Professional Documents
Culture Documents
New Hydrophobic IOL Materials and Understanding TH
New Hydrophobic IOL Materials and Understanding TH
net/publication/271533199
CITATIONS READS
38 15,544
2 authors, including:
Manfred Tetz
SEE PROFILE
Some of the authors of this publication are also working on these related projects:
All content following this page was uploaded by Manfred Tetz on 11 November 2015.
To cite this article: Manfred Tetz & Matthew R. Jorgensen (2015) New Hydrophobic IOL
Materials and Understanding the Science of Glistenings, Current Eye Research, 40:10, 969-981
REVIEW ARTICLE
1
Berlin Eye Research Institute, Berlin, Germany and 2Institute for Integrative Nanosciences, Dresden, Germany
ABSTRACT
An introduction to the history of intraocular lenses (IOLs) is given, leading up to modern hydrophobic
Downloaded by [183.89.85.100] at 21:24 21 October 2015
examples. The roles of hydrophobicity, hygroscopy, materials chemistry, and edge design are discussed in the
context of IOLs. The four major types of IOL materials are compared in terms of their chemistry and
biocompatibility. An example of a modern ‘‘hydrophobic’’ acrylic polymer with higher water content is
discussed in detail.
Keywords: Glistenings, hydrophilic, hydrophobic, intraocular lens, IOL material
Received 11 February 2014; revised 30 September 2014; accepted 12 October 2014; published online 16 January 2015
Correspondence: Prof. Dr. Manfred Tetz, Berlin Eye Research Institute, Augentagesklinik, Spreebogen, Alt-Moabit 101B, 10559 Berlin,
Germany. Tel: +49 30-39 80 98-0. Fax: +49 30-39 80 98-44. E-mail: info@atk-spreebogen.de
969
970 M. Tetz & M. R. Jorgensen
to have a lingering negative attitude towards hydro- The wide variety of modern IOLs has allowed
philic lenses despite the significant improvement surgeons the luxury of selecting a lens best suited to
in their performance over time. In the early 1990s, a patient’s particular situation. Unfortunately, this
Alcon began to emphasize foldable hydrophobic availability also has led to some confusion regarding
acrylic lenses. A departure from the three main the classification of IOLs and conflicting accounts
classes of IOL materials at the time, the company’s of their performance. Part of this confusion arises
AcrySof IOL performed well and was implanted in naturally from the evolution of IOLs, as new gener-
many eyes. Square edges on the AcrySof resulted in ations of products were distinguished in the market
low posterior capsule opacification (PCO) rates, and using terminology that is not necessarily appropriate
the haptics on the 1-piece design were less likely to get anymore. This article will explain some of the
caught on injection. Additionally, there was not fear of terminology surrounding modern IOLs. The strengths
discoloration as sometimes found in silicone lenses, and weaknesses of the four main types of lenses will
and calcification was generally not a problem.20–31 be summarized. Finally, focusing on the most modern
The rise of foldable lenses, the complications with hydrophobic IOLs, improvements in material chem-
the IOgel lens, and success of early flexible hydro- istry and design will be explained in detail.
phobic acrylics are some of the major factors that
have led to the creation of several versions of flexible
IOLs on the market, making hydrophobic acrylic HYDROPHOBICITY AND
the preferred optic material.32 HYGROSCOPY IN THE CONTEXT OF
Downloaded by [183.89.85.100] at 21:24 21 October 2015
IOLS
FIGURE 2 Schematic diagram of a hydrophobicity measurement. A water droplet is placed on the surface for testing, is carefully
photographed, and the angle between the surface and the droplet is measured. Hydrophilic surfaces have shallow angles (left) while
hydrophobic surfaces have larger angles (right).
degrees for super-hydrophobic surfaces, such as the to be biocompatible, a material must be able to
Lotus leaf.35,37 A material’s chemistry plays a large actually perform the task for which it was designed.
role on its hydrophobicity because the oxygen– A perfectly safe, stable, and square-edged IOL
hydrogen bonds in water are highly polar, meaning is not biocompatible if it cannot perform on target
that there are partial electric charges on the atoms. optically.
These partial charges tend to be attracted to opposite
charges, so water can dissolve salts (which consist of
charged atoms) and is attracted to materials that also PROS AND CONS OF IOL MATERIALS
have partially charged bonds. Polymers, unless spe-
cifically designed otherwise, consist primarily of non- Properties of Polymers: ‘‘All Non-Crystalline
polar carbon–carbon and carbon–hydrogen bonds, Polymers are Fluids’’
which is why they are not generally hydrophilic.
Although not often mentioned in ophthalmology To understand the difference between various acrylic
literature, hygroscopy better explains a material’s polymers used for IOLs, a brief diversion into poly-
tendency to absorb and hold water. A highly hygro- mer chemistry and nomenclature is needed. As acrylic
scopic material draws water into itself usually polymers form, the carbon–carbon double bonds of
accompanied by some physical changes such as the monomers open and join to each other to form the
swelling. Certainly, it is possible for a material to be polymer backbone. Any substituent groups attached
hydrophilic but not hygroscopic. For example, bare directly to the backbone are added as a prefix to the
Downloaded by [183.89.85.100] at 21:24 21 October 2015
silica glass is hydrophilic but because there is no ‘‘acrylate’’ base, while substituents attached further
internal room for the penetration of water, it is away from the backbone to the oxygen are separately
not hygroscopic. In the context of ophthalmology, designated. Note, for example, the differences
hydrophobicity has been used to describe both the between poly(methyl acrylate; PMA), poly(methyl
surface and interior of IOLs. In the following discus- methacrylate), and poly(ethyl methacrylate) are
sion, we differentiate between the interaction of an shown in Figure 3. When more than one type of
IOL’s surface with water (hydrophobicity) and the monomer is polymerized together, the resulting plas-
ability of IOLs to draw water into their interior tic can be called a copolymer.
(hygroscopy). The long chains of polymers and copolymers twist
and fold and link with each other in a messy fashion,
trying to find configurations that minimize the poly-
BIOCOMPATIBILITY mer’s internal energy. It should be noted that this
process continues, slowly, even when a polymer is in
Biocompatibility is a term used in many contexts, its solid-like glassy state. Technically, all non-crystal-
including IOLs. The biological response to a foreign line polymers are fluids. This fluidity changes with
material depends on both the chemistry of the temperature, which is a measure of a material’s
material and the design or morphology. Scales noted internal kinetic energy. At low temperatures, the
in the 1950s that for a material to be biocompatible, polymer chains move slowly, and the bulk material
it should be chemically inert, physically stable, non- flows so gradually that it appears solid and brittle.
carcinogenic, non-allergenic, capable of fabrication in Above a certain temperature, the long chains of glassy
the required form, and have no foreign body reac- polymers have enough energy to overcome local
tion.38 Within the context of ophthalmology, a material attractive and steric forces, and the polymer transi-
should also be optically transparent for long periods tions to a rubber-like or molten state. The temperature
of time, have a high resolving power or refractive at which this change from a solid-like fluid to a
index, and should block ultraviolet rays. Although all rubber-like fluid takes place is called the glass
of Scales’s recommendations remain true, we have transition temperature (Tg).
developed a deeper and more detailed view of what The stiffness of an acrylic polymer, which correlates
makes an IOL biocompatible today. Particular to IOLs, with its Tg, results from a compromise between the
there are certain features that enhance the biocom- bulkiness and flexibility of additions on the polymer
patibility of the lens in the eye. As discussed in detail backbone. Adding bulk increases the stiffness,
below, features of IOL design, such as edge square- while adding flexibility decreases it.39 For example,
ness and surface roughness, can affect PCO and cell PMMA is stiffer than PMA because addition of the
adhesion, which in turn influence biocompatibility. methyl group increases bulk but does nothing for
The material also plays an important role in IOL flexibility. On the other hand, poly(phenylethyl meth-
biocompatibility and overall lens quality, as discussed acrylate) or poly-PEMA (Tg = 26 C) is much softer
below in a comparison of the four major classes of IOL than PMMA (Tg = 105 C), even though it is bulky
materials. because a great amount of flexibility is given by the
In addition to the issues of biocompatibility-related rotating bonds in the long chain of atoms in the
stability and safety, it must be remembered that phenylethyl group.
Copyright ! 2015 Taylor & Francis Group, LLC
972 M. Tetz & M. R. Jorgensen
FIGURE 3 Examples of acrylic polymers. PMA, poly(methacrylate); PMMA, poly(methyl methacrylate); PEMA, poly(ethyl
methacrylate); poly-HEMA, poly(2-hydroxyethyl methacrylate). The hydroxyl group in poly-HEMA significantly increases this
polymer’s hygroscopy.
Downloaded by [183.89.85.100] at 21:24 21 October 2015
FIGURE 4 Model of two cross-linked chains of PHS copolymer. Due to the bulky side chains and uneven folding, the polymer
contains regions which are slightly more packed than others.
Polymer Hygroscopy and the Etiology of perspective of chemical potential equilibrium using
Glistenings concepts familiar to physical chemistry.42 Due to the
dynamics of the process of polymerization, it is clear
Although most polymer materials are not engineered that an entire piece of bulk material will not be
to contain water, all have a certain propensity to completely connected, with polymer chains perfectly
absorb it, or to be hygroscopic. In the case of the folded. Therefore, it is natural that there should be
AcrySof copolymer, discussed below, an equilibrium regions with the polymer more densely folded and
water content of about 0.1–0.5% has been observed, others that are loose. Figure 4 is a schematic of just
depending on the temperature.40 The equilibrium two folded polymer chains cross-linked to each other.
water content of various copolymers can also vary It can be readily seen that there arise areas that are
strongly depending on the concentration of salts (ionic much denser than others. Of course, polymers have
strength) in the surrounding solution, for example, the many more cross-linked chains than what is shown in
aqueous humor or storage media.41 The hygroscopic Figure 4, but the principle remains the same. Water
nature of polymers, and how this varies depending on diffuses into the polymer, due to normal equilibrium
temperature and ionic strength of the surrounding driving mechanisms, and then can collect in lower
solution, is well explained and understood from the density ‘‘pockets’’, especially when the equilibrium
Current Eye Research
New Hydrophobic IOL Materials 973
Downloaded by [183.89.85.100] at 21:24 21 October 2015
FIGURE 5 Examples of different grades of glistenings. (a) Relatively minor glistenings observed in an Alcon lens, unknown model,
9 years after implantation. (b) Moderate glistenings observed in an AcrySof lens 4 years after implantation. Note the visible surface
scratches. (c) Severe glistenings in an Acrysof lens 2 years after implantation.
hygroscopy of the polymer is quickly changed due to undesired defect nevertheless.48 Because glistenings
changes in temperature.42–44 These pockets can develop over time, they indicate a dynamic process
enlarge over time until they become discreet vacuoles within the lens/eye system with causes and long-term
of water visible as glistenings,45,46 or subsurface outcomes that are not entirely clear. There are some
nanoglistenings (SSNG), examples of which are cases where lenses with glistenings have been
shown in Figure 5. explanted,55 though this is rare.
Much has been said on the subject of glistenings, Table 1 lists several important physical properties
especially with respect to the popular AcrySof of common IOL materials for comparison, where Eq.
lens.14,44,46–51 Two studies, each including 115 eyes, Hygroscopy is the mass percent of water within the
showed the incidence of glistenings in the AcrySof lens at equilibrium, Cont. Angle is the angle between
MA60BM increasing over time since implantation. In the leading edge of a water droplet and the IOL
one, the incidence increased from 20% at 3 months to surface, Tensile Str. is the greatest amount of stress the
55% at 12 months.52 In the other 56% of IOLs showed material class can withstand while being pulled
glistenings at 14 months which increased to 87% at 28 without breaking, n is the refractive index, and Tg is
months.53 A third study including 42 eyes implanted the glass transition temperature.
with the MA60BM or MA30BA found 100% occur-
rence of glistenings after 4–46 months, postopera-
tively.54 While glistening studies have focused on the PMMA
AcrySof lenses, they have been observed in all the
early lens materials and were reported in PMMA at The first IOL, implanted in 1949, was made of PMMA.
least as early as 1984.50 Glistenings have been Thanks to its excellent tissue tolerance, its relatively
associated with less severe visual effects such as high refractive index (1.49), and optical clarity better
reduced contrast sensitivity but represent an than glass, PMMA has been used in a wide variety of
Copyright ! 2015 Taylor & Francis Group, LLC
974 M. Tetz & M. R. Jorgensen
implants for more than 70 years.56 Though PMMA has between silicone IOLs and discoloration.65
been almost completely replaced by modern foldable Unfortunately, there also were problems with surface
lenses, they still are available from a variety of calcification associated with certain eye conditions
companies. Primitive by today’s standards, the ori- that led to explantation.66 More modern silicone
ginal PMMA lenses still performed remarkably well. IOLs appear to have largely addressed these issues,
There have been reports of original lenses implanted some of which arose due to complications with
by Ridley remaining perfectly clear and centered for the way silicone IOLs were sterilized in permeable
more than 28 years.4 On the other hand, it also was packages.51,63
reported that some spontaneously dislocated into the
vitreous.11 Because PMMA is a rigid material and
original manufacturing was somewhat crude com- Hydrophilic (Hygroscopic) Acrylic
pared to today’s techniques, it has been shown to
irritate the sensitive uveal tissues, resulting in chronic Hydrophilic acrylic lenses, or, more accurately, hydro-
inflammation.10 However, inflammation is not the gel lenses, are highly hygroscopic, with an equilib-
most serious drawback of PMMA’s rigidity. Because rium water content ranging from 18 to 34% (or in
it is too rigid to fold, the lens cannot pass through certain cases even higher).67,68 Their contact angle
the small incisions allowed by phacoemulsification. measurements do not necessarily correspond to their
Widening of the incision is frustrating after removal hygroscopy, as the surface of the lenses could be
of the natural lens through a 2-mm opening and modified to alter their hydrophobicity. Hygroscopic
Downloaded by [183.89.85.100] at 21:24 21 October 2015
increases patient recovery time as well as surgically- acrylic polymers are made by modifying the side-
induced astigmatism. chains connected to the methacrylate backbone of
regular PMMA to contain hydroxyl components that
attract water, most commonly poly(2-hydroxyethyl
Silicone methacrylate), poly-HEMA, or just HEMA as shown
in Figure 3. The folding of poly-HEMA chains
Silicone also has a long history in medicine and depends significantly on the level of hydration, and
in cataract surgery. It was used even before folding so the physical and optical properties of the polymer
was common practice and was the first lens to be change as a function of water content. In the dry state,
widely used in procedures requiring folding, which hygroscopic lenses are hard and opaque.67 It is in this
started to become popular in 1984.13 In addition to the state that they are manufactured. As the lenses
immediate benefit of flexibility, properly formulated hydrate, they absorb water like a sponge, becoming
silicone can be manufactured by molding, is auto- soft and transparent.
clavable, and has superior biocompatibility.57 Due to their high water content, they have a lower
Polydimethoxysiloxane, a typical form of silicone, refractive index that increases the IOL thickness but
has a water content of less than 0.4%, though this also can reduce glare and other aberrations.
can be significantly higher for silicone hydrogels.58 Hydrogel lenses are, in a sense, more like biological
In studies comparing IOL edges, silicone had the tissue because they have such high water content.
squarest edges, which is a factor in their low PCO Cell adhesion tests have shown that, like with
rate.59–61 The chemistry of silicone’s surface also is a silicone, LECs are half as likely to adhere to the
factor, as lens epithelial cells (LECs) are half as likely polymer surface in vitro compared to PMMA.62 In a
to adhere to the surface than PMMA.62 Silicone also prospective study comparing 11 IOL lens surfaces in
has the most demanding fabrication, contributing to 653 eyes, hydrogel lenses had fewer foreign body
production costs of these IOLs. Perhaps, the most giant cells than both plain PMMA and silicone
significant problem with early silicone IOLs was their lenses.69 Unfortunately, the early experiences with
potential to develop a brown discoloration, first hydrophilic lenses were mostly negative. Early
reported by Milauskas in 1991.63 In response to the examples were too soft, potentially resulting in
initial report, the vice president of quality assurance secondary cataracts and other complications as
of one of the companies involved assured doctors that described previously.15–17 Probably, the most signifi-
the company had not found the discoloration in vitro, cant force limiting wider spread use of hydrophilic
and that it had examined one of the patients in IOLs is the risk of surface and internal calcification,
question only to find ‘‘an almost imperceptible, which has been reported and discussed exten-
unidentifiable imperfection on the periphery of the sively.18–20,22,66 Aversion to hygroscopic materials
lens’’.64 Milauskas was quick to reply that the lens in persists in spite of more recent reports that calcifi-
question had been examined by no less than five cation was more of a manufacturing and handling
people who all agreed that it was indeed brown.63 problem rather than a material problem that may
Ultimately, the exchanges between Milauskas, silicone have been addressed in second-generation hygro-
lens manufacturers, and other doctors recorded in the scopic IOLs.20 Recent studies surveying the edge
literature were enough to create a lasting association designs of IOLs across platforms have shown that
Current Eye Research
New Hydrophobic IOL Materials 975
EVOLUTION OF UNDERSTANDING IN
EDGE DESIGN
hygroscopic lenses tend to be the least reliable for the cause. At first, research into the relationship
having a square edge, resulting in higher rates of between IOL design and PCO included different lens
PCO.70,71 This difficulty in obtaining a square edge types and materials, without clear attention to the
likely results from the swelling that takes place in the effect of the optic edges. Early in vitro experiments
lens after it is lathe cut in the dry state. showed that plano-convex lenses were less likely to
develop PCO than optics with ridges, but in these
studies, lenses with both different materials and
Hydrophobic Acrylic different designs were used, making it difficult to
determine the cause of the effects.74 Several similar
Hydrophobic acrylics are modern foldable IOLs, the studies surfaced that focused on IOL design, and it
most widely used of which is the AcrySofÕ lens. became clear that the optic edge specifically played a
Foldable acrylic IOLs achieve polymer flexibility in a role in inhibiting PCO.75–77 In 1999, Nishi compared a
manner different than the hydrogels. In the case of square-edged AcrySof lens with a normal rounded
AcrySof, a blend of the monomers phenylethyl acryl- edge-PMMA lens and a PMMA lens that had been
ate (PEA) and PEMA are polymerized together (PEA/ specially made to have the same square edge as the
PEMA copolymer), along with a cross-linker butane- AcrySof. They found that the rounded-edge PMMA
diol diacrylate (BDDA), to create the desired stiffness. failed to inhibit migrating LECs and the AcrySof did,
Poly-PEA has a Tg of 3 C, poly-PEMA has a Tg of while the otherwise identical PMMA lens with the
26 C, while the final PEA/PEMA copolymer (which square edges inhibited LEC migration to the same
includes some BDDA cross-linking) is a compromise degree as the AcrySof.78 These results clearly showed
with a Tg of 14.0–15.5 C.72 The chemical structures of that the primary player in the fight against PCO in
the monomers involved in the AcrySof lens are shown these lenses was the optic edge over the material.
in Figure 6. By varying the flexibility of side chain More recently, there has been a series of papers
substituents, the relative proportion of different published, several with the author’s involvement, that
monomers, and controlling the amount of cross- define and evaluate the squareness or sharpness of
linking between polymer chains, hydrophobic acrylic IOL edges across the spectrum.61,71,75,79 If an image of
polymers with a wide range of physical properties are an IOL edge, such as the one shown in Figure 8(A), is
possible. overlapped with a section of a true square corner, the
Hydrophobic lenses have a remarkably good track difference in their areas is a measure of squareness, as
record. These are the most widely used IOLs and shown in Figure 8(B). Alternatively, the sharpness can
stand as tribute to the benefit of innovation in the field be measured from the radius of the corner (Figure 8C).
of cataract surgery and materials chemistry. Single- Of course, all macroscopic edges are round if one
and multi-piece hydrophobic IOLs can be folded and looks at high-enough magnification, so it is critical to
inserted through very small incisions while resisting choose magnifications that can distinguish between
damage. PCO is significantly lower than PMMA but different degrees of sharpness and which are of
generally a bit higher for hydrophobic acrylic lenses physiological importance. In the first published
compared with silicone, though this is usually clinic- work in a series of papers that define and evaluate
ally insignificant and may depend more on edge IOL edges, a set of otherwise equivalent PMMA
design than anything else.26,73 lenses were made with a varying amount of
Copyright ! 2015 Taylor & Francis Group, LLC
976 M. Tetz & M. R. Jorgensen
Downloaded by [183.89.85.100] at 21:24 21 October 2015
FIGURE 7 A comparison between the previous ‘‘gold standard’’ rounded edges of tumble polished PMMA lenses (a), with the square
edge of the Acrysof lens that resulted in lower PCO and a shift towards square-edged lenses (b).
FIGURE 8 Edge profile of a hydrophilic (hygroscopic) acrylic lens, and the method for measuring squareness and sharpness. (a) Raw
environmental scanning electronic microscopy image of the IOL edge. (b) The squareness of an edge can be measured by the area
enclosed between the IOL edge and a true square. (c) The edge sharpness can be measured from the radius of curvature on the corner
itself.
squareness to their edge by tumble polishing them for acrylic.61,71 Another group reported the sharpness of
different amounts of time.76 The squareness was 17 IOLs.75 Interestingly, sharpness and squareness are
evaluated from images at magnifications of 500 not necessarily the same thing as some lenses have
and 2000 taken using scanning electron microscopy very sharp but obtuse corners. In this case, a lens may
(SEM) by measuring the area contained between the look poor from the perspective of squareness but
edges of a true square angle and the sloping edges of favorable in terms of sharpness, so care must be used
the IOL. It was found that, as expected, the roundness in comparing the relative sharpness (or squareness)
of the edge increased as the amount of time in the across studies. Nevertheless, the major conclusions of
polishing machine increased. The lenses were tested studies using both methods remain the same: silicone
for LEC migration in vitro, and it was shown that has uniformly the squarest and sharpest corners,
lenses with a squareness above a certain threshold hydrophobic lenses are slightly worse (with a bit more
were effective in blocking cells from crossing the variation), and hydrophilic lenses have the greatest
boundary of the optic.79 Following this report, two deviation from the ideal with a wide range of results.
subsequent studies tested the edges of 40 lenses for Although some questions remain about the nature of
squareness, including those made from PMMA, sili- PCO prevention and IOL edges, a few things are clear:
cone, hydrophilic acrylic, and hydrophobic PCO correlates with optic edge design, square and
Current Eye Research
New Hydrophobic IOL Materials 977
sharp is better than rounded, and the degree of to be glistening-free when packaged in 0.9% saline
sharpness varies significantly among IOLs, making solution,80,81 is currently the only hydrophobic IOL
the issue of optic edge squareness an important factor material with a published composition that contains a
in overall IOL quality on a case-by-case basis. substantial hydrophilic component. Figure 9 shows
the components of this hydrophobic/hydrophilic
hybrid material, which is actually a copolymer of
PROPERTIES OF NEW HYDROPHOBIC three different monomers PEG-PEA/HEMA/styrene
IOLS (collectively called the PHS copolymer) and a cross-
linker. The material is flexible primarily due to the
As discussed above, IOL materials, chemistry, and long flexible chains of poly(ethylene glycol),
design have evolved with each generation of lens. which connect the phenyl ring to the polymer back-
Flexible hydrophobic acrylic IOLs dominate the bone. The styrene adds bulk and increases refractive
market because they perform very well, are cost- index, HEMA adds the ability to disperse water
effective, and do not have a negative history. The through the material, and EG-DMA increases struc-
more modern hydrophilic and silicone lenses gener- tural integrity.
ally also have great performance. However, there still
is room for improvement as the gold-standard lens
should have, as it ought to, a material and design that Equilibrium Water Content and Glistenings
minimizes potential complications in the patient
Downloaded by [183.89.85.100] at 21:24 21 October 2015
during and after surgery. The lens also should be One of the primary advantages of having a hydro-
defect-free; a defect might be any undesirable irregu- phobic polymer with a careful balance of HEMA is
larity that is big enough to be visible at the slitlamp. that glistenings can be eliminated. As discussed
The most modern hydrophobic IOLs approach this previously, water will enter all polymers to one
gold standard starting with the intelligent design of degree or another due to the principles of equilibrium.
their materials. An advance in IOL materials was Yet what happens to the water once inside? Is it more
made in going from silicone to hydrophilic, because a favorable for the water to collect in vacancies, or for it
cost-effective way to make a flexible biomimetic to be dispersed throughout the material? It is known
acrylic through the addition of water was used. An that the equilibrium water content of IOLs depends
advance was made moving to flexible hydrophobic on temperature and the ionic strength of the environ-
acrylic, because lenses could be flexible while avoid- ment.42,43 In the PEA/PEMA copolymer, when the
ing some of the issues specific to high water content equilibrium point for water is quickly reduced, the
lenses. Still, a problem has remained that all polymers ‘‘excess’’ water must go somewhere, and it might not
are hygroscopic to one degree or another, and the have enough time to diffuse out of the lens. On the
water which diffuses into the previous generation of other hand, the PHS copolymer contains a built-in
hydrophobic acrylic IOLs is not managed. Here, we mechanism to keep water dispersed and locked
will focus on one modern IOL polymer that is made of evenly throughout the material, so that when this
materials combining the best aspects of earlier hydro- equilibrium point fluctuates, ‘‘excess’’ water would
phobic and hydrophilic polymers, used in the Bausch rather connect to a water-friendly hydroxyl group
& Lomb enVista (previously the AVS EXACT) and the than collect in a dislocation surrounded by purely
Eternity IOL in Japan. This particular material, found hydrophobic polymer.
FIGURE 9 The components of a glistening-free hydrophobic acrylic. PEG-PEA: poly(ethylene glycol) phenyl ether acrylate; HEMA:
2-hydroxyethyl methacrylate; EG-DMA: ethylene glycol dimethacrylate. In the PEG-PEA component, the ethylene glycol subunit
(shown in red in online color version) is a repeating unit of variable length, making the link between the polymer backbone and the
phenyl group very long.
lenses in detail because information is limited, an copolymer materials.84 In this study, a very sharp tip
important factor may be that all of them are lathe cut. was brought onto the surface of the IOL, and the
It has been shown previously that lathe-cut lenses are displacement of the tip into the surface was recorded
less likely to develop glistenings, presumably because as a function of force applied.81 It was found that
a more homogeneous polymerization is possible the AcrySof, Hoya, and Sensar materials indented
when done under bulk conditions.42,44,55 by 800–1000 nm with an applied force of 60 mN, while
the PHS copolymer only indented by 150 nm. From
these measurements, and the dimensions of the tip,
Physical Properties of PHS Copolymer IOLs the surface hardness can be calculated. It was found
that the PHS copolymer has a hardness of 1.8 MPa,
The nature of the IOL surface, including its hydro- while the others had a hardness of 0.1–0.4 MPa.84
phobicity, resistance to damage, and smoothness are Although the study cited is limited by the number of
important lens parameters because it is the surface samples examined, it does highlight that the surface
that is the first point of interaction with the eye. hardness of IOLs and that this capability to resist
Hydrophobicity is a factor influencing cell adhesion to weakness, stress, and scratches can vary widely
the surface. One recent study tracked how the contact between materials.
angle of the PHS copolymer material changed over
time as it was hydrated in balanced salt solution at
37 C. It was found that the contact angle, which was Early Experiences with PHS Copolymer
initially 75 , stayed within a narrow window of 69–79 Lenses
over a period of 4 weeks, suggesting that the hydro-
phobicity of the lens surface remains stable over time The patent application for the PHS copolymer,
as the lens is hydrated. A similar stability was found invented by Advanced Vision Sciences, was published
for the Hoya AF-1 and Sensar AR40 materials, in in 2007.85 The material has been marketed and used
contrast to the PEA/PEMA copolymer material, in Japan since 2008 as the Eternity IOL. The rights to
which had a contact angle that dropped from about develop and market the material were acquired by
75 to 50 on hydration.82 Bausch & Lomb in 2009, after which a patent for an
Although it is not often discussed in the literature, updated design of the lens was filed in 2010.86 Finally,
scratching or tearing of IOLs can occur during the U.S. Food and Drug Administration approved the
insertion (as shown dramatically in an explanted material and new design marketed as enVista in 2012.
IOL in Figure 10 and more subtly in Figure 5b), only Although this material is still relatively new, there
to be discovered after the lens unfolds in the eye. have been a few clinical studies involving the lens.
Unless the crack or tear is very minor, so that it is The enVista directions for use include the results of
unlikely to result in negative visual outcomes,83 two clinical studies, one for the 3-piece design of the
the IOL must be explanted, often requiring an lens packaged in 10% saline and the other for the 1-
enlargement of the incision.51 Nanoindentation meas- piece design packaged in 0.9% saline. In both cases,
urements have been used to test the surface hardness there were no adverse events attributable to the IOL.87
of the AcrySof, Hoya AF-1, Sensar AR40, and PHS Interestingly, glistenings were observed in the IOLs
Current Eye Research
New Hydrophobic IOL Materials 979
stored in 10% saline solution. In a separate study towards a standard of quality that maximizes safety,
conducted in Japan, glistenings were observed in 40 minimizes PCO and calcification, and is defect-free.
eyes of 40 patients.44 Following the change to 0.9% Flexible hydrophobic acrylic lenses offer controlled
saline solution, no glistenings in this material have injection and unfolding, their manufacture allows the
been reported.72,81,88 It is hypothesized that in the case creation of a square optic edge, and they are not
of the lenses stored in 10% saline, the higher salt associated with calcification. The most modern flex-
content within the lens caused an osmotic pressure ible IOL materials contain a minimal amount of
difference after implantation, which leads to an influx equilibrium water content by design, a feature that
of water and the formation of glistenings. Naturally, appears to have the benefit of preventing glistenings.
the capability of any material formulation to disperse Modern IOLs have carefully fabricated edges to
water has limits. minimize PCO. All of these factors make modern
With regards to PCO, one study including 221 eyes acrylic lenses a welcome addition to the surgical
showed the YAG capsulotomy rate was 3.6% after armamentarium.
21 months.89 Another study reported an average
EPCO score of 0.032 ± 0.101 for 122 eyes 4 to
6 months after implantation with the enVista lens.88
Our own reported experience includes over 400 eyes,
DECLARATION OF INTEREST
with no significant IOL-related complications or
The authors report no conflicts of interest. The authors
glistenings.83 The biocompatibility of the enVista
alone are responsible for the content and writing of
Downloaded by [183.89.85.100] at 21:24 21 October 2015
15. Levy JH, Pisacano AM, Anello RD. Displacement of bag- 31. Brar GS, Grewal DS, Ram J, Singla M, Grewal SPS. Square-
placed hydrogel lenses into the vitreous following edge polymethylmethacrylate intraocular lens design for
neodymium:YAG laser capsulotomy. J Cataract Refract reducing posterior capsule opacification following paedi-
Surg 1990;16:563–566. atric cataract surgery: initial experience. Clin Exp
16. Werner L, Apple DJ, Escobar-Gomez M, Öhrström A, Ophthalmol 2008;36:625–630.
Crayford BB, Bianchi R, et al. Postoperative deposition of 32. Leaming DV. Practice styles and preferences of ASCRS
calcium on the surfaces of a hydrogel intraocular lens. members—2003 survey. J Cataract Refract Surg 2004;30:
Ophthalmology 2000;107:2179–2185. 892–900.
17. Werner L, Apple DJ, Kaskaloglu M, Pandey SK. Dense 33. Scatena LF, Brown MG, Richmond GL. Water at hydro-
opacification of the optical component of a hydrophilic phobic surfaces: weak hydrogen bonding and strong
acrylic intraocular lens: a clinicopathological analysis of 9 orientation effects. Science 2001;292:908–912.
explanted lenses. J Cataract Refract Surg 2001;27: 34. Feng L, Li S, Li Y, Li H, Zhang L, Zhai J, et al. Super-
1485–1492. hydrophobic surfaces: from natural to artificial. Adv Mater
18. Werner L, Hunter B, Stevens S, Chew JJL, Mamalis N. Role 2002;14:1857–1860.
of silicon contamination on calcification of hydrophilic 35. Ma M, Hill RM. Superhydrophobic surfaces. Curr Opin
acrylic intraocular lenses. Am J Ophthalmol 2006;141: Colloid Interface Sci 2006;11:193–202.
35–43. 36. Cao L, Hu H-H, Gao D. Design and fabrication of micro-
19. Izak AM, Werner L, Pandey SK, Apple DJ. Calcification textures for inducing a superhydrophobic behavior on
of modern foldable hydrogel intraocular lens designs. hydrophilic materials. Langmuir 2007;23:4310–4314.
Eye 2003;17:393–406. 37. Sugimura H, Ushiyama K, Hozumi A, Takai O.
20. Linnola RJ. Sandwich theory: bioactivity-based explan- Micropatterning of alkyl- and fluoroalkylsilane self-
ation for posterior capsule opacification. J Cataract Refract assembled monolayers using vacuum ultraviolet light.
Surg 1997;23:1539–1542. Langmuir 2000;16:885–888.
Downloaded by [183.89.85.100] at 21:24 21 October 2015
21. Hollick EJ, Spalton DJ, Ursell PG, Pande MV. Lens 38. Scales J. Discussion on metals and synthetic materials in
epithelial cell regression on the posterior capsule with relation to tissues; tissue reactions to synthetic materials.
different intraocular lens materials. Br J Ophthalmol 1998; Proc R Soc Med 1953;46:647–652.
82:1182–1188. 39. Krause S, Gormley JJ, Roman N, Shetter JA, Watanabe WH.
22. Hollick EJ, Spalton DJ, Ursell PG, Pande MV. Glass temperatures of some acrylic polymers. J Polym
Biocompatibility of poly(methyl methacrylate), silicone, Sci A 1965;3:3573–3586.
and AcrySof intraocular lenses: randomized comparison of 40. Miyata A, Yaguchi S. Equilibrium water content and
the cellular reaction on the anterior lens surface. J Cataract glistenings in acrylic intraocular lenses. J Cataract Refract
Refract Surg 1998;24:361–366. Surg 2004;30:1768–1772.
23. Linnola RJ, Werner L, Pandey SK, Escobar-Gomez M, 41. Stevenson WTK, Sefton MV. The equilibrium water content
Znoiko SL, Apple DJ. Adhesion of fibronectin, vitronectin, of some thermoplastic hydroxyalkyl methacrylate poly-
laminin, and collagen type IV to intraocular lens materials mers. J Appl Polym Sci 1988;36:1541–1553.
in pseudophakic human autopsy eyes: part 2: explanted 42. McQuarrie DA, Simon JD. Physical chemistry: a molecular
intraocular lenses. J Cataract Refract Surg 2000;26: approach. Sausalito: University Science Books; 1997.
1807–1818. 43. Omar O, Pirayesh A, Mamalis N, Olson RJ. In vitro
24. Nishi O, Nishi K, Wickström K. Preventing lens analysis of AcrySof intraocular lens glistenings in AcryPak
epithelial cell migration using intraocular lenses with and Wagon Wheel packaging. J Cataract Refract Surg 1998;
sharp rectangular edges. J Cataract Refract Surg 2000;26: 24:107–113.
1543–1549. 44. Werner L. Glistenings and surface light scattering in
25. Sundelin K, Friberg-Riad Y, Östberg A, Sjöstrand J. intraocular lenses. J Cataract Refract Surg 2010;36:
Posterior capsule opacification with AcrySof and poly(- 1398–1420.
methyl methacrylate) intraocular lenses: comparative 45. Ong MD, Callaghan TA, Pei R, Karakelle M. Etiology of
study with a 3-year follow-up. J Cataract Refract Surg surface light scattering on hydrophobic acrylic intraocular
2001;27:1586–1590. lenses. J Cataract Refract Surg 2012;38:1833–1844.
26. Abela-Formanek C, Amon M, Schauersberger J, Kruger A, 46. Tetz M, Werner L, Schwahn-Bendig S, Batlle JF. A
Nepp J, Schild G. Results of hydrophilic acrylic, hydro- prospective clinical study to quantify glistenings in a
phobic acrylic, and silicone intraocular lenses in uveitic new hydrophobic acrylic lOL; Poster presented at:
eyes with cataract: comparison to a control group. American Society of Cataract and Refractive Surgery
J Cataract Refract Surg 2002;28:1141–1152. Symposium on Cataract, IOL, and Refractive Surgery;
27. Kohnen T, Allen D, Boureau C, Dublineau P, Hartmann C, April 3–8, 2009; San Francisco, CA, USA.
Mehdorn E, et al. European Multicenter Study of the 47. Ballin N. Glistenings in injection-molded lens. J Am
AcrySof ReSTOR Apodized Diffractive Intraocular Lens. Intraocul Implant Soc 1984;10:473.
Ophthalmology 2006;113:578–584. 48. Dhaliwal DK, Mamalis N, Olson RJ, Crandall AS,
28. Nixon DR, Apple DJ. Evaluation of lens epithelial cell Zimmerman P, Alldredge OC, et al. Visual significance of
migration in vivo at the haptic-optic junction of a one-piece glistenings seen in the AcrySof intraocular lens. J Cataract
hydrophobic acrylic intraocular lens. Am J Ophthalmol Refract Surg 1996;22:452–457.
2006;142:557–562. 49. Chehade M, Elder MJ. Intraocular lens materials and
29. Toto L, Falconio G, Vecchiarino L, Scorcia V, Di Nicola M, styles: a review. Aust NZ J Ophthalmol 1997;25:255–263.
Ballone E, et al. Visual performance and biocompatibility 50. Nishihara H, Kageyama T, Ohnishi T, Koike M, Imai M,
of 2 multifocal diffractive IOLs: six-month comparative Shibuya A, et al. Glistenings in lathe-cut acrylic intraocular
study. J Cataract Refract Surg 2007;33:1419–1425. lens. Ganka Shujutsu 2000;13:227–230.
30. Zemaitiene R, Jasinskas V, Auffarth GU. Influence of three- 51. Davison JA. Clinical performance of alcon SA30AL and
piece and single-piece designs of two sharp-edge optic SA60AT single-piece acrylic intraocular lenses. J Cataract
hydrophobic acrylic intraocular lenses on the prevention of Refract Surg 2002;28:1112–1123.
posterior capsule opacification: a prospective, randomised, 52. Miyata A, Suzuki K, Boku C, Kinohira Y, Aramaki T,
long-term clinical trial. Br J Ophthalmol 2007;91:644–648. Ando M, et al. Glistening particles on the implanted
acrylic intraocular lens [Japanese]. Rinsho Ganka 1997;51: 72. McIntyre S, Werner L, Mamalis N. Hydrophobic Acrylic
729–732. IOLs: a primer. Cataract Refract Surg Today Europe 2011.
53. Peetermans E, Hennekes R. Long-term results of wagon 73. Abela-Formanek C, Amon M, Schild G, Schauersberger J,
wheel packed acrylic intra-ocular lenses (AcrySof). Bull Heinze G, Kruger A. Uveal and capsular biocompatibility
Soc Belge Ophtalmol 1999;271:45–48. of hydrophilic acrylic, hydrophobic acrylic, and silicone
54. Christiansen G, Durcan FJ, Olson RJ, Christiansen K. intraocular lenses. J Cataract Refract Surg 2002;28:50–61.
Glistenings in the AcrySof intraocular lens: pilot study. 74. Santos BA, Pastora R, DelMonte MA, O’Donnell Jr FE.
J Cataract Refract Surg 2001;27:728–733. Comparative study of the effects of optic design on lens
55. Werner L, Storsberg J, Mauger O, Brasse K, Gerl R, epithelium in vitro. J Cataract Refract Surg 1987;13:
Müller M, et al. Unusual pattern of glistening formation 127–130.
on a 3-piece hydrophobic acrylic intraocular lens. J 75. Nagata T, Watanabe I. Optic sharp edge or convexity:
Cataract Refract Surg 2008;34:1604–1609. comparison of effects on posterior capsular opacification.
56. Langer R, Tirrell DA. Designing materials for biology and Jpn J Ophthalmol 1996;40:397–403.
medicine. Nature [10.1038/nature02388] 2004;428:487–492. 76. Yamada K, Nagamoto T, Yozawa H, Kato K, Kurosaka D,
57. Cook CS, Peiffer Jr RL, Mazzocco TR. Clinical and Bissen Miyajima H, et al. Effect of intraocular lens design
pathologic evaluation of a flexible silicone posterior on posterior capsule opacification after continuous curvi-
chamber lens design in a rabbit model. J Cataract Refract linear capsulorhexis. J Cataract Refract Surg 1995;21:
Surg 1986;12:130–134. 697–700.
58. Blume I, Schwering PJF, Mulder MHV, Smolders CA. 77. Nagamoto T, Eguchi G. Effect of intraocular lens design on
Vapour sorption and permeation properties of poly migration of lens epithelial cells onto the posterior capsule.
(dimethylsiloxane) films. J Memb Sci 1991;61:85–97. J Cataract Refract Surg 1997;23:866–872.
59. Werner L, Müller M, Tetz M. Evaluating and defining the 78. Nishi O, Nishi K. Preventing posterior capsule opacifica-
sharpness of intraocular lenses: microedge structure of tion by creating a discontinuous sharp bend in the capsule.
Downloaded by [183.89.85.100] at 21:24 21 October 2015