You are on page 1of 26

Module 4

Hemodialysis Devices

M3 Dialysis Center, Inc


Hemodialysis Nurse Training Course
2014
Hemodialysis Nurse Training Course
M3 Dialysis Center, Inc.
MODULE 4: Hemodialysis Devices

Objectives
After completing this module, the learner will be able to:
1. Identify the purpose and characteristics of dialyzers.
2. Describe the purpose and chemical composition of dialysate.
3. Describe dialysate preparation and the three monitoring functions of the dialysate
delivery subsystem.
4. Describe the extracorporeal blood circuit functions and monitoring systems.

Definition of Terms
Adsorption
Biocompatibility
Cellulose membrane
Clearance
Convection
Dialysate
Dialysate Delivery System
Dialyzer
Diffusion
Extracorporeal Circuit
Hemodialysis Delivery System
Mass Transfer Coefficient
Membrane Surface Area
Modified-cellulose membrane
Molecular Weight Cut Off
Semi-permeable membrane
Synthetic membrane
Ultrafiltration
Ultrafiltration Coefficient

55
Hemodialysis Nurse Training Course
M3 Dialysis Center, Inc.
MODULE 4: Hemodialysis Devices

A dialyzer lets the patient’s blood - vital to use a membrane the


interact with dialysate through a semi- patient can tolerate
permeable membrane. A dialysate is a - can be tested by checking the
blend of treated water and chemicals; it patient’s blood for certain
removes wastes and fluid, and balances proteins and chemicals
electrolytes. A delivery system supplies - adsorption (attract and hold) of
fresh dialysate and removes used dialysate. proteins into the fiber wall is key
The dialyzer, dialysate, and delivery system to its biocompatibility; proteins
replace some tasks that failed kidneys can coat the surface so blood does
no longer do. not touch the “foreign”
Modern high-tech delivery systems membrane; the protein coating
include a blood pump, an ultrafiltration explains why reprocessed
pump, a dialysate conductivity monitor, (cleaned and reused) dialyzers
alarms, and pressure gauges. Better can be more biocompatible than
membranes, safety monitors, and the use of new ones.
computers have made dialysis safer. These
advances allow today’s staff to turn more of Note: Reprocessing dialyzers with bleach
their time to patients. Trained staff who can strip the protein coating off the
knows dialysis principles, equipment, and membrane.
procedures is the most vital monitors of
patient safety.

Dialyzers
Functions and Components
- have a blood and a dialysate
compartment
- Semi-permeable membrane
keeps the two compartments
apart; membrane is housed in a
plastic case, which holds the
dialyzer together and forms
Figure 1: Dialyzer design
pathways for blood and dialysate
to flow in and out
- Synthetic membranes are more
biocompatible than cellulose
Dialyzer Characteristics membranes; are also
 Biocompatibility hydrophobic (water repelling)
- means not harmful to biological  makes them better able
function to adsorb blood proteins
- immune cells in the blood reacts
to materials used to make
dialysis membranes

56
Hemodialysis Nurse Training Course
M3 Dialysis Center, Inc.
MODULE 4: Hemodialysis Devices

 Surface Area molecular weights; smaller


- a key to how well a dialyzer can molecules = lower ones
remove solutes - Dialyzers have molecular weight
- dialyzers with more surface area cutoffs ranging from 3,000 to
 expose more blood to more than 15,000 Da
dialysate  more solutes can be
removed from the blood  Ultrafiltration Coefficients (KUf)
- Total dialyzer surface area can - the amount of fluid that will
range from 0.5 – 2.4 square pass through the membrane in
meters. one hour, at a given pressure
Ultrafiltration (UF) - a way to
 Mass Transfer Coefficient (KoA) remove excess water from a
- is the ability of a solute to pass patient during hemodialysis by
through the pores of a dialyzer applying hydraulic pressure
- (in theory) is the highest possible (pressure applied to the blood or
clearance of a given dialyzer at dialysate compartment forces
infinite blood and dialysate flow water across the membrane)
- The higher the KoA, the more - dialysis machines can vary the
permeable the dialyzer. hydraulic pressure to control the
ultrafiltration rate (UFR) and
 Molecular Weight Cutoff amount of water removed
- determines the largest molecule - High pressure in the blood
that can pass through the compartment more fluid out
membrane of the blood and into the
- measured in daltons (Da) dialysate
- the average weight of a - Transmembrane pressure
molecule, expressed as the sum (TMP)- pressure difference
of the atomic weights of all the across the membrane
atoms in the molecule
- Larger molecules = higher ** blood compartment
pressure minus dialysate
Dialyzer compartment pressure
Surface Area KoA Kuf
Type (mL/h/mmHg)
- provided by the
2
F8 1.8 m 726 7.5 dialyzer’s manufacturer
- helps the staff member
Revaclear 1.4 m2 732 50 predict how much fluid
HF80S 1.8 m2 805 55 will be removed from
the patient during a
F160NR 1.5 m2 1064 50 treatment
F180NR 1.8 m2 1239 62
Table 1: VariationsTable 1: Variations of Dialyzers
of Dialyzers
57
Hemodialysis Nurse Training Course
M3 Dialysis Center, Inc.
MODULE 4: Hemodialysis Devices

Sample Problem:
A dialyzer with a KUf of 10 will Molecule Molecular Weight (Da)
remove 10 ml of fluid per hour for each
Albumin 66,000
mmHg of pressure.
B2 Microglobulin 11,600
Let’s say a dialyzer has a KUf of 10,
and a TMP of 100 mmHg, the patient Creatinine 113
would lose 1,000 ml of fluid per hour of
dialysis (10 x 100 = 1,000). Phosphorus (PO4²) 94.9
Z inc (Zn²+) 65.3
 Clearance (K)
- amount of blood that can be Nitric Oxide (NO3) 62
cleared of a solute in a given Urea 60
period of time
- how well dialyzers remove Calcium (Ca++) 40
solutes from the blood varies
Water (H2O) 18
from one another
- clearance rates for different Table 2: Molecular Weight of Solutes
molecules are given by the  Concentration
manufacturer for certain blood gradient
and dialysate flow rates  Size, weight and
- three main ways to remove charge of the solutes
solutes that affect a dialyzer’s
clearance: diffusion,  Convection
convection, and adsorption o as fluid crosses a semi
permeable membrane,
some solutes are pulled
 Diffusion along with it
o best way to remove small o solvent drag
(low molecular weight) o best way to remove
solutes larger solutes
o depends on: o Convective clearance
 Blood and dialysate depends on:
flow rates  Molecular weight
 Membrane surface cutoff o the
area and thickness membrane
 Number of pores  Membrane surface
 Solution temperature area
 Membrane resistance  Ultrafiltration rate
(UFR)

58
Hemodialysis Nurse Training Course
M3 Dialysis Center, Inc.
MODULE 4: Hemodialysis Devices

F8* HF80S F180NR Revaclear*

Urea 186/240 248 274 196/271


Creatinine 172/216 225 251 189/250

Phosphate 138/165 220 238 185/239


B12 76/83 155 168 144/170
B2M Not specified 0.65 Not specified 0.7

*Qb = 200/300, Qd = 500mL/min, UF= 0/60


Clearance = mL/min
Table 3: Clearance values of Various Dialyzers

** A sieving coefficient is for better


used to say how much solute is compatibility
expected to be removed by  adsorbed material
convection. can build up on the
membrane 
Sample Problem: ↓diffusion and
A sieving coefficient of 0.5 for a convection
solute means that 50% of the solute will  highly adsorptive
pass through the membrane to the membranes  less
dialysate side. The other 50% will be effective when they
adsorbed or rejected by the membrane. are reprocessed
many times; cannot
 Adsorption be evaluated by
o occurs when material testing dialyzers for
sticks to the dialyzer total cell volume (also
membrane called fiber bundle
o dialyzers adsorb volume)
materials, usually small
proteins, to some extent;
hydrophobic synthetic
membranes adsorb more
than cellulose
membranes
o Pros & Cons (in dialysis):
 adsorbed protein
keeps the membrane
away from the blood,

Figure2: Adsorption
59
Hemodialysis Nurse Training Course
M3 Dialysis Center, Inc.
MODULE 4: Hemodialysis Devices

 Total cell volume - an indirect Membranes


measure of changes in solute
transport for hollow fiber dialyzers
The semi permeable membrane
that are reused.
acts in some ways like the vessel wall of a
human nephron, because it is selective.
o A dialyzer’s adsorptive
- with microscopic pores, the
ability depends on:
membrane allows only certain
 Membrane material
solutes and water to pass
 Surface area
through
 How much material
- large substances such as protein
has already adsorbed
and blood cells won’t fit through
to the membrane
the small pores.

Dialyzer Design There are membrane factors that


affect removal of solutes and fluids during
Hollow fiber dialyzer - a clear plastic dialysis. These include the membrane
cylinder that holds thousands of fiber tubes material and characteristics of each
almost as thin as strands of hair dialyzer.
- fibers are held in place at each
end by polyurethane, Membrane Materials
clay-like “potting” - can affect diffusion
material that holds and UF as well as
fibers open so blood efficiency of dialysis
can flow inside them and the patient’s
- Hollow fiber dialyzers comfort during
allow for well- treatment
controlled, predictable
UF.  Cellulose
- fibers are rigid  no membranes
membrane compliance o made from
(change in shape or cotton-based
volume due to material that is
pressure); instead, hold spun into
almost the same hollow fibers
amount of fluid at high o have thin fiber
pressures as they do at walls (8-15
low pressures. microns)
- Resistance to blood flow Figure3: Dialyzer Design o size of
is low in hollow fiber dialyzers. molecules
cleared is
limited to about
3,000 Da

60
Hemodialysis Nurse Training Course
M3 Dialysis Center, Inc.
MODULE 4: Hemodialysis Devices

o slower removal of middle molecules,


molecules in the larger depends mainly on UF
molecular weight range, rates
such as beta-2- o do a good job of
microglobulin (ß2m = removing solutes up to
11,800) 15,000 Da
o surface areas range from o clears ß2m to some
0.5-2.1 meters extent
o have larger in vitro o membrane
(tested in a laboratory) biocompatibility ranges
urea and creatinine from good to very good
clearances compared to o are close to pure
synthetic dialyzers synthetics
o are the least
biocompatible; causes  Synthetic membranes
the most complement o are made from polymers
activation that are formed into
o least able to remove hollow fibers
solute by adsorption o The materials used in
synthetic membranes
 Modified cellulose are:
membranes  Polycarbonate
o hydroxyl group (OH-) are  Polyacrylonitrile
removed and replaced (PAN)
with acetate (cellulose  Polysulfone (PSF)
acetate), amino acids, or  Polymethylmethacryl
synthetic molecules ate (PMMA)
o have much
thicker fiber Cellulose Synthetic
walls = 22-40
microns Material Cotton-based Polymers
o use
Width Thin Thickest
convection,
(8-15 microns) (30-55 microns)
diffusion and
adsorption to Size Up to 3,000 Da Up to 15,000 Da
remove Accommodation
solutes
Biocompatibility Least Best
o clearance of
solutes, Adsorption Low High
especially
Table 4: Comparison of Cellulose and Synthetic membranes

61
Hemodialysis Nurse Training Course
M3 Dialysis Center, Inc.
MODULE 4: Hemodialysis Devices

o have the thickest fiber walls = 30-55 Determining Dialyzer


microns
o solutes are removed by convection, Clearance
diffusion, adsorption
o clearance of solutes, especially Manufacturers test dialyzers in vitro,
middle molecules, depends mainly using watery fluids that are thinner than
on UF rates blood. When measured during actual use
o remove solutes up to 15,000 Da, on patients, a dialyzer’s real clearance can
clearing ß2m to some extent differ from the manufacturers stated values
o membrane biocompatibility is very by ±10 – 30%. The clearance of urea (a
good small molecular weight solute) is most often
o highly adsorptive can quickly keep used to test the overall effectiveness of a
the blood from touching the dialyzer.
membrane
Dialysate
Measuring Dialyzer
Effectiveness Purpose of Dialysate
- a fluid that helps remove uremic
wastes, e.g. urea and creatinine,
A dialyzer’s effectiveness is checked
and excess electrolytes
by testing its clearance. Clearance is
- replaces needed substances,
expressed as the amount of blood (in mL)
such as calcium and bicarbonate,
that is completely cleared of a certain
which helps keep the body’s pH
solute in one minute of treatment, at a
balance
given blood flow rate (Qb) and dialysate
- patient’s blood  one side of
flow rate (Qd).
the membrane (blood
compartment)
Sample Problem:
- dialysate  on the other side of
A dialyzer has a stated urea
the compartment
clearance of 250mL/min at a Qb of 300 mL.
- Dialysate and blood never mix,
In one minute, 250 ml of blood would be
unless the membrane breaks
cleared of urea by the dialyzer. If 300mL of
- is prescribed to have a desired
blood is pumped through the dialyzer in
levels of solutes the patient
one minute, only 250 mL of blood will be
needs and none of the ones that
cleared of urea.
must be removed completely
- osmolality (solute particle
** The dialyzer’s surface area is
concentration) of dialysate
fixed. So either Qb or Qd must be increased
should closely match the blood
to increase clearance. Qb is always a factor
to keep too much fluid from
that limits clearance, since there is a limit to
moving across the membrane
how quickly blood can flow out of the
- concentration gradient created
patient’s vascular access.
decide the diffusion rates of

62
Hemodialysis Nurse Training Course
M3 Dialysis Center, Inc.
MODULE 4: Hemodialysis Devices

each solute across the  Calcium chloride


membrane  Glucose
- unwanted solutes leave the  Acetic acid (to lower the dialysate’s
blood and move into the pH)
dialysate; desired solutes stay in Bicarbonate concentrate
the blood  Sodium bicarbonate
- some solutes are added to  In some cases, sodium chloride
dialysate in amounts that can
cause them to enter the - Association for the
patient’s blood e.g sodium, Advancement of Medical
bicarbonate, and chloride Instrumentation (AAMI) has set
standard symbols to match the
concentrates

Almost all of the hemodialysis


machines today can use the varied dialysate
formulations. With the correct and precise
dilution of the concentrates to the
prescribed amount of water, right
concentration of electrolytes is achieved
that is vital to the cellular function.

Concentrate Proportioning Ratios


Typical Range of Substances in Dialysate
Dialysate
Figure 4: Concentration gradient & diffusion rates
 Sodium (Na+)
Composition of Dialysate - a major electrolyte of the body’s
- prescribed by the doctor blood plasma and interstitial
- starts out as two concentrated fluid
salt solutions: acid and - normal sodium concentration in
bicarbonate diluted with precise the blood = 135 – 145 mEq/L
amounts of treated water to - sodium concentration in the
make the final dialysate dialysate is most often kept in
- there are three formulations, the same range
which must match the right acid - level can be adjusted by a
concentrate with the right dialysate delivery system during
bicarbonate concentrate a treatment according to a
doctor’s prescription. This is
Acid Concentrate called sodium modeling.
 Sodium chloride
 Potassium chloride

63
Hemodialysis Nurse Training Course
M3 Dialysis Center, Inc.
MODULE 4: Hemodialysis Devices

Sodium Modeling
 creates more efficient fluid
shifts in the body
 removes fluid faster
 provides for better control of
blood pressure and fluid
removal
 can increase thirst and body
weight, and hypertension
between dialysis treatments

 Potassium (K+)
- a major electrolyte of the Figure 5: Dialysate Contents
intracellular fluid
- kept precise amounts within the - normal range of calcium in the
body on both sides of cell plasma = 8.5 – 10.5 mg/dL (4.5 –
membranes to send nerve 5.5 mEq/L)
signals - dialysate calcium = 2.5 – 3.5
- normal plasma potassium level = mEq/L
3.5 – 5.5 mEq/L
- dialysate range (depending on  Chloride (Cl‫)־‬
the patient’s needs) = 0 – 4 - concentration in dialysate
mEq/L depends on the contents of
chemicals such as sodium
 Magnesium (Mg++) chloride, potassium chloride,
- vital to the nerves and muscles magnesium chloride, and
- triggers enzymes that are key to calcium chloride
carbohydrate use - normal plasma chloride levels =
- normal plasma magnesium level 98 – 111 millimoles per liter
= 1.4 – 2.1 mEq/L (mM/L)
- dialysate range = 0.5 – 1.0 mEq/L - dialysate chloride ranges = 100 -
124 mEq/L
 Calcium (Ca++)
- found in the body in  Glucose (C6H12O6)
extracellular (outside the cells) - may be added to dialysate to
and intracellular (inside the prevent loss of serum glucose
cells) fluid and to reduce catabolism
- builds bones and teeth, helps - dialysate glucose range = 0 – 250
muscles move, is needed for mg/dL
blood clotting, and helps send - can be two to three times higher
nerve signals than normal (70-105 mg/dL),

64
Hemodialysis Nurse Training Course
M3 Dialysis Center, Inc.
MODULE 4: Hemodialysis Devices

Parts Parts Parts Parts


Type/Style
Acid Bicarb Water Dialysate

Drake-W illock 1.00 1.83 34

COBE Labs 1.00 1.72 42.28

Fresenius 1.00 1.225 32.775

Table 5: Concentrate Proportioning Ratios

which aids in UF because it has - added to dialysate to help


an osmotic (water-pulling) effect maintain patients’ pH; healthy
kidneys regulates bicarbonate to
 Bicarbonate (HCO3) keep the body’s pH within the
- used by the body to neutralize very tight limits necessary for
acids that are formed when cells cells to survive
metabolize proteins and other - can reduce dialysis-related
foods used for fuel problems e.g. hypotension,
- a buffer (a substance that tends muscle cramps, nausea, and
to maintain a constant pH in a fatigue after treatment
solution) even if an acid or base
is added Hemodialysis Delivery
- metabolic acidosis = too much
acid in the blood; typical with Systems
CKD patients since they in the
urine cannot excrete enough Purpose
acids - mixes and delivers dialysate
- pumps blood through the
Substance Concentration in Dialysate dialyzer
- monitors various dialysis
Sodium 135 to 145 mEq/L
parameters to ensure a safe
Potassium 0 to 4 mEq/L treatment
- monitors patient and machine
Calcium 2.5 to 3.5 mEq/L safety parameters including
Magnesium 0.5 to 1.0 mEq/L blood flow, dialysate flow,
dialysate temperature,
Chloride 100 to 124 mEq/L conductivity, venous and
Bicarbonate 32 to 40 mEq/L arterial pressure, blood in
dialysate leaks, patient blood
Glucose 0 to 250 mg /dL pressure, etc
Table 6: Typical Range of Substances in Dialysate
65
Hemodialysis Nurse Training Course
M3 Dialysis Center, Inc.
MODULE 4: Hemodialysis Devices

- two major subsystems: the Concentrate Proportioning Ratios


dialysate delivery system and - Two ways in making dialysate:
the extracorporeal blood circuit (both rely on a continuous supply
of fresh concentrate and water
 Dialysate Delivery System to a mixing chamber)
- controls the amounts of water
and chemicals in dialysate, and  Fixed-ratio pumps
checks its conductivity,  uses diaphragm or piston pumps
temperature, pH, flow rate, and to deliver set volumes of
pressure concentrate and water to a
- tests the dialysate for the mixing chamber
presence of blood  Servo-controlled mechanisms
 have conductivity control
 The Proportioning System sensors that constantly check
- dialysate is made by mixing fresh the dialysate’s total ion
concentrate with fixed amounts concentration
of treated water
- mixing is controlled by the Dialysate Flow Process
internal mechanical and - mixed  warmed  monitored
hydraulic design of the delivery for conductivity, temperature,
system pressure, and flow rate 
- water and concentrate ratio is dialysate
dependent on center’s policies - after dialysate leaves the
and procedures (See table on dialyzer  passes through a
concentrate proportioning blood leak detector  drain
ratios)

Figure 6: Single-patient dialysate delivery system


66
Hemodialysis Nurse Training Course
M3 Dialysis Center, Inc.
MODULE 4: Hemodialysis Devices

** Blood in the dialysate could mean mixture of the first concentrate


a tear in the membrane. So blood leak (most often acid) with water; the
detectors are often treated as other sensor measures the final
extracorporeal (outside the body) – alarms, dialysate after the second
even though they check the dialysate. concentrate is added
- multiple monitoring system/
 The Monitoring System redundant monitoring: use
- wrong dialysate = less effective conductivity sensors to make the
dialysis treatment or may cause dialysate itself and a second set
illness or death to a patient of sensors to check the mixtures,
- dialysate must be checked apart from the ones that control
throughout each treatment to the mixing
ensure that it is the right - conductivity is checked at the
concentration and temperature, point of mixing and again before
and that it is flowing at the right the dialysate enters the dialyzer
rate o used so two sensors
would have to fail before
 Conductivity a patient could be
- how much electricity the fluid harmed
will conduct - maybe stated in micromhos/cm,
- electrolytes or salts (chemicals in millimhos/cm,
dialysate except for glucose) microsiemens/cm, or
break apart in water to form millisiemens/cm
positive and negative charged - most dialysate delivery systems
ions have internal, preset
- electrolyte conductivity levels of conductivity limits
the dialysate is checked by the - when the dialysate
proportioning system concentration moves outside the
- is checked by placing a pair of
 Common type of conductivity
electrodes in the dialysate;
alarms:
voltage is applied to the
 Low conductivity - caused by:
electrodes, and the current is
o lack of concentrate in one
measured which gives the
or both of the
estimated total ion
concentrate jugs
concentration of the dialysate
 High conductivity - due to:
- a sensor cell may be used
o poor water flow to the
instead of electrodes
proportioning system
- hemodialysis delivery systems
o untreated incoming
have two or more independent
water
conductivity monitors - with
o use of the wrong
separate sensors and monitoring
dialysate concentrate
circuits: one sensor measures the

67
Hemodialysis Nurse Training Course
M3 Dialysis Center, Inc.
MODULE 4: Hemodialysis Devices

preset safe limits, it triggers a cold water, warming it before it


conductivity monitoring circuit; enters the heater
the circuit stops the flow of - most systems use a heater
dialysate to the dialyzer and controlled by a thermistor, a
shunts it to the drain called type of thermostat  a separate
bypass, thus keeping the wrong temperature monitor is placed in
dialysate from reaching the the dialysate path before the
patient dialyzer with preset limits, and
- the circuit also sets off audible works independently of the
and visual alarms to alert the heater control thermistor
staff - with some delivery systems, the
- the proportioning system must patient is the only “monitor” of
have enough of both low temperature
concentrates to complete the - whenever the temperature is
whole treatment either too hot or cold, a circuit
- the conductivity alarm should be sets off audible and visual alarms
working before each treatment, and triggers bypass to shunt
and the machine readings must dialysate to a drain
be checked against an - before each dialysis treatment,
independent meter check the dialysate temperature
alarm to ensure that it is working
 Temperature properly
- dialysate is kept in the range of
37°C to 38°C (98.6°F to 104°F)  Flow rate (dialysate)
for diffusion to be effective - flow to the dialyzer is controlled
by a flow pump
 too-hot dialysate = hemolysis
- may be preset or according to
 too-cool dialysate = patient is
the doctor’s prescription
cold and reduce diffusion =
- higher dialysate flow rates
↓dialysis efficiency
improves dialyzer efficiency,
- the high limit should be set at no though little improvement
higher than 41°C (105°F) and occurs above 800 mL/min
many alarm systems have a low - dialysate flow rates range from
setting, which should not be 0-1,000 mL/min
below 33°C (91°F)
 Dialysate flow rate audible and
- water must be heated to a
visual alarms may be set off by:
certain temperature before
o low water pressure
mixing with the concentrates
o dialysate pump failure
- some systems use a heat
o a blockage in the dialysate
exchanger before the heater, to
flow path
save energy  used dialysate
o a power failure
transfers its heat to the incoming

68
Hemodialysis Nurse Training Course
M3 Dialysis Center, Inc.
MODULE 4: Hemodialysis Devices

- a high/low conductivity, line clamps (to prevent further blood loss)


high/low pH, high temperature,  system may turn into a bypass mode 
or in some cases blood leak dialysate is shunted to the drain
alarm, can trigger the delivery
system to switch into bypass - Hemastix® (strip that reacts to
mode blood) should be used to check
- check the delivery system before the extent of the leak. The test
each treatment to be sure that must be taken where the
the bypass mode works properly dialysate leaves the dialyzer:
for all dialysate alarm conditions  If blood or pink color
can be seen in the
 Blood leak detector dialysate path, there
- used to check for blood in the is a major leak
used dialysate  Clear dialysate and a
- can sense very small amounts of positive Hemastix
blood, less than can be seen with test suggest a minor
the naked eye leak
- shines a beam of light through  Clear dialysate and a
the used dialysate and onto the negative Hemastix
photocell or photoresistor test mean a false
alarm
- depending on the center’s
procedures for a blood leak,
treatment should be stopped
without returning the patient’s
blood to prevent the
contaminated blood from
infecting the patient
- the blood leak detector’s basic
sensitivity is usually preset by
the manufacturer; adjustments
can be made within this limited
range.

Figure 7: Blood leak detector  pH


- a measure of how acidic or
Blood detection process: alkaline (basic) a solution is
dialysate with blood  pass through the - is based on the number of acid
beam of light  even a tiny amount of ions (hydronium ions) or alkali
blood will break the light beam detector (base) ions (hydroxyl ions) it
senses a break  audible and visual alarms contains
triggered  blood pump stops  venous

69
Hemodialysis Nurse Training Course
M3 Dialysis Center, Inc.
MODULE 4: Hemodialysis Devices

- A solution with:  Ultrafiltration Control


o An equal number of acid and
base ions is neutral and has a Ultrafiltration
pH value of 7.0  occurs during the treatment when the
o More acid ions is acidic and pressure on the blood side of the
the pH value will be less than dialyzer membrane is more positive
7.0 than the pressure on the dialysate
 White vinegar is an acid, side
with a pH of 2.9  this pushes fluid in the blood across
o More base ions is alkaline the membrane into the dialysate
and the pH will be greater compartment where it is then
than 7.0 expelled in the drain
 Bleach (sodium  the difference between these
hypochlorite) is alkaline pressures is the TMP
with a pH of 11.0
 The pH of blood is TMP and dialysate pressure
normally from 7.25-7.45;  determines how much fluid from the
a weak base blood is forced across the
- dialysate must have a pH close membrane
to blood so it does not change  before, dialysis machines used a
the blood pH; in general, the manual system of setting the TMP or
range of dialysate pH is from 7.0- a negative dialysate pressure for
7.4 achieving fluid removal but with the
- whether or not the delivery volumetric dialysis today, TMP is
system has a pH monitor, at the automatically calculated and set by
start of each treatment, an entering the desired fluid removal
external test must be done to (in mL) and the treatment time
ensure that the dialysate pH is in  fluid removal accuracy is more
a safe range precise than earlier UF control
- the most accurate pH measure systems due to:
uses a pH electrode, which puts o the in-vitro KUf values
out a small voltage when placed reported by dialyzer
in a solution; the voltage is read companies, however, in vivo
by a detection circuit that KUf is often somewhat lower
converts the signal into a pH (5% - 30%)
value and displays it o clotting of the dialyzer fibers
- test strips coated with a reduces the KUf by reducing
chemical that changes color the surface area of the
based on pH are another way to membrane
measure it o changes in the blood pump
speed lead to changes the
venous pressure

70
Hemodialysis Nurse Training Course
M3 Dialysis Center, Inc.
MODULE 4: Hemodialysis Devices

o an increase or decrease in monitor the flow rate of the


the dialysate flow, or a kink dialysate
or blockage in the dialysate
lines changes the dialysate  Extracorporeal Circuit
pressure - second major subsystem of the
hemodialysis delivery system
UF control systems - carries blood from the patient’s
 UF control is the means by which access to the dialyzer and back
the dialysis machine removes fluid to the access
from the patient and accurately - includes the:
measures it o arterial and venous blood
 Ultrafiltration rate (UFR) is the tubing
amount of fluid removed in a o blood pump
specific period of time o heparin pump
 Types: o dialyzer
o Volumetric control - uses o venous line clamp
two chambers that fill and o blood flow monitors
drain to control the volume o pressure monitors
of dialysate going to and o air monitors
coming from the dialyzer Components and Monitoring
o Flow control - uses sensors in o Blood tubing
the fluid path to and from - during treatment: blood through
the dialyzer to control and blood tubing flows from the

Figure 8: Volumetric Control System

71
Hemodialysis Nurse Training Course
M3 Dialysis Center, Inc.
MODULE 4: Hemodialysis Devices

Figure 9: Flow Control System

patient’s vascular access (arterial the tubing to the patient’s


needle)  dialyzer  back to needles or catheter ports
the patient’s access (venous  Dialyzer connectors:
needle) o Luer-Lok connectors at the
- with a small inner diameter of other end of the blood
the blood tubing causing only tubing segments connect the
about 100-250ml of blood is tubing to the dialyzer
outside the patient’s body at any o arterial blood tubing
time segment = arterial end of the
- two divisions: arterial (red) and dialyzer; venous blood tubing
venous (blue) segment = venous end of the
- typically, smooth on the inside dialyzer
to reduce clotting and air  Drip chamber / bubble trap:
bubbles o checks the arterial or venous
pressure in the blood circuit
Parts of blood tubing: o uses a monitoring line with
 Patient connectors: transducer protectors
o tip or Luer-Lok® connector o collects or “traps” any air
found at the end of the that accidentally gets into
arterial and venous blood the extracorporeal circuit
tubing segments connecting o keeps blood clots in the
extracorporeal circuit from

72
Hemodialysis Nurse Training Course
M3 Dialysis Center, Inc.
MODULE 4: Hemodialysis Devices

reaching the patient, by o connected to the machine’s


using a very fine mesh screen venous and/or arterial ports
(venous chamber) via a small tubing segment
 Blood pump segment: on top of the drip chamber
o a durable, pliable, larger o use membranes with a
diameter part of the arterial nominal pore size of 0.2
blood tubing threaded microns that are
through the blood pump hydrophobic when wetted,
roller to keep fluid from passing
 Heparin infusion line: through; if these filters get
o most often found on the wet, they prevent air flow
arterial blood segment just o wet or clamped venous
before the dialyzer transducer protector causes
o heparin may be given during TMP problems
treatment to the patient o a loose or damaged
through a very small transducer protector on a
diameter tube that extends pre-pump arterial drip
out of the blood tubing chamber port could allow air
 Saline infusion line: into the bloodline circuit
o allows saline to be given to o change wet transducer
the patient during dialysis protectors right away and
o most often found on the inspect the machine side of
arterial blood tubing the protector for
segment just before the contamination or wetting
blood pump, so saline can be (Centers for Disease Control
pulled into the circuit and Prevention); if a fluid
o if the saline infusion line is breakthrough is found on the
not clamped correctly, too removed transducer
much fluid or air can enter protector, the machine’s
the extracorporeal circuit internal transducer protector
 Transducer protectors (a back-up) must be
o a mechanical device inside inspected by a qualified
the machine that converts air technician
pressure into an electronic
signal, which display venous Other components of the extracorporeal
pressure, arterial pressure, circuit:
and TMP  Blood pump / blood flow rate
o damaged by moisture o moves blood from the
o barrier between blood in the patient’s arterial needle
tube and the transducer in through the blood tubing, to
the machine the dialyzer, and then back

73
Hemodialysis Nurse Training Course
M3 Dialysis Center, Inc.
MODULE 4: Hemodialysis Devices

to the patient through the o all blood pumps have a way


venous needle to allow hand cranking in
o a roller pump is the most case of emergency such as a
common type of blood power failure
pump; uses a motor that
turns a roller head  Extracorporeal pressure monitors
o speed of the roller head set o pressure in the
by the staff determines extracorporeal circuit
blood flow rate depends on blood flow rate
o blood flow rates (BFR) can be and resistance to the flow
varied between 0 mL/min o resistance occurs in every
and 600mL/min according to part of the extracorporeal
the prescription circuit: access needles or
o some machines count blood catheters, blood tubing, and
pump turns and calculate the dialyzer that are overcome
number of liters processed in by a blood pump
a treatment; knowing the o pressures are displayed in
number of liters prescribed millimeters of mercury
to be processed allows (mmHg) on a gauge, meter,
calculation of the blood flow or screen
rate: divide liters processed o extracorporeal pressure
by minutes of treatment monitoring is needed to
o for an effective treatment, calculate TMP and ensure
the blood flow rate must be patient safety
accurate and reflect the o pressure monitors have
doctor’s prescription upper and lower limits and
o if the rollers are too tight may be set; others, have a
(pump occlusion), the blood preset range within which
pump segment may crack or staff can choose a midpoint;
RBC may be destroyed; o when pressure exceeds the
otherwise, blood may escape high or low setting, the
out the back of the segment, system will trigger audible
reducing blood flow
below the prescribed
level
o Pump occlusion must be
checked periodically
and adjusted per the
manufacturer’s
instructions; checked
when the tubing size or
manufacturer changes
Figure 10: Blood pump
74
Hemodialysis Nurse Training Course
M3 Dialysis Center, Inc.
MODULE 4: Hemodialysis Devices

and visual alarms, stop the  Different pressure and safety


pump, and clamp the venous monitors:
line
o extracorporeal blood  Arterial pressure
pressure alarm must be  pressure from the patient’s
checked to ensure access to the blood pump
functionality before  also called pre-pump
treatment pressure
o a pre-pump or post-pump  with a fistula or graft = less
drip chamber may be placed than zero or negative arterial
on the arterial bloodline pressure due to the
o a monitoring line or pressure resistance from the vascular
gauge connection at each access and the pulling of the
drip chamber is used to blood pump
check arterial and/or venous  Predialyzer pressure
pressure in the  pressure between the blood
extracorporeal circuit pump and the dialyzer
 also called as post-pump
pressure, predialyzer
pressure, or post-pump
arterial pressure
 pressure in this segment
of the blood tubing is
greater than zero, or
positive
 monitored to detect
clotting in the dialyzer;
suspect clotting if there is
a large pressure
differential on each side
of the dialyzer
 Venous pressure
 pressure from the
monitoring site to the
venous return
 also called postdialyzer
pressure
 positive pressure readings
 Air detectors
o continuously check the
blood in the venous tubing
Figure 11: Pressure monitoring devices on arterial and venous
bloodlines segment for air and foam

75
Hemodialysis Nurse Training Course
M3 Dialysis Center, Inc.
MODULE 4: Hemodialysis Devices

Figure 12: Pressure alarm triggers

o checks for air at the venous calibrated by qualified


drip chamber or at the blood technicians
tubing just below it o when senses air, it triggers
o are ultrasonic devices that audible and visual alarms,
check for changes in a sound stop the blood pump, and
wave sent through a cross- clamp the venous blood
section of the blood path tubing to keep air from
o since sound travels faster getting into the patient’s
than air than liquid, any air in bloodstream
the blood will raise the speed o must always be used during
at which the sound wave the dialysis treatment and
passes through the blood, venous line clamps engaged
setting off an alarm with the tubing; must be
o sensitivity limits are most checked regularly before
often preset by the each treatment to ensure
manufacture, but can be functionality

76
Hemodialysis Nurse Training Course
M3 Dialysis Center, Inc.
MODULE 4: Hemodialysis Devices

injects the heparin


into the circuit
o heparin is stopped 30
minutes to 1 hour before
the end of the treatment
so blood clotting can go
back to normal
o heparin is infused into the
heparin line on the arterial
blood tubing before the
dialyzer and heparin lines
are placed after the blood
pump segment to avoid
Figure 13: Air Detector Alarm
negative pressure which can
draw air into the
 Heparin system extracorporeal circuit
o blood tends to clot when it o heparin pumps have variable
touches the artificial material speeds that can be set to the
of the lines and the dialyzer, physician’s prescription
thus, heparin, an o heparin is
anticoagulant, is used to
prevent clotting in the
extracorporeal blood circuit Sorbent Dialysis
o heparin may be given in a
variety of ways:  can be used for acute, home, and
 intermittently - (on chronic dialysis treatments
and off) during  needs no water treatment system
dialysis; a prescribed  does not contain water and
amount is injected concentrate proportioning pumps;
into the arterial uses premixed chemicals are added
bloodline at to 6 L of tap water
prescribed times
 bolus - the full
prescribed amount is
given all at once just
before the treatment
 continuous infusion -
a prescribed rate
throughout the
treatment; an
infusion pump is used
and the pump slowly
Figure 14: Heparin infusion line

77
Hemodialysis Nurse Training Course
M3 Dialysis Center, Inc.
MODULE 4: Hemodialysis Devices

Sorbent Dialysis Process: and is continuously displayed.

water and chemicals  Sorbent cartridge has four chemical


↓ layers which regenerates dialysate
cycled through a sorbent regenerative and treats water, purifying the 6L of
cartridge to purify the dialysate dialysate made with tap water\
↓  sorbent cartridge = continuous
dialysate (collected in a disposable bag) dialysate disinfection system =
↓ keeping bacteria and endotoxin
circulated to the dialyzer levels below 1 cfu mL and 0.5
↓ EU/mL, respectively
removal of electrolytes  can do short (3-5 hour) treatments
↓ at Qd up to 400 mL/min, or long and
used dialysate is then cycled through the slow (5-8 hour) treatment at Qd of
cartridge 200-300 mL/min, depending on
↓ which cartridge used
electrolytes & ultrafiltrates are chemically  using the sorbent cartridge = no
converted back into fresh dialysate continuous water source, floor
↓ drain, or water treatment system
storage bag  can be used anywhere with an
electrical outlet (or suitable
Each increase in the total volume of generator)
dialysate is a direct reflection of total UF
Conclusion
 Knowledge of
the hemodialysis
devices including
the dialyzer,
dialysate and the
delivery system
keeps the
patient from
danger.
 The delivery
system is
featured with
monitoring
capacities thus
enforcing safety
and quality of
each
hemodialysis
Figure 15: Sorbent regenerative cartrige

78
Hemodialysis Nurse Training Course
M3 Dialysis Center, Inc.
MODULE 4: Hemodialysis Devices

treatment.
 However, the dialysis staff is still the
most inevitable monitor of all to
keep patient safe. Alarms are of no
use if it was not checked and
verified for functionality before it is
being used.
 Moreover, it is crucial to master the
components of these dialysis
devices as they play a very
important role in hemodialysis
efficiency.
 If all these devices are maintained,
patients will definitely lead patients
to live full and active lives.
 Thus, utmost care, skill at finding
and troubleshooting problems and
one’s attention to detail will make
all the difference in patient’s
outcomes.
Figure 16: Sorbent dialysis system

79

You might also like