10’ Medical GAS ALARM 2 GAS Specification LCD AREA ALARM 2 GAS (0,V) @ Listed to UL1069 (Hospital Signaling and Nurse Call Equipment) and CSA_ €222 No, 2015-12 Sigal Equipment €© Product complies with NFPA 99a is manfactured nan 0-900 and ISO 13485 environment © Alarm shall incorporate a Liquid Crystal Display (LCD! to continuously indicate pressure or vacuum levels, Alarm shall be able to restart after a E reer power loss of 10 seconds © Alarm shall be able to restart after a power loss of 10 seconds, © Alarm shall be adjustable to display NFPA or IS cols astellasgas et tings including: PSI / KpA or inHg / mmHg. 6 Abn coditinssal be Went gabe lamsand vl LED Visual LED ulctrs shall be provi foreach service pee fx NORMAL rel for HIGH and LOW alarms serve: ren for NORMAL sed fr HIG ant LOW alana elo for TREND HIGH or TREND LOW conations Viol om sgn tall lsh even th aie alam sence and are con is comectd, Once corrected the vs sgl sal autom I * @ =Audible alarm shall be heard at 90 dBA © 3 ft, (NOTE: In the event of an al the ominous oul sasoud aan geen NORMAL e ally cancel, gas indicators shall change to HIGH or LOW red flashing alarm indica~ Ee Bi ton © Audible alarm silence button shal Be provided to silence the alarm. The audible alarm shall always remain in an alarm condition state and ener eS zed until the condition that initiated the alarm s corrected. ee | enone] : sia ao Seno © An alarm TEST button shal be provided to test the working conitions of Visual LED alarm indicators, audible warnings and gas set points. © Each gas sensor shall be able to communicate a continuous digital readout of 0-250 pst (0+1,724 KPal for pressure and 30" Hy (100-0 KPa) for 1 imeare | © Sensors shall have lca or remote mount capabilities (NOTE: Sensors ea Ve ibrated prior to shipment) 7 © Remote sensors shall have the capability to be located ‘up to 6000 ft or 1,524 m from the alarm (NOTE: Twisted ‘pai shielded wiring to be used UKET FR.03.02/VA/7770/2019 in application. PT. AMSAR PRIMA MANDIRI lok AAG No. 17 Rawakalong, Gunung Sindur, Bogor - Indonesia 16340 www.amsarmedical.com info@amsarmedical.com Telp. (021) 75674728 io’ Medical GAS ALARM 2 GAS Features Digital sensor for indicator display LCD panel operated by touch screen. Gas pressure sensor uses micro processor system, Indicator of damage appear immediately on the LCD, Interface with LAN / Building Management System Layar LCD/LCD Screen 10° [25.4 cml, digital sensor,2 text lines per gas. © Complies with NFPA"99 & Federal Cor mission (FCC) Part 15. UL/CSA Certifications Standard. Preasure sensitive LCD Touch screen, Adjustable Detection of Gas to Zone Location Auto Detection of Gas to Zone Location. Communicate to BMS. Adjustable audible alarm repeat. ° . . © Can be programmed to send emial / SMS if ei . . . reer F : munications Con Field adjustable pressure settings. Knockout for conduit installation, Equipped with local or remote sensors (Area), High visibility LED readouts. UL? Listed. Normally Open or Closed Contacts (Master) NEPA or ISO colors ~ Adjustable psi/KpA + inHg/mmHe- T I Adjustable. " 1 English/ Spanish Adjustable. © Assembled in the USA ISO 9001 & ISO 13485 Facility eoeecccccccce ml hel ge SUKET FR.03.02/VA/7770/2019 PT. AMSAR PRIMA MANDIRI lok AAG No. 17 Rawakalong, Gunung Sindur, Bogor - Indonesia 16340 www.amsarmedical.com info@amsarmedical.com Telp. (021) 75674728

You might also like