You are on page 1of 5

Nephrol Dial Transplant (2005) 20: 1951–1955

doi:10.1093/ndt/gfh926
Advance Access publication 5 July 2005

Original Article

Online measurement of haemoglobin concentration

Lindsay Chesterton, Stewart H. Lambie, Lisa J. Hulme, Maarten Taal,


Richard J. Fluck and Christopher W. McIntyre

Department of Renal Medicine, Derby City General Hospital, Derby, UK

Downloaded from https://academic.oup.com/ndt/article/20/9/1951/1850196 by guest on 06 May 2022


Abstract taken to achieve clinically important targets and give
Background. Regular monitoring of haemoglobin in early warning of suboptimal response to treatment.
chronic haemodialysis patients is essential to ensure
that targets for anaemia management are consistently Keywords: haemoglobin concentration;
achieved. Repeated blood sampling can be time on-line monitoring
consuming, invasive and, for pragmatic reasons, only
infrequently performed, often delaying therapeutic
change. On-line optical continuous monitoring of the
haemoglobin concentration would allow non-invasive
assessment of haemoglobin, and immediate therapeutic Introduction
changes could be implemented, thereby improving the
efficiency of anaemia management. This study aimed The measurement of haemoglobin (Hb) concentration
to evaluate the use of on-line haemoglobin concentra- is of great importance in the haemodialysis (HD)
tion measurement. population [1], with considerable resources of time,
Methods. Eleven dialysis monitors (IntegraÕ Hospal) effort and finance being expended to ensure that the
were calibrated using at least five haemoglobin samples patients gain the greatest benefit from maintenance
spread over at least 4 g/dl. Optical measurement of of their Hb at normal or near normal levels [2,3]. Hb
haemoglobin concentration is already incorporated measurements are not routinely measured at <4 weekly
into the dialysis monitor to allow the study of relative intervals, as this is the NKF-K/DOQI recommended
blood volume. Fifteen patients were studied with frequency for checking Kt/V [4]. This inevitably
paired haemoglobin measurements (i.e. dialysis moni- introduces a delay between any change in the patient’s
tor value and conventional laboratory assessment) Hb concentration and institution of an appropriate
taken at intervals over 7 months (mean 11.0±0.28 g/dl, therapeutic response. The greatest individual compo-
range 7.5–14.8). nent of this delay is likely to be the time between the
Results. Haemoglobin measured by HemoscanÕ cor- change in Hb occurring and the next routine blood
related well with the laboratory measurements round, but other components are also involved. These
(r2 ¼ 0.83, P<0.0001), indicating that the machine include the time between taking the blood sample
values are broadly comparable with laboratory figures. and processing it, the time between the result being
There was a mean underestimate of haemoglobin available and the appropriate physician being aware
by HemoscanÕ of 0.34%. There was no significant of it, and the time from then until the appropriate
deterioration in the quality of this correlation over the therapeutic response is actioned. All of these factors
study period (r2>0.8). impair the efficient detection and correction of anaemia
Conclusion. The ability of the dialysis monitor to in HD patients.
measure the optical concentration of haemoglobin Potentially, more frequent venesection could help
compared with conventional laboratory assessment is to improve this situation. There would, however, be a
both precise and accurate. Regular on-line assessment significantly increased cost in laboratory consumables,
of haemoglobin may allow more proactive microman- and inconvenience to nursing staff and patients.
agement of renal anaemia, with a reduction in the time Furthermore, increased numbers of blood tests can
result in worsening anaemia if a sufficient volume
of blood is withdrawn [5]. Optical measurement of Hb
Correspondence and offprint requests to: Dr S. H. Lambie,
Department of Renal Medicine, Derby City General Hospital,
(Hbopt) would avoid these problems.
Uttoxeter Road, Derby DE22 3NE, UK. Although optical monitoring of relative change
Email: slambie@doctors.org.uk in Hb concentration has been integrated into many
ß The Author [2005]. Published by Oxford University Press on behalf of ERA-EDTA. All rights reserved.
For Permissions, please email: journals.permissions@oupjournals.org
1952 L. Chesterton et al.
dialysis monitors for some time, this has not generally routine clinical blood sampling guidelines. The laboratory
been used to monitor absolute Hb concentrations [6,7]. values obtained were then compared with Hbopt recorded
The HemoscanÕ (Hospal, Mirandola, Italy) system by DialmasterÕ (Hospal, Mirandola, Italy) at the start of
allows calibration of Hbopt against local laboratory- dialysis. DialmasterÕ is a central computer system with
based measurement. If HemoscanÕ is to be used for the facility to record data from dialysis monitors linked to
Hb monitoring in the clinical arena, it is important to the system. Patients are identified by a card system, allowing
assess its accuracy and reliability. As Hb concentration for subsequent off-line analysis and interpretation of a large
changes over the course of a dialysis session, it would number of monitored variables. In this instance, a record
also be important to check that the initial measurement of Hbopt was studied, and the initial value was taken from
the first Hbopt recorded after dialysis initiation.
of Hbopt correlates closely with the pre-dialysis Hb
A total of 46 data points were collected (11 at initial
measured in the laboratory on a sample taken at the
alignment of the dialysis monitors, 10 from the middle of
initiation of dialysis. the study, nine at completion at the study and 16 from a
The accuracy and reliability of HemoscanÕ have comparison of recorded data with pre-dialysis samples).
not been tested in routine clinical use. The aim of this Autocalibration sufficient to analyse RBV is performed
study was to make an initial assessment of the clinical prior to every dialysis session. There are no manufacturer’s

Downloaded from https://academic.oup.com/ndt/article/20/9/1951/1850196 by guest on 06 May 2022


utility of on-line measurement of Hbopt. guidelines as to the frequency with which further laboratory
alignment should be performed.
Methods
Patients
Õ
Eleven Integra (Hospal, Mirandola, Italy) dialysis machines Fifteen patients were recruited from our chronic HD
were assessed. Relative blood volume (RBV) monitoring by population, mean age 65 years. All patients were undergoing
HemoscanÕ is fitted as standard to these machines. This standard HD. Patients were dialysed using bicarbonate
module allows continuous monitoring of RBV, the display buffering, a dialysate sodium concentration of 140 mmol/l
of signals in numerical and graphical mode, off-line data and a dialysate flow rate of 500 ml/min. HD used either
retrieval and periodic on-board alignment to the laboratory low-flux haemophan polycarbonate membranes (Hospal
Hb values, thereby accounting for any relative drift between HG 500–700) or mid-flux cellulose diacetate membranes
the two devices. Alignment with the local laboratory involves (Hospal Diacepal 20); no dialysers were reused, and a linearly
the simultaneous measurement of Hb by both HemoscanÕ decreasing ultrafiltration profile was used throughout. Those
and standard laboratory assessment, and consists of three known to have haemoglobinaemia or myoglobinaemia or
parts. First, the HemoscanÕ module collects data during a those prescribed rifampicin were excluded. These substances
dialysis session, laboratory data are then inserted and finally absorb light at the same wavelength as Hb and can interfere
standard linear regression is performed. Five separate pairs with the accurate measurement of Hbopt.
of samples are required to align the machines correctly Appropriate ethical approval was obtained from the
(not necessarily from the same patients). Sufficient spread local research ethics committee.
between the results is also required (at least 4 g/dl between
the maximum and minimum recorded values). All blood
samples were taken directly from the arterial line while a Statistical analysis
useable HemoscanÕ trace was available. Hb was subsequently All data were analysed using GraphPad Prism version
measured by the hospital laboratory Sysmex XE 2100Õ 3.00 for Windows (GraphPad Software, San Diego CA,
analyser (Sysmex, Kobe, Japan), which uses the sodium lauryl www.graphpad.com). Data are expressed as mean±SD
sulfate-haemoglobin method, and has a coefficient of varia- unless otherwise stated. Bland–Altman plots were created
tion of <1% for all red blood cell (RBC) parameters [8]. by plotting the difference between Hbopt and laboratory-
The coefficient of variation is higher at lower Hb levels, rising measured Hb, expressed as a percentage of the mean
to 3.9% for samples with a mean Hb of 6.9±0.27 g/dl. Once of the two measurements, against the mean of the two
sufficient numbers and spread of samples are available, measurements.
the IntegraÕ monitor internally aligns Hbopt against the
laboratory samples with standard linear regression. After it
has been performed, the displayed values of Hbopt and blood Results
volume are computed by applying the regression coefficients.
HemoscanÕ has a range of measurement from 7 to 14 g/dl, an
Hb measured by HemoscanÕ shows a strong correla-
accuracy of ±0.5 g/dl if aligned and a resolution of ±0.1 g/dl.
In the first phase of the study, further paired measure-
tion with that measured by SysmexÕ analyser
ments were made from a range of dialysis patients, using all (r2 ¼ 0.89, Figure 1) and is accurate as demonstrated
11 machines, at regular intervals over the course of 7 months, in the Bland–Altman plot (Figure 2). Mean over-
in order to ascertain the accuracy and reliability of the estimate of Hb by HemoscanÕ was 1.3±2.4%.
HemoscanÕ . Blood samples for laboratory analysis were Seven months later, precision and accuracy appear
taken mid-dialysis while a useable HemoscanÕ trace was undiminished (Figures 3 and 4) (r2 ¼ 0.97, P<0.0001).
available, and a simultaneous note of Hbopt was made. At this stage, there is a mean underestimate by
Ten further paired observations were taken in the middle HemoscanÕ of 1.0±1.5%. Two machines had had
of the study. adjustment of their software or other problems, and
In the second phase of the study, 16 blood samples were were no longer aligned. These monitors were removed
taken at dialysis initiation for laboratory analysis, as per from further analysis.
Online measurement of haemoglobin concentration 1953
15.0 15.0

Hb by Sysmex analyser (g/dl)


Hb by Sysmex analyser (g/dl)

12.5
12.5

10.0
10.0
n=9

n=11 7.5
7.5 10.0 12.5 15.0
7.5
Hb by Hemoscan (g/dl)

Downloaded from https://academic.oup.com/ndt/article/20/9/1951/1850196 by guest on 06 May 2022


7.5 10.0 12.5 15.0
Fig. 3. Linear regression plot of haemoglobin measured by the
Hb by Hemoscan (g/dl) SysmexÕ analyser vs HemoscanÕ 7 months after alignment,
Fig. 1. Haemoglobin measured by SysmexÕ analyser vs HemoscanÕ showing a significant correlation between the two measures
immediately after alignment, showing a high degree of correlation (r2 ¼ 0.97, P<0.0001).
between the two measures (r2 ¼ 0.90, P<0.0001).

20 20

Difference between Hemoscan


Difference between Hemoscan

percentage of the mean (%)


percentage of the mean (%)

and Sysmex analyser, as a


and Sysmex analyser, as a

15 15
10 10
5 5
0 0

−5 −5

−10 −10

−15 n=11 −15 n=9

−20 −20
8 9 10 11 12 13 14 15 8 9 10 11 12 13 14 15 16
Mean of Hemoscan and Sysmex
Mean of Hemoscan and Sysmex analyser (g/dl)
analyser (g/dl)
Fig. 4. Bland–Altman plot demonstrating the difference between
Fig. 2. Bland–Altman plot demonstrating the difference values detected by the HemoscanÕ and SysmexÕ analyser
between values detected by the HemoscanÕ and SysmexÕ analyser expressed as a percentage of the mean against the mean of Hb
expressed as a percentage of the mean against the mean of Hb measured by the HemoscanÕ and SysmexÕ analyser 7 months
measured by the HemoscanÕ and SysmexÕ analyser after post-alignment. There is a mean underestimate by HemoscanÕ of
alignment. There is a mean overestimate by HemoscanÕ of 1.0±1.5%.
1.3±2.4%.

HemoscanÕ was also accurate in the second phase


of the study. Laboratory Hb measured in the standard Discussion
pre-dialysis manner was comparable with the value
for Hbopt recorded by DialmasterÕ at the start of Initial results immediately post-alignment comparing
dialysis (Figure 5). In this case, HemoscanÕ under- Hbopt and Hb measured by the SysmexÕ XE 2100
estimated Hb by a mean of 0.7±4.1% compared with analyser show good precision and accuracy. This
the SysmexÕ analyser, though precision was slightly precision and accuracy are maintained in at least
reduced (r2 ¼ 0.75, compared with r2 ¼ 0.90 for all the medium term (over the 7 month period of this
paired samples). study). Overall results suggest that precision is
Finally, in an overall analysis (n ¼ 46), HemoscanÕ increased for higher values. Some of the scatter seen
shows good precision and accuracy (Figure 6). is due to inherent variability within the Sysmex XE
Mean overestimate by HemoscanÕ was 0.1±3.3%. 2100Õ analyser used as a comparator. While there
This is despite combining results from alignment, appears to be less agreement between the two methods
the middle and the end of the study, or from the at lower Hb levels, the documented precision of the
comparison of initial Dialmaster records of Hbopt, SysmexÕ analyser is also reduced for these concentra-
for blood samples and HemoscanÕ results taken tions [8]. Therefore, the scatter seen at lower Hb
simultaneously. concentrations may be due to the SysmexÕ analyser
1954 L. Chesterton et al.
Difference between Hemoscan
20 Our data suggest that while some precision is lost,
percentage of the mean (%)
and Sysmex Analyser, as a
15 this is still a clinically useful measure of Hb. Therefore,
10 off-line results will be available at the physician’s
5 convenience.
0 Measurement of Hbopt may also lead to further
−5 refinement of anaemia management. Measurement of
−10
Hbopt at every dialysis session would lead to early
n=16 detection of anaemia, and therefore should reduce the
−15
response time prior to appropriate clinical therapeutic
−20 action. This in turn should lead to an overall increase
8 9 10 11 12 13 14 15
in the total time that a patient spends with a satisfactory
Mean of Hemoscan and Sysmex
analyser (g/dl) Hb. Secondly, when considering the response to
changes in Hb and erythropoietin doses, currently
Fig. 5. Bland–Altman plot demonstrating the difference between available algorithms [9] depend on considering both
values detected by the HemoscanÕ and SysmexÕ analyser expressed
the absolute Hb concentration and the rate of change

Downloaded from https://academic.oup.com/ndt/article/20/9/1951/1850196 by guest on 06 May 2022


as a percentage of the mean against the mean of Hb measured by
the HemoscanÕ and SysmexÕ analyser using data recorded by of Hb concentration. Similar algorithms could be
DialmasterÕ . There is a mean underestimate by HemoscanÕ of constructed responding to smaller rates of change in
0.7±4.1%. Hbopt, potentially delivering a much more individually
tailored and more efficient erythropoietin regimen.
The full value of anaemia management using Hbopt
Difference between Hemoscan

20 compared with monthly laboratory-based measure-


percentage of the mean (%)
and Sysmex analyser, as a

15 ment would require evaluation by further study.


10 It is important to note that optical measurement
5 of Hb does have some limitations. Its use is restricted
0 in situations where substances that absorb light at
−5
the same wavelength as Hb may be present. These
include treatment with rifampicin, myoglobinaemia
−10
n=46
and haemoglobinaemia. Our data would indicate that
−15 repeated alignment of the HemoscanÕ module with
−20 the local laboratory should perhaps be performed every
7 8 9 10 11 12 13 14 15 16 3 months to ensure that drift does not occur.
Mean of Hemoscan and Sysmex Overall, this study suggests that HemoscanÕ is a
analyser (g/dl) reliable system for producing an accurate measure
Fig. 6. Bland–Altman plot demonstrating the difference between not only of RBV, but also of absolute Hb concen-
values detected by the HemoscanÕ and SysmexÕ analyser expressed tration. This has the potential to be a useful tool
as a percentage of the mean against the mean of Hb measured by
the HemoscanÕ and SysmexÕ analyser for all results taken during
with which to adjust erythropoietin dosages, enabling a
the study (alignment, middle, end and post hoc recorded results). more rapid response to any change in Hb, and hence
Overall, there is a mean overestimate by HemoscanÕ of 0.1±3.3%. more efficient use of a scarce resource.

rather than the HemoscanÕ module alone. In either Conflict of interest statement. C.W.M. has received an unrestricted
educational grant from Gambro Hospal. No other authors have
case, the magnitude of the fluctuations is sufficiently any conflict of interest to declare.
small that they would have little impact on clinical
decision making.
Results from the comparison of off-line analysis References
of the HemoscanÕ results with pre-dialysis blood
samples demonstrated slightly reduced precision and 1. Locatelli F, Conte F, Marcelli D. The impact of haematocrit
levels and erythropoietin treatment on overall and cardiovas-
accuracy compared with the initial alignment samples
cular mortality and morbidity—the experience of the
(r2 ¼ 0.75, as compared with r2 ¼ 0.90). This is still Lombardy Dialysis Registry. Nephrol Dial Transplant 1998; 13:
sufficiently reliable for the vast majority of clinical 1642–1644
situations, and is further supported by the ability 2. Burton C, Ansell D, Taylor H, Dunn E, Feest T. Management
of HemoscanÕ to measure Hbopt at every dialysis of anaemia in United Kingdom renal units: a report from the
session. Any inaccuracy would be automatically UK Renal Registry. Nephrol Dial Transplant 2000; 15:
re-checked after 2 or 3 days, and there will be less 1022–1028
3. Working Party for European Best Practice Guidelines for the
urgency to respond to single results if trends in Hb Management of Anaemia in Patients with Chronic Renal
can be observed. Recorded HemoscanÕ results by Failure. European best practice guidelines for the management
Dialmaster are suitably comparable with pre-dialysis of anaemia in patients with chronic renal failure. Nephrol Dial
Hb results measured in the conventional manner. Transplant 1999; 14 [Suppl 5]: 1–50
Online measurement of haemoglobin concentration 1955
4. I. NKF-K/DOQI Clinical Practice Guidelines for Hemodialysis 7. Moret K, Hassell D, Kooman JP et al. Ionic mass balance and
Adequacy: update 2000. Am J Kidney Dis 2001; 37 [Suppl 1]: blood volume preservation during a high, standard, and
S7–S64 individualized dialysate sodium concentration. Nephrol Dial
5. Napolitano LM. Scope of the problem: epidemiology of anemia Transplant 2002; 17: 1463–1469
and use of blood transfusions in critical care. Crit Care 2004; 8 8. Walters J, Garrity P. Performance evaluation of the Sysmex
[Suppl 2]: S1–S8 XE-2100Õ hematology analyzer. Lab Hematol 2000; 6: 83–92
6. Tonelli M, Astephen P, Andreou P et al. Blood volume 9. Richardson D, Bartlett C, Will EJ. Optimizing erythropoietin
monitoring in intermittent hemodialysis for acute renal failure. therapy in hemodialysis patients. Am J Kidney Dis 2001; 38:
Kidney Int 2002; 62: 1075–1080 109–117

Received for publication: 14.12.04


Accepted in revised form: 4.5.05

Downloaded from https://academic.oup.com/ndt/article/20/9/1951/1850196 by guest on 06 May 2022

You might also like