Ultrasonography
Janice S. Sung, MD
KEYWORDS
Breast Ultrasonography Imaging Image quality
KEY POINTS
Ultrasonography is an important modality that is frequently used in all aspects of breast imaging,
including breast cancer screening, the evaluation of palpable abnormalities, further characteriza-
tion of lesions seen mammographically, and for determining the method of percutaneous biopsy.
Understanding the basic technical aspects of ultrasonography equipment is critical to ensure high
breast ultrasonography image quality.
conclusions may be caused by technique or the imaged. During breast ultrasonography, the pa-
application or misapplication of image processing tient should be positioned with the ipsilateral arm
algorithms. Although there have been advances in over the head. In general, medial lesions should
Department of Radiology, Memorial Sloan-Kettering Cancer Center, New York, NY 10065, USA
E-mail address: sungj@mskcc.org
be scanned with the patient in the supine position, scale. Transducers with higher frequencies have
whereas lateral lesions should be scanned with improved contrast resolution.
the patient in a supine oblique position in order Spatial resolution is composed of both axial
to reduce the thickness of the breast. and lateral resolution. Axial resolution is the ability
to resolve structures along the axis of the ul-
TRANSDUCER SELECTION trasound beam or the Z plane (the depth). Axial
resolution depends on pulse length, which is
High-quality breast ultrasonography begins with determined by the frequency and the bandwidth
selection of an appropriate transducer. The trans- of the transducer. A transducer with a higher fre-
ducers used in breast imaging must have a high quency and a broader bandwidth has a shorter
frequency (between 10 and 15 MHz) because of wavelength and pulse length relative to a lower
the superficial nature of the breast and the need frequency transducer, which improves axial
to resolve small structures (Fig. 1). However, resolution.
higher frequency sound waves are more strongly Lateral resolution is the ability to resolve struc-
attenuated by tissue than lower frequency waves. tures in the X and Y planes that are at the same
Therefore there is a trade-off between higher reso- depth. Lateral resolution is related to the trans-
lution and reduced penetration. With proper posi- ducer beam width, and lateral resolution is not
tioning and the patient in the supine or supine adequate when lesions positioned side by side
oblique position, most breasts are only a few cen- are within the same beam width. Because a higher
timeters thick and high-frequency transducers frequency transducer has a narrower beam width
provide optimum image quality for all of the breast relative to a lower frequency transducer, lateral
tissue. When evaluating deep tissue in patients resolution is improved.
with particularly large breasts, it may be helpful
to have lower frequency transducers available to FOCAL ZONE
be used only in this specific situation.
Bandwidth is another consideration in trans- As discussed earlier, lateral resolution is related
ducer selection. The bandwidth is the spread of to the beam width. In addition to selecting a
frequency around the central frequency of the high-frequency transducer to maintain a narrow
transducer, and transducers with a broader band- beam width, the beam width can be further
width have improved resolution. Transducers used reduced by adjusting the focal zone.
in breast imaging may either be linear or matrix Linear array transducers have multiple piezo-
array, which affects image resolution. electric crystals or elements arranged side by
side. If a single element is used to both transmit
IMAGE RESOLUTION and receive the signal, the beam diverges quickly
after traveling a few millimeters, resulting in poor
The goal of equipment selection is to maximize im- lateral resolution because of beam divergence.
age resolution, which is the ability to distinguish Linear array transducers pulse adjacent elements
structures that are close together as separate le- simultaneously as a single element group to over-
sions. Image resolution is composed of contrast come beam divergence and then in succession to
and spatial resolution (Fig. 2). Optimal contrast form the image. Delaying the timing of firing of el-
resolution is necessary to differentiate subtle le- ements within a single element group adjusts the
sions from the surrounding breast tissue, and focal zone so that the beam is narrowed along
these lesions may have subtle variations of gray the long-axis plane. The specific time delay
Fig. 1. Transducer frequency. High-quality breast ultrasonography requires the use of high-frequency trans-
ducers. The same mass (arrows) is imaged using 9-MHz (A) and 12-MHz (B) transducers. The higher frequency
(12 MHz) transducer results in better detail of the mass and the surrounding breast parenchyma.
High-quality Breast Ultrasonography 521
Fig. 3. Focal zone. Images of the same mass with the focal zone set too deep (A, arrow) and at the appropriate
depth (B, arrow). When the focal zone is at the level of the mass, the margins and internal features can be better
characterized.
522 Sung
Fig. 4. Focal zone of superficial lesion. Because of transducer limitations, even the shallowest focal zone setting
may not achieve the narrowest beam width possible for lesions near the skin. Therefore a standoff pad may be
required to improve resolution. (A) A superficial hypoechoic mass (arrow). With the use of a standoff pad (B) al-
lowing optimal resolution at the level of the skin, a skin tract (arrow) is now seen, confirming the diagnosis of a
benign sebaceous cyst.
lesion, it may be beneficial to use standard ultraso- color Doppler.4 Power Doppler imaging may be
nography imaging, in which the posterior shadow- used to increased sensitivity compared with color
ing may help identify the lesion, and then apply Doppler in detecting small vessels and low flow.
spatial compound imaging for analysis once the The presence or absence or type of tumor
lesion has been detected. vascularity alone is not sufficient to characterize
a lesion as benign or malignant. Features suggest-
HARMONIC IMAGING ing malignant lesions include hypervascularity,
irregular branching central vessels, and more
Harmonic imaging is another signal processing than one vascular pole.5,6 Demonstrating internal
technique to improve contrast and lateral resolu- vascularity within a sonographic lesion confirms
tion and reduce artifacts.1 Harmonic imaging ap- that it is either solid or at least contains a solid
plies a filter to remove image harmonics, which component. Doppler imaging is most useful in dis-
are multiples of the transmitted frequency. This tinguishing a high-grade invasive cancer or meta-
technique is advantageous because much of the static lymph node, both of which can appear
artifact degrading image resolution is contained anechoic, from a simple or complicated cyst
within the lower frequency components of the (Fig. 8). Color Doppler may also be useful in differ-
beam (Fig. 7). entiating between debris within a duct and an in-
traductal mass.
COLOR DOPPLER AND POWER DOPPLER When using Doppler imaging, it is important to
apply light transducer pressure to prevent occlu-
Tumor angiogenesis plays a fundamental role in sion of slow flow within vessels. The position and
local tumor growth, invasion, and progression to size of the color box should also be focused to
metastases.2,3 As tumors outgrow their native the lesion of interest to maximize sensitivity to flow.
blood supply, hypoxia ensues, which induces
expression of multiple angiogenic factors such as REAL-TIME IMAGING
vascular endothelial growth factor. These factors
induce the growth of existing capillaries and Real-time imaging may be essential for accurate
the formation of abnormal vessels that are often interpretation in many situations. In patients with
tortuous and disordered, which may be seen on palpable complaints, real-time ultrasonography by
Fig. 5. Depth. The depth should be set so that the image focuses on the area of interest without including the
lung, which provides no useful information (A). When the depth is set appropriately (B), the features of the mass
(arrows) are better seen.
High-quality Breast Ultrasonography 523
Fig. 6. Gain. Lesions can be mischaracterized if the gain is set too low or too high. (A) The gain is set too low and
the lesion appears anechoic (arrow), suggesting a simple cyst. With the gain set correctly (B), low-level internal
echoes (arrow) are now seen in this complicated cyst. (C) The gain is set too high, creating artificial internal
echoes (arrow) mimicking a solid mass. When the gain is corrected (D), the lesion is seen to be a simple cyst
(arrow).
the interpreting radiologist may be prudent in cases biopsy is faster, more comfortable for the patient,
in which initial scanning by the technologist does and allows greater access to breast tissue, espe-
not identify an abnormality. Another situation in cially for far posterior and medial lesions that
which real-time imaging may be important is when may not be amenable to either stereotactic or
differentiating between mobile debris in a compli- MR imaging–guided biopsy. In addition, adequate
cated cyst versus a complex cystic mass (Fig. 9). sampling is more consistently obtained because
the needle can be seen traversing the target in
TARGETED ULTRASOUND: LESION real time.
CORRELATION When targeted ultrasonography is performed
to evaluate a mammographic finding, careful trian-
Targeted ultrasonography is often performed in gulation of lesion location should be performed.
order to identify a sonographic correlate to an Adjustments should be made when using cranio-
abnormality identified on mammography or caudal and mediolateral-oblique (MLO) views to
breast MR imaging before percutaneous biopsy. determine the clock axis for targeted ultrasonogra-
Ultrasonography-guided biopsy is the method of phy. The location of the lesion may be higher or
choice to sample any finding that is sonographi- lower than expected based on the MLO view for
cally evident. Compared with stereotactic or MR medial and lateral lesions respectively. In addition,
imaging–guided biopsy, ultrasonography-guided careful attention should be paid to both lesion depth
Fig. 7. Harmonic imaging. (A) The 12:00 axis of the right breast shows 2 subtle structures (arrow) that appear
almost isoechoic to the fat. (B) Harmonic imaging of the same area allows better delineation of the margins
and internal features (arrow), which allows the two lesions to be characterized as complicated cysts.
524 Sung
Fig. 8. Color Doppler. (A) A hypoechoic mass (arrow) in the far field with posterior enhancement, possibly rep-
resenting a complicated cyst. With color Doppler (B), vascularity (arrow) is seen within the mass, confirming
that it is solid and not a complicated cyst. Biopsy yielded invasive ductal carcinoma.
and the surrounding anatomic landmarks. The pa- enhancement.7–11 However, true ultrasonogra-
tient is positioned differently during breast ultraso- phy–MR imaging correlation can only be con-
nography, MR imaging, and mammogram, which firmed if follow-up MR imaging is performed,
affects lesion location. A superficial mass at the showing the localizing clip placed at the time of
fat-gland interface on mammography or MR imag- ultrasonography-guided biopsy within the area of
ing should also be at that interface on targeted ultra- enhancement on MR imaging. In one study, the
sonography (Fig. 10). Other findings such as an presumed sonographic correlate biopsied yielding
adjacent cyst or dilated duct may also be used as a benign, concordant diagnosis did not correspond
anatomic landmarks. with the lesion originally detected on MR imaging in
When sampling a potential sonographic corre- 12% of cases.10 For this reason, 6-month follow-up
late to a mammographic abnormality, a localizing MR imaging is recommended following benign
clip should be placed in the biopsied lesion using concordant biopsy of a sonographic correlate to
ultrasonography guidance. A postbiopsy mammo- an MR imaging–detected lesion.12
gram should be obtained to confirm correlation
between the biopsied lesion and the mammo- AMERICAN COLLEGE OF RADIOLOGY
graphic abnormality. ACCREDITATION
Targeted ultrasonography is often performed to
evaluate for a sonographic correlate to an MR The American College of Radiology offers a Breast
imaging finding in order to facilitate biopsy. A Ultrasound Accreditation Program. Accreditation
potential correlate is more frequently identified requires the application of many of the concepts
for enhancing masses compared with nonmass discussed in this article to obtain high-quality
Fig. 9. Real-time imaging. (A) In the right breast at the 6:30 axis is a cyst with a possible intracystic mass (arrow).
(B) When the patient is rolled into the left lateral decubitus position, the questionable mass moves into the
dependent portion of the cyst (arrow), consistent with mobile debris.
High-quality Breast Ultrasonography 525
Fig. 10. Lesion miscorrelation. (A) A superficial enhancing mass (arrow) in the superior breast. Targeted ultraso-
nography identified a mass (arrow) that was thought to represent a correlate (B). Although the sonographic mass
was of a similar size and clock axis to the MR finding, it was deep within the breast, whereas the MR mass was at
the fat-gland interface. Ultrasonography-guided biopsy yielded fibroadenoma. On 6-month follow-up MR imag-
ing, the clip (arrow) from ultrasonography-guided biopsy is seen deep to the MR mass (C). Subsequent biopsy
yielded invasive ductal carcinoma.
images. The center frequency of the transducer including breast cancer screening, the evaluation
must be at least 10 MHz. Unlike breast MR imag- of palpable abnormalities, further character-
ing accreditation, in which images from each indi- ization of lesions seen mammographically, and
vidual magnet must be submitted, a facility must for determining the method of percutaneous
only provide a single example of a cyst and solid biopsy. Understanding the basic technical as-
mass in the breast for ultrasonography accredita- pects of ultrasonography equipment is critical
tion regardless of the number of ultrasonography to ensure high breast ultrasonography image
units at that facility. Two orthogonal mammo- quality.
graphic views must be submitted with the lesion
circled and visible on both views. Sonographic REFERENCES
clinical images include the cyst and solid mass
on 2 orthogonal views without calibers, and 1 im- 1. Harvey CJ, Pilcher JM, Eckersley RJ, et al. Ad-
age with appropriate measurements. Images vances in ultrasound. Clin Radiol 2002;57(3):
should be taken along the longest axis of a lesion 157–77.
and then an orthogonal image. The longest axis 2. Schneider BP, Miller KD. Angiogenesis of breast
of the lesion is not necessarily in the radial or anti- cancer. J Clin Oncol 2005;23(8):1782–90.
radial plane, but may be in an oblique position. 3. Weidner N, Semple JP, Welch WR, et al. Tumor
Criteria for a simple cyst include anechoic, circum- angiogenesis and metastasis–correlation in inva-
scribed margin, and posterior enhancement. For sive breast carcinoma. N Engl J Med 1991;
submitted images of cysts, spatial compound im- 324(1):1–8.
aging should be avoided if the resulting image fails 4. Less JR, Skalak TC, Sevick EM, et al. Microvascular
to show posterior enhancement. architecture in a mammary carcinoma: branching
patterns and vessel dimensions. Cancer Res 1991;
SUMMARY 51(1):265–73.
5. Yang W, Dempsey PJ. Diagnostic breast ultrasound:
Ultrasonography is an important modality that is current status and future directions. Radiol Clin
frequently used in all aspects of breast imaging, North Am 2007;45(5):845–61, vii.
526 Sung
6. Mehta TS, Raza S, Baum JK. Use of Doppler ultra- suspicious on MRI. AJR Am J Roentgenol 2009;
sound in the evaluation of breast carcinoma. Semin 192(4):1128–34.
Ultrasound CT MR 2000;21(4):297–307. 10. Meissnitzer M, Dershaw DD, Lee CH, et al. Targeted
7. LaTrenta LR, Menell JH, Morris EA, et al. Breast le- ultrasound of the breast in women with abnormal
sions detected with MR imaging: utility and histo- MRI findings for whom biopsy has been recommen-
pathologic importance of identification with US. ded. AJR Am J Roentgenol 2009;193(4):1025–9.
Radiology 2003;227(3):856–61. 11. Beran L, Liang W, Nims T, et al. Correlation of tar-
8. Abe H, Schmidt RA, Shah RN, et al. MR-directed geted ultrasound with magnetic resonance imaging
(“Second-Look”) ultrasound examination for breast abnormalities of the breast. Am J Surg 2005;190(4):
lesions detected initially on MRI: MR and sono- 592–4.
graphic findings. AJR Am J Roentgenol 2010; 12. Sung JS, Lee CH, Morris EA, et al. Patient follow-up
194(2):370–7. after concordant histologically benign imaging-
9. Demartini WB, Eby PR, Peacock S, et al. Utility of tar- guided biopsy of MRI-detected lesions. AJR Am J
geted sonography for breast lesions that were Roentgenol 2012;198(6):1464–9.