You are on page 1of 6

QUESTIONS OF QUALITY

Where Can I Draw The Line?


R.D. McDowall, R D McDowall Ltd, Bromley, Kent, UK.

A question that keeps raising its head when working in a regulated laboratory is can chromatographers
integrate peaks manually? If they can, when can they do it? Also if they can manually integrate, when
should they not do it? 

One of the red rags to a regulatory change during a run, so getting the the functionality is essentially the same.
bull is the issue of manual integration right overall integration can be a Chromatographic integration begins
of chromatograms. If seen during an balancing act.  when the CDS samples the detector
inspection, you can almost see the The regulators are wrong in their output using an analogue to digital
wheels in the inspector’s brain turn in requirement for a procedure on manual converter (A/D). An integration method
mechanical precision as they question: integration. What is required instead is a in the CDS determines the frequency
are they testing into compliance? In a standard operating procedure (SOP) for of the detector data collection and
2014 FDA inspection of a laboratory chromatographic integration, of which analysis run time. Dependent on the
in Europe, one FDA 483 observation manual integration is an important CDS, either an integration method or a
stated: sub-set. Although the focus of this processing method will be automatically
No procedure exists describing how
column is manual integration please applied to the file of data slices to
to perform manual integration. 
do not lose sight of the bigger picture. calculate the peak areas or heights.
As such, we will not be discussing The processing method will contain the
A more serious non-compliance the various calibration methods that identities of the peaks of interest and
occurred in the FDA warning letter to could be applied to standards to their expected elution time windows.
Leiner Health Products (1):  quantify analytes in samples nor will These data files and their interpretation
In addition, our investigators
we be looking at analogue to digital constitute part of the raw data and
documented many instances with conversion because the latter has primary record for the analysis. We will
extensive manipulation of data been covered in an earlier Questions of discuss the definition of primary record
with no explanation regarding why Quality column (3). Furthermore we will in the next Questions of Quality column
the manipulation was conducted. not be considering the use of system (9) in light of the two MHRA data
This manipulation would include
changing integration parameters suitability tests (SSTs), again because integrity guidance documents issued
or relabelling peaks such that this column has already discussed this year (10,11).
previously resolved peaks would them (4–6). The paper by Hill et al. (7) The two most important parameters
not be integrated and included in on manual reintegration in bioanalysis is for integration are sampling rate
the calculation for impurities.
also a highly recommended publication and peak threshold. In combination
There have been many other cases to read on the subject. they determine the slices for peak
that I summarized in a recent Questions measurement but also suppress noise
of Quality column on the role of Back to Integration Basics to allow better peak measurement (8). 
chromatography data systems (CDS) Before we can discuss how to control • Peak width or sampling rate setting
in data falsification (2). What regulatory and manage manual integration it is governs how often the detector
guidance is there to help us? We need important to understand the basics output is sampled, which directly
to consider some basics of integrating of chromatographic integration itself. impacts the accuracy of peak area
chromatographic peaks before we The best book on the subject of measurement. If the sampling rate
can look at manual integration and the integration is by Norman Dyson (8) is too slow it could result in the
regulatory guidance surrounding it. The and, although the second edition was integration missing a small fast
reason for this is simple: if integration published in 1998, it is still applicable eluting peak or a valley between
parameters are set correctly and the and highly recommended if you two peaks. Conversely, too fast
chromatography is acceptable then want to understand chromatographic a data acquisition rate can be
there should be no need to reintegrate integration. The key integration managed by data bunching in the
manually in many cases. However, parameters are shown in Table 1 and CDS software. Typically you should
we also need to acknowledge that the main ones are discussed below. sample a peak at least 20–30 times
chromatographic systems are dynamic Note that some CDS suppliers may call across its width so that it can be
by their nature and separations can these parameters a different name but integrated accurately; for example,
2 LC•GC Europe June 2015
QUESTIONS OF QUALITY

low, too much of the peak tailing is


Figure 1: Key parameters for peak integration
counted. Conversely if the threshold
is set too high the peak finishes early,
1000
Peak Apex? which under-estimates the area of the
peak. 
• Once the peak has been detected
the next step is to determine when
it ends. Typically this is when the
signal returns to where the baseline
Millivolts

was before the peak started but if


there are several peaks eluting or
Noise?
the baseline is rising then the signal
Peak End? will not return to the baseline origin.
Peak Start?
Therefore baseline drift tolerance
determines how far the baseline can
0 drift from the original position. This
Baseline Drift?
Time
is in contrast to the peak threshold,
which determines how fast it can drift
away (8). 

From this and alongside a more


Table 1: Key CDS integration parameters for peak acquisition and measurement.
detailed discussion in Dyson’s book
Parameter Function (8) we can develop three basic rules of
• S
 ampling rate in Hertz (number of data points per second) that chromatographic integration:
the detector signal is sampled by the CDS A/D unit for a specific • Rule 1: Do NOT use default
method.
Sampling Rate integration parameters. Ensure that
• Some CDS may acquire at a high fixed rate of 100Hz regardless
(Peak Width) each set of integration parameters
of sampling rate setting (see bunching).
is tailored to an individual analytical
• Faster chromatography (UHPLC or capillary GC) needs a higher
sampling rate. procedure and do not use a one-
• A
 dding several consecutive data points to obtain an average time size-fits-all approach. For Beer-
Bunching slice value equivalent to a slower sampling rate. Lambert’s law to hold, the samples
(Smoothing) • Can also be used to reduce noise in a chromatogram. and standards must be consistently
• Achieved using the CDS software rather than via the A/D unit. integrated, otherwise the fundamental
• P
 arameter used to determine if a peak is detected or not. The comparison of absorbance versus
Peak Threshold
peak start is determined when the value or slope selected is concentration cannot be performed. 
exceeded (start threshold). • Rule 2: The function of a CDS is not
(Detection Threshold
or Slope Sensitivity) • Selecting too low a value results in integration of background to compensate for your poor method
noise; too large a value minimizes noise. This also results in small
development or separation. There is a
peaks not being identified and integrated.
Baseline Drift
belief in many laboratories that a CDS
• D
 etermines how far the baseline may be from the original position
Tolerance at the end of the peak. can be used to salvage an analytical
• S
 ets the minimum peak area or height for reporting run but this is not the case. Robust
Minimum Area chromatographic peaks. This parameter is independent of data and validated methods should
(Peak Area Reject) acquisition and the integration method and applies only to the reduce or eliminate this issue. 
integrated peak areas. • Rule 3: Understand what is
• D
 isables integration for a specific period in each injection. happening
Typically used at the start of an injection to avoid solvent front in the CDS. Just because you get
Integrate Inhibit effects or baseline perturbation.
a number from a CDS does not
• Note that the first signal after the end of the inhibition period is
mean you have to believe the result.
typically defined as a baseline, regardless of what is happening.
Use your eyes to look and your
brain to think. For example, look at
conventional high performance liquid peak from baseline noise. It is based the integration codes (BB or BV):
chromatography (HPLC) typically on the rate of change when the are these the ones expected? Are
requires a 0.5–1 Hz sampling rate, detector signal rises above a preset baselines positioned where they
faster capillary gas chromatography value in the integration method for should be and are they as expected?
(GC) has a sampling rate in the range peak start or falls in the case of peak Are retention times and peak shapes
5–20 Hz, and ultrahigh-pressure end. This needs to be set carefully consistent throughout the run? 
liquid chromatography (UHPLC) because if the baseline is noisy then
needs a sampling rate between noise is detected as peaks; if set too To complete this overview of CDS
20–50 Hz. Most A/D units in CDS are large the integration will miss small integration, please note what Dyson
rated at up to a 100 Hz sampling rate. peaks because the threshold is not says: improving the chromatography
• Peak threshold is the setting that triggered. The same process occurs must always take precedence over
discriminates the start and end of a at peak end but if the threshold is setting up the CDS (8):
www.chromatographyonline.com 3
QUESTIONS OF QUALITY

compliance hell.
Figure 2: Examples of peak shaving (2) and peak enhancing (3).
• Integration parameters are different
for standards and the unknown
1000 samples of the same run or between
replicate injections of the same
unknown sample in an analytical
batch.
• Evidence that the audit trail in the
system has been turned off and then
Millivolts

on again a few minutes later: what


changes have been performed that
are not recorded?

GMP Regulations and Regulatory


2 3
Guidance
0
1 Although there is great interest in data
integrity in laboratories working under
Time
the Good Manufacturing Practice
(GMP) regulations, there is a paucity of
either explicit regulation or regulatory
If a chromatographer finds it shows how the minor or second guidance to help chromatographers. 
necessary to tweak the parameters eluting peak in the chromatogram The only GMP regulations applicable
continuously in order to achieve
consistent measurement of
can be enhanced.  are the requirements for scientifically
standard samples, it is a clear sound analytical procedures in 21 CFR
indication that more work is needed Figure 2 shows the exaggerated 160(b) and for complete data under
to bring the instrument and analysis enhancement and shaving of peaks 21 CFR 211.194(a) (13). The latter topic
under control.  but sometimes all that is required was discussed in two earlier QOQ
In essence the likelihood of a is a small change to bring a non- columns (14,15). In addition, there is
Leiner‑style non-compliance citation conforming batch into compliance with data integrity audit under objective
(1) is looming. This is because a specification. Hence the need to trend 3 in the FDA’s Compliance Policy
excessive use of manual integration analytical results as required under Guide 7346.832 for pre-approval
to compensate for poor method the new revision to EU GMP for Quality inspections (16). Both the European
development or chromatographic Control laboratories (12) to highlight out Pharmacopoeia (chapter 2.2.46) (17)
separation increases risk to data of expectation (OOE) and out of trend and the United States Pharmacopoeia
integrity (accuracy), and increases time (OOT) results, as well as those that are <621> (18) discuss criteria for good
to generate, review, and release results. out of specification (OOS). Results out chromatography; there are no criteria
of trend or expectation may identify the for chromatographic integration or a
How Can Manual Integration actions of an individual analyst, which similar discussion on data integrity.
Result in Falsification? may warrant closer inspection. However, in the interest of scientific
Using manual integration to falsify In addition, inspectors and auditors soundness: can you justify and defend
chromatographic data can arise in a will look for a number of factors to your actions on the basis of good
number of ways, but the two main ways determine if there is unauthorized chromatographic science?
are:  manual manipulation of a peak: In my view, there is a need for
• Peak Shaving — Manually placing • Discovery of manual integration that regulatory agencies to issue guidance
the baselines to reduce the peak is not traceable or retrievable. A on the subject of integration with a
area on integration to enhance the chromatographer should be able to focus on manual integration for GMP.
analyte amount in a sample (only demonstrate that same results can be As they typically move at glacial
the standards are shaved) or reduce obtained if data files from a run are speeds, the likelihood of this occurring
the amount of analyte reported (by reprocessed – a red flag should be before the next ice age is probably
shaving the sample and not the raised if this cannot be done quickly. minimal. However, there is an area
standards). In Figure 2 the first eluting An inspector looking to see if the where regulators have been active on
peak has had the baseline manually CDS audit trail has been turned off the subject of manual reintegration and
adjusted to reduce the total peak temporarily to conduct falsification this is in bioanalysis of samples from
area. could also accompany this. non-clinical and clinical studies for the
• Peak Enhancing — Adjusting the • Electronic data are not available. registration of drugs.
baselines for integration to increase A focus on paper because the raw
the area of a peak. The enhancement data and the electronic records have Bioanalytical Regulatory
of sample areas over the standards been deleted or not saved means Guidance for Reintegration
can increase the calculated amount that data cannot be reprocessed and The FDA created guidance for
in the sample. If reduction is required, the result confirmed. The question bioanalytical method validation
the enhancement of the standards of falsification is raised and the following an AAPA-FDA conference
only is performed. Figure 2, line 3 laboratory is on the slippery slope to in Washington in 1990, where I was
4 LC•GC Europe June 2015
QUESTIONS OF QUALITY

involved as a co-chair, and the outcome and the final integration data should have a fighting chance for automatically
from that meeting was a scientific paper be documented at the laboratory and integrating your peaks and there is no
on the subject (19). After a follow-up should be available upon request.  need to intercede manually. However,
conference in 1999, the FDA issued we now need to turn our attention to
Guidance for Industry on Bioanalytical Although both the FDA and EMA manual integration. As we can see
Method Validation (20). In the section allow reintegration of chromatograms, from the regulatory citations and the
dealing with routine analysis of samples it must be under controlled conditions bioanalytical guidance documents,
there is mention of sample data and must be reported in the final what is needed is an SOP to control,
reintegration:  report along with who authorized the manage, and authorize integration —
•  An SOP or guideline for sample data reintegration and what the original both automatic and manual.
reintegration should be established. results were together with the need In an ideal world there would be
• This SOP or guideline should explain for audit trail entries and justification. a definition of manual integration.
the reasons for reintegration and how Personally, I prefer the European However, there is not a definition
the reintegration is to be performed. approach because the FDA guidance available for manual integration that I
The rationale for reintegration is over-bureaucratic. This is particularly could find. Neither in Dyson, nor on any
should be clearly described and the case as the FDA requirement for regulatory website. This was the point
documented.  documentation of the requestor and made by Hill et al. (7), who discussed
• The original and the reintegration data authorization by a manager will have the scope and terminology of manual
should be reported. to be outside of a CDS because there reintegration and they concluded that
are currently no functions available to it is essential to develop a consensus
Later in the document, in the section perform this. If required by regulators with respect to definitions. The problem,
on reports for routine studies, there is this needs to be included as a function they noted, is that that a consensus
another section on Documentation for in future versions of chromatography does not exist. 
Reintegrated Data (20), which, in part, data systems. This raises an important question: if
requires:  we can’t define manual integration how
• The method used for reintegration. What is Missing? can we write an SOP on integration as
• The reason for the reintegration. In reading both the GMP and GLP a whole? 
• The requestor of the reintegration and regulations, the European and US
the manager authorizing reintegration. Pharmacopoeia general chapters Scope of an Integration SOP
• Reintegration of a clinical or on chromatography, and the three In the spirit of the early pioneers,
preclinical sample should be bioanalytical method validation this column will present a possible
performed only under a predefined guidance documents, it strikes me approach to writing an SOP in the form
SOP. that there is a fundamental element of a flowchart that will consider some,
missing from the equation. The only but not all, options for both automatic
In 2013, the FDA issued a draft publications that discuss integration and manual integration. Of course,
revision of this Guidance for Industry are the bioanalytical method validation there is a little trepidation because one
where sample data integration was guidance documents. However, the definition of pioneer is finding new and
updated (please remember that focus in all three publications is on exotic ways to die (!) However, I hope
the guidance is still a draft), but the the application of a validated method that this column and my views stimulate
key requirements are essentially NOT on the validation of the analytical a debate about what constitutes manual
unchanged. There is a rearrangement method itself.  integration and we can get a definition
of the order listed above with the It is essential that validation of agreed upon. 
additional need to retain audit trail ANY analytical procedure includes In this discussion we need to
information from the CDS (21).  the integration parameters along balance sound science with regulatory
At last we are getting somewhere! with other methods measured in any compliance. Therefore, let me pose
We now have some guidance albeit method that utilizes chromatography. a question — which is worse: not
in the good laboratory practice (GLP) Note that the stimulus to the revision reintegrating when you know the results
arena. But wait! There is more! Not paper published in 2013 by Martin et are wrong or accepting wrong results
to be outdone by their American al. (23) for a life cycle approach did when you can see an unknown peak in
cousins, the EMA (European Medicines not discuss the inclusion of integration the sample chromatograms? You see
Agency) produced their own guidance parameters (24). I would strongly the dilemma? We know the outcomes
document on Bioanalytical Method suggest that the draft general chapters of both situations are undesirable, but
Validation (22) in 2011. Section 5.5 is include the requirement for stating as a result of inflexible procedures or
concise and devoted to the subject of integration parameters within the life fear of the regulations sound scientific
integration:  cycle of analytical procedures involving judgement cannot be exercised.
• Chromatogram integration and chromatography.  First, let us consider a “no manual
reintegration should be described in a integration allowed” option. Personally, I
SOP.  What is Manual Integration? think that this is an untenable situation,
• Any deviation from this SOP should So far we have discussed the regulatory particularly if a regulated laboratory
be discussed in the analytical report.  issues around peak integration, the is using recently developed methods,
• Chromatogram integration parameters main integration parameters, and the analyzing impurities where the limits of
and in case of re-integration, initial three rules of integration so that you quantification or detection are close to
www.chromatographyonline.com 5
QUESTIONS OF QUALITY

discussion, your laboratory SOP may


Figure 3: A suggested flow chart for manual intervention and manual integration.
have more areas depending on the
work performed. The outcome you
require is consistent and appropriate
Chromatographic Run
manual integration that is scientifically
defensible. Therefore, you need to avoid
Automatic situations where you have inconsistent
integration or inappropriate integration. 

Manual Intervention Versus


Manual
Manual Integration
Laboratory Complete
Investigation No Integration No Acceptable? Yes Analysis In Figure 3 options 1 and 2 are shown
OK?
as manual intervention and option 3 as
Yes manual integration. Let me clarify my
reasoning. 
Manual • Option 1: Peaks have slipped out of
Integration
Options a window and they are not correctly
identified. The automatic integration
is acceptable and all that is required
is to change the peak windows in the
integration method and reprocess.
1. Rename Peak areas are not changed by this
peaks or adjust
windows
approach. 
2. Adjustment of
integration • Option 2: Parameters in the
parameters 3. Manual
reposition of
integration or processing method
baselines need to be adjusted and then
applied to all injections in the run. An
example could be change of the peak
Manual Manual
threshold or minimum area to reduce
Intervention Reintegration the impact of baseline noise. Peak
areas may or may not be changed
under this option but there is no
baseline noise, or investigating stability acceptable we move to a second manual placement of the baselines by
of products. In these situations, manual decision point: is manual integration an analyst.
reintegration is scientifically sound and (whatever that term may cover) • Option 3: Manual placement of
defendable provided that it is covered permitted for this analytical procedure? baselines by the chromatographer is
in the integration SOP.  If not the next stage is a laboratory required because of a late running
A suggested flowchart for investigation. What methods could we peak or noise if undertaking an
consideration of integration is shown in consider for inclusion for no manual impurity analysis. Peaks areas will be
Figure 3. It begins with the completion integration? Perhaps measurement changed by the reintegration.
of the chromatographic run and the of active pharmaceutical ingredients
automatic integration of peaks by (APIs) or registered methods for Note that the three options in Figure 3
the original processing method. The finished product for the active are shown on three different levels. This
resulting chromatograms are reviewed ingredient? If these peaks cannot is my way of illustrating the descent
by the chromatographer to see if be integrated correctly are you out into manual integration hell. In the first
retention times and peak shapes are of control? We will consider if this option you may think of as purgatory,
as expected, the peak(s) have been is tenable when we have finished the second as limbo while the third
correctly identified, baseline placement discussing the flowchart.  option is; well, you get the idea. 
is as required by the analytical In my view, manual integration Options 1 and 2 are manual
procedure, and the sample is integrated must be specifically prohibited in the intervention but the baseline placement
consistently with the standard. There following circumstances: is performed by the CDS and is not
may be other criteria that an individual • Symmetrical peaks that have altered by a chromatographer. These
laboratory wishes to apply. We now acceptable baseline to baseline are the preferred options and easier
come to a decision point: is the run fitting following automatic integration. to justify scientifically. Option 3 is
acceptable? If yes the individual results • Enhancing or shaving peak areas where everything else has failed and a
and the reportable value are calculated to meet SST acceptance criteria chromatographer goes through individual
and all is well. All data at this point have or allowing a run to meet the test chromatograms and repositions the
been calculated automatically by the specification.  baselines where appropriate. This
CDS, the chromatographer is merely latter point is important, it is an exercise
confirming what the software has done Now we come to what constitutes of scientific judgement that needs to be
conforms to pre-defined expectations.  “manual integration”. Figure 3 presents backed up by the procedures within the
However, if the integration is not three options that I have selected for integration SOP.
6 LC•GC Europe June 2015
QUESTIONS OF QUALITY

It’s All Plain Sailing Now? using a scientifically sound, justifiable, comments made during the preparation
Let us return to the scenario we and transparent process. We are of this column.
discussed earlier at the top of the looking at consistency of an integration
flowchart in Figure 3. If the automatic technique among all trainees to ensure References
integration has failed because the a consistent approach, which can (1) Leiner Health Products, FDA Warning
Letter (Aug 2007)
peaks have slipped out of the retention be provided via a CDS by copying (2) R.D. McDowall, LCGC Europe 27(9),
windows (option 1 in the manual methods and chromatograms 486–492 (2014).
intervention section) do you want to to a training project or directory (3) C. Burgess, D.G. Jones, and R.D.
McDowall, LCGC International 10(12),
trigger a laboratory investigation? and allowing chromatographers
791–796 (1997).
Especially when the peak windows to integrate the same files. One (4) R.D. McDowall, LCGC Europe 23(7),
are readjusted and, after reintegration, outcome of the training is that trained 368–374 (2010).
the run now passes. The peaks chromatographers will think about (5) R.D. McDowall, LCGC Europe 23(11)
585–589 (2010).
are now correctly labelled and the where they place a baseline when (6) L. Kaminski et al., LCGC Europe 24(8),
peak areas are unchanged after the manually integrating a peak so that it is 418–422 (2011).
manual intervention. Although the FDA scientifically justified and complies with (7) H.M. Hill, D. Bakes, and I. Love,
Bioanalysis 6(9), 1171–1174 (2014)
classifies this situation under “manual the laboratory procedure.  (8) N. Dyson, Chromatographic Integration
integration”, there is no change to Another outcome of the training Methods. Second Edition (Royal Society
the actual measurement of the peaks is that all trainees must understand of Chemistry, Cambridge, UK, 1998).
that unauthorized manual integration (9) C. Burgess and R.D. McDowall, LCGC
of interest. Should this situation be
Europe, in preparation
classified as manual integration? outside of the scope of the SOP is (10) MHRA GMP Data Integrity Definitions
In my opinion no – this is a manual considered as data falsification. The and Guidance for Industry, Medicines
intervention but not reintegration. This more times that a run is reprocessed and Healthcare products Regulatory
Agency, London (January 2015).
is not intended to be word play but either via manual intervention or (11) MHRA GMP Data Integrity Definitions
a means of trying to define exactly manual integration the more quality and Guidance for Industry, Medicines
what the regulators want in light of assurance and regulatory scrutiny it and Healthcare products Regulatory
Agency, London (March 2015).
zero guidance and multiple citations will attract, that is, the larger the size
(12) EU GMP Chapter 6 on Quality Control
on the subject. Furthermore, because of the red rag. Unfortunately large red (2014).
there is no agreed definition of the rags are not very effective at stopping (13) FDA Current Good Manufacturing
term we have a problem – would regulatory bulls. Practice for Finished Pharmaceutical
Products, 21 CFR 211 (2009).
this be permitted? Would you take As a final comment, chromatographic (14) R.D. McDowall, LCGC Europe 26(6),
the Clint Eastwood approach to risk integration should be included in the 338–343 (2013).
management by feeling lucky or could data integrity self-inspections to ensure (15) R.D. McDowall, LCGC Europe 26(7),
389–392 (2013)
you justify that this is an acceptable that the training and procedure are (16) FDA Compliance Program Guide
practice?  being complied with in operational use. 7346.832 Pre Approval Inspections
Ideally the SST, standards, and QC (2010, effective 2012).
(17) European Pharmacopoeia 2.2.46
samples should be integrated the Summary
Chromatographic Separation Techniques
same way consistently throughout We have discussed manual integration (18) United States Pharmacopoeia <621>
the run; however, this may not be in the context of a regulated laboratory Chromatography (United States
the case, particularly near limits of and have found a number of problems. Pharmacopeial Convention, Rockville,
MD, USA).
quantification or detection. Let us Regulations and guidance are lacking (19) V.P. Shah et al., Pharmaceutical
look at a different situation that can in GMP and the only guidance on the Research 9, 588–592 (1992)
arise when operational efficiency is subject is for regulated bioanalytical (20) FDA Guidance for Industry, Bioanalytical
Methods Validation. FDA, Rockville, MD,
considered and different material laboratories, which can result in
USA (2001).
types are analyzed in a single run. an overly bureaucratic approach. (21) FDA Draft Guidance for Industry,
Although the method is validated for However, we lack an agreed definition Bioanalytical Methods Validation. FDA,
these different sample types, there can of what constitutes manual integration. Rockville, MD, USA (2013).
(22) European Medicines Agency, Committee
be differences in the chromatography Notwithstanding this minor problem, for Medicinal Products for Human Use.
that require different integration a suggested workflow for determining Guideline on Bioanalytical Methods
parameters. So how can you apply the how manual intervention and manual Validation. European Medicines Agency,
London, UK (2011).
same integration parameters across integration can be combined in an (23) G.P. Martin et al., Pharmacopoeial
the run? The classic example is mixing overarching SOP on chromatographic Forum 39(5), (September–October 2013)
APIs and stability samples where integration to meet regulatory concerns. (Available on-line at www.usp.org).
(24) R.D. McDowall, LCGC Europe 27(2),
the methods have different aims; for I hope this will stimulate a debate
91–97 (2014). 
example, determination of purity versus to define what constitutes manual
degradation. Perhaps the case of data integration and that regulatory agencies “Questions of Quality” editor Bob
integrity should win over operational will provide guidance on the subject — McDowall is Director at R.D. McDowall
efficiency and it would be better to eventually Ltd, Bromley, Kent, UK. He is also a
separate different types of analytical member of LCGC Europe’s editorial
procedure?  Acknowledgements advisory board. Direct correspondence
Regardless of the content of the I would like to thank Chris Burgess, about this column should be addressed
final SOP, chromatographers must be Howard Hill, Brian Jones, Mark to the editor‑in-chief, Alasdair Matheson,
trained to perform manual integration Newton, and Paul Smith for advice and at amatheson@advanstar.com
www.chromatographyonline.com 7

You might also like