l'raswutllya Astagt n rusttj :;, .sm iVuna if.ilotmrah :;, amt. Katlla Anggraem Agu stWJI :;
IJ Student at Physics Study Program JTB
lJ Research group : Nuclear Physics and Biophysics
11
Otolaryngology Specialist Consultant Doctor ofdr. Hasan Sadikin Hospital
Abstract
The jimdamental test of audiological evaluations is the pure tone test because it can define the
hearing threshold of several important ji-equencies in conversational speech. A personal computer (PC)
based pure-tone audiometer was made using MATLAB program that offers two choices, automatic and
manual testing. This was tested to 23 research subjects (18 patients and 5 controls) in Ear and Speech Clinic
dr. Hasan Sadikin Hospital Bandung. All the research subjects were tested twice, firstly with standard
audiometer and secondly with this PC-based audiometer. Each research subjects have two audiograms that
contain data of hearing threshold intensity at several speech frequencies. The comparison between the two
audiograms is summarized into 4 groups that depend on the several factors of the device and the research
subjects.
Keywords: Hearing impairment, Hearing threshold, Pure-tone, PC-based pure-tone audiometer, dr.
Hasan Sadikin Hospital
1033
International Conference on Mathematics and Natural Sciences (ICMNS)
November 29-30, 2006, Bandung-lndonesia
L'
15 . : :
The PC based audiometer software was
10 I '
5 : : : : t
made using MATLAB 6.5 and run with Notebook
0 -:--:~-- '---~~ +-7-·-~-:-. -~ Compaq Presario (Prosesor Intel® Pentium
125 250 500 1k 1k5 2k 3k 4k 6k ek
Mobile 1,73 GHz, 512 DDR SDRAM), and used
frequency (Hz)
Windows XP Professional as the operating
system. Figure-2 shows that the subject interacts
Figure- I b. The equivalent dB-SPL is for supra-aural earphone [I] with PC monitor, listens sound stimulus via
earphone, and responses the sound by clicking the
The threshold sensitivity in hearing test mouse button.
for air conduction in pure tone audiometer is PC-based audiometer programming and
evaluated using earphone. This test is capable to standard audiometers use Modified Hughson-
determine the degree of hearing loss (e.g. mild, Westlake Technique as follows: ( I) frequency
moderate, severe, or profound) but it is not to sequence started at I 000 Hz, then increased to
determine the type of hearing loss (e.g. 2000, 4000, 8000 Hz, and finall y decreased to
conductive, sensorineural, or mix) [3]. 500, 250, and 125 Hz. (2) initial intensity is 40 dB
Pure tone air conduction test procedures for each frequency, (3) if subject gives a response,
are: ( 1) case history I patient information, (2) test then intensity is decreased in I 0-15 dB steps until
environment, (3) listener position, (4) instruction there is no response, (4) if subject gives no
to the patient, (5) ear examination, (6) earphone response, then intensity is increased in 20 dB
placement, (7) ear selection, (8) frequency steps until there is a response. Then, intensity is
sequencing, (9) threshold procedures, and ( 10) decreased in 10 dB steps until there is no
false negative and positive responses [I]. response, (5) determination the threshold by
increased in 5 dB steps until subject gives a
III. Experimental Method response, then decreased in I0 dB steps until
subject gives no response, then increased in 5 dB
A. Self Created PC-Based Pure Tone steps until subject gives a response, (6) if a fal se-
Audiometer positive or false-negative response occurs then the
Pure tone threshold are regularly assessed subject must be re-instructcd, (7) a respons..: is
at octave and sometimes at half-octave intervals valid only if the time latency is con si st~.:nt. (S) the
between the range of 125 and 8000 Hz. The thresho ld is the lowest intensity which subjc:ct
output pure tones from standard audiometer at
gives 2 valid responses from 3 stimu lus.
RSHS were recorded (in l(dB-comp) scale) by
using a PC to be analyzed their maximum
B. PC-based Audiometer Test at RSHS
intensity and wave characteristics [5]. The tests were done at !!caring and
The valid recorded maximum intensity was Speech Impairment Clin ic at dr. Hasan Sadikin
written to Microso ft Excel. At each frequency, a Hospital Bandung on i'vlay 16m to 27 1h 2006. PC-
chart of l(dB-comp) as a function of l(dB-HL) was based audiometer tests for patients were done with
made to determine the conversion (regrl!ssion) assistance of an examiner, however lor controls
function from d13-HL to dB-comp to obtain the were done automatically (without an examiner).
next d13-comp [5]. Patients and controls were tested twice using
standard audiometer at RSHS ami then PC-based
audiometer.
103-1
International Conference on Mathematics and Natural Sciences (ICMNS)
November 29-30, 2006, Bandung-lndonesia
IV. Results
Table- I lists the results ofthreshold intensity as fu nction of frequency of8 patients from total 18 patients and 2 controls
from total5 controls (this is done because of space limitation). The s~~bject code P 16-1 means the subject is the fi rst
!·~·•;•l.l '"'h" n ... Jr-,1 ~~~ ~. , .,.! .,,,t.
p 16-2
p 17-3
p 18-2
p 18-3 ~Sf IS--------- -- - ~-? ...-.: ..-~~---l----~_?___~__}9__ _L_;!~ ___;___~Q___;___ ?_9___ ----~,S. __l___7Q___ _;, __~_Q_ ___Ll_O_.:___ ?.9.. _; __?,5._.!.___1?._
PC-bascd 35 : 30 ' 30 · 30 ; 40 ' 45 : 60 20 : 25 ' 30 ; 25 · 45 ; 65 ; 70
__¥Sf~--------- __ ).?____: ___)~ ___ !_)2____ j_ __~_Q___L_.29____ l_ __ 2,5___!___?_Q___ ____!?. __.l __ )Q___l_ __?_L_~_;___ ?Q___!_ __~_s. ___'_ ~g_
p 19-1
PC-Imcd 25 25 30 · 40 ~ 60 80 ' 80 20 20 : 30 ' 45 65 85 ' 80
p 24- 1
_nsns ______ --~?--~-2?____:____~Q_ __,__}_9__ ,____~Q___L?Q....!._?..9___ ---~-_.__?Q___:___?_Q_ ___;_i_0_;__?.9___~___?Q_-'-·2Q...
1
PC-bascd 30 ' 25 ' 15 ' 25 ; 50 : 65 ' 80 45 ' 70 ' 45 ' 35 ' 45 ' 70 80
p 24-2
p 27-3 _13-..§!-!L_____ ---~~... L .\?....L .!!L.J____t_Q__ _L __.\?.... L ..!2__L _1_9_______!Q...L...!.~...l...!.!!....L2._ t...?.... L~ ...-L..?.Q__
PC-bascd 5 : 5 ' 10 i IS ; 20 l 0 l 0 0 ·. 5 ; 15 . IS 1 15 : IS . 10
c 27-1
c 27-2
1035
'"•~~''"'' International Conference-on. .MathematicS-and Natural-Sc iences (ICMNS)
1"'.-----,..,_,~ooo
r-------.---,........,,.'--;oooo I ,'"' _ 1 ~ove~bey~<.\':3~, 200?, Banclung-lndonesia
i l lj i II i ! i II:
! ! : Ii
I1!11 .~ =1~1
tt-rn J~ ,
l_iJ&__tJl, ,~,I~
i 1 x-V! 1"---':\ 1
..
70
::.·~·~
~- &~01·>-{,: I : :;;.:=tt1~~7f?l~ ~~~ I .
0
II' .I ! i ·. I
- ~- L; I!:
i ! I' l
~:,. ~ II
30 !
I -lh--
i m
!jt!
I I . •• ~ 11.
1 : -. I
0""'
ao !
90
I i
"'
.... . ... L
1036