You are on page 1of 9

Oil-Water Emulsions: Particle Size Distributions

from Elastic Light Scattering Data

MICHAEL HOFER, JOSEF SCHURZ, AND O T T O GLATTER


Institut ffir Physikalische Chemic, Universitgit Graz, Heinrichstrafle 28, A-8010 Graz, Austria
Received July 16, 1987; accepted February 3, 1988

Particle size distributions are important quality criteria for pharmaceutical emulsions. Fat emulsions
for intravenous administration should have a typical particle size of a few hundred nanometers and not
overshoot an upper size limit of 2 urn. Particles with a higher diameter, suspected to cause adverse
reactions (emboli), can be found in a phase contrast microscope. Our method uses elastic light scattering,
i.e., the angular dependent scattering intensity to size particles in the range of 100 nm up to several
micrometers. The advantage of this method compared with quasi-elastic light scattering is the higher
resolution due to a higher information content of the experiment (much lower condition numbers of
the matrix). It takes into account the actual relative refractive index and uses the Lorenz-Mie theory as
a basis of the inversion procedure. This yields accurate number and intensity distributions, defined as
the relative number of particles and the relative scattering intensity in the interval R and R + dR. Since
the scattering intensity is roughly proportional to the sixth power of the radius, larger particles can be
found in the intensity distribution even if their number is in the ppm range. We can therefore supply
the producer with two important and well-defined kinds of information: (a) the number distribution,
which provides the parameters of the particles like mean, standard deviation, and skewness; (b) the
intensity distribution, which is a very sensitive control for undesired large particles. © 1989Academic
Press, Inc.

1. INTRODUCTION m o s t cases the particle size d i s t r i b u t i o n o f the


dispersed phase is a n i m p o r t a n t p a r a m e t e r for
T h e d i s p e r s i o n o f o n e i m m i s c i b l e liquid in the c o n t r o l a n d / o r the stability o f the e m u l -
a n o t h e r liquid in the f o r m o f small d r o p l e t s is sion.
called a n emulsion. W i t h the c o m p o n e n t s wa- Since oil particles in water have a relative
ter a n d oil two types o f e m u l s i o n are possible: refractive i n d e x m o f a b o u t 1.1 o r greater a n d
oil-water (O/W) and water-oil (W/O) since these particles are n e a r l y perfect spheres,
emulsions. W a t e r is the o u t e r or the i n n e r optical m e t h o d s for accurate particle sizing
phase, respectively. should use the L o r e n z - M i e t h e o r y (1, 2) as
E m u l s i o n s have b e c o m e increasingly i m - the basis o f the i n t e r p r e t a t i o n procedure.
p o r t a n t in m e d i c i n e , chemistry, p h a r m a c y , T h e m o s t c o m m o n optical techniques for
a n d m a n y o t h e r fields. F a t emulsions, for ex- particle sizing in the m i c r o m e t e r regime are:
a m p l e , are e m p l o y e d in m e d i c i n e for p a r e n -
teral n u t r i t i o n o r to t r a n s p o r t water-insoluble ( a ) Quasi-elastic light scattering, also re-
drugs, t h e r e b y a v o i d i n g u n d e s i r e d s e c o n d a r y ferred to as "laser D o p p l e r - s h i f t " or " p h o t o n
effects f r o m solvents or dissolving i n t e r m e d i - c o r r e l a t i o n " spectroscopy. C o n v e n t i o n a l de-
aries. T h e y are i m p o r t a n t for special poly- tection at a scattering angle o f 90 ° sizes par-
m e r i z a t i o n processes in c h e m i s t r y o r for sew- ticles f r o m s o m e n a n o m e t e r s u p to several m i -
age purification in e n v i r o n m e n t a l sciences. crometers. T h e m e t h o d gives q u i c k results b u t
Physiologists are interested in the b e h a v i o r o f with p o o r resolution ( 3 ) .
c h y l o m i c r a (which are n a t u r a l "fat-emulsions" ( b ) F r a u n h o f e r diffraction, which evaluates
in the b l o o d ) u n d e r different conditions. In particle size d i s t r i b u t i o n s f r o m the forward

147
0021-9797/89 $3.00
Copyright © 1989 by Academic Press, Inc.
Journal of Colloid and Interface Science, Vol. 127, No. 1, January 1989 All rights of reproduction in any form reserved.
148 HOFER, SCHURZ, AND GLATTER

scattered intensity. This method works well in II. THEORY


the size range of several micrometers up to the Although a full description of the method
millimeter regime, although relatively large can be found in (10), we briefly review the
errors occur for particles smaller than about most important features.
5 ~tm if their contrast is not high enough (m We are able to calculate size distributions
< 1.3 ) because of anomalous diffraction (4). from polydisperse systems if all particles have
(c) Microscopy, which is used to detect the same refractive index and the same shape
particles greater than several micrometers. but have different sizes defined by one size pa-
However, some special techniques such as rameter R. A simultaneous determination of
video-enhanced microscopy can be used to size and shape of polydisperse samples is not
size particles down to several hundred nano- possible; the shape of the particles must be
meters. assumed or known from other experiments.
We have already published a method for de-
Besides these optical techniques, other
termining such shapes from elastic light scat-
powerful methods of particle sizing are in use.
tering data for homogeneous and inhomoge-
Let us briefly sketch four of the most impor-
neous, spherical and nonspherical particles
tant ones:
(11, 12). The above restrictions should be re-
(a) Centrifugation methods (such as the membered although they are of little impor-
Joyce-Loebl disk centrifuge with optical de- tance in the case of emulsions where all par-
tector) are used to size particles in the range ticles are nearly perfect spheres constituted by
of several micrometers down to less than 0.1 the same material. Three size distributions are
/~m (5). commonly used:
(b) The Coulter counter sizes particles by (1) The number distribution Dn (R), giving
measuring the change in resistance of a the number of particles in the size range of R
charged orifice caused by the presence of non- and R + d R .
conducting particles. It is suitable for exam- (2) The mass or volume distribution
ining particles in the range of 1 to about 20 Dm(R), which is the mass or volume in the
/~m (6). interval R and R + d R .
(c) X-ray or neutron small angle scattering (3) The intensity distribution Di(R), de-
can be employed to size particles from several fining the scattered intensity at zero angle
nanometers up to about 100 nm (7). originating from the size interval R and R
(d) Field-flow fractionation techniques are + dR.
used to obtain high resolution from less than The angle-dependent scattering intensity
100 nm up to about 1/~m (8, 9). from all particles l ( h ) is related to the above
The method described below uses elastic distributions by the integrals
light scattering, i.e., the angle-dependent scat-
tering intensity for particle sizing in the range I(h) = Dn(R)ck(h, m, R ) d R [1]
fO °°
of about 0.1 /~m up to several micrometers
with high resolution. The Lorenz-Mie theory z(h)
is used as basis of the data inversion. The
method has the great advantage that very ac- = Dm(R)dp(h, m , R)4)(O, m, R ) - I / 2 d R
curate size distributions are obtained and that ~0 °°
[21
it is applicable to an arbitrary relative refractive
index m in the preresonant Mie region, which I(h)
nevertheless must be known a priori. In the
= Di(R)rb(h, m, R)49(O, m , R ) - ~ d R ,
following, independent and single scattering fO °°
is always assumed. [31
Journalof ColloidandInterfaceScience,Vol. 127, No. 1, January 1989
P A R T I C L E SIZE 149

where q~(h, m, R) represents the scattering of electromagnetic theory, we continued to use


function of a particle with size R and refractive it for the generation of a large array of scat-
index m. For convenience, the scattering angle tering functions of homogeneous spheres. This
0 has been replaced in Eqs. [1]-[3 ] by the array contains functions starting from R
wavelength-independent angular variable h, -- 12.5 nm up to R = 1050 nm (increment:
which is 12.5 n m ) f o r m = 1.001, 1.05, 1.075, 1.1, 1.15,
1.2, 1.25, and 1.3. These functions are used
41r
h = ~ sin 0/2. [4] to fit the measured scattering data with the
indirect transformation method.
R is the characteristic dimension of the par- The method would allow use of a spherical
ticle, i.e., the radius for spherical scatterers. In core-shell model to take into account the dif-
the Rayleigh-Debye-Gans approximation ferent refractive index of the detergent layer
(27rX-1R(m - 1) ~ 1) the three size distri- on the surface. This layer is very thin com-
butions are related by pared with the overall dimensions of the drop-
let and cannot be resolved by light scattering
Di(R) = R3Dm(R) = R6Dn(R). [51 experiments. This fact supports the use of ho-
Starting from the Rayleigh-Debye region mogeneous spheres.
(small scattering objects with low contrast), We have tested the method using simulated
deviations of 4(0, m, R) from R 6 with in- polydispersity profiles under various condi-
creasing size and refractive index clearly show tions (10). Some emulsions of the present se-
the Mie contributions to the scattering func- ries have been investigated with other tech-
tion (13). niques (quasi-elastic light scattering, micros-
It is impossible to solve Eqs. [ 1]- [ 3 ] ana- copy, Fraunhofer diffraction, and Coulter
lytically because no general analytical expres- counter). These methods gave similar overall
sion for 4(h, m, R) in the Lorenz-Mie theory shapes of the polydispersity profiles but were
exists. Therefore, numerical methods must be not capable of the high resolution of the elastic
applied to solve Eqs. [ 1]- [ 3 ] to obtain the de- light scattering technique.
sired particle size distribution. Discussion of The resolution of our method is the same
the mathematical details of the solution pro- as the resolution for Rayleigh-Debye scatter-
cedure (indirect transformation method) can ing experiments (7, 10). Sharp peaks with a
be found in (7, 10). It was necessary to modify size ratio down to 1:1.2 can be resolved. The
the method due to the fact that the Lorenz- precision necessary for such a resolution is in
Mie scattering functions should be used to ap- the order of several percentage points. All our
proximate the measured data instead of the experimental data were below this limit.
Rayleigh-Debye functions generated inter- To show the capabilities and the limitations
nally by the program. of our method under experimental conditions
For the generation of the Mie scattering we shall take as an example fat emulsions for
functions we have used the T-matrix algo- parenteral nutrition and focus our attention
rithms kindly sent to us by Professor P. W. on it in the following.
Barber. Since these programs can handle
spherical and spheroidal, homogeneous and III. STABILITY OF FAT EMULSIONS
inhomogeneous scattering objects (14-19) Emulsions are thermodynamically unstable
they are not optimized for the calculations of systems. They can be stabilized by emulsifiers,
scattering functions of spheres. Nevertheless, which create barriers to the coalescence of
since we have already performed many cal- droplets by reducing the interracial tension
culations with this program package, which between two immiscible liquids because they
proved to be accurate and "easy to handle" possess hydrophilic and hydrophobic regions
even for persons not involved in the "kernel" within the same molecule.
Journal of Colloid and Interface Science, Vol. 127,No. 1, January 1989
150 HOFER, SCHURZ, AND GLATTER

Some pure phospholipids, especially phos- provements and of the primary data handling
phatidylcholine, form neither good mechani- can be found in (26).
cal nor good electrical barriers to droplet co- All scattering curves were measured with
alescence. One therefore utilizes egg or soy this instrument in the angular range of 12
lecithin together with ionic lipids (e.g., phos- ~< 0 ~< 120 ° (0.003 ~< h ~< 0.026) at scattering
phatidylserine) for the stabilization of com- angles in three subdomains: 18 values were
mercial fat emulsions. Ionic lipids not only recorded from 12 to 30 ° , 12 from 30 to 60 ° ,
increase the surface potential but also produce and 14 from 60 to 120 ° in an equidistant h-
complex interfacial films with phosphatidyl- spacing. This is necessary in order to allow for
choline, which results in a liquid crystalline the solution of Eqs. [ 1] - [ 3 ] with the indirect
gel structure at the interface (20-22). transformation technique. At each scattering
In contrast to this, certain additives like angle 100 intensity values (integration time
dextrose or electrolytes reduce the zeta poten- = 0.2 s/value) were recorded and the weighted
tial considerably, leading to more unstable mean intensity was calculated. As statistical
systems (23). control of the averaging process we used the
The physical stability of emulsions and the x-square and the F test (27). The precision of
presence or absence of aggregation have im- our final data shown in Section V is in the
portant clinical consequences. Emulsions range of 0.5-3%.
containing droplets exceeding 6 # m in diam- FICA light scattering cuvettes are made of
eter are known to cause adverse reactions, crown glass. The minimum sample volume is
particularly embolism in the lungs (24, 25). about 2 ml. To avoid contributions due to re-
Therefore, in Europe, fat emulsions for intra- flections the inner wall of the cuvette is made
venous administration are not allowed to conical (see Fig. 1).
overshoot an upper size limit of 5 #m. More-
over, it has been noticed that aggregation,
creaming, and storage capacity are strongly
dependent on the mean diameter of the sample
as well as on the presence of a small amount
of micrometer-sized particles. Consequently,
accurate determination of particle size distri-
bution is very helpful in the refinement of C1 --A!
production processes and in stability evalua- GFt--~....... A2
tion and thus leads to higher security for the
patient. GF'2--~
C2--~--A3
IV. EXPERIMENTAL PROCEDURES
GC--
A. Experimental Setup
--F1
The experimental setup consists of a FICA - F2
-

50 light scattering photometer (Sofica Co.,


France) which has been modified in order to
C3--
allow precise measurements of scattering an-
gles down to about 12 °. The instrument was
D2
also equipped with additional electronic de-
vices for precise time integration and the pos- ~ - - L SC
sibility of transferring data automatically to a
computer via a standard serial interface (RS FIG. 1. Apparatusforthe purificationofwaterto produce
232C). A detailed description of these im- solvent free from dust and air-bubbles.
Journalof Colloidand InterfaceScience,Vol. 127, No. 1, January 1989
PARTICLE SIZE 151

Inversion procedures for obtaining the size tering intensity due to the high contrast (m
distributions were performed on a UNIVAC = 0.75 for air in water). Errors at low angles
1100 / 81 computer, using the modified indi- are more critical than those at high scattering
rect transformation technique. All polydis- angles since they influence the extrapolation
perse profiles are calculated up to a scattering of the scattering curve to 0 = 0. This part of
particle diameter of 2100 nm. The method is the scattering function contains almost all the
not restricted to this relatively low value but information about the maximum dimension
further extension would imply the measure- found in the sample, and hence, these errors
ment of the scattering function at scattering lead to a wrong polydispersity profile for large
angles lower than 12 ° , which is not possible particles. It is therefore necessary to prepare
with the present instrument. A new elastic light solvent and sample under dust-free conditions
scattering goniometer is under develop- without introducing air bubbles into the so-
ment now. lution. This procedure is usually fairly tricky;
it is still one of the main problems when elastic
B. Sample and Solvent Preparation light scattering measurements at small angles
The emulsions contain water, Soja oil (20 should be performed. In the following we de-
wt%), Lecithine (1.2%), Glycerine (2.5%), scribe an apparatus which enables the pro-
and NaOH to adjust the pH. They are pro- duction of sufficiently purified water.
duced with the usual technical procedure; i.e., Water can prove to be a somewhat insidious
Glycerine and water are mixed and heated to solvent in elastic light scattering experiments.
the right temperature, then mixed with Leci- It has the ability to trap a n d / o r generate air
thine and NaOH. This mixture is dispersed bubbles randomly. We have verified this be-
and homogenized after the addition of the oil, havior in a series of quasi-elastic light scatter-
diluted to the right concentration of the oil, ing experiments. With such an experiment it
and finally homogenized again in two steps. is possible to determine the hydrodynamic ra-
According to the sampling theorem of the dius of scatterers in solution in about 5 min
Fourier transformation the smallest h-value (30). Two cleaned and dried cuvettes were
(hmin) defines roughly the maximum dimen- filled with purified water and observed over 5
sion, which can be evaluated from the scat- days. At the beginning, both samples showed
tering function (see, for example, (28, 29)). small impurities in the size range of about half
Therefore, if we try to investigate large particles a micrometer, although the last filter of the
(of several micrometers), we should be able purification column had a pore size of 200
to measure the scattering function starting nm. During the time of observation, in the
from a scattering angle of at least 10 °. first sample the size of the air bubbles increased
The measured scattering function Im~ is the to about 4 ~m, whereas in the second sample,
sum of the scattering from the particles Ip and the hydrodynamic radius fluctuated in the vi-
the solvent Is. The contribution of the particles cinity of 1 #m. We have tried to improve the
decreases approximately with h -4 whereas a reproducibility but without success.
pure solvent ideally scatters light independent We therefore started to eliminate step by
of the scattering angle 0. At high 0 the two step all parts which could introduce impurities
quantities Ip and Is can therefore approach the and finally developed a closed system that
same order of magnitude. The subtraction (Ip yields the desired results. A schematic diagram
+ Is) - Is then introduces large errors iflp and of this apparatus is shown in Fig. 1: First, water
Is are not measured with high accuracy. from the piping is taken and distilled (D 1) di-
On the other hand, dust particles and micro rectly into a drop bottle (DB). This drop bottle
air bubbles either in the sample or in the sol- is then disconnected from the column and
vent cause relatively large errors in the poly- stored in a vacuum drying oven and evacuated
dispersity analyses because of the strong scat- down to 3 mbar. At this low pressure all air
Journal of Colloid and Interface Science, VoL 127, No. 1, January 1989
152 HOFER, SCHURZ, AND GLATTER

bubbles will vanish. The drop bottle will be (~20% w/w) was highly diluted (1:106). This
attached to the purification column and closed dilution shortens the stability of the samples
with a plug containing a filter of pore size 50 considerably. While the stock solution has a
nm. The cock (C 1) will be slowly opened now stability of several months, the stability of the
and water passes through two glass frits (GF1, diluted samples reduces to 1-2 days. This sta-
GF2) of pore sizes <40 #m and < 16 #m, re- bility was checked by repeating the measure-
spectively. This is managed in such a way that ment at low scattering angles. Unstable prep-
no water drops can be formed which would-- arations were rejected and the sample was
when falling--trap air bubbles. prepared once again.
Cock C2 controls the amount of water in There is no direct evidence that the size dis-
the graduated cylinder (GC). This cylinder is tribution is completely unchanged during the
directly connected with two filter holders, dilution procedure. For energetical reasons,
which are made of polycarbonate (F1, F2). however, it is impossible for the size of the oil
The membrane filter in the first holder has a droplets in the emulsion to be decreased by
pore size of 450 nm, the second a pore size of dilution. Any increase in size corresponds with
50 nm (regenerated cellulose, Sartorius, FRG). a decrease in stability. Only stable solutions
All inlets for pressure equalization (A 1, A2, were used in the experiments. The original
and A3) are closed with the same 50-nm samples were analyzed by phase contrast mi-
membrane filters in a polycarbonate filter croscopy. Since this method is not able to give
holder. Back pressure due to F1 and F2 will accurate polydispersity profiles it can be used
be overcome simply by increasing the height only to check for very large particles and for
of the water in the graduate cylinder (GC) to a rough estimate of the mean diameter of the
obtain a pressure of about 10 mbar. Thus, the oil droplets. At lower concentrations (dilution
water flows directly into the second distilling factor 100-1000), quasi-elastic light scattering
apparatus (D2), which is made of glass with was used to check for changes. Within the res-
no connecting pieces. The water (about 50 ml) olution capabilities of the above methods, no
in the 250-ml glass flask will be heated by change in polydispersity profiles was observed.
means of a heating mantle whereby the tem- For precise evaluation of the size distribu-
perature is kept at 90°C to avoid boiling. It tion it is necessary to know the ratio m of the
slowly evaporates and flows down into the light refractive indices of the oil phase and of water
scattering cell (LSC) along the wall to prevent in the emulsion. Determination of the indices
generation of drops. of the separated phases was done at the factory
The apparatus contains no glass joints. All before the emulsification process. The ratio m
connections are made of high-quality silicon was 1.107 _ 0.005. An error of 0.01 in m
tubing. C1, C2, and C3 are Teflon cocks so would not cause essential errors in the results.
that particles due to attrition and the need to We checked the refractive index of the oil
use grease may be avoided. phase after induced creaming in our labora-
This apparatus provides water sufficiently tory. Perfect agreement was found.
purified that it no longer has the capability of
trapping air bubbles. The water will now be V. RESULTS
so clean that even at a scattering angle of 15 o With elastic light scattering, we have inves-
it is impossible to measure a correlation func- tigated more than 100 samples of fat emulsions
tion in a quasi-elastic light scattering experi- for parenteral nutrition, using commercially
ment. This ensures that no scattering particles available, aged samples as well test prepa-
of appreciable size are present in the solvent. rations from Leopold & Co. (Graz, Austria)
Water prepared in this way was used to di- prepared with the aim of further improving
lute all samples to obtain sufficiently accurate the quality of the final emulsions for clin-
scattering data for Is and lp. The stock solution ical use.
Journal of Colloid and Interface Science, Vol. 127, No. 1, January 1989
PARTICLE SIZE 153

We discuss three examples from these test Z, ~ BL ,46


v I . B L 48
preparations, denoted by BL 46, BL 48, and
BL 51. In addition, we discuss one represen-
tative example from the bulk of commercially
available emulsions (SA 1) together with an
aged sample of the same producer (SA 2).
Figure 2 shows the measured scattering
curve of the three test preparations BL 46, BL
0 I I I
48, and BL 51 (circles), the fit (full line), and 30O 600 900 1200
the extrapolation to zero scattering angle R CnmJ

(dashed line). They have been plotted in the FIG. 3. Number distributions calculated via indirect
same logarithmic scaling in order to show transformation method from experimentaldata shown in
clearly the differences. BL 46 and BL 51 are Fig. 2. There are no significant deviations of mean and
very similar. In contrast to this, BL 48 shows skewness between the three samples.
a strong increase in scattering intensity in the
innermost part of the scattering curve. Since
of all distributions is small, indicating a good
large particles show a c o n s i d e r a b l e f o c u s i n g ef-
preparation. However, BL 51 is the best prep-
f e c t on the light--which means that they scat-
aration in the sense of monodispersity,
ter very efficiently to small scattering angles--
whereas it is difficult to recognize any differ-
BL 48 is suspected to have large oil droplets
ences between BL 46 and BL 48. No large par-
in the emulsion. Is it possible to recognize
ticles can be seen in this representation of the
them in a n u m b e r distribution?
scattering data although the innermost part of
Figure 3 depicts the n u m b e r distributions
the scattering curve of BL 48 (h ~< 0.01 or 0
D ~ ( R ) of BL 46, BL 48, and BL 51 evaluated
40 ° ) shows this strong increase in intensity
via the indirect transformation method. The
(see Fig. 2). This means that only a few large
mean of all three samples is very similar and
particles are present in sample BL 48.
lies in the vicinity of 150 nm. The skewness
The computed intensity distributions of BL
46, BL 48, and BL 51 are shown in Fig. 4.
5.5 Although BL 48 and BL 46 behave similarly
in the n u m b e r distribution (see Fig. 3 ) it now
turns out that BL 48 contains some large par-
5.5

.E
5.5
BL 48

17(~\\ , \
z,.5
46 7/\i\
! l ,
~ 5 F I ',

3.5 48

51 of "-Y ,-",~ ,
o 300 600 900 12oo
2"50.0 0.01 0.02 0.03-- R Into]
," h [nm-']
FIG. 4. Intensity distributions computed from experi-
FIG. 2. Measured scattering intensities (O) and fit mental data (Fig. 2). BL 46 and BL 51 showa monomodal
(--) extrapolated to zero angle (---) of three differently behavior; BL 48 is stronglyaffectedby contributions from
prepared fat emulsions for parenteral nutrition (angular large particles. Note that the shoulder at R ~ 900 nm is
range: 12 ~<0 ~< 120° or 0.003 ~<h ~<0.260). due to the limit of the experimental setup.
Journal of Colloid and Interface Science, Vol. 127, No. 1, J a n u a ry 1989
154 HOFER, SCHURZ, AND GLATTER

titles. Their concentration is so low that it is to


impossible to resolve them in the number dis-
tribution. However, their scattering power is
so high that they can be found easily in the
intensity distribution (Fig. 4). The shoulder 5
at R "~ 900 n m indicates the presence of oil
particles too large to be resolved with the pres-
ent experimental setup (0rain ~- 12°). The
i i
method cannot give the exact size of these 00 300 600 900
I
1200
largest particles, but their existance is indicated b R [nrn]
by the shoulder. For better resolution at large
FIG.6. Number distributionsevaluatedfrom scattering
R-values (up to several micrometers) one must data in Fig. 5 differjust slightlyfrom each other. The nar-
go down to scattering angles below 12 °. row shape of the distributions seems to indicate the good
The very small negative part of BL 46 at R quality of both samples.
"~ 900 nm is due to the fact that the indirect
transformation method does not impose ad-
ditional constraints such as nonnegativity on sition (31) has not yet led to improved solu-
the solution. However, we have found that the tions for our inverse problem.
quality of the solution of the inverse problem Let us now focus our attention on com-
can often be improved by the additional con- mercially available fat emulsions for parenteral
straint of nonnegativity (31), which is a phys- nutrition. We shall discuss one representative
ically based one, since there are no negative example of an emulsion still in clinical use
numbers of particles. This constraint avoids (SA 1) and another that is aged and no longer
oscillations around the abscissa without af- in clinical use (SA 2). For convenience, these
fecting the resulting solution essentially. A se- two examples come from the same producer.
ries of tests must be performed before this Figure 5 gives the measured scattering
constraint is used in the standard evaluation functions (circles), the fit of the indirect
routine. The use of singular value decompo- transformation method to the data (full line),
and the extrapolation to zero angle (dashed
line). The only remarkable differences be-
tween these two scattering functions is the
6.0 , slightly stronger falloff of SA 2 and a somewhat
different extrapolation to low scattering angles.
The number distribution in Fig. 6 shows
6+0 very narrow skewness of both samples with a
~o mean diameter in the vicinity of 120 n m in-
dicating a considerable monodisperse prepa-
5,0 ration. Up to R = 1050 nm no large particles
can be recognized in either SA 1 or SA 2, the
latter showing a small number of oil droplets
o in the region 300 ~ R ~< 500 nm. From the
intensity distribution in Fig. 7 evaluated from
2 scattering data in Fig. 5, two main features of
3.0 the aged sample SA 2 become obvious: first,
0.0 O.Ol 0102 0103
~. h Into-I] the shift of the maximum to higher R-values,
FIG. 5. Scatteringdata (©) and fit (--) of commercially
which can be interpreted in terms of droplet
available(SA 1) and aged sample (SA 2) of the same pro- coalescense, a process taking place in every fat
ducer. emulsion; second, the rise of larger particles
J o u r n a l o f C o l l o i d a n d Interface Science, Vol. 127, No. 1, January 1989
PARTICLE SIZE 155

lO Lan, K. H., and Ostrowsky, N., J. Chem. Phys.


82, 1551 (1985).
4. Jones, A. R., J. Phys. D. 10, 163 (1977).
5. Coil, H., and Searles, C. G., J. Colloid Interface Sci.
115, 121 (1987).
6. Groves, M. J., J. Dispersion Sci. Technol. 1, 97 (1980).
7. Glatter, O., J. Appl. Crystallogr. 13, 577 (1980).
8. Yang, F. J., Myers, M. N., and Giddings, J. C., J.
Colloid Interface Sci. 60, 574 (1977 ).
0 t 9. Yang, F.-S., Caldwell, K. D., and Giddings, J. C., J.
300 600 900 1200 Colloid Interface Sci. 92, 81 (1983 ).
-"- R [ n m ] 10. Glatter, O., and Hofer, M., J. Colloid Interface Sci.
122, 496 (1988).
1716. 7. Intensity distributions calculated from scattering
11. Glatter, O., Hofer, M., Jorde, Ch., and Eigner,
data in Fig. 5. The distribution of the aged sample (SA 2 )
W.-D., J. Colloid Interface Sci. 105, 577 ( 1985 ).
shows a shift to higher particle sizes with a relatively large
12. Glatter, O., and Hofer, M., J. Colloid Interface Sci.
number of oil droplets in the critical range for clinical
122, 484 (1988).
application. 13. Hofer, M., Thesis, University of Graz, 1987.
14. Barber, P. W., and Yeh, C.,Appl. Opt. 14, 2864 (1975).
15. Barber, P. W., IEEE Trans. Microwave Theory Tech.
(R > 900 nm) suspected to cause dangerous
MI-T-2S, 373 (1977).
adverse reactions like lung emboli when ad- 16. Barber, P. W., and Wang, D.-S,, AppL Opt. 17, 797
ministered intravenously. (1978).
The techniques described above can be use- 17. Wang, D.-S., and Barber, P. W., Appl. Opt. 18, 1190
fully applied to any kind of oil/water or water / (1979).
18. Wang, D.-S., Chen, C. H. H., Barber, P. W., and
oil emulsions provided that droplets do not
Wyatt, P. J., Appl. Opt. 18, 2672 (1979).
exceed several micrometers in size, which 19. Barber, P. W., Wang, D.-S., and Long, M. B., Appl.
would imply the measurement of the scatter- Opt. 20, 1121 (1981).
ing function at very small scattering angles (19 20. Ries, H. E., Matsumoto, M., Uyeda, N., and Suito,
< 5 °). One preferably uses Fraunhofer dif- E., in "Monolayers" (E. D. Goddard, Ed.), p. 286,
Advanced Chemistry Series, Vol. 144.
fraction methods (32) in the latter case.
21. Rydhag, L., and Wilton, I., J. Amer. Oil Chem. Soc.
Investigations using elastic light scattering 58, 830 (1981).
techniques in the Rayleigh or preresonant Mie 22. Becher, P., "Emulsions Theory and Practice." Rein-
regime are not restricted to simple polydis- hold, New York, 1965.
persity analyses. This kind of technique can 23. Black, C. D., and Popovich, N. G., Drug Intell. Clin.
Pharmacol. 15, 184 (1981).
also be used to study various features of emul-
24. Geyer, R. P., Watkin, D. M., Matthews, L. W., and
sion systems like stability properties or kinetics Store, J., Proc. Exp. Biol. 77, 872 (1951).
of droplet coalescence under different tem- 25. Fujita, T., Sumaya, T., and Yokohama, K., Eur. Surg.
perature, electrolyte, or additive concentra- Res. 3, 436 (1971).
tions. 26. Hofer, M., Diplomarbeit, University of Graz, 1984.
27. Brandt, S., "Statistical and Computational Methods
ACKNOWLEDGMENTS in Data Analysis." North-Holland, Amsterdam,
1970.
This work was supported by the ()sterreichischer Fonds 28. Bracewell, R., "Fourier Transform and Its Applica-
zur Frrderung der wissenschaftlichen Forschung. We thank tions." McGraw-Hill, New York, 1965.
Mrs. K. Schaupp and J. Kerbl from Leopold & Co, Graz, 29. Glatter, O., J. Appl. Crystallogr. 10, 415 (1977).
for generously providing the samples and for fruitful dis- 30. Pecora, R., "Dynamic Light Scattering." Plenum, New
cussions. We also thank Mrs. B. Miiller for drawing the York, 1985.
figures. 31. Lawson, C. L., and Hanson, R. J., "Solving Least
Squares Problems." Prentice-Hall, Englewood
REFERENCES Cliffs, NJ, 1974.
1. Lorenz, L., Videnskab. Selskab. Skrifter. 6, (i 890). 32. "Proceedings, International Symposium on 'Optical
2. Mie, G.,Ann. Phys. 25, 377 (1908). Particle Sizing: Theory and Practice,' Rouen,
3. Bertero, M., Brianzi, P., Pike, E. R., de Villers, G., 1987."

Journalof ColloidandInterfaceScience,VoL 127,No. 1,January1989

You might also like