Professional Documents
Culture Documents
3.1. Methodology
3.2. Limitation
The vortex tube that will be used in the research and for experimentation is the VT
Exair Model No. 3225, 708 SLPM, 25 SCFM The following is a picture of the vortex tube
that will be used along with its geometric parameters.
The experimental array components used in this study are shown in the figure below,
where the compressed air produced by the reciprocating compressor is stored in a reservoir
(capacity 220 L). The control valve and Filter Regulator (FR) unit regulate the flow rate and
pressure of the compressed air supplied to the VT at the inlet. Furthermore, the FR unit also
separates moisture from the air. The thermocouple measures temperature at three locations,
namely inlet, cold exit, and hot exit with an accuracy of ±0.1C. Pressure is measured using a
pressure gauge (±0.01 bar accuracy) on cold inlet and outlet. The rotameter shows the air
flow rate as it enters the VT and exits cold. Care has been taken to secure the 2 mm diameter
slot temperature probe to ensure its alignment with the VT axis and the cavity is sealed with
adhesive to avoid leakage.
The actual image of the experimental setup is shown in the next image with insulation
and cladding.
The table above shows a series of experiments that need to be carried out to reach the
optimal solution using the Taguchi method. The result for this set or the response parameter
is that Tc has been extracted from the experiment shown in the table below:
The value of ΔTc is calculated as the difference between the maximum and minimum
values of a certain parameter. The highest ΔTc value was observed for lc while the lowest
was for VT isolation conditions. shows that lc is the most important parameter for the cold
exit temperature difference VT. This is followed by inlet pressure. The isolation condition
was found to be the least important parameter.
The graph above can show the advantage of c for the temperature difference VT. The value at
the c level of 0.30 has the highest value of the S/N ratio when considering Tc. The S/N ratio
in the With Insulation (WI) condition is greater than in the Without Insulation (WOI)
condition. Also, the S/N ratio for an inlet pressure of 4 bar is higher than that for an inlet
pressure of 3 bar. This allows us to complete the optimal combination for the highest
performance VT as A3B1C2.
4.2. Regression Analysis
Regression analysis was performed to obtain a definitive mathematical correlation
between various VT performance parameters. And the contribution of individual factors on
system performance is also a key determinant for controlling the system. Therefore, the
researchers used ANOVA for the regression analysis. ANOVA is used to determine the role
of each factor on variations in system performance and to determine which factors have the
most influence. In this study, researchers used ANOVA with independent factors and
interactions between factors, the results of ANOVA from this experiment are shown in the
table below:
From the table above is the result of ANOVA without interaction between factors. the
table shows that the coolant mass fraction factor (µc) has the greatest contribution value in
influencing the value of ΔTc, namely the F value = 112.98 or the contribution value is 86.39,
while the insulation condition factor has the lowest contribution value, namely 2.02 %.
Because the table above is the result of factors without interaction, the researcher suspects
that the contribution can change if there is interaction between the factors. so that the
researcher conducted another ANOVA with the interactions shown in the table below:
From the table of factor interaction anova above, it shows that the factor that has the
most significant influence on the value of ΔTc is fixed, namely c with a contribution value of
86.40%. While the factor that has the least effect on this experiment is the interaction factor
between c and INSC with a contribution value of 0.02%. thus, the numerical model of this
ANOVA was used to predict the value of Tc with significant accuracy, as indicated by the
value of R2. The results of such predictions for the case without interaction, with the
interaction and the taguchi method are shown in the table below:
Experimental and predictive values have also been presented graphically in the graph
below to make it easier to understand the effects of considering interactions in statistical
modelling using regression and the Taguchi method.
From the graph above, it shows that the results of the regression analysis with
interactions show a better fit with the experimental results than the regression analysis
without interaction. This proves the usefulness of the regression equation obtained in this
analysis. Furthermore, the prediction results of the regression equation were validated
through a good match with the predicted results of the Taguchi method.
Referring to table 10 shows that the predicted value and the experimental value of the S/N
ratio are very close to each other. This indicates that the prediction has been successful. With
an error value of 8.84% between the predicted and experimental results, it shows that the
Taguchi method statistical assessment can be used to predict VT performance with significant
accuracy.
5. Conclusion
In this work, the process, and results of statistical assessment of the performance of the
VT temperature separation using the Taguchi method have been described. The breakthrough
achieved in this research is the use of the Mixed Level Design method from Taguchi, to
accommodate the number of different control factors. The results show that the prediction of
temperature separation is better when the interaction effects of control factors are considered.
The biggest contribution was shown by c (86.40%). Followed by the interaction of lc, Pi and
isolation conditions (6.56%). The Taguchi method has been able to obtain the optimal
combination of control factors for maximum temperature separation, which was further
verified by confirmatory experiments. The percentage deviation between the predicted and
experimental results was 8.84%, which was significantly accurate.