You are on page 1of 21

NCBI Bookshelf. A service of the National Library of Medicine, National Institutes of Health.

StatPearls [Internet]. Treasure Island (FL): StatPearls Publishing; 2022 Jan-.

Nuclear Medicine Computed Tomography Physics


Authors

Dibya Prakash1; Dawood Tafti2.

Affiliations
1 Indian College of Nuclear Medicine
2 Brooke Army Medical Center

Last Update: June 12, 2022.

Continuing Education Activity


Interdisciplinary imaging has significantly improved patient care. In nuclear medicine, computed tomography (CT) is
used with single-photon emission computed tomography (SPECT) and positron emission tomography (PET) to
improve the quantification of functional imaging by correcting for attenuation, scatter, and partial volume effects and
combinedly called SPECT-CT and PET-CT respectively. This article covers the historical developments of CT,
SPECT, and PET, their physics, the concept of hybrid imaging, and the benefits as well as limitations associated with
dual imaging. Various terms used in CT imaging, such as slip-ring technology, helical and spiral CTs, pitch, multiple-
detector CT technology, X-ray production, CT detectors, and Hounsfield unit, are discussed in detail with illustrative
images and examples. This activity highlights the tenets of PET, including sensitivity, spatial resolution, scatter and
random events, attenuation, and their corrections.

Objectives:

Describe the physics of computed tomography used in nuclear medicine for hybrid imaging.

Summarize the method of corrections used in positron emission tomography (PET) and single-photon emission
computed tomography (SPECT) data to generate fusion images.

Explain the artifacts, their cause, and possible solutions observed during hybrid imaging of PET-CT and
SPECT-CT.

Outline the advantages and disadvantages of PET-CT and SPECT-CT imaging.

Access free multiple choice questions on this topic.

Introduction
In nuclear medicine, computed tomography (CT) is used with positron emission tomography (PET) and single-photon
emission computed tomography (SPECT) machines, which are combinedly called PET-CT and SPECT-CT,
respectively. This dual-modality acquires co-registered images of structural and functional details in a single study
with datasets complementary to each other and highlights the abnormal functional images with accurate localization
of structures. CT imaging, when used in nuclear medicine, helps to improve the quantification of functional images by
correcting for attenuation, scatter, and partial volume effects. The images from both modalities are fused using
software, and standard uptake values (SUVs) are calculated.[1][2][3]

The concept of combining planar images of two modalities dates back to the 1960s, even before the invention of CT
machines. However, software for fusion imaging started to develop in the late 1980s.[4] The proper dual-modality
fusion approach was pioneered at the beginning of the 1990s by Hasegawa et al. at the University of California, who
used a diagnostic CT scanner with a SPECT camera to make the first SPECT/CT device. But the first commercial
SPECT-CT was introduced in 1999 by General Electric (The Hawkeye) using a low-power X-ray source with a
conventional SPECT camera.[5][6][7] 

A similar concept of combining PET with CT was suggested independently in 1991, and the first working prototype
PET/CT scanner was developed in 1998, but the first commercial PET-CT was available in 2001.[8][9]

The development of SPECT/CT has improved the sensitivity and the confidence that many nuclear medicine
investigations are interpreted.[10] It has been useful in diagnosing hyperparathyroidism, thyroid disorders, sentinel
node imaging, neuroendocrine imaging, bone disorders, brain imaging, infection imaging, and studies requiring
patient-specific dosimetry for dose planning. In many aspects, the combined SPECT/CT is a game-changer, initiating
a paradigm shift in SPECT utilization into the quantitative realm in nuclear emission tomography.[11]

Similarly, the PET-CT has been a game-changer in managing oncology, cardiology, and neurology patients, along
with many others. Head and neck cancers, thyroid cancers, lung cancers, breast cancers, esophageal cancers,
colorectal cancers, melanoma, lymphoma, and solitary pulmonary nodules have shown better accuracy when
diagnosed with PET-CT. The difference in accuracy between PET/CT and the PET alone or the CT alone for staging
and restaging is statistically substantial and averaging nearly 10 percent to 15 percent across all malignancies.

Function
The Computed Tomography Physics

Computed tomography (CT), also called computer-aided tomography (CAT), is a diagnostic imaging modality that
uses rotating X-ray tubes, detectors, and computers to generate three-dimensional images using a linear attenuation
map created from the object when X-rays pass through it. Sir Godfrey N. Hounsfield built the first CT scanner in a
British research laboratory in 1967.[12] 

In September 1971, the first clinical CT machine was installed at Atkinson-Morley Hospital in London.[13] Since
then, the technology, capabilities, and clinical applications have changed enormously, especially after introducing
multiple-row-detector CT scanners in 1998.[14] Since then, significant advancement has been made in scanner gantry
rotation speeds, temporal and spatial resolutions, detector coverage, and overall scan times.[15]

The initial CT scanner developed at EMI Corp (a British electronic firm) had one X-ray tube with one sodium iodide
scintillator detector and was called the first-generation CT scanner. The tube and the detector were linearly translated
and then rotated with a 1-degree increment to collect 180 views over 180 degrees, unlike 360 degrees for the current
scanners. The pencil beam had 3 mm width and 13 mm length and took around 5 or 6 minutes to complete the scan
(see "Narrow beam with translate-rotate motion"). To reduce the data acquisition time, the second-generation CT
scanners were introduced in late 1974 with the same translation-rotational motion but with a multi-detector system
that formed multiple narrow beams (see "Multiple-narrow beams with translate-rotate motion").[16]

In late 1975, with a 30-detector system, the acquisition time was reduced to 20 seconds. However, the heavy X-ray
tube and associated electronics were causing frequent misalignments in the hardware due to the fast translational and
rotational movements and bringing severe artifacts in the images. This need paved the way for 3rd generations of CT
scanners.

The third generation of CT scanners was introduced in late 1975, eliminating the translational motion and performing
the only rotational motion. The X-ray beam was widened as a fan beam to cover the whole patient in the field of view
(FOV) using an array of detectors and took less than 5 seconds to perform the acquisition.[16] 

The tube and detector assembly were stationary concerning each other while rotating the whole set-up around the
patient (see "Fan-beam, rotate only"). These models used a power supply to the X-ray tube with electric cables, and
the length of the cable restricted the continuous circular motion, which forced the scanner to move clockwise and
counterclockwise to acquire adjacent slices. The heavy gantry's continuous acceleration and deceleration limited the
scan speed to approximately 2 seconds per rotation. However, this motion caused detector stability issues and image
artifacts and led to the investigation of 4th generations of CT scanners. Though the third-generation scanners became
very popular, and today's state-of-the-art scanners are based on the third-generation CT scanners only with hardware
refinements such as slip-ring technology and multiple-detector rows.

The fourth generation of CT scanners was introduced in 1976, where the detector formed an enclosed stationary ring,
and the X-ray tube rotated in a circular motion around the patient inside that ring and achieved a 1-second scan
duration (see "Fan-beam with stationary circular detector"). However, the tube required a large ring of 170-180 cm to
maintain sufficient tube-patient distance. The full circle detector ring increased the cost of the scanner considerably;
however, the spatial resolution could be reached at around 4 mm. Also, the scatter correction problem could not be
solved in this generation.

The electron beam CT scanners (EBCT) was introduced in 1984 for ultrafast cardiac scans and completed the scan in
10 to 20 milliseconds to freeze the heart's motion. The system had no mechanically moving part and was sealed in a
vacuum, similar to cathode-ray tubes. The high-speed electron beam is focused and deflected by carefully built coils
to sweep around the target ring. The fan-beam X-rays are produced and collimated to the set of detectors positioned in
an arc of 216 degrees. 8 cm of coverage along the patient axis is obtained for the heart. However, the higher
equipment cost and poor image quality for general imaging limit the use of EBCTs.

Until 1987, cables were used to supply the power to X-ray tubes and communicate detector signals. The rotation was
coming to a halt every after 360 degrees, and the cables were looped onto a drum, released throughout the one
rotation, and then spooled back in the opposite direction. Scanning, braking, and reversal took at least 8 to 10 seconds
for each cycle, with only 1 or 2 seconds dedicated time for data acquisition. As a result, the temporal resolution was
poor compelling longer procedural duration. The slip ring technology, first introduced in 1987, eliminated interscan
delays and permitted continuous nonstop rotation. The slip ring is a circular ring that transfers power and data signals
from the stationary part to the rotating part of the scanner (see "Slip-ring technology").

The slip ring technology led to the foundation of spiral CTs, and the first, became available in 1990; however, its
clinical acceptance took another two years.[17] In a conventional CT scanner, the collimated X-ray beam is rotated in
a circular motion around the object or the patient to measure the attenuation profiles from many angular positions and
generate various image planes. The images are reconstructed, and the object is pushed for a small increment within
the gantry for the next measurement. This procedure is repeated, and the high-quality images are generated slice by
slice. This imaging method is called axial imaging or the step-and-shoot method and takes a longer duration. During
the scan, the patient should not move to avoid blurry images or motion artifacts. Shorter scan duration remained a
primary requirement to avoid motion-related artifacts, which led to the invention of spiral CT scanners.

The object is continuously moved through the gantry in Z-direction in spiral CTs. In contrast, the X-ray tube and
detector assembly move around the object in a circular rotatory motion in the XY plane to acquire the images
continuously (see "Spiral CT machines"). The volume covered in a particular number of spiral turns is represented by
a single data set. The spiral CTs are also called helical CTs. Technically or practically, there is no difference between
the two terms, and it is only the CT manufacturer naming it differently. The spiral CTs allowed the entire organ to be
imaged in a single breath-hold.

A concept of "pitch" is introduced in helical CTs to define the table movement speed and the covered slice thickness.
It is defined as "distance traveled by the table" in one 360 degrees gantry rotation divided by "the beam collimation."
The pitch has no unit. For example, suppose the table moves 5 mm in one second, the beam collimation is 5 mm, the
pitch stands at 1, and the entire X-ray helix is covered with X-ray exposure. Whereas, if the table moves 7.5 mm
during one rotation of the X-ray tube and the beam collimation is 5 mm, the pitch would be 1.5.
Whenever the pitch is more than 1, there is a gap in the X-ray helix; however, the entire volume is still covered. This
allows a faster scan with a lower radiation burden to the patient but certainly compromises image quality. A balance
between image quality and the speed of acquisition should be maintained. In another example, if the table moves 4
mm in one rotation and the beam collimation is 5 mm, the pitch would be 0.8. There would be an overlap in X-ray
exposure and the exposed area gets more radiation. The pitch values generally vary from 0.7 to 1.5, depending on the
study. 

Multislice CTs

The first modern multislice CTs (MSCT), or "multiple-detector-row CT" (MDCT) scanners, were introduced in late
1998; however, the concept was introduced in 1992 (twin CT).[18] It contains a rotating arc of detectors and the X-ray
tube(s), allowing volumetric coverage of data acquisition with thin slices. The detectors are placed along the Z-
direction in multiple rows (see "Multi-slice CT Scanners"). The initial scanners had 16-rows of detector elements with
1.25 mm or 1 mm width along the z-axis with total coverage of 20 mm. Different vendors used flexible combinations
of widths such as 1.25 mm, 2.5 mm (1.25 × 2), 5 mm (1.25 × 4), 10 mm (2.5 × 4), 20 mm (5 × 4) to create 4-slice data
channels, as handling of 16-slice data was not feasible at that time. As the capabilities increased, in 2003 - 2004, the
manufacturers introduced 6-, 8-, 16-, 32- and 40-slice models. In 2005, 64-slice models were also introduced with a
total z-axis coverage of 32-40 mm depending upon available slice thickness of 0.5 mm to 0.625 mm.[18] 

Subsequently, 2007 to 2008 saw 128-, 256-, 320-slice, and dual-source CT scanners with z-coverage of up to 16 cm.
[19] By 2012, Toshiba medical (now Canon Medical System) introduced a 640-slices CT scanner, which was digitally
enhanced from 320-slice to 640-slice with double-slice technology and coneXact™ reconstruction algorithm, though
the z-axis coverage remained the same at 16 cm.[20]

Currently, commonly available CT scanners are 16-, 32-, 40-, 64- and 128-slice, with less commonly available 256-
and 320-slice CT scanners as their cost is significantly higher. Commandeur et al. (2019) compared different state-of-
the-art models with their technical specifications and showed available tube rotation time up to 0.270 seconds,
temporal resolution up to 0.135 seconds, and slice thickness up to 0.5 mm depending on model to model. Many such
systems allow simultaneous acquisition of the whole organ, such as the heart, brain, blood vessels, joints, etc., due to
larger Z-coverage. The best reported in-plane spatial voxel dimensions of state-of-the-art MSCTs are 0.23–0.4 mm
and 0.3 mm in the Z-direction.[15] 

X-ray Production

German physicist Wilhelm Conrad Roentgen discovered the first X-ray tube in November 1895.[21] Though the
design of X-ray tubes has changed considerably since its inception, the basic principles remain the same. The early X-
ray tubes were unreliable and inefficient. In 1913, an American Physicist Dr. William David Coolidge, used the
thermionic emission phenomenon to improve the release of electrons from filament using electric current and
accelerated towards the anode to produce X-rays. This method of production is the prototype for the current
generation of X-ray tubes.

The electrons released from the heated cathode are accelerated towards the high voltage anode. This flow of
accelerated electrons is called tube current. The electrons interact with the anode and X-rays are produced in all
directions, but they are collimated to emerge out from a particular direction and form a beam. The number of
electrons liberated from the cathode depends upon the temperature of the filament, which is regulated by the electric
current to the filament.

For heating the filament, modern X-ray tubes receive 4 to 5 amperes of current. When there is no voltage difference
between the anode and cathode, the thermionic emissions remain in the vicinity of the cathode, forming a cloud of
electrons. These electrons travel only when there is a high electrical potential difference in kilovoltage (kV).
Once the potential difference is applied, the electrons rush toward the anode and generate kinetic energy while
moving towards the anode. The electron flow from the cathode to the anode forms an electric current and this is in the
range of 25 – 1000 milliampere (mA). The cathode is surrounded by a focusing cup, which deflects electrons at a
specific location on the anode, called the focal spot. The size of the focal spot is an important factor and affects the
image quality and sharpness of any X-ray imaging system.[22] A smaller focal spot gives better image quality.
Routine diagnostic CTs have 2.0 mm of focal spot size.

The tube voltage kV determines the energy of X-rays. The X-rays have energy varying from 0 keV to the maximum
or peak applied voltage (kVp). Since X-rays, less than 20 keV are not useful for imaging and only add to additional
radiation burden without any contribution to the image quality, they are stopped using a filter, usually made up of
aluminum, and placed at a tube window. Also, the amount of electricity (mA) used to heat the coil for a particular
duration (seconds) determines the total number of X-rays emitted. The product of these two terms is called mAs or
milliampere-second.

Tungsten coils have remained the choice of material to act as a cathode in a chemically and physically harsh
environment; however, a thin layer of tungsten on graphite substrate is commonly used as anode material.[21]
Tungsten allows heating temperatures beyond 2000 degrees Celsius and is a good conductor of heat & electricity. This
element has a high atomic number (Z =74). The good conduction of heat allows the anode to remove a large amount
of thermal energy generated during electron interactions.

The good electrical conductivity keeps electrons flowing freely and does not allow them to build upon the anode
surface to create repulsive conditions for the incoming electron stream. The higher Z materials tend to release
electrons easily. Since only 0.2 percent of thermionic electrons generate characteristic and bremsstrahlung X-rays, a
balance of 99.8 percent of electrons release their kinetic energy in heat, and therefore a huge amount of heat is
generated within the tube; this creates the need for a material having a high melting point (Tungsten has 3370 degrees
Celcius melting point). Gold (Z=79) and molybdenum (Z=42) are other traditional target materials. The anode and
cathode assembly are housed within an evacuated glass or ceramic tube (see "Production of X-rays"). The vacuum
within the tube helps prevent the interaction of electrons with any gas molecules.

CT Detectors

The X-rays generated from the X-rays tube are read by the detectors placed opposite to this in an arc formation. These
detectors are made up of an inorganic transparent ceramic detector and chosen based on their qualities of good light
output, relatively lower afterglow effect, and resistance to the damage created by the radiation over prolonged use.
Some of the commonly used materials are cesium iodide thallium activated [CsI (Tl)] crystals, cadmium tungstate
(CdWO), gadolinium oxysulfide (GdOS), yttria, and gadolinia with europium activated (Y, Gd)O: Eu or YGdEuO.
General Electric (GE) Healthcare has used (YGd)O in many CT scanners, whereas Siemens has used proprietary
material Ultra-fast Ceramics (UFC) and Stellar detectors.

The scintillation photon generated by the ceramic detectors is converted into electrical signals by photodiodes. These
semiconductor devices work similarly to photomultiplier tubes (PMTs) by converting light photons into the current.

Hounsfield Unit

When X-rays pass through the tissues, they are attenuated, and a linear attenuation map is generated from the 360
degrees of rotation of the X-ray tube and detectors. The same is used for reconstructing the image in CT technology.
The relationship between transmission intensities for pre and post-patient and the linear attenuation coefficient is as
follows:

I = I0 exponential to the power(-ux)                                                           


Where I0 and I are pre-and post-patient photon intensities, x is penetrating path distance along with the patient, and u
is linear attenuation coefficient. A reference linear attenuation coefficient of water is set to 1, and a relative number is
generated depending on the density of the material in consideration. This relative number is called as CT number or
Hounsfield Unit (HU) and is calculated with the formula below:

HU = [(umaterial - uwater)/uwater]  × 1000             

If we place the linear attenuation coefficient of water as µwater in the above equation,

HU = [(uwater - uwater)/uwater]  × 1000

therefore, the HU of water is 0.

The attenuation coefficient of vacuum is 0, and since air has relatively very small attenuation, it is also approximated
to 0. Therefore, equation 2 for air is:

HU = [(0-uwater)/uwater] × 1000 = -1000                

In most CT systems, the CT numbers vary from −1000HU to +1000HU for different tissues. Since the lung is mostly
filled with air, its HU value varies between -400HU to -900HU, the soft tissues have around -80HU to 80 HU, and the
softer and spongy bones (trabecular bones) have HU values between 100 to 300, whereas harder or cortical bones
have it in a range of 300HU to 1000HU. The metallic implants have an HU value of around 1000, whereas the HU
value for iodinated CT contrast is between 100 and 600.

The Physics of Hybrid Imaging

In many clinical conditions, functional imaging modalities such as SPECT and PET provide more information than
anatomical imaging modalities such as CT and MRI. But these modalities often lack detailed anatomical information.
In the 1980s, one of the solutions to this problem was to acquire images separately and merge them using computers.
However, these image sets often get misregistered and lead to errors in interpretation.

In the early 1990s, Hasegawa led a group which used a diagnostic CT scanner and conventional gamma camera with a
common patient table for sequential imaging of SPECT and CT. But the first commercial SPECT-CT was introduced
by General Electric (The GE Discovery VG Hawkey) in 1999, and now the SPECT-CT is available with different
multi-detector CT technology such as 2-, 4-, 6-, 16-, and 64-slice systems.[23] 

Similarly, combining PET with CT was also suggested in 1991, but the first working prototype PET/CT scanner was
developed in 1998. The first commercial PET-CT was available in 2001 with a single-slice CT system, but now the
PET-CTs are available with 128-slice of diagnostic CTs.[24]

SPECT-CT

CT imaging complements the functional information of SPECT in many ways, such as anatomically localization of
radiopharmaceutical distribution, attenuation and scatter correction, and determination of the impact of partial volume
effect because of the low spatial resolution of the SPECT system and converting quantitative SPECT images into
standardized uptake values (SUVs). The SPECT is a rotating gamma camera that determines the radiotracer
distribution and quantifies the same in 3-dimensional  images.

The gamma camera technology has not been changed much since its inception by Hall Anger in 1958. In 1963, Kuhl
and Edwards demonstrated the first SPECT design, and since then, many configurations have been tried, including
rotating patient chairs, translating detectors, etc.[23] But the SPECT using a rotating gamma camera was introduced
in the late 1980s (between 1977-1980).
The gamma camera, also called an Anger camera or a scintillation camera, consists of the detector (s), photomultiplier
tubes (PMT), a pulse-height analyzer, an anti-coincidence circuit including amplifiers, a positioning circuit, and a
computer system (see "The Gamma Camera"). The detector is covered with a collimator and coupled with an array of
photomultiplier tubes for positional information. The collimators are made up of lead or tungsten and come in
different configurations such as low-energy high resolution, low-energy all-purpose, high energy, and pin-hole
collimators for different users' needs.

Only photons moving parallel to the collimator holes reach the crystal and absorb it. The incident radiation is
converted into light photons by scintillator detectors, which are further converted into electrical signals by PMTs to be
ready by the subsequent electronics. The commonly used detector configuration is a dual detector system; however,
single or triple detector systems and organ-specific cameras are also available. The gamma camera allows static
planar imaging, whole-body scan with continuous slow-moving patient table, and the SPECT with detectors heads at
various angles such as 90 degrees, 120 degrees, and 180 degrees.

The spatial resolution of a gamma camera is limited by many factors, including the detector thickness, collimator
construction, the patient to collimator distance, the energy of the radioisotope, and the size and density of the organ or
the body being imaged. However, the sensitivity is limited mainly by collimators and keeps it around 100 to 200
counts/second/Mega Bequerrel (MBq) or 0.01 to 0.02 percent for gamma-energy of 140 keV. The energy resolution of
NaI (Tl) detectors has around 10 percent in the range of 0.1 to 1.0 MeV photons.

Photon Scatter

Approximately 20 to 50 percent of all detected events by the Anger camera are scattered within the body, and the
directional changes allow for mispositioning. This degrades the image quality and increases the background with a
decrease in contrast. This is one of the main reasons scatter correction techniques are required.

Photon Attenuation

The gamma-photons are attenuated within the body, on the collimators, and within the detector assembly. The
attenuation is dependent on photon energy and the electron density, i.e., the Z number of the attenuating material,
whereas the attenuation coefficient is the measurement of probability where a unit length of material attenuates the
photon.

In CT, the X-ray radiation beam is well collimated to fall on the detector and called narrow-beam collimation. In
Nuclear Medicine, the source is within the body and emitted in all directions and falls on the detector, which is only a
part of total emissions as the detector is in a particular direction. This uncollimated condition is called broad-beam
collimation and leads to lower effective attenuation. The conditions of broad-beam and narrow-beam are important
while applying attenuation and scatter corrections in SPECT reconstruction data. They possess a large impact on the
reconstructed image.

Corrections for Photon Attenuation and Scattering

The CT number or Hounsfield Unit (HU) from generated CT data is used for attenuation and scatter corrections in
SPECT-CT. Generally, both the corrections are performed in separate steps. The energy of X-ray photons and the
photons emitted by the radioisotopes has a different range of energies. The energy of gamma-photons used in imaging
has higher than X-ray photons.

The conversion of attenuation coefficient from CT number to the gamma-energy is a bilinear relation but behaves
linearly at a threshold value of 0. The bilinear scaling implementation requires preprocessing of CT data and the
creation of two distinct data sets separated from the threshold value where the relationship changes. This can be easily
performed by separating CT numbers at a threshold number and applying regression equations to each image set.
Then recombining both data sets finally corrects the data for attenuation.
The scatter correction can also be performed to improve the reconstruction accuracy. The techniques which have been
validated for SPECT scatter correction are the "Triple-energy-window (TEW) method, the transmission-dependent
scatter correction (TDSC) and using the calculations of physics of scattering. The TDSC and the scattering physics
calculations both use CT data to improve accuracy.[25][26][27][28]

PET-CT

The PET-CT scanners use positron-emitting radionuclides and are configured in sequential gantries or in-line gantries
for PET and CT machines and share the same patient couch. The Positron-emitting radionuclides, also called PET
radionuclides, are neutron-deficient and achieve stability by converting a proton into a neutron (n), a positron (a
positive electron), and an electron neutrino. The positron loses its kinetic energy and recombines with one of the
orbital electrons in surrounding tissues.

The rest mass-energy of the positron-electron pair (a transient atom called positronium) is converted into two 511 keV
annihilation photons, emitted in approximately opposite directions (approximately 180 degrees). This pair of gamma
photons are detected by the block of detectors placed in a circular ring. Various scintillation crystals such as bismuth
germanate (BGO), lutetium oxyorthosilicate (LSO), and gadolinium oxyorthosilicate (GSO) are used as PET
detectors.

A coincidence circuit is used to record the valid events within the coincidence timing window. Typically, the
coincidence timing window is 6-12 nanoseconds. The event detection is stored in a sinogram matrix to represent the
activity distribution at a particular angle and axial position. An algorithm is used to reconstruct the sinogram data and
represent the activity distribution inside the patient body. Therefore, the radiopharmaceutical distribution, which
follows certain physiological pathways and localizes within the body, can be mapped and measured.

For medical applications, commonly used positron-emitting radionuclides are Fluorine-18, Gallium-68, Oxygen-15,
Nitrogen-13, Carbon-11, and Rubidium-82. The energy range of these radioisotopes is 0.634 MeV for F-18 to 3.4
MeV for Rb-82. The physical properties of these radionuclides are summarized in table 1.

Table 1: Physical properties of commonly used PET radionuclides[29]

 *EC = Electron Capture

Sensitivity

The PET sensitivity is defined as the number of 511 keV photon pairs detected in unit time by the device from each
unit of radioactivity present in the source.[30] It is measured in counts per second per kilo becquerel or cps/kBq.
Since PET is a 3-dimensional imaging modality and uses electronic collimation of coincidence detection instead of
absorptive collimators used in SPECT, the sensitivity increases significantly.

The sensitivity also increases by adding more detector material in thickness and axial extent. With a 50 percent
increase in the thickness of LSO crystal (from 2 cm to 3 cm), the intrinsic sensitivity increases by 40
percent. Increasing a 30 percent axial extent increases the volume sensitivity by 78 percent. However, increasing axial
thickness increases the number of PMTs and thereby the cost of the scanner.

The 5-Ring GE discovery IQ Gen2 offers PET axial coverage of 26 cm, whereas the newly introduced Siemens
Biograph Vision Quadra offers 106 cm of the axial field of view (3.2 mm crystal element with 51 mm volumetric
resolution) and allows total body (vertex to thigh) imaging in one position. However, commonly used PET scanners
have 2 to 3 ring detectors, and a typical 3-ring GE Discovery IQ has a sensitivity of 7.78 cps/kBq.

Spatial Resolution

The positron physics, i.e., emissions and annihilation reactions, limits spatial, temporal, and contrast resolutions.
Ideally, starting from the positron range, our equipment should map the origin of positron decay, whereas, in PET
imaging, the radiopharmaceutical distribution map is generated for annihilated photons. This sets the lower limit in
spatial resolution of the PET camera for any radionuclide used in imaging.

The second limitation is that both annihilation photons are assumed to be collinear; however, they come out at an
angle of (180±0.25) degrees. For a typical PET scanner design, the non-colinearity can contribute to about 1.5 mm
full width at half maximum (FWHM) for F-18.

Along with these two limitations, the PET resolution is further degraded by PET detectors' material, size, and design.
This effect can contribute to about 2 mm resolution degradation. With detector pixel size in the range of 3.2 to 5 mm
and 15 to 25 mm in thickness, the PET can provide spatial resolution in the range of 3.5 to 5 mm at the machine level.

Scatter and Randoms

If photons of different origins are registered as related events due to scattering from body tissues, these unrelated,
erroneously counted events are called scatter events. The number of such events caused by scattering is called scatter
fraction. This is the ratio of scattered events to total events. Random events are those events that are unrelated to each
other but are counted as valid events due to coincidence timing window width.

These events are reduced or eliminated by modeling both processes. Scattered photons can be eliminated by
identifying the lost energy in the scattering process and applying simple energy threshold rejection criteria. However,
the random events can be counted from the singles rate, the coincidence timing window, or by direct counting of
delayed events acquired out of the timing window. The current detector systems typically lead to 30 to 40 percent
scattering fractions at the system level.[30]

The Attenuation and Attenuation Correction

On interaction with body tissues, the photons lose their energy by Compton scattering or the Photo-electric effect, and
many of them are eliminated due to significant loss of energy. This loss in the counting of annihilation photons is
called attenuation. An algorithm generated from CT transmission data compensates for such losses. However, the
mean X-ray energy used in a clinical set-up is in the range of 70 keV, which is far less than the annihilation photons'
energy of 511 keV and must be scaled to such a level.

Since the body has various tissues with different CT numbers, more than one scaling factor is applied to compensate
for attenuation corrections. Generally, the CT scan is performed before the PET data acquisition, though the order can
be changed, and an attenuation correction map is generated for the entire volume. The attenuation correction factors
(ACFs) are formed by reprojecting the scaled CT images and then interpolating from CT to PET spatial resolution.
However, serious mismatches can be generated for many reasons, including patient respiration, due to oral and
intravenous contrast and metallic implants.

Issues of Concern
Breathing Artifacts

The involuntary breathing motions cause the majority of PET or SPECT and CT misregistration issues. The CT
acquisition can acquire the images in a single breath-hold and fix the position of organs such as the lung, liver, spleen,
diaphragm, and heart, which move during respiration. But, PET or SPECT imaging (combinedly called emission
scans) takes a longer duration, and respiratory blurrings are bound to occur. This causes misalignment of both images,
particularly near the diaphragm area.

The same is also manifested in the attenuation correction factor map generated by the CT images. Many methods,
including altering breath-holding patterns, mid-inspiration, or end-expiration during CT acquisition, have been tried to
solve this problem. However, the best solution is simply acquiring the CT data without any instruction and allowing
the patient to take shallow breathing with instruction for lying still with no other motion than involuntary.[31]
Contrast Agent Artifacts

CT imaging with intravenous and oral contrast can help with anatomic localization. However, since the contrast is
denser in attenuation, it can falsely alter the attenuation map generated by the CT data and inaccurately represent
gamma photon attenuation in certain areas. This can result in artifacts in the final images.

Many solutions to this problem have been considered, including viewing non-attenuation corrected images and
correlating them with attenuation corrected images. However, this can be cumbersome on many software. Another
solution is acquiring low-dose non-contrast CT and using it attenuation correction. However, both solutions have their
own merits and demerits. A software-based solution has been tried to generate an attenuation correction algorithm that
does not generate artifacts, which is the most appealing solution to these artifacts.[32]

Orthopaedic Metallic Implants and Chemo port

The orthopedic metallic implants and in-situ chemo ports cause photopenic areas in PET and SPECT images. When
using CT data for attenuation correction, the concerned area is overestimated, and artifacts in the final image are
observed. Though it is easy to correlate such areas of the artifact, when clinically relevant lesions are adjacent to such
artifact, it becomes difficult to interpret, especially in cases of infection or loosening.[33] 

Dental implants or fillings can also cause such artifacts and obscure real lesions or create apparent lesions in a patient
with head & neck malignancies with involvement of the oral cavity or tonsils.[34][35]

Some other areas of concern that affect the image quality are patient size, positioning of arms, and calcified lymph
nodes.[32]

Clinical Significance
Applications of SPECT-CT

The SPECT-CT provides 10 percent better quantification accuracy than planar imaging, allowing its use for
individualized radionuclide treatment planning. Such implementations improve patient outcomes.

SPECT-CT improves the sensitivity and specificity in preoperative localization of parathyroid adenomas and
improves the detection rate in the asymptomatic group.[36] It also improves image contrast and quantification,
enabling the precise localization of intra-thoracic goiters.

In thyroid cancer management, accurate staging and risk stratification are essential. SPECT-CT allows more
precise characterization of the etiology, i.e., begins versus malignant.

SPECT-CT has significantly improved the diagnostic accuracy of neuroendocrine tumors. Also, this helps in
optimizing PRRT (Peptide-receptor radionuclide therapy) with patient-specific dosimetry.

In the case of skeletal scintigraphy, the SPECT-CT again increases the specificity and positive predictive value
in oncology patients. It also defines the extent of metastases better by reducing unnecessary diagnostic
procedures.

In orthopedic implants with suspected infection, the SPECT-CT has provided accurate information, which
allows differentiation of bone infection from soft tissue infection. It is also beneficial in diagnosing relapsed
osteomyelitis with post-traumatic structural bone alterations.[37][38]

In coronary artery disease, the SPECT-CT detects coronary artery stenosis and myocardial perfusion
abnormalities. In a selected group of patients, it improves the diagnostic accuracy of detecting CAD and
guidance regarding revascularization procedures. Also, the anatomical information provides advantages of
coronary artery calcium scoring along with perfusion data.[39]
The SPECT-CT provides many additional values in sentinel node imaging for breast cancer, melanoma, oral
cavity, penile, vulvar, prostatic, testicular, renal cell, cervical, endometrial, and lung malignancies.

In the case of lung ventilation-perfusion (VQ) scintigraphy, the authors, in a study, found that the SPECT alone
and SPECT-CT have a sensitivity of 97 percent, whereas the specificity of SPECT increased from 88 percent to
100 percent in the case of SPECT-CT.[40]

Applications of PET-CT

Undoubtedly, PET-CT has changed the management of oncological patients. A seemingly limitless variety of PET
tracers has covered all cancer hallmarks.[41] 18F-FDG (fluorodeoxyglucose) is the established radiotracer for
oncological applications; however, novel tracers such as 68Ga-FAPI (fibro-blast-activation-protein) alone have shown
its uptake in 28 kinds of cancers.[42] The enhanced knowledge of cancer characteristics allows the detection of
different metabolic or pathogenic pathways within cancer cells, resulting in the development of new
radiopharmaceuticals.

Apart from oncological applications, the PET-CT is very useful for non-oncological applications such as in the
assessment of myocardial viability, cognitive impairment/Alzheimer’s disease, infectious/inflammatory diseases,
assessment of sarcoidosis, the diagnosis of osteomyelitis, spondylodiscitis of non-postoperative origin, the
investigation of pyrexia of unknown origin (PUO), postoperative fever, recurrent sepsis, septic embolisms, detection
of synovitis in the shoulder, hip & sternoclavicular joints, assessment of the atheromatous plaque calcification along
with many more.[43]

Enhancing Healthcare Team Outcomes


The nuclear medicine specialty involves using radioisotopes for diagnosis, staging, therapy, and monitoring of
patients' responses. Whether diagnostic or therapeutic, it involves using radiopharmaceuticals, which prompt the best
practice of radiation protection to be applied at every stage to minimize and mitigate the risks of accidental radiation
exposure to staff, patients, their relatives, and the environment. 

The nuclear medicine team is comprised of nuclear medicine clinicians (which can also include mid-level providers
such as NPs or PAs), nuclear medicine physicists, radiological safety officers, nuclear medicine technologists, radio-
pharmacists, nurses, and the other auxiliary staff. The department itself does not generate referrals but receives them
from varied disciplines such as cardiology, oncology, nephrology, urology, endocrinology, pediatrics, neonatology,
surgery, respiratory medicine, orthopedics, psychiatry, etc. Interprofessional communication within the department
and inter-disciplinary coordination from varied specialties are crucial in proper patient care, especially in nuclear
medicine. Any error at any stage can bring significant consequences and thereby compromise patient safety. 

The nuclear medicine physician is a doctor who is extensively trained and licensed to report nuclear medicine scans
and provide therapies to the patients. They are also responsible for determining the nature of the procedure, the
radiopharmaceutical, the quantity to be used, and guiding the technologist to obtain certain image sets apart from
routine imaging.

A nuclear medicine physicist is a person who is expected to understand the complete science involved in imaging the
patients. They are responsible for understanding the complete diagnostic or therapeutic procedures used in imaging,
instrumentation, or other computer analysis. They are expected to recognize any divergence from normal operations
along with the responsibility for radiation dosimetry, safety, teaching, and any research in the department. 

The nuclear medicine technologist is responsible for acquiring image datasets in consultation with nuclear medicine
physicians and nuclear medicine physicists, along with assisting nuclear medicine physicists in other routine works
such as quality assurance of the imaging and other radiation safety and measuring equipment. The radio-pharmacist is
responsible for preparing radiopharmaceuticals, performing quality control, and injecting them in consultation with
nuclear medicine physician, along with the responsibility of maintaining records and safe storage of radioactive
materials. The nurse is responsible for patient preparation and assists nuclear medicine physicians in day-to-day
activities.

Patient safety is paramount in high-quality healthcare and is defined as the reduction of the risk of unnecessary harm
associated with healthcare to a minimum acceptable.[44] A study published in 2020 by Kasalak et al. included 147
patient safety incidents reported in nuclear medicine practice from the period between 2014 to 2019 and concluded
the majority of patient safety incidents in nuclear medicine occur in three main International Classification for Patient
Safety (ICPS) categories. These areas are medication, clinical administration, and clinical process or procedure.
[45] [Level 2] They can be considered key strategic areas for incident prevention and patient safety improvement.

The incidents which can be commonly reported in nuclear medicine are:

Extravasation of radiopharmaceutical or CT contrast agent despite check with a saline flush causing pain and
swollen arm. A rescheduling of study can be an option

Incomplete patient verification by the radiopharmaceutical administration staff resulting in a medical event as
per the United States National Regulatory Commission Regulations (USNRC) 10 Code of Federal Regulation
(CFR) Part 35.3045

Wrong administration of radiopharmaceuticals such as I instead of I-mIBG

Administration of CT contrast media despite reportedly informing history of an allergic reaction

Proper interprofessional communication within the department, attentiveness towards the assigned job, and knowing
the key areas of medical errors can be the key to avoiding such incidents. Knowing the science and the professional
skills is a basic need of any profession. A number of studies suggest that medical errors could have been prevented
under better conditions and better communications, improving outcomes and enhancing team performance.[46]
[47] [Level 1]

Review Questions

Access free multiple choice questions on this topic.

Comment on this article.

References
1. Wang R, Duan X, Shen C, Han D, Ma J, Wu H, Xu X, Qin T, Fan Q, Zhang Z, Shi W, Guo Y. A retrospective
study of SPECT/CT scans using SUV measurement of the normal pelvis with Tc-99m methylene diphosphonate. J
Xray Sci Technol. 2018;26(6):895-908. [PubMed: 30103368]
2. Kaneta T, Ogawa M, Daisaki H, Nawata S, Yoshida K, Inoue T. SUV measurement of normal vertebrae using
SPECT/CT with Tc-99m methylene diphosphonate. Am J Nucl Med Mol Imaging. 2016;6(5):262-268. [PMC free
article: PMC5069278] [PubMed: 27766184]
3. Kinahan PE, Fletcher JW. Positron emission tomography-computed tomography standardized uptake values in
clinical practice and assessing response to therapy. Semin Ultrasound CT MR. 2010 Dec;31(6):496-505. [PMC
free article: PMC3026294] [PubMed: 21147377]
4. Townsend DW. Positron emission tomography/computed tomography. Semin Nucl Med. 2008 May;38(3):152-66.
[PubMed: 18396176]
5. Townsend DW. Dual-modality imaging: combining anatomy and function. J Nucl Med. 2008 Jun;49(6):938-55.
[PubMed: 18483101]
6. Bocher M, Balan A, Krausz Y, Shrem Y, Lonn A, Wilk M, Chisin R. Gamma camera-mounted anatomical X-ray
tomography: technology, system characteristics and first images. Eur J Nucl Med. 2000 Jun;27(6):619-27.
[PubMed: 10901447]
7. Patton JA, Delbeke D, Sandler MP. Image fusion using an integrated, dual-head coincidence camera with X-ray
tube-based attenuation maps. J Nucl Med. 2000 Aug;41(8):1364-8. [PubMed: 10945529]
8. Townsend DW, Beyer T, Blodgett TM. PET/CT scanners: a hardware approach to image fusion. Semin Nucl Med.
2003 Jul;33(3):193-204. [PubMed: 12931321]
9. Townsend DW, Carney JP, Yap JT, Hall NC. PET/CT today and tomorrow. J Nucl Med. 2004 Jan;45 Suppl 1:4S-
14S. [PubMed: 14736831]
10. Ljungberg M, Pretorius PH. SPECT/CT: an update on technological developments and clinical applications. Br J
Radiol. 2018 Jan;91(1081):20160402. [PMC free article: PMC5966195] [PubMed: 27845567]
11. Bailey DL, Willowson KP. An evidence-based review of quantitative SPECT imaging and potential clinical
applications. J Nucl Med. 2013 Jan;54(1):83-9. [PubMed: 23283563]
12. Bercovich E, Javitt MC. Medical Imaging: From Roentgen to the Digital Revolution, and Beyond. Rambam
Maimonides Med J. 2018 Oct 04;9(4) [PMC free article: PMC6186003] [PubMed: 30309440]
13. Wesolowski JR, Lev MH. CT: history, technology, and clinical aspects. Semin Ultrasound CT MR. 2005
Dec;26(6):376-9. [PubMed: 16392657]
14. Flohr TG, Schaller S, Stierstorfer K, Bruder H, Ohnesorge BM, Schoepf UJ. Multi-detector row CT systems and
image-reconstruction techniques. Radiology. 2005 Jun;235(3):756-73. [PubMed: 15833981]
15. Commandeur F, Goeller M, Dey D. Cardiac CT: Technological Advances in Hardware, Software, and Machine
Learning Applications. Curr Cardiovasc Imaging Rep. 2018 Aug;11 [PMC free article: PMC6814142] [PubMed:
31656551]
16. Goldman LW. Principles of CT and CT technology. J Nucl Med Technol. 2007 Sep;35(3):115-28; quiz 129-30.
[PubMed: 17823453]
17. Kalender WA. Technical foundations of spiral CT. Semin Ultrasound CT MR. 1994 Apr;15(2):81-9. [PubMed:
8198822]
18. Goldman LW. Principles of CT: multislice CT. J Nucl Med Technol. 2008 Jun;36(2):57-68; quiz 75-6. [PubMed:
18483143]
19. Hurlock GS, Higashino H, Mochizuki T. History of cardiac computed tomography: single to 320-detector row
multislice computed tomography. Int J Cardiovasc Imaging. 2009 Apr;25 Suppl 1:31-42. [PubMed: 19145476]
20. Tan Z, Miao Q, Li X, Ren K, Zhao Y, Zhao L, Li X, Liu Y, Chai R, Xu K. The primary study of low-dose
pancreas perfusion by 640- slice helical CT: a whole-organ perfusion. Springerplus. 2015;4:192. [PMC free
article: PMC4409948] [PubMed: 25932375]
21. Behling R. X-ray sources: 125 years of developments of this intriguing technology. Phys Med. 2020
Nov;79:162-187. [PubMed: 32826172]
22. Jain A, Panse A, Bednarek DR, Rudin S. Focal spot measurements using a digital flat panel detector. Proc SPIE
Int Soc Opt Eng. 2014 Mar 19;9033(90335F) [PMC free article: PMC4188439] [PubMed: 25302004]
23. Hutton BF. The origins of SPECT and SPECT/CT. Eur J Nucl Med Mol Imaging. 2014 May;41 Suppl 1:S3-16.
[PubMed: 24218098]
24. Townsend DW, Beyer T. A combined PET/CT scanner: the path to true image fusion. Br J Radiol. 2002 Nov;75
Spec No:S24-30. [PubMed: 12519732]
25. Ichihara T, Ogawa K, Motomura N, Kubo A, Hashimoto S. Compton scatter compensation using the triple-
energy window method for single- and dual-isotope SPECT. J Nucl Med. 1993 Dec;34(12):2216-21. [PubMed:
8254414]
26. Meikle SR, Hutton BF, Bailey DL. A transmission-dependent method for scatter correction in SPECT. J Nucl
Med. 1994 Feb;35(2):360-7. [PubMed: 8295011]
27. Beekman FJ, Kamphuis C, Frey EC. Scatter compensation methods in 3D iterative SPECT reconstruction: a
simulation study. Phys Med Biol. 1997 Aug;42(8):1619-32. [PubMed: 9279910]
28. Kadrmas DJ, Frey EC, Karimi SS, Tsui BM. Fast implementations of reconstruction-based scatter compensation
in fully 3D SPECT image reconstruction. Phys Med Biol. 1998 Apr;43(4):857-73. [PMC free article:
PMC2808130] [PubMed: 9572510]
29. Conti M, Eriksson L. Physics of pure and non-pure positron emitters for PET: a review and a discussion.
EJNMMI Phys. 2016 Dec;3(1):8. [PMC free article: PMC4894854] [PubMed: 27271304]
30. Vandenberghe S, Moskal P, Karp JS. State of the art in total body PET. EJNMMI Phys. 2020 May 25;7(1):35.
[PMC free article: PMC7248164] [PubMed: 32451783]
31. Hicks R, Lau E, Binns D. Hybrid imaging is the future of molecular imaging. Biomed Imaging Interv J. 2007
Jul;3(3):e49. [PMC free article: PMC3097677] [PubMed: 21614291]
32. Blodgett TM, Mehta AS, Mehta AS, Laymon CM, Carney J, Townsend DW. PET/CT artifacts. Clin Imaging.
2011 Jan-Feb;35(1):49-63. [PMC free article: PMC4277262] [PubMed: 21237418]
33. Bujenovic S, Mannting F, Chakrabarti R, Ladnier D. Artifactual 2-deoxy-2-[(18)F]fluoro-D-glucose localization
surrounding metallic objects in a PET/CT scanner using CT-based attenuation correction. Mol Imaging Biol.
2003 Jan-Feb;5(1):20-2. [PubMed: 14499157]
34. Goerres GW, Hany TF, Kamel E, von Schulthess GK, Buck A. Head and neck imaging with PET and PET/CT:
artefacts from dental metallic implants. Eur J Nucl Med Mol Imaging. 2002 Mar;29(3):367-70. [PubMed:
12002712]
35. Kamel EM, Burger C, Buck A, von Schulthess GK, Goerres GW. Impact of metallic dental implants on CT-
based attenuation correction in a combined PET/CT scanner. Eur Radiol. 2003 Apr;13(4):724-8. [PubMed:
12664109]
36. Komek H, Yilmaz EE, Cakabay B, Altindag S, Ozdemir N, Can C, Kepenek F. Contrast Enhanced [99mTc]
MIBI SPECT/CT in Primary Hyperparathyroidism. Ann Ital Chir. 2018;89:379-384. [PubMed: 30049907]
37. Horger M, Eschmann SM, Pfannenberg C, Storek D, Dammann F, Vonthein R, Claussen CD, Bares R. The value
of SPET/CT in chronic osteomyelitis. Eur J Nucl Med Mol Imaging. 2003 Dec;30(12):1665-73. [PubMed:
14523585]
38. Filippi L, Schillaci O. Usefulness of hybrid SPECT/CT in 99mTc-HMPAO-labeled leukocyte scintigraphy for
bone and joint infections. J Nucl Med. 2006 Dec;47(12):1908-13. [PubMed: 17138732]
39. Gaemperli O, Kaufmann PA, Alkadhi H. Cardiac hybrid imaging. Eur J Nucl Med Mol Imaging. 2014 May;41
Suppl 1:S91-103. [PubMed: 24658682]
40. Gutte H, Mortensen J, Jensen CV, Johnbeck CB, von der Recke P, Petersen CL, Kjaergaard J, Kristoffersen US,
Kjaer A. Detection of pulmonary embolism with combined ventilation-perfusion SPECT and low-dose CT:
head-to-head comparison with multidetector CT angiography. J Nucl Med. 2009 Dec;50(12):1987-92. [PubMed:
19910421]
41. Lopci E, Fanti S. Non-FDG PET/CT. Recent Results Cancer Res. 2020;216:669-718. [PubMed: 32594402]
42. Kratochwil C, Flechsig P, Lindner T, Abderrahim L, Altmann A, Mier W, Adeberg S, Rathke H, Röhrich M,
Winter H, Plinkert PK, Marme F, Lang M, Kauczor HU, Jäger D, Debus J, Haberkorn U, Giesel FL. 68Ga-FAPI
PET/CT: Tracer Uptake in 28 Different Kinds of Cancer. J Nucl Med. 2019 Jun;60(6):801-805. [PMC free
article: PMC6581228] [PubMed: 30954939]
43. Jiménez-Bonilla JF, Banzo I. Non-oncological applications of PET/CT. Med Clin (Barc). 2018 May
23;150(10):387-389. [PubMed: 29196040]
44. Runciman W, Hibbert P, Thomson R, Van Der Schaaf T, Sherman H, Lewalle P. Towards an International
Classification for Patient Safety: key concepts and terms. Int J Qual Health Care. 2009 Feb;21(1):18-26. [PMC
free article: PMC2638755] [PubMed: 19147597]
45. Kasalak Ö, Yakar D, Dierckx RAJO, Kwee TC. Patient safety in nuclear medicine: identification of key strategic
areas for vigilance and improvement. Nucl Med Commun. 2020 Nov;41(11):1111-1116. [PMC free article:
PMC7556244] [PubMed: 32769813]
46. Schreiber M, Klingelhöfer D, Groneberg DA, Brüggmann D. Patient safety: the landscape of the global research
output and gender distribution. BMJ Open. 2016 Feb 12;6(2):e008322. [PMC free article: PMC4762108]
[PubMed: 26873042]
47. Institute of Medicine (US) Committee on Quality of Health Care in America. To Err is Human: Building a Safer
Health System. Kohn LT, Corrigan JM, Donaldson MS, editors. National Academies Press (US); Washington (DC):
2000. [PubMed: 25077248]
Figures

Narrow beam with translate-rotate motion. Contributed by Dibya Prakash, MSc. (Physics, Bioinformatics),
DMRIT, RSO (NM)
Multiple-narrow beams with translate-rotate motion. Contributed by Dibya Prakash, MSc. (Physics,
Bioinformatics), DMRIT, RSO (NM)
Fan-beam, rotate only. Contributed by Dibya Prakash, MSc. (Physics, Bioinformatics), DMRIT, RSO (NM)
Fan-beam with stationary circular detector. Contributed by Dibya Prakash, MSc. (Physics, Bioinformatics),
DMRIT, RSO (NM)
Slip-ring technology. Contributed by Dibya Prakash, MSc. (Physics, Bioinformatics), DMRIT, RSO (NM)
Tables

Radionuclide Decay Mode (in Mean Positron Range


Half-life Positron Energy in
percent) in water (in mm)
keV
(minutes)

F-18 109.8 634 0.6


Positron (96.9)

EC (3.1)

Ga-68 67.8 1.899 3.5


Positron (88.9)

EC (11.1)

C-11 20.4 960 1.2


Positron (99.8)

EC (0.2)

N-13 10.0 Positron (100) 1199 1.8

O-15 2.0 1732 3.0


Positron (99.9)

EC (0.1)

Rb-82 1.3 3378 7.1


Positron (81.8)

EC (13.1)

Copyright © 2022, StatPearls Publishing LLC.


This book is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/),
which permits use, duplication, adaptation, distribution, and reproduction in any medium or format, as long as you give appropriate credit to the original
author(s) and the source, a link is provided to the Creative Commons license, and any changes made are indicated.

Bookshelf ID: NBK582124 PMID: 35881729

You might also like