Philips Healthcare

Digital Detector
Pixium 4343RC
Level 2 Documentation
Repair Manual

This document and the information contained in it is strictly reserved for current Philips Healthcare ("Philips" personnel, Philips licensed repre‐
sentatives and Philips customers who have purchased a valid service agreement for use by the customer's designated in-house service em‐
ployee on equipment located at the customer's designated site. Use of this document by unauthorized persons is strictly prohibited. This docu‐
ment must be returned to Philips when the user is no longer licensed and in any event upon Philips' first written request.

ATTENTION: This page contains copyrighted materials that are confidential and/or proprietary. Any release or distribution of this material,
without permission, is a violation of law.
© 2019 Koninklijke Philips N.V.
ALL RIGHTS RESERVED

451298439931aa
CSIP statement

Proprietary Notice:

This document and the information contained in it is proprietary and confidential information of Philips Healthcare
("Philips") and may not be reproduced, copied in whole or in part, adapted, modified, disclosed to others, or disse-
minated without the prior written permission of the Philips Legal Department. Use of this document and the infor-
mation contained in it is strictly reserved for current Philips personnel and Philips customers who have a current
and valid license from Philips for use by the customer’s designated in-house service employee on equipment lo-
cated at the customer’s designated site. Use of this document by unauthorized persons is strictly prohibited. Re-
port violation of these requirements to the Philips Legal Department. This document must be returned to Philips
when the user is no longer licensed and in any event upon Philips’ first written request.

Warranty Disclaimer Language:

Philips provides this DOCUMENT without warranty of any kind, implied or expressed, including, but not limited to,
the implied warranties of merchantability and fitness for a particular purpose.

Limitations of Liability Language:

Philips has taken care to ensure the accuracy of this document. However, Philips assumes no liability for errors or
omissions and reserves the right to make changes without further notice to any products herein to improve reliabil-
ity, function, or design. Philips may make improvements or changes in the product(s) or program(s) described in
this document at any time.

2 CSIP Level 2 451298439931aa


ATTENTION: This page contains copyrighted materials that are confidential and/or proprietary.
Any release or distribution of this material, without permission, is a violation of law.
© 2019 Koninklijke Philips N.V.
ALL RIGHTS RESERVED
Replacement Fault Finding Procedures

marker

6 Replacement
Code No. 9897 010 0075x

Items Supplied
– Transport box
– Sealing bag
– Humidity protection bag (containing humidity protector and humidity indicator)
– Empty CD
– Detector exchange report

6.1 Procedure Digital Detector Pixium 4343RC


This replacement procedure is valid for BuckyDiagnost VS, DigitalDiagnost VM, and DigitalDiagnost
TH.

451298439931aa CSIP Level 2 37


ATTENTION: This page contains copyrighted materials that are confidential and/or proprietary.
Any release or distribution of this material, without permission, is a violation of law.
© 2019 Koninklijke Philips N.V.
ALL RIGHTS RESERVED
Procedure Digital Detector Pixium 4343RC Replacement

Unpacking the New Detector

WARNING
Hazard of contamination caused by a broken detector panel
If a collision causes the detector panel to break, CsI particles contaminated with Thallium may be
emitted to the environment.
In this case, follow the instructions given below:
• Put a dust mask on your face.
• Wear plastic gloves.
• Collect all dropped particles and put them, together with the detector, in a sealed package.
• Clean all surfaces that may be contaminated with the particles.
• Place the used dusters in the same sealed package.
• Clearly identify the package with the following sentence “WARNING: DAMAGED DETECTOR”.
• Send back the package containing the damaged detector, so that the supplier can take appropri-
ate actions to properly recycle the detector.
If you do not obey these instructions, there is a risk of death or serious injury.

• Open the transport box.

Fig. 42

• Remove the sealing bag with the detector inside


carefully from the transport box.
• Position the detector on a suitable surface.

Fig. 43

38 CSIP Level 2 451298439931aa


ATTENTION: This page contains copyrighted materials that are confidential and/or proprietary.
Any release or distribution of this material, without permission, is a violation of law.
© 2019 Koninklijke Philips N.V.
ALL RIGHTS RESERVED
Replacement Procedure Digital Detector Pixium 4343RC

NOTICE
To avoid condensation:
Do not open the inner plastic bag before the detector temperature has reached room temperature.
The acclimatization time takes at least 7 hours 30 minutes.
If you do not obey these instructions, there is a risk of property damage.

• Open the sealing bag carefully.


• Remove the detector completely from the seal-
ing bag.
• Make sure that the detector is not damaged.
• Put the detector onto a suitable surface until it is
needed for the installation in the Bucky unit.
• If the detector is damaged, refer to the section
Sending back the old/defective detector stated
below.

Fig. 44

Securing the System


For DigitalDiagnost VM

WARNING
Hazard of moving counterweight carriage
• Before you replace the digital detector, make
sure that the counterweight carriage is se-
cured with the two "red headed" screws (1)!
If you do not obey these instructions, there
1
is a risk of death or serious injury.

Fig. 45

451298439931aa CSIP Level 2 39


ATTENTION: This page contains copyrighted materials that are confidential and/or proprietary.
Any release or distribution of this material, without permission, is a violation of law.
© 2019 Koninklijke Philips N.V.
ALL RIGHTS RESERVED
Procedure Digital Detector Pixium 4343RC Replacement

For BuckyDiagnost VS

WARNING
Hazard of moving counterweight carriage
• Before replacing the digital detector, make
sure that the counterweight carriage is se-
cured with the four safety screws (1)!
If you do not obey these instructions, there
is a risk of death or serious injury.

• Switch on the system.


1 1
• Tilt the VS to the 90° position (horizontally).
• Move the counterweight carriage into its parking
position and fixate it with the four safety screws
(1), two on each side of the column.
icn-13-v-712-16079-130000-02-A.iso

Fig. 46

Removing the Covers

WARNING
Hazard of electrical shock through running system!
• Do not touch live components while the system power is not switched off.
If you do not obey these instructions, there is a risk of death or serious injury.

• Switch off the system.


• Make sure that the system cannot be switched on accidentally.
• Wait until the capacitors are discharged.
• Remove the top cover of the X-ray generator.
• Switch off the X-ray generator at ENF1.

• Loosen the two Allen screws (1).


• Remove the two Allen screws (2) if available. 2
• Remove the cover (3). 1

Fig. 47

40 CSIP Level 2 451298439931aa


ATTENTION: This page contains copyrighted materials that are confidential and/or proprietary.
Any release or distribution of this material, without permission, is a violation of law.
© 2019 Koninklijke Philips N.V.
ALL RIGHTS RESERVED
Replacement Procedure Digital Detector Pixium 4343RC

• Loosen the two nuts on each side of the two


front cover supports (1).
• Remove the two front cover supports.

Fig. 48

For DigitalDiagnost VM only


• Remove the handgrip (1) to get access to the
rear cover.

Fig. 49

For BuckyDiagnost VS Digital / DigitalDiagnost VM


• Remove the three screws (1).
• Remove the rear cover (2).

• Disconnect the earth wire VA X12 (3).


• Remove the grid.

Fig. 50

For DigitalDiagnost TH (BU3) 2


• Remove the rear cover (1) of the Bucky unit 3
digital (two screws).

• Disconnect the earth wire SA X12 (2).

Fig. 51

Removing the Amplimat Chamber


• Disconnect the cable VA X8.

451298439931aa CSIP Level 2 41


ATTENTION: This page contains copyrighted materials that are confidential and/or proprietary.
Any release or distribution of this material, without permission, is a violation of law.
© 2019 Koninklijke Philips N.V.
ALL RIGHTS RESERVED
Procedure Digital Detector Pixium 4343RC Replacement

• Remove the upper screw from the Amplimat


stop.
• Move the angle plate to the right side.

Fig. 52

• To get access to the connector VA PH X1, pull


out the 5-field Amplimat chamber (approximately
10 cm).

Fig. 53

• Remove the metal clamps (2, 3).


• Disconnect VA PH X1 (1).
• Remove the 5-field measuring chamber and put 1

it down securely.

2 3

Fig. 54

42 CSIP Level 2 451298439931aa


ATTENTION: This page contains copyrighted materials that are confidential and/or proprietary.
Any release or distribution of this material, without permission, is a violation of law.
© 2019 Koninklijke Philips N.V.
ALL RIGHTS RESERVED
Replacement Procedure Digital Detector Pixium 4343RC

• To protect the cables on the right side, pull out


the plastic strip so that all cables on the right
side are covered.

Fig. 55

• Disconnect the protective earth conductor


VA X11 from the detector plate.

Fig. 56

Removing the Old Detector

Important
It is not allowed to dispose of the detector. The old or defective detector must be sent back to the
company.

• Disconnect the following cables:


– VA X1
– VA X2
– VA X11
– VA X15

Fig. 57

451298439931aa CSIP Level 2 43


ATTENTION: This page contains copyrighted materials that are confidential and/or proprietary.
Any release or distribution of this material, without permission, is a violation of law.
© 2019 Koninklijke Philips N.V.
ALL RIGHTS RESERVED
Procedure Digital Detector Pixium 4343RC Replacement

• To release the detector plate inside the Bucky


unit 3, Pull out the black bolts on the left hand
side and on the right hand side and then turn
them for 90°.
• Pull out the metal plate carefully. It is the base
plate for the detector.

Fig. 58

• Pull out the detector (approximately 10 cm).


• Disconnect the following cables from the detec-
tor:
– VA-PX (POF TX-grey)
– VA-PX (earth)

Fig. 59

• Remove the detector carefully.


• Position the detector with the smooth side (with-
out screw holes) onto a suitable surface (for ex-
ample, a table covered with a towel). The screw
holes face towards the upper side.

Fig. 60

44 CSIP Level 2 451298439931aa


ATTENTION: This page contains copyrighted materials that are confidential and/or proprietary.
Any release or distribution of this material, without permission, is a violation of law.
© 2019 Koninklijke Philips N.V.
ALL RIGHTS RESERVED
Replacement Procedure Digital Detector Pixium 4343RC

Installing the New Detector


• Position the metal plate (1) at the bottom side of
the detector.
• Tighten the four screws (2, Allen key 4 mm).

Fig. 61

• Connect the protective earth conductor VA X15


of the detector plate to the detector (washer, ring
terminal, spring washer, nut).

Fig. 62

• Connect the power supply cable.

Fig. 63

451298439931aa CSIP Level 2 45


ATTENTION: This page contains copyrighted materials that are confidential and/or proprietary.
Any release or distribution of this material, without permission, is a violation of law.
© 2019 Koninklijke Philips N.V.
ALL RIGHTS RESERVED
Procedure Digital Detector Pixium 4343RC Replacement

Position the ring terminals of the protective earth


conductor horizontally (1) on the earth bolt on
the left side of the Bucky unit 3 digital. The ca-
bles face inwards to protect them against dam-
age when inserting the digital detector.

Fig. 64

NOTICE
• Be careful not to damage the protective earth conductors on the left side of the Bucky unit 3 digi-
tal while inserting the digital detector.
If you do not obey these instructions, there is a risk of property damage.

• Carefully insert the detector plate with the detec-


tor facing upwards into the Bucky unit 3 digital.
• Push the detector (1) carefully into the Bucky
unit 3 digital until 2/3 of the detector is inside.

Fig. 65

• Connect the cables at the detector:

– Earth wire VA P X11 (connection


between Bucky unit and detector plate, left
connector, not in the figure).
– Communication cable (Ethernet cable) VA P
X2 (1) to X (eth).
– Plastic optical fiber VA P B1 (2) to X
(TXgrey).
• Attach the POF cable with a cable tie to the
Ethernet cable (10 cm between plug and cable
for the Ethernet cable, 13 cm between plug and
cable for the POF cable). That way, a correct
strain relief is guaranteed. Fig. 66

46 CSIP Level 2 451298439931aa


ATTENTION: This page contains copyrighted materials that are confidential and/or proprietary.
Any release or distribution of this material, without permission, is a violation of law.
© 2019 Koninklijke Philips N.V.
ALL RIGHTS RESERVED
Replacement Procedure Digital Detector Pixium 4343RC

Installing the Amplimat Chamber

NOTICE
• Make sure that the wires are located under the housing of the 5–field measuring chamber.
If you do not obey these instructions, there is a risk of property damage.

• Insert the 5-field measuring chamber carefully.


• Connect VA PH X1 (1).
• Install the metal clamps (2, 3). 1

• Push the 5-field measuring chamber fully into


the Bucky unit 3 digital.

2 3

Fig. 67

• Reinstall the Amplimat stop.


• Attach the Amplimat cable in the plastic clamp
on the detector plate.

Fig. 68

• Push the detector completely into the Bucky unit


3 digital until the two levers snap in. It might be
possible to pull and turn the levers to let them
snap.

Fig. 69

Installing the Covers

NOTICE
• Make sure that the cables behind the rear cover do not obstruct the movement of the grid frame.
If you do not obey these instructions, there is a risk of property damage.

451298439931aa CSIP Level 2 47


ATTENTION: This page contains copyrighted materials that are confidential and/or proprietary.
Any release or distribution of this material, without permission, is a violation of law.
© 2019 Koninklijke Philips N.V.
ALL RIGHTS RESERVED
Procedure Digital Detector Pixium 4343RC Replacement

• Push the plastic strip for cable protection into its parking position.
• Connect the cables:
– X2
– X2out
– X5
– X8

For DigitalDiagnost TH (BU3) 1

• Connect the earth wire SA X12 (1).


2
• Install the rear cover (2) of the Bucky unit 3 digi-
tal
(two screws).

Fig. 70

For BuckyDiagnost VS Digital / 1


DigitalDiagnost VM

• Connect the earth wire VA X12 (1).


• Install the rear cover (2) of the Bucky unit 3 digi- 2
tal
(three screws).

Fig. 71

For DigitalDiagnost VM only


• Reinstall the handgrip (1).

Fig. 72

48 CSIP Level 2 451298439931aa


ATTENTION: This page contains copyrighted materials that are confidential and/or proprietary.
Any release or distribution of this material, without permission, is a violation of law.
© 2019 Koninklijke Philips N.V.
ALL RIGHTS RESERVED
Replacement Procedure Digital Detector Pixium 4343RC

Repackaging the Old Detector

WARNING
Hazard of contamination caused by a broken detector panel
If a collision causes the detector panel to break, CsI particles contaminated with Thallium may be
emitted to the environment.
In this case, follow the instructions given below:
• Put a dust mask on your face.
• Wear plastic gloves.
• Collect all dropped particles and put them, together with the detector, in a sealed package.
• Clean all surfaces that may be contaminated with the particles.
• Place the used dusters in the same sealed package.
• Clearly identify the package with the following sentence “WARNING: DAMAGED DETECTOR”.
• Send back the package containing the damaged detector, so that the supplier can take appropri-
ate actions to properly recycle the detector.
If you do not obey these instructions, there is a risk of death or serious injury.

• Open the humidity bag containing the humidity protector and the humidity indicator.

451298439931aa CSIP Level 2 49


ATTENTION: This page contains copyrighted materials that are confidential and/or proprietary.
Any release or distribution of this material, without permission, is a violation of law.
© 2019 Koninklijke Philips N.V.
ALL RIGHTS RESERVED
Procedure Digital Detector Pixium 4343RC Replacement

• Insert the detector into the new, unused sealing


bag.
• Insert the humidity protector and the humidity in-
dicator into the sealing bag.
• Push the detector completely into the sealing
bag.

Fig. 73

• Close the sealing bag hermetically with the


hook-and-loop fastener.

Fig. 74

50 CSIP Level 2 451298439931aa


ATTENTION: This page contains copyrighted materials that are confidential and/or proprietary.
Any release or distribution of this material, without permission, is a violation of law.
© 2019 Koninklijke Philips N.V.
ALL RIGHTS RESERVED
Testing Detector Test (Pixium 4343RC)

• Complete the detector exchange report.


• Store the data/pictures on the CD.
• Insert the packed detector, the completed detec-
tor exchange report and the CD into the trans-
port box.

Fig. 75

• Shut the transport box and send back the old de-
tector.

Fig. 76

6.1.1 Adjustment
Refer to the section Setting-to-work, chapter chapter Calibrating the Flat Detector (Pixium
4343RC) on page 28.

6.1.2 Testing
Refer to the section Testing, chapters chapter Detector Test (Pixium 4343RC) on page 51, chapter
Order of Amplimat Fields on page 55, chapter Amplimat Dose Test on page 60 and chapter
Image Quality Test on page 64.
mark

451298439931aa CSIP Level 2 51


ATTENTION: This page contains copyrighted materials that are confidential and/or proprietary.
Any release or distribution of this material, without permission, is a violation of law.
© 2019 Koninklijke Philips N.V.
ALL RIGHTS RESERVED