Professional Documents
Culture Documents
2 Workshop
Table of Contents
Introduction ........................................................................................................................................... 3
Introduction .......................................................................................................................................... 3
Exercise 1 ............................................................................................................................................ 3
Exercise 3a – Instrument Method for an Ion Chromatography Instrument (ICS 2100) ..................... 10
Exercise 5 – Finalizing the Sequence: Assigning the Methods to the Injections .............................. 34
Introduction
Introduction
This workshop shows the workflow of creating all the files required to start measuring samples in
Chromeleon, when eWorkflows are not used. This workflow consists of:
Setting up a Sequence using the Sequence Wizard
Setting up Instrument Methods
Setting up Processing Methods
Updating the Sequence
Starting the analysis by adding the Sequence to the Queue
The modifications to Processing Methods and Report Templates, as well as generating eWorkflows,
are topics of separate workshops.
Exercise 1
1. Open the Chromeleon Console by double clicking the Chromeleon 7 icon on the computer
This folder will be your personal training folder and contain all the exercises and results of this
training.
3. Create a folder Sequences in your personal training folder.
We are simulating a clean system which has not yet any objects that would be
required for setting up a sequence and running it. During this part of the exercise we
will setup a sequence just containing injections. During additional exercises, we will
generate Instrument Method and Processing inside this new sequence. This keeps
everything together and emphasises the Sequence as the parent object for all
samples and injections and associated items like the Instrument Method and the
Processing Method.
Figure 2: Starting the New Sequence Wizard from the Console’s Create menu
2. Click Sequence.
The New Sequence Wizard opens.
3. On the Instrument page (Figure 3) select the instrument for which you will create the
Instrument Method in Exercise 3 (ICS-2100, UltiMate3000_RS or TRACE 1300). The
following figures show the steps needed to create a Sequence for the UltiMate3000_RS. For
the ICS-2000 and the Trace 1300 similar dialogs will appear.
Click Next.
4. In the Unknown Injections page (Figure 4), enter the following parameters:
Number of Vials: 10 for UltiMate_3000_RS and ICS-2100; 8 for Trace 1300
Injections per Vial: 1
Start Position: RA1 for UltiMate_3000_RS and ICS-2100; 1 for Trace 1300
Injection Volume: 20 µL for UltiMate_3000_RS; 5 µL for ICS-2100 and Trace.
The pattern for the Injection Name is Sample #n.
Note: The allows additional options, e.g. automatic numbering of the samples using a
“User Index”.
Click Next.
5. In the Methods and Reporting dialog, leave the fields for Instrument Methods and Processing
Method empty (Figure 5). We will assign the methods later, once we have created them in the
next steps (Exercise 3 and Exercise 4).
6. Reports Templates are not yet created either, therefore we leave the corresponding field
empty.
7. When not present in the Sequence, Chromeleon 7 will apply pre-defined View Settings
automatically, so we can leave that field empty too. Click Next.
Figure 5: Sequence Wizard – Selection of Instrument and Processing Methods, Report Template and
View Settings
8. In the following dialog you can enter a comment to provide additional information about the
Sequence. Proceed with Next and then click Finish.
9. Save the sequence in the folder Sequences of your training folder. Give the sequence a name
following ISO standard date format <YYYY-MM-DD> followed by a running sequence number
suffix <_XX>. For example: 2017-12-31_01 (Figure 7).
10. The Sequence will be created and displayed in the Console’s Data category (Figure 8).
Figure 8: Sequence / Injection list as created by the Wizard for the UltiMate 3000
Note: The New Sequence Wizard only permits to create Unknowns. After the injections are created
the sample type can be easily changed, for example to Calibration Standard.
Note: You can use F9 to Fill Down for the Injections with the same information, so that it’s not
necessary to edit the lines one by one.
Note: At this point no Level can be defined for the Calibration Standards. This is because there is no
Processing Method present where the Level(s) could be created.
1. As a starting point, select the Sequence that you created in Exercise 2; Chromeleon will then
offer this Sequence as the storage place for the Instrument Method that you will create, and
this is what we want.
Figure 10: Starting the Instrument Method Wizard from the Console’s Create menu
Note that the values for the parameters in these exercises are just examples, not general
recommendations (e.g. for the Instrument Methods in your lab).
For instance, the value for Data Collection Rate should be selected in a manner that your
narrowest peaks should have about 20 data points. Oversampling your chromatogram does
not only increase the amount of data to be stored, it negatively affects performance, and it does NOT
make peak detection better, rather the opposite.
Figure 12 to Figure 22 show all the steps of the Instrument Method wizard.
Figure 12: Instrument Method Wizard – Selection of Run Time and Diagnostic Channels
Figure 13: Instrument Method Wizard – Input of Gradient Type and pump flow rate.
Figure 14: Instrument Method Wizard – Input of conditions for the Eluent Generator and CR-TC
Figure 17: Instrument Method Wizard – Definition of sample preparations steps (if required)
Figure 20: Instrument Method Wizard – Definition of temperatures for column and compartment
Figure 21: Instrument Method Wizard – Input of a possible Comment and finalization of the Wizard.
2. Click Finish and save the Instrument Method inside the Sequence that you created in
Exercise 2 using the name ICS-2100_01.
3. Once you have finished the Instrument Method Wizard, you can re-visit the pages for each
module via the so-called Device Views and modify the parameters (Figure 22).
Alternatively, you can modify the lines of the Instrument Method in the Script Editor
(Figure 23). You can also add new lines and add comments to the lines.
5. Change the concentration via the Device View for the Suppressor back to 30 mM. What do
you observe?
6. Perform a check for the Instrument Method and make corrections if errors occur (Warnings
have to be checked, but not necessarily to be corrected).
DAD:
Wavelengths: See Table 2
Peak Width at half height: 0.03 Minutes
Use Recommended Values: On
3-D Feld Min. Wavelength: 220 nm
3-D Feld Max. Wavelength: 400 nm
3-D Feld Bunchwidth: 1 nm
Slit Width: Wide
No timed events
3DField Off 1
Table 2: Definition of Detector Channels of DAD
Figure 24 to Figure 33 show all the steps of the Instrument Method Wizard.
Figure 25: Instrument Method Wizard – Selection of Run Time and Diagnostic Channels
Figure 26: Instrument Method Wizard – Pressure Limits and Flow Acceleration
Figure 27: Instrument Method Wizard – Definition of gradient (Equilibration Stage removed)
2. Click Finish in the wizard and save the Instrument Method inside the Sequence that you
created in Exercise 2 using the name UltiMate_3000_RS_01.
3. Once you have finished the Instrument Method Wizard, you can re-visit the pages for each
module via the so-called Device Views and modify the parameters.
4. Alternatively you can modify the lines of the Instrument Method in the Script Editor
(Figure 34). You can also add new lines and add comments to the lines.
7. Check the method for the Instrument Method and make corrections if errors
occur (Warnings have to be checked, but not necessarily to be corrected).
1. Click Finish in the wizard and save the Instrument Method inside the Sequence that you
created in Exercise 2 using the name Trace_1300_01..
2. Once you have finished the Instrument Method Wizard, you can re-visit the pages for each
module via the so-called Device Views and modify the parameters.
3. Alternatively you can modify the lines of the Instrument Method in the Script Editor
(Figure 46). You can also add new lines and add comments to the lines.
5. Check the method for the Instrument Method and make corrections if errors
occur (Warnings have to be checked, but not necessarily corrected).
1. As a starting point, select the Sequence that you created in Exercise 2; Chromeleon will then
offer this Sequence as the storage place for the Processing Method that you will create, and
this is what we want.
Figure 47: Starting the Processing Method Wizard from the Console’s Create menu
Note: The 3D Quantitative template layout includes Detection, Component Table, Calibration,
Unidentified Peak Group Table, Chromatogram Subtraction, Advanced Settings, SST/IRC, UV
Spectral Library Screening and UV Settings tab pages.
Other layouts are available to show/hide tabs/columns according to various data processing needs.
4. Save the Processing Method with the name Method1 inside the Sequence that you created in
Exercise 2
The Processing Method does not contain any processing parameter settings yet. Those will be added
and clarified in a separate workshop.
Note: Remember that you can use F9 to Fill Down for the Injections with the same information, so
that it’s not necessary to edit the lines one by one.
Figure 50: Sequence: Instrument and Processing Method assigned to all Injections
2. If necessary, remove sequences which you don’t want to run. Select these sequences and
click the Remove button.
3. Click Add. This opens a dialog, to add a Sequence to be run to the Queue (Figure 52).
Select the Sequence created in Exercise 4.
4. After adding the appropriate Sequence to the Queue, perform a Ready Check in
the Queue tab to check for errors and warnings.
1. From the Instruments category select the Instrument, then the Queue tab, – OR –
In the Data category select the running Sequence.
2. Click Stop. A window will be opened and you will be asked if you want to stop the Queue
Immediately, After current injection or After current sequence.
Choose Immediately.
(a) (b)
Figure 53: Two ways of stopping the Queue: (a) From the Queue tab (b) From the Sequence Status Bar
3. (In the Save dialog) create a new folder in your personal training folder called Templates.
4. (Re)name the sequence giving it a meaningful name like the type of analysis you want to
perform. Example: Aspirin in Tablets.
5. The new Sequence does not contain any raw data and the Injection Status is set to Idle.
Note: It is advisable to store empty sequence templates in a locked folder. This way the sequence
templates cannot be modified by mistake. To lock a folder right-click and select “Read-Only” from the
context menu, or tick the corresponding checkbox in the Sequence’s Properties dialog.
Note: You could also duplicate a Sequence by Copy-Paste. Discuss in the class what would be the
difference compared to Save As… in terms of the Data Audit Trail of the copy.