You are on page 1of 26

BRITISH STANDARD BS EN

196-21:1992

Methods of testing
cement —
Part 21: Determination of the chloride,
carbon dioxide and alkali content of
cement
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

The European Standard EN 196-21:1989 has the status of a


British Standard
BS EN 196-21:1992

Cooperating organizations

The European Committee for Standardization (CEN), under whose supervision


this European Standard was prepared, comprises the national standards
organizations of the following countries.

Austria Oesterreichisches Normungsinstitut


Belgium Institut belge de normalisation
Denmark Dansk Standardiseringsraad
Finland Suomen Standardisoimisliito, r.y.
France Association française de normalisation
Germany Deutsches Institut für Normung e.V.
Greece Hellenic Organization for Standardization
Iceland Technological Institute of Iceland
Ireland National Standards Authority of Ireland
Italy Ente Nazionale Italiano di Unificazione
Luxembourg Inspection du Travail et des Mines
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

Netherlands Nederlands Normalisatie-instituut


Norway Norges Standardiseringsforbund
Portugal Instituto Portuguès da Qualidade
Spain Asociación Española de Normalización y Certificación
Sweden Standardiseringskommissionen i Sverige
Switzerland Association suisse de normalisation
United Kingdom British Standards Institution

This British Standard, having


been prepared under the
direction of the Technical Sector
Board for Building and Civil
Engineering, was published
under the authority of the
Standards Board and comes Amendments issued since publication
into effect on
15 June 1992
Amd. No. Date Comments
© BSI 01-1999

The following BSI references


relate to the work on this
standard:
Committee reference B/516
Draft for comment 87/10353 DC

ISBN 0 580 20680 7


BS EN 196-21:1992

Contents

Page
Cooperating organizations Inside front cover
National foreword ii
Brief history 2
Foreword 2
Text of EN 196-21 5
National annex NA (informative) Test method for the
alkali content of cement to be used for the purpose of
calculating the reactive alkali content of concrete 16
National annex NB (informative) Committees responsible Inside back cover
National annex NC (informative) Cross-references Inside back cover
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

© BSI 01-1999 i
BS EN 196-21:1992

National foreword

This British Standard has been prepared under the direction of the Technical
Sector Board for Building and Civil Engineering. It is the English language
version of EN 196-21:1989 “Methods of testing cement; Determination of the
chloride, carbon dioxide and alkali content of cement”, published by the European
Committee for Standardization (CEN). EN 196-21 was drawn up by Technical
Committee 51, Cement and building limes, as Part 21 of a series on testing
cement which was accepted by CEN on 3 October 1989. The UK gave a negative
vote at the final voting stage, but under the CEN Rules now in force, is bound to
abide by the majority decision and to implement this Part. Thus, the European
test procedures have been introduced, where relevant, into the 1991 revisions of
British Standards for cement BS 12, BS 146, BS 4027, BS 4246, BS 6588 and
BS 6610. This British Standard supersedes clauses 16 and 17 of BS 4550-2 which
are withdrawn.
The UK negative vote on this Part was due to the fact that the European
methods for alkalis give values for Na2O equivalent of cement higher by
approximately 0.025 % absolute than those obtained using either the current BS
or the ASTM methods which have been used in the UK for the purpose of
calculating the “reactive” alkali content of concrete.
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

In consequence, the BS 4550-2:1970 test method is retained as national annex


NA to be used only for the determination of the alkali content of cement for the
purpose of calculations to check compliance with the maximum value of 3.0 kg/m3
of Na2O equivalent for the reactive alkali content of a concrete mix, recommended
when aggregate containing an alkali reactive constituent, or when an unfamiliar
cement/aggregate combination, is in use (see 4.2.4 of BS 5328-1:19911)).
However, the European test method is called up in 8.3 of BS 4027:19912) which
specifies the maximum permitted alkali content of sulfate-resisting Portland
cement, type Low Alkali.
National annex NB gives the committees responsible for the UK participation in
the preparation of this standard.
National annex NC gives details of corresponding British Standards for
international standards referred to in the European Standard.
A British Standard does not purport to include all the necessary provisions of a
contract. Users of British Standards are responsible for their correct application.
Compliance with a British Standard does not of itself confer immunity
from legal obligations.

Summary of pages
This document comprises a front cover, an inside front cover, pages i and ii,
the EN title page, pages 2 to 20, an inside back cover and a back cover.
This standard has been updated (see copyright date) and may have had
amendments incorporated. This will be indicated in the amendment table on
the inside front cover.

1) BS 5328 Concrete, Part 1:1991 Guide to specifying concrete


2) BS 4027:1991 Specification for sulfate-resisting Portland cement

ii © BSI 01-1999
EUROPEAN STANDARD EN 196-21
NORME EUROPÉENNE
EUROPÄISCHE NORM December 1989

UDC 666.94:691.54:620.1:543.84

Descriptors: Cements, tests, chemical analysis, determination of content, chlorides, carbon dioxide, alkali

English version

Methods of testing cement; Determination of the chloride,


carbon dioxide and alkali content of cement

Méthodes d’essais des ciments; Détermination Prüfverfahren für Zement; Bestimmung des
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

de la teneur en chlorures, en dioxide de carbone Chlorid-, Kohlenstoffdioxid- und Alkalianteils


et en alcalis dans les ciments von Zement

This European Standard was approved by CEN on 1989-06-16. CEN members


are bound to comply with the CEN/CENELEC Internal Regulations which
stipulate the conditions for giving this European Standard the status of a
national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national
standards may be obtained on application to the Central Secretariat or to any
CEN member.
This European Standard exists in three official versions (English, French,
German). A version in any other language made by translation under the
responsibility of a CEN member into its own language and notified to the
Central Secretariat has the same status as the official versions.
CEN members are the national standards bodies of Austria, Belgium,
Denmark, Finland, France, Germany, Greece, Iceland, Ireland, Italy,
Luxembourg, Netherlands, Norway, Portugal, Spain, Sweden, Switzerland and
United Kingdom.

CEN
European Committee for Standardization
Comité Européen de Normalisation
Europäisches Komitee für Normung

Central Secretariat: rue de Stassart 36, B-1050 Brussels

© CEN 1989 Copyright reserved to CEN members


Ref. No. EN 196-21:1989 E
EN 196-21:1989

Brief history Foreword


This European Standard was prepared by Technical The standard EN 196 on methods of testing cement
Committee CEN/TC 51 “Cement”, the Secretariat of consists of the following parts:
which is held by IBN. — Part 1: Determination of strength;
It is intended, in a revision of European Standard — Part 2: Chemical analysis of cement;
EN 196-2, to adopt the methods of test in this
— Part 3: Determination of setting time and
standard into EN 196-2, so that this standard will
soundness;
then contain all methods of test required for the
chemical analysis of cement. — Part 4: Quantitative determination of
constituents;
In accordance with the Common CEN/CENELEC
Rules, the following countries are bound to — Part 5: Pozzolanicity test for pozzolanic
implement this European Standard: cements;
Austria, Belgium, Denmark, Finland, France, — Part 6: Determination of fineness;
Germany, Greece, Iceland, Ireland, Italy, — Part 7: Methods of taking and preparing
Luxembourg, Netherlands, Norway, Portugal, samples of cement;
Spain, Sweden, Switzerland and United Kingdom. — Part 21: Determination of the chloride, carbon
dioxide and alkali content of cement.
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

2 © BSI 01-1999
EN 196-21:1989

Contents
Page Page
Brief history 2 7.7 Expression of results 13
Foreword 2 7.8 Repeatability and reproducibility 13
1 Object and field of application 5 8 Determination of the alkali content
2 References 5 (alternative method) 13
3 General requirements for testing 5 8.1 Principle 13
3.1 Number of tests 5 8.2 Reagents 13
3.2 Repeatability and reproducibility 5 8.3 Apparatus 14
3.3 Expression of masses, volumes 8.4 Plotting the calibration curve 14
and results 5 8.5 Procedure 14
3.4 Determination of constant mass 5 8.6 Expression of results 14
3.5 Preparation of the cement sample 5 8.7 Repeatability and reproducibility 15
3.6 Reagents 5 Figure 1 — Apparatus for the
3.7 Volumetric glassware 6 determination of the carbon dioxide content
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

(reference method) 8
4 Determination of the chloride content 6
Figure 2 — Apparatus for the
4.1 Principle 6
determination of the carbon dioxide content
4.2 Reagents 6 (alternative method) 10
4.3 Apparatus 6 Table 1 — Volumes of solutions for the
4.4 Procedure 6 preparation of calibration solutions and their
4.5 Expression of results 6 sodium oxide and potassium oxide
concentrations 12
4.6 Repeatability and reproducibility 7
Table NA.1 — Solutions for checking
5 Determination of the carbon purity of calcium solution 17
dioxide content reference method) 7
Table NA.2 — Solutions for calibrating
5.1 Principle 7
flame photometer 18
5.2 Reagents 7
5.3 Apparatus 7
5.4 Procedure 8
5.5 Expression of results 9
5.6 Repeatability and reproducibility 9
6 Determination of carbon dioxide
content (alternative method) 9
6.1 Principle 9
6.2 Reagents 9
6.3 Apparatus 9
6.4 Procedure 9
6.5 Expression of results 10
6.6 Repeatability and reproducibility 10
7 Determination of the alkali content
(reference method) 11
7.1 Principle 11
7.2 Reagents 11
7.3 Apparatus 11
7.4 Preparation of calibration solutions
and calibration graphs 11
7.5 Dissolution of the test portion 11
7.6 Procedure 12

© BSI 01-1999 3
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

4
blank
BS EN 196-21:1992

1 Object and field of application If the results of two determinations differ by more
than twice the standard deviation for repeatability,
This European Standard lays down the methods for
repeat the test and take the mean of the two closest
the determination of the chloride, carbon dioxide
values as the result.
and alkali content of cement.
3.4 Determination of constant mass
The standard describes the reference methods and,
in certain cases, an alternative method which can be Determine constant mass by making
considered as giving equivalent results. successive 15 min ignitions followed each time by
If other methods are used, their results shall be cooling and then weighing. Constant mass is
shown to be equivalent to the results given by the reached when the difference between two successive
reference methods. In the case of a dispute, only the weighings is less than 0,0005 g.
reference methods are applicable. 3.5 Preparation of the cement sample
This standard applies to cements and also to their Before starting the determinations, treat the
constituent materials, such as clinker and laboratory sample, taken in accordance with
blastfurnace slag. EN 196-7, as follows to obtain a sample for testing.
Take approximately 100 g of the sample using a
2 References sample divider or by quartering. Sieve this portion
EN 196-2, Methods of testing cement — Chemical on a 150 µm or 125 µm sieve until the residue
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

analysis of cement. remains constant. Remove the metallic iron from


EN 196-7, Methods of testing cement — Methods of the material retained on the sieve by means of a
taking and preparing samples of cement. magnet. Then grind the iron free fraction of the
retained material so that it completely passes
ISO 3534-1977, Statistics — Vocabulary and the 150 µm or 125 µm sieve. If this sample contains
symbols. particles of metallic iron such as those that may be
introduced accidentally during grinding, remove
3 General requirements for testing these iron particles completely using a magnetic
3.1 Number of tests stirrer in a suspension, for example in cyclohexane.
Transfer the sample to a clean dry flask with an
The number of tests for the various determinations
airtight closure and shake vigorously to mix it
(clauses 4 to 8) shall be two (see also 3.3).
thoroughly.
3.2 Repeatability and reproducibility
Carry out all operations as quickly as possible to
The standard deviation of repeatability gives the ensure that the sample is exposed to ambient air
closeness of agreement between successive results only for the minimum time.
obtained with the same method on identical test
3.6 Reagents
material, under the same conditions (same operator,
same apparatus, same laboratory and short time Use only reagents of analytical quality and distilled
interval3)). water, or water of equal purity, during the analysis.
The standard deviation of reproducibility gives the Unless otherwise stated % means percent by mass.
The concentrated liquid reagents used in this
closeness of agreement between individual results
standard have the following densities (r)
obtained with the same method on identical test
(in g/cm3 at 20 °C):
material but under different conditions (different
operators, different apparatus, different
laboratories and/or different times3)). hydrochloric acid 1,18 to 1,19
The standard deviations of repeatability and hydrofluoric acid 1,13
reproducibility are expressed in absolute percent. nitric acid 1,40 to 1,42
3.3 Expression of masses, volumes and results
perchloric acid 1,60 to 1,67
State masses in grams to the nearest 0,0001 g and
phosphoric acid 1,71 to 1,75
volumes from burettes in millilitres to the
nearest 0,05 ml. sulphuric acid 1,84
Express the results, given by the mean of two
determinations, as a percentage generally to two
decimal places.

3)
As defined in ISO 3534.

© BSI 01-1999 5
EN 196-21:1989

The degree of dilution is always given as a 4.3.4 Filter paper, coarse (pore diameter
volumetric sum, for example, dilute hydrochloric approximately 20 µm).
acid 1 + 2 means that 1 volume of concentrated 4.3.5 5 ml pipette
hydrochloric acid is to be mixed with 2 volumes of
water. 4.4 Procedure

3.7 Volumetric glassware Weigh (5 ± 0,05) g of cement and place in a 250 ml


beaker, add 50 ml of water and, while stirring with
The volumetric glassware shall be of analytical a glass rod, 50 ml of dilute nitric acid 1 + 2 (4.2.2).
accuracy, i.e. class A as defined in the ISO Heat the mixture to boiling, stirring occasionally,
standards on laboratory glassware. and boil for 1 min. Add 5 ml of silver nitrate
solution (4.2.5) by pipette (4.3.5) into the boiling
4 Determination of the chloride solution. Then boil for a maximum of 1 min and
content filter through a filter paper (4.3.4) washed before
4.1 Principle use with dilute nitric acid 1 + 100 (4.2.3) into
a 500 ml flask. Wash the beaker, glass rod and filter
This method gives the total halogen content except paper with dilute nitric acid 1 + 100 until the
for fluoride and expresses the result as Cl−. The volume of the filtrate and the washings is 200 ml.
cement sample is decomposed with boiling dilute Cool the filtrate and washings to below 25 °C.
nitric acid. Sulphides are oxidized into sulphates
Add 5 ml indicator solution (4.2.9) and titrate with
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

and do not interfere. The dissolved chloride is


precipitated using a known volume of a standard the ammonium thiocyanate solution (4.2.7) shaking
silver nitrate solution. After boiling, the precipitate vigorously until a drop of this solution produces a
is washed with dilute nitric acid and discarded. The faint reddish-brown colouration which no longer
filtrate and washings are cooled to less than 25 °C disappears on shaking. Record the volume V1.
and the residual silver nitrate is titrated with a If the chloride content of cement exceeds 0,17 %, it
standard ammonium thiocyanate solution using an will be necessary to start the test again with a
iron (III) salt as indicator. smaller quantity of cement.
4.2 Reagents Carry out the same procedure with no cement
sample and record the volume, V2, of ammonium
4.2.1 Concentrated nitric acid (HNO3). thiocyanate solution used in the blank titration.
4.2.2 Dilute nitric acid 1 + 2.
4.5 Expression of results
4.2.3 Dilute nitric acid 1 + 100.
Calculate the chloride content (in %) from the
4.2.4 Silver nitrate (AgNO3), dried at 150 °C. formula:
4.2.5 Silver nitrate solution 0,05 mol/l.
Dissolve 8,494 g of silver nitrate in water in
a 1 000 ml volumetric flask and make up to the
mark. Store the solution in a brown glass flask and (1)
protect it from the light.
4.2.6 Ammonium thiocyanate (NH4SCN).
4.2.7 Ammonium thiocyanate solution, where:
approximately 0,05 mol/l. Dissolve 3,8 g of
ammonium thiocyanate in water and make up
m1 is the mass of the cement test portion;
to 1 000 ml.
4.2.8 Ammonium iron (III) sulphate V1 is the volume of the ammonium
(NH4Fe(SO4)2.12H2O). thiocyanate solution used for the titration
4.2.9 Indicator solution. Add 10 ml of dilute nitric of the test solution;
acid 1 + 2 to 100 ml of a cold saturated solution of V2 is the volume of the ammonium
ammonium iron (III) sulphate in water. thiocyanate solution used for the titration
4.3 Apparatus of the blank solution.
4.3.1 Balance, capable of weighing to the The mean of the two results shall be rounded to the
nearest 0,0001 g. nearest 0,01 %.
4.3.2 10 ml burette, graduated to 0,1 ml.
4.3.3 Desiccator, containing anhydrous magnesium
perchlorate (Mg(ClO4)2).

6 © BSI 01-1999
EN 196-21:1989

4.6 Repeatability and reproducibility 5.3 Apparatus


The standard deviation of repeatability is 0,005 %. 5.3.1 Apparatus for the determination of the carbon
The standard deviation of reproducibility dioxide content. Figure 1 shows a typical piece of
is 0,010 %. apparatus which can be fitted with either a
cylindrical pressure container, a small electrical
5 Determination of the carbon dioxide compressor or a suitable suction pump which will
content (reference method) ensure an even flow of gas or air.
5.1 Principle The gas (air or nitrogen) entering the apparatus has
previously had its carbon dioxide removed by first
The cement sample is treated with phosphoric acid being passed through an absorbent tube or tower
to decompose the carbonate present. The carbon containing the carbon dioxide absorbent (5.2.5). The
dioxide liberated is entrained in a current of carbon apparatus consists of a 100 ml reaction flask (A)
dioxide-free gas or air through a series of absorption fitted with a three neck adaptor. Neck B1 is
tubes. The first two remove hydrogen sulphide and connected to a dropping funnel (O), neck B2 to a
water and the following then absorb carbon dioxide. connecting tube and neck C to a water cooled
Two absorption tubes, each containing a granular condenser. The funnel onto B1 and the connecting
absorbent for carbon dioxide and anhydrous tube onto B2 are joined together by means of a
magnesium perchlorate to retain the water formed Y-piece (P), so that the carbon dioxide-free air can
during the absorption reaction are weighed to
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

flow either through the connecting tube or the


determine the mass of carbon dioxide released. funnel by means of a Mohr clip (N). After the
5.2 Reagents condenser (L), the gas is passed through
5.2.1 Copper sulphate (CuSO4.5H2O). concentrated sulphuric acid (D), then through
absorption tubes containing the absorbent for
5.2.2 Copper sulphate solution, saturated. hydrogen sulphide (5.2.3) (E) and for
5.2.3 Absorbent for hydrogen sulphide. Place a water (5.2.4) (F) and subsequently through two
weighed quantity of dried pumice stone with a grain absorption tubes (G,H) which can be weighed and
size between 1,2 mm and 2,4 mm into a flat dish and which are three-quarters filled with the absorbent
cover with a volume of saturated copper sulphate for carbon dioxide (5.2.5) and a quarter with the
solution so the mass of the copper sulphate solution absorbent for water (5.2.4). The absorbent for
is approximately half that of the pumice stone. carbon dioxide (5.2.5) is placed upstream of the
Evaporate the mixture to dryness, while stirring absorbent for water (5.2.4) with respect to the gas
frequently with a glass rod. Dry the contents of the flow. Absorption tube (H) is followed by an
dish for at least 5 h in an oven at a temperature additional absorption tube (I), which also contains
of (155 ± 5) °C. Allow the solid mixture to cool in a the absorbent for carbon dioxide and water, which is
desiccator and store in an airtight bottle. fitted in order to protect absorption tube (H) against
5.2.4 Absorbent for water, anhydrous magnesium penetration by carbon dioxide and water from the
perchlorate (Mg(ClO4)2) with a particle size air.
between 0,6 mm and 1,2 mm. The absorption tubes (G) and (H) which are to be
5.2.5 Absorbent for carbon dioxide, synthetic weighed may have, for example, the following sizes.
silicates with a particle size between 0,6 mm
to 1,2 mm4) impregnated with sodium hydroxide External distance between branches 45 mm
(NaOH). Internal diameter 20 mm
5.2.6 Concentrated phosphoric acid (H3PO4).
Distance between the lower part of 75 mm
5.2.7 Concentrated sulphuric acid (H2SO4). the tube and the upper part of the
ground section
Tube wall thickness 1,5 mm

4)
The absorbent can be obtained ready for use.

© BSI 01-1999 7
EN 196-21:1989

5.3.2 Balance, capable of weighing to the Condition the closed absorption tubes (G) and (H)
nearest 0,0001 g. for 15 min in the balance case in order to achieve
5.3.3 Electric oven, which can be set at (105 ± 5) °C temperature equilibrium. Then weigh each tube
and at (155 ± 5) °C. separately. Shut off the flow of gas and attach the
tubes to the apparatus as shown in Figure 1.
5.3.4 Desiccator, containing anhydrous magnesium
perchlorate (Mg(ClO4)2). Wear protective gloves when carrying out this
operation.
5.4 Procedure
Then reopen the gas flow. After 10 min close
Weigh (1 ± 0,05) g of cement and place it in a absorption tubes (G) and (H), remove them, place
dry 100 ml distillation flask. Connect the flask to them in the balance case for 15 min and then weigh
the apparatus (5.3.1) as shown in Figure 1, but them separately. Repeat the passage of gas, removal
without the two absorption tubes (G) and (H). Pass and weighing of absorption tubes (G) and (H) for as
a current of carbon dioxide-free gas through the long as is required for the results of two successive
apparatus for 15 min at approximately 3 bubbles weighings of a tube not to differ by more
per second (bubble counter) via the connecting tube than 0,0005 g.
onto B2 (branch onto B1, Mohr clip closed). Release
the Mohr clip and remove the gas supply from the If the change in mass of the absorption tubes (G)
funnel (O). Add 30 ml concentrated phosphoric acid and (H) remains greater than 0,0005 g, renew the
absorbents in tubes (E) and (F).
into the dropping funnel and reconnect the gas
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

supply to fill the funnel (O).

Figure 1 — Apparatus for the determination of the carbon dioxide content (reference method)

8 © BSI 01-1999
EN 196-21:1989

Attach the weighed absorption tubes (G) and (H) to 6.2 Reagents
the apparatus, as shown in Figure 1. 6.2.1 Mercuric (II) chloride (HgCl2).
Open the funnel tap and allow the phosphoric acid
6.2.2 Absorbent for carbon dioxide, synthetic
to drop into the distillation flask (A). After the silicates with a particle size between 0,6 mm
reaction has ceased, heat the contents of the flask to and 1,2 mm5) impregnated with sodium hydroxide,
boiling and boil gently for 5 min. Maintain the gas
(NaOH).
flow through the apparatus until the flask has
cooled to room temperature. 6.2.3 Concentrated sulphuric acid (H2SO4).
Close absorption tubes (G) and (H), remove them 6.2.4 Dilute sulphuric acid 1 + 4.
and place them in the balance case for 15 min and 6.2.5 Absorbent for water, anhydrous magnesium
then weigh them separately. The increase in mass of perchlorate (Mg(ClO4)2) with a particle size
each tube is used for the calculation of the carbon between 0,6 mm and 1,2 mm.
dioxide content (see 5.5).
6.3 Apparatus
The carbon dioxide is practically completely
absorbed by tube (G). If the increase in mass of 6.3.1 Apparatus for the determination of the carbon
tube (H) exceeds 0,0005 g, renew the absorbent in dioxide content. The apparatus is shown in Figure 2.
tube (G) and start the test again. A small vacuum pump is used to generate reduced
pressure in the apparatus.
5.5 Expression of results
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

6.3.2 Balance, capable of weighing to the


Calculate the carbon dioxide content (in %) of the nearest 0,0001 g.
cement from the formula:
6.4 Procedure
Place (1 ± 0,05) g of cement into the 100 ml
(3) distillation flask A of the apparatus (6.3.1). Mix this
cement with a small (about 50 mg) amount of
where
mercuric (II) chloride (6.2.1) using a spatula and
then add enough water to form a slurry. Connect the
m2 is the mass of cement test portion; flask to the ground joint of the dropping funnel (O).
Then draw air for 15 min through the apparatus,
m3 is the increase in mass of tube G after passing the air through an absorption tower (J)
absorption; filled with absorbent (6.2.2) to remove the carbon
m4 is the increase in mass of tube H after dioxide before the air passes into the flask.
absorption. After stopping the flow of air add 25 ml to 30 ml of
sulphuric acid (6.2.4) from the dropping funnel (O)
The mean of the two results shall be rounded to the into the flask. Take care to ensure that some of the
nearest 0,01 %. acid remains in the dropping funnel as a seal.
If the carbon dioxide content calculated from Turn the vacuum pump on again, so that the current
equation (2) is less than 0,5 %, repeat the of air carries the liberated carbon dioxide through
determination with a cement sample weighing 2 g. the condenser (L) and the first two absorption
5.6 Repeatability and reproducibility tubes (F), filled with magnesium perchlorate (6.2.5)
The standard deviation of repeatability is 0,07 %. for the purposes of drying the air, to the two
The standard deviation of reproducibility is 0,10 %. previously weighed absorption tubes (G) filled with
absorbent (6.2.2). An absorption tube (I) filled with
6 Determination of the carbon dioxide magnesium perchlorate (6.2.5) and
absorbent (6.2.2) is fitted after these tubes in order
content (alternative method) to prevent penetration by the ambient air. A gas
6.1 Principle washing bottle (D) filled with sulphuric acid (6.2.3)
The carbon dioxide is driven off by means of is connected to this as a bubble counter.
sulphuric acid, absorbed by sodium hydroxide and
determined gravimetrically. Any hydrogen sulphide
present which may distort the results is absorbed by
mercuric (II) chloride.

5)
See page 6.

© BSI 01-1999 9
EN 196-21:1989

After about 10 min heat the contents of the flask to where


boiling and boil gently for 5 min. Maintain the air
flow through the apparatus until the flask has
m5 is the mass of the cement test portion;
cooled to room temperature. Close the taps and
remove the absorption tubes (G), place them in the m6 is the increase in mass of the absorption
balance case for 15 min in order to achieve tubes G after absorption.
temperature equilibrium and then weigh them.
The mean of the two results shall be rounded to the
6.5 Expression of results nearest 0,01 %.
Calculate the carbon dioxide content (in %) from the 6.6 Repeatability and reproducibility
formula:
The standard deviation of repeatability is 0,07 %.
The standard deviation of reproduciblity is 0,10 %.
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

Figure 2 — Apparatus for the determination of the carbon dioxide content


(alternative method)

10 © BSI 01-1999
EN 196-21:1989

7 Determination of the alkali content 7.3 Apparatus


(reference method) 7.3.1 Balance, capable of weighing to the
7.1 Principle nearest 0,0001 g.
A butane or propane flame is used to excite the 7.3.2 Calibrated burettes.
alkalis to emit their characteristic spectrum in the 7.3.3 Flame photometer, with which the intensities
visible range. The emission is proportional to the of the sodium line at 589 nm and the potassium line
alkali content at low concentrations. The influence at 768 nm can be measured: a sufficiently stable
of large quantities of calcium on the sodium apparatus shall be used.
determination is suppressed by means of phosphoric 7.3.4 Platinum dish.
acid.
7.3.5 Filter paper, medium pore size (pore diameter
7.2 Reagents approximately 7 µm).
7.2.1 General requirements 7.3.6 Platinum stirrer.
References to water mean distilled water or water of 7.4 Preparation of calibration solutions and
the same degree of purity with an electrical calibration graphs
conductivity of approximately 2 µS/cm. Use
reagents of analytical quality (see also 3.6); their Prepare the calibration solutions using the
alkali content shall be tested by means of this volumes of alkali stock solution, dilute hydrochloric
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

method. If the alkali content of a reagent acid 1 + 19 and dilute phosphoric acid 1 + 19 listed
exceeds 0,01 %, the batch concerned is unsuitable in Table 1. The volumes listed in lines 1 to 8 shall be
and shall therefore be replaced by another which made up to 1 000 ml with water. Store these
shall be tested in the same way. calibration solutions in polyethylene bottles.
7.2.2 Concentrated hydrochloric acid (HCl). Spray the calibration solutions into the flame of the
flame photometer (7.3.3). Spray the blank
7.2.3 Dilute hydrochloric acid 1 + 19. solution (Table 1, line 1) first and set the indication
7.2.4 Concentrated phosphoric acid (H3PO4). on the apparatus to 0. Then spray the other
calibration solutions in the order of increasing
7.2.5 Dilute phosphoric acid 1 + 19; store this
solution in a polyethylene flask. concentration (lines 2 to 8). Measure the intensities
for Na2O at 589 mm and for K2O at 768 nm.
7.2.6 Concentrated nitric acid (HNO3).
Plot graphs of the measured intensities against the
7.2.7 Concentrated perchloric acid (HClO4). corresponding concentrations of sodium oxide and
7.2.8 Concentrated hydrofluoric acid (HF). potassium oxide in the calibration solutions.
7.2.9 Sodium chloride (NaCl), dried at 105 °C to 7.5 Dissolution of the test portion
constant mass. 7.5.1 Cements completely soluble in acid
7.2.10 Potassium chloride (KCl), dried at 105 °C to (content of insoluble residue < 3 %)
constant mass. Weigh 0,1 g cement into a 50 ml beaker, make into
7.2.11 Alkali stock solution. Weigh about 0,566 g of a slurry with 10 ml of water and add 10 ml of dilute
sodium chloride and about 0,475 g of potassium hydrochloric acid 1 + 19. Warm the mixture until
chloride, transfer them to a 1 000 ml volumetric the cement has dissolved, breaking any lumps down
flask, add 100 ml each of dilute hydrochloric with a glass rod. Filter the suspension through the
acid 1 + 19 and dilute phosphoric acid 1 + 19, filter paper (7.3.5) into a 100 ml volumetric flask
dissolve and make up to the mark with water. This washing through with boiling water. Wash the filter
solution contains approximately 0,300 g each of paper and residue with boiling water until there is a
Na2O and K2O. The actual contents can be quantity of approximately 80 ml in the 100 ml
determined from the original quantities from the volumetric flask. Then allow the filtrate and
following formulae: washing water to cool to ambient temperature.
Add 10 ml of dilute phosphoric acid 1 + 19 to the
K2O (in g/l) = 0,6138 × actual mass of solution, make up to the mark and mix thoroughly.
potassium chloride in g (4)
Na2O (in g/l) = 0,5303 × actual mass of
sodium chloride in g (5)

© BSI 01-1999 11
EN 196-21:1989

7.5.2 Cements not completely soluble in acid 7.6 Procedure


Use the method described below where the content Spray the measuring solution produced as described
of insoluble residue, determined in accordance with in 7.5.1 or 7.5.2 into the flame of the flame
clause 9 of EN 196-2, exceeds 3 %. photometer (7.3.3). Measure the intensity of the
Weigh 0,2 g of cement into a platinum dish and sodium line at 589 nm and the potassium line
add 5 ml of concentrated nitric acid (7.2.6). Heat at 768 nm. Obtain the sodium oxide or potassium
the mixture, for example, on a hot-plate, and oxide concentration in the solution respectively by
evaporate to dryness. Disperse the residue from means of a linear interpolation from the intensities
evaporation in 5 ml of water, add 2 ml of and the associated concentrations of the calibration
concentrated perchloric acid6) (7.2.7) and then solutions measured as described in 7.4.
add 10 ml of concentrated hydrofluoric acid (7.2.8). Use the graphs plotted as in 7.4 to obtain the
Heat the mixture and evaporate to dryness. Prevent sodium oxide and potassium oxide concentrations of
overheating by frequent agitation by means of the the solution in mg/l or use the intensities and the
platinum stirrer (7.3.6). Add 40 ml of water associated concentrations of the calibration
and 20 ml of dilute hydrochloric acid 1 + 19 to the solutions with the next higher and the next lower
residue from evaporation and heat until the residue intensity for the calculation as follows.
has dissolved. Filter the suspension through the Calculate the sodium oxide (CNa2 O ) or potassium
filter paper (7.3.5) into a 200 ml volumetric flask
oxide (CK 2 O ) concentration of the sample from the
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

washing through with hot water. Wash the filter


paper and residue with hot water until the 200 ml intensities l Na2O or lK2O respectively using the
volumetric flask contains a volume of following formulae:
approximately 150 ml. Then allow the filtrate and
washing water to cool to ambient temperature.
Add 20 ml of dilute phosphoric acid 1 + 19 to the (6)
solution, make up to the mark with water and mix
thoroughly.

(7)

Table 1 — Volumes of solutions for the preparation of calibration solutions and their sodium
oxide and potassium oxide concentrations
Line Alkali stock solution Dilute hydrochloric acid Dilute phosphoric acid Na2O and K2O
(7.2.11) 1 + 19 1 + 19 concentrations
ml ml ml mg/l
1 — 100,0 100,0 Blank solution
2 3,3 99,6 99,6 1,0
3 8,3 99,1 99,1 2,5
4 16,7 98,3 98,3 5,0
5 25,0 97,5 97,5 7,5
6 33,3 96,6 96,6 10,0
7 41,7 95,8 95,8 12,5
8 50,0 95,0 95,0 15,0

6) Perchloric
acid vapours form explosive mixtures with organic materials. It is therefore necessary to take special precautionary
measures when working with perchloric acid: the use of fume cupboards flushed with water and a general ban on the use of
organic substances in the same fume cupboard.

12 © BSI 01-1999
EN 196-21:1989

where 7.8 Repeatability and reproducibility


The standard deviation of repeatability is:
CBn is the concentration of the sodium oxide or 0,01 % for the determination of Na2O;
potassium oxide respectively in the
calibration solution having a lower 0,02 % for the determination of K2O.
concentration than the measuring The standard deviation of reproducibility is:
solution in mg/l; 0,02 % for the determination of Na2O;
CBh is the concentration of the sodium oxide or 0,03 % for the determination of K2O.
potassium oxide respectively in the
calibration solution having a higher 8 Determination of the alkali content
concentration than the measuring (alternative method)
solution in mg/l;
8.1 Principle
IBn is the intensity of the calibration solution
having a lower concentration than the Cements which are completely soluble are treated
measured solution; with hydrochloric acid. Cements which are
incompletely soluble are first evaporated with
IBh is the intensity of the calibration solution hydrofluoric acid/sulphuric acid. The alkali contents
having a higher concentration than the of the solutions are determined by means of flame
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

measuring solution. photometry.


7.7 Expression of results 8.2 Reagents
Calculate the contents of sodium oxide or potassium 8.2.1 Concentrated hydrochloric acid (HCl).
oxide as percentages in the cement from the 8.2.2 Dilute hydrochloric acid 1 + 9.
following equations using the corresponding 8.2.3 Concentrated hydrofluoric acid (HF).
concentrations as determined in accordance
with 7.6. 8.2.4 Concentrated sulphuric acid (H2SO4).
8.2.5 Lithium chloride (LiCl); dried at 120 ºC.
Na2O = 0,1 CNa 2 O (8) 8.2.6 Sodium chloride (NaCl); dried at 105 ºC to
constant mass.
K2O = 0,1 CK2 O (9)
8.2.7 Potassium chloride (KCl); dried at 105 ºC to
where constant mass.
8.2.8 Stock solution7)

CNa 2 O is the sodium oxide concentration of Dissolve 0,610 g of lithium chloride, 0,2542 g of
the measuring solution as calculated sodium chloride and 0,1907 g of potassium chloride
by formula (6) (in mg/l); in water in a 1 000 ml volumetric flask and make up
to the mark.
CK2 O is the potassium oxide concentration
of the measuring solution as 8.2.9 Caesium chloride (CsCl).
calculated by formula (7) (in mg/l). 8.2.10 Aluminium nitrate (Al(NO3)3.9H2O).
8.2.11 Buffer solution7)
The mean of the two results for each oxide shall be
rounded to the nearest 0,01 %. Dissolve 50 g of caesium chloride and 250 g of
aluminium nitrate in water and make up
The total alkali content, A, is obtained by converting
to 1 000 ml.
the potassium oxide content to equivalent sodium
oxide from the formula:

A = Na2O + 0,658 K2O (10)

7)
These solutions can also be obtained made up ready for use.

© BSI 01-1999 13
EN 196-21:1989

8.3 Apparatus Spray the calibration solutions into the flame of the
8.3.1 Balance, capable of weighing to the flame photometer (8.3.3). Spray the blank
nearest 0,0001 g. solution 1 first and set the indication on the
apparatus to 0. Then spray the other
8.3.2 Electric oven, which can be set at (105 ± 5) °C calibration solutions in order of increasing
and at (120 ± 5) °C. concentration (2 to 7). Measure the intensities
8.3.3 Flame photometer, capable of measuring the for Na2O at 589 nm and for K2O at 768 nm. Plot
intensities of the sodium line at 589 nm and the graphs of the measured intensities against the
potassium line at 768 nm. The flame photometer corresponding concentrations of sodium oxide and
shall be operated with a propane-air flame at a potassium oxide in the calibration solutions.
relatively low temperature in order to avoid If a sufficiently stable photometer is used, it is only
distortion of the determination by alkaline earths. necessary to establish the calibration curves from
8.3.4 Filter paper, medium porosity (pore diameter time to time. However, the values measured for
approximately 7 µm). flasks 1 and 7 shall be checked for each analysis.
8.3.5 Platinum dish. 8.5 Procedure
8.3.6 Heating lamp. 8.5.1 Cements completely soluble in acid
8.3.7 Calibrated burette. (content of insoluble residue < 3 %)
Weigh 0,2 g cement into the platinum dish (8.3.5),
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

8.4 Plotting the calibration curve


make into a slurry with 3 ml of water and evaporate
In the case of cements which are completely soluble
to dryness after adding 20 ml of dilute hydrochloric
in hydrochloric acid, evaporate 20 ml of hydrochloric
acid 1 + 9. Add hot water and 2 ml of dilute
acid 1 + 9 to dryness, and, in the case of cements not
hydrochloric acid 1 + 9 to the residue and filter
completely soluble in acid, evaporate 15 ml of
through the filter paper (8.3.4) into a 100 ml
hydrofluoric acid (8.2.3) and 5 ml of sulphuric volumetric flask which already contains 10 ml of the
acid (8.2.4) to dryness for the purposes of preparing buffer solution (8.2.11). Wash the residue with hot
the calibration solutions for each calibration point.
water until the volumetric flask is almost filled to
In both cases dissolve the residue from evaporation the mark. Then cool to 20 °C and make up to the
with 2 ml of hydrochloric acid 1 + 9 and 3 ml of mark with water.
water. Transfer the solution to a 100 ml volumetric
Measure the solution in the flame
flask and add 10 ml of buffer solution (8.2.11). Add photometer (8.3.3). The scale values read off in
the following quantities of the stock solution (8.2.8) conjunction with the calibration curves (8.4) give
to the individual volumetric flasks using the
the concentrations of K2O and Na2O in mg/l
calibrated burette (8.3.7).
respectively.
8.5.2 Cements not completely soluble in acid
Volumetric flasks 1 2 3 4 5 6 7
Weigh 0,2 g cement into the platinum dish (8.3.5).
Stock solution (ml) 0 1 3 5 10 20 30 Make into a slurry with 3 ml of water and evaporate
Then make up the volumetric flasks to the mark after adding 5 ml of concentrated sulphuric
with distilled water: acid (8.2.4) and 15 ml of concentrated hydrofluoric
acid (8.2.3). Evaporate to dryness under the heating
With an original quantity of sample of 0,2000 g, the lamp (8.3.6). Mix the residue with hot water
values measured for flasks 1 to 7 correspond to a and 2 ml of hydrochloric acid 1 + 9 and continue the
K2O and Na2O content of: procedure as described in 8.5.1.
8.6 Expression of results
Volumetric 1 2 3 4 5 6 7
flasks Calculate the content of alkali as a percentage for an
original sample weight of 0,2000 g.
Na2O 0 0,07 0,20 0,34 0,67 1,35 2,02
The mean of the two results for each oxide shall be
content (%) rounded to the nearest 0,01 %.
K 2O 0 0,06 0,18 0,30 0,60 1,20 1,81 Convert the potassium oxide content to the
content (%) equivalent sodium oxide content using formula (10).
State the sum of the sodium oxide values as the total
alkali content A.

14 © BSI 01-1999
BS EN 196-21:1992

8.7 Repeatability and reproducibility


The standard deviation of repeatability is:
0,01 % for the determination of Na2O;
0,02 % for the determination of K2O.
The standard deviation of reproducibility is
0,02 % for the determination of Na2O;
0,03 % for the determination of K2O.
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

© BSI 01-1999 15
BS EN 196-21:1992

National annex NA (informative) NA.3.3 Aluminium solution, (20 g of Al per litre.)


Test method for the alkali content of Prepare this solution by dissolving super-purity
aluminium metal in nitric acid. The following
cement to be used for the purpose of alternative methods have been found to be
calculating the reactive alkali content convenient.
of concrete a) From a sheet of super-purity aluminium,
between 0.5 mm and 3 mm thick, cut two pieces,
NA.1 General one about 120 mm × 360 mm and the other
This method of test is intended to be used for the about 120 mm × 200 mm. Bend the larger piece
determination of the alkali content of cement only to fit inside a 2 l borosilicate glass or plastics
for the purpose of calculating the reactive alkali beaker, then bend the smaller piece to fit inside
content of the concrete mix when aggregate the larger one. Near the middle of the top edge of
containing an alkali reactive constituent, or when each piece drill a hole and twist a separate piece
an unfamiliar cement/aggregate combination, is in of thick aluminium wire through each one to
use (see 4.2.4 of BS 5328-1). make electrical connections. Clean both sheets by
washing successively in dilute hydrochloric
NA.2 Apparatus acid (1 + 9), water and industrial methylated
spirits. Allow the sheets to dry, weigh them
NA.2.1 Simple flame photometers, burning natural
to ± 0.05 g and place them in the beaker ensuring
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

or bottled gas and incorporating interchangeable


that they do not make contact with each other.
filters which select light of the appropriate
Cover the beaker with a plastics cover and take
wavelengths, are generally suitable. Use in
accordance with the manufacturers’ instructions. the wire leads out through holes in the cover.
Place the beaker and contents in a fume
NA.3 Reagents cupboard. Connect the lead from the outer
aluminium sheet to the positive terminal and the
Prepare reagents by the following methods and lead from the inner sheet to the negative terminal
store in well stoppered rigid polyethylene or of a source of direct current which can be varied
polypropylene bottles. from 0 V to 12 V8).
NA.3.1 Sodium stock solution, (100 mg of Na2O While stirring, add 200 ml of nitric acid
per litre.) Dry a quantity of analytical reagent to 800 ml of warm water (about 50 °C) in a
grade, or purer, NaCl at 250 °C, weigh 0.1886 g, separate beaker and transfer this solution to the
dissolve in water and dilute to 1 000 ml. beaker containing the electrodes. Switch on the
NA.3.2 Potassium stock solution, (100 mg of K2O power supply and increase the applied voltage to
per litre.) Dry a quantity of analytical reagent obtain a current of about 8 A. When it is
grade, or purer, KCl at 250 °C, weigh 0.1583 g, considered that sufficient aluminium has
dissolve in water and dilute to 1 000 ml. dissolved9), switch off and disconnect the power
supply. Remove, wash, dry and reweigh the
aluminium sheets together with any pieces that
may have become detached.
Calculate from the loss of mass the volume the
solution should occupy to contain 20.0 g of
aluminium per litre. Dilute to this volume with
water.
b) Cut sufficient super-purity aluminium sheet or
wire into small pieces and wash successively
with dilute hydrochloric acid (1 + 9), water and
industrial methylated spirits. Dry, weigh
out 20.0 g and transfer a little at a time, using
clean tongs, to a gently boiling solution of 100 ml
of nitric acid in 500 ml of water to which 2 ml of
saturated mercuric chloride solution have been
added.

8) There should be a suitably rated fuse in the circuit to protect the power supply.
9) With a current of 8 A approximately 20 g of aluminium is dissolved in 21/2 h. A lower current for a longer time is permissible.

16 © BSI 01-1999
BS EN 196-21:1992

Maintain the volume at about 600 ml by NOTE 1 For instruments using internal standardization, add
adding water as necessary. When all the metal the internal standard solution to the calibration solutions in
accordance with the manufacturers’ instructions.
has dissolved, cool the solution and dilute
NOTE 2 In some cases different Na 2O and/or K 2O
to 1 000 ml. concentration ranges may be required. Should this be necessary,
NA.3.4 Calcium solution, (6.3 g of CaO per litre.) modify the alkali contents of the calibration solutions as
appropriate but take the Al, Ca and HNO3 additions from
Weigh 11.25 g of analytical reagent grade, or purer. Table NA.2. The concentration of the scale-100 solution should
CaCO3 into a beaker, cover with 100 ml of water and not exceed that recommended by the instrument manufacturer.
carefully add 25 ml of HNO3. When solution is
complete dilute to 1 000 ml with water. NA.5 Procedure
Test for sodium and potassium contamination by NA.5.1 Determination of alkalis soluble in
the following procedure. nitric acid
Prepare solutions A, B and C, using the quantities NOTE This procedure is generally applicable to cements, and
given in Table NA.1. other materials, which are almost completely decomposed by
nitric acid solution.
With the sodium light filter in place, set the flame
Weigh out 500 mg of sample into a beaker made
photometer to read 0 with solution A and 100 with
from a material which will not yield sodium or
solution B. Spray solution C and record the reading.
potassium to the solution. Add about 50 ml of cold
Replace the sodium light filter by the potassium
water, disperse the powder by swirling and add 5 ml
light filter and repeat the procedure. If readings
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

of nitric acid. Swirl again, add hot water to


higher than 2 for Na2O (0.1 mg/l) or higher than 1
about 150 ml total volume, bring to the boil and
for K2O (0.1 mg/l) are obtained reject the calcium
simmer for 5 min to 10 min. Cool, transfer to
solution and prepare a new solution using a purer
a 500 ml volumetric flask, add 50 ml of the
source of CaCO3.
aluminium solution, dilute to the mark with
NA.4 Calibration water10) and mix thoroughly.
Allow to settle and pour off sufficient solution for the
Prepare calibration solutions using the quantities
flame determinations. If necessary, centrifuge or
given in Table NA.2.
filter the solution before spraying.
With the sodium light filter in place set the flame
photometer to read 0 with scale-zero solution
and 100 with scale-100 solution. Record the
readings obtained for each of the intermediate
calibration solutions, checking and adjusting if
necessary, the 0 and 100 settings between each
reading. Replace the sodium light filter by the
potassium light filter and repeat the calibration
procedure. Construct the calibration graphs of
instrument reading against milligrams per litre of
Na2O or K2O.
Table NA.1 — Solutions for checking purity of calcium solution
Solution Sodium stock Potassium stock Aluminium Calcium HNO3 Water to make
solution solution solution solution
ml ml ml ml ml ml

A 0 0 50 0 5 500
B 25 50 50 0 5 500
C 0 0 50 50 5 500

10) Beforemaking up to volume, add, where necessary, sufficient of the standard calcium solution to bring the concentration of
calcium oxide in the diluted solution to approximately 630 mg/l, and, for instruments using internal standardization, also add
the internal standard solution in accordance with the manufacturer’s instructions.

© BSI 01-1999 17
BS EN 196-21:1992

Table NA.2 — Solutions for calibrating flame photometer


Calibration Sodium Potassium Aluminium Calcium HNO3 Water to Equivalent
solution stock stock solution solution make concentrations
solution solution
Na2O K2 O

ml ml ml ml ml ml ml ml

Scale-zero 0 0 100 100 10 1 000 0 0


Scale-100 50 100 100 100 10 1 000 5 10
Nominal 20 5 10 50 50 5 500 1 2
Nominal 40 10 20 50 50 5 500 2 4
Nominal 60 15 30 50 50 5 500 3 6
Nominal 80 20 40 50 50 5 500 4 8
With the sodium light filter in place set the flame CAUTION. Perchloric acid fumes can react
photometer to read 0 with scale-zero solution explosively with some absorbent organic materials.
and 100 with scale-100 solution. Spray the sample Fume cupboards for use with perchloric acid should
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

solution and take several readings, checking and never be constructed in wood but polyvinyl chloride
adjusting if necessary, the 0 and 100 settings and stainless steel have been found suitable. All
between each reading. Replace the sodium light materials of construction should be non-absorbent
filter by the potassium light filter and again set the and interior surfaces should be free from crevices
flame photometer to read 0 with scale-zero solution and ledges that could prevent complete washing
and 100 with scale-100 solution. Spray the sample down. Guidance on the suitability of a fume
solution and take several readings, checking, and cupboard for use with perchloric acid should be
adjusting if necessary, the 0 and 100 settings sought from the manufacturer.
between each reading. Recommendations on all aspects of fume cupboards
If the sample gives a reading of more than 100 for are given in BS 725811).
either sodium or potassium, transfer 50 ml of Add 5 ml of nitric acid, cool and transfer to a 500 ml
sample solution to a 100 ml graduated flask, dilute volumetric flask. Add 50 mL of the aluminium
to the mark with scale-zero solution and re-spray. In solution, dilute to 500 ml with water12) and mix
exceptional cases a further dilution may be thoroughly. Then repeat the procedure in NA.5.1
required. from the second paragraph onwards.
NA.5.2 Determination of total alkalis Make blank determinations following the above
Weigh 500 mg of sample into a platinum basin and procedure with no sample present.
remove carbon or carbonaceous matter by careful
ignition at 925 ± 25 °C. Cool the sample and NA.6 Calculation
add 20 ml of water and 10 ml of perchloric acid. Stir, From the respective calibration graphs for Na2O
add 20 ml of hydrofluoric acid and evaporate the and K2O (see NA.4) convert flame photometer
solution to strong fumes of perchloric acid in a readings to concentrations (milligrams per litre).
mechanically ventilated fume cupboard. Add 20 ml For the determination of total alkalis deduct the
of water and 5 ml of perchloric acid and again blank determinations from the readings given
evaporate until fumes just cease to be emitted. by Na2O and K2O respectively.
Where dilution of the original solution has been
made apply the dilution factor.

11)
BS 7258 Laboratory fume cupboards.
12) Beforemaking up to volume, add, where necessary, sufficient of the standard calcium solution to bring the concentration of
calcium oxide in the diluted solution to approximately 630 mg/l, and, for instruments using internal standardization, also add
the internal standard solution in accordance with the manufacturer’s instructions.

18 © BSI 01-1999
BS EN 196-21:1992

Calculate the alkali contents as percentages by the


formula:
alkali oxide = 0.1D(C – B)
where

C is the concentration of alkali (milligrams


per litre);
B is the blank reading (when determining
alkalis soluble in nitric acid this is zero);
D is the dilution factor, i.e.
volume of diluted solution (ml)
------------------------------------------------------------------------------------
volume of aliquot taken (ml)
(where no dilution has been made this
equals one).
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

© BSI 01-1999 19
BS EN 196-21:1992

National annex NB (informative)

Committees responsible
The United Kingdom participation in the preparation of this European Standard was entrusted by the
Technical Sector Board for Building and Civil Engineering (B/-) to Technical Committee B/516 upon which
the following bodies were represented:

British Aggregate Construction Materials Industries


British Cement Association
British Precast Concrete Federation
British Ready Mixed Concrete Association
Cement Admixtures Association
Cementitious Slag Makers Association
Concrete Society
County Surveyors’ Society
Department of the Environment (Building Research Establishment)
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

Department of the Environment (Property Services Agency)


Department of Transport
Federation of Civil Engineering Contractors
Mortar Producers Association
National Rivers Authority
Quality Ash Association
Society of Chemical Industry
Water Services Association of England and Wales

National annex NC (informative)

Cross-references

Publication referred to Corresponding British Standard

BS EN 196 Methods of testing cement


EN 196-2 Part 2:1992a Chemical analysis of cement
EN 196-7 Part 7:1992 Methods of taking and preparing samples of cement
ISO 3534:1977 BS 5532 Statistical terminology
Part 1:1978 Glossary of terms relating to probability and general terms relating
to statistics
a
In preparation.

20 © BSI 01-1999
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

blank
|
|
|
|
|
|
|
|
|
BSI Ð British Standards Institution
|
|
|
|
|
|
| BSI is the independent national body responsible for preparing British Standards. It
|
| presents the UK view on standards in Europe and at the international level. It is
| incorporated by Royal Charter.
|
|
| Revisions
|
|
| British Standards are updated by amendment or revision. Users of British Standards
|
| should make sure that they possess the latest amendments or editions.
|
|
| It is the constant aim of BSI to improve the quality of our products and services. We
|
| would be grateful if anyone finding an inaccuracy or ambiguity while using this
| British Standard would inform the Secretary of the technical committee responsible,
|
| the identity of which can be found on the inside front cover. Tel: 020 8996 9000.
|
| Fax: 020 8996 7400.
|
|
| BSI offers members an individual updating service called PLUS which ensures that
|
| subscribers automatically receive the latest editions of standards.
|
|
| Buying standards
|
| Orders for all BSI, international and foreign standards publications should be
|
| addressed to Customer Services. Tel: 020 8996 9001. Fax: 020 8996 7001.
|
Licensed copy:Hyder Consulting, 13/02/2005, Uncontrolled Copy, © BSI

|
| In response to orders for international standards, it is BSI policy to supply the BSI
|
| implementation of those that have been published as British Standards, unless
|
| otherwise requested.
|
|
| Information on standards
|
| BSI provides a wide range of information on national, European and international
|
| standards through its Library and its Technical Help to Exporters Service. Various
|
| BSI electronic information services are also available which give details on all its
|
| products and services. Contact the Information Centre. Tel: 020 8996 7111.
|
| Fax: 020 8996 7048.
|
|
| Subscribing members of BSI are kept up to date with standards developments and
| receive substantial discounts on the purchase price of standards. For details of
|
| these and other benefits contact Membership Administration. Tel: 020 8996 7002.
|
| Fax: 020 8996 7001.
|
|
| Copyright
|
|
| Copyright subsists in all BSI publications. BSI also holds the copyright, in the UK, of
|
| the publications of the international standardization bodies. Except as permitted
| under the Copyright, Designs and Patents Act 1988 no extract may be reproduced,
|
| stored in a retrieval system or transmitted in any form or by any means ± electronic,
|
| photocopying, recording or otherwise ± without prior written permission from BSI.
|
|
| This does not preclude the free use, in the course of implementing the standard, of
|
| necessary details such as symbols, and size, type or grade designations. If these
|
| details are to be used for any other purpose than implementation then the prior
| written permission of BSI must be obtained.
|
|
| If permission is granted, the terms may include royalty payments or a licensing
|
| agreement. Details and advice can be obtained from the Copyright Manager.
|
| Tel: 020 8996 7070.
|
|
|
|
|
|
|
|
|
BSI |
|
389 Chiswick High Road |
|
London |
|
W4 4AL |
|
|
|
|
|
|

You might also like