DAS gain calibration has not been run since converter boards were changed
DUE TO Scanning hardware is hung : Power off gantry for 2 minutes .
Issue
Getting error Message 'DAS gain calibration has not been run since converter boards were changed' while running fast
cals.
Troubleshooting
This is often associated with errors in the log identifying the failing controller.
Isolated instances could be related to power faults. Repeated instances should be investigated.
Search the error log for repeated occurrences.
Troubleshoot by attempting manual communication with the offending controller using "ping","nbsClient", etc.
Examples:
ping orp
ping tgp
ping etc
ping stc
ping obcr
Try cycling power to the Table Gantry. Leave the power off for at least (2) minutes to give all power supplies a chance to
bleed-off to zero volts. Often this can be done by using the Gantry Enable switch found on many PDU's. If an FE or In-
House field engineer is on site it can also be done by turning the gantry off and then, after a minute or two, turn back
on using the gantry power switch. In some cases, the whole system will need to have the power cycled.
Resolution
https://psdb2.gehealthcare.com/knowledgecentral/viewContent.do?externalId=74213&sliceId=1 1/4
30/3/23, 11:42 DAS gain calibration has not been run since converter boards were changed DUE TO Scanning hardware is hung : Power off gantry …
Power off gantry for 2 minutes - PDU with external gantry power switch.
When cycling power to the Gantry, always leave it off for two minutes to bleed off all power before turning it back on. This
is especially important on VCT scanners. Failing to leave the gantry off long enough may cause the power supplies
to not start up at the correct voltages and the cooling fans will not work. Verify with the user that they hear the
fans restart after the gantry is powered back up.
On a CPDU, there is a PDU Auxiliary Gantry Power Switch located on the upper right rear of the CPDU as viewed from the
front of the PDU. See illustration 1 below.
Ill 1.
On a NGPDU, there is a green Gantry Enable Switch on the front upper right corner next to the big red Emergency Off
button. It looks like an indicator light, but if it is pushed, the gantry will lose power. Pushing it again restores power to the
gantry. See Illustration 2.
Ill 2.
Of course, if the service person is on site, they can also cycle gantry power using the service panel gantry power switch.
Note: This method is restricted to qualified service personnel only.
After gantry power is restored, you will need to reset the x-ray and drives power with the Gantry Reset button located on
the gantry control panel. With H2 gantries, you will also have to reset Scan Hardware inside Service tools. HP
gantries reset the Scan Hardware when the gantry reset button is pushed.
Ill 3.
https://psdb2.gehealthcare.com/knowledgecentral/viewContent.do?externalId=74213&sliceId=1 2/4
30/3/23, 11:42 DAS gain calibration has not been run since converter boards were changed DUE TO Scanning hardware is hung : Power off gantry …
To do this use the standard best practices, tools and procedures that fall within the scope of past service training
curriculum and employee qualifications.
Follow all Safety and Quality Procedures that are found in the Service Manuals.
It is the service persons’ responsibility to use the appropriate system documentation to effect repairs and prove the service
action had the desired or intended effect, with no adverse effects.
Error Code
With this problem or issue, you may find one or more of the following errors:
Error: 182026 Did not receive LAN Watchdog message from STC
Error: 230023025 Could not obtain the detector temperature prior to starting: status = -1.
Error: 230023027 Could not obtain the detector temperature prior to starting due to a communication timeout.
Error: 244540 Scan Seq Id: 0 Immediate Request cannot be honored since the firmware is in the DOWN state
Error: 244603 Converter board revisions have changed. Das Gain Calibrations may be invalid.
Error: 245011 The firmware wasn't available when a scan request was received
Error: 260007384 Scan ABORT -- DAS Has Detected An Error. The DAS has received an out-of-sequence command
https://psdb2.gehealthcare.com/knowledgecentral/viewContent.do?externalId=74213&sliceId=1 3/4
30/3/23, 11:42 DAS gain calibration has not been run since converter boards were changed DUE TO Scanning hardware is hung : Power off gantry …
Error: 230017818 ACCS CCal: WARNING: Operating point maybe too near the ZRATIO Limits
Error: 230023025 Could not obtain the detector temperature prior to starting: status = -1.
Error: 244520 Scanning hardware has been shut down. NOTE: This is a very generic error. You need to search the log for
a more specific error.
Error: 244540 Scan Seq Id: 0 Immediate Request cannot be honored since the firmware is in the DOWN state
Error: 244634 Subsystem State; DCB:up CCB:up TCB:up JEDI:up TGP/ORP:down DEVICES:up
Error: 245178 An Firmware Complete event was recieved when Scan Manager was not in a scanning state
Error: 245225 Scanmgr received a firmware down event while scanning. Scanning hardware reset might be requested.
Error: 245226 Scanmgr requested hardware reset. Scanning hardware might be down, or the communication with the
firmware might have failed.
Error: 260116313 Table Gantry Processor - Host (OC) LAN Communication Watchdog Timed out.