Professional Documents
Culture Documents
Orthopantomograph Op 3D: Service Manual
Orthopantomograph Op 3D: Service Manual
Service Manual
ENGLISH
Contents
1 Disclaimer.............................................................................................................................. 7
2 Introduction........................................................................................................................... 8
2.1 ORTHOPANTOMOGRAPH™ OP 3D..........................................................................................8
2.2 Intended use............................................................................................................................. 9
2.3 Abbreviations.............................................................................................................................9
2.4 Disposal and recycling.............................................................................................................. 9
2.5 Associated documentation........................................................................................................ 9
2.6 For the service manual user....................................................................................................10
3 Safety....................................................................................................................................11
3.1 Signal words............................................................................................................................11
3.2 General servicing precautions................................................................................................. 11
3.3 Electrical safety....................................................................................................................... 11
3.4 Electrostatic discharge............................................................................................................ 12
3.5 Device modifications during service........................................................................................ 12
3.6 Radiation safety....................................................................................................................... 12
3.7 Symbols that may appear on the device or its parts................................................................13
4 Device description............................................................................................................ 15
4.1 Mechanical description............................................................................................................15
4.1.1 Main parts................................................................................................................. 15
4.1.2 Device movements.................................................................................................... 16
4.1.3 Device covers............................................................................................................17
4.1.4 Device labels............................................................................................................. 20
4.1.5 Status indicator light..................................................................................................21
4.1.6 Patient positioning panel............................................................................................21
4.1.7 Emergency stop switch............................................................................................. 22
4.2 Electrical description............................................................................................................... 23
4.2.1 Circuit board locations...............................................................................................23
4.2.2 Circuit boards overview............................................................................................. 23
4.2.3 Position control..........................................................................................................24
4.2.4 Exposure logic........................................................................................................... 24
4.2.5 Power supply.............................................................................................................25
4.2.6 Fuses.........................................................................................................................26
4.3 Circuit boards..........................................................................................................................26
4.3.1 i2000 image capture board....................................................................................... 26
4.3.2 i3000 Main control board.......................................................................................... 29
4.3.3 i4000 power supply board.........................................................................................31
4.3.4 i4100 back panel board............................................................................................ 33
4.3.5 i6000 user interface board........................................................................................ 33
4.3.6 i8000 LED indicator board.........................................................................................34
4.3.7 i9000 chin rest sensor board.....................................................................................34
4.4 Device boot up sequence....................................................................................................... 35
4.5 Graphical User Interface (GUI)................................................................................................36
5 Diagrams............................................................................................................................. 37
5.1 Wiring diagram........................................................................................................................ 37
5.2 Fusing diagram........................................................................................................................39
5.3 Emergency stop...................................................................................................................... 39
ORTHOPANTOMOGRAPH OP 3D iii
5.4 Mains voltage distribution........................................................................................................39
5.5 Power distribution +24Vdc...................................................................................................... 40
5.6 Ethernet communication..........................................................................................................40
5.7 CAN communication............................................................................................................... 40
5.8 i2C self diagnosis communication........................................................................................... 40
5.9 Rotation movement................................................................................................................. 41
5.10 Linear movement...................................................................................................................41
5.11 Pivot movement.....................................................................................................................42
5.12 Z movement.......................................................................................................................... 42
5.13 Collimator movements...........................................................................................................43
5.14 Positioning lights....................................................................................................................43
5.15 Imaging chain - 2D/3D sensor.............................................................................................. 44
5.16 Imaging chain - 2D sensor.................................................................................................... 44
5.17 Exposure control 1................................................................................................................ 45
5.18 Exposure control 2................................................................................................................ 45
5.19 Exposure warning light.......................................................................................................... 46
5.20 Device control panel............................................................................................................. 46
5.21 Chin rest detection................................................................................................................47
5.22 PAN ceph sensor movement.................................................................................................47
6 Device settings.................................................................................................................. 48
6.1 Service mode access.............................................................................................................. 48
6.2 Setting menus......................................................................................................................... 49
6.2.1 Quality control........................................................................................................... 49
6.2.2 Calibrations................................................................................................................49
6.2.3 Settings..................................................................................................................... 50
6.2.4 Testing.......................................................................................................................55
6.2.5 Service...................................................................................................................... 63
6.2.6 Logs.......................................................................................................................... 64
6.2.7 About.........................................................................................................................65
7 Maintenance....................................................................................................................... 66
7.1 Firmware update..................................................................................................................... 66
7.2 Annual maintenance................................................................................................................68
7.2.1 Cleanliness of the device...........................................................................................68
7.2.2 Condition of the cables..............................................................................................68
7.2.3 Checking the tightness of linear movement bearings................................................. 69
7.2.4 Device movements and functionality..........................................................................71
7.2.5 Exposure indicator functionality................................................................................. 73
7.2.6 Changing the fuses................................................................................................... 73
7.2.7 Monitor calibration..................................................................................................... 74
7.3 Calibrations............................................................................................................................. 75
7.3.1 Calibration programs................................................................................................. 75
7.3.2 When to calibrate the device.....................................................................................76
7.4 Exposure tests.........................................................................................................................79
7.4.1 PAN exposure test.................................................................................................... 79
7.4.2 CEPH exposure test.................................................................................................. 81
7.4.3 3D exposure test.......................................................................................................82
8 Troubleshooting.................................................................................................................83
8.1 Troubleshooting - Common issues.......................................................................................... 83
8.2 Troubleshooting - Movement errors........................................................................................ 88
8.3 Image transfer issues.............................................................................................................. 89
8.4 Fail-safe mode.........................................................................................................................89
8.4.1 Fail-safe mode error messages................................................................................. 89
8.5 Error codes............................................................................................................................. 93
8.5.1 List of error codes..................................................................................................... 93
iv ORTHOPANTOMOGRAPH OP 3D
8.5.2 Acknowledging errors.............................................................................................. 134
8.6 Resetting the device IP configuration manually......................................................................134
8.6.1 Reset to DHCP (factory default setting)...................................................................134
8.6.2 Reset to Static IP address....................................................................................... 134
ORTHOPANTOMOGRAPH OP 3D v
9.10.3 Replacing CEPH secondary collimator opto sensor............................................... 220
9.10.4 Replacing the CEPH tubehead.............................................................................. 220
9.10.5 Replacing the CEPH tubehead scanning mechanism.............................................222
9.10.6 Replacing CEPH tubehead scan opto sensor........................................................ 223
9.10.7 Replacing the CEPH tubehead cables................................................................... 223
9.10.8 Replacing the straight part of the CEPH arm......................................................... 226
9.10.9 Replacing the arched part of the CEPH arm..........................................................230
9.10.10 Replacing the CEPH head support...................................................................... 231
9.10.11 Replacing the ear rods and nasion support......................................................... 232
9.11 Mechanical adjustments...................................................................................................... 233
9.11.1 2D/3D Sensor height adjustment...........................................................................233
9.11.2 2D Sensor adjustments......................................................................................... 233
9.11.3 PAN/3D Collimator adjustment.............................................................................. 236
9.11.4 CEPH primary collimator adjustment..................................................................... 238
9.11.5 CEPH secondary collimator adjustment................................................................. 240
9.11.6 Ear rod adjustment................................................................................................ 242
9.11.7 Patient positioning light adjustment........................................................................246
9.12 After service work............................................................................................................... 248
vi ORTHOPANTOMOGRAPH OP 3D
1 Disclaimer
1 Disclaimer
ORTHOPANTOMOGRAPH™ OP 3D Service Manual, 224067-PTU r1.
The manufacturer has no liability for consequential damage, personal injury, loss, damage or expense
directly or indirectly caused by the use of the product. No agent, distributor or other party is authorized to
give warranty or other liability on behalf of the manufacturer with respect to its products.
Manufacturer:
PaloDEx Group Oy
Nahkelantie 160
FI-04300 Tuusula
FINLAND
ORTHOPANTOMOGRAPH OP 3D 7
2 Introduction
2 Introduction
2.1 ORTHOPANTOMOGRAPH™ OP 3D
The ORTHOPANTOMOGRAPH™ OP 3D (later called device) is a dental X-ray device producing high quality
digital images of dentition, TM-joints, head and neck area and carpus. To take images, you need a suitable
workstation connected to the device and a dental imaging software to capture and manage the images.
The ORTHOPANTOMOGRAPH™ OP 3D can be used for the following procedures depending on the device
configuration and country-specific regulatory approvals:
Panoramic Imaging
• Standard panoramic
• Segmented panoramic
• Pediatric panoramic
• Bitewing
• TMJ, lateral projection
Cephalometric Imaging
• Cephalometric lateral projection
• Cephalometric pediatric lateral projection
• Cephalometric Posterior-Anterior (PA) projection
• Carpus view
3D CBCT Imaging
• FOV (Volume Height x Diameter)
◦ 5x5
◦ 6x8*
◦ 6x9
◦ 8x8*
◦ 9 x 11
◦ 9 x 14
• Resolution
◦ LDT - Low Dose Technology™
◦ Standard
◦ High
◦ Endo (5 x 5 only)
• Scout images for FOV position and height verification
• Dental cast imaging
* In some countries/areas where 3D volume dimensions are limited by local regulations FOV sizes 6 x 9 and
9 x 11 are replaced by FOV sizes 6 x 8 and 8 x 8.
NOTICE! This manual describes the service of a device with all available configurations and so may
contain instructions to parts that are not available with the serviced device.
CAUTION! USA only: Federal law restricts this device to sale by or on the order of a dentist or other
qualified professional.
To function, the device requires an additional workstation with MDD approved dental imaging software.
Every software used with the device must conform to the MDD and the relevant legal requirements in the
USA. The workstation must conform to all the device and dental imaging software requirements.
8 ORTHOPANTOMOGRAPH OP 3D
2 Introduction
2.2 Intended use
ORTHOPANTOMOGRAPH™ OP 3D is an X-ray device that is intended to be used for imaging of adult and
pediatric patients. The device can be configured to take panoramic, cephalometric, or 3D images of the
cranio-maxillofacial complex for use in diagnostic support. The device can also be configured to take carpus
images.
ORTHOPANTOMOGRAPH™ OP 3D must only be used and operated by dentist and other qualified
professionals.
2.3 Abbreviations
PCBA Printed Circuit Board Assembly
THA Tubehead Assembly
GUI Graphical User Interface
EMI Electromagnetic Interference
ESD Electrostatic Discharge
DUT Device Under Test
At least the following parts of the device should be recycled according to local and national regulations
regarding disposal of non-environmentally friendly materials:
• Tubehead (oil)
• All electronic circuits and electronic boards
• Plastic parts
2.5 Associated documentation
• ORTHOPANTOMOGRAPH™ OP 3D User and Installation manuals. Download the latest versions from
https://tech.dexis.com.
• Dental imaging software documentation
• ORTHOPANTOMOGRAPH™ OP 3D Sparepart Catalogue.
ORTHOPANTOMOGRAPH OP 3D 9
2 Introduction
10 ORTHOPANTOMOGRAPH OP 3D
3 Safety
3 Safety
3.1 Signal words
The following signal words and labels are used in this document:
WARNING! Indicates a hazardous situation which, if not avoided, could result in death or serious
injury.
CAUTION! Indicates a hazardous situation which, if not avoided, could result in minor or moderate
injury.
NOTICE! Highlights suggestions which will result in enhanced installation, reliability, or operation.
Not used for safety related hazards.
• Use only the original spare parts that are provided by the manufacturer when you repair the device or
replace any parts in it.
• Before you attempt to service the device make sure that you know how to operate it. Read the OP 3D
User manual.
• Attach the covers before handing the device over to the user.
• Ensure the electrical safety of the device by inspecting the proper grounding of user touchable metallic
parts before handing the device over to the user.
• Clean any accumulated dust inside the device.
• Attach all screws and parts that have been removed or loosened during the maintenance.
• Check the tightness of the screws in the device, especially in the carriage and the wall mount.
• For connection requirements and safety, refer to the OP 3D User and Installation manuals.
• Read and understand the warnings and precautions listed in the following sections of this service manual
and in the OP 3D User and Installation manuals.
• The device column is not a servicable part and it should not be opened or altered in any circumstances.
3.3 Electrical safety
• Disconnect the device from the main power supply before repairing or replacing any mechnical parts,
circuit boards or installing accessories.
• Disconnect the device from the main power supply before removing the covers.
• Disconnect the device from the main power supply before replacing circuit boards or other electrical
components.
• The circuit boards and electrical components of the device can contain capacitors with high voltages.
Wait ten (10) minutes, after disconnecting the device from the power supply, before handling the board
or component.
• If you have to leave the device unattended during service or maintenance, disconnect the device from
main power supply to protect people, who may touch the device, from electric shock.
• The device must only be used in areas that are provided with a protective earth connection to ensure an
equipotential ground connection.
• Assemble the device according to the instructions in OP 3D Installation manual, always use the correct
materials and original spare parts.
Do not compromise:
ORTHOPANTOMOGRAPH OP 3D 11
3 Safety
3.4 Electrostatic discharge
Electrostatic Discharge (ESD) can damage or destroy electronic components.
When you service the device take precautions to avoid electrostatic charge to build up and discharge
(ESD). Follow the local and national recommendations for the prevention of ESD. If no recommendations are
available, follow the guidelines below:
• Leave all new or replacement circuit boards and electrical parts in their protective packaging until the
boards are needed.
• Before handling circuit boards and electrical parts make sure that any static electricity charge that has
built up in your body is discharged.
• When examining and checking circuit boards use an antistatic wrist wrap which is connected to a ground
point through a 1 Mohm current limiting cable. Also use a cable to connect the device to the same
ground potential as the wrist wrap.
• When handling circuit boards hold them by their edges and do not touch any components or connectors.
• If an antistatic mat is used, connect the wrist wrap to the mat and the mat to the ground potential.
• To make sure that the wrist wrap is in good condition, check and wash it frequently.
CAUTION! Make sure that any modifications done to the device do not degrade the EMC
performance of the device.
• Never make unauthorized changes or modifications to the device or its parts.
• Never dismantle or remanufacture any part of the tubehead assembly or beam limiting device.
• Never adjust any part of the beam limiting device unless under the direction of the manufacturer.
• Ensure that the EMC performance of the device is not degraded when the device is upgraded, serviced
and repaired.
NOTICE! Ensure that the EMC performance is not degraded thoughout the whole service life of the
device. Refer to Electromagnetic Compatibility (EMC) tables in OP 3D User and Installation manuals.
3.6 Radiation safety
CAUTION! When you take test exposures, take adequate precautions to protect youself from
radiation. Stand behind a suitable radiation shield that is positioned at least 2 meters from the device.
Before you use or service the device familiarize yourself with local and national radiation safety standards
and requirements relating to dental x-ray equipment.
12 ORTHOPANTOMOGRAPH OP 3D
3 Safety
Manufacturer
Date of manufacture
Medical device
Serial number
Lot number
Caution
Radiation warning
Laser warning
Dangerous voltage
On or enabled
Off or disabled
Exposure switch
Ethernet
Focal spot
ORTHOPANTOMOGRAPH OP 3D 13
3 Safety
Do not reuse
Recyclable
Operating instructions
Refer to operating instructions for more information. The operating
instructions can be supplied electronically or in paper format.
This symbol indicates that the waste of electrical and electronic equipment
must not be disposed as unsorted municipal waste and must be
collected separately. Please contact an authorized representative of the
manufacturer for information concerning the decommissioning of your
equipment.
CE Mark
NRTL Mark
Conforms to AAMI ES60601-1:2006. Certified to CSA.
Importer
14 ORTHOPANTOMOGRAPH OP 3D
4 Device description
4 Device description
4.1 Mechanical description
4.1.1 Main parts
MAIN DEVICE
1. Column 2. Carriage 3. Upper shelf
4. Rotating unit 5. Sensor (2D/3D)* 6. Sensor (2D)*
7. PAN/3D Tubehead 8. Patient positioning panel 9. Status indicator light
10. Power switch (Back of the 11. Emergency stop switch 12. Exposure switch
device)
13. Mirror 14. Lower shelf tray 15. Patient handles
16. Head support 17. Head support locking lever 18. Chin rest
CEPHALOMETRIC UNIT
19. CEPH head support 20. CEPH tubehead 21. CEPH arm
22. Nasion support 23. Ear rod
* The sensor configuration and appearance may vary depending on the device configuration.
ORTHOPANTOMOGRAPH OP 3D 15
4 Device description
4.1.2 Device movements
16 ORTHOPANTOMOGRAPH OP 3D
4 Device description
4.1.3 Device covers
CARRIAGE COVERS
1. Upper shelf cover left 2. Upper shelf cover right 3. Carriage side cover left
4. Carriage side cover right 5. Finger shield plate 1 6. Finger shield plate 2
ORTHOPANTOMOGRAPH OP 3D 17
4 Device description
1. Rotating unit side cover - left 2. Rotating unit side cover - right 3. Rotating unit finger protection -
upper
4. Rotating unit finger protection - 5. Tubehead front cover 6. Tubehead back cover
lower
7. Tubehead plate 8. 2D/3D Sensor covers with 9. 2D Sensor covers with collars
collars (3D devices without CEPH) (3D devices with ceph)
18 ORTHOPANTOMOGRAPH OP 3D
4 Device description
1. Lower shelf upper cover 2. Lower shelf lower cover 3. Layer laser cover
4. Patient handle
1. Ceph tubehead top cover 2. Ceph tubehead front cover 3. Ceph tubehead back cover
4. Ceph head support mount 5. Ceph head support mount 6. Ceph head support upper cover
upper cover lower cover
ORTHOPANTOMOGRAPH OP 3D 19
4 Device description
4.1.4 Device labels
MAIN DEVICE LABELS
20 ORTHOPANTOMOGRAPH OP 3D
4 Device description
Device status indicator light, located on the top of the carriage, illuminates according the device status:
• YELLOW: Device is generating X-rays.
• BLUE: Device is in error state or pending user action. Check the GUI for details.
• GREEN: Device is ready for imaging.
• GRAY: Power down mode. Device is not ready for imaging, follow the instructions on the status message
bar on device GUI.
1. Carriage up/down slider. The carriage can be driven up and down by sliding a finger on the carriage
up/down slider. The finger movement detection works best when using the entire finger tip or two
fingers side by side.
The speed of the carriage movement depends on the distance the finger is moved. For faster carriage
speed, start from the upper/lower edge of the slider.
2. HOME button. Drives the rotating unit to HOME (Patient-In) and scan start positions. Pressing the
HOME button also wakes the device from Power save mode.
3. Test mode button. Disables radiation production of the device. You can also use the GUI to activate
the Test mode.
4. Patient positioning lights button. Turns the patient positioning lights on/off.
5. Exposure indicator (YELLOW). Device is radiating.
6. Status indicator (BLUE). Device is in error state or pending user action. Check the GUI for details.
7. Ready indicator (GREEN). Device is ready for imaging.
ORTHOPANTOMOGRAPH OP 3D 21
4 Device description
NOTICE! If carriage up/down slider backlight is not illuminated, the carriage up/down slider is in safe
mode and cannot be used. Wait until the light comes back on or wipe the panel clean if it is dirty and
dry if it is wet.
Pressing the emergency stop switch immediately terminates the imaging and all device movements and
resets all circuit boards.
NOTICE! An interrupted imaging process cannot be resumed. A new image needs to be taken.
22 ORTHOPANTOMOGRAPH OP 3D
4 Device description
4.2 Electrical description
4.2.1 Circuit board locations
1. i4000 Power Supply board
2. i2000 Image Capture
Board
3. i3000 Main Control Board
4. i4100 Back panel board
5. i5000 Cascade board (in
the Tubehead assembly)
6. i6000 User interface board
7. i7000 Tubehead interface
board (in the PAN/3D
and CEPH tubehead
assemblies)
8. i8000 LED indicator board
9. i9000 Chin rest sensor
board
ORTHOPANTOMOGRAPH OP 3D 23
4 Device description
4.2.3 Position control
Positioning of the device is done by stepper- and DC-motors, using feedback from optical, resistive and Hall
sensors. The control chain per actuator is described below.
Upper shelf linear and pivot movement, rotating unit movement
The movement is driven by a stepper motor. A resistive sensor determines the approximate position of the
actuator. A number of exact positions is determine by matching the approximate position with feedback
from an optical sensor changing state at regular intervals. Positioning between these points is based on by
counting steps fed to the stepper motor.
Collimator movement
The movement is driven by a stepper motor. An optical sensor is used to determine the Home position.
Driving to wanted position is done by first finding the Home position, and then counting steps fed to the
stepper motor.
Z-movement
The movement is driven by a DC motor. An optical sensor is used to determine the Home position. The Z-
position is determined by using the home position as reference and counting pulses generated by a Hall
sensor connected to the motor.
Imaging chain overview / Image data transfer
Image capturing is handled by the i2000 Image Capture Board. This board is responsible for both triggering
the sensor and capturing image data. The data is stored in the i2000 Image Capture Board until transferred
through the i3000 board to the image capturing workstation.
4.2.4 Exposure logic
Exposure logic is solely controlled by i4000 board. The i4000 board includes power supply, inverter and
filament power supply that are responsible for supplying the X-ray tubehead.
Tubehead selection
Prior to the exposure applicable tubehead is selected by logical control signal PAN_ENABLE_N/
KEF_ENABLE.
Filament pre-heat
Prior the high voltage generation at the tubehead the filament is pre-heated. The pre-heating process is
controlled and monitored by the i4000 board.
Exposure
After the filament pre-heat, the exposure is enabled. Exposure is monitored by the i4000 board by utilizing
the KVFB and MAFB signals and comparing them against the selected reference signals.
Backup timer
The exposure timing is controlled by the firmware. In the event of malfunction there is an added safety timer
implemented in hardware, which will cut the exposure in case EXPENA-signal is active too long.
24 ORTHOPANTOMOGRAPH OP 3D
4 Device description
4.2.5 Power supply
The mains power supply comes from the wall socket through the back panel to the i4100 board. For further
information, see i4100 back panel board on page 33. See also Power distribution +24Vdc on page 40.
See chapter Changing the fuses on page 73 for information on how to install the fuses.
ORTHOPANTOMOGRAPH OP 3D 25
4 Device description
If the device mains voltage needs to be changed, change the fuses accordingly. See chapter Changing the
fuses on page 73 for information on how to install the fuses.
4.2.6 Fuses
The allowed main fuse types are listed in the table below. See also Fusing diagram on page 39.
Main fuses
(5x20mm)
4.3 Circuit boards
Some circuit boards and their versions require specific Firmware version to function correctly. See the
compatibility table below.
26 ORTHOPANTOMOGRAPH OP 3D
4 Device description
Functionality
The i2000 board handles image capturing function of the device. The main functions of the i2000 board are
listed below.
• Image capturing from both the 2D/3D and 2D sensors
• Image storing to the image memory
• Image transmission to the modality workstation.
• Driving the rotating unit, sensor swap and collimator plates.
• Providing voltage for the sensors.
Indicators
LED # Color Default state Function Notes
H1 Green ON Configuration done
H2 Green ON Ethernet Link
H3 Green ON FPGA init done
H4 Green Blinking Ethernet RX/TX activity
H5 Green ON P1V35 PG OFF when device is in
the sleep mode
H6 Red OFF FPGA PIO LED5
H7 Green OFF PAN_P5V_D_3D ON when 3D imaging
mode selected and
device is reserved.
H8 Green OFF PAN_P5V_A_CEPH ON when 3D imaging
mode selected and
device is reserved and.
H9 Green OFF Panel supply P12/P6V PG ON when the device is
reserved
H10 Green OFF PAN_P12V ON when 3D imaging
mode selected and
device is reserved.
H11 Green ON P24V Board input voltage
H12 Red OFF FPGA FW_ERR ON when error state
H13 Red OFF FPGA CORE_ERR ON when error state
H14 Green Blinking FPGA PIO LED1 FPGA PIO LED0-2 forms
a group of running lights.
Fast blinking (1s round
time) when FPGA boots
up.
Slow blinking (8s round
time) when FPGA fully
operational.
ORTHOPANTOMOGRAPH OP 3D 27
4 Device description
Indicators
LED # Color Default state Function Notes
H15 Green ON P3V3 PG OFF when device is in
the sleep mode.
H16 Red OFF FPGA PIO LED4
H17 Green ON P5V PG OFF when device is in
the sleep mode
H18 Green Blinking FPGA PIO LED2 See H14
H19 Green OFF FPGA PIO LED3
H20 Green Blinking FPGA PIO LED0 See H14
H21 Red OFF Motor driver POR on ON when a motor
malfunction detected
H22 Green ON DDR_VTT OFF when device is in
the sleep mode
H23 Green Blinking MCU STAT0 MCU SW controlled LED
H24 Green ON NSTATUS
H25 Red OFF FPGA RESET
H26 Green OFF CPAN_P12V ON when PAN/CEPH
imaging mode selected.
H27 Red OFF MCU STAT1 SW controlled LED
H28 Green ON P24V_MOTOR +24Vdc power input for
motors.
H29 Green OFF PAN_P5V_D_CEPH ON when PAN/CEPH
imaging mode selected.
H30 Green OFF PAN_P5V_A_3D ON when PAN/CEPH
imaging mode selected.
28 ORTHOPANTOMOGRAPH OP 3D
4 Device description
Fuses
Ref Value Type Replaceable
F1 FUSE slow 4A 250Vac 5x20 Cooper Bussmann S506-4-R Yes
mm
F2 FUSE slow 6.3A 250Vac Cooper Bussmann S506-6.3-R Yes
5x20mm
F3 Resettable Fuse PPTC 16V, Cooper Bussmann No
0.25A, Itrip 0.5A PTS120616V025
Functionality
The main functions of the i3000 Main control board are:
• Main control of the system
• Linear and pivot motion drive and position sensing
• Exposure control
• Image data streaming between i2000 and MWS
• Audio output
• User interfaces
• Remote GUI through LAN
• i4000 and i2000 firmware and FPGA updates.
ORTHOPANTOMOGRAPH OP 3D 29
4 Device description
Indicators
LED # Color Default state Function Notes
H1 Green ON VAUX33 power on
H2 Green ON MWS Ethernet link
H3 Green ON i2000 Ethernet link
H4 Green ON/blink MWS Ethernet RX/TX activity
H5 Green OFF Motor in reset (POR)
H6 Green ON +24V supply connected to board
and
FUSE ok
30 ORTHOPANTOMOGRAPH OP 3D
4 Device description
Fuses
Ref Value Type Replaceable
F1 6,3AT/250Vac Cooper Bussmann S506-6.3-R Yes
F2 4A 250Vac slow (5x20mm) Cooper Bussmann Yes
S506-4-R
F3 1,1A/6V Resetable PTC fuse No
F4 1,1A/6V Resetable PTC fuse No
F5 0,25A/16V Resetable PPTC fuse No
F6 2A/6V Resetable PTC fuse No
F7 0,25A/16V Resetable PPTC fuse No
F8 0,25A/16V Resetable PPTC fuse No
Functionality
The i4000 board is one of the control boards of the device. The main features of the i4000 power supply
board are:
• Generate all main supplies to the device
• UI control
• Z-motor control and position sensing
• Laser motor control and position sensing
• Chin rest sensor interface
• Patient positioning lights control
• Scan motor control and positioning
• Filament control
• HV control
ORTHOPANTOMOGRAPH OP 3D 31
4 Device description
Indicators
LED # Color Default state Function
H1 Green ON +3V3HV is on (i4000 R20 and older)
+5V0HV is on (i4000 R21 and newer)
H2 Green ON -5V is on
H3 Green ON +20V Triac is on
H4 Green ON +17V is on
H5 Green ON HV395 is on
H6 Green ON PFC power is on
H7 Green ON P15V is on
H8 Green OFF Preheat is on
H9 Green ON P3V3 is on
H10 Green ON P5V is on
H11 Green ON +24V is on
H12 Red OFF Tubefail error
H13 Green ON +20V DCDC
H14 Yellow OFF Exposure on
H15 Yellow OFF Power detected
LA100 ON Mains voltage on
32 ORTHOPANTOMOGRAPH OP 3D
4 Device description
Fuses
Ref Value Type Replaceable
F1 FF8A/500Vac (6,3x32mm) Siba 70-125-40 8A Yes
F2 2A/6V Resetable PPTC fuse Littelfuse No
0805L200SLTHYR
Functionality
• Mains fuses
• Switch of permanent or non-permanent mains installation:
NOTICE! When the non-permanent installation is made, both L (live) and N (neutral) fuses must
be used. Switch SW1 must be set to 2F position. See Mains connection type configuration on
page 25.
NOTICE! When the permanent installation is made, only L (live) fuse must be used. Switch SW1
must be set to 1F position. See Mains connection type configuration on page 25.
• Mains switch is connected to this board
• Exposure switch connector
• Remote exposure switch connector
• Door switch connector
• External warning light connector
Indicators
LED # Color Default state Function Notes
H1 Green ON Green led
controlled by FW.
Fuses
See chapter Fuses on page 26.
Test points
J3 GND test clip
ORTHOPANTOMOGRAPH OP 3D 33
4 Device description
Functionality
The i8000 board generates exposure warning, ready and error status indications. The status is indicated
with colored LEDs.
Indicators
Refer to Status indicator light on page 21 for the indicators.
Functionality
The i9000 board senses the chin rest positions and calibration phantoms with 3 HALL sensors.
Indicators
34 ORTHOPANTOMOGRAPH OP 3D
4 Device description
BOOT UP
DEVICE SOFTWARE
Ping i2000
Diagnose i2000
Initialize movements
ORTHOPANTOMOGRAPH OP 3D 35
4 Device description
NOTICE! The available imaging modalities vary depending on the device configuration.
36 ORTHOPANTOMOGRAPH OP 3D
5 Diagrams
5 Diagrams
5.1 Wiring diagram
CEPH UNIT
NASIO POSITION
J1019
J7003
6
SCAN HOME i5000 CASCADE BOARD
SCAN (S) i7000 TUBE HEAD
2PH
INTERFACE BOARD CARPUS DETECTION
J5005 J5001
J1024 J1020
J1118A/B 5
J7002 J7000
HEAD SUPPORT 4
3
TUBE
J1103 CEPH FAN OPTION
THA 1 2
J1007
J1020A/B J1019A/B
1000 BASE-T J1002
J1001
J4017 J4004 J4016 J1101
J4015
J1100 USB
(oponal)
DOOR SW
EXP. SW J4102 J4106
REM.EXP. SW
7
J4103
J4005
EXP. SW
REM.EXP. SW J4104
J4013
8
J4105 J4101
J4019
EXP. WARN
STATUS LEDs
i4100 BACK
PANEL BOARD
i4000 POWER
J4006
LASER SELF
2PH SUPPLY BOARD DIAGNOSTICS
J4009
J4018
J1022
LASER-HOME
J4020
EMERGENCY E
STOP SWITCH
J4002
J1039 J1011
J4022
J8000
Z POSITION
Z MOTOR
i8000 LED
DC J1037 INDICATOR
J4025 J4003 J4010 J4007 BOARD
J1010
J1009
J1008
i9000 JAW
J1008
i6000 USER
SENSOR
MIRROR OPTO INTERFACE
BOARD
LASER LASER LASER LASER
BOARD
CRADLE & LOWER SHELF
ORTHOPANTOMOGRAPH OP 3D 37
5 Diagrams
UPPER SHELF
J3004
J3012
6 LINEAR ABS
J1017
J3001
4
J3005
J1018
J3008
PIVOT HOME J3011 J3013
1 J1104 J1105
PIVOT LINEAR
2
LINEAR
MOTION J1016 J1015 MOTION
2PH 2PH
ABS.
ROT. HOME ANGLE
J1107
J1106
ROTATING
2PH J1012 J1027 J1013
J2003
J2002 J2001 J2010 J2000 J2011 J2004
J2005
7 CUB HOME
J1031
SELF
J7005
J7001 J7003 J1032
CLB HOME
COL LEFT
J2003 J2013 J2012 J2015 J2017 J2007
J5003 J5002 BORDER J1035
2PH
DETECTOR
J5005 J5001 SWAP (D)
LIMIT 3D LIMIT PANCEPH
J1033 2PH
CRB HOME
J0601 J0100 J0300 J0200 J0400 J0500
COL RIGHT J0101 J0301 J0201 J0401 J0501
BORDER J1036
2PH 3D 2D
TUBE
38 ORTHOPANTOMOGRAPH OP 3D
5 Diagrams
5.2 Fusing diagram
i4000 POWER SUPPLY i2000 IMAGE CAPTURE
F3
i4100 BACK PANEL BOARD Motors
GENERATOR
F1 F6
L Logic
F1
MAINS INLET
N PFC
DC/DC
F2 i3000 MAIN CONTROL
SUPPLY F1
F12 SWITCH Motors
SW1 STAND-BY
SUPPLY F2
EXT. WARNING LIGHT F3 Logic
F11
5.3 Emergency stop
Emergency stop button
J1011
10 3 1 3
J4002
i4000
11 6 3 1
1A 2A 2 1
4 2 1 3 J4107 L N
MAINS CORD J4106
F1
L 2 2
LIVE
NEUTRAL N 1 1
PROTECTIVE EARTH F2
J4018 J4001
SW1
i4100 i4000
ORTHOPANTOMOGRAPH OP 3D 39
5 Diagrams
J3002 J3003
F1
13 13 1 1 1 1 P24V_MOT
i4100 i4000
8 8 2 2 P24V_MOT 2 2 F2
P24V ON F1
J4101 J4013 J4020 P24V J2000
J1005 24V
13 13
THA
8 8
J7000
5.6 Ethernet communication
BACK PANEL
Link Link Link
OK OK OK
LAN
J3006
J3007
J1002
i3000
J2011
i2000 or
MWS
5.7 CAN communication
J3010
J3009
J2010
J4009
1 1 1 1
i2000 i3000 i4000
2 2 2 2
10 THA
5 PAN/3D
+24Vdc 2 6 12 5
J4019
+5Vdc +5Vdc +5Vdc +5Vdc +24Vdc +24Vdc
1 1 1 1 i4000 2 2
i7000
J4016
J2004
J3001
2
J3008
2
J4006
2 2 6 9
J7000
i2000 i3000
3 3 3 3 13 10 THA
4 4 4 4 5 5 CEPH
+5V ON
40 ORTHOPANTOMOGRAPH OP 3D
5 Diagrams
5.9 Rotation movement
MOTOR
CAN BUS ERROR
J2010
FROM
i3000 J1012
J2002
M
i2000 2ph Step
J2001
ROT. HOME
J1027
ROT. ABS.
J1013
5.10 Linear movement
CAN BUS
J3010
FROM
i4000 J1015
J3011
M
i3000 2ph Step
J3004
J
LINEAR HOME
J1025
LINEAR ABS.
J1017
ORTHOPANTOMOGRAPH OP 3D 41
5 Diagrams
5.11 Pivot movement
CAN BUS
J3010
FROM
i4000 J1016
J3011
M
i3000 2ph Step
J3005
J
PIVOT HOME
J1026
PIVOT ABS.
J1018
5.12 Z movement
Z-ACTUATOR
DC
From i3000
J4002
i4000
J6001
J4007
UI PANEL
+ J1011 TACOMETER
i6000
Z-HOME
E
J1039
42 ORTHOPANTOMOGRAPH OP 3D
5 Diagrams
5.13 Collimator movements
5.14 Positioning lights
LASER HOME LASER LIFT
M
2ph Step
J1022
J4018 3D FOV
HEIGHT
CAN BUS J1008
J4010
J4009
FROM J1009
i3000
POSTERIOR
ANTERIOR
i4000
SHARP
LAYER
J1003
J4003
J1010
MID
SAGITTAL
ORTHOPANTOMOGRAPH OP 3D 43
5 Diagrams
J2013
SENSOR
7 LAN
1
J1002
J2011
J3007
J3006
J101
8 i2000 i3000 or
2 MWS
Status IMG.DATA
J2015
J0301
IMG.MEM
LVDS
2
1 H10
11
J2012
+12V OK
2D 12
SENSOR LAN
J1002
J2011
J3007
J3006
i2000 i3000 or
MWS
IMG.DATA
J2017
J400
IMG.MEM
LVDS
Status
44 ORTHOPANTOMOGRAPH OP 3D
5 Diagrams
5.17 Exposure control 1
J1005
J4019 1 1 1 J7000
2 2 2
3 3 3
4 4 4
5 5 5 THA PAN
7 7 7 i7000
11 11 11
8 8 8
1 1 1
J4005 J7002
2 2 2
J1004
i4000
J4016 1 1 J7000
2 2
3 3
4 4
5 5 THA CEPH
8 7 i7000
12 11
9 8
1 1
J4004 J7002
2 2
5.18 Exposure control 2
Exposure button
J4103 J4101 J4013
1 1 1
P5V
2 2 2
Remote exposure 3 3 i4000
P3V3
button J4014 6 6
1 7 7
2 8 8
Remote exposure
button
i4100
J4102
1
2
Door switch
J4102
7
8
ORTHOPANTOMOGRAPH OP 3D 45
5 Diagrams
i4000 7 7 J4105
8 8 1
12 12 2
F3 External
13 13 warning
lamp
J4007 J6001
1 1 i6000
2 2 Exposure
warning
8 8
9 9 Ready
6 6 Error
10 10
J4022 J8000
i8000
1 1
2 2 Exposure
warning
3 3
4 4 Ready
5 5 Error
6 6
J4007
FROM
J6001
i4000 i6000
i3000
46 ORTHOPANTOMOGRAPH OP 3D
5 Diagrams
J9001
6 5
10 6
J1014
FROM i3000 M
J2003
LIMIT
J1028 3D
LIMIT
J1029 PAN
ORTHOPANTOMOGRAPH OP 3D 47
6 Device settings
6 Device settings
The Service mode of the GUI is used when the device is being maintained. The Service mode gives you a
full access to functionalities such as the device logs and errors and firmware update.
The device has also an Installation mode with limited settings intended for the device installation.
When you have unlocked the Service mode, you can see both the Installation mode and the Service
mode items in the menu list.
1. Go to device settings on the device GUI.
3. The GUI informs that additional options have been added and goes to Service mode.
48 ORTHOPANTOMOGRAPH OP 3D
6 Device settings
6.2 Setting menus
6.2.1 Quality control
Shows a list of Quality Control (QC) programs, their completion status and the last completion date.
6.2.2 Calibrations
Shows a list of calibrations programs, their completion status and the last completion date.
For information on how to perform the calibrations, refer to the OP 3D Installation manual.
ORTHOPANTOMOGRAPH OP 3D 49
6 Device settings
6.2.3 Settings
DEVICE SETTINGS:
Friendly name Set a name for the device, which is shown next to the device settings
icon and on the imaging software.
3D Metal Artefact Enable or disable MAR, Metal Artefact Reduction. MAR is used to
Reduction reduce the effect of metals and other dense radiopaque objects on the
3D image.
(3D devices only)
It is recommended to keep MAR activated.
Force device language Change the language of the device GUI if needed.
It is recommended not to force the language of the GUI but change
the language profile of the workstation.
Default to Scout before 3D Enable or disable automatic Scout image activation when a 3D
imaging modality is selected. It is recommended to have the Scout
(3D devices only)
image taking enabled.
NOTICE! If the selection is set to "disabled", a Scout image
mode can still be activated manually in the 3D imaging
program selection view.
DAP corrections Change the DAP correction factor, which is used to calculate the
shown DAP value on the GUI.
The shown DAP value is calibrated in the factory and the correction
factor set to 1.0 as default and should not be altered in normal
conditions. The DAP correction factors are intended for fine tuning the
shown DAP value by service personnel based on DAP measurements.
NOTICE! Changing the DAP correction factor has no impact
on the device radiation output, it is used for DAP display
calibration purposes only.
There are separate DAP correction options for PAN, 3D and CEPH
imaging.
50 ORTHOPANTOMOGRAPH OP 3D
6 Device settings
PAN imaging using 2D Disable/enable PAN imaging using the 2D sensor. By default, the
sensor device uses the 2D sensor (if available) for Panoramic imaging.
NOTICE! When the sensor used for panoramic imaging is
changed, some calibrations need to be rerun.
Volume level Adjust the volume level (5-10) of the audible signals played by the
device.
NOTICE! Does not affect exposure warning signal.
Button volume level Adjust the volume level (0-10) of the audible signal played when a
button on the patient positioning panel is pressed.
The default volume level setting is 8.
Button sound Select the sound that is played when HOME, Test mode or Patient
positioning lights buttons on the patient positioning panel are
pressed.
Time zone Set the timezone to match the installation location. To change the time
zone, start to write the continent, capitol or state (US) and select the
correct selection from the drop down list.
ORTHOPANTOMOGRAPH OP 3D 51
6 Device settings
CALIBRATION INTERVALS:
Change the required calibration and QC interval for the User and for Service (full calibration) according to
the local regulations. The device will remind the user when user recalibration, service recalibration and new
QC retakes are required, according to this setting.
52 ORTHOPANTOMOGRAPH OP 3D
6 Device settings
INSTALLATION SETTINGS:
License If the user has purchased a licensable upgrade for optional imaging
programs, for example 9x14 3D FOV, they can be activated from this
field.
Type the license key to the field and click Save to activate the
program.
Reset device type When the device is permanently downgraded, for example CEPH
arm removed, click Reset to reset the device type to correspond to
the current configuration. The device can detect and update its type
automatically if modalities are added, but when they are removed, the
reset must be done manually.
When the device configuration has changed, device type reset
removes the device calibration data and the device must be fully
calibrated.
Select country Select the installation country of the device. The selection configures
the device according to country specific regulations.
NOTICE! This setting must ALWAYS be configured correctly
during the installation.
Available selections:
• Other: Default selection without any device restrictions.
• China: Restricts maximum selectable current to 15 mA.
• Germany: Spinal compensation for 2D QC programs.
• Canada-Ontario: Restricted 3D FOV sizes to 5 x 5, 6 x 8 and 8 x
8.
NOTICE! For PAN QC to be used with 1.8mm Cu filter and
CEPH QC to be used with 0.8mm Cu filter, select Germany.
Door Switch If an optional Door Switch has been installed to the device, select an
operation mode for it. The default state for Door Switch is Disabled.
There are two selectable operation modes for the door switch:
Prevent exposure - Prevents taking a new image while the door is
open.
Prevent and stop exposure - Prevents taking a new image and
stops the ongoing exposure if the door is opened.
ORTHOPANTOMOGRAPH OP 3D 53
6 Device settings
Demo mode Used for setting the device into demonstration mode where the
radiation production of the device is disabled while maintaining all
other functionalities.
NOTICE! Demo mode is intended for exhibition use only.
Z motor current limit (down) Adjust the maximum current used to drive the column down.
Increase the current if column movement down is not smooth and
stops constantly.
NOTICE! Available only in devices with i4000 circuit board
r18 and newer.
Device time Shows the time and date that the device has received from the
network.
Click Refresh to update the displayed Device time.
NOTICE! The time display doesn't update automatically, but
needs manual refresh.
Fetch new time from Click Start to fetch the time from the network again. This can be
network useful in cases where the device time is incorrect.
54 ORTHOPANTOMOGRAPH OP 3D
6 Device settings
Adjust the default mA and kV values for medium size patient preset. Other patient size selections are
changed relatively to the adjustment made.
Adjustments made to the 3D imaging program with standard resolution are applied relatively to other
resolution selections as well.
6.2.4 Testing
6.2.4.1 Exposure tests
The device has PAN, CEPH and 3D exposure test programs with the device movements disabled. These
programs can be used for radiation related measurements.
A radiation measuring equipment is needed for the radiation measurements. For more information on how to
do these tests, see chapter Exposure tests on page 79.
ORTHOPANTOMOGRAPH OP 3D 55
6 Device settings
6.2.4.2 Motion tests
Motion tests are used for testing the individual movements of the device. If the device has prompted an error
related to the device movements, use the movement tests to see if the cause for the error is fixed and to
verify that the device movements are smooth.
Main movements:
• Pivot
• Linear
• Rotator
Other movements:
• Laser
• Upper collimator
• Left collimator
• Right collimator
• 2D sensor
• CEPH secondary collimator
• CEPH tubehead
56 ORTHOPANTOMOGRAPH OP 3D
6 Device settings
Motion test page also shows an indication of the device position and graphs showing the device actuator
positions and opto sensor states. When the device is moved by hand or by the movement test programs,
the device position values and the icons for tubehead, sensor and upper shelf positions and graphs should
change accordingly. This information can be used to identify which part of the system is faulty in case of
main movement issues.
For more details on how to use these graphs to identify the source of movement issue, see chapter
Troubleshooting - Movement errors on page 88.
• Shows the current position of the device in numeral and graphic format.
• The picture of the device changes according to the position of the device and the values change
accordingly if there are no issues.
ORTHOPANTOMOGRAPH OP 3D 57
6 Device settings
Movement calibration tests can be used to run a single device movement at a time. This functionality can be
used to identify which device movement causes issues when the device movement calibration has failed and
to identify in which point the movement fails.
These tests are performed in the same manner as the regular device movement calibration and will produce
the same calibration error messages when an issue is found.
1. Select a movement to be tested.
2. Turn the device to HOME position by hand. The GUI shows graphs that tell when the device is in
correct position (all arrows on green).
58 ORTHOPANTOMOGRAPH OP 3D
6 Device settings
6.2.4.3 System tests
System tests are used for testing various system statuses and performance values. If the device has
prompted an error related to the system components, use the system tests to see if the cause for the error is
fixed or to help you determine the cause for the issue.
To start a test, press Start test button next to the selected test. After the test is complete the result is shown
on the GUI.
Door statuses
• Shows the detected status (OPEN / CLOSED) of the door switch (optional) and the mirror.
Sensor tests
• The test generates test images from the selected sensor.
• Frames 1 and 2 of the generated test image should look similar.
◦ If the frames 1 and 2 look different, the sensor may be faulty.
◦ If the frames 1 and 2 lack of gray shades, the sensor cables may be faulty.
ORTHOPANTOMOGRAPH OP 3D 59
6 Device settings
Bus tests
1. I2C bus test
• Test the functionality of I2C buses on the device to check if data is transferred correctly between
the circuit boards.
• If data is not transferred correctly between the circuit boards it may indicate that:
◦ Cable connection between the circuit boards is faulty
◦ Circuit board doesn't receive power
◦ Circuit board is faulty
• If there is an issue with data transfer within the system, the test shows the result with red color on
the failed section.
PASSED FAILED
PASSED FAILED
60 ORTHOPANTOMOGRAPH OP 3D
6 Device settings
• If there is an issue with the power distribution within the system, the circuit boards and their sections
turn red.
In this example, i2000 circuit board doesn't receive any power from i3000.
• Enabling device movements for the test helps you verify if the Ethernet cable between the i2000 and
i3000 board is faulty.
◦ Device movements require you to press the exposure button for the whole duration of the test.
ORTHOPANTOMOGRAPH OP 3D 61
6 Device settings
62 ORTHOPANTOMOGRAPH OP 3D
6 Device settings
6.2.5 Service
In the Service menu you can update the device firmware, drive the device to packing position, reset the
device calibrations to factory settings and give the device service commands.
Firmware update Update the device firmware. See more instructions in Firmware update
on page 66.
Factory settings If the device calibration data needs to be reset, click on the Reset
button and enter PIN code 1917. This will reset all device calibration
data but does not affect any other device setting such as user settings
and DAP correction factors.
Service command Allows the device to be given service commands through GUI. This
may be useful when the device is in error state where conventional
fixes do not work, for example when multiple circuit boards are
changed simultaneously and the board eeproms are out of sync.
Contact the manufacturer technical support for more details.
Drive column down Allows the column to be driven down for device maintenance if the
device is in error state.
Press Start button and press the exposure button down to drive the
column down. Release the exposure button to stop the device.
ORTHOPANTOMOGRAPH OP 3D 63
6 Device settings
6.2.6 Logs
In the Logs menu you can access device logs, error messages and diagnostics.
Device diagnostics Shows the active error message and 20 rows of log prior to the error for
easier problem solving.
Download Download the entire device log and data to the workstation.
The log files are named as in the following example:
OP3D_KT1234567_Log_YYYYMMDD_TTmm.tar
• OP3D = device type
• KT1234567 = device serial number
• YYYYMMDD = log storing date
• TTmm = log storing time
The device logs are exported in tar-format. This file format can be unpacked using file archiver software,
such as 7-zip. The .tar package contains several log files in a folder structure. The most relevant log file for
troubleshooting is var/log/error.
NOTICE! The log files can be opened with proper editor softwares, such as Notepad++.
NOTICE! If the system has the KaVo Connect cloud service installed, the device logs can be sent to
the manufacturer technical service for easier and faster troubleshooting using the manual reporting
tool.
64 ORTHOPANTOMOGRAPH OP 3D
6 Device settings
6.2.7 About
Exposure counters Shows the amount of exposures taken with the device.
ORTHOPANTOMOGRAPH OP 3D 65
7 Maintenance
7 Maintenance
The maintenance and calibration procedure intervals described here are minimum requirements and
recommendations. The maintenance and calibration procedures can be made more frequent and stringent
to comply with local regulations regarding the use and maintenance of dental X-ray devices.
7.1 Firmware update
The firmware update can be performed through the device GUI.
Update the firmware to the latest version always during annual maintenance and when functional device
parts are replaced.
Some circuit boards and their versions require specific Firmware version to function correctly. See the
compatibility table below.
4. Update the workstation SW, drivers and DTX Studio driver, if needed.
5. Go to device settings on the GUI.
66 ORTHOPANTOMOGRAPH OP 3D
7 Maintenance
ORTHOPANTOMOGRAPH OP 3D 67
7 Maintenance
7.2 Annual maintenance
Annual maintenance procedures must be performed annually to ensure that the performance and
functionality of the device remains consistent. Annual maintenance consists of visual inspections, functional
tests and calibrations.
68 ORTHOPANTOMOGRAPH OP 3D
7 Maintenance
NOTICE! You need to remove the upper shelf covers before this procedure. For instructions, see
Removing covers - UPPER SHELF on page 137.
1. Remove the screws (3 pcs) from the linear movement rail on the left side of the upper shelf.
2. Slide off the linear movement rail from the upper shelf.
3. Turn the linear movement screw to move movement sledge so that the screw hole on the upper shelf is
aligned with a bearing.
ORTHOPANTOMOGRAPH OP 3D 69
7 Maintenance
4. Check the tightness of the bearing through the screw hole. Tighten if needed.
8. Remove the screws (3 pcs) from the linear movement rail on the right side of the upper shelf
9. Slide the rail slightly out from the upper shelf until the larger holes on the rail are aligned with the screw
holes.
70 ORTHOPANTOMOGRAPH OP 3D
7 Maintenance
10. Turn the linear movement screw to move movement sledge so that the screw hole on the upper shelf is
aligned with a bearing.
11. Check the tightness of the bearing through the screw hole. Tighten if needed.
12. Repeat the process for all bearings.
13. Push the linear movement rail back in and attach and tighten the screws.
ORTHOPANTOMOGRAPH OP 3D 71
7 Maintenance
6. When the device movements stop check that the device stays firmly in position.
7. Press and hold down the exposure button.
8. Check the smoothness of the rotation, linear and pivot movements. Monitor the device operation
and observe for any excessive noises or evident vibrations of the device during its rotation and pivot
movements. Also, make sure that the rotation and pivot movements are continuous until the end of
exposure.
NOTICE! If the exposure button is released, the device movements should stop.
9. Press the HOME button to drive the device back to home position.
10. Monitor the device operation and observe for any excessive noises or evident vibrations of the device
during its rotation movement.
11. Do the same tests for PAN and CEPH imaging modalities. Monitor the operation of the 2D
(cephalometric) sensor. Are the movements smooth? Is excessive noise observed?
12. Drive the device Up/Down using the positioning panel.
13. Monitor the device during up and down movement and observe for any excessive noises or non-
smoothness of the movement.
NOTICE! Check that the upward movement stops at the set Z-movement limit.
14. Select PAN and 3D imaging modalities from the device GUI and press the Patient positioning lights
button. Check that the correct lights are lit according to the selected modality. For more information on
the positioning lights, see the OP 3D User manual.
16. Check that the device GUI indicates that the emergency stop switch has been pressed.
NOTICE! The device restarts after the emergency stop has been pressed.
17. Make sure that the device cannot be driven up/down, rotated by pressing the HOME button and that
an exposure cannot be taken while the emergency stop switch is activated.
18. Release the emergency stop switch by rotating it clockwise.
19. Check that the GUI indication disappears and that the movements work normally again.
20. Switch off the device from the power switch.
72 ORTHOPANTOMOGRAPH OP 3D
7 Maintenance
21. Check that the device and all of its parts are isolated from mains and all device functions are disabled.
7. When the exposure starts, check that all exposure warning indicators are working:
• Exposure warning indicator on the GUI
• Yellow exposure indicator light on the Positioning panel
• Yellow exposure indicator light on the Status indicator light on top of the carriage.
• Audible exposure warning signal
• Remote exposure switch (if in use)
• Warning light (if in use)
8. When the exposure ends, check that the indicator lights go out, exposure warning signal stops and a
program end tone is played.
9. Check that the taken image transfers to the imaging workstation.
NOTICE! See chapter Fuses on page 26 for details on the correct fuses.
To release a fuse cap, push it inwards and rotate counterclockwise with a flat head screwdriver.
Remove the fuse from the socket and replace it with a new one. Install the fuse by pushing the fuse cap in
and rotating it clockwise with a screwdriver.
The 10A and 15A fuses are physically of different sizes and have their own fuse caps. If you need to change
the fuse rating, also the fuse cap needs to be changed. The fuses are supplied with the corresponding fuse
caps.
ORTHOPANTOMOGRAPH OP 3D 73
7 Maintenance
7.2.7 Monitor calibration
It is recommended to calibrate or check a test image with the primarily used workstation monitor. Refer to
the monitor/software manuals or Display Calibration Kit for information on how to do this.
74 ORTHOPANTOMOGRAPH OP 3D
7 Maintenance
7.3 Calibrations
7.3.1 Calibration programs
The available calibration programs and their completion order may vary between device configurations. This
chapter describes the device calibrations for each device configuration. For information on how to perform
the calibrations, refer to OP 3D Installation manual.
Calibration program order per device configuration:
ORTHOPANTOMOGRAPH OP 3D 75
7 Maintenance
Circuit boards
Z-movement limits X
PAN QC X X X
3D QC X X X
CEPH QC X* X* X*
* If CEPH device
76 ORTHOPANTOMOGRAPH OP 3D
7 Maintenance
Collimator calibration X X
3D pixel calibration X* X
Lasers alignment X* X
PAN QC X X X X
3D QC X X X
CEPH QC X* X* X
ORTHOPANTOMOGRAPH OP 3D 77
7 Maintenance
Sec. col Pri. col Ceph arm Ceph head Ceph Ceph
assembly assembly half (either) support * tubehead tubehead
scan mech cables
CEPH QC X X X X X X
78 ORTHOPANTOMOGRAPH OP 3D
7 Maintenance
7.4 Exposure tests
The device offers PAN, CEPH and 3D exposure test programs with the device movements disabled. These
programs can be used for radiation related measurements.
A separate radiation measuring device/equipment is required for taking the radiation measurements.
NOTICE! The available programs depend on the device configuration.
1. Go to device settings.
1. Adjust the exposure test kV and mA settings using the + and - buttons.
ORTHOPANTOMOGRAPH OP 3D 79
7 Maintenance
2. Attach the measuring device in front of the sensor cover. The detector of the measuring device must be
positioned to the area where the X-ray beam hits. The center of this area is located approximately 2.5
cm from the left side of the sensor cover window.
If using 2D sensor:
The detector of the measuring device must be positioned to the area where the X-ray beam hits. The
2D sensor front cover has small markings on top and bottom edges of the sensor window that indicate
the aproximate location of the center of the beam.
NOTICE! The X-ray beam must fully hit the detector of the measuring device. See the
measuring device User manual for more information.
NOTICE! You may use a fluorescent screen to pinpoint the beam location more accurately.
NOTICE! When using the PAN testing program, the X-ray beam is adjusted wider (over 2 cm)
than in normal imaging programs to allow easier positioning of the measurement device.
3. Protect yourself from radiation.
4. Press and hold down the exposure button to start the test program.
80 ORTHOPANTOMOGRAPH OP 3D
7 Maintenance
5. When the exposure warning stops and the program end tone is played, the program is complete.
6. Review the measurement data and compare it to the values shown on GUI.
7. If the verification fails, ensure the alignment of the measurement device and redo the test program.
1. Adjust the exposure test kV and mA settings using the + and - buttons.
2. Attach the measuring device to the middle of the 2D sensor cover window using sticky tape. The
detector of the measuring device must be positioned to the area where the X-ray beam hits. The 2D
sensor front cover has small markings on top and bottom edges of the sensor window that indicate the
approximate location of the center of the beam.
NOTICE! You may use a fluorescent screen to pinpoint the beam location more accurately.
5. When the exposure warning stops and the program end tone is played, the program is complete.
6. Review the measurement data and compare it to the values shown on GUI.
7. If the verification fails, ensure the alignment of the measurement device and redo the test program.
ORTHOPANTOMOGRAPH OP 3D 81
7 Maintenance
1. Select the length of the radiation pulse, 5ms (LDT/standard), 10ms (Standard resolution) or Constant
(High/ENDO resolution).
2. Adjust the mA value using the + and - buttons.
3. Adjust the exposure time if using Constant mode.
4. Attach the measuring device to the middle of the sensor cover window.
7. When the exposure warning stops and the program end tone is played, the program is complete.
8. Review the measurement data and compare it to the values shown on GUI.
9. If the verification fails, ensure the alignment of the measurement device and redo the test program.
82 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
8 Troubleshooting
Image transfer speed very slow. Somewhere in the network, Make sure that the whole system
between the device and the (including network switches etc.)
acquisition workstation, Fast is using a Gigabit Ethernet (1Gb)
Ethernet (100Mb) connection is connection.
used.
Broadcom or Intel interface
controller are recommended.
Cannot connect to the device with Device and system IP 1. Check that the power LEDs
an imaging software. configurations are mixed up. and Ethernet LED are lit on
i3000 circuit board.
2. Connect Ethernet cable from
a workstation directly to
J3006 connector on i3000.
3. Reset the device IP address
according to Reset to Static
IP address on page 134.
4. Open a web browser.
5. Type address
https://192.168.1.10:9931
6. Device GUI opens up and
you are able to diagnose the
issue.
ORTHOPANTOMOGRAPH OP 3D 83
8 Troubleshooting
84 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
PAN or 3D images have stripe • The device is calibrated 1. Run Collimator calibration
artifacts. incorrectly. with all device covers
• Linear movement is not installed.
smooth due to worn out 2. Recalibrate the device.
mechanics. 3. Check that the linear
movement rails are clean and
no metal residue/chips are
present.
4. Replace linear movement
rails according to Replacing
the linear movement
mechanism on page 172.
Images have stripe artifacts. Device has been inactive for 4 1. Take following standard
or more weeks and needs to be panoramic images without a
"warmed up". patient:
• 63 kV, 3.2 mA
• 66 kV, 3.2 mA
• 70kV, 6.3 mA
• 85kV, 8 mA
2. Run all available QC
programs and check the
resulting images for any
artifacts.
Image quality not optimal due to Linear movement screw is 1. Test linear movement by
obstructed linear movement. installed improperly. hand. The movement should
be easy and smooth.
2. Loosen the linear movement
mechanism and reassemble
it according to Replacing the
linear movement mechanism
on page 172.
Images are elliptical or have Pivot movement screw is installed 1. Test pivot movement by
sawtooth pattern. improperly. hand. The movement should
be easy and smooth.
2. Loosen the pivot movement
mechanism and reassemble
it according to Replacing the
pivot movement mechanism
on page 179.
ORTHOPANTOMOGRAPH OP 3D 85
8 Troubleshooting
CALIBRATION ISSUES
Error message "Z motor current Current of the Z motor down 1. Adjust the Z motor current
limit down too low" is shown movement is set too low. limit (down) to a higher
during the Z-movement limits value. See chapter Settings
calibration. on page 50.
2. Redo the calibration.
CEPH calibrations fail constantly. CEPH arm is installed incorrectly 1. Check that all CEPH arm
or the column is not correctly screws are installed and tight,
attached to the floor causing the especially the arm alignment
device to move. screw with a conical washer.
2. Check that the CEPH arm is
aligned correctly according to
OP 3D Installation manual.
NOTICE! The arm
isn't horizontally level
if aligned correctly.
3. Check that the used floor
screws are not too small.
4. Check that all floor and wall
screws are installed and tight.
86 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
ERROR MESSAGES
GUI shows a message "Detach Issue in CEPH head support 1. Run system Voltage test and
carpus holder" even though the calibration. check that NASIO sensor is
carpus holder is not attached. functional.
2. Recalibrate the CEPH head
support.
MECHANICAL ISSUES
The device does not identify the Issue in CEPH head support 1. Recalibrate the CEPH head
CEPH head support or nasion calibration. support.
support position correctly.
ORTHOPANTOMOGRAPH OP 3D 87
8 Troubleshooting
Opto sensor, cable connections and the circuit boards are functional. The problem is most likely in the
movement wiper or the movement sensor.
1. Run movement calibration test for the movement in question. See chapter Motion tests on page 56 for
more information.
2. The test fails and stops at the position where error is encountered.
3. Check that the movement sensor is correctly attached to the device, there is no dirt or similar between
the wiper and the sensor and that the wiper presses on the sensor correctly.
B - Device position indicator does not move and the numeral values do not change
1. Activate Rotator, Pivot or Linear movement graph, depending on the movement for which the error
code was produced.
2. Check is the BLUE line drawn to the graph when the device is moved.
• If the line is not drawn at all, the circuit board controlling the movement or the cable from the circuit
board to the movement sensor is faulty.
• If the line is drawn but has bumps, cuts or is not drawn linearly, the issue is with the movement
sensor or the wiper.
3. Check is the GREEN line drawn to the graph when the device is moved.
• If the opto state does not change between 0 and 1 on the graph, the issue is with the opto sensor.
88 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
8.4 Fail-safe mode
If the device detects errors during the boot sequence, the device starts up in a fail-safe mode. This mode is
identified in the GUI view.
When the device enters fail-safe mode, the GUI shows the likely cause of the error that prevents the device
to start properly. You can also easily download the device logs from this view, for further investigation.
The device can identify at least the following issues:
• Sensor(s) not responding
• I2C bus is not working
• CAN bus is not working
• Ethernet bus between i2000 and i3000 circuit boards is not working
• Device serial number is not identified
• Device MAC address is not identified
• MCU on a main circuit board is not working
See chapter Fail-safe mode error messages on page 89 for instructions on how to fix the errors prompted
by the fail-safe mode.
In fail-safe mode, you can still access the device settings and enter Service mode to perform System tests
on the device. See chapter Testing on page 55 for more info.
NOTICE! In some cases, where the device serial number cannot be identified, the device shows
up on the imaging software with serial number KT9999999. Similarly, if the device's MAC address
cannot be identified, the default address of 00:16:BF:08:FF:FF is shown.
ORTHOPANTOMOGRAPH OP 3D 89
8 Troubleshooting
been solved. If you change multiple circuit boards at once the EEPROMs on the circuit boards may not
update correctly preventing the device to start at all.
CAN bus not working between Problem is caused most likely due to faulty connector soldering, faulty
i3000 and i2000 component on a circuit board or faulty connector pins.
Check cable connections:
• i3000 to i2000: J3009 - J2010
NOTICE! Use a bypass cable SP01572 between the
connectors above to see if the cable is faulty.
CAN bus not working between Problem is caused most likely due to faulty connector soldering, faulty
i3000 and i4000 component on a circuit board or faulty connector pins.
Check cable connections:
• i3000 to i4000: J3010 - J4009
NOTICE! Use a bypass cable SP01572 between the
connectors above to see if the cable is faulty.
Parts that may need to be replaced:
• i3000 SP01239 / 0.805.4968
• i4000 SP01231 / 0.805.4960
• CARRIAGE CABLE SET SP01230 / 0.805.4959
CAN bus not working between Problem is caused most likely due to faulty connector soldering, faulty
i3000 and both i2000 and i4000 component on a circuit board or faulty connector pins.
Check cable connections:
• i3000 to i4000: J3010 - J4009
• i3000 to i2000: J3009 - J2010
NOTICE! Use a bypass cable SP01572 between the
connectors above to see if the cable is faulty.
Parts that may need to be replaced:
• i2000 SP01245 / 0.805.4974
• i3000 SP01239 / 0.805.4968
• i4000 SP01231 / 0.805.4960
• MAIN CABLE SET SP01242 / 0.805.4971
• CARRIAGE CABLE SET SP01230 / 0.805.4959
90 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
I2C bus not working between Problem is caused most likely due to faulty connector soldering, faulty
i3000 and i2000 component on a circuit board or faulty connector pins.
Check cable connections:
• i3000 to i2000: J3008 - J2004
NOTICE! Use a bypass cable SP01572 between the
connectors above to see if the cable is faulty.
Parts that may need to be replaced:
• i2000 SP01245 / 0.805.4974
• i3000 SP01239 / 0.805.4968
• MAIN CABLE SET SP01242 / 0.805.4971
I2C bus not working between Problem is caused most likely due to faulty connector soldering, faulty
i3000 and i4000 component on a circuit board or faulty connector pins.
Check cable connections:
• i3000 to i4000: J3001 - J4006
NOTICE! Use a bypass cable SP01572 between the
connectors above to see if the cable is faulty.
Parts that may need to be replaced:
• i3000 SP01239 / 0.805.4968
• i4000 SP01231 / 0.805.4960
• CARRIAGE CABLE SET SP01230 / 0.805.4959
I2C bus not working between Problem is caused most likely due to faulty connector soldering, faulty
i3000 and both i2000 and i4000 component on a circuit board or faulty connector pins.
Check cable connections:
• i3000 to i4000: J3001 - J4006
• i3000 to i2000: J3008 - J2004
NOTICE! Use a bypass cable SP01572 between the
connectors above to see if the cable is faulty.
ORTHOPANTOMOGRAPH OP 3D 91
8 Troubleshooting
i2000 MCU firmware update failed Micro-controller unit on i2000 board is faulty.
Parts that need to be replaced:
• i2000 SP01245 / 0.805.4974
i3000 MCU firmware update failed Micro-controller unit on i3000 board is faulty.
Parts that need to be replaced:
• i3000 SP01239 / 0.805.4968
i4000 MCU firmware update failed Micro-controller unit on i4000 board is faulty.
Parts that need to be replaced:
• i4000 SP01231 / 0.805.4960
Imaging sensor(s) not responding No working sensors available. Problem is caused most likely due to
faulty cables between i2000 and sensor, faulty sensor or faulty i2000
board.
NOTICE! If the device has two sensors and this error
message appears it's more likely that i2000 board is faulty.
Check cable connections:
• i2000 to 2D/3D sensor: J2013 - J0101 (power)
• i2000 to 2D/3D sensor: J2013 - J0601 (power)
• i2000 to 2D/3D sensor: J2015 - J0301 (data)
• i2000 to 2D sensor: J2012 - J0201 (power)
• i2000 to 2D sensor: J2017 - J0401 (data)
92 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
Invalid or missing serial number Problem caused most likely by changing multiple main circuit boards
at once and their EEPROM hasn't updated correctly.
Invalid or missing MAC address No field service solutions. Contact the manufacturer for more support.
8.5 Error codes
When an error code appears on the GUI, the device will stop working and cannot be operated while the
error code remains on the display.
NOTICE! You can use a bypass cable SP01572 between the CAN and I2C connectors on the main
circuit boards to see if the cable connection between them is faulty.
UNCLASSIFIED
Code Problem Solution
E00002 Invalid error code, you Software error. Take logs and re-boot the device.
should not receive this.
See chapter Logs on page 64 for more details.
E30205 Error in wake up Clear the error and re-boot the device.
process.
ORTHOPANTOMOGRAPH OP 3D 93
8 Troubleshooting
E08403 Exposure button active Check that the switch is not pressed down.
on boot
Check cable connections:
• Exposure switch to i4100: J4103
• i4100 to i4000: J4101 - J4013
E08404 Remote exposure button Check that the switch is not pressed down.
active on boot
Check cable connections:
• Remote exposure switch to i4100: J4104
• i4100 to i4000: J4101 - J4013
94 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
ORTHOPANTOMOGRAPH OP 3D 95
8 Troubleshooting
96 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
ORTHOPANTOMOGRAPH OP 3D 97
8 Troubleshooting
E09021 Stepper thermal This error may appear during multiple device movements. Read
shutdown the device logs to see which motor has caused the error.
98 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
ORTHOPANTOMOGRAPH OP 3D 99
8 Troubleshooting
100 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
E09028 Stepper charge pump Read the device log to identify which main circuit board caused
failure the error.
ORTHOPANTOMOGRAPH OP 3D 101
8 Troubleshooting
102 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
PAN/3D:
Check cable connections:
• i4000 to PAN/3D Tubehead: J4019 - J1005 - THA
CEPH:
Check cable connections:
• i4000 to CEPH Tubehead: J4016 - THA
E10004 Invalid boost time Software error. Take logs and re-boot the device.
E10005 Invalid kV value See chapter Logs on page 64 for more details.
E10006 Invalid mA value
E10007 Invalid mA kV value on
prepare
ORTHOPANTOMOGRAPH OP 3D 103
8 Troubleshooting
PAN/3D:
Check cable connections:
• i4000 to PAN/3D Tubehead: J4019 - J1005 - THA
CEPH:
Check cable connections:
• i4000 to CEPH Tubehead: J4016 - THA
104 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
CEPH:
Check cable connections:
• i4000 to CEPH Tubehead: J4016 - THA
• i4000 to CEPH Tubehead: J4004 - THA
E10208 Calibration invalid pwm Software error. Take logs and re-boot the device.
E10211 Program unexpected See chapter Logs on page 64 for more details.
event
E10403 Calibration Unexpected
event
ORTHOPANTOMOGRAPH OP 3D 105
8 Troubleshooting
E11003 Invalid panel mode Software error. Take logs and re-boot the device.
E11004 Unknown event See chapter Logs on page 64 for more details.
E11005 Unknown reply
E11006 Receive buffer full
E11202 Too short reply
E11203 CRC check failed
E11204 Invalid token count
E11205 Invalid cropping settings
E11206 Capturix Failed
E11207 Unexpected capturix Parts that may need to be replaced:
boot
• i2000 SP01245 / 0.805.4974
106 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
E13011 FIFO0 overrun Software error. Take logs and re-boot the device.
E13012 FIFO1 overrun See chapter Logs on page 64 for more details.
E13013 Unkown status
ORTHOPANTOMOGRAPH OP 3D 107
8 Troubleshooting
E18404 Proximity disabled or no Software error. Take logs and re-boot the device.
pulses
See chapter Logs on page 64 for more details.
E18405 Drive aborted button
E18406 Calibration homing Parts that may need to be replaced:
aborted
• Z-OPTO SP01683 / 0.805.3411
108 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
E19003 Invalid state shift Software error. Take logs and re-boot the device.
See chapter Logs on page 64 for more details.
E19004 Self-diagnostic failed NOTICE! E19005 error may disappear when you power
E19005 Bus failing off the device and power it on again.
Check cable connections:
E19006 UI panel defected
• i6000 to i4000: J6001 - J4007
E19007 Continuous UI panel
error
Parts that may need to be replaced:
• LOWER SHELF UPPER COVER WITH CONTROL PANEL
SP01275 / 0.805.5004
• i4000 SP01231
E19402 ESD error ref Software error. Take logs and re-boot the device.
E19403 ESD error signal See chapter Logs on page 64 for more details.
E19404 ESD error proximity
E19405 ESD error bus
E19406 Slider disabled too large
change
E19407 Slider disabled all
channels active
E19408 Proximity active after Check that:
reset
• Patient positioning panel is clean and intact
E19409 Slider disabled proximity • Error doesn't reproduce after re-boot
active
Parts that may need to be replaced:
• LOWER SHELF UPPER COVER WITH CONTROL PANEL
SP01275 / 0.805.5004
ORTHOPANTOMOGRAPH OP 3D 109
8 Troubleshooting
E21207 Route - no valid route Software error. Take logs and re-boot the device.
found
See chapter Logs on page 64 for more details.
E21208 Goto ADC - Invalid start
position
E21209 Goto ADC - Opto hit too Actuator pushed externally
fast
Check cable connections:
• i3000 to Pivot movement sensor: J3005 - J1026
• i3000 to Pivot movement sensor: J3005 - J1018
110 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
ORTHOPANTOMOGRAPH OP 3D 111
8 Troubleshooting
E22207 Route - no valid route Software error. Take logs and re-boot the device.
found
See chapter Logs on page 64 for more details.
E22208 Goto ADC - Invalid start
position
E22209 Goto ADC - Opto hit too Actuator pushed externally
fast
Check cable connections:
• i3000 to Linear movement sensor: J3004 - J1025
• i3000 to Linear movement sensor: J3004 - J1017
112 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
ORTHOPANTOMOGRAPH OP 3D 113
8 Troubleshooting
114 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
E23211 Goto ADC - Edges end Parts that may need to be replaced:
not found
• ROTATING UNIT UPPER PART SP01330 / 0.805.5171
ORTHOPANTOMOGRAPH OP 3D 115
8 Troubleshooting
Check that:
• The pivot sensor wiper presses the sensor properly. If not, the
wiper is faulty.
Check that:
• The pivot sensor wiper presses the sensor properly. If not, the
wiper is faulty.
116 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
Check that:
• The pivot sensor wiper presses the sensor properly. If not, the
wiper is faulty.
E26207 Calibration invalid edge Software error. Take logs and re-boot the device.
See chapter Logs on page 64 for more details.
E26402 Position guard failed Move the upper shelf to HOME position by hand and re-boot the
device.
E26403 Timeout Software error. Take logs and re-boot the device.
See chapter Logs on page 64 for more details.
ORTHOPANTOMOGRAPH OP 3D 117
8 Troubleshooting
Check that:
• The linear sensor wiper presses the sensor properly. If not, the
wiper is faulty.
Check that:
• The linear sensor wiper presses the sensor properly. If not, the
wiper is faulty.
118 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
Check that:
• The linear sensor wiper presses the sensor properly. If not, the
wiper is faulty.
E27207 Calibration invalid edge Software error. Take logs and re-boot the device.
See chapter Logs on page 64 for more details.
E27402 Position guard failed Move the upper shelf to HOME position by hand and re-boot the
device.
E27403 Timeout Software error. Take logs and re-boot the device.
See chapter Logs on page 64 for more details.
ORTHOPANTOMOGRAPH OP 3D 119
8 Troubleshooting
E28207 Calibration invalid edge Software error. Take logs and re-boot the device.
See chapter Logs on page 64 for more details.
120 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
ORTHOPANTOMOGRAPH OP 3D 121
8 Troubleshooting
122 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
ORTHOPANTOMOGRAPH OP 3D 123
8 Troubleshooting
124 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
E54402 Button up during start Software error. Take logs and re-boot the device.
E54403 Unhandled event See chapter Logs on page 64 for more details.
E54404 Door opened during Door with the door switch installed was opened during the
exposure exposure and the exposure was stopped.
If the error occurs while door is shut, replace the door switch.
ORTHOPANTOMOGRAPH OP 3D 125
8 Troubleshooting
E57003 Received message with Software error. Take logs and re-boot the device.
no service registered
See chapter Logs on page 64 for more details.
E57004 Isr invalid service
E57005 Isr frame queue full
E57006 Isr interupt enabled failed
E57007 Buffer full force reboot
E57202 ACK timeout Update firmware to FW2.3.1 or newer. If error persists the error
E57203 ACK timeout dropped source is i2000, i3000 or i4000 circuit board.
Check cable connections:
E57205 New before end
• i3000 to i2000: J3009 - J2010
E57207 EOF
• i4000 to i3000: J4009 - J3010
E57404 Invalid ack
Parts that may need to be replaced:
• Carriage cable set SP01230
• Main cable set SP01242
• i2000 SP01245
• i3000 SP01239
• i4000 SP01231
E57204 Invalid sender Software error. Take logs and re-boot the device.
E57206 Mixed messages See chapter Logs on page 64 for more details.
E57208 Order
E57402 Cannot re enable
receiving
E57403 Extra ack
E57405 Failed to send message
retry
126 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
ORTHOPANTOMOGRAPH OP 3D 127
8 Troubleshooting
128 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
2D sensor:
Check that:
• H9 and H26 LEDs on i2000 light up (green) during
preparation. If not, i2000 is faulty.
ORTHOPANTOMOGRAPH OP 3D 129
8 Troubleshooting
2D sensor:
Check cable connections:
• i2000 to 2D sensor: J2017 - J0401 (data)
130 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
E62027 TCP DIRECT MODE Take logs and re-boot the device.
NOT ALLOWED
See chapter Logs on page 64 for more details.
ORTHOPANTOMOGRAPH OP 3D 131
8 Troubleshooting
132 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
ORTHOPANTOMOGRAPH OP 3D 133
8 Troubleshooting
8.5.2 Acknowledging errors
The device responds to error situations by reporting an error. Most errors are shown in a message box
on the GUI. Some errors can be reset by closing the message box and some require that the device is
rebooted.
If an error that requires rebooting occurs or the device fails to operate as described in the User manual:
1. Switch off the power.
2. Wait for a few seconds.
3. Switch on the power.
6. The device will beep 3 times to indicate that configuration change has been applied and the Ethernet
interface of the device is now configured as DHCP client.
7. Connect the Ethernet cable to network and let the device retrieve a new IP address.
Refer to the OP 3D Installation manual for more details on network configuration and how to configure the
communication link.
134 ORTHOPANTOMOGRAPH OP 3D
8 Troubleshooting
5. Within 30 seconds after the blue indicator light has started to blink, press and hold the patient
positioning lights button in patient positioning panel for approx.15 seconds.
6. The device will beep 3 times to indicate that configuration change has been applied and the Ethernet
interface of the device is now configured to a static IP address 192.168.1.10 and Subnet mask
255.255.0.0.
7. Connect the Ethernet cable to network or workstation.
8. Ensure the workstation has IP configuration from the same IP network address range as the device, for
example IP address 192.168.1.11 and Subnet mask 255.255.0.0.
Refer to the OP 3D Installation manual for more details on network configuration and how to configure the
communication link.
ORTHOPANTOMOGRAPH OP 3D 135
9 Part replacement instructions
The device cables have labels near the connectors that tell the connector number, for example J2016,
where the first 2 numbers indicate to which board they are connected to. In this example the cable connects
to board i2000 to a connector J2016.
The connector on the cable and the circuit board have always the same number.
136 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
NOTICE! Be careful not to damage the covers when removing them. Avoid excessive use of power.
2. Turn the rotating unit so that you can access the underside of the upper shelf.
3. Remove the screw from below the upper shelf.
4. Lift off the left cover half and then the right cover half from the upper shelf.
NOTICE! Be careful not to damage the clamps that hold the covers together.
ORTHOPANTOMOGRAPH OP 3D 137
9 Part replacement instructions
2. Turn the protruding parts of the lower shelf tray slightly and remove the lower shelf tray.
138 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
NOTICE! Be careful not to damage the mirror when you remove it.
ORTHOPANTOMOGRAPH OP 3D 139
9 Part replacement instructions
Detach the screw on the back panel cover and lift off the cover.
140 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
2. Pull the covers apart, turn them outwards and lift them off.
1. Remove the screws (6 pcs) from the top of the rotating unit.
ORTHOPANTOMOGRAPH OP 3D 141
9 Part replacement instructions
1. Remove the tubehead back cover by detaching the indicated screw and pull the cover outwards. The
screw has a screw cover cap over it.
1. Detach the indicated screws (4 pcs) that hold the cover on the tubehead frame.
NOTICE! Check the cable routing through the lips on the cover for easier re-installation.
142 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
1. Pull the left and right 2D/3D sensor collar cover halves downwards.
2. Pull the right side collar cover half downwards to unlock the teeth holding the covers together.
ORTHOPANTOMOGRAPH OP 3D 143
9 Part replacement instructions
2D ONLY DEVICES
1. Detach the 2D sensor cover screw.
2. Pull the sensor collar halves apart to detach them from the sensor.
3D DEVICES
1. Pull the 2D sensor collar cover downwards.
2. Detach the collar cover halves by lifting the left side cover upwards and pulling the covers apart.
144 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
4. Pull the sensor covers apart to detach them. To help unlocking the cover halves, squeeze the left side
cover edges while pulling the covers apart.
ORTHOPANTOMOGRAPH OP 3D 145
9 Part replacement instructions
3. Detach the positiong panel cable from the circuit board below the upper cover half.
NOTICE! Guide the positioning panel cable through the mounting clips on the cover when you
install the cover back.
146 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
4. Press on both sides to release the lower cover half and remove the lower cover.
1. Move the nasion support and ear rods to their outmost position.
2. Remove the head support mount top cover screws (2 pcs) and detach the cover.
3. Remove the screws (3 pcs) from the upper head support cover.
4. Pull the lower head support cover downwards from the ear rod slide openings to detach it from the
upper cover.
ORTHOPANTOMOGRAPH OP 3D 147
9 Part replacement instructions
6. Pull the left hand side of the upper head support cover over the head support mount.
7. Turn the head support to LAT position and remove the cover.
1. Remove the screw and the cover mounting washer and remove the tubehead upper cover.
NOTICE! Hold on to the tubehead back cover you remove the upper cover, otherwise the back
cover falls down.
148 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
3. Remove the screw that holds the tubehead front cover latch and slide the latch up.
NOTICE! Hold on to the tubehead front cover when you remove the latch to make sure that the
cover does not fall off.
4. Remove the tubehead front cover.
ORTHOPANTOMOGRAPH OP 3D 149
9 Part replacement instructions
1. Detach the emergency stop switch cable connector from the column.
NOTICE! Push the locking claws on the connector with a screwdriver to release it easier.
3. Unscrew the plastic collar holding the emergency stop switch in place.
150 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
Required tools:
• Hex key 5mm
NOTICE! Push the locking claws on the connector with a screwdriver to release it easier.
2. Unscrew the plastic collar holding the emergency stop switch in place.
ORTHOPANTOMOGRAPH OP 3D 151
9 Part replacement instructions
3. Measure the distance from the floor plate to the lower edge of the outer column cover.
• If more than 35 cm / 13.7 in, continue with the installation.
• If less than 35 cm / 13.7 in, the column must be driven up first. See instructions in Driving the Z-
motor up with a faulty sensor on page 155.
4. Isolate the device from mains.
5. Remove the Z-movement sensor housing screw.
152 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
6. Mark the current column tilt adjustment block location, for example by marking the locking screw
locations. This helps aligning the device back to its original position later.
8. Loosen the lower and tighten the upper tilt adjustment screw to give more room for the sensor housing
to move.
9. Pull the top part of the Z-movement sensor housing off the column and slide it off the device.
NOTICE! Bend the sensor housing to get it past the cable clamp.
ORTHOPANTOMOGRAPH OP 3D 153
9 Part replacement instructions
10. Get the new spare part and make sure that the metallic Z-movement sensor cable cover is evenly in
place in the cutout.
11. Slide the new Z-movement sensor housing into the device.
14. Adjust the column tilt adjustment block to its original location.
154 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
16. Route the mains cord, exposure button cable, Ethernet cable and other optional cables through the
cable clamp and reattach the cable clamp with 2 screws.
2. Place a suitable object, credit card, paper etc., into the groove on the spare part, next to the sensor.
ORTHOPANTOMOGRAPH OP 3D 155
9 Part replacement instructions
5. Remove the object from the sensor once the device has moved up to suitable height.
CAUTION! If the object is not removed during the Z-movement, the column keeps driving up
until it hits mechanical limit or the ceiling.
NOTICE! You can drive the device up/down normally after the initialization.
6. Detach the spare part sensor from the device and continue to remove the old sensor housing.
156 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
ORTHOPANTOMOGRAPH OP 3D 157
9 Part replacement instructions
CAUTION! After i4000 failure, wait for minimum 10 minutes before removing the circuit board. Some
components are charged for a while after the fuse F1 has blown and can cause electric shock if
touched.
NOTICE! Always update the device firmware before replacing the i4000 power supply board.
158 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
• Remove all carriage covers. See chapter Removing covers - CARRIAGE on page 138.
• Remove the horizontal lasers. See chapter Replacing the horizontal lasers on page 157.
NOTICE! Take a picture of the cables attached to the i4000 circuit board before you detach any of
them. This will help you to re-attach the cables.
2. Detach cable J4001 from upper right corner of the i4000 circuit board.
3. Proceed to detach the cables J4019, J4016, J4023, J4007, J4003, J4025, J4009, J4006, J4013, and
J4020 from the bottom of the i4000 circuit board.
4. Detach also the cables J4017, J4022, and J4002 on the left side at the bottom of the i4000 circuit
board.
NOTICE! If the device has a cephalometric arm, detach also cable J4005.
5. Loosen the screws (2 pcs) at the top and remove screws (2 pcs) the at the bottom of the i4000 board.
ORTHOPANTOMOGRAPH OP 3D 159
9 Part replacement instructions
NOTICE! The i4000 board plate is attached to the carriage with same screws as the i4000 fan
and its cover.
7. Put the new i4000 board plate and the i4000 fan in place.
8. Re-install the screws (4 pcs) holding the i4000 board and board plate in place.
9. Connect all the previously detached cables.
NOTICE! Connect cables to the connectors with the same number on the i4000 board.
160 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
ORTHOPANTOMOGRAPH OP 3D 161
9 Part replacement instructions
1. Detach the back panel plate by removing the indicated screws (2 pcs).
162 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
6. Install the new mains switch by following the steps in reverse order.
7. Check that the mains switch works properly.
2. Turn the flaps of the i8000 LED indicator board slightly to remove the board and its cover from the
carriage.
ORTHOPANTOMOGRAPH OP 3D 163
9 Part replacement instructions
3. Disconnect the cables going down from the i3000 main control board.
4. Detach the screw from the cable clamp and insert a screwdriver between the cable clamp and the
device body to lift the clamp.
164 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
ORTHOPANTOMOGRAPH OP 3D 165
9 Part replacement instructions
NOTICE! Detach the grounding cables from cables J3007, J3006, and J3006.
2. Remove the SD card from the old i3000 main control board and keep the SD card , you will need it
later.
3. Remove the screws (4 pcs) and lift off the i3000 main control board.
166 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
4. Put the new i3000 main control board to its place on the upper shelf.
NOTICE! Make sure that you put the new i3000 main control board to its place the right way.
Check that the numbers of connectors match the ones on the cables.
5. Attach the new board in place with the screws (4 pcs).
6. Insert the SD card from the old i3000 main control board to the SD card retainer of the new board.
7. Connect all the cables.
8. Run the required calibrations. See chapter When to calibrate the device on page 76.
9.7.3 Removing the i3000 main control board with the support plate
Before you begin:
• Remove the upper shelf covers. See chapter Removing covers - UPPER SHELF on page 137.
ORTHOPANTOMOGRAPH OP 3D 167
9 Part replacement instructions
1. Detach all cables (J3002, J3003, J3005, J3004, J3011, J3009, J3008, J3007, J3006, J3012, J3001,
J3010) from the i3000 main control board.
NOTICE! Detach also the grounding cables of cables J3007, J3006, and J3006.
168 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
1. Pull the speaker off the upper shelf. The speaker is not fixed in place with screws.
ORTHOPANTOMOGRAPH OP 3D 169
9 Part replacement instructions
2. Install the new linear movement sensor wiper assembly in its place by pushing it in.
NOTICE! Make sure that the wiper is locked firmly to its place.
3. Run the required calibrations. See When to calibrate the device on page 76.
2. Pry the linear movement sensor wiper off from the linear movement sledge.
170 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
3. Mark the exact place of the linear movement sensor using a marker pen.
5. Remove the old linear movement sensor and attach the new linear movement sensor to the exact same
spot.
NOTICE! Make sure that the sensor is evenly attached to the surface.
ORTHOPANTOMOGRAPH OP 3D 171
9 Part replacement instructions
3. Remove the opto sensor housing with the opto sensor from the upper shelf.
4. Replace the entire opto sensor housing with the opto sensor or replace the opto sensor with a new one.
5. Reinstall all parts in reverse order.
Required tools: Hex key 3mm and 5mm, spanner 15mm, pliers
1. Detach the cable from the linear movement motor.
2. Remove the retaining nut holding the linear movement screw in the coupler.
172 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
5. Pull the linear movement screw with the compression spring out of the coupler.
6. Remove the linear movement sledge with the linear movement screw from the upper shelf.
7. Detach the linear movement motor fixing screw.
8. Push the motor slightly backwards and remove it from upper shelf.
ORTHOPANTOMOGRAPH OP 3D 173
9 Part replacement instructions
9. Detach the screws (2 pcs) on linear bearing housing and remove the linear bearing housing.
10. Detach the linear movement rail screws (3+2 pcs) on both sides of the upper shelf.
11. Remove the linear movement rails from both sides of the upper shelf.
12. Attach the new linear movement rails on both sides of upper shelf.
13. Put the linear bearing housing in its place. Do not tighten the screw yet.
14. Put the new linear movement motor in its place.
NOTICE! Ensure that the motor sits on the black damper plate that is used for dampen the
vibration caused by the motor.
174 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
15. Check that the linear bearing housing and the motor are aligned.
16. Attach the screws (3 pcs) loosely to secure the motor and the linear bearing housing to upper shelf. Do
not tighten the screws yet.
17. Install the new linear movement sledge to upper shelf.
NOTICE! Lift the linear wiper slightly when you slide the linear movement sledge in so that you
do not damage the wiper.
18. Make sure that the small compression spring is in place at the end of the linear movement screw.
19. Attach and tighten the linear movement screw to the motor coupler.
ORTHOPANTOMOGRAPH OP 3D 175
9 Part replacement instructions
20. Fix the linear movement screw to the coupler with the retaining nut.
21. Check that the linear movement screw is able to move a bit in horizontal direction.
22. Move the linear movement back and forth by hand to make sure that the movement is smooth.
23. Tighten the top screws (2 pcs) of the motor and the linear bearing housing.
24. Tighten the side screw of the linear bearing housing gently, so that the linear bearing housing does not
move.
25. Attach the cable to the linear movement motor.
176 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
1. Detach the screws from the support bar at the back end of the upper shelf and then slide off the
support bar.
2. Detach the screws (2 pcs) from the pivot movement sensor wiper and lift it off.
3. Install the new pivot movement sensor wiper and attach it with the screws.
4. Slide the support bar back into its place and attach the screws.
5. Run the required calibrations. See When to calibrate the device on page 76.
ORTHOPANTOMOGRAPH OP 3D 177
9 Part replacement instructions
2. Mark the exact place of the sensor by using a marker pen before you remove it.
3. Remove the old pivot movement sensor and attach the new pivot movement sensor to the exact same
spot.
NOTICE! Make sure that the sensor is evenly attached to the surface.
7. Slide the support bar back into its place and attach the screws.
8. Run the required calibrations. See When to calibrate the device on page 76.
178 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
3. Remove the opto sensor housing with the opto sensor from the upper shelf.
4. Replace the entire opto sensor housing with the opto sensor or replace the opto sensor with a new one.
5. Reinstall all parts in reverse order.
ORTHOPANTOMOGRAPH OP 3D 179
9 Part replacement instructions
5. Remove the pivot movement motor housing screw and pull the motor out.
7. Detach the retaining nut from the coupler of the pivot motor.
180 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
8. Unscrew the coupler from the pivot movement screw. Use a spanner/wrench to turn the coupler and
hold the other end of the pivot movement screw in place with a spanner/wrench or pliers.
9. Pull the pivot movement screw out of the bearing housing and remove the spring.
NOTICE! To get more room to work you can also detach the cogwheel.
12. Turn the upper shelf so that the pivot movement sledge moves out to the open area between the
cogwheels (min. 45°).
ORTHOPANTOMOGRAPH OP 3D 181
9 Part replacement instructions
13. Remove the chain from the slot in the pivot movement sledge.
NOTICE! One end of the chain link inside the slot sits slightly further out. Pry the chain from that
side to get it out easier.
14. Drop the chain off the large cogwheel.
15. Slide the pivot movement sledge assembly off the rail.
18. Remove the pivot movement guiding rail screws (2 pcs) from the upper shelf.
19. Remove the pivot movement guiding rail and the pivot bearing housing from the upper shelf.
Installing the new pivot movement mechanism parts:
182 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
20. Place the he pivot movement guiding rail and the pivot bearing housing onto the upper shelf. The pivot
bearing housing has a tab that goes to the slot on the edge of the upper shelf.
21. Attach the rail of the pivot movement to the upper shelf with 2 screws.
22. Attach the pivot bearing housing to the upper shelf with 3 screws.
23. Slide the pivot movement sledge to the rail all the way next to the small cog wheel.
NOTICE! Make sure that the carrier moves smoothly in the rail.
ORTHOPANTOMOGRAPH OP 3D 183
9 Part replacement instructions
24. Put the chain around the large cogwheel. The chain master link must be on top of the chain and
located on the opposite side of the upper shelf than the pivot movement sledge as shown on the picture
below.
25. Move the pivot movement sledge adjacent to the large cogwheel and turn the upper shelf so that the
arrow on the cogwheel and the line on the sledge are facing each other.
26. Push the chain into the slot between the carrier and the cogwheel.
184 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
28. Attach and tighten the tension spring to tighten the chain.
29. Check that the arrow on the cogwheel and the line on the sledge are still facing each other to ensure
that the chain timing is correct.
32. Push the pivot movement screw through the bearing housing.
ORTHOPANTOMOGRAPH OP 3D 185
9 Part replacement instructions
33. Attach the coupler to the pivot movement screw. Use a spanner/wrench to turn the coupler and hold
the other end of the pivot movement screw in place with a spanner/wrench or pliers.
186 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
NOTICE! Turn the motor axle so that the flat side is facing the coupler retaining nut.
40. Move the upper shelf slowly back and forth and check that the pivot movement is smooth.
41. Reattach the pivot movement opto sensor housing to the upper shelf.
42. Reinstall all other removed parts and reattach all cables.
43. Run the required calibrations. See chapter When to calibrate the device on page 76.
ORTHOPANTOMOGRAPH OP 3D 187
9 Part replacement instructions
1. Disconnect the connector from the i9000 chin rest sensor board.
188 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
ORTHOPANTOMOGRAPH OP 3D 189
9 Part replacement instructions
6. Insert and tighten the screw that holds the laser in place.
190 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
Replace the upper lower shelf cover according to chapter Removing the lower shelf upper and lower cover
on page 146.
ORTHOPANTOMOGRAPH OP 3D 191
9 Part replacement instructions
6. Push the head support locking lever locking pin up from the bottom of the lower shelf and remove the
pin.
8. Remove the axle from the lower shelf. Press the clips on the holders inwards to release them from the
lower shelf.
192 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
10. Push the rotational damper into the end of the axle.
11. Attach the washer and the bushing to the other end of the axle.
12. Attach the axle to the lower shelf. Make sure that the axle holder clips lock the axle in place.
13. Attach the head support locking lever, axle spring, axle fixing screw and axle damper plate in reverse
order that they were removed.
14. Re-attach the tilt light connector and the i9000 board.
ORTHOPANTOMOGRAPH OP 3D 193
9 Part replacement instructions
2. Detach the connectors J4003 and J4007 from i4000 board on the carriage.
3. Detach the lower shelf fixing screws (3 pcs) from the carriage.
194 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
1. Detach the cables J2000, J2004, J2019 (main cable set), J2002, J2003, J2011 (Ethernet), J2015 (3D
sensor), J2017 (2D sensor), J2012, J2013.
ORTHOPANTOMOGRAPH OP 3D 195
9 Part replacement instructions
3. Remove the screws (3 pcs) holding the collimator on the tubehead; one on the right side, one on the left
side and one below the collimator.
196 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
10. Adjust the collimator if needed. See instructions in PAN/3D Collimator adjustment on page 236.
11. Re-attach all device covers and re-calibrate the collimator again.
2. Detach the THA cables from the tubehead and the i7000 board.
NOTICE! The cable routing and clamping differ between main cable versions.
ORTHOPANTOMOGRAPH OP 3D 197
9 Part replacement instructions
5. Install the new tubehead in reverse order and connect all cables and their grounding clamps as shown
before.
6. Attach the secondary collimator, if the device has a cephalometric arm. See chapter Replacing the
CEPH secondary collimator assembly on page 216.
7. The device must be completely recalibrated after this procedure. See chapter When to calibrate the
device on page 76.
1. Detach the tubehead hanger bar fixing screws (2 pcs) from the rotating unit.
198 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
1. Detach the grounding cable from the i2000 image capture board.
5. Detach the tubehead cable connectors J1004 and J1005 from the upper shelf.
ORTHOPANTOMOGRAPH OP 3D 199
9 Part replacement instructions
6. Remove the screw from the cable clamp and turn the clamp up.
7. Move the rotating unit so that it is in a 90 degree angle with the upper shelf and push it against the
column.
NOTICE! The tubehead must be in home position (on the right hand side of the device).
8. Push the cable clamp lid down and pull it open to release the main cable.
9. Remove the screws (4 pcs) on the linear sledge that hold the rotating unit.
200 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
10. Lift the rotating unit and slide it off from upper shelf. Lay the rotating unit carefully on the floor.
ORTHOPANTOMOGRAPH OP 3D 201
9 Part replacement instructions
NOTICE! The correct position of the cable is indicated with markings on the cable.
16. Turn the metallic cable holder down to keep the cable in place.
202 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
17. Lift the rotating unit back to the grooves in the upper shelf.
NOTICE! The Linear movement sledge and the rotating unit hanger must be straight for them to
fit together.
NOTICE! 2 persons are recommended for this task.
NOTICE! Check that the main cable does not get squeezed between the rotating unit and the
upper shelf.
18. Re-attach the rotating unit to the linear sledge with 4 screws.
20. Connect all connectors to i3000 and i2000 boards and attach the i2000 grounding cable.
ORTHOPANTOMOGRAPH OP 3D 203
9 Part replacement instructions
21. Attach the THA cables to the tubehead with clamps and connect them to i7000 board.
4. Remove the screws (2 pcs) that hold the main cable set support.
204 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
7. Insert a tool into the openings of the rotating unit upper part to push and release rotating sensor from its
fastening.
10. Re-assemble the device by following the steps above in reverse order and re-connect all cables.
11. Run the required calibrations. See chapter When to calibrate the device on page 76.
ORTHOPANTOMOGRAPH OP 3D 205
9 Part replacement instructions
206 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
NOTICE! Hold on to the sensor because it can fall when all screws are removed.
1. Detach the 2D sensor cover plate screws (2 pcs) and remove the cover plate.
ORTHOPANTOMOGRAPH OP 3D 207
9 Part replacement instructions
4. Detach the screws (6 pcs) from the 2D sensor and remove the sensor.
1. Detach the 2D sensor screws (4 pcs) and remove the sensor back cover.
NOTICE! Hold on to the sensor when you remove the screws so that the sensor cover does not
drop down.
208 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
2. Detach the cable clamps and detach the connectors from the 2D sensor.
4. Remove the rotating unit finger guard by pushing on the clips and pulling down the finger guard.
ORTHOPANTOMOGRAPH OP 3D 209
9 Part replacement instructions
6. Detach the cable holder and cable clamps from the upper part part of the rotating unit.
8. Slide the 2D sensor cable set away from the rotating unit.
NOTICE! Slide the cable set away from the rotating unit starting from the i2000 image board
end downwards as indicated by arrows in the picture.
9. Install the the new 2D sensor cable set by following the steps in reverse order.
10. Run the required calibrations, see When to calibrate the device on page 76.
210 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
1. Detach the 2D/3D sensor power cables and the grounding cable.
ORTHOPANTOMOGRAPH OP 3D 211
9 Part replacement instructions
5. Detach the sensor holder assembly screws (2 pcs on both sides of the i2000 circuit board).
6. Remove the old sensor holder assembly by sliding it off from the rotating unit.
7. Slide the new sensor holder assembly to the rotating unit.
8. Attach the new sensor holder assembly (2 screws on both sides of the i2000 circuit board).
9. Attach the previously removed sensor mount to the new sensor holder assembly.
10. Run the required calibrations. See chapter When to calibrate the device on page 76.
212 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
1. Turn the 2D sensor movement screw with a flat head screwdriver or 3mm hex key to move the 2D
sensor to the middle position.
1. Remove the i2000 image capture board. For instructions, see Replacing the i2000 image capture board
on page 195.
2. Remove the sensor holder assembly and sensor(s). For instructions, see Replacing the sensor holder
assembly on page 210.
3. Remove the tubehead. For instructions, see Replacing the tubehead on page 197.
4. Remove the rotating unit (upper part) as instructed in Replacing the main cable set on page 199.
5. Install the new rotating unit (upper part) by following the instructions in Replacing the main cable set on
page 199 in reverse order.
6. Install the tubehead back by following the instructions in Replacing the tubehead on page 197 in
reverse order.
ORTHOPANTOMOGRAPH OP 3D 213
9 Part replacement instructions
7. Install the sensor holder assembly and sensor(s) back by following the instructions in Replacing the
sensor holder assembly on page 210 in reverse order.
8. Calibrate the device. See chapter When to calibrate the device on page 76.
1. Push the motor mount at the end of the rotating unit frame inwards to loosen the belt and drop the belt
off the sprocket.
3. Pull the motor mount and spring off the rotating unit frame.
4. Put the new motor mount and spring to the rotating unit and connect the motor cable.
5. Push the motor mount inwards and lift the belt up around the sprocket.
NOTICE! Make sure that the belt is tight and has not fallen off the large sprocket.
6. Run the required calibrations. See chapter When to calibrate the device on page 76.
214 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
2. Clean the surface from any remaining adhesive with ethanol or isopropanol.
5. Attach the new window carefully to the cutout on the tubehead cover. Start by placing the top edge of
the window in place first.
6. Make sure that the window is correctly in the cutout and is not misaligned or over the edge.
7. Apply strong pressure on the edges of the window in circular motion.
ORTHOPANTOMOGRAPH OP 3D 215
9 Part replacement instructions
8. Apply more strong pressure on the bottom edge of the window for minimum 60 seconds.
1. Detach the screws (2 pcs above and one below) that hold the primary collimator.
216 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
ORTHOPANTOMOGRAPH OP 3D 217
9 Part replacement instructions
6. Detach the indicated screws (2 pcs) and remove the secondary collimator assembly from the tubehead.
7. Take the new secondary collimator assembly and detach the collimator plate by removing the indicated
screws (2 pcs) and loosening the vertical alignment screw (circled).
218 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
14. Adjust the secondary cephalometric collimator. See CEPH secondary collimator adjustment on page
240.
ORTHOPANTOMOGRAPH OP 3D 219
9 Part replacement instructions
15. Re-attach the tubehead covers and the upper shelf covers.
16. Run the required calibrations. See chapter When to calibrate the device on page 76.
220 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
2. Detach the tubehead cables by pushing on the locking clip and pulling the connector off.
3. Remove the screws (2 pcs) on top of the tubehead by using a 5mm hex key.
NOTICE! Hold on to the tubehead when you remove the screws. Notice that the tubehead is
heavy.
4. Slide the tubehead off from its retainer.
ORTHOPANTOMOGRAPH OP 3D 221
9 Part replacement instructions
Required tools:
• Hex key 3 mm
2. Remove the screws (6 pcs) that hold the scanning mechanism in place.
4. Follow the steps in reverse order to install the new scanning mechanism.
5. Install the CEPH tubehead back on.
6. Run the required calibrations, see When to calibrate the device on page 76.
222 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
2. Remove the CEPH tubehead scan opto sensor mounting plate screws (2 pcs).
1. Detach the mirror opto cable and loosen the mirror base plate screws (3 pcs).
ORTHOPANTOMOGRAPH OP 3D 223
9 Part replacement instructions
3. Detach the cable clamps of the ceph tubehead cables J1007, J7000, J7002 inside the carriage and
detach the connectors.
CAUTION! Make sure that the power cannot be swiched on while you carry out this procedure.
NOTICE! You must remove the carriage side covers and the mirror before you can replace this
part.
4. Detach the groove strip from the rear of the cephalometric arm.
5. Detach the cable clamps and detach the connectors from the tubehead.
224 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
9. Detach the cable holder from the tubehead scanning mechanism support.
10. Open the cable holder that hold the cable set inside the cephalometric arm.
NOTICE! Lightly push on the upper part of the holder and then pull it open.
ORTHOPANTOMOGRAPH OP 3D 225
9 Part replacement instructions
11. Detach screw holding the cable cover plate and pry the cover plate off.
12. Detach the lower connector inside the cephalometric arm opening and pull the cable off the cable
clamp.
13. Pull the tubehead cable set out from the carriage and through the opening on the cephalometric arm.
14. Push the new cable set inside the carriage and then insert the cable set to the cephalometric arm.
15. Attach all connectors and cable clamps of the new tubehead cable set and guide the cables through
the cable holders.
16. Run the required calibrations, see When to calibrate the device on page 76.
226 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
1. Detach the groove strip from the rear of the cephalometric arm.
2. Detach screw holding the cable cover plate and pry the cover plate off.
3. Detach the lower connector inside the cephalometric arm opening and detach the cable from the cable
clamp.
4. Detach the mirror opto cable and loosen the mirror base plate screws (3 pcs).
ORTHOPANTOMOGRAPH OP 3D 227
9 Part replacement instructions
7. Detach the tubehead cables (3-pin connector, 14-pin connector and grounding cable) inside the
carriage.
8. Pull the tubehead cable set out from the carriage and through the opening on the cephalometric arm.
228 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
NOTICE! There are holes on the bottom of the cephalometric arm that allow you to use a long
hex key to unscrew the screws.
10. Remove the lower cephalometric arm joint screws (2 pcs).
11. Remove the arm alignment screw with a conical washer to the cephalometric arm joint bar.
12. Turn the straight part of the cephalometric arm inwards and then pull it off.
NOTICE! A latch inside the straight part of the cephalometric arm locks the arm in place, push
on the latch to release the locking.
13. Install the new straight part of the cephalometric arm back in place by following the steps in reverse
order.
14. Run the required calibrations. See When to calibrate the device on page 76.
ORTHOPANTOMOGRAPH OP 3D 229
9 Part replacement instructions
1. Detach the upper connector (J1019A) from the carriage inside the cephalometric arm opening.
2. Detach the 4 bolts and washers that hold the arched part cephalometric arm in place.
5. Install the new CEPH arm flange. Tighten the installation bolts to 25 Nm torque.
6. Install the new arched part of the cephalometric arm in place.
7. Install the straight part of the cephalometric arm in place and connect all related cables and cable
clamps.
8. Run the required calibrations, see When to calibrate the device on page 76.
230 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
1. Detach the groove strip from the rear of the cephalometric arm.
2. Detach screw holding the cable cover plate and pry the cover plate off.
3. Detach the upper connector (J1019A) from the carriage and the cable clamps holding the cables inside
the cephalometric arm opening.
ORTHOPANTOMOGRAPH OP 3D 231
9 Part replacement instructions
4. Push on the retainer clip to release the head support mount lower cover and then pull it down.
NOTICE! This step is optional. You can pull the lower cover down to get more working space.
NOTICE! Hold on to the head support when you remove the screws.
232 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
9.11 Mechanical adjustments
NOTICE! The 2D/3D sensor covers must be removed before this procedure. For instructions, see
Removing the 2D/3D sensor covers on page 143.
1. Loosen the 2 sensor height locking screws.
2. Turn the adjustment screw on to adjust the height of the sensor. One turn is equivalent of 0.7 mm.
ORTHOPANTOMOGRAPH OP 3D 233
9 Part replacement instructions
2. Turn the height adjustment screw to adjust the height of the sensor. One turn is equivalent of 0.7 mm.
234 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
2. Turn the upper sensor position adjustment screw to adjust the sensor tilt. You can see the amount of tilt
adjusted from the alignment holes.
2. Move the sensor sideways by turning the sensor position adjustment screws (2 pcs).
ORTHOPANTOMOGRAPH OP 3D 235
9 Part replacement instructions
3. Check that both ends of the sensor have moved evenly from the sensor alignment holes. The holes are
equally visible when the sensor is vertical.
1. To move the collimator in UP/DOWN direction, turn the collimator height adjustment screw through the
cutout (hex key 2.5mm):
a) Clockwise to move the collimator downwards.
b) Counter clockwise to move the collimator upwards.
NOTICE! The GUI informs you how much the collimator needs to be adjusted and in which
direction. One full turn of the screw (360°) equals about 0.7 mm.
236 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
2. To rotate the collimator CLOCKWISE (CW)/COUNTER CLOCKWISE (CCW), loosen the 2 upper
locking screws (1), the lower locking screw (2) and turn the adjustment screw (3):
a) Clockwise to turn the collimator clockwise.
b) Counter clockwise to turn the collimator counter clockwise.
NOTICE! You may also loosen the screws 2 and 3, turn the collimator to the required position
by hand and then tighten the screws.
NOTICE! Use the angle scale on the tubehead to determine how much you have rotated the
collimator.
3. After the adjustments, tighten the locking screws and redo the collimator calibration to verify the
adjustment.
4. Install the detached covers.
ORTHOPANTOMOGRAPH OP 3D 237
9 Part replacement instructions
NOTICE! The tubehead rear cover may fall down after the tubehead upper cover is detached.
3. Pull the tubehead front cover locking piece off and detach the tubehead front cover.
238 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
b. Adjust the tilt by turning the tilt adjustment screw. The amount of tilt can be inspected from the scale
at the end of the collimator.
ORTHOPANTOMOGRAPH OP 3D 239
9 Part replacement instructions
b. Rotate the collimator assembly by hand. The amount of rotation can be inspected from the scale on
the tubehead.
4. After the adjustments, tighten the loosened screws and rerun the CEPH Tube Sweep Calibration to
verify the adjustment.
5. Install the detached covers.
240 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
b) Adjust the height of the collimator plate by turning the height adjustment screw.
b) Adjust the tilt of the collimator plate by turning the tilt adjustment screw.
ORTHOPANTOMOGRAPH OP 3D 241
9 Part replacement instructions
3. Tighten the locking screws and redo the collimator calibration to verify the adjustment.
4. Install the detached covers.
NOTICE! When reinstalling the PAN/3D tubehead rear cover you must first remove the
tubehead plate from the PAN/3D tubehead rear cover.
3. Hold the ear rod assembly in place and loosen the ear rod assembly rotation locking screw.
4. Rotate the ear rod assembly by hand according to the instructions given on the GUI during the CEPH
position calibration.
5. Tighten the locking screw.
6. Verify the correct ear rod assembly alignment by performing the CEPH position calibration.
7. Install the detached cover.
242 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
3. Remove the screws (3 pcs) from the upper head support cover.
4. Pull the lower head support cover downwards from the ear rod slide openings to detach it from the
upper cover.
ORTHOPANTOMOGRAPH OP 3D 243
9 Part replacement instructions
6. Pull the left hand side of the upper head support cover over the head support mount.
7. Turn the head support to LAT position and remove the cover.
8. Loosen the ear rod height locking screws (2 pcs) and adjust the eccentric screw to lift/lower the ear
rods as required.
NOTICE! The ear rod assembly has height indicators to be used as guides.
244 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
3. Remove the screws (3 pcs) from the upper head support cover.
4. Pull the lower head support cover downwards from the ear rod slide openings to detach it from the
upper cover.
6. Pull the left hand side of the upper head support cover over the head support mount.
7. Turn the head support to LAT position and remove the cover.
ORTHOPANTOMOGRAPH OP 3D 245
9 Part replacement instructions
8. Loosen the ear rod tilt locking screws (2 pcs) and adjust the eccentric screw to adjust the ear rod tilt as
required.
NOTICE! The ear rod assembly has a tilt amount indicator to be used as a guide.
9. Verify the correct ear rod tilt by performing the CEPH position calibration.
10. Install the detached covers.
NOTICE! Remember to press the lower head support cover back against the upper cover to
reattach it.
246 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
8. All PAN/3D patient positioning lights come on. Check that the Horizontal lights hit the horizontal grooves
and the Midsagittal light hits the vertical groove on the phantom to ensure that the positioning lights are
correctly aligned.
9.11.7.2 Tilt light
1. Detach the Layer laser cover from the lower shelf by detaching a screw on the underside and pulling it
out.
2. If you need to rotate the light, loosen the screw A. If you need to reposition the light assembly, loosen
the screw B and move the assembly sideways by turning the screw C.
ORTHOPANTOMOGRAPH OP 3D 247
9 Part replacement instructions
6. All PAN/3D patient positioning lights come on. Check that the Tilt light hits the vertical groove on the
phantom to ensure that the positioning light is correctly aligned.
9.11.7.3 CEPH FH light
1. The positioning light is held in place in a fixture with one screw. Loosen the screw to release the light.
248 ORTHOPANTOMOGRAPH OP 3D
9 Part replacement instructions
ORTHOPANTOMOGRAPH OP 3D 249
10 Safety tests for maintenance
10.2 Measurement equipment
Use test equipment that is specifically designed to be used with the IEC 62353 standard. For example: Fluke
ESA612 Electrical Safety Analyzer for IEC 62353.
Always follow the instructions supplied with the test equipment that you are using. Also, familiarize yourself
with the relevant electrical standards for this test equipment. The measurement equipment must comply to
the requirements as specified in the IEC 62353 standard.
250 ORTHOPANTOMOGRAPH OP 3D
10 Safety tests for maintenance
ORTHOPANTOMOGRAPH OP 3D 251
10 Safety tests for maintenance
Non-permanent installation:
252 ORTHOPANTOMOGRAPH OP 3D
10 Safety tests for maintenance
1. Measure the protective earth resistance between the applicable measurement points A and B:
Measurement point A - Power supply
NOTICE! With permanent installation, the resistance of protective earth connections in the
supply main may be taken into account.
ORTHOPANTOMOGRAPH OP 3D 253
10 Safety tests for maintenance
CEPH tubehead top cover screw: CEPH head support mount top cover screw
(either):
2. Measure the protective earth resistances with the opposite polarity of the measurement equipment.
3. Check that the protective earth resistance is less than 300 mΩ.
4. Record the highest value.
254 ORTHOPANTOMOGRAPH OP 3D
10 Safety tests for maintenance
ORTHOPANTOMOGRAPH OP 3D 255
PaloDEx Group Oy reserves the right to make changes in specification and features shown herein, or discontinue the product
described at any time without notice or obligation.
Contact your PaloDEx Group Oy representative for the most current information.
Headquarters
PaloDEx Group Oy
Nahkelantie 160 | FI-04300 Tuusula | FINLAND
Tel. +358 10 270 2000
https://dexis.com/contact-us