VERTEX CALIBRATION W.L.L
ees
3 g_ Tel 49144466 1126| Fax 19744017 2179 | Ema: ino@vertexcal com
By 8 (pated wwvertexcalbration.com
CALIBRATION CERTIFICATE
Certificate Number: VCS2023-154
\VC-RF-M-46, REV. NO: 01, REVDATE: MARCH 2022
Customer Name : ADVANCE CONSTRUCTION TECHNOLOGY SERVICES (ACTS)
Address Doha Qatar Delivery Date 9-Jan-2023,
CSR Number ‘cR/30/001 Calibration Date: 8-Jan-2023
Date Received S-Jan-2023, Suggested Due Date : 7-lan-2024
Description Load Cell With Display Indicator Validity 2 Tyear
Brand / Make Interface Manufacturing Received Condition : In Tolerance
Model / Type 121 1BAY-25KN-TB / 9320-1 TEDS Returned Condition: in Tolerance
Serial Number +1 513266A / 17026982 Calibration Status: Pass
Equipment ID Number: N/A Location Vertex Lab
Size / Range 2 25 kN Temperature 2 242°C
Resolution/Least Count: 1 kN. Humidity 51.8 %RH
Calibration Procedure: VC/CM/M/31 Page Number 2 1of2
DETAILS OF MASTER INSTRUMENTS USED FOR CALIBRATION
DESCRIPTION ‘ASSET NUMBER BRAND/MAKE MODEL/TYPE DUEDATE TRACEABILITY NUMBER
Load Cell \VER-CAL-LC-03, Moorehouse Psp i-dun-23 c1o1aaer 121
Barometric Pressure -VER-CAL-BAR-001 tech sp700 2A-Jun-23, 08-20-A.098837
Remarks:
1. Averages of minimum three readings are reported,
2. Suggested due date and test points were given based on customer requirements.
3. These results are obtained at the time of calibration
4. The user should determine the suitability of the instrument for its intended use.
5. Uncertainty of measurement at 95% confidence level by multiplying coverage factor k = 2.
6. Condition of instrument found satisfactory during receipt.
7. Our masters are directly calibrated through ISO/IEC 17025 accredited calibration lab having direct traceability with national / Intemational
standard,
8. Vertex Calibration Laboratory is accredited in accordance with the recognized International Standard as per ISO/IEC 17025:2017.
9. UUC- Unit Under Calibration
10,Pass/ Fail criteria is decided by comparing the error of the unit under calibration and its associated by measurement uncertainty with the
tolerance specified by the manufacturer / customer.
‘11-Accuracy, tolerance and acceptance criteria stated as per customer requierments.
‘12.Any hand written corrections (except @) or photocopies of the report invalidate this certificate.
‘1BResults reported outside the scope af accreditation are mark with asterisk (1
Calibrated By:
— if |veriex,
Rustum Tristan Jokico Ge Pate aS aa Mohideen Math
Calibration Technician ETE Rohe Cater_| (QNQC Engineer
Calration eee are esting due des wer submited/approve by the eutomer Any numberof factor may case an nstument to dt cut of tlerance before the net scheduled of
‘alorationRecabration cys shouldbe based frequency of sed envionment condition an catomers exabshed static accuracy The Ifematon on this apart peti ony to
‘he intranet cei
‘Tic ceiato may not be reproduce ther han nfl except ha pe wit apron fing boron, Thi eport sl not be vl i not companied with Veer Caton fiat.VERTEX CALIBRATION W.L.L
aliens
ee
: eg Tel:#876 46.46 1126| Fax: +974 4017 2179 Emal : info@verexcal com
Bp (yt Soe ‘wwwsertexcalbrationom
CALIBRATION RESULT
Certificate Number: VCS2023-154
Ye-RF-M46,REV.NO:01, REVDATE: MARCH 2022
Description Toad Call With Display Tndieate ‘CSR Number CSRBONOT
Calibration Procedure : VC/CM/M/31 : Calioation Date + Ban-2023,
Serial Number 5132664 / 1702692 Suggested Due Date TJan-2026
Equipment ID Numbe : N/A Model /Type 121 1BAY-25KN-TB / 9320-1 TEDS
Brand / Make Interface Manufacturing Size/Range 2 25KN
Temperature pare Location Vertex Lab
Humidity S1.8%RH Page Number 2of2
Force Accuracy
Nominal Load in kN er Reesing During ie Carationin kN average | Percentage Error Tolerance Estimated Uncertainty #
Runt Run 2 Run3 Reading in kN ba KN kv
° 0.000 0.000 0.000 0.000 - 7 - -
5 4987 4991 4985 4988 261 0063 | PASS 0.0058
10 9979 9986 9977 9980 210 0063 PASS oot19
15 4965 497314878 tage 0.234 0.063 PASS 00216
20 19953 | 1996119965 19.960 0236 0.063 PASS 0.0317
25 24986 2498524981 24984 0.056 0.063 PASS 0.0420
"= End of Calibration =
Where statements of conformity are made inthis report, the following decision rules are applied
PASS - Results + expanded uncertainty within limits / specification
FAIL- Results + expanded uncertainty exceeds limit / specification
INDETERMINATE - Results + expanded uncertainty overlaps limit / specification
Calibrated By: Fi Approved By:
i if vertex, bes,
+ 1S ppd cb ls unin :
Rustum Tristan Jokico L ¥ C.R.No: 125408, Doha « Qatar Mohideen Maajith
Calibration Technician ne QA/QC Engineer
"Thi eae yt be epoca ha in ul pt i pr wit apo of dng botany Taree shl nebo ot cope wth Vrs olin cl