where 0 < 7 < Tand k = 0,1,2,.... Equation (3-7) implies that the circuit holds
the amplitude of the sample from one sampling instant to the next. Such a data-hold
is called a zero-order hold, or clamper, or staircase generator. The output of the
zero-order hoid is a Staircasé function. Ii this Book, unless Otherwise stated, we
assume that the hold circuit is of zero order.
It will be seen later that the transfer function G,, of the zero-order hold may
be given by
ts
G
s
Zero-Order Hold. Figure 3-3 shows a sampler and a zero-order hold. The
input signal x(¢) is sampled at discrete instants and the sampled signal is passed
through the zero-order hold. The zero-order hold circuit smoothes the sampled
signal to produce the signal /i(r), which is constant from the last sampled value until
the next sample is available. That is,
‘We shall obtain a mathematical model of the combination of a real sampler and
zero-order circuit, as shown in Figure 3~4(a). Utilizing the fact that the integral of
impulse function is a constant, we may assume th o-order hold is an
“ategrator, and the input to the zero-order hold circui of impulses. Then
“amath -al model for the real sampler and zero- jay be constructed
as shown in Figure 3-4(b), where Gio(s) is the transfer function of the zero-order
{hold and x*(°) is the impulse sampled signal of x(")
Figure 3-4 (a) A real sampler and
sampler and transfer function Gan(s)
Consider the sampler and zero-order hold shown in Figure 3-4(a). Assume
that the signal x(*) iszero fort < 0. Then the output h,(t) is related to x(t) as follows:
Next, consider the mathematical model shown in Figure 3~4(b). The output
of this model must be the same as that of the real zero-order hold, or
By comparing Equations (3-12) and (3-13), we see that the transfer function of the
zero-order hold may be given by
Gils)
Note that, mathematically, the system shown in Figure 3~4(a) is equivalent to the
system shown in Figure 3-4(b) from the viewpoint of the input-output relationship.
‘That is, a real sampler and zero-order hold can be replaced by a mathematically
equivalent continuous-time system that consists of an impulse sampler and a transfer
function (1 — e~T)/s. The two sampling processes will be distinguished (as they are
in Figure 3-4) by the manner in which the sampling switches are drawn.