If you have problems
I you have problems using the unit, fllow these
‘general procedures to track down the cause from
the symptoms you are experiencing [fan ertor
‘message appears on the display, refer to the
specific error inthis cher.
IE the problem isnot resolved using the table
‘below, Contact technical support at
1-800-926-0505
fyouare
having this
“This might be te cause
INopower | Unieno plugged in; no power atthe
indleaion | outlets ineortet power (make sure
‘tet matches abel on back of
‘mi defective power cord
‘efctve power switch or fuse (3.
cal Technical Suppo,
Noteating | Speint or backup at set propery;
efctve heater or conrller (all
Technical Sopp)
Always Detective conller cll Techical
beatin: Seppo)
empeaure
preter than
Hverysiow | Empry tank or exemely low lguid
esting of | level sn tank (add liquide the tank
samples to improve heating)
Uni heaing up
slower than
ora
Defeatve contol call Techazal
Seppo)
Alar light on_| Setpoint or bake not et properly
eonstartly detective primary controler (al
‘Technical Suppor)
Drapiny not [Temperate F or C stn eo,
eonest temperate nt calibrated (perform
calibration procedure with
"hemeometr; defective eontller
(cal Technic! Suppor),
ate [Broken iveraal clesrcal
operation connection defective heater or
‘contol call Technical Suppor)
ror message:
er
temperature sensor or
(all Techicl Suppo,
ror mesg:
2
Fale of console (call Techaial
Seppo)
Dispay shows
power-up.
‘Wrong volage, Comtller nts
‘up propery eal Techni! Sop=
poe).
Specifications
These are nominal specifications. Fisher
reserves the right to ehange specifications or
designs at anytime without incuing
obligation.
Replacement Parts
WARNING: To avoid clectrical shock, always
disconnect from power supply before
‘maintenance and servicing. Refer servicing to
‘ualifed personae!
Powerreguinments 120 VAC 07230 VAC
10%
S00 60 He
Trem Par No
Gover 2 ter Phase Lid | SPNIOD308
susinles Steel Lid _| 15-462
Ambient conditions 4°Cw 135°C,
‘upto 73% reative humidity
Tenperatue range Cover open: Ambien 1 60
“C Cover closed: Ambient
wore
[Gover 2 er sallow and Sie
Plastic id | sPNvo2s0s
Stainless Steel Lid | 15-462-33
cover 10 liter Plastic Lid | SPNi02306
Stainles Steel Lid _| 15-462
[Gover 20and 28 ter Plas Lid | SPNLODS07
CContat 40,1°C between ambient Skinless Sel Lid | 15-462-35
and 100°C Hinge, Tor2 Ted) SPNTODIRS
Uniformity so2voasrc Ming. 2 ies) SPNIO2384
Hinge, id (ioral except? Her) SPNIO2386
Stability Los'c asc [Ming, rin 28L_and Dual Models) [SPNLO02347
Mex Alsiude 2000 Hinge, iw for 2S, 5, 10 & 20 liter) | SPNLO2«59
[lip thermomeer (or 21. shallow) [SPNIOD423
over Volage EC 664) [cin thermometer (al excep 2.8) | SP0D426
category [Microbicide, ¥ o7 Bote 1.641.337
‘Potltion Degree 2 (BC 664) [Pump Siphon SPNUOZ391
[Contoiee PC bord 200) SPNIO2AI9
CE Products the eevant EC harmonized [Cgatolee PC Boasd Q30V) SPNLOHO6
standards for safety TECIONO-ENSIOIO and [pepo a
45, 4-Tand ENVS0I40)
[contol safey
“Applies to 2807 units only
Sensor, RTD SPNUID3SI
Dithser Tray. 2 er
[ister Tray. 2L sallow or Wer
[Ditfcer Tray, 10 ter SPNIO2SS«
[Ditfser Tay, 20 of 25h SPNIO2S55
Fuse, Type 3AG 1.25425, 1DAmp | SPN4s920
Schemat
120 VAC Unit
Note: The Stainless stel water
‘bath cover was designed as an
alternate tothe soe through
Plastic (polycarbonate) cover.
[No exateral is impervious to
attack by all chemicals. The
Stainless lids are made of a
‘good quality 304 stainless sce}
‘whied is dhe same asthe tank
Clean with mile soap and wa
ter, Rinse with clear tap water.
‘Schematic
230 VAC Unit
= re
art No. 102370 Rev. K
Published 01/08
Printed in USA.