Professional Documents
Culture Documents
Process:
When manifesting a plan, it is important to consider and evaluate the different tissue
densities the particles of the beam are going to be traveling through. Inhomogeneities such as
metal prostheses, bone, and lung or any other air cavities are important to consider. Due to the
difference in tissue densities throughout the body, there are tools that can be utilized to correct
this difference. A heterogeneity correction can be applied to a plan to account for the change in
inhomogeneities. In this treatment planning project, two plans were developed to evaluate the
differences with the heterogeneity correction on and without the correction. The two plans that
were created were focused on a tumor that is in the left posterior and superior aspect of the lung.
Both plans had an AP/PA beam arrangement, 6X beam energy, and a 1-centimeter multi-leaf
Research:
When inhomogeneities are present, many events can occur that will affect the dose
distribution of a plan. Different densities that the beam will travel through can cause an increase
in scatter, a range of absorptions of the primary beam, and photon interactions causing electrons
to be set into motion. As the beam travels through a lung for example, there could be a loss of
charged particle equilibrium. Then when the beam is about to exit the air cavity, it must build up
again before it enters the tissue. This effect that the photon beam experiences is known as
interface effects.1 Due to this effect, underdosing can occur at both the distal and proximal
interfaces. The range of underdosing that can occur is based on the energy, cavity size, and the
location. To counteract the dose that is lost from the interaction of the beam with different
densities, increasing the field size as well as introducing lateral interfaces are two possible
When creating a breast plan, this interface effect can have a negative impact on the dose
distribution. When arranging tangent fields on a breast plan, there is lung tissue that is going to
be irradiated due to the nature of the body and field angles. Due to this, some parts of the breast
will end up receiving a higher dose because of the lower lung density.2 Another issue that could
arise is when treating either the bone or the soft tissue adjacent to the bone. When a beam is
entering into the soft tissue and then becomes in contact with the bone, the tissue receives an
increase of dose mostly from backscatter that is produced from the higher energy beams. The
bone also causes attenuation from the beam which then reduces the dose. The dose reduction
depends on the soft tissue thickness, bone thickness and density, and the beam energy. 3
High Z material that is used for hip prosthesis or dental fillings can alter the dose
distribution in a plan tremendously. Once the beam interacts with high Z material in any area of
the body, there is an abundance of scatter that is produced within the patient. Dental fillings for
example, can cause the mean dose of the plan to be higher. High doses can also appear around
the low dose streak coming from the high Z material. 4 Due to this scatter and unwanted high
dose, the patient could exhibit side effects to surrounding structures such as mucosal reactions in
the oral cavity. To help account for this scatter and dose inaccuracy caused from the dental
fillings, two to four millimeters of tissue equivalent material should be used to regain
There are two methods that are used to correct for inhomogeneities. The first method is a
one-dimensional method that is based on the path length that the beam travels from the source to
the point. During this path, the electron density information is gathered and applied to the ratio of
TAR and the power law method which are two examples of this type of correction. The second
method is a three-dimensional correction that gathers density data from a CT scan. This type of
method determines the change in the primary photon fluence and then calculates the scatter dose.
4
The equivalent tissue-air ratio is a method that is an example of this type of correction that can
be applied. It does not matter which correction is used, the equation D’=D x CF can be applied to
find the corrected dose (D’) for either method. The uncorrected dose (D) is multiplied by the
Findings:
Plan 1 that was created is utilizing the heterogeneity correction, resulting in a mean dose
of 103.3% and a max dose of 108.9% to the PTV. The maximum hot spot for this plan came to
119.2 % and 100% of the dose only covering 93.3543% of the PTV, as displayed in Figure 4.
With the heterogeneity correction, 105 monitor units were delivered from the AP beam and 255
from the PA beam totaling 360 monitor units. The mean dose for the organs at risk for Plan 1 are
as follows: spinal cord- 0.8%, heart-0.4%, left lung-19.1%, and right lung-0.6%. In Plan 1, the
Plan 2 does not use heterogeneity correction which resulted in a mean dose of 105.2%
and a max dose of 108.2% to the PTV. The maximum hot spot for this plan came to be 112%
with the 100% isodose line covering 100% of the PTV. Without the heterogeneity correction, the
AP beam required 136 monitor units and the PA beam needed 249, totaling 385 monitor units.
The mean dose for the organs at risk for Plan 2 are as follows: spinal cord- 1.3%, heart-0.7%, left
lung-19.1%, and right lung-0.7%. In Plan 2, the isodose lines are uniform in shape with the hot
Summary:
Both plans presented different dose distributions along the beam’s pathway. Plan 2
created a more uniform dose distribution throughout the patient because the planning system
assumed that all the tissue was of the same density. Another difference between the two plans
5
was the number of monitor units that were needed. This difference of 25 monitor units is because
the planning system considers all the tissues to be of the same density and assumes that it is
penetrating through more tissue than it is. In Plan 1, the shape of the isodose lines fluctuate
greatly compared to Plan 2 because they are changing as they interact with different tissue
densities. Plan 2 displays better PTV coverage although this is not an accurate representation
because with the heterogeneity correction, only 93.3543% of the PTV is receiving 100% of the
dose. The comparison of these two plans confirms that we should be using heterogeneity
correction when planning tumors that are within the lung. The clinic that I am located at, utilizes
Figure 4
References
1. McDermott P, Orton C. Dose distributions in two and three dimensions. The Physics &
2010:14.50-14.58
12, 2023.
3. Gibbons JP. Khan’s The Physics of Radiation Therapy. 6th Philadelphia: Wolters Kluwer
Health; 2020.
doi:10.1186/s40729-021-00372-5