CALIBRATION CERTIFICATE
Date of Service Vessel Name IMO No, Certificate No.
20 DEC 2021 SANDRO 9390525 VMSC-2112-988H
Place of Service Flag Class Service Order No.
SINGAPORE PANAMA KR VMSR-21-2318
TECHNICAL DESCRIPTION:
‘TEMPERATURE CALIBRATOR
Maker : DOVER Serial No. : MT 1408 T0214
Model : 650H Range 2 30-650°C
Calibrated On : 20 DEC 2021
Next Due Calibration : 19 DEC 2022
DESCRIPTION OF INSPECTION / TESTS:
THE CALIBRATION METGHOD WAS CARRIED OUT OF COMPARISON AGAINST THE REFERENCE
STANDARD TRACEABLE TO NATIONAL OF INTERNATIONAL STANDARDS
CALIBRATION RESULTS:
Reading Set from Reading from Reading from Ref, | eovute Deviation
Instrument Instrument Instrument sence Trop
c °c *C °c
30.0 30.0 30.4 ot Wl
700.0 700.0 100.2 02
200.0 200.0 200.3 03
300.0 300.0 300.3 | 03
400.0 400.0 400.4 04
500.0 500.0 500.4 04
600.0 00.0 ann 5 an
650.0 850.0 650.5 05
THE CALIBRATION RESULTS ARE SHOWN IN THE ABOVE TABLE. THE ESTIMATE UNCERTAINTY IS 40.30% FULL
SCALE FOR A CONFIDENCE LEVEL OF APPROXIMATELY 95% WITH A COVERAGE FACTOR OF K = 2
WE HEREBY CERTIFY THAT WE HAVE TESTED THE ABOVE INSTRUMENT IN ACCORDANCE WITH
STANDARD INSPECTION PROCEDURES AND THAT THE QUALITY IS IN FULL CONFORMANCE WITH THE
SPECIFICATIONS AND THE STANDARDS.