You are on page 1of 9

IS 1248 (Part 3) :2003

TM-Tmhm-mmwmvl
w7’13-mn mdit&Mfaa’nTaHTq

( ri%i%’1 qwt?m )

Indian Standard
DIRECT ACTING INDICATING ANALOGUE
ELECTRICAL MEASURING INSTRUMENTS
AND THEIR ACCESSORIES
PART 3 SPECIAL REQUIREMENTS FOR WATTMETERS AND VARMETERS

~ Third Revision)

ICS 17.220.20

Q BIS 2003
BUREAU OF INDIAN STANDARDS
MANAK BHAVAN, 9 BAHADUR SHAH ZAFAR MARG
NEW DELHI 110002

December 2003 Price Group 3


Measuring Equipment for Basic Electrical Quantities Sectional Committee, ET 12

FOREWORD
This Indian Standard (Part 3) (Third Revision) was adopted by the Bureau of Indian Standards, after the draft
finalized by the Measuring Equipment for Basic Electrical Quantities Sectional Committee had been approved
by the Electrotechnical Division Council.
This standard covers the special requirements for wattmeters and varmeters.
l-his standard was first published in 1958 and was revised in 1968 and 1983.
This standard is one of a series of nine Indian Standards on direct acting indicating analogue electrical measuring
instruments and their accessories. Other standards published in this series are as follows:

(Part 1) :2003 Definitions and general requirements @urth revision)


(Part 2) :2003 Special requirements for ammeters and voltmeters (third revision)
(Part 4): 2003 Special requirements for frequency meters (third revision)
(Part 5) :2003 Special requirements for phase meters, power factor meters and synchroscopes ([bird
revision)
(Part 6) :2003 Special requirements for ohmmeters (impedance meters) and conductance
meters (third revision)
(Part 7) :2003 Special requirements for multi-function instruments (third revision)
(Part 8) :2003 Special requirements for accessories (third revision)
(Part 9) :2003 Test methods (fhird revision)
In the preparation of this standard, assistance was derived from lEC 60051-3 (1984) ‘Direct acting indicating
analogue electrical measuring instruments and their accessories: Part 3 Special requirements for wattmeters and
varmeters (jourth edition)’, issued by the International Electrotechnical Commission (IEC).
For the purpose of deciding whether a particular requirement of this standard is complied with, the final value,
observed or calculated, expressing the result of a test or analysis, shall be rounded off in accordance with 1S 2: 1960
‘Rules for rounding off numerical values (revised)’. The number of significant places retained in the rounded off
value should be the same as that of the specified value in this standard.
IS 1248 (Part 3) :2003

Indian Standard
DIRECT ACTING INDICATING ANALOGUE
ELECTRICAL MEASURING INSTRUMENTS
AND THEIR ACCESSORIES
PART 3 SPECIAL REQUIREMENTS FOR WATTMETERS AND VARMETERS

( Third Revision)

1 SCOPE 0.05,0.1,0.2,0.3,0.5, I, 1.5,2,2.5,3 and 5.


1.1 This standard (Part 3) applies to direct acting Separate accuracy class to be stated on the dial if the
indicating wattmeters and varmeters having an same instrument is to be used for the measurement of
analogue display. ac and dc.
NOTE — For multi-function instruments, see 1S 1248 (Part 7),
4.3 Compliance with the Requirements of this
1.2 This part also applies to non-interchangeable
Standard
accessories [as defined in 2.1.15.3 of IS 1248
(Part l)]. The provisions of 4.3 IS 1248 (Part 1) shall be
applicable.
2 REFERENCES
2.1 The following standards are necessary adjuncts to 5 REFERENCE CONDITIONS AND INTRINSIC
this standard: ERRORS
IS No. Title
5.1 Reference Conditions
1248 Direct acting indicating analogue
electrical measuring instruments and 5.1.1 The reference values of the influence quantities
their accessories: shall be as given in Table 1 of IS 1248 (Pa~ 1) and
with Table 1.
(Part 1): 2003 Definitions and general requirements
(jourth revision) 5.1.2 The provisions of 5.1.2 of IS 1248 (Part 1) shall
(Part 7): 2003 Special requirements for multi- be applicable.
function instruments (third
5.1.3 Reference conditions different fkom those given
revision)
(Part 9): 2003 Test methods (third revision) in Table 1 of IS 1248 (Part 1) and Table 1 may be
specified, but they shall then be marked in accordance
3 DEFINITIONS with 8 of IS 1248 (Part 1).
The definitions given in 1S 1248 (Part 1) shall apply.
5.2 Limits .of Intrinsic Error; Fiducial Value
4 DESCRIPTION, CLASSIFICATION AND The provisions of 5.2 of IS 1248 (Part 1) shall be
COMPLIANCE applicable.
4.1 Description 5.2.1 Correspondence Between Intrinsic Error arui
Wattmeters and varmeters shall be described according Accuracy Class
to their method of operation as given in 2.2 of 1S 1248
The provisions of 5.2.1 of IS 1248 (Part 1) shall be
(Part 1).
applicable.
4.2 Classification
5.2.2 Fiducial Value
Wattmeters and varmeters shall be classified in one of
the accuracy classes denoted by the following class The fiducial value for a wattmeter or varmeter
indices: corresponds to the following.

1
IS 1248 (Part 3) :2003

Table 1 Reference Conditions and Tolerances, Additional to Those Given in Table 1 of


1S 1248 (Part 1) for Testing Purposes

(Clauses 5.1.1 and 5.1 .3)

SI Influence Quantity Reference Conditions Unless Tolerance Permitted for Testing


No. Otherwise Marked Purposes, Application for a Shsgle
Reference Value (see Note 1)
(1) (2) (3) (4)

i) Voltage component of measured Rated voltage or any voltage witbin the + 2 percent of the rated value
power referenee range, if any
1 1 1

ii) Current component of the Any current up to the rated mrrrent or up —


measured power to the upper limit of the reference range,

I I if any
I I
I
iii) Frequency of voltage and current Instruments using refereoce frequency *O. 1 pereent of the reference frequency
I components of the measured phase shit%ng devices
I

I
I
power

I
Other instruments 45 Hz to 65 Hz
I
* 2 percent of the referenee frequency

iv) Power factor cos $ = I or rated eos 1$(see Note 2) 0.01 lagging or leading* 0.01

v) Phase balance (for polyphase Symmetrical voltages and currents See Notes 3 and 4
instruments)

NOTES

1 This tolerance applies when a single reference value is specified in this table or is marked by the manufacturer. For a
reference range, no tolerance is allowed.

2 cos $ = 0.2 lagging for low power factor wattmeters.

sin+ for varmeters. Positive sign for lagging (inductive), negative sign for leading (capacitive).

3 The difference between any two line-to-line voltages and between ”line-to-neutral voltages shall not exceed 1 percent of the
average (lines-to-line, line-to-neutral voltages respectively).

Each of the currents in the phases shall differ by not more than 1 pereent from the average of the currents.
The angles between each of the currents and the corresponding phase-to-neutral voltages shall differ by not more than 2°.

4 Single phase testing of polyphase instruments is acceptable if permitted by the manufacturer.

5.2.2.1 The upper limit of the measuring range for the do not correspond directly to its electrical input
following: quantity.
a) Instruments with the mechanical and/or The class index is marked using Symbol E-10 given
electrical zero at one end of the scale; in Table 3 of IS 1248 (Part 1) [see 9 of IS 1248
b) Instruments with the mechanical zero outside (Part l)].
the scale irrespective of the position of the The sub-clause does not apply to a wattmeter or
electrical zero; and varmeter designed to be used in conjunction with a
c) Instruments with the electrical zero outside shunt, a series resistor (impedance) or an instrument
the scale irrespective .of the position of the transformer. These instruments are to be treated in
mechanical zero. accordance with 5.2.2.1 or 5.2.2.2 as appropriate.
The class index is marked using Symbol E-1 given in
6 NOMINAL RANGE OF U-SE AND
Table 3 of IS 1248 (Part l)[see 9 of IS 1248 (Part l)].
VARIATIONS
5.2.2.2 The sum of the electrical -values, irrespective
of sign, corresponding to the two limits of the 6.1 Nominal Range of Use
measuring range when both the mechanical and the The provisions of 6.1 of IS 1248 (Part 1) and Table 2
electrical zeros are displaced within the scale. shall be applicable.
The class index is marked using Symbol E-1 given in 6.2 Limits of Variations
Table 3 of IS 1248 (Part 1) [see8 of 1S 1248 (Part l)].
The provisions of 6.2 of IS 1248 (Part 1) and Table 2
5.2.2.3 The span for an instrument whose scale marks shall be applicable.

.
IS 1248 (Part 3) :2003

Table 2 Limits of tbe Nominal Range of Use and Permissible Variations Additional to
Those Given in Table 2 of IS 1248 (Part 1)

(Clauses 6.1 and 6.2)

Influence Quantity Limits of the Nominal Permissible Variation Expressed as a For the
Range of Use Unless Percentage of the Class Index Recommended
Otherwise Marked Tests, Ref to
Cl of IS 1248
(Part 9)

(1) (2) (3) (4)

W.tortion of ac Distortion factor Instruments using phase I 00 percent 3.7.3


;oltage or shifting devices 50/’
:urrent
Other instruments 20V0
mmponents of
he measured Peak factor 1 to 3 (see Note 2) Under eorrsideration
oower (see Note I)

Frequency of ac Instruments Reference frequency +1% I 00 pereent 3.8.1


voltage and using phase or lower I imit of refer%ce
xrrrent shitthsg devices range – I “/o and upper limit
;omponents of of reference range +10/0
the measured
Other Reference frequency +10%
power
instruments or lower limit of refetirrce
range – 100/’ and upper
limit of reference range
+ 1()%

Voltage component of the measured Reference voltage +15% 100 percent 3.9.1
power or lower limit of refirence
range – 150/0 and upper
limit of reference range
+150/0

Power factor for Class indices 0.3 Any: lagging or leading I 00 percent 3.10.1
wattmeters and smaller

Other Phase angle @ 60)’1


instruments lagging (inductive)

Power tl~ctor for Class indices 0.3 Any: lagging or leading 100 percent 3.10,2
varmeters and smaller

Class indices 0.5 Phase angle 0() . . 60” ‘j


and greater lagging (inductive)

Phase balance (for polyphase Disconnection of one 200 percent 3.12.1


instruments) current component of the
measured power

Interaction between measuring Disconnection of one 200 pereent 3.16


elements polyphase instruments voltage component of the
(see Note 4) measured power

Magnetic field of external origin 0.4 kA/m Class indices Class indices 3,5
0,3 and smaller 0.5 and greater

Electrodynamics 3 pereent of the 6 percent of


instruments if fiducial value the fiducial
not astatic and I (see Note 5) value (see
or not having a Note 5)
magnetic screen

Ferrodynamic 1.5 pereent of 3 pereent of


instruments if the fiducial the fiducial
not astatic and I value value (see
or not having (see Note 5) Note 5)
a magnetic
screen

All other 0.75 percent of 1.5 percent of


instruments the fiducial the fiducial
value value
(see Note 5) (see Note 5)

3
IS 1248 (Part 3) :2003

Table 2 (Concluded)

NOrES

1 For instruments having electronic devices in their measuring circuits.

2 The permissible variation due to a peak factor of other than W (corresponding to a sine wave)is included in the permissible variation due
to distortion of the measured power. For instruments having a peak factor capability greater than 3, the manufacturer shall state:

a) The instrument peak factor capability producing a variation of 100 percent of the class index.

b) The upper and lower limits of the frequency response (bandwidth) to 0.707 times the indication at the reference frequency.

c) The et~ective maximum rate of change of internal instrument ac amplitier response (slew rate), expressed in volts per second
using appropriate S.1. prefixes.

Peak factor relates to the total peak factor capability of the instrument and includes both the peak factor due to a distorted waveform and the
peak Factor due to spurious impulses (which maybe random or harmonically related to the fundamental frequency) containing negligible
average power.

3 Inductive power factor unless otherwise agreed between the manufacturer and the user.
4 It may sometimes be impossible to carry out test for interaction to between the measuring elements due to interconnection of the current
and/or voltage circuits (for example 272 element wattmeters and meters utilizing electronic devices to obtain their voltage from voltage
measuring circuits.
5 Not as a percentage of the class index.

.6.3 Conditions for the Determination of Variations be open-circuit.


The provisions of 6.3 of IS 1248 (Part l) and Table 2 For the recommended tests, see 5.8 of IS 1248
shall be applicable. (Part 9).

6.4 Determination of Variation Due lo Power Factor 7.2 Damping


The power factor shall be lagging for instruments of The provisions of 7.2 of IS 1248 (Part 1) shall
class indices 0.5 and greater. For instruments of class apply.
0.3 and smaller, the test shall be carried out using both
lagging and leading power factors. 7.2.1 Overshoot
The provisions of 7.2.1 of IS 1248 (Part 1) shall apply.
6.5 Special Tests for Variations
Special tests using combinations of the components 7.2.2 Response Time
of the measured quantity may be carried out when The provisions of 7.2.2 of IS 1248 (Part 1) shall apply.
considered necessary by agreement between the
manufacturer and the user. However, the requirements of 7.2..1 and 7.2.2 of
IS 1248 (Part 1) do not apply to the following types of
7 FURTHER ELECTRICAL AND MECHANICAL wattmeters and varmeters:
REQUIREMENTS
a) thermal instruments,
7.1 Voltage Tests, Insulation Test and Other Safety b) instruments having a freely suspended
Requirements moving element,
For a fixed wattmeter or varmeter having the upper c) instruments having a material pointer longer
limit of its measuring range of 1A to 10A and intended than 150 mm,
for use with a current transformer having a high over- d) instruments in which the power (active or
current capability, the current circuit shall not open reactive) corresponding to the upper limit of
when the ammeter is subjected to 30 times the nominal the measuring range is less than 10W or
secondary current of the associated current transformer 10 VAR, and
for a period of 2s. e) special-purpose instruments where other
response times may be required. Such
A portable wattmeter or varmeter intended for similar
instruments will be the subject of-agreement
use shall withstand 15 times the upper limit of its
between the manufacturer and the user.
measuring range for a period of 2s.
These wattmeters need not be functional after 7.2.3 Impedance of the External Measuring Circuit
application of this overload but shall not then The provisions of 7.2.3 oHS 1248 (Part 1) shall apply.

4
IS 1248 (Part 3) :2003

7.3 Self-Heating c) Instruments having a freely suspended


moving element.
The provisions of 7.3 of IS 1248 (Part I) shall apply,
7.4.2.1 The values of current and voltage for the
7.4 Permissible Overloads overloads of short duration shall be the product of the
relevant factor given in Table 3 and the value .of the
7.41 Continuous Overload upper limit of the electrical input quantity unless other
For the recommended test, see 5.6 of 1S 1248 (Part 9). values are stated by the manufacturer. The power factor
(reactive power factor) shall beat its reference value.
Wattmeters and varmeters, together with their non-
interchangeable accessory (ies), if any, except for 7.4.2.2 The full duration of each overload shall be
instruments fitted with a non-locking switch, shall be applied except when an automatic cut-out (fuse) fitted
subjected to a continuous overload of 120 percent of to the instrument has interrupted the circuit in less than
the rated values of current and voltage sequentially, the time specified in Table 3.
the other value being maintained at its rated value. The
The automatic cut-out shall be reset (or the fuse
duration of each application shall be 2h.
replaced) before the application of.the next overload.
After having cooled to the reference temperature, the
7.4.2.3 After having been subjected to the overloads
instrument together with its non-interchangeable
of short duration and after having cooled to the
accessory(ies), if any, shall comply with its accuracy
reference temperature, wattmeters and varmeters
requirements; however, the overload shall not be
repeated. whose mechanical zero is within the scale, together
with their non-interchangeable accessory(ies), if any,
The continuous overload test shall be carried out under shall comply with both of the “following requirements:
reference conditions except for current and voltage.
The power factor for wattmeters shall be cos $ = 1 and a) The deviation of the index from.the zero scale
the power factor for varmeters shall be sin $ = 1. mark, expressed as a percentage of the scale
length, shall not exceed the following value:
7.4.2 Overloads of Short Duration
1) 0.5 for instruments of class indices 0.3
For the recommended test, see 5.4 of IS 1248 (Part 9). and smaller, and
Wattmeters and varmeters, together with the non- 2) the class index for instruments of class
interchangeable accessory (ies), if any, shall be indices 0.5 and greater.
subjected to overloads of short duration. b) The wattmeter or varmeter together with its
non-interchangeable accessory(ies), if any,
However, these requirements do not apply to:
after adjustment of the zero (if necessary)
a) Thermocouple instruments, shall comply with the accuracy requirements;
b) Electrostatic instruments, and however, the overloads shall not be repeated.

Table 3 Overloads of Short Duration


(Clause 7.4.2.1)

Current Factor Voltage Factor Number of Duration of Each Interval Between


Overloads Overload(s) Successive Overload(s)
(1) (2) (3) (4) (5)
Class indices 0.5 and smaller

1 2 I 5 —

2 1 5 0.5 15

Class indices 1 and greater

10 1 9 0.5 60

10 1 I 5 —

1 2 I 5 —

NOTE — When two or three series of tests are specified, all the tests should be carried out in the order given. The overloads of short
duration are applied simultaneously to all the measuring elements of polyphase wattmeters and varmeters.

5
IS 1248 (Part 3) :2003

A wattmeter orvarmeter whose mechanical 8.5 Adjuster(s), Mechanical and/or Electrical


zero is outside the scale is considered to have
The provisions of 8.4 of 1S 1248 (Part 1) shall be
complied with this requirement if, after applicable.
having cooled to the reference temperature,
it has errors not exceeding those relating to 8.6 Effects of Vibration and Shock
its class index; however, the overloads shall
The provisions of 8.5 of 1S 1248 (Part 1) shall be
not be repeated.
applicable.
7.5 Limiting Values of Temperature
9 INFORMATION, GENERAL MARKINGS AND
‘The provisions of 7.5 of IS 1248 (Part i ) shall be SYMBOLS
applicable.
The provisions of 9 of IS 1248 (Part 1) shall be
7.6 Deviation from Zero applicable.

For the test method, see 5.9 of IS 1248 (Part 9). 10 MARKINGS AND SYMBOLS FOR
TERMINALS
If a wattmeter or varmeter has a zero position marked
on the scale, it shal I be tested for return to zero when 10.1 The provisions of 10.1 of IS 1248 (Part 1) shall
de-energised. The test shall be carried out under be applicable.
reference conditions.
10.2 The provisions of 10.2 of IS 1248 (Part 1) shall
7.6.1 A II Circuits Energized be applicable.

After a period of energization of 30s at the upper limit 10.3 The provisions of 10.3 of IS 1248 (Part 1) shall
of the measuring range, the deviation of the index from be applicable.
the zero scale mark, expressed as a percentage of the
10.4 Special Markings for Terminals
scale length, shall not exceed a value corresponding
to 50 percent of the class index. All terminals shall be marked so that they can be
uniquely identified.
7.6.2 Voltage Circuit(s) (My Energized
With the voltage circuit(s) only energized, the deviation
10.4.1 Single Element Instruments
of the index from the zero scale mark shall not exceed Wattmeters and varmeters having only two current
a value corresponding to 100 percent of the class index. terminals and two voltage terminals shall have the
current voItage terminals easily distinguishable. The
8 CONSTRUCTIONAL REQUIREMENTS current terminal which is normally associated with a
8.1 The provisions of 8.1 of IS 1248 (Part 1) shall be particular voltage terminal shall be identified by both
applicable. of them being marked with a common sign.

8.2 The provisions of 8.2 of IS 1248 (Part 1) shall be 10.4.2 Polyphase Instruments
applicable. For all polyphase wattmeters and varmeters, a
8.3 The provisions of 8.3 of IS 1248 (Part I) shall be connection diagram shall be provided, preferably
applicable. affixed to the case.
The identification of a terminal on the instrument and
8.4 Preferred Values
on the connection diagram shall be identical.
The upper limit of the measuring range for ammeters
The connection diagram shall show the intended
and voltmeters shall preferably be one of the following interconnection of the elements of the instrument with
values: the external circuit.
1, 1.2, 1.5, 2, 2.5, 3, 4, 5, 6, 7.5, and 8
11 TESTS TO PROVE COMPLIANCE WITH
or their decimal multiples and sub-multiples. THIS STANDARD

For multi-range instruments, at least one of the ranges The provisions of 11 of 1S 1248 (Part 1) shall be
shall preferably comply with this requirement. applicable.

6
Bureau of Indian Standards

BIS is a statutory institution established under the Bureau of Zndian Standards Act, 1986 to promote
harmonious development of the activities of standardization, marking and quality certification of goods
and attending to connected matters in the country.

Copyright

BIS has the copyright of all its publications. No part of these publications may be reproduced in any form
without the prior permission in writing of BIS. This does not preclude the free use, in the course of
implementing the standard, of necessary details, such as symbols and sizes, type or grade designations.
Enquiries relating to copyright be addressed to the Director (Publications), BIS.

Review of Indian Standards

Amendments are issued to standards as the need arises on the basis of comments. Standards are also reviewed
periodically; a standard along with amendments is reaffirmed when such review indicates that no changes are
needed; if the review indicates that changes are needed, it is taken up for revision. Users of .Indian Standards
should ascertain that they are in possession of the latest amendments or edition by referring to the latest issue of
‘BIS Catalogue’ and ‘Standards: Monthly Additions’.

This Indian Standard has been developed from Doc : No. ET 12 (5057).

Amendments Issued Since Publication

Amend No. Date of Issue Text Affected

BUREAU OF INDIAN STANDARDS


Headquarters :
ManakBhavan,913ahadur Shah Zafar Marg, New Delhi 110002 Telegrams : Manaksanstha
Telephones :23230131,23233375,2323 9402 (Common to all oftices)

Regional Offices : Telephone


Central : Manak Bhavan, 9 Bahadur Shah Zafar Marg 23237617
NEW DELHI 110002 { 23233841
Eastern : 1/14 C.I.T. Scheme VII M, V. I. P. Road, Kankurgachi 23378499,23378561
KOLKATA 700054 { 23378626,23379120
Northern : SCO 335-336, Sector 34-A, CHANDIGARH “160022 603843
{ 609285
Southern : C.I.T. Campus, IV Cross Road, CHENNAI 600113 22541216,22541442
{ 22542519,22542315
Western : Manakalaya, E9 MIDC, Marol, Andheri (East)
. . f2832 9295.28327858
MUMBAI 400093 1 2832 ?891 :28327892
Branches : AHMEDABAD. BANGALORE. BHOPAL. BHUBANESHWAR. COIMBATORE. FARIDABAE+
GHAZIABAD. GUWAHATI. HYDERABAD. JAIPUR. KANPUR. LUCKNOW. NAGPUR.
NALAGARH. PATNA. PUNE. RAJKOT. THIRWANANTHAPURAM. VISAKHAPATNAM.

Printed at Prabhat Otlset Press, New Delhi-2

You might also like