Professional Documents
Culture Documents
87
Report Committee
J. M. Boone (Chair), University of California Davis, Sacramento, California, USA
J. A. Brink, Massachusetts General Hospital, Boston, Massachusetts, USA
S. Edyvean, Public Health England, Oxfordshire, UK
W. Huda, Medical University of South Carolina, Charleston, South Carolina, USA
W. Leitz, Swedish Radiation Protection Authority, Stockholm, Sweden
C. H. McCollough, Mayo Clinic, Rochester, Minnesota, USA
M. F. McNitt-Gray, University of California Los Angeles, Los Angeles, California, USA
ICRU Sponsors
P. Dawson, UCL Hospitals, London, UK
P. L. M. DeLuca, University of Wisconsin, Madison, Wisconsin, USA
S. M. Seltzer, National Institute of Standards and Technology, Gaithersburg, Maryland, USA
The Commission wishes to express its appreciation to the individuals involved in the preparation of this
Report for the time and efforts that they devoted to this task and to express its appreciation to the
organizations with which they are affiliated.
All rights reserved. No part of this book may be reproduced, stored in retrieval systems or transmitted in
any form by any means, electronic, electrostatic, magnetic, mechanical photocopying, recording or
otherwise, without the permission in writing from the publishers.
British Library Cataloguing in Publication Data. A Catalogue record of this book is available at the British
Library.
Journal of the ICRU Vol 12 No 1 (2012) Report 87 doi:10.1093/jicru/ndt006
Oxford University Press
the development of guidance and recommendations The Commission has found its relationship with
can make an important contribution. all of these organizations fruitful and of substantial
benefit to the ICRU program.
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Abstract . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.1 Introduction . . . . . . . . . . . . . . . . . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. . 25
3.2 CT in Neuroradiology . . . . . . . . . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. . 25
3.3 CT in Thoracic Radiology . . . . . . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. . 28
3.4 CT in Abdominal Imaging . . . . . . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. . 29
3.5 CT in Pediatric Radiology. . . . . . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. . 31
3.6 CT in Clinical Use: Summary . . . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. . 33
5. CT X-Ray-Spectrum Characterization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. .. . .. . 47
5.2 Methods for HVL Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. .. . .. . 48
5.2.1 Conventional HVL Assessment in CT . . . . . . . . . . . . . . . . . . . . . . . . . . .. .. . .. . 48
5.2.2 Aluminum-Cylinder Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. .. . .. . 48
5.2.3 Real-Time Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. .. . .. . 48
5.2.4 Aluminum Cage with Real-Time Probe Method. . . . . . . . . . . . . . . . . . .. .. . .. . 50
5.3 Spectrum Assessment Using the Tube Potential and the HVL . . . . . . . . . . . . .. .. . .. . 50
5.3.1 Off-Angle HVL Assessment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. .. . .. . 52
5.4 Typical HVL Values in CT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. .. . .. . 52
6.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
6.2 Theoretical Methods for Predicting fair(z) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
6.3 Measurement of fair(z) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
6.3.1 Optically Stimulated Luminescence Systems. . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
6.3.2 TLD Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
6.3.3 Film . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
6.3.4 Computed-Radiography Detectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
6.3.5 Real-Time Radiation Detectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
6.3.6 Summary of fair(z) Measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
6.4 Measurement of fL(x) or fA(u) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
6.4.1 Measurement of fL(x) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
6.4.2 Measurement of fA(u) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
6.5 Planar Measurements of the CT Beam Profile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
6.6 Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
7. CT Dosimetry in Phantoms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
8.1 Introduction . . . . . . . . . . . . . . . . . . . .. .. . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. . 89
8.2 Absorbed Dose versus Patient Size .. .. . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. . 90
8.3 Size Metrics . . . . . . . . . . . . . . . . . . . . .. .. . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. . 91
8.4 Size-Specific Dose Estimates . . . . . .. .. . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. . 95
8.5 Summary . . . . . . . . . . . . . . . . . . . . . . .. .. . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. .. .. . .. . 97
References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
Journal of the ICRU Vol 12 No 1 (2012) Report 87 doi:10.1093/jicru/ndt007
Oxford University Press
Preface
For nearly three decades, the ICRU has devel- There have been a number of dosimetric quan-
oped authoritative Reports dealing directly with tities used for CT since the early 1980s, and these
radiological imaging. These include: Report 41, have undergone incremental changes over the past
Modulation Transfer Function of Screen-Film 3 decades. It has been recognized that in light of
Systems (1986); Report 54, Medical Imaging – The the large population doses associated with CT, and
Assessment of Image Quality (1995); Report 70, given the improved features of modern scanners,
Image Quality in Chest Radiography (2003); Report that current radiation-dosimetry methods are out
74, Patient Dosimetry of X Rays Used in Medical of date. Therefore, there is a need to make more ac-
Imaging (2005); Report 79, Receiver Operating curate measurements of more relevant parameters,
Characteristic Analysis in Medical Imaging (2008); including those that take into consideration scan-
and Report 82, Mammography – Assessment of specific parameters such as patient size and the
Image Quality (2009). This Report on computed length of the scan.
tomography (CT) continues that series and is One of the principal driving forces for the
intended to advance dosimetry and image-quality increased use of CT for clinical diagnosis is the
evaluation in this important application. phenomenal improvement in image quality that
Computed tomography has experienced rapid has occurred over the past 15 years. In addition to
growth in technological sophistication, and with increased spatial resolution, improvements in
these advancements there has been a commensur- detectors, x-ray tubes, and reconstruction algo-
ate increase in the types of clinical questions that rithms have led to significant improvements in the
can be addressed. In addition to profound improve- signal-to-noise properties in CT images. These
ments in image quality, the acquisition times for improvements challenge traditional measures of
routine CT examinations have dropped from several image quality. This Report seeks to amend,
minutes in the early era to a few tens of seconds improve, and update the methods for both dosim-
today, opening up new clinical uses, including pedi- etry and image-quality evaluation in CT. After
atric, cardiac, and thoracic imaging, for which reviewing current CT-dose metrics, a number of
organ or patient motion preclude the use of other updated measurement procedures are recom-
imaging modalities. The improvement in the diag- mended that capitalize on faster systems and on
nostic information that CT provides has led to a new phantom designs that allow more accurate as-
large increase in utilization, with well over 100 sessment of dose in patients. Measurement proce-
million studies performed worldwide annually. dures are introduced that allow rapid measurement
CT scans involve radiation exposure, and the of the x-ray beam in terms of both its quality and
radiation dose levels in most cases are higher the spatial distribution of the air kerma. Updated
than with other radiographic examinations. The methods for characterizing image quality, including
higher dose levels, coupled with the very large both spatial resolution and noise performance are
number of CT scans performed annually, has led recommended. Overall, this Report capitalizes on
to concerns about the associated radiation risks. recent developments in CT metrology combined with
The National Council on Radiation Protection and new measurement technology, with the intent of
Measurements has reported that 48 % of the providing more accurate characterization of the dose
average total dose to citizens in the U.S. is from and image-quality from modern, high-performance
medical-imaging procedures, and of this about CT systems.
49 % is due to CT. Thus, approximately 24 % of the
radiation burden to the U.S. population is from CT John M. Boone
examinations. Stephen M. Seltzer
Glossary
Artifact the appearance in the CT image of details not present in the scanned object.
Acquisition process of acquiring raw data by measuring the transmission of the x-ray
beam through an object.
Acquisition channel one or more electronically coupled detector arrays yielding one transmission
profile.
Attenuation reduction in the number of photons in an x-ray beam upon passage through
matter.
Automatic exposure control automatic adjustment of the tube current, depending on the attenuation of
the patient, to achieve a pre-determined level of image quality.
Axial CT the standard step-and-shoot acquisition mode of CT scanners. Also called se-
quential by some manufacturers.
Axis of rotation axis about which the x-ray tube rotates, coincident with the isocenter.
Back projection a mathematical procedure for CT reconstruction, based on projecting
the detected x-ray signal back along the measurement trajectory between
detector element and focal spot.
Beam filtration material, usually metal or plastic, placed in the x-ray-tube port that filters
the x-ray beam and preferentially removes lower-energy photons, sparing
some dose to the patient.
Beam hardening the result of filtration of a polychromatic beam by the absorption of lower-
energy photons in a patient or material, with a subsequent increase in
effective energy. Beam hardening can cause artifacts in CT images, so the
x-ray spectrum is pre-hardened in CT to reduce beam-hardening artifacts.
Beam-shaping filter same as bow-tie filter.
Bow-tie filter a metal filter that is placed in the x-ray-tube assembly, which is thicker at
the periphery and tapers gradually to a thin filter at the center of the beam.
The bow-tie filter reduces x-ray intensity incident peripherally upon the
patient, and compensates for the typically round contours of the patient’s
body to deliver a more homogeneous beam intensity to the detector arrays.
Breast computed tomography a method of examining the breast utilizing a specially designed CT scanner,
where the breast hangs pendant into the field of view with the woman lying
prone on the table.
Collimation geometrical limitation of the extent of the radiation beam in the z direction.
Computed-tomography dose integral along a line parallel to the axis of rotation (z) of the air-kerma
index (CTDI) profile;
ð
1 1
CTDI1 ¼ KðzÞdz;
nT 1
ð þ50 mm
x 1
CTDI100 ¼ KðzÞ dz;
nT 50 mm
4
Glossary
Detector efficiency the fraction of x-ray energy incident upon a detector that contributes to the
electronic signal.
Detector element (del) a single element of a detector array.
Detector row a row of detector elements, including their interspace material, arranged
along an arc centered on the axis of rotation; same as detector array.
Dose – length product (DLP) a parameter used as a surrogate measure for energy imparted to the patient
in a CT scan of length L:
DLP ¼ CTDIvol L:
5
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
b
p¼ :
nT
6
Glossary
Post-patient collimation geometrical limitation of the extent of the radiation beam in the z direction
by a slit device positioned between the patient and the detector.
Infrequently used for thin-section imaging.
Profile a profile is a collection of rays acquired at a specific gantry angle using the
detector array. The term profile is synonymous with projection or view.
Projection same as profile.
Quantitative computed the use of CT images and the corresponding HU values for quantitative
tomography (QCT) characterization of organs or tissues. QCT is most widely used in the deter-
mination of bone-mineral content, lung density, and treatment planning in
radiotherapy.
Ray the schematization of a narrow beam of x rays from the x-ray-tube focal spot
to a single detector element, giving rise to a detector-element reading. Each
view or projection is composed of numerous rays.
Reconstruction algorithm precisely defined computational procedure to produce CT images from
the acquired projection data. Filtered back projection has been used for most
image reconstruction in CT; however, in recent years, iterative-
reconstruction algorithms have been used as well.
Reconstruction filter for filtered back projection, the reconstruction filter is used to mathematical-
ly filter the projection data prior to back projection. Typically, the reconstruc-
tion filter is described in the frequency domain and consists of a ramp
component combined with an apodization filter to reduce the impact of
quantum noise at higher spatial frequencies. Reconstruction filters have
names that vary among CT-scanner manufacturers.
Region of interest (ROI) a region that is part of an image. The ROI is typically square or rectangular,
but can be of any arbitrary shape.
Rotation period time duration of a single 3608 rotation of the x-ray tube and detector array
(in third-generation CT geometry). Typical rotation periods for modern CT
scanners range from 0.25 s to 1.0 s.
Scan range the length of the body that is visualized, measured along the z axis for one
volume acquisition or for a series in sequential (axial) acquisition
(step-and-shoot mode).
Scan time the time interval between the beginning and the end of the acquisition of
data, i.e., time for either a complete spiral acquisition or for a single tube ro-
tation in the step-and-shoot mode.
Scattered radiation x rays whose trajectories have been redirected due to a scattering event in
the patient or CT hardware. Generally, scattered radiation can deposit
energy inside the patient but outside of the primary x-ray beam, can be
emitted into the CT scanner room, and can reach detector elements and
cause artifacts or increased noise in the reconstructed CT images.
Section in multi-slice scanning, the tomographic volume corresponding to a single
image in the reconstructed CT data set. Sections can be of different thick-
ness, usually ranging from 0.5 mm to 5 mm, and sometimes to 10 mm. CT
sections are also called CT slices.
Section thickness the thickness of a CT image, measured at the scanner isocenter.
Thicknesses range from 0.5 mm to 10 mm on most CT systems. Also called
slice thickness.
Sequential (axial) CT CT-scanning technique in which images are acquired at fixed z positions
interspersed by translation of the patient in the z direction. During transla-
tion, the x-ray beam is turned off. Sequential or axial CT imaging is the
basic step-and-shoot acquisition method of CT scanners.
7
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
CTDIw
CTDIvol ¼ :
p
center peri
where CTDI100 or CTDI100 refer to measurements of CTDI100 at the center
(center) or periphery ( peri) of the head or body phantom.
z axis axis parallel to the axis of rotation.
8
Abstract
Computed tomography has experienced a number of significant technological advances over the past
decade, and these have had pronounced impacts on the accuracy of radiation dosimetry and the assessment
of image quality. After reviewing CT technology and clinical applications, this Report describes and dis-
cusses existing dosimetry methods and then presents new methods for radiation dosimetry, including the
evaluation of beam quality, and measurement of CT-scanner output in air and in phantoms. Many of the
proposed dosemetric quantities can be measured quickly using a real-time ionization chamber, which is
introduced here. Traditional measurements of image quality for computed tomography rely upon simple and
subjective observations. A more rigorous approach is proposed, including routine use of the modulation-
transfer function for describing spatial resolution along all axes, and of the noise-power spectrum for des-
cribing the noise amplitude and texture properties of CT images. This Report focuses on new but practical
methods for the assessment of radiation dose and image quality for CT scanners.
9
Journal of the ICRU Vol 12 No 1 (2012) Report 87 doi:10.1093/jicru/nds001
Oxford University Press
1. Introduction
This Report centers on computed tomography mainly addresses measurement of the spectral prop-
(CT) scanners, which are used widely in diagnostic erties of CT scanners, including use of the new tech-
medicine around the world. The Report focuses on nology of real-time ionization-chamber systems.
dose assessment, as well as the characterization of Although the measurement of the half-value layer
image quality in CT. (HVL) in CT is generally performed infrequently,
A description of the basic technology underlying x-ray beam quality is an important property that
CT is provided in Section 2. The early systems, relates to both image quality and absorbed dose in
which used simple acquisition geometries requiring the patient. Proposed methods for HVL assessment
several seconds per slice, have evolved into modern considerably streamline previous approaches.
axial (sequential) and spiral (helical) acquisition Section 6 discusses the assessment of the x-ray
platforms that can acquire a large number of slices output (air kerma or air-kerma rate) of a CT
in one-third of a second or less. A number of differ- scanner as measured free-in-air in the absence of a
ent dose-reduction features have been added to CT phantom. Although a dosimetry phantom is an im-
scanners in recent years, and these are also dis- portant tool in the assessment of dose in CT, the
cussed in Section 2. x-ray output (air kerma) is a fundamental quantity
Section 3 describes the clinical utility of CT, that impacts the dose in the patient during CT. In
described by radiologists specializing in neurora- this Section, it is recognized that measuring the
diology, body CT, thoracic imaging, and pediatric output characteristics of the CT scanner in the
radiology. This Section introduces the reader to a absence of a phantom can be a practical method for
sampling of clinical applications of CT, demonstrat- routine quality assurance. This Section focuses on
ing why this modality is used so widely and so fre- the dose distributions in both the z dimension and
quently in diagnostic medicine. along the axial or x – y dimensions of the CT system.
Absorbed-dose levels have been an emerging The real-time dosimeter has an important role to
issue in regard to CT for the past decade or more. play in these measurements.
Sections 4 through 9 of the Report focus on various Section 7 introduces a newly designed phantom
aspects of radiation dose and dose assessment, that is long enough (600 mm) to capture the major-
recognizing recent advancements in CT scanner ity of the absorbed dose due to x-ray scatter that
design, new perspectives on the radiation dose dis- occurs in a phantom. This Section is harmonized
tribution in CT, and the important issue of patient with AAPM Report 111 (AAPM, 2010), and the
size and length of the CT scan. phantom was designed in close collaboration with
Section 4 describes the conventional measure- AAPM Task Group 200. With a phantom and real-
ments of CT-radiation dose, and is intended to fa- time dosimeter, an asymptotic curve that represents
miliarize the reader with traditional methods of the air kerma or absorbed dose at the center plane
radiation-dose assessment in CT. This Section high- of the CT scan can be determined as a function of
lights the technological baseline, and provides a the scan length in a single measurement. This
context from which the many CT-dosimetry recom- “rise-to-equilibrium” curve can be used to describe,
mendations contained in this Report can be in general, the dose distribution in CT under the
understood. simplified conditions of a homogeneous phantom.
This Report also introduces the use of a real-time In all x-ray imaging modalities, including CT, the
dosimeter for evaluation of the air-kerma rate. The size of the patient plays an important role in x-ray
use of a real-time dosimeter provides capabilities in dosimetry. Section 8 discusses methods by which
CT dosimetry that cannot be achieved using patient size, defined as the water-equivalent diam-
integration-mode radiation meters, and simplifies eter, can be used to adjust the value of CTDIvol
several of the proposed measurements. These new reported by the CT scanner to produce a more accur-
dosimeters are first discussed in Section 5. Section 5 ate dose estimate. A quantity is introduced in this
Section called the size-specific dose estimate. This and x-ray-tube current, the reconstruction kernel
Section is well aligned with AAPM Report 204 used, and other technical parameters are discussed.
(AAPM, 2011a), which has been endorsed by the The assessment of both spatial resolution and con-
AAPM, the ICRU, and the Image Gently campaign. trast resolution is described in some cases using the
Modern CT systems dynamically adjust the x-ray- CT phantom recommended in Section 7. The Report
tube current depending upon the dimensions of the concludes with Section 12, which makes specific
patient along the length of the scan, in a procedure recommendations with respect to the characteriza-
called automatic exposure control or x-ray-tube tion of both dose and image quality in CT.
current modulation. In order to compute dose when
x-ray-tube-current modulation is used, an image- 1.1 CT Manufacturers and Model Names
by-image assessment of both patient size and tube
There are only a few large manufacturers of CT
current becomes necessary to achieve good accuracy.
scanners, and there are some differences in the fun-
Section 9 describes the algorithms necessary for
damental technology offered by these companies. In
this, together with a computer program developed
some cases, the makes and models available differ,
to illustrate the trends in dose estimates as a func-
depending on country of placement. In this Report
tion of patient size and CT technical parameters.
when discussing CT characteristics, in many cases,
Image quality is an important characteristic of
the type of scanner (manufacturer and model) is
any imaging system, and is related to the associated
mentioned. In doing so, there is no intent for en-
absorbed-dose levels. In Section 10, characterization
dorsement, but rather to allow readers to under-
of the spatial resolution of CT scanners is discussed.
stand the type of scanner that produced a specific
Historical methods have utilized the point-spread
set of data. In general, this knowledge will be more
function, line-spread function, and edge-spread
useful to readers who work in the CT field, and who
function to compute the modulation transfer func-
are familiar with the subtle differences in the cap-
tion (MTF). This Section also discusses the use of
abilities of various CT-scanner types. The manufac-
the MTF to characterize the spatial resolution in
turers of the CT scanners mentioned in this report
the z dimension, moving away from the historical
are listed here alphabetically:
use of the slice-sensitivity profile.
Section 11 addresses the measurement of noise General Electric Medical Systems, Waukesha,
texture and amplitude in CT images. The 3D noise- Wisconsin.
power spectrum (NPS) is proposed to characterize Philips Healthcare, DA Best, The Netherlands.
the noise properties of CT images. The dependen- Siemens Medical Systems, Erlangen, Germany.
cies of the NPS on CT parameters such as x-ray- Toshiba Medical Systems, Irvine, California, USA.
tube potential, the product of gantry rotation time
12
Journal of the ICRU Vol 12 No 1 (2012) Report 87 doi:10.1093/jicru/nds002
Oxford University Press
For the first 75 years of x-ray imaging, the detec- Riddle, 1967). In the early 1960s, several independ-
tors used in diagnostic radiology, such as radio- ent investigations into the medical applications
graphic film or image intensifiers, provided of image-reconstruction techniques were carried
reasonably good visualization of high-contrast out (Cormack, 1963; Kuhl and Edwards, 1968;
objects. However, their ability to record small dif- Oldendorf, 1961). However, it was not until the late
ferences in transmitted x-ray signals was limited. 1960s and the work of Hounsfield that the first
Several factors contributed to the inability to clinically useful x-ray CT scanner was developed
resolve low-contrast signals. (Hounsfield, 1973).
First, large-area detectors record a large amount Despite its relatively low spatial resolution,1 CT
of scattered radiation, making small differences in was rapidly accepted into clinical practice because
x-ray transmission difficult to resolve. Second, the of its tomographic nature and superior contrast
superposition of the patient’s three-dimensional in- resolution (0.5 %) when compared with existing
formation onto a two-dimensional detector obscures radiographic images (approximately 5 %). In 1979,
low-contrast information. Geometrical tomography the Nobel Prize in Medicine was awarded to
provided some separation of overlapping structures, Hounsfield and Cormack for their contributions to
but its utility was limited to high-contrast struc- the development of x-ray CT (Hounsfield, 1980).
tures such as bones, airways, and iodinated
vessels. Finally, radiographic film, intensifying
(scintillation) screens, and image intensifiers can
exhibit non-linearities and/or non-uniformities that 2.1 CT-Scanner Design
are large compared with the small differences in
CT-scanner hardware is designed to determine
x-ray transmission of low-contrast objects.
effective x-ray attenuation coefficients at each point
Introduced clinically in the early 1970s, x-ray
within a volume of interest from transmission mea-
computed tomography (CT) overcame many of the
surements acquired at multiple angles through the
difficulties encountered in using large-area detec-
object. A set of transmission measurements
tors. First, the sequential irradiation of slabs of
through the object at a given angle is known as a
tissues and collimation at the detector markedly
projection, or view. These projection measurements
reduced the amount of scattered radiation mea-
are mathematically combined to form a two-
sured. Second, the reconstruction of a tomographic
dimensional representation of a three-dimensional
image eliminated much of the problem of overlap-
object. Figure 2.2 illustrates a single projection,
ping anatomy. X-ray CT was the first imaging mo-
which contains multiple rays (representing line
dality that allowed physicians to see the internal
integrals) through the patient, acquired as the
structure of a three-dimensional object in cross-
x-ray tube rotates around the patient.
section (see Figure 2.1). With CT, the use of high-
The x-ray tube is used to irradiate the patient
efficiency, low-noise detectors that could respond
with a diverging beam of x-rays, and transmission
linearly over a wide range of transmission values
measurements are acquired using a detector array
provided uniform sensitivity over a wide dynamic
on the opposite side of the patient. At each angular
range, producing excellent CT images across a wide
position, approximately 600 –900 transmission
variety of clinical applications.
measurements (rays) are acquired (with each de-
The Radon transform introduced the concept of
tector array), with a spacing of approximately
mathematically reconstructing the internal struc-
1 mm. The x-ray tube and detector array(s) rotate
ture of an object from multiple projections through
continuously, and the data-collection process is
the object (Radon, 1917). The techniques of image
reconstruction were pursued in a variety of disci- 1
Current spatial resolution limits are 1.0 cycle/mm for CT
plines including Bracewell’s work in the 1950s in compared with 3.5 cycle/mm for digital radiography and 7.0
radio astronomy (Bracewell, 1956; Bracewell and cycles/mm for digital mammography.
Figure 2.1. Sample CT images. High-spatial-resolution images can be generated in all the axial, coronal, and sagittal planes from a
single data acquisition.
mref ¼ I0 : ð2:1bÞ
14
Basics of Computed-Tomography Technology
the Hounsfield Unit,2 HU, is the name given to now in clinical use. Some scanners rapidly toggle
the gray-scale values in CT images. The HU is the position of the x-ray source (by magnetically
defined as: deflecting the electron beam inside the tube),
forming two channels from one physical detector. It
mx;y;z mw is also possible to add the signal from adjacent de-
HUx;y;z ¼ 1000 ; ð2:3Þ tector arrays together, for instance, a 64-detector
mw
array scanner with 0.625 mm detector widths can be
configured to acquire 32 1.25 mm, 16 2.50 mm,
where the HU for a voxel located at position (x,y,z),
or 8 5.00 mm channels. These systems are referred
HUx,y,z, corresponds to the measured effective linear
to as multiple-detector-row CT (MDCT) scanners.
attenuation coefficient for that voxel, mx,y,z, rescaled
Although the design details become more compli-
after being normalized to the effective linear attenu-
cated as the number of channels increases, the basic
ation coefficient of water, mw. Thus, the gray-scale
idea behind MDCT imaging remains the same.
values (CT numbers) in CT images have physical
Scanners with more data channels generally offer
meaning and are essentially rescaled linear attenu-
more coverage in the z-axis direction for each rota-
ation coefficients. With the definition given in Eq.
tion of the CT gantry, which results in even greater
(2.3), the HU of water is 0.0 and that of air is 21000.
scanning speed. However, this scanner geometry
These are the only two materials for which the
also requires more divergence of the x-rays along the
Hounsfield scale is defined and calibrated against.
z-axis direction (forming a cone angle), and allows
Since the early days of CT, a number of diffe-
more x-rays scattered within the patient to reach the
rent CT-scanner designs have been developed.
detectors, potentially creating anomalous signal in-
Currently, all commercial systems use a geometry
formation. Algorithms that take into account the
in which the x-ray tube and opposing detector
cone-beam geometry and increased scatter are
array are mounted onto a common frame (gantry)
required on these newer CT-scanner types.
that rotates on a mechanical “slip-ring” system,
In addition to the development of more numerous
which passes electrical signals across sliding metal
detector arrays with thinner detector elements,
contacts (Crawford and King, 1990; Kalender et al.,
gantry rotation periods have continued to decrease.
1990; Rigauts et al., 1990). Slip-ring designs elimi-
Current state-of-the-art systems have rotation
nated inter-scan delays that were once necessary to
periods in the range of from 0.27 s to 0.35 s per rota-
unwind cables between successive rotations of the
tion. With the use of partial-scan reconstruction tech-
gantry; this development enabled the continuous
niques (rotation of approximately 1808 plus the fan
gantry rotation and patient-table motion necessary
angle in the scan plane), this has allowed temporal
for helical (spiral) CT data acquisition. The axis of
resolutions per image in the range of from 130 ms to
rotation of the CT gantry is called the isocenter.
175 ms. Application of this technology to cardiac CT is
The rotational plane of the CT gantry is perpen-
now well established. For patients with high heart
dicular to the z axis of the system, which is typical-
rates (.70 beats/min) or irregular heart rhythms,
ly the cranial – caudal direction of the patient.
further improvements in temporal resolution are
In the late 1990s, a new CT design emerged for
required. One approach to this problem has been the
clinical use. Instead of acquiring a single ring of
development of a CT system equipped with two x-ray
transmission data with each rotation of the x-ray
tubes and two detector arrays, placed at approximate-
tube, the new CT scanners were able to acquire four
ly right angles to one another. Temporal resolutions
separate rings of data with each rotation. This was
as short as 75 ms are possible on such systems, which
made possible by the use of multiple rows of x-ray
are referred to as dual-source CT scanners.
detector arrays positioned side by side along the z
axis. The most important clinical advantage of these
scanners is the considerable decrease in routine
2.2 Scan-Acquisition Modes
scanning time while simultaneously producing
thinner image sections (higher z-axis resolution), There are two basic scan modes in CT, axial
i.e., faster scans with better spatial resolution. (sequential) scanning and helical (spiral) scanning.
Since the late 1990s, CT scanners have been Building on these are specialized modes of scan-
designed with an increasing number of detector ning for perfusion, fluoroscopy, cardiac, and dual-
channels or arrays. CT scanners with 4, 6, 8, 10, 16, energy imaging.
20, 32, 40, 64, 128, and 320 detector channels are
2.2.1 Axial CT Scanning
2
The Hounsfield Unit is not a unit in the usual sense. As it is so
basic to CT imaging, this nomenclature will be retained in this In the axial scan mode, the patient table remains
Report. stationary while the tube and detector array rotate
15
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
Figure 2.3. Illustration of the differences between axial (a) and helical (b) scan modes.
once around the patient, collecting the necessary where z is the “over-beaming” that is necessary in
data for image reconstruction. After one rotation, MDCT systems. The z portion of the beam corre-
the patient table is moved along the z axis to the sponds to the width of the penumbra on both sides
next position, and another set of scan data are of the active beam, which extends beyond the edges
acquired (see Figure 2.3a). This mode is also re- of the active detector arrays (nT) to reduce artifacts
ferred to as “step-and-shoot” scanning. If projec- in helical scanning. With a helical CT scanner, the
tions through the entire organ of interest (e.g., the ratio of the table feed b (a distance) per full gantry
heart or brain) can be acquired in one rotation, rotation (i.e., 3608) to the beam width is defined as
such as with 16 cm wide detector arrays, then no the pitch, p, thus
table translation is required.
In single-detector row CT systems, the image b
p¼ : ð2:5Þ
thickness is determined primarily by the collima- nT
tion of the x-ray beam along the z axis, and one
wide detector array (e.g., 13 mm) was used to In helical scanning, extra rotations of data acquisi-
acquire different slice thicknesses such as 1 mm, tion are required at the beginning and end of the
3 mm, 5 mm, 7 mm, and 10 mm thick CT slices, but helical scan in order to provide sufficient data for
not simultaneously. In MDCT scanners, the image image reconstruction at the edges of the prescribed
thickness is determined by the detector-element scan range. This is referred to as “over-ranging.” For
dimensions, which on modern systems are approxi- CT scanners with wider detector arrays (40 mm to
mately from 0.50 mm to 0.625 mm along the z axis, 160 mm), a significant amount of extra irradiation
as measured at isocenter (the physical detectors beyond the prescribed scan volume can occur.
are about two times wider, the difference being the Scanner manufacturers have addressed this by
geometric magnification from isocenter to the phys- using adaptive z-axis collimators (see Figure 2.4),
ical detectors). In MDCT, the data from adjacent which open and close at the beginning and end of
detector rows can be added together to give wider the scan to shield the patient from unnecessary ra-
image thicknesses. This provides the flexibility to diation (Christner et al., 2010; Deak et al., 2010).
reconstruct narrow and/or wider image thicknesses With all other technique factors held constant,
from the same measured projection data set, so a the use of larger pitch values in helical scanning
range of different slice thicknesses can be acquired will reduce patient dose; the average absorbed dose
simultaneously. for a given scan volume can be shown to be inverse-
ly proportional to pitch. Although the dose is
decreased, there is a penalty in terms of loss of
2.2.2 Helical (Spiral) Scan Mode image quality (see Table 2.1). The penalty depends
Helical (spiral) scanning involves continuous on whether the scanner has a single detector row
translation of the patient table with continuous or is a MDCT system.
x-ray rotation and data collection (see Figure 2.3b). In single-detector-row CT, as the pitch is
This decreases overall scan time, and can allow increased, the data sampling along z is more
scanning of the entire adult torso within a breath sparse, and the resultant image is wider ( poorer
hold. Helical scanning also allows for reconstruc- z-axis resolution). Image noise is not affected,
tions of overlapping slices at any z-axis position. however, as the same number of projections is
For an MDCT scanner with n arrays that have a always used to form an image. MDCT scanners use
thickness T (at isocenter), the beam width a as spiral interpolation algorithms that are different
measured at the isocenter is given by than those in single-detector-row CT and take ad-
vantage of the multiple rings of transmission data.
a ¼ nT þ z; ð2:4Þ For MDCT scanners, the width of the section
16
Basics of Computed-Tomography Technology
Figure 2.4. Adaptive collimation in helical CT. For helical CT scanning without adaptive z-axis beam collimation (left), half of the x-ray
beam width at both ends of the scan is not used because those projections do not have sufficient angular sampling for image
reconstruction, resulting in unnecessary absorbed dose in the patient. With adaptive beam collimation, collimators are used to
eliminate patient exposure at the edges of the scan, reducing the absorbed dose in the helical scan mode, especially for short scan
ranges or wide beam widths.
sensitivity profile (i.e., image thickness) remains Table 2.1. Effect of increased pitch.
relatively constant as the pitch changes. However,
Scanner type Tube Dosea Image Image
with all other technique factors held constant, such
current noise thickness
as occurs using a constant value of the product of
tube current, J, and tube-on time per rotation, t, Single-detector Constant Reduced Constant Larger
corresponding to a constant “mAs”-setting,3 image row
noise will increase as pitch increases. This is Multi-detector Constant Reduced Increased Constant
because the number of projections that can be used row
Multi-detector Increased Constant Constant Constant
while still maintaining the prescribed image thick-
rowb
ness decreases as pitch increases. However, some
MDCT scanners automatically adjust the photon a
As represented by the volume CT dose index (AAPM, 2008;
fluence, that is, Jt, to compensate for changes in IEC, 2009; McNitt-Gray, 2002).
b
the pitch setting, and therefore image thickness, Tube current, J (mA setting), adjusted as pitch, p, is changed to
noise, and absorbed dose remain constant as pitch yield the same effective photon fluence, Jt (“mAs” setting),
where the effective fluence is defined as Jt/p.
is varied. This approach allows the scanner oper-
ator to adjust pitch as needed, without having to
calculate and make the compensatory changes in 2.2.3 Repeated Scanning at the Same
Jt to maintain image-noise levels. The ratio of Jt Table Position
to pitch is referred to as the effective “mAs” or
“mAs”/slice, depending on the scanner manufactur- CT fluoroscopy is used to guide interventional
er (AAPM, 2011b). procedures, such as the placement of biopsy
needles to extract tissue samples for the assess-
ment of pathology. This near-real-time procedure is
3
During the development of CT scanners, certain jargon has useful for providing visual feedback for needle
been established in which the product of the tube current, J, guidance.
and the tube-on time per rotation, t, is referred to as the “mAs.” The CT perfusion study is used with an intravas-
Although such a variable as this product should not be named
using unit symbols (which symbols, in this case, should correctly
cular injection of an iodine-based contrast agent,
be written as mA s), the use of “mAs” appears to be so and a quantitative assessment of contrast kinetics
entrenched that it will occasionally be used in this Report to is made by scanning the same region for a relative-
avoid terms that might be obscure to the practitioner. A similar ly long period of time, which depends on the organ
issue concerns using “kV” to indicate the value of the or tissue of interest. Scan times of about 40 s are
accelerating potential, V, for the x-ray tube (which is commonly
expressed in units of kV). When needed, these quantities will be
typical for brain perfusion imaging, although scans
termed “mAs” setting and “kV” setting, which correspond to up to 3 min in duration can be necessary in abdo-
displays on most scanners. minal organs. In both of these scan modes, the
17
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
table is stationary and the same region of the models of Siemens CT scanners have two x-ray tubes
patient is repeatedly scanned, resulting in the and detector arrays, and these allow simultaneous ac-
buildup of absorbed dose in the tissues within and quisition of low- and high-V data. Other manufac-
around the scanned volume. In these acquisition turers have developed single x-ray tube systems that
modes, it is typical to operate the scanner with are capable of rapidly switching the tube potential in
a lower tube current, as very high absorbed order to acquire low- and high-V data in one rotation
doses can result from these repetitive scanning of the gantry. In switched-potential systems, the
procedures. x-ray pulse duration can be adjusted to balance the
x-ray intensity levels between the low- and high-V
pulses. In the future, it is anticipated that photon-
2.2.4 Dual-Energy CT
counting detector systems will be developed for CT;
Dual-energy CT scanning more easily allows the such systems will differentiate detected photon
delineation of regions in the patient corresponding energy into several energy channels, and thus dual-
to different densities and mean atomic number, energy scanning can be performed without switching
such as bone, iodine, or soft tissue. The acquired x-ray-tube potential.
dual-energy CT images are processed such that Dual-energy CT scanning can be used also to es-
subsequent computer-based enhancement or elim- timate so-called “monoenergetic” images, which are
ination of certain tissue types is accomplished. An mathematical combinations of the high- and low-V
example would be subtracting out the bones in a scans that approximate the CT image that would
neck CT angiogram to allow the physician to better be produced with a monoenergetic x-ray beam
evaluate the vasculature without having it (instead of the polyenergetic spectra that were ac-
obscured by overriding boney anatomy. Equation tually used). Dual-energy CT imaging also allows
(2.3) defines the relationship between the CT the correction for beam-hardening artifacts. Beam
number and the effective linear attenuation coeffi- hardening occurs as a polyenergetic x-ray spectrum
cient, and it is the differences in how meff decreases passes through a thick, dense structure such as the
for higher-energy x-ray spectra that allows discrim- bones in the skull (see Figure 2.5). As the x-ray
ination of materials of different atomic number. beam traverses the tissue, the overall x-ray fluence
Dual-energy scanning requires the acquisition of is of course reduced, but there is also a shift in the
two sets of projection data, each of which contain a shape of the x-ray spectrum as lower-energy x-ray
sufficient number of projections for image recon- photons are absorbed preferentially as a function of
struction. One set is acquired at a high x-ray-tube depth relative to higher-energy photons. This
potential, V, (e.g., 140 kV) and the other at a lower occurs because, in general, the effective linear
x-ray-tube potential (e.g., V ¼ 80 kV). From either
the reconstructed images or the raw projection
data, the effective atomic number of the tissue com-
position can be differentiated from the material
density. The reason for this is that the CT number
of any voxel is a rescaled measure of the linear
x-ray attenuation coefficient [Eq. (2.3)]. Thus, the
same amount of x-ray attenuation (i.e., the same
CT number) can be obtained for two different mate-
rials (e.g., iodine and bone) in typical single-energy
CT scanning, if the densities of the materials com-
pensate for the differences in atomic number. In
dual-energy CT, the projection data acquired at the
two x-ray tube potentials provide the information
needed to assess what fraction of the x-ray attenu-
ation is due to the material composition (atomic
number) and what fraction is due to mass density. Figure 2.5. Beam hardening. Three x-ray spectra, all produced
The dual-energy concept was initially described by at 120 kV, are shown, corresponding to the spectra at different
Godfrey Hounsfield in 1973 (Hounsfield, 1973), and points along the x-ray beam path in CT. These spectra are
was subsequently explored by several investigators interpolated measured data, and the energy resolution was not
sufficient to resolve the individual characteristic x-ray peaks.
(Alvarez and Macovski, 1976; Kalender et al., 1986)
Beam-hardening artifacts occur if the reconstruction algorithm
decades ago. Widespread clinical applications of does not take beam hardening into consideration (see inset,
dual-energy CT did not emerge until approximately arrows point to a dark region in the brain resulting from beam
2007 (Johnson et al., 2007; Primak et al., 2007). Some hardening).
18
Basics of Computed-Tomography Technology
attenuation coefficient is higher at lower x-ray en- noisier, reduced-dose projection data, while the
ergies. This process causes meff in Eq. (2.1a) to vary quality of the coronary angiogram was maintained
as the x-ray beam passes through the patient, cre- during the full-dose portion of the cardiac cycle.
ating slight errors in the CT numbers on the recon- This strategy evolved further to eliminate the con-
structed CT images. Artifacts that occur in the stant motion of the patient (and table) through the
image, often adjacent to highly attenuating struc- gantry, using axial prospectively triggered acquisi-
tures, are referred to as beam-hardening artifacts. tions (see Figure 2.6b), in which the tube current
was shut off in synchrony with the ECG signal
2.2.5 Cardiac CT during the scan, and also between table movements
(Hsieh et al., 2006; Stolzmann et al., 2008). Wider
Initially, CT imaging of the coronary arteries detector coverage, up to 160 mm for some systems,
was accomplished with unique CT scanner designs, also became available, such that the entire heart
for example, electron-beam CT. These systems could be imaged in one beat (Einstein et al., 2010;
were not, however, widely available. Cardiac CT Goma et al., 2011; Rybicki et al., 2008).
became more feasible clinically with the advent of With dual-source CT scanner technology, a high-
four-detector-row CT scanners. The technique pitch mode (so-called “flash” scan mode, pitch of
required that the CT scan continuously throughout 3.4, Siemens Healthcare) was developed to image
the cardiac cycle and the table be translated slowly the entire heart in one beat using approximately
along the z axis, such that every z-axis position 40 mm detector coverage (Achenbach et al., 2010;
was imaged during every phase of the cardiac Flohr et al., 2009; Lell et al., 2009; McCollough
cycle. The ECG trace was recorded simultaneously et al., 2009). For stable heart rates below approxi-
with the CT projections. This data redundancy was mately 65 beats/min, high-pitch scanning enabled
required because the reconstruction algorithm a reduction in the estimated effective dose from
sorted through the acquired data to determine approximately 12 mSv using 64-detector-row
which projection data were appropriate to recon- technology (Hausleiter et al., 2009) to below 1 mSv
struct a CT image at a given z position at a given (Achenbach et al., 2010; Raff et al., 2009;
phase in the cardiac cycle. This approach, referred Stolzmann et al., 2008).
to as retrospective gating (see Figure 2.6a), was
very dose inefficient compared with non-cardiac
CT exams.
With 16- and 64-detector-row scanner technology, 2.3 Scan-Acquisition Parameters and Their
the use of ECG-based tube-current modulation Effects on Image Quality
became common, whereby the tube current was CT image quality can be predictably affected by
reduced by a factor of 5 or more during those phases several scan-acquisition parameters. These para-
of the cardiac cycle in which image reconstruction meters include x-ray-tube potential, V (typically in
was not anticipated (Jakobs et al., 2002). In this units of kV), tube current, J (typically in units of
manner, ventricular and valve function could still be mA), rotation time, t (typically in units of s), pitch,
assessed over the entire cardiac cycle using the p, reconstructed CT image thickness, beam and de-
tector collimation, and scan mode.
The typical range of x-ray-tube potentials pro-
vided in CT systems is approximately from 80 kV to
140 kV. Within this range, the effective linear at-
tenuation coefficient for many tissues changes sub-
stantially. Although the use of lower V settings to
improve soft-tissue contrast has been recommended
(Huda et al., 2000a; 2000b), this approach can lead
to fairly severe beam-hardening artifacts, even in
small children (Cody et al., 2004). Higher V
(130 kV to 140 kV) is often used in CT imaging of
the bodies of very large patients.
Contrast resolution (i.e., the ability to distinguish
lesions differing only slightly in HU values) is
Figure 2.6. In retrospective gating (a), the x-ray beam is
strongly dependent on CT-image noise. Noise is
continuously on throughout the cardiac cycle, irradiating the
patient much longer than with prospective triggering (b), in the pseudo-random visual pattern of CT numbers
which the x-ray beam is turned on only for short periods of time observed in a CT image of a uniform object such as a
during the phase of interest within the cardiac cycle. water-filled cylinder. The so-called root-mean-square
19
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
(RMS) noise can be determined on a CT image by thicknesses with no additional scanning necessary,
computing and hence no additional absorbed dose in the
patient. If, however, the CT-acquisition protocol is
sffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
Pn 2 set to acquire eight 5.0 mm thick rings of transmis-
i¼1 HUi HU ; ð2:6Þ sion data, retrospective generation of thinner CT
s ¼
n1 images would not be possible. Although the thinner
1.25 mm images convey more spatial resolution
where s is the RMS noise, and HUi corresponds to along the z axis, they are substantially noisier than
the CT number in each of n voxels in a region with a corresponding 5.0 mm thick images for the same
mean voxel value of HU (mean CT number). A acquisition parameters.
simple software tool to calculate RMS noise in a
user-defined region of interest is commonly available
on all current CT scanners and on image-review
2.4 Dose-Reduction Techniques
workstations. Image noise cannot be measured ac-
curately in patient images (due to the presence of Modern dose-management strategies in CT
tissue heterogeneity) unless regions of known homo- imaging recommend that the appropriate radiation
geneous tissue composition are present, such as fluence level be used for the specific patient thick-
urine in the bladder or blood in large arteries with ness and the specific diagnostic task for which the
no contrast agent present. Noise is routinely mea- scan was ordered. Unlike film-based radiographic
sured on phantoms of uniform composition. imaging, a CT image never appears “over-exposed”
The image noise in a CT image is affected by the or “underexposed” in the sense of being too dark
reconstruction methods used for computing the CT or too light. The normalization of the acquired CT
image from the raw projection data, and in general projection data [Eq. (2.2)] factors out the incident
the reconstruction process tends to add noise correl- beam intensity used during acquisition, resulting
ation to the reconstructed image. The correlation in uniform gray scale images for low-dose and
structure of CT image noise can be characterized high-dose images alike. As a consequence, CT tech-
by computing the noise-power spectrum, and this nologists do not receive the feedback on technique
will be discussed in more detail in Section 11. selection that was available in radiography during
Despite the influences of reconstruction methods the screen-film era. Therefore, the CT operator needs
in CT images, image noise is principally a function to consider patient size when selecting parameters
of the photon counting statistics of the individual that affect absorbed dose and image quality, the most
transmission measurements, which are a function basic factor being the photon fluence used to acquire
of the photon fluence measured at the detector ele- the scan, which is most commonly adapted by adjust-
ments during the scan. Thus, in CT scanning, any- ing the tube-current–time product, Jt (mA s) (FDA,
thing done to increase the number of photons 2002; Linton and Mettler, 2003).
reaching the detector will in general result in lower
image noise and better contrast resolution.
2.4.1 Fixed Tube Current (Technique Charts)
Another factor that directly affects image noise,
and Patient Size
and thus affects contrast resolution, is the thick-
ness of the reconstructed CT image t. If the image For most CT applications, it is common to stand-
thickness is reduced by a factor of 2, (e.g., from ardize the tube potential and gantry rotation time:
5 mm to 2.5 mm), then pffiffiffi the CT image noise the fastest rotation time is typically used to reduce
increases by a factor of 2 (or about 40 %), because motion artifacts and the lowest plausible tube po-
only half of the x-ray photons contribute to the tential is selected to maximize image contrast, pro-
image in that case. In MDCT scanners, the detector vided that the tube-current limits are adequate to
configuration implemented during the scan acquisi- provide sufficient photon fluence (Funama et al.,
tion will define the range of CT image-thickness 2005; Huda et al., 2002a; McCollough et al., 2006;
values that are available for retrospective CT Nakayama et al., 2005; Siegel et al., 2004). Thus,
image reconstruction. For example, if a CT scan x-ray tube current, J, is the primary parameter
protocol on a 64-slice MDCT system calls for the ac- used to account for patient size (see Section 8).
quisition of 64 rings of transmission data, When all other parameters are kept constant, I has
0.625 mm thick, then this finely sampled data can a linear relationship with detected x-ray fluence
be used (after the scan) to retrospectively recon- rate, i.e., mðx;
_ uÞ ¼ k1 J, and the product of tube
struct thicker images such as 1.25 mm, 2.5 mm, or current, J, and exposure time, t has a linear rela-
5.0 mm thick images. Thus, several sets of image tionship with detected x-ray fluence; mðx; uÞ ¼ k2 J t,
data can be reconstructed having different image where k1 and k2 are constants of proportionality
20
Basics of Computed-Tomography Technology
that depend on numerous scanner and technique quality for the diagnostic task. The specific imple-
factors. mentations of AEC differ by the manufacturer.
Tube current is typically adjusted to patient Practically speaking, acceptable noise levels can
dimensions based on the overall attenuation of the change across patient-size ranges, and so size-
anatomy of interest. Weight is used in some set- based technique charts are sometimes used to pre-
tings to adjust tube current, although weight is scribe the AEC parameters (Kalra et al., 2004;
less useful than actual physical dimensions as a 2005; McCollough et al., 2006; Nyman et al., 2005;
quantity to adjust for x-ray attenuation (Boone Wilting et al., 2001).
et al., 2003; McCollough et al., 2002; Wilting et al.,
2001). Importantly, when considering CT images 2.4.3 Angular and Longitudinal X-Ray-
acquired across a range of patient sizes, it has been Tube-Current Modulation
shown that radiologists do not find the same
Angular tube-current modulation addresses (see
image-noise level acceptable for all patient sizes
Section 9) variation in the x-ray attenuation path
(Wilting et al., 2001). Radiologists tend to require
within a scan plane, for example, the differences in
lower image noise in children relative to larger
the anterioposterior (AP) and lateral dimensions of
patients because children often lack adipose tissue
the patient at a particular location along the z axis
between organs and tissue planes and have smaller
of the patient. The tube current is varied as the
anatomic dimensions (Boone et al., 2003; Kalra
x-ray tube rotates around to the patient, with
et al., 2004; McCollough et al., 2002; Wilting et al.,
higher current used for thicker projections and
2001). For body CT imaging, it has been found that
lower current for thinner projections through the
a reduction in Jt by a factor of from 4 to 5 from
patient’s anatomy. Attenuation information is
adult techniques is generally acceptable in infants,
determined from the previously acquired CT locali-
although for obese patients, an increase of at least
zer view, or on some systems the projection meas-
a factor of two is required (McCollough et al.,
urement acquired 1808 before the currently
2002). To achieve sufficient exposure levels for
acquired projection is used to modulate the tube
obese patients, the tube current can be increased,
current level; for a 0.30 s gantry rotation period,
the gantry rotation period can be increased, or the
that requires dynamic (150 ms) modulation hard-
pitch can be decreased (at constant Jt setting).
ware. X-ray-tube-current modulation along the z
axis of the patient takes into account variations in
2.4.2 Automatic Exposure Control attenuation among different regions (e.g., shoulders
versus abdomen) by varying the average Jt along
Modern CT systems have the capability to adjust
the patient’s long axis. To achieve the right com-
the x-ray-tube current in response to variations in
promise between image quality and absorbed dose,
patient attenuation (Gies et al., 1999; Haaga et al.,
the parameters that influence tube-current modu-
1981; Kalender et al., 1999; McCollough, 2005).
lation, which are scanner-manufacturer dependent,
Methods of adapting the tube current to patient at-
must be clearly communicated by the manufacturer
tenuation, known generically as automatic expos-
and properly chosen by the user. Both angular and
ure control (AEC), are analogous to AEC methods
z-axis tube-current modulation are effective
(also called photo-timing) in general radiography,
methods of achieving CT dose reduction.
and in CT have demonstrated substantial reductions
in absorbed dose when image quality is appropriately
2.4.4 Adjusting X-Ray-Tube Potential Based
specified. An exception to this trend occurs with obese
on Patient Size
patients, in which the radiation output of the system
is increased under most AEC system scenarios to Several investigators have studied the use of
ensure adequate image quality. In obese patients, lower-tube-potential CT imaging to improve image
much of the additional x-ray energy is absorbed by quality or reduce absorbed dose. The principle
excess adipose tissue, a relatively radio-insensitive behind lower-energy imaging is that the attenu-
tissue. In obese patients due to x-ray attenuation by ation coefficient of iodine increases as photon
peripheral adipose tissues, absorbed doses in internal energy decreases toward the iodine K-edge energy
organs are slightly reduced at the same tube of 33 keV. In CT exams involving the use of iodi-
current–time product compared with thinner nated contrast media, the superior enhancement of
patients (Schmidt and Kalender, 2002). iodine at lower tube potentials improves the conspi-
AEC is a broad term that encompasses not only cuity of hypervascular or hypovascular pathologies.
tube-current modulation (see Section 3), but also However, images obtained using lower tube poten-
refers to overall technique selection according to tials tend to be much noisier, mainly due to the
patient size in order to achieve appropriate image higher absorption of low-energy photons by the
21
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
patient, unless the tube current is adequately the patient. The tube assembly for modern CT
adjusted (FDA, 2002; Funama et al., 2005; Haaga scanners usually has between 1 mm and 3 mm alu-
et al., 1981; Huda et al., 2000a; Kalender et al., minum filtration with an additional flat filter of
1999; McCollough, 2005; Nakayama et al., 2005; 0.1 mm copper, giving a total filtration of between
Siegel et al., 2004). The x-ray tube output (for in- 6 mm and 9 mm of aluminum equivalent. As x-ray
stance, the measured air kerma at isocenter) changes tube and generator technology has advanced,
non-linearly as a function of V, when other para- allowing higher peak power levels, additional filtra-
meters are held constant (see Figure 2.7). In general, tion can be used without limiting the ability of the
there is a power-law function that describes x-ray system to achieve the necessary x-ray fluence rates
tube output as a function of x-ray-tube potential, V: for fast CT scanning protocols. MDCT scanners also
make much more efficient use of the x rays that are
Kair ¼ aðVÞb ; ð2:7Þ produced because of the larger beam widths that are
used. Compared with a single-detector-row CT
where Kair is the air kerma at isocenter, a a coeffi- scanner with a 5 mm beam width, an MDCT system
cient, and b the parameter of interest here. As seen with a 40 mm beam width (i.e., nT ¼ 40) is eight
in Figure 2.7 for four CT scanners from different times more efficient in terms of usable photon
manufacturers, the exponent b ranges from 1.8 to fluence rate per x-ray-tube power.
2.6. As a generalization, b increases with increasing Because most body regions that are scanned in
added filtration in the x-ray beam, and that is why clinical CT are approximately cylindrical, the x-ray
the curves in Figure 2.7 have decreasing overall path length is smaller at the periphery of the
output levels as the value of b increases. patient relative to the center. If uncompensated,
The optimal and most dose-efficient tube poten- the x-ray fluence reaching the detectors corre-
tial is dependent on the patient size and diagnostic sponding to the periphery of the patient would be
task (McCollough, 2005), considering the much greater than at the center of the patient. To
contrast-to-noise ratio in tissues with and without compensate for this, the x-ray beam in CT scanners
contrast agent (Yu et al., 2010). is attenuated more at the periphery and less at the
center. These bow-tie or wedge filters reduce un-
necessary patient absorbed dose by matching the
2.4.5 Beam-Shaping Filters
filter shape to the patient size and clinical applica-
CT scanners generally use a greater amount of tion. Most CT systems have two or three different
beam filtration than x-ray systems used for radiog- bow-tie filters, which are selected based on patient
raphy or fluoroscopy in order to reduce the beam size. In cardiac CT, the region of interest is cen-
hardening that occurs as the beam passes through tered within the thorax, and therefore the x-ray
fluence can be substantially reduced outside the
cardiac volume of interest using an appropriately
designed bow-tie filter, with no loss of diagnostic in-
formation. The use of beam-shaping filters on CT
scanners heightens the importance of properly cen-
tering the patient in the scan field.
22
Basics of Computed-Tomography Technology
focal-spot distribution. The algorithm might also noise-reduction techniques can be used to reduce
include a model of the quantum noise (i.e., Poisson noise levels in the projection data (Funama et al.,
distribution). The forward projection data are com- 2011; La Riviere, 2005; La Riviere et al., 2006; Li
pared with the originally measured projections, et al., 2004; Manduca et al., 2009; Silva et al., 2010;
and the differences are used to update the estimate Wang et al., 2006) or on the reconstructed CT
of the truth. This process is repeated (iterated) images (Bittencourt et al., 2011; Tipnis et al.,
until the differences between the actual and simu- 2010). These approaches can reduce noise, but do
lated projections are acceptably small. This tech- not in themselves reconstruct an image and do not
nique can produce images that more closely reduce artifacts or improve spatial resolution.
resemble the scanned object, and, in particular,
noise and artifacts can be reduced substantially
2.5 Summary
and spatial resolution improved, leading to the
potential for absorbed-dose reduction. CT technology has evolved over four decades to
Iterative-reconstruction methods or other image the point at which CT is one of the most widely
noise-reduction techniques do not, by themselves, used diagnostic imaging examinations performed.
reduce absorbed dose in CT scanning. Rather, by Its rapid scan time and isotropic resolution provides
improving the image quality through noise reduc- excellent anatomical detail, and - with the use of
tion, the technical factors that affect absorbed dose dynamic imaging and contrast-medium injection -
(tube potential and current, pitch, etc.) can be functional physiological information such as regional
adjusted to realize excellent image quality at blood volume and tissue perfusion can be assessed.
reduced dose levels. Because CT entails more absorbed dose than radiog-
In contrast to iterative-reconstruction methods, raphy, there have been a number of recent efforts to
which form an image from the projection data, reduce the radiation levels in clinical CT scanning.
23
Journal of the ICRU Vol 12 No 1 (2012) Report 87 doi:10.1093/jicru/nds003
Oxford University Press
Figure 3.1. Images from a 42-year-old with the new onset of seizures. Axial CT slices without contrast (left) could be interpreted as
normal. Following contrast administration (right), there are enhancing foci due to metastatic melanoma (arrows).
26
Computed Tomography in Clinical Use
Figure 3.3. Images from a 44-year-old with upper and lower extremity paralysis following trauma. (a) Axial CT images demonstrate a
burst fracture of the C6 vertebral body (black arrow) with spinal-canal narrowing (white arrow). (b) These findings are more clearly
demonstrated on reformatted sagittal images.
27
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
28
Computed Tomography in Clinical Use
Figure 3.8. Axial chest CT scan without contrast in an Figure 3.9. A 64-year-old male presented with chronic cough
asymptomatic 73-year-old man. This image shows a 1.4 cm and an abnormal chest radiograph. Chest CT without contrast
nodule in the left upper lobe. Comparison with a prior chest CT shows a heterogeneous mass-like opacity (arrow) in the right
performed 4 y earlier showed interval growth of the nodule. upper lobe. Further clinical evaluation led to a diagnosis of lung
cancer.
29
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
Figure 3.10. Chest CT, coronal reconstruction in a 52-year-old Figure 3.11. Images from a 35-year-old male with right-lower-
male smoker. The patient complained of increasing shortness of quadrant pain. Coronal CT reformatted image demonstrates the
breath. This CT image shows abnormal lung parenchyma, liver (L) and the kidney (K). There is a tubular structure seen in
especially at the apices (arrows) representing bullous the right lower quadrant of the abdomen (arrow), which
emphysema, a smoking-related lung disease. represents an inflamed appendix in this patient.
radiation. However, in many parts of the world, imposes an obstruction, which causes kidney swel-
ultrasound is often used as a screening exam, and ling due to the back-up of urine, with associated
CT is used for definitive diagnosis, particularly in intense flank pain.
situations in which ultrasound visualization is CT is well suited for the examination of the
obscured by overlying structures in the abdomen patient with acute blunt or penetrating abdominal
such as loops of air-filled bowel. With the rising trauma (Federle et al., 1981). In these situations,
number of obese individuals in many countries, CT the patient is often unstable, and it is essential to
is often used as the first-choice imaging modality rapidly diagnose any potential abnormality for im-
owing to the difficulty of obtaining diagnostic- mediate surgery or other interventional therapy.
quality ultrasound images of internal organs in This can include injuries to arteries causing intern-
obese individuals. CT gives a quick and comprehen- al bleeding, injuries to hollow viscous tissues (e.g.,
sive view of the abdomen and pelvis, and is the bowel), or injuries to solid organs. Figure 3.13
imaging method of choice in a number of clinical shows a CT image of a patient with blunt trauma
situations. Situations in which CT is helpful are from a motor-vehicle accident. The patient has a se-
described below. verely fractured spleen with active bleeding from
There are numerous etiologies of abdominal the spleen into the abdomen.
pain, for which CT is an ideal tool (Rao et al., CT is the preferred imaging modality for the
1997). CT can evaluate inflammatory conditions in detection of malignant or non-malignant abdom-
solid organs or hollow viscous regions within the inal masses (Freeny et al., 1988). CT can also
abdomen. These can include evaluation of pancrea- guide biopsy or treatment of these masses.
titis or diverticulitis and related complications. CT Figure 3.14 shows images of a patient with pan-
can evaluate other inflammatory conditions such creatic cancer; CT can detect the cancer, deter-
as acute appendicitis (see Figure 3.11). With this mine local extension, and identify liver and other
condition, the appendix becomes inflamed and metastases, resulting in tumor staging. Tumor
enlarged, usually due to a small appendicolith staging is important because it helps to guide
(small stone-like structure) obstructing the appen- oncologists as to what therapies should be consid-
dix. This can be seen as a fluid-filled enlarged ap- ered. CT imaging can also be used to follow oncol-
pendix in the right lower quadrant of the abdomen ogy patients to evaluate their response to tumor
(Figure 3.11). In some situations in which appendi- therapy.
citis is suspected, there can be other etiologies of CT is well suited to guide needle-biopsy place-
the patient’s pain. Figure 3.12 shows CT images of ment into abdominal masses, particularly those
a patient with right-sided pain due to a renal that are not amenable to biopsy under ultrasound
stone, which has lodged in the ureter. The stone guidance. Biopsy needles are placed in the
30
Computed Tomography in Clinical Use
Figure 3.12. Coronal CT image from a 45-year-old with right-sided flank pain. (a) This coronal CT image demonstrates the liver (L),
right kidney (K), and the bladder (B). There is dilatation of the right renal pelvis and ureter (arrow). (b) Another reformatted coronal
image demonstrating the liver (L), kidney (K), and the bladder (B). Note that there is a high-attenuation spherical structure
representing a kidney stone (arrow), which is obstructing the right ureter and causing the patient’s right-sided pain.
31
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
Figure 3.15. Images from a 68-year-old with metastatic colon cancer. (a) The axial CT scan shows a heterogeneous mass in the right
lobe of the liver (arrow). (b) This liver lesion was treated with radiofrequency ablation (RFA), and three electrodes used to perform the
RFA are seen in the liver.
32
Computed Tomography in Clinical Use
Figure 3.18. (a) Three-dimensional reconstruction of a complex (tri-plane) ankle fracture in a 14-year-old boy (arrows show multiple
fracture lines). These 3D images can be viewed from various angles, and allow the surgeon to accurately plan the surgical reduction of
this injury. (b) A post-operative radiograph shows anatomic bony alignment after fracture reduction and internal fixation with plates
and screws.
the standard imaging modality used in the evalu- fibrosis, can also be followed to assess its severity
ation of pediatric blunt abdominal trauma, usually and/or response to therapy.
from motor vehicle or bicycle accidents, falls, or CT scanning is very useful in evaluation of bony
sports injuries. CT can accurately define the extent abnormalities. Three-dimensional reformatted
and type of intra-abdominal and intra-pelvic injury images are particularly useful to the orthopedic
(Wootton-Gorges, 2010). CT can also be used in the surgeon managing spinal curvature, fractures,
diagnosis and management of children with ab- or bone tumors. Figure 3.18 shows a 3D reforma-
dominal tumors. Figure 3.17 shows a child with tion of a complex fracture of the ankle. Three-
Wilms tumor, the most common kidney and dimensional images allow the surgeon to accurately
intra-abdominal malignant tumor in children. plan the operative procedure before going to the op-
Although ultrasound is favored in many centers for erating room.
diagnosing appendicitis in children, CT has also
been useful in the diagnosis of appendicitis in chil-
dren. In some centers, using CT to diagnose appen-
3.6 CT in Clinical Use: Summary
dicitis has decreased the negative appendectomy
rate (Callahan et al., 2002). One study also CT was originally utilized for examining the
describes less delay in treatment and fewer compli- head and brain only, because the head could be
cations with the use of CT in over 2000 patients immobilized for long scan periods. Through four
with suspected appendicitis seen between 1997 and decades of technological advancement, CT scanners
2004 (Frei et al., 2008). can now acquire high-resolution images throughout
CT scanning of the chest in the pediatric popula- the body in a matter of a few seconds. The increased
tion is most frequently performed for trauma and image quality and decreased scan time has led to
for evaluation for metastatic tumor spread to the the use of CT for many examinations for which radi-
lungs or other tissues in the thorax. Evaluation of ography was used in the past. This has increased
serious pneumonia is another common reason phy- the utilization of CT imaging in most clinical disci-
sicians ask for CT scans of the chest. Less-frequent plines. Although the diagnostic accuracy of CT is
reasons for scanning the chest include congenital generally superior to radiographic (or other)
abnormalities, such as maldevelopment of the imaging procedures, the increased use of CT has
heart or large vessels in the chest, or primary also increased concerns about large populations
tumors in the chest. Lung disease, such as cystic receiving higher exposures to ionizing radiation.
33
Journal of the ICRU Vol 12 No 1 (2012) Report 87 doi:10.1093/jicru/nds004
Oxford University Press
routine clinical decision-making and protocol techniques. Although CTDI-based methods have
selection. proved useful in the past when the CT function
† The computation of CT doses for a specific was far more limited, it is clear that CT-dosimetry
examination type for a general population of methods need to advance to keep pace with the
patients ( pediatric patients, women undergo- technological advances of the scanners themselves.
ing CT evaluation for pulmonary embolism This is discussed in detail below.
during pregnancy, trauma patients, etc.) for Ideally, a single radiation-dose quantity in CT
clinical practice or research studies. would be usable by all parties of interest. However,
† Estimation of peak doses in specific organs different quantities might be necessary for the
(e.g., brain, skin) to assess the potential for needs of various stakeholders involved in CT, in-
deterministic effects in generic patients for cluding patients, radiologists and other physicians,
specific CT examinations (e.g., CT head perfu- medical physicists, lawyers, and others. For
sion, CT angiography). example, to monitor radiation output of a CT
† Assessment of the cumulative organ dose scanner over extended periods of time, a number of
from CT examinations and comparison of this different radiation measurements could be used.
value with the exposures from other artificial- However, for the assessment of potential fetal risk
or natural-radiation sources. from an extensive series of CT procedures, more
† Comparison of the estimated radiation risks specific dosimetric quantities would be required.
associated with different radiological methods Section 4.2 discusses existing dosimetric quantities
that are expected to provide similar medical that have been used for these various purposes for
outcomes. the past several decades.
(3) Dosimetry performed on specific individual CT dosimetry is in a state of flux as a result of a
patients that usually (but not always) occurs number of recent publications, including AAPM
after the examination has been performed, and Report 111 (AAPM, 2010). The purpose of this
typically includes the following situations: Section is to define the state of the art for CT dose as-
† Estimation of the absorbed dose in specific sessment. It should be noted that the CTDI method-
organs associated with a particular ology is a key part of International Electrotechnical
examination. Commission regulations, and it is therefore unlikely
† Estimation of the stochastic radiation risk that measurement protocols that are a part of many
associated with a specific CT examination. regulations worldwide will be changed in the short
† Assessment of the absorbed dose to a fetus term. Therefore, a review of current methodologies
when a pregnant woman is undergoing or has for assessing “dose” from a CT scanner will be
had a CT examination. described below.
† Absorbed dose in exposed tissues when there
is concern about the potential for the patient
to develop deterministic effects from a high- 4.2 CTDI-Based Metrics
dose CT procedure.
4.2.1 Basic Tools
(4) The assessment of CT dosimetric quantities to
evaluate large numbers of dose estimates for Figure 4.1 illustrates the basic shape of the
characterization of an institution’s CT-dose 100 mm “pencil chamber” used for over 30 y for CT
practices, including: dose measurements. The pencil chamber has a
† The use of dose-reporting software to charac- 100 mm active length and is designed to fit within
terize quantities such as the volume com- a solid phantom with holes placed at various loca-
puted tomography dose index (CTDIvol ), tions as illustrated in Figure 4.2. The design of the
dose – length product (DLP), or size-specific chamber produces a uniform response along its
dose estimates (SSDE) for classes of CT exam- axis. A radiation meter longer than the collimated
inations such as the head, chest, or abdomen/ radiation beam was required historically in CT
pelvis CT scans. dose assessment because it is difficult to position a
† Assessment of the DICOM object (RDSR small radiation meter precisely within the very
object—radiation dose structured report) for narrow collimated slice thicknesses, especially in
CT-dose registry programs. the era of single-slice scanners in which nT ranged
typically from 1 mm to 10 mm. With multi-detector
Increasing concern over the radiation dose asso- array (n) CT scanners with detector widths of T (in
ciated with modern CT scanners has led to the mm), the parameter nT describes the nominal
need for more accurate methods for CT-dose assess- beam width, and this is a very useful measure in
ment and for a wider range of CT-dose assessment CT acquisition protocols and dosimetry in general.
36
Overview of Existing CT Dosimetry Methods
Figure 4.2. Schematics of the CTDI phantom. (a) CTDI phantoms are cylinders comprised of PMMA. At a minimum, there is a
through-hole at the center and at one location on the periphery of the phantom. Various manufacturers have slightly different designs,
and some phantoms have several holes around the periphery. (b) The phantom is illustrated with a pencil chamber positioned in the
central hole. All remaining holes are plugged with PMMA rods during measurements. With the phantom placed on the patient table,
the 12 o’clock and 6 o’clock positions are illustrated.
37
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
measuring dose need to be addressed, as will be Another parameter becomes useful in this discus-
discussed later in this Report. sion when considering the clinical reality of a
The phantoms used for CT dosimetry over the series of axial CT scans that cover a length of the
years were defined by the report of AAPM patient. This is the scan interval, b, defined as the
Diagnostic Radiology Committee Task Force on CT table-translation distance between rotations. An
Scanner Phantoms (AAPM, 1977). The general interval of b ¼ nT produces “contiguous” scans,
design of the CT dosimetry phantom is illustrated leaving no gaps in the acquired image data. CTDI1
in Figures 4.2a and 4.2b. The phantoms are fabri- represents the average dose (i.e., air kerma) in a
cated from polymethyl methacrylate (PMMA), or defined phantom over a small interval +b/2 about
[C5H2O8]n, which has a density of approximately the center of the scan length that accumulates at
1.19 g/cm3. Although the design of the phantom z ¼ 0 from a series of N contiguous (b ¼ nT) scans
has several variations, in general there is a central covering a scan length L ¼ Nb, which is sufficient
hole and one or more peripheral holes, which are to produce scatter equilibrium at z ¼ 0 (symbolical-
through-bored with diameters of 12 mm, large ly denoted by L ! 1) (AAPM, 1990; IEC, 2009;
enough to allow the placement of the pencil COCIR, 2000; Shope et al., 1981).
chamber in any of the holes. When the pencil The CTDI1 is in principle measured in the axial
chamber is placed to make a measurement in one scan mode using a single 2p rotation of the x-ray
hole, PMMA rods are placed in all the other holes source around a stationary phantom, and allows
to displace the air with solid material. There are one to estimate the average absorbed dose at the
two phantoms in general usage for CT dosimetry, a center (z axis) of the scan volume resulting from
“body” phantom that is 320 mm in diameter, and a multiple gantry rotations, if the scan length is suf-
“head” phantom that is 160 mm in diameter. The ficiently long for the central dose to approach its
head phantom also doubles as a pediatric body asymptotic upper limit (McCrohan et al., 1987;
phantom, with an equivalent cross-sectional area of COCIR, 2000; Shope et al., 1981). To measure
the typical 2-year-old. Both of the PMMA phantoms CTDI1 in one axial scan, the measurement instru-
are 150 mm in length, although some phantoms ment needs to be long enough to integrate the ma-
that were 140 mm in length were manufactured and jority of the dose from scattered radiation that
used in some settings. propagates away from the scan plane.
4.2.2 CTDI1
In this Section, the methods to measure or calcu- 4.2.3 CTDIFDA
late various traditional CT dose descriptors are
described. The CTDI is currently the primary dose The multiple-scan average dose (MSAD) repre-
measurement concept in CT: sents the absorbed dose averaged over a small scan
interval, b, and requires multiple exposures for its
ð
1 1 direct measurement. The CTDI approach offers a
CTDI1 ¼ KðzÞdz; ð4:1Þ more convenient yet nominally equivalent method
nT 1
of predicting the MSAD, and requires only a single
where axial CT scan acquisition. In the early days of CT,
K(z) is the air kerma in the phantom as a func- this saved considerable time and x-ray-tube heat
tion of position on the z axis. loading.
n is the number of tomographic sections imaged Theoretically, the equivalence of the MSAD and
in a single axial scan. This is equal to the number the CTDI requires that all contributions from the
of data channels used in a particular scan. The tails of the absorbed-dose distribution be included in
value of n is less than or equal to the maximum the CTDI measurement. The exact integration limits
number of data channels available on the system. required to meet this criterion depend upon the width
T is the nominal width of the tomographic of the primary beam, the scattering medium, and the
section along the z axis imaged by one data x-ray beam energy. To standardize CTDI measure-
channel. In MDCT scanners, several detector ele- ments, the FDA (1984) introduced the integration
ments can be grouped together to form one data limits of +7T, where T is the nominal slice width.
channel. In single-detector-row (single-slice) CT, Interestingly, the original CT scanner, the EMI
the z-axis collimation at the isocenter determines Mark I, was a dual-detector-row system. Hence, the
the nominal beam width. nominal radiation-beam width was equal to twice the
38
Overview of Existing CT Dosimetry Methods
nominal slice width (i.e., 2 T). Thus, 20 mm, the integration length of 14T ¼ 17.5 mm is
ð 7T smaller than the primary-beam width and clearly
fSI beyond the intent of the CTDIFDA model. As n ¼ 1
CTDIFDA ¼ KðzÞdz: ð4:2Þ
nT 7T for the vast majority of CT scanners in use from
1975 to the late 1980s, the integration width was
For an ionization chamber that is calibrated in terms 14 times wider than the primary collimated beam,
of air kerma, the measured value is converted to so that the measurement of all of the primary radi-
absorbed dose in PMMA, using the ratio of the mass ation was intended. The limits of integration in the
energy-absorption coefficients: CTDIFDA mean that for a 3 mm beam width (T ¼
3 mm), the integration length of 14T was 42 mm;
fmen =rgmedium 98 mm for a 7 mm beam width; and 140 mm for a
fSI ¼ : ð4:3Þ
fmen =rgair 10 mm beam width. However, the variation in the
integration length (42 mm to 140 mm) implies that
For the typical energies used in CT, the value of fSI is differing extents of x-ray-scatter-tail integration
0.90 mGy/mGy to convert to absorbed dose in PMMA. would be measured with changes in T, even though
Here, fSI is the value of the so-called f-factor in units the propagation distances of scatter along the z
consistent with the International System of Units. A axis from the center of the primary beam (at z ¼ 0)
value of fSI of 1.06 mGy/mGy would be appropriate to is related more to the x-ray spectrum, phantom
convert to absorbed dose in tissue. Conversion to composition, and diameter than to the beam width
absorbed dose in PMMA is unique to CTDIFDA, and T (Boone, 2007). It is likely that the importance of
the more-recently defined CTDI quantities (discussed the role of scattered radiation in CT dosimetry was
below) are reported as air kerma, with no conversion not fully appreciated in the early years of CT
factor necessary. The original definition of CTDIFDA operation.
was to include the f-factor such that the resultant The scattering media for CTDI measurements
measure was in terms of the quantity absorbed dose; were also standardized by the FDA (1984). As men-
however, the more modern CTDI100 is explicitly tioned previously, these consist of the two PMMA
defined as a measurement of air kerma. In the follow- cylinders, 160 mm and 320 mm in diameter and
ing, the general term dose is kept because air kerma 150 mm in length. Figure 4.3 shows these two phan-
is closely related to absorbed dose in air at the x-ray toms in the measurement position.
energies used in CT.
For the CTDIFDA, the limits of integration were 4.2.4 The nT Term
expressed in terms of nT (such as +7nT), allowing Air kerma, K, is defined (ICRU, 2011) as
for the potential underestimation of the MSAD by
the CTDI. For modern MDCT scanners, such as a dEtr
K ¼ ; ð4:4Þ
16-slice scanner using nT ¼ 16 1.25 mm ¼ dm
Figure 4.3. Photographs of the CTDI head phantom and body phantom, with the cable from the pencil chamber visible. For CTDI
measurements, the cylindrical phantom is located concentrically in the gantry using laser lights for positioning. Along the z axis, the
phantom is positioned so the center of the axial (sequential) scan intersects the phantom at its midpoint. The pencil chamber is also
positioned so its midpoint is aligned with the center of the radiation beam (and hence with the phantom). In the photograph on the left,
the head phantom is placed on the table. This is a common geometry for an emergency-room CT scanner, in which many patients are
either unconscious or otherwise impaired. In most CT scanners outside of the emergency department, the head phantom is more
typically placed in the CT head holder, which is used preferably for head imaging with ambulatory patients.
39
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
where dEtr is the mean sum of the initial kinetic position (nearest to the tabletop), then this meas-
energies of all the charged particles liberated in a urement will be in general lower in value than
mass dm of a material by the uncharged particles with the pencil chamber placed in the 12 o’clock
incident on dm. position. Even though the measurement value is
Ionization chambers, including the long pencil the integral of the 2p rotation of the x-ray tube
chamber typically used for CT, are designed to around the phantom and chamber, the peripheral
produce accurate dosimetry readings only when measurement of air kerma is highly influenced by
they are completely and uniformly exposed, and this the part of the 2p scan when the x-ray source is on
is the manner in which most ionization chambers the side of the phantom at which the peripheral
are used and calibrated. When partial exposure of measurement is made. When the CT table is
the pencil chamber occurs, as in CT measurements, between the x-ray source and chamber as in the 6
a partial-volume correction is necessary. o’clock position, it will have a greater influence on
the measured air kerma than when the table is not
between the source and pencil chamber as in the
4.2.5 CTDI100
12 o’clock position (see Figure 4.2b).
The CTDI100 protocol calls for measurements to Unlike CTDIFDA, CTDI100 values are reported in
be made using a 2p rotation of the gantry in the terms of air kerma, and no conversions are per-
axial mode, with the PMMA cylindrical phantom formed to convert the air kerma to absorbed dose
placed concentrically with its center coincident in PMMA or any other medium.
with the isocenter of the CT scanner’s rotation, and CTDI100 represents the accumulated multiple-
with the center of the phantom along the z axis scan air kerma at the center of a 100 mm scan and
located at the CT scanning location (Figure 4.3) at underestimates the accumulated air kerma for
z ¼ 0. The pencil chamber is placed on the central longer scan lengths. It is thus smaller than the
axis of the phantom, and subsequently at the per- equilibrium air kerma or MSAD for most clinical
ipheral position. The measurements made under examinations, because most patient CT scans are
these conditions are referred to in this document as longer than 100 mm. The CTDI100, as does the
center peri
the CTDI100 and CTDI100 . Some head and body CTDIFDA, requires integration of the radiation dose
phantoms include multiple (e.g., 4) holes in the profile from a single axial scan over specific inte-
phantom at its periphery to allow assessment of gration limits. In the case of CTDI100, the integra-
peri
CTDI100 at various angles, and these values are tion limits are +50 mm, which corresponds to the
center
typically averaged. Because the CTDI100 meas- 100 mm length of the commercially available pencil
urement is only at the center of the phantom, and ionization chamber (Jucius and Kambic, 1980;
also at the center of the scanner’s field of view, this AAPM, 1990; 1993; McNitt-Gray, 2002). Pencil
measurement is relatively insensitive to the influ- chambers of lengths other than 100 mm have been
ence of the bow-tie filter, which is relatively homo- manufactured; however, the 100 mm version is
geneous at the center of the field of view. However, required for the CTDI100 measurement, thus
center
the CTDI100 measurement is indirectly affected ð 50 mm
by the bow-tie filter (see Section 2), because periph- 1
CTDI100 ¼ KðzÞdz: ð4:6Þ
erally scattered photons can reach the central axis nT 50 mm
and affect the measurement. The peripheral meas-
peri
urement, CTDI100 , however, interrogates the x-ray We note that historically the integrand in Eq. (4.6)
fan beam at a wider fan angle and therefore is has been written in terms of a generalized dosimet-
more sensitive to the attenuation properties of the ric quantity, such as D(z), but current protocols are
bow-tie filter. Because some CT scanners initiate specified in terms of air kerma, K(z). The use of a
the beam-on position at an arbitrary start angle, single, constant set of integration limits avoids the
peri
the peripheral measurements, CTDI100 , are typic- problem of overestimation for narrow slice widths
ally less reproducible than those for the center (e.g., ,3 mm) (AAPM, 1990) compared with
hole. CTDIFDA. It also resolves the problem of measure-
It is also noted that measurement of CTDI100 ments for a number of different scan lengths; the
requires that the phantoms be placed on the CT +7T limits for CTDIFDA were defined when mul-
table, and the measurements therefore include at- tiple thermoluminescent dosimeters (TLDs) were
tenuation of the table itself. Although the table has more commonly used, and the length of the TLD
little effect on the central measurement, it has a array could be varied. CTDI100 is typically mea-
measurable influence on the peripheral CTDI100 sured using a 100 mm long, 3 cm3 active volume
measurement; if the peripheral hole location where CT pencil ionization chamber in conjunction with
the pencil chamber is placed is in the 6 o’clock the two standard CTDI PMMA phantoms (AAPM,
40
Overview of Existing CT Dosimetry Methods
1990; FDA, 1984), using a single rotation about a overlap of the dose profiles, and the larger the
stationary patient table and phantom, i.e., in axial CTDIvol value. Because no averaging over the scan
(or sequential) acquisition mode. length L has been performed, CTDIvol (and also
CTDIw) is essentially an average over the area of
the central scan plane at z ¼ 0, with no appreciable
4.2.6 Weighted CTDI, CTDIw
z extent, namely over only a relatively small dis-
The dose in CT varies across the axial tance +b/2 about z ¼ 0 for axial scans, and is a pure
field-of-view, that is, across the plane of an individ- planar average at z ¼ 0 for helical scanning (Dixon,
ual axial CT scan. For example, for the 320 mm 2003). Thus, its name is somewhat of a misnomer.
diameter PMMA body phantom, the CTDI100 is Likewise, MSAD is a central-plane average about
typically a factor of two higher at the periphery z ¼ 0. Pitch is defined as the ratio of the table travel
than at the center of the field-of-view (average for per rotation, b, to the total active detector length
120 kV from four major manufacturers ¼ 1.99; coef- (nT) (IEC, 2009; McCollough and Zink, 1999), as in
ficient of variation ¼ 6 %); whereas for the head Eq. (2.5).
phantom (160 mm diameter), the CTDI100 is quite The CTDIvol provides a single CT dose quantity
similar at the center and peripheral locations based on a directly and easily measured quantity,
(average ratio 1.08; COV ¼ 2 %). The average which approximates the average absorbed dose (in
CTDI100 across the field-of-view at the central scan air) across the central scan plane for a 100 mm
plane is estimated by the weighted CTDI (CTDIw) scan length in a standardized (CTDI) phantom
(IEC, 1999; 2002; Leitz et al., 1995), where (IEC, 2009). CTDIvol is a useful dose metric for a
standardized phantom for a specific examination
1 center 2 peri protocol, because it takes into account the import-
CTDIw ¼ CTDI100 þ CTDI100 : ð4:7Þ
3 3 ant parameter of pitch. Its value is displayed pro-
CTDIw is intended as an average value, calculated spectively on the console of newer CT scanners,
from the CTDI100 center and peripheral measure- although it might be mislabeled as CTDIw on some
ments. The weights of 1/3 and 2/3 are from the as- older systems. The IEC consensus agreement on
sumption of a linear increase in CTDI100 with radial these definitions is used on most modern scanners
distance from the phantom central axis, and approx- (IEC, 2009).
imates the relative areas represented by the center
and edge values (Leitz et al., 1995). CTDIw is a 4.2.8 Limitations of CTDIvol
useful indicator of a CT scanner’s radiation output
(air-kerma levels measured in a specific phantom) Although CTDIvol is an attempt, given previous
for a specific “kV” and “mAs” setting. According to CTDI quantities, to better estimate the average
IEC requirements, CTDIw must use CTDI100 as absorbed dose (in air) in the central CT slice for an
described above (IEC, 2009). object of attenuation similar to that of the CTDI
phantom, it does not represent the average dose for
most patients, which differ from the phantom in
4.2.7 Volume CTDI, CTDIvol size, shape, and attenuation; moreover, the 100 mm
To represent the radiation levels (air kerma) in a integration limits omit a considerable fraction of
phantom for a specific scan protocol, which almost the scatter tails (Boone, 2007) and underestimates
always involves a series of scans, it is essential to the absorbed dose for typical body-scan lengths of
take into account any gaps or overlaps between the 250 mm or more. This is discussed at length in
x-ray beams from consecutive rotations of the x-ray Section 7. The CTDIvol as indicated on the scanner
source. This is accomplished with the use of a dose console remains unchanged whether the scan
descriptor known as the volume CTDI, or CTDIvol. length is 10 mm, 100 mm, or 1000 mm. CTDIvol
As CTDIw is the central plane (along the z axis) air estimates the dose for a 100 mm scan length only,
kerma for contiguous scans (b ¼ nT), CTDIvol con- as it is derived from the CTDI100, even though the
verts CTDIw to the central plane for an arbitrary actual central dose will increase as a function of
scan interval, b, as scan length, up to an asymptotically approached
equilibrium dose value (AAPM, 2010).
nT 1 Although CTDIvol is not an ideal direct measure-
CTDIvol ¼ CTDIw ¼ CTDIw ; ð4:8Þ ment of patient dose for the reasons discussed
b p
above, it is considered to be a useful measurement
where b is the table translation increment per axial of the pitch-corrected x-ray output properties (inte-
scan (IEC, 2009), and p is the pitch. The smaller grated air kerma for specific technique factors) of
the scan spacing relative to nT, the greater the each make and model of CT scanner. As will be
41
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
seen later in Sections 7 and 8, CTDIvol has shown the total nominal beam width. The overall dose
value in normalizing a number of quantities to the implications of over-ranging with regard to the
output characteristics of different CT scanners. dose –length product, DLP, depends on the scan
The usefulness of CTDIvol is further supported by length. For helical scans that are relatively short
the widespread availability of tools ( phantoms and (e.g., 100 mm), a scanner using a 40 mm collimated
pencil chambers) for measuring it, and on more beam width can have up to 40 % of the DLP in the
recent scanners, the IEC requires display of the over-ranging region. For a scan length of 300 mm
CTDIvol after a CT scan technique is set up, but with a 20 mm collimated beam, about 7 % of the
before it is initiated. CTDIvol is also commonly DLP is contributed by the over-ranging region.
reported after the CT scan, and more recently the Newer models of CT scanners have independent
CTDIvol is often stored as a secondary capture collimator systems (adaptive dose collimation, see
image (a bitmap image) in the patient’s CT image Section 2.2.2) that are designed to substantially
files in the Picture Archiving and Communication reduce the over-ranging dose, and these systems
System (PACS) or in the DICOM (digital image perform well in terms of reducing unnecessary dose
communication) standard. at the edges of the CT scan.
Table 4.1 shows the CTDIvol and DLP for typical
CT exams. The values are meant to be illustrative
4.3 Dose – Length Product only; they can vary by scanner model, vendor, insti-
tution, and image-quality requirements. Because
The total energy 1 deposited in the phantom by a the DLP is related to the total energy absorbed, it
given scan protocol can be estimated by multiplying depends only on the total number of rotations, and
CTDIvol by the directly irradiated mass of the is indifferent as to how these rotations are distribu-
phantom pr 2rL, where L is the scan length and r ted along z, i.e., DLP remains the same even if
is the mass density. Thus the table does not move (L ¼ 0). For a given tube-
potential setting and bow-tie filter, the DLP
1 ¼ rpr2 L CTDIvol : ð4:9Þ
depends on the product of the primary beam width,
a, and the total tube-current–time product used
Therefore, a surrogate measure of the total energy
for the entire scan series. If scan length is changed
absorbed in the phantom is the dose –length
by changing pitch without increasing the total
product (DLP) (IEC, 1999), where
“mAs” setting, DLP remains unchanged.
DLP ¼ CTDIvol L; ð4:10Þ
42
Overview of Existing CT Dosimetry Methods
Table 4.1. Illustrative values for CTDIvol and DLP for common CT exams for 4- and 16-channel MDCT.
Exam Beam Pitch Tube current time per Scan CTDIvol (mGy) DLP
collimation rotation (mA s) length (cm) (mGy cm)
Table 4.2. Effective-dose weighting factors from ICRP Table 4.3. DLP-to-E conversion coefficients, k, for various types
Publication 103 (ICRP, 2007), combined with the example of a of CT examinations. Conversion coefficients for head-and-neck
thoracic CT scan. Column 1 identifies the organ site of interest, assume the use of the 16 cm diameter CT head phantom. All
and column 2 gives the tissue-weighting factors (wT) defined in other conversion coefficients assume the use of the 32 cm
ICRP Publication 103. Column 3 represents the various organ diameter CT body phantom (Bongartz et al., 2004; Shrimpton
absorbed doses from a thoracic CT examination, determined et al., 2006).
using Monte Carlo methods. Column 4 is the product of wT and
the organ absorbed doses, resulting in the contribution to Region of body k/[mSv/(mGy cm)]
effective dose from that organ. Summing column 4 represents
the total effective dose of 9.3 mSv for this hypothetical CT Head and neck 0.0031
examination. Head 0.0021
Neck 0.0059
Organ site WT Organ dose E Chest 0.014
(mSv/mGy) (mGy) (mSv) Abdomen and pelvis 0.015
Trunk 0.015
Gonads 0.08 0.03 0.00
Bone marrow 0.12 6.70 0.80
Colon 0.12 0.24 0.03
Lung 0.12 24.00 2.88 considerations. Effective dose was never intended
Stomach 0.12 5.90 0.71 to be used for the assessment of an individual
Bladder 0.04 0.02 0.00 patient’s radiation dose, because the radiobiologic-
Breast 0.12 18.00 2.16
al weighting factors are not pertinent to a specific
Liver 0.04 9.00 0.36
Esophagus (thymus) 0.04 27.00 1.08 patient.
Thyroid 0.04 4.50 0.18 To minimize controversy over differences in
Skin 0.01 4.60 0.05 effective-dose values that are purely the result of
Bone surface 0.01 13.00 0.13 calculation methodology and data sources, a
Brain 0.01 0.20 0.00
Salivary glands (brain) 0.01 0.20 0.00
generic estimation method was proposed by the
Remainder 0.12 7.50 0.90 European Working Group for Guidelines on
Total 9.28 Quality Criteria in CT (IEC, 2009). In this method,
effective-dose values are calculated from the NRPB
dose-conversion coefficients derived from Monte
Carlo calculations of organ absorbed doses (Jones
4.4 Estimation of Effective Dose
and Shrimpton, 1991), using the ImPACT spread-
When comparing different CT scan protocols, or sheet (ImPACT, 2011). The effective doses were
when comparing absorbed-dose levels between compared with DLP values for the corresponding
x-ray radiography and CT, methods are required clinical exams to determine a set of coefficients, k,
that allow prorating the absorbed dose levels for a that are dependent only on the region of the body
generic patient. The effective dose, E, defined in being scanned (head, neck, thorax, abdomen, or
ICRP 60 (ICRP, 1991) and subsequently redefined pelvis) (see Table 4.3). Using this methodology, E
(ICRP, 2007), can be a useful tool in such compari- can be estimated from the DLP, which is reported
sons (see Table 4.2). Effective dose is not a physic- on most CT systems, according to
al dose, as its computation includes weighting
factors that are derived from radiobiological E k DLP; ð4:11Þ
43
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
Figure 4.4. The relationship between dose– length product, DLP, Figure 4.6. The DLP versus E for head CT scans. The fit shows
and the effective dose, E. The individual data points were a k coefficient of 0.0023 mSv/(mGy cm).
gathered from 46 different CT scanners. With the line
intersecting the y axis at zero, the inverse slope of the line
shown is the k coefficient for abdominal CT (k ¼ 0.017 mSv/
(mGy cm)).
44
Overview of Existing CT Dosimetry Methods
2006; Gray et al., 2005), and results of national Radiological Health. Users prescribing CT protocols
surveys, such as the NEXT study (Conway that result in doses above references values
et al., 1992) conducted by the Food and Drug should have an appropriate justification (Hart
Administration’s Center for Devices and et al., 1996).
45
Journal of the ICRU Vol 12 No 1 (2012) Report 87 doi:10.1093/jicru/nds005
Oxford University Press
5. CT X-Ray-Spectrum Characterization
evaluated graphically, interpolated among adjacent of a CT scanner HVL is typically performed infre-
data, or estimated by curve fitting of A(t). The HVL quently or not at all in most settings.
increases as the x-ray-tube potential increases, with
constant inherent þ added filtration. It is noted that
5.2.2 Aluminum-Cylinder Method
by convention, the HVL is defined in terms of air
kerma, and therefore its accurate measurement Kruger et al. (2000) described a method by which
requires that a properly calibrated air-ionization the HVL of a CT x-ray beam can be measured
chamber be used. If solid-state detectors are used, without the need for using the service mode of the
then corrections are necessary to convert the mea- system. This “ring” method uses a series of alumi-
sured values to air kerma. num cylinders of slightly different diameters that
Modern CT dosimetry is substantially informed can be nested in a concentric fashion to create a
by data produced using Monte Carlo radiation- number of different filtration thicknesses. The
transport techniques. In order to achieve desired 100 mm pencil chamber is aligned at the isocenter
levels of accuracy, methods require a good model of of the scanner with the aid of laser positioning,
the x-ray spectrum used. It has been shown that and a series of repeated axial CT scans is per-
x-ray spectrum modeling can produce adequate formed, with additional cylindrical filtration added
spectra if both the tube potential and the HVL of between measurements (see Figure 5.2a). The
the x-ray beam are known (Boone, 1986; 1988; table is positioned out of the gantry for this meas-
Duan et al., 2011; Turner et al., 2009). Therefore, urement. The ionization chamber is read out in the
for accurate CT dosimetry, characterization of the integration mode, and an axial CT scan with the
HVL of the CT scanner is necessary. HVL measure- same technique settings is repeated until all of the
ment is useful for research studies of CT beam measurements are made. The investigators used a
spectra, but can be a quality-control tool as well. number of cylinders, each with 2.0 mm thickness.
This approach was demonstrated to produce HVL
measurements of acceptable accuracy. The HVL
5.2 Methods for HVL Measurement values for 120 kV measured with the ring method
and the conventional service-mode method were
5.2.1 Conventional HVL Assessment in CT
not significantly different. The investigators also
The conventional assessment of the HVL in com- studied the influence of mispositioning the ioniza-
puted tomography requires that the service mode tion chamber away from isocenter, and found that
of the CT scanner be used in order to stop the rota- the ring technique was relatively insensitive to
tion of the x-ray tube. In this mode, the HVL meas- realistic positioning uncertainties in the field.
urement is performed as it would be for any Although the manual placement of the alumi-
radiographic system; measurements are made in a num cylinders between measurements can be time-
fixed geometry as increasing thicknesses of filter consuming, by obviating the need for the presence
material are added incrementally to determine of the service engineer during the measurements,
Kair(t) over a range of t [see Eq. (5.2)]. Figure 5.1 this approach has clear advantages over the con-
illustrates the experimental set-up for conventional ventional HVL-measurement method when service
HVL assessment. Although a number of different engineers are not routinely available.
measurement configurations are possible, in
Figure 5.1 the aluminum filters are placed directly
5.2.3 Real-Time Probes
on the gantry cowling, the x-ray tube is centered at
the 6 o’clock position, and the ionization chamber X-ray source rotation in a CT scanner creates a
is placed in the center of the field. The patient dynamic situation in which the radiation beam is
table is retracted out of the gantry so as not to directed toward the center from a circular trajec-
interfere with the measurements. A series of mea- tory. Combined with table translation along z, and
surements are then made with different thick- depending upon the experimental setup and if a
nesses of Al placed in the beam. phantom is involved, the air-kerma rate can vary
Few medical physicists have the passwords ne- rapidly and appreciably. The use of a conventional
cessary to enter the service mode of most CT scan- thimble chamber with integration periods of 0.1 s
ners, let alone the ability to use the system in precludes the measurement of the dynamic charac-
service mode in a safe and time-efficient manner. teristics of the CT x-ray beam. Therefore, the use of
Therefore, the conventional measurement of the a rapid-readout, real-time radiation meter is recom-
HVL using the service mode typically requires the mended for several classes of measurement in CT.
presence of the service engineer during the proced- For the purposes of characterizing the dynamically
ure. Because of this difficulty, the characterization changing x-ray beam in a CT scanner, a rate of
48
CT X-Ray Spectrum Characterization
Figure 5.1. The traditional measurement of the HVL for a CT scanner. (a) Serial measurements are made with the x-ray gantry held
stationary, a process typically requiring access to the service mode of the system. (b) A photograph showing a 0.6 cm3 ionization
chamber used for HVL measurements. A sheet of aluminum is positioned directly below the ionization chamber on the plastic cowling
of the scanner.
Figure 5.2. Measuring the HVL using a series of Al rings. (a) Schematic illustration of the method. (b) The concentric Al rings used for
this method.
1000 samples per second or higher is considered necessary to provide a signal measurement that is
necessary. similar or identical to air ionization chambers.
There are different types of technologies that can Although energy corrections are plausible in a
be brought to bear to achieve a high temporal free-in-air geometry for which x-ray scatter levels
bandwidth x-ray meter: (1) solid-state x-ray meters, are small, much of the utility of a real-time x-ray
and (2) air-ionization chambers with real-time dosimeter is in phantom measurements, in which
readout capabilities. Solid-state x-ray meters can scatter levels can be quite high. A large scattered-
be fabricated from a number of different detector radiation component introduces uncertainty with
materials. A simple approach is to couple a scintil- respect to the effective energy of the x-ray beam,
lating material to a photodiode using a fiberoptic confounding accurate energy-correction algorithms.
cable, assuring that the photodiode is outside of Nevertheless, it seems that real-time solid-state
the x-ray beam or is well shielded. A key require- x-ray detectors can be designed and calibrated for
ment for solid-state x-ray meters is that the x-ray use in real-time CT dosimetry.
scintillator material has a short decay constant. A real-time thimble chamber that uses air as the
Most solid-state x-ray meters demonstrate excellent x-ray detector presents a technological challenge if a
linearity as a function of air-kerma rate. One of the small detector volume is desired. The density of air is
biggest limitations of solid-state x-ray meters is three orders of magnitude smaller than many solid-
that in general they have a different response than state detectors, and therefore the electronic-signal
air and therefore energy-dependent corrections are levels generated in air-based detectors are much
49
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
smaller, leading to noisy signals. Nevertheless, the isocenter of the CT scanner using laser positioning,
use of air as the detection medium is appealing due with the patient table retracted out of the beam.
to the long history of air-ionization chambers in diag- An aluminum cage (see Figure 5.3) has been
nostic radiology applications. designed to have a number of aluminum filters
(eight in the configuration shown) of different
5.2.4 Aluminum Cage with Real-Time Probe thicknesses placed around the center of the cage in
Method which the real-time probe is positioned. One or
more axial CT scans are acquired, and the data
Most ionization chambers read out in the inte- derived from the real-time probe can then be used
gration mode, in which they integrate the charge to compute the HVL. The measurement geometry
produced over a time interval that is typically from is shown in Figure 5.4. Data generated from the
sub-seconds to several seconds. Real-time radiation real-time probe (in this case, Radcal Chamber
meters are only now becoming available, and these Model 10X6-0.6) using this technique are illu-
systems are capable of accurately measuring a radi- strated in Figure 5.5a. This signal trace is used to
ation beam of varying intensity over short periods produce the attenuation measurements described
of time ranging from 100 ms to 1000 ms. Thus, 1000 in Eq. (5.2), and is plotted in Figure 5.5b.
or more measurements per second can be made There are a number of commercially available
using real-time x-ray detector systems. real-time radiation meters available, including air-
When a real-time radiation meter is available, an ionization chambers and solid-state systems. Solid-
aluminum-cage method for measuring the HVL is state, real-time radiation meters include silicon-
feasible. Similar to the ring setup (see Section based systems as well as scintillator-based probes.
5.2.2), the real-time probe is positioned at the For the solid-state radiation meters used to deter-
mine the HVL, in general, energy-dependent correc-
tions are required.
Figure 5.4. The HVL cage illustrated in the context of a rotating CT gantry. (a) Schematic illustrating use of the cage. (b) Photograph
of a prototype HVL cage.
50
CT X-Ray Spectrum Characterization
Figure 5.5. Results from an Al-cage measurement. (a) The signal trace generated by the real-time probe in the cage geometry for HVL
assessment. This trace illustrates attenuated regions labeled a–h, with the interspersed unattenuated measurements. These data were
acquired on a General Electric VCT scanner for a tube potential of 120 kV, and with the medium bow-tie filter, using a prototype air-
ionization chamber (Radcal, Monrovia, CA, USA). (b) The attenuation curve computed from the data shown in (a). The attenuation
values corresponding to the specific signal components a–h are labeled. The dashed line corresponds to the conventional measurement
of HVL, and the solid line is that measured using the real-time air-ionization chamber and the Al cage.
51
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
anode, then the mF(E)tadd term should be the same 5.4 Typical HVL Values in CT
at all x-ray tube potentials because the inherent fil-
The HVLs typical in conventional radiography are
tration does not change within the x-ray tube as-
compared with those typical in CT in Figure 5.8;
sembly when the x-ray tube potential is changed.
measured HVLs from two scanner types are shown.
However, due to inherent inaccuracies of the spec-
Measured HVLs as a function of tube potentials for
trum model, the optimal estimated spectra f0 (E)
several CT models are provided in Table 5.1. It is
will have slightly different thicknesses tadd for dif-
clear from Figure 5.8 that the typical HVL in CT is
ferent tube potentials. Differences between actual
considerably larger than that generally used in con-
and estimated inherent þ added filtration also will
ventional radiographic imaging. The primary
occur when the precise elemental composition of
purpose for the harder beam (i.e., higher HVL) in
the filtration is not known.
CT is to reduce the absorbed dose in the patient,
and to reduce the impact of beam-hardening arti-
facts in CT images (see Section 2.4.5).
5.3.1 Off-Angle HVL Assessment HVL measurement in the clinical CT environment
can be useful as a quality-assurance tool, and is
The HVL described so far corresponds only to the
HVL along the central ray of the scanner, i.e.,
where the fan angle is zero (u ¼ 0). Thus, the influ-
ence of the bow-tie filter is not included in the HVL
assessment, as the bow-tie filter is conventionally
defined as having zero thickness at u ¼ 0. Because
the bow-tie filter adds additional x-ray absorber
material gradually toward the periphery of the fan
angle, the beam will be increasingly hardened by
this additional filter, and the HVL will increase. If
there were interest in measuring the off-angle
HVL, then the only direct method would be the
conventional method outlined earlier. Both the
Al-ring and the Al-cage methods outlined above are
designed to measure the HVL along the central ray
of the scanner, and they cannot measure the
off-angle HVL. An indirect method for estimating
the off-angle HVL is to perform the assessment at
the center of the field as described above, to use the
known bow-tie filter-thickness distribution to Figure 5.8. Typical values of the HVLs used in CT and
radiography. Measured HVL values for two commercial CT
further harden the beam using appropriate attenu-
scanners are shown. For the radiography curve, the HVLs were
ation coefficients, and then to compute the HVL. modeled (Boone and Seibert, 1997) using a total of 2.0 mm of
Estimation of the bow-tie thickness as a function of added filtration (lower solid line), and for the CT curve a total
angle will be described in Section 6. added filtration of 10.0 mm (upper solid line).
Figure 5.7. X-ray spectra generated using a spectrum model, the known tube potential, and the measured HVL. (a) A 120 kV spectrum,
which has a small amount of inherent filtration, resulting in a measured HVL of 5.1 mm of Al. (b) A 120 kV spectrum for an x-ray
system with more inherent filtration, resulting in a harder x-ray spectrum with a measured HVL of 9.0 mm of Al.
52
CT X-Ray Spectrum Characterization
Table 5.1 Typical values of the HVL in mm AL for several measured in a relatively easy and rapid manner, it
commercially available CT scanners (Mathieu et al., 2010). can and should become a useful parameter for moni-
These were measured at the isocenter for the “large” or “body”
toring beam quality as a part of a quality-assurance
bow-tie filter
program in CT, just as it is in radiography, fluoros-
X-ray GE VCT GE Siemens Toshiba copy, and mammography settings.
tube Lightspeed-16 Sensation 64 Aquilliona In a research setting, the estimated spectrum
potential f0 (E) determined for a given tube potential and
measured HVL can be used in Monte Carlo simu-
80 kV 5.4 5.9 6.2 4.6 lations to estimate organ dose or other dosimetric
100 kV 6.6 7.2 7.6 5.8
120 kV 7.7 8.3 8.7 6.7
quantities. For CT dosimetry, and in particular in
135 kV — — — 7.4 Monte Carlo based dosimetry studies, the use of
140 kV 8.6 9.2 9.7 — accurate x-ray-beam spectra is an important con-
sideration in the assessment of absorbed dose in
a
Premier One models. patients, phantoms, and in air. Moreover, as the
number of scanner models, typical tube poten-
likely to be of value as a parameter during scanner tials, and acquisition protocols continue to in-
acceptance-testing. The value of the HVL for a crease, it might become necessary to generate
given tube potential is an indicator of beam quality dosimetry tables based on generic beam-quality
(often used a surrogate for effective energy), which parameters, e.g., tube potential and HVL. This
should remain relatively stable over time and after approach is common in mammography but not
x-ray-tube changes. Thus, if the HVL can be yet in CT.
53
Journal of the ICRU Vol 12 No 1 (2012) Report 87 doi:10.1093/jicru/nds006
Oxford University Press
explained by straightforward physical and geomet- Glenwood, IL, USA) to provide for the measure-
rical principles. ment of fair(z) in CT. These OSL systems have ex-
cellent spatial resolution and are well-suited for
qualitative measurement of air-kerma profiles
6.3 Measurement of fair(z)
free-in-air along the z axis. The system uses a long
6.3.1 Optically Stimulated Luminescence strip of AlO2 film inside a light-tight plastic cylin-
Systems der that is 150 mm long and 10 mm in diameter. To
measure fair(z) directly, the dosimeter is placed in
Arrays of optically stimulated luminescence
air at isocenter; it is also designed to fit within a
(OSL) dosimeters were developed by a commercial
phantom. The OSL dosimeter is exposed under
manufacturer (CT Dosimeter, Landauer, Inc.,
typical CT operating conditions in the axial mode,
and is then sent back to the manufacturer to be
analyzed. The manufacturer’s reader uses a
0.2 mm slit width that samples every 0.05 mm and
uses a 20-point smoothing average to generate
output values. From the resulting profiles (see
Figure 6.4), one can obtain the raw values for
fair(z). In addition, derived parameters such as the
full-width at half-maximum can be calculated,
which can be compared with the nominal x-ray-
beam collimation to determine the extent of over-
beaming (see, e.g., Liu et al., 2005). Although the
energy response of the OSL dosimeter is reason-
ably constant over the x-ray energies involved, it is
not the same as that of ionization chambers.
Figure 6.1. The components of a modern CT scanner tube However, for OSL measurements made free-in-air
assembly. X rays emanating from the tungsten anode pass
at the isocenter of the CT scanner, the relative in-
through inherent þ added filtration, and are then attenuated by
the beam-shaping filter. Finally, collimator blades truncate the tensity of the x-ray-beam profile is of primary inter-
beam in the – z and þ z directions. est and not the absolute values; in this case, a
Figure 6.2. X-ray-tube geometry in a CT scanner. (a) The x-ray tube in a modern CT scanner rotates in an orientation such that the
plane of the anode is parallel to the plane of gantry rotation. (b) The geometry of the electron beam striking the anode. As the x-ray-
tube potential increases, the kinetic energy of the electrons incident on the anode increases, and so does the average depth of
penetration (DoP). As x rays are produced within the anode, they are attenuated by different thicknesses of anode material along the z
axis of the beam, with greater thickness on the anode side of the field (giving rise to the heel effect). This basic geometry can be used to
derive the theoretical shape of the air-kerma profile along the z axis.
56
CT Output Characteristics Measured in Air
57
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
Figure 6.4. Relative profiles, fair(z), measured for a Toshiba CT scanner operating at 120 kV using OSL detectors. The one-dimensional
view of the x-ray beam intensity distribution along the z axis is shown for: (a) a 32 mm nominal collimation setting, and (b) a 128 mm
nominal collimation setting.
extended out in front of the table, so the table pro- where FT is the Fourier transform, FT21 is the
vided mechanical translation but was not in the inverse Fourier transform, sinc(lprobe/n) is the
x-ray beam. The fair(z) measured with the real-time Fourier transform of a rect function of length lprobe,
probe relies on stable table velocity. The raw meas- and n is the frequency. Here the length, lprobe, is
urement of fair(z) using this approach also includes the active length of the chamber. When the same
the influence of the length (and sensitivity profile) real-time instrument is used, e.g., in annual x-ray
of the real-time probe. Let the true profile be fair(z) beam comparisons, serial comparisons among the
m
and the response function of the real-time probe be uncorrected profiles, fair ðzÞ, would be adequate for
m consistency checking.
q(z), then the measured profile fair ðzÞis given by
Ð1
m
fair ðzÞ ¼ 1 fair ðz0 Þ qðz z0 Þ dz0 : ð6:1Þ 6.3.6 Summary of fair(z) Measurements
The fair(z) in CT can be measured using a
number of different technologies with adequate-to-
If q(z) is small in z extent in comparison to fair(z),
m excellent results. There are tradeoffs with each
then fair ðzÞ ffi fair ðzÞ. In the special but realistic case
method, and it is ultimately up to the CT scientist
that q(z) is a rect function, the influence of the
or medical physicist to determine which approach
probe length, lprobe, can be removed by deconvolu-
is most appropriate for their specific application
tion in the frequency domain:
and institution. It is observed that with film proces-
m
sors rapidly becoming less available in the hospital
1 FT½ fair ðzÞ setting, the film-sensitometric approach will be
fair ðzÞ ¼ FT ; ð6:2Þ
sincðlprobe =nÞ more difficult to implement in coming years. CR is
58
CT Output Characteristics Measured in Air
Figure 6.7. Relative profile, fair(z), measured for a General Electric VCT scanner at 120 kV using a CR detector system. An image of the
profile is shown in the left panel, and the profile taken through this is plotted in the right panel. These data were not corrected for the
response of the CR plate.
59
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
and the necessary signal processing can be imple- (where u ¼ 0) the bow-tie filter has no effect on
mented using pre-programmed spreadsheet soft- fA(u), but there is still considerable x-ray beam fil-
ware. Although real-time radiation detectors are not tration material in the tube assembly.
widely available at this time, it is anticipated that Measuring fA(u) can be difficult, however,
as CT dosimetry progresses the use of real-time because the location of the x-ray source is not
dosimeters will become commonplace. obvious due to the CT-scanner cowling and because
physical positioning of the detector along a radius
of curvature is challenging. For most experimental
6.4 Measurement of fL(x) or fA(u) settings, therefore, a measurement of the air-
kerma profile along a line running through the iso-
6.4.1 Measurement of fL(x)
center is more convenient. The results of such mea-
The function fL(x) represents a linear measure- surements are related to fA(u), and are referred to
ment of air kerma across the field of view, and the here as fL(x), where the (horizontal) x axis is or-
function fA(u) represents an angular measurement thogonal to the central ray (where u ¼ 0) and gen-
of air kerma. The angle u is defined here as the erally runs through isocenter. Due to the rotational
angle from the central ray of the beam in the trans- symmetry in CT, it is recognized that measure-
verse x –y plane, such that at the center of the field ments along the (vertical) y axis are the same as
u ¼ 0. This angle is called the fan angle. those made along the (horizontal) x axis, as long as
Evaluation of the beam profile as a function of the the measurement is made orthogonal to the central
fan angle essentially characterizes the angle- ray. In all cases, however, this profile will be re-
dependent attenuation of the bow-tie filter that the ferred to here as fL(x).
scanner employs. This can also be called the bow- Turner et al. (2009) described a method for using
tie filter function. Beam-profile data are useful for the vertical adjustment capabilities of the CT couch
two primary reasons: (1) accurate CT dosimetry is (with an additional vertical stand) to make mea-
increasingly dependent upon Monte Carlo compu- surements along a line running through the isocen-
tations of dose-conversation coefficients, and the ter of the CT system, and this technique is
characterization of the bow-tie-filter function is es- illustrated in Figure 6.10. Ideally, this measure-
sential for accurate computer modeling in CT. For ment of fL(x) would be performed with a relatively
users of CT Monte Carlo approaches, the data small device (such as a small-diameter thimble
derived from a measurement of the air kerma as a chamber). In practice, Turner et al. (2009) used a
function of fan angle for a variety of CT scanners 100 mm pencil ionization chamber placed in air
are extremely useful. (2) For the clinical practition- and positioned perpendicular to the scan plane so
er, the fL(x) and fA(u) provide characterization of that only a portion of the chamber’s length was
one of the two dimensions of the air-kerma distri- irradiated. This method also requires that the CT
bution from x rays incident upon the patient. gantry be in a fixed, non-rotating position; a
Combined with fair(z), characterization of fA(u) (or 3 o’clock position was used with the central ray of
fL(x)) provides a two-dimensional understanding of
the x-ray output properties.
Because at least two bow-tie filters (body and
head) are used on most CT scanners and because of
the different beam spectra that are possible, the
measurement of all combinations can require 8, 12,
or more measurements for each CT scanner. Thus,
a simple method for performing the beam profile
measurement is desirable. The profile fA(u) requires
that the air kerma (or air-kerma rate) be measured
on a radius of curvature centered at the x-ray
source, such that beam divergence due to the
inverse-square law and solid-angle effects are fac-
tored out. That is, fA(u) should represent the
angular dependence of the air-kerma profile; fA(u)
profiles measured at different radii from the x-ray
source should vary only by a scalar constant, and
Figure 6.10. The experimental geometry used to characterize
thus fA(u) profiles normalized to unity at u ¼ 0 will the beam profile in the x direction. The position of the ionization
be independent of the distance from the x-ray chamber is moved vertically using the CT tabletop, with the
source. Note that at the center of the fan beam stationary x-ray source positioned in the 3 o’clock position.
60
CT Output Characteristics Measured in Air
Figure 6.11. Linear profiles for small and large bow-tie filters.
These results for fL(x) indicate the shape of the bow-tie filter Figure 6.12. The x-ray-beam profile, fL(x), for a General Electric
function as a function of linear distance across the field of view, Lightspeed 16 scanner measured with a pencil chamber. The
and these data include the 1/r fall-off as discussed in the text. bow-tie filter function for both the head and body filters is
shown, along with no filter (Huda, personal communication).
The use of the pencil chamber results in a 1/r fall-off.
61
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
This approach uses CR plates that are commonly train, at u ¼ 0, 2p, 4p, . . ., at which the x-ray tube
available at many medical centers. The gray scale is closest to the probe, maximizing the intensity in
produced by CR systems is not a linear function of regard to the inverse-square law and minimizing
the air kerma, and therefore a calibration proced- the influence of the bow-tie filter. The other point
ure is necessary if absolute values are required. in the rotation of the gantry at which the bow-tie
Evaluation of the characteristic curve of the CR filter has no effect is when u ¼ p, 3p, 5p, . . ., at
plate is a straightforward process (Lui et al., 2005). which the x-ray tube is on the opposite side of iso-
center from the x-ray probe.
Image-processing techniques are used
6.4.2 Measurement of fA(u)
(McKenney et al., 2011) to identify the peaks at
McKenney et al. (2011) report a method that uses even multiples of p in the signal trace (gray line in
a single radiation detector with real-time readout Figure 6.15), and the odd multiples of p are identi-
to measure fA(u) directly. The real-time detector fied as they are midway between adjacent maxima.
was placed in air, near the periphery of the CT From these two values of the signal amplitude
field of view, as shown in Figure 6.14. No other (averaged over several gantry rotations), the effect
object was placed in the x-ray beam, and the table of the inverse-square law in the absence of the
was retracted. The air-kerma rates were recorded bow-tie filter’s influence can be deduced, and this is
in real time (1 kHz) during CT acquisitions at shown as the black line in Figure 6.15. The align-
specified technique factors (tube potential, tube- ment of the odd-p and even-p locations along the
current – time product, etc.) over several gantry time-domain signal train allows the gantry angle
rotations; the results are shown in Figure 6.15. In and the fan angle to be deduced as a function of
this method, as the gantry rotates, the air-kerma time as well. Using the data given in Figure 6.15
rate is recorded at the location of the probe. The along with a well-defined mathematical construct
air-kerma rate changes due to the varying angle u (McKenney et al., 2011), the bow-tie filter function
through the bow-tie filter, and also because the dis- can be determined. The resultant bow-tie filter
tance between the probe and the x-ray tube function, fA(u), is shown in Figure 6.16 for four
changes, resulting in a change of intensity due to x-ray-tube potentials. This procedure does not
the inverse-square law. Referring to the geometry require precise probe positioning because the dis-
shown in Figure 6.14, there are two points in the tance between the x-ray probe and the isocenter
2p rotation of the gantry at which the bow-tie filter can be determined mathematically from the mea-
has no effect (i.e., at a fan angle, u ¼ 0), and in sured data. Thus, probe positioning does not
both cases this occurs when the x-ray tube, the iso- require careful physical alignment; the probe can
center, and the x-ray probe are co-linear. The first simply be placed at a convenient location near the
point corresponds to the maxima in the signal edge of the scanner’s field of view. The probe,
Figure 6.14. The measurement geometry for the characterization-of-beam-relative-attenuation (COBRA) method (McKenney et al.,
2011). Angles are defined in (a), and dimensions are defined in (b). A small ionization probe is placed completely within the x-ray beam
on the z axis, toward the periphery of the scanner’s field of view. The table is retracted from the field of view for this measurement.
62
CT Output Characteristics Measured in Air
63
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
64
CT Output Characteristics Measured in Air
Figure 6.19. Use of radiochromic self-developing film to determine the relative two-dimensional distribution. A photograph of the
self-developing film exposed at the isocenter of the CT scanner is shown in (a). The analog film was digitized, and the red channel was
used with appropriate characteristic-curve normalization to determine the relative two-dimensional radiation-intensity distribution,
shown in (b). An isometric plot of the distribution is shown in (c).
65
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
accurate air-kerma value is ambiguous because the Figure 6.8, both the spatial profile and the abso-
edges of the beam are ill-defined, and also because lute air kerma can be determined using a
the x-ray beam is inhomogeneous along z. Both of minimum of hardware: just a laptop computer,
these effects make the 100 mm pencil chamber an electrometer, and the real-time ionization
ill-suited tool for characterizing the absolute air chamber. The fair(z) measurement using this hard-
kerma for a CT x-ray tube head. ware is a robust characterization of the x-ray-tube
Given the above considerations, long-term output, and is likely to be all that is required for
recommendations assume that real-time air- CT-scanner output consistency measurements over
ionization chambers will eventually become time. Of course, complete measurement would
routine measurement instruments in the CT envir- include evaluation at all tube potentials, for all
onment. Using a small ionization chamber that is bow-tie filters, and for all clinically useful beam-
translated through the beam, as illustrated in collimation settings.
66
Journal of the ICRU Vol 12 No 1 (2012) Report 87 doi:10.1093/jicru/nds007
Oxford University Press
7. CT Dosimetry in Phantoms
The standard 160 mm and 320 mm diameter, most circumstances, a cylindrical phantom is used
150 mm long PMMA phantoms have been used for because it has rotational symmetry and is easier to
40 y in CT dosimetry, and serve as the basis of CT fabricate than more complicated shapes. Figure 7.1
dosimetric quantities such as CTDIvol, as discussed illustrates the basic geometry of a cylindrical
in Section 4. With the increasing sophistication of phantom with a diameter dphantom and a length
CT acquisition modes, as well as an interest in Lphantom. Different investigators have studied CT
better characterizing the absorbed dose arising from scanners with a large number of beam widths in the
the long scatter paths in CT, there is a need for a z dimension, and the parameter L in Figure 7.1
longer phantom. In the previous Section, character- represents the length of the phantom irradiated dir-
ization of the x-ray beam in air was discussed. ectly by the primary beam (scan length for axial or
When a phantom is introduced into the x-ray beam helical scans). Even though a number of custom CT
in CT, it becomes a major source of scattered radi- phantoms have been described in the literature, it is
ation that is responsible for absorbed dose in parts a common to incorporate holes that allow the inser-
of the phantom (and patient) away from the colli- tion of a radiation meter at the center or the periph-
mated primary beam. Hence, phantoms are an es- ery of the phantom.
sential tool in understanding the distribution of Figure 7.2 illustrates a single x-ray projection
absorbed dose in CT. through a 300 mm diameter cylindrical water
phantom, involving no x-ray-tube rotation around
the phantom. In this numerical simulation, the
7.1 Axial Dose Profiles in Phantoms x-ray beam from the General Electric VCT scanner
was modeled mathematically. The inset in
The axial or sequential CT scan involves the rota-
Figure 7.2 shows the primary x-ray beam in the
tion of the x-ray-tube head around the patient or
axial plane, and the individual profiles run along
phantom with no table motion during rotation. In
the z axis at the points marked a through e in the
this basic mode of operation, it is instructive to
figure. These data are useful to better understand
understand the extent and shape of the absorbed-
what happens to the primary x-ray beam as it is
dose profile along the z axis. It is realized that for
axial scanning, the absorbed dose in the patient is es- attenuated by the phantom. The simulation involves
sentially a summation of the absorbed-dose distribu- primary radiation only; the entrance-beam profile
tions from individual axial scans spaced at equal (curve a) is narrow because it strikes the phantom
intervals along z. Those profiles, which can be mea- at a position between the isocenter and the x-ray
sured as air kerma or absorbed dose, can be com- source where there is less beam divergence. As the
puted using Monte Carlo techniques or can be x-ray beam penetrates the phantom, it is both atte-
measured in physical phantoms using a number of nuated and diverging, hence the subsequent profiles
different detector systems. In many cases, only the are lower in amplitude and wider. These data are
relative shape of the absorbed-dose profile is of inter- from a single projection, but the x-ray beam profile
est, and the amplitude is arbitrarily normalized. In along the z dimension in an axial CT scan repre-
such cases, the quantity used to describe the profile sents the rotational summation of the data illu-
shape can be either air kerma or absorbed dose, and strated in Figure 7.2. At the isocenter (curve c), the
the generic term profile is used. For the purpose of dose distribution integrated over a 2p rotation of
absolute dose assessment, either the quantities of air the gantry will have a shape identical to curve c, but
kerma or absorbed dose are reported. away from isocenter the dose profiles will be the in-
The shape, size, and anatomical complexity of the tegral over angle (and hence depth of interaction) of
patient tend to complicate an understanding of the the different x-ray-beam profiles.
absorbed-dose distribution; therefore, it is routine to Monte Carlo simulations were performed using
use simpler phantom shapes for this assessment. In a number of cylindrical phantoms with different
68
CT Dosimetry in Phantoms
Figure 7.3. The DSF for infinitely long phantoms and a very narrow primary incident x-ray beam (0.01 mm). (a) DSFs for three regions
in the phantom, R1, R2, and R3, illustrated in the inset. These data are for a tube potential of 120 kV and a 30 cm diameter water
phantom using a General Electric body bow-tie filter. (b) The influence on DSF of phantom composition is illustrated for a 120 kV scan
in a 40 cm diameter phantom with a Siemens body bow-tie filter. The profile data are for region R1. (c) DSF profiles are illustrated for
phantoms of different diameter, for a 120 kV scan in a water phantom using a General Electric body bow-tie filter. The data show the
profiles for region R1. (d) DSFs are shown for two different beam spectra, showing greater scattering for higher-energy spectra. These
data are for region R1, using a General Electric body bow-tie filter, and a 30 cm diameter water phantom.
Figure 7.4. Axial absorbed-dose profiles measured by Nakonechny et al. (2005). Profiles at various primary-beam scan widths are
shown for (a) a Philips PQ5000 system, a single-detector array fourth-generation scanner, and (b) a Philips MX8000 Quad system, a
four-detector-array, third-generation CT scanner. The profiles were measured using a diamond-detector system (PTW, Freiberg,
Germany).
A Siemens Sensation 64 CT scanner was used to head phantom (160 mm diameter) are shown in
produce the absorbed-dose profiles given in Figure 7.5a for both the central and peripheral
Figure 7.5. Absorbed-dose profiles for the PMMA regions. Absorbed-dose profiles for the PMMA body
69
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
Figure 7.5. Absorbed-dose profiles for a Siemens Sensation 64 scanner. (a) for the 160 mm head, and (b) the 320 mm body PMMA
phantoms. The center and peripheral profiles are shown in both cases. These measurements were performed using a commercially
available array of optically stimulated luminescence detectors (McNitt-Gray et al., 2008). The curves were normalized to a maximum of
unity for the periphery measurement.
70
CT Dosimetry in Phantoms
Figure 7.7. Absorbed-dose profiles for a number of different CT beam widths for a Toshiba 256-channel CT scanner. These profiles were
measured in the axial (cone-beam) scanning mode, using PIN-photodiode sensors. (a) Profiles for a 320 mm PMMA body phantom at
the center, and (b) on the peripheral axis of the phantom (Mori et al., 2005), using a 900 mm long phantom. (c) The Mori et al. data
were fit using analytical methods (Dixon and Boone, 2011) with good results shown here.
Geleijns et al. (2009) evaluated the absorbed-dose measurements and Monte Carlo calculations were
profiles in a wide-cone-beam CT system. Figure 7.8 used to estimate the profile. The data shown in
gives the normalized absorbed-dose profile in a Figure 7.8 are from Monte Carlo simulations for the
320 mm diameter PMMA phantom for a Toshiba x-ray dose profiles at the center and periphery of the
320 cone-beam CT scanner, which has a nominal phantom. In the center, scatter tails at 10 %
beam width of 160 mm. In this study, both maximum approach +140 mm, for a full width of
71
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
Figure 7.8. Absorbed dose from Monte Carlo computations for a Figure 7.9. Normalized absorbed dose as a function of z for a
Toshiba 320-detector CT scanner with a body bow-tie filter. number of different scan lengths: 10 mm, 50 mm, 100 mm,
These profiles were computed (Geleijns et al., 2009) using a 150 mm, 200 mm, 300 mm, 400 mm, 500 mm, and 600 mm (from
nominal 160 mm beam width. The periphery and center profiles center to edge on the graph). These data were derived by
are illustrated for a 320 mm diameter, 350 mm long PMMA convolving the DSF computed from Monte Carlo simulations
phantom at 120 kV. with rect functions characterizing the length of the scan (i.e., L
in Figure 7.1), for a 320 mm diameter PMMA phantom at
120 kV, assuming a General Electric Lightspeed-16 body bow-tie
280 mm. Subtracting the primary-beam width of ap- filter.
proximately 150 mm, the scatter tails are seen to
have a full width of about 130 mm at the 10 % level. necessary (Dixon, 2003), is represented by a convolu-
Figures 7.3 through 7.8 illustrate CT absorbed-dose tion
Q z of the axial dose profile with a rect function,
profiles for a simple axial rotation from a number of L . When tube-current-modulation schemes are
different CT scanners and measurement conditions, used, the convolution is invalid and a superposition
performed by a number of different investigators. of profiles of varying amplitude is required.
These include both measured and Monte Carlo Figure 7.9 shows normalized cumulative absorbed-
derived profiles, using a number of different phan- dose distributions for a series of helical CT scans of
toms. The x-ray-primary-beam width ranged from a different scan length, produced by Monte Carlo simu-
0.01 mm nominal beam width (in a simulation) to a lation (Boone et al., 2009). These data were calculated
3 mm nominal width for a single-detector-array CT for a General Electric Lightspeed-16 system with a
system, and up to a 180 mm beam width for a full- tube potential of 120 kV and an infinitely long
cone-beam CT system. Despite the wide range of para- 320 mm diameter PMMA phantom. The scan lengths
meters studied in these examples, the absorbed-dose range from 100 mm to 600 mm in Figure 7.9. Note
profile from scattered radiation at the center along the that the amplitude of the cumulative absorbed-dose
z axis of the phantom averaged 154 mm in width at distribution increases as the scan length increases,
10 % amplitude (standard deviation of 33 mm). The even though the primary-beam contribution is the
primary-beam width influences the scatter-tail width same for all curves. The increase in the absorbed-
in a complicated fashion (i.e., a mathematical convolu- dose amplitude at the center of the scan length (z ¼
tion, as discussed below), and simple subtraction of 0) as a function of scan length, L, is due to the incre-
the beam width does not take these influences fully mental contributions of the long-range, low-
into account. This approach is nevertheless useful to amplitude scatter tails that were seen in Figures 7.3
illustrate the significant absorbed-dose profiles that through 7.8, accumulating at z ¼ 0.
result from scattered radiation in a phantom. These The cumulative absorbed-dose distributions from
observations also point to the need for a phantom Nakonechny et al. (2005) are shown for a number
longer than 150 mm if the scattered radiation dose is of different scan lengths, L, in Figure 7.10. These
to be characterized accurately. profiles were produced using a computer-modeled
single-scan absorbed-dose profile measured on a
Picker PQ5000 CT system, in an elliptical water-
7.2 Cumulative Absorbed-Dose Distributions
equivalent phantom. The profiles were computed
for Helical Scans
by convolving the axial absorbed-dose
Q z distributions,
For the typical whole-body CT scanner, helical f(z), with rect functions L for different scan
(spiral) acquisition techniques are commonly used to lengths L. The mathematical description of this
scan the patient, especially in the chest, abdomen, process is discussed in Section 7.3.2. Nakonechny
and pelvic regions. For a helical CT scan, the accumu- et al. note that the asymptote to the equilibrium
lated absorbed-dose distribution at the center of the dose, Deq, is reached in this phantom for a scan
scan length (from –L/2 to þL/2), smoothed as length .370 mm.
72
CT Dosimetry in Phantoms
73
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
given DSF and a relatively narrow collimated states that the absorbed-dose profile at the center of
primary beam (as in helical CT scanners), the the phantom for a single gantry rotation and no
absorbed dose at the center of the field of view table motion is the convolution of the DSF(z) and
along z increases as the scan length increases. the collimated beam width a.
Absorbed-dose profiles along the z axis for a number
of different scan lengths are shown in Figure 7.9. As
can be seen, the absorbed dose, DL(0), at z ¼0 along 7.3.2 Cumulative Absorbed-Dose
the dashed vertical line increases as the scan length Distribution, DL(z), for Multiple
L increases. As L increases, however, the absorbed Rotations Covering a Scan Length L
dose at the center of the scan will at some point The cumulative absorbed-dose distribution, DL(z),
reach an asymptotic limit, and this is the so-called from a helical scan of length L with a table advance
equilibrium dose, Deq. per gantry rotation of b, as illustrated in Figures 7.9
and 7.10, is described by the convolution
ð
7.3.1 Dose Profile for a Single Axial Rotation 1 þ1 Y z0
DL ðzÞ ¼ f ðz z0 Þ dz0
Consider a DSF generated using a very thin b 1 L
(0.010 mm) primary-beam impulse [approximating 1 Q z
¼ f ðzÞ : ð7:3Þ
a Dirac delta function, d(z)] projected through the z b L
collimator to the central axis giving a width a for
the primary beam full width at half maximum Equation 7.3 applies only in the absence of tube-
(FWHM). Then an analytical function describing current modulation during the scan (i.e., is valid for
the complete absorbed-dose profile, f(z), resulting constant x-ray-tube output). The convolution given
from a single axial rotation can be obtained by con- in Eq. (7.3) can be reduced to a simple integration:
volving the dose-spread function DSF ¼ d(z) þ h ð þL=2
dsf(z) obtained from Monte Carlo simulation with 1
DL ðzÞ ¼ f ðz z0 Þ dz0 : ð7:4Þ
the primary-beam intensity profile (Dixon and b L=2
Boone, 2010; 2011). Note that the upper case dose-
spread function, DSF(z), includes both the primary Evaluating this function at z ¼ 0, Eq. (7.4) further
component, d(z), and the scattered component, simplifies to
dsf(z), of the dose-spread function. The axial scan ð þL=2
profile is the convolution of the DSF and a rect func- 1
DL ð0Þ ¼ f ðz0 Þ dz0 ¼ p CTDIL ; ð7:5aÞ
tion of width a: b L=2
z
f ðzÞ ¼ DSFðzÞ A0 P where we recall that pitch is defined as p ¼ b/nT.
a This equation reduces further to CTDIL for a
z
¼ ½dðzÞ þ h dsf ð zÞ A0 P ; ð7:1Þ table advance of b ¼ nT (where p ¼ b/nT¼1). The
a dashed vertical line at z ¼ 0 in Figure 7.9 corre-
sponds to DL(0); this maximum absorbed-dose
where, assuming a point focal spot, d(z) replicates
value as a function of scan length L is shown in
the rectangular primary-beam profile
Figure 7.13. The gradual increase in absorbed dose
z as a function of scan length becomes asymptotic at
A0 P ;
a the larger scan lengths. This trend has been re-
ferred to as the rise-to-dose-equilibrium curve
with A0 the primary-beam amplitude, and h the (Dixon and Boone, 2010). To express the explicit de-
scatter-to-primary ratio (SPR) on that axis. pendence of DL(0) on L, the function h(L) is defined
Replacing d(z) by the focal-spot-intensity distribu- as h(L) ¼ DL(0); thus, these functions are mathem-
tion projected through the collimators produces a atical synonyms. Recognizing this, Eq. (7.5a) can be
more realistic primary-beam function (Dixon et al., written as
2005) without affecting the scatter component in
ð þL=2
Eq. (7.1). Equation 7.1 can be simplified to 1
hðLÞ ¼ f ðz0 Þ dz0 : ð7:5bÞ
Ð þa=2 0 0
b L=2
f ðzÞ ¼ a=2 DSFðz z Þ dz : ð7:2Þ
This formalism includes also the simplification that These and related curves have been reported by
the primary beam in air
Q can
z be described by a rect a number of different investigators. If the cumula-
function of width a, a . Equation 7.2 simply tive absorbed dose at z ¼ 0 is normalized to Deq, it
74
CT Dosimetry in Phantoms
1 h
HðLÞ ¼ þ ð1 lÞð1 eL=d1 Þ þ lð1 eL=d2 Þ :
1þh 1þh
ð7:9Þ
75
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
Figure 7.14. H(L) curves computed by convolution with measured dose profiles (Nakonechny et al., 2005). These curves show the trend
for (a) the single-detector array Philips PQ5000, and for (b) the four-detector array Philips MX8000. A 300 mm long, water-equivalent
elliptical (200 mm 300 mm) phantom was used in these measurements for a tube potential of 120 kV.
Figure 7.15. Graphs adapted from Mori et al. (2005) showing measured H(L) curves. These data are for a 900 mm long, 320 mm
diameter PMMA phantom, for a tube potential of 120 kV. Three different beam-collimation widths are shown in each plot, for (a) the
center, and (b) the periphery positions.
position in the 320 mm diameter PMMA 7.4 Phantom Design and Usage
phantom, the efficiency of measuring dose over a
Mori et al. (2005) studied the dependence of the
length of 100 mm (i.e., the CTDI100 methodology)
absorbed-dose profile on phantom length, in which
was approximately 62 % (see Figure 7.17) com-
he compared the absorbed-dose profiles in a
pared with an infinitely long measurement (or
phantom of length 140 mm with that in a 900 mm
CTDI1). Evaluating this in the context of the
long phantom. These results are shown in
analytical development above, the ordinate value
Figure 7.18 for the central axis. Near the center of
at a beam width of 100 mm is approximately 62
the scans, close to the primary beam, there is little
% in Figure 7.16a (see the dashed line with
difference between the two profiles; however,
arrow), in excellent agreement with the 62 %
farther from the center, the relative dose profiles
value determined independently.
diverge. These data suggest that a 140 mm long
76
CT Dosimetry in Phantoms
Figure 7.16. H(a) curves for wide-cone-beam CT. The solid circles are the data from Mori et al. (2005), as seen in Figure 7.15. The solid
line was derived using Eq. (7.9) (Dixon and Boone, 2010), with shape parameters from Monte Carlo determined DSFs similar to those
shown in Figure 7.3. (a) The H(a) curve for the center of an essentially infinitely long phantom, and (b) the corresponding curve at the
periphery. There was no curve fitting per se in these comparisons; the fit parameters for the center profile correspond to a 1/e scatter
length of 117 mm (d1), and a scatter-to-primary ratio of 13.
77
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
Figure 7.19. The ICRU/AAPM phantom design. The ICRU, in collaboration with AAPM Task Group 200, defined a polyethylene
phantom for the purposes of measuring the rise-to-dose-equilibrium function, h(L) or H(L). (a) The phantom design, and (b) a
photograph of the phantom. Due to the large mass of this large phantom, it was designed to be modular, with three different sections.
picture of an actual phantom. The phantom com- facilities of CT-scanner manufacturers, this
prises high-density polyethylene (mass density of phantom is practical. The phantom design repre-
0.97 g/cm3) and is 300 mm in diameter. The total sents a reasonable compromise between utility and
length of the phantom is 600 mm. Due to the large cost; it is long enough that accurate measurements
mass of the phantom (41 kg), it was necessary to of the equilibrium dose can be made, and—because
manufacture it in three separate sections. Because the stock material comes from extruded polyethyl-
a radiation detector is to be located at the center of ene boules—it is relatively inexpensive. Several of
the phantom along the z dimension, it was neces- these phantoms have been manufactured and dis-
sary to design the phantom without a seam at the tributed internationally, and some results mea-
center (z ¼ 0) to avoid the potential streaming of sured with the phantom are reported below.
primary radiation that could give erroneously high
readings. In order to keep the three sections
aligned, alignment pegs were included in the
7.5 AAPM Report 111 Recommendations
design. The three sections of the phantom are re-
for Assessment of H(L)
ferred to as section A, B, and C. Sections A and B
are bored through with holes necessary for place- The measurement of the rise-to-dose-equilibrium
ment of dosimeters (e.g., small thimble chambers) curve, H(L), is discussed at length in AAPM Report
at the center, periphery, and at an intermediate pos- 111 (AAPM, 2010) with the methodology described
ition between the center and edge of the phantom. in that report for estimating H(L) as indicated sche-
Figure 7.20 shows a more complete technical matically in Figure 7.21. A short thimble chamber,
drawing of the phantom. connected to an integrating electrometer, is placed
It is recognized that this phantom is too large near the center of a long cylindrical phantom on the
and heavy to be practical for routine dosimetry CT couch. A series of helical CT scans of different
measurements on CT scanners in the clinical envir- lengths are acquired, each scan centered on the
onment. However, for research in CT dosimetry, thimble chamber. In Figure 7.21, three different
and potentially for dose assessment in the testing scan lengths are shown, La, Lb, and Lc. The
78
CT Dosimetry in Phantoms
Figure 7.21. The measurement procedure described in AAPM Report 111 (AAPM, 2010). An integrating thimble chamber is placed in
the center hole (or periphery, not shown) of a long cylindrical phantom. A series of helical CT scans are performed, with the center
(along z) of the scan located at the center of the phantom. Each chamber reading (in scans a, b, and c) is plotted as a function of the
scan length, L. The three different scan lengths La, Lb, and Lc are shown on the left, and the measured air-kerma values are plotted on
the right. These data are obtained for a given scanner, tube potential, “mAs” setting, and phantom, resulting in the h(L) profiles. When
the asymptotic value Deq [or equivalently h(1)] is used for normalization, the H(L) profiles are produced.
measurements for each of these scan lengths corres- by Deq and therefore asymptotically approaches
pond to a specific point on the H(L) curve, indicated unity whereas the unnormalized function h(L),
in the graph. The rise-to-dose-equilibrium curve, asymptotically approaches Deq. Hence, h(L) depends
H(L), defined by Eqs. (7.6) and (7.10), is normalized on CT acquisition parameters and has the utility
79
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
80
CT Dosimetry in Phantoms
the x-ray beam approaches and then passes where the primary term becomes more prominent.
through the real-time detector (a small thimble ion- This is more apparent in peripheral dose profiles
ization chamber), and continues toward the other than in center profiles.
edge of the phantom. An idealized real-time air-
kerma rate of the radiation meter is shown in
Figure 7.22b. As the primary beam enters the CT
7.7 Measurement of h(L) in the Clinical
phantom on the left, the x-ray probe measures the
Environment
signal produced by distant x-ray scatter. As the
x-ray beam approaches the detector, the measured- Using the real-time probe located near the center
scatter intensity increases, due both to solid-angle (along the z axis) of the ICRU/AAPM, 600 mm
effects and the shorter attenuation path length long, 300 mm diameter polyethylene phantom
between the x-ray beam and the real-time probe. shown in Figure 7.19, the measurement of f(z) can
As the primary beam passes over the probe, it be determined directly. Such a measurement for a
records not only scattered radiation but the Siemens ASþ CT scanner is illustrated in
primary radiation as well. The x-ray beam then Figure 7.23a. This is similar to the idealized dose
continues toward the right edge of the phantom, profile shown in Figure 7.22c; however, there is an
producing an approximately symmetrical profile. obvious notched pattern in this measured profile
The real-time results, measured as air kerma per that is due to CT-couch attenuation. Although the
time, can be converted to air kerma as a function of modern CT couch is made from low-attenuation
position using the known sampling period Dt (i.e., carbon-fiber components, there is still some attenu-
Kair ¼ K_ air Dt) and the known velocity v of the ation by the patient table evident in Figure 7.23a.
table transport, where position ¼ t/v. After this Although the notches in the f(z) profile appear as a
transformation, the profile as illustrated in distraction, the integration of this curve from 2 L/2
Figure 7.22c represents the air-kerma profile, f(z). to þL/2 according to Eq. (7.5b) produces a mono-
Using Eq. (7.5b), the cumulative air kerma at the tonic function h(L) in which the influence of
center of the phantom for scan length L can be the table attenuation is largely eliminated (see
obtained by integrating f(z) from 2L/2 to þL/2, as Figure 7.23b). Thus, depending on the alignment of
shown graphically in Figure 7.22c. These data are the notched pattern (i.e., its position relative to z ¼
then compiled to compute h(L) as a function of L, as 0), the integration of f(z) can result in the apparent
shown in Figure 7.22d. Despite the convenience of cancellation of the notches. That is, notches on one
this method for the assessment of h(L), it is not pos- side of z ¼ 0 will cancel out the absence of a notch
sible to accurately assess the SPR, h, in Eq. (7.10), on the other side. Another interesting observation
and therefore, there is some imprecision in the exact in regard to Figure 7.23a is how sharp each of the
shape of the curve near the y axis (for L , nT), notches appears, given that for most of the curve
Figure 7.23. Air-kerma rate from a Siemens spiral-CT scan through a phantom. (a) The signal trace from the real-time probe for a
helical scan performed along the entire length of the ICRU/AAPM phantom is shown for a Siemens ASþ scanner operating at 120 kV.
The thimble chamber was positioned at the center of the 300 mm diameter polyethylene phantom. (b) The rise-to-dose-equilibrium
curve, h(L), is computed from the profile shown in (a) by integrating from the center out. The h(L) from the real-time trace (solid line)
compares well with individual measurements using AAPM Report 111 methods (solid circles).
81
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
Figure 7.24. Air-kerma rate from a General Electric helical CT-scan through the ICRU/AAPM phantom. (a) The real-time trace is
shown for a General Electric VCT scanner operating at 120 kV. Although the notches from the table produce a trace with substantial
inhomogeneities, integration from the center outwards of this profile tends to reduce the impact of the table notches. (b) The h(L) curve
for the General Electric VCT scanner, determined by center-to-edge integration of the profile shown in (a). As before, the individual
data points represent serial measurements using AAPM Report 111 methodology.
Figure 7.25. Air-kerma rate from a Philips Brilliance 16 helical-CT scan through the ICRU/AAPM phantom. These data were
measured for a tube potential of 120 kV. (a) The output of the real-time x-ray meter. (b) The h(L) curve computed from the trace in (a).
Individual data points using AAPM Report 111 serial measurement methods are shown.
the real-time radiation meter is measuring scat- Figure 7.24a shows a measured f(z) using the
tered radiation only. As the x-ray tube rotates and 0.6 cm3 real-time probe through the center of the
the x-ray beam intercepts the table, the primary ICRU/AAPM phantom for a General Electric VCT
x-ray intensity striking the phantom is rapidly scanner, and the corresponding h(L) curve is shown
reduced, also reducing the scatter signal even at in Figure 7.24b. Again, the notches from the
considerable distances away from the primary CT-table attenuation seen in Figure 7.24a do not
x-ray beam. appear in the h(L) curve in Figure 7.24b. The solid
A particularly important observation from points measured using the AAPM Report 111
Figure 7.23b is that the solid circles plotted are h(L) methods are shown as well, with excellent agree-
values measured using an integrating ionization ment between the two measurement approaches.
chamber according to the methodology of AAPM Analogous results for a Philips Brilliance 16 CT
Report 111. The agreement between these points scanner are shown in Figure 7.25a and 7.25b. The
and the solid curve measured using the real-time very good agreement between the two measurement
probe is excellent. methods for three different scanners as shown in
82
CT Dosimetry in Phantoms
Figures 7.23 – 7.25 demonstrates concordance dependent on table speed, it is not an ideal
between the AAPM Report 111 measurement measure of CT scanner output per se, as x-ray
methods and those using the real-time probe output should be independent of table speed. For
described here. measurements involving helical-CT scanning,
There is a sentiment in the CT community that however, CTDIvol can be considered to be a
CTDIw can be considered as a loose measure per- measure of pitch-normalized output (e.g., air
taining to the relative output of the CT scanner; kerma per effective “mAs” setting, where effective
CTDIw is measured as air kerma in a PMMA “mAs” is the tube-current–time product divided
phantom (160 mm or 320 mm in diameter) and by the pitch). Embracing this concept, the h(L)
increases with x-ray tube potential, tube current, curves shown for three different CT models and
and exposure time, and has only a slight depend- manufacturers in Figures 7.23 – 7.25 are normal-
ence on collimation at clinically realistic settings ized by the corresponding CTDIvol at the appro-
on MDCT scanners. Because CTDIvol is priate tube potential and for each machine type,
and subsequently corrected to the appropriate
tube-current–time product, and pitch for each
scan. These results are shown in Figure 7.26.
Given that these curves are from measurements
on three different CT scanners, the normalized
curves are remarkably similar.
Figure 7.27a shows f(z) curves for a Siemens
ASþ CT scanner measured for a number of x-ray
tube potentials using the real-time probe
method. As expected, at the same tube-current –
time value, the air kerma for the system at
higher x-ray-tube potentials (and hence the area
of the curves) increases. Integration of these
curves leads to the h(L) curves given in
Figure 7.27b. The h(L) curves can also be nor-
malized by the tube-potential-dependent values
of CTDIvol, and these results are shown in
Figure 7.27c. It is evident that this normaliza-
Figure 7.26. Three G(L) curves from three different CT
scanners, normalized by CTDIvol. This figure illustrates the
tion essentially eliminates the effects of the
three central-position h(L) curves illustrated in Figs. 7.23, 7.24, x-ray-tube potential, resulting in very similar
and 7.25, normalized by CTDIvol corresponding to the CT normalized curves for all four x-ray-tube poten-
scanner used. tials evaluated.
Figure 7.27. Real-time probe data versus x-tube potential for a Siemens’ ASþ CT scanner using the ICRU/AAPM phantom. (a) The
traces of the real-time x-ray detector for four different tube potentials. All traces were corrected to a current–time product of 100 mA s.
The amplitude for the higher tube potentials is greater, as anticipated. The different location of the notches in the four traces indicates
that the start angle changed between measurements. (b) The resulting h(L) curves from the traces for the four tube potentials shown in
(a). The asymptotic value reached at the right end of the curves corresponds to the total area of the f(z) traces shown in (a). (c) The
individual h(L) curves for the x-ray tube potentials shown in (b), normalized by the measured CTDIvol, and thus are Gc(L) curves per
Eq. (7.13). As seen before with the different CT scanners, these normalized G(L) curves essentially fall on one another.
83
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
84
CT Dosimetry in Phantoms
Figure 7.16a. Figure 7.29 shows the excellent is due to the influence of both phantom attenuation
agreement among measured, simulated, and ana- and the inverse-square law.
lytical results. As noted previously, although the Gx(L) curves
The function h(L) represents the raw measure- are plotted starting at L ¼ 0 in Figure 7.30b and
ment (using either the integrating-probe method of 7.30d, the data are unreliable for L , nT, which is
AAPM Report 111 or the real-time probe measure- about 40 mm for this scanner. For the 320 mm
ment approach) of absorbed dose as a function of diameter phantom (see Figure 7.30b), the center
scan length in a phantom, at either the center measurement is nearly always lower than that of
[hc(L)] or peripheral [hp(L)] locations. The h(L) the peripheral value, as would be expected for a
curves are dependent on output-related parameters large-diameter phantom. For the 160 mm diameter
such as tube potential, tube current, acquisition phantom (see Figure 7.30c), however, the center
time, pitch, etc. Combining the center and periph- and peripheral profiles intersect at a scan length of
eral measurements, the CTDIvol(L) can be derived: about 135 mm, with the center exceeding the per-
ipheral beyond this scan length. This demonstrates
1 2 the increasing build-up of scattered radiation at
CTDIvol ðLÞ¼ ha ðLÞ ¼ hc ðLÞ þ hp ðLÞ: ð7:12Þ
3 3 the center of the phantom as a function of scan
length.
Following Eq. (7.11), Eq. (7.12) can be divided by
Although the numerator, CTDIvol(L), in Eq. (7.13)
the conventional (L ¼ 100 mm) CTDIvol value, mea-
is from a measurement using a long phantom and
sured at (or scaled to) the same technique factors
the denominator, CTDIvol(L ¼ 100 mm), is from a
(tube potential, tube current, pitch, etc.) that were
measurement in the standard 150 mm diameter
used to measure the h(L) data, to yield:
PMMA phantoms, the value of Ga(L) at L ¼ 100 mm
CTDIvol ðLÞ 1 2 is approximately 1.0 for both the 320 mm diameter
¼ Ga ðLÞ ¼ Gc ðLÞ þ Gp ðLÞ: (see Figure 7.30b) and the 160 mm diameter (see
CTDIvol ð100 mmÞ 3 3
Figure 7.30c) phantoms. The implications of Eqs.
ð7:13Þ 7.11–7.12 and Figures 7.23–7.30 will be discussed
in the Summary section.
Figure 7.30a shows three 320 mm diameter PMMA
phantoms placed end-to-end on the table of a CT
scanner (Siemens ASþ). The G(L) curves acquired
7.9 The Radial Dose Profile
at 120 kV for both the center [Gc(L)] and the per-
ipheral profiles [Gp(L)] were evaluated using the Much of the focus so far has been on the air-
real-time-probe method for both the 160 mm and kerma or absorbed-dose distribution as a function of
320 mm diameter phantoms. The 320 mm diameter z in a phantom. In this Section, the radial distribu-
results are shown in Figure 7.30b and those for a tion is discussed. A Monte Carlo evaluation of an in-
160 mm diameter phantom in Figure 7.30c. At the finitely long, 320 mm diameter water phantom has
peripheral location [i.e., Gp(L)], some waviness in been performed, and the x-ray energy deposited ra-
the profile is seen (Figures 7.30b and 7.30c), which dially at 10 mm annular thicknesses was evaluated.
Figure 7.30. Data from a long PMMA phantom acquired at 120 kV. (a) A photograph of three 320 mm diameter PMMA phantoms,
placed end-to-end on a CT couch. (b) The center, Gc(L), and peripheral, Gp(L), curves are shown for the 320 mm diameter PMMA
phantom. The Ga(L) was computed as well, seen as the dashed line. Due to the normalization, the Ga(L) curve passes through a value
of 1.0 at L ¼ 100 mm. (c) The Gc(L), Gp(L), and Ga(L) curves are shown for the 160 mm diameter head phantom.
85
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
Figure 7.31. Radial dose profiles. (a) Monte Carlo generated profiles as a function of radial position (from center to edge) for a 320 mm
diameter water phantom in a beam generated at 120 kV. The energy deposited increases from center to edge, which is consistent with
basic principles of radiation attenuation. Because absorbed dose is energy deposited per mass, and the mass of an annulus of smaller
radius is smaller than that of an annulus toward the periphery of the phantom, the absorbed-dose profile is different. (b) The radial
absorbed-dose profiles for four different x-ray tube potentials, for a 320 mm diameter water phantom. In all cases, the absorbed dose at
the center of the phantom is about from 20 % to 25 % less than the maximum absorbed doses that occur near the periphery of the
phantom.
The deposited energy is shown as the lower curve in curve for either the center (x ¼ c) or peripheral (x ¼
Figure 7.31a. Although less energy is deposited p) phantom positions, or the 13 and 23 weighted
centrally in the phantom due to attenuation, the average value [x ¼ a; see Eq. (7.12)]; this is consist-
volumes (and hence masses) of the inner annuli ent with the definition h(L) ; DL(0). The amplitude
(and central cylinder) are also much smaller, so that of h(L) scales with the CT technique factors
the radial distribution of the equilibrium absorbed used, and therefore this function reflects the specific
dose (top profile in Figure 7.31a) is much more set of acquisition parameters (tube potential, tube
uniform than the distribution of total energy depos- current, acquisition time, pitch, etc.) used for the
ited. Figure 7.31b illustrates the equilibrium scan. In the limit of L ! 1, h(L)! Deq, and the
absorbed dose as a function of radial position for function H(L), defined as h(L)/Deq, approaches
four x-ray tube potentials. There is some variation unity. The normalized value of H(L) is useful in
in the absorbed dose as a function of radial position many applications (AAPM, 2010) and is also import-
in CT. However, compared with x-ray projection ant to the understanding of the energy deposited
imaging such as radiography and fluoroscopy, the by scattered radiation (i.e., Eqs. 7.9 and 7.10), and
radial equilibrium absorbed-dose distribution in CT the normalization process eliminates most of the
is relatively uniform, the peripheral-to-central-axis dependencies on technique factors.
ratio being about 1.4 for a tube potential of 120 kV It is recognized that the function h(L) is useful
in this 320 mm diameter phantom. because it retains the CT scanner output character-
istics that are dependent upon technique factors;
however, this means that there would be a different
h(L) curve for each combination of technique
7.10 Summary factors. To address this, the Gx(L) was defined in Eq.
In this Section, the body of knowledge produced (7.13), which is simply hx(L)/CTDIvol, where the
by many scientists working in the field of CT dos- definitions for x were defined previously. The utility
imetry has been presented and compared. Many of Gx(L) is that it retains the output characteristics
common themes have emerged. Collectively, this unique to the CT technique factors, but is normal-
work demonstrates that CT dosimetry in phantoms ized by a quantity that is available on the CT
has moved well beyond the traditional parameter of console (or in the DICOM header), the conventional
CTDIvol, and it also follows the developments pre- CTDIvol. Therefore, given the Ga(L) curve for a
sented in AAPM Report 111 (AAPM, 2010). given phantom diameter and composition, the
The function hx(L) was defined as a raw (unnor- planar dose average at z ¼ 0 can be computed
malized) measurement of the rise-to-equilibrium knowing the CTDIvol as well as the scan length L.
86
CT Dosimetry in Phantoms
Because the CTDIvol and the dose – length product known for the average diameter of the patient’s
(DLP) are reported on all modern CT scanners for anatomy being scanned. This would suggest that a
each CT scan, the value of L can be computed as family of Ga(L) functions for different phantom dia-
L ¼ DLP/CTDIvol [see Eq. (4.11)]; this means that meters, Gdia
a ðLÞ, could be useful for the clinical as-
the length-adjusted planar-dose average (at z ¼ 0) sessment of patient dose.
can be computed in a clinical setting when G(L) is
87
Journal of the ICRU Vol 12 No 1 (2012) Report 87 doi:10.1093/jicru/nds008
Oxford University Press
90
Patient Size-Specific Dose Estimation
Figure 8.2. Examples of three-dimensional voxel phantoms (Turner et al., 2010). Organ boundaries were hand segmented in most
cases.
91
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
Figure 8.5. Patient sizes. (a) The anterioposterior dimension as a function of patient age. (b) The transverse or lateral dimension as a
function of age (Kleinman et al., 2010).
92
Patient Size-Specific Dose Estimation
Figure 8.7. Effective diameter as a function of linear dimensions of the torso. The points represent data reported in Boone et al. (2003),
Kleinman et al. (2010), and ICRU Report 74 (ICRU, 2005). A second-order polynomial provides an approximate fit (solid line). (a) The
effective diameter as a function of the AP dimension. (b) The effective diameter as a function of the lateral dimension (fit is solid line).
initiated. For the localizer view, the AP dimension Although the lateral and the AP linear dimensions
can be measured from a lateral localizer scan, and can be measured on an axial CT image using soft-
the lateral patient dimensions can be evaluated in ware tools available on all modern CT scanners,
a PA view. By determining the AP or lateral dimen- the preferred approach for estimating size is to
sion of the patient on the localizer view, the effect- measure the area of the patient’s body directly
ive diameter of the patient can be estimated. As from the cross-section of the CT image. A relatively
will be shown below, these values can then be used simple approach for estimating the area of a
to estimate a conversion factor that in turn can be patient’s cross-section is described below.
used to compute the SSDE (AAPM, 2011) for the The automated assessment of the effective diam-
patient prior to the commencement of the CT scan. eter from an axial CT image is relatively straight-
In principle, this method would allow the technolo- forward. The dimensions of a given pixel (Dx, Dy)
gist to predict inappropriately high patient doses are found in the DICOM header, and the pixel area
prior to the CT scan, and make technique adjust- is computed: Ap ¼ Dx Dy. In the CT image, the
ments to reduce the absorbed dose. HU of water is approximately 0, and the range of
It is recognized that the localizer view on the CT HUs in most soft tissues ranges from about 2200
scanner is a projection image produced in fan-beam to þ200; however, the HU for lung tissues can be
geometry, and therefore some magnification in the much lower, and the HU of bone is much higher.
image is present. If the patient is not properly cen- Air surrounds the body and is also present in the
tered in the gantry, the linear dimensions mea- lung and in some body cavities. To calculate the ef-
sured on the image will be less accurate because of fective diameter from a single CT image, an algo-
the magnification. CT manufacturers have access rithm can be used to evaluate each HU value in
to the attenuation data measured during a localizer the image, and tally only those pixels that exceed a
view, and these data could also be used to compute certain threshold value, G. Although for air, HU
a more accurate estimate of patient size. However, ; 2 1000, the HU of air in some regions of the
this information is proprietary to each vendor, and image internal to the body can be higher than
achieving a common metric across vendors using 21000 due to x-ray scatter and other artifacts, sug-
attenuation measurements is not likely. gesting that a threshold value slightly higher than
In most clinical settings when dose estimates are 21000 should be used to segment tissues from air.
required (after the CT scan), the CT image itself A computer algorithm that counts the number of
should be used for the determination of patient pixels in the image exceeding a threshold value G
size. Unlike the localizer images, the CT image (such as G ¼ 2800) can be used to segment body
dimensions are accurate in all three dimensions. tissues from air. In this algorithm, N is first set to
93
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
zero, and then: In this case, for pixels other than for air, the
average pixel density relative to water is
yP
max xP
max
if ðimageðx; yÞ . GÞ; N ¼ N þ 1; ð8:1Þ X
y¼0 x¼0 Pave ¼ : ð8:5Þ
N
where G is the threshold value, image(x,y) is the
Note that if in Eq. (8.4), HU ¼ 0 for all pixels (cor-
HU value for pixel (x,y), and the summation is per-
responding to water), then the average density Pave
formed over the entire CT image. It is understood
would be unity. The water-equivalent patient diam-
that N ¼ N þ 1 refers to indexing the counter vari-
eter, dw, is given by
able N when the if condition is true. The number of
pixels, N, exceeding the threshold value in the CT rffiffiffiffiffiffiffi
4A
image is then multiplied by the area per pixel, Ap, dw ¼ Pave : ð8:6Þ
to compute the total cross-sectional area, A, of the p
segmented body region in that CT image,
Figure 8.8 illustrates this case for cylindrical
A ¼ N Ap : ð8:2Þ PMMA (r ¼ 1.19 g/cm3) phantoms with diameters
ranging from 100 mm to 320 mm. The estimation of
The effective diameter is then computed from the deff or dw for the patient from area measurements
segmented body area: made from the CT image is a direct measurement;
however, the estimation of deff from the lateral or
rffiffiffiffiffiffiffi AP dimensions assumes a characteristic (elliptical)
4A : ð8:3Þ
def f ¼ shape for the human torso. The use of dw is pre-
p ferred over deff, in general, especially for the thorax.
Another consideration regarding patient dimen-
This procedure is relatively robust, but it tends to sions relates to where along z should the effective
include the CT table as part of the body area; in diameter be measured. To study this, data from a
practice, this results in a small positive bias in the cohort of patients were evaluated, and the average
estimate of the effective diameter. One of the lim- diameter of the patient [using Eq. (8.3)] for each
itations of this procedure is that in some cases the CT image over the entire length of the scan was
body area in the CT images is truncated, that is, compared against the diameter of one CT image at
part of the patient’s anatomy is outside of the the midpoint of an abdominal-pelvis CT scan.
reconstructed CT image. In such cases, the method Figure 8.9a illustrates this relationship for the
described above will underestimate the effective most prevalent CT examination, the abdomen –
diameter. However, it might still be possible to pelvis scan. Although good correlation exists
evaluate either the lateral or AP dimension of the between these two dimensions of body size, there is
patient, which then can be used in the estimation concern about the 4.7 cm offset seen in the data.
of body size. Assuming that the average diameter is the superior
In some CT images (e.g., chest), an appreciable quantity as it takes into consideration all of the CT
amount of internal air is contained within the body images in the scan, these data suggest that the
dimensions; however, air does not attenuate x rays midpoint diameter for an abdomen –pelvis CT scan
appreciably, and therefore the external dimensions is not the best value to use for the SSDE.
of the body will overestimate the attenuation of Figure 8.9b illustrates the average-diameter versus
that axial section of the body. To compensate for midpoint-diameter relationship for the abdomen-
this, the water-equivalent patient diameter can be only component of the CT scan, and here better
used. agreement is seen, with no offset and an insignifi-
The assessment of the water-equivalent patient cant 2.6 % difference in slope. Figure 8.9c illus-
diameter, dw, requires an additional step in the al- trates this relationship for the pelvis component of
gorithm described above for deff. Essentially, the the CT scan, and an even better agreement is seen
average density of the pixels in the image that are with no offset and a slope of essentially 1.0.
flagged as being tissue (when HU . G) is also The data shown in Figure 8.9a indicate that
tallied: there should be some concern when using SSDE
8 corrections that are global and are not performed
yP
max xP
max < N ¼ N þ 1; on a CT slice-by-slice basis in the abdomen –pelvis
if ½imageðx; yÞ . G; HU þ 1000 :
y¼0 x¼0 :X ¼ X þ exam. This bias is considered significant, and
1000 further research is necessary to evaluate if the
ð8:4Þ midpoint diameter of the patient is sufficiently
94
Patient Size-Specific Dose Estimation
Figure 8.8. The assessment of the water-equivalent diameter. (a) An axial CT image of a small PMMA phantom. (b) The
water-equivalent diameter as a function of actual PMMA diameter. The higher density of PMMA relative to water accounts for the
slight upward bias of the data relative to the line of identity.
robust for SSDE assessment in abdomen – here (specifically, the top eight data sets as indi-
pelvis CT. cated in the key in Figure 8.10). The difference
between the AAPM Report 204 trend line and that
developed from this more inclusive database is very
small, and—given the dispersion in the data—the
8.4 Size-Specific Dose Estimates
differences are clearly insignificant. Therefore, the
The combined data from the numerous studies use of the curve as published in AAPM Report 204
described in Section 8.2 have been evaluated to is recommended.
develop a relationship between a dose-conversion The data illustrated in Figure 8.10 are for a tube
factor and the effective diameter of the patient. As potential of 120 kV. The data shown in Figure 8.11a
described by Turner et al. (2010), when size- correspond to the average absorbed dose as a func-
dependent scaling factors were normalized by the tion of effective diameter for x-ray-tube potentials
CTDIvol, many of the dependencies on scanner ranging from 80 kV to 140 kV, with each curve nor-
type, bow-tie filter, and x-ray-tube potential were malized to a setting of 100 mA s (Zhou and Boone,
effectively factored out. Therefore, the data used as 2008). As expected, for the same “mAs” setting,
input to this analysis were individually normalized higher tube potential leads to higher air-kerma and
by CTDIvol, and then combined. If the CTDIvol was absorbed-dose levels. The average absorbed dose
specifically reported, it was used for the normaliza- for smaller patients is higher than for larger
tion. For the data illustrated in Figure 8.10 (for a patients, consistent with the results of Figure 8.10.
tube potential of 120 kV), if CTDIvol data were not Figure 8.11b shows the curves for 80 kV to 140 kV,
available, the individual curves were normalized to but normalized to 1 mGy of air kerma at the iso-
a value of 1.09 for a 320 mm diameter patient. This center of the scanner. This normalization causes
factor of 1.09 [see Eq. (4.3)] converts from air the four curves to essentially collapse on one
kerma to absorbed dose in water, recognizing that another. Figure 8.11c shows the average absorbed
CTDIvol is measured in air, whereas the desired dose as a function of the effective diameter, but
dose information is for a water-equivalent patient. here each curve is normalized by the CTDIvol
The data points in Figure 8.10 show very similar assessed for the 320 mm diameter phantom. The
trends; however, there is greater dispersion in the curves in Figure 8.11c also show significant
data at smaller effective diameters. This is in part overlap, with better overlap at larger effective dia-
due to the fact that the normalization point is at meters and more dispersion at smaller effective
320 mm. The best-fit curve of the points is the solid diameters. Based upon the plots shown in
black line. The dashed line also plotted represents Figures 8.11b and 8.11c, normalization to the air
the results of AAPM Report 204 (AAPM, 2011), kerma at isocenter results in the best agreement
which was derived from 8 of the 12 data sets shown among these four curves. However, such air-kerma
95
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
96
Patient Size-Specific Dose Estimation
Figure 8.11. Average absorbed dose as a function of the effective diameter. (a) For four different tube potentials and 100 mA s on the
General Electric VCT scanner (Zhou and Boone, 2008). (b) Data of Fig. 8.11a normalized by the air kerma at isocenter of the scanner.
(c) Data of Fig. 8.11a normalized by the CTDIvol for the 320 mm PMMA phantom.
dose estimates. The diameter-versus-age information (200 mm to 300 mm) than the 100 mm scan length
used in Figure 8.12 was described in ICRU Report 74 that is the basis for CTDIvol (i.e., taking into con-
(ICRU, 2005). sideration the discussion of H(L) in Section 7), at
least for the eight studies that contributed to the
AAPM Report 204 data that are recommended for
8.5 Summary use in this Report. The SSDE values were computed
The data from a number of investigators have for the center of the scan field along the z axis, and
been combined and show that for the same CT the absorbed doses at the edges of the field will be
technique factors, the average absorbed dose is slightly lower (see Figures 7.9 and 7.10). The SSDE is
higher for smaller patients. The curves shown in considered to be more accurate than CTDIvol, and
Figures 8.10 and 8.12 clearly demonstrate these allows a relatively straightforward assessment of
trends. The CTDIvol is reported for air; however, absorbed dose in patients. The coefficients provided in
the dose-conversion coefficients to obtain SSDE Table 8.1 describe the equations to convert CTDIvol to
include the f-factor, fSI, from air to water absorbed SSDE when the deff or dw of the patient is known. A
dose. The SSDE dose-conversion coefficients also more complete discussion of the SSDE concept is
include the fact that abdomen scans are longer available in AAPM Report 204 (AAPM, 2011).
97
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
98
Journal of the ICRU Vol 12 No 1 (2012) Report 87 doi:10.1093/jicru/nds009
Oxford University Press
Figure 9.1. Illustration of tube-current modulation. (a) The so-called “wedding-cake” phantom. The individual section diameters are
10 cm, 13 cm, 16 cm, 20 cm, and 32 cm. (b) The x-ray-tube-current–time product resulting from the automatic exposure control on a
Siemens scanner, as a function of the reference “mAs”. (c) The x-ray-tube-current –time product resulting from the automatic exposure
control used by General Electric as a function of the noise index.
100
Automatic Exposure Control in CT
where fd32D
w
is the coefficient from AAPM Report 204
(AAPM, 2011) for a water-equivalent diameter of dw,
9.5 Examples of Slice-by-Slice CT Dose
derived from the tables in Report 204 (hence the “D”
Calculation
in the superscript nomenclature) for the 32 cm diam-
eter phantom. The CTDI32 vol is the CTDIvol reported When TCM is used, and when the patient has a
by the CT scanner; for this adult-abdomen applica- relatively non-uniform distribution of diameters
tion, all scanner manufacturers use the 32 cm diam- along the scan length, Eq. (9.3) will yield a more
eter phantom as reference, and that is indicated accurate SSDE than using the simpler assumption
explicitly here in the superscript. of constant or average tube current for a scan. For
For a situation in which the x-ray-tube current is implementation of Eq. (9.3) to be practical, auto-
constant, but the body region of the patient has mated software is needed to extract the pertinent
variable diameter along the scan length, the SSDE information from the DICOM header on each CT
101
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
image, and an automatic determination of the dw is developed to perform the SSDE calculation when
necessary for each CT image (see Section 8.3). To current-tube modulation is used.
realize the potential of this approach, software was Figure 9.4 illustrates such software. The entire
series of CT images in a patient scan is read into
computer memory, and the pertinent DICOM
header information is extracted. The CT data itself
are used to generate the PA (box 1 in Figure 9.4)
and lateral data (box 2) projections, illustrated in
the upper right of Figure 9.4. The effective “mAs”
is displayed in the lower left plot (box 3), and it is
observed that large variations in the effective
“mAs” were used in this CT scan. The lower middle
panel (box 4) illustrates the equivalent diameter
assessed from CT images along the length of the
scan. From this, the fdw conversion factor for the
SSDE estimate was computed from the average
diameter of each CT slice. Once the effective diam-
eter is known, the fdw conversion factor is computed
analytically from equations given in AAPM Report
204, which expedites this calculation. The com-
Figure 9.3. The CTDIvol as a function of effective “mAs” (J puted SSDE is shown plotted as a function of z on
Dt p 21) for one scanner (Siemens Definition ASþ), from 227 the lower right panel (box 5).
data sets. The slope of this curve (0.0674) determines the This software was used to evaluate parameters on
relationship between the CTDIvol and effective “mAs” for this 227 patients, and a number of trends can be
scanner, for a particular tube potential and bow-tie filter.
Figure 9.4. The output of a program written to evaluate the slice-by-slice characteristics of an abdominal-pelvic CT scan. The CT
images were used to generate the AP (box 1) and lateral (box 2) projection images shown. The three plots illustrate (box 3) the effective
“mAs,” (box 4) the water-equivalent patient diameter, and (box 5) the SSDE as a function of CT-slice position along the z axis. For each
of the four sections A, B, C, and D on each plot, summary statistics are provided in the data box in the upper left of the display.
102
Automatic Exposure Control in CT
identified as illustrated in Figures 9.5 through 9.9. that 77 % of the variance can be attributed to
The study involved the use of one CT scanner for patient diameter; the residuals between the mea-
one CT protocol (abdominal-pelvis examination), sured data and best-fit line are illustrated in
with automatic exposure control (including tube- Figure 9.6b. The variance of the CTDIvol in
current modulation). Figure 9.5 illustrates, as is to Figure 9.6a is 55.2 mGy2, and the variance in
be expected, that the effective “mAs” of the scan Figure 9.6b is 12.96 mGy2, confirming the 77 % re-
tracks reasonably well with the average patient duction in variance ([55.2 2 12.96]/55.2 ¼ 0.77).
diameter. The CTDIvol also tracks quite well with Relative to the CTDIvol, the SSDE tends to be
patient diameter, as shown in Figure 9.6a. This larger for smaller patients, and smaller for larger
figure illustrates that under the simplified condi- patients. Figure 9.7 shows the SSDE as a function
tions of this one-scanner/one-protocol study, of CTDIvol; lower values of CTDIvol correspond to
changes in the CTDIvol can be largely attributed to smaller patient diameters (as seen in Figure 9.6a),
patient diameter coupled with automatic exposure and the lower CTDIvol values for smaller diameter
control. The R 2 value of 0.77 in Figure 9.6a suggests patients are associated with higher SSDE conver-
sion coefficients. The increase in SSDE relative to
CTDIvol is shown as a function of average patient
diameter in Figure 9.8. For patients whose diameter
is smaller than 25 cm, the relative dose increase
(i.e., SSDE/CTDIvol ) is from 50 % to 60 % on
average, and for patients whose diameter is about
30 cm, the relative dose increase is about 20 %.
Patients in the 35 cm to 45 cm range in diameter
have a small dose reduction.
The data were further evaluated as shown in
Figure 9.9, where the ratio of the average pelvis
diameter to the average abdomen diameter was
assessed for male and females in the cohort of CT
patients. A significant difference in the ratio is
seen; women tend to have lower abdomen/pelvis
diameter ratios, and men have characteristically
larger ratios. These trends substantiate the very
general observation that men tend to gain weight
Figure 9.5. The effective “mAs” as a function of average patient in their abdomen whereas women tend to carry
diameter. These data were acquired using tube-current weight in their hips. These trends have gender-
modulation. The effective “mAs” increases non-linearly as a
dependent ramifications on dose assessment when
function of average patient diameter, typical of TCM techniques
on any modern CT scanner. tube-current modulation is used.
Figure 9.6. (a) The CTDIvol as a function of average patient diameter. Tracking with the effective “mAs” (as shown in Fig. 9.5), the
CTDIvol increases as a function of patient diameter. (b) The residual differences between the actual values and the curve fit (in
Fig. 9.6a). The mean residual value is 0.0 mGy, and the standard deviation is 4.88 mGy.
103
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
Figure 9.7. The SSDE as a function of CTDIvol. The dashed line Figure 9.9. The ratio of the average abdomen diameter to the
of identity is shown. As expected, the SSDE is higher than the average pelvis diameter for two populations, males and females.
CTDIvol (i.e., is above the line of identity) for smaller patients, A significant difference ( p , 0.05) is observed, demonstrating
and it is only slightly lower for larger patients. that men and women tend to have different body-shape
characteristics.
P
N organ
9.6 Organ-Dose Estimation Dorgan ¼ fD32w ½i Yz½i C32 Ji Dt p1 ; ð9:4Þ
i¼1
Organ-by-organ dose assessment is considered by
many to be the ultimate dose assessment; a com- where Yorgan
z is a dose-weighting function describing
plete description of organ doses is necessary for the the contribution of organ dose at position z for each
actual computation of effective dose, using (for organ of interest (liver, kidney, bladder, etc.) from
example) the tissue-weighting coefficients from the average absorbed dose in the plane at the z pos-
ICRP Publication 103 (ICRP, 2007). Although over ition corresponding to the index value i. Figure 9.10
the years, there have been many efforts using illustrates this concept. Note that z[i] need not be a
Monte Carlo calculations in three-dimensional linear function of i. Examples of realistic boney
phantoms to determine organ dose, the methods for landmarks in the torso include the superior edge of
computing SSDE on a slice-by-slice basis described the clavicle, the sternum, the pelvic crest, or any
above demonstrate that a practical tool for the esti- vertebral body. All other terms in Eq. (9.4) have
mation of absorbed dose on a slice-by-slice basis is been described previously.
presently available. It is further suggested that al- Using center and peripheral CTDI100 values
though Monte Carlo dose assessment is essential in averaged over tube potential (from 100 kV to
104
Automatic Exposure Control in CT
9.7 Summary
CT dosimetry has had to adapt rapidly to
changes in CT-scanner capabilities. Although con-
stant-tube-current protocols are still used for
Figure 9.10. A torso with boney landmarks, and with associated some applications, AEC techniques in CT lead to
representative organ positions. Using an organ-weighting dynamically adjusted x-ray-fluence rates during
system (Yorgan
z ), determined from a large patient cohort but the entire procedure in the majority of CT proto-
scaled to each individual patient, estimation of individual organ
cols. Dosimetry under these conditions is most ac-
doses would be possible even when AEC is used for a patient
with a non-uniform diameter distribution along z. curate when the effective “mAs” for each CT slice
generated in the image series is considered separ-
ately, and size corrections using the SSDE can
140 kV) and across four manufacturers’ systems1 then address both the specific x-ray-tube output
(ImPACT, 2004), the periphery-to-center CTDI100 for each CT image, as well as the water-
ratio in the 160 mm diameter head phantom is 1.08 equivalent diameter of the patient at that z pos-
(COV ¼ 2.8 %), and for the 320 mm diameter body ition. Although simpler methods such as using
phantom the ratio is 1.98 (COV ¼ 7.5 %). Thus, the one CT image at the middle of the CT scan to es-
weighting functions, Yorgan
z , would have to take into timate patient size can deliver reasonable esti-
consideration the average in-plane position of the mates for some scan protocols, or assuming that
organ, from center to edge; however, this is likely an average tube current was used over the entire
to be a second-order effect relative to the organ’s scan, a slice-by-slice approach to individual
position along the z axis. Furthermore, most patient dosimetry will likely yield the most accur-
organs are positioned neither at the center nor at ate dose estimate possible at this time. Such an
the extreme periphery, and so radial ( periphery- approach requires software that evaluates each
to-center) averaging is already included in the com- CT image in the CT scan. If such software is not
putation of CTDIw from CTDIcenter100 and CTDIperi100 . available, other methods for dose assessment
These observations suggest that there might be when AEC modes are employed will be adequate
little dependence on organ position in the (x – y) to determine SSDE in most clinical settings.
1
General Electric VCT, Philips Brilliance, Siemens’ Definition
AS, and Toshiba Aquilion Multi.
105
Journal of the ICRU Vol 12 No 1 (2012) Report 87 doi:10.1093/jicru/ndt001
Oxford University Press
108
Spatial Resolution in CT
Figure 10.2. Illustration of the spread functions used in imaging. (a) Input images (left to right) defining the point-spread function
(PSF), the line-spread function (LSF), and the edge-spread function (ESF). (b) Simulated degraded-output images showing raw image
data used for the measurement of the PSF, LSF, and ESF. The blurring seen in these functions is due to the imperfect resolution
properties of the imaging system being characterized. (c) Graphs showing the actual PSF, LSF, and ESF. The PSF(x,y) is a 2D function,
and the LSF(x) and ESF(x) are 1D functions.
side lobes seen in Figure 10.4b are the result of where rc is the center position of the PSF.
the edge-enhancement features of the convolution Rearrangement of the above equation and substi-
kernel used for filtered-back-projection reconstruc- tuting b ¼ 2arc gives
tion (Bushberg et al., 2012).
In the early work of Schneiders and Bushong
1=2
1
(1980), the technique of Judy was used with a a r þ b ¼ ln : ð10:7Þ
PSFðrÞ
custom computer program to compute the edge-
spread function from an image. The ESF was dif-
ferentiated numerically, producing the line-spread Using this approach, point-spread-function data
function. Fourier transform of the LSF resulted in were plotted as shown in Figure 10.5a. Data points
the (complex) optical-transform function, whose along a straight line describe a point-spread func-
modulus is the modulation-transfer function. This tion that is strictly Gaussian, whereas deviation
study was an early demonstration of computer- from the straight line suggests non-Gaussian be-
based numerical methods for the calculation of havior. The slope of the line shown in Figure 10.5a
spatial-resolution in CT. is the value of a in Eq. (10.7), determined by linear
Other investigators (Nickoloff and Riley, 1985) regression. By inserting the value of a into
developed an interesting approach toward charac- Eq. (10.6), the point-spread function is obtained
terizing the spatial resolution of CT. They assumed (see Figure 10.5b).
that the point-spread function could be approxi- Before the existence of modern-day picture
mated by the Gaussian distribution: archiving and communication systems (PACS),
access to the digital-image data from a CT scanner
or any other digital-imaging modality was limited,
2
ðrrc Þ2
PSFðrÞ ¼ ea ; ð10:6Þ and the CT-image-file format was often proprietary.
109
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
Figure 10.3. Early MTF characterization in CT by Judy (1976). (a). By angling an edge of PMMA in water by a small angle (u) relative
to the image pixel array, the sampling interval D was decreased geometrically, producing a smaller sampling interval cD. For this work,
u ¼ 1.28 and c ¼ 0.02. (b) The ESF computed from the oversampled edge. (c) The MTF computed from the LSF (determined by
numerically differentiating the ESF). This scanner had a cut-off frequency of about 0.30 mm21 at 5 % MTF.
110
Spatial Resolution in CT
Figure 10.4. Resolution measurements on an early CT scanner. (a) Bischof and Ehrhardt (1977) characterized the PSF in the CT field
of view, and the MTFs were computed at several locations. (b) An assumed symmetrical PSF was computed from the two-dimensional
MTF measurements.
Figure 10.5. Resolution assessment using a Gaussian model. (a) The analysis concept devised by Nickoloff and Riley (1985), as
described in the text. When the PSF data plotted in this manner fall on a straight line, the CT system demonstrates a PSF that can be
characterized by a Gaussian function. (b) The Gaussian PSF derived from the slope of the line in (a).
111
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
Figure 10.6. MTF analysis using a bar phantom. (a) The method described by Droege and Morin (1982) made use of the basic image
analysis tools available on CT scanners in the early 1980s, with regions of interest located on a bar phantom as shown. The mean and
standard deviation of the HU in each region of interest were used to compute the MTF. (b) The MTF measured using the Droege–
Morin method compares well with the more conventionally (i.e., wire PSF) measured MTF curve.
mathematically, creating the line-spread functions, subsequent calculation of the MTF) needs to be
illustrated in Figures 10.8b and 10.8c, for two dif- referenced and/or made in a consistent manner
ferent reconstruction kernels. The MTFs from this using similar regions in the image.
method are compared with those using the alumi- All modern CT scanners make use of filtered
num foil in Figure 10.8d (standard kernel) and in back projection (FBP), and many also utilize itera-
Figure 10.8e (bone kernel). Although some system- tive reconstruction techniques. There are a number
atic bias is apparent, the MTFs from these two of iterative algorithms in common use among the
phantoms are quite similar. commercial CT vendors, but most of the details of
For routine quality control in the clinical envir- these algorithms are proprietary. Nevertheless, in
onment, bar patterns designed specifically for CT general, iterative algorithms can exhibit non-linear
are commonly used (see Figures 10.9a and 10.9b). behavior that violates the measurement paradigm
This technique allows the user to determine the (i.e., a linear, stationary system) for the MTF.
limiting spatial resolution subjectively, by visual in- Specifically, iterative reconstruction methods can
spection. Although this technique is adequate for produce images in which the MTF exhibits a greater
spot checking the resolution properties of a CT or lesser dependence on spatial location (in compari-
scanner in the field, the evaluation of the son to FBP) as well as a dependence on the contrast
modulation-transfer function is considered to be a (difference in HU) of the material forming the edge.
more rigorous assessment of the resolution proper- Commercial CT scanners for medical applications
ties of the CT scanner. are designed to optimally image tissue, which
spans a range in Hounsfield units from about 2200
to þ200. Tissues and non-tissue materials encoun-
10.3.1 Limitations and Concerns in
tered outside this range can produce unwanted
Axial-MTF Assessment
effects on image quality, chiefly high-density
As mentioned in Section 10.2, a stationary objects of high atomic number (Z) that produce
imaging system is one in which the point-spread artifacts and non-linear responses. Therefore, reso-
function is constant over the field of view. Modern lution templates and phantoms that make use of
fan-beam and cone-beam CT systems are only ap- either high-Z or very dense materials can drive the
proximately stationary, as there are differences in scanner into non-linear regions of operation. The
the PSF from the center to edge in the recon- effect of dense or high-Z materials can also cause
structed image, and there is a slight angular de- beam hardening, another source of non-linearity in
pendence as well. Therefore, the location of the the reconstructed image. These observations
measurement of spread functions (and the suggest that phantoms for the measurement of
112
Spatial Resolution in CT
Figure 10.7. MTF analysis using an aluminum slit. (a) A thin Al foil is compressed between two tissue-equivalent blocks, and is aligned
parallel to the z axis, with a slight angle with respect to the x– y plane. (b) The image from which the LSF is deduced. (c) The LSF
synthesized using the angled-slit method, for three reconstruction kernels. (d) The corresponding MTF curves for the three kernels.
spatial resolution should employ relatively low-Z saturation should be avoided by not exceeding the
materials or foils thin enough to not exceed the 12-bit HU range between 21024 and þ3095. A
linear operating range of the scanner. A contrary histogram plot of Hounsfield units for the image
observation, however, is that high-Z or dense mate- where spatial resolution is to be measured can
rials produce edges or lines with a high signal-to- verify that HU saturation has not occurred.
noise ratio, which helps to produce high-precision Quantum noise tends to degrade the precision
MTF measurements. At the very minimum, HU and accuracy of measurements of spatial resolution.
113
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
Figure 10.8. An inverse-contrast slit method employing an air slit. Some results are compared with those from an aluminum slit as
described in Figure 10.7 (from Boone, 2000). (a) An air gap created between two pieces of solid phantom using paper spacers. (b) The
LSF determined from the air slit (AS) and the aluminum slit (Al), for the standard kernel. (c) The LSF determined from the air slit
(AS) and the aluminum (Al) slit for the bone kernel. (d) The MTFs assessed for the standard kernel. (e) The MTFs assessed for the
bone kernel.
Figure 10.9. Resolution assessment using the American College of Radiology CT phantom. (a) An image from the resolution section of
the ACR Phantom, reconstructed with the standard kernel on a General Electric VCT system. (b) The same raw data as in (a) but
reconstructed with the bone kernel, which has higher spatial resolution compared with that using the standard kernel.
To reduce the influence of quantum noise, tube- images are recommended, but with the following
current and exposure-time settings that give the considerations. The x-ray-tube focal spot can experi-
flexibility to use the small and large focal spots of ence blooming effects (an increase in apparent size)
the x-ray tube and yet provide high-precision at very high tube-current settings, and thus the use
114
Spatial Resolution in CT
of the highest tube-current settings is not recom- profiles are shown in Figure 10.10 for representative
mended unless these settings are used clinically. single-slice CT scanners of the late 1980s, for colli-
Large-diameter phantoms tend to increase the noise mated slice thicknesses ranging from 8 mm to
levels in the reconstructed image due to their 10 mm. A mathematical model of the slice-sensitivity
greater attenuation. Therefore, it is desirable to profile was also developed (solid lines in
characterize spatial resolution using a relatively Figure 10.10a–10.10c; see also Section 6.2).
small-diameter phantom. When the ESF is used to With the advent of helical (spiral) CT, Hu and Fox
measure spatial resolution, differentiation [Eq. (1996) developed analytical equations corroborated
(10.4b)] to produce the LSF (which leads to the by physical experiments to describe the slice-
MTF) adds considerable noise to the LSF profile, sensitivity profiles in single-slice helical scanners
whereas integration of the PSF to produce the LSF (see Figure 10.11a). Profiles with different colli-
[Eq. (10.3)] acts to reduce noise. These considera- mated slice thickness are shown in Figure 10.11b.
tions also have an impact on the choice of method- These investigators also computed the modulation-
ology used for MTF characterization in CT. transfer function for the z-axis resolution, calling it
Historically, spatial resolution in CT has been the longitudinal MTF (see Figure 10.11c).
described in units of cycles per centimeter or line Figure 10.12a shows a diagram of a common geo-
pairs per centimeter (i.e., cm21); however, with the metric strategy used to estimate slice thickness. A
improved spatial-resolution capabilities of modern number of evenly spaced small attenuating rods are
scanners, it is recommended that cycles per milli- aligned diagonally in a phantom, and a CT scan is
meter (i.e. mm21) be used instead in all forums of acquired of this array. Visual inspection of the CT
discussion, including the scientific literature and image allows the number of rods in the image to be
commercial, technical, and sales documents. counted (see the drawing in Figure 10.12b). For
the American College of Radiology (ACR) CT-
accreditation phantom, each rod corresponds to
10.4 Resolution Along the z Axis
0.5 mm of slice thickness. For example, counting 10
The z-axis resolution in CT has improved signifi- rods in the image then results in an estimate of a
cantly with the introduction of MDCT scanners. In 5.0 mm slice thickness. Figure 10.13a through
the era of single-slice CT scanners, slice thicknesses 10.13d illustrates CT images from the actual ACR
used for clinical imaging were typically 5 mm, phantom, for different thicknesses. This determin-
7 mm, or 10 mm, depending upon the clinical appli- ation of slice thickness allows rapid subjective as-
cation. On occasion, thinner slices, such as 1 mm or sessment of slice thickness for routine quality
2 mm, were used for specialty applications. As the assurance or for spot checking CT systems in the
voxel thickness dimension was on the order of from clinical environment. Although adequate for these
10 to 20 times that of the in-plane pixel dimensions, purposes, the determination of the line-spread func-
the slice-sensitivity profile was used to characterize tion along the z axis with subsequent conversion to
spatial resolution (essentially the line-spread the longitudinal MTF is considered to be a more
function). rigorous quantitative evaluation of the longitudinal
Gagne (1989) studied the shape of the slice- spatial resolution.
sensitivity profile and the air-kerma profile, both ex- Thin-section single-detector-array axial-CT acquisi-
perimentally and mathematically. Samples of these tion provides well-sampled data for CT-image
Figure 10.10. Slice-sensitivity profiles. A comparison is made between a geometric model (solid line) and the measured sensitivity
profile (symbols) for the z axis (see Gagne, 1989), for nominal CT slices between 8 mm and 10 mm. (a) For the General Electric 9800
scanner. (b) For the Siemens DR/H CT system. (c) For a stationary-detector CT system, the Picker 1200.
115
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
Figure 10.11. Measurements of slice-sensitivity profiles and the longitudinal MTF. (a) A comparison between measured data (symbols)
and model predictions (lines) for different collimations (from Hu and Fox, 1996). (b) The impact of x-ray-beam collimation on the
slice-sensitivity profile, for a helical acquisition on a single-detector-row CT scanner. (c) The z-axis (or longitudinal) MTF from the
slice-sensitivity profiles.
116
Spatial Resolution in CT
Figure 10.14. CT-acquisition geometry in the z dimension. (a) A single-detector-array, axial-scan geometry has a negligible cone angle,
and therefore there is complete sampling along the z dimension for contiguous SDCT scanning. (b) An MDCT system has a small
half-cone angle of approximately from 18 to 48, depending on the scanner configuration, which results in a slight under-sampling in the
z dimension for both axial and helical scans. (c) Full-cone-beam scanners, such as the Toshiba Aquilion 320, or flat-panel-based CT
systems make use of large cone angles; the severe under-sampling in the z dimension can give rise to cone-beam artifacts.
Full-cone-beam CT scanners are available that would be to use a large high-contrast sphere em-
use very wide cone angles for axial acquisition; for bedded in a homogeneous material as shown in
example, the Toshiba Aquilion 320 makes use of a Figure 10.15a. Such a phantom provides edges in
160 mm wide array of detectors along z, such that all three orthogonal dimensions, which can be used
the edge detectors (see Figure 10.14c) have a cone to characterize the resolution in each dimension.
angle approaching 88. Flat-panel-based CT scanners Although there is potential utility in the computa-
used in dental, orthopedic, and breast-imaging appli- tion of the 3D MTF, the parameters that influence
cations also have very large cone angles, in some the axial (x – y) resolution in CT tend to be quite
cases exceeding 208. With both MDCT and different from those that influence the longitudinal
full-cone-beam CT systems, the resolution along the z (z) resolution. For this reason, separate resolution
axis has increased dramatically compared with what measurements of the axial and longitudinal compo-
was practical with single-detector-array CT systems, nents are typically performed. In both cases, the
and this has been accompanied by much shorter line-spread function serves as a basis from which
scan times as well. Nevertheless, the increased z-axis to calculate the MTF.
resolution has also given rise to the potential for
cone-beam and other artifacts stemming from under- 10.5.1 Axial-Plane Resolution
sampling in z. Due to these considerations, CT reso-
It is straightforward to generate a line image in
lution along the z axis for MDCT and cone-beam CT
CT by scanning a high-contrast (thin foil) plane of
scanners can have substantial spatial dependence
material encompassed in a cylindrical phantom;
(non-stationarity).
this is illustrated in Figure 10.15b and is pictured
in Figure 10.15c. This phantom comprises 100 mm
10.5 Modern Resolution Metrics in CT
diameter PMMA, and has two sections: on top, the
The voxel dimensions of a CT image are no cylinder was cut down the center and machined
longer measured in fractions of a centimeter, but flat, and a thin metal foil was placed between the
rather in fractions of a millimeter. With this in two halves. The bottom section is homogeneous
mind, it is reasonable to use methods in CT that PMMA, but placed vertically between the two
have been used to characterize spatial resolution in cylinders is another metal foil. The upper section of
digital radiography systems, with the need to also the phantom is used to measure the LSF in the
address the 3D nature of the CT volume data set. axial (x – y) plane, and the foil separating the two
As mentioned in Section 10.1, a fully 3D character- cylinders vertically is used to measure the LSF in
ization of the spatial resolution of a CT scanner is the coronal (x –z) plane (effectively the same as the
possible, yielding the PSF3D(x,y,z). One approach sagittal plane).
117
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
Figure 10.15. Phantom for measuring MTF in both axial and azimuthal planes. (a) A high-contrast sphere embedded in a homogeneous
phantom allows the edge-spread function to be evaluated in any plane: the axial (x–y), the coronal (x–z), and the sagittal (y–z). (b) A
phantom composed of a flat, thin metal foil embedded in a 100 mm PMMA cylinder has orthogonal planes of foil for measuring the
MTFxy and the MTFz. (c) A photograph of the phantom schematized in (b), with a half-cylinder removed for demonstration.
Figure 10.16a illustrates a CT image of the top performance of CT scanners. Although the assess-
section of the MTF phantom, showing a bright line ment of spatial resolution in the clinical environ-
where the metal foil was placed between the two ment is qualitatively managed by visual inspection
half-cylinders. The oversampled LSF (Boone, 2000; of images of bar or wedge phantoms (see
Bushberg et al., 2012) was computed from the ROI Figure 10.9), scientific characterization of the
shown in Figure 10.16a, with the result illustrated spatial resolution in CT should make use of the
in Figure 10.16b. The Fourier transform is taken of MTF. The MTF can be determined in a number of
the LSF data, and after normalization the MTFs different ways, through the measurement of the
were computed for three different reconstruction point-spread function, the line-spread function, or
kernels, with the resultant MTFs shown in the edge-spread function, with appropriate manipu-
Figure 10.16c. lation of the data as discussed in this Section. The
in-plane (axial) MTF, or MTFxy, for a reconstructed
10.5.2 z-Axis Resolution image in a CT volume data set has similar reso-
lution in the x and y dimensions, allowing for radial
Resolution assessment along the z axis requires
averaging. The longitudinal MTF, or MTFz, is com-
a point, line, or edge running nearly perpendicular
puted and interpreted separately from the MTFxy.
to the z axis. Figure 10.15a illustrates a sphere
For simplicity and consistency, the oversampled
embedded in a homogeneous phantom, and—as
line-spread function approach is preferred, as
mentioned above—this approach provides edges in
described in Sections 10.5.1 and 10.5.2, and shown
all dimensions. However, for MDCT or cone-beam
in Figures 10.7b, 10.16a, and 10.17a. In most
CT systems with large cone angles used during ac-
cases, spatial resolution should be measured using
quisition, evaluation of the LSF(z) can be fraught
high-dose-technique settings to achieve low-noise
with artifacts. The alternative method demon-
images. Due to the high attenuation in large
strated in Figure 10.15b makes use of a thin metal
diameter phantoms, small-diameter phantoms
foil placed between two cylinders of PMMA. A
(such as the 100 mm diameter phantom pictured in
coronal image showing this foil as a bright line is
Figure 10.15c) are useful to achieve higher-preci-
shown in Figure 10.17a. The cylindrical phantom is
sion measurements of the MTF at a given tech-
slightly angled with respect to the axis of rotation
nique setting. The CT-acquisition parameters
in order to produce a slight angle in the vertical
described in Table 10.1 provide guidance in these
line of Figure 10.17a relative to the (x, z) pixel
respects. If the MTF with the small x-ray-tube
array in this coronal image. This angulation allows
focal spot is to be measured, the tube current
for the synthesis of the oversampled LSF, and
should be reduced and the scan time increased to
these data allow the computation of the pre-
enable small-focal-spot acquisition.
sampled MTF as shown in Figure 10.17b for three
Most clinical CT-scanner systems now have avail-
different reconstructed slice thicknesses.
able iterative-reconstruction techniques using stat-
istical or model-based reconstruction algorithms.
Images reconstructed using these methods, as
10.6 Summary
opposed to FBP, will demonstrate adaptive smooth-
Spatial resolution is a very important component ing in a manner that is dependent upon the specific
of image quality, requiring thorough and robust reconstruction algorithm. The adaptive behavior of
processes to characterize and quantify the these algorithms implies that the LSF (and MTF)
118
Spatial Resolution in CT
Figure 10.16. MTFs measured in the axial plane for an MDCT system. (a) A CT image of the top section of the phantom (see
Figure 10.15b) showing a line corresponding to the metal foil in the phantom. The box indicates the region of interest in which the data
points for the LSF were measured. (b) An oversampled LSF synthesized from the data in the box shown in (a). (c) MTFs shown for
different kernels on a General Electric 16-slice Lightspeed system.
Figure 10.17. Longitudinal MTFs measured on a modern MDCT system. (a) A coronal image of the phantom pictured in Figure 10.15c.
The longitudinal MTF is computed from vertical profiles sampled in the box shown. (b) The longitudinal MTFz shown for three slice
thicknesses. Smaller reconstruction thicknesses increase the spatial resolution in the z dimension.
will change depending on the local anatomy in the case, the spatial variation violates basic assump-
patient, depending on both position (x, y, z) and tions of stationary behavior, reducing the meaning-
contrast (signal difference) in the reconstruction. fulness of the MTF as a “global” metric. Similarly,
For iterative reconstruction, the spatial dependence iterative-reconstruction methods often involve
of the MTF can be greater or less than in FBP; adaptive, signal-dependent smoothing and regular-
some iterative methods have been developed specif- ization that impart a dependence of the MTF on
ically to yield a PSF that does not vary in space the contrast (difference in attenuation coefficient)
and others allow the PSF to vary. In the latter of the material forming a given edge. For example,
119
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
Table 10.1 Recommended parameters for the measurement of Recognizing these complexities in the spatial-
the MTF in CT. resolution characteristics of non-linear iterative re-
construction and the lack of a single “global” reso-
Parameter Value Tolerance or
note
lution metric, the MTF should be characterized in
a manner that is explicitly “local,” i.e., local in both
Tube potential 120 kV (typical) 80 kV to 140 kV location and contrast level.
Tube current 500 mAa +100 mA Given these constraints, it is recognized that
Time 1 sa +0.25 s characterizing the MTF of a CT scanner using the
Display field of view 200 mm +20 mm filtered-back-projection-reconstruction mode repre-
Phantom diameter 100 mm +5 mm
Phantom composition PMMA or other plastic
sents a reasonable estimate of the spatial resolution
materials of that scanner, but the local nature of the MTF
Number of lines in LSF 20-50 (about Zero padded to should be recognized. Furthermore, in most clinical
50 mm long) N ¼ 256 CT applications that employ iterative-reconstruc-
Angle of line to pixel 18 to 58 tion techniques, the iterative-reconstruction images
matrix
are combined with filtered-back-projection images
a
If the MTF for the small focal spot is being measured, the tube in a weighted manner (20 % iterative/80 % FBP,
current will need to be decreased to allow small-focal-spot 30 %/70 %, etc.), which can further complicate spe-
operation, and the rotation time can be increased to compensate. cification and emphasizes the need for rigorous
reporting of measurement conditions, reconstruc-
high-contrast edges can result in an evidently tion methods, and analysis techniques (including
higher MTF than that with low-contrast edges. dependence on location and contrast).
120
Journal of the ICRU Vol 12 No 1 (2012) Report 87 doi:10.1093/jicru/ndt002
Oxford University Press
Figure 11.1. Basic concepts of signal and noise. (a) A circular signal region located in the center of a noisy image. The detectability of
this disk is related to its signal-to-noise ratio. (b) The concepts relevant to the signal (s2 – s1) to noise (s) are illustrated. These idealized
curves (measured curves would be noisier) are computed from histograms of the gray-scale values in the signal and background regions
shown in (a).
Figure 11.2. Contrast-detail curves. (a) Image of a traditional CD phantom, for which the diameter of each disk decreases from right to
left and the contrast of the disks decreases from top to bottom. (b) An early depiction of the CD diagram for two different General
Electric CT scanners, the CT/T 7800 and the CT/T 8800. Adapted from Wagner et al. (1979).
images, Figure 11.3a typifies noise produced by keys of noise than the simple standard deviation; it
on the left side of the piano (low-frequency sound), describes the noise variance as a function of spatial
and Figure 11.3b typifies noise produced by keys frequency and therefore characterizes noise
further to the right on the piano (higher-frequency texture. When combined with dosimetric quan-
sound). With audible sound, the frequencies are tities, it can be used for comparisons among scan-
temporal frequencies measured, say, in units of s21, ners and among protocols, and has been shown to
whereas in images, the frequencies are spatial fre- be useful in translating protocols from one CT plat-
quencies, in mm21. form to another (Soloman et al., 2012). The
primary source of noise correlation in CT is essen-
tially the point-spread function, PSF3D, discussed
in Section 10. In MDCT imaging, the 3D point-
11.3 The Noise-Power Spectrum, NPS(f )
spread function induces noise correlation in all
The noise-power spectrum (NPS) is a useful three spatial dimensions. Because CT images are
measure that provides a more complete description produced by mathematical reconstruction (e.g.,
122
Noise Assessment in CT
Figure 11.3. Noise texture: the spatial-frequency distribution of the noise. Both images are from a Siemens CT scanner and have the
same standard deviation (s ¼ 21.5 HU); hence, the noise is equal in the two images. The different appearance of the two images is due
to the different reconstruction kernels used (Boedeker et al., 2007). (a) An image reconstructed using a B10 kernel, a low-pass kernel,
producing noise texture with lower-spatial-frequency noise. (b) An image reconstructed using a B50 kernel. The B50 kernel is a
high-pass kernel, producing CT images with greater spatial resolution but also with higher-frequency noise.
Figure 11.4. The noise-power spectrum, NPS. The solid curves represent an analytical prediction, and the open circles represent the
corresponding computer simulation. Adapted from Riederer et al. (1978). (a) The noise-power spectrum for a Hanning filter, where fN is
the Nyquist frequency. (b) The noise-power spectrum for a ramp filter.
filtered back projection), the PSF depends not only where the summation over i (and multiplication by
on acquisition parameters such as focal-spot size N 21) refers to averaging the NPS values over N
and detector dimensions, but also on the selection ROIs. The values of D correspond to the pixel spacing
of reconstruction parameters, principally the recon- (i.e., spacing between pixel centers in x or y) in a
struction kernel and the CT-slice thickness. given plane. Nx and Ny are the number of voxels in
The autocorrelation function describes noise cor- each dimension of the ROI used in the computation.
relation in the spatial domain. The Fourier trans- The terms fx and fy are here the spatial frequencies
form of the autocorrelation function is the so-called in the x and y dimensions, respectively, and the
Wiener spectrum (Dainty and Shaw, 1974; Faulkner DFT2D is the discrete Fourier transform in 2D.
and Morres, 1984), also known as the noise-power Seminal work by early researchers studying noise
spectrum, NPS. In the context of a two-dimensional properties in CT (Riederer et al., 1978) demon-
(2D) region of interest (ROI), let Ii (x,y) be the signal strated that the NPS characterizes the noise texture
in the i th ROI with I being the mean of Ii (x, y). of the CT image, and that in the axial plane the
Then, the 2D NPS is computed as NPS largely reflects the CT reconstruction kernel
used. Figure 11.4 illustrates analytically derived
1X N and computer-simulated NPSs for the Hanning
NPSð fx ; fy Þ ¼ DFT2D ½Ii ðx; yÞ Ii 2 Dx Dy filter (see Figure 11.4a) and the ramp filter (see
N i¼1 Nx Ny ;
Figure 11.4b). For an early single-detector-array CT
ð11:1Þ scanner, these investigators used the 2D NPS
123
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
because the scanner could acquire only one CT shown in the figure. Each 2D ROI is extracted from
image per axial scan; thus, no noise correlation the image and, after (optional) detrending (dis-
occurred in the z dimension. More recent work has cussed below) is used (see Figure 11.5b), the 2D
extended 2D NPS concepts to cone-beam CT Fourier transform is computed [see Eq. (11.1)] for
systems (Baek and Pelc, 2010; 2011a; 2011b). each ROI. After the considerable averaging of the
An example of how the 2D NPS is measured is N NPS data sets in the frequency domain, the
illustrated in Figure 11.5. A homogenous phantom mean 2D NPS is produced, and this is shown in
(e.g., water or plastic) is scanned using the desired Figure 11.5c. The fx and fy frequencies in the 2D
technique factors (tube potential, tube current, ro- NPS can be collapsed to a 1D radial frequency, fr,
tation time, etc.), producing a relatively homoge- by radially averaging using
neous image as seen in Figure 11.5a. ROIs are
qffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
sampled from the homogeneous image. Typically fx2 þ fy2 :
fr ¼ ð11:2Þ
that sampling is performed at a constant radius as
Figure 11.5. The NPS computation in CT. (a) ROIs sampled at a constant radius around the center of the phantom image. (b) An example
of detrending is shown. Here the raw ROI extracted from the image is fit with a low-order polynomial, which defines the background
trend in the data. Only the very low frequencies in the image can be addressed by the polynomial fit, due to its low order. The background
trend is subtracted from the raw ROI, resulting in the corrected ROI. (c) The two-dimensional Fourier transform is computed for the
corrected regions of interest (seen in b). In the axial plane as shown here, the NPS is approximately rotationally symmetric, and typically
has a shape resembling a torus. (d) The 1D NPS curve from radially averaging the 2D NPS. The initial positive slope of this curve results
from the ramp filtering that is used in filtered-back-projection reconstruction, and the negative slope at higher spatial frequencies occurs
due to the roll-off properties of the reconstruction kernel used to dampen high-frequency noise in the images.
124
Noise Assessment in CT
The typical appearance of the (axial) 2D NPS is experimental cone-beam CT scanner (Siewerdsen
that of a torus, as shown in Figure 11.5c. Radial et al., 2002). Figure 11.7a shows the 3D NPS in the
averaging of this 2D surface using Eq. (11.2) axial plane, and Figure 11.7b shows the 3D NPS in
results in a 1D NPS (see Figure 11.5d). The the coronal or sagittal plane. Due to rotational
positive-slope region of the NPS in Figure 11.5d is symmetry, the sagittal and coronal 3D NPS projec-
due to the ramp filtering, and the subsequent tions are essentially identical. Figure 11.7c illus-
fall-off at higher frequencies is due to the (mathem- trates a 3D rendering of the 3D noise-power
atical) reconstruction kernel that is used. Different spectrum, with a cutout view for clarity.
kernels are used that represent different trade-offs Figure 11.8 illustrates a 2D NPS obtained from a
between spatial resolution and image noise, but all Siemens clinical CT system. The NPS values for
clinical kernels produce some roll-off in response at four different reconstruction kernels (B10, B20,
high frequencies, similar to that shown in B30, and B40) are illustrated. When the NPS is
Figure 11.5d. Roll-off refers to the progressive re- computed directly from the CT scan of a homoge-
duction in the filter function at higher spatial fre- neous (usually cylindrical) object, some degree of
quencies, to reduce the impact of quantum noise on “cupping” is usually present, and this manifests as
the image. The 2D NPS can be displayed using 3D a spike in the curve at low frequencies in the NPS
plotting techniques, and a 2D NPS is shown in iso- (see Figure 11.8a). The cupping artifact is well
metric format in Figure 11.6. known in CT, and can be from beam hardening,
With modern MDCT and cone-beam CT systems, scattered radiation, or other phenomena. Cupping
there are many detector arrays in the z dimension refers to a low-frequency trend in the CT image in
that acquire the raw data simultaneously, and which HU values near the center of a homogeneous
these give rise to noise correlation in z. Thus, for object are lower than those near the periphery. To
these modern CT systems, the 3D NPS is necessary reduce the impact of cupping on the low-frequency
for fully characterizing the noise correlation in the portion of the NPS, a number of techniques can be
image data. The 3D NPS is a straightforward ex- used. The so-called detrending methods (Dobbins
tension of the 2D function (Tward and Siewerdsen, et al., 2006; Yang et al., 2008) can be used to sub-
2008), and is defined as tract out low-frequency trends in the spatial
domain. Figure 11.5b illustrates one type of
1P N
NPSðfx ;fy ;fz Þ ¼ DFT3D ½Ii ðx;y; zÞ I 2 Dx Dy Dz ; detrending, for example, in which a low-order poly-
N i¼1 i
Nx Ny Nz nomial fit to the data is subtracted out, reducing
some of the low-frequency components. Other
ð11:3Þ
researchers (e.g., Yang et al., 2008) have used
image subtraction to reduce the impact of cupping
where fz, Dz, and Nz refer to the frequency, voxel
on the NPS calculation. With this method, two sep-
spacing, and number of voxels used in the
arate acquisitions are made of a homogeneous (typ-
z dimension, and DFT3D is the discrete Fourier
ically cylindrical) test object, and these two data
transform in 3D. The 3D NPS is measured as
sets are subtracted from one another. With this ap-
described in Figure 11.5; however, the ROIs become
proach, inhomogeneities in the phantom are sub-
volumes of interest (VOIs), and the computations
tracted away also, and only noise remains. In 2D,
are performed for all three dimensions.
the image subtraction is represented as
Figure 11.7 illustrates different views of a fully
3D NPS computed from CT images acquired on an Kðx; yÞ ¼ IA ðx; yÞ IB ðx; yÞ ; ð11:4Þ
125
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
Figure 11.7. A 3D NPS computed for a cone-beam CT scanner. (a) The axial plane (fx – fy) of the 3D NPS shown for fz ¼ 0. (b) The
sagittal plane of the 3D NPS, shown for fy ¼ 0. (c) Three-dimensional rendering of the 3D NPS.
Figure 11.8. Two-dimensional NPS for a Siemens Somatom 64-slice CT scanner, adapted from Boedeker et al. (2007). (a) The 2D NPS
for four reconstruction kernels indicated in the inset. The low-frequency spike is indicated by the arrow. (b) The 2D NPS for the same
four reconstruction kernels, but using 2D image-subtraction methods [see Eq. (11.4)] to suppress the low-frequency spike.
illustrates a significant reduction in the low- It is usually true for commercial CT scanners that
frequency spike that is seen in Figure 11.8a, and Dx ¼ Dy, and hence for a square ROI Nx ¼ Ny.
this was achieved using the 2D image-subtraction However, the voxel spacing in the z dimension (Dz)
method of Eqs. (11.4 and 11.5). is usually not equal to that in the x or y dimension.
For full 3D data sets, image subtraction is per- Consequently, the value of Nz can be different from
formed using Nx or Ny, depending upon the size of the cylinder
scanned and the length (along z) of the scan itself.
Kðx; y; zÞ ¼ IA ðx; y; zÞ IB ðx; y; zÞ ; ð11:6Þ Figure 11.9 illustrates the influence of phantom
diameter on the NPS. These NPS curves were com-
where the A and B subscripts refer to independent- puted using 3D data-set subtraction techniques
ly acquired measurements of the same volume (x, y, [Eqs. (11.6 and 11.7)] to reduce artifacts in the low-
z). The 3D NPS is then computed using frequency regions of the NPS. For the curves
shown in Figure 11.9a, the same technique factors
N
P 2 (80 kV, 7 mA, 16.6 s) were used for polyethylene
DFT3D ½Ki ðx;y;zÞK i phantoms of four different diameters. In this case,
NPSðfx ;fy ;fz Þ ¼ 1 i¼1 Dx Dy Dz :
the largest-diameter phantom attenuates the most
N 2 Nx Ny Nz x-ray photons, and hence fewer photons reach the
ð11:7Þ detector compared to the smaller phantoms.
126
Noise Assessment in CT
Figure 11.9. NPS curves computed from images of four polyethylene cylinders of different diameters acquired on a prototype
cone-beam breast CT scanner. Three-dimensional image-subtraction methods [see Eqs. (11.6 and 11.7)] were used to suppress
low-frequency artifacts. (a) Images were acquired with constant technique settings, so that the air kerma at isocenter (in the absence of
a phantom) was the same for each of the four cylinders imaged. Due to the increased attenuation of the larger-diameter cylinders, the
NPS is greatest with the larger-diameter cylinder, and decreases with smaller phantom diameters. (b) For images acquired in this case,
the absorbed dose at the center of the phantoms was kept approximately constant, i.e., the incident air kerma used for the
larger-diameter cylinder was greater than for the smaller cylinder. With these acquisition parameters, the smallest-diameter cylinder
has the largest-amplitude NPS curve, and the noise level decreases with increasing phantom diameter.
Therefore, the noise variance (area of the NPS noticeable higher-frequency noise characteristics,
curve in this display format) is greatest for the i.e., the torus has a larger diameter.
largest-diameter phantom. In Figure 11.9b, The sagittal and coronal NPS curves in
however, the absorbed dose in each phantom was Figure 11.10a and 11.10b show virtually identical
designed to be approximately equal, requiring a trends, due to the approximate radial symmetry of
higher x-ray-tube current for the larger phantoms. the NPS in these planes. These 3D functions can
In this case, the smallest phantom was scanned be condensed by radially averaging the 3D NPS in
using the lowest tube current and had the highest the axial plane, through the use of Eq. (11.2). After
noise levels. The integrated volume of the 3D NPS performing the radial averaging, which collapses
curve is equal to the overall noise variance, s2, in the fx and fy axes to a single fr axis, the information
the image: illustrated in Figure 11.10 can be condensed as
ððð seen in Figure 11.11. This figure illustrates the
s2 ¼ NPSð fx ; fy ; fz Þdfx dfy dfz : ð11:8Þ (still 3D) NPS topography (fz and fr) for two differ-
ent reconstruction kernels on the Siemens CT plat-
form. The B10s kernel (see Figure 11.11a) is a
“soft” kernel and produces “smoother” images, with
an NPS that has more lower-frequency content,
11.4 Demonstration of NPS Utility and the B80s kernel (see Figure 11.11b) is a
In this Section, a number of trends that are “sharper” kernel that passes more higher-frequency
evident from NPS analysis are discussed. noise. Compared with B10s, images reconstructed
The ICRU/AAPM polyethylene phantom (see with the sharper B80s kernel will have better
Figure 7.19) was used in this work. Figure 11.10 spatial resolution (MTF) but will also have more
illustrates the axial (fx – fy), sagittal (fz –fy), and noise at higher frequencies.
coronal (fx – fz) views of the 3D NPS for two differ- The NPS plots shown in Figure 11.11 contain sig-
ent reconstruction kernels on a Siemens ASþ nificant amounts of information, which can be dis-
scanner. In the axial displays, zero frequency is at played more concisely as shown in Figure 11.12.
the center of the torus, and higher frequencies Figure 11.12a illustrates the radially averaged NPS
advance radially (see axes in Figure 11.5c). The for the B10s kernel; the three plots in
images in Figure 11.10a correspond to a reconstruc- Figure 11.12a correspond to the data regions high-
tion kernel (B10s) that provides considerable smooth- lighted with the horizontal bars in Figure 11.11a.
ing, thereby reducing much of the high-frequency The 1D NPS profiles for kernel B80s are shown in
noise in the image data sets. The images in Figure 11.12b. These plots illustrate curves of the
Figure 11.10b correspond to a high-resolution recon- 3D NPS at fz ¼ 0, fz ¼ 12 fN, and fz ¼ fN, where fN is
struction kernel (B80s), and the NPS demonstrates the Nyquist frequency (see Section 10.3). It is
127
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
Figure 11.10. Full displays of the 3D NPS. (a) The 3D NPS for a B10s kernel on a Siemens scanner. These data were acquired using
the ICRU/AAPM phantom. The B10s kernel produces relatively smooth images by attenuating the higher-frequency content in the
images. This manifests in the 3D NPS as a small-diameter torus in the axial (fx – fy) display. The center of the torus corresponds to fx ¼
fy ¼ 0, with higher frequencies advancing away from the center. The coronal and sagittal NPS displays show the elongation of the basic
torus in the orthogonal dimensions. (b) The 3D NPS for a high-pass kernel (B80s). The diameter of the torus is greater than that in (a),
corresponding to the shift to higher spatial frequencies of the NPS with this kernel.
recognized that the full family of curves produced the NPS of the B80s kernel is seen to have signifi-
from the data illustrated in Figure 11.11a would cantly more area compared with that with the B10s
span the spaces between the curves illustrated kernel, and the area of these curves is proportional
in Figure 11.12a. It is important to note that to s2 [Eq. (11.8) applies here as well, except that
the ordinate of Figure 11.12b is scaled differently the integration over z has already been performed].
from that in Figure 11.12a, which is necessary Figure 11.14 illustrates the 3D NPS data,
because the B80s kernel not only produces higher- showing the frequency fz as a function of the radial-
frequency image noise than does the B10s filter (it ly averaged frequency fr, for three different recon-
is shifted to the right), but it also has higher noise structed slice thicknesses (1 mm, 2 mm, and
(the variance is larger) for the same acquisition 5 mm). The maximum frequency computed (fN)
parameters. along the fz axis in these plots corresponds to
Figure 11.13 gives NPS plots for the (Siemens) (2t)21, where t is the slice thickness. Figure 11.15a
kernels B10s and B80s, and two others, for which shows the NPS amplitude as a function of the
the data illustrated in Figure 11.11a and 11.11b radial frequency fr, integrated over fz. Similarly, the
were integrated over fz, i.e., 3D NPS data can be integrated horizontally (see
Figure 11.14) over the radial frequency fr, as
ð fN
ð fN
NPSð fr Þ ¼ NPSð fr ; fz Þdfz ; ð11:9Þ
fN NPSð fz Þ ¼ NPSð fr ; fz Þdfr : ð11:10Þ
fN
128
Noise Assessment in CT
Figure 11.11. Radially averaged 3D NPS plots, obtained from the symmetrical (in fx and fy) 3D NPS in Fig. 11.10. Here the NPS
amplitude is shown using a color scale. The plots extend from f ¼ 0 to the Nyquist frequency for each axis. (a) NPS data from the
low-pass B10s kernel. (b) NPS of the high-pass B80s kernel.
Figure 11.12. NPS(fr) for three locations along the fz axis: fz ¼ 0, fz ¼ 12 fN, and fz ¼ fN, where fN is the Nyquist frequency. NPS(fr) for (a)
the B10s kernel and (b) the B80s kernel.
(126 mA s) produces about 25 % of the x-ray over fz. These scans used the same set of technique
fluence compared with the higher current –time factors [tube potential, tube current, slice thick-
product (500 mA s), and the noise variance scales ness, etc., and with unit pitch ( p ¼ 1)]; however,
inversely, as is apparent in this figure, i.e., the area the noise variance in the helical scan is slightly
under the 126 mA s curve is about four times greater than that of the axial scan.
greater than that of the 500 mA s curve. Figure 11.18 shows NPS curves calculated at
Figure 11.17 compares NPS curves for axial (se- four different radii. These are 3D NPS data inte-
quential) and helical (spiral) acquisition with the grated over fz [Eq. (11.9)] for the four different
Siemens ASþ platform, using the ICRU/AAPM radii. As mentioned previously, the areas under
polyethylene phantom. Figure 11.17 shows the these curves are proportional to s2. The trends in
NPS resulting from the integration of the 3D NPS this figure show that the NPS computed from VOIs
129
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
near the center of the field (e.g., r ¼ 30 mm) have comparing protocols (including reconstruction-
slightly higher noise variance than measurements kernel selection) among CT scanners from different
at larger radii (e.g., r ¼ 97 mm). Although the vendors. Figure 11.19 gives an important compari-
radial dependence of the amplitude of the NPS is son of NPS curves between two different vendors:
small, these results do suggest that the use of a General Electric shown on the top row and
standard radius for NPS evaluation would be more Siemens on the bottom row. The low-pass (softer)
consistent for comparisons across other parameters kernels are illustrated in the left column of plots,
(kernel, scanner type, tube current, etc.). and the high-pass (sharper) kernels are illustrated
As there is a need to develop more uniform inter- in the right column of plots. Using the NPS,
pretation of performance among CT-scanner plat- Solomon et al. (2012) performed a numerical com-
forms, the NPS can also serve an important role in parison among the performances of the different re-
construction kernels of these two vendors. These
data provide important information with respect to
matching protocols across these two specific CT
scanner platforms. Matching the reconstruction
kernels between vendors as done by Solomon et al.
(2012) is a practical example of the utility of the
NPS(f ) in the clinical environment, over and above
the straightforward assessment of noise.
Figure 11.14. Examples of the radially averaged 3D NPS for slice thicknesses of (a) 1 mm, (b) 2 mm, and (c) 5 mm. The plots extend
from f ¼ 0 to the Nyquist frequency for each axis.
130
Noise Assessment in CT
Figure 11.15. The 2D NPS from the integral of the 3D NPS. (a) The 2D NPS computed from the 3D NPS by integration over fz. In this
display, as a function of fr, the area of the NPS area is proportional to the noise variance. The results for the 1 mm slice thickness show
approximately twice the noise variance compared with that for the 2 mm slice, as would be expected based upon the relative number of
x-ray quanta that contribute. These curves were produced by vertically summing the data shown in Fig. 11.14 and then multiplying the
sum by Dfz. (b) The 2D NPS computed from the 3D NPS by integration over fr, as a function of fz. Curves for four different slice
thicknesses are shown. Smaller slice thicknesses are capable of displaying higher spatial frequencies in the fz dimension.
Figure 11.16. The NPS for two products of tube current and scan Figure 11.18. Dependence of the NPS on the radius of
time. Here, the area of the lower-dose curve (126 mA s) is measurement. Four curves are given for the radii of
approximately four times that of the higher-dose curve (500 mA s). measurement indicated. Although the differences are small,
these data suggest that a consistent radius of measurement
should be used to preserve NPS measurement precision.
g2 MTF 2 ð f Þ
Figure 11.17. Comparison of the 2D NPSs for axial and helical NEQð f Þ ¼ SNR2out ð f Þ ¼ ; ð11:11Þ
CT. NPSs were integrated over fz. For the same technique
NPSð f Þ
factors as with the axial scan (and with pitch, p ¼ 1), the helical
(spiral) NPS has slightly greater noise variance as seen by the where g is used to normalize units, and is typically
larger area under the curve. equal to the mean gray scale (HU in CT) of the ROIs
used to compute the noise-power in the image.
imaging system passes signal. The NPS(f ) has Equation (11.11) is general and has been used ex-
been discussed at length in Section 11, and it tensively in the analysis of planar radiographic
describes how an imaging system passes noise (or, and mammography systems. Investigators in CT
more specifically, noise variance). The concepts of (Wagner and Brown, 1982) measured the frequency-
MTF(f ) and NPS(f ) can be brought together to de- dependent NEQ(f) for a second-generation CT
scribe the frequency-dependent SNR(f ); however, it system, and their result is illustrated in Figure 11.20.
131
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
Figure 11.19. Comparisons of the 2D NPS from General Electric and Siemens CT systems. These measured NPS curves have utility in
clinical protocol development across vendor platforms. Adapted from Soloman et al. (2012). (a) The NPS for the low-pass General
Electric reconstruction kernels. (b) The NPS for the high-pass General Electric kernels. (c) The NPS for the low-pass Siemens kernels.
(d) The NPS for the high-pass Siemens kernels.
132
Noise Assessment in CT
The ratio, therefore, between the MTF 2 and NPS An alternative to the use of the DQE(f ) for the
greatly mitigates the influence of the kernel, and study of noise versus dose in CT is possible,
demonstrates the fundamental NEQ(f) [i.e., SNR2out ] however, and the phantom described in Section 7
performance of the CT scanner. (see Figures 7.19a and 7.19b) is an important tool
for this. The phantom was designed as both a dos-
imetry phantom and a tool for measuring the
NPS(f ). For a specific set of acquisition parameters
11.6 Dose-Normalized NPS(f ) (tube potential, tube current, rotation time, pitch,
field of view, bow-tie filter, reconstruction kernel,
Large differences are seen in the values of etc.), the NPS(f ) is characterized at a standard air
NPS(f ) as a function of the tube-current– time kerma at the center of the phantom. The proposed
product; see Figure 11.16, where results for a factor measurement procedure is as follows:
of 4 difference in tube-current – time product (and
radiation dose) are illustrated. This observation sug- (1) The center section of the ICRU/AAPM
gests that the NPS(f), if measured at a standard phantom, a cylinder of polyethylene 200 mm long
dose value, might provide an excellent approach for and 300 mm in diameter, is positioned on the CT
comparing noise versus dose among CT scanner table with a thimble chamber located in the center
models. In radiography and other planar-imaging hole of the phantom, and also at the center along
modalities, the detective quantum efficiency, DQE(f), the z axis of the phantom as shown in
is a quantity that essentially describes the dose effi- Figure 7.22a.
ciency of the detector system. That is, for a given (2) Air-kerma measurements are made for an
dose, what SNR2out ð f Þ does the imaging system axial or helical scan acquired throughout the entire
produce? The DQE(f) is simply the NEQ divided by length of the phantom, using either an integrating
the incident fluence (Dainty and Shaw, 1974): electrometer or an electrometer with real-time read
out as described in Section 5.2.3. The scan is per-
SNR2out ð f Þ g2 MTF 2 ð f Þ formed at the desired technique factors (tube po-
¼ DQEð f Þ ¼ ; tential, rotation time, pitch, and x-ray-tube current
SNR2in ð f Þ qNPSð f Þ
of J1, etc.), and the air kerma, K1, measured at the
ð11:12Þ center of the phantom during the scan is recorded.
The x-ray-tube current is adjusted to a new value
where q is here the x-ray-photon fluence incident on
J2, with the intention of producing a specific mea-
the detector used to produce the image for which
sured air kerma, Kset, at the center of the phantom.
NPS(f) was measured. For Poisson noise, SNRin
pffiffiffi Because air kerma is linearly proportional to tube
¼ q, and hence SNR2in ð f Þ ¼ q. The incident photon
current when all other technique parameters are
fluence in CT has little spatial correlation, and
fixed, the selected tube current J2 can be deter-
hence its frequency distribution is “white,” meaning
mined using
that the input SNRin(f) has constant amplitude as a
function of spatial frequency. That is why the q term
on the right side of Eq. (11.12) has no explicit fre- Kset
J2 ¼ J1 ; ð11:13Þ
quency dependence. K1
Despite the utility of the DQE(f ) in planar x-ray
imaging systems, the meaning of q is difficult to in- where Kset could be a typical value of air kerma
terpret for CT systems (Tapiovaara and Wagner, used in body CT, such as 10 mGy.
1985) due to the presence of the bow-tie filter, (3) After J2 is selected, a second CT scan of the
which spatially modulates the value of q in the entire length of the phantom is acquired as before,
field of view. In addition, unlike the simple digital- and the measured air kerma, K2, is recorded. The
image corrections (e.g., flat-field) performed with value of K2 should be very close (within from 2 % to
digital radiography systems, the reconstruction 5 %) to the desired air kerma, Kset. Using this ac-
algorithms used in CT include corrections for quisition, images are then reconstructed using the
beam-hardening, x-ray scatter, and other effects desired reconstruction parameters (slice thickness,
that consider the presence of an object in the kernel, display field of view, etc.).
scanner. CT scanners are simply not designed to (4) The CT images produced in step 3 above are
produce optimal images when no object is present, used to evaluate the 3D NPS(f), and—after radially
and therefore it would be impractical to use scans averaging using Eq. (11.2) and integration over fz
in air to characterize CT scanner performance. using (Eq. 11.9)—NPS curves similar to those shown
Given these observations, the use of the DQE(f ) for in Figures 11.13, 11.15a, and 11.16 are produced and
CT appears to be limited at this time. used for comparative analysis.
133
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
With the above procedure performed on any CT The NPS(fr) curves, measured from the full 3D
manufacturer’s platform, NPS(f ) curves can be NPS with subsequent radial averaging [Eq. (11.2)]
compared at essentially the same air-kerma levels and integration over fz [Eq. (11.9)] should be
at the center of a standard phantom. This approach adopted for the evaluation of the noise performance
allows for a direct comparison of the noise perform- of CT scanners. NPS curves, similar to those shown
ance among two or more whole-body CT scanners. in Figures 11.13 and 11.15 –11.17, are considered to
be most useful in the evaluation of the noise prop-
erties of MDCT systems. This is because, in this
11.7 Summary
display format, the area under each curve is pro-
Contrast resolution in CT has traditionally been portional to s2. The methods described in Section
measured using so-called low-contrast test objects, 11.6 provide guidance in this manner. The 300 mm
which are widely available in commercial phantoms diameter polyethylene ICRU/AAPM phantom,
(see Bushberg et al., 2012). This method requires scanned to produce an air kerma of 10 mGy at the
specialized phantoms and relies on subjective center, is appropriate for body-imaging-protocol as-
(human-observer) interpretation of patterns similar sessment. A 200 mm long, 180 mm diameter poly-
to the CD phantom illustrated in Figure 11.2a. ethylene phantom would be appropriate for head
Moreover, in addition to the somewhat qualitative CT protocols, with 50 mGy air kerma as the target
observation used as a measure, vendors can report level.
the measurement made under differing conditions, These recommendations apply not only to the
obscuring the ability to make meaningful compari- clinical evaluation of CT-scanner noise-versus-dose
sons among different scanners. This situation performance, but also extend to data provided by
clearly needs to be updated to more modern, quan- CT manufacturers in commercial, technical, and
titative, and objective metrics. marketing documents. It is anticipated that the
Thus, it is recommended that quantitative meas- measurement of NPS(f ) as outlined above will sup-
urement of the NPS, using the middle section plant the use of visual assessment of low-contrast
of the 300 mm diameter polyethylene ICRU/AAPM test phantoms in acceptance-testing and quality-
phantom evaluated at a radius of approximately assurance procedures for CT.
100 mm, be used to quantify noise in CT systems.
134
Journal of the ICRU Vol 12 No 1 (2012) Report 87 doi:10.1093/jicru/ndt003
Oxford University Press
This Section summarizes the most significant isocenter only. Characterization of CTDIvol on
recommendations discussed in this Report. Sections newly installed CT platforms provides continuity
2–4 are introductory in nature, and were provided with measurements from previous CT scanners,
to acquaint those less familiar with current CT even as new CT dosimetry measures are introduced
technology, clinical usage, and conventional CT and used with more frequency. The rationale for
dosimetry associated with this modality. this recommendation is also found in Sections 7
and 8 of this Report, in which the CTDIvol is used
to scale the size-specific dose estimate, SSDE, and
12.1 Radiation-Dose Assessment in CT to normalize rise-to-equilibrium curves. If the
recommendations of Section 12.1.3 below are fol-
Sections 4 – 9 discuss matters pertaining to radi- lowed, CTDIvol measurements performed after the
ation dosimetry, and recommendations associated initial acceptance testing become unnecessary;
with each of these Sections will be provided here in however, if the procedures in Section 12.1.3 are not
order. In the discussion below, the term acceptance followed, then continued periodic evaluation of
testing refers to the initial measurements per- CTDIvol during quality assurance is recommended.
formed on a newly installed CT scanner, usually
prior to clinical use. Quality assurance is the 12.1.2 CT X-Ray-Spectrum Characterization
process of periodic evaluation of a CT scanner, and
is often performed annually or after a major service The CT x-ray spectrum is dependent on the tube
event (such as an x-ray-tube change). potential and its waveform, and the amount of
For dosimetry in CT, many Monte Carlo and total filtration in the x-ray beam at the center of
other studies describe the absorbed dose in a par- the fan beam. All modern CT systems make use of
ticular organ or site (e.g., liver, bladder, fetus) as a high-frequency generators that produce very little
function of air kerma at the isocenter of the voltage ripple in the tube potential, so this is a
scanner. When cast in this manner, the dose coeffi- minor factor in today’s systems. Non-invasive
cients take on the units of absorbed dose per air x-ray-tube-potential meters can be used to deter-
kerma, typically in units of mGy/mGy. mine the accuracy of the tube-potential reading on
the console. The total filtration can be evaluated
with knowledge of both the tube potential and the
12.1.1 Existing CT-Dosimetry Methods
measured half-value layer, HVL.
Historical acceptance testing and quality-control In general, the current literature tends to report
procedures for CT systems typically involve the dose coefficients (e.g., absorbed dose in an organ or
assessment of the CTDIvol for several standard pro- phantom location as a function of CTDIair, the air
tocols. The set of ACR quality-control recommenda- kerma at isocenter) for the specific manufacturer
tions is one example. In addition, based on IEC and model of the CT scanner (Boone et al., 2004;
(2009) recommendations, CTDIvol is displayed on Zhou and Boone, 2008). However, the numbers of
the console of all modern CT scanners. Due to its manufacturers and models of CT scanners are
broad utility and ubiquitous availability, it is growing, and scientific reports in the future will
strongly recommended that the CTDIvol be mea- have more impact if, in addition to reporting dosim-
sured at acceptance testing using both the 160 mm etry values, they report dose-conversion values as a
diameter and 320 mm diameter PMMA phantoms, function of x-ray-tube potential and HVL of the
at a clinically relevant tube-current – time product x-ray beam for the central ray. Although different
(e.g., 100 mA s) over the range of tube potentials CT-scanner vendors use slightly different source-
used by the CT scanner. Measurement of the to-isocenter distances and bow-tie filter designs, in
CTDIvol in air CTDIair is also strongly recom- general these influences will likely have a small
mended, and this measurement is made at the impact on the dose-conversion factors. Therefore,
136
Summary of Recommendations
practical for routine testing of CT systems in the below. An additional function of the phantom meas-
field. However, this phantom does have two import- urement is to evaluate the constancy of the tube-
ant roles to play: current–time product required to produce a stand-
(1) It is strongly recommended that the full ard air-kerma value at the center of the phantom,
600 mm long phantom be available at manufac- which would serve as an additional quality-control
turer’s CT-testing facilities, and that thorough mea- parameter. For example, if an effective “mAs” of Jt
surements (at both the center and the peripheral is required for an air kerma of 10 mGy at the
locations) be performed using the phantom, such center of the phantom at one time, and 0.80 Jt is
that H(L) and related functions [h(L), G(L), etc.] required for the same air kerma at a later time,
can be provided for each CT-scanner model. then this is cause for concern and suggests that the
Comprehensive measurements should be measured x-ray-tube output be further investigated.
at all tube-potential settings, with all appropriate
bow-tie filters, and over a realistic range of collima-
tion settings. The use of the real-time-probe meas-
12.1.5 Patient SSDE
urement procedure as outlined in Section 7 will
expedite these measurements. Normalization of the Section 8 describes methods that can be used to
acquired h(L) curves by CTDIvol leads to G(L) compute conversion coefficients that can be used
curves that would be useful for the assessment of with the known CTDIvol to estimate the average
patient dose in the clinical setting. The large mass absorbed dose to patients for a scan of standard
of the full phantom should not be an impediment length. The SSDE provides a more accurate esti-
for use in a factory environment. mate of patient absorbed dose, and utilizes data
A subset of the tests with the 600 mm long that are available in the CT-image data set; specif-
phantom are strongly recommended to be made by ically, the recorded CTDIvol and the water-
CT manufacturers for each CT model using only equivalent diameter of the patient that can be
the 200 mm long center section of the ICRU/AAPM estimated directly from the CT images. The math-
phantom. For example, with a thimble chamber ematical methods pertinent to the conversion
placed in the center hole of the phantom at the process are described in Section 9.
center of the hole along z, the air kerma should The effective diameter of the patient was defined
be measured for a helical acquisition (with pitch for quantifying patient “size,” when an axial CT
p 1) performed over the entire length of the image or set of CT images is available. The effect-
phantom. This measurement should be performed ive diameter, deff, is appropriate for abdomen and
for all x-ray-tube potentials at a standard tube- pelvis imaging, for which only a small amount of
current – time product (e.g., 100 mA s). In addition, internal air exists in body cavities. For thoracic CT
at a standard tube potential (e.g., 120 kV), this scans, however, the large amount of air in the
measurement should be repeated for the three lungs needs to be accounted for, and therefore the
most commonly used collimation settings (e.g., at water-equivalent diameter, dw, should be used for
Cmax, 12 Cmax, and 14 Cmax, for a scanner with a the thorax. Given the similarity between the calcu-
maximum collimation of Cmax). The data generated lations for deff and dw, it is highly recommended
by this measurement protocol will allow for direct that the water-equivalent diameter be used for
comparison in the field, using the same 200 mm computing the SSDE, for the head, neck, thorax,
long phantom and measurement protocol. These abdomen, and pelvis.
data also provide a means to relate the partial- The SSDE can also be used for estimating dose
phantom measurements to the G(L) curves after the localizer view is acquired, but prior to the
acquired in the 600 mm long phantom [e.g., CT scan. In this case, the lateral and/or AP dimen-
G600(L)/G200(L)]. sion of the patient needs to be used to compute dw.
(2) For the evaluation of CT scanners in the clin- Even after the CT images become available, one or
ical environment, it is strongly recommended that more edges of the patient in axial CT images might
only the middle section of the ICRU/AAPM be cut-off (i.e., a portion of the patient’s anatomy is
phantom be used. This phantom is 200 mm long outside of the field of view) due either to clinically
and 300 mm in diameter, and is slightly lighter appropriate use of small-display fields of view or
than the ubiquitous 320 mm diameter PMMA when a very large patient is scanned. This situ-
phantom. The principal function of the phantom is ation prevents the accurate assessment of the area
to measure image noise: the noise-power spectrum, of the patient in the CT image, and hence compro-
NPS, at a constant air kerma at the center hole of mises the ability to estimate of dw from the CT
the phantom, as described in Section 11.6. This will image. In this case, the localizer view provides the
be discussed in greater detail in Section 12.2.2 necessary data for estimating patient size using the
137
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
PA or lateral dimension of the patient, one of which an average SSDE (and standard deviation) for each
is generally visible on the localizer image. region should be reported. It is recognized that the
To overcome the problem of determining dw when absorbed dose associated with a given CT slice
a small reconstructed CT field of view prevents as- along z is not only dependent on the dw and effect-
sessment of the patient’s complete anatomy, it is ive “mAs” for that slice, but that the scatter tails
strongly recommended that CT manufacturers from adjacent CT slices contribute a significant
provide pre-calculated dw values for each CT- fraction of dose as well. However, dw and effective
section image during the reconstruction process. “mAs” do not change dramatically from slice to
The CT-scanner hardware would compute the full slice along the z axis. Section 9.5 demonstrates a
field-of-view (FOV) CT images, determine the dw software-based approach for the computation of
for each axial image, and report this in the DICOM SSDE on a CT image-specific basis.
header. The full-FOV images need not be stored
unless this is desired by the user. This dw computa-
tion includes many elements of the standard recon-
12.2 Other Considerations in Patient
struction process for smaller-FOV images, and so
Dosimetry in CT
the additional computational burden would be
light. Proprietary or vendor-specific algorithms for The current state of science does not allow the
computation of dw are discouraged, because all estimation of the radiation risk for a specific
aspects of patient dosimetry should be transparent patient, but it is known generally that the higher
and verifiable by individuals responsible for CT the absorbed dose the higher the risk of adverse
dose assessment at a given institution. If dw were health effects. The effective dose, E, defined by the
available for each axial CT image in the DICOM ICRP (ICRP, 2007) for the purpose of radiation pro-
header, then image cut-off due to the use of a small tection and discussed in Section 4, is based on the
FOV would not be an issue, and subsequent dose absorbed dose in sensitive organs and tissues.
assessment would be expedited considerably. When the DLP is known, so-called k-factors can be
used to estimate E [see Eq. (4.11)]. However, such
information cannot be used to evaluate individual
12.1.6 Automatic Exposure Control in CT
risk because effective dose applies only to a popula-
The protocol for computing the SSDE is tion of adults, averaged for males and females, and
described in Section 8, and the most basic computa- is intended to provide a tool for risk management
tion of SSDE is given by Eq. (9.1). This equation (limitation and optimization) for regulatory
assumes a patient of constant diameter and a CT purpose. Organ and tissue absorbed doses for indi-
scan with constant tube current. Although this ap- viduals undergoing a radiological procedure such
proach is adequate for constant-tube-current oper- as CT are no doubt fundamental, but the relation-
ation and for general prospective-patient dose ships that govern the radiation response of a specif-
calculations (i.e., for research or clinical CT proto- ic patient (depending on age, sex, size, health
cols for which patient size can only be assumed), history, genetic nature, etc.) are still largely
the complexities of size dependence in CT dosim- unknown.
etry combined with the increasing use of tube- Although effective dose is not suitable for the as-
current-modulation-acquisition modes require a sessment of the risk for individual patients, it can
more sophisticated approach. Furthermore, be useful when comparing different diagnostic mo-
Figure 8.9a shows that for the abdomen –pelvis dalities and techniques. The assessment of organ
scan, a significant difference can occur when a doses does allow, however, the use of organ-, sex-,
single value of dw is computed from a single CT and age-specific coefficients provided in the BEIR
image near the center of the scan (in z), when com- VII report (NAS/NRC, 2013) for the estimation of
pared with the slice-by-slice determination of the the putative risk to a smaller population and thus
average dw. Therefore, for comprehensive patient- not all adults, but not an individual. Although
dose estimation, it is highly recommended that the Monte Carlo calculations of organ absorbed dose
average dose be computed as the average SSDE are useful in developing CT-dose coefficients for
value across all images in the CT scan using the generic populations (Zhang et al., 2012), they are
CT-image specific value of dw and the not necessary for the individual patient. A soft-
CT-image-specific effective “mAs” or the CTDIvol ware-based approach to assessing the absorbed
[see Eq. (9.3)]. Thus, when computational capabil- dose for a specific CT procedure is described in
ities are available, the SSDE should be computed Section 9.6 and is illustrated in Figure 9.4. It is
on a CT-image-specific basis for a given region in expected that the image-by-image computation of
the body (e.g., abdomen, pelvis, thorax, head) and the SSDE would be useful in organ-dose estimation
138
Summary of Recommendations
using probability-density functions (see Figure used reconstruction kernels for that scanner.
9.10) that describe the fraction of an organ’s Measurement of the in-plane MTFxy as a function
volume that resides in each CT section. The prob- of the display field of view is also strongly recom-
ability-density-function concept can be developed mended. The longitudinal MTFz should be mea-
using gender-specific organ templates ( probability sured instead of the slice-sensitivity profile; in most
of a specific organ being located along the z axis), applications the slice-sensitivity profile should be
which can be aligned (in z) with a specific patient’s retired as a CT-resolution assessment along the
anatomy in the CT image using boney landmarks z axis.
(e.g., vertebral bodies). To increase precision, the MTF should be mea-
sured in images produced with low noise levels.
Therefore, in addition to using a high tube-current–
12.3 Image-Quality Metrics time setting for the scan, a small-diameter phantom
is useful because of its lower attenuation. A small-
Image quality in CT is a complex topic; in this
diameter prototype phantom is illustrated in
Report the emphasis is on the development and use
Figure10.15b and is pictured in Figure 10.15c.
of more quantitative metrics for image-quality as-
The measurement of the MTFxy in CT images
sessment rather than historical, qualitative
that are reconstructed using non-linear iterative
methods. For both spatial resolution and noise as-
methods needs to be performed with caution, as the
sessment, proposed methods require access to the
MTF construct assumes a linear, stationary
CT-image digital data as well as the ability to ma-
system. Therefore, it is highly recommended for
nipulate them with custom software. Although
general quality assurance that the MTF be charac-
some of the mathematical details of these algo-
terized using images that are reconstructed by fil-
rithms have been outlined in this Report, it is
tered back projection. In such cases, it is
anticipated that both open-source and commercial
understood that these MTFs represent an upper
software will eventually be available for the recom-
bound on the spatial resolution of the scanner.
mended analyses discussed here.
Iterative-reconstruction techniques generally trade-
off spatial resolution to reduce noise in an adaptive
12.3.1 Spatial Resolution in CT
(local) manner, and any attempt to quantify the
A reasonable goal in characterizing spatial reso- MTFs when iterative-reconstruction methods are
lution in CT is to perform the assessment in 3D. used should be considered to be valid only within a
However, there are considerable differences in the small (local) region in the image.
resolution properties between the in-plane MTFxy
and the longitudinal MTFz, and the factors respon-
12.3.2 Noise Assessment in CT
sible for these differences are well understood.
Therefore, the independent characterization of the The noise texture in CT is well characterized by
MTFxy and MTFz is currently recommended; the 3D NPS. The measurement of MTF requires
however, the full 3D MTF in CT can be of value in points, lines, or edges in the phantom for resolution
certain settings. measurement, but the evaluation of the NPS is
The in-plane CT image is (in general) rotational- performed using a homogeneous phantom with no
ly symmetric, in the sense that the x and y dimen- specific structures. This makes phantom fabrica-
sions are subject to identical factors that influence tion much easier than for the case of the MTF.
spatial resolution when filtered-back-projection However, the phantom for NPS assessment should
methods are used for image reconstruction. Therefore, approximate the size of patients because noise (and
the modulation-transfer function (MTFxy) can be com- noise texture) is directly related to the size of the
puted for the axial plane, and the frequency-domain phantom. To address this, the ICRU Report
variables (fx and fy) can be described using a single Committee and AAPM Task Group 200 developed a
variable fr. The measurement of the in-plane MTFxy is 300 mm diameter, polyethylene phantom that
fundamentally influenced by the reconstruction serves as both a dosimetry phantom and a
kernel used, whereas the longitudinal MTFz is almost phantom for the measurement of the 3D NPS (see
independent of the kernel and is more dependent on Figures 7.19 and 7.20). It is highly recommended
CT-section thickness. that the fully 3D NPS be measured using volumes
It is highly recommended that during acceptance of interest [see Eq. (11.3)]; methods for radially
testing of a newly installed CT scanner, both the averaging [Eq. (11.2)] and integration over fz [see
in-plane and longitudinal MTFs be measured as Eq. (11.9)] should be used for the presentation of
outlined in Section 10. The in-plane MTFxy should the axial-image NPS. Given the slight dependence
be characterized using a few of the most commonly of the NPS on the radius of measurement (see
139
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
Figure 11.18), it is highly recommended that a use of the real-time probe for rapid single-
radius of approximately 100 mm be employed when acquisition characterization of H(L). A series of dif-
the 300 mm diameter ICRU/AAPM polyethylene ferent normalizations leading to h(L) and G(L)
phantom is used. were also discussed. What is still needed is for
Noise and absorbed dose are intrinsically linked these metrics to be included in a practical CT-dos-
because they are fundamentally related to the x- imetry methodology. Although characterization of
ray fluence used during image acquisition; conse- H(L) is the first step toward understanding the im-
quently, measuring the NPS in the absence of portant dosimetric consequence of scan length, a
absorbed-dose information is of little value. To straightforward recipe for how the calculation of an
address this, the protocol specified in Section 11.6 individual’s dose should be computed is the subject
is highly recommended. In that protocol, the center of future research. A likely correction factor, V(L),
section of the ICRU/AAPM phantom is scanned is envisioned:
over its entire length using helical (spiral) or axial
(sequential) scanning at a fixed air-kerma level at HðLÞ
VðLÞ ¼ : ð12:1Þ
the center of the phantom. The measurement of Hð100 mmÞ
the NPS at standardized air-kerma levels in the
phantom is meant to provide practitioners with the It is recognized that H(L) describes the dose (at the
means to compare the NPS among different CT periphery or center) at the center slice (i.e., at z ¼
scanners, and to monitor a specific scanners’ per- 0) of a CT scan of length L. Future studies should
formance periodically. For consistency testing, NPS develop methods that lead to accurate estimates of
curves similar to those shown in Figure 11.16 the dose along the entire length of the CT scan.
provide a more quantitative approach for noise The convolution method described in Eq. (7.4) gives
assessment in CT, compared with subjective guidance in this manner, and other investigators
visual assessment of a low-contrast test object. are studying numerical approaches to this as well
Comparison of NPS curves among CT scanners (see (Li et al., 2012).
Figure 11.19) is also an important aspect in proto-
col development in an environment in which mul-
12.4.3 Incorporation of Scan-Length
tiple models (and manufacturers) of CT scanners
Corrections to the SSDE
are used.
The SSDE methods described in Section 8 and in
AAPM Report 204 (AAPM, 2011) allow the estima-
12.4 Future Directions tion of SSDE based upon the known CTDIvol and a
conversion coefficient. The conversion coefficients
12.4.1 Real-Time Probe [see Figure 8.10 and Eq. (9.1)] were developed by
The modern CT system is a dynamic platform: combining the work of a number of investigators,
not only is the beam direction changing rapidly, but and include the effect of scan length for standard
the beam intensity also can change rapidly. Given studies and the conversion from air kerma to
this dynamic environment, it is unrealistic to con- absorbed dose in tissue. Although the method is
tinue to use radiation meters that operate only in considered sufficiently accurate for current pur-
an integration mode. The dynamic nature of CT poses, it is not robust in the case of large deviations
requires dynamic, real-time-measurement technol- in scan length because of the implicit assumptions.
ogy. The real-time probe, discussed in Section 5.2.3, It should be recognized that both the SSDE and
is a necessary tool for the evaluation of CT dose in the H(L) provide tools that address the size of the
the modern era. A number of measurement proto- patient over the length of their CT scan;
cols have been described in this Report that rely the SSDE addresses the diameter of the patient,
upon the use of a real-time radiation dosimeter. and the H(L) curve suggests a process [e.g., poten-
Such probes are certain to become an essential tool tially using Eq. (12.1)] for adjusting patient dose
for the efficient and comprehensive characteriza- based upon the length of the CT scan. Because the
tion of CT scanners. conversion coefficients associated with SSDE in-
corporate the assumption of a “standard” scan
length, it is questionable whether or not the appli-
12.4.2 Rise-to-Equilibrium Curve, H(L)
cation of an additional scan-length correction
AAPM Report 111 (AAPM, 2010) describes a factor would improve the accuracy of the existing
method for characterizing the rise-to-equilibrium SSDE methods. Furthermore, the shape of the
curve, H(L). Section 7 of this Report extends the H(L) curve is diameter-dependent [see Figure 7.28
approach of AAPM Report 111 by introducing the for related G(L) curves], and the curve shape is
140
Summary of Recommendations
also dependent upon location in the axial plane studies might prospectively develop conversion
(i.e., center or peripheral positions; see coefficients that address differences both in diam-
Figure 7.29). Despite these complexities, future eter and scan length within the same construct.
141
Journal of the ICRU Vol 12 No 1 (2012) Report 87 doi:10.1093/jicru/ndt005
Oxford University Press
References
AAPM (1977). American Association of Physicists in Baek, J., and Pelc, N. J. (2011a). “Effect of detector lag
Medicine. Phantoms for Performance Evaluation and on CT noise power spectra,” Med. Phys. 38,
Quality Assurance of CT Scanners. AAPM Report No. 2995–3005.
1 (American Association of Physicists in Medicine, Baek, J., and Pelc, N. J. (2011b). “Local and global 3D
College Park, MD). noise power spectrum in cone-beam CT system with
AAPM (1990). American Association of Physicists in FDK reconstruction,” Med. Phys. 38, 2122 –2131.
Medicine. Standardized Methods for Measuring Bischof, C. J., and Ehrhardt, J. C. (1977). “Modulation
Diagnostic X-ray Exposures, AAPM Report No. 31 transfer function of the EMI CT head scanner,” Med.
(American Association of Physicists in Medicine, Phys. 4, 163 – 167.
College Park, MD). Bittencourt, M., Schmidt, B., Seltmann, M., Muschiol, G.,
AAPM (1993). American Association of Physicists in Ropers, D., Daniel, W., and Achenbach, S. (2011).
Medicine. Specification and Acceptance Testing of “Iterative reconstruction in image space (IRIS) in
Computed Tomography Scanners, AAPM Report No. cardiac computed tomography: initial experience,”
39 (American Association of Physicists in Medicine, Int. J. Cardiovasc. Imaging 27, 1081–1087.
College Park, MD). Boedeker, K. L., Cooper, V. N., and McNitt-Gray, M. F.
AAPM (2008). American Association of Physicists in (2007a). “Application of the noise power spectrum in
Medicine. The Measurement, Reporting and modern diagnostic MDCT: Part I. Measurement of
Management of Radiation Dose in CT, AAPM Report noise power spectra and noise equivalent quanta,”
No. 96 (American Association of Physicists in Phys. Med. Biol. 52, 4027–4046.
Medicine, College Park, MD). Boedeker, K. L., and McNitt-Gray, M. F. (2007b).
AAPM (2010). American Association of Physicists in “Application of the noise power spectrum in modern
Medicine. Comprehensive Methodology for the diagnostic MDCT: Part II. Noise power spectra and
Evaluation of Radiation Dose in X-Ray Computed signal to noise,” Phys. Med. Biol. 52, 4047– 4061.
Tomography, AAPM Report No. 111 (American Bongartz, G., Golding, S., Jurik, A., Leonardi, M., van
Association of Physicists in Medicine, College Park, MD). Persijn van Meerten, E., Rodrı́guez, R., Schneider, K.,
AAPM (2011). American Association of Physicists in Calzado, A., Geleijns, J., Jessen, K., Panzer, W.,
Medicine. Size-Specific Dose Estimates (SSDE) in Shrimpton, P., and Tosi, G. (2004). European
Pediatric and Adult Body CT Examinations, AAPM Guidelines for Multislice Computed Tomography
Report No. 204 (American Association of Physicists in (European Commission).
Medicine, College Park, MD). Boone, J. M. (1986). “Equivalent spectra as a measure of
Aberle, D. R., Adams, A. M., Berg, C. D., Black, W. C., beam quality,” Med. Phys. 13, 861 –868.
Clapp, J. D., Fagerstrom, R. M., Gareen, I. F., Boone, J. M. (1988). “The three parameter equivalent
Gatsonis, C., Marcus, P. M., and Sicks, J. D. (2011). spectra as an index of beam quality,” Med. Phys. 15,
“Reduced lung-cancer mortality with low-dose com- 304– 310.
puted tomographic screening,” N. Engl. J. Med. 365, Boone, J. M. (2007). “The trouble with CTDI100,” Med.
395– 409. Phys. 34, 1364 –1371.
Achenbach, S., Marwan, M., Ropers, D., Schepis, T., Boone, J. M. (2009). “Dose spread functions in computed
Pflederer, T., Anders, K., Kuettner, A., Daniel, W. G., tomography: A Monte Carlo study,” Med. Phys. 36,
Uder, M., and Lell, M. M. (2010). “Coronary computed 4547–4554.
tomography angiography with a consistent dose below Boone, J. M. (2010). “Method for evaluating bow tie filter
1 mSv using prospectively electrocardiogram-triggered angle-dependent attenuation in CT: Theory and simu-
high-pitch spiral acquisition,” Eur. Heart. J. 31, lation results,” Med. Phys. 37, 40– 48.
340– 346. Boone, J. M., and Seibert, J. A. (1997). “An accurate
Alvarez, R. E., and Macovski, A. (1976). method for computer-generating tungsten anode x-ray
“Energy-selective reconstructions in X-ray computer- spectra from 30 to 140 kV,” Med. Phys. 24, 1661–1670.
ized tomography,” Phys. Med. Biol. 21, 733 –744. Boone, J. M., Geraghty, E. M., Seibert, J. A., and
Baek, J., and Pelc, N. J. (2010). “The noise power spec- Wootton-Gorges, S. L. (2003). “Dose reduction in pedi-
trum in CT with direct fan beam reconstruction,” Med. atric CT: A rational approach,” Radiology 228,
Phys. 37, 2074 –2081. 352– 360.
Bracewell, R. N. (1956). “Strip integration in radio Deak, P. D., Smal, Y., and Kalender, W. A. (2010).
astronomy,” Aust. J. Physc. 9, 198. “Multisection CT protocols: Sex- and age-specific con-
Bracewell, R. N., and Riddle, A. C. (1967). “Inversion of version factors used to determine effective dose from
fan-beam scans in radioastronomy,” Astrophys. J. 150. dose–length product,” Radiology 257, 158 –166.
Brenner, D. J., and Hall, E. J. (2007). “Computed tomog- Dixon, R. L. (2003). “A new look at CT dose measure-
raphy—an increasing source of radiation exposure,” ment: beyond CTDI,” Med. Phys. 30, 1272– 1280.
N. Engl. J. Med. 357, 2277– 2284. Dixon, R. L., and Ballard, A. C. (2007). “Experimental
Brooks, R. A., and Chiro, G. D. (1976). “Statistical limita- validation of a versatile system of CT dosimetry using
tions in x-ray reconstructive tomography,” Med. Phys. a conventional ion chamber: Beyond CTDI100,” Med.
3, 237 –240. Phys. 34, 3399– 3413.
Burgess, A. E. (1999). “The Rose model, revisited,” Dixon, R. L., and Boone, J. M. (2010). “Cone beam CT
J. Opt. Soc. Am. A, 16, 633– 646. dosimetry: a unified and self-consistent approach in-
Callahan, M. J., Rodriguez, D. P., and Taylor, G. A. cluding all scan modalities—with or without phantom
(2002). “CT of appendicitis in children,” Radiology 224, motion,” Med. Phys. 37, 2703– 2718.
325– 332. Dixon, R. L., and Boone, J. M. (2011). “Analytical equa-
Christner, J. A., Kofler, J. M., and McCollough, C. H. tions for CT dose profiles derived using a scatter kernel
(2010a). “Estimating effective dose for CT using dose- of Monte Carlo parentage with broad applicability to CT
length product compared with using organ doses: dosimetry problems,” Med. Phys. 38, 4251–4264.
Consequences of adopting International Commission Dixon, R. L., Munley, M. T., and Bayram, E. (2005). “An
on Radiological Protection publication 103 or dual- improved analytical model for CT dose simulation with
energy scanning,” Am. J. Roentgenol. 194, 881 –889. a new look at the theory of CT dose,” Med. Phys. 32,
Christner, J. A., Stierstorfer, K., Primak, A. N., 3712–3728.
Eusemann, C. D., Flohr, T. G., and McCollough, C. H. Droege, R. T., and Morin, R. L. (1982). “A practical
(2010b). “Evaluation of z-axis resolution and image method to measure the MTF of CT scanners,” Med.
noise for nonconstant velocity spiral CT data recon- Phys. 9, 758 –760.
structed using a weighted 3D filtered backprojection Duan, X., Wang, J., Yu, L., Leng, S., and McCollough,
(WFBP) reconstruction algorithm,” Med. Phys. 37, C. H. (2011). “CT scanner x-ray spectrum estimation
897– 906. from transmission measurements,” Med. Phys. 38,
Christner, J. A., Zavaletta, V. A., Eusemann, C. D., 993– 997.
Walz-Flannigan, A. I., and McCollough, C. H. (2010c). EC (1999). European Commission. 1999 European
“Dose reduction in helical CT: Dynamically adjustable Guidelines on Quality Criteria for Computed
z-axis X-ray beam collimation,” Am. J. Roentgenol. 194, Tomography, Report EUR 16262 (European
W49– W55. Commission, Luxembourg).
COCIR (2000). European Coordination Committee of the Einstein, A. J., Elliston, C. D., Arai, A. E., Chen, M. Y.,
Radiological Electromedical and Healthcare IT Mather, R., Pearson, G. D. N., DeLaPaz, R. L.,
Industry. Radiation Exposure in Computed Nickoloff, E., Dutta, A., and Brenner, D. J. (2010).
Tomography, 2nd ed., Nagel, H. D., Ed. (COCIR, “Radiation dose from single-heartbeat coronary CT
Frankfurt). angiography performed with a 320 –detector row
Cody, D. D., Moxley, D. M., Krugh, K. T., O”Daniel, J. C., volume scanner,” Radiology 254, 698 –706.
Wagner, L. K., and Eftekhari, F. (2004). “Strategies for Faulkner, K., and Moores, B. M. (1984). “Analysis of
formulating appropriate MDCT techniques when x-ray computed tomography images using the noise
imaging the chest, abdomen, and pelvis in pediatric power spectrum and autocorrelation function,” Phys.
patients,” Am. J. Roentgenol. 182, 849 –859. Med. Biol. 29, 1343– 1352.
Cohen, G., and DiBianca, F. A. (1979). “The use of FDA (1984). United States FDA Code of Federal
contrast-detail—dose evaluation of image quality in a Regulations. 21 CFR 1020.33 - Computed tomography
computed tomographic scanner,” J. Comput. Assist. (CT) equipment, US Government Printing Office,
Tomogr. 3, 189– 195. Washington, DC.
Conway, B. J., McCrohan, J. L., Antonsen, R. G., Rueter, FDA (2002). Food and Drug Administration. “FDA public
F. G., Slayton, R. J., and Suleiman, O. H. (1992). health notification: Reducing radiation risk from com-
“Average radiation dose in standard CT examinations puted tomography for pediatric and small adult
of the head: Results of the 1990 NEXT survey,” patients,” Pediatr. Radiol. 32, 314– 316.
Radiology 184, 135– 140. Federle, M. P., Goldberg, H. I., Kaiser, J. A., Moss, A. A.,
Cormack, A. M. (1963). “Representation of a function by Jeffrey, R. B., Jr., and Mall, J. C. (1981). “Evaluation
its line integrals, with some radiological applications,” of abdominal trauma by computed tomography,”
J. Appl. Phys. 34, 2722– 2727. Radiology 138, 637– 644.
Crawford, C. R., and King, K. F. (1990). “Computed tom- Flohr, T. G., Leng, S., Yu, L., Aiimendinger, T., Bruder,
ography scanning with simultaneous patient transla- H., Petersilka, M., Eusemann, C. D., Stierstorfer, K.,
tion,” Med. Phys. 17, 967 –982. Schmidt, B., and McCollough, C. H. (2009).
Dainty, J. C., and Shaw, R. (1974). Image Science, “Dual-source spiral CT with pitch up to 3.2 and 75 ms
Principles, Analysis, and Evaluation of Photographic- temporal resolution: Image reconstruction and assess-
Type Imaging Processes (Academic Press, London). ment of image quality,” Med. Phys. 36, 5641– 5653.
144
References
Freeny, P. C., Marks, W. M., Ryan, J. A., and Traverso, L. NRPB-R289 (National Radiological Protection Board,
W. (1988). “Pancreatic ductal adenocarcinoma: Diagnosis Chilton, Didcot, Oxon, England).
and staging with dynamic CT,” Radiology 166, 125–133. Hausleiter, J., Meyer, T., Hermann, F., Hadamitzky, M.,
Frei, S. P., Bond, W. F., Bazuro, R. K., Richardson, D. M., Krebs, M., Gerber, T., McCollough, C. H., Martinoff, S.,
Sierzega, G. M., and Reed, J. F. (2008). “Appendicitis Kastrati, A., Schomig, A., and Achenbach, S. J. (2009).
outcomes with increasing computed tomographic “Estimated radiation dose associated with cardiac CT
scanning,” Am. J. Emerg. Med. 26, 39–44. angiography,” JAMA 301, 500– 507.
Frush, D. P., and Applegate, K. (2004). “Computed tom- Hounsfield, G. N. (1973). “Computerized transverse axial
ography and radiation: Understanding the issues,” scanning (tomography): Part 1. Description of system,”
J. Am. Coll. Radiol. 1, 113–119. Br. J. Radiol. 46, 1016– 1022.
Funama, Y., Awai, K., Nakayama, Y., Kakei, K., Nagasue, Hounsfield, G. N. (1980). “Nobel Award address.
N., Shimamura, M., Sato, N., Sultana, S., Morishita, Computed medical imaging,” Med. Phys. 7, 283 –290.
S., and Yamashita, Y. (2005). “Radiation dose Hsieh, J. (2003). Computed Tomography: Principles,
reduction without degradation of low-contrast detect- Design, Artifacts and Recent Advances, SPIE Press
ability at abdominal multisection CT with a low-tube Monograph PM114 (Society of Photo-Optical
voltage technique: Phantom study,” Radiology 237, Instrumentation Engineers, Bellingham, WA).
905– 910. Hsieh, J., Londt, J., Vass, M., Li, J., Tang, X., and
Funama, Y., Taguchi, K., Utsunomiya, D., Oda, S., Okerlund, D. (2006). “Step-and-shoot data acquisition
Yanaga, Y., Yamashita, Y., and Awai, K. (2011). and reconstruction for cardiac x-ray computed tomog-
“Combination of a low-tube-voltage technique with raphy,” Med. Phys. 33, 4236–4248.
hybrid iterative reconstruction (iDose) algorithm at Hu, H. (1999). “Multi-slice helical CT: scan and recon-
coronary computed tomographic angiography,” struction,” Med. Phys. 26, 5 –18.
J. Comput. Assist. Tomogr. 35, 480. Hu, H., and Fox, S. H. (1996). “The effect of helical pitch
Gagne, R. M. (1989). “Geometrical aspects of computed and beam collimation on the lesion contrast and slice
tomography: sensitivity profile and exposure profile,” profile in helical CT imaging,” Med. Phys. 23,
Med. Phys. 16, 29– 37. 1943–1954.
Geleijns, J., Salvado Artells, M., de Bruin, P. W., Matter, Huda, W. (2002a). “Dose and image quality in CT,”
R., Muramatsu, Y., and McNitt-Gray, M. F. (2009). Pediatr. Radiol. 32, 709– 713; discussion 751– 754.
“Computed tomography dose assessment for a 160 mm Huda, W. (2002b). “Effective doses to adult and pediatric
wide, 320 detector row, cone beam CT scanner,” Phys. patients,” Pediatr. Radiol. 32, 272– 279.
Med. Biol. 54, 3141– 3159. Huda, W., Atherton, J. V., Ware, D. E., and Cumming,
Gies, M., Kalender, W. A., Wolf, H., and Suess, C. (1999). W. A. (1997). “An approach for the estimation of effect-
“Dose reduction in CT by anatomically adapted tube ive radiation dose at CT in pediatric patients,”
current modulation. I. Simulation studies,” Med. Phys. Radiology 203, 417– 422.
26, 2235–2247. Huda, W., Scalzetti, E. M., and Levin, G. (2000a).
Goma, C., Ruiz, A., Jornet, N., Latorre, A., Pallerol, “Technique factors and image quality as functions of
R. M., Carrasco, P., Eudaldo, T., and Ribas, M. (2011). patient weight at abdominal CT,” Radiology 217,
“Radiation dose assessment in a 320-detector-row CT 430– 435.
scanner used in cardiac imaging,” Med. Phys. 38, Huda, W., Scalzetti, E. M., and Roskopf, M. (2000b).
1473–1480. “Effective doses to patients undergoing thoracic com-
Gorny, K. R., Leitzen, S. L., Bruesewitz, M. R., Kofler, J. M., puted tomography examinations,” Med. Phys. 27,
Hangiandreou, N. J., and McCollough, C. H. (2005). 838– 844.
“The calibration of experimental self-developing Huda, W., Ravenel, J. G., and Scalzetti, E. M. (2002a).
Gafchromic HXR film for the measurement of radi- “How do radiographic techniques affect image quality
ation dose in computed tomography,” Med. Phys. 32, and patient doses in CT?,” Semin Ultrasound CT MR
1010–1016. 23, 411– 422.
Gray, J. E., Archer, B. R., Butler, P. F., Hobbs, B. B., Huda, W., Sajewicz, A. M., Ogden, K. M., Scalzetti,
Mettler, F. A., Jr., Pizzutiello, R. J., Jr., Schueler, E. M., and Dance, D. R. (2002b). “How good is the ACR
B. A., Strauss, K. J., Suleiman, O. H., and Yaffe, M. J. accreditation phantom for assessing image quality
(2005). “Reference values for diagnostic radiology: in digital mammography?,” Acad. Radiol. 9, 764 –
Application and impact,” Radiology 235, 354 –358. 772.
Haaga, J. R., Miraldi, F., MacIntyre, W., LiPuma, J. P., Huda, W., Lieberman, K. A., Chang, J., and Roskopf, M.
Bryan, P. J., and Wiesen, E. (1981). “The effect of mAs L. (2004). “Patient size and x-ray technique factors in
variation upon computed tomography image quality as head computed tomography examinations. II. Image
evaluated by in vivo and in vitro studies,” Radiology quality,” Med. Phys. 31, 595– 601.
138, 449– 454. Huda, W., Sterzik, A., Tipnis, S., and Schoepf, U. J.
Hanson, K. M. (1979). “Detectability in computed tomo- (2010). “Organ doses to adult patients for chest CT,”
graphic images,” Med. Phys. 6, 441 –451. Med. Phys. 37, 842 –847.
Hart, D., Hillier, M. C., Wall, B. F., Shrimpton, P. C., and ICRP (2007). International Commission on Radiological
Bungay, D. (1996). Doses to Patients from Medical Protection. The 2007 Recommendations of the
X-ray Examinations in the UK: 1995 Review, Report International Commission on Radiological Protection,
145
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
ICRP Publication 103. Ann. ICRP 37(2– 4). (Elsevier Kalra, M. K., Maher, M. M., Toth, T. L., Schmidt, B.,
Health Science, Oxford, UK). Westerman, B. L., Morgan, H. T., and Saini, S. (2004).
ICRU (2011). International Commission on Radiation “Techniques and applications of automatic tube
Units and Measurements. Fundamental Quantities current modulation for CT,” Radiology 233, 649– 657.
and Units for Ionizing Radiation, ICRU Report 85a. Kalra, M. K., Rizzo, S., Maher, M. M., Halpern, E. F.,
J. ICRU 11. (Oxford University Press, Oxford). Toth, T. L., Shepard, J. A., and Aquino, S. L. (2005).
Image Gently (2011). “Image Gently and CT Scans,” http “Chest CT performed with z-axis modulation:
://www.pedrad.org/associations/5364/ig/?page=614 Scanning protocol and radiation dose,” Radiology 237,
ImPACT (2004). ImPACT Organization. “ImPACT ctdi 303– 308.
tables,” http://www.impactscan.org/ctditables.htm Kleinman, P. L., Strauss, K. J., Zurakowski, D., Buckley,
ImPACT (2011). ImPACT Organization. “ImPACT’s ct K. S., and Taylor, G. A. (2010). “Patient size measured on
dosimetry tool” http://www.impactscan.org/ctdosimetry CT images as a function of age at a tertiary care chil-
.htm dren”s hospital,” Am. J. Roentgenol. 194, 1611–1619.
IEC (2009). International Electrotechnical Commission. Kruger, R. L., McCollough, C. H., and Zink, F. E. (2000).
Medical Electrical Equipment. Part 2-44: Particular “Measurement of half-value layer in x-ray CT: A com-
Requirements for the Basic Safety and Essential parison of two noninvasive techniques,” Med. Phys. 27,
Performance of X-Ray Equipment for Computed 1915–1919.
Tomography. IEC publication No. 60601-2-44. ed. 3.1 Kuhl, D. E., and Edwards, R. Q. (1968). “Rapid brain
(International Electrotechnical Commission Central scanner with self contained computer and CRT display
Office, Geneva, Switzerland). for both rectilinear and transverse section viewing,”
Jakobs, T. F., Becker, C. R., Ohnesorge, B., Flohr, T., J. Nucl. Med. 9, 332.
Suess, C., Schoepf, U. J., and Reiser, M. F. (2002). La Riviere, P. J. (2005). “Penalized-likelihood sinogram
“Multislice helical CT of the heart with retrospective smoothing for low-dose CT,” Med. Phys. 32,
ECG gating: Reduction of radiation exposure by 1676–1683.
ECG-controlled tube current modulation,” Eur. Radiol. La Riviere, P. J., Bian, J., and Vargas, P. A. (2006).
12, 1081–1086. “Penalized-likelihood sinogram restoration for com-
Johns, H. E., and Cunningham, J. R. (1974). The Physics of puted tomography,” IEEE Trans. Med. Imaging 25,
Radiology, 3rd ed. (Charles C. Thomas, Springfield, IL). 1022–1036.
Johnson, T. R., Krauss, B., Sedlmair, M., Grasruck, M., Larson, D. B., Johnson, L. W., Schnell, B. M., Goske, M.
Bruder, H., Morhard, D., Fink, C., Weckbach, S., J., Salisbury, S. R., and Forman, H. P. (2011a). “Rising
Lenhard, M., Schmidt, B., Flohr, T., Reiser, M. F., and use of CT in child visits to the emergency department
Becker, C. R. (2007). “Material differentiation by dual in the United States, 1995– 2008,” Radiology 259,
energy CT: Initial experience,” Eur. Radiol. 17, 793– 801.
1510–1517. Larson, D. B., Johnson, L. W., Schnell, B. M., Salisbury,
Jones, D. G., and Shrimpton, P. C. (1991). Survey of CT S. R., and Forman, H. P. (2011b). “National trends in
Practice in the UK. Part 3: Normalised Organ Doses CT use in the emergency department: 1995–2007,”
Calculated using Monte Carlo Techniques, Report Radiology 258, 164– 173.
NRPB-SR250 (National Radiological Protection Board, Lehmann, L. A., Alvarez, R. E., Macovski, A., Brody,
Chilton, UK). W. R., Pelc, N. J., Riederer, S. J., and Hall, A. L.
Jucius, R. A., and Kambic, G. X. (1980). “Measurements (1981). “Generalized image combinations in dual KVP
of computed tomography x-ray fields utilizing the digital radiography,” Med. Phys. 8, 659 –667.
partial volume effect,” Med. Phys. 7, 379– 382. Leitz, W., Axelsson, B., and Szendro, G. (1995).
Judy, P. F. (1976). “The line spread function and modula- “Computed tomography dose assessment: A practical
tion transfer function of a computed tomographic approach,” Radiat. Prot. Dosim. 57, 377– 380.
scanner,” Med. Phys. 3, 233 –236. Lell, M., Marwan, M., Schepis, T., Pflederer, T., Anders,
Kalender, W. A. (2011). Computed Tomography: K., Flohr, T., Allmendinger, T., Kalender, W., Ertel, D.,
Fundamentals, System Technology, Image Quality, Thierfelder, C., Kuettner, A., Ropers, D., Daniel, W. G.,
Applications, 3rd ed. (Publicis Publishing, Erlangen, and Achenbach, S. (2009). “Prospectively ECG-triggered
Germany). high-pitch spiral acquisition for coronary CT angiog-
Kalender, W. A., Perman, W. H., Vetter, J. R., and Klotz, raphy using dual source CT: Technique and initial ex-
E. (1986). “Evaluation of a prototype dual-energy com- perience,” Eur Radiol, 19, 2576–2583.
puted tomographic apparatus. I. Phantom studies,” Li, T., Li, X., Wang, J., Wen, J., Lu, H., Hsieh, J., and
Med. Phys. 13, 334 –339. Liang, Z. (2004). “Nonlinear approach to sinogram
Kalender, W. A., Seissler, W., Klotz, E., and Vock, P. smoothing for low-dose x-ray CT, ” IEEE Trans. Nucl.
(1990). “Spiral volumetric CT with single-breath-hold Sci. 51, 2505–2513.
technique, continuous transport, and continuous Linton, O. W., and Mettler, F. A., Jr. (2003). “National
scanner rotation,” Radiology 176, 181 –183. conference on dose reduction in CT, with an emphasis
Kalender, W. A., Wolf, H., and Suess, C. (1999). “Dose re- on pediatric patients,” Am. J. Roentgenol. 181,
duction in CT by anatomically adapted tube current 321– 329.
modulation. II. Phantom measurements,” Med. Phys. Liu, H. L., Liu, R. R., Reeve, D. M., Shepard, S. J., and
26, 2248–2253. Willis, C. E. (2005). “Measurement of CT radiation
146
References
profile width using CR imaging plates,” Med. Phys. 32, “Experimental validation of a method characterizing
2881–2887. bow tie filters in CT scanners using a real-time dose
Manduca, A., Yu, L., Trzasko, J. D., Khaylova, N., Kofler, probe,” Med. Phys. 38, 1406–1415.
J. M., McCollough, C. M., and Fletcher, J. G. (2009). McNitt-Gray, M. F. (2002). “AAPM/RSNA Physics
“Projection space denoising with bilateral filtering and Tutorial for Residents: Topics in CT. Radiation dose in
CT noise modeling for dose reduction in CT,” Med. CT,” RadioGraphics 22, 1541–1553.
Phys. 36, 4911– 4919. McNitt-Gray, M., Angel, E., Turner, A., Stevens, D.,
Mathieu, K. B., McNitt-Gray, M. F., Zhang, D., Kim, H. Primak, A., and Cagnon, C. (2008). “CTDI normalized
J., and Cody, D. D. (2010). “Precision of dosimetry- to measured beam width as an accurate predictor of
related measurements obtained on current multidetec- dose variations for multidetector row CT (MDCT) scan-
tor computed tomography scanners,” Med. Phys. 37, ners across all manufacturers,” Meeting of the
4102–4109. Radiological Society of North America (RSNA),
McCollough, C. H. (2003). “Patient dose in cardiac com- Chicago, IL, (abstract).
puted tomography,” Herz 28, 1– 6. Mori, S., Endo, M., Nishizawa, K., Tsunoo, T., Aoyama,
McCollough, C. H. (2005). “Automatic exposure control in T., Fujiwara, H., and Murase, K. (2005). “Enlarged lon-
CT: Are we done yet?,” Radiology 237, 755 –756. gitudinal dose profiles in cone-beam CT and the need
McCollough, C. H. (2006). “It is time to retire the com- for modified dosimetry,” Med. Phys. 32, 1061–1069.
puted tomography dose index (CTDI) for CT quality Nakayama, Y., Awai, K., Funama, Y., Hatemura, M.,
assurance and dose optimization,” Med. Phys. 33, Imuta, M., Nakaura, T., Ryu, D., Morishita, S.,
1189 –1191. Sultana, S., Sato, N., and Yamashita, Y. (2005).
McCollough, C. H. (2006). “Standardization in CT ter- “Abdominal CT with low tube voltage: Preliminary
minology: A physicist’s perspective,” Radiology 241, observations about radiation dose, contrast enhance-
661– 662. ment, image quality, and noise,” Radiology 237,
McCollough, C. H. (2009). “Re: Maximizing dose reductions 945–951.
with cardiac CT,” Int. J. Cardiovasc. Imaging 25, 647. Nakonechny, K. D., Fallone, B. G., and Rathee, S. (2005).
McCollough, C. H., and Zink, F. E. (1999). “Performance “Novel methods of measuring single scan dose profiles
evaluation of a multi-slice CT system,” Med. Phys. 26, and cumulative dose in CT,” Med. Phys. 32, 98–109.
2223–2230. NAS/NRC (2006). National Academies/National Research
McCollough, C. H., Zink, F. E., Kofler, J., Matsumoto, Council. Health Effects from Exposure to Low Levels of
J. S., Thomas, K. B., and Hoffman, A. D. (2002). “Dose Ionizing Radiation, BEIR VII Phase 2 (National
optimization in CT: Creation, implementation and clin- Academy Press, Washington, DC).
ical acceptance of size-based technique charts,” NCRP (2009). National Council on Radiation Protection
Radiology 225(P), 591. and Measurements. Ionizing Radiation Exposure of
McCollough, C. H., Bruesewitz, M. R., and Kofler, J. M., the Population of the United States, NCRP Report No.
Jr. (2006). “CT dose reduction and dose management 160 (National Council on Radiation Protection and
tools: Overview of available options,” Radiographics Measurements, Bethesda, Maryland).
26, 503– 512. Newman, B., and Callahan, M. J. (2011). “ALARA (as
McCollough, C., Leng, S., Schmidt, B., Allmendinger, T., low as reasonably achievable) CT 2011—executive
Eusemann, C., and Flohr, T. (2009). “Use of a pitch summary,” Pediatr. Radiol. 41 Suppl 2, 453– 455.
value of 3.2 in dual source, cardiac CT angiography: Nickoloff, E. L., and Riley, R. (1985). “A simplified ap-
Dose performance relative to existing scan modes,” proach for modulation transfer function determina-
(Annual Meeting of the Radiological Society of North tions in computed tomography,” Med. Phys. 12,
America, Chicago, IL) (abstract). 437– 442.
McCollough, C. H., Leng, S., Yu, L., Cody, D. D., Boone, Nickoloff, E. L., Dutta, A. K., and Lu, Z. F. (2003).
J. M., and McNitt-Gray, M. F. (2011). “CT dose index “Influence of phantom diameter, kVp and scan mode
and patient dose: They are not the same thing,” upon computed tomography dose index,” Med. Phys.
Radiology 259, 311 –316. 30, 395– 402.
McCrohan, J. L., Patterson, J. F., Gagne, R. M., and Nyman, U., Ahl, T. L., Kristiansson, M., Nilsson, L., and
Goldstein, H. A. (1987). “Average radiation doses in a Wettemark, S. (2005). “Patient-circumference-adapted
standard head examination for 250 CT systems,” dose regulation in body computed tomography. A prac-
Radiology 163, 263– 268. tical and flexible formula,” Acta Radiol. 46, 396– 406.
McGahan, J. P., and Dodd, G. D. (2001). “Radiofrequency Ogden, K. M., and Huda, W. (2008). “The effect of meas-
ablation of the liver,” Am. J. Roentgenol. 176, 3–16. urement location on in-air CTDI values,” Med. Phys.
McGahan, J. P., Loh, S., Fitzgerald, E., Koppie, T., 35, 2657.
Evans, C. P., Dall’Era, M., and Li, C. S. (2011). Oldendorf, W. H. (1961). “Isolated flying spot detection of
“Pretreatment imaging can be used to select imaging radio-density discontinuities displaying the internal
guidance, ultrasound alone versus CT plus ultrasound, structural pattern of a complex object,” IRE Trans
for percutaneous renal radiofrequency ablation,” Bio-Med Elect BME 8, 68– 72.
Am. J. Roentgenol. 197, 1244– 1250. Primak, A. N., Fletcher, J. G., Vrtiska, T. J., Dzyubak, O.
McKenney, S. E., Nosratieh, A., Gelskey, D., Yang, K., P., Lieske, J. C., Jackson, M. E., Williams, J. C., Jr.,
Huang, S. Y., Chen, L., and Boone, J. M. (2011). and McCollough, C. H. (2007). “Noninvasive
147
RADIATION DOSE AND IMAGE-QUALITY ASSESSMENT IN COMPUTED TOMOGRAPHY
differentiation of uric acid versus non-uric acid kidney scanners from different manufacturers,” Med. Phys.
stones using dual-energy CT,” Acad. Radiol. 14, 39, 6048–6055.
1441–1447. Stolzmann, P., Leschka, S., Scheffel, H., Krauss, T.,
Radon, J. (1917). “Über die Bestimmung von Funktionen Desbiolles, L., Plass, A., Genoni, M., Flohr, T. G.,
durch ihre Integralwerte längs gewisser Wildermuth, S., Marincek, B., and Alkadhi, H. (2008).
Mannigfaltigkeiten,” Ber. Verh. Sächs. Akad. Wiss. “Dual-source CT in step-and-shoot mode: Noninvasive
Leipzig, Math.-Nat. kl. 69, 262– 277. coronary angiography with low radiation dose,”
Raff, G. L., Chinnaiyan, K. M., Share, D. A., Goraya, T. Radiology 249, 71–80.
Y., Kazerooni, E. A., Moscucci, M., Gentry, R. E., and Tapiovaara, M. J., and Wagner, R. F. (1985). “SNR and
Abidov, A. (2009). “Radiation dose from cardiac computed DQE analysis of broad spectrum X-ray imaging,” Phys.
tomography before and after implementation of radiation Med. Biol. 30, 519 –529.
dose-reduction techniques,” JAMA 301, 2340–2348. Tipnis, S., Ramachandra, A., Huda, W., Hardie, A.,
Rao, P. M., Rhea, J. T., Novelline, R. A., McCabe, C. J., Schoepf, J., Costello, P., Flohr, T., and Sedlmair, M.
Lawrason, J. N., Berger, D. L., and Sacknoff, R. (2010). “Iterative reconstruction in image space (IRIS)
(1997). “Helical CT technique for the diagnosis of ap- and lesion detection in abdominal CT,” Proc. SPIE
pendicitis: Prospective evaluation of a focused appen- 7622, 76222K-1–76222K-12.
dix CT examination,” Radiology 202, 139 –144. Tucker, D. M., Barnes, G. T., and Chakraborty, D. P.
Riederer, S. J., Pelc, N. J., and Chesler, D. A. (1978). (1991a). “Semiempirical model for generating tungsten
“The noise power spectrum in computed x-ray tomog- target x-ray spectra,” Med. Phys. 18, 211– 218.
raphy,” Phys. Med. Biol. 23, 446– 454. Tucker, D. M., Barnes, G. T., and Wu, X. Z. (1991b).
Rigauts, H., Marchal, G., Baert, A. L., and Hupke, R. “Molybdenum target x-ray spectra: A semiempirical
(1990). “Initial experience with volume CT scanning,” model,” Med. Phys. 18, 402 –407.
J. Comput. Assist. Tomogr. 14, 675 –682. Turner, A. C., Zhang, D., Kim, H. J., DeMarco, J. J.,
Rose, A. (1973). Vision: Human and Electronic (Plenum Cagnon, C. H., Angel, E., Cody, D. D., Stevens, D. M.,
Press, New York). Primak, A. N., McCollough, C. H., and McNitt-Gray,
Rybicki, F. J., Otero, H. J., Steigner, M. L., Vorobiof, G., M. F. (2009). “A method to generate equivalent energy
Nallamshetty, L., Mitsouras, D., Ersoy, H., Mather, R. spectra and filtration models based on measurement
T., Judy, P. F., Cai, T., Coyner, K., Schultz, K., for multidetector CT Monte Carlo dosimetry simula-
Whitmore, A. G., and Di Carli, M. F. (2008). “Initial tions,” Med. Phys. 36, 2154– 2164.
evaluation of coronary images from 320-detector row Turner, A. C., Zankl, M., DeMarco, J. J., Cagnon, C. H.,
computed tomography,” Int. J. Cardiovasc. Imaging Zhang, D., Angel, E., Cody, D. D., Stevens, D. M.,
24, 535– 546. McCollough, C. H., and McNitt-Gray, M. F. (2010).
Schmidt, B., and Kalender, W. A. (2002). “A fast voxel- “The feasibility of a scanner-independent technique to
based Monte Carlo method for scanner- and patient- estimate organ dose from MDCT scans: Using CTDIvol
specific dose calculations in computed tomography,” to account for differences between scanners,” Med.
Physica Medica 18, 43–53. Phys. 37, 1816– 1825.
Schneiders, N. J., and Bushong, S. C. (1980). “Computer Tward, D. J., and Siewerdsen, J. H. (2008). “Cascaded
assisted MTF determination in CT,” Med. Phys. 7, systems analysis of the 3D noise transfer characteristics
76–78. of flat-panel cone-beam CT,” Med. Phys. 35, 5510–5529.
Shope, T. B., Gagne, R. M., and Johnson, G. C. (1981). “A Wagner, R. F., and Brown, D. G. (1982). “Overview of a
method for describing the doses delivered by transmis- unified SNR analysis of medical imaging systems,”
sion x-ray computed tomography,” Med. Phys. 8, IEEE Trans. Med. Imaging 1, 210– 213.
488– 495. Wagner, R. F., Brown, D. G., and Pastel, M. S. (1979).
Shrimpton, P. C., Hillier, M. C., Lewis, M. A., and Dunn, “Application of information theory to the assessment of
M. (2006). “National survey of doses from CT in the computed tomography,” Med. Phys. 6, 83–94.
UK: 2003,” Br. J. Radiol. 79, 968– 980. Wang, J., Li, T., Lu, H., and Liang, Z. (2006). “Penalized
Siegel, M. J., Schmidt, B., Bradley, D., Suess, C., and weighted least-squares approach to sinogram noise reduc-
Hildebolt, C. (2004). “Radiation dose and image tion and image reconstruction for low-dose x-ray computed
quality in pediatric CT: Effect of technical factors and tomography,” IEEE Trans. Med. Imaging 25, 1272–1283.
phantom size and shape,” Radiology 233, 515 –522. Wilting, J. E., Zwartkruis, A., van Leeuwen, M. S.,
Siewerdsen, J. H., Cunningham, I. A., and Jaffray, D. A. Timmer, J., Kamphuis, A. G., and Feldberg, M. (2001).
(2002). “A framework for noise-power spectrum analysis “A rational approach to dose reduction in CT:
of multidimensional images,” Med. Phys. 29, 2655–2671. Individualized scan protocols,” Eur. Radiol. 11,
Silva, A. C., Lawder, H. J., Hara, A., Kujak, J., and 2627–2632.
Pavlicek, W. (2010). “Innovations in CT dose reduction Winer-Muram, H. T. (2006). “The solitary pulmonary
strategy: Application of the adaptive statistical itera- nodule,” Radiology 239, 34–49.
tive reconstruction algorithm,” Am. J. Roentgenol. 194, Wootton-Gorges, S. L. (2010). “Pediatric trauma,” pp 75–
191– 199. 97 in Essentials of Pediatric Radiology—A
Solomon, J. B., Christianson, O., and Samei, E. (2012). Multimodality Approach (Cambridge University Press,
“Quantitative comparison of noise texture across CT New York).
148
References
Yang, K., Kwan, A. L. C., Huang, S.-Y., Packard, N. J., Primak, A. N., McCollough, C. H., and McNitt-Gray,
and Boone, J. M. (2008). “Noise power properties of a M. F. (2009). “Variability of surface and center position
cone-beam CT system for breast cancer detection,” radiation dose in MDCT: Monte Carlo simulations
Med. Phys. 35, 5317 –5327. using CTDI and anthropomorphic phantoms,” Med.
Yu, L., Li, H., Fletcher, J. G., and McCollough, C. H. Phys. 36, 1025 –1038.
(2010). “Automatic selection of tube potential for radi- Zhang, D., Cagnon, C. H., Villablanca, J. P., McCollough,
ation dose reduction in CT: A general strategy,” Med. C. H., Cody, D. D., Stevens, D. M., Zankl, M.,
Phys. 37, 234 –243. Demarco, J. J., Turner, A. C., Khatonabadi, M., and
Zeman, R. K., Herlihy, V., Branham, T. A., Bhargavan, McNitt-Gray, M. F. (2012). “Peak skin and eye lens
M., Bush, K. M., and McCollough, C. H. (2011). “Can radiation dose from brain perfusion CT based on
experienced CT radiologists use technique parameters Monte Carlo simulation,” Am. J. Roentgenol. 198,
to predict excessive patient dose? An analysis of the 412– 417.
ACR CT accreditation database,” J. Am. Coll. Radiol. Zhou, H., and Boone, J. M. (2008). “Monte Carlo evalu-
8, 275 –280. ation of CTDI1 in infinitely long cylinders of water,
Zhang, D., Savandi, A. S., Demarco, J. J., Cagnon, C. H., polyethylene and PMMA with diameters from 10 mm
Angel, E., Turner, A. C., Cody, D. D., Stevens, D. M., to 500 mm,” Med. Phys. 35, 2424– 2431.
149
EXCELLENCE
EXCELLENCE
IN EVIDENCE
IN EVIDENCE
Discover
Discovera awealth
wealth
ofofknowledge
knowledge
OXFORD JOURNALS ONLINE ARCHIVE
OXFORD JOURNALS ONLINE ARCHIVE
> uncover backfiles of content
> uncover
dating from 1849
backfiles of content
dating from 1849
> instant access to 146 years
of >research
instant access to 146 years
of research
> seamless access from Volume 1,
> seamless
Issue 1 access from Volume 1,
Issue 1
> access by IP range, Athens
or >Shibboleth
access by IP range, Athens
or Shibboleth
> COUNTER-compliant usage
> COUNTER-compliant usage
statistics
statistics
www.oxfordjournals.org/collections/archives 1
Radiation Protection Dosimetry
Previously published special issues available from Oxford University Press
The current year and two previous years’ issues are available from Oxford University Press. All issues
published prior to 2000 are available from the Periodicals Service Company, psc@periodicals.com.
Please visit the web site for pricing and ordering information
www.rpd.oxfordjournals.org