(QPTENTECH] “herons toe Earth Testor WTD-20KWe
aiitowtoney "Meanie
[Certificate No. 41323 February 17, 2021 Page: 172
Cal ficate
To
(Megabras MTD-20KWe.
2161106
Date of calibration ‘oari7ia024
Calibration procedure
‘The calibration procedure used was based on the recommendations of the IEC 17025 Standard, inthe technical instruction
L014 (rev 3.1) and in the manuals of operation of the involved instruments,
Used Standards
RESISTANCE DECADE BOX |, YOKOGAWA model 2786-20 Serial number 01017, with calibration certificate
No. 174181101 issued by IPT on 06/19/2019, validity for 2 year.
Fluke 8845A, 6 % digits, digital multimeter, serie N° 1869003 with calibration cerifcate N° 309598, issued on January 25, 2019
by Everett in accordance with the Fluke Calibration Quality System document QSD 117.0., validity for 2 year
Uncertainties
‘The conventional true values are affected with the following uncertainties:
Voltage : +0,5%
Resistance: +0,5%
‘The uncertainty expanded told iis based on a standardized uncertainty combined multiplied by an inclusion factor K=2, for a
level of trust of approximately 95%.
Enviromental conditions
Room temperature: Relative humidity:
252°C ser5% |
ible error ; + (2% read value + 1 % ofthe full range ).
Maximum
‘True Value (V) Error(V) | admissible
error + (V)
504 504 a 301
100.6 100.5 “04 401
150.8 1506 | 02 5.02
‘only be able to be integral
ie reproduction of t iment