Professional Documents
Culture Documents
Radiography Test Procedure Part - 1: Surface Preparation
Radiography Test Procedure Part - 1: Surface Preparation
–1
This Radiography Test Procedure is written with reference to ASME
BPVC Sec-V, Article 2 & 22 for film base radiography of castings
and welds with X-rays/Gamma Rays using industrial radiographic
film.
Note: Whole procedure in one blog can lead to a very long blog,
hence the procedure is written in parts.
Surface Preparation
For Casting: Castings Surfaces should not have surface
irregularities like dents & surface deformity that the surface
irregularities don’t hide the discontinuity or be confused with the
image of any discontinuity on the resulting radiograph.
Back-scattered Radiation
Radiation which is scattered more than 90 deg with respect to
the incident beam, that is, backward in the general direction of
the radiation source.
Backscatter Radiation
SYSTEM OF IDENTIFICATION
A system shall be used to produce permanent identification on each
radiograph traceable to the
contract,
component,
weld or weld seam, or
part numbers, as appropriate.
Example : Identification System
Caution: In any case, this information shall not obscure the area
of interest.
INTENSIFYING SCREENS
Intensifying screen: a material that converts a part of the
radiographic energy into light or electrons and that, when in contact
with a recording medium during exposure, improves the quality of
the radiograph, or reduces the exposure time required to produce a
radiograph or both.
Lead foil screens are commonly used in direct contact with the
films, and, depending upon their thickness, and composition of
the specimen material.
screen used in front of the film acts as a filter, absorbing the
scattered radiation arising from the specimen, thus improving
radiographic quality.
the back lead screens are used as protection against back-
scattered radiation and their thickness is only important for this
function.
For radiography using radioactive sources, the minimum
thickness of the front lead screen should be 0.005 in. (0.13
mm) for iridium-192, and 0.010 in. (0.25 mm) for cobalt-60.
Copper Screen
Fluorescent Screens
Fluorescent intensifying screens have a cardboard or plastic
support coated with a uniform layer of inorganic phosphor
(crystalline substance). Phosphors convert photon energy to light.
The support and phosphor are held together by a radiotransparent
binding material.
Types of IQI
Material: IQIs shall be selected from either the same alloy material
group or grade as identified in SE- 1025 for hole type or SE-747 for
wire type, or from an alloy material group or grade with less
radiation absorption than the material being radiographed.
However with regard to (a) and (b) above, when the actual
material/weld thickness is measured, IQI selection may be
based on these known values.
Better IQI sensitivity is shown in any row of the Table below which is
above the equivalent sensitivity row. If the designated IQI and hole
are not represented in the table, the next thinner IQI row from Table
T-283 may be used to establish equivalent IQI sensitivity.
In Radiography quality level 2-2T the first number of the quality level
designation refers to IQI thickness expressed as a percentage of
specimen thickness; the second number refers to the diameter of
the IQI hole that must be visible on the radiograph, expressed as a
multiple of penetrameter thickness, T.
For Example :
n=2
if 2T hole of the IQI is visible on the radiograph
ES = 2 x (√2T/2T)
=2%
If you want to find Value of “n” for a given Hole type
IQI, it is given by the formula
Formula to calculate
“n” value for given IQI and Thickness
Note 2: The quality level required using wire IQI’s shall be equivalent
to the 2-2T level of Practice E1025 unless a higher or lower quality
level is agreed upon between purchaser and supplier.
RADIOGRAPHIC DENSITY
film density: The quantitative measure of diffuse optical light
transmission (optical density, blackening) through a developed film.
Density Limitations
1.8 minimum for single film viewing for radiographs made with
an X-ray source
2.0 minimum for radiographs made with a gamma-ray source.
1.3 minimum density in each film of for composite viewing of
multiple film exposures, each
The maxim.um density shall be 4.0 for either single or
composite viewing.
Density Variation
1. vary by more than minus 15% or plus 30% from the density
through the body of the designated hole-type IQI adjacent to
the essential hole or adjacent to the essential wire of a wire-
type IQI,
2. exceed the minimum/maximum allowable density ranges as
provided above when calculating the allowable variation in
density, the calculation may be rounded to the nearest 0.1
within the range specified above.
(b) When the requirements of (a) above are not met, then an
additional IQI shall be used for each exceptional area or areas and
the radiograph retaken.
(c) When shims are used with hole-type IQIs, the plus 30% density
restriction of (a) above may be exceeded, and the minimum density
requirements given above do not apply for the IQI, provided the
required IQI sensitivity mentioned below are met.
Required Sensitivity
Radiography shall be performed with a technique of sufficient
sensitivity to display the designated hole-type IQI image and
the essential hole, or the essential wire of a wire-type IQI as
required in table T- 276 or Table T-283 in case of using
equivalent IQI.
For wire-type IQIs, the essential wire shall be visible within the
area of interest representing the thickness used for
determining the essential wire, inclusive of the allowable
density variations described above.
MONITORING DENSITY OF RADIOGRAPHS
or
Calibration of Densitometers
Densitometers shall be calibrated at least every 3 months
during use as follows:
The step wedge calibration film shall have been verified within
the last year by comparison with a national standard step tablet
unless, prior to first use, it was maintained in the original light-
tight and waterproof sealed package as supplied by the
manufacturer.
(a) The density of the steps on a step wedge comparison film shall
be verified by a calibrated densitometer.
References:
ASME BPVC Sec-V 2019
Article 1
Article 2
Article 22, SE-94