Professional Documents
Culture Documents
51000-18
© Hach Company, 1998, 1999. All rights reserved. Printed in the U.S.A. hm/dp 1/99 2ed
Rev1, 5/00
Trademarks of Hach Company
Operation ............................................................................................................................................................. 13
Page 3
51000-18TOC.fm Table of Contents
Table of Contents
Maintenance ........................................................................................................................................................ 83
Page 4
Table of Contents 51000-18TOC.fm
Table of Contents
Page 5
51000-18TOC.fm Table of Contents
Table of Contents
Page 6
Table of Contents 51000-18TOC.fm
Certification
The APA 6000™ Alkalinity Process Analyzer has been tested and is certified
as indicated to the following instrumentation standards:
Product Safety
The APA 6000 Alkalinity Analyzer is ETL listed to UL Standard 3101-1
(Listing # H0492805390) and is ETL certified to CSA Standard C22.2 No. 142
(Certification # H0492805390).
Immunity
EN 50082-2 (Generic Immunity Standard) per 89/336/EEC EMC:
Emissions
EN 50081-2 (Generic Emissions Standard) per 89/336/EEC EMC:
Page 7
51000-18 certification.fm Certification
Certification
This device complies with Part 15 of the FCC Rules. Operation is subject
to the following two conditions:
(1) This device may not cause harmful interference, and (2) this device must
accept any interference received, including interference that may cause
undesired operation.
This equipment has been tested and found to comply with the limits for a
Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference when
the equipment is operated in a commercial environment. This equipment
generates, uses, and can radiate radio frequency energy and, if not installed
and used in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference, in which case the user
will be required to correct the interference at his own expense.
1. Disconnect the APA 6000 Alkalinity Analyzer from its power source to
verify that it is or is not the source of the interference.
2. If the AC power supply for the APA 6000 Alkalinity Analyzer is connected
into the same outlet as the device with which it is interfering, try
another outlet.
3. Move the APA 6000 Alkalinity Analyzer away from the device receiving
the interference.
4. Reposition the receiving antenna for the device receiving the interference.
Page 8
Certification 51000-18 certification.fm
Safety Precautions
Please read this entire manual before unpacking, setting up,
or operating this instrument. Pay particular attention to all danger and
caution statements. Failure to do so could result in serious injury to
the operator or damage the equipment.
To ensure the protection provided by this equipment is not impaired,
do not use or install this equipment in any manner other than that
which is specified in this manual.
Precautionary Labels
Read all labels and tags attached to the instrument. Personal injury or
damage to the instrument could occur if not observed.
! Section 2.6 Replacing the Soda Lime in the Standard Solution Scrubbers on page 54
Page 9
51000-18 safety.fm Safety Precautions
Safety Precautions
! Section 7.3.3 Recorder Output Connections on page 134
Page 10
Safety Precautions 51000-18 safety.fm
Specifications
Performance
Range: 1 to 1000 mg/L for total alkalinity as calcium carbonate; 5 to
1000 mg/L for P alklanity as calcium carbonate
Response Time: For single sample stream instruments, less than 10 minutes
for 90% response to step change at sample inlet.
Relays: Two SPDT relays with a maximum contact rating of 5 A resistive load
at 240 VAC. Additional relays available through optional Signal Output
Module.
Features
Instrument Enclosure: NEMA 4X (Indoor) and IEC 529 (IP 66) with
provision for air purge
Reagent Enclosure: Fully enclosed latch cases with protected ducting for
reagent and standard lines.
Page 11
51000-18 specification.fm Specifications
Specifications
Sample/Environmental Requirements
Operating Temperature: 5 to 50 °C (41 to 122 °F)
Sample Pressure: 2.5 to 100 psig (0.17 to 6.9 bar) to at Basic Water
Conditioning Filter; 0.5 to 30 psig (0.035 to 2 bar) maximum at sample inlet
block.
Grab Sample:
Temperature: 5 to 50 °C
Page 12
Specifications 51000-18 specification.fm
Operation
DANGER
Handling chemical samples, standards, and reagents can be dangerous. Review the necessary Material
Safety Data Sheets and become familiar with all safety procedures before handling any chemicals.
DANGER
La manipulation des échantillons chimiques, étalons et réactifs peut être dangereuse. Lire les Fiches de
Données de Sécurité des Produits (FDSP) et se familiariser avec toutes les procédures de sécurité avant
de manipuler tous les produits chimiques.
PELIGRO
La manipulación de muestras químicas, estándares y reactivos puede ser peligrosa. Revise las fichas
de seguridad de materiales y familiarícese con los procedimientos de seguridad antes de manipular
productos químicos.
GEFAHR
Das Arbeiten mit chemischen Proben, Standards und Reagenzien ist mit Gefahren verbunden. Es wird dem
Benutzer dieser Produkte empfohlen, sich vor der Arbeit mit sicheren Verfahrensweisen und dem richtigen
Gebrauch der Chemikalien vertraut zu machen und alle entsprechenden Materialsicherheitsdatenblätter
aufmerksam zu lesen.
PERICOLO
La manipolazione di campioni, standard e reattivi chimici può essere pericolosa. La preghiamo di prendere
conoscenza delle Schede Techniche necessarie legate alla Sicurezza dei Materiali e di abituarsi con tutte
le procedure di sicurezza prima di manipolare ogni prodotto chimico.
Page 13
51000-18 operation stopper.fm Operation
Visit http: //www.hach.com
Section 1 Introduction
1. As you unpack the analyzer and its accessories, refer to the packing lists
in Section 1.2 to ensure all the hardware required to install and operate
the analyzer arrived with the shipment.
3. See Section 7.7 on page 147 to bring the sample from the process
stream to the APA. Only a qualified technician should complete the steps
in this section.
Page 15
51000-18 introduction.fm Introduction
Section 1
4. Refer to Section 7 on page 123 to complete the mounting and electrical
installation of the APA. Only a qualified technician should complete the
steps in this section.
Page 16
Packing Lists 51000-18 introduction.fm
Section 1
The analyzer uses a colorimetric method to determine concentration.
Data retrieval and communication with the instrument is done through
the AquaTrend® Interface keypad and display, mounted in the door of
the analyzer.
The instrument case meets NEMA 4X (indoor) and IP66 industrial enclosure
requirements. The front door is constructed of corrosion-resistant materials
and opens for a full view of the analyzer’s modular components (refer to
Figure 3 on page 19). The analyzer can be used on a bench or mounted on a
wall or panel. Electrical connections are made through the back of the
analyzer. Sample and drain connections are made with standard tubing and
fittings on either side of the instrument.
Reagents and standards are plumbed from the reagent enclosures on the
side of the instrument (see Figure 2). All tubing enters the instrument at the
ports at the bottom of the instrument bezel.
The analyzer has two analog outputs of 0-20 or 4-20 mA, which can be used
to drive a chart recorder or a process controller. Recorder span minimum and
maximum values can be programmed using the AquaTrend Interface keypad.
Two programmable alarm circuits provide relay closures, both normally open
and normally closed, for two selectable concentration set points. Set points
can be programmed anywhere within the overall range. System warning and
system alarm features may be used to provide alarm relay closures indicating
a need for operator attention.
If you need more than two relays or two analog outputs, external Signal
Output Modules (SOM) may be purchased.
See Sections 1.4.4, 1.4.5, 1.4.6 and 1.4.7 for details about these kits
Instrument Enclosure
The instrument enclosure houses various modules, the instrument’s power
supply and control circuitry (see Figure 3). Within the enclosure, a module
panel separates sensitive electrical components from the wetted parts.
All openings to the module panel are gasketed to protect against leakage.
Page 17
51000-18 introduction.fm Instrument Overview
Section 1
The module panel folds out to allow access to the electrical connections at the
back of the instrument.
Electrical access points (power hookups, alarm outputs, etc.) are behind
the module panel in a grounded, covered area. A plastic upper cover
shields the power supply and control circuit board. Do not remove this cover!
The metal lower cover is removable for access to electrical connections
during installation.
Page 18
Instrument Overview 51000-18 introduction.fm
Section 1
Autoburette Module
The analyzer uses a positive-displacement autoburette to measure and move
fluids through the various analytical components in the instrument. The
autoburette module consists of a clear acrylic block which houses a ceramic
piston, a double seal for the piston, a linear actuator, a two-way solenoid
valve, and control circuitry.
Colorimeter Module
Samples are measured in a temperature-controlled single-wavelength flow
cell. The light source is an LED, which passes through a 600 nm bandpass
filter. The path length is 8.75 mm (0.344 inches).
Mixer Module
The mixer module mixes reagents, sample, and standards.
Module Panel
Page 19
51000-18 introduction.fm Instrument Overview
Section 1
Data can be viewed in numeric or graphical format and reviewed on trend
graphs for details of measurement history. With the addition of the Hach Serial
I/O Module, data can be downloaded to a computer or printer.
A “network” is the connection of one or more devices (at least one must be a
sensor) to a master AquaTrend. This master AquaTrend serves as the
network control panel.
Page 20
Instrument Overview 51000-18 introduction.fm
Section 1
Each AquaTrend network device manual (AquaTrend, Digital Display Module,
Serial I/O Module, and Signal Output Module) provides setup instructions for
that component. The table below describes the maximum number of devices
that can be managed by one Master AquaTrend on a system. Additional
power supplies may be needed for devices other than the SIO.
One AquaTrend can link up to eight sensors (including the Alkalinity Analyzer)
and other devices in a variety of configurations. If you want to link any
additional sensors, consult the sensors’ manuals and the AquaTrend Interface
Manual. These manuals also contain all the information needed for setting up
an AquaTrend network and other network devices.
Page 21
51000-18 introduction.fm Instrument Overview
Section 1
Page 22
Instrument Overview 51000-18 introduction.fm
Section 1
APA Alkalinity Reagent 1 is an acid titrant that reacts with the alkalinity in
the sample.
The fifth reagent, APA Alkalinity Cleaning Solution, cleans the system during
the Prime and Instrument Clean functions. This solution assures the system
stays free of precipitate buildup. Initially, a different cleaning solution (Cat. No.
26974-53) may be used to make sure all parts are wetted.
Figure 7 APA 6000 Analyzer Reagent, Standard, and Cleaning Solution Containers
Page 23
51000-18 introduction.fm Instrument Overview
Section 1
Page 24
Instrument Overview 51000-18 introduction.fm
Section 1
51169-00
Tefzel Tubing
51178-00
51340-00
51345-00
51357-00
51131-00 51129-00
51356-00 51132-00 51130-00
51359-00
51293-00
2. The sample and indicator are delivered to the mixer and blended.
5. The mixture is delivered to the detector, dispersing the acid through the
mixture and providing a characteristic titration profile.
Page 25
51000-18 introduction.fm Theory of Operation
Section 1
Some of the sample/indicator mixture is then aspirated from the mixer and
delivered to the detector, replacing the previous clear fluid. The remainder of
the mixture is aspirated from the mixer, followed by a slug of acid. This
acidified mixture is delivered to the detector via the reaction coil. As it travels
through the reaction coil, the acid becomes dispersed through the
sample/indicator mixture on both sides of the acid slug, forming a pH gradient.
There will be an small volume (band) of fluid on both sides of the acid peak
where the acid exactly neutralizes the alkalinity present in the sample, and the
pH indicator changes color.
As the fluid flows through the detector, the absorbance at 600 nm is recorded,
producing a characteristic titration peak. The peak width is determined at two
specific absorbance values. One width is determined at the absorbance
representing pH 8.3, and the other at the absorbance representing pH 4.5.
The first width represents the phenolphthalein alkalinity and the second width
represents the total alkalinity of the sample. The instrument automatically
calibrates using two standards (0 and 2000 mg/L as CaCO3) to establish a
multi-point calibration curve.
Optional Signal Output Modules (SOM) provide the analyzer with additional
relay and recorder/controller outputs.
Serial Input/Output (SIO) modules can connect the analyzer to a computer
via an RS232 interface.
Digital Display Modules (DDM) provide a remote display of a given sample
stream’s measurement.
Remote AquaTrends allow for remote operation of the analyzer and
remote display of sample measurements. Portable AquaTrends provide
temporary operation/control of analyzers that do not have a built-in
AquaTrend® Interface.
Page 26
Method of Analysis 51000-18 introduction.fm
Section 2 Startup
2.1 Introduction
It is assumed that the analyzer has already been plumbed to the sample
source and a drain, as well as wired to a power source and any other
electrical wiring associated with 4-20 mA connections, relay connections, and
network connections by the installation technician according to the Hardware
Installation section. This section presents the steps, in sequence, an operator
should take to begin operation. This section will;
• describe the steps the operator will perform in getting the reagents and
standards set up.
• list the steps required to prime the sample lines and reagents.
Page 27
51000-18 startup.fm Startup
Section 2
1
2
4
DISPLAY EXIT
3 5
MENU ENTER
Allows you to navigate through the various menus, to change settings, and to move the
1 cursor when observing graph data.Table 4 summarizes the functions of the arrow keys in
each menu view.
Displays the data display screen. Once the data is displayed, pressing the DISPLAY key
2
toggles between data presented in numerical form and graph form.
3 Displays the main menu from any menu level or from the data display.
Accesses the data display screen from the main menu. From any other menu level, press the
4 EXITkey to display the previous menu screen. The EXIT key can also be used to abandon any
alphanumeric entry or selection process.
The arrow icons and their positions in the arrow field in the top right corner of
the screen (item 6 on Figure 11) indicate which of the arrow keys are available
for the screen you are in. Figure 10 shows the screen categories and Table 4
summarizes the function of the arrow keys for each screen category.
Page 28
The AquaTrend 51000-18 startup.fm
Section 2
5. Select STANDBY.
6. Press ENTER.
Note: This puts the analyzer into AquaTrend screens appear in several different formats (see Figure 10).
Standby Mode and terminates all See Figure 11 and Figure 12 for descriptions of each field.
operations. When the analyzer is
restarted, it will automatically start When the MAIN MENU is displayed on the AquaTrend screen, press the EXIT
performing measurements. or display key to return to the data display mode. Consecutive presses of the
EXIT key from any menu takes you back, one screen at a time, to the data
display screen. Pressing DISPLAY from any menu immediately returns you to
the measurement display screen.
The AquaTrend will display data from the APA 6000 Alkalinity Analyzer except
when the operator uses it for specific purposes. When viewing data, press the
DISPLAY key to alternate between displaying the data in numeric and
graphical mode.
• When viewing data in graph mode, you can see a history of readings
over the last 60 minutes, 24 hours or 30 days. The AquaTrend,
however, displays data for only one measurement channel at a time in
the graphic mode.
For details on selecting the preferred graph time and scaling, see
Section 3.2.1 on page 70.
1. Menu Screen 2. Numeric Data 3. Graph Display Screen 4. Pop-up Entry Screen
Display Screen
Page 29
51000-18 startup.fm The AquaTrend
Section 2
Page 30
The AquaTrend 51000-18 startup.fm
Section 2
15
2
AquaTrend™
14
3 13
12
6
11
7 Alarms
8 9 10
Table 6
Number Field Information Shown in Field
1 Status Bar Shows the active menu
2 Channel assignment Channel letter of the measurement shown in the graph
3 Graph data, upper limit (y-axis) Upper numeric limit of trend graph data (Y max)
4 Cursor Y-data Y value (measurement) value at cursor location
5 Units Unit of the measurement on the displayed channel (ppm or mg/L)
6 Species Chemical form of the measurement on the displayed channel
7 Graph data, lower limit (y-axis) Lower numeric limit of trend graph data (Y min)
Lower time limit If reviewing data with cursor, this fields is replaced by the
8 Graph time, lower limit (x-axis)
(oldest data) date and/or time, depending on selected time range.
9 Cursor time/date Time or date of data at the cursor position
Current time If reviewing data with cursor, this field is replaced by the
10 Graph time, upper limit (x-axis)
(latest data) date and/or time depending on selected time range.
Recalls value of data at its current position. Use the left and right arrow keys to
11 Graph cursor move the cursor to different times on the graph for data review. Move the cursor
all the way to the right or press EXIT to return to the current reading.
Line graph of 60 minutes, 24 hours of data, or one month of data. These
12 Graph Area
settings are selectable in the DISPLAY SETUP menu.
Depending on the sensor type and configuration, the display shows
13 Sensor name measurement channel and name or the sensor location, or a combination of
both.
14 Units and chemical form Units and chemical form of the measurement on the channel
Direction of these arrow icons correspond to enabled arrow keys for each
15 Arrow Field
screen.
16 Current Data Most recent measurement (same as numeric display)
Page 31
51000-18 startup.fm The AquaTrend
Section 2
5. A pop-up screen will prompt you to enter the new time. The numbers
representing the hour will be highlighted. Press the up arrow key to
increase the hour setting and press the down arrow key to decrease the
setting. When the proper hour is displayed, press the right arrow key and
set the minutes in the same manner.
Note: The AquaTrend uses a 24-hour clock (military time), so there is no selection for
a.m. or p.m. If the current time is 4:00 p.m., for example, you would enter the
24-hour format equivalent of 16:00.
Page 32
The AquaTrend 51000-18 startup.fm
Section 2
5. A pop-up box will prompt you to enter the new date. The numbers
representing the month will be highlighted. Press the up arrow key to
increase the month setting and press the down arrow key to decrease the
setting. When the proper month is displayed, press the right arrow key
and set the day. Repeat to set the last two digits of the current year.
5. A pop-up box will prompt you to select the day. Use the up or down arrow
key to highlight the desired day.
6. Press ENTER to save the change or press EXIT to cancel. The day will
change automatically. Check to be sure the clock is set at the correct time.
Page 33
51000-18 startup.fm The AquaTrend
Section 2
5. A pop-up box will appear with SELECT DATE FORMAT. Use the up and
down arrow keys to select a date format: MM/DD/YY is month/day/year,
and DD/MM/YY is day/month/year.
4. Use the up or down arrow key to select the language you want to use and
press ENTER. All screens will be displayed in this language.
Page 34
The AquaTrend 51000-18 startup.fm
Section 2
4. Use the up or down arrow key to select EDIT NAME and press ENTER.
5. Use the left and right arrow keys to choose the character to change and
use the up and down arrow keys to cycle through letters, numbers and
symbols until you reach the character you want.
Note: In order to distinguish between devices on the network, use a unique and
descriptive sensor name. The sensor name can be up to 12 characters long in any
combination of alphanumeric characters. The first character must be a letter or the
# sign.
Page 35
51000-18 startup.fm The AquaTrend
Section 2
• setting a password;
The default password is HACH. The default security level is NONE. This
security level allows access to all functions.
5. Enter the current password for access to the AquaTrend menus. Pressing
the right and left arrow keys moves the cursor right and left; pressing the
up and down arrow keys scrolls through the available characters. The
default password is HACH.
Page 36
The AquaTrend 51000-18 startup.fm
Section 2
TOTAL lockout means the operator is unable to change any of the analyzer’s
settings or cause the analyzer to perform any operations (priming, cleaning,
calibrating, etc.) from the keyboard. Any user will be able to review all settings,
but will be unable to modify them if they are locked out.
PARTIAL lockout allows user access to the following menus and functions:
Setting the security level to NONE allows access, without a password, to any
of the menus except SECURITY SETUP.
5. Enter the password (default is HACH). Use the left and right arrow keys to
choose the character to change and use the up and down arrow keys to
cycle through letters, numbers and symbols until you reach the character
you want.
Page 37
51000-18 startup.fm The AquaTrend
Section 2
5. Enter the password (this is necessary only if the security level is set to
Partial or Total).
6. Choose the duration of time you want the keypad to remain unlocked:
30, 60, 120 or 360 minutes.
Page 38
The AquaTrend 51000-18 startup.fm
Section 2
Important Note: Changing units causes the data log for the measurement to
be erased.
6. Select the units you want to use and press ENTER, or press EXIT to cancel
Note: When units are changed, the display shows new units with the next
measurement.
Page 39
51000-18 startup.fm The AquaTrend
Section 2
2.3.1 Alarms
Each APA 6000 Analyzer provides 14 alarm signals for your system. Two of
these, ALA and ALB, are permanently attached to the analyzer’s internal
alarm outputs, labeled Relay A and Relay B. The remaining alarm signals,
AL1 through AL12, can be attached to relays in optional Signal Output
Modules (SOMs).
SOMs can be purchased separately, and provide two relays and two analog
outputs each. For information on attaching alarms to relays and recorders to
analog outputs, see Section 7.3.4 on page 136 and Section 8 on page 155.
Page 40
Outputs and Relays 51000-18 startup.fm
Section 2
To set up the alarms:
ALA and ALB are internal analog outputs connected to relays in the APA
analyzer and may be used to actuate the air backflush and submersible
pump cleaning cycles. AL1 through AL12 may be connected to relays in
an SOM.
Note: Setup for Air Backflush and Sample Pump is described in Section 2.4.1.1 on
page 49 and Section 2.4.1.2 on page 50.
Page 41
51000-18 startup.fm Outputs and Relays
Section 2
8. If HIGH, LOW or RATE was selected in Step 7, the following setup options
are available:
10. Press ENTER to save the alarm setting, or press EXIT to cancel.
Page 42
Outputs and Relays 51000-18 startup.fm
Section 2
Page 43
51000-18 startup.fm Outputs and Relays
Section 2
5. Select the analog output you want to use and press ENTER.
Select RC [1 -12] if you want to configure a SOM output to a recorder.
Select RCA or RCB if you want to configure one of the APA analog outputs
as a recorder output.
OR
12. Modify the minimum value by using the arrow keys and press ENTER
when the value entry is complete.
Page 44
Outputs and Relays 51000-18 startup.fm
Section 2
14. Modify the maximum value by using the arrow keys and press ENTER
when the value entry is complete.
17. To change the recorder output action when the analyzer is unable to
perform measurements in the selected measurement range, select
MEAS FAILURE: XXXX and press ENTER.
18. Select HOLD, GO MAX or GO MIN to indicate what the output action
should be when the analyzer cannot perform a selected measurement.
GO MAX sets the output to 20 mA, GO MIN sets the output to 4 mA, and
HOLD keeps the output at the current value. Press ENTER.
19. To complete the recorder setup for RCA or RCB, see Section 2.3.3.3. To
complete the recorder setup for RC1-12 see Analog Output Settings in
the SOM Manual.
Page 45
51000-18 startup.fm Outputs and Relays
Section 2
Only RCA, RCB, RC1, and RC2 can configured as controller outputs to
actuate controllers.
1. Follow Steps 1-7 in Section 2.3.3.1 to access the controller setup menu,
then choose CONTROLLER in Step 7.
2. To change the measurement that the analyzer uses to monitor for its
control action, select MEASUREMENT: XXXX and press ENTER.
4. To turn off the control output, enable automatic control or enable manual
control, select ENABLE: XXXX and press ENTER.
7. To change the dead band value, select DEAD BAND: XXXX and
press ENTER.
8. Modify the dead band value by using the arrow keys. Press ENTER.
9. To change the proportional band value, select PROP BAND: XXXX and
press ENTER.
10. Modify the proportional band value by using the arrow keys.
Press ENTER.
Page 46
Outputs and Relays 51000-18 startup.fm
Section 2
11. To change the reset value, select RESET: XXXX and press ENTER.
12. Modify the reset value by using the arrow keys. Press ENTER.
13. To change the rate value, select RATE: XXXX and press ENTER.
14. Modify the rate value by using the arrow keys. Press ENTER.
15. To change the set point value, select SET POINT: XXXX and press ENTER.
Modify the set point value by using the arrow keys. Press ENTER.
16. To change the output action setting, select OUTPUT ACTION: XXXX and
press ENTER.
18. To change the recorder or controllers output action when the analyzer is
unable to perform measurements in the selected measurement range,
select MEAS FAILURE: XXXX and press ENTER.
19. Select HOLD, GO MAX or GO MIN to indicate what the output action
should be when the analyzer cannot perform a selected measurement.
GO MAX sets the output to 20 mA, GO MIN sets the output to 4 mA, and
HOLD keeps the output at the current value. Press enter.
Page 47
51000-18 startup.fm Outputs and Relays
Section 2
If you wish to adjust (trim) the milliamp output range for the internal
recorder/controller outputs (RCA or RCB) or if you want to change the range,
select OUTPUT A SETUP or OUTPUT B SETUP. Press ENTER.
• To adjust the mA output at the bottom of the scale, select ZERO SCALE
TRIM and press ENTER. Use the up/down arrow to select FINE or
COARSE ADJUSTMENT. Press ENTER. The new trim mA value will appear.
Select ACCEPT NEW VALUE when done and press ENTER.
• To adjust the mA output at the top of the scale, select FULL SCALE TRIM
and press ENTER. Adjust as described above.
Page 48
Sample Conditioning Setup 51000-18 startup.fm
Section 2
1. Press the MENU key on the AquaTrend to start from the MAIN MENU.
2. Select the SENSOR MENU by pressing the down arrow key. Press ENTER.
Note: An alarm relay should be used to actuate the air backflush valve.
5. Select the alarm output that drives the relay which controls the air
backflush (see Section 7.3.2 on page 133 for installation.)
8. Ensure that ENABLED is set to YES. If not, select ENABLE and press
ENTER. Select YES and press ENTER.
Page 49
51000-18 startup.fm Sample Conditioning Setup
Section 2
5. Select the alarm output that drives the relay which controls the sample
pump (see Section 7.3.2 on page 133 for installation).
8. Ensure that ENABLED is set to YES. If not, select ENABLE and press
ENTER. Select YES and press ENTER.
Page 50
Sample Conditioning Setup 51000-18 startup.fm
Section 2
2. Remove the bottle vent cap assembly and tubing from the top of the
existing solution bottle.
3. Rinse the straw surrounding the tubing with deionized water and shake off
any residual water before inserting it into the new solution. This avoids
possible contamination. Be sure the tubing extends to the bottom of the
straw. If not, adjust it and re-mark the tubing.
4. Insert the straw into the new bottle and screw the vent cap on. Be sure the
mark on the tubing is still located where the tubing exits the bottle. This
assures the inner tube has remained at the bottom of the bottle.
6. Use the applicable prime function to fill the tubing with new solution
(see Section 2.7.2 on page 55 for priming procedures).
7. Set the solution level (Section 2.5.3 on page 53) using the
AquaTrend keypad.
Page 51
51000-18 startup.fm Changing Reagents and Standards
Section 2
1. Remove the bottle vent cap assembly from the top of the existing
standard solution bottle. Keep the scrubber and reagent line in the same
position relative to the cap.
2. Rinse the straw surrounding the tubing with deionized water and shake
off any residual water before inserting it into the new solution. This
avoids possible contamination. Be sure the tubing extends to the bottom
of the straw.
3. Insert the vent cap assembly into the new standard bottle. Make sure the
scrubber tubing is above the level of the liquid. Make sure the reagent line
to the instrument is at the bottom of the bottle.
5. Use the applicable prime function to fill the tubing with new solution
(see Section 2.7.2 on page 55 for priming procedures).
6. Set the solution level (Section 2.5.3 on page 53) using the
AquaTrend keypad.
Page 52
Changing Reagents and Standards 51000-18 startup.fm
Section 2
7. If you have refilled all the solutions at once, select RESET ALL TO 100%
and press ENTER. Go to Step 9.
If you have not refilled all the solutions, select the solution(s) you
are changing.
8. Press ENTER and a pop-up box will show a default setting of 100%. If
necessary, adjust the setting by pressing the up and down arrow keys.
9. Press the ENTER key to accept the value or press EXIT to cancel.
10. Repeat Steps 7-9 for each solution that was changed.
Page 53
51000-18 startup.fm Changing Reagents and Standards
Section 2
1. Gently remove the scrubber cap from the glass tube containing the soda
lime. Be sure to keep the inlet and outlet tubes in their proper positions.
3. Add new soda lime to the glass tube (a funnel may help). Be sure to leave
enough room to replace the top.
4. Gently insert the scrubber cap back into the glass tube. Be sure the seal
is tight.
Page 54
Replacing the Soda Lime in the Standard Solution Scrubbers 51000-18 startup.fm
Section 2
1. Set the flow rate with the ¼-in. ball valve. When the lever is perpendicular
to the valve body, the valve is fully closed; when the lever is parallel with
the valve body, the valve is fully open.
2. Check the flow by pulling the ¼-in. OD tube from the push-in fitting at the
drain tee (the tubing from the sample inlet block to the drain tee). If
desired, measure the flow rate by directing the sample into a large
measuring container (flask or cylinder) over a set amount of time. Keep
the end of the tube at the same height as the push-in fitting.
5. Select RESTART.
6. Press ENTER.
Page 55
51000-18 startup.fm Starting Analyzer Operation
Section 2
Priming the analyzer assures that the standards and/or reagent fill the tubing
going to the rotary valve. When reagents, standards or cleaning solutions are
replaced, perform the applicable priming operations.
• PRIME ENTIRE SYSTEM NOW. This will prime all the lines from
the solution bottles and sample lines. Use this option when all the
reagent bottles are replaced. The system prime will take about
12 minutes. The analyzer will display various messages about the
progress of the cycle.
• PRIME REAGENTS NOW. Use this option when only Reagent 1 and/or
Reagent 2 is replaced.
• PRIME STANDARDS NOW. Use this option when only the Standards
are replaced.
7. A pop-up window will indicate that priming will begin after the current
operation cycle. When priming starts, a pop-window will indicate the time
remaining to complete the prime cycle.
Page 56
Starting Analyzer Operation 51000-18 startup.fm
Section 2
6. The default is for the analyzer to calibrate once every week. To change
this, select one of the following:
Note: When auto-calibrate is set to 12 hours, the analyzer calibrates at the set
auto-cal time and again 12 hours later.
• Select AUTO-CAL TIME to specify the clock hour you want the
instrument to calibrate (use a 24-hour clock). Use the up and down
arrow keys to enter an hour between 1 and 24. For example, if you
want the instrument to calibrate at 2:00 p.m., enter 14.
Page 57
51000-18 startup.fm Starting Analyzer Operation
Section 2
6. The default is for the analyzer to perform a calibration check once every
day. To change this, select one of the following:
Note: When auto-calibrate check is set to 12 hours, the analyzer checks the calibration
at the set auto-cal time and again 12 hours later.
• Select AUTO-CAL CHECK TIME to specify the clock hour you want the
instrument to check the calibration (use a 24-hour clock). Use the up
and down arrow keys to enter an hour between 1 and 24. For
example, if you want the instrument to calibrate at 2:00 p.m., enter 14.
Page 58
Starting Analyzer Operation 51000-18 startup.fm
Section 2
7. Set the warning limit to 0 to disable this feature. Otherwise, use the arrow
keys to choose a percent value by which the calibration value can vary.
For example, entering 05 will cause the analyzer to issue a warning if the
point being checked is different by ±5% of the current calibration value.
Page 59
51000-18 startup.fm Starting Analyzer Operation
Section 2
7. Set the recal limit to 0 to disable this feature. Otherwise, use the arrow
keys to choose a percent value by which the calibration value can vary.
For example, entering 05 will cause the analyzer to recalibrate if the point
being checked is different by ±5% or greater of the current calibration
Page 60
Starting Analyzer Operation 51000-18 startup.fm
Section 2
8. Select the how often you want the analyzer to automatically clean itself
and press ENTER.
Note: Using the default settings is strongly recommended by the manufacturer and
should only be changed if sample components cause measurement problems.
If cleaning frequency is decreased, verify that the cleanliness and operation of the
analyzer is not degraded.
Page 61
51000-18 startup.fm Instrument Clean Function Setup
Section 2
6. Enter the registration number that came with your sample stream
expansion kit.
7. Select the number of sample streams the analyzer is plumbed for and
press ENTER.
Page 62
Adding Sample Streams 51000-18 startup.fm
Section 2
Note: The sequence will terminate at the first unused step. Succeeding steps
will be cleared.
Page 63
51000-18 startup.fm Adding Sample Streams
Section 2
This option is used when the instrument is measuring more than one sample.
Analyzing two or three sample streams creates more than eight values that
can be displayed. Since the AquaTrend only has eight channels, the operator
must choose which eight values should be displayed.
6. Use the arrow keys to scroll to the desired measurement type and press
ENTER. An asterisk will appear next to the enabled measurement type.
This measurement may then be connected to a channel on the
AquaTrend (see Section 2.9.4)
Page 64
Adding Sample Streams 51000-18 startup.fm
Section 2
For example, APA Alkalinity analyzers can measure from 1-3 sample streams.
The alkalinity measurements for each sample stream can be displayed as five
types of alkalinity. To display any or all of these alkalinity types requires that
each measurement type be assigned a channel on the AquaTrend (see
Section 2.9.3). These could be called 1TAlk, 1PAlk, and 1HyAlk. To assign a
channel to a measurement, use the AquaTrend menu as described below.
2. Use the up or down arrow key to select the AQUATREND MENU and
press ENTER.
5. Select ATTACH, and press ENTER. A list of available sensors will appear.
• If the sensor has more than one measurement available, select the
desired measurement and press ENTER.
The message OPERATION COMPLETE will appear. The screen will return to
the display setup menu upon completion of the measurement to channel
setup. See Section 3.2.1 on page 70 for more information about viewing
process analysis data.
Page 65
51000-18 startup.fm Adding Sample Streams
Section 2
Attaching a measurement to channel with one sensor
Page 66
Adding Sample Streams 51000-18 startup.fm
Section 2
2. Use the up or down arrow key to select the AQUATREND MENU and
press ENTER.
5. Select DETACH, and press ENTER. A list of attached channels will appear.
Page 67
51000-18 startup.fm Adding Sample Streams
Visit http: //www.hach.com
Section 3 Normal Operation
3.1 Measurements
The specific steps of a normal measurement cycle are detailed in Table 11.
The measurement cycle carries out a colorimetric analysis to make an
accurate measurement of alkalinity in the sample stream (also see
Section 1.5 on page 25).
During normal operation the 10-port Rotor Valve rotates to various
positions to move sample and reagents through the analyzer. Table 10 lists
the port numbers and their function for each APA 6000™ Alkalinity Process
Analyzer. Table 11 describes the valve position and autoburette actions during
a normal cycle.
Table 10 Port Functions For Alkalinity Analysis
Port # Function Port # Function
1 Detector 6 Standard 2
2 Sample 1 7 Reagent 1
3 Sample 2 8 Reagent 2
4 Sample 3 9 Mixer
5 Standard 1 10 Grab sample
Page 69
51000-18 normal operation.fm Normal Operation
Section 3
When viewing more than one channel, the display will not identify the sensor
that is sending measurement data. If the APA is not networked, the
measurement channel is associated only with that specific APA. If the APA is
networked, you may see the same channel numbers displayed, even though
the channels show data from different analyzers. For example, you may see
two 1ALKALINITY channels displayed. To identify the instrument that is
sending the displayed data, highlight the measurement channel and press
ENTER. The display will show all the information for that measurement. Press
ENTER again to return to the multiple display screen.
A cursor may be used to examine the graph data in detail. The lower and
upper concentration limits are programmable as are the time intervals.
2. 24 hour - displays the last 24 hours of data where each point is based on
an average measurement value during a 12-minute interval.
3. 30 days - displays the last 30 days of data, where each point is based on
an average measurement value during a 6-hour interval.
Page 70
Process Analysis 51000-18 normal operation.fm
Section 3
4. Select CHANNEL GRAPH SETTINGS from the DISPLAY SETUP MENU and
press ENTER. A list of available channels will appear.
6. Select TIMEBASE: and press ENTER. The pop-up screen shows: 60 min,
24 hrs, 30 days.
7. Select the desired time base and press ENTER to confirm the new setting
or press EXIT to cancel. The selected graph time will appear for a specific
channel on the x-axis on the graph each time a graph is displayed.
Page 71
51000-18 normal operation.fm Process Analysis
Section 3
4. Select CHANNEL GRAPH SETTINGS from the DISPLAY SETUP MENU and
press ENTER. A list of available channels will appear.
7. Use the arrow keys to change the displayed values. Press ENTER to
accept the value or EXIT to cancel.
8. Repeat Step 6 and Step 7 to change the YMIN: displayed value. Enter
negative numbers by changing the first character to a negative sign with
the arrow keys.
Time/date data for the cursor location will appear below the x-axis.
The measurement value at the cursor location will appear in the second field
from the top along the y-axis. The measurement value is an average of
several measurements. To examine the specific measurement data for the
point at the cursor location, press ENTER. The display will show the time, the
maximum and minimum concentration values and the average value. Press
EXIT to return to the graph.
Page 72
Process Analysis 51000-18 normal operation.fm
Section 3
Note: To filter sample, use a 60-cc syringe (Cat. No. 22587-00) and a 5-µm Acrodisc
(Cat. No. 22261-00). Draw sample into the syringe, attach the filter, then dispense
sample through the filter.
Page 73
51000-18 normal operation.fm Grab Sample Analysis
Section 3
8. Select MEASURE GRAB SAMPLE. The display will show PREPARING
SYSTEM FOR GRAB SAMPLE. PLEASE WAIT.
9. The display will read POUR GRAB SAMPLE. Remove the cap from the grab
sample block. Pour at least 100 mL of filtered sample into the grab sample
block (see Figure 17). Replace the cap.
10. The display will read MEASURING GRAB SAMPLE NOW. A pop-up menu
will display the time remaining to complete the grab sample analysis.
When the grab sample measurement is complete, the grab sample
reading will appear on the screen. The top reading is total alkalinity and
the bottom reading is P alkalinity
Page 74
Grab Sample Analysis 51000-18 normal operation.fm
Section 3
5. Select GRAB SAMPLE and press ENTER.
7. A pop-up box will appear with the most recent grab sample data, including
the time and date of the reading. Select REVIEW NEXT GRAB to step
through the eight most recent grab sample measurements. After the last
grab sample, press enter to return to the previous menu. Press Exit to
return to the grab sample history menu.
8. Pages that read ppm CaCO3 represent T alkalinity and pages that read just
ppm represent P alkalinity.
Page 75
51000-18 normal operation.fm Calibration Options
Section 3
2. Select SENSOR MENU and press ENTER.
The analyzer will wait until any current measurement or priming cycle is
complete before doing the calibration check. Calibration checks typically takes
30 minutes to complete. The maximum calibration check time is 40 minutes.
Page 76
Calibration Options 51000-18 normal operation.fm
Section 3
6. A pop-up box will appear with the calibration date and time of the seven
most recent calibrations. Select REVIEW NEXT CAL to step through the
“pages” of the calibration history. After the last calibration, press ENTER to
return to the previous menu. To exit the CAL HISTORY MENU, press EXIT.
The calibration is a multiple point calibration, but the display will only show
the lowest and highest point values, which correspond to PT1 and PT2.
P alkalinity values will appear on the second line; T alkalinity values will
appear on the third line. These values represent what the new point would
have read on the previous curve.
Page 77
51000-18 normal operation.fm Calibration Options
Section 3
6. A pop-up box will appear with the calibration check date and time of the
seven most recent calibrations. Select REVIEW NEXT CAL CHECK to step
through the “pages” of the calibration check history. After the last
calibration check, press ENTER to return to the previous menu. To exit the
CAL CHECK HISTORY MENU, press EXIT.
The display shows the calibration point checked in the first column and
the value measured during the calibration check in the second column.
The second line shows the T alkalinity value and the third line shows the
P alkalinity value.
Page 78
Calibration Options 51000-18 normal operation.fm
Section 3
The most recent alarm/warning is listed first. A bell icon to the left of the
warning indicates the alarm/warning is active. Selecting a specific
alarm/warning and pressing ENTER provides more detailed information about
that alarm/warning (i.e., time, date, value).
Page 79
51000-18 normal operation.fm Alarm Log Review
Section 3
To clean the wetted parts of the analyzer, the system default is set to run a
normal clean cycle every day. However, you may want to use this option in
other circumstances. During operation air bubbles may develop inside the
autoburette, tubing or the colorimeter. This is especially true of new
instruments where system components are not thoroughly wetted. In addition,
clean samples require more cleaning cycles because clean water does not
contain enough surfactants to keep the system thoroughly wetted. To clear
bubbles from the system, use the Normal Instrument Clean function. If
bubbles persist in a new analyzer, use the alternate Cleaning Solution
(Cat. No. 26974-53) for the first weeks of operation.
Note: The alternate cleaning The cleaning option dispenses the cleaning solution (an acid) through all the
solution is a weak surfactant and wetted parts in the analyzer. The cleaning operation has several functions:
will not eliminate metal precipitate
in the system. • It is used to prime the system (cleaning solution is pulled directly into
the autoburette).
Note: This alternative solution should only be used for a break-in period.
Page 80
Cleaning the Instrument 51000-18 normal operation.fm
Section 3
Note: Using the default settings is 7. Select one of the following:
strongly recommended by the
manufacturer and should only be • To clean the entire system immediately, select CLEAN SYSTEM NOW
changed if sample components and press ENTER. The analyzer will complete the current cycle, then
cause measurement problems. If perform the cleaning operation.
cleaning frequency is decreased,
verify that the cleanliness and • The system default is set to clean the analyzer every day.
operation of the analyzer is To change this default, select AUTO-CLEAN and press ENTER.
not degraded. Select the how often you want the analyzer to automatically clean
itself and press ENTER.
• To change the type of cleaning cycle, select NORMAL or EXTENDED.
Press ENTER. The analyzer will complete the current cycle, then
perform the cleaning operation.
If the air temperature surrounding the instrument becomes higher than the
temperature control point of the analyzer, the analyzer can still function but
performance may fall outside of specifications. Recalibrating at the air
temperature allows the measurements to remain accurate. Best accuracy is
achieved if a constant ambient temperature is maintained, even if it is above
the analyzer setpoint. If the analyzer is calibrated and measures at the same
temperature, readings will be accurate. If the control temperature changes
between calibration and measurement, accuracy will be reduced.
Page 81
51000-18 normal operation.fm Temperature Control
Section 3
If the analyzer will be shut down for an extended time, it is best to run a clean
cycle just before shutdown to fill all the wetted parts with cleaning solution and
help prevent fouling/biogrowth in the system. See Section 3.6 on page 80
for cleaning.
4. Select SENSOR STATUS and press ENTER. A pop-up menu will appear.
Use the arrow keys to select STANDBY. Press ENTER.
4. Select SENSOR STATUS and press ENTER. A pop-up menu will appear.
Use the arrow keys to select RESTART. Press ENTER.
Page 82
Shutting the Analyzer Down (Off or Standby) 51000-18 normal operation.fm
Maintenance
DANGER
Some of the following manual sections contain information in the form of warnings, cautions and notes
that require special attention. Read and follow these instructions carefully to avoid personal injury and
damage to the instrument. Only personnel qualified to do so, should conduct the maintenance tasks
described in this portion of the manual.
DANGER
Certains des chapitres suivants de ce mode d’emploi contiennent des informations sous la forme
d’avertissements, messages de prudence et notes qui demandent une attention particulière. Lire et suivre
ces instructions attentivement pour éviter les risques de blessures des personnes et de détérioration de
l’appareil. Les tâches d’entretien décrites dans cette partie du mode d’emploi doivent être seulement
effectuées par le personnel qualifié pour le faire.
PELIGRO
Algunos de los capítulos del manual que presentamos contienen información muy importante en forma de
alertas, notas y precauciones a tomar. Lea y siga cuidadosamente estas instrucciones a fin de evitar
accidentes personales y daños al instrumento. Las tareas de mantenimiento descritas en la presente
sección deberán ser efectuadas únicamente por personas debidamente cualificadas.
GEFAHR
Einige der folgenden Abschnitte dieses Handbuchs enthalten Informationen in Form von Warnungen,
Vorsichtsmaßnahmen oder Anmerkungen, die besonders beachtet werden müssen. Lesen und befolgen
Sie diese Instruktionen aufmerksam, um Verletzungen von Personen oder Schäden am Gerät zu
vermeiden. In diesem Abschnitt beschriebene Wartungsaufgaben dürfen nur von qualifiziertem Personal
durchgeführt werden.
PERICOLO
Alcune parti di questo manuale contengono informazioni sotto forma d’avvertimenti, di precauzioni e di
osservazioni le quali richiedono una particolare attenzione. La preghiamo di leggere attentivamente e di
rispettare quelle istruzioni per evitare ogni ferita corporale e danneggiamento della macchina. Solo gli
operatori qualificati per l’uso di questa macchina sono autorizzati ad effettuare le operazioni d’istallazione
e di manutenzione descritte in questa parte del manuale.
Page 83
51000-18 maintenance stopper.fm Maintenance
Visit http: //www.hach.com
Section 4 Routine Operator Maintenance
Page 85
51000-18 routine operator maintenance.fm Routine Operator Maintenance
Section 4
Hach recommends changing both standards at the same time. To change any
of these solutions see Section 2.5 on page 51.
Page 86
Reagent Consumption 51000-18 routine operator maintenance.fm
Section 4
Use the following procedure, Figure 18, and Figure 19 to replace the
autoburette seals:
1. Place the instrument in standby, then turn it off (see Section 3.8 on
page 82). This is when the piston is fully forward.
4. Remove the lock ring from the autoburette body; set aside.
Solenoid Valve
Page 87
51000-18 routine operator maintenance.fm Replacing Autoburette Seals
Section 4
5. Gently slide autoburette body assembly off of the piston by pulling it away
from the module base.
6. Remove seal gland nut using spanner wrench and set aside (see
Figure 20).
10. Push 2 new small O-rings onto each of the 2 new piston seals to make a
seal/O-ring assembly.
Page 88
Replacing Autoburette Seals 51000-18 routine operator maintenance.fm
Section 4
11. Insert one seal/O-ring assembly into the seal gland of pump block with the
Teflon base against the bottom of the seal gland.
12. Insert the large O-ring into the bottom of the seal gland, seating with the
seating tool.
13. Insert the seal spacer into the seal gland, with cross-drilled holes and
groove oriented toward the bottom of the seal gland.
14. Insert one large O-ring into the seal gland and seat against the
seal spacer.
Note: It is best to push on the 16. Install the seal gland nut until finger tight. Wet the piston with a towel
autoburette body assembly with the moistened with deionized water. Slide the pump body assembly partially
piston assembly tool after each onto the piston and tighten the seal gland nut, using a spanner wrench,
quarter turn of the seal gland nut to until the plunger on the force gauge bottoms out just before the piston
assure that the seals don’t become moves (see Figure 21).
over tightened.
17. Seat the autoburette body fully into the autoburette motor housing.
18. Reinstall the knurled housing lock ring until it is full tightened.
Lock Ring
Page 89
51000-18 routine operator maintenance.fm Replacing Autoburette Seals
Section 4
Note: There should NOT be any 19. Reinstall all four tubing connections.
movement of the autoburette body
once the lock ring is fully in place. 20. Reinstall the reagent line and bottle cap into the cleaning solution bottle.
Be sure the reagent line is at the bottom of the bottle.
1. Place the instrument in standby, then turn it off (see Section 3.8 on
page 82).
3. Break the siphon to the reagent bottles and standard bottles as follows:
a) Disconnect the reagent line from the reagent vent cap insert. b) Lift the
reagent line so the bottom of the reagent line is above the liquid level in
each reagent/standard bottle. c) Re-collar the reagent line in the vent cap
insert to prevent it from moving around.
Valve Stator
1
2 10
3
9
8
Spacer
4
5
6
Valve Rotor
Valve Body
Page 90
Maintaining the Valve Rotor and Stator 51000-18 routine operator maintenance.fm
Section 4
Note: Avoid contaminating the 4. Before removing the valve head, remove tubing to ports 4, 7 and 10 for
standards. wrench clearance (see Figure 23). There is no need to drain sample line 3
as the sample was turned off in step 2. The line to port 7 is the grab
sample line. This will not drip if there is no fluid in the grab sample port.
1 1
2 2
10 10
3
3
9
9
4
4
8
8
7
7
5 5
6 6
Note: Do not the scratch the 5. Using a 7/64-in. Allen wrench, loosen the three recessed socket head cap
surfaces of the rotor and stator. screws in the valve stator (valve head). Do not completely remove the
screws, but back them out far enough to remove the head. These screws
will help hold a ring spacer that lies under the valve head or stator. Be
careful not to lose this spacer. Pull the valve head away from the valve
module and inspect surfaces of the rotor and stator.
Note: A small amount of fluid may 6. Inspect the slot in the rotor for any plastic shavings or sample debris. If
drain from the stator. If the volume is these are visible, remove the rotor from its holder for cleaning. Carefully
larger, be sure the siphon to the clean this slot with a small stylus, or a soft bottle brush, taking care not to
reagent bottles was broken. scratch the valve surface.
7. Inspect the valve surface of the valve head/stator. If this surface is scored,
replace the valve head along with the rotor.
Reinstallation
Note: If you are not sure what 1. Before reinstalling the rotary valve, dry all surfaces with paper toweling.
position the valve is in, briefly turn Reposition the valve head with port 1 at 12 o’clock (top center) on the
the instrument on. The valve will valve body. Make sure the locating pins in the valve body engage the
move to port 1. Immediately shut the locating pin holes on the back side of the valve head/stator.
instrument off.
2. Hold the valve stator in place as you engage the three socket head cap
screws. Take care to evenly tighten these three screws. You will feel an
even resistance as the valve stator tightens down because a spring
washer is compressed during the rotor/stator mating. Tighten these
screws in even increments until tight (100 to 120 inch-ounces of torque).
Note: Do not over-tighten these 3. Re-attach lines 4, 7, and 10 to the valve head/stator. If the valve stator is
valve fittings. These fitting are replaced, connect the remaining lines to the new stator.
finger-tight fittings. Torque these
fitting to no more than 18 inch- 4. Return the raised reagent lines of the reagent/standard bottles to their
ounces (0.13 Joules of torque). original positions.
Page 91
51000-18 routine operator maintenance.fm Maintaining the Valve Rotor and Stator
Section 4
5. Start the analyzer. Go to the prime menu and select the
PRIME ENTIRE SYSTEM option (see Section 2.7.2 on page 55).
Recalibrate the instrument (see Section 3.4 on page 75).
2. Remove the fittings on either end of the tube from their ports. Note
whether it is a ¼-28 flat bottom fitting or a 10-32 rotary valve fitting.
3. Remove the fittings from the ends of the tubing and discard.
5. If the tube is an external tube, slide the rubber grommet onto the tubing
and insert into the appropriate position in the manifold on the inside
bottom of the instrument case.
6. Route the tubing to its appropriate place on the outside and inside of the
instrument (refer to Figure 26).
7. Install the nut onto the tubing, followed by the spacer and ferrule, as
shown in Figure 24 or Figure 25.
Page 92
Maintaining the Valve Rotor and Stator 51000-18 routine operator maintenance.fm
Section 4
Ferrule Tubing
#10-32 Nut
Detail A
Page 93
51000-18 routine operator maintenance.fm Maintaining the Valve Rotor and Stator
Section 4
1 10
2 APA 6000
HACH COMPANY
LOVELAND, CO
HOLDING COIL 2 1
5 4
2-Way 3 16
VALVE
OUT
MIXER 2
6 1 AUTO-
IN
BURETTE
7
15
CO AIN
CL
ST
Wa
MI
ST
RE
RE
DR
XE
LO
AN
AN
EA
AG
s
AG
te
RD
RIM
DA
DA
NE
E
E
NT
NT
8
R
R
R
RA
ET
D2
D1
2
IN
ER
9 11 12 13 14
Page 94
Maintaining the Valve Rotor and Stator 51000-18 routine operator maintenance.fm
Section 4
2. Put the instrument in standby, then turn the instrument off (see
Section 3.8 on page 82).
3. Notice the mixer housing is split. Loosen the two Phillips screws that hold
the top of the module housing in the module block.
4. Hold the part of the housing next to the block while pulling the top of the
housing out from the block. The part of the housing that holds the stir bar
should stay attached to the instrument. Let the top part of the housing
hang from the attached tubing.
5. Use a cotton swab soaked with a dilute bleach solution to clean the mixing
chamber and the module housing top. Rinse well with deionized water.
7. Reassemble the housing, making sure the O-ring and stir bar are in place.
8. Tighten the Phillips screws that hold the top of the housing to the block.
Page 95
51000-18 routine operator maintenance.fm Cleaning the Mixer Module
Section 4
Mounting Screw
2 Places
To Mixer Drain
To Rotary Valve
Stir Bar
20549-59 O-ring
40788-00
Page 96
Cleaning the Mixer Module 51000-18 routine operator maintenance.fm
Section 5 Warnings, Alarms, and
Troubleshooting
• Ensure the reagent line in the guide tube goes to the bottom of the
reagent 1 bottle.
• Check tubing at the rotary valve and make sure the tubing is filled with
fluid, not air.
• If ports on the rotary valve have been removed recently, refer to the
plumbing diagram and ensure the ports were reconnected properly.
• If reagent or standard is not delivered to the valve, check for kinks in the
tubing. Replace any tubing that is kinked.
• If reagents or standards are degrading, check the seal on the bottle caps
and the tubing connections. Replace if necessary.
• Watch the rotor valve movement during a cycle. If fluid is moving from the
common port to more than one outlet port, cross-leakage in the rotary
valve is occurring. Inspect the rotary valve per Section 4.5 on page 90. It
is likely that stator will need to be replaced.
• If fluid in the holding coil is not observed, the holding coil may be blocked.
Replace the valve module.
Page 97
51000-18 warnings alarms and troubleshooting.fm Warnings, Alarms, and Troubleshooting
Section 5
• If air is moving through the piston seals during operation, replace the
seals per Section 4.4 on page 87.
• If there is fluid leakage around the autoburette body, check the seals
using the Force Gauge Tool (see Section 4.4 on page 87). Tighten the
seals if they are loose. If they are not loose, replace them.
• Check the tubing for kinks which may cause reduced or blocked flow.
Reduced or blocked flow may also be caused by fittings that are
over-tightened.
• Examine the fittings for leakage and tighten if necessary. The fittings
should be fingertight.
Page 98
General Troubleshooting Procedure 51000-18 warnings alarms and troubleshooting.fm
Section 5
• The following tables describe the possible warnings and alarms, along
with causes, analyzer actions and corrective actions. Use these tables
and Section 5.1 to return the analyzer to proper operation.
Page 99
51000-18 warnings alarms and troubleshooting.fm Specific Alarm/Warning Messages
Section 5
Table 18 APA 6000 System Alarms (continued)
Alarm/Warning Cause Analyzer Action Corrective Action
The dipswitch is set to
Unknown dipswitch setting on inform the analyzer Verify dipswitch setting is correct. If
INVALID DIPSWITCH
APA main board. which chemistry problem persists, replace the main board.
to run.
NO xyz module
Internal network bus error; Verify module is installed correctly and
RESPONSE (where
software time-out occurred connections are secure. Restart the
xyz = colorimeter, Analyzer shuts down
while waiting for response analyzer. Replace module if problem
mixer, autoburette, or
from xyz module. persists.
rotary valve module)
An xyz module was not
Verify correct module is securely installed.
detected on the internal
NO xyz MODULE Analyzer shuts down Check cable connections. Verify correct
network bus, but is required for
chemistry is selected via dipswitch.
the selected chemistry.
An extra xyz module was
Verify correct chemistry is selected via
EXTRA xyz MODULE detected, but is not required Analyzer shuts down
dipswitch. Replace with correct module.
for the selected chemistry.
This xyz module type was Verify correct module software is being
WRONG MODULE
detected, but is not required Analyzer shuts down used. Verify correct chemistry is selected
(xyz)
for the selected chemistry. via dipswitch. Replace with correct module.
Cannot properly install xyz
INSTALL ERROR xyz Analyzer shuts down Replace module.
module.
Internal network bus error;
NO INT NEUR software time-out occurred Verify modules are securely installed.
Analyzer shuts down
RESPONSE while waiting for response Check cable connections.
from module.
Verify module is installed correctly. Check
Communication failed with xyz backplane cable connections. Restart
xyz NV FAIL Analyzer shuts down
module. analyzer. Replace module if problem
persists.
Page 100
Specific Alarm/Warning Messages 51000-18 warnings alarms and troubleshooting.fm
Section 5
Table 19 APA 6000 System Warnings (continued)
Alarm/Warning Cause Analyzer Action Corrective Action
Module xyz
temperature detector
Analyzer continues to If ambient temperature is well above this,
xyz TEMP VERY LOW measures below -10 °C
operate check the cable connections.
(clears when it is >-
5 °C).
Module xyz
temperature detector
Occurs on startup of analyzer and will
measure more than
clear after temperature control warms up
1 °C below the Analyzer continues to
xyz TEMP MARG LOW module to 50 °C. If ambient temperature
temperature setpoint. operate
is less than 5 °C, analyzer may not be
Warning clears when
able to get to temperature control range.
module temperature is
within 0.5 °C.
Module xyz
temperature detector Temperature is maintained at 40 °C by
measure more than 1°C heating. If ambient temperature is
above the temperature Analyzer continues to greater than 50 °C, this warning is
xyz TEMP MARG HIGH
setpoint. Warning clears operate displayed. If temperature changes are
when module rapid, analyzer may have trouble
temperature is within calibrating.
0.5 °C.
Module xyz
temperature detector
Analyzer continues to If ambient temperature is well below this,
xyz TEMP VERY HIGH measures above 58 °C
operate check all cable connections.
(clears when it is <
55 °C).
Module xyz
temperature is not
increasing to control Analyzer continues to Check cable connections. Shut heater off
xyz TEMP NOT MOVING
temperature setpoint operate and retry every 20 minutes.
with heater on full
power.
Analyzer restarts and tries
Unexpected xyz module to continue operation. If Verify module xyz is installed correctly.
xyz RESET reset (i.e., power loss or problem persists, it displays Check backplane cable connections.
low voltage). EXCESSIVE RE-INITS Replace module if problem persists.
alarm and shuts down.
Analyzer restarts and tries
to continue operation. If Check to ensure cable connections on
VALVE NO HOME Rotary valve cannot find
problem persists, it displays valve module are secure. Replace
SIGNAL the home location.
EXCESSIVE RE-INITS module if problem persists.
alarm and shuts down.
Analyzer restarts and tries
Valve home signal does
to continue operation. If Check to ensure cable connections on
not clear when valve
VLV HOME SIG ON problem persists, it displays valve module are secure. Replace valve
moves away from its
EXCESSIVE RE-INITS module if problem persists.
home position.
alarm and shuts down.
Analyzer restarts and tries
Rotary valve failed to Check valve for proper tightness and
to continue operation. If
arrive at its required obstructions to the valve rotor. Valve
VLV MOVE ERR problem persists, it displays
port after motor may be wearing out; replace valve
EXCESSIVE RE-INITS
request signal. module if problem persists.
alarm and shuts down.
Check valve for proper tightness and
Valve module required
Analyzer continues to obstructions to the valve rotor. Valve
VALVE STEPS ERROR extra steps to get to
operate motor may be wearing out; replace valve
final destination.
module if problem persists.
Page 101
51000-18 warnings alarms and troubleshooting.fm Specific Alarm/Warning Messages
Section 5
Table 19 APA 6000 System Warnings (continued)
Alarm/Warning Cause Analyzer Action Corrective Action
Analyzer restarts and tries
to continue operation. If Check for clogged lines in the analyzer.
Error in position
problem persists, and Check module cable connections.
BURETTEx HOME ERR detected since last
EXCESSIVE RE-INTS Replace Autoburette module if problem
cycle
alarm results and the persists.
analyzer shuts down.
Analyzer restarts and tries
Check for clogged lines in the analyzer.
to continue operation. If
The home position has Check module cable connections.
BURETTEx HOME DRIFT problem persists, it displays
drifted excessively. Replace Autoburette module if problem
EXCESSIVE RE-INITS
persists.
alarm and shuts down.
Home position is when autoburette is at
The module cannot find Analyzer restarts and tries
the fully dispensed position. Check
its home position (NO to continue operation. If
BURETTEx NO HOME module cable connections. If autoburette
HOME) or cannot move problem persists, it displays
BURETTEx HOME ON moves, check optical switch cable. If not,
away from its home EXCESSIVE RE-INITS
check motor cable connections. If
position (HOME ON). alarm and shuts down.
problem persists, replace Autoburette.
Autoburette system
hysteresis and Analyzer continues to Be sure seals are not too tight. If problem
BURETTEx HYST WARN
blacklash is greater operate persists, replace Autoburette module.
than 28 µL.
Analyzer main board Analyzer continues to May be problem with main board. Call
WATCHDOG TIMEOUT
processor times out. operate Hach Service.
Internal neuron cannot
INT NEUR HS TIME OUT communicate with main Analyzer shuts down If problem persists, call Hach Service.
board.
Analyzer tries to reconnect
Unexpected internal Cycle analyzer power. If problem
INT NEUR RESYNC with neuron and continue to
neuron reset occurred. persists, call Hach Service.
operate.
Analyzer tries to reconnect
Unexpected external Cycle analyzer power. If problem
EXT NEUR RESYNC with neuron and continue to
neuron reset occurred. persists, call Hach Service.
operate.
The AquaTrend was If problem persists, check cable
busy when analyzer Analyzer continues to connections. If communications between
AT NOT READY
tried to send operate the AquaTrend and analyzer are
a message. intermittent, call Hach Service.
If problem persists, check cable
The AquaTrend did not
Analyzer continues to connections. If communications between
AT NV FAIL respond after analyzer
operate the AquaTrend and analyzer are
sent message.
intermittent, call Hach Service.
External network bus
error - loss of Analyzer continues to
SIO NV FAIL Check cable connections to SIO.
communication with operate
SIO.
Voltage readings are made at the
VREF VOLT ERROR VDD
System voltage(s) are Analyzer continues to detector module. Check cable
VOLT ERROR VCC
outside specified limits. operate. connection at back panel. Ensure
VOLT ERROR
detector module is securely installed.
Attempt to perform
Analyzer continues to Be sure standard bottles contain
FAILED TO CAL CHECK check of calibration
operate. standard.
curve failed.
Page 102
Specific Alarm/Warning Messages 51000-18 warnings alarms and troubleshooting.fm
Section 5
Table 19 APA 6000 System Warnings (continued)
Alarm/Warning Cause Analyzer Action Corrective Action
A check of the
calibration curve
Analyzer continues to
CAL CHECK WARNING showed the check point Recalibrate the analyzer.
operate.
was outside the
specified range limit.
Analyzer continues to
Unable to complete a Ensure calibration standard bottles are
CAL FAILURE operate using previous
calibration. not empty.
calibration curve.
Analyzer continues to
Unable to complete a Ensure calibration standard bottles are
CAL STD OUT operate using previous
calibration. not empty.
calibration curve.
Detector A/D counts are
Analyzer continues to Replace colorimeter/detector module if
LED OUTPUT LOW below recommended
operate. problem persists.
operating range.
Detector A/D counts are
Analyzer continues to Replace colorimeter/detector module if
LED OUTPUT HIGH above recommended
operate. problem persists.
operating range.
Slope of calibration
MARG CAL (TALK Analyzer continues to
curve is outside nominal Check accuracy of calibration standards.
SLOPE) operate using new slope.
default slope.
Slope of calibration
MARG CAL (PALK Analyzer continues to
curve is outside nominal Check accuracy of calibration standards.
SLOPE) operate using new slope.
default slope.
P alkalinity calibration Analyzer continues to
CK STD2 P ACCURACY Check accuracy of calibration standards.
curve is questionable. operate using new slope.
Page 103
51000-18 warnings alarms and troubleshooting.fm Specific Alarm/Warning Messages
Section 5
5.3 Diagnostics
The diagnostic menu provides options to help the operator diagnose and
solve potential problems.
Page 104
Diagnostics 51000-18 warnings alarms and troubleshooting.fm
Section 5
7. Select the desired state for the relay and press ENTER.
Page 105
51000-18 warnings alarms and troubleshooting.fm Diagnostics
Section 5
Page 106
Diagnostics 51000-18 warnings alarms and troubleshooting.fm
Section 5
Page 107
51000-18 warnings alarms and troubleshooting.fm Diagnostics
Section 5
Page 108
Diagnostics 51000-18 warnings alarms and troubleshooting.fm
Installation
DANGER
Some of the following manual sections contain information in the form of warnings, cautions and notes
that require special attention. Read and follow these instructions carefully to avoid personal injury and
damage to the instrument. Only personnel qualified to do so, should conduct the installation/maintenance
tasks described in this portion of the manual.
DANGER
Certains des chapitres suivants de ce mode d’emploi contiennent des informations sous la forme
d’avertissements, messages de prudence et notes qui demandent une attention particulière. Lire et suivre
ces instructions attentivement pour éviter les risques de blessures des personnes et de détérioration de
l’appareil. Les tâches d’installation et d’entretien décrites dans cette partie du mode d’emploi doivent être
seulement effectuées par le personnel qualifié pour le faire.
PELIGRO
Algunos de los capítulos del manual que presentamos contienen información muy importante en forma de
alertas, notas y precauciones a tomar. Lea y siga cuidadosamente estas instrucciones a fin de evitar
accidentes personales y daños al instrumento. Las tareas de instalación y mantenimiento descritas en la
presente sección deberán ser efectuadas únicamente por personas debidamente cualificadas.
GEFAHR
Einige der folgenden Abschnitte dieses Handbuchs enthalten Informationen in Form von Warnungen,
Vorsichtsmaßnahmen oder Anmerkungen, die besonders beachtet werden müssen. Lesen und befolgen
Sie diese Instruktionen aufmerksam, um Verletzungen von Personen oder Schäden am Gerät zu
vermeiden. In diesem Abschnitt beschriebene Installations- und Wartungsaufgaben dürfen nur von
qualifiziertem Personal durchgeführt werden.
PERICOLO
Alcune parti di questo manuale contengono informazioni sotto forma d’avvertimenti, di precauzioni e di
osservazioni le quali richiedono una particolare attenzione. La preghiamo di leggere attentivamente e di
rispettare quelle istruzioni per evitare ogni ferita corporale e danneggiamento della macchina. Solo gli
operatori qualificati per l’uso di questa macchina sono autorizzati ad effettuare le operazioni di
manutenzione descritte in questa parte del manuale.
Page 109
51000-18 installation stopper.fm Installation
Visit http: //www.hach.com
Section 6 Sample Conditioning
! DANGER DANGER
This instrument should be Cet appareil doit être installé par du personnel technique qualifié, afin d’assurer
installed by qualified technical le respect de toutes les normes applicables d’électricité et de plomberie.
personnel to ensure adherence to
all applicable electrical and PELIGRO
plumbing codes. Este instrumento debe ser instalado por personal técnico capacitado para
assegurar el cumplimiento con todos los códigos eléctricos y de plomería
aplicables.
GEFAHR
Um zu gewährleisten, daß alle elektrischen und sanitärinstallationstechnischen
VDE-Vorschriften und gegebenefalls die Zusatzvorschriften der zuständigen
Elektrizitäts- und Wasserwerke erfüllt werden, darf dieses Gerät nur von
geschultem Fachpersonal installiert werden.
PERIGO
Este instrumento deve ser instalado por pessoal técnico qualificado para
asegurar o cumprimento de todas as normas elétricas e de canalização
aplicáveis.
! DANGER DANGER
This instrument is not intended Cet instrument n’est pas conçu pour une utilisation avec des échantillons
for use with flammable samples inflammables ou des échantillons contenant des hydrocarbures.
or those containing
hydrocarbons. PELIGRO
Este instrumento no está destinado para uso con muestras inflamables o que
contengan hidrocarburos.
GEFAHR
Dieses Gerät nicht verwenden bei Tests mit brennbaren Lösemitteln oder
Lösemitteln mit Kohlenwasserstoffen.
PERIGO
Este instrumento não é feito com o fim de ser empregado com amostras
inflamáveis ou aquelas que contêm hidrocarbonetos.
Page 111
51000-18 sample conditioning.fm Sample Conditioning
Section 6
The 22-µm Y-strainer filter protects the analyzer’s components and tubing by
allowing only particles smaller than 22 µm to pass on to the instrument
(a 22-µm particle is at the limit of visibility by the human eye). Use the ¼-inch
ball valve (refer to Figure 31 on page 117, item 21) to restrict and control the
flow rate of filtered sample from the filter through the sample inlet block (item
20). Install another on/off valve upstream of the filter for occasions when the
sample needs to be turned off or on. The filter housing valve (item 2) can be
opened to flush the unfiltered side of the filter element, or adjusted in any
partially open position for continuous bypass.
For samples that tend to clog the analyzer filter or tubing, additional
conditioning steps are required (see Section 6.2).
Some different applications for bypassing unfiltered sample through the filter
are as follows:
• Low to high flow bypass will let occasional sample upsets or slugs of
algae bypass through the filter rather than clogging the filter.
• High flow bypass will allow some self-cleaning of the filter as well as
provide the previously mentioned benefit of keeping the filter clear when
occasional sample upsets or slugs of algae occur.
• Unfiltered bypass flow may be used as a fast sample loop. If the sample is
a distance away from the instrument, the sample can be pumped through
a pipe to the filter. Unfiltered sample can then be returned back to the
sample site with pipe. This creates a fast flowing sample loop which can
minimize lag time between the process stream and the analyzer,
depending on the pump flow rate and pipe size selected. Use this concept
with clean or dirty water.
Page 112
Sample Conditioning 51000-18 sample conditioning.fm
Section 6
If the 22 µm filter clogs frequently, even when using the high flow unfiltered
bypass option, an air backflush kit may solve these problems. The 22-µm filter
will not collapse when using the air backflush option.
Installation
See Section 2.4.1.1 on page 49 to enable this option. The analyzer
automatically controls the timing of the air backflush by opening/closing the air
valve (item 28 in Figure 31) supplied with the kit. Electrical installation of the
air backflush is illustrated in Figure 28, Figure 29, and Figure 30. Only a
qualified technician should perform the electrical installation.
See Figure 31 on page 117 for plumbing the air backflush kit. Install the gray
female T-fitting (item 25) between the filter housing and the ¼-inch ball valve
(replace item 7 of the basic conditioning kit with item 25). The other
components of the air backflush kit attach to item 25. All the other
components are part of the Basic Water Conditioning Kit which must be
installed to route sample to the analyzer.
Page 113
51000-18 sample conditioning.fm Adding To or Changing the Basic Sample Conditioning
Section 6
Page 114
Adding To or Changing the Basic Sample Conditioning 51000-18 sample conditioning.fm
Section 6
Page 115
51000-18 sample conditioning.fm Adding To or Changing the Basic Sample Conditioning
Section 6
Table 20 Parts List for Figure 31
Item Number Description Catalog Number
1 Screw, 6-32 x 1.5, pan head (2) 5593-11
2 Screw, 6-32 x 2.5, pan head (2) 19495-00
3 Fitting, elbow, ½-in. ID tubing x ¾-in. NPT (1) 30618-00
4 Fitting, reducer, ¾-inch NPT x ¼-inch FNPT (3) 31560-00
5 Teflon Tubing, 0.170 ID x 0.250 OD (10 feet) 45462-00
6 Tygon Tubing, ¾-in. OD x 1/8-in. wall (2 feet) 45543-00
7 Clamp, conduit hanger, 1-in. (2) 47349-00
8 Coupling, ¾-in. NPT x ¾-in. NPT (3) 49620-00
9 Manifold Grommet (1) 51036-00
10 Tee Fitting, 1 x 1 SOC x ¾-in. FNPT (1) 51080-00
11 Hex Nut Fitting, for 0.062 OD tubing (1) 51129-00
12 Ferrule Fitting, for 0.062 OD tubing (1) 51130-00
13 Flangeless Nut Fitting, for 0.062 OD tubing (1) 51131-00
14 Ferrule Fitting, for 0.062 OD tubing (1) 51132-00
15 Clean Water Label (affix to sample line) (1) Part of 51139-00
16 Tefzel Tubing, 0.030 ID x 0.062 OD (3 feet) 51178-00
17 PVC Pipe, 1-in. (1) 51239-00
18 Male Connector for ¼-in. tubing (3) 51246-00
19 Tube Fitting, ¼-in. tube x stem elbow (1) 51362-00
20 Sample Inlet Block (1) 51365-00
21 Ball Valve, ¼-in. NPT, PVC (1) 51395-00
22 Strainer Housing, ¾-in., without filter (1) 51396-00
23 Filter Element, 450 mesh (1) 51397-00
24 Teflon Thread Tape 70608-24
25 Female Tee Fitting, ¾-i.h NPT, gray (1) 31553-00
26 In-line Check Valve, for ¼-in. OD tubing 51172-00
27 Fitting, Tube, ¼-in. OD to ¼-in. OD 51179-00
28 2-way Air Solenoid Valve 51181-00
Page 116
Adding To or Changing the Basic Sample Conditioning 51000-18 sample conditioning.fm
Section 6
Figure 31 Assembly for Air Backflush and High-flow Unfiltered Sample Bypass
Page 117
51000-18 sample conditioning.fm Adding To or Changing the Basic Sample Conditioning
Section 6
Page 118
Adding To or Changing the Basic Sample Conditioning 51000-18 sample conditioning.fm
Section 6
To turn the submersible sample pump off and on each cycle, you can program
one of the analyzer alarm relays to turn the power to the submersible sample
pump on and off. Use the analyzer alarm relays to control an appropriate
power relay which supplies power to the pump in use. See Section 2.4.1.2 on
page 50 to enable this option, and Figure 33 for installation. Also refer to
Section 7.3.2 on page 133 for information about the alarm contact
requirements and Figure 34, Figure 35, and Figure 36 for typical sample
pump connections.
Page 119
51000-18 sample conditioning.fm Adding To or Changing the Basic Sample Conditioning
Section 6
Figure 35 Sample Pump Control Using An Intrinsically Isolated Solid State Relay*
Page 120
Adding To or Changing the Basic Sample Conditioning 51000-18 sample conditioning.fm
Section 6
Figure 36 Sample Pump Control Using the Voltage From the APA 4-20 mA Recorder Output For the
Control Voltage On a Solid State Relay*
1. Follow the steps for setting up alarm relays in Section 2.4.1.2 on page 50.
Choose one of the configured channels that receives sample from the
sample stream which the sample pump controls. For instance, if
measuring from sample stream 1, select that measurement as the alarm
relay between the recorder output and the solid state relay.
Page 121
51000-18 sample conditioning.fm Adding To or Changing the Basic Sample Conditioning
Section 6
2. Remove the two screws holding the sample inlet block to the side of
the analyzer.
3. Place the sample inlet block (item 20) from the kit on top of the original
inlet block. Install the next length screw from the kit through the holes in
both blocks and fasten to the side of the analyzer.
4. Open the analyzer door. Remove a blank rubber grommet from the
bulkhead on the same side of the analyzer as the sample inlet block.
5. Thread the tubing (item 16) through the manifold grommet (item 9). Push
the grommet into the empty hole in the bulkhead.
6. Thread the tubing out through one of the holes next to the original sample
tubing in the back of the analyzer.
7. Remove the plug at port 3 of the rotary valve. Put a hex nut fitting (item
11) and ferrule (item 12) on the tubing. Install it in port 3 of the rotary
valve. See Section 4.5.1 on page 92 for an illustration of this assembly.
8. Put the flangeless nut fitting (item 13) and ferrule fitting (item 14) on the
tube. Put the Sample 2 label on the tubing. Screw the nut and ferrule into
the sample inlet block.
If the APA 6000 relays are used to control air backflush or submersible pump
cleaning, it is possible to use these options on the added sample streams.
The APA internal alarm relays can be used in parallel to control them. See
Section 2.4.1.1 on page 49 and Section 2.4.1.2 on page 50 to set up these
options. Also refer to Figure 31 on page 117 and Figure 54 on page 151.
Page 122
Installing Multiple Sample Streams 51000-18 sample conditioning.fm
Section 7 Hardware Installation
To control the analyzer from a location other than the instrument location,
connect a Remote AquaTrend. The Remote AquaTrend can be mounted on a
bench, pole or panel as far away as 1320 feet (400 m).
AquaTrend™
DISPLAY EXIT
MENU ENTER
Page 123
51000-18 hardware installation.fm Hardware Installation
Section 7
7.2 Mounting
Secure the rear of the instrument to the bench or table with two ¼-20 UNC
screws to prevent the instrument from tipping while the module panel is open.
3. Install the top two 3/8-16 UNC hexbolts, leaving a ½-inch gap under each
bolt head.
4. Slide the bracket keyholes onto the bolts and lower the frame so the bolts
settle in the narrow part of the keyholes.
Note: Be careful to not crimp or 5. Slide the instrument into the panel cutout so the back of the bezel is flush
compress the analyzer tubing when with the front of the panel.
panel-mounting the analyzer.
6. Install two #10-16 x 0.75-in. long self-tapping screws into the top
mounting holes (see Figure 39).
7. Install four #10-16 x 0.75-in. long, self-tapping screws into the side
mounting holes. Secure the reagent brackets with these screws also (see
Figure 39).
Page 124
Mounting 51000-18 hardware installation.fm
Section 7
8. Make all electrical connections described in Section 7.3.
10. Mount the grab sample block to the front panel routing the tubing through
panel and into the tubing guide.
Page 125
51000-18 hardware installation.fm Mounting
Mounting
Page 126
Section 7
17.086" (433.97 mm)
14.50" (368.30 mm)
14.00" (355.60 mm)
10.00" (254.00 mm)
.50" (12.70 mm) 4.00" (101.60 mm)
.31" (7.87 mm) 6.88" (174.77 mm)
1.90"
Cutout 8 Places
for
Instrument
1.150"
(29.21 mm)
15.00" (381.00 mm)
.725"
(18.42 mm)
.362"
.438" (9.21 mm)
(11.11 mm)
Cutout
(55.58 mm)
for
51000-18 hardware installation.fm
2.188"
Grab
Sample
.156" Block
(3.96 mm)
2 Places
.375"
(9.52 mm)
Section 7
Panel
Page 127
51000-18 hardware installation.fm Mounting
Section 7
GEFAHR
Alle elektischen Anschlüsse müssen van qualifizierten Techniekem
durchgeführt werden, damit die Übereinstimmung mit allen elektrischen
Anschlussdaten gewärhrleistet ist.
PELIGRO
Todas as lagções eléctricas devem ser feitas por técnicos qualificados, para
assegurar o cumprimento de todos os regulamentos em vigor.
Four access holes for electrical connections are provided along the back of
the instrument enclosure (see Figure 44). Remove the cover seals by
loosening the wing nuts from inside the instrument.
This product is designed to meet NEMA 4X (INDOOR) and IEC 529 I.P. 66
enclosure requirement when the front door is securely latched. Achieving
these protection levels, however, depends on the use of proper hardware in
the installation of this product. Standard off-the-shelf NEMA-approved conduit
hardware is required for proper installation of all electrical connections.
The power cord or power cable is not supplied with the APA 6000
Analyzer and must be selected to meet applicable electrical codes for
the installation site.
Page 128
Electrical Connections 51000-18 hardware installation.fm
Section 7
Hach recommends that care be taken with hook-up configurations other than
shown in these illustrations. First, be sure your configuration satisfies local
electrical codes and second, ensure that your configuration does not cause
susceptibility to lightning, power surges, or transients upsets.
Figure 41 Looking Inside the APA 6000, with Hinged Module Panel Down
Door Open
THIS CLASS A DIGITAL APPARATUS CET APPAREIL NUMERIQUE DE LA WARNING: FOR CONTINUED ADVERTISSEMENT: POUR UNE
NEMA 4X (INDOOR), IIP66
Instrument
PROTECTION AGAINST FIRE PROTECTION CONTINUE CONTRE
MEETS ALL REQUIREMENTS OF THE CLASSE A RESPECTE TOUTES LES HAZARD REPLACE ONLY WITH LE FEU REMPLACER LE FUSIBLE
CANADIAN INTERFERENCE CAUSING EXIGENCES DU REGLEMENT SUE LA SAME TYPE AND RATING OF PAR UN DU MEME TYPE ET
REGULATIONS. MATERIEL BROUILLEUR DU CANADA. FUSE DE MEMES CARACTERISTIQUES
! F1 & F2
On/Off NETWORK
A B
RECORDER
A
+ -
B
+ -
A
ALARM
NO COM NC NO COM NC
B
BLACK
1
WHITE
2
T, 1.6A, 250V
Switch J4?
Steel
Restraini
Cable
Module Panel
(Shown in
Open Position)
Page 129
51000-18 hardware installation.fm Electrical Connections
Section 7
Figure 42 Looking Inside the APA 6000, Removing the Customer Access Cover
Wingnuts
Door Open
THIS CLASS A DIGITAL APPARATUS CET APPAREIL NUMERIQUE DE LA WARNING: FOR CONTINUED ADVERTISSEMENT: POUR UNE
NEMA 4X (INDOOR), IIP66
Instrument
PROTECTION AGAINST FIRE PROTECTION CONTINUE CONTRE
MEETS ALL REQUIREMENTS OF THE CLASSE A RESPECTE TOUTES LES HAZARD REPLACE ONLY WITH LE FEU REMPLACER LE FUSIBLE
CANADIAN INTERFERENCE CAUSING EXIGENCES DU REGLEMENT SUE LA SAME TYPE AND RATING OF PAR UN DU MEME TYPE ET
REGULATIONS. MATERIEL BROUILLEUR DU CANADA. FUSE DE MEMES CARACTERISTIQUES
! F1 & F2
On/Off NETWORK
A B
RECORDER
A
+ -
B
+ -
A
ALARM
NO COM NC NO COM NC
B
BLACK
1
WHITE
2
T, 1.6A, 250V
Switch J4?
Page 130
Electrical Connections 51000-18 hardware installation.fm
Section 7
Figure 43 Looking Inside the APA 6000, Locating and Identifying the Customer Connections
J11
Network Terminator
(enabled)
Door Open
A B
A
+ -
B
+ -
A
NO COM NC NO COM NC
B
BLACK
1
WHITE
2
! Switch
J11
J10
Power
Terminal
Strip
Ground Stud
Module Panel is not J7 Ground J8 J9
shown in this illustration. Network Stud Recorder Alarm
Terminal Terminal Terminal
Strip Strip Strip
Drain
Page 131
51000-18 hardware installation.fm Electrical Connections
Section 7
External
5 Amp
Circuit Breaker
THIS CLASS A DIGITAL APPARATUS CET APPAREIL NUMERIQUE DE LA WARNING: FOR CONTINUED ADVERTISSEMENT: POUR UNE
NEMA 4X (INDOOR), IIP66 PROTECTION AGAINST FIRE PROTECTION CONTINUE CONTRE
MEETS ALL REQUIREMENTS OF THE CLASSE A RESPECTE TOUTES LES HAZARD REPLACE ONLY WITH LE FEU REMPLACER LE FUSIBLE
CANADIAN INTERFERENCE CAUSING EXIGENCES DU REGLEMENT SUE LA SAME TYPE AND RATING OF PAR UN DU MEME TYPE ET
REGULATIONS. MATERIEL BROUILLEUR DU CANADA. FUSE DE MEMES CARACTERISTIQUES
! F1 & F2
J11
Permanent
Connection (Conduit)
#8 Ground Stud
Instrument On/Off Switch
Approved NEMA
style strain relief
THIS CLASS A DIGITAL APPARATUS CET APPAREIL NUMERIQUE DE LA WARNING: FOR CONTINUED ADVERTISSEMENT: POUR UNE
NEMA 4X (INDOOR), IIP66 PROTECTION AGAINST FIRE PROTECTION CONTINUE CONTRE
MEETS ALL REQUIREMENTS OF THE CLASSE A RESPECTE TOUTES LES HAZARD REPLACE ONLY WITH LE FEU REMPLACER LE FUSIBLE
CANADIAN INTERFERENCE CAUSING EXIGENCES DU REGLEMENT SUE LA SAME TYPE AND RATING OF PAR UN DU MEME TYPE ET
REGULATIONS. MATERIEL BROUILLEUR DU CANADA. FUSE DE MEMES CARACTERISTIQUES
! F1 & F2
J11
Page 132
Electrical Connections 51000-18 hardware installation.fm
Section 7
GEFAHR
Die Ralaisanschlüsse sind entweder für Niederspannung (< 30V) oder für
Hochspannung (>30 V) vorgesehen. Wenn sowohl Nieder-als auch
Hochspannung an die Relais angeschlossen wird, kann es zu einem
Stromschlag kommen.
PERIGO
A área de ligaçõ do relé é designada só para ligações de baixa voltagem (<30 V)
ou só para ligações de alta voltagem (>30 V). Dá-se um curto-circuito se fizerem-
se simultaneamente ligações de baixas e altas voltagens na área de ligaçõa do
relé.
Note: If voltages >30 Vrms are The Analyzer contains two alarm relays designed for use with either high
connected to the relay contacts, you voltage (greater than 30V-RMS and 42.2V-PEAK or 60 VDC) or low voltage
must limit current to the relay (less than 30V-RMS and 42.2V-PEAK, or less than 60 VDC), but not a
contacts to 5 amps and be able to combination of both high and low voltage. Hach recommends using these
remove power from the relays relays to control air backflush or submersible pumps.
locally in case of an emergency or
for servicing of the product. This can
Wire alarm relays to the connectors labeled Alarm A or Alarm B on the back
be accomplished with an external
switch and a 5-amp fuse or with a electrical panel (see Figure 45). SPDT relay contacts are unpowered and
switched 5-amp circuit breaker. rated to carry up to 5 amperes at 240 VAC with a resistive load. If high-current
or inductive-load devices are to be controlled by analyzer alarms, use a
heavy-duty slave relay actuated by the alarm relay. Both normally open and
normally closed contacts are available. The connector will accept 22- to 12-
gauge wire, insulated as appropriate for the voltage. Size the wires according
to the load and local electrical code regulations.
If more relays are desired, supplement your system with the AquaTrend
Signal Output Module (SOM). Each SOM provides two additional relays.
See Section 8.1 on page 155.
Page 133
51000-18 hardware installation.fm Electrical Connections
Section 7
Approved NEMA
style strain relief
NETWORK RECORDER
A B A
ALARM
B
F1 & F2
J11
Relay
Permanent
Connection
(Conduit)
#8 Ground Stud
2 Places
GEFAHR
Spannungsversorgungsanlagen führen evlt. Erdpotential, wenn eine schlechte
Systemerdung vorgenommen wurde. Vosicht ist geboten, wenn das
angeschlossene Gerät an verschiedenen Orten geerdet ist. Die Anweisungen in
diesem Abschnitt sind genau zu befolgen.
PERIGO
Os sistemas eléctricos de aparelhos podem densenvolver tensões de massa se
forem utilizadas técnicas de ligação à terra fracas. Para reduzir a possibilidade
de perigo de choques, há que ter cuidado em ligar o equipamento à terra em
lugares diferentes. Siga escrupulosamente as instruções contidas neste
manual.
Page 134
Electrical Connections 51000-18 hardware installation.fm
Section 7
The analyzer contains two isolated 4-20 mA analog outputs which can drive
a recorder or control a process (see Figure 46). Recommended output
impedance is 3.6 to 500 ohms. Route 22-to 12-gauge recorder wires through
the access hole below the terminal strip and connect them to the terminal
strip J4, labelled “Recorder”. Use shielded, twisted-pair cable to connect
the analyzer to the recorder. Tie the shield to earth ground only at the
analyzer end to avoid ground loops. To ensure the analyzer meets
NEMA 4X and I.P. 66 enclosure ratings, use only the supplied NEMA-style
enclosure strain relief fittings.
If more analog outputs are desired, you can supplement your system with the
AquaTrend Signal Output Module (SOM). Each SOM provides two additional
analog outputs.
THIS CLASS A DIGITAL APPARATUS CET APPAREIL NUMERIQUE DE LA WARNING: FOR CONTINUED ADVERTISSEMENT: POUR UNE
NEMA 4X (INDOOR), IIP66 PROTECTION AGAINST FIRE PROTECTION CONTINUE CONTRE
MEETS ALL REQUIREMENTS OF THE CLASSE A RESPECTE TOUTES LES HAZARD REPLACE ONLY WITH LE FEU REMPLACER LE FUSIBLE
CANADIAN INTERFERENCE CAUSING EXIGENCES DU REGLEMENT SUE LA SAME TYPE AND RATING OF PAR UN DU MEME TYPE ET
REGULATIONS. MATERIEL BROUILLEUR DU CANADA. FUSE DE MEMES CARACTERISTIQUES
! F1 & F2
J11
Strip Recorder
Page 135
51000-18 hardware installation.fm Electrical Connections
Section 7
GEFAHR
Spannungsversorgungsanlagen führen evlt. Erdpotential, wenn eine schlechte
Systemerdung vorgenommen wurde. Vosicht ist geboten, wenn das
angeschlossene Gerät an verschiedenen Orten geerdet ist. Die Anweisungen in
diesem Abschnitt sind genau zu befolgen.
PERIGO
Os sistemas eléctricos de aparelhos podem densenvolver tensões de massa se
forem utilizadas técnicas de ligação à terra fracas. Para reduzir a possibilidade
de perigo de choques, há que ter cuidado em ligar o equipamento à terra em
lugares diferentes. Siga escrupulosamente as instruções contidas neste
manual.
Note: The network supports The APA analyzer can be connected into a network to communicate with other
distances of up to 1650 ft sensors/analyzers and accessory modules. Refer to Table 22 to determine
(500 meters) with a maximum what wiring connections are necessary.
distance between any two devices
of 1320 ft (400 meters).
Page 136
Electrical Connections 51000-18 hardware installation.fm
Section 7
7.3.4.1 Connecting the APA 6000 Analyzer To a Serial Input/Output Module (SIO)
Refer to the Installation section of the SIO Module Manual for SIO electrical
connections. In Network Connections, the SIO manual refers you to connect
into the network via the appropriate sensor manual. At this point, proceed to
Section 7.3.4.5 on page 142 of this manual. See Figure 47 for an illustration
of the connection.
Page 137
51000-18 hardware installation.fm Electrical Connections
Section 7
7.3.4.2 Connecting the APA 6000 Analyzer to a Signal Output Module (via the PS1201)
3. Connect the SOM(s) relays and analog outputs as directed in the SOM
Manual in SOM Power and Network Connections.
Page 138
Electrical Connections 51000-18 hardware installation.fm
Section 7
1. Make electrical connections at the SIO. Refer to the SIO Manual for
electrical connections. The Network Connections section of the SIO
Manual refers you to connect into the network via the appropriate sensor
manual. At that point, proceed to Step 2, below.
2. Connect the other end of the SIO network cable to the PS1201 Power
Supply Module as directed in the PS1201 Manual in the SIO to PS1201
Connection section.
Page 139
51000-18 hardware installation.fm Electrical Connections
Section 7
7.3.4.4 Connecting the APA 6000 Analyzer to Other Sensors, an SIO Module and SOMs
Use the following guidelines to connect the analyzer into the network.
• The APA 6000 has one opening for the network cabling connection. If only
connecting two APA 6000 analyzers together with no other sensors, refer
to Section 7.3.4.5 on page 142 and connect them directly to each other.
Page 140
Electrical Connections 51000-18 hardware installation.fm
Section 7
Page 141
51000-18 hardware installation.fm Electrical Connections
Section 7
The Cable Termination Kit is supplied with the SIO, SOM, or PS1201
Modules. If connecting the APA 6000 to another APA 6000, order this kit and
cable along with the additional APA 6000.
2. Open the APA 6000 Door and remove the screws on the Hinged Panel
holding the modules to access the electrical connections at the back of
the instrument. Refer to Figure 41 on page 129.
3. Remove the two lower wingnuts on the lower back panel, then remove the
lower terminal strip cover. Refer to Figure 42 on page 130.
4. Remove the Wire Access Plug below the Network terminal strip where the
network cable will be routed into the APA 6000. Refer to Figure 43 on
page 131.
5. Take one end of the network cable and strip 6 inches (15 cm) of outer
insulation from the cable.
6. Put the strain relief and sealing washer (supplied with the cable) over the
cable and slide the strain relief and sealing washer 7 inches down from
the end of the cable. Tighten the strain relief nut until the cable does not
slip through the strain relief. See Figure 51 on page 144.
Note: Connect shield at one end of 7. Remove ¼ inch of insulation from the white and green wires. Either
cable only. If connecting to another insulate the shield wire with appropriate length of heat shrink tubing or
analyzer, then connect shield in only remove ¼ inch of insulation from the shield wire.
one of the analyzers. Insulate the
non-connected end with tape or 8. Add a insulated quick connect to each wire using a crimping tool.
heat-shrink tubing.
9. Pull the cable through the APA 6000 wire access opening until the strain
relief /sealing washer is pulled flush with the opening. From inside the
analyzer, screw the locknut onto the strain relief and tighten.
10. Connect the white wire to the APA 6000 position “A” on the Network
terminal block. Connect the green wire to position “B” on the Network
terminal block. If connecting the shield wire, connect the ground wire of
the cable to the terminal stud near the network terminal block.
11. If the AquaTrend in the analyzer will be the master AquaTrend or if the
analyzer is not connected to a network, connect network terminator by
Page 142
Electrical Connections 51000-18 hardware installation.fm
Section 7
placing the jumper over pins 2 and 3 on the J11 connector (see
Figure 50). If this is not the master AquaTrend, place the jumper over pin
2 (See J11 Disabled, below).
12. Replace the lower terminal strip cover over the electrical connections and
install the two wingnuts to secure the cover.
13. Close the Hinged Panel holding the modules and reinsert and tighten
screws holding the module door shut.
THIS CLASS A DIGITAL APPARATUS CET APPAREIL NUMERIQUE DE LA WARNING: FOR CONTINUED ADVERTISSEMENT: POUR UNE
NEMA 4X (INDOOR), IIP66 PROTECTION AGAINST FIRE PROTECTION CONTINUE CONTRE
MEETS ALL REQUIREMENTS OF THE CLASSE A RESPECTE TOUTES LES HAZARD REPLACE ONLY WITH LE FEU REMPLACER LE FUSIBLE
CANADIAN INTERFERENCE CAUSING EXIGENCES DU REGLEMENT SUE LA SAME TYPE AND RATING OF PAR UN DU MEME TYPE ET
REGULATIONS. MATERIEL BROUILLEUR DU CANADA. FUSE DE MEMES CARACTERISTIQUES
! F1 & F2
J11
3 3 3
1 1 1
Page 143
51000-18 hardware installation.fm Electrical Connections
Section 7
J11
Network Terminator
(enabled)
Door Open
A B
A
+ -
B
+ -
A
NO COM NC NO COM NC
B
BLACK
1
WHITE
2
! Switch
J11
J10
Power
Terminal
Strip
Ground Stud
Module Panel is not J7 Ground J8 J9
shown in this illustration. Network Stud Recorder Alarm
Terminal Terminal Terminal
Strip Strip Strip
Fuse Holders
THIS CLASS A DIGITAL APPARATUS CET APPAREIL NUMERIQUE DE LA WARNING: FOR CONTINUED ADVERTISSEMENT: POUR UNE
NEMA 4X (INDOOR), IIP66 PROTECTION AGAINST FIRE PROTECTION CONTINUE CONTRE
MEETS ALL REQUIREMENTS OF THE CLASSE A RESPECTE TOUTES LES HAZARD REPLACE ONLY WITH LE FEU REMPLACER LE FUSIBLE
CANADIAN INTERFERENCE CAUSING EXIGENCES DU REGLEMENT SUE LA SAME TYPE AND RATING OF PAR UN DU MEME TYPE ET
REGULATIONS. MATERIEL BROUILLEUR DU CANADA. FUSE DE MEMES CARACTERISTIQUES
! F1 & F2
J4?
Page 144
Electrical Connections 51000-18 hardware installation.fm
Section 7
GEFAHR
Wenn der Gerätestrom und der Strom an den Relaiskontakten mehr als 30 Volt
beträgt, muss die Stromzuführung unterbrochen werden, bevor die
Eingangsssicherungen des Geräte eingesetzt bzw. entfernt werden. Ein
Fernleistungschalter oder ein Abschalter müssen benutzt werden. Durch den
Stromschalter des Analysators wird der Strom nicht von den Sicherungen
genommen.
PERIGO
Se a corrente no instrumento e a corrente nos contractos do relé é superior a 30
Volts, a alimentação tem de ser desligada antes de remover ou instalar os
fusíveis de entrada do instrumento. Deve-se usar um interruptor de corrente
remoto ou um interruptor no circuito. O interruptor do analisacor não remove
corrente dos fusíveis.
The analyzer contains two T, 1.6 A, 250 V, AC-line (5mm x 20 mm) fuses
(Cat. No. 52083-00. These are located on the power supply board in the back
of the instrument. They protrude through the power supply cover on the lower
right side and are labeled F1 and F2 (see Figure 51). It is not necessary to
remove the power supply cover to replace the fuses.
3. Replace the fuse (T 1.6 A, 250 V, Cat. No. 52083-00) inside the carrier.
4. Turn the fuse carrier ¼ turn clockwise to reinstall it in the fuse holder.
Page 145
51000-18 hardware installation.fm Fuse Replacement
Section 7
Note: Chemicals are present in the 1. Put the analyzer in standby (see Section 3.8 on page 82). Turn off power
pump, valve and reagent tubing. to the instrument by toggling the on/off switch at the top center of the
Before doing any maintenance, module panel to off.
review all MSDSs and use
recommended safety measures to 2. Break the siphon of any reagent/cleaning tubing that is connected to the
prevent exposure to potential
module by removing the tubing from the bottles (see Section 2.5 on
chemical hazards.
page 51). Shut off the sample stream to the analyzer.
4. Loosen the two captive screws that secure the module to the instrument
panel with a Phillips screwdriver.
Loosen these
captive screws to
remove modules.
Page 146
Changing Modules 51000-18 hardware installation.fm
Section 7
The air purge connection is located in the middle of the back of the
instrument. To connect an air supply, remove the plug in the quick connect
fitting using the Quick Connect Release tool. Connect ¼-in. poly tubing by
pushing the tubing into the fitting. You should feel two distinct “stops” as you
push. If the tubing is not seated, it will pop out when air pressure is applied.
Use only dry instrument air at 15 scfh.
CAUTION PRUDENCE
This instrument is not intended Cet instrument n’est pas conçu pour une utilisation avec des échantillons
for use with flammable samples inflammables ou des échantillons contenant des hydrocarbures.
or those containing
hydrocarbons. CAUDAD
Este instrumento no está destinado para uso con muestras inflamables o que
contengan hidrocarburos.
VORSICHT
Dieses Gerät darf nicht für Tests mit brennbaren Proben oder Proben, die
Kohlenwasserstoffe enthalten, benutzt werden.
ATENÇÃO
Este instrumento não é feito com o fim de ser empregado com amostras
inflamáveis ou aquelas que contêm hidrocarbonetos.
DANGER DANGER
This instrument should be Cet appareil doit être installé par du personnel technique qualifié, afin d’assurer
installed by qualified technical le respect de toutes les normes applicables d’électricité et de plomberie.
personnel to ensure adherence to
all applicable electrical and PELIGRO
plumbing codes. Este instrumento debe ser instalado por personal técnico capacitado para
assegurar el cumplimiento con todos los códigos eléctricos y de plomería
aplicables.
GEFAHR
Um zu gewährleisten, daß alle elektrischen und sanitärinstallationstechnischen
VDE-Vorschriften und gegebenefalls die Zusatzvorschriften der zuständigen
Elektrizitäts- und Wasserwerke erfüllt werden, darf dieses Gerät nur von
geschultem Fachpersonal installiert werden.
PERIGO
Este instrumento deve ser instalado por pessoal técnico qualificado para
asegurar o cumprimento de todas as normas elétricas e de canalização
aplicáveis.
Page 147
51000-18 hardware installation.fm Air Purge
Section 7
Note: Plumbing connections should Run sample lines with small diameter tubing to the analyzer to minimize
be made only by qualified personnel lag time (the time required for sample to flow from sample point to the
to assure conformance to all analyzer). Route the tubing as directly as possible between the sampling
applicable plumbing codes. point and the analyzer. Long sample lines and low flow rates result in
significant lag time between actual process line conditions and the sampling
point of the instrument.
Install sample line taps into larger process pipes to minimize the chances of
ingesting sediment from the pipe line bottom or air bubbles from the top. A tap
projecting into the center of the pipe is ideal (see Figure 53).
Air (Typical)
Sediment (Typical)
For samples that tend to clog the analyzer filter or tubing, additional
conditioning steps are required (see Section 6.2 on page 112).
The 22-µm Y-strainer filter protects the analyzer’s components and tubing
by allowing only particles smaller than 22 µm to pass on to the instrument
Page 148
Plumbing Installation 51000-18 hardware installation.fm
Section 7
(a 22-µm particle is at the limit of visibility by the human eye). Use the ¼- inch
ball valve to restrict and control the flow rate of filtered sample from the filter
through the sample inlet block. The customer should install another on/off
valve upstream of the filter for occasions when the sample needs to be turned
off or on. The ball valve on the filter housing can be opened to flush the
unfiltered side of the filter element, or adjusted in any partially open position
for continuous bypass.
The filtered sample then flows through the sample inlet block, from which the
instrument draws sample when needed. The analyzer is very flexible
concerning the sample flow rate through the sample bypass block. Within
reasonable bounds, the instrument is not affected by how fast the sample is
flowing through the block (it should be between 25 and 1000 mL per minute).
From the sample inlet block, the filtered sample flows to the filtered sample
bypass drain tee. It is very important to mount the bypass drain tee 6 inches to
3 feet (15 to 90 cm) higher than the sample inlet block. When properly
mounted, the vacuum breaker design of this bypass drain tee prevents
negative sample pressure and establishes the needed positive sample
pressure for sample flow through the analyzer. Connect the filtered sample
bypass drain tee to 1-inch PVC pipe, and run the pipe to the drain location.
PVC pipe is generally less expensive than hose or tubing, and is more
permanent and trouble-free than flexible hose or tubing.
1. Assemble the parts of the system in the configuration shown in Figure 54.
2. All threaded fittings need to be sealed with two wraps of the teflon
tape provided.
3. Unscrew the top of the filter housing. Insert the filter element (either end
first) and push until it stays in place. Replace the filter housing top.
4. When pushing tubing into the push-in fittings, make sure you push the
tubing all the way in. Two distinct “clicks” should be felt when pushing in
the tubing. The first is when the tubing clears the gripper ring, and the
second is when the tubing bottoms out in the fitting.
Note: This may be a little difficult. To 5. Be sure to cut the ¼-inch O.D. tubing for use in the push-in fittings with a
help grip the tubing, try using rubber sharp knife so the ends are round, cleanly cut and not cut at an angle.
gloves or other material that grips
and forcefully push the tubing into 6. The push-in fittings are designed to use with soft ¼-inch OD plastic tubing
the fitting. Two stops should be felt; as included in the kit. Tubing material such as PTFE or HDPE is
otherwise the tubing will not be fully
recommended. The push-in fitting will not grip hard plastic or metal tubing
inserted and will leak.
and the tubing will slip out.
Page 149
51000-18 hardware installation.fm Plumbing Installation
Section 7
1. Set filtered sample bypass flow by adjusting the ¼-inch ball valve, item 21
in Figure 54. The valve is fully closed when the lever is perpendicular to
the valve body and fully open when the lever is parallel with the valve
body. Hach recommends a filter bypass flow of 200–500 mL/min., but
25-1000 mL/min. is acceptable (see discussion in Determining Sample
Lag Time, Section 7.7.2 on page 152).
2. Check the flow by pulling the ¼-OD tube from the push-in fitting at the
drain tee (the tubing from the sample inlet block to the drain tee). If
desired, measure the flow rate by directing the sample into a large
measuring container (flask or cylinder) over a set amount of time. Keep
the end of the tube at the same height as the push-in fitting.
Page 150
Plumbing Installation 51000-18 hardware installation.fm
Section 7
Figure 54 Sample Line and Basic Water Conditioning Assembly
Page 151
51000-18 hardware installation.fm Plumbing Installation
Section 7
The lag time for sample to flow through the 0.25 O.D. X 0.17 ID tubing
supplied with the instrument must also be taken into account. The internal
volume of a 10 foot length of the tubing is 44.6 mL. Therefore, if the flow rate
is 100 mL/min, it will take (44.6 mL)/(100 mL/min) = 0.45 minutes for the
sample to flow through 10 feet of tubing.
The total lag time for the sample to reach the instrument would be the lag time
due to the filter plus the lag time through the tubing. Add the results for the
filter and 10 foot length of tubing as calculated above for the flow rates
indicated in Table 24.
Table 24
Lag time for 10 ft of Total Filtered Sample
Flow rate (mL/min) Filter lag time (min) Total lag time (min)
Tubing (min) Used L/day (gal/day)
100 6.8 0.45 7.2 144 (38)
200 3.4 0.22 3.6 288 (76)
500 1.4 0.10 1.5 720 (190)
Page 152
Plumbing Installation 51000-18 hardware installation.fm
Section 7
Drain Block
Note: It is imperative to mount this This assembly can be installed with the sample bypass and run into the same
drain point below the instrument drain or plumbed to a different drain than the sample bypass. Refer to
drain block. Failure to do so may Figure 55.
result in improper functioning of the
instrument or backflow of waste into The cabinet overflow drain is located in the middle of the tubing manifold and
the instrument cabinet.
is sized to accommodate a 1-in. ID hose. This drain is used if an upset inside
the instrument results in a significant leak. The supplied hose from this drain
port must be connected and run to a drain which is at or below the level of the
drain fitting.
The reagent enclosures each have an overflow fitting in the bottom of the
case. You can plumb these to the drain by punching out the knock-out in the
fitting, attaching a 3/8-inch ID hose to each fitting and running it to a drain
which is lower in elevation than the bottom of the reagent enclosure.
Page 153
51000-18 hardware installation.fm Plumbing Installation
Section 7
Figure 56 PVC Drain Assembly for the Instrument/Grab Sample Block Drain
Page 154
Plumbing Installation 51000-18 hardware installation.fm
Section 8 Network Menu Configuration
1. Ensure step 1 and step 2 of Section 8.1 have been completed. After
completion of these two steps it is assumed that electrical connections to
the SOM relays and analog outputs have been made.
2. Set up and test the SOM Relays and Analog Outputs per Section 2 of the
SOM Manual.
8.1.3 APA 6000 Menu Options for SOM Relays & Analog Outputs
Now you must assign the relay and analog outputs of the SOM to specific
alarm and recorder outputs of the APA 6000. Perform the steps in Alarm to
Relay Connections and Analog Output Settings in the SOM Manual. Then go
to Section 2.3 on page 40 to select the desired options.
Page 155
51000-18 network menu configuration.fm Network Menu Configuration
Section 8
3. Verify the computer can transmit to SIO. If a computer will be used for
remote operation of the APA 6000, verify commands can be sent to the
analyzer. This can be done by transmitting xRID, where x = 1 to 8,
corresponding to channels 1 to 8. Follow this command with a carriage
return from the computer.
To find out what channel number data is being sent, turn the analyzer off and
on. On power up, the APA 6000 will send a message similar to the one below.
This is the power up message. The number between brackets is the SIO
channel number for the APA analyzer.
If there is no return response, troubleshoot the SIO per the SIO Manual.
Page 156
Using an APA With an SIO Module 51000-18 network menu configuration.fm
Section 8
8.2.3 RS232 Data Format: APA 6000 Menu Options for SIO
The APA 6000 sends out measurement readings, alarms, and warnings
through the SIO. This section instructs how to select a particular format.
For a more detailed explanation of the format types, refer to Section 8.2.4.
Note the selections under SIO1 SETUP and SIO2 SETUP are the same. Up to
two SIOs can be connected to the APA 6000. If installing a second SIO or the
analyzer is connected to the second SIO on the network, select SIO2 SETUP.
Page 157
51000-18 network menu configuration.fm Using an APA With an SIO Module
Section 8
First occurrence of any alarm or warning will be sent out. The time/date of the
alarm or warning is printed followed by either ALRM or WARN to specify it is
an alarm or warning. The specific alarm or warning is then printed on the next
line. When the alarm or warning condition clears it will be sent preceded by
END: to indicate the condition is cleared. Some alarms and warnings provide
information only and do not have an end (i.e. a fail to calibrate message is
printed out only when the condition occurs). If a alarm or warning condition is
active, then a *A for alarm or a *W for warning is appended to the
measurement readings.
When sending commands through the SIO, the channel number must
precede the command. For example, if the analyzer is connected to SIO
channel 1 and you wish to restart the analyzer, 1SRT should be sent. The
analyzer will reply with OK and the SIO inserts the channel assigned for the
analyzer (i.e. [1]OK will be the response to the computer).
Page 158
Using an APA With an SIO Module 51000-18 network menu configuration.fm
Section 8
Table 25 APA 6000 Remote RS232 Commands
Description of Remote Command Command Description of Command/Settings
Restart system SRT Restarts the system
Put instrument in standby HLT Puts the analyzer in standby
Perform a calibration CAL Performs a calibration at end of the current cycle
Perform a clean cycle CLN Performs a clean cycle at end of the current cycle
Prime entire system PMA Primes all the reagent and sample lines now
Prime reagents PMR Primes reagents lines at end of current cycle
Prime standards PMS Primes standard lines at end of current cycle
Prime electrolyte PME Refreshes electrolyte in the system at end of current cycle
Set/query Auto-Calibration setting
CTM Sending just first three characters CTM will result in current auto-cal.
settings to be returned. Add the equal sign and one of the characters
listed in the Command column to the left to set new auto-calibration
interval.
CTM=4,x Calibrate every 12 hours.
x = 0 to 23 indicates time of day to calibrate (using a 24-hour clock).
If x=22, the analyzer calibrates at 10:00 p.m. and a.m.
CTM=5,x Calibrate daily.
x = 0 to 23 indicates time of day to calibrate (using a 24-hour clock).
CTM=x,y Auto-Calibrate Weekly.
x = 6 to 12 which signifies the day of week to calibrate where Monday=6
and Sunday=12. y = 0 to 23 to indicate time of day to calibrate
(using 24-hour clock).
Recall Calibration RCL=A Sending just first three characters recalls the calibration values from the
last calibration. Adding “=A” will recall all calibrations stored (up to 7).
Reset to Default calibration RDC Resets calibration curve to factory default values.
Recall Grab Sample RGB=A Sending just first three characters recalls the grab sample value(s) from
the last grab sample cycle.
Adding “=A” will recall all grab samples stored (up to 8).
Get/Set Auto-Refresh settings ELI=x Sending just first three characters results in current settings being
returned. Adding =x will result in following settings where x =1=6 hours,
x=2=12 hours and x=3=24 hours.
Set Type of Auto-Clean cycle ACT=x Sending just first three characters results in current settings being
returned. Adding =x will result in following settings:
x=0=Normal Clean, x=1= Extended Clean
Get/Set Auto-Clean settings CLI=x Sending just first three characters results in current settings being
returned. Adding =x will result in following settings:
x =1=6 hours, x=2 =12 hours and x=3=24 hours.
x=4 to 10 signifies clean weekly where the day of the week to clean is 4
for Monday to 10 for Sunday.
Get/Set sample sequence SEQ Returns “SEQUENCE=123...1” which represents the order in which the
samples are measured.
SEQ=123.. Can have up to 8 values in sequence. The numbers after the equal sign
represent the order in which sample measurements are performed.
Measure Grab sample GRB Perform a grab sample measurement
Recall Grab Sample RGB=A Sending just first three characters recalls the grab sample value(s) from
the last grab sample cycle.
Adding “=A” will recall all grab samples stored (up to 8).
Page 159
51000-18 network menu configuration.fm Using an APA With an SIO Module
Section 8
Table 25 APA 6000 Remote RS232 Commands (continued)
Description of Remote Command Command Description of Command/Settings
Get Instrument ID RID Nitrate V1.0 C1.0
Ammonium V1.0 C1.0
Amm and Nitr V1.0 C1.0
Hardness V1.0 C1.0
Alkalinity V1.0 C1.0
Output latest alarms/warnings RAR Output up to last 10 alarms/warning messages since last “RAR”
command. Only enabled if SIOx setup in Query mode.
Output latest readings. RMR Output last measurement reading for all cycles (grab and normal
measure-ment readings & calibrations). Only enabled if SIOx setup
in Query mode.
Reset all fluid levels to 100% RFL Resets all standards, reagents and cleaning solutions to 100%.
Query/Set reagent levels. RG1=xxx Send just first three characters if requesting level.
The third character is the number of the reagent.
Add “=” followed by reagent level in percent if setting to new level.
Allowed values are between 0 and 100%.
Query/Set standard levels. SD1=xxx Send just first three characters if requesting level.
The third character is the number of the standard.
Add “=” followed by standard level in percent if setting to new level.
Allowed values are between 0 and 100%. The percent sign is optional.
Query/Set pump wash fluid level. WSH=xxx Send just first three characters if requesting level.
Add “=” followed by fluid level in percent if setting to new level.
Allowed values are between 0 and 100%. The percent sign is optional.
Request all Instrument settings and ALL Sends out all instrument settings.
information.
Set/Query Instrument name. RIN=xxx Send first three characters to get instrument name. Default name is
“#1 APA 6000”. Add equal sign plus text to set new name. Maximum
length of name is 12 characters. Note: all letters will be converted to
upper case.
Test internal recorder channel A. RCA=x Set x = F or H or Z for output recorder full-scale, half-scale or zero
respectively. Send just first three characters to stop test.
Test internal recorder channel B. RCB=x Set x = F or H or Z for output recorder full-scale, half-scale or zero
respectively. Send just first three characters to stop test.
Test internal relay A. RLA=x Set x = 0 or 1 for relay off or on. Send just first three characters to stop
test.
Test internal relay B. RLB=x Set x = 0 or 1 for relay off or on. Send just first three characters to stop
test.
Get Software Version numbers. RVN Get version numbers of instrument and its modules. For Alkalinity V1.0
C1.0
EXT NEURON: EXT0101
INT NEURON: INT0101
(and following if used:)
VALVE MODULE: VLV0103
DETECT MODULE: DET0102
BURETTE1 MODULE: PMP0102
BURETTE 2 MODULE: PMP0102
MIXER MODULE: PMP0102
Query alarm/warning logs TAL Transmits all messages in current alarm log
Clear alarm/warning logs WAL Clears alarm logs.
Page 160
Using an APA With an SIO Module 51000-18 network menu configuration.fm
Section 8
Page 161
51000-18 network menu configuration.fm Using an APA With a Digital Display Module (DDM)
Visit http: //www.hach.com
Section 9 PID Control
The APA 6000 Analyzer uses a PID (Proportional, Integral (reset), Derivative
(rate) control algorithm. Each of the analyzer settings in Section 2.3.3.2 on
page 46 are described below.
Page 163
51000-18 pid control.fm PID Control
Visit http: //www.hach.com
Appendix A Bench Method Procedure
The precision of the APA 6000 analyzer is specified at 3%, however the
precision of the bench method is estimated to be 10%. This may lead to
apparent discrepancies between measurements obtained on the instrument
and with the bench procedure.
Contact with carbon dioxide in the air reduces the sample P alkalinity with
time. While the measurements given by the APA analyzer are representative
of the sample P alkalinity where it is sampled, the results from the bench
method may only be representative of the sample P alkalinity after some of it,
over time, has reacted with carbon dioxide in the air.
Also note that alkalinity measurements on the instrument are made at 50 °C,
while the bench method is performed at room temperature. Depending on the
sample matrix, this may contribute to apparent differences between
instrument and bench method results.
1. Select the sample 2. Insert a clean 3. Turn the delivery knob 4. Use a volumetric pipet
volume and Sulfuric Acid delivery tube into the to eject a few drops of to measure the sample
(H2SO4) Titration Cartridge titration cartridge. titrant. Reset the counter volume from Table 1.
corresponding to the Attach the cartridge to to zero and wipe the tip. Transfer the sample into
expected alkalinity the titrator body. a clean 250-mL
Note: For added convenience
concentration as mg/L Erlenmeyer flask. Dilute
Note: See General use the TitraStir Stir Plate.
calcium carbonate (CaCO3) See General Description to about the 100-mL
Description, Step-by-Step
from Table 1. in the Digital Titrator Digital Titrator Manual, Step 3 mark with deionized
Manual for assembly in Step-by-Step. water, if necessary.
Note: See Sampling and
Storage following instructions, if necessary.
these steps.
Table 1
Range Sample Titration Cartridge Catalog Digit
(mg/L as CaCO3) Volume (mL) (H2SO4) Number Multiplier
10-50 100 0.1600 N 14388-01 0.1
50-100 50 0.1600 N 14388-01 0.2
100-500 100 1.600 N 14389-01 1.0
500-1000 50 1.600 N 14389-01 2.0
Page 165
51000-18 bench method procedure.fm Bench Method Procedure
Appendix A
5. Add the contents of 6. If the solution turns 7. Calculate: 8. Add the contents of
one Phenolphthalein pink, titrate to a colorless one Bromcresol Green-
Digits Required x Digit
Indicator Powder Pillow end point. Place the Methyl Red Indicator
Multiplier = mg/L CaCO3 P
and swirl to mix. delivery tube tip into the Alkalinity Powder Pillow to the flask
solution and swirl the flask and swirl to mix.
Note: A solution of one pH
while titrating with sulfuric
8.3 Buffer Powder Pillow and Note: Four drops of Methyl
one Phenolphthalein Powder acid. Record the number
Purple Indicator Solution may
Pillow in 50 mL of deionized of digits required. be substituted for the
water is recommended as a Note: If the solution is Bromcresol Green-Methyl
comparison for determining colorless before titrating with Red Indicator Powder Pillow.
the proper end-point color. sulfuric acid, the Titrate from green to a gray
Phenolphthalein (P) Alkalinity end point (pH 5.1).
Note: Four drops of
Phenolphthalein Indicator is zero; proceed with Step 8.
Note: Four drops of
Solution may be substituted Bromcresol Green-Methyl
for the Phenolphthalein Red Indicator Solution may
Indicator Powder Pillow. be substituted for the
Bromcresol Green-Methyl
Red Indicator Powder Pillow.
Page 166
Bench Method Procedure 51000-18 bench method procedure.fm
Appendix A
Page 167
51000-18 bench method procedure.fm Bench Method Procedure
Appendix A
6. Check your results. The sum of the three alkalinity types will equal the
total alkalinity.
Page 168
Bench Method Procedure 51000-18 bench method procedure.fm
Appendix A
Required Reagents
(varies with sample characteristics)
Alkalinity Reagent Set (about 100 tests) ...................................................................................................... 22719-00
Includes: (1) 942-99, (1) 943-99, (1) 14388-01, (1) 14389-01
Required Apparatus
Digital Titrator.............................................................................................................................each.......... 16900-01
Flask, Erlenmeyer, 250 mL ........................................................................................................each.............. 505-46
Page 169
51000-18 bench method procedure.fm Bench Method Procedure
Visit http: //www.hach.com
Glossary
Master AquaTrend - Acts as the main network data control center for setup
and operation of the network. The network can contain as many as nine
AquaTrends but only one can be the Master AquaTrend. Only through the
Master AquaTrend can you perform network operations.
network - The connection of one or more devices (of which at least one is a
sensor or analyzer) to a Master AquaTrend.
pop-up box - A selection or message box at the end of a menu chain. This
box either provides a choice of options—selectable by moving the cursor and
pressing ENTER—or allows you to use the arrow keys to enter an
Page 171
51000-18 glossary.fm Glossary
Glossary
SCADA - Supervisory Control and Data Acquisition. General term for any
automated (and usually PLC or computer-controlled) system for control and
data acquisition.
sensor - Any one of the Hach devices designed to take measurements for
specific parameters and for use with the AquaTrend Network System (e.g., the
1720D Turbidimeter)
Signal Output Module (SOM) - Network device that provides two relays,
each with normally open and normally closed contacts rated at a maximum of
5 A/250 VAC, and two analog outputs, selectable through the AquaTrend
keypad at either 4-20 or 0-20 mA. Sensor alarms can be configured to trigger
one the relays in the SOM and/or a recorder, or controller signal from a
sensor can be attached to the analog outputs of the SOM.
The SOM can either be a stand-alone device (model 51250) or can be internal
to the AquaTrend/SOM (model 51350).
Page 172
Glossary 51000-18 glossary.fm
Glossary
1720D 1720D
PS1201 PS1201
1650 feet (500 meters) Total maximum distance for network communication.
SOM PS1201
Other
Device
SOM
1650 feet (500 meters) Total maximum distance for network communication.
PS1201 DDM
Other Device
Master Other
AquaTrend Device
1720D
1650 feet (500 meters) Total maximum distance for network communication.
Distances in excess of 1650 feet (500 meters) require a repeater.
100 feet (30.3 meters) Maximum distance between the power supply and the device it is powering.
1320 feet (400 meters) Maximum distance between two network devices.
Page 173
51000-18 glossary.fm Glossary
Visit http: //www.hach.com
General Information
Page 175
51000-18 general information stopper.fm General Information
Replacement Parts
Instrument Configurations
Description unit Cat. No.
APA 6000™ Alkalinity Analyzer, with AquaTrend® Interface ......................................................each.......... 51000-10
APA 6000™ Alkalinity Analyzer, without AquaTrend® Interface .................................................each.......... 51000-12
Sample Conditioning
Air Backflush Kit, automatic, 115 Vac ........................................................................................each.......... 51068-00
Air Backflush Kit, automatic, 230 Vac ........................................................................................each.......... 51068-01
Basic Sample Conditioning Kit w/ 22-µm filter ...........................................................................each.......... 51045-00
Filter, 0.2-µm membrane ............................................................................................................each.......... 51042-00
APA 6000 Micro Filter System, model 9700, 115 V ...................................................................each.......... 51339-00
APA 6000 Micro Filter System, model 9700, 230 V ...................................................................each.......... 51339-01
Filter Element, 450 mesh ...........................................................................................................each.......... 51397-00
Filter, support element ...............................................................................................................each.......... 51034-00
O-ring, Viton, filter seal ..............................................................................................................each.......... 51038-00
Page 176
Replacement Parts 51000-18 replacement parts.fm
Replacement Parts
Replacement Parts
Description unit Cat. No.
AquaTrend Instrument Manual ...................................................................................................each.......... 51350-18
Binder, D-ring, for AquaTrend manual .................................................................................each.......... 52001-77
Autoburette Module, with 2-way valve only ................................................................................each.......... 51020-02
Autoburette Block .......................................................................................................................each.......... 51108-00
Bulkhead Union, 0.25 OD tubing to tubing (for air purge bulkhead) ..........................................each.......... 51104-00
Cable, backplane .......................................................................................................................each.......... 51161-00
Cable, AquaTrend ......................................................................................................................each.......... 51164-00
Cable Termination Kit .................................................................................................................each.......... 52156-00
Colorimeter, APA 6000 Alkalinity................................................................................................each.......... 51026-01
Colorimeter Temperature Control Block .....................................................................................each.......... 51116-00
Faceplate Assembly, APA 6000..................................................................................................each.......... 51022-00
Fitting, Ferrule, 0.062 OD...........................................................................................................each.......... 51130-00
Fitting, Ferrule, with ring, 0.062 OD ...........................................................................................each.......... 51132-00
Fitting, Union, Tefzel, ¼-28 ........................................................................................................each.......... 51359-00
Flangeless Hex Nut Fitting, 0.062 OD........................................................................................each.......... 51129-00
Flangeless Nut Fitting, 0.062 OD ...............................................................................................each.......... 51131-00
Fuse, 1.6 A, 250 V, 5 mm x 20 mm, UL, CSA, VDE approved...................................................each.......... 52083-00
Grab Sample Port Cap...............................................................................................................each.......... 51317-00
Grab Sample/Drain Block Assembly ..........................................................................................each.......... 51503-00
Grommet, manifold, APA 6000 (for tubing).................................................................................each.......... 51036-00
Hex Nut Valve Plug, 10-32 .........................................................................................................each.......... 51393-00
Manual, Alkalinity, APA 6000......................................................................................................each.......... 51000-18
Mixer Module Assembly, APA 6000, 3-mL .................................................................................each.......... 51021-01
Plug, air purge, for ¼-in. tubing ..................................................................................................each.......... 51140-00
Plug, manifold, APA 6000...........................................................................................................each.......... 51224-00
Plug, flat bottom, Tefzal, ¼-28....................................................................................................each.......... 51293-00
Power Switch, 4A, 250 VAC .......................................................................................................each.......... 51155-00
Quick Reference Card................................................................................................................each.......... 51000-44
Reagent Bottle Vent Cap Assembly ...........................................................................................each.......... 51215-00
Reagent Housing Assembly.......................................................................................................each.......... 51048-00
Restraining Cable, steel, w/eyelets (for module panel) ..............................................................each.......... 51417-00
Rotary Valve, 10-port, with actuator ...........................................................................................each.......... 51150-00
Sample Inlet Block .....................................................................................................................each.......... 51365-00
Seal, oil tight, 0.88-in. diameter .................................................................................................each.......... 42210-00
Scrubber, CO2, for standard bottles ...........................................................................................each.......... 51431-00
Scrubber, CO2, soda lime absorbant ........................................................................................ 500g.......... 27476-00
Stator, for 10-port valve (replaces 51150-00).............................................................................each.......... 51347-00
Stir Bar, micro, 3 mm x 10 mm...................................................................................................each.......... 20549-59
Teflon Tubing, 0.170 ID x 0.250 OD .............................................................................................feet.......... 45462-00
Teflon Thread Tape, 0.25-in wide ...............................................................................................1 roll.......... 70608-24
Temperature Control Block Assembly ........................................................................................each.......... 51046-01
Tubing, Tefzel, 0.030 ID x 0.062 OD, 2000 psi .............................................................................feet.......... 51178-00
Valve Body Replacement, 10-port valve ....................................................................................each.......... 51507-00
Valve Module Assembly, APA 6000............................................................................................each.......... 51029-01
Washer, sealing, ½-in. NPT .......................................................................................................each.......... 10338-14
Page 177
51000-18 replacement parts.fm Replacement Parts
Replacement Parts
Kits
Description unit Cat. No.
Installation Kit ..........................................................................................................................each.......... 51040-00
Fitting, elbow, ½ ID tubing x ¾ NPT PP (2) .........................................................................each.......... 30618-00
Fitting, reducer, ¾ MNPT x ¼ FNPT PVC (3)......................................................................each.......... 31560-00
Tubing, Teflon, 0.170 x 0.250 OD .............................................................................. 3 m (10 ft.).......... 45462-00
Tubing, Tygon, ¾-in OD x 1/8-in. wall, black ............................................................. 1.8 m (6 ft.).......... 45543-00
Clamp, for hanging conduit, 2.5 cm (1 in.) (2) .....................................................................each.......... 47349-00
Coupling, ¾-in. NPT x ¾-in. NPT (3)...................................................................................each.......... 49620-00
Tee Fitting, 1-in. x ¾-in. x 1-in. PVC (2)...............................................................................each.......... 51080-00
Drain Pipe, pre-cut, 1-in. PVC (2) ........................................................................... 15 cm (6 in.).......... 51239-00
Fitting, ¼-in. tube connector, push-fit (3) .............................................................................each.......... 51246-00
Ball Valve, ¼-in. NPT, PVC (1) ............................................................................................each.......... 51395-00
Filter Housing, ¾-in., w/o filter .............................................................................................each.......... 51396-00
Filter Element, 22 µm, 450 mesh (1) ...................................................................................each.......... 51397-00
Teflon Thread Tape, 0.25-in wide (1 roll) .............................................................................each.......... 70608-24
Page 178
Replacement Parts 51000-18 replacement parts.fm
How to Order
U.S.A. Customers
By Telephone: By Mail:
6:30 a.m. to 5:00 p.m. MST Hach Company
Monday through Friday P.O. Box 389
(800) 227-HACH (800-227-4224) Loveland, Colorado 80539-0389 U.S.A
Information Required
• Hach account number (if available) • Billing address
• Your name and phone number • Shipping address
• Purchase order number • Catalog number
• Brief description or model number • Quantity
International Customers
Hach maintains a worldwide network of dealers and distributors.
To locate the representative nearest you, send E-mail to
intl@hach.com or contact:
Page 179
51000-18 general information section.fm How to Order
Repair Service
In Canada:
Hach Sales & Service Canada Ltd.
1313 Border Street, Unit 34
Winnipeg, Manitoba
R3H 0X4
(800) 665-7635 (Canada only)
Telephone: (204) 632-5598
FAX: (204) 694-5134
E-mail: canada@hach.com
Page 180
Repair Service 51000-18 general information section.fm
Warranty
Page 181
51000-18 general information section.fm Warranty
Index
Page 182
Index 51000-18IX.fm
Index
Bench mount ................................................... 124 connections .............................................136, 142
Drain lines ....................................................... 152 printer connections .......................................... 156
electrical connections ...................................... 128
network connections ...............................136, 142 O
panel mount .................................................... 124
Operating environment ....................................16, 123
plumbing .......................................................... 147
power connections .......................................... 130 Operation
recorders ......................................................... 134 normal ............................................................... 69
sample pump ................................................... 119 starting .............................................................. 55
sample streams .........................................62, 122 theory of ............................................................ 25
wall mount ....................................................... 124 Optional equipment ................. 20, 26, 155, 161, 172
Output ......................................... 40, 43, 44, 46, 138
K 4-20 mA .......................................................17, 27
adjusting ............................................................ 48
Keypad ..............................................................27–28
hold ................................................................. 104
Keys ......................................................................... 28 internal output test ........................................... 106
Kit RS 232 ....................................................157, 158
installation ...................................................17, 24 sensor ............................................................... 65
maintenance .........................................17, 24, 85
reagent .............................................................. 17
P
sample conditioniong ........................................ 48
tool ..............................................................17, 24 Password ................................................................. 36
PID settings .....................................................46, 163
L Port .......................................................................... 69
Lag time ................................................................. 152 Power connectionsl ................................................ 130
Language selection ................................................. 34 Priming ..................................................................... 55
Printer connections ........................................156, 157
M Printer format ......................................................... 157
Maintenance ............................................................ 85 Proportional band .....................................43, 46, 163
autoburette ........................................................ 85
changing modules ........................................... 146 R
changing solutions ......................................51, 53 Reagents ....................................................23, 51, 85
colorimeter ........................................................ 85 changing ............................................................ 86
kit ....................................................................... 24 consumption ...................................................... 86
mixer module ..................................................... 95 priming .............................................................. 55
scheduled .......................................................... 85 Recorders .......................................... 40, 43, 44, 134
tubing ................................................................ 92 adjusting output (trim) ....................................... 48
valve rotor and stator ........................................ 90 internal output test ........................................... 106
Measurement ........................................................... 69 Relays .....................................................40, 105, 133
attaching to a channel ....................................... 65
Replacement parts ................................................. 176
channel .............................................................. 67
Restarting the instrument ......................................... 82
Module
alarm messages ................................................ 99 Rotor ...........................................................69, 90, 97
autoburette ...........................................18, 85, 87 alarm messages .............................................. 101
changing modules ........................................... 146 RS 232 commands ................................................ 158
colorimeter ..................................................18, 85 RS 232 output ........................................................ 157
mixer ....................................................18, 85, 95
rotary valve ........................................................ 18 S
rotor valve ............................................85, 90, 97
Sample conditioning .................................48–50, 112
Mounting the analyzer ........................................... 124 air backflush .....................................49, 113, 133
sample pump ....................................50, 119, 133
N setup .........................................................48, 149
Network ................................................................... 20 ultrafiltration ..................................................... 118
computer connections ..................................... 156 Sample flow ............................................................. 55
configuration .................................................... 155 Sample lag time ..................................................... 152
Page 183
51000-18IX.fm Index
Index
Sample lines .......................................................... 148 Specifications ........................................................... 11
Sample pump .........................................50, 119, 133 Standards ..........................................................23, 51
Sample sequencing ................................................. 63 priming .............................................................. 56
Sample streams Standby ..............................................................29, 82
adding .......................................................62, 122 Stator ..........................................................85, 90, 97
sequencing ........................................................ 63
Security .................................................................... 36 T
level ................................................................... 37 Temperature control ................................................ 81
password ........................................................... 36 alarm messages .............................................. 101
unlocking the keypad ........................................ 38
Theory of operation .................................................. 25
Setpoint ....................................................43, 47, 163
Time setting ............................................................. 32
Setup
Tools ...........................................................24, 88–89
alarms ............................................................... 40
analyzer name ................................................... 35 Trim .......................................................................... 48
auto-calibration .................................................. 57 Troubleshooting ...............................................97–102
auto-clean ......................................................... 61 Tubing ................................................................97, 98
date ................................................................... 33 replacement ...................................................... 92
date format ........................................................ 34 replacement lengths .......................................... 94
day of the week ................................................. 33
language ........................................................... 34 U
measurement units ............................................ 39
Ultrafiltration ........................................................... 118
overview ............................................................ 15
sample conditioning .......................................... 48 Units of measure ................................................31, 39
sample measurement channels ........................ 65 Unlocking the keypad .............................................. 38
sample streams ................................................. 62
security ........................................................36–38 W
time ................................................................... 32 Warnings
unlocking the keypad ........................................ 38 alarm log ........................................................... 79
Shutdown ................................................................. 82 clearing the alarm log ........................................ 79
SIO .............................. 20, 137, 139, 156, 157, 172 sensor .........................................................41, 42
Software version .................................................... 108 troubleshooting .................................................. 97
Solution levels ......................................................... 53 warning messages ............................................ 99
SOM ..................................... 20, 138, 139, 155, 172 Waste disposal ...................................................... 154
Wiring topologies ................................................... 173
Page 184
Index 51000-18IX.fm