You are on page 1of 4

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles

for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

Designation: C1875 − 18

Standard Practice for


Determination of Major and Minor Elements in Aqueous
Pore Solutions of Cementitious Pastes by Inductively
Coupled Plasma Optical Emission Spectroscopy (ICP-OES)1
This standard is issued under the fixed designation C1875; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.

1. Scope Middle Distillate Fuels by Inductively Coupled Plasma


1.1 This practice describes a procedure for collection, Atomic Emission Spectrometry (ICP-AES)
sample preparation and analysis of aqueous pore solutions E1479 Practice for Describing and Specifying Inductively
obtained from cementitious materials at different hydration Coupled Plasma Atomic Emission Spectrometers
times when analyzed by ICP-OES for the six most common
readily soluble elements aluminum, calcium, potassium, 3. Terminology
silicon, sodium and sulfur. 3.1 Definitions:
1.2 The values stated in SI units are to be regarded as 3.1.1 Unless otherwise specified, for definitions of terms
standard. No other units of measurement are included in this used in this practice refer to Terminology C1738/C1738M.
standard.
1.3 This standard does not purport to address all of the 4. Summary of Practice
safety concerns, if any, associated with its use. It is the 4.1 This document describes practice for determining the
responsibility of the user of this standard to establish appro- concentration of multiple dissolved components in the pore
priate safety, health, and environmental practices and deter- solution of cementitious pastes as a function of time using
mine the applicability of regulatory limitations prior to use. inductively coupled plasma optical emission spectroscopy
1.4 This international standard was developed in accor- (ICP-OES). A cementitious material is mixed with water at a
dance with internationally recognized principles on standard- desired water to solid ratio using a Practice D1129 certified
ization established in the Decision on Principles for the constant mixer. Aliquots of the mix are divided into centrifuge
Development of International Standards, Guides and Recom- tubes and stored in a water bath or environmental chamber at
mendations issued by the World Trade Organization Technical constant temperature. The samples centrifuged at different
Barriers to Trade (TBT) Committee. times. The extracted pore solution is filtered and diluted with a
5 % nitric acid solvent. The prepared test solutions flow into
2. Referenced Documents the plasma torch of the ICP instrument where it is vaporized
2.1 ASTM Standards: and ionized, producing characteristic atomic wavelength emis-
--`,`,````,``,,,,`,,`,`,``,,,,,`-`-`,,`,,`,`,,`---

C25 Test Methods for Chemical Analysis of Limestone, sion spectra. The spectra are dispersed by a high resolution
Quicklime, and Hydrated Lime grating and the intensities of the individual wavelengths are
C114 Test Methods for Chemical Analysis of Hydraulic measured. Elemental concentrations are determined by relating
Cement the emission intensities at specific wavelengths for an unknown
C1738/C1738M Practice for High-Shear Mixing of Hydrau- sample to analytical curves made from reference standards of
lic Cement Pastes known composition. The concentration of the elements in the
D1129 Terminology Relating to Water solutions can then be calculated. Internal standards compensate
D1193 Specification for Reagent Water for variations in test sample flow rates in the instrument.
D5673 Test Method for Elements in Water by Inductively
Coupled Plasma—Mass Spectrometry 5. Significance and Use
D7111 Test Method for Determination of Trace Elements in 5.1 The chemical composition of the liquid in cementitious
pastes is an important indicator of the solid component
1
reactivity at early times, being influenced by the content and
This test method is under the jurisdiction of ASTM Committee C01 on Cement
rate of reaction of readily soluble alkali components, lime, and
and is the direct responsibility of Subcommittee C01.23 on Compositional Analysis.
Current edition approved July 1, 2018. Published August 2018. DOI: 10.1520/ other soluble phases. Monitoring the solution composition with
C1875-18 time can provide valuable diagnostic information about cement

Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States

Copyright ASTM International


Provided by S&P Global under license with ASTM
1Licensee=PT Pertamina (Persero)/5987886001, User=Rullyanti, Novie
No reproduction or networking permitted without license from S&P Global Not for Resale, 02/07/2023 01:07:12 MST
C1875 − 18
quality and reactivity to supplement other sources of charac- 7.6 Calibration Blank Solution—5 % (v/v) HNO3.
terization data. This practice is intended to aid in the interpre-
7.7 Nitric Acid—5 % (v ⁄v).
tation of the concentrations of readily soluble components in
cement paste solutions, which may include portland cement, 7.8 Check Standards—Quality control certified NIST trace-
limestone, fly ash, ground granulated blast furnace slag, or able samples that are stable and representative of the samples
other components. It provides guidance for measuring the time of interest. These check standard samples represent ground
dependence of the concentrations of one or more components, truth samples and are used to check that the measurement
on an elemental basis, including, but not limited to, aluminum, process is in control.
calcium, potassium, silicon, sodium, and sulfur. 7.9 Internal Standards—Internal standards are used to cor-
6. Apparatus rect for instrument drift and physical interferences. Refer to
instrument operating manual for suggested internal standards.
6.1 Spectrometer—Spectrometer-An inductively coupled Add internal standards to blanks, samples, and standards in a
plasma emission spectrometer of the simultaneous or sequen- like manner.
tial type including RF generator, torch, nebulizer, connector
tube, spray chamber, recommended peristaltic pump and host 7.10 Membrane Filter—Nylon 25 mm diameter, 0.2 µm.
computer. Sample uptake is done by self-aspiration, or with a 7.11 Syringe and Needle—Capacity of at least 2 mL.
peristaltic pump.
7.12 Centrifuge Tubes—Clean plastic centrifuge tubes the
6.2 Mass Flow Controllers—A mass-flow controller to regu- appropriate size for centrifuge.
late the nebulizer gas may be used as recommended by the
instrument manufacturer. 7.13 Glassware—Volumetric flasks complying with the re-
quirements of ISO 1042, made of borosilicate glass complying
6.3 Balance—Top loading or analytical, with automatic tare, with the requirements of ISO 3585. Glassware should be
with capacity of at least 500 g and a sensitivity of at least 0.1 cleaned before each use by soaking in nitric acid and then
g to weigh paste mix components. rinsing thoroughly with deionized water.
6.4 Centrifuge—Use a centrifuge capable of accommodat-
7.14 Pipette—Calibrated volumetric pipettes and pipette
ing two or more 50 mL centrifuge tubes at rotational speeds up
tips, for dispensing.
to 5000 min-1.
6.5 Timer—Calibrated timer for use of tracking blending 8. Hazards
and reaction times.
8.1 The toxicity or carcinogenicity of each reagent used in
7. Reagents and Materials this practice has not been precisely defined; however, each
7.1 Ordinary laboratory apparatus are not listed, but are chemical should be treated as a potential health hazard.
assumed to be present. Adequate precautions should be taken to minimize exposure of
personnel to chemicals used in this practice.
7.2 Purity of Reagents—At a minimum, reagent grade or
better chemicals shall be used in all tests. 8.2 Gases under high pressure are used in this practice. Use
only apparatus rated for handling the high gas pressures that
7.3 Purity of Water—References to water shall be under- occur in this practice.
stood to mean deionized water. Tap water may be used only to
prepare cement paste mixes if desired. If tap water is used for 9.1 Powder Homogenization—A uniform distribution of
mixing the composition of the water itself must be separately powder components is essential to ensure representative sam-

--`,`,````,``,,,,`,,`,`,``,,,,,`-`-`,,`,,`,`,,`---
measured as a test sample as described in Section 11. pling during the mixing process, especially for blended powder
components such as portland cement and fly ash. Powders
7.4 Argon—High purity grade (99.99 %).
should therefore be homogenized for 30 minutes with a
7.5 Calibration Standards—Single-element or multi- planetary mixer or similar device.
element calibration reference solutions that combine appropri-
ate volumes of the stock solutions or weighed reagent-grade 10. Sample Preparation and Collection
chemicals in acid-rinsed volumetric flasks. Dilute to the
volume of the flask using a 5 % (v/v) nitric acid solution. Prior 10.1 High Shear Mixing:
to preparing the multi-element solutions, analyze each stock 10.1.1 Blend the solid and liquid components to make a
solution separately to check for strong spectral interference and paste according the standard practice described in Practice
the presence of impurities. Take care when preparing the D1129.
multi-element solutions to verify that the components are 10.1.2 Set a timer to mark the end of mixing. The time on
compatible and stable (i.e., that they do not interact to cause this timer will serve as the total reaction time for the remainder
precipitation) and that none of the elements present will of the procedure.
contribute to mutual spectral interference. Calibration refer- 10.1.3 Stir the paste with the spatula for no more than 15 s
ence solutions must be verified initially using a quality control ensure homogeneity.
sample and monitored periodically for stability. Mixed calibra- 10.1.4 Transfer (25 6 1) g of the paste into each of twelve
tion standards will vary, depending on the number of elements centrifuge tubes with secured lids in the minimum amount of
being determined. time possible, preferably less than 2 min.

Copyright ASTM International


Provided by S&P Global under license with ASTM
2Licensee=PT Pertamina (Persero)/5987886001, User=Rullyanti, Novie
No reproduction or networking permitted without license from S&P Global Not for Resale, 02/07/2023 01:07:12 MST
C1875 − 18
10.1.5 Place the twelve centrifuge sample tubes upright in 12. Interfaces
sample holders and place them in the temperature control 12.1 Refer to ICP-OES operating manual for instructions on
device (water bath or environmental chamber). minimizing and checking for interferences.
10.2 Liquid Solution Extraction:
10.2.1 Refer to manufacturerâs manual for instructions 13. Preparation of ICP-OES Apparatus
on safe and optimal operation of centrifuge. 13.1 Prepare and operate the spectrometer in accordance
10.2.2 Set centrifuge parameters to 5000 rpm for approxi- with the manufacturerâs instructions. The present practice
mately 1 min, and load one centrifuge tube filled with (25 6 1) assumes that good operating procedures are followed. Design
g of solid to serve as a counterweight. differences between spectrometers make it impractical to
10.2.3 Within the first 10 min on the timer, select one of the specify the required steps in detail.
samples and place it into the centrifuge opposite the preloaded 13.2 Peristaltic Pump—If a peristaltic pump is used, inspect
counterweight. Begin centrifuging. Record the time indicated the pump tubing and replace it, if necessary, before starting
on the timer when the centrifuge started. each day. Verify the solution uptake rate and adjust it to the
10.2.4 Immediately after the centrifuge comes to a complete desired rate.
stop, remove sample centrifuge tube. 13.3 ICP Excitation Source—Initiate the plasma source at
10.2.5 Using a needle tip syringe, extract pore solution from least 30 min before performing an analysis. During this warm
sample without disturbing solid material in tube. up period, nebulize either water or a dilute (5 %) acid solution.
10.2.6 Invert syringe and cautiously remove needle tip and NOTE 3—Some instrument manufacturers recommend even longer
add a 0.20 µm nylon membrane filter at the syringe opening. warm-up periods to minimize changes in the slopes of calibration curves.

--`,`,````,``,,,,`,,`,`,``,,,,,`-`-`,,`,,`,`,,`---
10.2.7 Transfer pore solution from syringe through filter 13.4 Wavelength Profiling—Perform any wavelength profil-
into a clean empty plastic sample container. Label and note ing that is specified in the normal operation of the instrument.
centrifugation time recorded in 10.2.3. 13.5 Operating Parameters—Assign the appropriate oper-
10.2.8 Repeat 10.2.2 through 10.2.7 at different desired ating parameters to the instrument task file so that the desired
times, based on the running time being kept on the timer set in elements can be determined. Parameters to be included are
10.1.2. element, wavelength(s), background correction points
NOTE 1—Many cement pastes will begin to harden after three hours, or (optional), inter-element correction factors, and integration
perhaps even sooner, making extraction of the liquid by a centrifuge time. Multiple integrations (typically three) are required for
difficult or impossible. each measurement. A typical measurement time is 10 s.

11. Preparation of Test Solutions and Standards 14. Calibration and Standardization
11.1 Preparation of Test Sample: 14.1 Allow the ICP-OES instrument to warm up for at least
11.1.1 Transfer appropriate amount of a filtered sample into 30 min. Operate the spectrometer according to the operation
a volumetric flask then dilute to volume with 5 % HNO3. manual for the instrument.
Repeat for each filtered pore solution from 10.2. 14.2 Calibrate the instrument for the elements chosen over a
suitable concentration range by atomizing the calibration blank
NOTE 2—Appropriately dilute samples to within the calibration range of
each chosen element. and mixed calibration standard solutions, including any inter-
nal standards, and recording their concentrations and signal
11.2 Preparation of Calibration Standard: intensities. A minimum of six standards and a blank should be
11.2.1 Blank—Prepare a blank solution by adding same used for calibration with one of the standards. The computer
concentration of diluted acids as used in the sample dissolution software will establish the slope, intercept and correlation
(usually 5 % nitric acid). coefficients for each element. Use manufacture suggested
11.2.2 Calibration Standards—repare known concentrations wavelengths each element. The calibration correlation coeffi-
of single-element or multi- element calibration reference solu- cient shall be equal to or greater than 0.990. It is recommended
tions by combining appropriate volumes of the stock solutions that the calibration blank and standards be matrix matched with
in acid-rinsed volumetric flasks. Add sufficient 5 % nitric acid the same acid concentration contained in the samples. Analyze
to dilute to volume. Take care when preparing the multi- appropriate reference solutions to validate the calibration of the
element solutions to verify that the components are compatible instrument before proceeding to the sample analysis.
and stable (they do not interact to cause precipitation) and that 14.3 Alternatively, calibrate according to the manufacturer’s
none of the elements present exhibit mutual spectral interfer- instructions if equivalent results are achieved.
ence. Calibration reference solutions must be verified initially
using a quality control sample and monitored periodically for 15. Analysis by Inductively Coupled Plasma Optical
stability. Emission Spectroscopy
11.3 Preparation of Check Standard: 15.1 Aspirate the prepared samples into the calibrated
11.3.1 Prepare the same way as the calibration standards instrument using the same conditions established for the
11.2.2 specifically using NIST traceable certified reference calibration procedure. Rinse sufficiently to prevent carryover
solutions. between samples.

Copyright ASTM International


Provided by S&P Global under license with ASTM
3Licensee=PT Pertamina (Persero)/5987886001, User=Rullyanti, Novie
No reproduction or networking permitted without license from S&P Global Not for Resale, 02/07/2023 01:07:12 MST
C1875 − 18
15.2 Run a blank and a check standard (certified NIST 16.3.2 The result obtained for the LCS shall fall within the
traceable solution) every few samples or as established to be limit of 610 % of the expected value, or meets the desired
necessary for the instrument. Analyze a blank and check limits. If the result is not within these limits, reporting of
standard at the end of each run. The concentration shall be results should be halted until the problem causing the deviation
within 610 % of the expected value and the internal standard is identified and resolved. An indication of the occurrence
should fall within established limits of deviation. If the should accompany the reported results.
concentration is out of range, correct the problem, recalibrate
the instrument and rerun the samples in question. 17. Calculations
15.3 Perform the corrections and calculations, including 17.1 To calculate the concentration of each element in the
dilution factors, using the instrument host computer. original undiluted solution, multiply the measured concentra-
tion of that element by the dilution factor.
16. Quality Control

--`,`,````,``,,,,`,,`,`,``,,,,,`-`-`,,`,,`,`,,`---
16.1 Confirm the performance of the instrument and cali- 18. Report
bration curve by analyzing a quality control sample. 18.1 Data to report shall include the following:
16.2 Internal Standard: 18.1.1 All sample and standard information including
16.2.1 Select the appropriate internal standard element(s) source details;
for the capabilities (mass availability, sensitivity) of the ICP- 18.1.2 Sample analysis results;
OES instrument available. Yttrium is often used as an internal 18.1.3 Quality assurance and quality control data;
standard for this purpose. The element(s) chosen for the 18.1.4 Identity of laboratory analysts;
internal standard should not be a component of the test 18.1.5 Information on instrumentation operation and equip-
specimen or calibration standard. Add internal standards to ment used;
blanks, samples, and standards in a like manner. 18.1.6 Any other information that may help identify the
specimen preparation conditions or measurement conditions.
16.3 Laboratory Check Standard (LCS):
16.3.1 To ensure that the test method is in control, analyze 19. Precision and Bias
a certified NIST traceable standard solution every few samples
or as established to be necessary for the instrument. The 19.1 This practice is provided for users to develop their own
standard added to 5 % HNO3 (v/v) should be appropriate for precision and bias statements when using this method.
the type of samples analyzed and should produce results of
sufficient precision to ensure meaningful assessment of accu- 20. Keywords
racy. The LCS must be taken through all the steps of the 20.1 atomic emission; cementitious; chemical analysis; hy-
analytical method including sample preservation and pretreat- dration; ICP; inductively coupled plasma; optical emission;
ment. pore solution; spectroscopy

ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/

Copyright ASTM International


Provided by S&P Global under license with ASTM
4Licensee=PT Pertamina (Persero)/5987886001, User=Rullyanti, Novie
No reproduction or networking permitted without license from S&P Global Not for Resale, 02/07/2023 01:07:12 MST

You might also like