You are on page 1of 17

Energy Dispersive, X-ray determination of elements in a sample, independent of

their chemical form. It is built on the fact that elements that


Fluorescence Analysis are irradiated with high-energetic X-rays have a certain
probability of emitting characteristic X-rays, the energies of
which are unique for each element. In the energy-dispersive
Peter Wobrauschek and Christina Streli (ED) systems, the emitted X-rays are detected via their
Atominstitut, Vienna University of Technology, energies. The use of the EDXRF technique has accelerated
Vienna, Austria since the 1960s as a result of the development of liquid
nitrogen cooled solid-state detectors, nuclear electronics,
Eva Selin Lindgren and small computers. Nowadays compact light-weight
electrically cooled detectors are available, which, together
Rector’s Office, University of Borås, Borås, Sweden with air-cooled low-power X-ray tubes, are perfectly suited
for handheld spectrometers. EDXRF is multielemental
and nondestructive and can be applied to large as well as
1 Introduction 1
small samples of different composition and character. If
2 Physical Principles 2 conditions are optimized, minimum detection limits can
2.1 Interactions of X-rays with Matter 2 be below the nanogram absolute or micrograms per gram
2.2 Principles in Quantitative Evaluation of concentration level for small laboratory instruments and
Element Concentration 3 into the femtogram or nanograms per milliliter region
2.3 Interelement Effects 4 for more advanced instrumentation (total reflection X-ray
2.4 Dependence of Detection Limits and fluorescence (TXRF), synchrotron radiation).
Penetration Depth on X-ray Energy EDXRF spectrometers exist at many degrees of sophis-
and Source Characteristics 5 tication, ranging from advanced laboratory instruments
3 Energy-Dispersive X-ray Fluorescence to small portable instruments for field observations. They
Instrumentation 6 can be designed for analysis of bulk material or for scan-
ning and elemental mapping of small areas. The use of
3.1 General Features 6
X-ray optics like single or polycapillaries and curved
3.2 X-ray Sources 6 mirrors in Kirkpatrick–Baez geometry lead to X-ray spot
3.3 Energy-dispersive X-ray Fluorescence sizes of micrometers and below ideally suited for micro-
Spectrometer Designs 8 XRF. Mapping with high spatial resolution is thus possible
3.4 Detector Characteristics and New opening investigations down to cell dimension.
Developments 9 Typical applications for EDXRF are analysis of
3.5 Micro-X-ray Fluorescence 10 agricultural material, medical samples, archaeological
3.6 Portable Systems 10 and historical objects, painting and fine art objects, and
environmental samples such as soil, ores, water, and
3.7 Combination with Other Techniques 10
aerosol particles.
4 Sample Handling and Quantification in
Practice 10
4.1 Sample Preparation 10
4.2 Quantitative Analysis of Major or/and
1 INTRODUCTION
Trace Constituents 12
4.3 Surface Analysis and Microanalysis 14 EDXRF is an analytical method for qualitative as well
as quantitative determination of elements in a sample,
5 Conclusions: Needs for Future
independent of their chemical form. It is built on the
Developments 14
fact that elements that are irradiated with high-energy X-
Acknowledgments 15 rays have a certain probability of emitting characteristic
Abbreviations and Acronyms 15 X-rays. The energy, E, of this radiation is uniquely
Related Articles 15 dependent on the element, which emits the characteristic
References 15 X-rays by Moseley’s law, E = K(Z − s),(1) where K and
s are constants that vary with the spectral series and Z
is the atomic number of the element. In ED systems, the
emitted X-rays are detected by their energies, whereas
Energy-dispersive X-ray fluorescence (EDXRF) is an in wavelength-dispersive X-ray fluorescence (WDXRF)
analytical method for qualitative as well as quantitative systems, the physical equivalent to energy, the wavelength

Encyclopedia of Analytical Chemistry R.A. Meyers (Ed.)


Copyright  2010 John Wiley & Sons Ltd
2 X-RAY SPECTROMETRY

of the characteristic photon is measured. This statement Owing to the heavy weight of the components, there are
can be explained by the simple relation no portable WD spectrometers available.
The use of the EDXRF technique has accelerated
E = hν, λ.ν = c > E = hc/λ > E(keV) since the 1960s as a result of the development of solid-
state detectors, nuclear electronics, and small computers.
= 1.2396/λ(nm) (1)
Today about 14 000 EDXRF spectrometers are in use,
many of which belong to research laboratories. This
from which it is seen that the wavelength and energy of
number is rapidly growing because handheld portable
the X-rays are reciprocal to each other.
spectrometers combining new detector technologies for
WDXRF utilizes the properties of single crystals to
charge collection and cooling are available. The compo-
diffract X-rays of a specific wavelength under a specific
sition of fine art objects, scrap metals, wall painting,
angle (Braggs law) and the intensity of the diffracted
or environmental samples can be analyzed on the spot,
radiation is counted by a simple radiation counter e.g.
within seconds. A vast number of scientific publications
Geiger–Muller counter. As characteristic radiation for
one element has a fixed well-defined value λ only at exist on the subject of EDXRF and its applications.
specific angles, a high intensity of the refracted radiation Authoritative textbooks on the principles and instrumen-
is measured. So using a goniometer with the Bragg crystal tation for EDXRF are those of Bertin,(2) Jenkins et al.,(1)
as the dispersive element in its center and a detector at Van Grieken and Markowicz,(3) and Beckhoff et al.(4)
the circumference, a continuous rotation will result in
intensive peaks whenever the condition of Braggs law
2 PHYSICAL PRINCIPLES
2d sin θ = n λ (2)

is fulfilled. Theta θ is the refraction angle, λ is the charac- 2.1 Interactions of X-rays with Matter
teristic wavelength, and d is the distance between lattice
planes of the single crystal. By continuous rotation of the When a beam of X-rays (originating e.g. from X-ray
detector from 0 to 180° , angle 2θ, while the crystal is in tubes) impinges on a sample, it will interact with
position θ, collects the photons reflected. This sequential the atoms in the sample by three processes, namely,
measurement leads to a spectrum ‘‘intensity versus wave- the photoelectric effect and coherent (Rayleigh) and
length’’ allowing determination of elements and concen- incoherent (Compton) scattering. For the photoelectric
tration in the sample because the intensity is proportional effect to occur, the energy of the impinging photons has
to the concentration of the respective element. to be large enough to create a vacancy in one of the
In the case of EDXRF, the detector crystal, made of shells of the studied elements. The photoelectric effect
lithium-drifted silicon Si(Li) or hyperpure Ge, takes on may result in emission of characteristic X-rays, but once
two tasks together, namely, dispersion according to the a vacancy has been created in an inner shell the atom
energy and counting of the photons of individual energy. can also de-excite by emission of Auger electrons. The
This collection process results in a spectrum ‘‘intensity probability that characteristic X-rays will be emitted –
versus energy’’ of the characteristic radiation emitted and not an Auger electron – varies from one element
from the sample. More details about energy-dispersive to another and is described as the fluorescence yield.
detectors are presented later. Typical applications For elements of low atomic number, emission of Auger
for EDXRF are analyses of agricultural material, electrons dominates, whereas emission of characteristic
medical samples, archaeological and historical objects, X-rays is more likely for elements of high atomic number.
and environmental samples such as soil, water, and A sketch of the XRF process leading to emission of a
aerosol particles. EDXRF working is simultaneous, characteristic X-ray is shown in Figure 1. On the basis of
multielemental, and nondestructive, and can be applied Moseley’s law the wavelength or energy of the emitted
to small samples (about 0.1 mg). For scanning objects and radiation is characteristic for the element and thus the
mapping small areas in a specimen, EDXRF instruments element can be identified.
with narrow beams of X-rays can be used. The scattering processes are of two types, incoherent
WDXRF is widely used for routine analysis and several and coherent. The probabilities of both kinds of scattering
hundred thousand instruments are distributed in steel vary with photon energy and the composition of the
production, cement industry, geologic sample analysis, sample. The scattered X-rays are an important part of
worldwide. In scientific research, the high resolution is the background radiation on which the characteristic
appreciated allowing precision measurements in cases peaks are superimposed. Figure 2 shows a typical X-ray
where the line overlaps represent a problem in EDXRF. spectrum with characteristic and scattered radiation.
ENERGY DISPERSIVE, X-RAY FLUORESCENCE ANALYSIS 3

Characteristic Incident monochromatic or contains a range of different energies.


X-ray photon Polychromatic X-ray sources can, however, be treated
Ei = EK − EL E0 > EK as a sum (or an integral) of monochromatic sources,
and the discussion below is therefore limited to the case
of monochromatic excitation. For this case, with further
limitations to homogeneous samples and in the absence of
interelement effects, a simple relationship can be obtained
between the mass thickness (or areal density), mi , of a
specific element and the measured intensity Ii of its
characteristic X-ray peak:
Photoelectron
EPE = E0 − EK Ii = Io Ki mi Di (3)

Figure 1 Schematic drawing of the XRF process in an atom. where I0 is the intensity of the primary beam of energy E0 ,
An incident photon of energy E0 transfers its energy to an which, in the case that the source is a secondary-target X-
electron in the K-shell. The electron is expelled from the atom
and leaves a vacancy. An electron from the L-shell can transit ray tube arrangement, is influenced by the anode material
to the K-shell to fill the vacant site. The difference in binding and operating voltage and current; Ii is the intensity of
energy between the two shells, EK − EL , is used by the atom to characteristic X-rays (K- or L-radiation) of element ‘‘i’’
emit a characteristic X-ray of energy E. Note that the distances of energy Ei .; Di is a correction factor for attenuation
between the atomic shells are not drawn to scale but shown of both the incident X-rays and the characteristic X-
schematically only.
rays of element ‘‘i’’ in the sample; and Ki is a factor that
depends on the setup of the instrument and basic physical
2.2 Principles in Quantitative Evaluation of Element processes.
Concentration Ki thus depends on instrumental factors and on
fundamental physical parameters related to the energies
The quantitative evaluation of element concentra- of the incident and characteristic X-rays. The important
tions in EDXRF depends on whether the source is parameters to be considered are

MoKa
inc.
coh.
MoKb
inc.
coh.
Ca

K
He Sr
Pb
As
Zn Pb Rb
Mn Br

P S Ar Cu

Cl

Cr Ni

0 100 200 300 400


Channels
Figure 2 X-ray spectrum of a biological standard reference material, SRM1571 (National Institute of Standards and Technology
(NIST)), ‘‘orchard leaves’’, measured in a secondary-target EDXRF spectrometer of the geometry of Figure 3. The sample thickness
is 30 mg cm−2 . The excitation conditions were as follows: tube voltage 55 kV and current 10 mA. The characteristic Kα lines are
denoted by chemical symbols, except for Pb, for which both Lα and Lβ are marked. Note the high intensities of the coherently and
incoherently scattered Kα and Kβ radiation of the secondary target, which, in this spectrum, is Mo. The X-ray spectrum has been
recorded with an Si(Li) detector.
4 X-RAY SPECTROMETRY

• the geometry of the XRF spectrometer; be done automatically by means of computer programs.
• the attenuation of the exciting and characteristic In evaluation of net peak areas in a spectrum, the problem
radiation along the beam paths and in the detector of interference between spectral lines will, however, have
window; to be addressed. In a number of cases, the K-lines from
• the detector efficiency for radiation of energy Ei ; one element will overlap or coincide with the L-lines
• the probability of producing a vacancy with incident of another element, as is illustrated in Figure 2 by the
radiation of energy E0 in the relevant shell of element overlap of As Kα with Pb Lα. In other cases, the Kα
‘‘i’’; from one element will coincide with the Kβ from another.
• the fluorescence yield of element ‘‘i’’. By knowing relative ratios, Kα/Kβ, Lα/Lβ, it is possible
to resolve the overlapping peaks. This problem is not
Note, however, that the essential feature of Equa- so severe in WDXRF as the crystal resolution is in the
tion (3) is that – under the assumptions of homogeneity range of a few electron volts at 5.9 keV, whereas the
and absence of interelement effects in the sample – the energy resolution of an ED detector is, for this energy
factor Ki is independent of the mass thickness, mi , of value, at its best around 130 eV at 5.9 keV. However,
the analyzed element and of the attenuation correc- this procedure of spectrum deconvolution will reduce the
tion, Di .Ki can be determined through calibration of precision of the measurement to an extent determined by
the instrument with suitable calibration standards under the statistics of the measurement.
well-defined conditions.
The amount of analyte is often written as 2.3 Interelement Effects
m i = Wi m (4) In a sample composed of many elements, each element
will interact with the radiation from all other elements. If
where Wi is the weight fraction mi is the mass thickness the matrix elements have characteristic radiation that is
of element ‘‘i’’ in the sample, and m is the sample mass. on the short wavelength side of the absorption edge
The attenuation correction, Di , in Equation (3) has the of the element of interest, element ‘‘i’’, conditions
form given in Equation (5): may be favorable for the enhancement of the lines
of element ‘‘i’’. The combined effects of the matrix
1 − exp −[(µ(E0 ) csc ϕ1 + µ(Ei ) csc ϕ2 ) · m] elements on each other are often called interelement
Di = (5)
[µ(E0 ) csc ϕ1 + µ(Ei ) csc ϕ2 ] · m or absorption–enhancement effects. Enhancement effects
are especially important for cases in which the analyte is
In this expression, µ(E0 ) and µ(Ei ) are the mass present at trace level and the enhancing elements at
attenuation coefficients for the exciting and characteristic percentage levels. In order to illustrate this, consider
radiation of energies E0 and Ei , respectively; they can the determination of one element, ‘‘i’’, by measuring its
be expressed as weighted sums of the individual mass Kα radiation in a matrix consisting of lighter as well as
attenuation coefficients for all elements in the sample. heavier elements. In this case, the lighter elements will
ϕ1 is the glancing angle between the exciting radiation not be able to enhance the characteristic Kα radiation
and the sample surface plane and ϕ2 the glancing angle of the element ‘‘i’’ because their characteristic radiation
between the characteristic radiation and the sample will not be energetic enough to create the necessary
surface plane. For two special cases, the attenuation vacancy in the K-shell of element ‘‘i’’ (see Figure 1).
correction and thus Equation (3) simplifies tremendously, Thus, it is necessary to look at the effects of the heavy
namely, for ‘‘thin’’ and ‘‘infinitely thick’’ samples. elements in the sample. If their Kα energies are lower
In some cases, it is also necessary to take into than that of the absorption edge for the K-shell in
account the fact that the characteristic radiation produced element ‘‘i’’ they will not cause enhancement. If, on
by an element as well as the scattered radiation can the other hand, their Kα lines have higher energies than
cause fluorescence of another element (interelement and that of the K-shell absorption edge they will enhance
related effects, as discussed below). This effect leads to the the Kα radiation of element ‘‘i’’. However, the general
enhancement of one spectral line, the absorption already problem of determining whether or not enhancement
having been taken into consideration. Furthermore, to will occur is not so difficult for homogeneous samples,
perform quantitative evaluation, one needs to know since absorption edge energies for different shells and
whether impurities are present in the spectrometer characteristic X-ray energies for all elements are known
material and the properties of the detector.(1 – 3) and readily available in tabulated form in textbooks or
Equations (3) and (4) require the net peak areas from the web.
of the characteristic peaks in the X-ray spectrum to A situation similar to the interelement effect occurs
determine element concentrations in a sample. This can when the backscattered exciting and/or fluorescent
ENERGY DISPERSIVE, X-RAY FLUORESCENCE ANALYSIS 5

radiation excites the elements in a thick sample. These Table 1 Depth of penetration in micrometers of characteristic
effects are of relatively large importance for analysis of radiation from sulfur, chromium, and yttrium in a matrix of
trace elements embedded in organic matrices, since the iron (density 7.87 g cm−3 ) and graphite (density 2.25 g cm−3 ).
The penetration depth for reduction of X-ray intensities to
medium-weight and heavy elements are usually present in two-thirds of the original has been calculated according to
low concentrations and the incoherent scattering in these the formula I = I0 exp(−µρ x2/3 ). The elements that generate
cases is dominating.(5) (Compare the spectrum from the characteristic radiation are assumed to be present at trace levels,
organic specimen as given in Figure 2.) This fundamental which is why the density of the matrix is not affected
problem described above is the same for EDXRF and Characteristic Matrix Mass x2/3 (µm)
WDXRF and must be taken into consideration in both radiation attenuationa
techniques. (µ, cm2 g−1 )
SKa Fe 1170 0.44
2.4 Dependence of Detection Limits and Penetration CrKa Fe 119 4.3
Yka Fe 58.5 8.8
Depth on X-ray Energy and Source Characteristics Ska C 208 8.7
The intensity of a beam of X-rays passing through a Cr Ka C 15.3 120
Yka C 0.813 2200
sample is attenuated by the above-mentioned photoelec-
tric and scattering processes according to the exponential a
J.W. Robinson (ed.), Handbook of Spectroscopy, CRC Press, Boca
law: Raton, Vol. 1, 1974.

I (E) = I0 (E) exp(−[µ(E)ρx]) (6) on whether corrections for attenuation have to be made
when quantifying element concentrations. Concepts such
where I0 (E) is the original intensity of the beam; I (E) is as ‘‘thin or thick’’ samples are always seen in relation to
the intensity after the beam has traveled a distance x in the energy of the X-rays of interest.
the sample; ρ is the density of the sample; and µ(E) is the
mass attenuation coefficient of the sample material. All
these considerations are valid for both WDXRF as well 2.4.2 Minimum Detection Limit
as EDXRF.
A widely used definition for the minimum detection limits
Since the three attenuation processes that contribute to
(MDL) in EDXRF is the concentration (or, alternatively,
the total attenuation of the beam are independent of each
the amount of mass) needed for the number of counts
other, the mass attenuation coefficient can be written as
in the characteristic
√ peak of an element to be equal to
a sum of coherent, incoherent, and photoelectric mass
or larger than 3 B, where B is the number of counts √
attenuation coefficients. Each process has a functional
in the background below the peak of the element ( B
dependence of the photon energy, E, shown explicitly in
corresponds approximately to one standard deviation).
Equation (6).
Specifically

2.4.1 Penetration Depth 3 √


MDL = · B (7)
k
A relevant question in XRF analysis is how deeply the
X-rays will penetrate into the material. A quantitative where k is the corresponding sensitivity constant
measure of this can be given in terms of the penetration measured as the net number of counts in the characteristic
depth, which is usually defined as the path length needed peak per unit concentration and time.
for the intensity of the X-rays to be reduced by a Since the attenuation of X-rays is stronger for soft X-
given factor. Since the mass attenuation coefficients, rays than for hard X-rays, the background in many designs
µ(E), in Equation (6) vary widely with X-ray energy and of EDXRF spectrometers is rather flat for energies
material composition, penetration depths are indeed very below the scattered radiation. Best values for MDL
different for hard and soft X-rays and vary appreciably are usually obtained for medium-weight elements. From
with sample composition. Table 1 shows some typical Equation (7), it can be seen that the general problem of
values for penetration depths for characteristic X-rays lowering detection limits can be solved by finding ways of
of some elements embedded in an iron and carbon increasing the intensity of the characteristic peak while
matrix.(5) In this example, the penetration depth varies decreasing the intensity of the background radiation.
by 3–4 orders of magnitude. Generally, X-rays of low A substantial portion of the background radiation is due
energy do not penetrate very far in a matrix of heavy to incoherent and coherent scattering of the primary
material, whereas high-energy X-rays in a light material radiation in the sample itself. These scattered peaks will,
have high penetration. This fact has a large influence by means of secondary processes in the sample, give rise
6 X-RAY SPECTROMETRY

to a general increase in the background also at lower is informative. For reasons of comparison, a more
energies. Furthermore, the scattered radiation will also appropriate parameter is the MDL achieved with each
contribute to an increase in the background level owing to instrument.
incomplete charge collection and other processes in the
solid-state detectors. Therefore, sample properties that
influence scattering processes, for example, thickness and 3 ENERGY-DISPERSIVE X-RAY
average atomic number, have an essential influence on FLUORESCENCE INSTRUMENTATION
the detection limits in an EDXRF spectrometer.
The counts in both peak and background are time 3.1 General Features
dependent, but in different ways: √ the peak increases
linearly with time, whereas B increases with the A typical feature in element analysis in practice is that
square root of time. This means that the detection samples may contain major as well as trace concentrations
limit will be reduced as analysis time is increased, and of different elements. Samples such as alloys, mineral
the analyst can use this fact if elements present in low ores, and ceramics usually contain a number of elements,
concentrations are of special interest. The price paid for the concentrations of which may vary by several orders
improvement in MDL is, of course, a longer running time of magnitude.
for the spectrometer. For many laboratory instruments, In an EDXRF spectrometer, the important parts to
collection times in the range 200–1000 s for each X-ray consider are the X-ray source, the beam and spectrometer
spectrum are chosen. geometry, and the detector. An obvious goal for the
In practice, for bulk samples of organic material, MDLs designer of XRF spectrometers is to maximize the
are of the order of 0.1 ppm for laboratory instruments for intensity of the characteristic X-rays from the elements
monochromatic sources with energy matching the binding while keeping the background radiation as low as possible.
energy for the K-shell of the element of interest.(6) In Most ED spectrometers have an elemental range from
bulk geological samples, MDLs are generally somewhat Na to U and, if special detectors with ultrathin windows,
larger. Measured as absolute amounts (for example in polymers of 300 nm supported by a Si grid are used,
mass units), the lowest MDLs can reach the picogram the range can be extended to still lighter elements such
region for thin samples (especially with TXRF(7) ) and the as Be, B, and C.(9) There are many different designs
femtogram region for microbeams that utilize a focused of ED spectrometers for laboratory and field work.(10)
synchrotron beam as source.(8) Combinations of WDXRF and EDXRF systems are also
MDLs in the 10−10 g region are reached for many used, in which the better resolution of WD instruments
elements in thin and thick samples in laboratory EDXRF is utilized for the lightest elements and the simplicity and
spectrometers and MDLs with TXRF yield 10−12 g using higher sensitivity of ED systems is utilized for elements
laboratory sources. above Ca. The cost of an ED system is about US$ 70 000
and upward.
2.4.3 Sensitivity
3.2 X-ray Sources
A concept that is often used to describe the performance
of an EDXRF method is the sensitivity, which is defined In photon-induced XRF, the most common sources are
as the change in instrument response (net number of X-ray tubes, which can be used directly for broadband
counts in a characteristic peak) per unit change in the excitation or in combination with filters or secondary
concentration of a measured element. Thus, sensitivity in fluorescers to obtain semimonochromatic radiation. In
X-ray instruments can be given as count rate per parts per some spectrometers, especially portable instruments for
million. The sensitivity concept is useful for comparing field work, and for problems in which only one or a few
analyses of different elements in a given spectrometer. elements are of interest (for example, in routine process
It varies to a high degree with the difference in energy control), radioisotopes are still useful.
between the incident and the characteristic radiation and
is highest if the energy of the incident radiation is only 3.2.1 X-ray Tubes
slightly greater than the energy of the K-edge for the
element in question. In X-ray tubes, the radiation spectrum is influenced
However, sensitivity in XRF instruments (expressed by two major processes that occur when electrons are
as count rate per parts per million) is a measure of the accelerated by the high voltage from the cathode to the
excitation and detection capability for each element with anode. One process is the creation of characteristic X-
a specific XRF instrument. Usually, sensitivity is not used rays in the anode material itself, which occurs when the
for comparing different instruments but, nevertheless, kinetic energy of the impinging electron is larger than
ENERGY DISPERSIVE, X-RAY FLUORESCENCE ANALYSIS 7

the binding energy of a particular shell in the anode Secondary


material. The other process is the creation of continuous target X-ray
bremsstrahlung, which is created when the electrons are tube
slowed down in the anode. The bremsstrahlung spectrum
contains energies from zero up to the kinetic energy of the
electrons, which is equal to the electron charge times the
tube voltage. Accordingly, the spectrum from the X-ray
tube contains a continuous distribution overlaid with the
Detector
characteristic radiation from the anode material about
100 times higher in intensity compared to bremsstrahlung
in that energy range. By varying parameters such as anode
material and thickness, tube voltage, and current settings, Collimator
the broadband spectrum from the X-ray tube can be
Sample
optimized for the elements of interest. Some special X-
ray tubes are equipped with dual (W/Sc or W/Mo alloy) Figure 3 Preferred geometry of a secondary-target EDXRF
or multiple anodes mechanically exchangeable in the high spectrometer. The broadband radiation from an X-ray tube
vacuum of the tube. excites a secondary target of a pure specimen, most often a
Furthermore, the spot size of the electron focus on the metal. The characteristic Kα and β radiation from the secondary
target is used as semimonochromatic X-ray source to excite
anode material is an important factor defining the emitted
the different elements in the sample. A detector registers the
brightness. The applicable electric power depends on the radiation from the sample. Note that the three X-ray beams are
size as the heat dissipation is the limiting factor. X-ray mutually orthogonal and that the sample plane is horizontal.
tubes with standing anodes have, for Mo and W, typically
3 kW (60 kV and 50 mA) if a focus of about 0.4 × 12 mm2
is used. Broad focus tubes with 7 × 7 mm2 can dissipate directions of primary beam path, secondary-target beam
up to 40–50 kW. Rotating anodes where the anode is path, and path from sample to detector. The material in
formed like a wheel and rotated at 10 000 rpm allows up the secondary target can be chosen so that its character-
to 18 kW for a 0.5 × 10 mm2 electronic focus. Low-power istic radiation is just above the K-edge of the elements
X-ray tubes of 50–75 W (50 kV and 1–1.5 mA) with a of most interest in the sample. The anode material in
focal size of 100 µm × 100 µm emit a high flux of photons the X-ray tube is generally chosen to have a high output
and can be ideally combined with X-ray optical elements over a wide range of energies in order to allow different
to focus the beam to micrometer dimensions with a flux secondary-target materials to be used. Most ideal are
of 1 × 1010 photons/s and are perfect for microanalysis combinations where the characteristic radiation from the
in the lab. These X-ray tubes are air cooled and of anode is with its energy just above the absorption edge of
light weight, ideally suited for the design of handcarried the target material e.g. Rh anode–Mo secondary target.
portable spectrometers.(11) Also the tube radiation is linearly polarized after 90° scat-
As the MDL for a given element is dependent on the tering on the secondary target in this orthogonal geometry
intensity of the characteristic as well as the background (all axes have an angle of 90° to each other), thus not
radiation (Equation 7), there is often a desire to use contributing to spectral background in the triaxial geom-
monochromatic or quasi-monochromatic radiation. This etry of Figure 3. Although the intensity of the primary
can be achieved by use of filters and/or secondary-target beam is reduced considerably by the secondary-target
arrangements. Filters inserted between the X-ray tube arrangement, the technique usually gives low detection
and the sample will allow transmission of the character- limits for many elements. This is mainly because of the
istic line of the filter but will attenuate the low-energy part low background in the X-ray spectra (see Figure 2). Other
of the tube radiation by a substantial factor. Thus, filtering techniques to modify the output from an X-ray tube
may change the shape of the tube spectrum and improve utilize low-Z elements (carbon or Plexiglas) scatterers
MDLs for elements lighter than the filter material. By to polarize the entire spectrum of X-rays from the tube
varying filter material and filter thickness, this method or multilayers to obtain an intensive purely monochro-
can give a high flexibility in the use of tube excitation. matic beam, as demonstrated for TXRF spectrometers
In secondary-target arrangements, the radiation from (see Streli et al.(8) ).
the X-ray tube is allowed to excite a target made of a mate- For microbeam EDXRF, microfocus low-power X-ray
rial whose characteristic radiation (only the intensive Kα tubes with anode spots of the order of 100 µm2 are now
and Kβ lines are emitted) is used as a source to excite available, as are high-intensity rotating anode sources
the sample. Figure 3 shows an arrangement of this kind, (18 kW) with various focal spot sizes (see Beckhoff
often called triaxial geometry due to the three orthogonal et al.(4) ).
8 X-RAY SPECTROMETRY

3.2.2 Radioactive Sources Kirkpatrick–Baez optics with two mirrors leading to


sensational beam dimensions of micrometers and even
Radioisotopes as X-ray sources utilize the γ-rays from down to 30 nm. This is ideally suited for elemental analysis
the decay of the radioisotope or the characteristic X-rays of cell material and inclusions in geological formations.
emitted from the decay product. The output intensity Also micro-XRF in human bone, radioactive particles
is lower than that which can be achieved by an X-ray and paintings were reported Janssens et al.(16)
tube, but the advantages are that the intensity of the
source is stable and no electric power supply is needed.
Radioisotopes are therefore practicable in portable field 3.3 Energy-dispersive X-ray Fluorescence
instruments. The drawback is the limited half-lives for Spectrometer Designs
some of the most commonly used isotopes (for example As already pointed out, there are three main points to
2.7 years for 55 Fe with 5.9- and 6.4-keV X-rays (Mn Kα consider for achieving good performance in an EDXRF
and Kβ lines) and 1.3 years for 109 Cd with 22.2- and system, namely the X-ray source, the detector, and the
25.5-keV X-rays (Ag Kα and Kβ lines) plus 88.2 keV design of the spectrometer. A few things of importance for
γ-rays). Also 241 Am with 59.5 keV γ-rays is useful for the design are mentioned here, namely, beam geometry,
higher-Z element excitation and this isotope has a half- choice of materials, and compactness of the design.
life of 433 years which is convenient. The most common The problems are discussed from the perspective of
geometries of the radioisotope sources are either that of a a secondary-target setup but are easy to generalize for
point source or that of an annular source. The sensitivity other arrangements.
and MDLs are, for some instruments, almost identical to
those for X-ray tube secondary-target arrangements.(12)
3.3.1 Beam Geometry

3.2.3 Synchrotron Radiation A large part of the background signals in X-ray spectra
depends on primary and secondary scattered radiation in
Synchrotron radiation is the ideal source for EDXRF, the sample. This can be illustrated in the secondary-target
since it can be monochromatized to the desired XRF arrangement shown in Figure 3. In such a system,
energy with high brilliance. Furthermore, it is linearly contributions to the background in the final spectrum are
polarized in the plane of the electron or positron orbit, obtained from (i) the scattered primary radiation from
which is useful for decreasing scattered radiation and the tube; (ii) the scattering of the quasi-monochromatic
thus reducing the background. Synchrotron radiation beam from the secondary target; (iii) the radiation from
at the National Synchrotron Light Source ((NSLS), the sample; (iv) detector artifacts like tailing stemming
Brookhaven National Laboratory) with collimated and from incomplete charge collection; (v) escape and sum
focused X-rays of high energy has been used in peaks; and (vi) Compton backscattering from high-energy
an X-ray microscope, mapping trace elements with photons leading to high background in the lower energy
a spatial resolution of 10 × 10 µm2 and with MDLs region. Instrumental blank spectrum from collimators can
into the femtogram region for thin specimens.(13) be a problem in the detection of ultratrace elements if the
However, synchrotron radiation is expensive and is not instrument lines are the same as in the analyzed sample.
generally available in research laboratories. TXRF with The primary radiation from the X-ray tube is very
synchrotron radiation has brought a major breakthrough intense and, although it will be scattered twice (in the
in the surface analysis of Si wafers at ultratrace levels. secondary target and the sample), it is likely to contribute
Detection limits of absolute 4 fg for Ni corresponding to the background unless special care is taken. The
to the more used units in surface contamination radiation from the secondary target will be scattered
of 5 × 108 at cm−2 were achieved in Hamburg at upon reaching the sample and constitutes the four large
Hasylab Deutsches Elektronensynchrotron (DESY) with scattered peaks shown in Figure 2. These peaks have been
monoenergetic synchrotron radiation from a bending observed to be responsible for part of the flat background
magnet beamline.(14) Also at Stanford Synchrotron in the low-energy region by multiple scattering in the
radiation laboratory (SSRL) TXRF measurements were sample and by incomplete charge collection effects in the
performed showing the low-Z element detection capacity detector.
with optimized excitation condition.(15) From basic physical principles, it is well known that
In a critical review of synchrotron radiation total scattering is at a minimum when the scattering angle is
reflection X-ray fluorescence (SR-TXRF), the advantages 90° . Thus, it is important that the impinging and emitted
and disadvantages can be found in Streli et al.(8) radiation form an angle of 90° with respect to each other.
Synchrotron radiation is most effective combined with This will be achieved if the secondary target and the
X-ray optical elements like single or polycapillaries or sample are both oriented at angles of incidence of 45° with
ENERGY DISPERSIVE, X-RAY FLUORESCENCE ANALYSIS 9

respect to the beam directions. Because the radiation from snout inside, are designed for this purpose. In combination
the X-ray tube is scattered twice – in the secondary target with special detectors having ultrathin windows, elements
and in the sample – and since the reason for scattering from Be upward can be detected. To evacuate the sample
at 90° is to utilize both minimum scattering cross section area comprises two other advantages: the scatter of the
and the production of a linearly polarized beam, it is also exciting radiation in air is avoided, which improves the
of importance that the geometry of the three beam paths scattered background, and the Ar K-lines will not be
is a triaxial (xyz) geometry as shown in Figure 3 and not measured since Ar is not present in vacuum.
an xyx planar geometry.(17)
3.4 Detector Characteristics and New Developments
3.3.2 Choice of Materials – Compact Designs
For ED spectrometers, liquid-nitrogen-cooled Si(Li)
In an EDXRF spectrometer, the major part of the primary detectors are the most widely used, having their perfor-
radiation emitted by the tube and secondary target will mance for X-rays of an energy ranging from 1 to about
not only be directed toward the sample but will also 25 keV and described by the efficiency of the detector.
interact with the spectrometer. This interaction will take For K-lines from heavier elements (30–100 keV) the
place by means of the same processes as in the sample planar high-purity germanium detectors are preferable
itself. Different materials have different abilities to scatter owing to their better detection efficiency in this region;
and absorb radiation, and the purity of the material will the formerly available Ge(Li) is not produced any longer.
have an influence on whether characteristic lines from The classical LN2-cooled detector is bulky because of
impurities in the material close to the beam paths will the storage vessel (Dewar) for the cooling medium. An
pass into the detector. Thus, it is important that the elegant way to get rid of the Dewar is the use of elec-
materials in objects such as filters and collimators are of trical cooling by Peltier elements. These Peltier-cooled
the highest possible purity. detectors operate not at 77 K but around 240 K, which is
Another important aspect to consider in an XRF setup reached by one- or two-stage Peltier elements working
are the actual geometric dimensions of the various parts of with a power of 3–4 V and 1–2 A.
the spectrometer. Since the general goal is to have efficient The fundamental working principle of an ED detector
and high sensitivity in the instrument, long beam paths are is the absorption of the characteristic photons in the
of disadvantage. In air, a long beam path for the radiation intrinsic zone of the Si crystal. The energy absorbed
from sample to detector will contribute to the attenuation in the semiconductor material creates a certain number
of characteristic radiation from the sample, and thus of electron–hole pairs, linearly correlated to the energy
significantly reduce detection efficiency for light elements of the photons. These charges, created in the intrinsic
in particular. Thus, a geometry as compact as possible zone of the Si crystal, form a charge pulse collected at
with short beam paths is desirable. However, the material the respective electrodes on the contact layers across
in the housing of the spectrometer has to be chosen so the surface of the Si crystal and are proportional
that unwanted radiation can be stopped efficiently. The to the energy of the absorbed photon. This p–i–n
combination of compact geometry and high stopping (positive–intrinsic–negative)-type structure of the Si
efficiency generally means that light materials such detector collects the charges along the parallel electric
as aluminum should be avoided or that the radiation field lines between p- and n- silicon and the contact layers
channels have to be clad with higher-Z materials having under reverse high voltage. A new idea to change the
good absorption properties. way of collecting charges is the sideward drift.(18) In this
Another aspect to consider in relation to Figure 3 is the case, the positive electron-collecting electrode is small
position of the sample. In many spectrometers, the sample and located either at the center of the crystal or slightly
position is vertical, which means that samples have to be eccentric at the outside of the detector structure, and
self-supporting. By choosing an arrangement in which the has the electric field lines parallel to the surface of the
sample plane is horizontal and the detector looks down or detector crystal. According to the collection mode, these
up at the sample, a more versatile instrument is achieved, detectors are termed drift detectors. The advantage is
in which the sample can also be in nonself-supporting high energy resolution, high throughput, and operation
forms, for example, as liquids and powders. With a at 220–250 K. The charge cloud is collected by a charge-
vertical sample position, such specimens are difficult to sensitive preamplifier and transformed into a voltage
irradiate directly by the beam. Measurements of light signal that is amplified through the classical electronic
elements require a special environment – either vacuum amplification chain of preamplifier and amplifier/shaper,
or He flushing – to avoid absorption of the low energy and converted from the analog signal into a digital
radiation. Thus vacuum chambers holding all components one by the ADC (analog-to-digital converter). Modern
of the spectrometer, in particular, sample and detector electronics use direct conversion and digitize the preamp
10 X-RAY SPECTROMETRY

signal immediately (DSP (digital signal processing)) spatial resolution mapping. In a particularly interesting
leading to best possible energy resolution in the range geometry, placing a second polycapillary in front of
of 125–130 eV at 5.9 keV, which is the accepted reference the detector and aligning the system to have both foci
energy emitted from a 55 Fe source. Another new feature matching on the surface of the sample where the X-ray
of a modern ED detector is the extension of the detectable beam impinges, leads to the so-called confocal geometry
range of elements to the light elements like Be, B, C, N, with the possibility to scan layer by layer starting from
O, F, Na, Mg, and Al. This is possible because, instead the surface and getting a 3D image at the end. Even
of the Be entrance window, polymer structures of 300- absorption effects can be corrected as the depth is well
nm thickness supported by a Si grid (Moxtek window known. A typical voxel size is 20 × 20 × 20 µm3 . Since
trade mark (TM)) can withstand air pressure and are not the spectra in each measured position is known, an
absorbing as strongly as 8-µm Be which normally is in elemental presentation in 3D is finally possible (chemical
use. Also, the contact layers on the crystal are extremely imaging).(23)
thin giving an adequate efficiency in the energy region of An excellent review of microanalysis is given in Ref. 16.
the light elements. All measurements for light elements Figure 4 shows an example of 2D imaging of a bone
have to be performed in vacuum to avoid absorption of sample.
the fluorescence radiation on the path to the detector.
The dimensions of pin-type detectors range from 10 to
3.6 Portable Systems
100 mm2 with a thickness of 3–4 mm, the silicon drift
detectors (SDD) have typically 10–100 mm2 area but Developments in excitation sources, detectors, and micro-
are between 300 and 500 µm in thickness. As a result processor technology have facilitated the design of ED
of the thinner crystal dimension, a poorer efficiency of configurations capable of analyzing most elements with
the SDD is observed. It continuously drops from about detection limits below the milligram per kilogram level.
10 keV compared to the Si pin where the decrease of In the last 20 years, portable XRF systems have also been
the efficiency starts roughly at 20 keV. For benchtop and developed and are now commercially available, making
portable spectrometers, scintillation counters and gas- it possible to take the spectrometer to the sample rather
filled proportional counters are sometimes used, often than the other way round. A detailed review has been
in combination with filters to compensate for the lack published by West and Potts.(24)
of energy resolution. Other types of detectors have Special applications find the portable systems in the
been developed, for example, mercury iodide detectors field of archaeometry and cultural heritage investigations
at room temperature for high- and medium-Z-element as the technique is nondestructive. Further in-field
analysis.(19) measurements and screening of hazardous substances
High-purity Si detectors are now able to separate are possible with handheld systems.
K fluorescence lines down into the 1-keV region and
compare well with compound semiconductor detectors
such as CdZnTe and HgI2.(20) An overview of the present 3.7 Combination with Other Techniques
status of several semiconductor detectors can be found in A comprehensive review on recent technological advan-
the article by McGregor and Hermon,(21) and Beckhoff ces is given by Tsuji et al. in Tsuji et al.(25)
et al.(4)
Arrays of SiLi, GeHP, or SDD are available with
up to 100 elements, increasing the solid angle of
detection drastically – obviously extremely expensive 4 SAMPLE HANDLING AND
instruments.(22) QUANTIFICATION IN PRACTICE

3.5 Micro-X-ray Fluorescence 4.1 Sample Preparation


The use of X-ray optical elements such as single or A typical feature in element analysis in practice is that
polycapillaries, where X-ray total reflection occurs on the samples may contain high as well as trace concentrations
inner wall side in of a thin-walled tube, on the inner wall of different elements. Samples such as alloys, mineral
side X-ray total reflection occurs is leading to a focussed ores, and ceramics usually contain a number of elements
beam of high intensity. A bunch of several hundred whose concentrations may vary by several orders of
thousands of single capillaries form the polycapillary with magnitude.
intensities several orders of magnitude higher than a EDXRF analysis can be performed on a vast number
pinhole of the same dimension. Scanning in steps of of samples of very different composition and status.
micrometers according to the spot dimensions allows high Examples of common samples are
ENERGY DISPERSIVE, X-RAY FLUORESCENCE ANALYSIS 11

BE I
II
III
V

IV

100 µm
(a)

11400 370
Ca Sr

0 0

4000 1200
Zn Pb

0 0
(b)
Figure 4 Elemental (Ca, Sr, Zn, and Pb) 2D image of a bone sample.

• geological – for example minerals, rock, soil, and 4.1.1 Solid Samples
sediments;
• biological – roots, leaves, stems, moss, lichens, algae, Homogeneous samples for which the concentrations of
fish, and animals; the elements of interest are significantly larger than the
detection limits (usually 0.1–1 ppm) are conveniently
• medical – blood, hair, skin, muscle, brain, teeth, bone,
analyzed without any sample preparation. This is a great
enzymes, and metalloproteins;
advantage, since a minimum of sample treatment will
• industrial – metals, ores, paper, and pulp;
minimize the risk of contamination. Many samples belong
• cultural – pottery, paintings, statues, coins, books, and
to this category.
clothes;
Some samples may have sufficiently high element
• environmental – water, air pollutants, waste, and
concentrations but may be inhomogeneous and contain
pollen;
grains of different sizes. In view of the poor penetration
• pharmaceuticals;
depths for X-rays from light elements, the grain size
• food stuff and drinking water, wine, or brandy.
may affect the quantitative evaluation. For such samples,
either grinding into a fine powder or digestion and dilution
There is no strict division between the different may be necessary.
categories, but the above list gives an idea of the
diversity of the samples concerning range of elemental 4.1.2 Liquid Samples
concentrations – from traces in biological material to
major components in minerals – as well as the range of Trace elements in water samples can be present in
densities involved. The material may exist in different different forms, for example as insoluble particulates or
forms, such as airborne particles, liquids, and solids. as colloidal or soluble species. Suspended material can be
12 X-RAY SPECTROMETRY

separated by filtration, and the deposits on the filter can For polychromatic X-ray sources, for example broad-
be directly analyzed in the X-ray spectrometer. For water, band excitation from X-ray tubes, Equation (3) is not
many element concentrations are below MDLs, especially valid. In this case, the intensity, Ii , can be described by
for heavy metals. The colloidal and soluble fractions may an integral over the distribution of excitation energies.
therefore need preconcentration, which may be achieved In practice, the approach often used is to calibrate the
by evaporation by heating or by freeze-drying. Care must instrument with standards as similar as possible to the
be taken, however, to avoid loss of volatile elements investigated samples.
in these processes. A freeze-dried residue may be acid- Liquids are a special case of homogeneous samples.
digested. A number of other chemical preconcentration For most liquid samples, quantification is facilitated by
procedures have been used on natural waters, snow, and the ability to spike the sample with internal standards
rain, with acceptable precision and recovery rates.(26) before digestion or further treatment. By choosing two
Many of these techniques have the advantage that they internal standards with different attenuation properties,
increase element concentrations, and at the same time, the attenuation for all elements in the sample can be
giving rise to thin film samples that are easy to evaluate obtained together with information on the sample mass
quantitatively. For many fluids – and in particular if the thickness. In TXRF, samples are prepared on a reflector
TXRF technique is used – the fluid can be deposited as thin films achieved after the drying process of the liquid
directly onto a carrier and dried at low temperature. solution where the addition of one internal standard was
TXRF shows excellent results for the analysis of Pb, Hg done. The quantification is easily done with reference
in water where 10 µg L−1 MDL can be reached. to this internal standard added to the solution, using the
Details on sample preparation are given in Sample calibration data for each element expressed as sensitivities
Preparation For X-Ray Fluorescence Analysis achieved from a set of multielement standards with
varying concentrations. The equation reads as follows:

4.2 Quantitative Analysis of Major or/and Trace Cx = Ix /Sx × Sstd /Istd × Cstd (8)
Constituents
As discussed in Section 3.2, quantification involves where Cx denotes concentration of the unknown; Ix , the
knowledge of some factors that are different for intensity of the unknown element; Istd , the intensity of
each spectrometer (geometrical properties, detector and the standard; Sx , the sensitivity of unknown element; Sstd ,
source characteristics, etc.) and some parameters that are the sensitivity of standard; and Cstd , the concentration of
dependent on physical processes. In a practical case, both standard.
these aspects have to be taken into account. In this case, after the calibration of the spectrometer is
established, the sample is spiked with the known amount
of internal standard and dried on a reflector serving as
4.2.1 Homogeneous Samples: General Approach sample carrier, and after EDXRF analysis, the evaluation
of the intensities will lead immediately to quantitative
Samples analyzed in spectrometers with monochromatic results of the unknown elements in the sample.
or nearly monochromatic X-ray sources are usually rather
easy to work with, since Equations (3–5) and (6–8)
4.2.2 Heterogeneous Samples
can be used for quantitative evaluation. A convenient
approach is to calibrate the spectrometer with known For heterogeneous samples, quantitative evaluation is
standards, which are either thin or of known mass more difficult, and this is particularly the case for X-rays
thickness. Calibration thus gives information on the from light elements, for which the penetration depths
factors Ki for different characteristic X-ray energies, of the characteristic X-rays are small. Another problem
for a certain element. These factors can be stored for with heterogeneous samples is the profile of the X-ray
a special configuration of a given spectrometer and used beam itself. In the ideal case, the beam intensity should
for element quantification of the unknown samples. If be constant over the beam area and drop abruptly to
an unknown sample is considered to be thin for some zero at the edges of the beam. In reality, the beam
elements (Di = 1), the amounts of these elements can be intensity does not behave so well, at least not for the
readily evaluated by using Equation (3). If the sample is of radiation from X-ray tubes. Owing to the geometrical
intermediate or infinite thickness, the attenuation factors effects and different path lengths for different parts of
Di will have to be evaluated either from measurements the radiation from the tube to the sample, the beam
of mass thickness or from information contained in may not have a homogeneous intensity distribution.
the coherently and incoherently scattered radiation, as This has also been shown to be the case for secondary-
discussed below. target arrangements.(27) Thus, for laboratory instruments,
ENERGY DISPERSIVE, X-RAY FLUORESCENCE ANALYSIS 13

knowledge of intensity variations within the beam is spectrum from a biological sample, in which Mo has been
essential for heterogeneous samples. Alternatively, one used as secondary target, it is seen that the incoherently
can achieve good accuracy for heterogeneous samples scattered radiation is very pronounced. In a biological or
and beams by scanning the sample, so that effects due to other kind of light-element sample in which medium
sample and beam inhomogeneity can be leveled out. and heavy elements are present at trace levels, the
incoherently scattered radiation is essentially a measure
of the mass seen by the impinging beam. The argument
4.2.3 Methods for Quantification in Practice
for this is that for the light elements (except hydrogen),
Many samples can be regarded as ‘‘thin’’ for medium- e.g. carbon, nitrogen, and oxygen, which are the main
and high-Z elements, but in reality, few samples are thin constituents in most biological samples, the mass of an
with respect to light elements of low atomic number atom is closely proportional to its atomic number (Z).
Z (see Table 1). Thus, one will often have to evaluate Furthermore, the binding energies of the electrons in
the attenuation properties of the sample, at least for the light elements are small compared with the energy
some elements. Many methods have been developed to of the exciting radiation. Thus, the exciting radiation
assist quantification. Some of these use an experimental essentially sees a cloud of free electrons, and the
approach, whereas others build on the fundamental probability of incoherent (Compton) scattering will be
parameters of the physical processes involved, often proportional to the number of electrons, which, in turn,
in combination with knowledge after calibration of the is approximately proportional to the sample mass. This
instrumental performance. A few methods to illustrate approach is also valid for some elements of importance
the principles in quantitative evaluations for nonliquid in geochemical samples, such as Mg, Si, and Ca (ratios
samples are described. of relative atomic mass to atomic number: 2.03, 2.01,
The first method, which was one of the earliest, is 2.00) but less so for Na, Al, and K (ratios of whose
the emission–transmission method.(1,2) A precondition is relative atomic mass to atomic number are in the range
that the sample is of ‘‘intermediate thickness’’, meaning 2.09–2.06). For samples composed of heavy elements,
that the intensity of the characteristic X-rays grows with the ratios of atomic mass to atomic number will change
sample thickness, but it does not grow linearly. The owing to increased neutron excess in the atomic nuclei.
method is based on an independent measurement of Furthermore, the binding energy of the inner electrons
the factor exp[(µ(E0 ) csc ϕ1 + µ(Ei ) csc ϕ2 )m], which is increases and the free electron incoherent scattering
contained in the attenuation correction, Di . This factor, cross section is no longer such a good approximation.
which measures the sample attenuation, can be directly Secondary scattering effects in the material may also
evaluated from three measurements: appear. Several practical methods, however, have been
developed, which make use of the information contained
1. First the characteristic radiation from an element ‘‘i’’ in both coherently and incoherently scattered radiation
in the sample is measured. for thick samples of geochemical origin. Giauque et al.(28)
2. Under identical conditions, the same characteristic and Giauque(29) have used the ratios of the incoherently
radiation is measured from the combination of the scattered radiation from Ag Kα and Kβ radiation together
sample and a target of element ‘‘i’’ positioned behind with the coherently scattered Ag Kα to ascertain sample
the sample. mass thickness and photoelectric cross sections for thick
3. In the third measurement, the characteristic radiation geochemical samples. Also, in the case of WDXRF,
is measured from the target of element ‘‘i’’ alone, the scattered Compton lines from the X-ray tube have
without the sample. proved to be useful as internal standards in light-element
samples.(30)
It is not difficult to show that these measurements The general aim of quantification in EDXRF is to
will give knowledge of the factor exp [(µ(E0 ) csc ϕ1 + arrive at the sample composition from the information
µ(Ei ) csc ϕ2 )m], which is needed in Equation (5). If this contained in the spectrum. In an XRF spectrum, the
factor is known, it is straightforward to calculate mi of medium-heavy and heavy elements in the sample will
element ‘‘i’’ from measurement of Ii and knowledge of the reveal their presence by their characteristic radiation
instrumental factors Ki and I0 according to Equation (3). as well as by their contribution to the coherently and
Another well-known method is to make use of the incoherently scattered radiation. Thus, a first estimate
information contained in the coherent and incoherent of the relative abundance of these elements can be made
scattering of the incident (exciting) radiation. Under directly from the spectrum. As discussed earlier, however,
certain conditions, the intensity of the scattering peaks this first estimate will be significantly modified by
or combinations of these can be used as internal considering attenuation and interelement effects, etc. To
standards.(1,2,19 – 21) From Figure 2, which is an X-ray evaluate the mass thickness of the light elements – which
14 X-RAY SPECTROMETRY

do not show up by their characteristic X-rays – their 5 CONCLUSIONS: NEEDS FOR FUTURE
contribution to the scattering peaks can be determined DEVELOPMENTS
from the spectrum by subtraction of the scattering due
to the medium-heavy and heavy elements. From this Although EDXRF has many advantages, in that it is
evaluation, the total mass thickness from light as well as multielemental, nondestructive, and needs a minimum
medium and heavy elements can be calculated. amount of sample treatment, it has two distinct disadvan-
In an iterative process, in which all the information tages:
from the elements through their characteristic as well
as scattered X-rays is used, the concentrations can be • For bulk samples it is not very sensitive. MDLs for
recalculated until the concentration values obtained agree laboratory equipment are in the order of 0.1 ppm.
with the experimental spectrum. Measured in mass units, it has MDLs of the order
In the practical situation, the analyst is dependent on of a fraction of a nanogram for the best cases in
computer programs in which an X-ray library is stored realistic samples. For synchrotron radiation sources
together with information on the actual spectrometer and special designs (TXRF; grazing emission X-ray
(obtained by calibration samples). Spectrum evaluation fluorescence (GEXRF)), MDLs may be reduced by
programs are available in commercial instruments several orders of magnitude, but not for all kinds of
and from research groups. A more-or-less continuous samples. A substantial part of the lack of sensitivity for
development of these programs takes place, in which bulk samples is due to effects caused by the presence
peak shapes and background profiles are described by of the large scattered peaks entering the solid-state
model functions.(31) detectors.
• In comparison with microprobes with charged parti-
4.3 Surface Analysis and Microanalysis cles – electrons or protons – microprobes of X-rays
with high lateral resolution are difficult to obtain. This
Quantification of thin surface layers or microstructures is because charged particles can be steered by electric
essentially makes use of the same fundamental methods and magnetic fields, whereas X-rays will have to be
as discussed above, namely, adding of an internal focused with mirrors or other kinds of optical devices.
standard for liquids, use of external reference material, The advantage of X-rays over charged particles is that
and use of scattered radiation. For thin self-supporting they deposit less energy into the sample and thus will
homogeneous solid surface layers, quantification can cause less heating and loss of volatile material.
be achieved by use of scattered radiation, calibration
constants, or fundamental parameters. Within a few years, EDXRF has mutated into a
In microelectronics, layers and implants need to be very versatile analytical tool owing to the development
prepared in the nanometer region. In these cases, the Si of sources, optics, and detectors. The applicability
wafer, as the basic material, is the best substrate for TXRF of EDXRF has been drastically enhanced in various
in order to use the angle-dependent fluorescence signal fields. Quantification is no longer a big problem
of the layer or implant to derive information on layer since the knowledge of primary X-ray spectra and
thickness, element and density – in case of implants the fundamental parameters has drastically improved. The
concentration profile can be derived by angle scanning improvement in computing capability and speed has led
from 0° to 2–3 times the critical angle. This technique to faster quantification and data evaluation. Software
is sensitive in the nanometer range and allows excellent packages are available as freeware, performing spectrum
insight to the new design of ultrashallow junctions. Both, deconvolution to obtain net counts for each element of the
laboratory sources and synchrotron radiation can be sample and also solving the rather complicated equations
used for these applications, called grazing incident X-ray related to all interaction processes rapidly in seconds.
fluorescence (GIXRF).(32) The elemental mapping capacity (2D and 3D) with
In the case of heterogeneous microstructures smaller high power sources opens new applications in medicine,
than the X-ray beam diameter, quantification by means biology, environment, and cultural heritage investiga-
of external references is difficult in spectrometers based tions, where quantification is not the most relevant
on X-ray tubes owing to the inhomogeneous beam profile information.
as discussed above, and the difficulty in positioning EDXRF has reached a degree of maturity with regard
the samples within the beam. In these cases, internal to accuracy and precision of the measurements. New
standards such as the scattering peaks from the sample will sources – X-ray optics and various excitation geometries,
give reasonably good average values, although structures like TXRF, GIXRF, and GEXRF – have shown a wide
with resolution below 10 µm are presently difficult to field of applications in layer and implant characterization.
study with laboratory equipment.(33) In combination with high-resolution single crystals
ENERGY DISPERSIVE, X-RAY FLUORESCENCE ANALYSIS 15

in double-crystal arrangements, energy increments of p–i–n Positive–Intrinsic–Negative


100 meV are possible; this can be used to study the SDD Silicon Drift Detectors
chemical state by measuring the fluorescence signal SR-TXRF Synchrotron Radiation Total Reflection
in dependence of the incident photon energy around X-ray Fluorescence
the absorption edge of the element considered. X-ray SSRL Stanford Synchrotron Radiation
absorption near edge spectroscopy (XANES) is used to Laboratory
determine the chemical state of the element, extended TM Trade Mark
X-ray absorption fine structure (EXAFS) is used to TXRF Total Reflection X-ray Fluorescence
determine the coordination number and the distance to WDXRF Wavelength-Dispersive X-ray
the neighboring atoms. Fluorescence
Further developments of X-ray sources like free XANES X-ray Absorption Near Edge
electron lasers X-ray free electron laser (XFEL) and Spectroscopy
Linac coherent light source (LCLS) will be soon XFEL X-ray Free Electron Laser
available for the scientific community and will require
the development of ultrafast detector technology and
electronics to make use of the high brilliant sources for
EDXRS. RELATED ARTICLES
Future development, which would imply break-
throughs on the above items, would be Environment: Water and Waste (Volume 4)
Sample Preparation for Elemental Analysis of Biological
• improvements in detector technology to minimize Samples in the Environment • X-ray Fluorescence
effects that cause increased background, for instance Spectroscopic Analysis of Liquid Environmental Samples
trapping and polarization in the detector material;
• development of laboratory X-ray lasers in order to Forensic Science (Volume 5)
obtain parallel, polarized, and intense beams. X-ray Fluorescence in Forensic Science
For both these developments it is easy to observe that
not only EDXRF but also a vast field of other analytical Peptides and Proteins (Volume 7)
methods would benefit from them. X-ray Crystallography of Biological Macromolecules

Steel and Related Materials (Volume 10)


ACKNOWLEDGMENTS X-ray Fluorescence Spectrometry in the Iron and Steel
Industry
One of the authors, E. Selin Lindgren, is grateful to
Ing. Paulis Standzenieks for assistance in preparing the Nuclear Methods (Volume 14)
manuscript. Particle-induced X-ray Emission (PIXE)

Radiochemical Methods (Volume 14)


ABBREVIATIONS AND ACRONYMS Nuclear Detection Methods and Instrumentation

X-ray Spectrometry (Volume 15)


ADC Analog-to-Digital Converter
Portable and Handheld Systems for Energy-dispersive
DESY Deutsches Elektronensynchrotron
X-ray Fluorescence Analysis • X-Ray Fluorescence Anal-
DSP Digital Signal Processing
ysis, Sample Preparation For • Total Reflection X-ray
EDXRF Energy-Dispersive X-ray Fluorescence
Fluorescence • Wavelength-dispersive X-ray Fluores-
ED Energy-Dispersive
cence Analysis • X-ray Techniques: Overview
EXAFS Extended X-ray Absorption Fine
Structure
GEXRF Grazing Emission X-ray Fluorescence
GIXRF Grazing Incident X-ray Fluorescence REFERENCES
LCLS Linac Coherent Light Source
MDL Minimum Detection Limits 1. R. Jenkins, R.W. Gould, D. Gedcke, Quantitative X-ray
NIST National Institute of Standards and Spectrometry, Marcel Dekker, New York, 1981.
Technology 2. E.P. Bertin, Principles and Practise of X-ray Spectrometric
NSLS National Synchrotron Light Source Analysis, Plenum Press, New York, 1975.
16 X-RAY SPECTROMETRY

3. R.E. Van Grieken, A.A. Markowics (eds), Handbook of Dispersive Spectrometer’, Nucl. Instrum. Methods, 165,
X-ray Spectrometry, 2nd edition, Marcel Dekker, New 63–65 (1979).
York, 2002.
18. P. Lechner, C. Fiorini, R. Hartmann, J. Kemmer, N. Kra-
4. B. Beckhoff, B. Kanngießer, N. Langhoff, R. Wedell, use, P. Leutenegger, A. Longoni, H. Soltau, D. Stötter,
H. Wolff (eds), Handbook of practical X-ray Fluorescence R. Stötter, L. Strüder, U. Weber, Silicon drift detectors for
analysis, Springer, Berlin, Heidelberg, New York, 2006. high count rate X-ray spectroscopy at room temperature,
5. A.G. Karydas, T. Paradellis, ‘Calculation of the Enhan- Nucl. Instrum. Methods A, 458, 281–287, (2001).
cement of Characteristic Radiation Due to Scattering 19. J. Nissenbaum, A. Holzer, M. Roth, M. Schieber, ‘X-ray
in Radioisotope XRF Using Exact Differential Cross- Fluorescence Analysis of High Z Materials with Mercuric
sections’, X-ray Spectrom., 22(4), 208–215 (1993).
Iodide Room Temperature Detectors’, Adv. X-ray Anal.,
6. R.P. Pettersson, E. Selin Lindgren, ‘Energy-dispersive X- 24, 303–309 (1980).
ray Fluorescence Analysis’, in Surface Characterization,
20. P. Jalas, H. Sipila, S. Eranen, K. Leinonen, ‘Near Room
A User’s Sourcebook, eds. D. Brune, R. Hellborg, H.J.
Temperature Energy-dispersive Detectors Processed on
Whitlow, O. Hunderi, Wiley-VCH, Weinheim, 136–153,
1997. High-purity Silicon’, X-ray Spectrom., 24(2), 63–69 (1995).

7. P. Wobrauschek, Total reflection x-ray fluorescence 21. D.S. McGregor, H. Hermon, ‘Room-temperature Com-
analysis – a review, X-ray Spectrom., 36, 289–300 (2007). pound Semiconductor Radiation Detectors’, Nucl.
Instrum. Methods, 295, 101–124 (1997).
8. C. Streli, P. Wobrauschek, F. Meirer, G. Pepponi, Total
reflection X-ray fluorescence and grazing incidence X-ray 22. www.canberra.com, www.ketek.net, 2009.
spectrometry – Tools for micro- and surface analysis. A 23. B. Kanngießer, W. Malzer, I. Reiche, ‘A New 3D Micro
review, J. Anal. Atom. Spectr., 23, 792–798 (2008). X-ray Fluorescence Analysis Set-up- First Archaeometric
9. www.moxtek.com, 2009. Applications’, Nucl. Instrum. Methods, B 211/2, 259–264
10. www.panalytical.com, www.rigaku.com, www.spectro. (2003).
com, www.edax.com, www.oxford-instruments.com, 24. Ph. Potts, M. West (eds), Portable X-ray Fluorescence
www.thermofisher.com, 2009. Spectrometry, RCS Publishing, 2008.
11. homepage of oxford instruments: www.oxfordxtg.com, 25. K. Tsuji, J. Injuk, R. Van Grieken, X-ray Spectrometry
2009.
Recent Technological Advances, Wiley, 2004.
12. V. Valkovic, A. Markowics, N. Haselberger, ‘Review of
26. B. Holynska, ‘Sampling and Sample Preparation in
Recent Applications of Radioisotope Excited X-ray
EDXRS’, X-ray Spectrom., 22(4), 192–198 (1993).
Fluorescence’, X-ray Spectrom., 22(4), 199–207 (1993).
13. K.W. Jones, W.M. Kwiatek, B.M. Gordon, A.L. Han-son, 27. J. Boman, P. Standzenieks, E. Selin, ‘Beam Profile Studies
J.G. Pounds, M.L. Rivers, S.R. Sutton, A.C. Thompson, in an X-ray Fluorescence Spectrometer’, X-ray Spectrom.,
J.H. Underwood, R.D. Giauque, Y. Wu, ‘X-ray Micros- 20(6), 337–341 (1991).
copy Using Collimated and Focussed Synchrotron 28. R.D. Giauque, ‘A Novel Method to Ascertain Sample
Radiation’, Adv. X-ray Anal., 31, 59–68 (1987). Mass Thickness and Matrix Effects for X-ray Fluorescence
14. C. Streli, G. Pepponi, P. Wobrauschek, C. Jokubonis, Element Determinations’, X-ray Spectrom., 23(4),
G. Falkenberg, G. Zaray, J. Broekaert, U. Fittschen, 160–168 (1994).
B. Peschel, Recent results of synchrotron radiation 29. R.D. Giauque, F. Asaro, F.H. Stross, T.R. Hester, ‘High
induced total reflection X-ray fluorescence analysis
Precision Non-destructive X-ray Fluorescence Method
at HASYLAB Beamline L, Spectrochim. Acta, 61B,
Applicable to Establishing the Provenance of Obsidian
1129–1134 (2006).
Artifacts’, X-ray Spectrom., 22(1), 44–53 (1993).
15. C. Streli, G. Pepponi, P. Wobrauschek, N. Zöger,
30. H.A. van Sprang, M.H.J. Bekkers, ‘Determination of Light
P. Pianetta, K. Baur, S. Pahlke, L. Fabry, C. Mantler,
Elements Using X-ray Spectrometry. Part 1 – Analytical
B. Kanngiesser, W. Malzer, Analysis of low Z elements
on Si wafer surfaces with synchrotron radiation induced Implications of Using Scattered Tube Lines’, X-ray
total reflection X-ray fluorescence at SSRL, Beamline Spectrom., 27(1), 31–36 (1997).
3–3: comparison of droplets with spin coated wafers, 31. B. Vekemans, K. Janssens, L. Vincze, F. Adams, P. Van
Spectrochim. Acta, 58B, 2105–2112 (2003). Espen, ‘Analysis of X-ray Spectra by Iterative Least
16. K. Janssens, F. Adams, A. Rindby (eds), Microscopic X- Squares (AXIL): New Developments’, X-ray Spectrom.,
ray Fluorescence Analysis, Wiley, 2000. 23(6), 278–285 (1994).
17. P. Standzenieks, E. Selin, ‘Background Reduction of X- 32. G. Pepponi, C. Streli, P. Wobrauschek, N. Zoeger,
ray Fluorescence Spectra in a Secondary Target Energy K. Luening, P. Pianetta, D. Giubertoni, M. Barozzi,
ENERGY DISPERSIVE, X-RAY FLUORESCENCE ANALYSIS 17

M. Bersani, Nondestructive dose determination and 33. R.W. Ryon, H.E. Martz, J.M. Hernandez, J.J. Haskins,
depth profiling of arsenic ultrashallow junctions with E.J. Suoninen, ‘Selection of the Method’, in Surface
total reflection X-ray fluorescence analysis compared to Characterization, A User’s Sourcebook, eds. D. Brune,
dynamic secondary ion mass spectrometry, Spectrochim. R. Hellborg, H.J. Whitlow, O. Hunderi, Wiley-VCH,
Acta B, 59, 1243–1249 (2004). Weinheim, 14–31, 1997.

You might also like