You are on page 1of 8

Analytica Chimica Acta 505 (2004) 247–254

Method to reduce the memory effect of mercury in the analysis of


fish tissue using inductively coupled plasma mass spectrometry
Chris F. Harrington∗ , Sheila A. Merson, Tulay M. D’ Silva
University of Leicester, Department of Biochemistry, Biocentre, University Road, Leicester, LE1 7RH, UK

Received 9 September 2003; accepted 1 October 2003

Abstract

The routine determination of mercury (Hg) by inductively coupled plasma mass spectrometry (ICP-MS) is affected by a pronounced
memory effect in the sample introduction system. This results in long washout times for the analyte, which affects the accuracy and reliability
of the analytical procedure. By using a combination of flow injection sample introduction and a sulfur-containing compound in the carrier
solution, it was possible to decrease the memory effect of mercury to that for the internal standard (rhodium). The carrier solution contained
2-mercaptoethanol (2-ME) and the developed method was evaluated using three different fish tissue certified reference materials: CRM 464
(BCR, Brussels); DORM-1; and DORM-2 (NRC, Canada). The samples were mineralized using a combination of concentrated nitric acid and
hydrogen peroxide and heating in a closed microwave oven. The developed flow injection ICP-MS procedure gave values for total mercury
in all three CRM materials in agreement with the certified concentration range. Cold vapour atomic fluorescence spectrometry (CV-AFS)
confirmed the results from the developed method. The developed flow injection method had a detection limit (defined as three times the
standard deviation of the blank concentration) for mercury of 5.1 ␮g l−1 .
© 2003 Elsevier B.V. All rights reserved.

Keywords: Flow injection; Inductively coupled plasma mass spectrometry; Total mercury; Fish tissue; Certified reference materials; Memory effect

1. Introduction cury by reduction of inorganic mercury. Thus, cold vapour


(CV) generation coupled to atomic fluorescence spectrome-
Monitoring of mercury (Hg) in the tissue of edible fish is try (AFS) [3], atomic absorption spectrometry (AAS) [4,5],
extremely important because consumption of contaminated or inductively coupled plasma mass spectrometry (ICP-MS)
fish has caused serious neurological damage to newborn ba- [6,7], are techniques that give low limits of detection and
bies and adults [1,2]. Mercury tends to accumulate in fish can easily be automated. However, these methods are also
tissue, particularly, in the form of methylmercury, which is specific to a single species of mercury and affected by the
about 10 times more toxic than inorganic mercury [2]. The sample matrix. This is because organomercury compounds
consumption of fish is therefore a significant exposure route are not amenable to the formation of elemental mercury (the
for humans to mercury and a number of regulatory organi- cold vapour), so the sample preparation steps must convert
zations monitor mercury levels in fish tissue on sale to con- all the organomercury present in the fish tissue to a single
sumers. inorganic form. This can be facilitated by bromination of
A number of analytical methods can be used to measure the sample digests prior to analysis, which converts all the
total mercury concentrations in a variety of different sample organomercury compounds present to inorganic mercury.
matrices. The best analytical methods, in terms of the lowest Inductively coupled plasma mass spectrometry using
detection limit, are based on determination of elemental mer- conventional sample introduction with a peristaltic pump is
widely used for the determination of trace metals in a wide
variety of different sample matrices. However, a significant
∗ Corresponding author. Present address: Metal Speciation and Bio-
and difficult problem to overcome when using ICP-MS for
transformation Research Group, Leicester School of Pharmacy, De Mont-
fort University, The Gateway, Leicester LE1 9BH, UK.
total mercury analysis is the severe memory effect that is
Tel.: +44-116-2577121; fax: +44-116-2577135. apparent for mercury in the instrument, which has been
E-mail address: charring@dmu.ac.uk (C.F. Harrington). attributed to a combination of sample introduction, spray

0003-2670/$ – see front matter © 2003 Elsevier B.V. All rights reserved.
doi:10.1016/j.aca.2003.10.046
248 C.F. Harrington et al. / Analytica Chimica Acta 505 (2004) 247–254

chamber and nebulisation effects [8]. The consequences to determine which could be used to overcome the memory
of these effects include non-linear calibration graphs, long effect of mercury. A system utilizing 2-mercaptoethanol
washout times, decreasing sensitivity with time, and signals (2-ME) provided linear calibration graphs, without the
dependent on the matrix. need to use the standard additions approach, and was
A number of different approaches have been used with validated using three fish tissue CRMs. The results were
ICP-MS analysis to eliminate the mercury memory effect. confirmed by using CV-AFS, both with and without bromi-
For the analysis of potable water, adding gold to the samples nation, which confirmed that the sample digestion method
and standards as a stabilising agent decreased the memory was effective in quantitatively converting the organomer-
effect [9]. Other work monitoring mercury in sediments cury compounds present into the required inorganic
used a combination of flow injection sample introduction form.
and a Triton X-100/ammonia/ethylenediaminetetracetic
acid (EDTA) carrier solution to eliminate the memory ef-
fects associated with mercury determination and to obtain 2. Experimental
reproducible linear calibration plots [8]. This type of carrier
solution was also used for total mercury analysis of hu- 2.1. Reagents, standard solutions and samples
man blood, urine and nitric acid fish digests using ICP-MS
with conventional sample introduction [10]. A similar mix- Deionized water (>18 M) was obtained from an Elga
ture of reagents were added to the standards and samples Maxima water purification unit (Elga, Marlow, UK). Ni-
measured by conventional ICP-MS during the certification tric acid (Ultrapure, Baker, USA), hydrogen peroxide
of a cod tissue CRM (BCR CRM 422 Cod Muscle) [3]. (Ultrapure, Romil, UK) and 25% tetramethylammonium
This study showed that aqueous calibration standards do hydroxide (TMAH) (GPR, Aldrich, Dorset, UK), were
not produce linear calibrations, resulting in low results and used for tissue digestion. Nitric acid (Ultrapure, Baker),
that isotope dilution calibration could give a low result if 2-mercaptoethanol (Aldrich), Triton X-100 (BDH, Poole
the spike material was not added prior to digestion. Deter- Dorset, UK), ammonia (AnalaR, BDH) and ethylenedi-
mination of the total mercury in human hair and sediment aminetetracetic acid (Aldrich), were used as carrier solu-
CRMs by isotope dilution ICP-MS using hydrobromic tions. Tin(II) chloride (Analytical Reagent, Fisher Scientific,
acid as a wash step after each standard and sample pro- Loughborough, UK), potassium bromide (AnalaR), potas-
duced results in good agreement with the certified reference sium bromate (AnalaR), and hydroxylamine hydrochlo-
values [11]. ride (99%, Aldrich) were used for cold vapour generation
The sample preparation methods used for both methylmer- and bromination. Mercury, rhodium and thallium stock
cury and inorganic mercury have been detailed in a recent standards were obtained as 1000 mg l−1 solutions (Alfa,
European certification [12] of the two tuna reference mate- Johnson–Matthey, Royston, UK) and diluted as required.
rials, CRM 463 and CRM 464. The most commonly used The accuracy of the method was evaluated using three fish
preparation method involved acid digestion, usually nitric tissue CRMs: tuna fish CRM 424 (BCR, Brussels), and dog
acid, with heating via reflux or in a microwave oven. The fish tissue DORM-1 and DORM-2 (NRC, Ottawa).
majority of the analytical methods employed involved cold
vapour generation and either atomic absorption (CV-AAS) 2.2. Instrumentation
or atomic fluorescence (CV-AFS) detection to determine
total mercury. The FI-ICP-MS measurements were performed using
This paper describes the development of a rapid and an Elan 5000A ICP-MS instrument (Perkin–Elmer (PE),
accurate method for the determination of total mercury in Beaconsfield, UK). The plasma was generated using a
fish tissue by flow injection coupled to inductively coupled forward power of 1050 W and gas flow rates of: coolant
plasma mass spectrometry (FI-ICP-MS). Flow injection 15 l min−1 ; auxiliary 0.8 l min−1 ; and nebulizer 0.9 l min−1 .
analysis was assessed because it allows for the on-line addi- A cross-flow nebulizer and nickel sampler and skimmer
tion of the internal standard and it can be used to introduce cones were used throughout the study. The flow injection
small aliquots of solution into the instrument. This reduces apparatus comprised a PE AS90 auto-sampler and a PE
the possibility of experiencing signal stability problems due FIAS 400, made up of two peristaltic pumps, a five-port
to deposition of the matrix from the sample digests on the switching valve, and a 500 ␮l sample loop.
cones and limits the amount of mercury entering the system, The CV-AFS measurements were performed using a sys-
which helps to control the potential memory effect. Two tem comprising an auto-sampler, a continuous flow vapour
sample digestion methods were investigated, both based on generator, a Merlin fluorescence spectrometer and a control
the use of a closed microwave oven to reduce losses of mer- computer (PS Analytical, Orpington, Kent, UK).
cury by volatilization. The first used nitric acid/hydrogen A closed microwave system (MSP 1000, CEM, Buck-
peroxide, the second employed tetramethylammonium hy- inghamshire, UK) was used for the TMAH and nitric
droxide. Quantitation of the total mercury present was car- acid/hydrogen peroxide sample digestions. This particu-
ried out using three different flow injection carrier solutions, lar system utilizes 100 ml polyethylene vessels and the
C.F. Harrington et al. / Analytica Chimica Acta 505 (2004) 247–254 249

temperature and pressure were monitored throughout the 3. Results and discussion
digestion.
3.1. The mercury memory effect
2.3. Sample preparation
To demonstrate exactly what form the memory effect takes
The samples (approximately 0.5 g dry weight) were oxi- when nebulising a mercury solution through a double pass
dized in a closed microwave oven using concentrated nitric Scott-type spray chamber, a 5 ␮g l−1 mercury(II) chloride
acid (6 ml) in the first step, followed by addition of hydro- standard made up in water was continuously aspirated into
gen peroxide (4 ml) in the next. During the first stage the the ICP-MS instrument. The mercury response gradually in-
samples were heated to 100 ◦ C for 10 min. The second step creases with time and does not achieve a maximum (Fig. 1).
used the same parameters for 5 min. Before opening the mi- A solution of mercury(II) chloride at the same concentration
crowave oven, the solutions were fan cooled to 30 ◦ C and was made up in 2-mercaptoethanol (0.05%) solution and as-
opened when the internal pressure had dissipated. The di- pirated for the same time period. The mercury signal in this
gests were filtered (0.45 ␮m) and made up to a final volume case does reach a maximum, before returning to the baseline
of 100 ml giving the solutions a matrix of 7% HNO3 and fairly rapidly after aspiration has ceased (Fig. 1). This clearly
2% H2 O2 . The second digestion procedure involved heating illustrates how effective the addition of 2-mercaptoethanol
the samples (approximately 0.5 g) with 25% tetramethylam- can be in eliminating the memory effect for mercury in the
monium hydroxide (6 ml) in a microwave oven for 20 min. ICP-MS sample introduction system at this concentration
Once cooled, the digests were filtered (0.45 ␮m) and diluted level.
to 100 ml with deionised (DI) water. Once prepared, the di-
gests were stored in a refrigerator until analysed, usually 3.2. Choice of a suitable internal standard
within 5 days of preparation.
The use of internal standardization can improve analyti-
2.4. Flow injection inductively coupled plasma mass cal figures of merit such as precision and accuracy, and its
spectrometry implementation using a FI approach is simple and robust.
The element to use as an internal standard should behave
A simple FI system, described previously for the multi- in the instrument as similarly as possible to the analyte
elemental analysis of food digests [13,14], was employed of interest. Two candidate internal standards, rhodium and
throughout the present work. The system comprised on-line thallium were assessed by comparing calibration linearity
addition to the sample of rhodium (10 ␮g l−1 ) as inter- for the three candidate carrier solutions described below
nal standard, a carrier solution and a 500 ␮l injection (Section 3.3). Rhodium was considered because of its pre-
loop. The flow rates used were: 6.1 ml min−1 for the car- vious successful use by us in the multielemental analysis of
rier; 4.4 ml min−1 for the sample; and 6.9 ml min−1 for the food samples [13,14] and thallium because of its similar-
internal standard. The FI unit was coupled to the ICP-MS ity in mass number to mercury, which has been shown to
system via a piece of 830 mm × 0.5 mm i.d. PTFE tubing. improve precision [15].
The three different carrier solutions investigated were nitric The results from these studies (not reported) showed that
acid (0.1 M), Triton X-100/EDTA/ammonia (0.1/0.1/0.1% Rh gave better calibration linearity and short-term precision
v/v) and 2-mercaptoethanol (0.05% v/v). than Tl, may be because Rh has a first ionisation potential
closer than Tl to that of Hg [16]. The internal standard used
2.5. Cold vapour atomic fluorescence spectrometry in the evaluation of carrier solutions and for method vali-
(CV-AFS) dation was Rh at a concentration of 10 ␮g l−1 which was
added on-line to the samples and standards.
The sample digests were diluted (1 + 9) with hydrochlo-
ric acid (3 M) and 1 g of brominating agent (0.1 M potas- 3.3. Evaluation of different carrier solutions
sium bromide, 0.02 M potassium bromate) was added to
each aliquot and the solutions left to stand for a minimum of Triton X-100/ammonia/EDTA solution has been used pre-
30 min (in this way the organic mercury is oxidized to inor- viously as a carrier of analysis for total mercury in biofluids
ganic mercury). An excess of hydroxylamine hydrochloride by conventional ICP-MS [10] and as a diluent in the de-
was added to remove the excess bromination reagent. The termination of mercury in sediments by FI-ICP-MS [8] and
solutions were analyzed using a continuous flow vapour gen- fish by conventional ICP-MS [3]. Nitric acid has been used
eration atomic fluorescence detector, utilizing tin(II) dichlo- as a carrier solution for the determination of organomercury
ride as the reducing agent. Calibration standards in the range in biological reference materials by FI-ICP-MS [17]. The
1–5 ␮g l−1 were prepared in the same reagent as the sam- use of these two carrier systems was compared to the use of
ples and a water proficiency testing solution (Aquacheck, 2-mercaptoethanol (0.05% v/v) solution, to determine which
WRc, UK) was analyzed concurrently with the samples, as overcame the mercury memory effect most effectively. This
a quality control check. was assessed and compared for all three carrier systems by
250 C.F. Harrington et al. / Analytica Chimica Acta 505 (2004) 247–254

Fig. 1. Mercury signals for continuously aspirated 5 ng g−1 standards made up in deionised (DI) water or 2-mercaptoethanol (2-ME) solution.

evaluating the mercury and internal standard peak shapes, rier [9]. The authors felt that this method did not satisfy
the calibration linearity, the short-term precision and the the analytical performance criteria of the United Kingdom
detection limit (defined as three times the standard devia- Drinking Water Inspectorate (DWI).
tion of the blank concentration), for standards in the range The use of Triton X-100/EDTA/ammonia (0.1/0.1/0.1%
1–40 ␮g l−1 . v/v) solution as the FI carrier produced the peak profiles
The highest peaks for mercury were obtained using the shown in Fig. 2b. Tailing was observed for the mercury peaks
nitric acid carrier. The detection limit was 0.8 ␮g l−1 , which and the signal only just returns to the baseline before the
was the lowest for the three methods tested. However, the next replicate is added. In most cases, the figures of merit
calibration linearity was not so satisfactory (r 2 = 0.975, are the same as with the nitric acid carrier: the calibration is
n = 5) and the short-term precision for three injections very close to a straight line (r 2 = 0.999, n = 5); the detec-
of a 40 ␮g l−1 standard was the poorest of the three sys- tion limit is higher (8.7 ␮g l−1 ); but the short-term precision
tems investigated (CV = 9.5%). The mercury peak profiles is about the same (CV = 9.1%). In previous work using
(Fig. 2a) for the three injections tailed badly, the signal did this solution as a diluent for the determination of mercury
not reach the baseline before the next injection, and each in sediments [8] and fish tissue [3], the use of standard ad-
subsequent injection showed greater signal intensity than the ditions combined with internal standardization was required
previous one. The recovery for a 10 ␮g l−1 check standard to overcome the matrix effects and provide linear calibration
was repeatedly below the required quality control limit of graphs.
90%, indicating problems with calibration stability. These The use of 2-mercaptoethanol was suggested after it was
observations are characteristic of the mercury memory ef- shown to improve the liquid chromatographic properties of
fect and indicate that nitric acid is not a suitable carrier for organomercury compounds and eliminate injection problems
the determination of mercury at this concentration level. encountered with liquid chromatography-ICP-MS [18]. A
Nitric acid has been reported previously as a carrier so- 0.05% (v/v) solution of 2-mercaptoethanol was used as the
lution for the determination of organomercury in biological carrier, it generated the smallest mercury peaks of the three
reference materials using FI-ICP-MS [17]. In that work no carrier solutions evaluated (Fig. 2c). The reason for this is
memory effect was reported, or is evident in the peak profiles unclear at present. However, the peak profiles were more
presented, presumably because cysteine acetate is present in acceptable for quantitation, because the analyte signal re-
the sample solutions as a residue from the sample extrac- turns to the baseline level at the same time as the internal
tion method used. However, a method for the determination standard (rhodium 10 ␮g l−1 ). The figures of merit were sat-
of mercury in potable water showed that conventional anal- isfactory for routine applications: the calibration was linear
ysis of a potable water sample spiked at a concentration of (r2 = 0.999, n = 5); the detection limit was 5.1 ␮g l−1 ; and
1 ␮g l−1 with mercury, gave a short-term precision (CV) of the short-term precision was the lowest (CV = 6.7%) of the
22% for 10 replicate injections using nitric acid as the car- three carriers.
C.F. Harrington et al. / Analytica Chimica Acta 505 (2004) 247–254 251

Fig. 2. (a) Signal profile for three replicate injections of a 40 ng g−1 standard of inorganic mercury mixed on-line with rhodium internal standard, using
nitric acid (0.1 M) as carrier. (b) Signal profile for three replicate injections of a 40 ng g−1 standard of inorganic mercury mixed on-line with rhodium
internal standard, using Triton 100-X, EDTA and ammonia solution as carrier. (c) Signal profile for three replicate injections of a 40 ng g−1 standard of
inorganic mercury mixed on-line with rhodium internal standard, using 2-mercaptoethanol solution as carrier.
252 C.F. Harrington et al. / Analytica Chimica Acta 505 (2004) 247–254

Fig. 2. (Continued ).

The peak profile for the internal standard (Rh) using any species (dogfish (DORM-1 and 2) and tuna (TUNA 464))
of the three of the different carriers shows a characteristic were investigated using two reagent systems. The first sys-
negative response at the start and end of each peak. This tem evaluated used TMAH (25% v/v) solution and has
results from a segmented flow of Rh introduced into the been successfully used by other workers for the quantitative
system when the two-way switching valve moves between extraction of mercury and methylmercury from fish tissue
the open and closed positions. This has been observed in using open focused microwave assisted digestion [5]. The
other studies using the same FI set-up [13,14]. second reagent system investigated forms part of the routine
analytical protocol in our laboratory for the determination
3.4. Development of a suitable sample digestion procedure of trace metals in food and has also been used for the de-
termination of total mercury in sediments using an open
For quantitative recovery of the total mercury from the focused microwave assisted digestion [8]. In this two-step
fish tissue matrix, the material needs to be completely sol- approach, concentrated nitric acid is used initially followed
ubilized by the digestion reagent. Two different reagent by hydrogen peroxide.
systems with closed microwave heating were evaluated. A Digestion with TMAH and analysis by FI-ICP-MS
second requirement for full quantitative analysis is that all yielded recoveries of 75–91% (Table 1) for the three CRMs
the organomercury compounds present in the sample are compared to the certified mean values. Only the result for
converted to the inorganic form. This is necessary because DORM-2 was in agreement with the certified range. The
it has been reported that the chemical form of mercury in precision for the analysis of three replicate digests of the
solution influences the final quantitative result when using same material was <10% for DORM-2 and TUNA 464, but
ICP-MS analysis [3]. This conversion was confirmed by for the lower concentration DORM-1 material, the precision
CV-AFS once the most appropriate digestion procedure was 15%. The exact reason for the low recovery for mercury
had been established. The use of a closed microwave sys- when analysing the TMAH digests using the FI-ICP-MS
tem allowed for rapid heating of the digestion vessels and method is not clear. Further analysis of the same digests
mitigated against any loss of volatile mercury containing using the CV-AFS method gave acceptable results (not
species during sample preparation. shown), so the difference is unlikely to be due to inefficient
Three fish tissue certified reference materials represent- digestion. The probable cause is an interaction between the
ing high and low total mercury content (DORM-2, TUNA TMAH in the digest and the 2-mercaptoethanol used as the
464 and DORM-1, respectively) and two different fish carrier, which either reduces the effectiveness of the carrier
C.F. Harrington et al. / Analytica Chimica Acta 505 (2004) 247–254 253

Table 1
Analysis of fish tissue CRM digests by flow injection inductively coupled plasma mass spectrometry, after tetramethylammonium hydroxide digestion in
a closed microwave oven (n = 3)
CRM Mean concentration (mg kg−1 ) Standard deviation (mg kg−1 ) CV (%) Certified range (mg kg−1 ) Recovery (%)

DORM-1 0.60 0.09 15 0.724–0.872 75


DORM-2 4.24 0.20 5 4.38–4.90 91
TUNA 464 4.36 0.30 7 5.14–5.34 83

Table 2
Analysis of fish tissue CRMs by flow injection inductively coupled plasma mass spectrometry, after nitric acid/hydrogen peroxide digestion in a closed
microwave oven (n = 3)
CRM Mean concentration (mg kg−1 ) Standard deviation (mg kg−1 ) CV (%) Certified range (mg kg−1 ) Recovery (%)

DORM-1 0.73 0.05 7 0.724–0.872 91


DORM-2 4.46 0.18 4 4.38–4.90 96
TUNA 464 4.98 0.20 4 5.14–5.34 95

in decreasing the memory effect or facilitates formation of gests was better than 10% for each material. The signal peak
a different mercury species, which behaves differently to profiles for analysis of the DORM-1 digest using the devel-
the standard. oped FI-ICP-MS method are shown in Fig. 3. The profile
The recoveries for all three CRM materials digested for mercury is similar to that for the internal standard and
by heating in a closed microwave oven with nitric acid/ the mercury signal returns to the baseline rapidly after each
hydrogen peroxide (Table 2), followed by analysis using of the three injections, indicating that the memory effect has
FI-ICP-MS, were between 91 and 96%, compared to the been eliminated. The three injection replicates also show
mean certified values. All three of the determined mercury good short-term repeatability. Analysis of the same digests
concentrations agreed with the certified range for each ma- using the CV-AFS system produced very similar recoveries
terial, indicating complete recovery of the total mercury (Table 3) for the DORM-2 and TUNA 464 materials, but a
present in the samples. The precision for three replicate di- much higher recovery for the low-level DORM-1 CRM.

Fig. 3. Signal profile for a nitric acid/hydrogen peroxide digest of DORM-1 fish tissue CRM, with on-line addition of internal standard (Rh).
254 C.F. Harrington et al. / Analytica Chimica Acta 505 (2004) 247–254

Table 3
Analysis of fish tissue CRMs by cold vapour atomic fluorescence spectrometry after nitric acid/hydrogen peroxide digestion in a microwave oven
CRM Bromination Mean concentration Standard deviation CV (%) Certified range Recovery (%)
(mg kg−1 ) (mg kg−1 ) (mg kg−1 )
DORM-1 No 0.97 0.03 3 0.724–0.872 122
DORM-2 No 4.38 0.11 3 4.38–4.90 94
TUNA 464 No 5.06 0.15 3 5.14–5.34 97
DORM-1 Yes 0.99 0.04 4 0.724–0.872 124
DORM-2 Yes 4.68 0.18 4 4.38–4.90 101
TUNA 464 Yes 5.06 0.19 4 5.14–5.34 97
Digests brominated or unbrominated prior to analysis (n = 3).

3.5. Evaluation of the developed protocol (rhodium; 10 ␮g l−1 ), rather than the more time-consuming
standard additions calibration method. The protocol used
The nitric acid/hydrogen peroxide, closed microwave di- 2-mercaptoethanol (0.05% v/v) solution as the flow injec-
gestion method was shown to be compatible with the devel- tion carrier and a nitric acid/hydrogen peroxide digestion
oped FI-ICP-MS system and so was adopted for use. The method, which was applicable to fish tissue samples contain-
digestates were also brominated prior to CV-AFS analysis, ing low (DORM-1) and high (DORM-2, TUNA 464) levels
to check whether the digestion step had completely oxidized of total mercury.
the methylmercury present to inorganic mercury. The results
(Table 3) were exactly the same as without bromination, in-
dicating that the digestion step had fully converted all the Acknowledgements
mercury species present to the inorganic form.
The developed FI procedure allows for the routine use The work described in this report was supported under
of ICP-MS for the determination of mercury in fish tis- contract with the UK Department of Trade and Industry, as
sue samples and requires no further sample manipulation part of the National Measurement System Valid Analytical
than that which is required for conventional trace metal Measurement (VAM) Programme.
analysis of food samples. It could be possible to add
the sulfur-containing ligand directly to the samples and References
standards and run the system as a conventional ICP-MS
method. However, this would require extra sample ma- [1] A. Kudo, S. Miyahara, Water Sci. Technol. 23 (1991) 283.
nipulation, would entail the continuous aspiration of the [2] R.P. Mason, J.M. Benoit, Organomercury compounds in the envi-
sulfur-containing compound into the instrument and may ronment, in: Organometallic Compounds in the Environment, second
not overcome the memory effect of high concentrations of ed., P.J. Craig, Wiley, New York, 2003, p. 57.
[3] M.J. Campbell, G. Vermeir, R. Dams, P. Quevauviller, J. Anal. At.
mercury that could be present in some samples. The use Spectrom. 7 (1992) 617.
of FI allows for rapid sample analysis, on-line addition of [4] J.L. Manzoori, M.H. Sorouraddin, A.M.H. Shabani, J. Anal. At.
an internal standard and reduces the amount of mercury Spectrom. 13 (1998) 305.
entering the ICP-MS system. [5] C.M. Tseng, A. De Diego, F.M. Martin, D. Amouroux, O.F.X.
The limit of detection (LOD, 5.1 ␮g l−1 ) established for Donard, J. Anal. At. Spectrom. 12 (1997) 743.
[6] U. Vollkopf, A. Gunsel, A. Janssen, At. Spectrosc. 11 (1990) 135.
this method is in agreement with that reported for the deter- [7] M.T. Wei, S.J. Jiang, J. Chin. Chem. Soc. 46 (1999) 871.
mination of mercury in sediment (10 ␮g l−1 ) using FI and a [8] A. Woller, H. Garraud, F. Martin, O.F.X. Donard, P. Fodor, J. Anal.
similar instrument [8]. This could be improved to an LOD of At. Spectrom. 12 (1997) 53.
0.1 ␮g l−1 using cold vapour generation and detection using [9] J. Allibone, E. Fatemian, P.J. Walker, J. Anal. At. Spectrom. 14
(1999) 235.
the same make and model of instrument [6]. The use of iso-
[10] J.A. Moreton, H.T. Delves, J. Anal. At. Spectrom. 13 (1998) 659.
tope dilution in conjunction with cold vapour ICP-MS would [11] J. Yoshinaga, M. Morita, J. Anal. At. Spectrom. 12 (1997) 417.
be expected to give the lowest LOD because of the measure- [12] P. Quevauviller, I. Drabaek, H. Muntau, B. Griepink, BCR Informa-
ment of isotope ratios across the peak, and this approach tion; the Certification of the Contents (Mass Fractions) of Total and
achieved a 70 times improvement in the LOD (0.07 ␮g l−1 ) Methyl Mercury in Two Tuna Fish Materials CRMs 463 and 464,
European Commission, 1999.
[7].
[13] C.F. Harrington, B.E. Fairman, T. Catterick, Spectrosc. Eur. 9 (1997)
10.
[14] C.F. Harrington, S. Elahi, P. Ponnampalavanar, T.M. D’Silva, At.
4. Conclusions Spectrosc. 20 (1999) 174.
[15] F. Vanhaecke, H. Vanhoe, R. Dams, Talanta 39 (1992) 737.
[16] A.G. Sharpe, Inorganic Chemistry, Longman, Essex, UK, 1981.
A simple, fast and effective FI-ICP-MS method was devel- [17] D. Beauchemin, K.W.M. Siu, S.S. Berman, Anal. Chem. 60 (1988)
oped, which allows for the production of linear calibration 2587.
graphs using only the on-line addition of an internal standard [18] C.F. Harrington, T. Catterick, J. Anal. At. Spectrom. 12 (1997) 1053.

You might also like