Professional Documents
Culture Documents
Journal of Chromatography B
journal homepage: www.elsevier.com/locate/chromb
a r t i c l e i n f o a b s t r a c t
Article history: The purpose of this work was to develop and evaluate a fast, robust and specific UPLC–MS/MS screening
Received 1 July 2013 platform for the determination and quantification of a variety of commonly used drugs of abuse in
Accepted 14 December 2013 urine, i.e. a high-throughput quantitative analysis. Substances in the drug classes opioids, central nervous
Available online 25 December 2013
system stimulants and benzodiazepines and related agents were included in addition to cannabis and
pregabalin, a total of 35 different analytes. Based on the concentrations and the physico-chemical proper-
Keywords:
ties of the substances, three UPLC–MS/MS methods were developed in parallel. Prior to analysis, sample
UPLC–MS/MS
preparation consisted of two different simple dilutions with 60 and 100 L urine, respectively, using a
Opioids
Benzodiazepines
Tecan Freedom Evo pipetting robot platform. A Waters Xevo TQ-S tandem quadrupole mass spectrometer
Central nervous system stimulants coupled to a Waters I-class UPLC was used for quantitative analysis of one quantitative and one qualifying
Drugs of abuse MRM transition for each analyte, except for tramadol for which the metabolite O-desmethyl-tramadol
Screening was included in the MRM method to confirm tramadol identity. Deuterated analogs were included as
internal standards. The between-assay relative standard deviations varied from 2% to 11% and the limits
of quantification were in the range 1–200 ng/mL for the various analytes. After development and initial
testing, the method has been successfully implemented and routinely used at our hospital for quantitative
screening of drugs of abuse in more than 35,000 urinary samples.
© 2013 Elsevier B.V. All rights reserved.
1570-0232/$ – see front matter © 2013 Elsevier B.V. All rights reserved.
http://dx.doi.org/10.1016/j.jchromb.2013.12.014
84 S. Hegstad et al. / J. Chromatogr. B 947–948 (2014) 83–95
enzymatic hydrolysis of the urine before analysis [10]. Recently Al- Method 1 Methods 2 and 3
Saffar et al. published an analytical method of 26 new psychoactive
drugs within the group of legal highs, with a 1:5 dilution [11]. 60 µL urine 100 µL urine
Total dilution 1:250 Enzymatic hydrolysis
9 -Tetrahydrocannabinol carboxylic acid (THCA) and Total dilution 1:10
buprenorphine are particularly challenging analytes due to
more than 10-fold–100-fold lower concentrations than most other
drugs and medications in urine. In addition, buprenorphine has an UPLC method 1 UPLC method 2 UPLC method 3
Initial conditions: Initial conditions: (THCA)
unfavorable fragmentation in the ion source [12]. Therefore, these 98% formic acid (0.1 %) 98% formic acid (0.1%) Initial conditions:
drugs are often not included in methods aiming to identify several in water in water 40% formic acid (0.1%)
drugs in parallel. For example, only THCA and not buprenor- 2% methanol 2% acetonitrile in water
60% acetonitrile
phine was included in method presented by Eichhorst et al. [10].
Although direct LC–MS/MS methods of THCA as the only analyte
in urine have been reported [13,14], we have not identified any MS/MS method 1 MS/MS method 2 MS/MS method 3
previous studies describing a method for the detection of THCA ESI + ESI + (THCA) ESI –
and buprenorphine in addition to other common drugs of abuse
Fig. 1. Sample preparation and UPLC–MS/MS conditions.
using simple sample dilution and LC–MS/MS.
Another challenge in the development of an analytical urinary
method is the presence of glucuronide metabolites. Often, this morphine 6-glucuronide (M6G), oxazepam, ritalinic acid and
issue has been approached by applying glucuronide hydrolysis to 9 -tetrahydrocannabinol carboxylic acid (THCA) were from Chi-
the sample, either enzymatically, alkaline or by acid hydrolysis. ron (Trondheim, Norway); fentanyl, hydromorphone, ketamine,
However, as these procedures are often incomplete, variable con- methamphetamine, oxycodone, pethidine, tramadol and zolpi-
centrations of the parent substance and transformation to other dem were from Sigma-Aldrich (Saint Louis, MO); pregabalin
analytes might be the result, with additional uncertainty in the clin- and zopiclone were from Sequoia (Berkshire, UK); desmethyl-
ical interpretation of the results [15–18]. During the last decade, diazepam (DMD) was from Synfine Research (Ontario, Canada);
glucuronide metabolites are more regularly included in the analy- ephedrine was from Norsk Medisinaldepot AS (Oslo, Norway); O-
ses, such as morphine 3-glucuronide and morphine 6-glucuronide desmethyltramadol (O-DM-tramadol) was from Grunenthal GmbH
[19,20]. One advantage by including glucuronides is that the pres- (Aachen, Germany) and methadone was from St. Olav University
ence of them represents additional evidence for intake of the parent Hospital Pharmacy (Trondheim, Norway). The internal standards
substance. On the other hand, glucuronide standards are often 7-AC-d4 , amphetamine-d3 , BECG-d3 , codeine-d3 , methadone-d3 ,
expensive and the supply situation is not always stable. methamphetamine-d5 , MDMA-d5 , 6-MAM-d3 , morphine-d3 ,
The Department of Clinical Pharmacology at St. Olav Univer- M3G-d3 and oxazepam-d5 and were from Lipomed AG (Arlesheim,
sity Hospital analyses about 50,000 urine samples per year. Most Switzerland); 7-AF-d3 , DMD-d5 , ephedrine-d6 , ethylmorphine-d5 ,
of these samples are from opioid addicted individuals treated fentanyl-d5 , hydromorphone-d3 , MDA-d5 , OH-alprazolam-d5 ,
with buprenorphine or methadone in a rehabilitation program. oxycodone-d6 , ritalinic acid-d9 , zolpidem-d6 and zopiclone-d8
The samples are analyzed to confirm the use of buprenorphine were from Chiron (Trondheim, Norway); buprenorphine-d4 was
or methadone and to monitor possible intake of other drugs of from Synfine Research (Ontario, Canada); pregabalin-d4 was from
abuse, either illegally consumed or in some cases prescribed. The C/D/N Isotopes Inc. (Quebec, Canada); tramadol-d6 was from
quantitative results are (after adjustment based on the concentra- Toronto Research Chemical Inc. (Ontario, Canada) and THCA-d3
tion of creatinine in the samples) used to differentiate new intake was from Cerilliant (Round Rock, TX). The enzyme ß-glucuronidase
from residual drug excretion caused by a former intake [21,22]. (Type 2 HP-2 from Helix pomatia, ≥100,000 units/mL) was obtained
Based upon these prerequisites, the aim of the present study was from Sigma-Aldrich (Saint Louis, MO). External quality control
to develop a simple, robust and specific UPLC–MS/MS screening samples (Liquichek Urine Toxicology Control, Level C4) were
platform for the determination and quantification of a variety of purchased from Biorad, (Irvine, CA).
commonly used drugs of abuse in urine, suitable for implementa- LC–MS grade methanol and acetonitrile were purchased from
tion as a high-throughput routine method. Due to the wide range of Merck (Darmstadt, Germany) and formic acid 98%, ammonium
physico-chemical properties of the analytes, the differences in uri- acetate 99% and acetic acid 100% were from VWR (Leuven,
nary concentrations between analytes and the desire of performing Belgium). All water used was provided from a Millipore A10 Synthe-
glucuronide hydrolysis for some analytes but not for others, we sis filtering system (Billerica, MA). 96-well sample collection plates
have developed two methods for sample preparation with three (1 mL) and polypropylene cap mat round wells for 96-well plates
different UPLC–MS/MS methods (Fig. 1). were purchased from Waters (Milford, MA).
Table 1
Analyte and internal standard transition ions and associated mass spectrometric parameters (cone voltage, collision energy) for method 1 (ESI+).
Analyte Transition (m/z) Cone voltage (V) Collision energy (eV) Internal standard
7, the calibrators and quality control (QC) samples were pre- pipetting robot before each analysis. The solutions used for calibra-
pared in three different standard solutions. For tramadol and tor and QC samples were stored at −20 ◦ C. The internal standard
zopiclone the calibrators and QC samples were prepared in blank was diluted with methanol/water (50:50, v/v) and contained
urine, adjusted to pH 8.0 and pH 5.0, respectively. For practi- 150 ng/mL 7-AC-d4 , 75 ng/mL 7-AF-d3 , 50 ng/mL amphetamine-d3 ,
cal purposes ketamine, O-DM-tramadol, pethidine and zolpidem 50 ng/mL BECG-d3 , 100 ng/mL codeine-d3 , 100 ng/mL ephedrine-
were prepared in the same calibrators and QC samples as tra- d6 , 25 ng/mL methadone-d3 , 15 ng/mL methamphetamine-d5 ,
madol. For the remaining compounds in method 1, calibrators 50 ng/mL MDMA-d5 , 150 ng/mL morphine-d3 , 150 ng/mL M3G-
and QC samples were prepared in blank urine (pH 6.2–6.7). The d3 , 150 ng/mL oxycodone-d6 , 2500 ng/mL pregabalin-d4 , 75 ng/mL
three different standards solutions were prepared in four sets of ritalinic acid-d9 , 100 ng/mL tramadol-d6 , 100 ng/mL zolpidem-d6
calibrators and three sets of QC. The three different standards solu- and 100 ng/mL zopiclone-d8. The solution was stored at 4 ◦ C until
tions (20 L of each) were mixed by a Tecan Freedom Evo 200 use.
86 S. Hegstad et al. / J. Chromatogr. B 947–948 (2014) 83–95
Table 2
Analyte and internal standard transition ions and associated mass spectrometric parameters (cone voltage, collision energy and) for method 2 (ESI− for THCA/THCA-d3, ESI+
for all other analytes).
Analyte Transition (m/z) Cone voltage (V) Collision energy (eV) Internal standard
Methods 2 and 3: Analytes included in methods 2 and 3 are pre- 2.4. UPLC conditions
sented in Table 2. All analytes were included in one set of calibrators
and QC samples (buprenorphine, DMD, ethylmorphine, fentanyl, A Waters Acquity UPLC I-Class FTN system (Waters, Mil-
MDA, 6-MAM, OH-alprazolam, oxazepam, PMA, PMMA, and THCA). ford, MA) was used for separation, applying an Acquity HSS
The solutions used for calibrator and QC samples were stored at T3-column (2.1 mm ID × 100 mm, 1.8 m) maintained at 50 ◦ C.
−20 ◦ C. The mobile phases and their gradient profiles are shown in
The internal standard was diluted with methanol/water (50:50, Table 3. The total cycle time for all three methods combined
v/v) and contained 50 ng/mL buprenorphine-d4 , 300 ng/mL DMD- was 7.9 min. The pre-inject and post-inject washes were per-
d5 , 300 ng/mL ethylmorphine-d5 , 20 ng/mL fentanyl-d5 , 300 ng/mL formed with methanol/acetonitrile/isopropanol/water/formic acid
hydromorphone-d3 , 300 ng/mL OH-alprazolam-d5 , 300 ng/mL (25:25:25:24:1, v/v) for 3 s and 12 s, respectively. The injection vol-
MDA-d5 , 150 ng/mL 6-MAM-d3 , 300 ng/mL oxazepam-d5 and ume was 5 L for all three methods. A column filter (in-line filter,
150 ng/mL THCA-d3 . The solution was stored at 4 ◦ C until use. 0.2 m, Waters) was used in the front of the analytical column in
methods 2 and 3.
Table 3
Mobile phases, total cycle times and flow rates used in the UPLC separation of the analytes.
Time (min) 0.1% formic Methanol (%) Time (min) 0.1% formic Acetonitrile (%) Time (min) 0.1% formic Acetonitrile (%)
acid in water acid in water acid in water
(%) (%) (%)
0 98 2 0 98 2 0 40 60
0.5 75 25 0.8 65 35 0.8 2 98
1.9 20 80 1.5 2 98 1.3 2 98
1.91 2 98 1.9 2 98 1.31 40 60
2.4 2 98 1.91 98 2
2.41 98 2 2.0 98 2
2.5 98 2
A/mobile phase B. This was done for each of the analytes individ- 2.6.2. Limits of quantification and detection
ually. The dwells times for the MRM channels were automatically The limit of quantification (LOQ) was determined by diluting
calculated by the software, given the premises that the peaks were standard samples of different concentrations, denoted LOQ sam-
expected to elute over 2–3 s, and a desired number of 20 data points ples. LOQ samples were run in one replicate on ten different days.
determining the peak. This resulted in a MS method cycle time of A standard sample with a concentration identical to the LOQ sample
0.10–0.15 s, and dwell times ranging from 4 to 63 ms. The inter- was included in the calibration curve. The signal-to-noise criteria
channel delay time was set to 3 ms and the inter-scan delay time (using the peak-to-peak algorithm) for the LOQ samples were ≥10
was 3 ms at all times. and precision and bias of calculated concentrations were within
The signal of amphetamine was unexpectedly low at the lowest ±20%. The limit of detection (LOD) was determined by further dilu-
calibration level compared to methamphetamine due to fragmen- tion until the signal-to-noise ratio reached 3 (S/N ≥ 3).
tation of amphetamine in the ion source to the fragment ion m/z 91.
Reducing the ion source fragmentation of amphetamine increased 2.6.3. Precision and bias
the signal of the lowest calibration level. Due to the high sensi- Within-assay precision was estimated by analysis of ten sep-
tivity of the system it was for several compounds necessary to arate replicates of QC samples at three concentrations in a single
detune the analytical signal to obtain linear calibration curves. This assay. Between-assay precision and accuracy were determined by
was done by increasing cone voltage or collision energy, or mak- analysis of one replicate at three QC concentrations on ten different
ing use of a second transition with less intensity or using the C13 days.
isotope as precursor ion of the analyte. MRM transitions, cone volt-
age and collision energy for the analytes and the internal standards
2.6.4. Matrix effects
used in the presented methods are shown in Tables 1 and 2. The
Matrix effects (ME) were evaluated by the method pro-
optimized/detuned MS conditions are presented in the supplemen-
posed by Matuszewski et al. [25]. The analyte signal in the
tary Tables 1 and 2. System operation and data acquisition were
spiked methanol/water solution (10:90, v/v; method 1 or 60:40,
controlled using Mass Lynx 4.1 software (Waters). All data were
v/v; method 2 and 3) was compared with the analyte sig-
processed with the Target Lynx quantification program (Waters).
nal in the matrix. Six replicates of urine from six different
The analytes were identified by comparing the retention time and
individuals were analyzed. The concentrations corresponded to
the ratio between the two MRM transitions of the corresponding
the lowest QC level of the analytes. Spiked QC samples in
calibrator and QC samples. The retention times deviated less than
methanol/water or urine (methods 2 and 3) were prepared as
±1% from the average ratio, and the maximum deviation was set
described in Section 2.3, with internal standards and analytes
to 20%. In addition, the identity of tramadol was verified by the
added after the hydrolysis step. The ME (in percent) was calculated
presence of O-DM-tramadol.
as ME% = (Peak intensitymatrix /Peak intensitymethanol/water ) × 100.
Supplementary material related to this article can be found,
Relative ME (CV%) expresses the precision of peak intensity
in the online version, at http://dx.doi.org/10.1016/j.jchromb.
in matrix. ME% corrected with internal standard was cal-
2013.12.014.
culated as = [(Peak intensitymatrix /Peak IS intensitymatrix )/(Peak
intensitymethanol/water /Peak IS intensitymethanol/water )] × 100. Rela-
2.6. Method validation tive ME corrected to IS (CV %) expresses the precision of Peak
intensity/Peak IS intensity in matrix.
The validation was done according to guidelines given by Peters
et al. [23] and the Food and Drug Administration [24]. 2.6.5. Specificity
Specificity of the methods was investigated by selecting
drugs often detected in urinary samples at our laboratory with
2.6.1. Calibration curves almost the same MH+ and MH− ions (±2 atomic mass units)
The four-point calibration curves (three replicates of each as the analytes. Substances meeting the molecular weight (MW)
standard) were based on peak area ratios of the analyte relative criteria for the analytes included in method 1 were analyzed
to the internal standard. For amphetamine the peak height was by methods 2 and 3, and vice versa. The drugs/metabolites
used instead of the area due to interfering peaks at the transition tested were alimemazine (MW 298), carbamazepine (MW 236),
m/z 135.9 > 91 in some of the authentic specimens tested during carbamazepine-10, 11-epoxide (MW 252), chlorprothixene (MW
method development. Quadratic calibration curves were used with 316), citalopram (MW 324), clomipramine (MW 315), clon-
linear weighting (1/x). Origin was included for all analytes except azepam (MW 316), clozapine (MW 327), desmethylclomipramine
amphetamine, methamphetamine and methadone, for which ori- (MW 301), desmethyldoxepin (MW 265), desmethylflunitrazepam
gin was excluded. (MW 299), desmethylmianserin (MW 250), diazepam (MW 284),
88 S. Hegstad et al. / J. Chromatogr. B 947–948 (2014) 83–95
duloxetine (MW 297), felbamate (MW 238), flunitrazepam (MW M6G are hydrolyzed to a higher degree [18] we also decided to
313), fluoxetine (MW 309), ketobemidone (MW 247), levome- analyze these metabolites in urine not undergoing hydrolysis. In
promazine (MW 329), meprobamate (MW 218), mianserin (MW contrast to C6G, benzodiazepine glucuronides such as oxazepam
264), midazolam (MW 326), mirtazapine (MW 265), moclobemide glucuronide are efficiently hydrolyzed by H. pomatia [27]. There-
(MW 269), N-desmethylcodeine (MW 285), nortriptyline (MW fore, and due to the fact that of most glucuronidated standards were
263), O-desmethylvenlafaxine (MW 263), olanzapine (MW 312), very expensive and the supply situation could not be guaranteed
paroxetine (MW 329), pentazocine (MW 285), pimozide (MW 462), by the manufacturers, we decided to include buprenorphine, DMD,
pipothiazine (MW 476), promethazine (MW 284), reboxetine (MW ethylmorphine, hydromorphone, OH-alprazolam, oxazepam and
313), sertraline (MW 306), tetrahydrocannabinol (MW 314) and THCA in the method with hydrolysis. The remaining compounds
warfarin (MW 308). The concentrations tested ranged from 30 to were included in method 2 due to limitation of sensitivity with the
15,000 ng/mL for the various analytes. sample dilution of 1/250 used in method 1.
Conditions for the enzymatic hydrolysis of buprenorphine,
2.6.6. Carry-over effects DMD, ethylmorphine, OH-alprazolam, oxazepam and THCA
To evaluate possible carry-over effects by UPLC–MS/MS, were systematically examined in authentic urine specimens.
authentic specimens containing amphetamine (>90,000 ng/mL), Parameters tested were enzyme concentration (10,000, 25,000 and
buprenorphine (>2000 ng/mL), codeine (>10,000 ng/mL), DMD 50,000 U/mL), incubation time (40, 60, 90 and 120 min), pH in buffer
(>3500 ng/mL), ethylmorphine (>2000 ng/mL), methadone (pH 4.3, 4.8 and 5.3) and incubation temperature (60, 65 and 70 ◦ C).
(>90,000 ng/mL), methamphetamine (>100,000 ng/mL), 6-MAM The most satisfactory results were achieved when 100 L of urine
(>500 ng/mL), morphine (>100,000 ng/mL), OH-alprazolam was buffered to pH 4.8, incubated with 2500 U of enzyme at 65 ◦ C
(>2000 ng/mL), oxazepam (>100,000 ng/mL) and THCA for 60 min (data not shown).
(>10,000 ng/mL) were analyzed. A standard with double con- Another objective was to dilute the samples as much as possible
centration of the highest calibrator was used for the other analytes. with a satisfactory signal/noise ratio (>40–50) at the lowest cali-
The specimens or standards were run followed by one diluted bration level. A high sample dilution will decrease ME and extend
matrix blank. column life time. In method 1 the samples were diluted 1/250,
Carry-over (%) was determined by comparing the found concen- while the analytical signals of THCA, 6-MAM and buprenorphine
tration in the blank samples versus the concentration of the lowest in methods 2 and 3 limited the sample dilution to 1/10. For sev-
calibrator. eral analytes the signal of the highest standard was saturated. It
was therefore necessary to decrease the signal by increasing cone
2.6.7. Stability voltage and collision energy, and using either another transition
To evaluate stability, QC samples (1 mL) were prepared in urine with less intensity or the C13 -transition. Supplementary Tables 1
and kept frozen at −20 ◦ C for four months in ScreenMates tubes, and 2 show optimized and detuned parameters/transitions for the
1.4 mL (Thermo Scientific, Hudson, NH). Stability of THCA in urine analytes.
(10 mL) was evaluated after storage at 4 ◦ C for 1 month in Pyrex In method 1, carry-over was seen after injections of
tubes, 11.5 mL (Corning, Lowell, MA). The external quality control highly concentrated samples or standards (∼100,000 ng/mL) for
C4 (145 ng/mL) was used to verify stability of THCA. The stabil- amphetamine, methadone and methamphetamine when a high pH
ity of diluted urine samples stored at 4 ◦ C in the autosampler was mobile phase and a Waters Acquity BEH C18 column was used. By
evaluated by re-injecting QC samples after three days. changing the mobile phase to 0.1% formic acid in water, carry-over
in the first blank was reduced to <20% of calibrator 1 response. In
2.6.8. Application of methods addition, to obtain satisfactory chromatography for the most polar
In order to identify possible methodological problems such as compounds, the column was changed to a Waters Acquity HSS T3-
false positive/negative results, variation of internal signal, stability column. Further experiments confirmed that the carry-over was
of retention time, disturbing peaks, approximately 600 consecutive found to be on-column and not in the autosampler. Thus, we con-
routine samples were analyzed with the new screening methods cluded that formic acid minimized the interactions between basic
and compared to the results found with our previous routine LC–MS drugs, being positively charged, and the column used.
screening methods [26]. In the analysis of the remaining substances, carry-over was
shown for THCA, and another approach had to be used to reduce
3. Results and discussion this effect. Starting the UPLC gradient at 2% acetonitrile gave a carry-
over in the first blank equal to 200% of calibrator 1 after injection
3.1. Method development of a high THCA standard (10,000 ng/mL). By increasing the initial
acetonitrile concentration from 2% to 60%, carry-over was reduced
Originally, the aim of this work was to develop one method for to 10% of calibrator 1. However, an initial concentration of 60%
all analytes. Due to various considerations and sometimes mutually acetonitrile was not suitable for the remaining compounds, due
exclusive requirements, such as the strength of the analytical signal to high elution strength. The remaining substances were therefore
related to the LOQ for each substance, the width and level of the analyzed with two separate UPLC methods (methods 2 and 3), as
calibration ranges, the desire to hydrolyze glucuronides for some described in Fig. 1. The pre-column filter used in methods 2 and
analytes but not for others, and the availability and cost of reference 3 was replaced after approximately every 600 injections and the
substances, two pre-treatment methods and three UPLC–MS/MS life-time of the analytical column was around 2000 injections.
methods were developed (Fig. 1).
The yield of enzymatic glucuronide hydrolysis is depend- 3.2. Method validation
ing on various parameters like the chemistry of the specific
glucuronidated metabolite, the enzyme strain used, incubation 3.2.1. Calibration curves
temperature, time and pH. As hydrolysis of C6G by H. pomatia is Calibration curves were made for each analyte in the concen-
low (22% at 60 ◦ C and 4 h [18]), measuring glucuronidated codeine tration ranges listed in Tables 4 and 5. As analytes with high
instead of hydrolyzing the sample will enhance the reproducibility as well as with low concentrations were included in the same
and extend the time of detection, thereby facilitating the inter- assay, not all compounds had linear calibration curves. A weighted
pretation of which opiate has been ingested. Although M3G and (1/x) second order regression line was therefore applied for each
S. Hegstad et al. / J. Chromatogr. B 947–948 (2014) 83–95 89
Table 4
Calibration concentrations, correlation coefficient (R2 ), limit of detection (LOD), limit of quantification (LOQ), within-assay precision, between-assay precision and bias for
analytes included in method 1.
Analyte Calibration R2 (n = 4) LOD (ng/mL) LOQ (ng/mL) Theoretical Within-assay Between-assay Between-day
concentrations concentration precision, precision, bias (%)
(ng/mL) (ng/mL) (%CV) (n = 10) (%CV) (n = 10)
7-AC 100, 250, 500, 2000 0.9963 6 20 150 6.0 7.6 4.9
500 5.4 6.3 3.7
1600 4.1 6.0 1.9
7-AN 100, 250, 500, 2000 0.9948 6 20 150 6.4 7.9 3.9
500 5.5 6.4 0.8
1600 4.6 7.2 3.0
Amphetamine 100, 500, 2000, 10,000 0.9982 6 50 150 3.2 3.2 8.2
1500 2.2 4.6 5.3
8000 5.2 5.9 0.7
BECG 100, 500, 2000, 10,000 0.9995 1 10 150 3.1 5.1 9.5
1500 4.5 6.8 3.4
8000 4.7 5.7 −0.4
Codeine 100, 500, 2000, 10,000 0.9978 12 20 150 5.5 5.4 8.9
1500 4.6 6.3 6.0
8000 6.0 6.1 3.1
C6G 500, 1000, 2000, 10,000 0.9975 8 50 750 4.1 4.3 7.6
1500 4.6 4.4 9.9
8000 4.3 5.6 2.9
Ephedrine 100, 500, 1000, 2000 0.9985 6 20 150 3.9 4.6 11.0
300 3.3 4.9 7.3
1600 4.1 4.7 0.1
Ketamine 100, 500, 2000, 10,000 0.9981 6 20 150 6.5 5.6 9.0
1500 3.0 6.8 5.4
8000 3.0 7.1 1.0
Methadone 200, 500, 2000, 10,000 0.9991 25 40 300 2.5 8.3 8.2
1500 2.1 4.7 5.1
8000 4.4 4.0 1.4
Methamphetamine 100, 500, 2000, 10,000 0.9992 3 10 150 3.6 3.6 11.4
1500 3.7 4.3 7.1
8000 5.7 3.5 2.7
MDMA 100, 250, 500, 1000 0.9985 3 10 150 3.1 4.4 9.4
300 2.7 3.3 5.7
800 3.8 3.7 1.1
Morphine 100, 500, 2000, 10,000 0.9992 6 20 150 4.5 6.2 14.0
1500 4.8 6.1 5.3
8000 5.8 7.4 2.7
M3G 500, 2000, 10,000, 50,000 0.9985 8 50 750 3.6 2.3 4.9
8000 3.3 4.5 3.2
40,000 4.8 4.5 0.6
M6G 100, 500, 2000, 10,000 0.9986 25 50 150 3.9 2.6 4.4
1500 3.3 3.4 3.3
8000 4.5 4.6 −1.4
O-DM-tramadol 100, 250, 500, 1000 0.9846 1 10 150 3.2 4.9 8.0
300 3.2 3.6 7.8
800 2.0 2.5 2.6
Oxycodone 100, 500, 2000, 10,000 0.9986 3 20 150 4.6 3.8 8.1
1500 4.8 7.1 5.2
8000 4.0 4.6 2.5
Pethidine 100, 500, 2000, 10,000 0.9982 6 20 150 7.1 7.7 6.8
1500 2.8 7.6 5.6
8000 3.7 7.9 2.3
Pregabalin 500, 1000, 5000, 10,000 0.9987 30 100 750 3.3 5.7 7.8
3000 2.5 4.7 5.0
8000 4.2 4.1 2.5
Ritalinic acid 1000, 5000, 20,000, 50,000 0.9994 60 200 1500 4.4 8.1 6.7
15,000 5.0 5.6 0.02
40,000 4.8 6.5 0.2
90 S. Hegstad et al. / J. Chromatogr. B 947–948 (2014) 83–95
Table 4 (Continued)
Analyte Calibration R2 (n = 4) LOD (ng/mL) LOQ (ng/mL) Theoretical Within-assay Between-assay Between-day
concentrations concentration precision, precision, bias (%)
(ng/mL) (ng/mL) (%CV) (n = 10) (%CV) (n = 10)
Tramadol 100, 500, 2000, 10,000 0.9979 3 10 150 3.9 3.8 5.5
1500 3.9 4.7 2.6
8000 3.2 3.3 3.1
Zolpidem 100, 500, 2000, 10,000 0.9981 6 20 150 6.2 6.6 9.1
1500 3.9 7.1 5.0
8000 3.8 5.4 1.0
Zopiclone 100, 500, 2000, 10,000 0.9991 3 50 150 4.1 3.4 10.6
1500 3.2 3.2 6.2
8000 3.6 3.2 5.6
compound to better fit the observed data points. The methods negligible. Consequently, for these analytes, the intercepts with
were initially developed with inclusion of the origin in the regres- the y axis were almost identical irrespective of whether the
sion line for the calibration curves. Due to response in blank point of origin was included or not. We therefore in these cases
matrix, exclusion of the origin caused a slightly higher inter- chose to keep the origin in the regression line for the calibra-
cept of the calibration curve with the y axis for amphetamine, tion curve. The correlation coefficients for the regression line
methamphetamine and methadone, as compared to the inclusion were above 0.99 for all the compounds (Tables 4 and 5). The
of the origin. Therefore, origin was excluded for these sub- MRM chromatograms of the lowest calibration level are shown in
stances. For all other compounds, background responses were Figs. 2 and 3.
Table 5
Calibration concentrations, correlation coefficient (R2 ), limit of detection (LOD), limit of quantification (LOQ), within-assay precision, between-assay precision and bias for
analytes included in methods 2 and 3.
Analyte Calibration R2 (n = 4) LOD (ng/mL) LOQ (ng/mL) Theoretical Within-assay Between-assay Between-day
concentrations concentration precision, precision, bias (%)
(ng/mL) (ng/mL) (%CV) (n = 10) (%CV) (n = 10)
Buprenorphine 1, 10, 100, 1000 1.000 0.3 1 1.5 3.2 4.9 2.3
150 5.4 4.8 0.2
800 4.6 3.1 −0.9
Ethylmorphine 100, 500, 1000, 2000 0.999 2 10 150 1.9 4.7 1.3
300 4.4 6.1 −0.1
1600 4.4 3.1 −3.2
Fentanyl 10, 50, 200, 1000 1.000 0.6 2 15 2.1 4.4 2.0
200 3.5 3.7 −0.6
800 4.8 2.4 −0.5
Hydromorphone 100, 500, 2000, 10000 1.000 15 50 150 1.7 3.3 1.3
1500 2.6 3.7 −0.3
8000 4.8 1.5 0.4
MDA 100, 250, 500, 1000 0.999 3 20 150 2.2 4.4 0.7
300 3.7 4.4 −0.6
800 5.6 3.4 −0.5
Oxazepam 100, 500, 2000, 10,000 1.000 3 20 150 2.6 4.6 2.1
1500 3.5 5.4 −2.0
8000 6.2 3.6 3.2
PMA 100, 250, 500, 1000 0.998 6 50 150 2.6 8.6 −8.6
300 5.0 8.2 −10.8
800 6.9 5.4 −3.1
PMMA 100, 250, 500, 1000 0.999 2 10 150 3.0 11.0 −2.7
300 4.7 8.2 −6.3
800 5.6 4.0 0.1
Fig. 2. Method 1; MRM chromatograms of the analytes with quantitative transitions from the lowest concentration calibrator.
92 S. Hegstad et al. / J. Chromatogr. B 947–948 (2014) 83–95
Fig. 3. Methods 2 and 3; MRM chromatograms of the analytes with quantitative transitions from the lowest concentration calibrator.
3.2.2. Limit of detection and quantification, precision and bias and a retention time closer to PMA and PMMA, ethylmorphine-d5
The LOD, LOQ, within-assay precision, between-assay precision was preferred instead of MDA-d5 . Obviously, the use of deuterated
and bias for the analytes are presented in Tables 4 and 5. The pre- PMA and PMMA as internal standards could have reduced the ME.
cision was <15% and bias was ±15% at LOQ for all analytes. The Since our method is used for screening, only, the observed matrix
within-assay CVs were 2–8%, the between-assay CVs were 2–11%, effects were nevertheless considered acceptable.
and bias of the QC samples varied between −11% and +14%. Accord-
ing to the Bioanalytical method validation guidelines published
by the U.S. Food and Drug Administration [24], precision and bias 3.2.4. Specificity
below 15% (20% at the LOQ level) are acceptable. There were no Hydromorphone, morphine and N-desmethylcodeine were sep-
indications that precision and bias were affected for the compounds arated chromatographically using both methods. In addition,
included when using detuned MS parameters. In fact, we have also O-desmethylvenlafaxine and tramadol were baseline separated in
previously demonstrated high reproducibility and robustness for method 1. Desmethylclomipramine and desmethylflunitrazepam
oxazepam and 3-hydroxy-diazepam using detuned MS parameters gave an interfering peak for codeine. This was not investigated
in the routine setting [28]. any further, since these metabolites are excreted in urine at low
concentrations [29,30].
Table 6
Matrix effect (ME) for analytes included in method 1.
Analyte Theoretical concentration (ng/mL) ME (%) Relative ME (% CV) ME Corrected with ISa (%) Relative ME corrected with IS (% CV)
sample has been analyzed directly after a highly concentrated sam- analytes were found to be stable in the autosampler for three days
ple (>10,000 ng/mL) of the same analyte. at 4 ◦ C.
Table 7
Matrix effect (ME) for analytes included in methods 2 and 3.
Analyte Theoretical concentration (ng/mL) ME (%) Relative ME (% CV) ME Corrected with ISa (%) Relative ME corrected with IS (% CV)
Buprenorphine 1.5 60 21 93 7
DMD 75 88 8 100 5
Ethylmorphine 150 70 16 90 8
Fentanyl 15 76 10 94 6
Hydromorphone 150 68 8 117 6
MDA 150 50 25 94 8
6-MAM 10 95 14 115 10
OH-alprazolam 75 100 4 100 5
Oxazepam 150 89 6 93 4
PMA 150 52 20 67 12
PMMA 150 46 22 59 14
THCA 15 94 15 101 14
a
IS = internal standard.
94 S. Hegstad et al. / J. Chromatogr. B 947–948 (2014) 83–95
1. The most obvious reason for this finding is the difference in pre-column. For method 1 the pre-column and analytical column
sample dilution between the methods. However, by using another are changed simultaneously after approx. 10,000 injections. The
LC-gradient and negative ESI mode, the internal standard of THCA lifetime of the column used for methods 2 and 3 was significantly
was not affected in the same manner as in method 2. Major ion sup- increased by using a pre-column (up to 10,000 injections). How-
pression reduces sensitivity of the method and may lead to false ever, the pre-column must be replaced after 400–600 injections
negative results. A minimum recovery level of signal of internal to prevent peak tailing. Carry-over of buprenorphine (in the con-
standards was therefore set to 20%. centration range 1–3 ng/mL) was gradually observed in samples
In a few samples (<1% of all samples positive for amphetamine), analyzed after specimens containing > 1000 ng/mL buprenorphine.
an interfering peak was observed in the MRM chromatogram for These samples are routinely reanalyzed in order to avoid false pos-
amphetamine. In order to investigate this issue further, a few urine itive test results of buprenorphine. The methods were accredited
samples were analyzed by accurate mass MS/MS using a Xevo- according to ISO 17,025 by the Norwegian Accreditation Services
G2-S Q-TOF (Waters). The disturbing compound was identified (http://www.akkreditert.no/en/) in April 2013.
to origin from gabapentin and its ion-source fragmentation (data
not shown). To separate amphetamine and gabapentin chromato- 4. Conclusions
graphically, an alternative UPLC method involving a Waters Acquity
BEH C18 column, and ammonium formate pH 10.1/methanol as In conclusion, the presented methods have facilitated screening
mobile phases can be used if needed. of a large number of drugs of abuse with simple sample prepa-
In addition, a few false positive PMA results at low concen- ration, good sensitivity and specificity, and short run-times. The
trations (100–300 ng/mL) were observed in some samples. There methods have been routinely used at our hospital for eight months,
was no presence of PMMA in these samples whereas retention analyzing more than 35,000 samples. The implementation of the
time and ion ratio of PMA corresponded to those of the calibra- methods has eliminated or highly reduced the need of reagents
tors/QCs. We suspected these positive samples to be false because and toxic extraction solvents, significantly improved the efficiency,
PMMA was not present in them, as PMA is usually produced and considerably reduced the manual workload at our routine
from PMMA in the body after intake of PMMA. Due to lack of laboratory.
experience with these compounds we decided to use the same
alternative UPLC method as for amphetamine, to further elucidate
Acknowledgements
this matter. In addition, the cut-off value for reporting PMA as pos-
itive was increased from 100 to 300 ng/mL, thereby solving the
The authors would like to thank Hilde Havnen, Audhild Naavik
issue.
and Lisa Sylten for technical assistance during method develop-
For a number of compounds the urinary concentrations in
ment and validation.
several samples were above the highest calibration level. The
analytes with the highest proportions of samples above the high-
est calibration level were methadone (40%), methamphetamine References
(10%), codeine (10%), amphetamine (6%), oxazepam (6%) and
[1] F.T. Peters, Clin. Biochem. 44 (2011) 54.
morphine (6%). Although the analytical results are used to dis-
[2] H.H. Maurer, Ther. Drug Monit. 32 (2010) 324.
tinguish between a previous and a recent intake based upon a [3] J.J. Pitt, Clin. Biochem. Rev. 30 (2009) 19.
decrease in the (creatinine-adjusted) concentration compared to [4] A.D. de Jager, N.L. Bailey, J. Chromatogr. B: Anal. Technol. Biomed. Life Sci. 879
(2011) 2642.
the previous sample, it is not necessary to quantify concentra-
[5] C. Yuan, C. Heideloff, M. Kozak, S. Wang, Clin. Chem. Lab. Med. 50 (2012) 95.
tions precisely at these high levels. As far as any two samples [6] E. Gustavsson, M. Andersson, N. Stephanson, O. Beck, J. Mass Spectrom. 42
from the same individual are obtained with at least one or a (2007) 881.
few days’ interval (depending on the specific drug), such high [7] J.C. Van De Steene, W.E. Lambert, J. Am. Soc. Mass Spectrom. 19 (2008) 713.
[8] F.T. Peters, D. Remane, Anal. Bioanal. Chem. 403 (2012) 2155.
concentrations could always be evaluated as a new intake. The [9] J.O. Thorngren, F. Ostervall, M. Garle, J. Mass Spectrom. 43 (2008) 980.
same practice is described by Berg et al. [34], where only low [10] J.C. Eichhorst, M.L. Etter, N. Rousseaux, D.C. Lehotay, Clin. Biochem. 42 (2009)
concentrations (within the linear range) were used for interpre- 1531.
[11] Y. Al-Saffar, N.N. Stephanson, O. Beck, J. Chromatogr. B: Anal. Technol. Biomed.
tation. Life Sci. 930 (2013) 112.
The work of Eichhorst et al. [10] included a similar number [12] D. Thieme, H. Sachs, M. Thevis, J. Mass Spectrom. 43 (2008) 974.
and types of analytes and the samples were injected once instead [13] N. Stephanson, H. Dahl, A. Helander, O. Beck, Ther. Drug Monit. 24 (2002) 645.
[14] C. Chebbah, O.J. Pozo, K. Deventer, P. Van Eenoo, F.T. Delbeke, Rapid Commun.
of thrice, as we did. However, our methods included morphine Mass Spectrom. 24 (2010) 1133.
and codeine glucuronides in addition to buprenorphine, and were [15] P. Wang, J.A. Stone, K.H. Chen, S.F. Gross, C.A. Haller, A.H. Wu, J. Anal. Toxicol.
validated for both lower (for some compounds) and higher concen- 30 (2006) 570.
[16] W.C. Duer, S. McFarland, J. Anal. Toxicol. 31 (2007) 419.
tration ranges. In contrast to Eichhorst et al. we use two transitions
[17] S. Fu, J. Lewis, H. Wang, J. Keegan, M. Dawson, J. Anal. Toxicol. 34 (2010) 243.
for each compound to confirm identity except tramadol, for which [18] L.P. Hackett, L.J. Dusci, K.F. Ilett, G.M. Chiswell, Ther. Drug Monit. 24 (2002)
the metabolite O-desmethyl-tramadol was included for confirma- 652.
[19] J.O. Svensson, M. Andersson, E. Gustavsson, O. Beck, J. Anal. Toxicol. 31 (2007)
tion. In addition, the within-assay and between-assay CVs were
81.
substantially lower with our methods, indicating a more reliable [20] D.M. Shakleya, R. Dams, R.E. Choo, H. Jones, M.A. Huestis, J. Anal. Toxicol. 34
quantitative determination of the analytes. (2010) 17.
[21] A.A. Westin, M.A. Huestis, K. Aarstad, O. Spigset, J. Anal. Toxicol. 33 (2009) 610.
[22] M.A. Huestis, E.J. Cone, J. Anal. Toxicol. 22 (1998) 445.
3.3. Experience from routine use [23] F.T. Peters, O.H. Drummer, F. Musshoff, Forensic Sci. Int. 165 (2007) 216.
[24] Guidance for industry, Bioanalytical Method Validation, May 2001, http://
The methods have been run daily for eight months and both www.fda.gov/downloads/Drugs/GuidanceComplianceRegulatoryInformation/
Guidances/ucm070107.pdf
the simple pre-treatment of the samples and the robust data [25] B.K. Matuszewski, M.L. Constanzer, C.M. Chavez-Eng, Anal. Chem. 75 (2003)
processing ensure consistent results between many analytes. Two 3019.
separate analytical columns are routinely used; in addition meth- [26] G. Bagoien, G. Morken, K. Zahlsen, T. Aamo, O. Spigset, J. Clin. Psychopharmacol.
29 (2009) 248.
ods 2 and 3 samples are injected twice on the same column. [27] R. Meatherall, J. Anal. Toxicol. 18 (1994) 382.
Due to a daily number of routine samples of about 250, the fil- [28] S. Hegstad, E.L. Oiestad, U. Johansen, A.S. Christophersen, J. Anal. Toxicol. 30
ter in front of the columns for both methods was replaced by a (2006) 31.
S. Hegstad et al. / J. Chromatogr. B 947–948 (2014) 83–95 95
[29] A. Wohlfarth, N. Toepfner, M. Hermanns-Clausen, V. Auwarter, Anal. Bioanal. [32] D.E. Moody, K.M. Monti, A.C. Spanbauer, J. Anal. Toxicol. 23 (1999) 535.
Chem. 400 (2011) 737. [33] M.H. Jamerson, J.J. McCue, K.L. Klette, J. Anal. Toxicol. 29 (2005) 627.
[30] M. Forsman, I. Nystrom, M. Roman, L. Berglund, J. Ahlner, R. Kronstrand, J. Anal. [34] T. Berg, E. Lundanes, A.S. Christophersen, D.H. Strand, J. Chromatogr. B: Anal.
Toxicol. 33 (2009) 491. Technol. Biomed. Life Sci. 877 (2009) 421.
[31] S. Golding Fraga, J. Diaz-Flores Estevez, C. Diaz Romero, Ann. Clin. Lab. Sci. 28
(1998) 160.