You are on page 1of 126

FieldMan

Measurement Device for


Electromagnetic Fields
User manual
Narda Safety Test Solutions GmbH
Sandwiesenstraße 7
72793 Pfullingen, Deutschland

® Names and Logo are registered trade-


marks of Narda Safety Test Solutions GmbH -
Trade names are trademarks of the owners.

© 2022
Order no.: 2460/98.21
Issue: 01/10.2022, A ...
Previous issue: --/--, A ...
Subject to change.
Our normal terms of warranty and delivery
apply.
Table of Contents

1 About this product . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7


1.1 About the measurement of electromagnetic fields . . . . . . . . . 8
1.2 About the Narda FieldMan . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.3 Supported probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

2 General safety instructions. . . . . . . . . . . . . . . . . . . . . . 11


2.1 Using these operating instructions . . . . . . . . . . . . . . . . . . . . 12
2.2 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.3 Non-intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.4 Dangers due to electromagnetic fields . . . . . . . . . . . . . . . . . 12
2.5 Correct handling of rechargeable batteries. . . . . . . . . . . . . . 13

3 About this operating manual . . . . . . . . . . . . . . . . . . . . 15


3.1 Structure of the operating instructions . . . . . . . . . . . . . . . . . 16
3.2 Language of the user interface . . . . . . . . . . . . . . . . . . . . . . . 16
3.3 Characters and symbols used . . . . . . . . . . . . . . . . . . . . . . . 17

4 Connection and commissioning. . . . . . . . . . . . . . . . . . 19


4.1 Unpacking. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.2 Equipping the case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.3 Device overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.4 Power supply/rechargeable battery operation . . . . . . . . . . . 24
4.5 Connect and test probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.6 Initial commissioning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

Narda FieldMan 3
5 Controls and display elements . . . . . . . . . . . . . . . . . . . 29
5.1 Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.2 LED display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.3 Display overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5.4 Display of probe info on the display. . . . . . . . . . . . . . . . . . . 34
5.5 Status bar symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.6 Symbols of function keys . . . . . . . . . . . . . . . . . . . . . . . . . . . 36

6 General operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
6.1 The operating concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
6.2 Basic operating steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
6.3 The display in measuring mode. . . . . . . . . . . . . . . . . . . . . . 42

7 Device settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
7.1 Setting types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
7.2 Personalizing device settings . . . . . . . . . . . . . . . . . . . . . . . 45
7.3 General Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
7.4 Legal Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
7.5 Expert Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
7.6 Position and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
7.7 Connectivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
7.8 Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
7.9 Setups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
7.10 Data Logger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56

8 Operating modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
8.1 Overview of operating modes and probe types . . . . . . . . . . 58
8.2 Field Strength . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
8.3 Spatial Average . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
8.4 Timer Logging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
8.5 Spectrum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
8.6 Shaped Time Domain . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83

4 FieldMan Narda
9 Data Logger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
9.1 Structure of the Data Logger . . . . . . . . . . . . . . . . . . . . . . . . 92
9.2 Managing projects (Project) . . . . . . . . . . . . . . . . . . . . . . . . . 93
9.3 Managing measurements (Measurements) . . . . . . . . . . . . . 94
9.4 Managing screenshots (Screenshots) . . . . . . . . . . . . . . . . . 96
9.5 Managing media (Media) . . . . . . . . . . . . . . . . . . . . . . . . . . . 96

10 Cross-mode functions . . . . . . . . . . . . . . . . . . . . . . . . . . 97
10.1 Post-Processing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
10.2 Selection of a standard . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
10.3 Offset correction – Zero . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
10.4 Commenting on measurement results . . . . . . . . . . . . . . . . 103
10.5 Alarm function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
10.6 Acoustic intensity display (HF probes only) . . . . . . . . . . . . 105
10.7 Peak marker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
10.8 Display of overdriven measured values . . . . . . . . . . . . . . . 107

11 PC software and updates . . . . . . . . . . . . . . . . . . . . . . 109


11.1 Narda-TSX PC software . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
11.2 Firmware update. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111

12 Care, maintenance and disposal . . . . . . . . . . . . . . . . 113


12.1 Cleaning the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
12.2 Replacing/removing rechargeable batteries . . . . . . . . . . . . 115
12.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116

13 Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
13.1 EU Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . 118
13.2 UKCA Declaration of Conformity . . . . . . . . . . . . . . . . . . . . 118
13.3 Regulatory Compliance Mark (RCM) . . . . . . . . . . . . . . . . . 118
13.4 FCC and IC statement . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
13.5 Radio approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
13.6 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119

Narda FieldMan 5
6 FieldMan Narda
1 About this product

In this chapter you will find some useful information about


measuring electromagnetic fields, about the Narda
FieldMan and about the probes.

1.1 About the measurement of electromagnetic fields


(page 8)
1.2 About the Narda FieldMan (page 8)
1.3 Supported probes (page 9)

Narda FieldMan 7
1 About this product

1.1 About the measurement of


electromagnetic fields
In the modern world, people live and work virtually permanently in the vicinity of technical
equipment that generates electromagnetic fields. With increasing research into their effect
on humans, there is an increasing awareness of the problem and availability of information
on this topic. Limits have long since been defined by various bodies to protect users from
exposure to emissions.

1.2 About the Narda FieldMan


The Narda FieldMan measures non-ionizing radiation and low frequency fields with the
highest accuracy. With its digital measuring probes for electric and magnetic field
strengths, the instrument covers the range from static and low-frequency fields in medical
and industrial applications to mobile phone frequencies and millimeter waves. Both
broadband HF probes with a flat frequency response and “shaped probes” are available,
which evaluate the field strength according to a personal protection standard. Probes
with integrated FFT analysis allow spectral measurements and analysis in the time
domain up to frequencies of 400 kHz. All probes have a digital interface that transmits the
measurement data to the base unit without interference. This eliminates the need for cal-
ibration of the base unit.
The newly developed, extremely powerful Narda-TSX PC software is available for evalu-
ating and documenting the measurement results, media and other information. It is
Narda's new software platform that will support other Narda products in the future in addi-
tion to FieldMan. The firmware update is also performed via the PC software
(see 11 PC software and updates on page 109).

8 FieldMan Narda
1.3 Supported probes

Applications and standards


The FieldMan enables precision measurements to ensure the safety of people, espe-
cially in working environments where high electric or magnetic field strengths are
expected. A key task is to demonstrate compliance with general safety regulations such
as FCC, IEEE, ICNIRP or the EMF Directive 2013/35/EU. Examples of measurement
environments are:
• Broadcast transmitters (e.g. IEC 62577)
• Radio base stations (e.g. IEC/EN 62232)
• Induction heating and melting (e.g. EN 50519)
• Household appliances (e.g. IEC/EN 62233)
• Electric welding equipment (e.g. IEC/EN 62822)
• Railroad environment (e.g. EN 50500)
• Vehicle environment (e.g. IEC 62764)
• Power supply systems (e.g. IEC/EN 62110)
• Medical electrical equipment (e.g. IEC/EN 60601)
• TEM cells and absorber chambers for the verification of electromagnetic compatibility
(EMC)

1.3 Supported probes


A variety of isotropic field probes are available for the FieldMan. All of them transmit their
information and measurement data as a digital signal to the FieldMan, either via an elec-
trical USB interface or via an optical COM interface. In this way, interference is almost
completely eliminated compared to high-impedance analog interfaces. The specially
developed screw connectors and electrical contacts are extremely robust and resistant.
The probes are automatically detected after connection to the FieldMan.
The characteristics of the probes are listed in the respective data sheet. The following
probe types are supported:
• Broadband probes with digital interface for FieldMan (P/N 2462/xx)
• Selective probes with digital interface for FieldMan (P/N 2463/xx)
• EHP-50F and EHP-50G field analyzers with optical interface (P/N 2404/xxx)
• HP-01 magnetometer with optical interface (P/N 2405/xxx)

Narda FieldMan 9
1 About this product

10 FieldMan Narda
2 General safety
instructions

This chapter contains important information on the safe use


of the FieldMan. You should therefore read this chapter
carefully and follow the instructions given.

2.1 Using these operating instructions (page 12)


2.2 Intended use (page 12)
2.3 Non-intended use (page 12)
2.4 Dangers due to electromagnetic fields (page 12)
2.5 Correct handling of rechargeable batteries (page 13)

Narda FieldMan 11
2 General safety instructions

2.1 Using these operating instructions


⇒ Read these operating instructions carefully and completely before working with the
device.
⇒ Keep these operating instructions at hand so that they are always available to all us-
ers when working with the device.
⇒ Only pass the device on to third parties together with these operating instructions.

2.2 Intended use


The Narda FieldMan is used to measure and evaluate electromagnetic fields.
⇒ Only operate the device under the conditions and for the purposes for which it was
designed.
⇒ Pay particular attention to the technical specifications in the data sheet for the device.
Intended use also includes the following:
⇒ Observe the national accident prevention regulations at the deployment location.
⇒ The device may only be operated by appropriately qualified and trained personnel.

2.3 Non-intended use


The Narda FieldMan is not a warning device that actively warns of the existence of
dangerous fields by means of visual or acoustic signals.
⇒ Always consider the device as a measuring device, never as a warning device.
⇒ Only approach unknown field sources while carefully observing the current measured
value display.
⇒ In case of doubt, use an additional warning device such as RadMan or Nardalert
from Narda Safety Test Solutions.

2.4 Dangers due to electromagnetic fields

2.4.1 Strong fields


Very strong electromagnetic fields occur near many radiation sources.
⇒ Observe safety barriers and markings.
⇒ In particular, persons with electronic implants must keep away from dangerous areas.

12 FieldMan Narda
2.5 Correct handling of rechargeable batteries

2.4.2 Undervaluation of the field strength


Metallic stickers in the sensor area of the probe can lead to measurement errors,
especially to undervaluation of the electromagnetic field strength.
⇒ Attach all types of stickers to the black probe shaft only.
⇒ An automatic self-test function can detect any probe faults, alerts the user and does
not allow further measurement in case of serious faults. The sensor system is also
checked for open circuits and short circuits in the process. In case of doubt, a test
generator (see accessories in the data sheet) can also be used for checking.
⇒ If a malfunction is suspected, take the unit out of operation and contact your Narda
service center. Addresses can be found at the end of these operating instructions and
on the Internet at www.narda-sts.com.

2.5 Correct handling of rechargeable


batteries
Rechargeable batteries can overheat, explode or ignite if handled incorrectly.
⇒ Observe the following notes when handling rechargeable batteries:
– Always handle the rechargeable battery with care.
– Do not drop, damage or expose the rechargeable battery to excessively high
temperatures.
– Never store the rechargeable battery under very high temperatures (e.g. in the
car) for more than one to two days.
– Never leave the discharged rechargeable battery in the unused measuring
instrument for a long time.
– Never store the rechargeable battery for more than six months without charging it
in between.
– Close the rechargeable battery compartment again after replacing the
rechargeable battery and never operate the FieldMan with the rechargeable
battery compartment open.
– Defective rechargeable batteries are special waste and must therefore not be
disposed of with the usual household waste.
– Observe the national regulations for the disposal of rechargeable batteries.

Narda FieldMan 13
2 General safety instructions

14 FieldMan Narda
3 About this operating
manual

This chapter contains information on the structure and


layout of the operating instructions.

3.1 Structure of the operating instructions (page 16)


3.2 Language of the user interface (page 16)
3.3 Characters and symbols used (page 17)

Narda FieldMan 15
3 About this operating manual

3.1 Structure of the operating instructions


These operating instructions are divided into the following main sections:
• General operation: Here you will find basic operations such as switching on and off.
• Device settings (DEVICE SETTINGS): Device settings describes all settings that
relate to the device in general.
• Operating modes: The operating modes and the operating mode-specific settings
(MEASUREMENT SETTINGS) are described here.
• DATA LOGGER: In this chapter you will find all of the information about retrieving
stored measurements.
• Cross-mode functions: Some functions affect several or all operating modes.
These functions are described here.

3.2 Language of the user interface


In these operating instructions, screen contents are presented in English. However, the
user interface of the FieldMan can also be displayed in other languages (see “Select the
language”).
⇒ Please note that in this case the displayed terms differ from those described.

16 FieldMan Narda
3.3 Characters and symbols used

3.3 Characters and symbols used


In these operating instructions, various elements are used to indicate special text mean-
ings or especially important text passages.

3.3.1 Symbols and warning words in warning


instructions
In accordance with American National Standard ANSI Z535.6-2006, the following warn-
ing instructions, symbols and warning words are used in this document:

In combination with the warning words CAUTION, WARN-


ING and DANGER, the general danger symbol warns of a
risk of serious injuries. Observe all the following instruc-
tions to avoid injuries or death.

Indicates a danger that will result in damage to or destruc-


NOTICE tion of the device.

CAUTION Indicates a danger with low to moderate risk of injury.

Indicates a danger that may result in death or serious


WARNING injury.

Indicates a danger that will result in immediate death or


DANGER severe injury.

Structure of the warning instructions


All warnings are structured as follows:

WARNING WORD
Type and source of the danger
Consequences resulting from non-observance
⇒ Action for danger avoidance

Narda FieldMan 17
3 About this operating manual

3.3.2 Symbols and text markings in the document


Symbol Meaning
✓ Prerequisite
Indicates a prerequisite that must be satisfied before one of the
following actions is performed, e.g.:
✓ The device is switched off.
⇒ Action
Indicates a single action, e.g.:
⇒ Switch on device.
1. Sequence of actions
2. Indicates a sequence of actions that must be performed in the
3. specified order.

↳ Result
Indicates the result of an action, e.g.:
↳ The device starts a self-test.
Bold text Controls and display elements
Identifies controls and display elements, e.g.:
⇒ Press ENTER key.

3.3.3 Terms and abbreviations


Term Meaning
Rechargeable Rechargeable battery
battery
Battery Non-rechargeable battery
SK1 – SK4 Softkey 1 – Softkey 4

18 FieldMan Narda
4 Connection and
commissioning

This chapter provides information about the connection


options of the Narda FieldMan, gives instructions for
commissioning and describes the connection options.

4.1 Unpacking (page 20)


4.2 Equipping the case (page 21)
4.3 Device overview (page 22)
4.4 Power supply/rechargeable battery operation (page 24)
4.5 Connect and test probe (page 25)
4.6 Initial commissioning (page 27)

Narda FieldMan 19
4 Connection and commissioning

4.1 Unpacking
ATTENTION
Device/accessories damaged during transport
Commissioning of damaged equipment/accessories may cause consequential
damage.
⇒ After unpacking, check the device and all accessories for any damage that may
have occurred during shipping.
⇒ In case of damage to the device or components, contact your Narda service center.
The addresses of your Narda service center can be found at the end of these
operating instructions manual and on the Internet at www.narda-sts.com.

4.1.1 Packaging
The packaging is designed so that it can be reused if not damaged during a prior trans-
port.
⇒ Do not throw away the packaging and use the original packaging for all further trans-
portation.

4.1.2 Package contents


⇒ Check the delivery for completeness.
• Narda FieldMan
• Rechargeable battery (already inserted)
• Case
• Probe (type and quantity according to order)
• Charging power supply
• Shoulder strap
• USB cable
• Bag with 5 colored marking rings for broadband probes
• PC software (free download from www.narda-sts.com)
• General safety instructions
• Operating instructions on USB stick
• Calibration certificate (for probes)

20 FieldMan Narda
4.2 Equipping the case

4.2 Equipping the case


In addition to the components included in the scope of delivery, numerous other option-
ally available accessories can be stored in the supplied case. The following diagrams
show the distribution in the top and bottom of the case.

1 16
11
2
7
6

3 9
8

12

10
4 14
15
5 13

Fig. 1: Case with equipment

1 FieldMan 9 Country adapter for FieldMan


2 Small parts and bag with marking rings charging power supply
3 Shoulder strap (2244/90.49) 10 EHP-50 table-top tripod (2404/
4 EHP-50F 90.02)
5 Suitable for all probes 11 Folding bag for USB cable, optical
cables, safety instructions and USB
6 HP-01 stick
7 EHP/HP-01 charging power supply 12 RadMan belt bag (2250/92.06)
with country adapter (2259/92.08)
13 O/E converter (2260/90.06)
8 FieldMan charging power supply
(2259/92.29) 14 Tripod extension (2244/90.45)
15 Test generator (2244/90.38)
16 Broadband probes

Narda FieldMan 21
4 Connection and commissioning

4.3 Device overview


1

6 11

2 7

12
8

3
9
13
10

Fig. 2: Device overview FieldMan

22 FieldMan Narda
4.3 Device overview

No Description
.
1 Connector plug for probes
With USB-C socket
2 Display
Display area with LEDs for the operating and charging status (top) and the softkey
fields.
3 On/Off key
4 Keypad
With the soft keys, cross key, Back and Save key
5 Tripod connection
With thread1) for vertical mounting on a tripod
6 USB-C socket
For charging the rechargeable battery, remote control and data transmission
7 Optical port
For connecting components (e.g. EHP-50F/G, HP-01)
8 Micro SD card
For saving measurement data, setups, user standards and media files
9 Network connection
Gigabit Ethernet LAN connection for remote control and data transmission
10 Auxiliary
MMCX connector (for future applications)
11 Flap
• Closed: to store the device with small probes
• Opened: to store the device with large probes
Ultrasonic range finder (flap open)
For distance measurement in conjunction with the GNSS/Range Finder option
12 Tripod connection
With thread1) for horizontal mounting on a tripod
13 Battery compartment
⇒ To change the rechargeable battery (see 12.2 Replacing/removing recharge-
able batteries on page 115 ).

1) 1/4-inch 20-thread UNC female thread

Narda FieldMan 23
4 Connection and commissioning

4.4 Power supply/rechargeable battery


operation
This device uses a rechargeable Li-ion battery pack, which is installed and pre-charged
in the device. The included USB-C Power Delivery charger (12 V, 3 A) enables fast
charging of the built-in rechargeable batteries. A full charge takes place within 4 hours,
with 80% of capacity reached after just 2.5 hours.

4.4.1 Operation with charger connected


Measuring with the charger connected can influence the measurement properties.
This means that compliance with the technical data is no longer guaranteed.

With the above restriction, the FieldMan can be powered from any USB interface that
supplies a current of at least 1 A. Operation is then also possible without a built-in
rechargeable battery.
⇒ Use a floating power supply for long-term operation, e.g. a suitable power bank (PD
12 V, 2.5 A, BC1.2 and QC3.0).

⇒ To ensure compliance with the permissible radio interference emissions, use only
the supplied charging power supply or the car charger (optional accessory) from
Narda.

4.4.2 Operation with rechargeable battery


⇒ Fully charge the rechargeable battery pack before commissioning.

Charging rechargeable batteries


If the device is not expected to be used for several months:
⇒ Charge the device before storage to avoid deep discharge.
Starting the charging process:
1. Plug the included USB-C charging power supply with the matching country adapter
into a power outlet.
2. Connect the FieldMan with the supplied USB-C cable.
↳ The charging process begins.

Not all USB-C cables support fast charging with Power Delivery.

24 FieldMan Narda
4.5 Connect and test probe

Display of rechargeable battery and charge status


The charge status is indicated by the status LED:
• Rechargeable batteries are being charged: Status LED lights up red.
• Rechargeable batteries are charged: The device automatically switches to trickle
charge, status LED lights up green.
In addition, the display shows the rechargeable battery capacity as well as the charging
or rechargeable battery mode:
• Device on: Rechargeable battery capacity in percent and charge status are
displayed in the upper right corner (see 5.5 Status bar symbols on page 35).
• Device off: When charging begins, the current charging process, the capacity and
the charging mode are displayed for approx. 1 minute.
– Fast Charging: Fast charging
– Charging: Normal charging
⇒ Activate the display again by pressing any key.

4.5 Connect and test probe


Probes can be connected directly to the probe connector on the top of the FieldMan or to
the optical input:
• Connection at the probe connector: All digital HF and LFFieldMan probes
• Connection at the optical input: HP-01 and EHP-50F/G field analyzers.

4.5.1 Connect the probe to the probe connector


1. Hold the probe so that the lug on the probe connector faces the groove in the socket
on the top of the instrument.
NOTICE! Do not push in the connector by force. Check position of lug and groove.
2. Insert the probe connector straight into the probe socket all the way from the top.
3. Tighten the locking nut finger-tight. Do not use tools.

Removing the probe:


⇒ Loosen the locking nut and pull the probe upwards and out at the metal plug.

Narda FieldMan 25
4 Connection and commissioning

4.5.2 Connecting the probe to the optical input


The HP-01 and EHP-50F/GF field analyzers are connected to the optical input of the
FieldMan. A suitable optical cable is included with the probes.
Connecting a probe:
1. Go to GENERAL SETTINGS > CONNECTIVITY > OPTICAL INTERFACE and select
Probe.
2. Connect a suitable probe via the optical cable to the optical connector on the
FieldMan (see Fig. 2: 7).

⇒ For more information on using a probe via the optical link, refer to the operating in-
structions for the probe.

4.5.3 Test probe


After connecting a probe to the probe connector, a self-test is automatically started. The
digital part, the analog part and the measuring sensors of the probe are checked for cor-
rect function. If a fault is detected that could affect the measurement, a corresponding
message is displayed and measurements are not possible.

WARNING
Probe is not working properly
Due to a defective probe, any high radiation levels that may be present cannot be
detected.
⇒ Observe the information displayed during and after the self-test of a probe. If there
is any doubt about the correct functioning of the probe, a test source (see accesso-
ries in the data sheet) can be helpful to check the probe.
⇒ In order to make the correct probe selection, you should always obtain knowledge
of the expected frequency and field strength before starting a measurement.

4.5.4 Attaching marking rings


Broadband probes are provided with a groove for a marking ring in the area of the con-
nector. Probes are generally supplied with a black ring. Users of different probes are
offered the possibility to distinguish the probes more easily by attaching colored rings.
The scope of delivery includes 5 colored rings for this purpose.

26 FieldMan Narda
4.6 Initial commissioning

4.6 Initial commissioning

4.6.1 Switch on the device

NOTICE
Damage to a device covered in dew during commissioning
A device stored or transported at low temperatures may become covered in dew
when brought into a warm room. If it is put into operation in this condition, it may
be damaged.
⇒ To prevent damage, wait until all dew on the surface of the device has evaporated.
⇒ The device is not ready for operation until it has reached the operating range of ‒20
to +50 °C.

The device is switched on via the Power key on the left side of the housing.
⇒ Press Power key for > 1 s to switch on the device.
↳ After the system start, the device is ready for operation.
During initial commissioning, then change the following settings if necessary:
• The language of the interface
• Local time
• Time zone (with GNSS option)
• Country of use (for WiFi/Bluetooth option)

4.6.2 Selecting the language of the interface


1. Press SK4 to open the device settings.
2. Under GENERAL SETTINGS > Language, select the language
(see 7.3.1 DISPLAY on page 45).

4.6.3 Setting the local time and time zone


1. Press SK4 to open the device settings.
2. Under POSITION AND TIME > DATE / TIME > Time, set the time,
under... > Time Zone, set the time zone (see 7.6.3 DATE / TIME on page 49).

Narda FieldMan 27
4 Connection and commissioning

4.6.4 Selecting the country of use (for WiFi/


Bluetooth option)
If the WiFi/Bluetooth option is available, a selection list of countries for which a radio
license exists appears during initial commissioning and after a factory reset. Then select
the country in which the device will be operated. The country selection configures the
radio module for this region and selects the corresponding radio standard.

The device may only be operated with the correct country setting! Operation of the
radio module in countries not listed is not permitted.

Select the country of use:


✓ The device is being operated for the first time or a factory reset has been performed.
⇒ On the start screen, select the country of use from the drop-down menu and confirm
with OK.
Changing the country setting at a later date:
A factory reset must be performed for this purpose.
1. Press SK4 to open the device settings.
2. Perform the reset under GENERAL SETTINGS > DEVICE > Factory Reset
(see 7.3.5 DEVICE on page 47).
3. On the start screen, select the country of use from the drop-down menu and confirm
with OK.

28 FieldMan Narda
5 Controls and display
elements

In this chapter you will find descriptions of the device with its
connections and keys as well as the display.

5.1 Controls (page 30)


5.2 LED display (page 32)
5.3 Display overview (page 33)
5.4 Display of probe info on the display (page 34)
5.5 Status bar symbols (page 35)
5.6 Symbols of function keys (page 36)

Narda FieldMan 29
5 Controls and display elements

5.1 Controls

3
4 6
5

30 FieldMan Narda
5.1 Controls

No Description
.
1 Status LED
Operating display and charge status (see 5.2 LED display on page 32).
2 Power key
• Switch device on/off: Press and hold key > 1 s
• Screen saver: With the device switched on, briefly press the key
(see 6.2.2 Using the screen saver on page 39)
• Hardware reset: Press key for a very long time (> 10 s)
3 Softkeys SK1... SK4: (from left to right)
Performs different functions depending on the associated screen display. The la-
beling mainly takes the form of symbols (see 5.6 Symbols of function keys on
page 36).
4 Back key
Change back to the next higher menu level. However, a paused measurement re-
mains in HOLD mode (display is paused).
A long key press resets the measured values (Reset Results).
5 Cross key ◄
Navigation in menus and input fields, move marker horizontally, change result type
for large measurement display, peak marker to left.
Cross key ►
Navigation in menus and input fields, move marker horizontally, change result type
for large measurement display, peak marker to right.
Cross key ▲
Navigation in menus and selection lists, change values in selection fields, move
peak marker to the next higher peak.
Cross key ▼
Navigation in menus and selection lists, change values in selection fields, move
peak marker to the next lower peak.
Cross key ●
Confirmation of a selection (OK key).
6 Save key
A short key press saves the displayed measurement results, if desired also with a
screen copy.
A long key press saves a screen copy as a PNG file.

Narda FieldMan 31
5 Controls and display elements

5.2 LED display


The device status is indicated via the 3-color status LED above the display.
When the device is switched on:
✓ The status LED immediately lights up red and signals the device response.
✓ The screen remains dark.
✓ Shortly after, the boot screen is displayed with the Narda logo and FieldMan lettering.
✓ After the operating system is loaded, the status LED lights up according to the
following conditions:

USB connected

Device status Display


Device off Red: Device charging rechargeable battery
Green: Rechargeable battery is full
Device on Red: Device charging rechargeable battery
Green: Rechargeable battery is full
Red flashing fast: Fault, has priority over red

USB not connected

Device status Display


Device off LED off
Device on Yellow: Power-On status
Red flashing slowly: Rechargeable battery almost empty (oper-
ating time < 15 min), has priority over white
Red flashing fast: Fault, has priority over red flashing slowly

Alarms or remote operation are indicated on the display and not via the status LED.

32 FieldMan Narda
5.3 Display overview

5.3 Display overview


The display is divided into different display areas:
1
2
3

No Area Description
.
1 Status bar Time, alarm, GPS, BT, WiFi, charge status, mute,
date, air temperature, humidity, distance meter
2 Probe information Probe model, charge status, correction frequency, fre-
quency range/span, field type, filter setting, pulse de-
tection (Pulse Detection = On, short integration time
e.g. for radar signals)
3 Information about the Result type, detector
measured values
4 Main display of the mea- Isotropic measured value (RSS) and measuring unit,
sured values measured values for spatial axes (if available)
5 Additional display of the Statistical measured values over the entire period of
measured values the measurement, depending on the operating mode
6 Measurement sequence Graphical display of measured values in the time or
or measurement graphic frequency range, bar display, or information on the
measurement process
7 Function keys (softkeys) SK1 – SK4

Narda FieldMan 33
5 Controls and display elements

5.4 Display of probe info on the display


The most important probe information and its settings are shown on the display during
the measurement. This means that the result of a measurement can also be clearly
assessed on the basis of a screenshot. Some examples of the probe info field are
explained below.
HF broadband probes

1
2 3

1 Probe model 4 Correction frequency


2 Probe frequency range 5 Reference frequency
3 Field type, filter setting, Pulse when
pulse detection is switched on (probe-
dependent)

EHP-50F/G and HP-01


1 3
2 4

1 Probe model 3 Probe rechargeable battery charge


2 Configured span 4 Field type and range

BFD-400
1
2 3
4

1 Probe model 3 Field type and range


2 Configured span incl. low-pass filter 4 Pass or stop filter
setting

34 FieldMan Narda
5.5 Status bar symbols

5.5 Status bar symbols


Symbol Function Symbol Function Symbol Function
Rechargeable WiFi switched on WiFi connected
battery charge
full
Rechargeable Bluetooth Bluetooth
battery charge switched on connected
half full
Rechargeable GNSS switched GNSS position
battery charge on, no position found
low yet
Rechargeable Speaker muted GNSS signal
battery charge lost, position fro-
critical zen
Rechargeable Alarm switched Memory card
battery is on missing
charging
Ambient tem- Relative humid- Distance mea-
perature ity surement

Narda FieldMan 35
5 Controls and display elements

5.6 Symbols of function keys


Symbol Function Symbol Function Symbol Function
Operating Zoom out Add text com-
mode1) ment
Measurement Axis selection Edit text com-
settings ment
Menu Exit Upper case

Pause Edit or rename Lower case

Start Confirm Backspace

Stop Delete Move data to an-


other project
Marker Activate timer Jump to next re-
on/off cord

Peak marker Apply current Jump to previous


on/off time record

Zoom Record voice Restart


on/off comment audio/video

Zoom in or add Edit voice com- Stand-by (only


ment for TIMER LOG-
GING)

1) Initials indicate the active operating mode (e.g. FS = FIELD STRENGTH)

36 FieldMan Narda
6 General operation

In this chapter you will find information on the general


operation of the FieldMan. Operating mode-specific
operating instructions can be found in the descriptions of
the respective operating mode.

6.1 The operating concept (page 38)


6.2 Basic operating steps (page 39)
6.3 The display in measuring mode (page 42)

Narda FieldMan 37
6 General operation

6.1 The operating concept


The combination of the context-dependent softkeys (SK1 - SK4) and the permanently
assigned keys (cross key, Back, Save) enables intuitive and fast operation of the Narda
FieldMan.

6.1.1 Basic functions of the softkeys


The softkeys SK1 - SK4 provide access to all instrument and measurement settings as
well as to all measurement functions. At the top level, the softkeys SK1, SK2 and SK4
have identical functions in all operating modes. SK3 has an operation-specific function
(e.g. starting the measurement)

SK1 SK2 SK3 SK4


Operating mode Measurement Operating mode- Device settings
selection settings specific function

Display of the active function


The symbol above a softkey always shows the function that is activated when the softkey
is pressed (i.e. not the one that is currently active), e.g.:

Key Function
The symbol indicates that pressing the softkey will turn on the
marker (i.e. not that it is already on).

6.1.2 Functions of the fixed keys


Key Function
Back: Return from all submenus to the next higher level.

Save: Save measurements (short press) and screenshots (long


press).
Cross key: Navigate in menus, select entries, change values,
change cursor position in a graphical display
● key: Confirm selection
Power: Short press activates the screen saver

38 FieldMan Narda
6.2 Basic operating steps

6.2 Basic operating steps


Even if some functions and displays differ in the various operating modes, the operation
is basically identical. These identical functions and displays are described below. Differ-
ences and details can be found in the descriptions of the operating modes and the menus
(e.g. DATA LOGGER).

6.2.1 Switching the device on/off


The device is switched on and off via the Power key on the left side of the housing.
Switch device on/off:
⇒ Press Power key and hold for > 1 s.

6.2.2 Using the screen saver


When the screen saver is activated, the screen is turned off, thus saving power. However,
the device continues to measure normally. The screen saver can either be activated
immediately by briefly pressing the Power key or configured via a timer in the device set-
tings.
Activate/deactivate screen saver immediately:
✓ The device is switched on.
⇒ Press the Power key briefly (< 1 s).
↳ The screen is switched off.
The operating indicator (yellow LED) remains lit.
⇒ Press any key briefly to turn the screen back on.
Configure screen saver:
1. Press SK4 to open the device settings.
2. Select GENERAL SETTINGS > DISPLAY > Screen Saver.
3. Select the desired time after which the screen saver is to be activated without
pressing any keys (see 7.3.1 DISPLAY on page 45).
⇒ If the screen saver was automatically turned off after the selected time, press any key
briefly to turn the screen back on.

6.2.3 Selecting an operating mode


⇒ Press SK1 , use the ▲▼ keys to highlight an entry, and use the ● key to select
it.
⇒ For information on the operating modes, see 8 Operating modes on page 57.

Narda FieldMan 39
6 General operation

6.2.4 Changing the measurement settings in an


operating mode
✓ The desired operating mode is active.
⇒ Press SK2 , use the ▲▼ keys to highlight an entry, and use the ● key to select
it.
⇒ For information on the menus and functions, see the descriptions of the respective
operating mode.
Simplified representation
In the following descriptions, this simplified representation is used for changes in the
measurement settings:
⇒ MEASUREMENT SETTINGS: ... (corresponding action).

6.2.5 Starting and pausing/stopping a


measurement
⇒ Press SK3.

6.2.6 Changing the device settings


⇒ Press SK4 .
⇒ For information on device settings, see 7 Device settings on page 43.

6.2.7 Exiting a submenu


⇒ Press key.

6.2.8 Saving a measurement


⇒ Briefly press the key.
↳ A message confirms that the measurement has been saved.
⇒ To manage stored measurements, see 9 DATA LOGGER on page 91.

40 FieldMan Narda
6.2 Basic operating steps

6.2.9 Taking a screenshot


⇒ Press the key for approx. 2 s.
↳ A message confirms that the screenshot has been taken.
⇒ To manage saved screenshots see 9 DATA LOGGER on page 91.

Narda FieldMan 41
6 General operation

6.3 The display in measuring mode


The displays in the operating modes are always structured in the same way. The dis-
played measured values, evaluations and graphical representations then differ depend-
ing on the operating mode, the probe used and the selected settings.
⇒ Detailed information on the displays can be found in the descriptions of the operating
modes.

3
4

No. Area Values


1 Selection field • Display type: ACTUAL, AVERAGE...
• Detector type: RMS, Peak
2 Main field Measured value display
X/Y/Z only for ACTUAL, the unit can be selected
3 Status bar When the measurement is running, a bar indicates the prog-
ress of the averaging.
4 Supplementary Evaluations
field MAX, MIN, MAXAVG, AVG and the current averaging time
until the set time is reached.
5 Graphic display • Temporal or spectral display of the measurement results
• Bar graph of the instantaneous value.

42 FieldMan Narda
7 Device settings

The menu can be used to change the device settings and to


display the information about the measuring instrument and
a connected probe. In addition, the menu allows access to
user-defined setups and stored measurement results in
DATA LOGGER.

7.1 Setting types (page 44)


7.2 Personalizing device settings (page 45)
7.3 GENERAL SETTINGS (page 45)
7.4 LEGAL NOTICES (page 47)
7.5 EXPERT SETTINGS (page 47)
7.6 POSITION AND TIME (page 48)
7.7 CONNECTIVITY (page 50)
7.8 INFORMATION (page 51)
7.9 SETUPS (page 53)
7.10 DATA LOGGER (page 56)

Narda FieldMan 43
7 Device settings

7.1 Setting types


On this device, a distinction is made between Device settings and Measurement set-
tings.

7.1.1 Device settings


Device settings include all parameters that affect the general device behavior or the form
of the display, but do not affect the measurement itself. The device always remembers
the last used device settings, stores them on the SD card and loads them the next time
the device is started. Device settings, with the exception of EXPERT SETTINGS, are not
saved in a setup.
Open device settings:
⇒ Press SK4.
For information on device settings, see the following sections:
• GENERAL SETTINGS on page 45
• EXPERT SETTINGS on page 47
• POSITION AND TIME on page 48
• CONNECTIVITY on page 50
• INFORMATION on page 51
• SETUPS on page 53
• DATA LOGGER on page 91

7.1.2 Measurement settings


These include all parameters that influence a measurement. They are grouped under the
softkey SK2 and are supplemented by the settings under EXPERT SETTINGS (see
7.5 EXPERT SETTINGS on page 47). Only measurement settings are used for saving
and loading a setup.
Open measurement settings:
⇒ Press SK2 in the respective operating mode.
Information on the measurement settings can be found in the description of the respec-
tive operating modes.

7.1.3 Resetting settings


Via Factory Reset (see 7.3 GENERAL SETTINGS on page 45), both setting types can
be reset to the delivery state.

44 FieldMan Narda
7.2 Personalizing device settings

7.2 Personalizing device settings


Measuring instruments are often used by several people. The FieldMan offers a quick
and easy way to personalize settings, setups and measurement results through its
removable memory card. We therefore recommend every user to use their own personal
memory card.
Using a new memory card is completely hassle-free:
⇒ Insert new memory card.
↳ The required directory structure is created automatically.

Narda recommends the use of high quality memory cards for industrial applications.

7.3 GENERAL SETTINGS


In this submenu, settings are made for the device behavior that affect the screen display,
acoustic notifications, and the power-on and power-off behavior of the device.

7.3.1 DISPLAY
Parameter Default Description
Language English Language of the user interface
Color Scheme Normal Color scheme of display
Auto-Brightness Control On Automatic adjustment of screen
brightness (via light sensor)
Brightness Level 40% Setting of screen brightness
Screen Saver 10 min Screen saver. Reactivation of screen
by pressing any key.
Temperature Unit °C Temperature unit

Narda FieldMan 45
7 Device settings

7.3.2 SAVE OPTIONS


Parameter Default Description
Add Voice Comment Off Add a voice comment when saving.
Add Text Comment Off Add a text comment when saving.
Default Text ... Default text when adding text com-
ments.
Auto Save Screenshot Off When saving manually, a screen
copy is automatically created.

7.3.3 SOUND
Parameter Default Description
Mute Off Muting of the speaker, symbol in the
status bar
Sound Level 50% Playback volume
Audible RF Indicator Off Acoustic intensity display with 3 sen-
sitivity settings (parameter appears
only with HF probes)

7.3.4 ALARM
Parameter Default Description
Alarm Function Off Alarm when limit exceeded
⇒ For description, see 10.5 Alarm
function on page 104.
E-Field Unit V/m Unit of E-field
E-Field Limit 61 V/m Limit of E-field
B/H-Field Unit A/m Unit of B/H field
B/H-Field Limit 0.16 A/m Limit of B/H field
Limit for Results in % 100% For evaluating probes and SHAPED
TIME DOMAIN

46 FieldMan Narda
7.4 LEGAL NOTICES

7.3.5 DEVICE
Parameter Default Description
Format Memory Card FORMAT Format SD card after confirmation
Automatic Shutdown 30 min Power off during inactivity
Calibration Reminder When due Can be activated on expiry or in ad-
vance, or generally deactivated
Factory Reset RESET Sets all adjustable parameters to the
as-delivered values

7.4 LEGAL NOTICES


In this submenu you will find information about the radio approvals, and the Electronic
Label (E-Label).
To access the E-Label:
⇒ Press SK4, ▼ key and OK.
⇒ For more information, see 13.5 Radio approval on page 119.

7.5 EXPERT SETTINGS


This submenu is probe-dependent and can only be selected if an HF probe or BFD-400
is connected. There are no adjustment options for other probes.

7.5.1 POST-PROCESSING
In this submenu, settings can be made for HF probes that require precise knowledge of
the prevailing field. If the frequency of a field source is known and is specified here as the
reference frequency, a more accurate display result can be obtained by applying the cor-
rection factor stored in the probe, or the percentage of the limit that has been reached can
be displayed in % relative to a safety standard (see 10.1 Post-Processing on page 98).

HF probes are broadband probes and cannot measure selectively at the set frequency.
In the case of incorrect frequency specification or multifrequency fields, these settings
can therefore lead to incorrect results.

Narda FieldMan 47
7 Device settings

Parameter Default Description


Apply Correction Factor Off Apply correction factor on/off
CF
Reference Frequency Fmin Reference frequency for correction
(probe dependent) factor and for % of standard
% of Standard Off Conversion to display in % of stan-
dard for measured field strengths at
a reference frequency
Standard ICNIRP 1998 Occ Applied standard

7.5.2 Band-Filter
In this submenu, settings for a bandpass or bandstop filter with variable center frequency
can be made for suitable probes (currently only BFD-400). After switching on, the filter
has an effect only in the broadband modes FIELD STRENGTH, SPATIAL AVERAGE
and TIMER LOGGING.

Parameter Default Description


Band-Filter Off Switch band filter on/off
Filter Type Band-pass Filter type
Center Frequency 50 Hz Center frequency of the filter

7.6 POSITION AND TIME


In this submenu, settings are made which affect the position determination by means of
the satellite receiver and the integrated ultrasonic distance sensor. The system time of
the device can also be set or synchronized with the very accurate satellite time. The
results of the GNSS position determination are also displayed.

48 FieldMan Narda
7.6 POSITION AND TIME

7.6.1 DISTANCE METER (optional)


Parameter Default Description
Distance Meter Off Distance meter on/off
Status at Power-on Off Activate function automatically after
device start
Distance Unit Meter (m) Distance unit

7.6.2 LOCATION SERVICES (optional)


Parameter Default Description
GNSS Receiver Off GNSS receiver on/off
Status at Power-on Off Activate function automatically after
device start
GNSS Format DMS GNSS format
Latitude 48°27'29.718''N Current latitude
Longitude 9°13'49.425''E Current longitude
Altitude (MSL) 440 m Current altitude
Accuracy H/V 4.3/8.0 m Approximate horizontal and vertical
position accuracy
Systems GPS/ Galileo/ GNSS systems to be used
GLONASS

7.6.3 DATE / TIME


Parameter Default Description
Date Format YYYY-MM-DD Date format
Date 2022-08-18 Current date
Time Format 24 h Time format
Time 13:30:59 Current time
Time Zone UTC+01:00 Time zone
Daylight Saving Time Off Daylight saving time on/off
Sync with GNSS SYNC Synchronization of time with GNSS

Narda FieldMan 49
7 Device settings

7.7 CONNECTIVITY
In this submenu, settings are made for remote operation via wireless and wired connec-
tions.

7.7.1 SERIAL INTERFACE


Parameter Default Description
USB On USB on/off
Optical Interface Probe Optical interface selection:
• Off
• Remote: Remote control function
• Probe: Probe connection

7.7.2 NETWORK
Parameter Default Description
TCP Remote Port 5025 Port for TCP Remote
Host Name fieldman_<S/N> Predefined host name

7.7.3 WIRELESS INTERFACE (optional)


This group only appears if the chargeable WiFi/BT option has been activated.

Parameter Default Description


WiFi Off WiFi on/off
Bluetooth Off Bluetooth on/off

50 FieldMan Narda
7.8 INFORMATION

7.7.4 ETHERNET
Parameter Default Description
Ethernet Off LAN on/off
With LAN = Off the following lines
are grayed out
DHCP On If DHCP is enabled, the following
lines are grayed out
IPv4 Address 192.168.128.128 Settings menu, individually scrollable
digit groups
Subnet Mask 255.255.255.0 Settings menu default masks 1...32,
default 24
IPv6 Address fc00::192:168:128: Settings menu, individually scrollable
128 digit groups
Prefix 7 Prefix length of the IPv6 address
MAC Device-dependent Individual MAC address of the Field-
Man

7.8 INFORMATION
This submenu displays device and probe-specific information. In addition, the user or a
calibration laboratory can update the third-party calibration date and recommended cali-
bration interval for all probes except the HP-01 and EHP-50F/G. For probes with self-test
capability, the test can be performed here.

7.8.1 DEVICE INFORMATION


Parameter Default Description
Device Self-Test START Starts self-test to be performed in
separate window (also performed at
each restart)
Product Name FieldMan Product name
Part Number 2461-01 Article number
Serial Number A-0051 Serial number
Firmware Version V1.0.0 Firmware version

Narda FieldMan 51
7 Device settings

Parameter Default Description


Bootloader Version V1.0.1 Bootloader version
Image Version V1.0.2 The image contains all SW compo-
nents
<option name> P/N of the option Display only if option is enabled

7.8.2 PROBE INFORMATION


Parameter Default Description
Probe Self-Test START Starts self-test to be performed in
separate window (also performed
each time the probe is connected)
Product Name EFD-0691 Probe product name
Part Number 2462/14 Probe article number
Serial Number A-0051 Probe serial number
Field Type E Field type
Shaping Standard No shaping Standard
Lower Frequency Limit 0.1 MHz Lower cutoff frequency
Upper Frequency Limit 6000 MHz Upper cutoff frequency
Firmware Version V1.0.0 Firmware version of the probe
Bootloader Version V1.0.0 Bootloader version of the probe
Temperature Sensor 15.7 °C Probe temperature
Narda Calibration Date 2022-12-15 Last calibration by Narda
Third-Party Calibration – Calibration date of other suppliers
Date can be entered here by the user.
Recalibration after 24 Number can be edited here by the
(months) user.
<option name> P/N of the option Display only if option is enabled

52 FieldMan Narda
7.9 SETUPS

The information under Narda Calibration Date or Third-Party Calibration Date


(most recent date counts) and Recalibration after (months) are used as a reminder
function for an upcoming calibration. The values are stored in the EEPROM of the
connected probe. No values can be changed for the EHP-50F/G and HP-01 probes.
There is therefore no Third-Party Calibration Date here and the recommended
calibration interval is always 24 months.

7.9 SETUPS
A setup contains all changeable settings that can influence a measurement and its mea-
surement result. This includes all settings under MEASUREMENT SETTINGS (see in
the respective operating mode) and EXPERT SETTINGS (see 7.5 EXPERT SETTINGS
on page 47).
Using a setup can greatly simplify and speed up the preparation of recurring measure-
ments. In addition, faulty measurements due to incorrectly selected settings can be ruled
out.
Since the selection options offered and the saved setups depend on the type of probe
connected in each case, the submenu can only be activated when the probe is plugged
in. The device setups are stored separately according to the 4 defined probe groups on
the SD card of the device.

7.9.1 SD card directory structure


The setups are stored in the directory structure specified below. A maximum of 99 setups
can be created per subdirectory.

Directory Setups
Setups / RF / HF probes
(high frequency broadband probes)
Setups / HP / HP-01
(probes for static fields)
Setups / EHP / EHP-50F, EHP-50G
(FFT field analyzers)
Setups / LF / BFD-400, 100 cm2 and 3 cm2
(and other low frequency probes)

Narda FieldMan 53
7 Device settings

In the delivery state, the directories are empty. Measurement settings are saved to one
of the above directories whenever the instrument is turned off, a probe is disconnected
or unplugged, or the user saves the current measurement settings as a setup with the
probe connected.

The last general device settings used (see 7.3 GENERAL SETTINGS on page 45) are
also saved on the SD card when the device is switched off. Since the settings are
probe-independent, they are stored in the Device Settings directory. When the device
is switched on, these settings are loaded or – if the file is missing – the factory settings
are loaded.

7.9.2 Switching on the device without a probe


connected
Without a probe connected, only the probe-independent Last Device Settings are loaded.
The last measurement settings Last Settings are loaded only after a connected probe
has been detected.

7.9.3 Switching on the device with a probe


connected
After switching on the device, the last used device and measurement settings are
restored by loading the two settings files. If one of the files is not available, the factory
settings are loaded instead.

7.9.4 Probe change during operation


If a probe is unplugged while the device is running, the current measurement settings are
saved as Last Settings for this probe type. When reconnecting the same probe type, the
same Last Settings will be used. If another probe type is detected instead, the Last Set-
tings for this probe type will be loaded.

54 FieldMan Narda
7.9 SETUPS

7.9.5 Creating and managing setups


Open the setup menu:
✓ The desired probe is connected.
✓ To create a new setup, make the desired settings.
⇒ Open MAIN MENU > SETUPS.
↳ Only the setups created for the currently connected probe type are displayed in al-
phabetical order.

Create a new setup:


⇒ Press SK4.
↳ A new setup is created under the default name Setup [n] with the current settings.
The digit [n] is incremented up automatically.

Rename a setup:
1. Highlight a setup with ▲▼, then press SK2.
↳ The editor opens.
2. Edit the setup title:
– Cross key: Navigate to the character area and select with the central key.
– SK1: Cancel editing
– SK2: Switch upper/lower case letters
– SK3: Delete character to the left of the cursor
– SK4: Accept input and finish

Load a setup:
⇒ Use ▲▼ to highlight a setup and select it with the central key.
↳ The parameters are loaded and the measurement screen is displayed.

Delete a setup:
1. Use ▲▼ to highlight a setup, then press SK1.
2. Confirm deletion with SK4 or cancel with SK1.

Delete multiple setups:


Setups can only be deleted individually in the device.
⇒ To delete multiple setups, remove the SD card and edit the setup directories exter-
nally.
⇒ Alternatively, format the SD card.
ATTENTION! This will cause all stored data to be lost.

Narda FieldMan 55
7 Device settings

7.10 DATA LOGGER


⇒ For a description of the settings for the DATA LOGGER, see 9 DATA LOGGER on
page 91.

56 FieldMan Narda
7.10 DATA LOGGER

8 Operating modes

In this chapter, after an overview of the operating modes,


you will find detailed information on each operating mode,
with the specific settings and the measurement procedures.

8.1 Overview of operating modes and probe types


(page 58)
8.2 FIELD STRENGTH (page 60)
8.3 SPATIAL AVERAGE (page 67)
8.4 TIMER LOGGING (page 71)
8.5 SPECTRUM (page 78)
8.6 Shaped Time Domain (page 83)

Narda FieldMan 57
8 Operating modes

8.1 Overview of operating modes and


probe types
Depending on the connected and recognized probe type, the following operating modes
are available for selection in the FieldMan:

Operating mode (abbre- HF probes HP-01 EHP-50F/G BFD-400-1


viation1)) (DC/LF) (LF) (LF)
FIELD STRENGTH (FS) X X X X
SPATIAL AVERAGE (SA) X X X X
TIMER LOGGING (TL) X X X X
SPECTRUM (SP) – X X X
SHAPED TIME DOMAIN – – X X
(ST)
1) Name displayed on the soft key in connection with the mode symbol

The operating mode is selected and displayed on the device via SK1. SK1 is only
available for selection when a valid probe has been detected.

FIELD STRENGTH (FS)


Measurement of field strength or flux density and numerical representation of the mea-
surement results for a defined frequency range, which may be broadband or band-lim-
ited. A time curve (Time Curve) can be optionally activated for the measurement results,
which is implemented as a rolling memory.
⇒ See 8.2 FIELD STRENGTH on page 60.
SPATIAL AVERAGE (SA)
Allows spatial averaging of measured field strength values from multiple measurement
positions and is mostly required for measuring thermal effects with HF probes. Averaging
is done via the square of the field strength values (RMS averaging). In addition, a delay
can be set for the measurement start. This allows the person performing the measure-
ment to move away from the measuring location to avoid influencing the field.
⇒ See 8.3 SPATIAL AVERAGE on page 67.

58 FieldMan Narda
TIMER LOGGING (TL)
Time-controlled broadband measurement of field strength values in a definable period of
time. With the RMS detector type, all measured values (samples) acquired in a measure-
ment interval are condensed into isotropic min/avg/max values that characterize the
measurement interval. With the peak detector type (BFD-400 only), the acquired mea-
sured values (samples) are compressed to an isotropic max value.
⇒ See 8.4 TIMER LOGGING on page 71.
SPECTRUM (SP)
FFT analysis of the measurement signal in the selected frequency range with display of
the frequency spectrum, frequency marker evaluation and display of the broadband
level. With the % unit selected, a normalized representation of the frequency response is
shown, related to the limit values of a selected standard.
⇒ See 8.5 SPECTRUM on page 78.
SHAPED TIME DOMAIN (ST)
Evaluation method in the time domain for the gapless recording of the weighted peak or
RMS values with display of the exposure index in % related to a selectable personal pro-
tection standard. Whether the peak or RMS detector is used is defined in the measure-
ment method standards IEC/EN 62233 for household appliances (Weighted RMS) or
IEC/EN 62311 for other appliances (Weighted Peak).
⇒ See 8.6 Shaped Time Domain on page 83.

Narda FieldMan 59
8 Operating modes

8.2 FIELD STRENGTH


The FIELD STRENGTH operating mode is used to measure the field strength or flux den-
sity and enables the numerical display of the measurement results for a defined fre-
quency range. This frequency range can be broadband or band-limited. A time curve
(Time Curve) can be activated as a rolling memory for the measurement results.
This is the most commonly used mode for broadband or band-limited measurements.

8.2.1 Measurement settings


The MEASUREMENT SETTINGS menu provides quick access to the most important
measurement settings for the selected operating mode.

Function1) Description
Reset Results Reset measurement results
Zero Offset correction (HP-01 probe only)
⇒ See 10.3 Offset correction – Zero on page 102.
Frequency Span Frequency range
Field Electric or magnetic field
Range Measuring range
Unit Unit
Low Cut Filter Filter for suppression of low frequencies (low cut filter)
Detector Detector type
Time Curve Show/hide time curve display
Time Span Timeline for time curve display
Single Run Automatic stop of the measuring sequence
(only visible if Time Curve = On)
Pulse Detection Short integration time for HF measurement, e.g. for more
accurate measurement of short radar pulses (only visible
with suitable HF probes)
Averaging Time Time for moving averaging
1) The available parameters depend on the selected operating mode and the connected probe
type.

60 FieldMan Narda
8.2 FIELD STRENGTH

8.2.2 Numerical display


Two fields are available for numerical display:
• Main field: Displays current or static measured values depending on the selected
result type and detection type
• Supplementary field: Always displays static values

Detector
Select a detector:
⇒ MEASUREMENT SETTINGS: Select a detector under Detector.
↳ The selection option depends on the probe used:

Detector Description Availability


RMS All measured values are recorded and displayed All probes
as RMS values.
Peak All measured values are recorded and displayed BFD-400-1
as peak values. probe only

Result type
The result types are always displayed in the supplementary field. To display them in the
main field, the result type must be selected in the selection field using the cross key.
Select a result type:
⇒ Press the ◄ ► keys repeatedly until the desired result type is displayed.
↳ The selection option depends on the selected detector:

Detector Result types Display in supple- Evaluation


mentary field
RMS ACTUAL - Current measured value
MINIMUM MIN Smallest value
MAXIMUM MAX Highest value
AVERAGE AVG Averaged value
MAX AVERAGE MAXAVG Highest averaged value
Peak ACTUAL - Current measured value
MINIMUM MIN Smallest peak value
MAXIMUM MAX Highest peak value

Narda FieldMan 61
8 Operating modes

The display in detail


1
2

4
5

Fig. 3: Numerical display in the FIELD STRENGTH operating mode


1 Result type selection field
2 Selected detector (selection in MEASUREMENT SETTINGS)
3 Main field:
– Left: Result type ACTUAL (current isotropic value and three axes)
– Right: Other result type (maximum value and current value)
4 Display status of measurement (e.g. HOLD) and averaging time until the set value
was reached.
5 Supplementary field with statistical values

Result types AVG and MAXAVG


The time window defined by the Averaging Time (see 8.3 SPATIAL AVERAGE on
page 67) for the formation of the moving average for the result types AVG and MAXAVG
shifts continuously. As long as not all measured values are available to fill the time win-
dow, the averaging time is reduced to the actual acquisition time. The progress of the
averaging is displayed in the AVG TIME field and in the progress bar between the main
field and the supplementary field. After the set averaging time has been reached, the sta-
tus display disappears. For MAXAVG, a valid value can only be displayed after the aver-
aging time has elapsed.

Reset the measured values:


⇒ Press and hold key (> 2 s).
– oder –
1. Press SK2.
↳ Reset Results is already selected as the first entry.
2. Press ● key.
↳ The measured values are reset and the menu is closed.

62 FieldMan Narda
8.2 FIELD STRENGTH

8.2.3 Bar display


The bar represents the ACTUAL values with logarithmic scaling. The left and right limits
are specified by the measuring range limits of the connected probe. MIN and MAX are
displayed as marker lines.
Activate the bar display:
To display the bar graph, the time curve graph must be turned off.
⇒ MEASUREMENT SETTINGS: Deactivate Time Curve.
Start and stop the measurement:
⇒ Press SK3.

The display in detail

2 4

Fig. 4: Bar display in the FIELD STRENGTH operating mode


1 Current measured value
2 MIN value
3 MAX value
4 Overall bar scaled to measuring range

Narda FieldMan 63
8 Operating modes

8.2.4 Time curve display


The time curve display shows the isotropic result of the field strength running from right
to left with a rolling memory. The individual axes are not displayed graphically.
Activate the time curve display:
⇒ MEASUREMENT SETTINGS: Activate Time Curve.
↳ Since the curve memory is always active in the FIELD STRENGTH operating
mode (even with Time Curve = Off), the complete curves of ACTUAL and, if ap-
plicable, AVG since the last operating mode change or reset appear immediately.
Start and stop the measurement:
⇒ Press SK3.

The display in detail


1

3
4
2

Fig. 5: Time curve display in the FIELD STRENGTH operating mode


1 Y-axis with automatic scaling
2 X-axis manually adjustable via Time Span
3 Yellow curve: Time curve of the current measured values (ACTUAL)
4 Blue curve: Time curve of the averaged measured values (AVG)

The display of the AVG curve is omitted if the peak detector was selected instead of
the RMS detector for LF probes.

64 FieldMan Narda
8.2 FIELD STRENGTH

Using the marker


The measurement results can be evaluated via the marker.
Switch on the marker:
1. Press SK3 to stop the display (HOLD).
↳ SK1 now shows the marker symbol.
2. Press SK1 to show the marker.
↳ A circle shows the marker position in the diagram (1), the time and measured
value are indicated on the right above the diagram.

3. Use the ◄ ► keys to move the marker.


4. Press SK1 to turn the marker off again.

If the display was stopped via HOLD, the measurement continues in the background.
As soon as the display is restarted, there is therefore continuous recording and
updating of the time curve display.

Using the peak marker


The peak values can be evaluated via the peak marker.
⇒ For information on the peak marker, see 10.7 Peak marker on page 106.

Narda FieldMan 65
8 Operating modes

Stopping the measurement sequence automatically


If required, the measurement sequence can be stopped automatically after a run via the
Single Run function.
Activate automatic stop:
⇒ MEASUREMENT SETTINGS: Activate Single Run.
↳ When activated, the results memory is deleted and a measurement sequence is
performed over the set time period. After the time period is reached, the measure-
ment is stopped completely. The measurement sequence can be restarted at any
time via a reset.

66 FieldMan Narda
8.3 SPATIAL AVERAGE

8.3 SPATIAL AVERAGE


The SPATIAL AVERAGE operating mode enables the spatial averaging of measured
field strength values over 1 to a maximum of 100 measuring positions. It can also be use-
ful to measure at a single measuring position, in order to record the field strength in a
specific spatial volume there, averaged over the desired measuring time, by continuous
movement of the measuring probe.
The spatial averaging is determined by quadratic averaging over all measured positions
and displayed as a horizontal line in the bar graph. The measurement per position can be
stopped automatically after a preset time or manually. An adjustable delay allows the per-
son performing the measurement to leave the measurement site before the measure-
ment starts.

8.3.1 Measurement settings


The MEASUREMENT SETTINGS menu provides quick access to the most important
measurement settings for the selected operating mode.

Function1) Description
Zero Offset correction (HP-01 probe only)
⇒ See 10.3 Offset correction – Zero on page 102.
Frequency Span Frequency range
Field Electric or magnetic field
Range Measuring range
Unit Display unit
Low Cut Filter Filter for suppression of low frequencies (low cut filter)
Detector Detector type
Measurement Time Stop measurements:
• Manual: The measurement must be ended manually.
• 5 s, 10 s...: The measurement stops after the
preselected time.

Narda FieldMan 67
8 Operating modes

Function1) Description
Start Delay Start type of measurement:
• No delay: The measurement starts immediately
• 5 s, 10 s...: The measurement starts after the selected
time. A countdown shows the time until the start.
Pulse Detection Short integration time for HF measurements, e.g. for more
accurate measurement of short radar pulses (only visible
with suitable HF probes).
1) The available parameters depend on the operating mode and the probe type.

8.3.2 Measurement
✓ Automatic or manual stop was selected.
✓ Immediate or delayed start was selected.
1. Press SK3.
↳ The measurement starts immediately or after the selected time.
2. Press SK3 to stop the measurement manually or wait for the preselected time to
elapse. Even with automatic stop, a measurement in progress can be stopped
manually.
↳ After the measurement has been completed, a selection window appears:
– Next Position: Start the measurement at the next position
– Repeat Position: Start a new measurement at the same position
– Cancel All: Delete all measurements. The deletion must be confirmed again.
– SAVE AND FINISH: Save measurements and end measurement.
3. Select an entry. Cancel All must be confirmed again.
↳ The function is executed immediately and the selection window is closed.

Set start with or without delay:


1. MEASUREMENT SETTINGS: Select Start Delay.
2. Select the desired setting and press the ● key:
– No delay: The measurement starts immediately
– Time: The measurement starts after the selected time. A countdown shows the
time until the start.
3. Exit the menu with the Back key.

68 FieldMan Narda
8.3 SPATIAL AVERAGE

8.3.3 Numerical display


Two fields are available for numerical display:
• Main field: Displays the current measured values isotropically and for the three axes
• Supplementary field: Displays the averaged values of the last measured position as
well as over all positions

Detector
In the SPATIAL AVERAGE operating mode, RMS is always used.

Result type
In the SPATIAL AVERAGE operating mode, ACTUAL is always used.

The display in detail

4
5

Fig. 6: Numerical display in the SPATIAL AVERAGE operating mode


1 Result type always ACTUAL, detector always RMS
2 Current isotropic measured value
3 Current measured values of the three axes
4 Number of the position and current time over which averaging is currently being
performed
5 Evaluations:
– AVG: Average value of the displayed position
– SPATIAL AVG: Spatial average of all measured positions

Narda FieldMan 69
8 Operating modes

8.3.4 Bar display


In the vertical bar graph, the successively measured values for each measuring position
are displayed from left to right. The spatial average is shown as a horizontal line. To eval-
uate all positions, you can scroll to the left and right with the arrow keys after the mea-
surement has been completed.

The display in detail

Fig. 7: Bar display in the SPATIAL AVERAGE operating mode


1 Average value of the respective position
2 Spatial average over all measured positions

Evaluating the bar display


✓ The measurement is completed (SAVE AND FINISH).
⇒ Select a measuring position with the ◄ ► keys.
↳ In the bar display: The selected bar is highlighted brighter.
↳ In the supplementary field: The position as well as the associated measured value
(AVG) is displayed. Up to 13 bars can be displayed simultaneously in the graph.
If there are more than 13 positions, the bar display can be scrolled using the ◄ ►
keys.

70 FieldMan Narda
8.4 TIMER LOGGING

8.4 TIMER LOGGING


The TIMER LOGGING operating mode enables time-controlled measurement of field
strength values in a definable time period of up to 100 hours. During recording, the prog-
ress is displayed via a time bar. After the measurement has been completed, the saved
result can be viewed as a time curve display in the data logger (see 9 DATA LOGGER on
page 91).

8.4.1 Measurement settings


The MEASUREMENT SETTINGS menu provides quick access to the most important
measurement settings for the selected operating mode.

Function1) Description
Zero Offset correction (HP-01 probe only)
⇒ See 10.3 Offset correction – Zero on page 102.
Start Time Timer start time
Timer Duration Timer runtime
Storage Interval Storage interval
Frequency Span Frequency range
Field Electric or magnetic field
Range Measuring range
Unit Display unit
Low Cut Filter Filter for suppression of low frequencies (low cut filter)
Detector Detector type (only available with BFD-400-1 probe)
Pulse Detection Short integration time for HF measurements, e.g. for more
accurate measurement of short radar pulses (only visible
with suitable HF probes)
Averaging Time Averaging time
1) The available parameters depend on the operating mode and the probe type.

A maximum of 32,000 storage intervals can be recorded. If a short storage interval is


set with a long runtime, one of the values is automatically corrected. Example: With a
storage interval of 1 s, the maximum possible runtime is 8 h 53 min 20 s.

Narda FieldMan 71
8 Operating modes

8.4.2 Sequence of a time-controlled measurement


1. Select timer settings
• Start Time
• Timer Duration
• Storage Interval
⇒ See 8.4.1 Measurement settings on page 71.

2. Activate comment functions (if required)


• Add Text Comment
• Add Voice Comment
⇒ See 7.3 GENERAL SETTINGS on page 45.

3. Activate timer
⇒ Press SK3.
↳ When the comment function is activated, the user is prompted for a voice com-
ment and/or text input.
↳ The device waits until the specified start time for the measurement.
↳ The setting for automatic shutdown (Automatic Shutdown) is ignored and the
device remains switched on (see 7.3.5 DEVICE on page 47).

The following actions are possible during this time:


Start recording immediately:
⇒ Press SK3.
Deactivate timer again:
⇒ Press SK1.
↳ In this case, previously recorded voice or text comments will be deleted.

72 FieldMan Narda
8.4 TIMER LOGGING

Activate stand-by until recording starts:


⇒ Press SK2.
↳ The device is put into a power-saving sleep mode (stand-by), wakes up again at
the specified time, and then performs the measurements. A certain lead time is re-
quired to reactivate the connected probes.
⇒ During standby, the device can be woken up at any time by pressing any key and will
be in the same state as before standby.

This function is only available for long periods of 5 minutes or more and for probes with
electrical interface.

4. Timer is started
• When recording is started, the previously recorded measurement data is reset.
• Then the measurement data are stored cyclically until the measuring duration (Timer
Duration) is reached.
The following actions are possible during this time:
End measurement prematurely:
⇒ Press SK3.
↳ Even if the measurement is terminated prematurely, all intervals measured up to
that point are saved.
↳ Afterwards, the timer is automatically deactivated again.

The rechargeable battery life can be extended by activating the screen saver. This
makes longer measuring times possible.
⇒ See 7.3.1 DISPLAY on page 45.

5. After the measurement


The device switches off automatically after the set measuring time (Timer Duration) has
elapsed (regardless of the setting for Automatic Shutdown).
Automatic shutdown does not take place in the following cases:
• The user has ended the measurement prematurely.
• A key was pressed directly after the measuring period (Timer Duration) elapsed.

Narda FieldMan 73
8 Operating modes

8.4.3 Numerical display


Two fields are available for numerical display:
• Main field: Displays current or static measured values depending on the selected
result type and detection type
• Supplementary field: Always displays static values

Detector
A selection is only available for the BFD-400-1 probe. For all other probes, RMS is spec-
ified.
Select a detector:
⇒ MEASUREMENT SETTINGS: Select a detector under Detector.

Detector Description Availability


RMS All measured values are recorded and displayed All probes
as RMS values.
Peak All measured values are recorded and displayed BFD-400-1
as peak values. probe only

Result type
The result types are always displayed in the supplementary field. To display them in the
main field, the result type must be selected in the selection field using the cross key.
Select a result type:
⇒ Press the ◄ ► keys repeatedly until the desired result type is displayed.
↳ The selection option depends on the selected detector:

Detector Selection via Main field Supplementary


cross key field
Top Bottom
RMS ACTUAL ACTUAL XYZ MAX, MIN,
(all probes) MINIMUM MIN ACTUAL MAXAVG, AVG

MAXIMUM MAX ACTUAL


AVERAGE AVG ACTUAL
MAX AVERAGE MAXAVG ACTUAL

74 FieldMan Narda
8.4 TIMER LOGGING

Detector Selection via Main field Supplementary


cross key field
Top Bottom
Peak ACTUAL ACTUAL XYZ MAX, MIN
(BFD-400-1
only) MINIMUM MIN ACTUAL
MAXIMUM MAX ACTUAL

The display in detail


1
2

Fig. 8: Numerical display in the TIMER LOGGING operating mode


1 Result type selection field
2 Selected detector (selection in MEASUREMENT SETTINGS)
3 Main field:
– Left: Result type ACTUAL.
The current isotropic value and the three axes are displayed.
– Right: Other result type, here as an example AVERAGE.
The selected result type is displayed and ACTUAL is always displayed as well
4 Supplementary field

The measured values are always displayed, regardless of whether data is being
recorded.

Narda FieldMan 75
8 Operating modes

8.4.4 Bar display


The bar display graphically shows the progress of the measurement. In addition, the start
time, timer duration and storage interval are specified numerically.

1
2
3

Fig. 9: Progress display via time bar: Activated timer (left), after start (right)

No Activated timer Timer started


.
1 Animated symbol for activated timer Animated symbol for remaining time
2 Start time of recording Remaining time
3 Time bar of recording duration
4 Recording duration and storage interval
5 Softkey for cancel, stand-by and start Softkey for stop

8.4.5 Time curve display in the DATA LOGGER


When viewing saved results via the DATA LOGGER, the results are displayed graphi-
cally as a progression over time. In the process, the up to 32,000 measurement intervals
are condensed to a maximum of 240 measurement intervals to enable display and eval-
uation with the Marker and Peak Marker. In case of later data evaluation on the PC,
detailed evaluation in full time resolution is possible.
The time axis displayed on the device is labeled with the total measurement duration in
the format hh:mm:ss. When the marker is switched on, the maximum value that occurred
in the (condensed) measurement interval is displayed as ACT value together with the
time of the interval end.

76 FieldMan Narda
8.4 TIMER LOGGING

The display in detail


⇒ To call up the time curve display in the DATA LOGGER, see 9 DATA LOGGER on
page 91.

1 2

3
4

Fig. 10: Graphic display in the TIMER LOGGING operating mode after calling it up in the DATA
LOGGER
1 Unit
2 Time and ACT measured value with marker switched on
3 Graphical display of the current values
4 Measuring duration in the format hh:mm:ss

Using the marker


1. Press SK1 to turn on the marker.
↳ A circle shows the marker position in the diagram, the time and measured value
are indicated on the right above the diagram.
2. Use the ◄ ► keys to move the marker.
3. Press SK1 to turn the marker off again.

Using the peak marker


The peak values can be evaluated via the peak marker.
⇒ For information on the peak marker, see 10.7 Peak marker on page 106.

Narda FieldMan 77
8 Operating modes

8.5 SPECTRUM
The SPECTRUM operating mode enables the FFT analysis of the measurement signal
in the selected frequency range with display of the frequency spectrum, evaluation via a
frequency marker and display of the broadband level. Either one of the single axes X, Y
or Z or Isotropic can be selected as the measuring axis. In addition, a selectable limit
value curve can be displayed in the spectrum. When the unit % is selected, the spectrum
is normalized to the field strength limits of a personal protection standard.

8.5.1 Measurement settings


The MEASUREMENT SETTINGS menu provides quick access to the most important
measurement settings for the selected operating mode.

Function1) Description
Reset Results Reset measurement results
Zero Offset correction (HP01 probe only)
⇒ See 10.3 Offset correction – Zero on page 102.
Result Type Result type (see Result type on page 79)
Frequency Span Frequency range
Frequency Scale Linear or logarithmic scaling
Field Electric or magnetic field
FFT Points Determines the frequency resolution of the FFT analysis
(number of frequency values = FFT points / 2)
Range Measuring range
Axis Spatial axis of the probe shown in the graph
Unit Display unit
Low Cut Filter Filter for suppression of low frequencies (low cut filter)
Detector Detector type: RMS, Peak
Standard Displays a selection list of the standards according to
whose limits normalization is performed or the limit pro-
gression is displayed in the spectrum (see
8.5 SPECTRUM on page 78).
Number of Averages Number of spectra over which averaging is performed
1) The available parameters depend on the selected operating mode and the connected probe
type.

78 FieldMan Narda
8.5 SPECTRUM

8.5.2 Measuring and evaluating a spectrum


The measurement starts immediately after selecting the operating mode.
The ◄ ► keys can be used to move and evaluate the marker at any time during the mea-
surement. However, to use the peak marker, the display must be stopped.
Stop and start measurement:
⇒ Press SK3.
In the graphical display, the measured values can be evaluated using the Peak Marker
function (see 8.5.4 Graphic display on page 81).

8.5.3 Numerical display


Two fields are available for numerical display:
• Main field: Displays current or statistical broadband values depending on the
selected result type and detection type
• Supplementary field: Always displays statistical broadband values

Detector
A selection is only available for the BFD-400-1 probe. For all other probes, RMS is spec-
ified.
Select a detector:
⇒ MEASUREMENT SETTINGS: Select a detector under Detector.

Detector Description Availability


RMS All measured values are recorded and displayed All probes
as RMS values.
Peak All measured values are recorded and displayed BFD-400-1
as peak values. probe only

Result type
In SPECTRUM mode, the control pad is used for the marker. The selection of the result
type is therefore made in the measurement settings.
Select a result type:
⇒ MEASUREMENT SETTINGS: Select the desired result type under Result Type.
↳ The selection option depends on the detector and the probe.

Narda FieldMan 79
8 Operating modes

When the result type is selected in the measurement settings, the result type displayed
in the graphical display also switches over automatically.

Detector Selection via mea- Main field Supplementary


surement settings field
Top Bottom
RMS ACTUAL ACTUAL XYZ MAX, AVG
(no MAXIMUM MAX ACTUAL
HF probes)
AVERAGE AVG ACTUAL
Peak ACTUAL ACTUAL XYZ MAX
(BFD-400-1 MAXIMUM MAX ACTUAL
only)

The display in detail


1
2

Fig. 11: SPECTRUM operating mode: Numerical display of the broadband values (left) and the
highest spectral line in % of the selected standard (right).

1 Selected result type (selection in MEASUREMENT SETTINGS)


2 Selected detector (selection in MEASUREMENT SETTINGS)
3 Main field: Result type ACTUAL
– Left: Current isotropic value and the three axes
– Right: Current isotropic value and the three axes for the highest line of the
normalized spectrum with frequency indicated

Reset the measured values:


⇒ MEASUREMENT SETTINGS > Reset Results > ● key.
↳ The measured values are reset and the menu is closed.

80 FieldMan Narda
8.5 SPECTRUM

8.5.4 Graphic display


Properties of the graphical representation:
X-axis
• The entire frequency range up to the set span is always displayed (MEASUREMENT
SETTINGS > Frequency Span).
• The frequency scaling can be set to either linear from 0 Hz or logarithmic
(MEASUREMENT SETTINGS > Frequency Scale).
Y-axis
• The field strength is displayed on a logarithmic scale.

Measuring curve
• Representation: Maximum of one measuring curve (X, Y, Z axis or isotropic)
• If standard is selected: The associated limit value curve is additionally displayed.
The field strength axis is scaled automatically. Some standards are only defined from
higher frequencies and can therefore only be displayed partially or not at all.
• % unit selected: Spectrum is displayed normalized, related to the selected standard.
• If no standard is selected (Standard = None): A message prompts to select a
standard.

Narda FieldMan 81
8 Operating modes

The display in detail

1 4
2 5

3 6

Fig. 12: Graphical display of the spectrum


1 Type of measuring curve (X, Y, Z-axis or isotropic)
2 Selected standard
3 Marker position
4 Marker functions (see Using the marker on page 82)
5 Values at the marker position
6 Limit curve of the selected standard
7 Spectrum

Using the marker


⇒ Use the ◄ ► keys to move the marker to the desired position on the measuring
curve.

Using the peak marker


The highest spectral lines can be evaluated via the peak marker.
✓ The peak marker can only be selected after the display has been stopped.
⇒ For information on the peak marker, see 10.7 Peak marker on page 106.

82 FieldMan Narda
8.6 Shaped Time Domain

8.6 Shaped Time Domain


In the SHAPED TIME DOMAIN operating mode, an evaluation method is used in the
time domain, which ensures gapless recording of the weighted peak or RMS values
related to a selectable personal protection standard. The result is displayed as an expo-
sure index in %.
The applied measurement method is described in international standards and is gener-
ally referred to as Weighted Peak Method or Weighted RMS Method respectively.
In addition to the numerical result display, a logarithmically scaled bar display shows the
exposure index graphically over the entire measurement range of the probe. Instead of
the bar display, a time curve graph can also be displayed.

The SHAPED TIME DOMAIN operating mode is only available for the EHP-50F/G and
BFD-400-1 probes.

8.6.1 Measurement settings


The MEASUREMENT SETTINGS menu provides quick access to the most important
measurement settings for the selected operating mode.

Function1) Description
Reset Results Reset measurement results
Field Electric or magnetic field
Range Measuring range
Low Cut Filter Selection of the lower cutoff frequency, e.g. to reduce the
(BFD-400-1 only) influence of the earth's magnetic field
Method Method for weighting:
• WPM-TD (IEC/EN 62311): Weighted Peak Method
(fixed for the EHP-50F/G probe)
• WRM-TD (IEC/EN 62233): Weighted RMS Method
Weighting Displays a selection list of standards according to whose
limits the evaluation is performed.
Time Curve Graphic display:
• Bar display: Time Curve = Off
• Time curve graph: Time Curve = On

Narda FieldMan 83
8 Operating modes

Function1) Description
Time Span Time domain (X axis)
Single Run Measurement ends after set Time Span
1) The available parameters depend on the selected operating mode and the connected probe
type.

8.6.2 Numerical display


Two fields are available for numerical display:
• Main field: Displays current or static measured values, depending on the selected
result type and detection type, as well as the selected evaluation standard
• Supplementary field: Always displays static values

Detector
The Peak or RMS detector to be used is determined by the selected measurement
method (see 8.6.1 Measurement settings on page 83, Method).
Evaluation curve
The limits to be used for calculating the exposure index are determined by selecting a
standard (see 8.6.1 Measurement settings on page 83, Weighting).

Result type
The result types are always displayed in the supplementary field. To display them in the
main field, the result type must be selected in the selection field using the cross key.
Select a result type:
⇒ Press the ◄ ► keys repeatedly until the desired result type is displayed.

Detector1) Selection via Main field Supplementary


cross key field
Top Bottom
RMS + ACTUAL ACTUAL XYZ MAX, MIN
Peak MINIMUM MIN ACTUAL
MAXIMUM MAX ACTUAL
1) In this mode of operation, the EHP-50F/G probe has only the Peak detector type,
on the EHP-50G the WPM option must be enabled.

84 FieldMan Narda
8.6 Shaped Time Domain

The display in detail


1
2
3

Fig. 13: Numerical display in the SHAPED TIME DOMAIN operating mode
1 Selected result type (selection via cross key)
2 Selected detector (selection in MEASUREMENT SETTINGS)
3 Selected evaluation standard
4 Main field:
– Left: Result type ACTUAL.
The current isotropic value and the three axes are displayed.
– Right: Other result type, here as an example MAXIMUM.
The selected result type is displayed and ACTUAL is always displayed as well
5 Supplementary field:
– Statistical values MIN and MAX

Narda FieldMan 85
8 Operating modes

8.6.3 Bar display


The bar represents the ACTUAL values with logarithmic scaling. The left and right limits
are specified by the measuring range limits of the connected probe. MIN and MAX are
displayed as marker lines.
Activate the bar display:
To display the bar graph, the time curve graph must be turned off.
⇒ MEASUREMENT SETTINGS: Select Time Curve = Off.

The display in detail


5

2
1

Fig. 14: Bar display in the SHAPED TIME DOMAIN operating mode
1 MIN value numerical and graphical
2 MAX value numerical and graphical
3 Current measured value (ACTUAL)
4 Overall bar scaled to measuring range
5 Measurement method and weighting standard

86 FieldMan Narda
8.6 Shaped Time Domain

8.6.4 Time curve display


The time curve graph represents the isotropic result of the exposure index running from
right to left with a rolling memory.
Activate the time curve display:
⇒ MEASUREMENT SETTINGS: Activate Time Curve.
↳ Since the curve memory is always active in the SHAPED TIME DOMAIN operat-
ing mode (even with Time Curve = Off), the complete curve progression since the
last operating mode change or reset appears immediately.

1
3

Fig. 15: Time curve display in the SHAPED TIME DOMAIN operating mode
1 Selected standard
2 Measured values in % of standard
3 Result type
4 Time curve of the measurement
5 Time axis

Narda FieldMan 87
8 Operating modes

Using the marker


✓ The display is in HOLD mode.
1. Press SK1 to turn on the marker.
2. Use the ◄ ► keys to move the marker to the desired position on the measuring
curve.
3. Press SK1 to turn the default marker off again.

2
3
4

1 5

Fig. 16: Marker in the SHAPED TIME DOMAIN operating mode


1 SK1: Switch marker on/off
2 Movement symbols for standard markers
3 Measured values at the marker position
4 Marker position
5 SK4: Switch peak marker on/off

If the display was stopped via Hold, the measurement continues in the background. As
soon as the display is restarted, there is therefore continuous recording and updating
of the time curve display.

Using the peak marker


The peak values can be evaluated via the peak marker.
⇒ For information on the peak marker, see 10.7 Peak marker on page 106.

88 FieldMan Narda
8.6 Shaped Time Domain

Stopping the measurement sequence automatically


If required, the measurement sequence can be stopped automatically after a run via the
Single Run function.
Activate automatic stop:
⇒ MEASUREMENT SETTINGS: Activate Single Run.
↳ When activated, the results memory is deleted and a measurement sequence is
performed over the set time period. After the time period is reached, the measure-
ment is stopped completely. The measurement sequence can be restarted at any
time via a reset.

Narda FieldMan 89
8 Operating modes

90 FieldMan Narda
9 DATA LOGGER

The data memory manages the stored measurement


results, screenshots and media (photos and videos).
Selected data can be displayed and evaluated with the Data
Viewer. All acquired data is stored on the SD card of the
device, organized by projects.

9.1 Structure of the DATA LOGGER (page 92)


9.2 Managing projects (PROJECT) (page 93)
9.3 Managing measurements (MEASUREMENTS) (page 94)
9.4 Managing screenshots (SCREENSHOTS) (page 96)
9.5 Managing media (MEDIA) (page 96)

Narda FieldMan 91
9 DATA LOGGER

9.1 Structure of the DATA LOGGER


All content (measurements, screenshots, media) is managed in projects. Up to 99 proj-
ects can be created, edited and deleted. To call up the saved content, first select the
desired project. Then the associated content can be selected.
Open the data memory:
✓ The device is in the measuring mode of any operating mode.
⇒ Open the device settings with SK4, then select DATA LOGGER.
↳ The main menu opens.

1
2
3
4

5
6

Fig. 17: Main menu DATA LOGGER


1 Selected project
2 Number of saved measurements in the selected project
3 Number of saved screenshots in the selected project
4 Number of stored media in the selected project
5 Selection to display the project at startup
6 Free storage space

92 FieldMan Narda
9.2 Managing projects (PROJECT)

9.2 Managing projects (PROJECT)


⇒ Select PROJECT in the main menu.
↳ The submenu opens.

9.2.1 Creating a new project


⇒ Press SK4.
↳ The project is created and named Project n. n is incremented up continuously (to
99).

9.2.2 Renaming a project


1. Select the project and press SK2.
2. Change the project title in the editor.
To do this, use the ◄ ► and ▲▼ keys to select the desired characters in the
character area and accept them with the ● key.

Move the cursor in the title window:


Move one character to the left in the title window
Move one character to the right in the title window

Functions of the keys SK1 to SK4:


Cancel input
Toggle upper/lower case
Delete last character
Save changes

9.2.3 Deleting a project


1. Select the project and press SK1.
2. Confirm the deletion.

Narda FieldMan 93
9 DATA LOGGER

9.3 Managing measurements


(MEASUREMENTS)
In the MEASUREMENTS menu, the measurements stored in a project can be called up
and evaluated, supplemented with comments or deleted.
Select a measurement:
✓ The desired project has been selected
(see 9.2 Managing projects (PROJECT) on page 93).
1. Mark MEASUREMENTS and select it with the ● or ► key.
↳ The submenu opens.
2. Mark the desired measurement.

You now have the following options:


• Open and evaluate the measurement
• Add a text comment
• Add a voice comment
• Delete the measurement

9.3.1 Open and evaluate a measurement


⇒ Open the marked measurement with the ● or ► key.
↳ The measurement is displayed.
Additional functions are available via the keys SK1 to SK4:
Turn on the marker
Change measurement settings
(limited selection, e.g. change the measurement unit)
Show previous measurement of the selected project
Show next measurement of the selected project

Using the marker


1. Press SK1 to turn on the marker.
2. Use the ◄ ► keys to move the marker to the desired position on the measuring
curve.
3. Press SK1 to turn the default marker off again.
⇒ For more information on the marker functions, see the descriptions in the operating
modes.

94 FieldMan Narda
9.3 Managing measurements (MEASUREMENTS)

Using the peak marker


The peak values can be evaluated via the peak marker.
⇒ For information on the peak marker, see 10.7 Peak marker on page 106.

9.3.2 Adding a text comment


✓ The measurement is marked in the overview page (MEASUREMENTS).
1. Press SK3.
↳ The editor for entering a comment opens.
2. Use the ◄ ► and ▲▼ keys to select the desired characters in the character area and
accept them with the ● key.
3. Press SK4 to save the comment.

Move the cursor in the title window:


Move one character to the left in the title window
Move one character to the right in the title window

Functions of the keys SK1 to SK4:


Cancel input
Toggle upper/lower case
Delete last character
Save changes

9.3.3 Adding a voice comment


✓ The measurement is marked in the overview page (MEASUREMENTS).
1. Press SK4.
↳ The recording starts immediately.
2. Press SK4 to stop the recording.
↳ The recording is played back.
3. After that, the following functions of the keys SK1 to SK4 are available:
Delete recording
Repeat recording

Narda FieldMan 95
9 DATA LOGGER

Play recording
Save recording
4. Press SK4 to save the recording.
↳ In the overview page (MEASUREMENTS), the symbol for the voice comment
appears behind the selected measurement.

9.3.4 Deleting a measurement


✓ The measurement is marked in the overview page (MEASUREMENTS).
⇒ Press SK1 and confirm the deletion.
↳ The measurement is deleted.

9.4 Managing screenshots


(SCREENSHOTS)
In the SCREENSHOTS menu, the screenshots stored in a project can be called up and
deleted.
Select a screenshot:
✓ The desired project has been selected
(see 9.2 Managing projects (PROJECT) on page 93).
1. Mark SCREENSHOTS and select it with the ● or ► key.
↳ The submenu opens.
2. Mark the desired screenshot.
Show the marked screenshot:
⇒ Press the ● or ► key to open the marked screenshot.
⇒ SK3 key: Show previous screenshot
⇒ SK4 key: Show next screenshot
⇒ Press the Back key to return to the menu.
Delete the marked screenshot:
1. Press SK1.
2. Confirm deletion with SK4 or cancel with SK1.

9.5 Managing media (MEDIA)


In conjunction with a smartphone and the associated app, images and videos can be
stored and accessed in the media area. This function is not available at the moment.

96 FieldMan Narda
9.5 Managing media (MEDIA)

10 Cross-mode functions

This chapter describes functions and settings that affect


several operating modes or are not mode-specific.

10.1 Post-Processing (page 98)


10.2 Selection of a standard (page 100)
10.3 Offset correction – Zero (page 102)
10.4 Commenting on measurement results (page 103)
10.5 Alarm function (page 104)
10.6 Acoustic intensity display (HF probes only) (page 105)
10.7 Peak marker (page 106)
10.8 Display of overdriven measured values (page 107)

Narda FieldMan 97
10 Cross-mode functions

10.1 Post-Processing
⇒ See 7.5.1 POST-PROCESSING on page 47.
This menu contains settings for post-processing the measurement data of broadband HF
probes. These setting options are thus limited to the operating modes of FIELD
STRENGTH, SPECTRUM and TIMER LOGGING. Two types of post-processing are
available:
• The application of a correction factor
• The conversion to display in % of the standard for measured field strengths at a
reference frequency
For evaluating probes, only the correction factor can be applied.

10.1.1 Applying a correction factor


⇒ See 7.5.1 POST-PROCESSING on page 47: Apply Correction Factor CF
When calibrating the HF probes, the measuring error at the defined calibration frequen-
cies is recorded and stored in the probe. If the correction is applied (Apply Correction
Factor CF = On), then the measuring error is taken into account in the measurement
result and thus a higher measurement accuracy is achieved. However, the correction can
only be applied meaningfully when measuring at a single prevailing frequency, which
must also be known. If this frequency does not exactly correspond to a calibration fre-
quency, it can still be entered as a reference frequency (Reference Frequency) and the
corresponding correction factor is then determined by interpolation from the adjacent cor-
rection factors.

10.1.2 Setting the reference frequency


⇒ See 7.5.1 POST-PROCESSING on page 47: Reference Frequency
Reference Frequency provides a selection list of probe-specific calibration frequencies.
In addition, the input of any frequency is possible with a resolution of up to 1 kHz.
The last set value in MHz is always displayed, even if the frequency no longer matches
the probe after a probe change. Only when the calibration factor or the display in % of the
standard are activated is the frequency list updated by reading from the probe data and
set to the lowest value.

98 FieldMan Narda
Set the reference frequency:
1. Select Reference Frequency.
↳ A list opens with the calibration frequencies stored in the probe. The last selected
value is automatically highlighted.
2. If necessary, select another value from the list.
In addition to the values from the list, any intermediate value can be set with a resolution
of 1 kHz.

Set any value:


1. After selecting Reference Frequency, press the SK1 softkey.
2. Set the digit and value with the cross key.
3. Press SK4 to transfer the new value to the selection list.
↳ The pen symbol indicates a frequency where calibration has not been performed
and where the calibration factor is determined by interpolation. There can be only
one entry with pen symbol at a time. This remains in the list until another fre-
quency value is selected and is then deleted again.

HF probes are broadband probes and cannot measure selectively at the set frequency.
In the case of incorrect frequency specification or multifrequency fields, these settings
can therefore lead to incorrect results.

10.1.3 Display in % of standard


Non-evaluating probes measure the field strength. The % of Standard unit is only avail-
able for selection for these probes if the conversion for a specific personal protection
standard has been activated in the menu. The measured value display in % is related to
the power density and not to the field strength.
Activate conversion:
✓ DEVICE SETTINGS > EXPERT SETTINGS
1. Select Standard.
2. Reference frequency of the field (Reference Frequency) indicated.
3. Activate % of Standard: Select Lowest Limit or Reference Frequency.

The reference to the lowest limit value of a standard (Lowest Limit) is recommended
when the frequency of the field source is not known.

The Lowest Limit setting is not available if the use of the correction factor is switched
on.

Narda FieldMan 99
10 Cross-mode functions

Reference Frequency Correction Factor Lowest Limit

The reference frequency If the correction factor is ap- When referring to the small-
(Reference Frequency) is plied at the same time, the est (strictest) limit value,
displayed in the probe info entry CF: 2.45 GHz ap- REF: Lowest Limit is dis-
field, e.g. with REF: 2.45 pears instead. played.
GHz.
The standard selected in the menu is also shown in the display (in the yellow field above
the isotropic measured value).

10.2 Selection of a standard


There is a wide range of national and international standards setting exposure limits for
electric and magnetic fields. They can differ in frequency range and range of effect (influ-
ence on the body).
The standards (limit value curves) to be available for selection on the FieldMan can be
configured via the Narda-TSX PC software. Predefined standards are available, which
can be extended by self-created standards.
User-defined standards are saved on the SD card during configuration and are marked
on the device with a preceding asterisk. Predefined standards, on the other hand, are
stored in the FieldMan.

The standards for evaluative measurements of the SHAPED TIME DOMAIN operating
mode are implemented in the measuring probes and can only be changed by a firmware
update of the manufacturer.

100 FieldMan Narda


10.2 Selection of a standard

10.2.1 Standards for HF probes


The HF probes of the FieldMan measure broadband frequencies and cannot be set to a
specific frequency. The measurement result can be displayed as the percentage of the
limit value that has been reached if the frequency of a field source is known or in relation
to the lowest limit value above the frequency of a standard. The settings are made in the
menu under EXPERT SETTINGS.
A selection of predefined standards appears below. The selection options in the Field-
Man may differ from this and depend on the device configuration.

Standard1) Frequency range


2013/35/EU LAL 1 Hz to 10 MHz
2013/35/EU HAL 1 Hz to 10 MHz
2013/35/EU Limbs 1 Hz to 10 MHz
2013/35/EU Thermal 100 kHz to 300 GHz
26. BImSchV 2013 (1a) 1 Hz to 10 MHz
26. BImSchV 2013 (1b) 100 kHz to 300 GHz
EMFV 2016 LAL 1 Hz to 10 MHz
EMFV 2016 HAL 1 Hz to 10 MHz
EMFV 2016 Limbs 1 Hz to 10 MHz
EMFV 2016 Thermal 100 kHz to 300 GHz
FCC 96-326 GP 300 kHz to 100 GHz
FCC 96-326 Occ 300 kHz to 100 GHz
GB8702-2014 GP 1 Hz to 300 GHz
ICNIRP 1998 GP 1 Hz to 300 GHz
ICNIRP 1998 Occ 1 Hz to 300 GHz
ICNIRP 2010 GP 1 Hz to 10 MHz
ICNIRP 2010 Occ 1 Hz to 10 MHz
ICNIRP 2020 GP 100 kHz to 300 GHz
ICNIRP 2020 Occ 100 kHz to 300 GHz
IEEE 2019 Unrest NS 1 Hz to 5 MHz
IEEE 2019 Restrd NS 1 Hz to 5 MHz
IEEE 2019 Unrestricted 100 kHz to 300 GHz
IEEE 2019 Restricted 100 kHz to 300 GHz

Narda FieldMan 101


10 Cross-mode functions

Standard1) Frequency range


SC6-2015 Unctrld NS 3 kHz to 10 MHz
SC6-2015 Ctrld NS 3 kHz to 10 MHz
SC6-2015 Uncontrolled 0.1(1) MHz to 300 GHz
SC6-2015 Controlled 0.1(1) MHz to 300 GHz
1) LAL: Low, ALs; HAL: High ALs; Limbs: Limbs ALs; Thermal: ALs Thermal; GP: General Public;
Occ: Occupational; Unrest: Unrestricted; Restrd: Restricted; Unctrld: Uncontrolled; Ctrld:
Controlled

10.2.2 Standards for LF probes


For the LF probes, the standard is used to display the limit curve graphically in the spec-
trum or to display the measurement result as a normalized spectrum. The selection is
made under MEASUREMENT SETTINGS > Standard. The same table applies as for
HF probes. The selection None is used to switch off the display of a limit value curve in
the SPECTRUM operating mode.

10.3 Offset correction – Zero


⇒ See measurement settings in the operating modes.
Manual offset correction is only required for the HP-01 probe. The correction process is
started by triggering Zero in the measurement settings and is recommended before mea-
suring small static fields. For offset correction of the HP-01, the sensor tip must be placed
in a zero field chamber before the correction process is started, in order to exclude the
influence of external fields and especially that of the earth's magnetic field.
Perform the offset correction:
1. Press SK2, then select Zero.
↳ A dialog box opens.
2. Follow the instructions in the dialog box.
3. Continue with SK4 or cancel with SK1.
↳ The progress and completion of the offset correction are displayed.
4. Follow the instructions in the dialog box.
5. Press the Back key to return to the measuring mode.

102 FieldMan Narda


10.4 Commenting on measurement results

10.4 Commenting on measurement results


✓ DEVICE SETTINGS > GENERAL SETTINGS
⇒ See 7.3.2 SAVE OPTIONS on page 46.
In this menu you can set whether a comment is to be recorded with the saving of mea-
surement results via voice or text input. Both functions can also be activated.
If the respective type of comment is activated and a measurement result (not a screen-
shot) was saved via the save key, the menu navigation for the voice comment opens first
and then the menu navigation for the text input via the editor. In the TIMER LOGGING
operating mode, the prompt for entering a voice or text comment already appears after
pressing Enable Timer, i.e. before the actual start of measurement.

Text comments and voice comments can also be added or changed at any time later
in DATA LOGGER.

10.4.1 Activating voice comments


Activate voice comments:
1. Add Voice Comment = On
2. Return to the measuring interface.
Enter a voice comment:
⇒ See 9.3.3 Adding a voice comment on page 95.

10.4.2 Activating text comments


For simplified text input, a predefined text (Default Text) can be used and added to
during input.
Activate text comments:
1. Add Text Comment = On
2. Edit Default Text if necessary.
3. Return to the measuring interface.
Enter a text comment:
⇒ See 9.3.2 Adding a text comment on page 95.

Narda FieldMan 103


10 Cross-mode functions

10.5 Alarm function


⇒ See 7.3.4 ALARM on page 46: Alarm Function
In this menu, alarm thresholds can be set and the alarm function can be activated. After
activating the alarm function, an alarm symbol appears in the status bar of the display.
When the alarm threshold is exceeded, a conspicuous acoustic signal sounds with a
visual alarm message. If the mute function is activated (Mute = On), only the visual alarm
is available.

Fig. 18: Display of alarm function

104 FieldMan Narda


10.6 Acoustic intensity display (HF probes only)

10.6 Acoustic intensity display (HF probes


only)
⇒ See 7.3 GENERAL SETTINGS on page 45: Audible Field Indicator
In this menu, the acoustic display can be activated and the sensitivity can be adjusted by
selecting the gain. In principle, the audio frequency increases as the field strength level
increases in order to be able to detect field maxima or leaks on HF lines.

Setting Scope of application


Off Inoperative
High Gain Up to 20 V/m or 20%
Normal Gain Up to 80 V/m or 80%
Low Gain Up to 240 V/m or 240%

If another unit of measurement is used, the specified values must be converted into V/
m accordingly.

Narda FieldMan 105


10 Cross-mode functions

10.7 Peak marker


In addition to the marker, the peak marker is available for evaluating the peak values in
the following operating modes.
• FIELD STRENGTH
• SPECTRUM
• SHAPED TIME DOMAIN
• TIMER LOGGING (after calling the measurement in the DATA LOGGER)
For this purpose, up to 25 peak values are recorded and sorted in descending order.
Number 1 is the highest peak value. The selected peak value is displayed within the
cross key symbol.

Marker type On/Off Symbols Representa- Control via cross


tion key
Off On on the display
Marker SK1 ◄► ◄►

Peak Marker SK4 ◄►


▲▼

Use the Peak Marker:


✓ The marker is switched on (see the descriptions in the respective operating modes).
1. Press SK4 to switch on the peak marker.
2. Move the marker:
a. Use the ◄ ► keys to move the marker to the nearest peak on the left or right.
b. Use the ▲▼ keys to move the marker to the next higher or next lower peak.
↳ The selected peak value is displayed within the cross key
symbol.
3. Press SK4 to turn the peak marker off again.

106 FieldMan Narda


10.8 Display of overdriven measured values

10.8 Display of overdriven measured values


Overdriving can lead to damage or even destruction of a measuring probe. In addition,
during overdriving, the measurement results are undervalued due to the occurring com-
pression and thus recorded as too low. Measurement data stored in data records are
identified by override symbols.
A currently overdriven state is indicated by a red instead of a yellow display field in con-
nection with an upward pointing triangle symbol

Due to previous overdriving, the statistical measured values of MAX, MAXAVG and AVG
are also undervalued and are therefore displayed in red after overdriving has occurred.
In this case, the values in the measurement settings should be reset.

Narda FieldMan 107


10 Cross-mode functions

108 FieldMan Narda


10.8 Display of overdriven measured values

11 PC software and updates

In this chapter you will find instructions for using the Narda-
TSX PC software as well as for updating the firmware of the
Narda FieldMan.

11.1 Narda-TSX PC software (page 110)


11.2 Firmware update (page 111)

Narda FieldMan 109


11 PC software and updates

11.1 Narda-TSX PC software


The newly developed, extremely powerful Narda-TSX PC software is available for evalu-
ating and documenting the measurement results, media and other information. It is
Narda's new software platform that will support other Narda products in the future in addi-
tion to FieldMan. The PC software is also required for firmware updating and enables
remote control of the FieldMan.
For connection to a PC, the FieldMan has the following interfaces:
• Serial COM interface via the USB-C port
• Network interface via the Ethernet port
• Serial COM interface via the optical port

11.1.1 Establishing a USB connection


The device comes with a USB cable that can be used for power supply and connection
to a PC.
Establish a USB connection:
1. Under GENERAL SETTINGS > CONNECTIVITY > SERIAL INTERFACE, activate
the USB function.
2. Connect the device at the USB-C port (see Fig. 2: 6) to a USB port on the PC using
the supplied USB cable.
↳ The first time you connect, the FieldMan is automatically set up on the PC.
↳ After setup, the FieldMan can be used.
⇒ Instructions for further use of the device can be found in the online help for the Narda-
TSX PC software.

11.1.2 Establishing a network connection


An Ethernet cable with RJ45 connector is required to establish a network connection.
Establishing a network connection:
1. Under GENERAL SETTINGS > CONNECTIVITY > ETHERNET, activate the
Ethernet function. Select the appropriate network settings and activate DHCP or
enter the address settings manually.
2. Connect the device using a suitable Ethernet cable at the Ethernet port (see Fig. 2: 9)
to a router or network hub.
3. Instructions for further use of the device can be found in the online help for the Narda-
TSX PC software.

110 FieldMan Narda


11.1.3 Establishing an optical connection
In principle, an optical connection can also be established with a PC and the PC software
can be used via this connection. However, due to the higher transmission speed, a USB
or Ethernet connection is much better suited for this purpose. However, for remote mea-
surements with HF probes, the optical connection via fiber optic cable and optical-electri-
cal converter to USB is the best solution, because in this way interference is avoided.
Establish an optical connection:
1. Under GENERAL SETTINGS > CONNECTIVITY > OPTICAL INTERFACE, select
the Remote function.
2. Connect the device at the optical connector (see Fig. 2: 7) to a PC via the optical
cable.
↳ The first time the device is connected, it is automatically set up on the PC.
↳ After the setup, the device can be used.
3. For information on remote control of the device, refer to the online help for the Narda-
TSX PC software.

11.2 Firmware update


The firmware update of the Narda FieldMan is performed using the Narda-TSX PC soft-
ware.
⇒ The description of how to perform the firmware update can be found in the online help
for the Narda-TSX PC software.

Narda FieldMan 111


11 PC software and updates

112 FieldMan Narda


11.2 Firmware update

12 Care, maintenance and


disposal

This chapter describes how to clean the device and replace


the rechargeable batteries.

12.1 Cleaning the device (page 114)


12.2 Replacing/removing rechargeable batteries (page 115)
12.3 Disposal (page 116)

Narda FieldMan 113


12 Care, maintenance and disposal

12.1 Cleaning the device


ATTENTION
Penetrating liquids
Liquids that get inside can damage or destroy the device.
⇒ Make absolutely sure that no liquids get into the interior of the device.

ATTENTION
Solvents
Solvents can attack the surfaces of the device.
⇒ Do not use solvents to clean the basic unit, probe and Netzteil/Ladegerät.

To clean the device:


1. Clean the device with a soft cloth. As cleaning agent, we recommend the use of
lukewarm water to which a drop of liquid detergent has been added.
2. To avoid streaks and stains, wipe the still damp parts of the device with a dry cloth.

114 FieldMan Narda


12.2 Replacing/removing rechargeable batteries

12.2 Replacing/removing rechargeable


batteries

CAUTION
Defective rechargeable battery
A defective rechargeable battery can lead to injuries and damage to the device.
⇒ Do not operate the device if the rechargeable battery is damaged.
⇒ Replace a defective rechargeable battery immediately.

⇒ Replace the rechargeable battery when the capacity noticeably decreases.


You can obtain a replacement rechargeable battery as a spare part from your Narda
sales partner (order number 2460/90.07).

Replace the rechargeable battery:


✓ The device is switched off.
1. Open the 3 screws on the rechargeable battery
compartment with a screwdriver and remove
the cover.

2. Unlock the connector at the top and pull it off.


3. Insert the new rechargeable battery and place
it in the handle.
4. Replace the cover and screw it tight.

⇒ Fully charge the rechargeable battery before use.

Narda FieldMan 115


12 Care, maintenance and disposal

12.3 Disposal

12.3.1 Disposal of end-of-life equipment


The crossed-out wheeled garbage can symbol indicates that this product is subject to the
European WEEE Directive 2012/19/EU on the disposal of waste electrical and electronic
equipment and must be disposed of separately from household waste in accordance with
your national regulations.
In the European Union, all electronic measuring systems purchased from Narda after
August 13, 2005 can be returned at the end of their useful life.
⇒ Further information is available from your Narda sales partner.

12.3.2 Disposal of removable batteries


Batteries must not be disposed of in household waste, but must be disposed of sepa-
rately from the product in accordance with the applicable regulations. Batteries can be
returned for free at the corresponding collection points, your dealer or directly via Narda.
⇒ Please discharge the batteries before disposal.

12.3.3 Deletion of private data


⇒ Remove the micro SD memory card before passing on or disposing of the device.
There is no personal data on the FieldMan itself.

116 FieldMan Narda


12.3 Disposal

13 Conformity

In this chapter you will find important information on


declarations of conformity and approvals.

13.1 EU Declaration of Conformity (page 118)


13.2 UKCA Declaration of Conformity (page 118)
13.3 Regulatory Compliance Mark (RCM) (page 118)
13.4 FCC and IC statement (page 118)
13.5 Radio approval (page 119)
13.6 Technical data (page 119)

Narda FieldMan 117


13 Conformity

13.1 EU Declaration of Conformity


Narda STS hereby declares that this equipment is in compliance with Directives 2014/53/
EU,
2014/35/EU and 2011/65/EU, as well as product standards EN 301 489-1 V2.2.3,
EN 301 489-17 V3.2.4, EN 61326-1:2021, EN 61010-1:2010.
The full text of the EU declaration of conformity is available at www.narda-sts.com

13.2 UKCA Declaration of Conformity


Narda STS hereby declares that this equipment complies with the product standards
EN 301 489-1 V2.2.3, EN 301 489-17 V3.2.4, EN 61326-1:2021, EN 61010-1:2010 and
the following regulations:
• Radio Equipment Regulations
• Electrical Equipment (Safety) Regulations
• The Restriction of the Use of Certain Hazardous Substances in Electrical and
Electronic Equipment Regulations
The full text of the UKCA declaration of conformity is available at www.narda-sts.com.

13.3 Regulatory Compliance Mark (RCM)


.

13.4 FCC and IC statement


This device complies with part 15 of the FCC Rules. Operation is subject to
the following two conditions:
(1) This device may not cause harmful interference, and
(2) this device must accept any interference received, including interfe-
rence that may cause undesired operation.

Per RSS-Gen, Section 8.4 this device complies with Industry Canada li-
cense-exempt RSS standard(s). Operation is subject to the following two
conditions:
(1) This device may not cause harmful interference, and
(2) this device must accept any interference received, including interfe-
rence that may cause undesired operation of the device.

118 FieldMan Narda


13.5 Radio approval

13.5 Radio approval


This device contains components for radio communication via WiFi and Bluetooth and
therefore represents a radio system. To use radio communication, the device option
2460/95.12 must be purchased and enabled.
⇒ For initial setup and country configuration, see 4.6 Initial commissioning on page 27.

Country Approval number


China CMIIT ID: 2022DJ18564
Japan 020-220193
USA FCC ID: 2A77Y-246001A
Canada IC: 28882-246001A

13.6 Technical data


Parameter Value
Frequency range 2.4 GHz – 2.495 GHz
Bandwidth 20 MHz
Max. transmitting power ≤ 17.5 dBm (56.2 mW)

Narda FieldMan 119


13 Conformity

120 FieldMan Narda


Index

Index

A Strong fields 12
Abbreviations 18 Undervaluation of the field
Acoustic intensity display 46 strength 13
Add voice comment (Data Logger) 95 Data Logger
Alarm Add text comment 95
Function 104 Delete measurement 96
Settings 46 Manage measurements 94
Altitude 49 Manage media 96
Applications 9 Manage screenshots 96
Attaching marking rings 26 Projects 93
Automatic shutdown 47 Structure 92
Voice comment
B Add 95
Band filter 48 Date 49
Bluetooth 50 Device
Brightness 45 Clean 114
Switch on 27
C Device information 51
Device overview 22
Center frequency 48
Device settings
Charge status, rechargeable battery 25
Change 40
Charging the rechargeable battery 24
Overview 44
Clean, device 114
Personalize 45
Color scheme 45
DHCP 51
Comment on measurement results 103
Display
Comment, measurement results 103
Function key symbols 36
Connection
In measuring mode 42
Ethernet 110
LCD 33
Optical 111
LED 32
USB 110
Probe info 34
Controls 30
Status bar symbols 35
Correction factor
Distance measurement 49
Apply 98
Settings 47
E
Create, setup 55
Create/manage setups 55 Electromagnetic fields 8
Equip case 21
D Ethernet 51
Exit submenu 40
Dangers
Measurement error 13

Narda FieldMan 121


Index

F LCD display 33
Factory reset 47 LED display 32
Factory settings 47 Longitude 49
Field Strength 60
Fields, electromagnetic 8 M
Filter type 48 MAC address 51
Firmware update 111 Manage measurements 94
Fixed keys 38 Manage media 96
Format (memory card) 47 Manage projects 93
Format memory card 47 Manage screenshots 96
Function keys, symbols 36 Manage, setup 55
Measurement
G Delete (Data Logger) 96
GNSS Save 40
Settings 49 Start/stop 40
Synchronization 49 Measurement error 13
GPS Measurement settings
Accuracy 49 Change 40
Settings 49 Explanation 44
System 49 Measuring mode, display 42
Mute 46
H
N
Host name 50
Network 50
I Network connection 110
Information Non-intended use 12
Device 51
O
Probes 52
Intended use 12 Offset correction 102
Interface Operating concept 38
Optical 50 Operating instructions, structure 16
Serial 50 Operating mode
Wireless 50 Field Strength 60
IP address 51 Overview 58
Shaped Time Domain 83
K Spatial Average 67
Keys Spectrum 78
Fixed 38 Timer Logging 71
Softkeys 38 Operating steps, basic 39
Operation with rechargeable battery 24
L Optical connection 111
Optical interface 50
Language, user interface 16
Overdriven measured values 107
Latitude 49

122 FieldMan Narda


Index

Overview of operating modes and probe Take automatically 46


types 58 Take manually 41
SD card directory structure 53
P SD card structure 53
Package contents 20 SD card, structure 53
Packaging 20 Select country 28
Peak marker 106 Select country of use 28
Percent of standard 47 Select language 45
Personalize, device settings 45 Select operating mode 39
Post-process measurement data 98 Serial interface 50
Post-Processing 98 Set default text 46
Post-processing of measurement Set reference frequency 98
data 98 Set time 27
Power supply 24 Set time zone 27
Probe Shaped Time Domain 83
Connect 25 Shutdown (automatic) 47
Test 26 Signs and symbols 17
Probe change during operation 54 Softkeys 38
Probe connection 25 Software update 111
Probe info on the display 34 Spatial Average 67
Probe information 52 Spectrum 78
Probe test 26 Standards 9
Probe types, overview 58 Device settings 47
Display in % 99
R HF probes 101
Rechargeable battery LF probes 102
Charge state 25 Select 100
Charging 24 Status bar, symbols 35
Correct handling of 13 Strong fields 12
Dispose 116 Subnet mask 51
Operation 24 Switching on, commissioning 27
Replace 115 Switching on/off, basic operation 39
Rechargeable battery/device Symbols 18
disposal 116 Function keys 36
Reference frequency 47 Status bar 35
Reset, settings 44 Symbols and signs 17
Resetting settings 44 Synchronization with GNSS 49

S T
Screen saver TCP 50
Settings 45 Terms 18
Use 39 Text comment
Screenshot Activate 103

Narda FieldMan 123


Index

Add (Data Logger) 95


Set 46
Text markings 18
Time 49
Timer Logging 71

U
Undervaluation of the field strength 13
Unpacking 20
Update 111
USB connection 110
USB on/off 50
Use
Intended 12
Unintended 12
User interface, language 16

V
Voice comment
Activate 103
Set 46
Volume 46

W
Warnings
Structure 17
Warning words 17
WiFi, wireless interface 50

Z
Zero 102

124 FieldMan Narda


Narda Safety Test Solutions GmbH
Sandwiesenstrasse 7
72793 Pfullingen, Germany
Phone +49 7121 97 32 0
info@narda-sts.com

Narda Safety Test Solutions


North America Representative Office
435 Moreland Road
Hauppauge, NY11788, USA
Phone +1 631 231 1700
info@narda-sts.com

Narda Safety Test Solutions S.r.l.


Via Benessea 29/B
17035 Cisano sul Neva, Italy
Phone +39 0182 58641
nardait.support@narda-sts.it

Narda Safety Test Solutions GmbH


Beijing Representative Office
Xiyuan Hotel, No. 1 Sanlihe Road, Haidian
100044 Beijing, China
Phone +86 10 6830 5870
support@narda-sts.cn

www.narda-sts.com

2460/98.21
01/10.2022, A ...
English

You might also like