2. Replace PC 1921 Monitoring.
3. Replace PC 1781 Fresh Gas Pressure Transducer.
4 Disable valves feedback signal inactive after activation of disable valves signal. 1. Replace PC 1921 Monitoring.
5 Fresh Gas Safety Valve indicated to be closed 5 s after activation of disable 1. Replace PC 1919 Exp. Channel.
valves signal.
2. Replace PC 1921 Monitoring.
1 Fresh Gas Safety Valve indicated to be open at start of test. 1. Replace PC 1919 Exp. Channel.
2. Replace PC 1781 Fresh Gas Pressure Transducer.
3. Replace PC 1920 Control.
2 Fresh Gas Safety Valve indicated to be open on entry in state where response 1. Replace PC 1919 Exp. Channel.
to activation of signal disable valves shall be tested.
6 - 56
6. TROUBLESHOOTING
66 76 980_02
3 Disable valves feedback signal inactive 5 s after activation of disable valves 1. Replace PC 1920 Control.
signal.
4 Disable valves feedback signal active 5 s after deactivation of disable valves 1. Replace PC 1921 Monitoring.
signal.
2. Replace PC 1920 Control.
5 Failed to reach a pressure below 5 cmH2O within 1 s from attempt to open 1. Replace PC 1920 Control.
Fresh Gas Safety Valve, which indicates that the valve did not open properly.
2. Replace PC 1919 Exp. Channel.
6 Pressurization peak pressure above 140 cmH2O during two consecutive 1. Replace PC 1919 Exp. Channel.
pressurization cycles, which indicates that the Fresh Gas Safety Valve did not
open.
2. Replace Fresh Gas Safety Valve.
3. Replace PC 1781 Fresh Gas Pressure Transducer.
7 Reading of current calibration value failed 1. Replace Fresh Gas Safety Valve.
OR
New calibration value outside valid range (0 to 255 DAC steps)
OR
Writing of new calibration failed.
Notice: This is failure may occur during first adjustment attempt.
2. Replace PC 1919 Exp. Channel.
6 - 57
6. TROUBLESHOOTING
66 76 980_02
10 Pressurization cycle did not finish within 10 s. 1. Check for leakage. Follow the step by step instruction in chapter 5 "Service
Notice: A pressurization cycle lasts until opening of Fresh Gas Safety Valve is procedures", section “Leakage detection”.
indicated or Pressurization peak pressure reaches its upper pressure limit or
until Pressurization time reaches its upper limit
11 Pressure > 25 cmH2O not reached within 10 s during the final part of the test 1. Check for leakage. Follow the step by step instruction in chapter 5 "Service
where SW controlled opening of Fresh Gas Safety Valve shall be checked. procedures", section “Leakage detection”.
12 Central Gas Supply AIR pressure outside interval 2500 hPa to 6500 hPa 1. Check central gas AIR and O2 supply.
OR
Central Gas Supply O2 pressure outside interval 2500 hPa to 6500 hPa.
2. Replace PC 1906 Gas Block.
6 - 58
6. TROUBLESHOOTING
66 76 980_02
1 VAP Bypass Valve pressurization stage, Fresh Gas end pressure ≤ 10 cmH2O 1. Check/replace Double channel plate.
OR
VAP Bypass Valve pressurization stage, Reflector end pressure ≤ 5 cmH2O
after 7000 ms pressurization.
2 VAP Bypass Valve pressure difference check stage, Fresh Gas/Reflector end 1. Check Vaporizer bypass valve PV5.
pressure difference < 30 % of VAP Bypass Valve pressure difference check
stage, Fresh Gas end pressure.
2. Replace Vaporizer Bypass Pilot Valve EMV4.
3. Replace PC 1907 Valve drivers.
3 VAP Bypass Valve pressure equalization stage, Fresh Gas/Reflector end 1. Check Vaporizer bypass valve PV5.
pressure difference > 10 % of VAP Bypass Valve pressure equalization stage,
Fresh Gas end pressure after 500 ms equalization.
2. Replace Vaporizer Bypass Pilot Valve EMV4.
3. Replace PC 1907 Valve drivers.
4 Pressure(s) too low during VAP 1 Valves pressurization stage. 1. Check/replace Double channel plate.
5 Fresh Gas/Reflector pressure difference too low during VAP 1 Valve pressure 1. Check Vaporizer 1 inlet/outlet valves PV1/PV2.
difference check stage.
2. Replace Vaporizer 1 Selection Pilot Valve EMV6.
3. Replace PC 1907 Valve drivers.
6 Fresh Gas pressure too low during VAP 1 Valve pressure difference check 1. Check/replace Double channel plate.
stage.
7 Fresh Gas pressure too low to measure during VAP 1 Valves pressure 1. Check/replace Double channel plate.
6 - 59
equalization stage.
6. TROUBLESHOOTING
66 76 980_02
8 Fresh Gas/Reflector pressure difference too high during VAP 1 Valves 1. Check/replace Vaporizer 1.
pressure equalization stage.
2. Check Vaporizer 1 inlet/outlet valves PV1/PV2.
3. Check/replace Double channel plate.
4. Replace Vaporizer 1 Selection Pilot Valve EMV6.
5. Replace PC 1907 Valve drivers.
9 Pressure(s) too high during VAP 1 Valves pressure equalization stage. 1. Check Vaporizer 1 inlet/outlet valves PV1/PV2.
2. Replace Vaporizer 1 Selection Pilot Valve EMV6.
3. Replace PC 1907 Valve drivers.
10 Pressure(s) too low during VAP 2 Valves pressurization stage. 1. Check/replace Double channel plate.
11 Pressure(s) too low during VAP 2 Valves pressurization stage. 1. Check/replace Double channel plate.
12 Fresh Gas/Reflector pressure difference too low during VAP 2 Valve pressure 1. Check Vaporizer 1 inlet/outlet valves PV1/PV2.
difference check stage.
2. Replace Vaporizer 1 Selection Pilot Valve EMV6.
3. Replace PC 1907 Valve drivers.
13 Fresh Gas/Reflector pressure difference too low during VAP 2 Valve pressure 1. Check Vaporizer 1 inlet/outlet valves PV1/PV2.
difference check stage.
2. Replace Vaporizer 1 Selection Pilot Valve EMV6.
3. Replace PC 1907 Valve drivers.
6 - 60
6. TROUBLESHOOTING
66 76 980_02
14 Fresh Gas pressure too low during VAP 2 Valve pressure difference check 1. Check/replace Double channel plate.
stage.
15 Fresh Gas pressure too low during VAP 2 Valve pressure difference check 1. Check/replace Double channel plate.
stage.
16 Fresh Gas pressure too low to measure during VAP 2 Valves pressure 1. Check/replace Double channel plate.
equalization stage.
2. Check Vaporizer 2 inlet/outlet valves PV3/PV4.
3. Replace Vaporizer 2 Selection Pilot Valve EMV5.
4. Replace PC 1907 Valve drivers.
17 Fresh Gas pressure too low to measure during VAP 2 Valves pressure 1. Check/replace Double channel plate.
equalization stage.
2. Check Vaporizer 2 inlet/outlet valves PV3/PV4.
3. Replace Vaporizer 2 Selection Pilot Valve EMV5.
4. Replace PC 1907 Valve drivers.
18 Fresh Gas/Reflector pressure difference too high during VAP 2 Valves 1. Check/replace Vaporizer 2.
pressure equalization stage.
2. Check Vaporizer 2 inlet/outlet valves PV3/PV4.
3. Check/replace Double channel plate.
4. Replace Vaporizer 2 Selection Pilot Valve EMV5.
5. Replace PC 1907 Valve drivers.
19 Fresh Gas/Reflector pressure difference too high during VAP 2 Valves 1. Check/replace Vaporizer 2.
pressure equalization stage.
2. Check Vaporizer 2 inlet/outlet valves PV3/PV4.
6 - 61
6. TROUBLESHOOTING
66 76 980_02
20 Pressure(s) too high during VAP 2 Valves pressure equalization stage. 1. Check Vaporizer 1 inlet/outlet valves PV1/PV2.
2. Replace Vaporizer 1 Selection Pilot Valve EMV6.
3. Replace PC 1907 Valve drivers.
21 Pressure(s) too high during VAP 2 Valves pressure equalization stage. 1. Check Vaporizer 1 inlet/outlet valves PV1/PV2.
2. Replace Vaporizer 1 Selection Pilot Valve EMV6.
3. Replace PC 1907 Valve drivers.
22 Central Gas Supply AIR pressure outside interval 2500 hPa to 6500 hPa 1. Check central AIR and O2 gas supply.
OR
Central Gas Supply O2 pressure outside interval 2500 hPa to 6500 hPa.
2. Replace PC 1906 Gas Block.
2 Fresh Gas Module AIR pressure < 2500 hPa. 1. Check central AIR gas supply.
2. Replace Fresh Gas Module AIR.
6 - 62
6. TROUBLESHOOTING
66 76 980_02
4 Fresh Gas Module N2O pressure < 2500 hPa. 1. Check central N2O gas supply.
2. Replace Fresh Gas Module N2O.
5 CON Reflector Gas Module Flow Transducer offset outside interval 0 mV to 40 With Monitoring flow transducer test (CFI 15):
mV.
1. Replace Reflector Gas Module.
Without Monitoring flow transducer test (CFI 15):
1. Replace PC 1920 Control.
6 CON Fresh Gas Module O2 Flow Transducer offset outside interval 0 mV to 40 With Monitoring flow transducer test (CFI 16):
mV.
1. Replace Fresh Gas Module O2.
Without Monitoring flow transducer test (CFI 16):
1. Replace PC 1920 Control.
7 CON Fresh Gas Module AIR Flow Transducer offset outside interval 0 mV to With Monitoring flow transducer test (CFI 17):
40 mV.
1. Replace Fresh Gas Module AIR.
Without Monitoring flow transducer test (CFI 17):
1. Replace PC 1920 Control.
6 - 63
6. TROUBLESHOOTING
66 76 980_02
9 CON Expiratory Flow Transducer offset outside interval 2480 mV to 2520 mV. With Monitoring flow transducer test (CFI 19):
1. Replace Patient Cassette.
Without Monitoring flow transducer test (CFI 19):
1. Replace PC 1920 Control.
10 Reflector Gas Module false flow correction outside interval 225 to 775. 1. Check that Reflector Gas Module is properly connected.
2. Replace Reflector Gas Module.
3. Replace PC 1920 Control.
11 Fresh Gas Module O2 false flow correction outside interval 225 to 775. 1. Check that Fresh Gas Module O2 is properly connected.
2. Replace Fresh Gas Module O2.
3. Replace PC 1920 Control.
12 Fresh Gas Module AIR false flow correction outside interval 225 to 775. 1. Check that Fresh Gas Module AIR is properly connected.
2. Replace Fresh Gas Module AIR.
3. Replace PC 1920 Control.
13 Fresh Gas Module N2O false flow correction outside interval 225 to 775. 1. Check that Fresh Gas Module N2O is properly connected.
2. Replace Fresh Gas Module N2O.
3. Replace PC 1920 Control.
6 - 64
6. TROUBLESHOOTING
66 76 980_02
15 Fresh Gas Module O2 dynamic common mode correction outside interval - 1. Check that Fresh Gas Module O2 is properly connected.
1371 to 1371.
Notice: When this CFI occurs, Fresh Gas Module O2 dynamic common mode
correction is reset to 0.
2. Replace Fresh Gas Module O2.
3. Replace PC 1920 Control.
16 Fresh Gas Module AIR dynamic common mode correction outside interval - 1. Check that Fresh Gas Module AIR is properly connected.
1371 to 1371.
Notice: When this CFI occurs, Fresh Gas Module AIR dynamic common mode
correction is reset to 0.
2. Replace Fresh Gas Module AIR.
3. Replace PC 1920 Control.
17 Fresh Gas Module N2O dynamic common mode correction outside interval - 1. Check that Fresh Gas Module N2O is properly connected.
1371 to 1371.
Notice: When this CFI occurs, Fresh Gas Module N2O dynamic common mode
correction is reset to 0.
2. Replace Fresh Gas Module N2O.
3. Replace PC 1920 Control.
18 No docked Patient Cassette detected. 1. Check the Patient Cassette is properly mounted.
2. Replace Patient Cassette.
6 - 65
6. TROUBLESHOOTING
66 76 980_02
1 Central Gas Supply AIR pressure outside interval 2500 hPa to 6500 hPa 1. Check central AIR and O2 gas supply.
OR
Central Gas Supply O2 pressure outside interval 2500 hPa to 6500 hPa.
2. Replace PC 1906 Gas Block.
2 Central Gas Supply N2O pressure outside interval 2500 hPa to 4500 hPa. 1. Check central N2O gas supply.
2. Replace PC 1906 Gas Block.
3 No message indicating that the noise measurement is finished has been 1. Replace PC 1919 Exp. Channel.
received from Exp. Channel subsystem within 5 s after start command was
sent.
4 No acknowledgement message has been received from Exp. Channel 1. Replace PC 1919 Exp. Channel.
subsystem within 1 s after message with set test result command was sent.
5 No acknowledgement message has been received from Exp. Channel 1. Replace PC 1919 Exp. Channel.
subsystem within 2 s after zeroing command message was sent.
6 Expiratory Flow Transducer k-value outside interval 0.92 to 1.08. 1. Contact MAQUET HSC via MSupport.
7 Sum of flow from all gas modules / (expiratory flow * k) for AIR/O2 stage is 1. Contact MAQUET HSC via MSupport.
outside interval 0.92 to 1.08.
8 Sum of flow from all gas modules / (expiratory flow * k) for AIR/N2O stage is 1. Contact MAQUET HSC via MSupport.
outside interval 0.92 to 1.08.
6 - 66
6. TROUBLESHOOTING
66 76 980_02
9 Sum of flow from all gas modules / (expiratory flow * k) for O2/N2O stage is 1. Contact MAQUET HSC via MSupport.
outside interval 0.92 to 1.08.
10 Sum of flow from all gas modules / (expiratory flow * k) for AIR stage is outside 1. Contact MAQUET HSC via MSupport.
interval 0.92 to 1.08.
11 Sum of flow from all gas modules / (expiratory flow * k) for O2 stage is outside 1. Contact MAQUET HSC via MSupport.
interval 0.92 to 1.08.
12 Sum of flow from all gas modules / (expiratory flow * k) for Reflector stage is 1. Contact MAQUET HSC via MSupport.
outside interval 0.92 to 1.08.
13 No docked Patient Cassette detected. 1. Check the Patient Cassette is properly mounted.
2. Replace Patient Cassette.
3. Check PC 1918 Exp. Channel Connector and associated cabling to PC 1919
Exp. Channel. Replace if necessary.
14 Metrics (measured values) from Exp. Channel subsystem are not available 1. Replace Patient Cassette.
onboard MON subsystem.
Notice: This may be due to an error in Exp. Channel subsystem or possibly
the Patient Cassette.
2. Replace PC 1919 Exp. Channel.
15 MON Reflector Gas Module Flow Transducer offset outside interval 0 mV to 40 With Control flow transducer test (CFI 5):
mV.
1. Replace Reflector Gas Module.
Without Control flow transducer test (CFI 5):
1. Replace PC 1921 Monitoring.
6 - 67
6. TROUBLESHOOTING
66 76 980_02
17 MON Fresh Gas Module AIR Flow Transducer offset outside interval 0 mV to With Control flow transducer test (CFI 7):
40 mV.
1. Replace Fresh Gas Module AIR.
Without Control flow transducer test (CFI 7):
1. Replace PC 1921 Monitoring.
18 MON Fresh Gas Module N2O Flow Transducer offset outside interval 0 mV to With Control flow transducer test (CFI 8):
40 mV.
1. Replace Fresh Gas Module N2O.
Without Control flow transducer test (CFI 8):
1. Replace PC 1921 Monitoring.
19 MON Expiratory Flow Transducer offset outside interval 2480 mV to 2520 mV. With Control flow transducer test (CFI 9):
1. Replace Patient Cassette.
Without Control flow transducer test (CFI 9):
1. Replace PC 1921 Monitoring.
20 AIR/O2 stage, CON Fresh Gas Module AIR flow outside interval 29.4 l/min to With CFI 21:
30.6 l/min.
1. Replace Fresh Gas Module AIR.
Without CFI 21:
1. Replace PC 1920 Control.
6 - 68
6. TROUBLESHOOTING
66 76 980_02
21 AIR/O2 stage, MON Fresh Gas Module AIR flow outside interval 29.4 l/min to With CFI 20:
30.6 l/min.
1. Replace Fresh Gas Module AIR.
Without CFI 20:
1. Replace PC 1921 Monitoring.
22 AIR/O2 stage, CON Fresh Gas Module O2 flow outside interval 29.4 l/min to With CFI 23:
30.6 l/min.
1. Replace Fresh Gas Module O2.
Without CFI 23:
1. Replace PC 1920 Control.
23 AIR/O2 stage, MON Fresh Gas Module O2 flow outside interval 29.4 l/min to With CFI 22:
30.6 l/min.
1. Replace Fresh Gas Module O2.
Without CFI 22:
1. Replace PC 1921 Monitoring.
24 AIR/O2 stage, CON Fresh Gas Module N2O flow outside interval -0.096 l/min With CFI 25:
to 0.096 l/min.
1. Replace Fresh Gas Module N2O.
Without CFI 25:
1. Replace PC 1920 Control.
25 AIR/O2 stage, MON Fresh Gas Module N2O flow outside interval -0.096 l/min With CFI 24:
to 0.096 l/min.
1. Replace Fresh Gas Module N2O.
6 - 69
26 AIR/O2 stage, CON Expiratory Flow Transducer flow outside interval 90 % to With CFI 27:
110 % of the total flow from all Gas Module measured by the CON.
1. CFI indicating patient cassette error may be due to other failures in the
system, e.g. gas delivery from the gas modules. In case of multiple CFI, where
patient cassette error is indicated, start troubleshooting the other indicated
parts.
2. Replace Patient Cassette.
Without CFI 27:
1. Replace PC 1920 Control.
27 AIR/O2 stage, MON Expiratory Flow Transducer flow outside interval 90 % to With CFI 26:
110 % of the total flow from all Gas Module measured by the MON.
1. CFI indicating patient cassette error may be due to other failures in the
system, e.g. gas delivery from the gas modules. In case of multiple CFI, where
patient cassette error is indicated, start troubleshooting the other indicated
parts.
2. Replace Patient Cassette.
Without CFI 26:
1. Replace PC 1921 Monitoring.
28 AIR/N2O stage, CON Fresh Gas Module N2O flow outside interval 29.4 l/min With CFI 29:
to 30.6 l/min.
1. Replace Fresh Gas Module AIR.
Without CFI 29:
1. Replace PC 1920 Control.
29 AIR/N2O stage, MON Fresh Gas Module N2O flow outside interval 29.4 l/min With CFI 28:
to 30.6 l/min.
6 - 70
6. TROUBLESHOOTING
66 76 980_02
30 AIR/N2O stage, CON Fresh Gas Module O2 flow outside interval -0.096 l/min With CFI 31:
to 0.096 l/min.
1. Replace Fresh Gas Module O2.
Without CFI 31:
1. Replace PC 1920 Control.
31 AIR/N2O stage, MON Fresh Gas Module O2 flow outside interval -0.096 l/min With CFI 30:
to 0.096 l/min.
1. Replace Fresh Gas Module O2.
Without CFI 30:
1. Replace PC 1921 Monitoring.
32 AIR/N2O stage, CON Fresh Gas Module AIR flow outside interval 29.4 l/min to With CFI 33:
30.6 l/min.
1. Replace Fresh Gas Module N2O.
Without CFI 33:
1. Replace PC 1920 Control.
33 AIR/N2O stage, MON Fresh Gas Module AIR flow outside interval 29.4 l/min to With CFI 32:
30.6 l/min.
1. Replace Fresh Gas Module N2O.
Without CFI 32:
1. Replace PC 1921 Monitoring.
6 - 71
6. TROUBLESHOOTING
66 76 980_02
35 AIR/N2O stage, MON Reflector Gas Module flow outside interval -0.096 l/min With CFI 34:
to 0.096 l/min.
1. Replace Reflector Gas Module.
Without CFI 34:
1. Replace PC 1921 Monitoring.
36 AIR/N2O stage, CON Expiratory Flow Transducer flow outside interval 90 % to With CFI 37:
110 % of the total flow from all Gas Module measured by the CON subsystem.
1. CFI indicating patient cassette error may be due to other failures in the
system, e.g. gas delivery from the gas modules. In case of multiple CFI, where
patient cassette error is indicated, start troubleshooting the other indicated
parts.
2. Replace Patient Cassette.
Without CFI 37:
1. Replace PC 1920 Control.
37 AIR/N2O stage, MON Expiratory Flow Transducer flow outside interval 90 % to With CFI 36:
110 % of the total flow from all Gas Module measured by the MON subsystem.
1. CFI indicating patient cassette error may be due to other failures in the
system, e.g. gas delivery from the gas modules. In case of multiple CFI, where
patient cassette error is indicated, start troubleshooting the other indicated
parts.
2. Replace Patient Cassette.
Without CFI 36:
6 - 72
6. TROUBLESHOOTING
66 76 980_02
38 O2/N2O stage, CON Fresh Gas Module AIR flow outside interval -0.096 l/min With CFI 39:
to 0.096 l/min.
1. Replace Fresh Gas Module AIR.
Without CFI 39:
1. Replace PC 1920 Control.
39 O2/N2O stage, MON Fresh Gas Module AIR flow outside interval -0.096 l/min With CFI 38:
to 0.096 l/min.
1. Replace Fresh Gas Module AIR.
Without CFI 38:
1. Replace PC 1921 Monitoring.
40 O2/N2O stage, CON Fresh Gas Module O2 flow outside interval 29.4 l/min to With CFI 41:
30.6 l/min.
1. Replace Fresh Gas Module O2.
Without CFI 41:
1. Replace PC 1920 Control.
41 O2/N2O stage, MON Fresh Gas Module O2 flow outside interval 29.4 l/min to With CFI 40:
30.6 l/min.
1. Replace Fresh Gas Module O2.
Without CFI 40:
1. Replace PC 1921 Monitoring.
42 O2/N2O stage, CON Fresh Gas Module N2O flow outside interval 29.4 l/min to With CFI 43:
30.6 l/min.
6 - 73
43 O2/N2O stage, MON Fresh Gas Module N2O flow outside interval 29.4 l/min to With CFI 42:
30.6 l/min.
1. Replace Fresh Gas Module N2O.
Without CFI 42:
1. Replace PC 1921 Monitoring.
44 O2/N2O stage, CON Expiratory Flow Transducer flow outside interval 90 % to With CFI 45:
110 % of the total flow from all Gas Module measured by the CON subsystem.
1. CFI indicating patient cassette error may be due to other failures in the
system, e.g. gas delivery from the gas modules. In case of multiple CFI, where
patient cassette error is indicated, start troubleshooting the other indicated
parts.
2. Replace Patient Cassette.
Without CFI 45:
1. Replace PC 1920 Control.
45 O2/N2O stage, MON Expiratory Flow Transducer flow outside interval 90 % to With CFI 44:
110 % of the total flow from all Gas Module measured by the MON subsystem.
1. CFI indicating patient cassette error may be due to other failures in the
system, e.g. gas delivery from the gas modules. In case of multiple CFI, where
patient cassette error is indicated, start troubleshooting the other indicated
parts.
2. Replace Patient Cassette.
Without CFI 44:
1. Replace PC 1921 Monitoring.
6 - 74
6. TROUBLESHOOTING
66 76 980_02
47 AIR stage, MON Fresh Gas Module AIR flow outside interval 29.4 l/min to 30.6 With CFI 46:
l/min.
1. Replace Fresh Gas Module AIR.
Without CFI 46:
1. Replace PC 1921 Monitoring.
48 AIR stage, CON Fresh Gas Module O2 flow outside interval -0.096 l/min to With CFI 49:
0.096 l/min.
1. Replace Fresh Gas Module O2.
Without CFI 49:
1. Replace PC 1920 Control.
49 AIR stage, MON Fresh Gas Module O2 flow outside interval -0.096 l/min to With CFI 48:
0.096 l/min.
1. Replace Fresh Gas Module O2.
Without CFI 48:
1. Replace PC 1921 Monitoring.
50 AIR stage, CON Fresh Gas Module N2O flow outside interval -0.096 l/min to With CFI 51:
0.096 l/min.
1. Replace Fresh Gas Module N2O.
Without CFI 51:
6 - 75
51 AIR stage, MON Fresh Gas Module N2O flow outside interval -0.096 l/min to With CFI 50:
0.096 l/min.
1. Replace Fresh Gas Module N2O.
Without CFI 50:
1. Replace PC 1921 Monitoring.
52 AIR stage, CON Reflector Gas Module flow outside interval -0.096 l/min to With CFI 53:
0.096 l/min.
1. Replace Reflector Gas Module.
Without CFI 53:
1. Replace PC 1920 Control.
53 AIR stage, MON Reflector Gas Module flow outside interval -0.096 l/min to With CFI 52:
0.096 l/min.
1. Replace Reflector Gas Module.
Without CFI 52:
1. Replace PC 1921 Monitoring.
54 AIR stage, CON Expiratory Flow Transducer flow outside interval 90 % to 110 With CFI 55:
% of the total flow from all Gas Module measured by the CON subsystem.
1. CFI indicating patient cassette error may be due to other failures in the
system, e.g. gas delivery from the gas modules. In case of multiple CFI, where
patient cassette error is indicated, start troubleshooting the other indicated
parts.
2. Replace Patient Cassette.
Without CFI 55:
1. Replace PC 1920 Control.
6 - 76
6. TROUBLESHOOTING
66 76 980_02
56 O2 stage, CON Fresh Gas Module AIR flow outside interval -0.096 l/min to With CFI 57:
0.096 l/min.
1. Replace Fresh Gas Module AIR.
Without CFI 57:
1. Replace PC 1920 Control.
57 O2 stage, MON Fresh Gas Module AIR flow outside interval -0.096 l/min to With CFI 56:
0.096 l/min.
1. Replace Fresh Gas Module AIR.
Without CFI 56:
1. Replace PC 1921 Monitoring.
58 O2 stage, CON Fresh Gas Module O2 flow outside interval 29.4 l/min to 30.6 With CFI 59:
l/min.
1. Replace Fresh Gas Module O2.
Without CFI 59:
1. Replace PC 1920 Control.
59 O2 stage, MON Fresh Gas Module O2 flow outside interval 29.4 l/min to 30.6 With CFI 58:
6 - 77
l/min.
6. TROUBLESHOOTING
66 76 980_02
60 O2 stage, CON Fresh Gas Module N2O flow outside interval -0.096 l/min to With CFI 61:
0.096 l/min.
1. Replace Fresh Gas Module N2O.
Without CFI 61:
1. Replace PC 1920 Control.
61 O2 stage, MON Fresh Gas Module N2O flow outside interval -0.096 l/min to With CFI 60:
0.096 l/min.
1. Replace Fresh Gas Module N2O.
Without CFI 60:
1. Replace PC 1921 Monitoring.
62 O2 stage, CON Expiratory Flow Transducer flow outside interval of 90 % to With CFI 63:
110 % of the total flow from all Gas Module measured by the CON subsystem.
1. CFI indicating patient cassette error may be due to other failures in the
system, e.g. gas delivery from the gas modules. In case of multiple CFI, where
patient cassette error is indicated, start troubleshooting the other indicated
parts.
2. Replace Patient Cassette.
Without CFI 63:
1. Replace PC 1920 Control.
63 O2 stage, MON Expiratory Flow Transducer flow outside interval of 90 % to With CFI 62:
110 % of the total flow from all Gas Module measured by the MON subsystem.
6 - 78
6. TROUBLESHOOTING
66 76 980_02
64 Reflector stage, CON Fresh Gas Module AIR flow outside interval 29.4 l/min to With CFI 65:
30.6 l/min.
1. Replace Fresh Gas Module AIR.
Without CFI 65:
1. Replace PC 1920 Control.
65 Reflector stage, MON Fresh Gas Module AIR flow outside interval 29.4 l/min to With CFI 64:
30.6 l/min.
1. Replace Fresh Gas Module AIR.
Without CFI 64:
1. Replace PC 1921 Monitoring.
66 Reflector stage, CON Fresh Gas Module N2O flow outside interval -0.096 l/min With CFI 67:
to 0.096 l/min.
1. Replace Fresh Gas Module N2O.
Without CFI 67:
1. Replace PC 1920 Control.
67 Reflector stage, MON Fresh Gas Module N2O flow outside interval -0.096 l/min With CFI 66:
to 0.096 l/min.
1. Replace Fresh Gas Module N2O.
6 - 79
6. TROUBLESHOOTING
66 76 980_02
68 Reflector stage, CON Reflector Gas Module flow outside interval 29.4 l/min to With CFI 69:
30.6 l/min.
1. Replace Reflector Gas Module.
Without CFI 69:
1. Replace PC 1920 Control.
69 Reflector stage, MON Reflector Gas Module flow outside interval 29.4 l/min to With CFI 68:
30.6 l/min.
1. Replace Reflector Gas Module.
Without CFI 68:
1. Replace PC 1921 Monitoring.
70 Reflector stage, CON Expiratory Flow Transducer flow outside interval -1 * 90 With CFI 71:
% to -1 * 110 % of the total flow from all Gas Module measured by the CON
subsystem.
1. CFI indicating patient cassette error may be due to other failures in the
system, e.g. gas delivery from the gas modules. In case of multiple CFI, where
patient cassette error is indicated, start troubleshooting the other indicated
parts.
2. Replace Patient Cassette.
Without CFI 71:
1. Replace PC 1920 Control.
71 Reflector stage, MON Expiratory Flow Transducer flow outside interval -1 * 90 With CFI 70:
% to -1 * 110 % of the total flow from all Gas Module measured by the MON
subsystem.
6 - 80
6. TROUBLESHOOTING
66 76 980_02
1 Measurement stage, 10 s pressure < 10 cmH2O 1. Check for leakage. Follow the step by step instruction in chapter 5 "Service
OR procedures", section “Leakage detection”.
Measurement stage, 10 s pressure < Pressurization stage, Start pressure.
2 Leakage > 300 ml/min. 1. Check for leakage. Follow the step by step instruction in chapter 5 "Service
procedures", section “Leakage detection”.
4 Cisothermal system < 2.60 ml/cmH2O 1. Check the Barometric pressure and redo the Pressure transducer test and
OR Flow transducer test.
Cadiabatic system > Cisothermal system. Note: This failure might occur if the AUTO ventilation test is done without a
CO2 absorber.
6 - 81
6. TROUBLESHOOTING
66 76 980_02
6 Leakage > 150 ml/min 1. Check for leakage. Follow the step by step instruction in chapter 5 "Service
AND procedures", section “Leakage detection”.
Leakage ≤ 300 ml/min.
8 Central Gas Supply AIR pressure outside interval 2500 hPa to 6500 hPa 1. Check central AIR and O2 gas supply.
OR
Central Gas Supply O2 pressure outside interval 2500 hPa to 6500 hPa.
2. Replace PC 1906 Gas Block.
9 The absorber switch indicates that no CO2 Absorber is docked. 1. Check that the Absorber is mounted correctly and the switch is in upright
position.
2. Check/replace the Absorber Sensor.
10 Locking of Fresh Gas Safety Valve failed. 1. Replace PC 1919 Exp. Channel.
11 Gas leakage into system > 150 ml/min. 1. Check/replace O2 Flush Valve.
1 Measurement stage, Reflector end pressure < 30 cmH2O. 1. Replace the manual tubing and/or Manual Ventilation Bag.
2 Locking of Fresh Gas Safety Valve failed. 1. Replace PC 1919 Exp. Channel.
6 - 82
6. TROUBLESHOOTING
66 76 980_02
4 Pressurization stage, Fresh Gas end pressure deviates 0.5 cmH2O or more 1. Replace the manual tubing and/or Manual Ventilation Bag.
from 33 cmH2O after 12 s pressurization attempt.
5 Leakage > 300 ml/min. 1. Replace the manual tubing and/or Manual Ventilation Bag.
7 Manual Ventilation Bag flow trigger detection time > 0.5 s, i.e. flow from 1. Check/replace Manual Ventilation Valve.
Manual Ventilation Bag did not exceed 6 l/min within 0.5 s from opening of
Manual Ventilation Valve.
2. Replace Manual Ventilation Valve Actuator.
8 Measurement stage, Reflector end pressure > Measurement stage, Reflector 1. Replace Reflector Gas Module Nozzle Unit.
start pressure + 1 cmH2O.
2. Replace Fresh Gas Module AIR Nozzle Unit.
10 Central Gas Supply AIR pressure outside interval 2500 hPa to 6500 hPa 1. Check central AIR and O2 gas supply.
OR
Central Gas Supply O2 pressure outside interval 2500 hPa to 6500 hPa.
2. Replace PC 1906 Gas Block.
6 - 83
6. TROUBLESHOOTING
66 76 980_02
1 PGA indicates that it has detected an error when PGA status is checked in the 1. Check for information in the Technical alarms log. For recommended
beginning of the test. actions, refer to Troubleshooting, section Technical error codes and messages.
2. Replace Patient Gas Analyzer AION.
2 CGA indicates that it has detected an error when CGA status is checked in the 1. Check for information in the Technical alarms log. For recommended
beginning of the test. actions, refer to Troubleshooting, section Technical error codes and messages.
2. Replace Control Gas Analyzer AION.
3 PGA indicates that it has detected an error when PGA status is checked in the 1. Check for information in the Technical alarms log. For recommended
end of the test. actions, refer to Troubleshooting, section Technical error codes and messages.
2. Replace Patient Gas Analyzer AION.
4 CGA indicates that it has detected an error when CGA status is checked in the 1. Check for information in the Technical alarms log. For recommended
end of the test. actions, refer to Troubleshooting, section Technical error codes and messages.
2. Replace Control Gas Analyzer AION.
5 Failed to read PGA error status within 20 s in the beginning of the test. 1. Check/replace Patient Gas Analyzer AION cable.
2. Replace Patient Gas Analyzer AION.
6 Failed to read CGA error status within 20 s in the beginning of the test. 1. Check/replace Control Gas Analyzer AION cable.
2. Replace Control Gas Analyzer AION.
7 Failed to read PGA error status within 20 s in the end of the test. 1. Check/replace Patient Gas Analyzer AION cable.
2. Replace Patient Gas Analyzer AION.
8 Failed to read CGA error status within 20 s in the end of the test. 1. Check/replace Control Gas Analyzer AION cable.
6 - 84
6. TROUBLESHOOTING
66 76 980_02
9 PGA zero calibration attempt has not finished within 50 s. 1. Make sure that no other gas than air was sampled through the Gas Analyzer
zero calibration gas intake.
2. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
3. Check/replace Patient Gas Analyzer AION cable.
4. Replace Patient Gas Analyzer AION.
10 CGA zero calibration attempt has not finished within 50 s. 1. Make sure that no other gas than air was sampled through the Gas Analyzer
zero calibration gas intake.
2. Redo the test and check for information in the Technical alarms log. For
recommended actions, refer to Troubleshooting, section Technical error codes
and messages.
3. Check/replace Control Gas Analyzer AION cable.
4. Replace Control Gas Analyzer AION.
11 PGA has reported error during zero calibration attempt. 1. Redo the test and check for information in the Technical alarms log. For
recommended actions, refer to Troubleshooting, section Technical error codes
and messages.
2. Replace Patient Gas Analyzer AION.
12 PGA has reported occlusion during zero calibration attempt. 1. Check for an obstruction in Patient Gas Analyzer Zero calibration gas inlet
line (from FLOW-i front until the Patient Gas Analyzer Zero gas inlet).
2. Replace Patient Gas Analyzer AION.
16 Central Gas Supply O2 pressure outside interval 2500 hPa to 6500 hPa. 1. Check central O2 gas supply.
2. Replace PC 1906 Gas Block.
17 O2 concentration check CGA Fresh Gas Sampling Point O2 concentration CGA Fresh Gas Sampling Point O2 concentration below 92%:
outside interval 92 % to 108 %.
1. Check for leakage between the Fresh Gas Elbows and the Control Gas
Analyzer inlet.
2. Calibrate Control O2 sensor.
3. Replace Control O2 sensor.
4. Replace Control Gas Analyzer Inlet Valve EMV14.
CGA Fresh Gas Sampling Point O2 concentration above 108%:
1. Calibrate Control O2 sensor.
2. Replace Control O2 sensor.
18 Pressure drop check #1 CGA Fresh Gas Sampling Point pressure drop outside CGA Fresh Gas Sampling Point pressure drop below 35 hPa:
interval 35 hPa to 80 hPa.
1. Check for leakage between the Fresh Gas Elbows and the Control Gas
Analyzer inlet.
2. Replace Control Gas Analyzer Inlet Valve EMV14.
3. Replace Control Gas Analyzer AION.
6 - 86
6. TROUBLESHOOTING
66 76 980_02
19 Pressure drop check #3 CGA Fresh Gas Sampling Point pressure drop outside CGA Fresh Gas Sampling Point pressure drop below 35 hPa:
interval 35 hPa to 80 hPa.
1. Check for leakage between the Fresh Gas Elbows and the Control Gas
Analyzer inlet.
2. Replace Control Gas Analyzer Inlet Valve.
3. Replace Control Gas Analyzer AION.
CGA Fresh Gas Sampling Point pressure drop above 80 hPa:
1. Check for an obstruction between the Fresh Gas Elbows and the Control
Gas Analyzer inlet.
2. Replace Control Gas Analyzer AION.
20 Pressure drop check #2 CGA Inspiratory Sampling Point pressure drop outside CGA Inspiratory Sampling Point pressure drop 5 hPa:
interval 5 hPa to 30 hPa.
1. Check for leakage between the fresh gas sampling filter S1 in Patient
Cassette Docking and the Control Gas Analyzer inlet.
2. Replace Control Gas Analyzer AION.
CGA Inspiratory Sampling Point pressure drop above 30 hPa:
1. Check for an obstruction between the fresh gas sampling filter S1 in Patient
Cassette Docking and the Control Gas Analyzer inlet.
2. Replace Control Gas Analyzer Inlet Valve EMV14.
3. Replace Control Gas Analyzer AION.
21 The measured pressures from Pressure drop check #1 and Pressure drop 1. If several Check failures appear in this test; perform recommended actions
check #3 differs more than 5%. for these failures.
6 - 87
6. TROUBLESHOOTING
66 76 980_02
22 Pressure drop check #1 PGA Y-piece Sampling Point pressure drop outside PGA Y-piece Sampling Point pressure drop below 25 hPa:
interval 25 hPa to 125 hPa.
1. Check leakage in the Patient Gas Analyzer sampling line, i.e. sampling line
adapter (at the Y-piece) to the Water trap.
2. Check for leakage between the Patient Gas Analyzer inlet and the Water
trap.
3. Replace Patient Gas Analyzer AION.
PGA Y-piece Sampling Point pressure drop above 125 hPa:
1. Check for an obstruction in the Patient Gas Analyzer sampling line, i.e.
sampling line adapter (at the Y-piece) to the Water trap.
2. Check for for an obstruction between the Patient Gas Analyzer inlet and the
Water trap.
3. Replace Patient Gas Analyzer AION.
23 Flow check #1 CGA Fresh Gas Return Point flow < 150 ml/min. 1. Check for an obstruction between the Control Gas Analyzer outlet and R2 at
the AFGO valve.
2. Replace Control Gas Analyzer Outlet Valve EMV16.
3. Replace Control Gas Analyzer AION.
24 Flow check #3 CGA Fresh Gas Return Point flow < 150 ml/min. 1. Check for an obstruction between the Control Gas Analyzer outlet and R2 at
the AFGO valve.
2. Replace Control Gas Analyzer Outlet Valve EMV16.
3. Replace Control Gas Analyzer AION.
25 Flow check #2 CGA Inspiratory Return Point flow < 140 ml/min. 1. Check for an obstruction between the Control Gas Analyzer outlet and gas
return filter R1 in Patient Cassette Docking.
6 - 88
6. TROUBLESHOOTING
66 76 980_02
26 Flow check #1 PGA Expiratory Return Point flow < 80 ml/min. 1. Check for an obstruction between the Patient Gas Analyzer outlet and gas
return filter R3 in Patient Cassette Docking.
2. Replace Patient Gas Analyzer AION.
27 Stable conditions not reached within 5 s during CGA/PGA zero flow pressure During Standby; verify that the Flow value from each gas analyzer is stable at
measurement. 0 ml/min. Note: A zero reference check is performed approximately once per
Stable conditions means that the pressure stays within 1 hPa during the minute and will cause the flow to temporarily rise.
measurement window of 2 s, and that flow stays within 1 ml/min variation
during a measurement window of 2 s. In addition, the flow value must be < 20
ml/min in this stage of the test.
If the value from the Patient Gas Analyzer is implausible or fluctuating:
1. Replace Patient Gas Analyzer AION.
If the value from the Control Gas Analyzer is implausible or fluctuating:
1. Replace Control Gas Analyzer AION.
28 PGA Zero Calibration Gas Inlet pressure drop outside interval of 5 hPa to 20 PGA Zero Calibration Gas Inlet pressure drop < 5 hPa:
hPa
OR
CGA Zero Calibration Gas Inlet pressure drop outside interval of 5 hPa to 20
hPa.
1. Check for leakage in the Patient Gas Analyzer zero calibration line (from
FLOW-i front until the Patient Gas Analyzer Zero calibration gas inlet).
2. Replace Patient Gas Analyzer AION.
PGA Zero Calibration Gas Inlet pressure drop > 20 hPa:
1. Check for for an obstruction in the Patient Gas Analyzer zero calibration line
(from FLOW-i front until the Patient Gas Analyzer Zero calibration gas inlet).
2. Replace Patient Gas Analyzer AION.
6 - 89
6. TROUBLESHOOTING
66 76 980_02
29 Patient O2 Sensor offset outside interval -3 % to 3 %. 1. Perform span calibration according to the instructions in chapter Service
procedures.
2. Replace Patient O2 Sensor.
30 Patient O2 Sensor noise > 6 %. 1. Perform span calibration according to the instructions in chapter Service
procedures.
2. Check/Replace Patient O2 Sensor cable.
3. Replace Patient O2 Sensor.
31 Control O2 Sensor offset outside interval -3 % to 3 %. 1. Calibrate the Control O2 Sensor via Service & Settings / Calibration.
2. Replace Control O2 Sensor.
32 Control O2 Sensor noise > 6 %. 1. Calibrate the Control O2 Sensor via Service & Settings / Calibration.
2. Check/Replace Control O2 Sensor cable.
3. Replace Control O2 Sensor.
33 Metric for CGA not available (invalid metric detected). 1. If several Check failures appear in this test; perform recommended actions
for these failures.
2. Check/replace Control Gas Analyzer AION cable.
6 - 90
34 Metric for PGA not available (invalid metric detected). 1. If several Check failures appear in this test; perform recommended actions
for these failures.
2. Check/replace Patient Gas Analyzer AION cable.
3. Replace Patient Gas Analyzer AION.
4. Replace PC 1921 Monitoring.
35 PGA indicates occlusion in the beginning of the test. 1. Check for an obstruction in the Patient Gas Analyzer sampling line, i.e.
sampling line adapter (at the Y-piece) to the Water trap.
2. Check for an obstruction between the Patient Gas Analyzer inlet and the
Water trap.
3. Replace Patient Gas Analyzer AION.
36 PGA indicates occlusion in the end of the test. 1. Check for an obstruction in the Patient Gas Analyzer sampling line, i.e.
sampling line adapter (at the Y-piece) to the Water trap.
2. Check for an obstruction between the Patient Gas Analyzer inlet and the
Water trap.
3. Replace Patient Gas Analyzer AION.
37 Control O2 Transducer indicates error status "No device". 1. Check/replace the ribbon cable.
2. Replace Control O2 Sensor Pm1116.
3. Replace PC 1920 Control.
38 Control O2 Transducer indicates error status "Internal error". 1. Replace Control O2 Sensor Pm1116.
39 Control O2 Transducer indicates error status "Out of spec.". 1. Perform Control O2 Sensor Pm1116 calibration.
2. Replace Control O2 Sensor Pm1116.
6 - 91
6. TROUBLESHOOTING
66 76 980_02
1 Battery power path check onboard Power Control subsystem failed. 1. Check/replace FUSE F1 and FUSE F2 on PC 1903 External Connectors.
Notice: Indicates a broken battery power path (for instance a non-connected
battery or a burnt fuse).
2. Check cable between the battery and PC 1903 External Connectors.
2 Battery power path check onboard Power Control subsystem has not finished 1. Contact MAQUET HSC via MSupport.
10 s after it was started.
3 Battery connection error indicated by Power Control subsystem during Initial 1. If several Check failures appear in this test; perform recommended actions
battery test. for these failures.
2. Replace PC 1923 Power Control.
4 The internal resistance measurement performed by Power Control subsystem 1. Replace Power backup battery.
indicates that the battery is bad.
5 Failed to read Battery capacity (invalid metric detected). 1. Replace PC 1923 Power Control.
6 Failed to read Battery capacity (invalid metric detected). 1. Replace PC 1923 Power Control.
7 Occurs in conjunction with following Info entries: 1. If several Check failures appear in this test; perform recommended actions
Power Control Battery status = "Invalid", for these failures.
Power Control Battery status = "Weak",
Power Control Battery status = "Empty".
2. Replace PC 1923 Power Control.
6 - 92
6. TROUBLESHOOTING
66 76 980_02
9 This error is persistent and will remain until a corrective action is logged in a 1. Check Power backup battery cables (i.e. Battery and Battery sense cables
Service Report under Service & Settings. The action can be one of: between Power backup battery and PC 1903 External Connectors).
1. Replaced internal battery.
2. Reset internal battery errors.
2. Replace Power backup battery.
10 This error is persistent and will remain until a corrective action is logged in a 1. Check Power backup battery cables (i.e. Battery and Battery sense cables
Service Report under Service & Settings. The action can be one of: between Power backup battery and PC 1903 External Connectors).
1. Replaced internal battery.
2. Reset internal battery errors.
2. Replace Power backup battery.
11 This error is persistent and will remain until a corrective action is logged in a 1. Check Power backup battery cables (i.e. Battery and Battery sense cables
Service Report under Service & Settings. The action can be one of: between Power backup battery and PC 1903 External Connectors).
1. Replaced internal battery.
2. Reset internal battery errors.
2. Replace Power backup battery.
12 This error is persistent and will remain until a corrective action is logged in a 1. Check Power backup battery cables (i.e. Battery and Battery sense cables
Service Report under Service & Settings. The action can be one of: between Power backup battery and PC 1903 External Connectors).
1. Replaced internal battery.
2. Reset internal battery errors.
2. Replace Power backup battery.
6 - 93
6. TROUBLESHOOTING
66 76 980_02
13 This error is persistent and will remain until a corrective action is logged in a 1. Check Power backup battery cables (i.e. Battery and Battery sense cables
Service Report under Service & Settings. The action can be one of: between Power backup battery and PC 1903 External Connectors).
1. Replaced internal battery.
2. Reset internal battery errors.
2. Replace Power backup battery.
14 This error is persistent and will remain until a corrective action is logged in a 1. Check Power backup battery cables (i.e. Battery and Battery sense cables
Service Report under Service & Settings. The action can be one of: between Power backup battery and PC 1903 External Connectors).
1. Replaced internal battery.
2. Reset internal battery errors.
2. Replace Power backup battery.
15 Battery connection error detected (BATTERY_CONNECTION_ERROR_1). 1. Check Power backup battery cables (i.e. Battery and Battery sense cables
between Power backup battery and PC 1903 External Connectors).
2. Check/replace Power backup battery fuses on PC 1903 External
Connectors.
3. Check that the PC 1903 External Connectors is properly connected to the
PC 1900 Main Back-plane.
4. Replace PC 1923 Power Control.
16 Battery connection error detected (BATTERY_CONNECTION_ERROR_2). 1. Check Power backup battery cables (i.e. Battery and Battery sense cables
between Power backup battery and PC 1903 External Connectors).
2. Check/replace Power backup battery fuses on PC 1903 External
Connectors.
3. Check that the PC 1903 External Connectors is properly connected to the
PC 1900 Main Back-plane.
4. Replace PC 1923 Power Control.
17 Battery connection error detected (BATTERY_CONNECTION_ERROR_3). 1. Check Power backup battery cables (i.e. Battery and Battery sense cables
6 - 94
18 Battery connection error detected (BATTERY_CONNECTION_ERROR_4). 1. Check Power backup battery cables (i.e. Battery and Battery sense cables
between Power backup battery and PC 1903 External Connectors).
2. Check/replace Power backup battery fuses on PC 1903 External
Connectors.
3. Check that the PC 1903 External Connectors is properly connected to the
PC 1900 Main Back-plane.
4. Replace PC 1923 Power Control.
19 Battery connection error detected (BATTERY_CONNECTION_ERROR_5). 1. Check Power backup battery cables (i.e. Battery and Battery sense cables
between Power backup battery and PC 1903 External Connectors).
2. Check/replace Power backup battery fuses on PC 1903 External
Connectors.
3. Check that the PC 1903 External Connectors is properly connected to the
PC 1900 Main Back-plane.
4. Replace PC 1923 Power Control.
20 Battery connection error detected (BATTERY_CONNECTION_ERROR_6). 1. Check Power backup battery cables (i.e. Battery and Battery sense cables
between Power backup battery and PC 1903 External Connectors).
2. Check/replace Power backup battery fuses on PC 1903 External
Connectors.
3. Check that the PC 1903 External Connectors is properly connected to the
PC 1900 Main Back-plane.
4. Replace PC 1923 Power Control.
6 - 95
6. TROUBLESHOOTING
66 76 980_02
2 The serial number read from the vaporizer is not equal to the serial number N/A
read from the during the Initial vaporizer test.
Notice: This check failure results in test result "Not performed" for the final
vaporizer test.
6 - 96
6. TROUBLESHOOTING
66 76 980_02
4 VAP disabled at start of Vaporizer initial test. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
5 Open VAP Lid detected at start of Vaporizer initial test. 1. Handling error.
2. Clean surface that reflects the PC 1931 Vaporizer Lid Sensor.
3. Vaporizer service.
6 Open VAP Lid detected during Vaporizer initial test. 1. Handling error.
2. Clean surface that reflects the PC 1931 Vaporizer Lid Sensor.
3. Vaporizer service.
7 VAP liquid level < 5 % (nearly empty) during Vaporizer initial test. 1. Handling error.
2. Vaporizer service.
8 VAP not initialized during Vaporizer initial test. 1. Move Vaporizer 1 to slot 2. If the corresponding failure occurs also in slot 2;
Vaporizer service.
2. Check/replace PC 1909 Vaporizer Adapter.
3. Replace PC 1920 Control.
9 VAP evacuation failed during Vaporizer initial test. 1. Move Vaporizer 1 to slot 2. If the corresponding failure occurs also in slot 2;
Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 2; replace Scavenging valve.
3. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 2; replace Pressure valve.
4. Replace PC 1907 Valve Drivers.
6 - 97
6. TROUBLESHOOTING
66 76 980_02
10 VAP pressure increase detected during Internal VAP test stage in Vaporizer 1. Move Vaporizer 1 to slot 2. If the corresponding failure occurs also in slot 2;
initial test. Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 2; replace Scavenging valve.
3. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 2; replace Pressure valve.
4. Replace PC 1907 Valve Drivers.
11 Internal VAP test not finished in time during Vaporizer initial test. 1. Move Vaporizer 1 to slot 2. If the corresponding failure occurs also in slot 2;
Vaporizer service.
2. Check/replace PC 1909 Vaporizer Adapter.
3. Replace PC 1920 Control.
12 Internal VAP test failed during Vaporizer initial test. 1. Vaporizer service.
13 VAP pressurization failed during Vaporizer initial test. 1. Move Vaporizer 1 to slot 2. If the corresponding failure occurs also in slot 2;
Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 2; replace Pressure valve.
3. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 2; replace Scavenging valve.
4. Check/calibrate Vaporizer Pressure Regulator REG1.
5. Replace PC 1907 Valve Drivers.
14 Removal of docked VAP detected during Preparation stage in Vaporizer 1. Handling error.
parallel leakage test.
15 Removal of docked VAP detected during Measurement stage in Vaporizer 1. Handling error.
parallel leakage test.
6 - 98
6. TROUBLESHOOTING
66 76 980_02
16 Removal of docked VAP detected during Evacuation stage in Vaporizer 1. Handling error.
parallel leakage test.
17 Disabling of VAP detected during Preparation stage in Vaporizer parallel 1. Check for information in the Technical alarms log. For recommended
leakage test. actions, refer to Troubleshooting, section Technical error codes and messages.
18 Disabling of VAP detected during Measurement stage in Vaporizer parallel 1. Check for information in the Technical alarms log. For recommended
leakage test. actions, refer to Troubleshooting, section Technical error codes and messages.
19 Disabling of VAP detected during Evacuation stage in Vaporizer parallel 1. Check for information in the Technical alarms log. For recommended
leakage test. actions, refer to Troubleshooting, section Technical error codes and messages.
20 Opening of VAP Lid detected during Preparation stage in Vaporizer parallel 1. Handling error.
leakage test.
Notice: This check failure results in test result "Not performed" for the final
vaporizer test.
2. Clean surface that reflects the PC 1931 Vaporizer Lid Sensor.
3. Vaporizer service.
21 Opening of VAP Lid detected during Measurement stage in Vaporizer parallel 1. Handling error.
leakage test.
Notice: This check failure results in test result "Not performed" for the final
vaporizer test.
2. Clean surface that reflects the PC 1931 Vaporizer Lid Sensor.
3. Vaporizer service.
22 Opening of VAP Lid detected during Evacuation stage in Vaporizer parallel 1. Handling error.
leakage test.
Notice: This check failure results in test result "Not performed" for the final
6 - 99
vaporizer test.
6. TROUBLESHOOTING
66 76 980_02
23 Vaporizer parallel leakage test, Preparation stage VAP pressure outside 1. Move Vaporizer 1 to slot 2. If the corresponding failure occurs also in slot 2;
interval 500 hPa to 1700 hPa. Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 2; replace Pressure valve.
3. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 2; replace Scavenging valve.
4. Check/calibrate Vaporizer Pressure Regulator REG1.
5. Replace PC 1907 Valve Drivers.
24 Vaporizer parallel leakage test,Measurement stage VAP pressure outside 1. Move Vaporizer 1 to slot 2. If the corresponding failure occurs also in slot 2;
interval 500 hPa to 1700 hPa. Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 2; replace Pressure valve.
3. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 2; replace Scavenging valve.
4. Check/calibrate Vaporizer Pressure Regulator REG1.
5. Replace PC 1907 Valve Drivers.
25 Vaporizer parallel leakage test, Measurement stage VAP Liquid System 1. Move Vaporizer 1 to slot 2. If the corresponding failure occurs also in slot 2;
leakage > 20 ml/min. Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 2; replace Scavenging valve.
26 Vaporizer parallel leakage test, Measurement stage VAP Liquid System 1. Move Vaporizer 1 to slot 2. If the corresponding failure occurs also in slot 2;
leakage < -15 ml/min. Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 2; replace Pressure valve.
6 - 100
6. TROUBLESHOOTING
66 76 980_02
27 Vaporizer parallel leakage test, Evacuation stage VAP pressure above 700 1. Move Vaporizer 1 to slot 2. If the corresponding failure occurs also in slot 2;
hPa for Desflurane or 200 hPa for other anesthetic agents after 10 s Vaporizer service.
evacuation attempt.
2. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 2; replace Scavenging valve.
3. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 2; replace Pressure valve.
4. Replace PC 1907 Valve Drivers.
28 Vaporizer parallel leakage test failed. 1. Perform recommended actions for other failures in this test.
29 The Vaporizer parallel leakage test for the vaporizer did not finish within 160 s. 1. Contact MAQUET HSC via MSupport.
30 The Vaporizer parallel leakage test for the other vaporizer did not finished 1. Contact MAQUET HSC via MSupport.
within 160 s.
32 A docked vaporizer was detected during Initial vaporizer test 1. Handling error.
AND
No docked vaporizer was detected during Final vaporizer test.
33 Vaporizer not disabled during Initial vaporizer test 1. Check for information in the Technical alarms log. For recommended
AND actions, refer to Troubleshooting, section Technical error codes and messages.
Disabled vaporizer detected during Final vaporizer test.
6 - 101
6. TROUBLESHOOTING
66 76 980_02
36 AUTO ventilation leakage test result ≠ PASSED 1. Perform recommended actions for the failures in the AUTO and/or MAN
OR ventilation leakage test.
MAN ventilation leakage test result ≠ PASSED.
37 AUTO or MAN ventilation leakage test not performed when Parallel vaporizer 1. Handling error.
leakage test has finished.
38 Central Gas Supply AIR pressure < 1800 cmH2O 1. Check central AIR and O2 gas supply.
OR
Central Gas Supply O2 pressure < 1800 cmH2O.
2. Replace PC 1906 Gas Block.
39 VAP flow limitation active after 50 s wait during VAP Safety Valve check. 1. Vaporizer service.
40 CGA inactive after 60 s wait during VAP Safety Valve check. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
6 - 102
6. TROUBLESHOOTING
66 76 980_02
41 CGA error detected during Initial stage. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
42 CGA error detected during VAP Safety Valve opened stage #1. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
43 CGA error detected during VAP Safety Valve opened stage #2. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
44 CGA error detected during VAP Safety Valve closed stage. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
45 VAP Safety Valve opened stage #1, AA concentration outside interval 3 % to 5 1. Check fresh gas sampling (double sampling).
%.
2. Vaporizer service.
3. Replace Control Gas Analyzer AION.
46 VAP Safety Valve closed stage, AA concentration not below 0.5 % (other 1. Vaporizer service.
anesthetic agent than Desflurane used).
2. Replace Control Gas Analyzer AION.
47 VAP Safety Valve closed stage, AA concentration not below 2 % (Desflurane 1. Vaporizer service.
was used).
2. Replace Control Gas Analyzer AION.
48 VAP Safety Valve opened stage #2, AA concentration outside interval 3 % to 5 1. Check fresh gas sampling (double sampling).
%.
2. Vaporizer service.
6 - 103
50 Vaporizer pressure has not reached inside the interval 1000 to 1400 hPa within 1. Move Vaporizer 1 to slot 2. If the corresponding failure occurs also in slot 2;
9 s during the first pressurization in the test. Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 2; replace Scavenging valve.
3. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 2; replace Pressure valve.
4. Replace PC 1907 Valve Drivers.
51 Vaporizer pressure has not been continuously inside the interval 1000 to 1400 1. Move Vaporizer 1 to slot 2. If the corresponding failure occurs also in slot 2;
hPa for 1 s within a total period of 10 s during the first pressurization in the test. Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 2; replace Scavenging valve.
3. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 2; replace Pressure valve.
4. Replace PC 1907 Valve Drivers.
52 Vaporizer pressure has not reached inside the interval 1000 to 1400 hPa within 1. Move Vaporizer 1 to slot 2. If the corresponding failure occurs also in slot 2;
9 s during the second pressurization in the test. Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 2; replace Scavenging valve.
3. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 2; replace Pressure valve.
4. Replace PC 1907 Valve Drivers.
6 - 104
6. TROUBLESHOOTING
66 76 980_02
54 Vaporizer pressure has not reached below 700 hPa for Desflurane (200 hPa 1. Move Vaporizer 1 to slot 2. If the corresponding failure occurs also in slot 2;
for other anesthetic agents) within 5 s during the first evacuation in the test. Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 2; replace Scavenging valve.
3. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 2; replace Pressure valve.
4. Replace PC 1907 Valve Drivers.
55 Vaporizer pressure has not been continuously below 700 hPa for Desflurane 1. Move Vaporizer 1 to slot 2. If the corresponding failure occurs also in slot 2;
(200 hPa for other anesthetic agents) for 5 s within a total period of 10 s during Vaporizer service.
the first evacuation in the test.
2. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 2; replace Scavenging valve.
3. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 2; replace Pressure valve.
4. Replace PC 1907 Valve Drivers.
56 Vaporizer pressure has not reached below 700 hPa for Desflurane (200 hPa 1. Move Vaporizer 1 to slot 2. If the corresponding failure occurs also in slot 2;
for other anesthetic agents) for 5 s within a total period of 10 s during the Vaporizer service.
second evacuation in the test.
2. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 2; replace Scavenging valve.
6 - 105
6. TROUBLESHOOTING
66 76 980_02
57 Vaporizer pressure has not been continuously below 700 hPa for Desflurane 1. Move Vaporizer 1 to slot 2. If the corresponding failure occurs also in slot 2;
(200 hPa for other anesthetic agents) for 5 s within a total period of 10 s during Vaporizer service.
the second evacuation in the test.
2. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 2; replace Scavenging valve.
3. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 2; replace Pressure valve.
4. Replace PC 1907 Valve Drivers.
58 Vaporizer reported error during Initial stage. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
59 Vaporizer reported error during VAP Safety Valve opened stage #1. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
60 Vaporizer reported error during VAP Safety Valve opened stage #2. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
61 Vaporizer reported error during VAP Safety Valve closed stage. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
62 Disabling of vaporizer detected during Initial stage. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
63 Disabling of vaporizer detected during first pressurization. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
6 - 106
6. TROUBLESHOOTING
66 76 980_02
64 Disabling of vaporizer detected during VAP Safety Valve opened stage #1. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
65 Disabling of vaporizer detected during Evacuation stage #1. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
66 Disabling of vaporizer detected during second pressurization. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
67 Disabling of vaporizer detected during VAP Safety Valve closed stage. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
68 Disabling of vaporizer detected during VAP Safety Valve opened stage #2. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
69 Disabling of vaporizer detected during Evacuation stage #2. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
70 VAP Injection Detector check failed (No spray detected during VAP Safety 1. Vaporizer service.
Valve Open check).
71 VAP Safety Valve check and VAP Injection Detector check not performed 1. Perform recommended actions for other failures in this test.
(Vaporizer initial test not completed).
72 VAP Safety Valve check and VAP Injection Detector check not performed 1. Perform recommended actions for other failures in this test.
(Vaporizer parallel leakage test not completed).
6 - 107
6. TROUBLESHOOTING
66 76 980_02
2 The serial number read from the vaporizer is not equal to the serial number N/A
read from the during the Initial vaporizer test.
Notice: This check failure results in test result "Not performed" for the final
vaporizer test.
3 VAP disabled at start of Vaporizer initial test. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
4 VAP disabled at start of Vaporizer initial test. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
5 Open VAP Lid detected at start of Vaporizer initial test. 1. Handling error.
2. Clean surface that reflects the PC 1931 Vaporizer Lid Sensor.
3. Vaporizer service.
6 Open VAP Lid detected during Vaporizer initial test. 1. Handling error.
2. Clean surface that reflects the PC 1931 Vaporizer Lid Sensor.
3. Vaporizer service.
7 VAP liquid level < 5 % (nearly empty) during Vaporizer initial test. 1. Handling error.
2. Vaporizer service.
8 VAP not initialized during Vaporizer initial test. 1. Move Vaporizer 2 to slot 1. If the corresponding failure occurs also in slot 1;
6 - 108
Vaporizer service.
6. TROUBLESHOOTING
66 76 980_02
9 VAP evacuation failed during Vaporizer initial test. 1. Move Vaporizer 2 to slot 1. If the corresponding failure occurs also in slot 1;
Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 1; replace Scavenging valve.
3. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 1; replace Pressure valve.
4. Replace PC 1907 Valve Drivers.
10 VAP pressure increase detected during Internal VAP test stage in Vaporizer 1. Move Vaporizer 2 to slot 1. If the corresponding failure occurs also in slot 1;
initial test. Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 1; replace Scavenging valve.
3. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 1; replace Pressure valve.
4. Replace PC 1907 Valve Drivers.
11 Internal VAP test not finished in time during Vaporizer initial test. 1. Move Vaporizer 2 to slot 1. If the corresponding failure occurs also in slot 1;
Vaporizer service.
2. Check/replace PC 1909 Vaporizer Adapter.
3. Replace PC 1920 Control.
12 Internal VAP test failed during Vaporizer initial test. 1. Vaporizer service.
13 VAP pressurization failed during Vaporizer initial test. 1. Move Vaporizer 2 to slot 1. If the corresponding failure occurs also in slot 1;
Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
6 - 109
14 Removal of docked VAP detected during Preparation stage in Vaporizer 1. Handling error.
parallel leakage test.
15 Removal of docked VAP detected during Measurement stage in Vaporizer 1. Handling error.
parallel leakage test.
16 Removal of docked VAP detected during Evacuation stage in Vaporizer 1. Handling error.
parallel leakage test.
17 Disabling of VAP detected during Preparation stage in Vaporizer parallel 1. Check for information in the Technical alarms log. For recommended
leakage test. actions, refer to Troubleshooting, section Technical error codes and messages.
18 Disabling of VAP detected during Measurement stage in Vaporizer parallel 1. Check for information in the Technical alarms log. For recommended
leakage test. actions, refer to Troubleshooting, section Technical error codes and messages.
19 Disabling of VAP detected during Evacuation stage in Vaporizer parallel 1. Check for information in the Technical alarms log. For recommended
leakage test. actions, refer to Troubleshooting, section Technical error codes and messages.
20 Opening of VAP Lid detected during Preparation stage in Vaporizer parallel 1. Handling error.
leakage test.
Notice: This check failure results in test result "Not performed" for the final
vaporizer test.
6 - 110
21 Opening of VAP Lid detected during Measurement stage in Vaporizer parallel 1. Handling error.
leakage test.
Notice: This check failure results in test result "Not performed" for the final
vaporizer test.
2. Clean surface that reflects the PC 1931 Vaporizer Lid Sensor.
3. Vaporizer service.
22 Opening of VAP Lid detected during Evacuation stage in Vaporizer parallel 1. Handling error.
leakage test.
Notice: This check failure results in test result "Not performed" for the final
vaporizer test.
2. Clean surface that reflects the PC 1931 Vaporizer Lid Sensor.
3. Vaporizer service.
23 Vaporizer parallel leakage test, Preparation stage VAP pressure outside 1. Move Vaporizer 2 to slot 1. If the corresponding failure occurs also in slot 1;
interval 500 hPa to 1700 hPa. Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 1; replace Pressure valve.
3. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 1; replace Scavenging valve.
4. Check/calibrate Vaporizer Pressure Regulator REG1.
5. Replace PC 1907 Valve Drivers.
24 Vaporizer parallel leakage test,Measurement stage VAP pressure outside 1. Move Vaporizer 2 to slot 1. If the corresponding failure occurs also in slot 1;
interval 500 hPa to 1700 hPa. Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 1; replace Pressure valve.
6 - 111
6. TROUBLESHOOTING
66 76 980_02
25 Vaporizer parallel leakage test, Measurement stage VAP Liquid System 1. Move Vaporizer 2 to slot 1. If the corresponding failure occurs also in slot 1;
leakage > 20 ml/min. Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 1; replace Scavenging valve.
26 Vaporizer parallel leakage test, Measurement stage VAP Liquid System 1. Move Vaporizer 2 to slot 1. If the corresponding failure occurs also in slot 1;
leakage < -15 ml/min. Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 1; replace Pressure valve.
27 Vaporizer parallel leakage test, Evacuation stage VAP pressure above 700 1. Move Vaporizer 2 to slot 1. If the corresponding failure occurs also in slot 1;
hPa for Desflurane or 200 hPa for other anesthetic agents after 10 s Vaporizer service.
evacuation attempts.
2. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 1; replace Scavenging valve.
3. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 1; replace Pressure valve.
4. Replace PC 1907 Valve Drivers.
28 Vaporizer parallel leakage test failed. 1. Perform recommended actions for other failures in this test.
29 The Vaporizer parallel leakage test for the vaporizer did not finish within 160 s. 1. Contact MAQUET HSC via MSupport.
30 The Vaporizer parallel leakage test for the other vaporizer did not finished 1. Contact MAQUET HSC via MSupport.
within 160 s.
6 - 112
6. TROUBLESHOOTING
66 76 980_02
32 A docked vaporizer was detected during Initial vaporizer test 1. Handling error.
AND
No docked vaporizer was detected during Final vaporizer test.
33 Vaporizer not disabled during Initial vaporizer test 1. Check for information in the Technical alarms log. For recommended
AND actions, refer to Troubleshooting, section Technical error codes and messages.
Disabled vaporizer detected during Final vaporizer test.
36 AUTO ventilation leakage test result ≠ PASSED 1. Perform recommended actions for the failures in the AUTO and/or MAN
OR ventilation leakage test.
MAN ventilation leakage test result ≠ PASSED.
6 - 113
6. TROUBLESHOOTING
66 76 980_02
38 Central Gas Supply AIR pressure < 1800 cmH2O 1. Check central AIR and O2 gas supply.
OR
Central Gas Supply O2 pressure < 1800 cmH2O.
2. Replace PC 1906 Gas Block.
39 VAP flow limitation active after 50 s wait during VAP Safety Valve check. 1. Vaporizer service.
40 CGA inactive after 60 s wait during VAP Safety Valve check. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
41 CGA error detected during Initial stage. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
42 CGA error detected during VAP Safety Valve opened stage #1. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
43 CGA error detected during VAP Safety Valve opened stage #2. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
44 CGA error detected during VAP Safety Valve closed stage. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
45 VAP Safety Valve opened stage #1, AA concentration outside interval 3 % to 5 1. Check fresh gas sampling (double sampling).
%.
2. Vaporizer service.
3. Replace Control Gas Analyzer AION.
6 - 114
6. TROUBLESHOOTING
66 76 980_02
47 VAP Safety Valve closed stage, AA concentration not below 2 % (Desflurane 1. Vaporizer service.
was used).
2. Replace Control Gas Analyzer AION.
48 VAP Safety Valve opened stage #2, AA concentration outside interval 3 % to 5 1. Check fresh gas sampling (double sampling).
%.
2. Vaporizer service.
3. Replace Control Gas Analyzer AION.
50 Vaporizer pressure has not reached inside the interval 1000 to 1400 hPa within 1. Move Vaporizer 2 to slot 1. If the corresponding failure occurs also in slot 1;
9 s during the first pressurization in the test. Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 1; replace Scavenging valve.
3. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 1; replace Pressure valve.
4. Replace PC 1907 Valve Drivers.
51 Vaporizer pressure has not been continuously inside the interval 1000 to 1400 1. Move Vaporizer 2 to slot 1. If the corresponding failure occurs also in slot 1;
hPa for 1 s within a total period of 10 s during the first pressurization in the test. Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 1; replace Scavenging valve.
3. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
6 - 115
52 Vaporizer pressure has not reached inside the interval 1000 to 1400 hPa within 1. Move Vaporizer 2 to slot 1. If the corresponding failure occurs also in slot 1;
9 s during the second pressurization in the test. Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 1; replace Scavenging valve.
3. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 1; replace Pressure valve.
4. Replace PC 1907 Valve Drivers.
53 Vaporizer pressure has not been continuously inside the interval 1000 to 1400 1. Move Vaporizer 2 to slot 1. If the corresponding failure occurs also in slot 1;
hPa for 1 s within a total period of 10 s during the second pressurization in the Vaporizer service.
test.
2. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 1; replace Scavenging valve.
3. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 1; replace Pressure valve.
4. Replace PC 1907 Valve Drivers.
54 Vaporizer pressure has not reached below 700 hPa for Desflurane (200 hPa 1. Move Vaporizer 2 to slot 1. If the corresponding failure occurs also in slot 1;
for other anesthetic agents) within 5 s during the first evacuation in the test. Vaporizer service.
2. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 1; replace Scavenging valve.
3. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 1; replace Pressure valve.
4. Replace PC 1907 Valve Drivers.
55 Vaporizer pressure has not been continuously below 700 hPa for Desflurane 1. Move Vaporizer 2 to slot 1. If the corresponding failure occurs also in slot 1;
(200 hPa for other anesthetic agents) for 5 s within a total period of 10 s during Vaporizer service.
the first evacuation in the test.
6 - 116
6. TROUBLESHOOTING
66 76 980_02
56 Vaporizer pressure has not reached below 700 hPa for Desflurane (200 hPa 1. Move Vaporizer 2 to slot 1. If the corresponding failure occurs also in slot 1;
for other anesthetic agents) for 5 s within a total period of 10 s during the Vaporizer service.
second evacuation in the test.
2. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 1; replace Scavenging valve.
3. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 1; replace Pressure valve.
4. Replace PC 1907 Valve Drivers.
57 Vaporizer pressure has not been continuously below 700 hPa for Desflurane 1. Move Vaporizer 2 to slot 1. If the corresponding failure occurs also in slot 1;
(200 hPa for other anesthetic agents) for 5 s within a total period of 10 s during Vaporizer service.
the second evacuation in the test.
2. Swap Vaporizer 1 and Vaporizer 2 Scavenging Valves. If the corresponding
failure occurs also in slot 1; replace Scavenging valve.
3. Swap Vaporizer 1 and Vaporizer 2 Pressure Valves. If the corresponding
failure occurs also in slot 1; replace Pressure valve.
4. Replace PC 1907 Valve Drivers.
58 Vaporizer reported error during Initial stage. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
59 Vaporizer reported error during VAP Safety Valve opened stage #1. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
6 - 117
6. TROUBLESHOOTING
66 76 980_02
61 Vaporizer reported error during VAP Safety Valve closed stage. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
62 Disabling of vaporizer detected during Initial stage. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
63 Disabling of vaporizer detected during first pressurization. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
64 Disabling of vaporizer detected during VAP Safety Valve opened stage #1. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
65 Disabling of vaporizer detected during Evacuation stage #1. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
66 Disabling of vaporizer detected during second pressurization. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
67 Disabling of vaporizer detected during VAP Safety Valve closed stage. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
68 Disabling of vaporizer detected during VAP Safety Valve opened stage #2. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
69 Disabling of vaporizer detected during Evacuation stage #2. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
6 - 118
6. TROUBLESHOOTING
66 76 980_02
71 VAP Safety Valve check and VAP Injection Detector check not performed 1. Perform recommended actions for other failures in this test.
(Vaporizer initial test not completed).
72 VAP Safety Valve check and VAP Injection Detector check not performed 1. Perform recommended actions for other failures in this test.
(Vaporizer parallel leakage test not completed).
1 One or several technical alarm(s) active. 1. Check for information in the Technical alarms log. For recommended
actions, refer to Troubleshooting, section Technical error codes and messages.
6 - 119
6. TROUBLESHOOTING
Initial troubleshooting
In case of Technical alarms, always start troubleshooting with a System Checkout:
Switch off the system. This will disable active Technical alarms that may interfere with the
System Checkout.
Startup the system and run a full System Checkout. The System Checkout can correct some
errors and thus remove the cause of the Technical alarm. If the Technical alarm remains,
proceed with troubleshooting according to the table below.
66 76 980_02 6 - 120
6. TROUBLESHOOTING
66 76 980_02
2 Power failure The -12 V supply voltage measured by the A/D converter on PC 1921 Monitoring 1. Replace PC 1923 Power control.
is too high. The alarm is activated if the voltage is above limit for at least 5 2. Replace PC 1921 Monitoring.
seconds. Limit: -10.8 V.
Error in the ±12 V converter on PC 1923 Power control or A/D converter error on
PC 1921 Monitoring.
3 Power failure The +12 V supply voltage measured by the A/D converter on PC 1921 Monitoring 1. Replace PC 1923 Power control.
is too low. The alarm is activated if the voltage is below limit for at least 5 2. Replace PC 1921 Monitoring.
seconds. Limit: 10.8 V.
Error in the ±12 V converter on PC 1923 Power control or A/D converter error on
PC 1921 Monitoring.
4 Power failure The +12 V supply voltage measured by the A/D converter on PC 1921 Monitoring 1. Replace PC 1923 Power control.
is too high. The alarm is activated if the voltage is above limit for at least 5 2. Replace PC 1921 Monitoring.
seconds. Limit: 13.2 V.
Error in the ±12 V converter on PC 1923 Power control or A/D converter error on
PC 1921 Monitoring.
5 Power failure Power control has detected an error in the power supply. Recommended action for AC_WEAK:
AC_WEAK means that the analog control voltage from PC 1923 Power control to 1. Check cables between PC1923 Power control
the AC/DC Converter, has reached its maximum allowable value (5 V). This and AC/DC Converter.
occurs when the output voltage from the AC/DC Converter does not respond 2. Replace AC/DC Converter.
correctly to the control voltage, i.e. the AC/DC output voltage does not increase
when the control voltage is increased. 3. Replace PC 1923 Power control.
AC_TOO_HIGH means that the voltage from the AC/DC Converter is too high.
Recommended action for AC_TOO_HIGH:
Additional information: 1. Replace AC/DC Converter.
1 = AC_WEAK 2. Replace PC 1923 Power control.
6 - 121
2 = AC_TOO_HIGH
6. TROUBLESHOOTING
66 76 980_02
6 Power failure Error in PC 1923 Power control CPU Flash memory (failed to erase, write or 1. Replace PC 1923 Power control.
format memory area).
Additional information:
1 = FLASH_REDUCED
2 = FLASH_WRITE
3 = FLASH_NOT_FORMATTED
7 Ventilation control error Resolution for any Gas Module (digital signal from Gas Module) mismatches with 1. Replace PC 1920 Control.
reported set resolution from Control node. The criterion is only checked if gas 2. Replace PC 1921 Monitoring.
supply conditions are correct. 3. Replace Gas Module.
Additional information:
In case of Gas Module error, the faulty Gas Module (Air / O2 / N2O / O2 Ref) will
be identified in the Technical alarms log.
8 Ventilation control error Signal corresponding to alarm name (VALVES_DISABLED) is continuously active 1. Replace PC 1919 Exp. channel.
for at least 5.5 seconds during normal operation conditions, e.g. No high airway 2. Replace PC 1920 Control.
press situation. 3. Replace PC 1938 Mains connections.
9 Safety Valve open Signal corresponding to alarm name (SAFETY_VALVE_OPEN) is continuously 1. Check connection to or replace Safety Valve.
active for 5.5 ±2 seconds during normal operation conditions, e.g. No high airway 2. Replace PC 1920 Control.
press situation and signal DISABLE_VALVES.L not active.
1. Safety Valve error.
2. SW control error.
10 Ventilation control error Internal CAN communication error. Communication with control CAN node fails. 1. Replace PC 1920 Control.
Alarm activated by Monitoring node if the Control node connection is lost.
Connection timeout: 10-13 seconds. The timeout is chosen to prevent a sub
system reboot from generating a technical alarm.
11 Re-specify alarm limits Monitoring memory error. Data stored in Monitoring node persistent memory is 1. Replace memory backup battery if also TE 20
lost or corrupted. Checks performed on a periodic basis. The alarms can occur if is activated.
Monitoring memory backup battery is removed or at system SW upgrade. 2. Restore factory default in S&S.
3. Replace PC 1921 Monitoring.
12 Ventilation control error Internal CAN communication error. Communication with Exp channel CAN node 1. Replace PC 1919 Exp. channel.
fails. Alarm activated by Monitoring node if the Exp. channel node connection is
lost. Connection timeout: 10-13 seconds. The timeout is chosen to prevent a sub
6 - 122
14 Audible backup alarm Buzzer error (buzzer on PC 1921 Monitoring). 1. Replace PC 1921 Monitoring.
error
15 Audible backup alarm Backup alarm capacitor error (capacitor on PC 1921). Backup alarm capacitor 1. Replace PC 1921 Monitoring.
error test at system startup fails. The warm cold start logic will not work.
Additional information:
0x0 = OK
0x01 = charge level to low
0x02 = charge level to high
0x04 = discharged too fast
0x08 = not discharging
0x10 = discharging during charging
0x20 = 5Vc diode broken
16 Communication error Internal communication error. Activated by the Monitoring node if its I2C 1. If the alarm appears after a System software
communication fails. The TE code is combined with a Device ID and set of installation, restart the system.
different error codes specifying in detail which type of error that occurred. The 2. Replace defective part according to Device ID
details must be retrieved via the technical log.The alarm indicates a permanent stated in the Technical alarms log. Refer to
failure. separate table for TE 16/TE 502/TE 601.
This alarm can however also be activated temporarily after a System software
installation.
Additional information:
Device ID on defective part will be stated in the Technical alarms log.
Refer to separate table for TE 16/TE 502/TE 601
17 Fan malfunction Fan error. The fan rotation speed is not correct. 1. Check if anything obstructs the fan.
2. Replace the fan.
Additional information: 3. Replace PC 1923 Power control.
The actual fan rotation speed when the error occurred is available in the
Technical alarms log.
18 Audible backup alarm Buzzer error (buzzer on PC 1921 Monitoring). Buzzer test at system startup fails. 1. Replace PC 1921 Monitoring.
error
20 Memory backup battery Memory backup battery voltage level too low on PC 1921 Monitoring. 1. Replace memory backup battery on PC 1921
depleted Limit: 3.0 ±0.1 V. Monitoring.
6 - 123
6. TROUBLESHOOTING
66 76 980_02
21 Ventilation error Internal CAN communication error. Communication with control CAN node fails. 1. Replace PC 1920 Control.
Alarm activated by Monitoring node if the node Control fails to connect.
Connection timeout: 10 seconds.
22 Control panel error Internal CAN communication error. Communication with panel CAN node fails. 1. Replace PC 1922 Panel.
Alarm activated by Monitoring node if the node Panel fails to connect. Connection
timeout: 45 seconds
23 Ventilation error Internal CAN communication error. Communication with Exp Channel CAN node 1. Replace PC 1919 Exp. channel.
fails. Alarm activated by Monitoring node if the node Exp. flow fails to connect.
Connection timeout: 10 seconds.
24 Ventilation control error Received range from Exp flow meter node does not match the set patient range 1. Replace PC 1920 Control.
received from Control node for duration of 5.5 seconds. Check is not done if 2. Replace PC 1919 Exp. channel.
VALVES_DISABLED is activated.
25 Barometric sensor error Alarm is activated if measured barometric pressure is above 1070 mbar. 1. Calibrate the barometer.
2. Replace PC 1921 Monitoring.
26 Barometric sensor error Alarm is activated if measured barometric pressure is below 690 mbar. 1. Calibrate the barometer.
2. Replace PC 1921 Monitoring.
27 APL pressure exceeded Measured airway pressure in manual mode is above: 1. Perform a System Checkout.
20 cmH2O for set APL pressure [SP, 12] cmH2O
set APL pressure + 7 cmH2O for set APL pressure [13, 49] cmH2O
set APL pressure + 10 cmH2O for set APL pressure (50, 80] cmH2O
28 System date and time Real time clock circuit error. Clock circuit cannot hold valid data or contains 1. Restart system.
error invalid data. 2. Replace PC 1921Monitoring.
Additional information:
0 = set date
1 = read RTC
2 = valid date time
29 Monitoring error Monitoring memory error. Data stored in Monitoring node persistent memory is 1. Restart system
lost or corrupted. Checks performed at startup. The alarms can occur if 2. If error occur in conjunction with SW update,
Monitoring memory backup battery is removed or at system SW upgrade. reinstall SW.
3. Replace PC 1921 Monitoring (if not memory
backup battery low).
6 - 124
6. TROUBLESHOOTING
66 76 980_02
30 Alarm limit error, check Monitoring memory error. Mismatch in alarm limit data. Checks performed on 1. Replace PC 1921 Monitoring.
alarm profile periodic basis.
31 Communication error Monitoring checks that PC 1900 Main back-plane contains valid data, i.e. 1. Replace PC 1900 Main back-plane (likely).
correctly configured in production and when installing new options. Main back- 2. Replace PC 1921 Monitoring (easiest).
plane memory contains valid serial number of the system and installed options.
This error code is most likely caused by a
malfunction on PC 1900 Main back-plane.
However, as PC 1900 is time consuming to
replace, it may be easier to start with
replacement of PC 1921 Monitoring.
32 Power failure The supply voltage +12V_UNREG_CORE, measured by the Monitor via A/D 1. Replace PC 1923 Power control.
converter on PC 1923 Power control, is too low. The alarm is activated if the
voltage is below limit for at least 5 seconds. Limit: 10.0 V.
33 Power failure The supply voltage +12V_UNREG_CORE, measured by the Monitor via A/D 1. Replace PC 1923 Power control.
converter on PC 1923 Power control, is too high. The alarm is activated if the 2. Replace AC/DC Converter.
voltage is above limit for at least 5 seconds. Limit: 16.0 V.
34 Power failure The supply voltage +24V_INSP_VALVE_1_2, measured by the Monitor via A/D 1. Replace PC 1923 Power control.
converter on PC 1923 Power control, is too low. The alarm is activated if the 2. Replace Gas Module 1 or 2 (i.e. Reflector Gas
voltage is below limit for at least 5 seconds. Limit: 21.6 V. Module or Fresh Gas Module O2).
Additional information:
Insp Valve 1 = Reflector Gas Module
Insp Valve 2 = Fresh Gas Module O2
35 Power failure The supply voltage +24V_INSP_VALVE_1_2, measured by the Monitor via A/D 1. Replace PC 1923 Power control.
converter on PC 1923 Power control, is too high. The alarm is activated if the
voltage is above limit for at least 5 seconds. Limit: 26.4 V.
Additional information:
Insp Valve 1 = Reflector Gas Module
Insp Valve 2 = Fresh Gas Module O2
6 - 125
6. TROUBLESHOOTING
66 76 980_02
36 Power failure The supply voltage +24V_INSP_VALVE_3_4, measured by the Monitor via A/D 1. Replace PC 1923 Power control.
converter on PC 1923 Power control, is too low. The alarm is activated if the 2. Replace Gas Module 3 or 4 (i.e. Fresh Gas
voltage is below limit for at least 5 seconds. Limit: 21.6 V. Module AIR or Fresh Gas Module N2O).
Additional information:
Insp Valve 3 = Fresh Gas Module AIR
Insp Valve 4 = Fresh Gas Module N2O
37 Power failure The supply voltage +24V_INSP_VALVE_3_4, measured by the Monitor via A/D 1. Replace PC 1923 Power control.
converter on PC 1923 Power control, is too high. The alarm is activated if the
voltage is above limit for at least 5 seconds. Limit: 26.4 V.
Additional information:
Insp Valve 3 = Fresh Gas Module AIR
Insp Valve 4 = Fresh Gas Module N2O
38 Power failure The supply voltage +24V_ANALYZER_CON, measured by the Monitor via A/D 1. Replace PC 1923 Power control.
converter on PC 1923 Power control, is too low. The alarm is activated if the 2. Replace Control Gas Analyzer AION (CON).
voltage is below limit for at least 5 seconds. Limit: 21.6 V.
39 Power failure The supply voltage +24V_ANALYZER_CON, measured by the Monitor via A/D 1. Replace PC 1923 Power control.
converter on PC 1923 Power control, is too high. The alarm is activated if the
voltage is above limit for at least 5 seconds. Limit: 26.4 V.
40 Power failure The supply voltage +24V_ANALYZER_MON, measured by the Monitor via A/D 1. Replace PC 1923 Power control.
converter on PC 1923 Power control, is too low. The alarm is activated if the 2. Replace Patient Gas Analyzer AION (MON).
voltage is below limit for at least 5 seconds. Limit: 21.6 V.
41 Power failure The supply voltage +24V_ANALYZER_MON, measured by the Monitor via A/D 1. Replace PC 1923 Power control.
converter on PC 1923 Power control, is too high. The alarm is activated if the
voltage is above limit for at least 5 seconds. Limit: 26.4 V.
42 Power failure The supply voltage +12V_UNREG_VAP_1, measured by the Monitor via A/D 1. Replace PC 1923 Power control.
converter on PC 1923 Power control, is too low. The alarm is activated if the 2. Remove/replace the Vaporizer that is
voltage is below limit for at least 5 seconds. Limit: 10.0 V. mounted in Vaporizer slot 1.
Additional information:
Applicable only when a Vaporizer is docked in slot 1.
6 - 126
6. TROUBLESHOOTING
66 76 980_02
43 Power failure The supply voltage +12V_UNREG_VAP_1, measured by the Monitor via A/D 1. Replace PC 1923 Power control.
converter on PC 1923 Power control, is too high. The alarm is activated if the 2. Replace AC/DC Converter.
voltage is above limit for at least 5 seconds. Limit: 16.0 V.
Additional information:
Applicable only when a Vaporizer is docked in slot 1.
44 Power failure The supply voltage +12V_UNREG_VAP_2, measured by the Monitor via A/D 1. Replace PC 1923 Power control.
converter on PC 1923 Power control, is too low. The alarm is activated if the 2. Remove/replace the Vaporizer that is
voltage is below limit for at least 5 seconds. Limit: 10.0 V. mounted in Vaporizer slot 2.
Additional information:
Applicable only when a Vaporizer is docked in slot 2.
45 Power failure The supply voltage +12V_UNREG_VAP_2, measured by the Monitor via A/D 1. Replace PC 1923 Power control.
converter on PC 1923 Power control, is too high. The alarm is activated if the 2. Replace AC/DC Converter.
voltage is above limit for at least 5 seconds. Limit: 16.0 V.
Additional information:
Applicable only when a Vaporizer is docked in slot 2.
46 Power failure The supply voltage +12V_MINI_VALVES, measured by the Monitor via A/D 1. Replace PC 1923 Power control.
converter on PC 1923 Power control, is too low. The alarm is activated if the 2. Check/replace valves connected to PC 1907
voltage is below limit for at least 5 seconds. Limit: 10.8 V. Valve drivers.
3. Replace PC 1907 Valve drivers.
47 Power failure The supply voltage +12V_MINI_VALVES, measured by the Monitor via A/D 1. Replace PC 1923 Power control.
converter on PC 1923 Power control, is too high. The alarm is activated if the
voltage is above limit for at least 5 seconds. Limit: 13.2 V.
48 Power failure The supply voltage +12V_UNREG_DISPLAY, measured by the Monitor via A/D 1. Replace PC 1923 Power control.
converter on PC 1923 Power control, is too low. The alarm is activated if the 2. Replace Control panel.
voltage is below limit for at least 5 seconds. Limit: 10.0 V. 3. Replace PC 1922 Panel.
49 Power failure The supply voltage +12V_UNREG_DISPLAY, measured by the Monitor via A/D 1. Replace PC 1923 Power control.
converter on PC 1923 Power control, is too high. The alarm is activated if the 2. Replace AC/DC Converter.
voltage is above limit for at least 5 seconds. Limit: 16.0 V.
6 - 127
6. TROUBLESHOOTING
66 76 980_02
50 Power failure The central supply voltage node +12V_UNREG on PC 1923 Power control, 1. Replace PC 1923 Power control.
measured by the Monitor via A/D converter on PC 1923, is too low. The alarm is 2. Replace AC/DC Converter.
activated if the voltage is below limit for at least 5 seconds. Limit: 10.0 V.
51 Power failure The central supply voltage node +12V_UNREG on PC 1923 Power control, 1. Replace PC 1923 Power control.
measured by the Monitor via A/D converter on PC 1923, is too high. The alarm is 2. Replace AC/DC Converter.
activated if the voltage is above limit for at least 5 seconds. Limit: 16.0 V.
52 Gas analyzer error Patient Gas Analyzer error. Patient Gas Analyzer is not calibrated correctly or 1. Calibrate Patient Gas Analyzer.
calibration is lost 2. Replace Patient Gas Analyzer.
Additional information:
Additional error codes reported by the Patient Gas Analyzer in the Technical log.
Refer to separate table for gas analyzer alarms.
53 Gas analyzer error Patient Gas Analyzer error. Flow through the Patient Gas Analyzer exceeds 1. May occur if an external suction device is
normal value (200 ml/min or 0 ml/min during STANDBY). used.
2. Check Patient Gas Analyzer pump.
Additional information: 3. Replace Patient Gas Analyzer.
Additional error codes reported by the Patient Gas Analyzer in the Technical log.
Refer to separate table for gas analyzer alarms.
54 Gas analyzer error Patient Gas Analyzer error. Gas concentrations are outside specified range. 1. Check delivery of agents/Vaporizer accuracy.
2. If 1 correct, calibrate Patient Gas Analyzer.
Additional information:
Additional error codes reported by the Patient Gas Analyzer in the Technical log.
Refer to separate table for gas analyzer alarms.
55 Gas analyzer error Patient Gas Analyzer error. Gas concentrations are outside specified range. 1. Check delivery of agents/Vaporizer accuracy.
2. If 1 correct, calibrate Patient Gas Analyzer.
Additional information:
Additional error codes reported by the Patient Gas Analyzer in the Technical log.
Refer to separate table for gas analyzer alarms.
6 - 128
6. TROUBLESHOOTING
66 76 980_02
56 Gas analyzer error Patient Gas Analyzer error. Zero calibration not correctly performed. Error report 1. Check Zero calibration sampling line.
must be repeated 5 times in a row before the alarm is activated. 2. Check Zero Calibration Valve.
1. Occlusion in Zero calibration sampling line. 3. Replace Patient Gas Analyzer.
2. Occlusion in Zero Calibration Valve.
Additional information:
Additional error codes reported by the Patient Gas Analyzer in the Technical log.
Refer to separate table for gas analyzer alarms.
57 APL valve malfunction APL knob or potentiometer HW failures. Activated when any of the following 1. Perform APL valve calibration
conditions are fulfilled: 2. Check/replace APL valve.
a. The APL value read from Control deviates more than 2 cmH2O from calculated
value on Monitoring for a time span of at least 5 seconds.
b. The APL potentiometer value is outside valid range.
The APL pressure will be set automatically to 15 cmH2O in infant patient category
and to 20 cmH2O in adult patient category. The operator will be informed about
the possibility to switch to Automatic Ventilation mode or Emergency Ventilation
mode.
58 Gas analyzer error Patient Gas Analyzer error. Refer to the Technical log for further information. 1. Restart system.
There are around 50 internal AION alarms which activates this alarm (these are 2. Check sampling tubing’s for occlusion.
often filtered in FLOW-i) 3. Check/ replace connection cable.
Possible causes: 4. Replace Patient Gas Analyzer.
1. Indicates a problem with the internal sampling tubing’s (90).
2. Connection cable problem.
3. Hardware problem.
Additional information:
Additional error codes reported by the Patient Gas Analyzer in the Technical log.
Refer to separate table for gas analyzer alarms.
59 Vaporizer error 1. Failed Vaporizer 1 test in SCO or as separate test. 1. If generated during SCO or separate tests;
2. Vaporizer 1 error. check for information in the Test results log. For
3. VAP_DOCK.L signal line error. recommended actions, refer to Troubleshooting,
section System Checkout (14. Vaporizer 1 final
4. DISABLE signal line from CON failed. test).
2. If generated during operation; isolate error
Additional information: cause by checking for other TE and perform
6 - 129
MON detects that Vaporizer 1 has been disabled (by CON). recommended action for other TE.
6. TROUBLESHOOTING
66 76 980_02
60 Vaporizer error 1. Failed Vaporizer 2 test in SCO or as separate test. 1. If generated during SCO or separate tests;
2. Vaporizer 2 error. check for information in the Test results log. For
3. VAP_DOCK.L signal line error. recommended actions, refer to Troubleshooting,
section System Checkout (15. Vaporizer 2 final
4. DISABLE signal line from CON failed. test).
2. If generated during operation; isolate error
Additional information: cause by checking for other TE and perform
MON detects that Vaporizer 2 has been disabled (by CON). recommended action for other TE.
61 Power button error Power button error. There are two parallel contact blades in the switch. The 1. Replace Power button.
output signals are compared to determine switch position and to detect a possible
malfunction in the switch. Activated if the power button signals differs during two
consecutive checks. A check is done every 5 seconds. Can only be activated in
STANDBY.
62 Emergency ventilation Emergency ventilation switch error. There are two parallel contact blades in the 1. Replace Emergency ventilation switch.
On/Off error switch. The output signals are compared to determine switch position and to
detect a possible malfunction in the switch. Activated if the emergency switch
signals differs during two consecutive checks. A check is done every 5 seconds.
63 Anesthesia delivery 1. Disturbances in communication between CON and Vaporizer 1. 1. Check/replace concerned PC boards:
error 2. Disturbances in communication between MON and Vaporizer 1. - PC 1909 Vaporizer adapter
3. Disturbances in communication between CON and MON. - PC 1920 Control
4. Disturbances in communication between Vaporizer 1 CPU and FPGA. - PC 1921 Monitoring.
2. Vaporizer service.
Additional information:
Detect inconsistencies between requested and delivered anesthetic agent flow
setting into Vaporizer 1. MON detects differences between flow setting data from
CON and from Vaporizer 1 (FPGA).
Information in Technical alarms log:
100 = Internal VAP error.
101 = Internal VAP error.
200 = Internal VAP error.
201 = Internal CON error.
202 = Internal error.
6 - 130
6. TROUBLESHOOTING
66 76 980_02
64 Anesthesia delivery 1. Disturbances in communication between CON and Vaporizer 2. 1. Check/replace concerned PC boards:
error 2. Disturbances in communication between MON and Vaporizer 2. - PC 1909 Vaporizer adapter
3. Disturbances in communication between CON and MON. - PC 1920 Control
4. Disturbances in communication between Vaporizer 2 CPU and FPGA. - PC 1921 Monitoring.
2. Vaporizer service.
Additional information:
Detect inconsistencies between requested and delivered anesthetic agent flow
setting into Vaporizer 2. MON detects differences between flow setting data from
CON and from Vaporizer 2 (FPGA).
Information in Technical alarms log:
100 = Internal VAP error.
101 = Internal VAP error.
200 = Internal VAP error.
201 = Internal CON error.
202 = Internal error.
65 Power failure The supply voltage +12V_UNREG_EXPANSION, measured by the Monitor via 1. Check/replace Gas Pressure Transducer
A/D converter on PC 1923 Power control, is too low. The alarm is activated if the used for measuring the gas pressure in cylinder
voltage is below limit for at least 5 seconds. Limit: 10.0 V. on the Backup Gas Trolley.
2. Check/replace Controller unit for the LED
lamp (the table illumination lamp).
3. Replace PC 1923 Power control.
4. Replace PC 1903 External connectors.
5. Check/remove external device connected to
connector N41 of PC 1903 External connectors.
66 Power failure The supply voltage +12V_UNREG_EXPANSION, measured by the Monitor via 1. Replace PC 1923 Power control.
A/D converter on PC 1923 Power control, is too high. The alarm is activated if the 2. Replace AC/DC Converter.
voltage is above limit for at least 5 seconds. Limit: 16.0 V.
67 Power failure The supply voltage +12V_UNREG_EXP_SAF_VALVES, measured by the 1. Replace PC 1919 Exp channel.
Monitor via A/D converter on PC 1923 Power control, is too low. The alarm is 2. Check/replace Fresh Gas Safety Valve.
activated if the voltage is below limit for at least 5 seconds. Limit: 10.0 V. 3. Check/replace APL/PEEP Valve.
4. Replace PC 1923 Power control.
68 Power failure The supply voltage +12V_UNREG_EXP_SAF_VALVES, measured by the 1. Replace PC 1923 Power control.
Monitor via A/D converter on PC 1923 Power control, is too high. The alarm is 2. Replace AC/DC Converter.
activated if the voltage is above limit for at least 5 seconds. Limit: 16.0 V.
6 - 131
6. TROUBLESHOOTING
66 76 980_02
69 Battery error The 'Battery replaced' date is compared to the system clock. This comparison 1. Replace power backup battery and confirm
shows that the power backup battery's operational life span is exceeded. replacement date in Service report (S&S).
70 Patient Cassette error Primarily an indication of Patient Cassette error. 1. Replace Patient Cassette.
Continuous check of: 2. Check electric connection between PC 1917
- EEPROM in the Patient Cassette and PC 1918 Exp.
- EEPROM Checksum channel connector. The position of PC 1918 can
- Thermistor be slightly adjusted to match the contacts on
- Cassette Current PC 1917.
- Cassette Power. 3. Replace PC 1918 Exp. channel connector.
4. Replace PC 1919 Exp. channel.
If any of the checks fail, this alarm is activated. 5. Check heating foil in cassette docking.
May also be caused by other errors. Technical alarm log 'Patient Cassette power
failure' indicates e.g. also PC 1919 Exp. Channel error, cassette docking heating
foil error and power error.
Additional information:
Information in Technical alarms log:
- Patient Cassette EEPROM read/write error.
- Patient Cassette EEPROM checksum error.
- Patient Cassette thermistor error.
- Patient Cassette power failure.
6 - 132
6. TROUBLESHOOTING
66 76 980_02
77 Battery error An error in the Power control system which causes that no valid power backup In case of error codes:
battery exists. - BATTERY_QUICKLY_DRAINED_ERROR
The problem may be due to the battery itself, PC 1923 Power control, PC 1903 - BATTERY_SHORTED_ERROR
External connectors, or the cables between the battery and PC 1903 External - BATTERY_LOW_VOLTAGE_ERROR
connectors.
- BATTERY_HIGH_VOLTAGE_ERROR
TE 77 errors with error code: BATTERY_CONNECTION_ERROR_nn are not
persistent and are reset by a system restart. - BATTERY_SYNC_ERROR
- BATTERY_OVERCHARGE_ERROR
The other TE 77 errors are persistent and will remain until a corrective action is 1. Check power backup battery cables (i.e.
logged in a Service Report under Service & Settings. The action can be one of: Battery and Battery sense cables between
battery and PC 1903 External connectors).
1. Replaced internal battery.
2. Replace power backup battery.
2. Reset internal battery errors.
81 Control panel Internal CAN communication error. Communication with different CAN nodes 1. Replace PC 1922 Panel.
disconnected fails. Alarm activated by Monitoring node if the Panel node connection is lost.
Connection timeout: 2.5 seconds.
6 - 133
6. TROUBLESHOOTING
66 76 980_02
83 Airway pressure sensor Pressure transducer error. Difference between compensated Fresh gas pressure 1. Check for occlusion or leakage in Inspiratory
error and Insp. pressure more than 15 cmH2O: pressure and Fresh Gas pressure sampling lines
a. During six consecutive breaths 2. Replace PC 1781 I (inspiratory pressure)
or 3. Replace PC 1781 F (fresh gas pressure)
b. 30 seconds when no inspiration is detected (applicable only in Manual 4. Replace Gas Modules, one at a time.
ventilation).
The alarm will not be activated if any of the following alarms are activated:
a. CHK_TUBING.
b. EXCESSIVE_LEAKAGE.
c. VAPORIZER_1_DISCONNECTED_ERROR.
d. VAPORIZER_2_DISCONNECTED_ERROR.
e. GAS_SUPPLY_PRESS_O2_TOO_LOW and
GAS_SUPPLY_PRESS_AIR_TOO_LOW at the same time.
Possible causes:
1. Pressure sampling line failure.
2. Pressure transducer error.
3. Too high pressure from Gas Module.
85 Backup cylinder Air cylinder pressure transducer error. Air cylinder pressure outside specified 1. Check/replace the cylinder pressure
pressure unknown: Air range -2000 to 27000 kPa (-20 bar to 270 bar) for more than 5.5 seconds. transducer.
2. Check the cable connection of the Cylinder
Pressure Transducer to PC 1916 Cylinder
Pressure Connection.
3. Check the cable connection of PC 1916
Cylinder Pressure Connection to PC 1903
External connectors.
4. Replace PC 1903 External connectors.
6 - 134
6. TROUBLESHOOTING
66 76 980_02
86 Backup cylinder O2 cylinder pressure transducer error. O2 cylinder pressure outside specified 1. Check/replace the cylinder pressure
pressure unknown: O2 range -2000 to 27000 kPa (-20 bar to 270 bar) for more than 5.5 seconds. transducer.
2. Check the cable connection of the Cylinder
Pressure Transducer to PC 1916 Cylinder
Pressure Connection.
3. Check the cable connection of PC 1916
Cylinder Pressure Connection to PC 1903
External connectors.
4. Replace PC 1903 External connectors.
87 Backup cylinder N2O cylinder pressure transducer error. N2O cylinder pressure outside specified 1. Check/replace the cylinder pressure
pressure unknown: N2O range -2000 to 27000 kPa (-20 bar to 270 bar) for more than 5.5 seconds. transducer.
2. Check the cable connection of the Cylinder
Pressure Transducer to PC 1916 Cylinder
Pressure Connection.
3. Check the cable connection of PC 1916
Cylinder Pressure Connection to PC 1903
External connectors.
4. Replace PC 1903 External connectors.
6 - 135
6. TROUBLESHOOTING
66 76 980_02
88 Airway pressure sensor Pressure transducer error. Difference between compensated Fresh gas pressure 1. Check for occlusion or leakage in Expiratory
error and Exp. pressure more than 15 cmH2O during 6 consecutive breaths or max 30 pressure and Fresh Gas pressure sampling lines
seconds. 2. Replace PC 1781 E (expiratory pressure)
Alarm check is not performed if: 3. Replace PC 1781 F (fresh gas pressure)
a. O2 Flush is active. 4. Replace Gas Modules, one at a time.
b. Safety Valve is opened.
c. Manual ventilation is enabled.
The alarm will not be activated if any of the following alarms are activated:
a. CHK_TUBING.
b. EXCESSIVE_LEAKAGE.
c. VAPORIZER_1_DISCONNECTED_ERROR.
d. VAPORIZER_2_DISCONNECTED_ERROR.
e. FRESH_GAS_INSUFFICIENT_ALARM
f. GAS_SUPPLY_PRESS_O2_TOO_LOW and
GAS_SUPPLY_PRESS_AIR_TOO_LOW at the same time.
Possible causes:
1. Pressure sampling line failure.
2. Pressure transducer error.
3. Too high pressure from Gas Module.
6 - 136
6. TROUBLESHOOTING
66 76 980_02
Technical
Error code Alarms Text Possible cause Recommended action
501 Memory backup battery Memory backup battery voltage level too low on the PC 1920 Control. 1. Replace memory backup battery on PC 1920
depleted Limit: 3.0 ±0.1 V. Control.
502 Communication error Internal communication error. Activated by the Control node if its I2C 1. Replace defective part according to Device ID
communication fails. The TE code is combined with a Device ID and set of stated in the Technical alarms log. Refer to
different error codes specifying in detail which type of error that occurred. The separate table for TE 16/TE 502/TE 601.
details must be retrieved via the technical log.
The alarm indicates a permanent failure.
Additional information:
Device ID on defective part will be stated in the Technical alarms log.
Refer to separate table for TE 16/TE 502/TE 601.
503 Ventilation control error Control memory error. Data stored in Control node persistent memory is lost or 1. Restart system. If error occurs in conjunction
corrupted. This alarm can only be activated at node startup. The alarms can occur with SW update, reinstall SW.
if Control memory backup battery is removed or at system SW upgrade. 2. Replace PC 1920 Control (if not memory
backup battery low).
505 Gas analyzer error Control gas analyzer error. Control gas analyzer is not calibrated correctly or 1. Calibrate Control gas analyzer.
calibration is lost. 2. Replace Control gas analyzer.
Additional information:
Additional error codes reported by the Control gas analyzer in the Technical log.
Refer to separate table for gas analyzer alarms.
506 Gas analyzer error Control gas analyzer error. Internal pneumatic malfunction. Flow through the 1. Check Control gas analyzer pump.
Control gas analyzer exceeds normal value (200 ml/min or 0 ml/min during 2. Replace Control gas analyzer.
STANDBY).
Additional information:
Additional error codes reported by the Control gas analyzer in the Technical log.
Refer to separate table for gas analyzer alarms.
6 - 137
6. TROUBLESHOOTING
66 76 980_02
507 Gas analyzer error Control gas analyzer error. Gas concentrations are outside specified range. 1. Check delivery of agents/Vaporizer accuracy.
2. If 1 correct, calibrate Control gas analyzer.
Additional information:
Additional error codes reported by the Control gas analyzer in the Technical log.
Refer to separate table for gas analyzer alarms.
508 Gas analyzer error Control gas analyzer error. Gas concentrations are outside specified range. 1. Check delivery of agents/Vaporizer accuracy.
2. If 1 correct, calibrate Control gas analyzer.
Additional information:
Additional error codes reported by the Control gas analyzer in the Technical log.
Refer to separate table for gas analyzer alarms.
509 Gas analyzer error Control gas analyzer error. Zero calibration not correctly performed. Error report 1. Check Zero calibration sampling line.
must be repeated 5 times in a row before the alarm is activated. 2. Check Zero Calibration Valve.
1. Occlusion in Zero calibration sampling line. 3. Replace Control gas analyzer.
2. Occlusion in Zero Calibration Valve.
Additional information:
Additional error codes reported by the Control gas analyzer in the Technical log.
Refer to separate table for gas analyzer alarms.
510 Ventilation control error Control memory error. Data stored in Control node persistent memory is lost or 1. Restart system.
corrupted. This alarm can only be activated at node startup. The alarms can occur 2. Replace PC 1920 Control (if not memory
if Control memory backup battery is removed or at System SW upgrade. backup battery low).
511 Gas analyzer error Control gas analyzer error. Refer to the Technical log for further information. 1. Restart system.
There are around 50 internal AION alarms which activates this alarm (these are 2. Check sampling tubing’s for occlusion.
often filtered in FLOW-i). 3. Check/ replace connection cable.
Possible causes: 4. Replace Control gas analyzer.
1. Indicates a problem with the internal sampling tubing’s.
2. Connection cable problem.
3. Hardware problem.
Additional information:
Additional error codes reported by the Control gas analyzer in the Technical log.
Refer to separate table for gas analyzer alarms.
6 - 138
6. TROUBLESHOOTING
66 76 980_02
512 Ventilation control error Control memory error. Data stored in Control node persistent memory is lost or 1. Restart system. If error occurs in conjunction
corrupted. This alarm can only be activated at node startup. The alarms can occur with SW update, reinstall SW.
if Control memory backup battery is removed or at System SW upgrade. 2. Replace PC 1920 Control (if not memory
backup battery low).
514 Gas analyzer error: O2 Control O2 Sensor error. 1. Check fresh gas supply to Control O2 sensor.
1. Broken connection to Control O2 sensor. 2. Check connection to Control O2 sensor.
2. Control O2 sensor HW error. 3. Calibrate Control O2 sensor.
3. Control O2 sensor out of calibration and/or exposed to low oxygen 4. Replace Control O2 sensor.
concentration or excessive pressure.
Additional information:
201 = no contact with device.
202 = internal failure (status code 'X').
203 = device operates outside spec (status code 'E'). Calibration can solve the
problem.
See Servomex Pm1116 manual for more info on status codes.
515 Gas analyzer error: O2 Control O2 Sensor error. Control O2 Sensor out of calibration. 1. Calibrate Control O2 Sensor.
Additional information:
204 = device is not calibrated.
See Servomex Pm1116 manual for more info on status codes.
516 Fresh gas O2 conc. Gas concentration delivery error. Checked by using Control gas analyzer and 1. Check quality of AIR and O2 gas supply.
Error comparing measured concentrations with set concentrations. Lower limit: 19 % 2. Calibrate Control O2 Sensor.
ABS absolute or 80% of set value or when momentary O2 concentration <10 %
ABS. 3. Check/replace Gas Module O2.
The alarm (TE 516) shall not be activated for measured O2 concentration values 4. Check/replace Gas Module AIR or N2O.
above 75 % ABS.
1. Insufficiency O2 supply. If Backup gas trolley or Backup gas holder is
2. Control O2 Sensor out of calibration. connected, this error message may indicate
3. Fresh Gas Module O2 error (low delivery). incorrect connection of backup gas supply.
4. Fresh Gas Module AIR or N2O error (high delivery).
6 - 139
6. TROUBLESHOOTING
66 76 980_02
517 Fresh gas O2 conc. Gas concentration delivery errors. Checked by using Control Gas Analyzer and 1. Check/replace Gas Module O2.
Error comparing measured concentrations with set concentrations. Upper limit 115 % of 2. Check/replace Gas Module AIR or N2O.
set value. 3. Calibrate Control O2 Sensor.
1. Gas Module O2 error (high delivery).
2. Gas Module AIR or N2O error (low delivery). If Backup gas trolley or Backup gas holder is
3. Control O2 Sensor out of calibration. connected, this error message may indicate
incorrect connection of backup gas supply.
520 Anesthesia delivery 1. Fresh gas flow error (from the Gas Modules). 1. If SCO fails; perform recommended action
error 2. Fresh gas to Vaporizer 1 leakage. according to SCO result.
3. Control Gas Analyzer error.
4. Vaporizer 1 error.
Additional information:
Control Gas Analyzer detects high anesthetic agent concentration. Alarm is
initiated by CON.
521 Anesthesia delivery 1. Fresh gas flow error (from the Gas Modules). 1. If SCO fails; perform recommended action
error 2. Fresh gas to Vaporizer 2 leakage. according to SCO result.
3. Control Gas Analyzer error.
4. Vaporizer 2 error.
Additional information:
Control Gas Analyzer detects high anesthetic agent concentration. Alarm is
initiated by CON.
522 Anesthesia delivery 1. Fresh gas flow error (from the Gas Modules). 1. Check liquid level.
error 2. Control Gas Analyzer error. 2. Vaporizer service.
3. Vaporizer 1 error.
Additional information:
Control Gas Analyzer detects low anesthetic agent concentration. Alarm is
initiated by CON.
The alarm is not active during the first 4 minutes after opening in inlet/outlet
valves to Vaporizer 1.
6 - 140
6. TROUBLESHOOTING
66 76 980_02
523 Anesthesia delivery 1. Fresh gas flow error (from the Gas Modules). 1. Check liquid level.
error 2. Control Gas Analyzer error. 2. Vaporizer service.
3. Vaporizer 2 error.
Additional information:
Control Gas Analyzer detects low anesthetic agent concentration. Alarm is
initiated by CON.
The alarm is not active during the first 4 minutes after opening in inlet/outlet
valves to Vaporizer 2.
6 - 141
6. TROUBLESHOOTING
66 76 980_02
Additional information:
Device ID on defective part will be stated in the Technical alarms log.
Refer to separate table for TE 16/TE 502/TE 601.
603 Key press timeout Key/button/knob indicated as pressed or stuck for at least one minute. HW related 1. Replace PC 1925 Display CPU.
error with touch screen, knob or membrane buttons.
604 Audible alarm error Control Panel loudspeaker error. Activated if no loudspeaker alarm is detected 1. Replace loudspeaker.
when an alarm is active. Condition must be fulfilled during three consecutive 2. Replace PC 1924 Display Connection.
checks. 3. Replace PC 1925 Display CPU.
605 Control panel error Panel memory error. Data stored in Panel node persistent memory is lost or 1. Restart system. If error occurs in conjunction
corrupted. Checks performed at startup. The alarms can occur if Panel memory with SW update, reinstall SW.
backup battery is removed or at system SW upgrade. 2 Replace PC 1922 Panel (if not memory
backup battery low).
606 Control panel high The temperature inside the Control Panel, monitored by a transducer integrated in 1. Replace PC1925 Panel (if room temperature
temperature the CPU on PC 1925, is above specified range: 55 °C ±5 °C. is max. 35 °C)
607 Communication error Video signal level (LVDS level) outside specified range. 1. Check/replace Control Panel Cable
This could detect a broken graphic driver on the panel. 2. Replace PC 1925 Display CPU
3. Replace PC 1924 Display Connection.
608 Power failure The +5 V voltage on PC 1925 Display CPU is outside specified range: +4.5 V to Measure voltage on PC 1924 Display
+5.5 V. Connection, connector P2, pin 4&6 (+5 V ±10%):
1. If voltage is correct, replace PC 1925 Display
CPU.
2. If voltage if outside range, replace PC 1924
Display Connection.
6 - 142
6. TROUBLESHOOTING
66 76 980_02
609 Backlight malfunction The backlight current on PC 1925 Display CPU is outside specified range: 1. Measure the voltage on PC 1925 Display
100 mA - 1800 mA. Indicates broken backlight. CPU, connector P3 (10-16 V).
2. Replace Backlight inverter.
3. Replace PC 1925 Display CPU.
4. Replace backlight.
611 Communication error CAN communication error between PC 1922 Panel and PC 1925 Display CPU. 1. Check/replace Control Panel Cable.
No response from PC 1925 within 3 seconds. 2. Replace PC 1924 Display Connection.
3. Replace PC 1925 Display CPU.
4. Replace PC 1922 Panel.
613 MAN/AUTO switch MAN/AUTO ventilation switch error. There are two parallel contact blades in the 1. Replace Manual ventilation switch.
malfunction switch. The output signals are compared to determine switch position and to
detect a possible malfunction in the switch.
614 Memory backup battery Memory backup battery voltage level too low on the PC 1922 Panel. 1. Replace memory backup battery on PC 1922
depleted Limit: 3.0 ±0.1 V. Panel.
6 - 143
6. TROUBLESHOOTING
66 76 980_02
Additional information:
Information in Technical alarms log:
- PC1919 EEPROM read/write error.
- PC1919 EEPROM checksum error.
6 - 144
6. TROUBLESHOOTING
66 76 980_02
CONTROL:
256 CTRL PC 1920 Control
257 CON_VALVE_DRIVERS_IO PC 1907 Valve Drivers I/O
260 CON_VALVE_DRIVERS_EEPROM PC 1907 Valve Drivers
6 - 145
6. TROUBLESHOOTING
66 76 980_02
PANEL:
512 PANEL PC 1922 Panel
513 PAN_VALVE_DRIVERS_IO PC 1907 Valve Drivers I/O
514 PAN_VALVE_DRIVERS_ADC PC 1907 Valve Drivers A/D Converter
515 PAN_O2_FLUSH_IO PC 1927 O2 Flush I/O
516 PAN_O2_FLUSH_EEPROM PC 1927 O2 Flush
517 PANEL_ADC PC 1922 Panel
EXPIRATORY FLOWMETER:
1024 EXP_CHAN PC 1919 Expiratory channel
1025 EXP_CASSETTE PC 1917 Expiratory channel cassette
Error # ERROR
1 CHK_SUM_ERR
2 NR_OF_BYTES_IN_CHKSUM_ERR
3 CHKSUM_FILL_BYTE_ERR
4 WRITABLE_RECORDS_FAILURE
5 WRITABLE_RECORDS_DIFFER
6 COMM_ERR
7 VALIDATION_ERR
8 WCR_VALIDATION_ERR
9 WRITE_NOT_PERMITTED
10 OTHER_ERR
6 - 146
6. TROUBLESHOOTING
66 76 980_02
Additional information:
Vaporizer general error alarm. CPU failed to write in the flash memory or
the FPGA SW is incompatible. Vaporizer is disabled, agent delivery not
possible.
901 951 Vaporizer error 1. Bubbles in the level meter tube. 1. Wait a few minutes until the bubbles are
2. Liquid level detectors malfunction. evacuated.
3. Poor cable connection inside Vaporizer. 2. Vaporizer service.
4. PC 1912, PC 1913 or PC 1910 malfunction.
Additional information:
Alarm when the Vaporizer fails to read agent liquid level.
902 952 Vaporizer error 1. One or both Vaporizer chamber temperature sensors broken. 1. Vaporizer service.
2. PC 1910 malfunction.
Additional information:
Vaporizer chamber temperature sensor error alarm. Sensors 1 and 2 have
too large deviation. Vaporizer is disabled, agent delivery not possible.
904 954 Vaporizer error 1. Vaporizer injector leakage, e.g. due to mechanical failure. 1. Switch to Standby mode. Remove and re-
2. PC 1914, PC 1926 or PC 1910 malfunction. insert the vaporizer. Perform a System
3. Poor cable connection between these boards inside Vaporizer. Checkout. This may clear the alarm.
2. Vaporizer service.
Additional information:
Spray detected when the Vaporizer injector is supposed to be closed.
Persistent error # 1 (high leakage) or # 2 (low leakage) will be set.
Vaporizer is disabled, agent delivery not possible.
6 - 147
6. TROUBLESHOOTING
66 76 980_02
905 955 Vaporizer error 1. Vaporizer injector leakage. 1. Switch to Standby mode. Remove and re-
2. PC 1914, PC 1926 or PC 1910 malfunction. insert the vaporizer. Perform a System
3. Poor cable connection between these boards inside Vaporizer. Checkout. This may clear the alarm.
2. Vaporizer service.
Additional information:
Persistent error # 1 (high injector leakage) or # 2 (low injector leakage) is
detected in the Vaporizer at power-on. The vaporizer will be disabled once
entering run mode, agent delivery not possible. The persistent error can be
cleared by the system during SCO.
906 956 Vaporizer error 1. Vaporizer input power voltage out of range 1. Vaporizer service.
2. PC 1910 or PC 1911 malfunction.
Additional information:
Vaporizer hardware error alarm. Internal voltages +1.2 V_FPGA, +12
V_VAP or +12 V_UNREG out of range. Vaporizer is disabled, agent
delivery not possible.
907 957 Vaporizer error 1. Anesthetic agent condensation in the chamber. 1. Place the Vaporizer inside i fume hood for at
2. PC 1914, PC 1926 or PC 1910 malfunction. least 30 minutes. This can vent agent droplets
3. Poor cable connection between these boards inside Vaporizer. from the chamber.
2. Vaporizer service.
Additional information:
Vaporizer spray signal error alarm. SPRAY signal level <625 mV. Normal
level (when there is no spray) is 800 mV. Vaporizer is disabled, agent
delivery not possible.
908 958 Vaporizer error 1. Agent liquid collected in the chamber (in front of spray sensor). 1. Place the Vaporizer inside i fume hood for at
2. PC 1914, PC 1926 or PC 1910 malfunction. least 30 minutes. This can vent agent collected
3. Poor cable connection between these boards inside Vaporizer. in the chamber.
2. Vaporizer service.
Additional information:
Vaporizer spray sensor error alarm. Calibration or test of spray signal
during the first 5 s of power-up failed. Vaporizer is disabled, agent delivery
not possible.
6 - 148
6. TROUBLESHOOTING
66 76 980_02
909 959 Vaporizer error Case 1) ERROR TE 902/952 is also presented: 1. Vaporizer service.
a) Vaporizer chamber temperature sensor # 1 is broken
b) PC 1910 malfunction.
Case 2) ERROR TE 902/952 is not presented:
PC 1910 failed in driving the heater element.
Additional information:
Vaporizer chamber temperature sensor # 1; temperature >60˚C. Vaporizer
is disabled, agent delivery not possible.
910 960 Vaporizer error 1. Malfunction in DC/DC converter or its peripheral components on PC 1. Vaporizer service.
1910.
Additional information:
Vaporizer heater voltage error alarm; heater voltage >47 V. Vaporizer is
disabled, agent delivery not possible.
911 961 Vaporizer error 1. CPU error on PC 1910. 1. Vaporizer service.
2. PC 1910 malfunction.
Additional information:
Vaporizer injector pulse error alarm; FPGA detects that injector control
pulse is >15 ms. Vaporizer is disabled, agent delivery not possible.
912 962 Vaporizer error 1. High agent liquid pressure. 1. Check/adjust Pressure Regulators REG1-3.
2. Vaporizer pressure transducer malfunctions. 2. Vaporizer service.
3. Loose pressure transducer cable connections inside Vaporizer.
4. PC 1910 malfunction.
Additional information:
Vaporizer pressure transducer error alarm; agent liquid pressure >220 kPa.
Vaporizer is disabled, agent delivery not possible.
6 - 149
6. TROUBLESHOOTING
66 76 980_02
913 963 Vaporizer error 1. Heating foil temperature sensor error. 1. Vaporizer service.
2. Poor heating foil cable connection inside Vaporizer.
3. Heating foil not correctly mounted onto the vaporizing chamber.
4. Vaporizer chamber temperature sensors 1 and 2 (on PC 1910) not
correctly mounted on the chamber.
5. PC 1910 malfunction.
Additional information:
Heating foil temperature sensor error alarm; temperature >170˚C.
Vaporizer is disabled, agent delivery not possible.
914 964 Vaporizer error 1. CPU SW corrupt. 1. Upgrade Vaporizer SW.
2. Vaporizer service.
Additional information:
Vaporizer software corrupt alarm. The CPU application has an incorrect
checksum and cannot start. Vaporizer is disabled, agent delivery not
possible.
915 965 Vaporizer error 1. Application code database is not complete. 1. Vaporizer service.
Additional information:
Vaporizer database corrupt alarm. Database, read from the Vaporizer
status register, is incomplete. Vaporizer is disabled, agent delivery not
possible.
916 966 Vaporizer error 1. Vaporizer pressure valve error. 1. Check Pressure Regulator REG1 output
2. Vaporizer scavenging valve error. pressure. If REG1 is outside specified range,
3. Vaporizer docking valve error. check Pressure Regulators REG2-3 output
pressure (REG2-3 supplies pressure to REG1).
4. Vaporizer pressure sensor or the peripheral circuits failed. If required, adjust REG1-3, refer to chapter
Service procedures.
Additional information: 2. Check/replace Vaporizer pressure valve.
Vaporizer pressure too low alarm. Liquid pressure <50 kPa after 3.75 3. Check/replace Vaporizer scavenging valve.
seconds when vaporizer pressure valve is open. Alarm is initiated by CON. 4. Vaporizer service.
Vaporizer is disabled, agent delivery not possible.
6 - 150
6. TROUBLESHOOTING
66 76 980_02
917 967 Vaporizer error 1. Vaporizer pressure valve error. 1. Check Pressure Regulator REG1 output
2. Vaporizer scavenging valve error. pressure. If REG1 is outside specified range,
3. Vaporizer docking valve error. check Pressure Regulators REG2-3 output
pressure (REG2-3 supplies pressure to REG1).
4. Vaporizer pressure sensor or the peripheral circuits failed. If required, adjust REG1-3, refer to chapter
Service procedures.
Additional information: 2. Check/replace Vaporizer pressure valve.
Vaporizer pressure too high alarm. Repeated Vaporizer high pressure 3. Check/replace Vaporizer scavenging valve.
alarm. Vaporizer liquid pressure above 170 kPa even if it has been lowered 4. Vaporizer service.
3 times within 20 s during normal operation. Alarm is initiated by CON.
918 968 Vaporizer error 1. Vaporizer pressure valve error. 1. Check Pressure Regulator REG1 output
2. Vaporizer scavenging valve error. pressure. If REG1 is outside specified range,
3. Vaporizer docking valve error. check Pressure Regulators REG2-3 output
pressure (REG2-3 supplies pressure to REG1).
4. Vaporizer pressure sensor broken or poor cable connection. If required, adjust REG1-3, refer to chapter
5. PC 1910 malfunction. Service procedures.
2. Check/replace Vaporizer pressure valve.
Additional information: 3. Check/replace Vaporizer scavenging valve.
Vaporizer sensor error alarm. Vaporizer pressure evacuation failure. Liquid 4. Vaporizer service.
pressure still over 1700 mbar after 1.4 s evacuation during normal
operation. Alarm is initiated by CON. Vaporizer is disabled, agent delivery
not possible.
919 969 Vaporizer error 1. Vaporizer pressure valve error. A dialogue will inform the user that the Vaporizer
2. Vaporizer scavenging valve error. has been unlocked but remains pressurized.
3. Vaporizer docking valve error. Service of the FLOW-i unit including concerned
Vaporizer is recommended.
4. Vaporizer pressure sensor broken or poor cable connection.
1. Check/replace Vaporizer scavenging valve.
5. PC 1910 malfunction.
2. Vaporizer service.
Additional information:
Vaporizer pressure evacuation error alarm. Vaporizer pressure is not below
200mbar (700 mbar for Desflurane) after 10 s evacuation. Alarm is initiated
by CON. Vaporizer is disabled, agent delivery not possible.
6 - 151
6. TROUBLESHOOTING
66 76 980_02
920 970 Vaporizer error 1. PC 1910 malfunction. Check if communication with another Vaporizer
2. I2C communication error in Vaporizer. in the same slot works:
3. I2C communication error in FLOW-i system. 1. If yes; Vaporizer service
2. If no; replace PC 1909 Vaporizer adapter
Additional information:
Vaporizer communication check error alarm. Write check failure in
communication with Vaporizer. The written temperature or safety valve
message data is not the same as the read data. Alarm is initiated by CON.
Vaporizer is disabled, agent delivery not possible.
921 971 Vaporizer error 1. PC 1910 malfunction. Check if communication with another Vaporizer
2. I2C communication error in Vaporizer. in the same slot works:
3. I2C communication error in FLOW-i system. 1. If yes; Vaporizer service
2. If no; replace PC 1909 Vaporizer adapter
Additional information:
Vaporizer communication error alarm. Alarm is initiated by CON. Vaporizer
is disabled, agent delivery not possible.
922 972 Vaporizer error 1. Failed Vaporizer test in SCO or as separate test. 1. If generated during SCO or separate tests;
2. CPU error on PC 1910. check for information in the Test results log. For
3. FPGA error in Vaporizer. recommended actions, refer to Troubleshooting,
section System Checkout (14. Vaporizer 1 final
4. VAP_DOCK.L signal line error. test).
5. DISABLE signal line from MON failed. 2. If generated during operation; isolate error
cause by checking for other TE and perform
Additional information: recommended action for other TE.
Vaporizer disable alarm. CON detects that the Vaporizer disable signal is
active (by MON). Vaporizer is disabled, agent delivery not possible.
923 973 Vaporizer error 1. Vaporizer could be disconnected when pressurized. 1. Check Vaporizer locking mechanism.
2. HW or SW failure in Vaporizer. 2. Check/replace PC 1909 Vaporizer adapter.
3. Vaporizer communication error. 3. Vaporizer service
Additional information:
Vaporizer disconnected when pressurized. VAP_DOCK signal goes high at
the same time that communication is lost. This means that the Vaporizer
was removed when pressurized. Alarm is initiated by CON.
6 - 152
6. TROUBLESHOOTING
66 76 980_02
924 974 Vaporizer error 1. Vaporizer SW incompatible with System SW. 1. Install compatible SW (System SW and/or
Vaporizer SW).
Additional information: 2. Vaporizer service.
Vaporizer SW is not compatible with System SW. Vaporizer SW version
higher than System SW. Alarm is initiated by CON. Vaporizer is disabled,
agent delivery not possible.
925 975 Vaporizer lid leakage 1. High Vaporizer leakage. 1. Vaporizer service.
2. Check/replace Vaporizer pressurize valve.
Additional information: 3. Check/replace Vaporizer scavenging valve.
Vaporizer lid leakage. After each pressurizing sequence, the Vaporizer
pressure valve is kept closed for 15 s to detect any high leakage in
Vaporizer. Alarm is initiated by CON.
926 976 Vaporizer error 1. CPU error on PC 1910. 1. Check/replace PC 1909 Vaporizer adapter.
2. FPGA in Vaporizer failed. 2. Vaporizer service.
3. VAP_DOCK.L signal line failure.
4. PC 1910 error.
Additional information:
DOCK.L signal is forced high by the Vaporizer without any error bit set.
FPGA has asserted "VAP_DOCK.L" to high because it did not get any
response from the CPU regarding a detected error. The Volume Reflector
and Fresh gas line is “flushed” through the Fresh Gas Safety Valve.
Vaporizer is disabled, agent delivery not possible.
6 - 153
6. TROUBLESHOOTING
66 76 980_02
Error Data
code Description limits Filter Possible cause Recommended action
1 Internal CPU error FATAL 1. SW download failed. 1. Re-install AION SW.
2. AION error. 2. Replace AION.
2 RAM error FATAL AION error. 1. Replace AION.
3 FLASH error FATAL 1. SW download failed. 1. Re-install AION SW.
2. AION error. 2. Replace AION.
4 EEPROM checksum error BM FATAL AION error. 1. Replace AION.
5 EEPROM checksum error GMB FATAL AION error. 1. Replace AION.
6 EEPROM checksum error PM FATAL AION error. 1. Replace AION.
7 SW error GMB FATAL AION error. 1. Replace AION.
8 SW Error BM. Incompatible AION AION error. 1. Replace AION.
type, BM type and GMB type
9 Input voltage error 1. Power supply 24 V. 1. Check power supply.
2. AION error. 2. Replace AION.
10 7 V outside specification 1 AION error. 1. Replace AION.
11 Reference voltage outside 1 AION error. 1. Replace AION.
specification
12 Valve voltage outside specification 1 AION error. 1. Replace AION.
13 Chopper motor voltage outside 1 AION error. 1. Replace AION.
specification
14 IR voltage low 1 AION error. 1. Replace AION.
15 IR voltage high 1 AION error. 1. Replace AION.
16 EEPROM write failure BM FATAL AION error. 1. Replace AION.
17 EEPROM write failure GMB FATAL AION error. 1. Replace AION.
18 EEPROM write failure PM FATAL AION error. 1. Replace AION.
19 System overheated FATAL 1. FLOW-i overheated. 1. Check fan/environmental temperature.
6 - 154
Control.
6. TROUBLESHOOTING
66 76 980_02
51 Receive data error 1 1. Ribbon cable error. 1. Check/replace the ribbon cable.
2. AION error. 2. Replace AION.
3. MON / CON error. 3. Replace PC 1921 Monitoring / PC 1920
Control.
52 GMB data error 1 AION error. 1. Replace AION.
60 Reduced accuracy mode timeout FATAL AION error. 1. Replace AION.
61 Zero reference measurement failed 1. Occluded sampling line. 1. Check sampling line.
2. AION error. 2. Replace AION.
62 Zero reference measurement failed N/A N/A (Artema info).
due to high CO2
63 Atmospheric pressure measurement 3 1. High pressure at AION inlet/low pressure at AION 1. Check FLOW-i pressure values.
failed outlet. 2. Replace AION.
2. AION error.
64 Zero reference measurement not Occlusion in sampling tubes. 1. Check tubing.
performed due to occlusion/flow
error
65 O2 Reference measurement failed 15 – 27% 1. Patient O2 Sensor Pm1111E calibration error. 1. Perform Patient O2 Sensor Pm1111E
O2 in 2. Patient O2 Sensor Pm1111E needs adjustment. calibration (Pre Service SW).
room air 3. High/low ambient O2 concentration. 2. Perform Patient O2 Sensor Pm1111E
4. Patient O2 Sensor Pm1111E error. adjustment (Pre Service SW).
Note: PGA only 3. Check tubing.
4. Replace Patient O2 Sensor Pm1111E.
66 No GMB flow during reference 1 Sampling line leakage. 1. Check sampling line.
measurement
67 Failed valve test during reference AION error. 1. Replace AION.
measurement (valve closed)
70 Self test error 1 (pump flow error) 4 1. Occluded sampling line/leakage. 1. Check sampling line.
2. AION error. 2. Replace AION.
71 Self test error 2 (Pressure error in 4 1. Occluded sampling line/leakage. 1. Check sampling line.
GMB) 2. AION error. 2. Replace AION.
72 Self test error 3 (O2 error) 4 N/A N/A (Artema info).
6 - 156
6. TROUBLESHOOTING
66 76 980_02
109 Data limit error Flow -20 – 500 1. High pressure at AION inlet/low pressure at AION 1. Check FLOW-i pressure values.
ml/min outlet. 2. Replace AION.
2. AION error.
110 Data limit error Temp BM 1. FLOW-i overheated. 1. Check fan/environmental temperature.
2. AION error. 2. Replace AION.
111 Data limit error Temp GMB 1. FLOW-i overheated. 1. Check fan/environmental temperature.
2. AION error. 2. Replace AION.
112 Data limit error Breath rate 2 – 100 N/A N/A (ignored by FLOW-i).
b/min
113 Data limit error Temp PM 1. FLOW-i overheated. 1. Check fan/environmental temperature.
2. AION error. 2. Replace AION.
140 Unspecified accuracy CO2 10.4% 1. Absorber error. 1. Check/replace absorber.
2. AION error. 2. Replace AION.
141 Unspecified accuracy N2O 102.2% AION error. 1. Replace AION.
142 Unspecified accuracy O2 100% 1. Patient O2 Sensor Pm1111E calibration error. 1. Perform Patient O2 Sensor Pm1111E
2. Patient O2 Sensor Pm1111E needs adjustment. calibration (Pre Service SW).
3. Sampling line leakage. 2. Perform Patient O2 Sensor Pm1111E
4. Patient O2 Sensor Pm1111E error. adjustment (Pre Service SW).
Note: PGA only. 3. Check tubing.
4. Replace Patient O2 Sensor Pm1111E.
143 Unspecified accuracy Halothane N/A N/A (ignored by FLOW-i).
144 Unspecified accuracy Enflurane N/A N/A (ignored by FLOW-i).
145 Unspecified accuracy Isoflurane N/A N/A (ignored by FLOW-i).
146 Unspecified accuracy Sevoflurane N/A N/A (ignored by FLOW-i).
147 Unspecified accuracy Desflurane N/A N/A (ignored by FLOW-i).
148 Unspecified accuracy Pressure 700 – 1. High pressure at AION inlet/low pressure at AION 1. Check FLOW-i pressure values.
1200 hPa outlet. 2. Replace AION.
2. AION error.
149 Unspecified accuracy Flow 0 – 250 1. High pressure at AION inlet/low pressure at AION 1. Check FLOW-i pressure values.
ml/min outlet. 2. Replace AION.
6 - 158
2. AION error.
6. TROUBLESHOOTING
66 76 980_02
Additional information:
During startup, AION performs a number of self tests
(sampling, pump, power supply, etc). This error code
indicates that the self tests have not passed.
33004 Incompatible AION HW 1. Wrong AION SW version. 1. Install correct AION SW version.
configuration 2. Wrong AION HW configuration. 2. Replace AION.
Additional information:
This error code indicates HW configuration error, but
the cause may be that the SW is incompatible.
33005 Incompatible AION SW version Wrong AION SW version. 1. Install correct AION SW version.
6 - 159
6. TROUBLESHOOTING
66 76 980_02
AION acronyms
BM Base Module including AION PC board.
GMB Gas Measurement Bench including PC board.
PM Pump Module.
6 - 160
6. TROUBLESHOOTING
The LEDs on PC 1907, as described here, indicates that a control signal to the concerned valve
is enabled.
LED # Valve LED # Valve
D35 VAP 1 Lock Pilot Valve EMV7 D42 VAP Bypass Pilot Valve EMV4
D36 VAP 1 Selection Pilot Valve EMV6 D43 VAP 2 Pressure Valve EMV11
D37 VAP 1 Scavenging Valve EMV10 D44 SCO Pilot Valve EMV3
D38 VAP 1 Pressure Valve EMV9 D45 Manual Vent. Pilot Valve EMV2
D39 CGA inlet Valve EMV14 D46 AFGO Valve EMV18
D40 VAP 2 Selection Pilot Valve EMV5 D47 VAP 2 Scavenging Valve EMV12
D41 VAP 2 Lock Pilot Valve EMV8 D48 CGA outlet Valve EMV16
66 76 980_02 6 - 161
6. TROUBLESHOOTING
66 76 980_02 6 - 162
6. TROUBLESHOOTING
PC 1920 Control
66 76 980_02 6 - 163
6. TROUBLESHOOTING
PC 1921 Monitoring
PC 1922 Panel
LED indicators on PC 1922 Panel are not visible (covered by the connector panel) and cannot
be used during troubleshooting.
66 76 980_02 6 - 164
6. TROUBLESHOOTING
66 76 980_02 6 - 165
6. TROUBLESHOOTING
The system will not start The Emergency ventilation 1. Turn off the switch.
up. The ON/OFF switch is switch is active. 2. Replace the switch in case of
not lit. malfunction.
The battery fuses F1 and F2 1. Check/replace fuses F1 and
on PC 1903 External F2 on PC 1903. Always
connectors are blown or replace both fuses at the
malfunction on PC 1903. same time.
2. Replace PC 1903.
The system is ON, but not Malfunction in the Power 1. Check/replace the Power
possible to switch OFF. button. button.
Poor cable connections. 1. Check Power button
connection to PC 1928 Power
On/Off.
2. Check PC 1928 cable
connector to P29 on PC 1900
Main back-plane.
The system will only run The AC/DC Converter will 1. Check cables and cable
on power backup battery. not supply power to PC 1923 connectors between AC/DC
Power control. Converter and PC 1923.
2. Check/replace the AC/DC
Converter.
The system will not run on The power backup battery 1. Check power backup battery
power backup battery. will not supply power to PC cables and cable connectors.
1923 Power control. 2. Check power backup battery.
3. Check/replace fuses F1 and
F2 on PC 1903. Always
replace both fuses at the
same time.
4. Replace PC 1923.
During a Leakage check Drive gas supply is added 1. Check/replace the O2 Flush
the message: ‘Leakage into the patient circuit. valve.
into the system appears’
Fresh gas or reflector gas is 1. Check all four gas modules.
added to patient circuit. Check that nozzle units are
properly inserted and not
damaged.
The Mechanical APL knob Poor adjustment of the 1. Adjust regulator according to
does not respond with a Mechanical APL pressure instructions in chapter Service
distinct feeling or gives an regulator REG5. procedures.
incorrect pressure level.
Mushroom valve in APL 1. Check/replace mushroom
actuator defective or not in valve in APL actuator.
correct position.
Defective drive gas supply 1. Check tubing.
tube to emergency ventilation
section.
66 76 980_02 6 - 166
6. TROUBLESHOOTING
The set APL value will not When Emergency Ventilation 1. Set Mechanical APL to
correspond to the value is switched Off, the minimum.
displayed on the Control Mechanical APL must be set
panel. to minimum. If the
Mechanical APL is set to a
higher pressure, the
Emergency APL Actuator
may not be de-pressurized
through the Excess pressure
vent in the regulator. This
may affect the normal APL
function.
The APL knob is not 1. Perform an APL knob
calibrated. Calibration will be calibration according to
lost e.g. during a System SW instructions in chapter Service
installation. procedures.
The APL potentiometer is 1. Check/replace the APL
defective. potentiometer.
PC 1907 Valve drivers is 1. Check/replace PC 1907.
defective.
66 76 980_02 6 - 167
NOTES
66 76 980_02
7. PREVENTIVE MAINTENANCE
66 76 980_02 7-1
7. PREVENTIVE MAINTENANCE
General
Only personnel trained and authorized by MAQUET shall be permitted to perform installation,
service or maintenance of the FLOW-i anesthesia system.
Preventive maintenance of the FLOW-i anesthesia system must be performed by personnel
trained and authorized by MAQUET at least once a year, or every 5000 hours of operation,
whichever comes first.
The power backup battery shall be replaced every three years.
The memory backup batteries on PC 1920, PC 1921 and PC 1922 shall be replaced every
five years.
It is recommended that a regular cleaning should be performed before carrying out Preventive
maintenance. Refer to instructions in the User's Manual.
In some parts of the Preventive maintenance, as described in this chapter, access to Service
& Settings (S&S) is required.
Some optional equipment used with the FLOW-i, but not covered by this Service Manual, may
also demand maintenance actions. Refer to the optional equipments documentation.
Required equipment
Standard service tools.
Regulator tester, P/N 66 85 355.
Lubricants should normally not be used when servicing the unit. If lubricants must be used,
use only very small amounts of grease with P/N 95 73 700. This special grease is O2
compatible.
Calibration gas kit including gas bottle, regulator and tubing. P/N 52 07 415.
Gas evacuation system.
Barometer (or information about the actual barometric pressure).
Access to the Service & Settings (S&S) is required.
‘Maintenance Kit FLOW-i 12 Months’ containing all parts needed during the maintenance.
Handling PC boards
Avoid contact with PC boards when performing maintenance inside the FLOW-i.
The PC boards contain components that are highly sensitive to static electricity.
Those who come into contact with circuit boards containing sensitive components must take
certain precautions to avoid damaging the components (ESD protection).
When working with ESD sensitive components, always use a grounded wrist band and
grounded work surface. Adequate service tools must also be used.
PC boards (spare parts) must always be kept in protective packaging for sensitive electronic
devices.
PC boards must not be inserted or removed while mains power or backup battery power is
applied to the PC boards.
Remove and insert the PC boards very carefully to avoid damage to the connectors.
66 76 980_02 7-2
7. PREVENTIVE MAINTENANCE
66 76 980_02 7-3
7. PREVENTIVE MAINTENANCE
66 76 980_02 7-4
7. PREVENTIVE MAINTENANCE
Preparations
Make a general inspection/visual check of the FLOW-i anesthesia system for external defects
or damages. Replace defective or damaged parts.
Make sure that gas supply hoses, connectors and color coding are correct.
Check and note Serial number, System software version and Operating time.
Check that a 'FLOW-i anesthesia system – User's Manual' corresponding to the installed
System software version is present. Also check that operating manuals for optional equipment
connected to the FLOW-i are present.
Check if there are any unexpected 'Technical alarms' in the Logs.
Service & Settings is required (Logs > Service logs > Technical alarms).
Perform a System Checkout to make sure that the FLOW-i work properly before the
maintenance.
Switch Off the FLOW-i.
Disconnect the mains power cable.
Disconnect the gas supplies (wall and/or cylinder).
Remove patient tubing (incl. gas sampling tube and water trap) and the manual breathing bag.
If fitted, remove bacteria filter(s) from the patient cassette.
66 76 980_02 7-5
7. PREVENTIVE MAINTENANCE
Parts to be cleaned
WARNING! Ensure that the EVAC system is correctly connected. Running the
system with a poor connection may result in anesthetic agent being emitted into the
operation environment. Leakages in the EVAC system will not be detected by the
System Checkout.
66 76 980_02 7-6
7. PREVENTIVE MAINTENANCE
Parts to be replaced
The letters A – M in the text below refers to the description of the Maintenance kit contents
above.
66 76 980_02 7-7
7. PREVENTIVE MAINTENANCE
66 76 980_02 7-8
7. PREVENTIVE MAINTENANCE
E – Volume Reflector
1. Replace Reflector Socket Silicone
Seal.
66 76 980_02 7-9
7. PREVENTIVE MAINTENANCE
F – EVAC
1. Replace EVAC silicone tube (angled
tube)
66 76 980_02 7 - 10
7. PREVENTIVE MAINTENANCE
J – Fan
1. Replace Dust Filter
66 76 980_02 7 - 11
7. PREVENTIVE MAINTENANCE
66 76 980_02 7 - 12
7. PREVENTIVE MAINTENANCE
Parts to be checked
WARNING! With mains power supply connected to the unit, there are energized
electrical components inside the equipment. All personnel must exercise extreme
caution when in the vicinity of this equipment if maintenance, fault tracing or
adjustments are performed with mains power supply connected and with the
FLOW- covers opened or removed.
CAUTION: The anesthetic gas and the calibration gas contain substances that may
be detrimental to your health. During use of such gases, assure that the FLOW-i is
connected to an effective gas evacuation system, e.g. the hospital’s AGSS system.
System status:
Mains power cable disconnected
Gas supply disconnected
System switched off.
1. Check cables
Check that the following power cables and connectors are not damaged or loose:
Mains power
AC/DC Converter
Power backup battery
Isolation Transformer
Check that the Control Panel cable and connectors are not damaged or loose.
For units with an optional Patient Monitor; check that the Patient Monitor cables and
connectors (mains power and communication) are not damaged or loose.
For units with an optional Backup Gas Trolley; check that the cable and connector (cable
connected to PC 1903 External Connectors) is not damaged or loose.
System status:
Mains power cable connected
Gas supply connected
System switched off.
66 76 980_02 7 - 13
7. PREVENTIVE MAINTENANCE
System status:
Mains power cable connected
Gas supply connected
System switched to Standby.
5. Fan
Check that the Fan works. Make sure that the cooling air flow is directed into the unit. The fan
rotation is monitored and the system will not start with a malfunction in the fan.
7. Control Panel
Check the screen readability:
Transparency
Surface
Brightness (backlight). Using S&S, check backlight operating time. Estimated lifetime (with
acceptable brightness level) for the lamps is 30000 hours. The time meter must be reset
after replacement of the backlight lamps. The replacement date is set using Service &
Settings (S&S > Service report > Replaced backlight).
Check if pixels on the screen are defective. A few defective pixels can be accepted. Check
that defective pixels are not concentrated to a small area thus reducing the readability in this
area.
Perform the checks listed below and make sure that the software responds to these actions.
Check the touch screen functions. Press buttons on different parts of the screen.
Check the Rotary knob. Turn and press the Rotary knob.
Check the Membrane buttons. Press all buttons. Note that some of the buttons are active
with the system in running mode only. E.g. the Audio pause button can only be checked
during an alarm condition.
8. Barometric pressure
Check that the barometric pressure value
(Service & Settings > Calibration >
Barometer > Current barometer pressure)
corresponds to the actual barometric
pressure value at the local site. The current
pressure may not differ more than ±5% from
the actual barometric pressure.
If required, enter a New pressure value.
Barometer calibration is described in chapter
'Service procedures'.
66 76 980_02 7 - 14
7. PREVENTIVE MAINTENANCE
System status:
Mains power cable connected
Gas supply connected
System switched to a ventilation mode.
System status:
Mains power cable connected
Gas supply connected
System switched off.
Re-assemble the system (covers, doors and optional equipment that has been removed).
66 76 980_02 7 - 15
7. PREVENTIVE MAINTENANCE
12. Carrier
Check the Control Panel arm friction. Adjust if required.
Check the Patient Monitor mounting friction (option).
Check that the wheels and wheel breaks work properly.
Check that the drawers work properly.
Check that the lift (C30 only) works properly.
Check that any optional equipment is properly mounted onto the carrier:
Auxiliary O2 and suction module.
Patient Monitor mounting.
Adjustable arm for patient monitor parameter modules. Check arm friction and module
bracket friction. Adjust if required.
Side rails.
Manual breathing bag support arm.
Additional table. Check arm friction and table friction. Adjust if required.
Rear handle.
13. Vaporizer
On each Vaporizer to be used on the
system:
Check that the O-ring (1) is mounted in
the Docking valve and that the O-ring is in
good condition.
66 76 980_02 7 - 16
7. PREVENTIVE MAINTENANCE
Safety inspection
Check that country specific warning labels are present.
Check that accessible fuses have correct values.
Perform a leakage current test. The use of an Electrical Safety Tester is recommended. The
leakage current test is a standard procedure regulated by IEC/EN 60601-1 or corresponding
national standards. Allowable values and test methods are defined in the standard IEC/EN
60601-1 Class 1, Type B.
Create and save a Service report in Service & Settings. Report performed actions:
Performed preventive maintenance. This will also reset the preventive maintenance time
counter in the Status menu.
Replaced expiratory membrane. This will also reset the remaining capacity meter for the
APL/PEEP Valve membrane in the Status menu. If the unit is equipped with several Patient
Cassettes, the remaining capacity meter must be reset on each cassette. Repeat the
Service report > Replaced expiratory membrane on each cassette and enter the cassette
S/N in the text field.
Replaced control panel backlight. This will also reset the Backlight operating time counter in
the Status menu.
Replaced power backup battery. This will also update the power backup Battery replaced
date in the Status menu.
Replaced memory backup batteries.
Note in the units log book that a Preventive Maintenance has been performed.
66 76 980_02 7 - 17
NOTES
66 76 980_02
8. INDEX
66 76 980_02 8-1
8. INDEX
Index
TBD
66 76 980_02 8-2
9. SERVICE MANUAL REVISION HISTORY
66 76 980_02 9-1
9. SERVICE MANUAL REVISION HISTORY
Page
Section Comment
X 1-1 General The Service Manual is based on
System version 1.2.
X 1-2 Hazard notices Warning regarding pressurized
components.
X 2-2 General Model C40 now included.
X 2-13 AFGO AFGO – new option.
X 2-15 Backup gas holder Backup gas holder – new option.
X 2-16 Patient monitor mountings Previously called ‘Patient monitor’,
but covers only the mechanical
mountings.
X 3-58 Emergency APL Pressure Scale (SP-30-80) removed from
Regulator the Emergency Ventilation panel.
X 3-80 PC 1922 Panel RS232 serial port (FCI 2) enabled.
X 3-83 AFGO AFGO – new option.
X 3-90 Backup gas holder Backup gas holder – new option.
X 3-92 Patient monitor mountings Revised description.
X 4-2 Preparations Warning regarding pressurized
components.
X 4-3 Replacing electrical Information regarding replacement
components of electrical components
X 4-5 Disconnecting and connecting Information on how to connect
internal tubing pressure transducer tubing.
X 4-8 Front door Information on how to reassemble
the unit and close the front door.
X 4-9 Side cover Information on how to remove the
side cover.
X 5-2 Tools CO2 always displayed in %.
X 5-2 Tools Tools menu now also available in
operational mode.
X 5-6 Software option installation New section in the manual.
X 5-10 APL knob calibration Information clarified.
X 5-11 Barometer calibration Previously not described in the
manual.
X 5-12 Gas analyzer and O2 sensor Calibration of gas analyzers now
calibration included in Service & Settings.
Information adapted to the new
procedure (pages 5-12 – 5-19).
X 5-20 Touch screen calibration New section in the manual.
X 5-21 AFGO leakage check New section in the manual.
66 76 980_02 9-2
9. SERVICE MANUAL REVISION HISTORY
Corrected
Changed
Added
Page
Section Comment
X 5-32 Pressure regulators REG1– Information adapted to the new
REG4 calibration procedure.
X 5-34 Mechanical APL calibration Information adapted to the new
design and calibration procedure.
X 5-41 Installation of optional New section in this manual.
equipment General information regarding
installation of optional equipment.
X 5-42 Interchangeability – Control New section in this manual – new
panel and Patient monitor option.
X 5-46 Model C40 – Prerequisites and New section in this manual.
handling General information regarding
Model C40.
X 5-48 Functional check of backup gas New section in this manual.
supply Information regarding functional
checks of the optional backup gas
supply.
X 6-3 Troubleshooting – System Information adapted to new and
Checkout changed tests in the System
Checkout procedure (pages 6-3 –
6-39).
X 6-40 Troubleshooting gas analyzers Information clarified.
X 6-41 Technical Error codes New layout to avoid fold-out pages
(pages 6-41 – 6-78).
X 6-49 Technical Error codes TE 59 and TE 60: Information
clarified.
X 6-52 Technical Error codes TE 77: Information clarified.
X 6-57 Technical Error codes TE 516: Information clarified.
X 6-58 Technical Error codes TE 517: Information clarified.
X 6-70 Technical Error codes TE 922 and TE 972: Information
clarified.
X 7-14 AFGO leakage check New check included in Preventive
Maintenance.
X 7-15 Backup gas trolley and Backup Information extended to cover
gas holder Backup gas holder.
New check of gas tube
connections included.
X 10-8 Service & Settings menu New version of the diagram,
adapted to System version 1.2.
Previously included in chapter 5.
Service procedures.
X 10-9 FLOW-i System Overview New version of the System
Overview, adapted to System
version 1.2.
66 76 980_02 9-3
9. SERVICE MANUAL REVISION HISTORY
Page
Section Comment
X 1-1 General The Service Manual is based on
System version 2.1.
X 2-6 Functional units AFGO valve is now an option.
X 2-9 Functional units New design of Oxygen connector
block.
X 2-10 Functional units System checkout valve removed.
X 2-14 Functional units AFGO valve is now an option.
X 2-17 Optional equipment New options added to the list.
X 2-19 Electronic subsystems Event logging removed from
Monitoring.
X 2-20 Electronic subsystems Event and Diagnostics logging
moved to Panel.
X 3-10 Gas control section EMV3 changed due to removal of
System checkout valve.
X 3-28 Connector block and Safety AFGO valve is now an option. The
valve valve is replaced by the Fresh gas
connector block.
X 3-56 Emergency manual New design of Oxygen connector
ventilation section block.
X 3-60 Emergency manual New design of Oxygen connector
ventilation section block.
X 3-61 System checkout System checkout valve removed.
X 3-83 AFGO AFGO valve is now an option.
X 4-4 Replacing PC boards Information stored on PC boards –
new table.
X 4-5 Special grease Information regarding special
grease rephrased.
X 5-4 System software installation Information updated and revised.
X 5-10 Logs New log structure. Information
previously not included in the
manual.
X 5-10 Tests New functions added. Information
previously not included in the
manual.
X 5-15 Calibrations Calibration chapter re-structured
(pages 5-15 – 5-27).
X 5-17 Barometer Barometer calibration procedure
changed.
X 5-24 Patient O2 sensor Description of the calibration
procedure revised (extended).
66 76 980_02 9-4
9. SERVICE MANUAL REVISION HISTORY
Corrected
Changed
Added
Page
Section Comment
X 5-28 Service functions New functions added. Information
previously not included in the
manual.
X 5-29 Service report New functions added. Information
previously not included in the
manual.
X 5-30 Startup configuration Information previously not included
in the manual.
X 5-29 Settings New functions added. Information
previously not included in the
manual.
X 5-32 Leakage detection Chapter name changed from
‘Manual leakage check’ to
‘Leakage detection’.
X 5-33 Test case 2 – CO2 Absorber Note regarding compressible
volume added.
X 5-52 Maximum accessory Table updated.
configuration
X 5-59 Functional check of backup Note 1 new.
gas supply
X 6-2 General Chapter Troubleshooting is based
on System version 2.1.
X 6-3 System Checkout System checkout procedure
changed in System version 2.1.
Information chapter ‘System
Checkout’ updated and re-
structured (pages 6-3 – 6-17).
X 6-18 Test results log The Test result logs are re-
structured and extended in System
version 2.1. Chapter ‘Test results
log’ shows interpreting of the logs
(pages 6-18 – 6-20).
X 6-21 Test results log Test result log table with CFI-
number, Possible cause and
Recommended action (pages 6-21
– 6-119). Separate index on page
6-20.
X 6-120 Technical error codes and Technical error code table updated
messages (pages 6-121 – 6-160).
Information is changed/new for the
following TE-codes:
16
24
27
29
57
69
70
77
83
88
66 76 980_02 9-5
9. SERVICE MANUAL REVISION HISTORY
Corrected
Changed
Added
Page
Section Comment
510
512
516
606
900 – 976. Vaporizer log
information revised.
X 7-2 Required equipment Information regarding special
grease rephrased.
X 10-8 Service & Settings menu New version of the diagram,
adapted to System version 2.1.
X 10-9 FLOW-i System Overview System Overview for FLOW-i units
with S/N below 1500.
X 10-0 FLOW-i System Overview System Overview for FLOW-i units
with S/N above 1500.
66 76 980_02 9-6
10. DIAGRAMS
66 76 980_02 10 - 1
`êáíáÅ~ä=`~êÉ=^_
`êáíáÅ~ä=`~êÉ=^_
`êáíáÅ~ä=`~êÉ=^_
`êáíáÅ~ä=`~êÉ=^_
User category: SSU, SSP
Access code: * * * *
Language: Local
Overview
Central gas pressure PATIENT PART PART -PS LOGS System checkout: -1. APL to SP -Current Patient System Valves - Connect SERVICE TASKS: -Date format - Airway PAGE 1 Waveform 1: -Factory default VENTILATION VENTILATION Startup Startup 1(4) Startup 2(4) Startup 3(4) Startup 4(4) -Current code Date -ISO
-Air Software version VERSION VERSION NO. -AFGO Alarms -O2 flush -2. APL to 30 barometer gas analyzer -Manual to AION... -Performed preventive -Time format pressure (Pressure) -Pressure 1(2) 1(2) -Start button Patient range: -Factory default Activate N2O -Enter new code -Day -SEV
VENTILATION
-N2O AION ARTICLE -PRVC Ventilation settings -Insp. & Exp. valves -3. APL to 80 pressure Ventilation maintenance -Decimal - Flow -Ppeak -Adjust : -Adult/ Infant Gas mix -Patient range -Confirm new code -Month -DES
-O2 -Serial number -SIMV (PC) Events -Vaporizer 1 -New Control -Fresh gas -Replaced expiratory separator - Dynamic -PEEP Waveform 2 ... % AUTO AUTO Ventilation mode: -O2/N2O -Year
SERIAL NO. Vaporizer 1 -Volume setting
-Hardware revision + PS -Vaporizer 2 barometer gas analyzer safety membrane -Language compliance -Pmean -Flow -Factory default -Factory default Volume setting: -Volume control -O2/Air -Breath cycle setting
-Software version
Backup gas pressure -Operating time -SIMV (VC) SERVICE LOGS -Internal tests pressure -AFGO -Replaced backlight -Medical gas - CO2 -Pplat -Volume -Tidal volume -Pressure control -Infant age Time
Patient cassette -FPGA version
-Air -Gas measurement bench + PS Test results -Barometer Control -PEEP/APL -Replaced internal battery colour code - Gas supply -CO2 Lower/ Upper Lower/ Upper -Minute volume -Pressure support O2 concentration -Adult age -Hour
Vaporizer 1 -Boot software
-N2O -Serial number Technical alarms -Gas supply pressure O2 sensor -Reset internal battery pressure (Flow) -O2 Lower Lower -PRVC - ...% -Minute
Vaporizer 2
-O2 -Hardware revision Configuration -Pressure transducer Vaporizer errors - Barometric -RR -Agent -MVe -MVe Breath cyde -SIMV (PC) + PS -Second
Circuit boards Vaporizer 2 -Replaced memory Basic
-Operating time Installation -Safety valve valves pressure -I:E -Blank -RR -RR setting: -SIMV (VC) + PS Vaporizer slot
-Software version backup batteries -Tidal Volume
Vaporizer 1 -Pump module Service reports -Vaporizer inlet/outlet valve -Vaporizer 1 -Ti -PEEP -PEEP -Insp. time/ -1
-FPGA version -Write comment and -Resp. Rate
-Agent -Serial number -Flow transducer in/out -Ti/Ttot Waveform 3 -i : E ratio Parameter Widgets -2
-Operating time -Boot software signature -PEEP
-Liquid level -AUTO ventilation leakage -Vaporizer 2 -Flow Upper Upper for choosen
-Status -Calibration dates -MAN ventilation leakage in/out (Volume) -Volume Infant age ventilation mode Fresh gas flow -O2 conc
-Ppeak -Ppeak
-Factory View latest -Gas analyzer -Vaporizer -MVe -Agent -MVe -MVe Adult age - ... l/min
Vaporizer 2 -Last calibration Time selection -Battery Bypass -MVi -CO2 I:E
-RR -RR
-Agent -Last calibration check -Start: day, month, year. -Technical alarms -Vaporizer 1 -VTi -O2 -I:E
-PEEP -PEEP
-Liquid level -End: day, month, year. pressure -VTe -Blank -T pause
-Status sensor Control: -Vaporizer 1 -Vèe -T insp. rise
Apnea time Apnea time
O2 -Serial number -Control disable vaporizers Scavenging Waveform 4 (only in (only in
Battery status -Last calibration -Vaporizer 1 -Vaporizer 2 (Lungmech. -Flow Trigger
Pressure support) Pressure support)
-Remaining time -Vaporizer 2 pressure calc.) -Volume -Trigger sensitivity
-Voltage -APL/PEEP valve -Vaporizer 2 -C dyn -CO2 -insp. cycle off
CONTROL GAS GAS
- Battery replaced -Control pressure transducer Scavenging -O2 2(2) 2(2)
Software version
-Control safety valve -Vaporizer 1 lock -Blank -Agent -Factory default -Factory default
Serial number AION -Vaporizer inlet/outlet valve -Vaporizer 2 lock -Blank
-Serial number GAS
System software version -AFGO PAGE 2 Lower/ Upper Lower/ Upper -Activate N2O
-Hardware revision
Operating time -AUTO ventilation leakage (Pressure) Waveform 5 Lower Lower -Gas mix
-Operating time
Next preventive- -MAN ventilation leakage -Ppeak -Flow -CO2 Et -CO2 Et -O2 conc.
-Gas measurement bench
maintenance -PEEP -Volume -O2 insp. -O2 insp. -Vaporizer slot
-Serial number
Monitoring: -Pmean -CO2 -O2 Et -O2 Et -Fresh gas flow
-Hardware revision
Patient cassette status -Water trap -Pplat -O2 -DES insp. -DES insp.
-Operating time
- APL/PEEP valve -Mains connection -Agent -DES Et -DES Et
-Pump module
membrane -Power failure (Flow) -Blank -ISO insp. -ISO insp.
-Serial number
-Remaining capacity -Operating time -Barometer -I:E -ISO Et -ISO Et
-First day of use -Calibration dates -Gas supply pressure -Ti Waveform 6 -SEV insp. -SEV insp.
-Operating time -Factory -Monitoring pressure transducer -Ti/Ttot -Flow -SEV Et -SEV Et
-Last calibration -Monitoring safety valve -Volume
Control panel status -Last calibration check -Gas analyzer (Volume) -CO2 Upper Upper
-Back light operating time -Battery -MVi -O2 -CO2 insp. -CO2 insp.
O2 sensor -Monitoring disable vaporizers -VTi -Agent -CO2 Et -CO2 Et
-Serial number -Technical alarms -VTe -Blank -O2 insp. -O2 insp.
-Last calibration -Absorber switch -Vèe -O2 Et -O2 Et
Loops -DES insp. -DES insp.
(Lungmech. -On -DES Et -DES Et
Panel: calc.) -Off -ISO insp. -ISO insp.
-Panel audio - dyn -ISO Et -ISO Et
-Blank Rotameter: -SEV insp. -SEV insp.
Leakage check: -On -SEV Et -SEV Et
-Multiple pressure leakage -Off
MAC
-Internal leakage
Enable MAC
-AFGO leakage
presentation:
-AUTO ventilation leakage On/ Off
-MAN ventilation leakage
FLOW-i Serial Number: Below 1500 Date: 2011.10.03 System Overview
A B C D E F G H
XXX-
Xxxxx = Internal serial communication
PC 1900 Main Back-plane
26
Backup Gas Holder 33 To connector
P42 on PC 1903
Backup Gas Trolley 32 Gas Control From AFGO Pilot Manual Ventilation switch APL Vaporizer (VAP1) Vaporizer (VAP 2) Gas Analyzer Section 10
XXX
Xnn = Connector on
Cylinder Gas
Valve EMV18 MAN PC 1900 Main Back-plane
Inlet Filter
External Connectors Section 6 Vaporizer Drive Docking
6 Vaporizer Drive Docking
Cylinder Gas Inlet Filter
O2 Cylinder Gas
N2O O2 Cylinder Pressure Regulator AUTO PC Lid Gas Filter Valve PC Lid Gas Filter Valve Zero calibration gas inlet
Safety Fill 1931 Sensor Safety Fill 1931 Sensor
Cylinder Gas Gas One-
Pressure N2O Cylinder
way Valve 3 P62 P70
Valve Valve S2 R2 Sampling
block
(supplied from Emergency
Ventilation section) Power On/Off and 21
PC1910 PC1910
P64
Transducer OV9 OV9
Gas Pressure Pin Index Vaporizer Gas Vaporizer S R Vaporizer PC boards
P63
O2 Exp. Pressure Gas P27 Control Gas Analyzer
Regulator Heating Foil Heating Foil
"8051"
Escape
1
REG21 PC 1907 Escape AION (CON)
"8051"
from Patient S1 R1
CPU
SV21 T T I2C-
CPU
REG23 OV21 Valve Drivers Pipe Pipe
O2 In 650 kPa Cassette P60 P61 P65 P66 From Absorber Sensor
Filter Vaporizing Filter Vaporizing Control 0 X 1 Panel
I/O-port ”001”
OV23 O2 Cylinder O2 Cylinder Gas O2 Sensor
Gas Pressure Pe P67 PP3 Chamber IN Chamber IN 2 E2Prom ”001”
Pin Index Safety Valve EMV2 Servomex
E2Prom ”001”
PC 1912
PC 1913
PC 1912
R1
PC 1913
Liquid S1
650 kPa O2 O2 N2O AIR P57 I/O port “000” Container VI Container VI
S
R2
R 5 From O2
N2O Cylinder E P58 S2 Flush Valve From
I2C Vap 1
PC 1916
I2C Vap 2
ADC ”01” P28 Power
Gas Safety Valve Cylinder Pressure Connection P56 Vaporizer Vaporizer Control Gas Control Gas Emergency
PC1781 EMV4 PV5 button
P69 Injector Injector Analyzer Analyzer Ventilation
P55 E2Prom “111” P P Power
AIR/N2O AIR/N2O Liquid Agent SV2 Liquid Agent SV2 Inlet Outlet Switch
Cylinder Gas
On LED
PC 1911
PC 1911
Inlet Filter
N2O Cylinder Gas Cylinder Gas
Pressure Pressure
P54 I2C-
Panel
I2C-
Monitor
P50
MOD1
P57
PV3
Filter
Vaporizer Vaporizer Filter Vaporizer Vaporizer
Valve Valve Control Gas Analyzer P29
P53 EMV5 PV4 Drain Pressure Safety Valve Drain Pressure Safety Valve
Transducer Regulator I2C- Plug Transducer Plug Transducer
P68 Water trap
Patient Monitor Control P14
P56
Sampling line
Water trap
receptacle
P149 P151 P154 P153 P152
P259
Pressure Supply Pilot Valves: N30 Trap W1
Gas One-way Valve 650 kPa EMV2 = Manual Ventilation Pilot Valve
EMV7 PP1 Valve P15
AIR/N2O Cylinder Detector
EMV3 = System Checkout Pilot Valve
Gas Safety Valve EMV4 = Vaporizer Bypass Pilot Valve P54 EMV9 P81 P30 P82 N1A
Vaporizer 1 Zero calibration gas inlet
AIR connected in diagram EMV5 = Vaporizer 2 Selection Pilot Valve EMV8 PP2 Scavenging PC 1909
2
EMV6 = Vaporizer 1 Selection Pilot Valve P66 Vaporizer Adapter (supplied from Emergency N1B P1B P1A
Valve
AIR/N2O EMV7 = Vaporizer 1 Lock Pilot Valve
P53 PV1 PV2 PV3 PV4
Ventilation section)
EMV8 = Vaporizer 2 Lock Pilot Valve
PV3
PV4
PV1
PV2
P31 Patient Gas Analyzer CAN 1 P108
E1 EMV10 AION (MON)
PV2 Vaporizer 2 Patient CPU I2C-Onboard
0 1 I2C- E2Prom “000”
To connector P42 on PC 1903 External Connectors Pressure O2 Sensor Exp. 168
P60
Valve Servomex 2 chan.
PV5 Pm1111E O2
Instant O2 Supply Pressure T4 3 I2C-
V2 E2Pot “01”
Lift 27 Gas Distribution Section 1 REG4
200 kPa
Pressure Regulator TestPoint 4 EMV11 Vaporizer 2
Scavenging
PV5
W2
W3
4
Monitor
Flush P61
Valve Vaporizer Bypass Valve Inspiratory Pressure
R3 5 from Patient
Lift REG3 O2 P32 Pi
Pilot Valve Pressure E2 Cassette
actuator O2 200 kPa Patient Gas Analyzer NTC
PC 1906
Regulators REG2/REG3 Pressure Test
Point 2/3
EMV12
5 Vaporizer Fresh Gas Pressure
from Fresh Gas P111
Gas Block Drive Gas Drive T2/T3 Pf
Supply OV4 Vaporizer Pressure Vaporizer from Vaporizer Insp. Pressure CPU T P8 Display/
S top
Re set
P50 P50
REG1 Exp. Pressure 8051
120 kPa Regulator Drive Gas To PC 1781 To PC 1781 I2C- PC 1925 Touch Trend s
MOD2 MOD1
on PC 1919 Onboard
on PC 1907 Display Panel Insp. Pressure
Auxiliary 29 Pressure
Pressure Test
V1 E2Prom ”000” CPU
S ave
s creen
End
ca s e
Fresh Gas Pressure
Transducer Transducer
Power O2 In Transducer
O2 2 Drive Gas Supply Section Point 1
T1 P9 P10
P110 PC 1919 Exp. Channel
Outlets 22
3
Central Gas Exp.
Pi
Supply P OV1 Reflector Gas Insp.
Backlight
N2A
1A Sample
AUT. 2 A 1 T P Reflector Gas 13 Pe
Filter Sample
R1
P1 P4 Inverter S creen
config.
Menu
Filter N2B
Patient Monitor NTC P48 Supply R3
Holder
Holder
PC 1924
Display
P2B P2A
P2 Connection P3
CAN 1 N116
AIR Out Reflector Volume 12 S1
Filter
O2 Flush
N7 comm. 1 5200
on PC 1907 P104 9 PLD
PV7
Valve Drivers PC 1918
R2
Pressure I2C-
Gas Inlets / Gas Outlets
2A
AUT. 4 A AIR In Transducer
AIR Fresh Gas Fresh Gas
Exp.
Channel Patient Control 48 MB Flash
Power 2 4 Pressure Connector
Cassette
Outlets Central Gas
P OV2 T P Supply Pf To PC 1781
AFGO
Pilot Valve P46 19 Power Section LED Lamp
I2C- I2C 128 MB SDRAM
Supply
O2
on PC 1919 EMV18 Docking Vap 1 Mux 1 MB SRAM
AFGO From connector I2C- Battery
Pressure Fresh Gas Valve PV7 P250 on Fuse 6.3 A AC/DC Vap 2
Filter
4
Transformer Cylinder CAN 1
N2O In Gas Module Docking NTC pressure
PC 1938 7 Pressure P42
Serial transducer
5
P = Pressure measuring point
Absorber Sensor
Filter
Auxiliary O2 P17 APL/PEEP 11 R = Gas return point E2Prom ”000”
Input/Output ports
Silencer Valve Coil S = Gas sampling point RS232 (FCI 2) N136 ADC ”00”
Emergency PP3
T = Temperature measuring point LVDS
N135 64 MB Flash
Ventilation Mechanical Emergency
PP3
S1 - R1 = Control Gas Analyzer
To AC/DC Power Inlet
in External connectors
To connector P146 on
PC 1923 Power Control
O2 l/min. APL APL Actuator CO2 Vaporizer regulating
on Insp (future option) Power backup battery VGA N134
128 MB SDRAM
2 MB SRAM
Manual
Suction
Emergency 10 Ventilation Absorber S2 - R2 = Control Gas Analyzer USB N133
O2 (from O2 Out) AIR (from AIR Out) Valve Fresh Gas measurement Battery
Power In Power Out Ventilation Switch Excess RJ12
pressure Gas analyzer zero Actuator To connector S3 - R3 = Patient Gas Analyzer Ethernet N132 PC 1922 Panel 25
vent calibration gas intake P67 on PC 1907 Y-piece measurement
1 2 1 3 0 0 S3
Valve Drivers Pe/Pf/ = Pe: Expiratory pressure
Isolation 30 AGS Flow EVAC 17 I
Pi/Pr Pf: Fresh gas pressure N5A
Indicator Pi: Inspiratory pressure
Transformer Fresh Gas to Safety Valve
Pr: Reflector pressure N5B P5B P5A
REG5 W1 - W3 = Water trap connections
Manual
Ventilation ADC “01”
To connector P151 on Oxygen Flow Emergency APL
PC 1928 Power On/Off Pressure Regulator Bag AFGO R2 I2C-
Power ADC “00”
12 34
Emergency O2 Flush
CPU
Transformer
Manual Manual P144 T
6
8051
Switching
Relay
EVAC Reservoir Ventilation AFGO
PV7
PATIENT
P142
Ventilation Valve From connector P249 on
PLD
Transformer
Section 15 Section 14 PV7 PC 1938 Mains connection P146 PC 1923 Power Control
Fresh Gas from Vaporizer
FLOW-i Serial
Version: 2.1
Number: Above 1500 Date: 2011.10.03 System Overview
A B C D E F G H
XXX-
Xxxxx = Internal serial communication
PC 1900 Main Back-plane
To connector 26
Backup Gas Holder 33 P42 on PC 1903 Backup Gas Trolley 32 Gas Control From AFGO Pilot Manual Ventilation switch APL Vaporizer (VAP1) Vaporizer (VAP 2) Gas Analyzer Section 10
XXX
Xnn = Connector on
Cylinder Gas
External Valve EMV18 MAN PC 1900 Main Back-plane
Inlet Filter
Section 6 Vaporizer Drive Docking
6 Vaporizer Drive Docking
Cylinder Gas Inlet Filter
Connectors O2 Cylinder Gas
N2O O2 Cylinder Pressure Regulator AUTO PC Lid Gas Filter Valve PC Lid Gas Filter Valve Zero calibration gas inlet
Safety Fill 1931 Sensor Safety Fill 1931 Sensor
Cylinder Gas Gas One-
Pressure way Valve 3 P62 P70
Valve Valve S2 R2 Sampling
block
(supplied from Emergency
Ventilation section) Power On/Off and 21
PC1910 PC1910
P64
Transducer OV9 OV9
N2O Cylinder Pin Index Vaporizer Gas Vaporizer S R Vaporizer PC boards
P63
O2 Exp. Pressure Gas P27 Control Gas Analyzer
Gas Pressure Heating Foil Heating Foil
"8051"
Escape
1
REG21 PC 1907 Escape AION (CON)
"8051"
from Patient S1 R1
CPU
Regulator T T I2C-
CPU
OV21 SV21 Valve Drivers Pipe Pipe
O2 In 350 kPa
650 kPa Cassette P60 P61 P65 P66 From Absorber Sensor
Filter Vaporizing Filter Vaporizing Control 0 X 1 Panel
I/O-port ”001”
OV23 O2 Cylinder O2 Sensor
Gas Pressure O2 Cylinder Gas Pe P67 Chamber IN Chamber IN 2 E2Prom ”001”
Pin Index Safety Valve E2Prom ”001” Servomex
PC 1912
PC 1913
PC 1912
R1
PC 1913
Gas One-way Valve Liquid S1
SV23 O2 O2 N2O AIR P57 I/O port “000” Container VI Container VI
S
R2
R 5 From O2
E S2 Flush Valve From
I2C Vap 1
650 kPa PC 1916
I2C Vap 2
ADC ”01” P28 Power
Cylinder Pressure Connection P56 Vaporizer Vaporizer Control Gas Control Gas Emergency
N2O Cylinder PC1781 EMV4 PV5 button
P69 Injector Injector Analyzer Analyzer Ventilation
Gas Safety Valve P55 E2Prom “111” P P Power
AIR/N2O AIR/N2O Liquid Agent SV2 Liquid Agent SV2 Inlet Outlet Switch
Cylinder Gas
On LED
Inlet Filter
PC 1911
PC 1911
N2O Cylinder Gas Cylinder Gas
Pressure Pressure
P54 I2C-
Panel
I2C-
Monitor
P50
MOD1
P57
PV3
Filter
Vaporizer Vaporizer Filter Vaporizer Vaporizer
Valve Valve Control Gas Analyzer P29
P53 EMV5 PV4 Drain Pressure Safety Valve Drain Pressure Safety Valve
Transducer Regulator I2C- Plug Transducer Plug Transducer
P68 Water trap
Patient Monitor Control P14
P56
Sampling line
Water trap
receptacle
P149 P151 P154 P153 P152
P259
SV22 Pressure Supply Pilot Valves: EMV7 PP1 N30 Trap W1
Gas One-way Valve EMV2 = Manual Ventilation Pilot Valve Valve P15
650 kPa Detector
EMV4 = Vaporizer Bypass Pilot Valve
AIR/N2O Cylinder EMV5 = Vaporizer 2 Selection Pilot Valve P54 EMV9 P81 P30 P82 N1A
Gas Safety Valve Vaporizer 1 Zero calibration gas inlet
EMV6 = Vaporizer 1 Selection Pilot Valve EMV8 PP2 Scavenging PC 1909
2
EMV7 = Vaporizer 1 Lock Pilot Valve P66 Vaporizer Adapter (supplied from Emergency N1B P1B P1A
AIR connected in diagram Valve
AIR/N2O EMV8 = Vaporizer 2 Lock Pilot Valve
P53 PV1 PV2 PV3 PV4
Ventilation section)
PV3
PV4
PV1
PV2
P31 Patient Gas Analyzer CAN 1 P108
E1 EMV10 AION (MON)
PV2 Vaporizer 2 Patient CPU I2C-Onboard
0 1 I2C- E2Prom “000”
To connector P42 on PC 1903 External Connectors Pressure O2 Sensor Exp. 168
P60
Valve Servomex 2 chan.
PV5 Pm1111E O2
Instant O2 Supply Pressure T4 3 I2C-
V2 E2Pot “01”
Lift 27 Gas Distribution Section 1 REG4
200 kPa
Pressure Regulator TestPoint 4 EMV11 Vaporizer 2
Scavenging
PV5
W2
W3
4
Monitor
Flush P61
Valve Vaporizer Bypass Valve Inspiratory Pressure
R3 5 from Patient
Lift REG3 O2 P32 Pi
Pilot Valve Pressure E2 Cassette
actuator O2 200 kPa Patient Gas Analyzer NTC
PC 1906
Regulators REG2/REG3 Pressure Test
Point 2/3
EMV12
5 Vaporizer Fresh Gas
from Vaporizer
Fresh Gas Pressure
from Fresh Gas P111
Gas Block Drive Gas Drive T2/T3 Pf
Filter N2B
Patient Monitor NTC P48 Supply Reflector R3
Holder
Holder
PC 1924
Display
P2B P2A
P2 Connection P3
S1 CAN 1 N116
Reflector
AIR Out
Filter
2A
AUT. 4 A AIR In Transducer
Fresh Gas Fresh Gas
Exp.
Channel Patient Control 48 MB Flash
O2 Flush
AIR 2
Power
Central Gas
Supply 4
Pressure Connector
Cassette 19 Power Section I2C- I2C 128 MB SDRAM
PV7
T P To PC 1781 P46 LED Lamp
R2
Outlets P OV2 Pf
Supply
O2
on PC 1919 Volume X
Docking Vap 1 Mux 1 MB SRAM
Reflector From connector I2C- Battery
Fresh Gas Reflector Fuse 6.3 A AC/DC Vap 2
Pressure Fresh Gas P250 on
Filter
4
Transformer
N2O In Gas Module Docking 7 Cylinder
Pressure P42
CAN 1
NTC pressure
PC 1938 Serial transducer
R
Absorber Sensor
Filter R = Gas return point
Auxiliary O2 P17 11 S = Gas sampling point
E2Prom ”000”
Input/Output ports
APL/PEEP RS232 (FCI 2) N136 ADC ”00”
Silencer Valve Coil PP3 T = Temperature measuring point
S1 - R1 = Control Gas Analyzer LVDS 64 MB Flash
Vaporizer regulating N135
PP3 on Insp (future option)
To AC/DC Power Inlet
in External connectors
To connector P146 on
PC 1923 Power Control Emergency CO2 S2 - R2 = Control Gas Analyzer Power backup battery VGA N134
128 MB SDRAM
2 MB SRAM
Manual Fresh Gas measurement
Suction
Emergency
APL Actuator
Ventilation Absorber S3 - R3 = Patient Gas Analyzer USB N133
O2 (from O2 Out) AIR (from AIR Out) Valve Y-piece measurement Battery
Power In Mechanical
Power Out
Ventilation APL Actuator To connector
P67 on PC 1907
Pe/Pf/ = Pe: Expiratory pressure
Pi/Pr Pf: Fresh gas pressure
RJ12
Ethernet N132 PC 1922 Panel 25
1 2 1 3 Pi: Inspiratory pressure S3
To connector P151 on O2 l/min. Valve Drivers Pr: Reflector pressure
Isolation 30 AGS Flow EVAC 17 PC 1928 Power On/Off Excess
Gas analyzer
W1 - W3 = Water trap connections N5A
Indicator pressure
Transformer 0 10
vent zero calibration To connector
Vaporizer
P5A
O2 Flush
P5B
I gas intake P70 on PC 1907 28 N5B
From
PV7
Manual Valve Drivers
R2
ADC “01”
0 Ventilation
Bag
AFGO I2C-
AFGO
Power ADC “00”
12 34 Pilot Valve
REG5 Emergency EMV18
CPU
Transformer Emergency Oxygen Flow Emergency APL
Manual Manual AFGO Additional PATIENTPATIENT P144 T
6
8051
Switching
Relay
EVAC Reservoir Ventilation Switch Pressure Regulator
Ventilation
Valve PV7 Fresh Gas
P142
Ventilation Outlet
To Safety
PLD
Valve
66 76 980_02
© Maquet Critical Care AB, 2010-2011. All rights reserved. MAQUET reserves the right to modify the design and specifications contained herein without prior notice.
Order No.: 66 76 980. Revision 02, October 2011. Price group: 7.