Professional Documents
Culture Documents
Chapter 3: De-Oiling Produced Water
Chapter 3: De-Oiling Produced Water
3.2 Theory
Water is almost always produced along with the crude oil. Usually when a new field is
discovered, the production is almost “dry”. The oil contains only a small percentage of water.
This soon changes. Each year the production contains more and more water, especially where the
water is pumped back into the formation. Eventually far more water is produced than oil.
Emulsions are formed, as the oil and water are co-produced. Even though the liquids are
immiscible, they become thoroughly mixed when produced to the surface and flow through the
production equipment. Additional agitation results as the liquids experience turbulence through
all the various pipes and fittings. All of this energy breaks both the oil and water into many tiny
droplets. Solids and naturally occurring surfactants often stabilize these droplets. The stabilized
droplets of water result in wet oil or regular emulsions. The stabilized droplets of oil create dirty
water or reverse emulsions.
Regular Emulsion — Wet oil, or water-in-oil mixture
Reverse Emulsion — Dirty water, or oil-in-water mixture
Insoluble oil, which is the oil that can be removed with conventional water clarifiers.
Soluble oil, which is not a true hydrocarbon but does show up in the IR and Gravimetric
readings as true oil.
Products containing phosphoric or phosphorus acid are regularly used in total fluid to remove
soluble oil. Caution needs to be exercised when using the phosphoric acid products. Phosphoric
acid will form calcium phosphate scale when used in produced waters that have significant
amounts of calcium present. Phosphorus acid will not react with calcium, and that is why it is
mostly widely used. Phosphorus acid is extremely corrosive and incompatible with 316 SS.
Hastalloy metals are recommended to be used in the injection system for any acid products.
Caution must be taken when using these products. The phosphorus acid can eat quickly through
metal if leaks are not attended to promptly. An injection quill must be used so that the acids can
be delivered directly into the produced fluids. This will keep the piping internals from corrosion
attack.
To determine if soluble oil is present in the produced water, a silica gel extraction test must be
run. This test is run with the regular Freon extraction test. After the Freon extraction, the sample
is run on the IR and is recorded as total ppm. The Freon extract is dispensed into a beaker, and
silica gel is added to the Freon extract. The sample is stirred for approximately one minute. The
sample is then allowed to sit. The silica gel will fall to the bottom of the beaker and separates
from the Freon. The Freon is filtered again and collected in the IR cuvette and placed in the IR
machine again. This reading will be insoluble oil ppm only. The insoluble oil ppm is then
subtracted from the total IR ppm, and the result is considered the soluble oil portion.
Insoluble materials will separate with time. Density differences will drive the contaminants out
of the water.
3.2.2.2 Distillation
Water can be boiled away from the contaminants. This very expensive method is only used in
special circumstances.
Oil Field Chemicals Training Manual
3-4
Property of Nalco Energy Services
Confidential & Proprietary – DO NOT DUPLICATE
Chapter 3: De-oiling Produced Water
3.2.2.3 Centrifuge
Spinning motion speeds up the process of natural separation by increasing the effective gravity.
3.2.2.4 Filtration
Solids and liquids can be physically trapped and removed from the water.
3.2.2.5 Coalescing
Coalescing is the action of combining droplets of oil together with the aid of a surface.
Tiny gas bubbles tend to attach themselves to solid and liquid contaminants and cause them to
float to the surface for easy removal.
In the oil and water separation process, there are four basic steps. These steps can be classified as
production, dehydration, water clarification, and reuse or disposal.
3.3.1 Production
Production equipment includes everything that brings the oil and water to the surface. This
includes producing wells, headers, automatic well test equipment (AWT), and flowline. All of
this equipment is a source of agitation that thoroughly mixes the oil and water (Figure 1).
3.3.2 Dehydration
The dehydration equipment includes all the various tanks that help the oil and water separate.
Some tanks, called free water knockouts (FWKO) simply provide time for the easy, or free
water, to separate. Next, a Heater Treater (HT) is commonly used to heat the crude oil emulsion
(as described in section 3.2). Heat helps break the emulsion and lowers the percentage of water
in the oil to pipeline specifications.
Dry
Emulsion Oil
FWKO Heater Treater
Water
Surge
Tank
Media Filters Gas Flotation Water
To Injection/Disposal
Figure 1
3.4.1 Coagulation
The processes of coagulation and flocculation are very important ways to separate contaminants
from dirty water. Often the two terms are used interchangeably, but knowing the differences will
give a much clearer understanding of water clarification.
Coagulation is the process of encouraging small droplets of oil to merge or coalesce to form
larger droplets.
As the droplets increase in size, they separate from the water phase many times more easily. In
order to promote coagulation there are several stabilizing factors that must be overcome.
Usually droplets of oil and particles of solids possess a natural negative charge, which is the
result of various negative ions and dissociated organic acids. These charges result in a force of
repulsion often measured as zeta potential. Crude oil contains a variety of organic acids called
napthenic acids. These not only increase zeta potential, but also increase the dispersability of oil
in water by a process known as surfactant stabilization. Until chemicals are added to neutralize
the zeta potential and decrease surfactant stabilization, coagulation is usually difficult and very
slow.
Solids can interfere with both oil dehydration and water clarification. They give rise to steric
stabilization, which simply means the solids act as a physical barrier preventing contact between
droplets. Solids often come up from the producing formation as sand, silt, and clay. Solids can
also be formed by the process of corrosion, scaling, and microbiological activity. Many of these
solids tend to coat the surfaces of the various droplets and hinder the coagulation process.
Chemicals having a degree of surfactant nature are able to remove solids from droplet surfaces
and enhance coagulation.
Even after the zeta potential has been neutralized and the solids have been treated, there remains
the problem of surface tension. Surface tension acts like a skin around each droplet that must be
broken before the droplets can coalesce. Chemicals that can minimize or reduce surface tension
make collisions between adjacent droplets more likely to result in coagulation.
If the zeta potential is not completely neutralized, there will remain a small force of repulsion
between the contaminants preventing coagulation. However, naturally occurring forces of
attraction, known as Van Der Waals forces, often become strong enough to hold the tiny
contaminants in close proximity. Thus, the zeta potential is neutralized; a point is reached where
tiny flocs begin to form. Depending on the amount of solids, the floc will float or sink, and the
contaminants will begin to separate from the water.
Since flocs are very loose agglomerations, a small amount of agitation can easily break the floc
and redisperse the oil and solids. Fortunately, flocs can be stabilized and their size increased by
polymer bridging. Polymer bridging uses very high molecular weight molecules to physically tie
the individual and flocculated contaminants together. Once linked by polymers, the flocs are
stabilized and able to survive more vigorous agitation.
3.4.3 Adsorption
In a filter or coalescer, oil and solids are removed from water by grains of media. Media can be
sand, garnet, coal, or even ground-up walnut shells. Any of these materials will remove
contaminants by trapping them between the various grains of media. Very small contaminants
will be able to find their way through passages between the media. However, if the media is
treated with a highly cationic polymer, many of the small negatively charged contaminants will
be attracted to and adsorbed onto the media. Thus, adsorption can be used to increase the
efficiency of filters and coalescers.
Metal salts are excellent choices for neutralizing zeta potential. Metallic ions with a multiple
positive charge are the most efficient. Zinc chloride and aluminum chloride are the most
common examples.
3.5.1.2 Polyamines
Like metallic salts, polyamines help neutralize zeta potential and speed up the coagulation
process. The significant advantages of polyamines are the abundance of positively charged
amine sites plus the surfactant and polymer bridging qualities that can be built into the polymer.
Polymers with high-charge density can be used as alternatives to metals. In filtration
applications, highly charged polyamines easily attach themselves to the filter media and still
have enough cationic charge potential to help remove oil and solids from the water. Higher
molecular weight solution polymers have the ability to help as flocculants through polymer
bridging. In all of the polyamines, there is a degree of surfactancy not found in metal salts, which
helps reduce surface tension and increases solids wetting.
Reverse emulsion breakers are polyamines proven especially useful for resolving oil-in-water
emulsion. These products incorporate surface-tension-reducing properties into the polyamine
chemistry and are often blended with metallic salts that speed coagulation. Other chemistries
utilized in total fluid are anionic polyacrylamides and dithiocarbamates. These are common to
production systems in the Gulf of Mexico and are used during well flow-back. These products
can also be used continuously in total fluid to improve interface and water quality in separators
with water cuts typically above 50%.
Oil external latex polymers are made from water-soluble monomers, suspended as small droplets
within a surrounding organic liquid. Droplet formation is done before polymerization. This
permits much larger polymers to be formed within the individual droplets without the problem of
long polymer chains forming very thick or solid products. Thus, for polyamines that become
more efficient as their molecular weight increases, a latex version of the chemistry is often
chosen.
Oil external latex polymers can be either anionic or cationic and have molecular weights in the
millions. When oil external latex is used, the polymer must be inverted to release the various
charged molecules. Inversion is accomplished through the use of surfactants, which disperse the
organic liquid and release the water droplets containing the desired chemical. This inversion
process requires time and agitation and is sometimes a problem for oil field applications.
However, the added efficiencies obtained with the latex products often make up for the extra
effort.
In some applications, latex polymers are extremely efficient, and only very low dosages are
needed. Often one quart of chemical can be too much. In order to make these low dosage rates
practical from the application standpoint, the latex is actually diluted with water to make a less
concentrated product. Of course, the polymer inverts in the presence of water, and although the
product concentration is low, the viscosity is often high. The high viscosity is due to the very
high molecular weight of the polymer. Both anionic and cationic latex products can be pre-
inverted.
Water external latex polymers offer an alternative way to make high molecular weight polymers.
Water instead of oil is used as the external or continuous phase of the latex process. Water
external latex products depend on the use of surfactants to isolate the droplets. When the
polymerization process is complete, the surfactants continue to stabilize the droplets and prevent
the droplets from coalescing. This technology is less common than external latexes. However,
these polymers are easy to pump, disperse quickly in produced water, and are often preferred for
field applications.
The following is a brief evaluation guide, intended to help in the selection of water clarifiers for
oil field systems. An outline of methods of selection for various systems, such as Wemco’s, gas
floatation cells, corrugated plate interceptors, and Vortols are included, along with some useful
tips taken from experienced experts in the field. (These methods should be noted and used in
conjunction with present system practices to reach the best conditions for chemical screening.)
3.6.1.2 Procedure
Add 100 mls of water to be tested directly into sample bottles. Always try to avoid collecting a
large volume of water at the sample source and filling bottles later.
Usually the water contains iron that is observed to come out of solution readily. It is advisable to
inject test chemicals into empty bottles and then fill bottles with produced water and cap
immediately. This reduces oxidation and helps in the selection of the most effective compound.
Add neat test chemicals to the bottles. Usually a low treat rate is used (1 to 50 ppm), however
treating rates as high as 1,000 ppm may be necessary. The undiluted chemicals can be added at
these low levels using a 10 micro liter dispenser.
Guard against variations in water quality by including a standard in each set of tests. A standard
can be a blank (no chemical added) or any chemical that shows promise.
Agitate. The amount of agitation to use is hard to determine, usually 50 to 100 shakes (by hand)
is used, unless the distance between the chemical injection point and the treating vessel is
unusually short.
After running a treatment performance curve with the incumbent product, select the most
effective dosage rate and begin comparisons of chemicals in your kit. Best candidates are then
subjected to decreasingly lower injection rates. Some compounds, while highly priced, may be
active at a lower dose; allow for these differences in bench testing.
3.6.1.3 Evaluation
Results of the test are taken by visual observation. Observe the water clarity of each bottle by
looking through the bottle at a printed surface such as a paint pen. Also consider the consistency
of the floc produced. A very fine floc that easily disperses throughout the water will not perform
well in most systems. A very thick sticky floc can form a pad in the vessel that the oil skimmings
return to. Observe the bottles within the first five minutes, as oxidation will rapidly occur.
Oil Field Chemicals Training Manual
3-12
Property of Nalco Energy Services
Confidential & Proprietary – DO NOT DUPLICATE
Chapter 3: De-oiling Produced Water
The above are comparable within one set of readings and should not be compared with those
taken at other times.
It is also important to note the appearance at the interface. Note if there is a clean oil layer
floating on the water or whether there is a floc present. A brisk twist or tilting of the bottle is
used to disturb the water, and an observation should be made as to whether a floc is thrown down
or not. Some water clarifiers containing metallic salts may cause a floc problem that is
undesirable in the system. Amines and quaternary compounds tend to be less troublesome.
Polyacrylates and dithiocarbamates can often help with interface and water quality in turbulent
under-designed vessels. Inject these products into total fluid at 5 to 15 ppm before initial
separation to improve the water quality leaving the vessel. This ultimately makes for easier
treatment by sending cleaner water to the CPI, DGF, and IGF cells. Bottle Test these chemicals
using produced fluids before initial separation (total fluids).
Citric acid can be added to the water at 50 ppm to reduce oxidation during the test period. At the
end of the test, allow the water to settle for 30 seconds prior to pulling a sample for oil-in-water
content. Actual effective treatment dosage, found in this manner, should closely correlate with
that required in the system. (Test volume of water required with the bench model is 3 liters; for
example, 45 μl of straight chemical would be required for a dosage rate of 15 ppm.) Products
that produce a stable form of oil immediately and continue to lift oil over the test period would
normally be in the final selection group.
Products selected should work very quickly to break any reverse emulsion that may be present in
the produced water. Vortols are very effective when all oil present is free and oil droplets are
above a critical particle size. Vortols are ineffective on oily waters that contain untreated reverse
When selecting a product for use in CPIs or similar systems, remember the turbulence, which in
most cases is minimal. In a Bottle Test, little agitation, at times 50 slow rolls or less, may be
appropriate in an attempt to simulate system turbulence. Products such as polyacrylates,
carbonates and those containing salts, such as aluminum sulfate or aluminum chloride, may be
very useful.
Directing produced water into a pit or large vessel permits contaminants to separate naturally.
Solids will sink and organic liquids will float when the amount of agitation or turbulence is
reduced.
Gas flotation systems, especially induced and dispersed systems, are becoming very common.
They are fast and very efficient.
The most commonly used chemicals are dispersion polymers and copolymers. When these
dispersion polymers are applied correctly, achieving complete inversion, they are able to remove
oil very efficiently. They help form stabilized floc and help create a lasting froth on top of the
water. As the contaminants begin to ride on top of the foam, they can be quickly and easily
removed.
In some systems, reverse emulsions or heavy amounts of dispersion enter the water clarification
system. These systems are frequently old and in poor repair. Metallic ions or highly charged
polyamines are frequently used in combination with dispersion polymers and polyacrylamides.
The cationic coagulants destabilize the contaminants and improve the speed of oil removal by the
flocculant.
Systems with high concentrations of brine frequently have high amounts of dissolved iron. This
iron can precipitate as it comes in contact with oxygen. Precipitated iron plugs injection wells
and can cause a sheen when water is discharged from offshore platforms. For these systems,
oxidizers or dithiocarbamates are used to precipitate the iron before the water leaves the
clarification equipment.
Many systems improve water quality by forcing water through beds of sand. In these systems
highly charged epi/DMA/NH3 and aniline formaldehyde polymers are used. Cationic charged
polymers help charge the sand filter media and attract the negatively charged contaminants.
Adding the chemical directly in front of the filter or coalescer does this most efficiently. By
putting the chemical as close to the media as possible, the interaction of the chemical with the
media is maximized.
3.8 Monitoring
When chemicals are applied to the customer’s system, water quality must be tested and
compared to previous or desired performance. The following techniques are commonly used to
help quantify the benefits of a new water clarifier product.
3.8.2 Turbidity
Turbidity meters are fairly common in water clarification plants. Testing samples is fairly easy.
A water sample is put in a glass cuvette and a NTU reading is taken. Readings start at zero for
distilled water and rise as the sample contains higher and higher quantities of light-scattering
impurities.
3.8.3 Infrared
Infrared spectrometers are sometimes preferred to visible spectrometers for determining oil ppm.
This is because some oils have very little visible color or darkness. Especially in the analysis of
soluble organics, an IR spectrometer is very useful. This testing uses Freon or hexane solvent to
Excessive oil — One of the first things to check in the system is the water outlet of the three
phase vessels. Samples should be taken and visually compared to what is normally coming
out of the vessels. It may be necessary to run IRs on some of the outlets. A sample with
excessive oil or excessively oily water could be an indication that the particular vessel is
experiencing some sort of upset and needs to be corrected. Check the vessel for dump valve
operation and interface pads. Demulsifiers can also dirty the water when more water-soluble
chemistries are blended together. Sending excessive oil out with the water will overload the
water treating system and cause an upset. Some separators have an isolated water section that
can trap oil on the surface. This accumulation of oil will have to be skimmed over and
removed before the water coming out of the vessel will improve.
Lack of floatation — Check the floatation cell and determine if the gas dispersion is
sufficient. Mechanical troubleshooting of the vessel will require checking the motors for
proper rotation and rpm of the impellers. Also the depurators must be free of any solids or
sludge buildup to ensure proper gas bubble dispersion. Salt buildup around the shaft is an
indication that oxygen is entering the vessel. For water circulation models, check the
eductors for plugging that would reduce proper floatation by not inducing sufficient
dispersed gas into the unit. Circulation pumps that have leaking seals are a sign of oxygen
intrusion into the float cell. Significant amounts of solids in the water as well as chemical
interference can also kill the foam in a float cell.
Gas blanket — Having the proper gas blanket is very critical to the efficient operation of a
float cell or any water treating vessel or tank A proper ounce gauge must be used to ensure
that a two- to four-ounce gas blanket is maintained at all times. On some vessels, the gas
blanket can fluctuate and allow oxygen to get in on the down cycle. All seals and hatches on
the water treating should be maintained and no leakage allowed. If gas can escape from a
hatch, it can also be an avenue for oxygen introduction. Some operators are under the false
impression that an oil blanket on top of water will prevent oxygen intrusion. However,
oxygen is much more soluble in oil than in water and will rapidly diffuse through an oil
layer. Water circulation or injection pumps are also sources of oxygen intrusion in a system.
Oxygen can be drawn into leaking pump suctions and seals. If a pump is leaking water, no
matter what the injection pressure, oxygen is being induced in the system If oxygen should
enter a vessel, it will react with the iron that is present in the water and cause iron oxide
formation. Oxidizing iron is one of the most frequently found problems in poor water quality.
The oxidized iron begins to form a solid. Oil can attach to it and make it difficult for the
water clarifier to lift out the iron-oxide oil-covered particle. Also iron oxide alone can make a
Finding the source of the contamination and preventing it from getting into the wet oil tank is
the best way to clear up the problem. If this is not possible, treating the wet oil tank with acid
or some other chemistry may clean up the problem. In some cases, changing the water
clarifier to one that can handle the contamination can be utilized. Care needs to be taken
when selecting a clarifier to handle contamination so that it does not create severe floc
problems. Other things to look for are corrosion inhibitors that petition into the water phase,
soaps from washing down the deck, work-over fluids from wells recently completed or
acidized, pads and interfaces in vessels pulling under, and solids from a particular well or
vessel that are feeding into the system. In all cases, a total systems approach is always the
best way to begin to attack a water-treating problem.
The suggested process to commission the water treating system is to utilize a filtration
company like Cetco filtration. They can be set up on the Wemco or last treating vessel outlet
and filter the water before it goes overboard. This gives you time to select the proper
demulsifier and water clarifier to treat the system. After implementing the water clarifier
program and obtaining sufficient water quality going overboard, the filtration equipment can
be disconnected and sent in. This process ensures no out-of-compliance water ever goes
overboard during the start-up of the water system.
If the filtration company is not an option for start-up of the water treating system, the
chemical program may need to be started up before the oil sales is above the 1% BS&W
level. If you begin to experience water quality problems, the water system can still be shut in
for a short time, and the platform will not have to be shut in due to high BS&W at the Lact
unit. During the initial start-up and unloading of the wells, some of the completion waters
and solids may have accumulated in some of the vessels and still be lying in the bottom.
When the water dumps are opened for the first time, the water and solids will enter the water
system and make the water harder to treat than normal and may cause upsets. This is why
filtration is preferred. However, if this is not available, you will need to open these vessels
slowly and feed them into the water system carefully.
Start-up of the water treating system needs to be thought through and planned beforehand.
This ensures that water treating will have no negative impact on the environment or the
whole treating system. It also ensures uptime on the platform.