You are on page 1of 2
QUALITY GLOBAL CALIBRATION AND INSPECTION SERVICES ‘AERB RECOGNISED CALIBRATION LABORATORY FOR RADIATION MONITORING INSTRUMENTS. NABL ACCREDITED CALIBRATION LABORATORY AS PER ISO/IEC 17025:2017 Cc: Date of Issue:03/06/2021 Certificate No.RT-1415 Customer Details : M/S RUBY HALL CLINIC 40-Sasoon Road,Pune-411001 Instrument Type : Radiation Survey Meter Instrument Manufacturer + Rotem-GM. Model : RAM GENE-MK2 Serial Number(s) 2 12818-097 Instrument Received on + 25/05/2021 Environmental Conditions P25i3°C ra Calibration Date + 03/06/2021 Calibration Due Date + 02/06/2023 . *Standard reference value was measured with the help ofa 100ccyeference graphite ionisation chamberin conjuction with a electrometer, traceable to Radiation Standards Section, RSSD, Bhabha Atomic Research Centre, Mumbai, (Certificate No. - BARC/RSSD/RSS/CAL/C-120/2017) Method of Calibration + Placing Radiation Survey Meter in front of collimated beam from Co-60 source and checking reading at various distances & durations. # This certificate shall not be reproduced in full without the approval of the calibration laboratory Regd. office : Plot NO. B.G.A.-1/4, Bhosari General Block MIDC, Bhosari, Pune - 411026. Maharashtra Contact no: (020) 27121843, 27119490, 8600100704 E-mail : tanmaypangare@gmail.com, bkp_gndt@yahoo.in visit us at : www. qualtyndt.org Page 1 of 2