(QPTENTECH] “herons toe Earth Testor WTD-20KWe aiitowtoney "Meanie [Certificate No. 41323 February 17, 2021 Page: 172 Cal ficate To (Megabras MTD-20KWe. 2161106 Date of calibration ‘oari7ia024 Calibration procedure ‘The calibration procedure used was based on the recommendations of the IEC 17025 Standard, inthe technical instruction L014 (rev 3.1) and in the manuals of operation of the involved instruments, Used Standards RESISTANCE DECADE BOX |, YOKOGAWA model 2786-20 Serial number 01017, with calibration certificate No. 174181101 issued by IPT on 06/19/2019, validity for 2 year. Fluke 8845A, 6 % digits, digital multimeter, serie N° 1869003 with calibration cerifcate N° 309598, issued on January 25, 2019 by Everett in accordance with the Fluke Calibration Quality System document QSD 117.0., validity for 2 year Uncertainties ‘The conventional true values are affected with the following uncertainties: Voltage : +0,5% Resistance: +0,5% ‘The uncertainty expanded told iis based on a standardized uncertainty combined multiplied by an inclusion factor K=2, for a level of trust of approximately 95%. Enviromental conditions Room temperature: Relative humidity: 252°C ser5% | ible error ; + (2% read value + 1 % ofthe full range ). Maximum ‘True Value (V) Error(V) | admissible error + (V) 504 504 a 301 100.6 100.5 “04 401 150.8 1506 | 02 5.02 ‘only be able to be integral ie reproduction of t iment