You are on page 1of 37

AIRVO™ 2

Technical Manual
This page has intentionally been left
blank.
BEFORE YOU START
This Technical Manual is intended for clinical engineering / technical personnel. It defines the technical
specifications, setup, servicing and troubleshooting information, for the AIRVO 2 humidifier. It applies to all
lots manufactured after 2016-06-06.

OTHER REFERENCES
• Refer to the AIRVO 2 User Manual for detailed instructions for use.

• Watch the training videos on the AIRVO website www.fphcare.com/airvo.

• Download the AIRVO 2 Simulator App to learn how to use the AIRVO 2.
You can change settings, simulate faults and test your skills. Available
from the Apple, Google Play and Windows App stores.

• Visit the Fisher & Paykel education & resources website www.fphcare.com/education to find self-paced
online courses and local training events.

• If the unit is ever used by multiple patients, the unit must be cleaned and disinfected between patients
according to instructions in the Disinfection Kit Manual (900PT600).

• For further assistance, please contact your Fisher & Paykel Healthcare representative.

TABLE OF CONTENTS
1. General information .................................................................................................................................. 4
Package contents ........................................................................................................................................................................ 4

AIRVO 2 and accessories ......................................................................................................................................................... 5

2. Setting up AIRVO 2 for first use .......................................................................................................... 6


Advanced settings ...................................................................................................................................................................... 8

3. Acceptance/performance checks ....................................................................................................... 14


Acceptance/performance checks form..........................................................................................................16

4. Servicing ....................................................................................................................................................... 17
5. Spare Parts ................................................................................................................................................... 18
Appendix A: IEC 60601-1-2 EMC tables ................................................................................................. 21
Appendix B: User interface flow charts ................................................................................................. 23
Appendix C: Default values ......................................................................................................................... 25
Appendix D: Troubleshooting Guide ....................................................................................................... 26
Appendix E: Tube and Chamber Kit labels ........................................................................................... 32
Appendix F: Error Flowchart ...................................................................................................................... 33
Appendix G: Transport (Low Power) Mode User Instructions........................................................ 35

3
1. GENERAL INFORMATION
The AIRVO 2 is a humidifier with integrated flow generator that delivers warmed and humidified respiratory
gases to spontaneously breathing patients through a variety of patient interfaces.

PACKAGE CONTENTS

AIRVO 2 humidifier AIRVO 2 User Manual AIRVO 2 Swingtag


(PT101xx)

Oxygen inlet extension kit Disinfection Kit Air filter (x2) Power cord
(900PT422) (900PT600) (900PT913) (900PT410xx)

WARNING
UNDER NO CIRCUMSTANCES SHOULD THE
AIRVO 2 BE OPENED OR ANY OF THE SIX
FASTENING SCREWS ON THE UNDERNEATH
SIDE OF THE DEVICE BE LOOSENED.

OPENING THE UNIT WILL AFFECT THE


OXYGEN SEALS INSTALLED INSIDE, WHICH
WILL COMPROMISE THE SAFETY OF THE
DEVICE.

4
AIRVO 2 AND ACCESSORIES

MUTE ON/OFF (STANDBY)


Heated
breathing
Patient
tube
interface
UP
DISPLAY
DOWN

MODE

HEATED BREATHING TUBE


OXYGEN
CONNECTION PORT
INLET PORT
HOSPITAL
STAND MEASUREMENT POINT OF
POLE DISPLAYED DEW POINT
MOUNTING TEMPERATURE
TRAY
CHAMBER PORTS

Water chamber

SERIAL PORT

POWER CORD FILTER COVER FINGER


and HEATER GUARD
CONNECTOR PLATE

AIRVO 2
(PT101xx)

AUTO-FILL WATER
CHAMBER (MR290)
AIR FILTER (with adapter fitted)

Note: Some accessories may not be available in all countries.


Please contact your local Fisher & Paykel Healthcare representative for more information.

5
2. SETTING UP AIRVO 2 FOR FIRST USE
1. REMOVE THE AIRVO 2 FROM ITS PACKAGING
Place the AIRVO 2 on the 900PT405 pole mounting tray, on the
900PT421 hospital stand.

2. CONNECT THE POWER CORD


Plug the power cord connector into the socket on the back of the
AIRVO 2.

3. ATTACH THE OXYGEN INLET EXTENSION KIT


Refer to the instruction sheet included with the kit itself.

4. ATTACH WATER CHAMBER AND HEATED BREATHING TUBE


The water chamber and heated breathing tube must be connected to
carry out the following setup and testing procedures.
If you have not been supplied with a reusable HC360 water chamber, you
can use an MR290 chamber instead.

6
5. SWITCH ON UNIT
Switch on the unit by pressing the On/Off button.

6. WARM-UP
The unit will begin to warm up.

“Warm-up” symbol

7. READY FOR USE


The “Ready for use” symbol means that the system is ready for the
patient to use.

“Ready for use” symbol

7
ADVANCED SETTINGS
When you see the “Warm-up” or “Ready for use” symbols, hold a
combination of three buttons (Up, Down and Mute) for 5 seconds, to
view and change advanced settings.
This button combination is for use by clinical engineering / technical
personnel only.

AIRVO 2 / myAIRVO 2 MODE


You can change the unit from “AIRVO” (hospital) mode to “myAIRVO”
(home / long-term care) mode, eg. for patients going home.
Contact Fisher & Paykel Healthcare for a myAIRVO 2 User Manual.

To change the mode:

Hold the Up and Down buttons for 3 seconds to


“unlock” the setting.

Use the Up button to select myAIRVO 2.


myAIRVO 2

Press the Mode button to confirm the change and/or move on to the
next screen. Note that the unit will reset itself if it is switched between
AIRVO 2 and myAIRVO 2 modes.

LANGUAGE
You can set the AIRVO 2 / myAIRVO 2 to one of 22 language settings:
English Nederlands Svenska Polski pl ‫ةيبرعلا‬ ar
Deutsch Português 简体中文 [simp.] Русский ru Türkçe tr
Español Dansk 繁体中文 [trad.] ‫תירבע‬ he
Français Suomi Ελληνικά el
Italiano Norsk 日本語 Română ro

To change the language:

Hold the Up and Down buttons for 3 seconds to


“unlock” the setting.

Use the Up and Down buttons to select the desired


language.

Press the Mode button to confirm the change and/or move on to the
next screen.

8
ENVIRONMENT SETTINGS (FOR DEFAULT MODE)
A clinician may change the “Environment Settings”, to customise
individual AIRVOs for different environments (eg. intensive care, general
care areas, emergency departments). The “Environment Settings”
37 60 95
chosen will put limits on the “Patient Settings” that the operator can
choose when in normal use.
31 10 21 This screen defines the “Environment Settings” for the AIRVO 2 when in
Default Mode (ie. non-“Junior Mode”).

Minimum dew-point The lowest target dew-point temperature that


temperature (°C) the operator will be able to select.
Possible Settings: 31, 34, 37 °C
If this is set to 31, the operator can select a target
dew-point temperature between 31 and 37. ie. 31,
34 or 37 (°C).
If the patient is tracheostomised, a clinician may
wish to set this value to 37, so that the operator
can only select a target dew-point temperature
between 37 and 37, ie. only 37 (°C).
Note: The maximum dew-point temperature
setting is always 37 °C in Default Mode.
Minimum flow (L/min) The lowest flow that the operator will be able to
select.
Possible Settings: 10 to 60 in increments of 5 L/min,
always less than or equal to Maximum Flow setting.
Example: If this is set to 10, the operator will be
able to select flows down to 10 L/min.
If this is set to 25, the operator will be able to
select flows down to 25 L/min.
Maximum flow (L/min) The highest flow that the operator will be able
to select.
Possible Settings: 10 to 60 in increments of 5 L/min,
always greater than or equal to Minimum Flow setting.
Example: If this is set to 60, the operator can
select flows up to 60 L/min.
If this is set to 35, the operator can select flows up
to 35 L/min.
Maximum oxygen fraction (%) The highest oxygen fraction that the operator
may set the unit to.
37 60 95 Possible settings: 30 - 100% in increments of 5% O2.
The unit will alarm if the measured oxygen
fraction exceeds this value.
31 10 21 Note: Even if this ‘Maximum oxygen fraction‘
setting is set to 100%, any time the measured
oxygen fraction exceeds 95%, the oxygen reading
will pulse red and the device will beep.
Minimum oxygen fraction (%) The lowest oxygen fraction that the operator
may set the unit to.
Possible settings: 21 or 25% O2.
When set to 25% the unit will alarm if the
measured oxygen fraction is below this value. This
allows detection of oxygen being disconnected.
Note that, for Oxygen display, this is a measurement only, not a control setting. The
operator changes the measured oxygen fraction by altering the AIRVO 2 target
flow setting and the flow of oxygen connected to the unit (e.g. from a flowmeter)
- there is no closed-loop control.

To change the environment settings:

Hold the Up and Down buttons for 3 seconds to


“unlock” the first setting.

Use the Up and Down buttons to change the setting,


37 60 95
then press the Mode button to progress to the next
setting.
31 10 21

Press the Mode button to confirm the change and/or move on to the
next screen.
9
ENVIRONMENT SETTINGS (FOR JUNIOR MODE)
This screen defines the “Environment Settings” for the AIRVO 2 when in
Junior Mode.
34 25 100
95 Junior Mode Enable/Disable When this option is enabled (default), the
operator can enter Junior Mode from the Home
Enabled
Screen, by holding the Mode button for 5
34 2 21 seconds.

When this option is disabled, entering Junior


Disabled mode is not possible.
Consider disabling this option if the unit will never
be used on pediatric patients.
Dew-point The only dew-point setting in Junior Mode is
temperature (°C) 34 °C.
Minimum flow (L/min) The lowest flow that the operator will be able to
select.
Possible Settings: 2 to 25 in increments of 1 L/min,
always less than or equal to Maximum Flow setting.
If this is set to 10, the operator will be able to
select flows down to 10 L/min.
Maximum flow (L/min) The highest flow that the operator will be able
to select.
Possible Settings: 2 to 25 in increments of 1 L/min,
always greater than or equal to Minimum Flow setting.
If this is set to 15, the operator can select flows up
to 15 L/min.
Maximum oxygen fraction (%) The highest oxygen fraction that the operator
may set the unit to.
37 60 95 Possible settings: 30 - 100% in increments of 5% O2.
The unit will alarm if the measured oxygen
fraction exceeds this value.
31 10 21 Note: Even if this ‘Maximum oxygen fraction‘
setting is set to 100%, any time the measured
oxygen fraction exceeds 95%, the oxygen reading
will pulse red and the device will beep.
Minimum oxygen fraction (%) The lowest oxygen fraction that the operator
may set the unit to.
Possible settings: 21 or 25% O2.
When set to 25% the unit will alarm if the
measured oxygen fraction is below this value. This
allows detection of oxygen being disconnected.
Note that, for Oxygen display, this is a measurement only, not a control setting. The
operator changes the measured oxygen fraction by altering the AIRVO 2 target
flow setting and the flow of oxygen connected to the unit (e.g. from a flowmeter)
- there is no closed-loop control.

To change the environment settings:

Hold the Up and Down buttons for 3 seconds to


“unlock” the first setting.

Use the Up and Down buttons to change the setting,


then press the Mode button to progress to the next
setting.

Press the Mode button to confirm the change and/or move on to the
next screen..

10
HOSPITAL LOCKS ENABLED/DISABLED SETTING
This screen defines the “Hospital Locks Enabled/Disabled Settings” for
the AIRVO 2 when in either Default Mode or Junior Mode.
By changing this setting, you can determine whether the temperature
lock in the Target Temperature screen and the flow lock in the Target
Flow screen are enabled or disabled.

To change the hospital locks settings:

Hold the Up and Down buttons for 3 seconds to


“unlock” the setting.

Use the Up and Down buttons to change the setting.

Press the Mode button to confirm the change and/or move on to the
next screen.

11
FLOW INCREMENT SETTINGS
This screen defines the “Flow Increment Settings” for the AIRVO 2 when
in either Default Mode or Junior Mode.
You can define the flow rate above which the increments are 5 L/min
and below which the increments are 1 L/min.
25

To change the environment settings:

Hold the Up and Down buttons for 3 seconds to


“unlock” the setting.

Use the Up and Down buttons to change the setting.


25 Flows > 25 L/min will increment in steps of 5 L/min
Flows < 25 L/min will increment in steps of 1 L/min

Press the Mode button to confirm the change and/or move on to the
next screen.

DISINFECTION STOP-GATE SETTINGS


This screen defines the “Disinfection Stop-Gate Settings” for the
AIRVO 2 when in either Default Mode or Junior Mode.

Confirmation required
At startup, disinfection number and time
information is shown, confirmation from the
operator is required before progressing to
warmup

Confirmation not required (default)


At startup, disinfection number and time
information is shown before progressing to
warmup

To change the environment settings:

Hold the Up and Down buttons for 3 seconds to


“unlock” the setting.

Use the Up and Down buttons to change the setting.

Press the Mode button to confirm the change and/or move on to the
next screen.

12
OXYGEN INPUT SETTINGS
This screen defines the “Oxygen Input Settings” for the AIRVO 2 when
in either Default Mode or Junior Mode.
The 100% setting is for hospital oxygen supplies, liquid oxygen or
standard bottled oxygen and is the default setting for AIRVO 2.
The 95% setting is for use with oxygen concentrators.

To change the environment settings:

Hold the Up and Down buttons for 3 seconds to


“unlock” the setting.

Use the Up and Down buttons to change the setting.

Press the Mode button to confirm the change and/or move on to the
next screen.

TRANSPORT (LOW POWER) MODE SETTINGS


This screen defines the “Transport (Low Power) Mode Settings” for the
AIRVO 2 when in either Default Mode or Junior Mode.

Transport (Low Power) Mode enabled


Transport (Low Power) mode may be activated by
the operator. (see “Transport (Low Power) mode”
user instruction in Appendix G).

Transport (Low Power) Mode disabled (default)


Transport (Low Power) mode may not be
activated by the operator..

To change the environment settings:

Hold the Up and Down buttons for 3 seconds to


“unlock” the setting.

Use the Up and Down buttons to change the setting.

Press the Mode button to return to the “Warm-up”/”Ready for use”


screen. You can now conduct the Performance/Acceptance checks..

13
3. ACCEPTANCE/PERFORMANCE CHECKS
This section contains performance checks which can be carried out on the AIRVO 2, however there is
no manufacturer requirement to carry out these checks on a routine basis. These checks test the basic
functions of the unit, the operation of the flow sensor and the audible alarm signal.
The Acceptance/Performance Checks Form on page 16 can be used to record the result from these checks.

SENSOR CHECKS

To ensure quality and patient safety, Fisher & Paykel Healthcare undertakes stringent testing to each
and every unit manufactured. The AIRVO 2’s sensors, measuring temperature, flow and oxygen, have
been carefully designed to exacting criteria, and are calibrated and tested in our controlled work
environment to strict limits. Furthermore, the AIRVO 2 itself carries out regular self-checks during
normal use, comparing sensor readings against expected values.

Do not perform additional tests of internal sensor accuracy, as these often produce erroneous
results due to limitations of the external test environment and equipment used, particularly given the
temperature, humidity content and/or flow of the gases being delivered by the AIRVO 2.

ACCEPTANCE/PERFORMANCE CHECKS
The acceptance/performance checks should be conducted under the following ambient conditions:
Temperature: 22 ± 2°C, Humidity: 50 ± 5% RH.

The following equipment is required:

AIRVO 2 humidifier Heated breathing tube and chamber kit Nasal interface
(from 900PT501 or 900PT551*/561* kit) (OPT842/44/46 or
OPT942/44/46)

*900PT551/900PT561 kits are not available in USA

A. HEATERPLATE TEST
1. Add 150 mL of room temperature (not hot) water to the water chamber and fit the chamber onto the
heater plate of the device. Fit the chamber tightly on to the chamber ports.
2. Connect the heated breathing tube to the Heated Breathing Tube Connection port. Connect the nasal
cannula interface to the heated breathing tube.
3. Turn on the device, by pressing the power button for 2 seconds. Warm-up bars will be displayed as the
unit warms up. Ensure the flow is set to 30 L/min.
4. Check that the “Ready for use” symbol (a “tick” or “check”) is displayed within 30 minutes.

B. “CHECK FOR LEAKS” TEST


After the “Ready for use” symbol is displayed, the “Check for leaks” sensor test can be tested as follows:
1. Remove the chamber completely from the unit.
2. Check that the display shows the “Check for leaks” error (in the appropriate language) and that the
audible alarm sounds, within 2 minutes.
3. Reconnect the chamber and check that this flashing display disappears, the audible alarm stops and the
display reverts back to the Warm-up/Ready-for-use screen.

14
C. “CHECK FOR BLOCKAGES” TEST
After completing the “Check for leaks” test, the “Check for blockages” test can be tested as follows:
1. Disconnect the cannula from the Heated Breathing Tube.
2. Completely block the end of the Heated Breathing Tube with your hand.
3. Check that the display shows the “Check for blockages” error (in the appropriate language) and that the
audible alarm sounds, within 30 seconds.
4. Unblock the end of tube and check that this flashing display disappears, the audible alarm stops and the
display reverts back to the previous display.
5. Reconnect the cannula to the Heated Breathing Tube.

D. “CHECK TUBE” TEST


After completing the above flow tests, the Tube Missing alarm can be tested as follows:
1. Remove the Heated Breathing Tube (pull the blue sleeve up first).
2. Check that within 10 seconds the display flashes the Breathing Tube symbol and the alarm sounds.
3. Refit the Heated Breathing Tube, check the alarm stops and that the display reverts back to the previous
display.

E. “POWER OUT” TEST


After the “Ready for use” symbol is displayed, the “Power Out” test can be tested as follows:
1. Pull out the power plug from mains power.
2. Check that the audible alarm sounds for at least 120 seconds.

Note: If any of the tests above fail, do not place the unit into service and please contact your F&P
Healthcare representative.

15
AIRVO™ 2 ACCEPTANCE/PERFORMANCE CHECKS FORM
This document should be used in conjunction with the AIRVO 2 technical manual. Results from the acceptance/performance checks can be recorded in the table
below. If any of the tests below fail, please contact your Fisher & Paykel Healthcare representative.

AIRVO 2 serial number: _________________________ Unit asset number: ________________________

Hospital/Dept.: ________________________________ Date (DD/MM/YYYY): _____ /______ / _________

Screen display Acceptance criteria Time Result

HEATERPLATE TEST < 30 minutes Pass/Fail

16
“CHECK FOR LEAKS” TEST
Check for
< 2 minutes Pass/Fail
Leaks

“CHECK FOR BLOCKAGES” TEST


Check for
< 30 seconds Pass/Fail
Blockages

“CHECK TUBE” TEST < 10 seconds Pass/Fail


Check Tube

­­­­­
“POWER OUT” TEST N/A > 120 seconds Pass/Fail

Test Engineer name: ____________________________ Signature: _____________________________


4. SERVICING
AIRVO 2 and myAIRVO 2 humidifiers do NOT require routine servicing or calibration.
The only checks that can be carried out are the Acceptance/Performance Checks in the previous section,
and the Electrical Safety Test detailed below.

WARNING
UNDER NO CIRCUMSTANCES SHOULD THE
AIRVO 2 BE OPENED OR ANY OF THE SIX
FASTENING SCREWS ON THE UNDERNEATH
SIDE OF THE DEVICE BE LOOSENED. OPENING
THE UNIT WILL AFFECT THE OXYGEN
SEALS INSTALLED INSIDE, WHICH WILL
COMPROMISE THE SAFETY OF THE DEVICE.

ELECTRICAL SAFETY TESTS


To test for electrical safety perform the following electrical safety tests and any others required by local
regulations.

Inspection Check the power cord for damage - cuts, stretching, wear, adequate cable restraint, bent
pins. Replace with F&P approved cord if necessary.
Insulation Resistance Use a 500 VDC insulation tester to measure the resistance between the mains plug phase pin
and the heaterplate* - it should be > 10 Mohm. Repeat test from the mains plug neutral pin to
the heaterplate*.
* Note: The exposed surface of the heaterplate is anodised (high resistance). Contact MUST
be made to the bottom lip of the heaterplate at the front of the device to make proper
connection - depress the finger guard and slip the tester probe beneath the heaterplate to
ensure contact to unanodised aluminium.

STORAGE AND DISPOSAL


Refer to AIRVO 2 User Manual.

17
5. SPARE PARTS
5.1 POWER CORDS

900PT410AZ (Aus/NZ)

900PT410EW (European)

900PT410UK (UK)

900PT410US (US/Canada)

900PT410KR (Korea)

900PT410JP (Japan)

900PT410ZH (China)

18
5.2 FOR AIRVO

900PT911
Non-Return Valve

900PT912
AIRVO 2 Filter Holder

900PT913
Air Filter (2-pack)

900PT422
Oxygen Inlet Kit

900PT407
Serial Port Cover (10-pack)

900PT408
AIRVO Outlet O-Ring (10-pack)

5.3 FOR DISINFECTION

900PT600
Disinfection Kit

900PT601
Disinfection Filter
(2-pack)

900PT602
Cleaning Sponge Stick
(20-pack)

900PT603
Disinfection Storage Cover
(20-pack)

19
5.4 HARDWARE

900PT405
Pole Mounting Tray

900PT421
Mobile Pole Stand /
900PT420
Mobile Pole Stand Extendable

900PT428
C-Clamp 25mm

900PT423
Hook for 900PT421

900PT426
Plastic Basket

900PT427/L
Oxygen Bottle Holder/Large

900PT411
UPS Mounting Kit

900PT409
3M Dual Lock Pad (4 pairs)

20
14 of 17
APPENDIX A: IEC60601-1-2 EMC TABLES
Appendix A: IEC60601-1-2 EMC tables
Guidance and manufacturer's declaration - electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should
ensure that it is used in such an environment.

Emissions test Compliance Electromagnetic environment - guidance


RF emissions The device uses RF energy only for its internal function. Therefore, its RF
Group 1 emissions are very low and are not likely to cause any interference in
CISPR 11 nearby electronic equipment.

RF emissions
Class B Applicable for countries with 100-115V and 220-240V mains voltage.
CISPR 11

Harmonic emissions
Class A
IEC 61000-3-2
The device is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
Voltage fluctuations/ power supply network that supplies buildings used for domestic purposes
flicker emissions
Complies
15 of 17
IEC61000-3-3

Guidance and manufacturer's declaration - electromagnetic immunity


The device is intended for use in the electromagnetic environment specified below. The customer or the user of device should ensure that it is used in
such an environment.

Immunity IEC60601 Electromagnetic environment


Compliance level
test test level - guidance
Electrostatic
discharge (ESD) ±2kV ,±4kV, ±6kV contact ±2 kV,±4kV, ±6 kV contact Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the
relative humidity should be at least 30%.
±2 kV ,±4kV, ±8 kV air ±2 kV ,±4kV, ±8 kV air
IEC61000-4-2

±2 kV for power
Electrical fast
transient/burst supply lines ±2 kV
Mains power quality should be that of a typical
±1 kV for input/output See note 2 below commercial or hospital environment.
IEC61000-4-4
lines

±1 kV differential mode ±1 kV
Surge
Mains power quality should be that of a typical
±2 kV common mode ±2 kV commercial or hospital environment.
IEC 61000-4-5

<5 % UT <5 % UT
(>95 % dip in UT) (>95 % dip in UT)
Voltage dips, short for 0.5 cycle for 0.5 cycle
interruptions and
voltage variations
on power supply 40 % UT 40 % UT
input lines (60 % dip in UT) (60 % dip in UT) Mains power quality should be that of a typical
for 5 cycles for 5 cycles commercial or hospital environment. If the user of
the device requires continued operation during
power interruptions, it is recommended the device
70 % UT 70 % UT be powered from an uninterruptible power supply
IEC 61000-4-11 (30 % dip in UT) (30 % dip in UT) or a battery.

for 25 cycles for 25 cycles

<5 % UT <5 % UT
(>95 % dip in UT) (>95 % dip in UT)
TM
for 5 sec
Airvo for Series
5 sec
Humidifiers
3 A/m
Models PT101 and PT100
3 A/m
Power frequency
(50/60 Hz)
Technical Manual Rev D (Part Number 185044926)
Power frequency magnetic fields should be at
magnetic field levels characteristic of a typical location in a typical
commercial or hospital environment.
IEC 61000-4-8

NOTE 1: UT is the a.c. mains voltage prior to application of the test level.

NOTE 2: This testing is not necessary for the safe operation of the device.

21
16 of 17

Guidance and manufacturer's declaration - electromagnetic immunity


The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should ensure that it is used in such
an environment.

IEC60601 test
Immunity test Compliance level Electromagnetic environment - guidance
level
Portable and mobile RF communications equipment should be
used no closer to any part of the device, including cables, than the
recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.

Recommended separation distance

Conducted RF 3 Vrms 3 Vrms d = 1.2 √P

IEC 61000-4-6 150 kHz to 80 MHz

Radiated RF 3 V/m 3 V/m d = 1.2 √P 80 MHz to 800 MHz

IEC 61000-4-3 80 MHz to 2,5 GHz

d = 2.3 √P 800 MHz to 2.5 GHz

where P is the maximum output power rating of the transmitter in


watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in metres (m).

Field strengths from fixed RF transmitters, as determined by an


electromagnetic site survey,ª should be less than the compliance
level in each frequency range b.

Interference may occur in the vicinity of equipment marked with


the following symbol:

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and
people.

a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the location in which the device is used exceeds the applicable RF
compliance level above, the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the device.

b 17 of 17
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Recommended separation distances between portable and mobile RF communications equipment


and the device

The device is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device
can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the device as recommended below, according to the maximum output power of the communications equipment.

Separation distance according to frequency of transmitter


AirvoTM Series Humidifiers
m
Models PT101 and PT100
Rated maximum output power Technical Manual Revto D
150 kHz 80 (Part
MHz Number 185044926)
80 MHz to 800 MHz 800 MHz to 2.5 GHz

of transmitter
W d = 1.2 √P d =1.2 √P d = 2.3 √P

0.01 0.12 0.12 0.23


0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23

For transmitters rated at maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation
applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.

NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and
people.

22
APPENDIX B: USER INTERFACE FLOW CHARTS

AIRVO 2

Heated breathing
tube detected

Last Last
Disinfection: Disinfection:
#4 #5

17 days 2 days 34 25 90 95%


14 hours 4 hours
ago ago
34 2 21

Hold
5s 37 60 90 60 100%
AIRVO 2 English 30
Junior Mode
V6.72 31 10 21 10

Hold 5 Hold Hold Hold Hold Hold Hold Hold Hold


3s 3s 3s 3s 3s 3s Hold 3s
s 3s 3s
New Target
Settings Press Press Press Press Press Press Press Press
37 60 90 34 25 90 100%
AIRVO 2 English 30
31 10 21 34 2 21

Transport
Mode

90
After 3 scrolls
Hold
21
5s

Press Press Press Press


90

21
onitor patient O2

Hold Hold
3s 3s

37 60

37 30
Wait for time out 120s 31 10

(180s with Hosp locks


disabled)

90
After 3 scrolls
Hold 21
3s
Press
Oxygen is Press Press Press
(Hosp
left on 60

30
37 90
locks 37
disabled) 31 10
onitor patient O2
21

Disconnect
Oxygen

Hold
5s
[Machine Data]

Hold
5s

Wait for time out 120s


Transport
(180s with Hosp locks
Mode
disabled)

Press

19:55
19:23

Hold
timer reaches 5s
00:00

Non-Transport
Mode

Press

23
FILTER CHANGE DUE

Any key
Wait for
press
time out
300s

Air Filter
change due Press Press
Replace now?

Now
Later
1000 hours use -
Replace filter
Press
Air Filter
change due
Replace now?

Now
Later

Press

Air Filter
change due
Replace now?

Now
Later

Filter timer remains Filter timer is reset


at 1000 hrs + to 0 hrs

Return

DISINFECTION

Disinfection tube detected

1000 hours use


hospital model
- see replace filter

Press Press

Wait for
time out 60s

24
APPENDIX C: DEFAULT VALUES
The following values are default settings set by the factory:

Parameter Model Mode Value Unit


Default set dewpoint AIRVO 2 or myAIRVO 2 Default 37 °C
temperature
AIRVO 2 or myAIRVO 2 Junior 34 °C
Max set dewpoint temperature AIRVO 2 or myAIRVO 2 Default 37 °C
AIRVO 2 or myAIRVO 2 Junior 34 °C
Min set dewpoint temperature AIRVO 2 or myAIRVO 2 Default 31 °C
AIRVO 2 or myAIRVO 2 Junior 34 °C
Default set flow AIRVO 2 Default 30 L/min
myAIRVO 2 Default 25 L/min
AIRVO 2 or myAIRVO 2 Junior 15 L/min
Max set flow AIRVO 2 or myAIRVO 2 Default 60 L/min
AIRVO 2 or myAIRVO 2 Junior 25 L/min
Min set flow AIRVO 2 or myAIRVO 2 Default 10 L/min
AIRVO 2 or myAIRVO 2 Junior 2 L/min
AIRVO 2 Default or Junior 95 %
Default upper oxygen limit
myAIRVO 2 Default or Junior 90 %
AIRVO 2 Default or Junior 100 %
Max upper oxygen limit
myAIRVO 2 Default or Junior 90 %
Min upper oxygen limit AIRVO 2 or myAIRVO 2 Default or Junior 30 %
Default lower oxygen limit AIRVO 2 or myAIRVO 2 Default or Junior 21 %
Max lower oxygen limit AIRVO 2 or myAIRVO 2 Default or Junior 25 %
Min lower oxygen limit AIRVO 2 or myAIRVO 2 Default or Junior 21 %
Language AIRVO 2 or myAIRVO 2 Default or Junior English
Hospital Locks Enabled/Disabled AIRVO 2 Default or Junior Enabled
Flow Increment Crossover AIRVO 2 or myAIRVO 2 Default or Junior 25 L/min
AIRVO 2 Default or Junior 100 %
Oxygen Input
myAIRVO 2 Default or Junior 95 %
Disinfection Stop-Gate AIRVO 2 Default or Junior Confirmation not required

Transport (Low Power) Mode AIRVO 2 Default or Junior Disabled


Day/Night mode myAIRVO 2 Default or Junior Day

25
APPENDIX D: TROUBLESHOOTING GUIDE

This Troubleshooting Guide is intended for technical users, including clinical/biomedical engineers and
technical personnel, of the AIRVO™ 2 humidifier. It applies to all AIRVO 2 humidifiers manufactured after
2016-06-06.
If this troubleshooting guide does not resolve your issue, please contact your local Fisher & Paykel
Healthcare representative.

D.1 AIRVO does not turn on

A. Press and hold the ON/OFF button for at least 2 seconds.

B. Is the AIRVO 2 plugged into mains power?

C. Is the power cord securely inserted into the back of the AIRVO 2?

D. Is the power cord damaged?


• If yes, replace the damaged cord. See Section 5.1 for a 900PT410xx replacement power cord.
• If no, try using another power cord.

E. Connect the AIRVO 2 into another power outlet.

F. Connect a different electrical device into the same power outlet. Turn on the device to confirm that the
power outlet is working.
G. The AIRVO 2 may be ‘on’ with a broken display.
Turn the AIRVO 2 on without the heated breathing tube and check that the audible alarm activates.

D.2 Power out (black screen)


The auditory alarm will sound for at least 120 seconds.

The most likely cause is a dislodged or disconnected power cord.


A. Please follow the instructions in Section D.1.
Note: Press “audio pause” button to permanently silence the alarm ( ).
The device will not automatically restart.

D.3 “Check tube”Fig. 1 or “E38“


A. Is the heated breathing tube attached correctly?
• Even if it appears to be, unplug and reconnect the heated breathing
tube.
B. Is the heated breathing tube visibly damaged?
• Check the electrical pins and the tube itself.
Figure 1
C. Try using a new heated breathing tube.

26
D.4 “Check for blockages”Fig. 2 or “E121”

D.4.1 WATER CHAMBER AND NON-RETURN VALVE


A. Have the silicone flaps of the non-return valve, found inside the left-hand
chamber port, been displacedFig. 3?
• If yes, return them to the correct position using a non-sharp tool, such Figure 2
as a pair of non-sharp tweezersFig. 5.
Note: If the Non-return valve is damaged or missing, replace with part
900PT911. Upon replacement, ensure the spine is sitting verticallyFig. 5.
If placed horizontally, this may cause the bottom flap to open due to
gravityFig. 4. This may cause both “Check for leaks” and “Check for blockages”
warnings. Figure 3 Figure 4
B. Is the MR290 water chamber overfilled above the black line?
• If yes, replace with a new water chamber. Contact your local Fisher &
Paykel Healthcare representative about the faulty chamber.

D.4.2 HEATED BREATHING TUBE


Figure 5
A. Is the heated breathing tube visibly blocked or kinkedFig. 6?

D.4.3 PATIENT INTERFACE AND AIRVO MODE

A. Is the patient interface visibly blocked or kinked?

B. Should the unit be in Junior modeFig. 7?


• If the AIRVO is in Default mode and the 900PT531 Junior tube or the
900PT561 AirSpiral™ tube is used with the OPT316 and OPT318 cannula
Figure 6
interfaces it may generate a “Check for blockages” alarm.
See Appendix E for identification of the Default and Junior Tube and
Chamber Kits, according to their labels.

C. Are you using an unsuitable cannula?


• The OPT312 and OPT314 cannot be used with the AIRVO 2.
See the User Manual for information regarding patient interfaces.

D.4.4 AIR FILTER

A. Is the air filter significantly discolored/dirty? Figure 7


• Replace with part 900PT913.

Note: A promptFig. 8 for filter change will occur at the start of the Disinfection
Cycle once the AIRVO 2 has counted 1,000 hours of use. Choose ‘Now’ or
‘Later’Fig. 9 by using the “up” or “down” buttons and press the “mode” button
( ) to confirm. Selecting ‘Now‘ will zero the counter. Selecting “Later“ Figure 8
will activate the prompt at the start of the next Disinfection Cycle.

B. Is there a foreign object blocking the air filter or filter holder?

D.4.5 CONDENSATION

Please see Section D.15.

D.4.5 ALTITUDE Figure 9

A. The AIRVO 2 is designed to operate at an altitude below 2,000 meters.

27
D.5 “Check for leaks”Fig. 10 or “E122”

The most likely cause is a missing water chamber or the existing chamber has
not been pushed into place correctly.

D.5.1 WATER CHAMBER


Figure 10
A. Is the water chamber fitted correctly? Even if it appears to be:
• Remove the water chamber.
• Push the chamber on firmly, until the finger guard “clicks” into placeFig. 11.
Warning: The heater-plate and base of the water chamber may be hot.

D.5.2 HEATED BREATHING TUBE

A. Is the heated breathing tube attached to the device correctly?


Even if it appears to be: Figure 11
• Disconnect the heated breathing tube.
• Check that the black O-ring is in placeFig.12.
If the O-ring is damaged or missing, replace with part 900PT408.
• Reconnect the heated breathing tube.

B. Confirm that the heated breathing tube is not visibly damaged.

D.5.3 PATIENT INTERFACE

A. Is the patient interface correctly fitted to the heated breathing tube? Figure 12
• Even if it appears to be, disconnect and reconnect the patient interface.
It should make a “click” sound when it is connected properly.
B. Should the unit be in Default (adult) mode?
• If the AIRVO is in Junior mode and the 900PT501 Default tube or the
900PT551/561 AirSpiral tube is used with the OPT842/44/46/70, RT013
or OPT942/44/46/70/80 interfaces, it may generate a “Check for leaks”
alarm.
See Appendix E for identification of the Default and Junior Tube and
Chamber Kits, according to their labels.

D.5.4 AIR FILTER & FILTER COVER

A. Is the air filter and filter cover (at the back of the device) correctly fitted, as
per the User Manual?

28
D.6 “O2 too low”Fig. 13
The measured oxygen level has fallen below the allowed limit.

A. Adjust the level of oxygen from the oxygen source as necessary, i.e. increase
the oxygen flow rate through the oxygen flow meter.

B. Is the oxygen source (wall or cylinder flow meter) turned on?

C. Is the oxygen source empty or faulty? Figure 13

D. Is the “AIRVO 2 oxygen inlet kit”Fig. 14 installed correctly, as per the


instructions included with part 900PT422 and confirmed that there are no
kinks in the “AIRVO 2 oxygen inlet kit” oxygen tubing?

E. Is the oxygen source tubing correctly and securely fitted to the AIRVO 2?
Figure 14
F. Allow the device to sufficiently warm up; rapid changes in temperature can
affect the sensor.
G. Is the minimum oxygen limit set to 25%?
• A prompt will appear with an option to change this lower limit to 21 %.
Select “Yes” or “No” by using the “Up“ and “Down” buttons. Press the
“mode” button ( ) to confirm selectionFig. 15.
See Section 2 - Advanced Settings to change this lower oxygen limit.
Figure 15

D.7 “O2 too high”Fig. 16


The measured oxygen level has risen above the allowed limit.

A. Adjust the level of oxygen from the oxygen source as necessary, i.e. decrease
the oxygen flow rate through the oxygen flow meter.

B. See Section 2 - Advanced Settings to change this lower oxygen limit. Figure 16

D.8 “Cannot reach target flow” Fig. 17


A. Press the “mode” button ( ) to continue normal operation.

B. Is the target flow setting too high for the patient interface?
• Check the swing tag/User Manual for the appropriate flow range for each
patient interface.

C. Follow steps in Section D.4 — “Check for blockages”.

Figure 17
D. If warning persists, turn down set flow-rate.

E. Is the altitude above 2,000 m?


The AIRVO 2 is designed to operate at an altitude below 2,000 meters.

29
D.9 “Cannot reach target temperature”Fig. 18
The most likely cause is operating the AIRVO 2 at a high flow rate in a cold room.
Consider decreasing the target flow setting.

A. Press “mode” button ( ) to continue.


Note: The humidity level may be compromised.
B. Is the ambient room temperature below 18 °C (64 °F)?
• If yes, proactive management of condensation may be required. Figure 18
See Section D.12 on prevention and management of condensation.

D.10 “Check water”Fig. 19


A. Is the water bag empty?
If yes, refill or replace the water bag and press the “mode” button ( ) to
reset the alarm.
B. Is the water chamber empty?
• If yes, replace the water chamber as it may be damaged.
Warning: The heater-plate and base of the water chamber may be hot.
C. Is there a kink in the fluid line, preventing water from flowing into the Figure 19
chamber?
D. Open the vent cap near the water bag spike. This allows the pressure to
equalize, letting the water flow into the water chamber.

D.11 “Check operating conditions” Fig. 20


This alarm may be caused by a sudden change in ambient room temperature,
e.g. storing the unit in a cold place, then using it in a warm place.
A. Is the ambient room temperature less than 10 °C (50 °F) or greater than
30 °C (86 °F)?
B. Leave the unit running for 30 minutes.
Switch the unit off, then restart. Figure 20

Disinfection Fig. 21
D.12. “Check for blockages”Fig. 2
A. Is the disinfection filter blocked or wet?

B. Check the integrity of the heater wire inside the disinfection tube. If it is
bunched up, try a new disinfection tube.

D.13. “Check for leaks”Fig. 10


A. Reconnect the disinfection tube.
Ensure that the disinfection tube is firmly secured to the left hand chamber
port, and the disinfection filter is firmly secured to the right hand chamber
port.

D.14. Warm up takes too long Figure 21


i.e., If the countdown timer is not displayed within 30 minutes.

A. Reconnect the disinfection tube, as described in Section D.13.A.

B. Check O2 is disconnected.

C. Turn off the AIRVO 2 and restart after 30 minutes.


30
D.15 Condensation

D.15.1 PREVENTION OF EXCESSIVE CONDENSATION


A. Is the AIRVO 2 being used in ambient conditions between 18 – 28 °C
(64 - 82 °F)?
• If the room is less than 18 °C (64 °F), condensation is more likely to
occur.

B. Is there a local source of cooling acting on the heated breathing tube?


• A fan to cool the patient,
• An air-conditioning unit, vent or an open window?
• Are you able to remove or minimize these sources of cooling, e.g.
redirect the fan, cooling the patient, away from the heated breathing
tube?

D.15.2 CONDENSATION MANAGEMENT


Figure 22
A. Implement
 a system to check the heated breathing tube for condensate
regularly.
B. Is
 the AIRVO 2 placed below head heightFig. 22?
• This will allow condensate to drain towards the water chamber, away
from the patient.
C. If
 condensation is present, drain it back into the water chamberFig. 23:
• Disconnect the patient interface from the heated breathing tube.
• Drain the tube by lifting the patient end of the tube, allowing the
condensate to run into the water chamber. Figure 23
• At higher target flow rates, it may be necessary to first reduce the target
flow rate to 30 L/min or below, to ensure the condensate drains into the
water chamber.
D. If condensate persists, consider turning the target temperature down.
• A lower target temperature will decrease the humidity output of the
AIRVO 2, decreasing the level of condensation.
Note: The temperature and humidity level delivered to the patient will also
be reduced.

D.16 ExxxFig. 24

A. Follow the instructions in Appendix F if a fault with an error code is


displayed on the AIRVO screen.

Figure 24

31
APPENDIX E: TUBE AND CHAMBER KIT LABELS

AIRVO 2
TM

900PT531
Tube and Chamber Kit (Junior)

INFANT PEDIATRIC

Single
Use Rx Only
0123
Fisher & Paykel Healthcare Ltd. 15 Maurice Paykel Place, East Tamaki,

Made in New Zealand


Auckland 2013, New Zealand Fisher & Paykel Healthcare Ltd. Unit 16,
Cordwallis Park, Clivemont Road, Maidenhead, Berkshire, SL6 7BU, UK

Label for Default (adult) Label for Junior


Tube and Chamber Kit Tube and Chamber Kit

Label for AirSpiral Label for Adult & Junior AirSpiral


Tube and Chamber Kit Tube and Chamber Kit

Note: Some accessories may not be available in all countries. 900PT551/900PT561 are not available in USA.
Please contact your local Fisher & Paykel Healthcare representative for more information.

32
APPENDIX F: ERROR FLOWCHART
The following pages provide troubleshooting advice for fault / error / “E” codes that may appear during use of the AIRVO 2.

Error Range/s: 1-10 11-33, 38-41, 44-50, 34-37, 42-43, 70-73, 61-69, 74, 91-106
141-158 75-78
Error Type: Motor Heated breathing tube Disinfection tube Chamber

Write down the Write down the error code Write down the error code Write down the error code and
error code and what and what happened before it and what happened before it what happened before it appeared.
happened before it appeared. appeared. Is the Airvo 2 being used in a cold or
appeared. Had the user just connected or How far through the hot room? Was it stored overnight
Was the unit dropped, replaced the breathing tube? disinfection cycle was the in a cold place? Did the user add hot
shaken or wet? Was there a “Check tube” error user? Was anything changed water to the chamber instead of room-
shown? (filter, tube, etc) half-way temperature water?
through?

Turn off Airvo 2. Check that the tube is


connected properly. Check that the tube, Check that the chamber is
Disconnect Airvo 2 connected properly to the Airvo 2,
from mains supply, disinfection filter and air
filter on back of the unit are that it isn’t overfilled and that hot
wait 5 seconds, then water hasn’t been added.
reconnect. connected properly.
Turn off Airvo 2.
Turn on Airvo 2. Disconnect Airvo 2 from mains
supply, wait 5 seconds, then
reconnect. Turn off Airvo 2.
Turn on Airvo 2. Turn off Airvo 2.
Disconnect Airvo 2 from mains Disconnect Airvo 2 from mains supply,
supply, wait 5 seconds, then wait 30 minutes for temperatures to
If problem reconnect. stabilize, then reconnect.
persists: Turn on Airvo 2 (ie. start the Turn on Airvo 2.
return Airvo 2 to If problem persists: disinfection cycle again).
Fisher & Paykel Repeat this process with a new
Healthcare. tube.

If problem persists:
If problem persists: Return Airvo 2 to Fisher & Paykel
Repeat this process with Healthcare.
another tube.

If problem
solved:
return faulty tube
to Fisher & Paykel If problem solved:
Healthcare. return faulty disinfection
tube to Fisher & Paykel
Healthcare.

If problem persists: If problem persists:


Return Airvo 2 to Fisher & Return Airvo 2 to Fisher &
Paykel Healthcare. Paykel Healthcare.

33
Error Range/s: 81-85, 111, 171 121-122 131-135 161 181-192
Error Type: Internal Error Flow Temperature Brownout Oxygen sensor

Write down the Write down the Write down the Check Write down the error
error code and what error code and what error code and what other electrical code and what happened
happened before it happened before it happened before it equipment nearby before it appeared.
appeared. appeared. appeared. and try another Is the Airvo 2 being used in
Did the Airvo 2 display a How long had the power outlet. a cold or hot room? Was it
“Check leaks” or “Check Airvo 2 been running? stored overnight in a cold
blockages” alarm? Is the Airvo 2 being place?
used in a hot or cold
environment?
Turn off Airvo 2.
Disconnect Airvo 2 If problem
from mains supply, persists:
wait 5 seconds, then Turn off Airvo 2. Return Airvo 2 to
Turn off Airvo 2. Fisher & Paykel
reconnect. Disconnect Airvo 2 Disconnect Airvo 2 Check that
Turn on Airvo 2. from mains supply, Healthcare. the Air Filter is
from mains supply,
wait 5 seconds, then wait 5 seconds, then securely fitted to
reconnect. reconnect. the back of the
Turn on Airvo 2. Turn on Airvo 2. AIRVO 2.

If problem persists: If problem persists:


return Airvo 2 to Fisher Return Airvo 2 to Fisher Turn off Airvo 2.
Check that the air & Paykel Healthcare.
& Paykel Healthcare. filter on the back of the Disconnect Airvo 2 from
unit is fitted correctly. mains supply, wait 30
Check that the tube and minutes for temperatures
chamber are both fitted to stabilize, then reconnect.
correctly. Turn on Airvo 2.
If this error occurred
during a disinfection
cycle, check that the
disinfection tube and
filter are both installed
properly.
If problem
persists:
Return Airvo 2 to
Fisher & Paykel
Healthcare.
If problem persists:
Return Airvo 2 to Fisher
& Paykel Healthcare.

34
APPENDIX G: TRANSPORT (LOW POWER) MODE
USER INSTRUCTIONS
To be used in conjunction with the AIRVO 2 User Manual.

REQUIREMENTS FOR USING TRANSPORT (LOW


POWER) MODE
To ensure correct operation of the AIRVO 2 during Transport (Low
Power) Mode, the power supply used during transport must be capable
of delivering the “Supply voltage/current” values listed in Section 5 of
the AIRVO 2 User Manual, for a period of at least 20 minutes.

TO ENTER TRANSPORT (LOW POWER) MODE


1. CHECK THAT UNIT IS READY FOR USE
Check that the unit is already operating and connected to a patient and
that the “Ready for Use” symbol is displayed showing that the system
is ready for the patient to use.

“Ready for use” symbol

2. ACCESS THE TARGET DEW-POINT TEMPERATURE SCREEN


Press the Mode button to advance to the target dew-point temperature
screen.

The current target dew-point temperature setting will be displayed.

37
°C
When you enter Transport (Low Power) Mode, the unit will enter a low-
power, low-humidity mode.

3. ACTIVATE TRANSPORT (LOW POWER) MODE


To activate Transport (Low Power)
Mode:

Press and hold the “Audio Pause” button


for 5 seconds.

The unit will enter Transport (Low Power)


Transport Mode Mode.
The Transport (Low Power) Mode symbol
Low (far left) is displayed.
Humidity!
To decrease power consumption, the
heater-plate will be switched off and
humidity output will decrease.

The target dew-point temperature


19:59 display will be replaced with a 20-minute
countdown timer.

5. DURING TRANSPORT (LOW POWER) MODE


When you return to the Summary screen, you will see that the dew-
point temperature display has been replaced with the Transport (Low
Power) Mode symbol and the countdown timer.
The time shown is the time left before the unit automatically reverts
19:34 back to normal operation and its original target dew-point temperature
settings.

35
6. TO DEACTIVATE TRANSPORT (LOW POWER) MODE
When the countdown reaches 0:00, the unit automatically reverts back
to normal operation and its original target dew-point temperature
settings.
To exit Transport (Low Power) Mode before the countdown reaches
0:00, follow the instructions below.

To deactivate Transport (Low Power) Mode

Press the Mode button to move to the target dew-


point temperature screen.

The countdown timer and Transport (Low Power)


15:42 Mode symbol are displayed

00:00

Press and hold the “Audio Pause” button for 5


seconds to deactivate Transport (Low Power) Mode.

New Target
Settings
New Target
Settings 37 30
°C L/min
The previous target settings for dew-point
temperature and flow will be restored.

37 30
°C L/min
The Transport (Low Power) Mode symbol has now

37
°C
disappeared. The target dew-point temperature
screen now displays the target dew-point
temperature setting.

Press the Mode button to advance through the target


flow and oxygen screens to the Summary screen.

7. TRANSPORT (LOW POWER) MODE DEACTIVATED


After Transport (Low Power) Mode is deactivated, the current output
dew-point temperature displayed on the Summary screen will have
decreased from its value before Transport (Low Power) Mode. The
AIRVO 2 will warm up until it reaches its target dew-point temperature
setting.

36
For more information please contact
your local Fisher & Paykel Healthcare representative
Manufacturer China Spain
Fisher & Paykel Healthcare Ltd Tel: +86 20 32053486 Tel: +34 902 013 346
15 Maurice Paykel Place Fax: +86 20 32052132 Fax: +34 902 013 379
East Tamaki, Auckland 2013
Finland Sweden
PO Box 14 348 Panmure Tel: +358 (0)405 406618 Tel: +46 8 564 76 680
Auckland 1741 Fax: +46 (0)8 36 6310 Fax: +46 8 36 63 10
New Zealand France Switzerland
Tel: +64 9 574 0100 Tel: +33 1 6446 5201 Tel: 0800 83 47 63
Fax: +33 1 6446 5221 Fax: 0800 83 47 54
Fax: +64 9 574 0158
Email: info@fphcare.co.nz Germany Taiwan
Web: www.fphcare.com Tel: +49 7181 98599 0 Tel: +886 2 8751 1739
Fax: +49 7181 98599 66 Fax: +886 2 8751 5625
India Turkey
Tel: +91 80 4284 4000 Fisher Paykel Sağlık Ürünleri
Fax: +91 80 4123 6044 Ticaret Limited Şirketi,
Irish Republic Alinteri Bulvari 1161/1 Sokak
Australia (Sponsor) Tel: 1800 409 011 No. 12-14, P.O. Box 06371 Ostim,
Fisher & Paykel Healthcare Pty Limited Ankara, Turkey
36–40 New Street, PO Box 167 Italy
Ringwood, Melbourne Tel: +39 06 7839 2939 Tel: +90 312 354 34 12
Victoria 3134, Australia Fax: +39 06 7814 7709 Fax: +90 312 354 31 01
REF 185047911 REV F © 2016-06 Fisher & Paykel Healthcare Limited

Tel: +61 3 9879 5022 Japan UK


Fax: +61 3 9879 5232 Tel: +81 3 5117 7110 Fisher & Paykel Healthcare Ltd
Fax: +81 3 5117 7115 Unit 16, Cordwallis Park
Austria Clivemont Road, Maidenhead
Tel: 0800 29 31 23 Korea
Tel: +82 2 6205 6900 Berkshire SL6 7BU, UK
Fax: 0800 29 31 22
Fax: +82 2 6309 6901 Tel: +44 1628 626 136
Benelux Fax: +44 1628 626 146
Tel: +31 40 216 3555
Northern Ireland
Fax: +31 40 216 3554 USA/Canada
Tel: 0800 132 189
Tel: +1 800 446 3908
Brazil or +1 949 453 4000
Russia
Tel: +55 11 2548 8002 Fax: +1 949 453 4001
Tel and Fax: +7 495 782 21 50

0123
One or more products in this manual are covered by or for use under one or more of the following patents, design registrations or design patents in the corresponding countries listed below.
Australia: AU2006325613, AU2011245770, AU2009251939, AU2011245771, AU2006266564, AU2005273119, AU2003204474, AU756477, AU2003244579, AU2007214363, AU2004202274, AU2009243444,
AU2004216567, AU2003219624, AU345792, AU355510, AU355511.
Europe: EP1973577 (DE, FR, GB), EP1396277 (DE, FR, GB, IT), EP1484161 (DE, FR, GB, IT), EP1495855 (DE, FR, GB, IT), EP1481702 (DE, FR, GB, SE), EP1471962 (DE, FR, GB, SE), EP2799103
(DE, ES, FR, GB, IT, NL, SE, TR), CRD no:002372292-0001, CRD no:002372292-0002, CRD no:002100289-0001, CRD no:002100289-0002, CRD no:002100289-0003, CRD no:002132464-0001,
CRD no:002132464-0002, CRD no:002452839-0001, CRD no:002452839-0002, CRD no:002554857-0001, CRD no:002554857-0002.
USA: US8741220, US8616202, US8453641, US8186345, US7111624, US8091547, US8550072, US6953354, US6598604, US7468116, US7493902, US8267092, US8220463, US8997747, US7458615, USD747792.
One or more of the products in this manual are also covered by or for use under one or more pending patent applications and/or design applications.

www.fphcare.com

You might also like