Professional Documents
Culture Documents
BNN30104
BIOCHEMISTRY
AND
BIOMOLECULAR
TECHNOLOGY
What is a microscope? How does it work?
Fig 3. Microscopic
Anatomy of the
Retina.
The Human Eye
• For an image to be seen clearly, it must be spread on
the retina at a sufficient visual angle.
• Unless the light falls on non-adjacent rows of retinal
cells (a function of magnification and the spreading of
the image), we are unable to distinguish closely-lying
details as being separate (resolution).
• Further, there must be sufficient contrast between
adjacent details and/or the background to render the
magnified, resolved image visible.
“SIMPLE” MICROSCOPE
• Because of the limited ability of the eye’s lens to
change its shape, objects brought very close to
the eye cannot have their images brought to focus
on the retina.
• The accepted minimal conventional viewing
distance is 10 inches or 250 millimeters (25
centimeters).
• More than five hundred years ago, simple glass
magnifiers were developed.
• These were convex lenses (thicker in the center
than the periphery).
• The specimen or object could be focused by use
of the magnifier placed between the object and
the eye.
• These “simple microscopes”, along with the
cornea and eye lens, could spread the image on
the retina by magnification through increasing the
visual angle on the retina.
“SIMPLE” MICROSCOPE
• Conversely, it may be (and often is) all too easy to degrade an image
through improper technique or poor equipment.
• Essentially, this is how a microscope functions.
• Light from a lamp passes through a substage condenser and then through
a transparent specimen placed over an opening in the stage.
• Light is then gathered by the objective.
• The objective, together with the built-in tube lens, focuses the image of the
specimen at the level of the fixed diaphragm of the eyepiece.
• The image is then seen by the observer as if it were at a distance of
approximately 10 inches (250 millimeters) from the eye.
1665 – English
physicist, Robert
Hooke looked at a
sliver of cork
through a
microscope lens and
noticed some
"pores" or "cells" in
it.
•Compound/Light Microscope
•Dissection/Stereo Microscope
•Scanning Electron Microscope (SEM)
•Transmission Electron Microscope (TEM)
Important optical component of the
microscope: (1) OBJECTIVE
• Its basic function is to gather the light passing through the specimen and then to project an
accurate, real, inverted IMAGE of the specimen up into the body of the microscope.
• Other related functions of the objective are to house special devices such as
• an iris for darkfield microscopy,
• or a phase plate for phase contrast microscopy.
• The objective must have the capacity to reconstitute the various points of the specimen into the
various corresponding points in the image, sometimes called the “anti-points”.
• The objective must be constructed so that it will be focused close enough to the specimen so that it
will project a magnified, real image up into the microscope.
Important optical component of the
microscope: (2) EYEPIECE
• Its basic function is to “look at” the focused, magnified real
image projected by the objective and magnify that image a
second time as a virtual image seen as if 10 inches from
the eye.
• In recording, a Photo eyepiece “picks up” the real image
projected by the objective a second time as a real image
able to be captured by a camera.
• The eyepiece houses a fixed diaphragm.
• It is at the plane of that fixed diaphragm that the image
projected by the objective will be “seen”.
• On the shelf of the fixed diaphragm, the eyepiece can be
fitted with scales or markers or pointers or crosshairs that
will be in simultaneous focus with the focused image.
Stereo microscope
• an optical microscope variant designed
for low magnification observation of a
Sample
• using light reflected from the surface of an object rather than
transmitted through it.
• uses two separate optical paths with
two objectives and eyepieces
• provide slightly different viewing angles
to the left and right eyes - produces a 3D visualization
LIGHT microscope
• The objective lens located on the nosepiece have a short focal length
and are close to the target object where it collects light and focuses
the image of the object into the microscope.
• The second lens, in the eyepiece,
has a longer focal length
and further enlarges the image
Scanning electron microscope (SEM)
Revolving Nosepiece
Arm
Objective Lens
Stage
Stage
Clips Coarse adjustment knob
Diaphragm
Fine adjustment knob
Light
Base
Ocular lens
base
body tube
light
The proper way to focus a microscope is
to start with the lowest power objective
lens first and while looking from the
side, crank the lens down as close to the
specimen as possible without touching
it. Now, look through the eyepiece lens
and focus upward only until the image
is sharp. If you can't get it in focus,
repeat the process again.
Once the image is sharp with the low
power lens, you should be able to
simply click in the next power lens and
do minor adjustments with the focus
knob. If your microscope has a fine
focus adjustment, turning it a bit should
be all that's necessary. Continue with
subsequent objective lenses and fine
focus each time.
Rotate to 40x objective, locate desired
portion of specimen in the center of the
field. Refocus very carefully so that the
specimen is focused as sharply as
possible. (Do not
alter focus for the
Following steps )
Partially rotate so that 40x and 100x
objectives straddle the specimen.
Place a small drop of oil on the slide in
the center of the lighted area. (Take care
not to dribble on the stage.)
Put the small drop
of oil directly over
the area of the
specimen to be
Examined.
Rotate so that the 100x oil immersion
objective touches the oil and clicks into
place.
Focus only with fine focus. Hopefully,
the specimen will come into focus easily.
Do not change focus dramatically.
Clean up!: When you have finished for
the day, wipe the 100x oil immersion
objective carefully with lens paper to
remove all oil. Wipe oil from the slide
thoroughly with a Kimwipe. Cleanse
stage should any oil have spilled on it.
Recap the immersion oil container
securely, replace in drawer.
Note: this table is intended as a simple
Optical Microscopy Visible Light Detect reflected light (opaque ~200 nm Surface or volume
samples) or transmitted light (can probe through
(transparent samples). transparent
Light focused using lenses. materials)
Near-Field Optical Visible Light Detect reflected light (opaque ~10 nm Surface or volume Biological samples.
Microscopy (NSOM) samples) or transmitted light (can probe through
(transparent samples). transparent
Uses an aperture very close to the materials)
sample surface.
X-Ray Microscopy (TXM, X-Rays Image derived from x-ray scattering or ~20 nm Surface or volume Can be tuned to specific frequencies
SXM, STXM) interference patterns. (x-rays can to provide element identification and
X-rays focused using a “zone plate” penetrate some mapping.
(Fresnel lens). materials)
Scanning Electron Electrons Detect electrons back-scattered by ~1 nm Surface Sample must be in a vacuum.
Microscopy (SEM) the sample.
Electrons focused using
electromagnets.
Transmission Electron Electrons Detect electrons scattered as they ~0.05 nm Volume Samples must be <100 nm thick.
Microscopy (TEM, STEM) move through the sample.
Electrons focused using
electromagnets.
Focused Ion Beam (FIB) Ions Detect ions back-scattered by the ~10 nm Surface Due to the large masses of the ions,
sample. this probe can be destructive to the
Ions focused using electromagnets. surface of the sample. Therefore, it
can also be used to etch the
sample.
Scanning Tunneling Cantilever Tip Detect the quantum tunneling current ~0.1 nm Surface Sample must be conductive material
Microscopy (STM) of electrons from the sample to the and must be in a vacuum.
probe tip. Can be used to manipulate atoms
on the sample surface.
Atomic Force Microscopy Cantilever Tip Detect the electrostatic force between ~0.1 nm Surface Can be used to manipulate atoms
(AFM) the sample and the probe tip. on the sample surface.
Magnetic Force Cantilever Tip Detect the magnetic force between ~10 nm Surface Sample must be ferromagnetic or
Microscopy (MFM) the sample and the probe tip. paramagnetic.