Problems of Forensic Sciences 2009, vol. LXXVII, 7–28 ISSN 1230-7483

QUALITY ASSURANCE ASPECTS OF IDENTIFICATION WITH


CHROMATOGRAPHIC – MASS SPECTROMETRIC METHODS

Maciej J. BOGUSZ
Consulting Toxicologist, GTFCh, Sinsheim, Germany

Abstract
The factors affecting identification of compounds with chromatography – mass spectrometry were discussed from the quality as-
surance point of view. The criteria of identification of compounds, as formulated by several international or national organisa-
tions, were discussed. Various recommendations or guidelines defining minimal attributes necessary for identification have been
published. These documents concern both chromatographic and mass spectrometric aspects of analysis. There is general agree-
ment concerning the kind of parameters to be controlled, e.g. retention time, number of ions, or ion intensity ratios. However, the
exact numerical criteria, formulated by different bodies or organisations, show distinct variations. The strictest requirements are
set in those areas where the results of identification may lead to legal sanctions, such as in forensic toxicology, doping analysis, or
food control. Requirements concerning quality assurance in the field of identification with GC-MS or LC-MS vary according to
the particular discipline and final task of the analysis and are continuously changing in relation to actual knowledge.

Key words
Identification; Quality assurance; Mass spectrometry; GC-MS; LC-MS.

Received 3 December 2008; accepted 20 January 2009

1. Introduction – Is there a difference between confirmation and


identification?
Identification of a particular compound is a key is- – What are the requirements for suitable databases?
sue in analytical toxicology. It precedes other steps, – What are the criteria for rejecting substances?
such as quantification, and, above all, interpretation. – What is the probability of correctness of identifica-
According to de Zeeuw and Franke [10], the pro- tion?
cess of identification starts with the comparison of an De Zeeuw and Franke distinguished three types of
unknown substance with a database containing refer- identification: Structure elucidation through identifi-
ence substances with exactly defined properties. The cation of a pure compound by powerful spectrometric
process of identification must continue, using several methods, confirmation as a result of successful com-
criteria, until only one substance remains on the list. parison of the properties of the expected substance
These authors have formulated the following ques- with a reference substance, and recognition as a result
tions, relevant for chromatographic – mass spectro- of positive matching of the properties of the unknown
metric procedures: substance with a reference database. In recommenda-
– How can analytical properties be compared (e.g. tions formulated by the Food and Drug Administration
how to compare mass spectra)? (FDA) [23], confirmation is similarly defined as “un-
– What should be considered an “adequate” match? ambiguous identification of a compound’s presence by
8 M. J. Bogusz

comparison to a reference standard (mass spectromet- taken with utmost care, particularly in the case of some
ric)”. However, the term “confirmation” may be used “exotic” compounds proposed by the instrument soft-
in a different way. In forensic toxicology and doping ware as candidates for identification. However, in most
control, the usual analytical strategy consists of two identification procedures used in legally sensitive fields,
steps; screening procedure, usually performed with such as doping analysis, forensic toxicology, or food
highly sensitive but less specific methods (mainly im- quality monitoring, tentative identification through the
munoassays, or tandem-MS using only one transition), GC-EI-MS library must be followed by a confirmation
and confirmation of the positive result of the screening step, using a reference compound analysed in identical
procedure, which is done with methods of the highest conditions to the studied specimen. In full-scan GC-
possible specificity, such as full-scan GC-MS [46], or MS-EI, at least four selective ions, including – if possi-
LC-MS-MS in MRM mode using three transitions ble – the molecular ion, should be present in proper
or product spectra [26, 48]. A similar definition was abundance ratios. If possible, mass spectra belonging
given by the U.S. Department of Agriculture [58] for to matrix compounds and affecting identification
pesticide identification: “Confirmation: Verification should be filtered out. This may greatly enhance the
of a previous analyte identification that is performed identification power of GC-MS screening [52].
by another analytical system”. De Zeeuw [9] warned However, it must be stressed that the decision con-
against using confirmation procedures as the only proof cerning preliminary selection-identification based on
of identification and postulated the use of various statistical analysis of a mass spectrum match with ref-
identification procedures, not only mass spectrometry. erence library spectra should be always done by the
Nevertheless, procedures based on hyphenation of chro- expert toxicologist. It is s/he who decides, depending
matographic separation with mass spectrometric de- on the case, when the identification level has been
tection are regarded as a sufficient tool for unequivo- achieved [9, 42, 43].
cal identification, if particular requirements are met. Liquid chromatography-mass spectrometry is po-
These requirements, such as establishing the analytical tentially better suited to screening than GC-MS since it
threshold, appropriately high for the particular task, covers a much broader spectrum of compounds of dif-
are formulated by responsible organisations or bodies ferent polarities. However, this technique has several
on a national or international level. important limitations, both on the chromatographic
and mass spectrometric side. LC, despite its universal-
ity, is a far less selective separation technique than
2. Screening procedures applied to capillary GC. On the other hand, mass spectra of com-
identification pounds obtained with various LC-MS instruments
may vary greatly in regard to relative abundance of
Mass spectrometric screening procedures, often fragment ions, even if exactly identical nominal condi-
used as a first step of identification, may be divided tions are applied [3]. This makes direct use of mass
into two main groups: general non-target screening, spectral data acquired in another laboratory very diffi-
and target substance-oriented screening. cult. For these reasons, LC-MS is an excellent tech-
nique for targeted screening or confirmation, but it is
2.1. General screening not easily applicable to a broad-spectrum search, e.g.
“general unknown” analysis. Nevertheless, in recent
In the case of non-target screening, which is often years, several spectral libraries, comprising hundreds
referred to in clinical and forensic toxicology as “gen- of compounds, have been developed using LC-MS
eral unknown” analysis, a capillary GC coupled with [45, 65], LC-MS-MS [31, 64], and LC-MS-TOF [11,
electron impact ionisation mass spectrometer (GC-EI- 21, 38]. Reviews of this topic have been carried out
MS) is a logical choice. Full-scan mass spectra, ob- recently by Gergov [20] and by Marquet [30].
tained with EI-MS, are not very dependent on the in-
strument and chromatographic conditions applied. 2.2. Targeted screening
Comprehensive libraries of reference EI mass spectra,
comprising thousands of substances, are available Targeted screening and subsequent confirmation is
nowadays [35, 36, 59]. This makes tentative identifi- employed if a particular compound or defined set of
cation possible through a library search – even without compounds is to be identified. This is the case in the
a reference compound. In cases where a reference search for scheduled, controlled substances, e.g. drugs
compound for comparative analysis is not available, of abuse [61], doping substances [19, 54], or food con-
a decision concerning the spectrum identity should be taminants [60]. In this situation, very specific meth-

Problems of Forensic Sciences 2009, vol. LXXVII, 7–28


Quality assurance aspects of identification with chromatographic... 9

ods, such as GC-MS-MS and LC-MS-MS, both in 3. Methodical considerations relevant for
MRM mode and also GC-TOF-MS or GC-MS-SIM, identification
are applicable. Identification is achieved through com-
parison of the chromatographic mobility and the pres- There are numerous possible factors which may af-
ence of particular fragment ions in intensities corre- fect identification by mass spectrometry, at various
sponding to preset values, observed in reference stan- stages of the identification procedure; the influence of
dards. Limited, task-oriented reference libraries have coextracted matrix compounds, incomplete chromato-
been built for particular groups of compounds. Com- graphic separation of compounds, general condition of
prehensive reference data libraries, e.g. for pesticides, instrument or detector saturation and peak distortion
based on LC-MS, may comprise several hundred sub- through excess of the substance. All these factors may
stances. change the abundances of ions in a particular mass
The definition of positive identification or confir- spectrum and cause mismatching with consequent
mation after LC-MS screening may vary. Some au- false-positive or false-negative identification [33, 34].
thors have mentioned high coincidence of at least two There are several approaches to optimising the identi-
of three spectra generated at three fragmentation ener- fication process with mass spectrometry. These ap-
gies [28], others a match of at least 60% in a reverse-fit proaches may be divided into optimisation of sample
library search [62] or similarity of at least five of the pretreatment, optimisation of chromatographic separa-
most abundant ions [4]. tion, and optimisation of mass spectrometric detection
For some halogen-containing compounds, such as itself.
chloramphenicol, alternative identification strategies
have been used. Bogusz et al. [2] applied LC-MS-MS 3.1. Optimisation of sample pretreatment for mass
(negative ionisation ESI, MRM mode). Three transi- spectrometric analysis
tions of deprotonated molecule (m/z 321) were moni-
tored. The following criteria of positive identification Optimal, matrix-oriented sample pretreatment is of
of chloramphenicol were formulated: Rt of target with- critical importance in the GC-EI-MS identification
in ± 1% of internal standard (deuterated analogue of procedure. It may involve cleavage of conjugates, deri-
target), the presence of three product ions originating vatisation, or several clean-up steps. Among isolation
from the precursor, and intensity ratios of product ions procedures used in GC-MS screening and identifica-
in the range of ± 2SD of the mean control values, i.e. tion, liquid-liquid extraction [35] and solid-phase ex-
± 25%. Mottier et al. [37] took advantage of the pres- traction (SPE) [16], have mainly been used. In recent
ence of two chlorine atoms in the chloramphenicol years, solid-phase microextraction (SPME) has been
molecule and used two precursor ions: m/z 321 and increasingly frequently applied; it is a solvent-free
m/z 323. The transitions of each isotopic form to prod- method and is particularly useful for screening of vola-
uct ions m/z 152 and 257 were monitored. The identifi- tile compounds of natural or synthetic origin [e.g. 5, 6,
cation criteria were as follows: Rt of target within ± 1% 24, 41].
of internal standard (deuterated analogue of target), For some compounds, particularly large molecules
the variability of the intensity ratios of product ions in with haptenogenic properties, immunoaffinity extrac-
the range of 15–25% of the mean control values. tion procedures have been developed. Ho et al. [26]
Since the intensity and sometimes pattern of frag- isolated insulins originating from various sources (hu-
mentation in LC-MS is related to the particular tech- man, bovine, and porcine) from horse plasma by im-
nique and conditions used, it has generally been postu- munoaffinity precipitation with antibody-coated mag-
lated that the reference standard should be analysed in netic beads followed by molecular weight centrifugal
exactly the same conditions as the examined sample, extraction. The obtained extracts were analysed by
ideally in the same analytical run. LC-MS-MS, using a QTrap mass spectrometer. For
Requirements concerning quality of mass spectro- screening purposes, single transition to one common
metric screening procedures have been reviewed and tyrosine immonium ion was monitored. Confirmation
discussed by various authors, such as Drummer [15], was performed with at least three transitions character-
Stimpfl et al. [50, 51], and Maurer [32] in relation to istic for the particular insulin, also taking retention
clinical and forensic toxicology, and by Thevis and time into account. This method was applied in doping
Schänzer [54] in relation to doping analysis. control analysis, as a tool of differentiation of equine
insulin from exogenous insulin.

Problems of Forensic Sciences 2009, vol. LXXVII, 7–28


10 M. J. Bogusz

3.2. Optimisation of chromatographic separation a UPLC-MS-MS-based screening method of 130 sub-


stances (diuretics, central nervous system stimulants
Capillary GC is a mature technique in relation to and opiates) for direct injections of urine samples.
the efficiency of chromatographic separation. Never- Samples were injected on a reversed phase column
theless, there are recent examples showing that this connected to a fast polarity switching and rapid scan-
field of chromatography may still be refined, in order ning tandem mass spectrometer with an electrospray
to facilitate identification in complex mixtures or matri- interface. The software used to evaluate the results
ces. Van der Lee et al. [60] presented targeted screen- produced reports, showing adverse analytical findings
ing for 106 scheduled pesticides and contaminants in in the sample. One 96-well plate could be analysed
animal feed matrix. The procedure was based on sol- within 16 h.
vent extraction, gel permeation chromatography clean- Several applications of UPLC-MS concerned pes-
up, and two-dimensional GC with full-scan TOF-MS. ticide residue analysis. UPLC-TOF-MS was applied to
All compounds were automatically detected at a level the rapid qualitative and quantitative analysis of 100 pes-
exceeding 50 ng/ml. Two-dimensional GC followed ticides targeted in strawberries [53]. Accurate mass
by TOF-MS was also applied by Banerjee et al. [1] for measurement of positive and negative ions allowed
multiresidue analysis of pesticides in grapes. A combi- their extraction following “full mass range data acqui-
nation of a non-polar and a polar capillary column con- sition” with negligible interference from background
nected in series was used. The method resolved co- or co-eluting species observed during UPLC gradient
elution problems as observed in full scan one-dimen- separation (in a cycle time of 6.5 min per run). In an-
sional GC-MS analysis and allowed chromatographic other study [44], 90 pesticides were screened in fruit
separation of 51 pesticides within 24 min run time with juices by UPLC-MS-MS (MRM) after simple aceto-
library-searchable mass spectrometric confirmation. nitrile extraction. The separation was achieved in an
An automated direct sample introduction technique 11 min run.
coupled to comprehensive two-dimensional gas chro-
matography-time of flight mass spectrometry was ap- 3.3. Optimisation of mass spectrometric detection
plied to the development of a screening method for
polychlorinated dibenzo-p-dioxins/dibenzofurans and Accurate mass determination emerged as one of the
polychlorinated biphenyls in fish oil [27]. Kolbrich et most promising identification tools in contemporary
al. [29] applied two-dimensional GC-EI-MS (SIM) mass spectrometry. The measured elemental mass al-
with cryofocusing for determination of methylene- lows calculation of the elemental formula of the ion,
dioxyethylamphetamine (MDEA), MDMA, and its me- and therefore confirmation of its identity in the case of
tabolites, 3,4-methylenedioxyamphetamine (MDA), a match with a known compound in a library search
4-hydroxy-3-methoxymethamphetamine (HMMA), and [21] or identification of an unknown ion. The mass ac-
4-hydroxy-3-methoxyamphetamine (HMA) in human curacy may be expressed in absolute values, usually in
plasma. This assay provided low limits of quantifica- millidaltons (mDa), or in relative values, e.g. in parts
tion and the chromatographic system should be suit- per million (ppm). These values are calculated as fol-
able for application to other analytes and to other lows:
complex matrices.
In the case of HPLC, recent years have brought dis- mDa: [(m/z)measured – (m/z)calculated] ´ 1000, {1}
tinct progress through introduction of fine-particle ppm: 106 ´ [(m/z)measured – (m/z)calculated]/(m/z). {2}
columns (particle size below 2 mm). This technique,
known as ultra-performance liquid chromatography Mass accuracy depends on the instrument used; for
(UPLC), enables much better separation in a shorter a standard quadrupole or ion-trap instrument, a mass
time and was immediately introduced in combination accuracy of 50–100 mDa is possible, for TOF or QTOF
with mass spectrometry for identification purposes. In below 10 mDa, for FTICR-MS and orbitrap – 1 mDa
the field of doping analysis, UPLC has enabled high- or less. Some new triple-quadrupole instruments show
throughput analysis, particularly of diuretic com- accurate mass capabilities below 1 mDa [40].
pounds. Ventura et al. [63] detected 34 scheduled di- Besides forensic toxicological screening [20, 22]
uretics and other doping agents in urine extracts, using high-resolution mass spectrometry has been applied in
UPLC-MS-MS in negative and positive ionisation other screening and identification procedures. Hernan-
mode. Total analysis time was 5 min, and the method dez et al. [24] developed a procedure based on gas
fulfilled the requirements established by the World chromatography coupled to high-resolution time-of-
Anti-Doping Agency. Thörngren et al. [56] developed flight mass spectrometry (GC-TOF-MS) for identifi-

Problems of Forensic Sciences 2009, vol. LXXVII, 7–28


Quality assurance aspects of identification with chromatographic... 11

cation of 60 organic pollutants in water after solid- lowed by centrifugation was used; urine specimens
phase microextraction. The identification was based were only spiked with internal standard and centri-
on evaluating the presence of up to five representative fuged. The chromatographic method consisted of on-
m/z ions per analyte, measured at high mass accuracy, line extraction with an Oasis HLB column and iso-
and determination of their Q/q (Q, quantitative ion; cratic separation on a Chromolith RP-18e column. The
q, confirmative ion) intensity ratio. This strategy led to drugs were detected with QTRAP hybrid tandem MS
the detection of target compounds in several water in positive and negative ionisation. One transition was
samples at low ppb levels. monitored for each drug. IDA was applied; if the sig-
Several hundred crop protection products were re- nal was greater than the preselected value, the EPI
cently banned from use by European and US environ- spectrum was recorded. Unequivocal identification was
mental agencies and this became an issue for envi- achieved for most of the drugs. In the case of anabolic
ronment monitoring laboratories. Since it was difficult steroids, highly similar EPI spectra were observed, and
to monitor all these compounds with GC-MS, LC-MS, the authors found the method unsuitable for this group
or LC-MS-MS, Thurman and Ferrer [57] developed of compounds. Drees et al. [14] compared the identifi-
a non-targeted screening procedure based on a combi- cation ability of MRM ratios and MRM-IDA, using
nation of LC-TOF-MS, LC-IT-MS, and LC-QTOF- QTrap hybrid triple quadrupole/linear ion trap MS.
MS. The Merck Index and ChemIndex, both commer- 14 selected drugs of abuse were examined at various
cially available, were applied as reference databases. concentrations. Both methods performed well at low
The identification strategy consisted of four steps: concentrations; at very high concentrations, the MRM-
full-scan LC-TOF-MS analysis; library search for em- IDA method gave a better chance of confirmation.
pirical formulas and any A+2 isotopes (for halogens A combination of LC-MS-MS and/or GC-MS-NPD
or S); LC-IT-MS or LC-Q-TOF-MS-MS for structure was used by Thevis et al. [55] to identify compounds
elucidation; and final comparative standard analysis. present in confiscated black market drug preparations,
Accurate mass determination may be very useful in the containing mainly anabolic steroids or sexual stimu-
analysis of such compounds, which cannot yield char- lants. The drugs were dissolved in methanol, and sub-
acteristic fragment ions in tandem mass spectrometric jected to LC-ESI-MS-MS examination on a QTrap
analysis, particularly in MRM mode. Nielen et al. [39] instrument. For each drug, three transitions were mon-
compared the performance of various high resolution itored. For the GC-MS-NPD examination, an inlet
LC-MS techniques for hormone residue analysis in splitter was applied, and the injected sample was sepa-
food samples. The authors used three types of instru- rated on two identical columns; one was connected
ments: LC-QTOF-MS, LC-FTIR-MS, and LC-FT-Orbi- with an EI-MS detector, the other with a nitrogen-
trap-MSn. The authors stressed the need for optimal phosphorous detector (NPD).
sample pretreatment and cleanup when using instru- Relative ion intensities (ion ratios) are usually ap-
ments with lower accuracy (around 20 ppm) since the plied as one of the identification features [2, 37]. Feng
LC resolution and MS resolution are strongly interre- et al. [18] developed an LC-MS-MS procedure for si-
lated and have a major impact on mass accuracy. In- multaneous determination of 30 different drugs of
struments with higher mass resolution may accom- abuse in urine. Three transitions were monitored for
modate a less completely resolved chromatographic each drug. The intensity ratios of the two fragment
separation. This may be illustrated by the analysis of ions to most abundant fragment ion were calculated
stanozolol; an inexplicable fragment observed in QTOF- and used to confirm the identity. These ratios should
MS (m/z 161.1223 at mass resolution 5000 FWHM) be in the range of ± 3SD as determined in the valida-
analysis appeared in LC-FTIR-MS as a doublet of ions tion procedure. Concheiro et al. [8] published an LC-
having minor mass differences (m/z 161.1073 and MS-MS study on the simultaneous determination of
m/z 161.1324, mass resolution 250,000). various drugs of abuse in urine. Two transitions were
Besides high-resolution accurate mass MS, other monitored for each drug. Relative ion intensities were
instrumental solutions have recently been applied for tested, in within-day and between-day mode at two
identification purposes. The most promising was the concentration levels. The variability of results was
application of information dependent acquisition much higher in between-day experiments than in
(IDA) in combination with an enhanced product ion within-day experiments. The authors concluded that
(EPI) spectrum. This technique was used by Stanley et relative ion intensity should be treated with caution as
al. [48] for screening of 32 acidic drugs in equine an identification criterion, and strongly recommended
plasma and 11 neutral drugs in equine urine. For plas- analysis of standard samples on the same day as real
ma, acetonitrile/internal standard precipitation fol- samples in order to fulfil the established criteria.

Problems of Forensic Sciences 2009, vol. LXXVII, 7–28


12 M. J. Bogusz

Stein and Heller [49] published a very important ked, demonstration of zero false positive rate, dem-
study on the factors responsible for false positive identi- onstration of 10% false negative rate at the toler-
fication with mass spectrometry. The following experi- ance level, demonstration of ruggedness and non-
ments were done: From the reduced NIST/EPA/NIH interference of other drugs or matrix components;
mass spectral library [59], comprising 96,464 entries, ev- – method description, also containing structure and
ery tenth spectrum (roughly 9600 spectra) was chosen as full spectrum of target compound, spectral data on
a search spectrum. These search spectra were eliminated at least three structurally-specific ions that com-
from the library before performing the search, so all li- pletely define the parent molecule or more if non-
brary spectra that matched the search spectrum for specific ions are included, proposed fragment ions
a given set of constraints were false positives. Search ex- structures, consistent with fragmentation patterns,
periments were performed using different a number of justification for specificity of selected ions or scan
peaks for matching, the number ranging from 1 to range, confirmation and operational criteria, as
8 peaks. The abundance window was set at 0.25, i.e. well as quality control section;
the search and library spectrum at given m/z was con- – confirmation criteria. In the confirmation proce-
sidered a match if the difference in abundance was less dure, the comparison standard(s) should be ana-
than 25%. It was shown that the number of peaks as lysed contemporaneously, in the presence of ex-
well as peak abundance clearly reduced the false posi- tracted matrix if appropriate. Any of the following
tive probability. The authors concluded that more than MS chromatograms may be used: TIC, RIC, SIM,
three fragment ions (in full scan or SIM mode) should or SRM. Flow injection analysis is discouraged.
be used for identification and their m/z values should The chromatographic peak should exceed a sig-
be carefully selected. Highly probable peak correla- nal-to-noise threshold of 3:1, and the retention time
tions such as 14 amu difference (methyl group loss) or should not differ from the retention time of the
18 amu (water loss) should be avoided. standard by more than 2% for GC-MS or 5% for
LC-MS. Mass spectral matching criteria vary de-
pending on the MS acquisition mode: in the full-
4. Legal and regulatory aspects of scan and partial-scan MS1, the spectrum should
identification contain at least three structurally-specific ions, and
the spectrum should visually match the spectrum of
Several professional organisations as well as na- the standard. An acceptability range of ± 20% for
tional and international agencies have formulated rec- relative abundance of major ions is recommended.
ommendations and guidelines concerning criteria for Library search algorithms should not be used to
identification with mass spectrometric methods. These confirm identity. All structurally-specific ions
documents may be divided into two main groups. The mentioned in the method description should be
first group constitutes legal regulations at a national present and their relative abundances should corre-
and international level, such as those issued by US Hu- spond to those of the standard. The presence of
man Health Services for clinical laboratories [12] and other unrelated prominent ions should be explained
for workplace drug testing [13] and by the European (e.g. from matrix compounds). If background sub-
Union [17]. The second group comprises recommen- traction has been used, it should be specified and
dations and guidelines of professional organisations at indicated. In the MS1 SIM, relative abundances of
a national and international level, such as the World three structurally-specific ions should match the
Anti-Doping Agency – WADA, College of American standard within ± 10% (absolute). In the case of
Pathologists – CAP, Food and Drug Administration – four or more structurally-specific ions, the match
FDA, US Department of Agriculture, Society of Fo- level is ± 15% (absolute). Relative abundances for
rensic Toxicologists – SOFT [47], German Society of more than three ions, which include less specific
Toxicological and Forensic Chemistry – GTFCh [22]. ions such as isotopes or due to loss of water, should
The FDA has published final guidance for the de- match the comparison standard within ± 10%. In
velopment, evaluation, and application of mass spec- the full-scan and partial-scan MSn, the spectrum
trometric methods for confirming the identity of ani- should visually match the spectrum of the standard,
mal drug residues [23]. According to the guidance, the and there should be general correspondence be-
confirmatory mass spectrometric procedures should tween relative abundances obtained in sample and
address each of the following points: standard. All structurally-specific ions mentioned
– a validation package from the originating labora- in the method description should be present. If
tory containing replicate samples, original and spi- a structurally-specific precursor ion completely dis-

Problems of Forensic Sciences 2009, vol. LXXVII, 7–28


Quality assurance aspects of identification with chromatographic... 13

sociates to product ions, the appearance of at least For full scan MS, the presence of a minimum of
two structurally-specific product ions should be four diagnostic ions with relative intensity of more
sufficient in MSn+1. The presence of other unrelated than 10% in the reference spectrum is obligatory. The
prominent ions should be explained (e.g. from ma- molecular ion should be included if its intensity is
trix compounds). If background subtraction has above 10%.
been used, it should be specified and indicated. In For SIM, the molecular ion should preferably be in-
MSn SRM, if a precursor ion is completely dissoci- cluded. The signal-to-noise ratio for each diagnostic
ated, the relative abundance of structurally-specific ion should be ³ 3:1.
product ions should match the comparison standard Maximum permitted tolerances for relative ion in-
within ± 10% in cases where two ions have been tensities are presented in Table I. For identification,
monitored, and within ± 20% in cases where three the EC has defined an identification points system. In
or more ions have been monitored; this system, at least four points are required for confir-
– quality control should include the following points: mation of identity. The number of points earned by
establishing of system suitability, running of at particular techniques are shown in Table II. According
least one control and one fortified control sample, to these criteria, e.g. four characteristic ions are needed
control of possible carryover. if GC-MS or LC-MS is applied, or one precursor and
The European Commission [17] has published re- two products for GC-MS-MS or LC-MS-MS. Nielen
quirements concerning performance of analytical et al. [39] proposed additional identification criteria to
methods and interpretation of results. In this docu- the four-point classification, based on high-resolution
ment, performance criteria for mass spectrometric de- mass spectrometry (Table III).
tection were formulated. It was stated that on-line or The College of American Pathologists (CAP) in
off-line chromatographic separation is a prerequisite their “Chemistry and Toxicology Checklist” of the
for mass spectrometric confirmation. Laboratory Accreditation Program formulated re-
For both GC-MS and LC-MS, the minimum ac- quirements, relevant for identification with chromato-
ceptable retention time for the target compound should graphic-mass spectrometric methods [7].
be at least twice the retention time corresponding to In the case of single-stage GC-MS or LC-MS, the
the void volume of the column (i.e., the dead time, or identification should be done on the basis of ion ratios,
Rt0). The relative retention time of the analyte (ratio using at least two ion ratios whenever possible. The
target : internal standard) should not differ from that of suggested tolerance limit for GC-MS is ± 20% of those
the reference standard more than ± 0.5% for GC and of calibrators, for LC-MS the limit is ± 30%. If only
± 2.5% for LC. one ratio of two characteristic ions is available, it may
Mass spectrometric detection should be carried out be acceptable if there are other identifying characteris-
in full-scan mode, SIM, as well as MS-MSn techniques tics, e.g. retention time. The internal standard should
such as SRM or other techniques in combination with be identified with at least one ion ratio. If full scan MS
appropriate ionisation modes. is used, the laboratory should set and validate its own

TABLE I. MAXIMAL PERMITTED TOLERANCES IN ABUNDANCE RATIOS IN COMPARISON WITH THE STANDARD

Relative intensity EC [17] EC [17] CAP [7] WADA [66] WADA [66]
(% of base peak)
GC-EI-MS GC-CI-MS, LC-MSn GC-EI-MS GC-CI-MS,
GC-MSn, GC-MSn,
LC-MS, LC-MS;
LC-MSn LC-MSn
> 50% ± 10% ± 20% ± 20% ± 10% absolute ± 10% absolute
> 20–50% ± 15% ± 25% ± 25% ± 20 % relative ± 25% relative
> 10–20% ± 10% ± 30% ± 30% ± 5% absolute ± 10% absolute
10% ± 50% ± 50% ± 50%

FDA [23] used flat rate of permitted tolerances: within ± 20% of relative abundance for MS scan and MSn SRM (three transitions, and
± 10% for MS-SIM and MSn SRM (three transitions). PDP [58] used flat rate of ± 20% of relative abundance (absolute difference) for all
MS techniques.

Problems of Forensic Sciences 2009, vol. LXXVII, 7–28


14 M. J. Bogusz

threshold of “spectral match” or fit for identification mation, and quantitation of pesticide residues using
purposes. GC and LC with mass spectrometric detection [58].
This SOP was written in order to combine the require-
TABLE II. IDENTIFICATION POINTS EARNED BY ments of all MS and MS-MS procedures used in pesti-
VARIOUS MS TECHNIQUES ACCORDING cide residue analysis into a single document.
TO EC [17] In the case of full-scan GC-MS, the spectra should
be recorded in the range of 20 to 500 amu. A minimum
MS technique Identification points of three structurally-specific ions, preferably includ-
earned per ion ing molecular ion, meeting the signal-to-noise 3:1 ra-
LR-MS 1.0 tio are required. Isotopic cluster ions may be used as
n one of three significant ions. The relative intensity ra-
LR-MS precursor ion 1.0
tios of each ion should be within ± 20% of the ratios
n
LR-MS product ion 1.5 observed in the reference standard. The use of library
HR-MS 2.0 search software for EI analysis is mandatory. The use
n of library searches in “soft” ionisation techniques,
HR-MS precursor ion 2.0
such as GC-CI-MS is discouraged. Chromatographic
n
LR-MS product ion 2.5 criteria (retention time accuracy) for GC-MS were not
formulated in this document.
LR-MS – low resolution mass spectrometry; HR-MS – high res-
olution mass spectrometry. In GC-MS-MS analysis, the retention time of the
target compound should not differ more than ± 0.05
For tandem MS (GC-MS-MS or LC-MS-MS) in min from the reference standard or ± 0.01 RRt. Two
SRM mode, at least one transition and one ion ratio transitions from one precursor ion or from two precur-
should be monitored, together with retention time. sors should be monitored. The relative intensity ratios
However, if enough ions of sufficient abundance exist, of each ion should be within ± 20% of the ratios ob-
two or more ion ratios should be monitored. Ion ratios served in the reference standard, and the abundance of
determined from full scan analysis are an acceptable each ion should exceed signal-to noise 3:1.
identification method and should fulfil the same crite- For LC-MS analysis, the apparatus should be capa-
ria as for SRM mode. Tolerance limits should be ade- ble of scanning 50 to 1200 amu in full scan mode. The
quate to the method employed and should be sup- retention time of the target compound should not differ
ported by references or own data. In another approach, more than ± 0.5 min from the reference standard or
two-fold acceptance criteria of data are applied for at ± 0.01 RRt. A minimum of three structurally-specific
least three ion ratios and a scoring system according to ions, preferably including a protonated or deproto-
EC requirements [17]. The tolerances of ion ratios dif- nated molecule, meeting the signal-to-noise 3:1 ratio
fer according to the abundance of ions (see Table I). are required. Isotopic cluster ions may be used as one
The US Pesticide Agency has published a standard of three significant ions. The relative intensity ratios of
operating procedure (SOP) for identification, confir- each ion should be within ± 20% of the ratios observed
in the reference standard. In LC-MS-MS analysis, chro-

TABLE III. PROPOSAL FOR ADDITIONAL LC-MS CRITERIA TO BE IMPLEMENTED IN EC RECOMMENDATIONS [17]
MADE BY NIELEN ET AL. [39]

Goal Mass resolution Mass accuracy Remarks


(FWHM) (mDa)
Screening 10,000 ± 50 (window) Relative retention time 2.5%
Confirmation 10,000 5 1.5 identification points/ion or product ion; at least one
ion ratio; relative retention time 2.5%
N.B. – LC/biogram: 1 additional identification point
HR confirmation 20,000 5 2 identification points/ion or product ion; at least one
ion ratio; relative retention time 2.5%
MS-MS 10,000 5 Confirm postulated structure by NMR and/or confirm
identification of accurate masses at mass resolution 70,000 (FWHM)
unknowns

Problems of Forensic Sciences 2009, vol. LXXVII, 7–28


Quality assurance aspects of identification with chromatographic... 15

matographic requirements (Rt or RRt) are the same as than 5% in the reference spectrum, the ion must be
for LC-MS, whereas mass spectrometric requirements present in the unknown. The concentrations of de-
are identical to GC-MS-MS. tected compounds should be comparable with those in
The World Anti-Doping Agency (WADA) [66] has the reference sample.
formulated criteria which must be fulfilled in order to If three diagnostic ions are not available, a second
identify a prohibited, scheduled substance in the urine derivative shall be prepared, or a second ionisation or
or blood of an examined athlete using chromatograph- fragmentation technique, based on different physical
ic-mass spectrometric procedures. These criteria are principles shall be used. In any case a minimum of two
divided into separation and detection requirements diagnostic ions is mandatory for each mass spectrum.
and are as follows: In MS-MS, the data can be acquired either in full
Chromatographic separation requirements. For capil- scan or selected reaction monitoring (SRM) mode.
lary GC, the retention time (Rt) of the analyte shall not The precursor ion should be present in both modes.
differ by more than 1% or ± 0.2 min (whichever is When monitoring more than one product ion, the rela-
smaller) from that of the same substance in a spiked tive intensities of any of the ions shall not differ by
urine sample. For HPLC, the Rt of the analyte shall not more than the amount given in the table from the rela-
differ by more than 2% or ± 0.4 min (whichever is tive intensities acquired from a spiked reference sam-
smaller) from that of the same substance in a spiked ple. The signal-to-noise for the least intense ion should
urine sample. These criteria may be relaxed, if the shift be greater than 3. For a diagnostic ion with a relative
in Rt may be explained (e.g. by sample overload). abundance of less than 5% observed in the reference
Mass spectrometric requirements. Full scan mode: full spectrum, the ion must be present in the studied sam-
scan or partial scan mode is the preferred approach to ple. Table IV shows a summary of requirements as
identification. A partial scan may begin at an m/z value formulated by various organisations.
greater than any abundant ion due to the derivatising
agent or chemical ionisation reagent. All diagnostic
ions with a relative abundance greater than 10% in the 5. Conclusions
reference spectrum obtained from a reference material
must be present in the spectrum of the unknown peak. The criteria of identification of compounds with
The relative abundance of three diagnostic ions shall chromatography-mass spectrometry are certainly not
not differ by more than the defined amount (see Ta- carved in stone, like the Ten Commandments. They
ble I) from the relative intensities of the same ions ob- are rather a moving target, changing their position in
served in the reference spectrum (obtained from relation to actual knowledge, with new requirements
spiked urine, reference collection sample, or reference continuously being set, which vary according to the
material). It is not permissible to collect additional particular discipline and final task of analysis. It is un-
ions and select those ratios that are within defined tol- derstandable that the strictest requirements are set in
erance. those areas, where the results of identification may
If computer-based mass spectral library searching carry with them legal sanctions. For these reasons, sev-
or matching is used, the results should be reviewed by eral organisations have issued more or less detailed
a qualified scientist. conditions, recommendations, or guidelines, defining
If three diagnostic ions with a relative abundance minimal attributes necessary for identification. These
greater than 5% are not available, a second derivative, documents concern both chromatographic and mass
yielding different diagnostic ions shall be prepared, or spectrometric aspects of analysis.
a second ionisation or fragmentation technique, based There is general agreement concerning the kind of
on a different physical principle shall be used. In any parameters to be controlled, such as retention time,
case, a minimum of two diagnostic ions is mandatory number of ions, or ion intensity ratios. However, the
for each mass spectrum. exact numerical criteria, formulated by different bod-
In SIM mode, at least three diagnostic ions must be ies or organisations, show distinct variations.
acquired, with a signal-to-noise ratio > 3 for the least These recommendations, which are products of
intense ion. The relative intensities of ions shall not common knowledge in a given discipline, should be
differ by more than the defined amount (see Table I) treated as a momentary, frozen picture of the present
from the relative intensities of the same ions observed situation in the field of analytical identification. As
in the reference spectrum (obtained from spiked urine, Hill wrote: “regulatory guidances are not a substitu-
reference collection sample, or reference material). te for good science, they do however provide a fra-
For diagnostic ions with a relative abundance of less mework that should initiate the questions why and

Problems of Forensic Sciences 2009, vol. LXXVII, 7–28


16 M. J. Bogusz

TABLE IV. SUMMARY OF IDENTIFICATION REQUIREMENTS AS FORMULATED BY VARIOUS ORGANISATIONS

Method/parameter FDA [23] EC [17] WADA [66] PDP [58] CAP [7]
GC-MS/Rt tolerance 2% 1% or 0.2 0.05
GC-MS/RRt tolerance 0.5% 0.01
LC-MS/Rt tolerance 5% 2% or 0.4 0.5
LC-MS/RRt tolerance 2.5% 0.1
MS scan minimal number of ions 3 4 3 3 3
MS SIM/minimal number of ions 3 4 3 3 3
n
MS scan minimal number of ions 3 4 3 3
n
MS SRM number of transitions 2 2 2 2 2
HRMS minimal number of ions 2
Signal-to-noise threshold 3 3 3 3 3

how” [25]. It will be always be incumbent upon the ex- 6. Buszewski B., Ulanowska A., Ligor T. [et al.], Identifica-
pert involved in identification to answer the question: tion of volatile organic compounds secreted from cancer
“Did I identify this compound according to the best tissues and bacterial cultures, Journal of Chromatogra-
available knowledge?” in addition to the question: phy B 2008, 868, 88–94.
“Did I fulfil all formal conditions, necessary for identi- 7. College of American Pathologists, Commission on Labo-
ratory Accreditation, Laboratory Accreditation Program,
fication?” At this point, professional ethics meets pro-
Chemistry and Toxicology Checklist 2006, 57–59.
fessional competence.
8. Concheiro M., De Castro A., Quintela O. [et al.], Determi-
nation of illicit drugs and their metabolites in human urine
by liquid chromatography tandem mass spectrometry in-
References cluding relative ion intensity criterion, Journal of Ana-
lytical Toxicology 2007, 31, 573–580.
1. Banerjee K., Patil S. H., Dasgupta S. [et al.], Optimization 9. De Zeeuw R. A., Substance identification: The weak link
of separation and detection conditions for the multiresi- in analytical toxicology, Journal of Chromatography B
due analysis of pesticides in grapes by comprehensive 2004, 811, 3–12.
two-dimensional gas chromatography-time-of-flight 10. De Zeeuw R. A., Franke J. P., General unknown analysis
mass spectrometry, Journal of Chromatography A 2008, [in:] Forensic Science, Bogusz M. J. [ed.], Handbook of
1190, 350–357. analytical separations, vol. 2, Elsevier Science, Amster-
2. Bogusz M. J., Hassan H., Al-Enazi E. [et al.], Rapid deter- dam 2000.
mination of chloramphenicol and its glucuronide in food 11. Decaestecker T. N., Vande Casteele S. R., Walle-
products by liquid chromatography-electrospray negative macq P. E. [et al.], Information-dependent acquisition-
ionization tandem mass spectrometry, Journal of Chro- mediated LC-MS/MS screening procedure with semi-
matography B 2004, 807, 343–356. quantitative potential, Analytical Chemistry 2004, 76,
3. Bogusz M. J., Maier R. D., Kruger K. D. [et al.], Poor 6365–6373.
reproducibility of in-source collisional atmospheric pres- 12. Department of Health and Human Services, Health Care
sure ionization mass spectra of toxicologically relevant Fi nanc ing Ad min is tra tion, Pub lic Health Ser vice,
drugs, Journal of Chromatography A 1999, 844, 409–418. Clinical Laboratory Improvement Amendments of 1988;
4. Bristow A. W., Nichols W. F., Webb K. S. [et al.], Evalua- Final rule, Fed eral Reg is ter 1992, 57, 967–1087
tion of protocols for reproducible electrospray in-source [http://www.cms.hhs.gov/CLIA].
collisionally induced dissociation on various liquid chro- 13. Department of Health and Human Services, Substance
matography/mass spectrometry instruments and the de- Abuse and Mental Health Services Administration, Man-
velopment of spectral libraries, Rapid Communications in datory guidelines and proposed revisions to mandatory
Mass Spectrometry 2002, 16, 2374–2386. guidelines for federal workplace drug testing programs,
5. Brown S. D., Rhodes D. J., Pritchard B. J., A validated Federal Register 2004, 71, 19644–19732 [http://work-
SPME-GC-MS method for simultaneous quantification place.samsha.gov].
of club drugs in human urine, Forensic Science Interna- 14. Drees J. C., Sasaki T. A., Stone J. A. [et al.], The advan-
tional 2007, 171,142–150. tages and limitations of MRM vs. full scan MS/MS for

Problems of Forensic Sciences 2009, vol. LXXVII, 7–28


Quality assurance aspects of identification with chromatographic... 17

drug confirmation using LC/MS/MS, Poster K35 on the 27. Hoh E., Lehotay S. J., Mastovska K. [et al.], Evaluation of
55th Conference of American Society of Mass Spectrome- automated direct sample introduction with comprehen-
try, Indianapolis 2007. sive two-dimensional gas chromatography/time-of-flight
15. Drummer O. H., Requirements for bioanalytical proce- mass spectrometry for the screening analysis of dioxins in
dures in postmortem toxicology, Analytical and Bio- fish oil, Journal of Chromatography A 2008, 1201, 69–77.
analytical Chemistry 2007, 188, 1495–1503. 28. Hough J. M., Haney C. A., Voyksner R. D., Evaluation of
16. Drummer O. H., Chromatographic screening techniques electrospray transport CID for the generation of search-
in systematic toxicological analysis, Journal of Chroma- able libraries, Analytical Chemistry 2000, 72, 2265– 2270.
tography B 1999, 733, 27–45. 29. Kolbrich E. A., Lowe R. H., Huestis M. A., Two-dimen-
17. EU Commission, Commission decision of 12 August sional gas chromatography/electron-impact mass spec-
2002 implementing Council Directive 96/23/EC concern- trometry with cryofocusing for simultaneous quantifi-
ing the performance of analytical methods and the inter- cation of MDMA, MDA, HMMA, HMA, and MDEA in
pretation of results, Official Journal of the European human plasma, Clinical Chemistry 2008, 54, 379–387.
Communities 2002, L 221, 8–34. 30. Marquet P., Identification and confirmation criteria for
18. Feng J., Wang L., Dai I. [et al.], Simultaneous determina- LC-MS, [in:] Applications of LC-MS in toxicology,
tion of multiple drugs of abuse and relevant metabolites Polettini A. [ed.], Pharmaceutical Press, London 2006.
inn urine by LC-MS-MS, Journal of Analytical Toxicol- 31. Marquet P., Saint-Marcoux F., Gamble T. N. [et al.],
ogy 2007, 31, 359–368. Comparison of a preliminary procedure for the general
19. Georgakopoulos C. G., Vonaparti A., Stamou M. [et al.], unknown screening of drugs and toxic compounds using
Preventive doping control analysis: liquid and gas chro- a quadrupole-linear ion-trap mass spectrometer with a liq-
matography time-of-flight mass spectrometry for detec- uid chromatography-mass spectrometry reference tech-
tion of designer steroids, Rapid Communications in Mass nique, Journal of Chromatography B 2003, 789, 9–18.
Spectrometry 2007, 21, 2439–2446. 32. Maurer H. H., Hyphenated mass spectrometric tech-
20. Gergov M. Forensic screening with liquid chromatogra- niques – indispensable tools in clinical and forensic toxi-
phy-mass spectrometry [in:] Forensic science. Handbook cology and in doping control, Journal of Mass Spectro-
on analytical separations, vol. 6, Bogusz M. J. [ed.], metry 2006, 41, 1399–1413.
Elsevier Science, Amsterdam, Boston 2008. 33. Maurer H. H., Forensic screening with GC-MS, [in:] Fo-
21. Gergov M., Boucher G. B., Ojanpera I. [et al.], Toxicolog- rensic science. Handbook on analytical separations, vol. 6,
ical screening of urine for drugs by liquid chromatogra- Bogusz M. J. [ed.], Elsevier, Amsterdam, Boston 2008.
phy/time-of-flight mass spectrometry with automated li- 34. Maurer H. H., Peters F. T., Analyte identification using li-
brary search based on elemental formulas, Rapid Commu- brary searching in GC-MS and LC-MS, [in:] The encyclo-
nications in Mass Spectrometry 2001, 15, 521–526. pedia of mass spectrometry, vol. 8, Niessen W. M. A. [ed.],
22. German Society of Toxicological and Forensic Chem- Elsevier, Amsterdam, Oxford 2006.
istry, Anlage zu den Richtlinien der GTFCh zur Quali- 35. Maurer H. H., Pfleger K., Weber A., Mass spectral library
tätssicherung bei forensisch-toxikologischen Unter - of drugs, poisons, pesticides, pollutants, and their metab-
suchungen, Toxichem + Krimtech 1998, 65, 18–24, olites, Wiley-VCH, Weinheim 2007.
[http://www.gtfch.org/tk/tk67_1/akqual.pdf GTFCh]. 36. McLafferty F. W., Registry of mass spectral data, Wiley,
23. Guidance for industry, Mass spectrometry for confirma- New York 2001.
tion of the identity of animal drug residue, U.S. Depart- 37. Mottier P., Parisod V., Gremaud E. [et al.], Determination
ment of Health and Human Services, Food and Drug of the antibiotic chloramphenicol in meat and seafood
Administration, Center for Veterinary Medicine, May 1, products by liquid chromatography-electrospray ioniza-
2003. tion tandem mass spectrometry, Journal of Chromato-
24. Hernandez F., Portolés T., Pitarch E. [et al.], Target and graphy B 2003, 994, 75–84.
nontarget screening of organic micropollutants in water 38. Nielen M. W., Bovee T. F., van Engelen M. C. [et al.],
by solid-phase microextraction combined with gas chro- Urine testing for designer steroids by liquid chromatogra-
matography/high-resolution time-of-flight mass spectro- phy with androgen bioassay detection and electrospray
metry, Analytical Chemistry 2007, 79, 9494–9504. quadrupole time-of-flight mass spectrometry identifica-
25. Hill H. M., Chromatography in a regulated environment, tion, Analytical Chemistry 2006, 7, 424–431.
[in:] Bioanalytical separations. Handbook of analytical 39. Nielen M. W. F., van Engelen M. C., Zuiderent R. [et al.],
separations, vol. 4, Wilson I. D. [ed.], Elsevier Science, Screening and confirmation criteria for hormone residue
Amsterdam 2003. analy sis using liquid chromatography accurate mass
26. Ho E. N. M., Wan T. S. M., Wong A. S. Y. [et al.], Doping time-of-flight, Fourier transform ion cyclotron resonance
control analysis of insulin and its analogues in equine and orbitrap mass spectrometry techniques, Analytica
plasma by liquid chromatography-tandem mass spec- Chimica Acta 2007, 586, 122–129.
trometry, Journal of Chromatography A 2008, 1201, 40. Niessen W. M. A., High-resolution mass spectrometry
183–190. and accurate mass determination, [in:] The encyclopedia

Problems of Forensic Sciences 2009, vol. LXXVII, 7–28


18 M. J. Bogusz

of mass spectrometry, vol. 8, Niessen W. M. A. [ed.], 53. Taylor M. J., Keenan G. A., Reid K. B. [et al.], The utility
Elsevier, Amsterdam, Oxford 2006. of ultra-performance liquid chromatography/electrospray
41. Pontes M., Marques J. C., Câmara J. S., Screening of vola- ionization time-of-flight mass spectrometry for multi-res-
tile composition from Portuguese multifloral honeys us- idue determination of pesticides in strawberry, Rapid
ing headspace solid-phase microextraction-gas chroma- Com mu ni ca tions in Mass Spec trom e try 2008, 22,
tography-quadrupole mass spectrometry, Talanta 2007, 2731–2746.
74, 91–103. 54. Thevis M., Schänzer W., Examples of doping control
42. Rivier L., Criteria for the identification of compounds by analysis by liquid chromatography-tandem mass spec-
liquid chromatography-mass spectrometry and liquid trometry: ephedrines, beta-receptor blocking agents, di-
chromatography-multiple mass spectrometry in forensic uretics, sympaticomimetics, and cross-linked hemo-
and doping analysis, Analytica Chimica Acta 2003, 492, globins, Journal of Chromatographic Science 2005, 43,
69–82. 22–31.
43. Rivier L., Identification and confirmation criteria for 55. Thevis M., Schrader Y., Thomas A. [et al.], Analysis of
LC-MS, [in:] Applications of LC-MS in Toxicology, confiscated black market drugs using chromatographic
Polettini A. [ed.], Pharmaceutical Press, London 2006. and mass spectrometric approaches, Journal of Analyti-
44. Romero-González R., Garrido Frenich A., Martinez cal Toxicology 2008, 32, 232–240.
Vidal J. L., Multiresidue method for fast determination of 56. Thörngren J. O., Ostervall F., Garle M., A high-through-
pesticides in fruit juices by ultra performance liquid chro- put multicomponent screening method for diuretics,
matography coupled to tandem mass spectrometry, Ta- masking agents, central nervous system (CNS) stimulants
lanta 2008, 76, 211–225. and opiates in human urine by UPLC-MS/MS, Journal of
45. Saint-Marcoux F., Lachatre G., Marquet P., Evaluation of Mass Spectrometry 2008, 43, 980–992.
an improved general unknown screening procedure using 57. Thurman E. M., Ferrer I., Identification of Unknown En-
liquid chromatography-electrospray mass spectrometry vironmental Contaminants Using Multidimensional LC-
and high performance liquid chromatography-diode array MS Strategies Involving TOF-MS, Ion-Trap MSn, and
detection, Journal of the American Society for Mass Spec- Q-TOF-MS-MS, [in:] The encyclopedia of mass spec-
trometry 2003, 14, 14–22. trometry, Hyphenated methods, vol. 8, Niessen W. M. A.
46. Segura J., Ventura R., Marcos J. [et al.], Doping sub- [ed.], Elsevier, Amsterdam, Oxford 2006.
stances in human and animal sport, [in:] Forensic science. 58. U.S. Department of Agriculture, Agriculture Marketing
Handbook of analytical separations, vol. 6, Bogusz M. J., Service, Science & Technology, Pesticide data program,
[ed], Elsevier Science, Amsterdam 2008. January 1, 2007 [http://www.ams.usda.gov/AMSv1.0/
47. Society of Forensic Toxicologists, SOFT/AAFS forensic getfile?dDocName=STELPRDC506150].
toxicology laboratory guidelines 2006, 1–24, [http: 59. U.S. Department of Commerce, NIST/EPA/NIH mass
//www.soft-tox.org/]. spectral library, Wiley, New York 2005.
48. Stanley S. M. R., Wee W. K., Lim B. H. [et al.], Direct-in- 60. Van der Lee M. K., van der Weg G., Trang W. A. [et al.],
jection screening for acidic drugs in plasma and neutral Qualitative screening and quantitative determination of
drugs in equine urine by differential-gradient LC-LC cou- pesticides and contaminants in animal feed using compre-
pled MS/MS, Journal of Chromatography B 2007, 848, hensive two-dimensional gas chromatography with time-
292–302. of-flight mass spectrometry, Journal of Chromatogra-
49. Stein S. E., Heller D. N., On the risk of false positive iden- phy A 2008, 1186, 325–339.
tification using multiple ion monitoring in qualitative 61. Van Thuyne W., Van Eenoo P., Delbeke F. T., Compre-
mass spectrometry: Large-scale intercomparisons with hensive screening method for the qualitative detection of
a comprehensive mass spectral library, Journal of the narcotics and stimulants using single step derivatization,
American Society for Mass Spectrometry 2006, 17, Journal of Chromatography B 2007, 857, 259–265.
823–835. 62. Venisse N., Marquet P., Duchoslav E. [et al.], A general
50. Stimpfl T., General unknown screening using GC-MS, unknown screening procedure for drugs and toxic com-
[in:] The encyclopedia of mass spectrometry, Hyphen- pounds in serum using liquid chromatography-electro-
ated methods, vol. 8, Niessen W. M. A. [ed.], Elsevier, spray-single quadrupole mass spectrometry, Journal of
Amsterdam, Oxford 2006. Analytical Toxicology 2003, 27, 7–14.
51. Stimpfl T., Vycudilik W., Automatic screening in post- 63. Ventura R., Roig M., Montfort N. [et al.], High-through-
mortem toxicology, Forensic Science International 2004, put and sensitive screening by ultra-performance liquid
142, 115–127. chromatography tandem mass spectrometry of diuretics
52. Stimpfl T., Demuth W., Varmuza K. [et al.], Systematic and other doping agents, European Journal of Mass Spec-
toxicological analysis: Computer-assisted identification trometry 2008, 14, 191–200.
of poisons in biological materials, Journal of Chromatog- 64. Weinmann W., Gergov M., Goerner M., MS/MS libraries
raphy B 2003, 789, 3–7. with triple-quadrupole tandem mass spectrometers for

Problems of Forensic Sciences 2009, vol. LXXVII, 7–28


Quality assurance aspects of identification with chromatographic... 19

drug identification and drug screening, Analysis 2000, 28,


934–941.
65. Weinmann W., Wiedemann A., Eppinger B. [et al.],
Screening for drugs in serum by electrospray ioniza-
tion/collision-induced dissociation and library searching,
Journal of the American Society for Mass Spectrometry
1999, 10, 1028–1037.
66. World Anti-Doping Agency (WADA), Identification cri-
teria for qualitative assays incorporating chromatography
and mass spectrometry, WADA Technical Document
TD2003IDCR 2004, 1–5.

Corresponding author
Maciej Bogusz
Forensicher Toxikologe, GTFCh
Burghaldeweg 51
D 74889 Sinsheim
e-mail: mbogusz@web.de

Problems of Forensic Sciences 2009, vol. LXXVII, 7–28


Problems of Forensic Sciences 2009, vol. LXXVII, 7–28

ZAPEWNIANIE JAKOŒCI W IDENTYFIKACJI METODAMI


CHROMATOGRAFICZNYMI SPRZʯONYMI ZE SPEKTROMETRI¥ MAS

1. Wprowadzenie tania zakresu mas [46] lub LC-MS-MS w trybie monitoro-


wania wielu reakcji (MRM) przy u¿yciu trzech przejœæ
Problematyka identyfikacji poszczególnych sk³adni- lub widma fragmentacyjnego [26, 48]. Podobna definicja
ków jest kluczowym zagadnieniem analityki toksykolo- zosta³a podana przez Ministerstwo Rolnictwa Stanów
gicznej. Poprzedza ona kolejne etapy, takie jak analiza Zjednoczonych [58] w odniesieniu do identyfikacji pes-
iloœciowa, a przede wszystkim interpretacja uzyskanych tycydów: „potwierdzenie: weryfikacja wczeœniejszej iden-
wyników. tyfikacji analitu wykonywana innym systemem anali-
Wed³ug De Zeeuw i Frankego [10] proces identyfi- tycznym”. De Zeeuw [9] ostrzeg³ przed stosowaniem
kacji rozpoczyna siê od porównania nieznanego zwi¹zku procedur potwierdzaj¹cych jako jedynego dowodu iden-
z bazami danych zawieraj¹cymi substancje wzorcowe tyfikacji i postulowa³ stosowanie ró¿nych procedur identy-
o dok³adnie zdefiniowanych w³aœciwoœciach. Proces iden- fikacji, a nie tylko spektrometrii mas. Niemniej jednak,
tyfikacji musi byæ kontynuowany przy u¿yciu ró¿nych je¿eli spe³nione s¹ odpowiednie wymagania, procedury
kryteriów, a¿ do momentu, kiedy na liœcie pozostanie tyl- oparte na po³¹czeniu rozdzia³u chromatograficznego i de-
ko jeden zwi¹zek. Autorzy ci sformu³owali istotne dla tekcji przy u¿yciu spektrometrii mas s¹ uznawane za wy-
procedur chromatografii sprzê¿onej ze spektrometri¹ mas starczaj¹ce narzêdzie do jednoznacznej identyfikacji.
pytania: Wymagania te, takie jak np. ustalenie progu analitycz-
– W jaki sposób mog¹ byæ porównane w³aœciwoœci ana- nego, odpowiednio wysokiego do konkretnego zadania,
lityczne (np. jak porównaæ widma masowe)? zosta³y sformu³owane przez w³aœciwe organizacje lub in-
– Co powinno byæ uwa¿ane za „w³aœciwe” dopasowa- stytucje na szczeblu krajowym lub miêdzynarodowym.
nie?
– Czy istnieje ró¿nica miêdzy potwierdzeniem i identy-
fikacj¹? 2. Procedury przesiewowe stosowane
– Jakie s¹ wymagania dotycz¹ce odpowiednich baz da- w identyfikacji
nych?
– Jakie s¹ kryteria odrzucenia zwi¹zku? Procedury przesiewowe wykorzystuj¹ce spektrome-
– Jakie jest prawdopodobieñstwo poprawnej identyfi- triê mas s¹ czêsto stosowane jako pierwszy etap identyfi-
kacji? kacji. Mo¿na je podzieliæ na dwie g³ówne grupy: ogólne
De Zeeuw i Franke wyró¿nili trzy nastêpuj¹ce rodzaje nieukierunkowane metody przesiewowe oraz metody
identyfikacji: okreœlenie struktury poprzez identyfikacjê ukierunkowane na konkretne zwi¹zki.
czystego zwi¹zku przy zastosowaniu nowoczesnych me-
tod spektrometrycznych, potwierdzenie bêd¹ce wynikiem 2.1. Ogólne analizy przesiewowe
udanego porównania w³aœciwoœci nieznanego zwi¹zku
ze zwi¹zkiem odniesienia oraz rozpoznanie, jako rezultat W przypadku analiz nieukierunkowanych, które s¹
odpowiedniego dopasowania, w³aœciwoœci nieznanego czêsto okreœlane w toksykologii klinicznej i s¹dowej jako
zwi¹zku z baz¹ danych. W zaleceniach sformu³owanych analiza „ogólnie nieznanych substancji”, logicznym wy-
przez Agencjê ds. ¯ywnoœci i Leków (FDA) [23] po- borem jest zastosowanie metody kapilarnej GC po³¹czo-
twierdzenie zdefiniowano podobnie, jako „jednoznaczn¹ nej ze spektrometri¹ mas z jonizacj¹ elektronow¹ (GC-
identyfikacjê obecnoœci zwi¹zku poprzez porównanie do EI-MS). Widma masowe uzyskane technik¹ EI-MS nie
substancji wzorcowej (spektrometria mas)”. Jednak¿e ter- s¹ zale¿ne od tego, jaki aparat i warunki chromatograficz-
min „potwierdzenie” mo¿e byæ stosowany w innym zna- ne zosta³y zastosowane. Dostêpne s¹ obecnie komplekso-
czeniu. W toksykologii s¹dowej i kontroli antydopingo- we biblioteki widm masowych uzyskanych technik¹ EI
wej strategia analityczna zazwyczaj sk³ada siê z dwóch obejmuj¹ce tysi¹ce zwi¹zków [35, 36, 59]. Biblioteki te
etapów. Pierwszym etapem jest procedura przesiewowa umo¿liwiaj¹ wstêpn¹ identyfikacjê zwi¹zku poprzez ich
wykonywana najczêœciej bardzo czu³ymi, ale mniej spe- przeszukiwanie, nawet bez zwi¹zku odniesienia. W przy-
cyficznymi metodami (g³ównie testami immunochemicz- padku, kiedy nie mo¿na uzyskaæ zwi¹zku odniesienia do
nymi lub tandemow¹ spektrometri¹ mas przy u¿yciu analizy porównawczej, decyzja dotycz¹ca identycznoœci
tylko jednego przejœcia). Kolejnym jest potwierdzenie widma masowego powinna byæ podejmowana z ogromn¹
pozytywnego wyniku metody przesiewowej, które jest rozwag¹. Szczególnie wa¿ne jest to w przypadku niektó-
wykonywane metodami o najwiêkszej mo¿liwej specy- rych „egzotycznych” zwi¹zków proponowanych przez
ficznoœci, takimi jak technika GC-MS w trybie przemia- oprogramowanie urz¹dzenia jako pretendentów do iden-
Zapewnianie jakoœci w identyfikacji metodami chromatograficznymi... 21

tyfikacji. Jednak w wiêkszoœci procedur identyfikacji jak GC-MS-MS i LC-MS-MS, obie w trybie MRM, a tak-
stosowanych w dziedzinach zwi¹zanych z prawem, jak ¿e GC-MS-TOF lub GC-MS-SIM. Identyfikacja odbywa
np. analiza œrodków dopingowych, toksykologia s¹dowa siê poprzez porównanie w³aœciwoœci chromatograficz-
lub kontrola jakoœci ¿ywnoœci, wstêpna identyfikacja dziê- nych i obecnoœci specyficznych jonów fragmentacyjnych
ki bibliotece GC-EI-MS musi byæ potwierdzona w na- o intensywnoœciach odpowiadaj¹cym zaprogramowanym
stêpnym etapie przy u¿yciu substancji odniesienia anali- wartoœciom zarejestrowanym dla substancji wzorcowych.
zowanej w identycznych warunkach jak badana próbka. Dla poszczególnych grup zwi¹zków stworzono biblioteki
W pe³nym widmie masowym otrzymanym technik¹ GC- odniesienia zorientowane na konkretne zadania. Biblio-
MS-EI powinny byæ obecne co najmniej cztery wybrane teki kompleksowe, np. biblioteka pestycydów opracowa-
jony we w³aœciwych proporcjach intensywnoœci, w tym – na w oparciu o metodê LC-MS, mog¹ obejmowaæ kilka-
je¿eli to mo¿liwe – jon molekularny. Widma masowe po- set substancji.
chodz¹ce od t³a oraz przeszkadzaj¹ce w identyfikacji po- Definicje pozytywnej identyfikacji lub potwierdzenia
winny zostaæ odjête. Mo¿e to znacznie zwiêkszyæ mo¿- w wyniku analizy przesiewowej technik¹ LC-MS mog¹
liwoœci techniki GC-MS [52]. siê ró¿niæ. Niektórzy autorzy wymieniaj¹ wysokie po-
Nale¿y podkreœliæ, ¿e decyzja dotycz¹ca wstêpnej dobieñstwo co najmniej dwóch z trzech widm otrzyma-
identyfikacji w oparciu o analizê statystyczn¹ widma ma- nych przy trzech energiach fragmentacji [28], inni nato-
sowego dopasowanego z biblioteki widm powinna byæ miast dopasowanie co najmniej 60% odwrotnego do-
zawsze wykonywana przez bieg³ego toksykologa. To on, pasowania uzyskane podczas przeszukiwania biblioteki
w zale¿noœci od przypadku, decyduje, czy zosta³ osi¹g- [62] lub podobieñstwo przynajmniej piêciu najbardziej
niêty odpowiedni stopieñ identyfikacji [9, 42, 43]. intensywnych jonów [4].
Chromatografia cieczowa po³¹czona ze spektrometri¹ Dla niektórych zwi¹zków zawieraj¹cych chlorowce,
mas jest technik¹ potencjalnie bardziej odpowiedni¹ do takich jak np. chloramfenikol, zastosowano alternatywne
celów przesiewowych ni¿ GC-MS, gdy¿ obejmuje znacz- strategie identyfikacji. Bogusz i in. [2] wykorzystali tech-
nie szersze spektrum zwi¹zków o ró¿nej polarnoœci. Jed- nikê LC-MS-MS z ujemn¹ elektronow¹ jonizacj¹ (ESI,
nak technika ta ma kilka istotnych ograniczeñ dotycz¹- tryb MRM). Monitorowali oni trzy przejœcia pozbawio-
cych zarówno chromatografii, jak i zwi¹zanych ze spek- nej protonu cz¹steczki (m/z 321). Sformu³owali nastêpu-
trometri¹ mas. LC, pomimo swojej uniwersalnoœci, jest j¹ce kryteria pozytywnej identyfikacji chloramfenikolu:
w porównaniu do kapilarnej GC znacznie mniej selek- Rt badanego zwi¹zku w zakresie ± 1% wzorca wewnêtrz-
tywn¹ technik¹ rozdzia³u. Z drugiej strony widma ma- nego (deuterowana pochodna badanego zwi¹zku), obec-
sowe zwi¹zków zarejestrowane za pomoc¹ ró¿nych apa- noœæ trzech jonów potomnych pochodz¹cych od jonu
ratów LC-MS mog¹ siê znacznie ró¿niæ wzglêdn¹ inten- macierzystego oraz stosunki intensywnoœci jonów po-
sywnoœci¹ jonów fragmentacyjnych, nawet je¿eli sto- tomnych w zakresie ± 2SD œredniej wartoœci kontrolnej,
sowane s¹ identyczne parametry analizy [3]. Sprawia to, czyli ± 25%. Mottier i in. [37] wykorzystali zalety wyni-
¿e bardzo trudno jest bezpoœrednio wykorzystywaæ wid- kaj¹ce z obecnoœci dwóch atomów chloru w cz¹steczce
ma masowe uzyskane w innym laboratorium. Dlatego te¿ chloramfenikolu i monitorowali dwa jony macierzyste:
LC-MS jest doskona³¹ technik¹ analizy ukierunkowanej m/z 321 oraz m/z 323. Monitorowali oni przejœcia ka¿de-
lub potwierdzaj¹cej, ale nie³atwo j¹ zastosowaæ w anali- go izotopu do jonów potomnych o m/z 152 i 257. Kryteria
zie szerokiego spektrum zwi¹zków, np. w analizie „ogól- identyfikacji by³y nastêpuj¹ce: Rt badanego zwi¹zku w za-
nie nieznanych substancji”. Pomimo tego, w ostatnich la- kresie ± 1% wzorca wewnêtrznego (deuterowana po-
tach za pomoc¹ technik LC-MS [45, 65], LC-MS-MS chodna badanego zwi¹zku) oraz zmiennoœæ stosunków
[31, 64] i LC-MS-TOF [11, 21, 38] opracowano kilka bi- intensywnoœci jonów potomnych w zakresie 15–25% œred-
bliotek widm obejmuj¹cych setki zwi¹zków chemicz- niej kontrolnej wartoœci.
nych. Prace przegl¹dowe dotycz¹ce tej tematyki zosta³y Poniewa¿ intensywnoœæ i niekiedy sposób fragmenta-
niedawno opublikowane przez Gergova [20] oraz Mar- cji w technice LC-MS s¹ zale¿ne od zastosowanych wa-
queta [30]. runków i urz¹dzenia, powszechnie zak³ada siê, ¿e wzo-
rzec odniesienia powinien byæ analizowany równolegle
2.2. Analizy ukierunkowane w dok³adnie takich samych warunkach, jak badana prób-
ka, a najlepiej w tym samym nastrzyku.
Analizy ukierunkowane i ich dalsze potwierdzenie s¹ Wymagania dotycz¹ce jakoœci procedur przesiewo-
stosowane, je¿eli ma byæ zidentyfikowany konkretny wych wykorzystuj¹cych technikê spektrometrii mas zo-
zwi¹zek lub okreœlona grupa zwi¹zków. Ma to miejsce sta³y omówione przez ró¿nych autorów, takich jak Drum-
w przypadku analiz w kierunku substancji kontrolowa- mer [15], Stimpfl i in. [50, 51] oraz Maurer [32] w odnie-
nych, jak np. œrodków uzale¿niaj¹cych [61], substancji sieniu do toksykologii klinicznej i s¹dowej, a tak¿e przez
dopingowych [19, 54] lub zanieczyszczeñ ¿ywnoœci [60]. Thevisa i Schänzera [54] w odniesieniu do analizy œrod-
Stosowane s¹ wtedy bardzo specyficzne metody, takie ków dopingowych.

Problems of Forensic Sciences 2009, vol. LXXVII, 7–28


22 M. J. Bogusz

3. Rozwa¿ania metodyczne istotne dla 3.2. Optymalizacja rozdzia³u chromatograficznego


identyfikacji
W odniesieniu do sprawnoœci rozdzia³u chromatogra-
Istnieje wiele czynników, które mog¹ mieæ wp³yw na ficznego, kapilarna GC jest dojrza³¹ technik¹. Niemniej
identyfikacjê technik¹ spektrometrii mas. Na ró¿nych jednak pojawi³y siê ostatnio przyk³ady ukazuj¹ce, ¿e ta
etapach procedury identyfikacji nale¿¹ do nich: wp³yw dziedzina chromatografii mo¿e byæ nadal udoskonalana
równolegle ekstrahowanych sk³adników matrycy, nie- w celu u³atwienia identyfikacji w z³o¿onych mieszaninach
pe³ny rozdzia³ chromatograficzny zwi¹zków, ogólny stan lub matrycach. Van der Lee i in. [60] przedstawili metodê
aparatu, a tak¿e nasycenie detektora i zniekszta³cenie przesiewow¹ ukierunkowan¹ na oznaczanie 106 pesty-
piku poprzez nadmiar substancji. Wszystkie te czynniki cydów i zanieczyszczeñ w paszach dla zwierz¹t. Proce-
mog¹ wp³ywaæ na zmianê intensywnoœci jonów w kon- dura ta by³a oparta na ekstrakcji rozpuszczalnikiem,
kretnym widmie masowym i powodowaæ niedopasowa- oczyszczaniu technik¹ chromatografii ¿elowej i analizie
nie, a w konsekwencji b³êdnie pozytywn¹ lub b³êdnie dwuwymiarow¹ GC z pe³nym przemiataniem widma
negatywn¹ identyfikacjê [33, 34]. Istnieje kilka elemen- TOF-MS. Wszystkie zwi¹zki by³y wykrywane automa-
tów optymalizacji procesu identyfikacji technik¹ spek- tycznie na poziomie powy¿ej 50 ng/ml. Dwuwymiarowa
trometrii mas. Mo¿na je podzieliæ na optymalizacjê przy- GC oraz TOF-MS by³y równie¿ stosowane przez Baner-
gotowania próbki, optymalizacjê rozdzia³u chromatogra- jee i in. [1] w analizie wielosk³adnikowych pozosta³oœci
ficznego oraz optymalizacjê parametrów pracy spektro- pestycydów w winogronach. Wykorzystano tutaj kombi-
metru mas. nacjê szeregowo po³¹czonych kolumn kapilarnych – nie-
polarnej i polarnej. Metoda ta rozwi¹zywa³a problemy
3.1. Optymalizacja przygotowania próbki do analizy zwi¹zane z koelucj¹, obserwowane w trybie pe³nego ska-
technik¹ spektrometrii mas nowania w jednowymiarowej analizie GC-MS i umo¿li-
wia³a rozdzia³ chromatograficzny 51 pestycydów w czasie
Decyduj¹ce znaczenie w procedurze identyfikacji analizy wynosz¹cym 24 min wraz z potwierdzeniem po-
technik¹ GC-EI-MS ma optymalne, dostosowane do ro- przez przeszukiwanie biblioteki widm masowych. Zauto-
dzaju materia³u, przygotowanie próbki. Przygotowanie matyzowana technika bezpoœredniego wprowadzenia
takie mo¿e siê sk³adaæ z etapu hydrolizy sprzê¿onych próbki w po³¹czeniu z obszern¹ dwuwymiarow¹ chroma-
zwi¹zków, derywatyzacji lub kilku etapów oczyszczania. tografi¹ gazow¹ sprzê¿on¹ ze spektrometrem mas z anali-
W metodach przesiewowych i identyfikacyjnych tech- zatorem czasu przelotu zosta³a zastosowana do opraco-
nik¹ GC-MS wœród procedur izolacji stosowana by³a wania przesiewowej metody oznaczania polichlorowa-
g³ównie ekstrakcja ciecz-ciecz [35] i ekstrakcja do fazy nych dibenzo-p-dioksyn (dibenzofuranów) i polichloro-
sta³ej (SPE) [16]. W ostatnich latach coraz czêœciej jest wanych bifenyli w oleju z ryb [27]. Kolbrich i in. [29] za-
wykorzystywana mikroekstrakcja do fazy sta³ej (SPME) stosowali dwuwymiarow¹ technikê GC-EI-MS (SIM)
ze wzglêdu na fakt, ¿e nie wymaga ona u¿ywania rozpu- z zatê¿aniem kriogenicznym w celu oznaczania w ludzkiej
szczalników i jest szczególnie przydatna w analizie prze- surowicy metylenodioksyetyloamfetaminy (MDEA),
siewowej lotnych zwi¹zków pochodzenia naturalnego MDMA i jej metabolitu 3,4-metylenodioksyamfetaminy
lub syntetycznego [np. 5, 6, 24, 41]. (MDA), 4-hydroksy-3-metoksymetamfetaminy (HMMA)
Dla niektórych zwi¹zków, a w szczególnoœci du¿ych oraz 4-hydroksy-3-metoksyamfetaminy (HMA). Metoda ta
cz¹steczek o haptenogennych w³aœciwoœciach, zosta³y zapewnia³a niskie granice oznaczalnoœci, a tego typu sys-
opracowane procedury ekstrakcji powinowactwa immu- tem chromatograficzny wydaje siê odpowiedni do zasto-
nologicznego. Ho i in. [26] wyosobnili z koñskiego oso- sowania w analizie innych zwi¹zków i z³o¿onych matryc.
cza insuliny pochodz¹ce z ró¿nych Ÿróde³ (ludzi, byd³a W przypadku metody HPLC ostatnie lata przynios³y
i œwiñ) przy u¿yciu str¹cania powinowactwa immunolo- wyraŸny postêp zwi¹zany z wprowadzeniem drobnoziar-
gicznego z magnetycznymi kulkami powlekanymi prze- nistych kolumn (wielkoœci cz¹stek poni¿ej 2 mm). Tech-
ciwcia³ami z nastêpuj¹cym rozdzia³em przez wirowanie nika ta, znana jako ultrasprawna chromatografia cieczo-
frakcjon uj¹ce. Otrzymane ekstrakty by³y analizowane wa (UPLC), a umo¿liwiaj¹ca znacznie lepszy rozdzia³
z u¿yciem aparatu LC-MS-MS wyposa¿onego w spektro- w krótszym czasie, zosta³a natychmiast zastosowana do
metr mas typu QTrap. Dla celów przesiewowych monito- celów identyfikacyjnych w po³¹czeniu ze spektrometri¹
rowano przejœcie do jednego wspólnego iminiowego mas. UPLC umo¿liwia wysokowydajn¹ analizê w dzie-
jonu tyrozyny. Potwierdzenie zosta³o dokonane przez co dzinie badañ œrodków dopingowych, a w szczególnoœci
najmniej trzy przejœcia charakterystyczne dla danej insu- œrodków moczopêdnych. Ventura i in. [63], u¿ywaj¹c
liny, bior¹c pod uwagê równie¿ czas retencji. Metoda ta urz¹dzenia UPLC-MS-MS pracuj¹cego w trybie ujemnej
by³a stosowana w kontroli antydopingowej w celu roz- i dodatniej jonizacji, wykrywali w ekstraktach z moczu
ró¿nienia koñskiej insuliny od insuliny pochodzenia eg- 34 wybrane diuretyki i inne œrodki dopingowe. Ca³ko-
zogennego. wity czas analizy wynosi³ 5 min, a metoda pod ka¿dym

Problems of Forensic Sciences 2009, vol. LXXVII, 7–28


Zapewnianie jakoœci w identyfikacji metodami chromatograficznymi... 23

wzglêdem spe³ni³a wymogi ustalone przez Œwiatow¹ z potrójnym kwadrupolem charakteryzuj¹ siê mo¿liwoœ-
Agencjê Antydopingow¹. Thörngren i in. [56] opracowa- ci¹ okreœlenia masy z dok³adnoœci¹ poni¿ej 1 mDa [40].
li przesiewow¹ metodê oznaczania 130 substancji (diu- Poza zastosowaniem w toksykologicznych badaniach
retyków, œrodków pobudzajcych oœrodkowy uk³ad ner- przesiewowych [20, 22] spektrometria mas o wysokiej
wowy i opiatów) opart¹ na technice UPLC-MS-MS rozdzielczoœci zosta³a wykorzystana w innych procedu-
z bezpoœrednim nastrzykiwaniem prób moczu. Próbki rach przesiewowych i identyfikacyjnych. Hernandez i in.
by³y wprowadzane na kolumnê z odwróconym uk³adem [24] opracowali procedurê opart¹ na technice chromato-
faz pod³¹czon¹ do szybkoskanuj¹cego tandemowego grafii gazowej sprzê¿onej ze spektrometri¹ mas o wyso-
spektrometru masowego z jonizacj¹ przez elektrorozpy- kiej rozdzielczoœci z analizatorem czasu przelotu (GC-
lanie, umo¿liwiaj¹cego szybk¹ zmianê polaryzacji rejes- TOF-MS) do identyfikacji w wodzie 60 zanieczyszczeñ
tracji jonów. Oprogramowanie wykorzystywane do opra- wyosobnionych za pomoc¹ mikroekstrakcji do fazy sta-
cowania wyników sporz¹dza³o raporty ukazuj¹ce nie- ³ej. Identyfikacja dla jednego analitu by³a oparta na oce-
prawid³owe wyniki analityczne w próbce. 96-do³kowa nie obecnoœci do piêciu reprezentatywnych jonów o m/z
p³ytka mog³a byæ przeanalizowana w ci¹gu 16 h. zmierzonych przy wysokiej dok³adnoœci masy i okreœle-
Kilka aplikacji UPLC-MS dotyczy³o analizy pozosta- niu dla nich stosunków intensywnoœci Q/q (Q jon iloœ-
³oœci pestycydów. Technika UPLC-TOF-MS zosta³a za- ciowy; q, jon potwierdzaj¹cy). Strategia ta doprowadzi-
stosowana do szybkiej ukierunkowanej analizy jakoœ- ³a do wykrycia w kilku próbkach wody analizowanych
ciowej i iloœciowej 100 pestycydów w truskawkach [53]. zwi¹zków w niskich stê¿eniach na poziomie ppb.
Dok³adny pomiar masy dodatnich i ujemnych jonów Problemem dla laboratoriów kontroli œrodowiska sta-
umo¿liwia³ ich ekstrakcjê z nastêpuj¹cym po niej „zbie- ³o siê kilkaset œrodków ochrony roœlin, których stosowa-
raniem danych z ca³ego zakresu mas”. Na oznaczenia nie nie zosta³o niedawno zakazane przez europejskie i ame-
mia³y wp³ywu interferencje z t³a i równoczeœnie koeluo- rykañskie agencje ochrony œrodowiska. Poniewa¿ trudno
wane sk³adniki obserwowane podczas rozdzia³u gradien- by³oby kontrolowaæ wszystkie te zwi¹zki technikami
towego technik¹ UPLC (w cyklu wynosz¹cym 6,5 min na GC-MS, LC-MS lub LC-MS-MS, Thurman i Ferrer [57]
jedn¹ analizê). W innych badaniach [44] analizowano opracowali nieukierunkowane procedury przesiewowe
technik¹ UPLC-MS-MS (MRM) 90 pestycydów wyosob- w oparciu o techniki ³¹czone LC-TOF-MS, LC-IT-MS
nionych z soków owocowych prost¹ ekstrakcj¹ acetoni- i LC-QTOF-MS. Jako referencyjne bazy danych zasto-
trylem. Rozdzia³ zosta³ osi¹gniêty w czasie analizy wy- sowano komercyjnie dostêpne bazy The Merck Index
nosz¹cym 11 min. i ChemIndex. Strategia identyfikacji sk³ada³a siê z czte-
rech etapów: analizy LC-TOF-MS z pe³nym przemiata-
3.3. Optymalizacja parametrów spektrometru mas niem widma, przeszukiwania biblioteki pod k¹tem wzo-
rów empirycznych i któregoœ z izotopów A+2 (dla halo-
Dok³adne okreœlenie masy wydaje siê jednym z naj- genów lub S), okreœlenia struktury za pomoc¹ technik
bardziej obiecuj¹cych narzêdzi identyfikacji we wspó³- LC-IT-MS lub LC-Q-TOF-MS-MS i koñcowych analiz
czesnej spektrometrii mas. Zmierzenie masy cz¹stecz- porównawczych. Dok³adne okreœlenie masy mo¿e byæ
kowej pozwala zidentyfikowaæ nieznany jon lub okreœliæ bardzo przydatne w analizie zwi¹zków, które nie tworz¹
jego wzór chemiczny, a tym samym potwierdziæ po- charakterystycznych jonów fragmentacyjnych w analizie
dobieñstwo do znanego sk³adnika w przypadku prze- tandemow¹ spektrometri¹ mas, szczególnie w trybie MRM.
szukiwania biblioteki [21]. Dok³adnoœæ masy mo¿e byæ Nielen i in. [39] porównali skutecznoœæ ró¿nych wysoko-
wyra¿ona w wartoœciach bezwzglêdnych, zazwyczaj rozdzielczych technik LC-MS w analizie pozosta³oœci
w milidaltonach (mDa) lub w wartoœciach wzglêdnych, hormonów w próbkach ¿ywnoœci. Autorzy ci stosowali
np. w czêœciach na milion (ppm). Wartoœci te oblicza siê trzy rodzaje instrumentów: LC-QTOF-MS, LC-FTIR-
w nastêpuj¹cy sposób: MS i LC-FT-Orbitrap-MSn. Podkreœlali oni potrzebê op-
tymalnej wstêpnej obróbki i oczyszczenia próbki dla in-
mDa: [(m/z)zmierzone – (m/z)obliczone] ´ 1000, {1} strumentów o mniejszych dok³adnoœciach (ok. 20 ppm),
ppm: 106 ´ [(m/z)zmierzone – (m/z)obliczone] / (m/z). {2} poniewa¿ rozdzielczoœæ LC i rozdzielczoœæ MS s¹ silnie
ze sob¹ powi¹zane i maj¹ istotny wp³yw na dok³adnoœæ
Dok³adnoœæ masy zale¿y od rodzaju zastosowanego pomiaru masy. Instrumenty o wy¿szej rozdzielczoœci mas
aparatu. Dla standardowego instrumentu z analizatorem mog¹ z³agodziæ skutki nieca³kowitego rozdzia³u chro-
w postaci kwadrupola lub pu³apki jonowej mo¿liwe jest matograficznego. Mo¿e to byæ zilustrowane analiz¹ sta-
uzyskanie dok³adnoœci mas 50–100 mDa, dla TOF lub nozololu; nieokreœlony fragment obserwowany dla
QTOF poni¿ej 10 mDa, natomiast dla FTICR-MS i in- QTOF-MS (m/z 161,1223 przy rozdzielczoœci mas
strumentów z analizatorem typu orbitrap – 1 mDa lub 5000 FWHM) okaza³ siê w analizie LC-FTIR-MS duble-
mniej. Niektóre z niedawno skonstruowanych aparatów tem jonów o niewielkich ró¿nicach mas (m/z 161,1073
oraz m/z 161,1324 przy rozdzielczoœci mas 250000).

Problems of Forensic Sciences 2009, vol. LXXVII, 7–28


24 M. J. Bogusz

Oprócz spektrometrii mas o wysokiej rozdzielczoœci sywnoœci dwóch jonów fragmentacyjnych do jonu frag-
do celów identyfikacji zastosowano niedawno inne roz- mentacyjnego o najwiêkszej intensywnoœci. Jak okreœ-
wi¹zania instrumentalne. Najbardziej obiecuj¹ce jest lono w procedurze walidacyjnej, wskaŸniki te powinny
u¿ycie warunkowo zale¿nej akwizycji (IDA) w po³¹cze- mieœciæ siê w zakresie ± 3SD. Concheiro i in. [8] opubli-
niu z wzmocnionym widmem jonu potomnego (EPI). kowali badania dotycz¹ce jednoczesnego oznaczania ró¿-
Technika ta zosta³a wykorzystana przez Stanleya i in. nych œrodków odurzaj¹cych w moczu technik¹ LC-MS-
[48] w celu wykrycia 32 zwi¹zków o charakterze kwaœ- MS. Dla ka¿dego zwi¹zku monitorowali dwa przejœcia.
nym w osoczu koñskim i 11 zwi¹zków o charakterze obo- Zmiennoœæ wewn¹trzgrupowa i miêdzygrupowa wzglêd-
jêtnym w moczu. W przypadku osocza przeprowadzo- nych intensywnoœci jonów by³a badana przy dwóch po-
no str¹canie acetonitrylem zawieraj¹cym wzorzec wew- ziomach stê¿eñ. Zmiennoœæ miêdzygrupowa by³a znacz-
nêtrzny, a nastêpnie odwirowywanie; próby moczu, do nie wiêksza ni¿ wewn¹trzgrupowa. Autorzy stwierdzili,
których dodawano wzorzec wewnêtrzny, by³y tylko od- ¿e stosunek intensywnoœci jonów jako kryterium identy-
wirowywane. Metoda chromatograficzna polega³a na fikacji nale¿y traktowaæ z du¿¹ ostro¿noœci¹, a tak¿e zde-
ekstrakcji przy u¿yciu kolumny Oasis HLB i izokratycz- cydowanie zalecili, aby w celu spe³nienia okreœlonych
nym rozdziale na kolumnie Chromolith RP-18e bezpo- kryteriów wraz z rzeczywistymi próbkami analizowaæ
œrednio ze sob¹ po³¹czonych. Zwi¹zki by³y wykrywane w tym samym dniu próbki wzorcowe.
hybrydowym tandemowym spektrometrem mas typu Stein i Heller [49] opublikowali bardzo wa¿ne bada-
QTRAP pracuj¹cym w trybie dodatniej i ujemnej joniza- nia dotycz¹ce czynników odpowiedzialnych za b³êdnie
cji. Dla ka¿dego zwi¹zku monitorowano jedno przejœcie. dodatnie identyfikacje za pomoc¹ spektrometrii mas.
Zastosowano technikê IDA, a je¿eli sygna³ by³ wiêkszy Przeprowadzono nastêpuj¹ce eksperymenty: ze zreduko-
ni¿ wybrana wartoœæ, rejestrowano widmo EPI. Dla wanej biblioteki widm masowych NIST/EPA/NIH [59]
wiêkszoœci zwi¹zków osi¹gniêto jednoznaczn¹ identyfi- zawieraj¹cej 96464 zapisów co dziesi¹te widmo (oko³o
kacjê. Autorzy stwierdzili, ¿e metoda ta nie nadaje siê do 9600 widm) zosta³o wybrane jako widmo do identyfika-
analizy sterydów anabolicznych, poniewa¿ dla tej grupy cji. Widma te zosta³y wyeliminowane z biblioteki przed
zwi¹zków obserwowano bardzo podobne widma EPI. wykonaniem przeszukiwania, co spowodowa³o, ¿e wszyst-
Drees i in. [14] porównali mo¿liwoœci identyfikacyjne kie widma dopasowane z biblioteki dla identyfikowane-
stosunków MRM i MRM-IDA przy zastosowaniu hy- go widma z danego zestawu ograniczeñ by³y b³êdnie po-
brydowego spektrometru mas QTrap w postaci potrójne- zytywne. Eksperymenty przeszukiwania zosta³y wyko-
go kwadrupola (liniowej pu³apki jonowej). Analizowali nane przy u¿yciu ró¿nej liczby pików do dopasowywania
oni 14 wybranych œrodków odurzaj¹cych w ró¿nych stê- w liczbie od 1 do 8. Intensywnoœæ okna zosta³a okreœlona
¿eniach. Obie metody dobrze sprawdza³y siê w niskich na 0,25, co oznacza, ¿e widma poszukiwane oraz te z bi-
stê¿eniach, natomiast przy bardzo wysokich stê¿eniach blioteki uznawano za pasuj¹ce przy danym m/z, je¿eli
metoda MRM-IDA charakteryzowa³a siê lepszymi mo¿- ró¿nice w intensywnoœci by³y mniejsze ni¿ 25%. Stwier-
liwoœciami potwierdzania. dzono, ¿e liczba pików, jak równie¿ intensywnoœci piku,
Po³¹czenie technik LC-MS-MS i (lub) GC-MS-NPD wyraŸnie zmniejszaj¹ prawdopodobieñstwo uzyskania
by³o u¿ywane przez Thevisa i in. [55] w celu oznaczenia wyników b³êdnie dodatnich. Autorzy stwierdzili, ¿e do
zwi¹zków obecnych w narkotykach skonfiskowanych na identyfikacji powinny byæ stosowane wiêcej ni¿ trzy jony
nielegalnym rynku, zawieraj¹cych g³ównie steroidy ana- fragmentacyjne (w trybie przemiatania ca³ego widma lub
boliczne i œrodki o dzia³aniu pobudzaj¹cym seksualnie. w trybie SIM), a ich wartoœci m/z powinny byæ starannie
Narkotyki by³y rozpuszczane w metanolu i poddawane dobrane. Nale¿y unikaæ korelacji z wysoce prawdopo-
analizie technik¹ LC-ESI-MS-MS przy zastosowaniu dobnymi pikami, takimi jak ró¿nica 14 amu (utrata grupy
aparatu QTrap. Dla ka¿dego zwi¹zku monitorowano trzy metylowej) lub 18 amu (utrata wody).
przejœcia. W przypadku badania technik¹ GC-MS-NPD
stosowano wielodro¿ny zawór dozuj¹cy, dziêki któremu
dozowana próbka by³a rozdzielana na dwóch identyczny- 4. Prawne i regulacyjne aspekty identyfikacji
ch kolumnach; jedna kolumna by³a pod³¹czona do detek-
tora EI-MS, a druga do detektora azotowo-fosforowego Kilka bran¿owych organizacji, jak równie¿ agencje
(NPD). krajowe i miêdzynarodowe, sformu³owa³y zalecenia
Obliczanie wzglêdnych intensywnoœci jonów (stosu- i wytyczne dotycz¹ce kryteriów identyfikacji z zastoso-
nków jonów) jest zwykle stosowane jako jedna z cech waniem metod spektrometrii mas. Dokumenty te mo¿na
identyfikacyjnych [2, 37]. Feng i in. [18] opracowali podzieliæ na dwie g³ówne grupy. Pierwsza grupa wynika
procedurê jednoczesnego oznaczania 30 ró¿nych œrod- z regulacji prawnych na szczeblu krajowym i miêdzy-
ków odurzaj¹cych w moczu technik¹ LC-MS-MS. Dla narodowym, jak wydane przez s³u¿bê zdrowia Stanów
ka¿dego zwi¹zku monitorowali trzy przejœcia. W celu Zjednoczonych dokumenty dotycz¹ce laboratoriów kli-
identyfikacji obliczano i wykorzystywano stosunki inten- nicznych [12] oraz testowania œrodków odurzaj¹cych

Problems of Forensic Sciences 2009, vol. LXXVII, 7–28


Zapewnianie jakoœci w identyfikacji metodami chromatograficznymi... 25

w miejscu pracy [13], a tak¿e opublikowane przez Uniê ne intensywnoœci powinny odpowiadaæ intensywnoœ-
Europejsk¹ [17]. Druga grupa obejmuje zalecenia i wy- ciom w standardach. Nale¿y wyjaœniæ obecnoœæ in-
tyczne zawodowych organizacji na szczeblu krajowym nych niepowi¹zanych widocznych jonów (np. z sk³ad-
i miêdzynarodowym, takich jak Œwiatowa Agencja An- ników matrycy). Je¿eli zastosowano odejmowanie t³a,
tydopingowa – WADA, Zrzeszenie Patologów Amery- nale¿y to zaznaczyæ i okreœliæ. W przypadku MS1
kañskich – CAP, Agencja ds. ¯ywnoœci i Leków – FDA, SIM wzglêdne intensywnoœci trzech strukturalnie spe-
Ministerstwo Rolnictwa Stanów Zjednoczonych, Stowa- cyficznych jonów powinny pasowaæ do wzorca w za-
rzyszenie Toksykologów S¹dowych – SOFT [47] czy kresie ± 10% (bezwzglêdna). W przypadku czterech
Niemieckie Towarzystwo Chemii Toksykologicznej i S¹- lub wiêkszej iloœci strukturalnie specyficznych jonów
dowej – GTFCh [22]. dopasowanie powinno byæ na poziomie ± 15% (bez-
FDA opublikowa³o koñcowe wytyczne dotycz¹ce wzglêdna). Wzglêdne intensywnoœci dla wiêcej ni¿
rozwoju, oceny i stosowania metod spektrometrii mas do trzech jonów, które zawieraj¹ mniej specyficznych
potwierdzania to¿samoœci pozosta³oœci leków dla zwie- jonów, takich jak izotopy lub pochodz¹ce z utraty
rz¹t [23]. Zgodnie z wytycznymi, procedury wykorzystu- wody, powinny byæ dopasowane do wzorca w za-
j¹ce spektrometriê mas powinny zawieraæ ka¿dy z nastê- kresie ± 10%. W pe³nym i czêœciowym przemiataniu
puj¹cych punktów: MSn widmo powinno wizualnie odpowiadaæ widmu
– pakiet walidacyjny z opracowuj¹cego metodê labo- standardu i powinna istnieæ ogólna zgodnoœæ miêdzy
ratorium zawieraj¹cy powtórzenia próbek, oryginal- wzglêdnymi intensywnoœciami uzyskanymi dla prób-
nych i wzbogaconych, pokazuj¹cy zerowy odsetek ki i standardu. Wszystkie strukturalnie specyficzne
b³êdnie dodatnich i 10% b³êdnie ujemnych wyników jony, o których mowa w opisie metody, powinny byæ
przy za³o¿onym poziomie tolerancji, ukazuj¹cy od- obecne. Je¿eli strukturalnie specyficzny jon macie-
pornoœæ metody i brak interferencji ze strony innych rzysty ca³kowicie rozpada siê do jonów potomnych,
zwi¹zków lub sk³adników matrycy; pojawienie siê przynajmniej dwóch strukturalnie spe-
– opis metody zawieraj¹cy tak¿e wzór i pe³ne widmo cyficznych jonów potomnych powinno byæ wystar-
analizowanego zwi¹zku, dane dotycz¹ce widm dla co czaj¹ce w MSn+1. Nale¿y wyjaœniæ obecnoœæ innych
najmniej trzech strukturalnie specyficznych jonów, niepowi¹zanych widocznych jonów (np. z sk³adni-
które ca³kowicie okreœlaj¹ macierzyst¹ cz¹steczkê lub ków matrycy). Je¿eli zastosowano odejmowanie t³a,
wiêcej jonów, je¿eli s¹ uwzglêdnione niespecyficzne nale¿y to zaznaczyæ i okreœliæ. W przypadku MSn
jony, proponowane wzory jonów fragmentacyjnych SRM, je¿eli jon macierzysty jest ca³kowicie frag-
wraz z œcie¿kami fragmentacji, uzasadnienie specy- mentowany, wzglêdna intensywnoœæ strukturalnie
fiki wybranych jonów lub zakresu przemiatania wid- specyficznych jonów potomnych powinna odpowia-
ma, kryteria potwierdzenia i kryteria operacyjne, jak daæ wzorcowi w granicach ± 10% w przypadku, gdy
równie¿ czêœæ dotycz¹c¹ kontroli jakoœci; monitorowane by³y dwa jony lub ± 20% w przy-
– kryteria potwierdzenia. W procedurze potwierdza- padku, gdy monitorowane by³y trzy lub wiêcej jonów;
j¹cej wzorzec (lub wzorce) porównywany powinien – kontrola jakoœci powinna zawieraæ nastêpuj¹ce punk-
byæ analizowany równoczeœnie, w razie potrzeby ty: ustalenie dopasowania systemu, analizê przynaj-
w obecnoœci ekstrahowanej matrycy. Mo¿e byæ zasto- mniej jednej próby kontrolnej i jednej wzbogaconej
sowany ka¿dy z nastêpuj¹cych chromatogramów otrzy- próby kontrolnej oraz kontrolê ewentualnego prze-
manych technik¹ spektrometrii mas: TIC, RIC, SIM niesienia.
lub SRM. Odradza siê wstrzykow¹ analizê przep³ywo- Komisja Europejska [17] opublikowa³a wymagania
w¹. Pik chromatograficzny powinien byæ wiêkszy ni¿ dotycz¹ce skutecznoœci metod analitycznych i interpreta-
próg stosunku sygna³u do szumu 3:1, a czas retencji nie cji wyników. W dokumencie tym zosta³y sformu³owane
powinien siê ró¿niæ od czasu retencji wzorca wiêcej ni¿ niezbêdne kryteria dla wykrywania technik¹ spektro-
o 2% dla GC-MS lub o 5% dla LC-MS; metrii mas. Stwierdzono, ¿e rozdzia³ chromatograficzny
– kryteria dopasowania widm masowych ró¿ni¹ siê w uk³adzie bezpoœrednim (on-line) lub poœrednim
w zale¿noœci od trybu akwizycji spektrometru mas: (off-line) jest wstêpnym warunkiem dla potwierdzenia
w pe³nym i czêœciowym przemiataniu MS1 widmo technik¹ spektrometrii mas.
powinno zawieraæ co najmniej trzy strukturalnie spe- W obu technikach – GC-MS i LC-MS – minimalny do
cyficzne jony i byæ dopasowane wizualnie do widma zaakceptowania czas retencji dla badanego sk³adnika po-
wzorca. Zalecany jest dopuszczalny zakres ± 20% winien wynosiæ co najmniej dwukrotnoœæ czasu retencji
wzglêdnej intensywnoœci g³ównych jonów. Algoryt- odpowiadaj¹cego martwej objêtoœci kolumny (tj. czasu
my wyszukiwania biblioteki nie powinny byæ wyko- martwego lub Rt0). Wzglêdny czas retencji analitu (sto-
rzystywane w celu potwierdzenia to¿samoœci. Powin- sunek analit : wzorzec wewnêtrzny) nie powinien siê ró¿-
ny byæ obecne wszystkie strukturalnie specyficzne niæ od otrzymanego dla wzorca wiêcej ni¿ ± 0,5% dla
jony, o których mowa w opisie metody, a ich wzglêd- techniki GC i ± 2,5% dla LC.

Problems of Forensic Sciences 2009, vol. LXXVII, 7–28


26 M. J. Bogusz

Detekcja technik¹ spektrometrii mas powinna byæ spe³niaæ takie same kryteria, jak w trybie SRM. Limity
przeprowadzana w trybie przemiatania ca³ego zakresu tolerancji powinny byæ adekwatne do zastosowanej me-
mas, SIM oraz technikami MS-MSn, takimi jak SRM lub tody i wspierane przez bibliograficzne lub w³asne dane.
innymi technikami w po³¹czeniu z odpowiednim sposo- W innym podejœciu stosowane s¹ dwuza³o¿eniowe kry-
bami jonizacji. teria uznania danych dla co najmniej trzech stosunków
Dla pe³nego przemiatania MS obowi¹zkowa jest obec- jonów i systemu punktacji zgodnie z wymogami Komisji
noœæ w widmie odniesienia minimum czterech jonów dia- Europejskiej [17]. Tolerancje stosunków jonów ró¿ni¹
gnostycznych z wzglêdn¹ intensywnoœci¹ powy¿ej 10%. siê w zale¿noœci od intensywnoœci jonów (patrz tabela I).
Jon molekularny powinien zostaæ uwzglêdniony, je¿eli Agencja ds. Pestycydów Stanów Zjednoczonych
jego intensywnoœæ jest wiêksza ni¿ 10%. opublikowa³a standardowe procedury operacyjne (SOP)
W trybie SIM powinien zostaæ uwzglêdniony jon mo- dotycz¹ce identyfikacji, potwierdzenia i iloœciowego
lekularny. Stosunek intensywnoœci sygna³-szum dla ka¿- oznaczenia pozosta³oœci pestycydów za pomoc¹ technik
dego jonu diagnostycznego powinien wynosiæ ³ 3:1. GC i LC z detekcj¹ za pomoc¹ spektrometrii mas [58]. Te
W tabeli I zamieszczono maksymalne dopuszczalne standardowe procedury operacyjne zosta³y napisane
granice tolerancji dla wzglêdnych intensywnoœci jonów. w celu po³¹czenia w jednym dokumencie wymagañ
Do identyfikacji Komisja Europejska zdefiniowa³a sys- wszystkich procedur dotycz¹cych technik MS i MS-MS
tem punktowy. W systemie tym w celu potwierdzenia stosowanych w analizie pozosta³oœci pestycydów.
to¿samoœci s¹ wymagane co najmniej cztery punkty. W przypadku pe³nego przemiatania GC-MS widma
W tabeli II jest ukazana liczba punktów uzyskanych spe- powinny byæ rejestrowane w zakresie od 20 do 500 amu.
cyficzn¹ technik¹. Zgodnie z tymi kryteriami, potrzebne Wymagane s¹ co najmniej trzy strukturalnie specyficzne
s¹ np. cztery charakterystyczne jony w przypadku, je¿eli jony, w tym najlepiej jony molekularne spe³niaj¹ce wy-
stosowane s¹ techniki GC-MS i LC-MS lub jeden jon ma- magania stosunku sygna³u do szumu 3:1. Obwiednie izo-
cierzysty i dwa potomne w przypadku GC-MS-MS lub topowe mog¹ byæ stosowane jako jeden z trzech istot-
LC-MS-MS. Nielen i in. [39] zaproponowali do cztero- nych jonów. Wzglêdne wskaŸniki intensywnoœci ka¿-
punktowej klasyfikacji dodatkowe kryteria identyfikacji dego jonu powinny mieœciæ siê w zakresie ± 20% stosun-
w oparciu o spektrometriê mas wysokiej rozdzielczoœci ków obserwowanych dla wzorca. Dla analizy EI obo-
(tabela III). wi¹zkowe jest korzystanie z oprogramowania s³u¿¹cego
Zrzeszenie Amerykañskich Patologów (CAP) w „Spi- do przeszukiwania biblioteki. Odradza siê przeszukiwa-
sie kontrolnym chemii i toksykologii” z Laboratoryjnego nie biblioteki w technikach „miêkkiej” jonizacji, np. GC-
Programu Akredytacji sformu³owa³o wymagania istotne MS-CI. W dokumencie tym nie zosta³y sformu³owane
dla identyfikacji metodami chromatograficznymi po³¹- kryteria chromatograficzne (dok³adnoœæ czasu retencji)
czonymi ze spektrometri¹ mas [7]. W przypadku jedno- dla GC-MS.
stopniowych technik GC-MS i LC-MS identyfikacja po- W analizie GC-MS-MS czas retencji badanego
winna odbywaæ siê w oparciu o proporcje jonów z wy- zwi¹zku nie powinien ró¿niæ siê o wiêcej ni¿ ± 0,05 min
korzystaniem, o ile to mo¿liwe, co najmniej stosunku od wzorca lub ± 0,01 RRt. Nale¿y monitorowaæ dwa
dwóch jonów. Proponowany limit tolerancji dla GC-MS przejœcia z jednego lub dwóch jonów macierzystych.
wynosi ± 20% wartoœci uzyskanych dla wzorców, nato- Wzglêdne stosunki intensywnoœci ka¿dego jonu powin-
miast dla LC-MS limit ten wynosi ± 30%. Je¿eli dostêpny ny mieœciæ siê w zakresie ± 20% stosunków obserwowa-
jest tylko jeden stosunek dwóch charakterystycznych jo- nych dla wzorca, a intensywnoœæ ka¿dego jonu powinna
nów, mo¿e on byæ wystarczaj¹cy w przypadku, gdy ist- przekraczaæ stosunek sygna³u do szumu 3:1.
niej¹ inne w³aœciwoœci identyfikuj¹ce, np. czas retencji. W analizie LC-MS aparat powinien mieæ mo¿liwoœæ
Wzorzec wewnêtrzny powinien byæ identyfikowany na skanowania w zakresie od 50 do 1200 amu w trybie
podstawie co najmniej jednego stosunku jonów. W przy- pe³nego przemiatania widma. Czas retencji badanego
padku, kiedy w technice MS stosowane jest przemiatanie zwi¹zku nie mo¿e ró¿niæ siê o wiêcej ni¿ ± 0,5 min od
ca³ego zakresu mas, laboratorium powinno okreœliæ i zwa- wzorca lub ± 0,01 RRt. Wymagane s¹ co najmniej trzy
lidowaæ swoje w³asne progi „dopasowania widm” lub do- strukturalnie specyficzne jony, w tym najlepiej protono-
stosowaæ je do celów identyfikacji. wane lub deprotonowane cz¹steczki, spe³niaj¹ce waru-
W tandemowej spektrometrii mas (GC-MS-MS lub nek stosunku sygna³u do szumu 3:1. Obwiednie izo-
LC-MS-MS) pracuj¹cej w trybie SRM powinno byæ mo- topowe mog¹ byæ stosowane jako jeden z trzech istot-
nitorowane co najmniej jedno przejœcie jonów i jeden sto- nych jonów. Relatywna intensywnoœæ stosunku ka¿dego
sunek jonów wraz z czasem retencji. Jednak¿e je¿eli jonu powinna mieœciæ siê w zakresie ± 20% stosunków
istnieje du¿o jonów o wystarczaj¹cej intensywnoœci, po- obserwowanych dla wzorca. W analizie LC-MS-MS wy-
winny byæ monitorowane dwa lub wiêcej stosunki jonów. magania chromatograficzne (Rt lub RRt) s¹ takie same
Stosunki jonów ustalone podczas pe³nego przemiatania jak w przypadku LC-MS, natomiast wymagania dotycz¹-
widma s¹ akceptowaln¹ metod¹ identyfikacji i powinny ce spektrometrii mas s¹ identyczne jak dla GC-MS-MS.

Problems of Forensic Sciences 2009, vol. LXXVII, 7–28


Zapewnianie jakoœci w identyfikacji metodami chromatograficznymi... 27

Œwiatowa Agencja Antydopingowa (WADA) [66] (otrzymywanym z wzbogaconego moczu, pobranej prób-
sformu³owa³a kryteria, które musz¹ byæ spe³nione w celu ki odniesienia lub materia³u referencyjnego). Dla jonów
identyfikacji substancji sklasyfikowanych jako niedoz- diagnostycznych o wzglêdnej intensywnoœci mniejszej
wolone w moczu lub krwi badanych sportowców za po- ni¿ 5% obserwowanej w widmie referencyjnym, jony te
moc¹ procedur chromatograficznych po³¹czonych ze musz¹ byæ obecne w badanej próbce. Stê¿enia wykrytych
spektrometri¹ mas. Kryteria te s¹ podzielone na wymaga- zwi¹zków powinny byæ porównywalne z tymi w próbce
nia dotycz¹ce rozdzia³u i wykrywania w opisany ni¿ej wzorcowej.
sposób: Je¿eli nie s¹ dostêpne trzy jony diagnostyczne, nale¿y
Wymagania rozdzia³u chromatograficznego. W przypad- przygotowaæ inn¹ pochodn¹ lub zastosowaæ drug¹ tech-
ku kapilarnej GC czas retencji (Rt) analitu nie powinien nikê jonizacji lub fragmentacji opart¹ na innych zasadach
siê ró¿niæ o wiêcej ni¿ 1% lub ± 0,2 min (w zale¿noœci, fizycznych. W ka¿dym przypadku obowi¹zkowe dla ka¿-
która wartoœæ jest mniejsza) w odniesieniu do tej samej dego widma masowego s¹ minimum dwa jony diagnos-
substancji we wzbogaconej próbce moczu. Dla techniki tyczne.
HPLC, Rt analitu nie powinien ró¿niæ siê o wiêcej ni¿ 2% W technice MS-MS dane mog¹ byæ zbierane w trybie
lub ± 0,4 min (w zale¿noœci, która wartoœæ jest mniejsza) przemiatania ca³ego widma lub monitorowania wybra-
w odniesieniu do tej samej substancji we wzbogaconej nych reakcji (SRM). Jon macierzysty powinien byæ obec-
próbce moczu. Kryteria te mog¹ zostaæ z³agodzone, je¿e- ny w obu trybach. Podczas monitorowania wiêcej ni¿
li mo¿na wyjaœniæ zmiany Rt (np. przez prze³adowanie jednego jonu potomnego wzglêdne intensywnoœci które-
próbki). gokolwiek z jonów nie powinny siê ró¿niæ o wiêcej ni¿
Wymagania spektrometrii mas. Tryb pe³nego przemiata- wartoœæ podana w tabeli I w stosunku do wzglêdnych in-
nia widma: preferowanym sposobem identyfikacji jest tensywnoœci zarejestrowanych dla wzbogaconej próbki
tryb pe³nego lub czêœciowego przemiatania widma. Czê- odniesienia. Stosunek sygna³u do szumu dla najmniej in-
œciowe przemiatanie mo¿e rozpocz¹æ siê przy wartoœci tensywnego jonu powinien byæ wiêkszy ni¿ 3. Dla jonów
m/z wiêkszej ni¿ jakiekolwiek intensywne jony pocho- diagnostycznych o relatywnej intensywnoœci mniejszej
dz¹ce z odczynnika derywatyzuj¹cego lub reagenta wy- ni¿ 5% obserwowanej w widmie referencyjnym, jony te
wo³uj¹cego jonizacjê chemiczn¹. Wszystkie jony diag- musz¹ byæ obecne w badanej próbce. Tabela IV przed-
nostyczne o wzglêdnej intensywnoœci wiêkszej ni¿ 10% stawia podsumowane wymogi sformu³owane przez ró¿ne
obserwowane w widmie odniesienia uzyskanym z ma- organizacje.
teria³u referencyjnego musz¹ byæ obecne w widmie nie-
znanego piku. Wzglêdna intensywnoœæ trzech jonów
diagnostycznych nie powinna ró¿niæ siê o wiêcej ni¿ 5. Wnioski
okreœlona wartoœæ (patrz tabela I) relatywnych inten-
sywnoœci tych samych jonów obserwowanych w widmie Kryteria identyfikacji zwi¹zków przy zastosowaniu
referencyjnym (otrzymywanym z wzbogaconego moczu, chromatografii po³¹czonej ze spektrometri¹ mas z pew-
pobranej próbki odniesienia lub materia³u referencyj- noœci¹ nie s¹ wyryte w kamieniu jak dziesiêæ przykazañ.
nego). Nie jest dopuszczalne zbieranie dodatkowych jo- S¹ raczej jak ruchomy cel zmieniaj¹cy swoj¹ pozycjê
nów i u¿ywanie ich stosunków pomimo, ¿e mieszcz¹ siê w zale¿noœci od aktualnej wiedzy i ci¹g³ego ustalania no-
w zakresie zdefiniowanych tolerancji. wych wymogów, które ró¿ni¹ siê w zale¿noœci od specy-
Je¿eli stosowane jest komputerowe wyszukiwanie lub ficznej dyscypliny i koñcowego zadania analizy. Jest
dopasowanie przez biblioteki widm masowych, uzyskane zrozumia³e, ¿e najsurowsze wymogi okreœlone s¹ w tych
wyniki powinny byæ sprawdzane przez wykwalifikowa- dziedzinach, w których wyniki identyfikacji prowadz¹ do
nego specjalistê. sankcji prawnych. Z tych powodów kilka organizacji
Je¿eli nie s¹ dostêpne trzy jony diagnostyczne o wzglêd- wyda³o mniej lub bardziej szczegó³owe warunki, zalece-
nej intensywnoœci wiêkszej ni¿ 5%, nale¿y przygotowaæ nia lub wytyczne, okreœlaj¹ce minimalne parametry ko-
kolejn¹ pochodn¹ dostarczaj¹c¹ inne jony diagnostyczne nieczne do identyfikacji zwi¹zku. Dokumenty te odnosz¹
lub zastosowaæ drug¹ technikê jonizacji lub fragmentacji siê do aspektów analizy dotycz¹cych zarówno chromato-
opart¹ na innych zasadach fizycznych. W ka¿dym przy- grafii, jak i spektrometrii mas.
padku obowi¹zkowe dla ka¿dego widma masowego s¹ Istnieje powszechna zgoda w sprawie rodzaju pa-
minimum dwa jony diagnostyczne. rametrów, które maj¹ byæ kontrolowane, jak np. czas re-
W trybie SIM musz¹ byæ zarejestrowane co najmniej tencji, liczba jonów lub stosunki intensywnoœci jonów.
trzy jony diagnostyczne o stosunku sygna³u do szumu > 3 Jednak¿e dok³adne liczbowe kryteria sformu³owane
dla najmniej intensywnego jonu. Wzglêdne intensywnoœ- przez ró¿ne instytucje lub organizacje charakteryzuj¹ siê
ci jonów nie powinny ró¿niæ siê o wiêcej ni¿ okreœlone pewnymi ró¿nicami.
wartoœci (patrz tabela I) relatywnych intensywnoœci tych Zalecenia te, bêd¹ce wynikiem wspólnej wiedzy
samych jonów obserwowanych w widmie referencyjnym w danej dyscyplinie, nale¿y traktowaæ jako chwilowy,

Problems of Forensic Sciences 2009, vol. LXXVII, 7–28


28 M. J. Bogusz

zamro¿ony obraz obecnej sytuacji w dziedzinie identyfi-


kacji analitycznej. Jak napisa³ Hill: „informacje regulacy-
jne nie s¹ substytutem dla dobrej nauki, stanowi¹ one
jednak ramy, które powinny zainicjowaæ pytania: dla-
czego i w jaki sposób?” [25]. Ekspert zaanga¿owan w ba-
dania identyfikacyjne musi przede wszystkim odpowie-
dziec na pytanie: „czy zidentyfikowa³em ten zwi¹zek
zgodnie z najlepsz¹ wiedz¹?” oprócz odpowiedzi na py-
tanie: „czy spe³ni³em wszystkie formalne warunki nie-
zbêdne dla identyfikacji?”. W tym miejscu zawodowa
etyka styka siê z zawodowymi kompetencjami.

Problems of Forensic Sciences 2009, vol. LXXVII, 7–28