Professional Documents
Culture Documents
for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: C1875 − 18
C25 Test Methods for Chemical Analysis of Limestone, sion spectra. The spectra are dispersed by a high resolution
Quicklime, and Hydrated Lime grating and the intensities of the individual wavelengths are
C114 Test Methods for Chemical Analysis of Hydraulic measured. Elemental concentrations are determined by relating
Cement the emission intensities at specific wavelengths for an unknown
C1738/C1738M Practice for High-Shear Mixing of Hydrau- sample to analytical curves made from reference standards of
lic Cement Pastes known composition. The concentration of the elements in the
D1129 Terminology Relating to Water solutions can then be calculated. Internal standards compensate
D1193 Specification for Reagent Water for variations in test sample flow rates in the instrument.
D5673 Test Method for Elements in Water by Inductively
Coupled Plasma—Mass Spectrometry 5. Significance and Use
D7111 Test Method for Determination of Trace Elements in 5.1 The chemical composition of the liquid in cementitious
pastes is an important indicator of the solid component
1
reactivity at early times, being influenced by the content and
This test method is under the jurisdiction of ASTM Committee C01 on Cement
rate of reaction of readily soluble alkali components, lime, and
and is the direct responsibility of Subcommittee C01.23 on Compositional Analysis.
Current edition approved July 1, 2018. Published August 2018. DOI: 10.1520/ other soluble phases. Monitoring the solution composition with
C1875-18 time can provide valuable diagnostic information about cement
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
--`,`,````,``,,,,`,,`,`,``,,,,,`-`-`,,`,,`,`,,`---
measured as a test sample as described in Section 11. pling during the mixing process, especially for blended powder
components such as portland cement and fly ash. Powders
7.4 Argon—High purity grade (99.99 %).
should therefore be homogenized for 30 minutes with a
7.5 Calibration Standards—Single-element or multi- planetary mixer or similar device.
element calibration reference solutions that combine appropri-
ate volumes of the stock solutions or weighed reagent-grade 10. Sample Preparation and Collection
chemicals in acid-rinsed volumetric flasks. Dilute to the
volume of the flask using a 5 % (v/v) nitric acid solution. Prior 10.1 High Shear Mixing:
to preparing the multi-element solutions, analyze each stock 10.1.1 Blend the solid and liquid components to make a
solution separately to check for strong spectral interference and paste according the standard practice described in Practice
the presence of impurities. Take care when preparing the D1129.
multi-element solutions to verify that the components are 10.1.2 Set a timer to mark the end of mixing. The time on
compatible and stable (i.e., that they do not interact to cause this timer will serve as the total reaction time for the remainder
precipitation) and that none of the elements present will of the procedure.
contribute to mutual spectral interference. Calibration refer- 10.1.3 Stir the paste with the spatula for no more than 15 s
ence solutions must be verified initially using a quality control ensure homogeneity.
sample and monitored periodically for stability. Mixed calibra- 10.1.4 Transfer (25 6 1) g of the paste into each of twelve
tion standards will vary, depending on the number of elements centrifuge tubes with secured lids in the minimum amount of
being determined. time possible, preferably less than 2 min.
--`,`,````,``,,,,`,,`,`,``,,,,,`-`-`,,`,,`,`,,`---
10.2.7 Transfer pore solution from syringe through filter 13.4 Wavelength Profiling—Perform any wavelength profil-
into a clean empty plastic sample container. Label and note ing that is specified in the normal operation of the instrument.
centrifugation time recorded in 10.2.3. 13.5 Operating Parameters—Assign the appropriate oper-
10.2.8 Repeat 10.2.2 through 10.2.7 at different desired ating parameters to the instrument task file so that the desired
times, based on the running time being kept on the timer set in elements can be determined. Parameters to be included are
10.1.2. element, wavelength(s), background correction points
NOTE 1—Many cement pastes will begin to harden after three hours, or (optional), inter-element correction factors, and integration
perhaps even sooner, making extraction of the liquid by a centrifuge time. Multiple integrations (typically three) are required for
difficult or impossible. each measurement. A typical measurement time is 10 s.
11. Preparation of Test Solutions and Standards 14. Calibration and Standardization
11.1 Preparation of Test Sample: 14.1 Allow the ICP-OES instrument to warm up for at least
11.1.1 Transfer appropriate amount of a filtered sample into 30 min. Operate the spectrometer according to the operation
a volumetric flask then dilute to volume with 5 % HNO3. manual for the instrument.
Repeat for each filtered pore solution from 10.2. 14.2 Calibrate the instrument for the elements chosen over a
suitable concentration range by atomizing the calibration blank
NOTE 2—Appropriately dilute samples to within the calibration range of
each chosen element. and mixed calibration standard solutions, including any inter-
nal standards, and recording their concentrations and signal
11.2 Preparation of Calibration Standard: intensities. A minimum of six standards and a blank should be
11.2.1 Blank—Prepare a blank solution by adding same used for calibration with one of the standards. The computer
concentration of diluted acids as used in the sample dissolution software will establish the slope, intercept and correlation
(usually 5 % nitric acid). coefficients for each element. Use manufacture suggested
11.2.2 Calibration Standards—repare known concentrations wavelengths each element. The calibration correlation coeffi-
of single-element or multi- element calibration reference solu- cient shall be equal to or greater than 0.990. It is recommended
tions by combining appropriate volumes of the stock solutions that the calibration blank and standards be matrix matched with
in acid-rinsed volumetric flasks. Add sufficient 5 % nitric acid the same acid concentration contained in the samples. Analyze
to dilute to volume. Take care when preparing the multi- appropriate reference solutions to validate the calibration of the
element solutions to verify that the components are compatible instrument before proceeding to the sample analysis.
and stable (they do not interact to cause precipitation) and that 14.3 Alternatively, calibrate according to the manufacturer’s
none of the elements present exhibit mutual spectral interfer- instructions if equivalent results are achieved.
ence. Calibration reference solutions must be verified initially
using a quality control sample and monitored periodically for 15. Analysis by Inductively Coupled Plasma Optical
stability. Emission Spectroscopy
11.3 Preparation of Check Standard: 15.1 Aspirate the prepared samples into the calibrated
11.3.1 Prepare the same way as the calibration standards instrument using the same conditions established for the
11.2.2 specifically using NIST traceable certified reference calibration procedure. Rinse sufficiently to prevent carryover
solutions. between samples.
--`,`,````,``,,,,`,,`,`,``,,,,,`-`-`,,`,,`,`,,`---
16.1 Confirm the performance of the instrument and cali- 18. Report
bration curve by analyzing a quality control sample. 18.1 Data to report shall include the following:
16.2 Internal Standard: 18.1.1 All sample and standard information including
16.2.1 Select the appropriate internal standard element(s) source details;
for the capabilities (mass availability, sensitivity) of the ICP- 18.1.2 Sample analysis results;
OES instrument available. Yttrium is often used as an internal 18.1.3 Quality assurance and quality control data;
standard for this purpose. The element(s) chosen for the 18.1.4 Identity of laboratory analysts;
internal standard should not be a component of the test 18.1.5 Information on instrumentation operation and equip-
specimen or calibration standard. Add internal standards to ment used;
blanks, samples, and standards in a like manner. 18.1.6 Any other information that may help identify the
specimen preparation conditions or measurement conditions.
16.3 Laboratory Check Standard (LCS):
16.3.1 To ensure that the test method is in control, analyze 19. Precision and Bias
a certified NIST traceable standard solution every few samples
or as established to be necessary for the instrument. The 19.1 This practice is provided for users to develop their own
standard added to 5 % HNO3 (v/v) should be appropriate for precision and bias statements when using this method.
the type of samples analyzed and should produce results of
sufficient precision to ensure meaningful assessment of accu- 20. Keywords
racy. The LCS must be taken through all the steps of the 20.1 atomic emission; cementitious; chemical analysis; hy-
analytical method including sample preservation and pretreat- dration; ICP; inductively coupled plasma; optical emission;
ment. pore solution; spectroscopy
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/