Professional Documents
Culture Documents
Direct Crude Oil To GCMS
Direct Crude Oil To GCMS
ION SOURCE
CO2
Fig. 1. Schematic drawing of the GC-IRMSsystem. I = Sample inlet capillary; 2 = reference inlet capillary; 3 = sample line; 4 = refer-
ence line; 5 = splitter union; VI = heart split valve; V2 = reference injection valve.
CO2 gas, ions at masses 44 (12C160160), 45 It is possible to select a whole section of the chroma-
(1sc160160 ) 12C160170) and 46 (12C’60180, togram, or just isolated components from the origi-
13C’70160) are produced. nal mixture, for isotopic analysis in the mass spec-
These ions are focussed by a stack of electrostatic trometer .
lenses into an ion beam and accelerated down a The heart cutting process involves the combina-
4000-V potential gradient into a 7.5 KGauss magnet- tion of a mechanical and pressure switching system.
ic field provided by a permanent magnet. Ions of When the heart split valve (VI) is in the closed posi-
masses 44, 45 and 46 are separated in the magnetic tion the effluent of the GC column is directed wholly
analyser, deflected through 90” along separate circu- down the sample line (3). When the heart split valve
lar paths with radii of curvature of approximately 9 is in the open position, the effluent at the GC column
cm. They are collected simultaneously within 3 Fara- is forced wholly towards the GC detector by the back
day cups where the currents produced are proportio- pressure, which is applied at the exit of the sample
nal to the number of ions collected. line capillary by the helium make up gas [He(l)].
The helium make up gas [He(2)] ensures efficient
The interface flushing of the heart split valve.
The interface consists of a continuous flow dual in- Material eluting from the sample line capillary is
let. High purity helium gas is drawn continuously completely combusted to CO2 and water on passage
through the inlet into the high vacuum enclosure of through a furnace tube filled with CuO granules and
the mass spectrometer, from the sample line open maintained at 900 “C. The helium make up gas He-
split (OS,_&) and the reference line open split (OS&. (1) gives a high linear velocity of the chromatograph-
The open splits prevent any disturbance within the ic components through the furnace tube, in order to
ion source from pressure transients which may occur maintain chromatographic resolution.
during the mechanical switching of valves in both the A cryogenic trap, maintained at -100 “C, is fitted
reference and sample lines. along the sample inlet capillary to remove water
The sample to be analysed is injected onto the col- from the sample gas. Pulses of CO2 reference gas are
umn of the gas chromatograph and its components injected via the on/off valve V2. The reference gas
are chromatographically separated. At the exit of pulses are injected such that they do not coincide
the column the components may be directed towards with any sample peaks. The reference gas is precali-
the GC detector or towards the MS sample inlet line. brated and its 13C and “0 values are calculated on
trendsin analyticalchemistry, vol. 9, no. IO, 1990 333
the international PDB standard scale. As in conven- carrier gas and injections were performed in split
tional IRMS, the isotopic ratios 45/44 and 46144 for mode. The temperature programme was from 35 “C
both the sample and reference CO gases are mea- to 150 “C at 4 “Clmin. Due to the large concentra-
sured by the mass spectrometer; Q452CO, and d4%02 tion of methane in the natural gas samples, 1~1 gas
are calculated from: was injected at a split ratio of 3O:l for the analysis of
methane while a splitless injection of 4 ,ul was used
(45/44)sample-(4Y44)reference x looo for the analysis of the remaining components.
d45co 2 =
(45W)reference The MS conditions for these analyses were 200,~A
trap current, 100 eV ionisation energy.
Data collection is on an IBM PS2 model 50 com-
(46/44)sample-(46/44)reference x looo puter system.
6%0, =
(46/44)reference For calculation purposes a CO, reference gas was
automatically introduced into the IRMS system in a
The contribution from 170 is corrected for, ac- series of pulses before and after the array of chroma-
cordin to formulations established by Craig”, and tographic peaks of interest.
the 6 ,BC is thus obtained. System control, data ac-
quisition and data processing are provided by an Results and discussion
IBM PS/2 Model 50 which communicates with the in- Reproducibility of standard mixture
strument via the RS232C interface. The reproducibility of the instrument was tested
by analysing a standard mixture made up of C,,-C,,
Experimental conditions n-alkanes together with pristane and phytane, dis-
The GC-IRMS analysis was performed on a VG solved in hexane. A volume of 1 ,ul of the standard
Isochrom II system interfaced to a Dani 8510 gas mixture was injected at 1:50 split ratio, allowing 50
chromatograph. For standard samples, oils and satu- nanograms of each component on the column. The
rated fractions, the GC was fitted with a fused silica standard mixture was analysed using both the ‘heart
OV-1 column (25 mm x 0.22 mm I.D.). Helium (0.8 cutting’ process and allowing the whole run through
bar) is used as a carrier gas and the injections were the mass spectrometer. There were no significant
performed in split mode. For whole oil analysis, the differences found between the two techniques. The
GC system was programmed from -10 “C to 300 “C results from five parallel analyses are shown in Table I.
at 4 “Urnin and held isothermally at 300 “C for 20 These analyses show good reproducibility with a
min. For analysis of the saturated hydrocarbon frac- standard deviation of 0.08-0.17, based on five runs
tion, the temperature programme was from 60 “C to (Table I). The same compounds were also analysed
300 “C at 4 “Urnin and held isothermally at 300 “C on a conventional dual inlet instrument after sealed
for 20 min. tube combustion (Table I). These analyses give a
For gas analysis, the gas chromatograph was fitted standard deviation on five runs of 0.05-0.30 on the
with a 10 m X 0.32 mm I.D. Poraplot Q capillary col- different alkane standards. Comparison between the
umn from Chrompak. Helium (3 bar) is used as a results for single components on the dual inlet instru-
TABLE I. Stable carbon isotope ratios of standard alkane mixture by GC-IRMS and dual inlet IRMS
ment and a mixture of these components on the TABLE III. Stable carbon isotope ratios of n-alkanes in North
GC-IRMS instrument shows only minor variation. Sea crude oils analysed by GC-IRMS
Most of the components have mean 613C values
Compound 613C,,, (X0) Standard deviation
which show less than 0.3%0 variation between the (5 parallels)
two techniques (Table I).
Ekofisk Statfjord Ekofisk Statfjord
oil oil oil oil
Analysis of natural gas samples
Natural gas samples from an onshore U.K. field nC6 -28.08 -29.59 0.37 0.18
were analysed and isotopic composition of C, to C, nc7 -27.93 -29.18 0.18 0.38
EKOFISK OIL
ot.\.. I~.,I~~,,~..,,..,,..,...,..,,,,.J
42 ‘43 44 45 46 47 48 49 50 51
Time (minutes) TABLE V. Stable carbon isotope ratios of n-alkanes in satu-
rated hydrocarbon fractions of North Sea crude oils analysed by
Fig. 2. Isotope gas chromatogram of Ekofisk oil using Mass 44 GC-IRMS
from the MS on Isochrom II.
Compound 613CPna (X0) Standard deviation
(5 parallels)
Ekofisk Statfjord Ekofisk Statfjord
oil oil oil oil
which will have a greater effect for the whole oil
analysis than the background material in the satu- G3 -28.38 -29.58 0.10 0.03
rated fractions. Determinations of the stable carbon 44 -28.24 -29.63 0.05 0.11
TABLE VI. Stable carbon isotope ratios of acyclic isoprenoid iso-alkanes. It is too early to give an exact definition
alkanes in saturated hydrocarbon fractions of North Sea crude of how complex a sample can be and yet still give
oils analysed by GC-IRMS
good results. A rule of thumb is that there should be
Compound 6t3C,oa (%o) Standard deviation good separation between two peaks to give good re-
(5 parallels) sults. The case where one compound is only repre-
sented as a shoulder on the peak of another com-
Ekofisk Statfjord Ekofisk Statfjord
oil oil oil oil pound will not give reliable results.
Another problem is the quantity of material
iC,s -24.27 -30.02 0.17 0.03
needed. Small peaks will yield carbon isotope data,
iC,, -24.48 -30.20 0.29 0.13
-25.95 -29.91 0.15 0.02 but these may be quite erroneous. The amount of
iCis
iCi, material will vary from compound to compound and
(Pristane) -27.79 -30.73 0.17 0.08 depends on the amount of CO, generated. If the
iC2a peaks are small, extra caution should be exercised in
(Phytane) -27.28 -30.68 0.24 0.11
interpreting the isotopic values given by the instru-
ment.
4 J. M. Hayes, R. Takigiku, R. Ocampo, H. J. Callot and P. 10 D. E. Matthews, E. Ben-Galin and D. M. Bier, Anal.
Albrecht, Nature, 329 (1987) 48-51. Chem., 51(1979) 80-84.
5 I. Gilmour, P. K. Swart and C. T. Pillinger, Org. Geochem., 11 H. Craig, Geochim. Cosmochim. Acta, 12 (1957) 133-149.
5 (1984) 665-670. 12 M. A. Northam, in B. M. Thomas et al., Petroleum Geo-
6 H. H. Chung, W. B. Sheri and P. L. Grizzle, Geochim. Cos- chemistry in Exploration of the Norwegian Shelf, Graham &
mochim. Acta, 45 (1981) 1803-1815. Trotman, 1985, pp. 93-100.
7 W. H. Stahl, Geochim. Cosmochim. Acta, 42 (1978)
1573-1577. Dr. Malvin Bjoray is managing director of Geolab Nor AIS, Hor-
8 M. Sano, Y. Yotsui, H. Abe and S. Sasaki, Biomed. Mass nebergveien 5, 7038 Trondheim, Norway since 1984; Keith Hall
Spectrom., (1976) l-3. does chromatographic consultancy work in geochemical, medical
9 D. E. Matthews and J. M. Hayes, Anal. Chem., 50 (1978) and chemical R&D; Janine Jumeau is Product Manager for VG
1465-1473. Isotech Ltd., Aston Way, Cheshire CWIOOHT, U.K.