Professional Documents
Culture Documents
On The Nature and Reduction of The Displacement Artifact In: Flow Images
On The Nature and Reduction of The Displacement Artifact In: Flow Images
G. NISHIMURA,JOHN1. JACKSON,
DWIGHT AND JOHNM. PAULY
In flow-imaging experiments with 2-D Fourier transform sequences, the time difference
between phase encoding and readout leads to a potentially misleading displacement artifact.
This artifact arises in regions of rapid flow and high shear, and manifests as an intensity
distortion in addition to a bulk shift. We have studied methods of mitigating the artifact,
including offset-echo acquisition, backward-evolvingphase encoding, moment-compensated
phase encoding, and projection-reconstruction imaging. Experiments on flow phantoms
verified the nature and reduction of this displacement artifact. Of the four methods studied,
the projection-reconstruction sequence proved to be the most effective, completely elim-
inating the artifact. 0 1991 Academic Press, Inc.
INTRODUCTION
We first consider the simplest case of a single spin positioned at (xa, yo) after ex-
citation and flowing at constant velocity with components ( v x , v,,). Using a conven-
48 I 0740-3194/91 $3.00
Copyright 0 1991 by Academic P r q Inc.
All rights of reproduction in any form reserved.
482 NISHIMURA, JACKSON, AND PAULY
121
The phase behavior of the flowing spin is therefore the same as that of a stationary
+
spin at yo v ~7/2);
( hence the y-component of the spin’s reconstructed position will
be based on the spin’s location at the center of the phase-encoding interval. While the
y-component depends on the phase-encoding interval, the x-component depends on
+
the spin’s position at the gradient echo-that is, at xo v,TE, where TE is the time
from the excitation to the gradient echo. Here, we assume both the excitation pulse
and readout dephasing interval are brief. We also ignore the blurring effect from motion
during the readout interval ( 4 ) . The displacement artifact arises because the x- and
y-positions are encoded at different times. Given the streaming nature of flow, we can
determine the extent of displacement by referencing time with respect to the center
of the phase encoding, 7/2, and considering the time difference from that point.
In this case then, the time difference is AT = TE - 7 / 2 and the displacement is
(VAT, 0 ) .
We now extend this analysis by considering a collection of flowing spins, as in
the case of a straight vessel oriented at some angle 8 with respect to the readout axis
(Fig. 1 ). Examining the behavior in the primed coordinate system with y’ aligned
with the longitudinal axis of the vessel, a spin positioned at (xb, y b ) at the center
+
of the phase-encoding interval will be reconstructed at (xb V A T cos f? sin 8, yb
+ VATsin f? sin f?), where v is the flow velocity (assumed parallel to the y’ axis) and
AT is the time from the center of the phase-encoding lobe to the gradient echo. Because
of the streaming nature of flow along y’, the artifact can be studied by considering
only the displacement in the x’direction. This x’-displacement is u A T cos 8 sin 8 (or
(VAT/2 ) sin 20), which is clearly a maximum when 8 = 45 O .
FIG.I . Oblique vessel oriented at angle 0 to the readout (x) and phase-encoding ( y ) axes. Flow is assumed
to be along the y’ axis.
DISPLACEMENT ARTIFACT IN FLOW IMAGES 483
In projective vessel imaging, the resultant projection profile will depend on the
distribution of velocities across the lumen because the profile is based on the line
integral through the lumen. With plug flow, the uniform velocity distribution leads
to the same amount of displacement for all spins in the vessel, and therefore the
displacement causes a bulk shift of the vessel with no distortion of the projection
profile.
With parabolic flow, v ( x ’ , z) = 2u,, ( 1 - (x” + z 2 ) / R 2 )where
, R is the vessel
radius and v,, is the average velocity. In this case, the distribution of velocities leads
to a distribution of displacements, creating a more subtle effect on the resultant line
integrals. Figure 2 plots the computed vessel projection profile (along x‘) for the case
of parabolic flow through a circular lumen. The ideal semiellipse profile when v,,, = 0
becomes distorted when v,, = 20 and 40 cm/s ( A T = 4 ms, 0 = 45”, R = 0.32 cm
( 8 in.)). The asymmetry of the profile with increasing velocity is a result of a “piling-
up” effect in the direction of the shift as faster flowing material, displaced by a greater
amount, reconstructs on top of adjacent slower flowing material ( 5 ) . The vessel profile
boundaries do not necessarily shift because of the lower velocities near the vessel wall.
Clearly, the distortion becomes appreciable when the displacement reaches the order
of a pixel width.
More generally, the effect seen in Fig. 2-in which the signal piles up-becomes
more pronounced in regions of high shear. Where a steep velocity gradient exists,
regions of relatively high velocities closely border regions of low velocities, resulting
in a “discrete” displacement. As an extreme example, we consider a 1-D velocity
profile consisting of three regions of plug flow (Fig. 3). The middle region will be
shifted by an amount greater than that of the two outer regions; therefore, the projection
FIG.2. Simulated vessel projection profiles: Ideal semiellipse profile of circular lumen when v, = 0 cm/
s (solid line) becomes distorted for v,,
= 20 cm/s (dashed line) and u,, = 40 cm/s (dotted line). Parabolic
velocity distriblltion assumed with AT = 4 ms, 0 = 45”, and R = 0.32 cm ( in.).
484 NISHIMURA, JACKSON, AND PAULY
velocity
profile
projection profile
FIG.3. Extreme shear example: Given a velocity distribution comprised of three regions of plug flow, the
middle region displaces more than the outer regions, creating a sharp peak and a void in the resultant
projection profile.
profile will show a sharp peak in the right transition region where the shift has resulted
in an overlap, and a void in the left transition region where the shift has rendered a
gap. This effect is similar to a water/fat chemical shift artifact in which fat shifts with
respect to water, leaving a void and creating a bright region of overlap.
ARTIFACT-REDUCTION METHODS
FIG.4. Comparison of test sequences: (a) Offset echo, (b) backward-evolvingphase encodin&(c) moment-
compensated phase encoding, (d) projection reconstruction.
the RF pulse with each higher-order phase encode, maintaining a fixed TE (Fig. 4b).
To analyze the impulse response in this case, the focus is on the interval, parameterized
by T ~which
, vanes with each phase encode. Thus, the phase upon completion of the
phase-encoding interval is
t 31
= Y.o[
YO + %( 7 - ?)I7. 141
for the positive phase encodes ( -Gyo applies for the negative phase encodes). The
impulse response amounts to a blur in this variable-duration scheme, not a simple
shift as in the variable-amplitude scheme. However, one simple interpretation of Eq.
[4] is that the effective y-position, yo -t vY(7 - ~ ~ / 2 although
), varying with each
phase-encoding step, is based nearer the end of the phase-encoding interval for the
lower-order phase encodes.
* + G,( 1 - a') -
T L TL
G ~ ( Y- G,TTE [71
2 2
for the first moment. Thus moment compensation for displacement reduction cor-
responds to making the first moment large, not zero. To determine the waveform
parameter values, we typically know Gyr = A (the desired phase-encoding area) and
G,, (the maximum gradient amplitude). Letting G2 = -GI = G,, and rearranging
Eqs. [ 61 and [ 71, expressions for r and (Y are
and
1 A
ff=--- [91
2 2Gma,r'
Note that LY < i , which implies that the second lobe will be longer than the first.
A trade-off with this type of moment compensation is the long gradient interval r
involved. To examine how severe this is, we let /3 = (TE - r ) be the time from the
end of phase encoding until the gradient echo. Therefore /3 represents how far forward
+
in time the phase-encoding lobes must compensate. Substituting /3 r for TE in Eq.
[ 8 1, it can be derived that
A
Gmax
For comparison, conventional uncompensated single-lobe phase encoding would use
a gradient of amplitude Gm,, and duration rc to achieve the desired gradient area A .
Thus,
Substituting this expression into Eq. [lo], we obtain an expression relating T to 7,:
However, even in the easiest case in which p = 0, the moment-compensated lobes are
2.4 1 times longer than the conventional uncompensated lobe.
EXPERIMENTAL RESULTS
FIG.5. 2-D Projection-reconstruction impulse response to a flowing spin: The flowing spin is assumed to
have velocity in the x-direction corresponding to motion of about 1.5 pixels during readout. The response
reveals asymmetric blur extending in all directions with energy shifted in the x-direction. Magnitude response
is shown next to an ideal impulse at x = 0 for comparison.
axis. Partial k-space reconstruction ( 9 ) was applied for the highly offset gradient-echo
acquisitions. To restrict the field-of-view along the primary flow direction, all sequences
used the same mildly selective asymmetric RF pulse with its main lobe offset toward
the readout to shorten the TE and to reduce phase shifts (6). Table 1 lists the TE,
echo offset, and AT for the five sequences.
The reference sequence was a 2-D FT sequence, velocity-compensated along the
readout direction, with a 2.15-ms trapezoidal phase-encoding lobe adjacent to the 6.2-
ms readout interval (TE = 6.8 ms, echo offset = 25%, AT = 2.6 ms).
A highly offset gradient-echo 2-D F T sequence served as one of the test sequences
(TE = 3.8 ms, echo offset = 42%, AT = 1.5 ms). With an echo offset of 42%, partial
k-space reconstruction becomes desirable for improving the detail of smaller structures.
The backward-evolving phase-encoding sequence (TE = 3.8 ms, echo offset = 42%,
AT = 1.0 ms) was approximated in the following manner. For the lower-order phase
TABLE 1
Test Sequence Parameters
encodes, the amplitude of a 1.0-ms triangular lobe adjacent to the readout interval
was incremented until the maximum amplitude of 1 G/cm was attained. For the
higher-order phase encodes, the lobe became trapezoidal at the maximum amplitude
as the additional area under the lobe was gained by increasing the interval toward the
RF pulse. The readout gradient was the same as in the highly offset echo sequence.
The moment-compensated sequence (TE = 6.6 ms, echo offset = 42%, AT = 0
ms) possessed a longer TE because of the additional gradient action needed to create
the large first moment. The phase-encoding gradient consisted of two lobes of equal
amplitude, the first of 1.7 ms and the second of 3.3 ms. This gave an effective AT of
0 ms, assuming constant velocity during the sequence. The same readout gradient
was used as in the previous two sequences.
For the PR sequence (TE = 1.0 ms, echo offset = 50%, AT = 0 ms), 256 radial
lines spaced over a 27r angular extent were acquired and reconstruction was performed
by interpolating the data onto a 2-D Cartesian grid prior to a 2-D FFT.Data acquisition
commenced at the base of the gradient ramp to minimize phase shifts and TE. The
readout interval for each radial k-space “spoke” was 3.1 ms.
Figures 6 and 7 show images obtained on straight and branching flow phantoms.
All of the images shown are magnified by 3X. In Fig. 6 , we can compare the results
FIG.6. Oblique tube phantom: (a) Reference Sequence, flow OFF; ( b ) reference sequence, flow ON from
bottom (uavc = 60 cm/s). Signal displacement causes signal pile up toward the right. Test sequences: (c)
Offset echo, (d) backward-evolving phase encoding, (e) moment-compensated phase encoding, (f) projection
reconstruction. Images (c) and (d) show progressively reduced artifact. Images (e) and ( f ) show no artifact.
490 NISHIMURA, JACKSON, AND PAULY
FIG. 7. Branching phantom: ( a ) Reference sequence, flow OFF, ( b ) reference sequence, flow ON from
bottom (nave= 40 cm/s). Artifact appears as bright oblique line beginning at the origin of the left branch
and as a diminished signal at the inner wall of the left branch. Test sequences: (c) Offset echo, ( d ) backward-
evolving phase encoding, (e) moment-compensated phase encoding, ( f ) projection reconstruction. Images
(c), ( d ) , (e) show progressively reduced artifact with better “filling” of the inner wall of the left branch.
Image ( f ) shows no artifact.
6d) as the signal on the tube’s left side is nearly comparable with that on the right
side. Both the moment-compensated phase-encoding sequence (Fig. 6e) and the 2-D
PR sequence (Fig. 6f) reveal no shift artifact.
The second experiment is of a branching phantom (Fig. 7) with a diameter of 0.47
cm ( h in.) in the main segment. Compared to the reference 2-D F T image with no
flow (Fig. 7a), the reference 2-D FT image with flow (average velocity of 40 cm/s)
entering the main segment from below is shown in Fig. 7b. The artifact in Fig. 7b
appears as an unusually bright line beginning near the origin of and extending into
the left branch. Correspondingly, there is signal loss along the inner wall of the left
branch. Such an artifact is attributable to the high shear near the inner walls of the
branch (10)-the distortion manifests as a region of diminished signal next to a dis-
tinctly bright region as discussed earlier. We observe some reduction in this artifact
in the test sequence images of Figs. 7c-7f. Both the highly offset echo image (Fig. 7c)
and the backward-evolving phase-encoding image (Fig. 7d) show a small reduction
in the intensity and extent of the artifact. A more marked reduction is seen in the
moment-compensated image (Fig. 7e). The artifact is completely absent in the 2-D
PR image (Fig. 7f) as this image compares favorably with the reference image obtained
without flow.
DISCUSSION
REFERENCES
I. G. K. SCHULTHESS AND C. B. HIGGINS, Radiology 157, 687 (1985).
2. D. G. NISHIMURA, A. MACOVSKI, AND J. M. PAULY,IEEE Trans. Med. Imaging MI-5, 140 ( 1986).
3. D. N. FIRMIN, G. L. NAYLER, P. J. KILNER,AND D. B. LONGMORE, Magn. Reson. Med. 14,230 ( 1990).
4. V. J. WEDEEN,R. E. WENDT,AND M. JEROSCH-HEROLD, Magn. Reson. Med. 11, 114 ( 1989).
5 . D. 0. KUETHEAND R. J. HERFKENS, Magn. Reson. Med. 10,57 (1989).
6. D. G. NISHIMURA, A. MACOVSKI, J. I. JACKSON,B. S. Hu, C. A. STEVICK, AND L. AXEL,Magn. Reson.
Med. 8,96 (1988).
7. A. KUMAR,D. WELTI,AND R. R. ERNST,J. Magn. Reson. 18,69 (1975).
8. P. C. LAUTERBUR, Nature 242, 190 (1973).
9. D. C. NOLL,D. G. NISHIMURA, AND A. MACOVSKI, IEEE Trans. Med. Imaging MI-10, 154 ( I 99 1).
10 F. J. WALBURN AND P. D. STEIN,J. Biornech. 14(9), 601 (1981).
11. D. C. NOLL,J. M. PAULY,D. G. NISHIMURA, AND A. MACOVSKI, in “Proceedings, 9th SMRM-WIP,”
p. 1314, 1990.
12. J. I. JACKSON, C. H. MEYER,D. G. NISHIMURA, AND A. MACOVSKI, in “Proceedings. 9th SMRM,” p.
428. 1990.