You are on page 1of 8

Biomaterials 23 (2002) 1995–2002

Influence of fluoride content and pH on the corrosion resistance of


titanium and its alloys
Nicolas Schiffa,*, Brigitte Grosgogeata, Michele Lissaca, Francis Dalardb
a
School of dentistry, University Claude Bernard of Lyon 1, Rue Guillaume Paradin, 69372 Lyon Cedex 08, France
b
Electrochemical and Physico-Chemical Laboratory of Materials and Interfaces, Polytechnic Institut of Grenoble, UMR 5631,
"
INPG-CNRS, Rue de la piscine 38042, Saint Martin D’heres Cedex, France
Received 30 January 2001; accepted 21 September 2001

Abstract

In the last decade, new titanium alloys have been developed in different areas of dentistry, such as TiAl6V4 and NiTi.
Concurrently, treatments using fluoride supplementation, such as odontology fluoride containing gels, have also been widely used in
odontology.
The aim of this study was to compare the NiTi, NiTiCo and TiAl6V4 alloys with the titanium, regarding the corrosion resistance
in Fusayama Meyer artificial saliva with different pH and fluoride contents. In the experimental methods, the corrosion potential
was measured over time, then a linear sweep voltametric analysis measured the polarization resistances and corrosion currents.
Our results have shown that TiAl6V4 alloys have a good corrosion resistance as good as that of titanium in Fusayama Meyer
saliva and acid salivary solution, except NiTi and NiTiCo alloys.
Conversely, we noticed, as for the titanium, a remarkable localized corrosion phenomenon of those new alloys in fluoride and
acid–fluoride salivary solutions. The fluoride ions could cause the breakdown of the protective passivation layer that normally exists
on the titanium and its alloys, leading to pit corrosion. r 2002 Published by Elsevier Science Ltd.

Keywords: Titanium; Corrosion; Titanium alloys; Fluoride; pH; Salivary media

1. Introduction The purpose of the present study is therefore to


compare the electrochemical properties of the TiAl6V4,
Titanium and the titanium-based alloy TiAl6V4 are NiTi and NiTiCo alloys with those of pure titanium, as
chosen for dental implantology because of their a function of the pH and fluoride content of the
resistance to corrosion and their biocompatibility [1–3]. medium.
Similarly, in orthodontics, NiTi memory alloys are
also widely used since they can create a less traumatic
force while still producing very good results [4–7].
2. Materials and methods
In parallel to these practices, fluoride therapy is now
recognized as being one of the principal methods in
2.1. Materials
preventing dental caries.
While numerous studies have been carried out on the
The origin and composition of the titanium, titanium
electrochemical corrosion of titanium and titanium-
alloy and nickel titanium alloys studied are shown in
based alloys, few have rarely taken into account the
Table 1.
influence of fluoride and pH on pure titanium [8–15],
The four samples were selected from metal already
and to our knowledge, the electrochemical behavior of
prepared for implantology and orthodontics use. The
TiAl6V4 and NiTi in a fluoridated medium has never
samples were shaped into a cylindrical form, as shown in
been investigated.
Fig. 1, in order to constitute the cap of a rotating disk
electrode.
*Corresponding author. Tel.: +33-476-86-19-63. The metal part was inserted in a polytetrafluoroethy-
E-mail address: nicolas.schiff@wanadoo.fr (N. Schiff). lene (PTFE) ring, adaptable to the working electrode.

0142-9612/02/$ - see front matter r 2002 Published by Elsevier Science Ltd.


PII: S 0 1 4 2 - 9 6 1 2 ( 0 1 ) 0 0 3 2 8 - 3
1996 N. Schiff et al. / Biomaterials 23 (2002) 1995–2002

Table 1
Origin and composition of materials

Material Composition (wt%) Supplier

Pure titanium grade 2 Ti: base, O: 0.18, N: 0.05, STRAUMANN


C: 0.06, H: 0.013 St Ismier, Switzerland
TiAl6V4 Ti: base, Al: 6, V: 4 CEZUS
Ugine, France
NiTi Sentalloy Ni: 50.4, Ti: 49.6 G.A.C
Tours, France
NiTiCo (experimental alloy) Ni: 51.7, Ti: 45, Co: 3.3 ENSCP
Paris, France

Table 2
Metal P.T.F. E. pH and fluoride concentration of different test media

pH Fluoride
content
Resin Medium 1: Fusayama Meyer saliva 5.3 F
Medium 2: acidified saliva 2.5 F
Medium 3: fluoridated saliva 5.3 1 g/l
Fig. 1. Rotating disk electrode. Medium 4: fluoridated–acidified saliva 2.5 1 g/l

In order to prevent infiltration of the electrolyte, the and Fluocaril, Synthelabo, Le Plessis-Robinson,
space between the metal and the PTFE was filled with a France) is about 4, and that after a meal, the buccal
resin having inert electrochemical properties. cavity pH can fall below this value in confined
The exposed surface of the metal disk constituted the zones [17]. We, therefore, chose the most unfavorable
working electrode (the areas concerned were 0.158 cm2 situation.
for pure titanium and TiAl6V4, 0.005 cm2 for NiTi (3). The third medium was identical to the reference
Sentalloy, and 0.036 cm2 for NiTiCo). medium but was enriched with fluoride ions (Prolabo,
The samples were mechanically cleaned with abrasive Paris, France) with a concentration of 0.1% (i.e., 1 g/l or
strips (BF03 0.3 mm, Tacussel, Villeurbanne, France). 1000 ppm), which corresponds to the average fluoride
concentration of the odontological gels mentioned
2.2. Test solutions above. The pH was still around 5.3.
(4) The last medium was both fluoridated and at
(1). The electrolyte reference used was Fusayama the same time acidified: 0.1 g/l of fluoride ions and
Meyer artificial saliva. All the ingredients were supplied pH 2.5.
by Sigma Chemical Company, St. Louis, USA. The Each of the four materials, described above, titanium,
composition of this solution, which closely resem- TiAl6V4, NiTi and NiTiCo, was tested in the four
bles natural saliva, is KCl (0.4 g/l), NaCl (0.4 g/l), different media. Fluoride ions concentrations and pH
CaCl2  2H2O (0.906 g/l), NaH2PO4  2H2O (0.690 g/l), are shown in Table 2.
Na2S  9H2O (0.005 g/l), Urea (1 g/l). The pH was
measured with an XC601 type glass electrode (Radio- 2.3. Electrochemical setup
meter Analytical, Villeurbanne, France) convected to a
PHM 220 type pH meter (Radiometer, Analytical, We used a glass electrochemical cell (Roucaire,
Villeurbanne, France). The pH of this reference saliva Courtaboeuf, France), with the thermostat set (Bio-
corresponding to our first test medium was 5.3. block, Illkirch, France) at 37.070.11C. The cover of the
(2). The second medium used had the same contents cell was pierced with openings to facilitate the insertion
as the first, but the pH was lowered by adding lactic acid of different electrodes (Radiometer Analytical, Villeur-
(Prolabo, Paris, France). This acid was chosen in order banne, France): TC 100 type satured calomel electrode
to obtain conditions that were as close as possible to the (SCE), platinium counter electrode and a working EDI
clinical reality since this acid is naturally released by 101 type rotating disk electrode. The entire three-
bacteria in the buccal cavity [16]. The pH was fixed electrode assembly was placed in a Faraday cage to
arbitrarily at 2.5, knowing that the pH of the limit noise disturbance, and was then connected to an
commercially available fluoridated odontological gels EG&G PAR 273 A type computer controlled potentio-
(e.g., Fluogel, Dentoria, Le Plessis-Robinson, France stat (Perkin Elmer, Evry, France).
N. Schiff et al. / Biomaterials 23 (2002) 1995–2002 1997

In our investigation, a rotating electrode, with a FUSAYAMA MEYER SALIVA


rotating speed of 500 rpm was used. ACIDIFIED SALIVA
FLUORIDATED SALIVA
0.2 FLUORIDATED-ACIDIFIED SALIVA
2.4. Technique for studying corrosion resistance 0

Corrosion potential was measured over time, then the -0.2

E(V/SCE)
polarization curves (i ¼ f ðEÞ) were plotted once the -0.4
potential had stabilized in order to determine the
-0.6
corrosion resistance of the titanium and titanium alloys
in different media [18]. -0.8
The corrosion potential was measured for a period of
-1
24 h. Voltametry was then performed at a scanning rate 0 5 10 15 20 25
Time(h)
of 0.25 mV/s in a potential range of 800–1200 mV/
SCE. Fig. 2. Corrosion potential evolution against time for Titanium in
Since the surface areas of our four samples were different media.
different, the results are presented according to unit
area. The densities of corrosion current values are given
in A/cm2. Table 3
The polarization curves were used to determine, by Corrosion potentials (mV/SCE) of the samples in different media after
24 h
the Stern–Geary method, the density of corrosion
current (icorr) and the polarization resistance Corrosion potential Ti gr 2 TiAl6V4 NiTi NiTiCo
(Rp ¼ ðDE=DiÞ) at the end of corrosion potential mV/SCE (Ecorr)
exposure after 24 h. The polarization resistance values Fusayama Meyer saliva 25 0 100 420
are given in kO cm2. Acidified saliva 80 50 250 880
The different experimentations ensured a perfect Fluoridated saliva 300 370 300 410
Fluoridated–acidified saliva 850 800 525 410
control of the physicochemical parameters. This permits
the duplication under identical experimental conditions,
therefore leading to comparable and reproducible
results with a maximum error of 10–15% for the values. potential indicated the passivation of the pure titanium
In order to confirm the influence of fluoride, the grade 2 in this medium [20–23].
current density was monitored over time for the pure In the acidified medium, the shape of the Ti gr 2 curve
titanium grade 2 sample by applying a potential of changed very little and the potential obtained was only
100 mV/SCE for 2 h in Fusayama Meyer saliva with very slightly negative (Fig. 2), suggesting that the
different fluoride concentrations. material retained its protective passive layer [24].
In the fluoridated medium, the potential of the pure
2.5. Surface analysis titanium grade 2 decreased, reaching 300 mV/SCE after
24 h (Fig. 2). This suggested that the characteristics of
Two of the alloys, NiTi and NiTiCo, were observed the passive film change in this medium, leading to a
using a LAECA CAMBRIDGE LTD-type scanning reduced corrosion resistance [25].
electron microscope. The samples were then subjected to Finally, in the fluoridated–acidified medium, the
chronoamperometry at a constant potential of 100 mV/ corrosion potential of the pure titanium grade 2
SCE for 5 h in two different media: Fusayama Meyer dropped sharply to a value of approximately 850 mV/
saliva and fluoridated–acidified medium. SCE (Fig. 2), suggesting the complete disappearance of
the protective passive layer and a consequent reduction
of the material corrosion resistance properties [10,15].
3. Results and discussion All the values are reported in Table 3.
Thus, the chronoamperometric examinations per-
3.1. Measurement of corrosion potential of the different formed on the pure titanium grade 2 enabled us to
materials determine the induction time prior to the current density
increase for different fluoride concentrations in Fusaya-
3.1.1. Pure titanium grade 2 ma Meyer saliva. We found that the higher the fluoride
We noted that the corrosion potential of the pure concentration in the medium, the shorter is the
titanium grade 2 stabilized at 25 mV/SCE in Fusayama induction time. This effect can be seen from the results
Meyer saliva medium (Fig. 2). This value is similar to shown on a semi-logarithmic scale in Fig. 3.
that obtained by Grosgogeat [19] in the same experi- These results confirmed the negative influence of
mental conditions. The continued increase in corrosion fluoride on the electrochemical behavior of pure
1998 N. Schiff et al. / Biomaterials 23 (2002) 1995–2002

1% FLUORIDATED SALIVA FLUORIDATED-ACIDIFIED SALIVA


2% FLUORIDATED SALIVA FLUORIDATED SALIVA
5 10
-7 ACIDIFIED SALIVA
0 FUSAYAMA MEYER SALIVA
-7
4 10 -0.1

E(V/SCE)
-0.2
i(A/cm²)

-7
3 10

-7
-0.3
2 10
-0.4
1 10-7
-0.5
0
-2.5 -2 -1.5 -1 -0.5 0 -0.6
0 5 10 15 20 25
Log(t)
TIME(h)
Fig. 3. Current evolution against time for titanium, in different
fluoride concentration media, at 100 mV/ECS (semi-logarithmic scale). Fig. 5. Corrosion potential evolution against time for NiTi Sentalloy
in different media.

FUSAYAMA-MEYER SALIVA FLUORIDATED-ACIDIFIED SALIVA


FLUORIDATED-ACIDIFIED SALIVA FLUORIDATED SALIVA
FLUORIDATED SALIVA ACIDIFIED SALIVA
0.2 ACIDIFIED SALIVA -0.2 FUSAYAMA MEYER SALIVA
-0.3
0
-0.4
-0.2
-0.5
E(V/SCE)

E(V/SCE)

-0.4 -0.6
-0.7
-0.6
-0.8
-0.8
-0.9
-1 -1
0 5 10 15 20 25 0 5 10 15 20 25
TIME(h)
TIME(h)
Fig. 4. Corrosion potential evolution against time for TiAl6V4 alloy in Fig. 6. Corrosion potential evolution against time for NiTiCo alloy in
different media.
different media.

titanium grade 2, as previously demonstrated by than pure titanium grade 2 and TiAl6V4, confirming the
Toumelin–Chemla [15], Boere [8] and Reclaru [12]. results of Rondelli [30–31] who obtained a value of
We, therefore, decided to compare this metal to 160 mV/SCE in the same experimental conditions.
titanium-based alloys tested in the same conditions. In acidified saliva medium, this alloy is deteriorated,
with a corrosion potential of 250 mV/SCE (Table 3),
3.1.2. TiAl6V4 suggesting an increase in the corrosion rate of the
For the TiAl6V4 alloy, the shape of the corrosion material in this medium.
potential curves over time was the same (Fig. 4). In In a fluoridated saliva medium, the alloy also
Fusayama Meyer saliva, the corrosion potential was deteriorated, reaching a potential of approximately
approximately 0 mV/SCE (Table 3), a value similar to 300 mV/SCE (Table 3), which was roughly the same
that obtained by Grosgogeat [19], Khan [26] and as in the case of pure Ti gr 2 and TiAl6V4.
Ricardo [25], also indicating that the metal underwent In the fluoridated–acidified saliva medium, the
passivation. potential dropped sharply, as observed for pure Ti gr
In the other three media, the results and conclusions 2 and the TiAl6V4 alloy, reaching a value of 550 mV/
were the same as for pure titanium grade 2. SCE (Table 3).

3.1.3. NiTi 3.1.4. NiTiCo


For the NiTi alloy, the shape of the curve in Lastly, the shape of the curve obtained for the NiTiCo
Fusayama Meyer saliva also indicated that a passive alloy (Fig. 6) and the very negative corrosion potential
layer formed on the metal [27–29] (Fig. 5), even though value of 400 mV/SCE (Table 3) after 24 h corrosion in
the corrosion potential was negative: 100 mV/SCE Fusayama Meyer saliva medium suggested a poorer
(Table 3). This result indicates that NiTi is less noble corrosion resistance [32]. The metal became less noble,
N. Schiff et al. / Biomaterials 23 (2002) 1995–2002 1999

suggesting that the characteristics of the passive film


change. This result is contrary to the findings of Chern
[33] who reported a more noble behavior for this
material. This difference may be due to the machining
of the sample and in particular the heat treatment.
Then, in acidified saliva medium, the NiTiCo alloy
deteriorated even more, with the value dropping to
880 mV/SCE (Table 3).
The fluoridated saliva medium was found to have the
same influence on corrosion (410 mV/SCE) (Table 3)
as in the case of the other alloys.
On the other hand, this was the alloy on which the
effect of the fluoridated–acidified saliva medium seemed
to be the least negative (400 mV/SCE) (Table 3),
though we should bear in mind that the corrosion
resistance of this alloy was supposed to be considerably
poorer in Fusayama Meyer saliva medium than in those
Fig. 8. Polarization curves for TiAl6V4 alloy in different media.
of the other samples.

3.2. Polarization curves


To the best of our knowledge, no studies have been
3.2.1. Pure titanium grade 2 and TiAl6V4 performed on TiAl6V4 in fluoridated medium.
Similar polarization curves were obtained for pure
titanium grade 2 and the TiAl6V4 alloy (Figs. 7 and 8). 3.2.2. NiTi and NiTiCo
The values given in Table 4 suggested that densities of The results obtained for NiTi and NiTiCo alloys were
corrosion currents of these two materials increased, different from those obtained with pure titanium and
indicating that their corrosion resistance decreased with TiAl6V4 (Figs. 9 and 10). As summarized in Table 4, the
medium changes: acidified, fluoridated or fluoridated– corrosion current density obtained in Fusayama Meyer
acidified. saliva, fluoridated saliva and acidified saliva media
The polarization resistance values in Table 5 indicate is higher, suggesting an insufficient electrochemical
an inversely proportional change: as the medium resistance.
changed to acidified, then to fluoridated and then to However, it is interesting to note that these two alloys
fluoridated–acidified, the values decreased, showing an seem to be less affected by the fluoridated–acidified
increase in corrosion rate. saliva medium than were pure titanium and TiAl6V4:
These different results are confirmed by different the corrosion current densities were also lower and the
studies [8,10,12] performed on pure titanium in Fusaya- polarization resistance values were of a similar magni-
ma Meyer saliva and acid–fluoridated saliva media. tude (Tables 4 and 5). These findings confirm our first
results, namely, that in Fusayama Meyer saliva and
acidified saliva media, the corrosion resistance of the
NiTi and NiTiCo alloys was low compared with that of
pure titanium and TiAl6V4.
On the other hand, in fluoridated saliva and
fluoridated–acidified saliva media, the polarization
resistance values of NiTi and NiTiCo alloys did not
drop as sharply as they did in that of pure titanium and
the TiAl6V4 alloy.

3.3. Surface analysis

It was decided to perform surface analyses on the


NiTi and NiTiCo alloys since they displayed the highest
risk of corrosion. The samples were observed using
scanning electron microscopy (SEM)[34]. The surface
micrographs obtained are shown in Figs. 11–14.
They show a general change in the surface of the NiTi
Fig. 7. Polarization curves for titanium in different media. and NiTiCo in Fusayama Meyer saliva medium, while
2000 N. Schiff et al. / Biomaterials 23 (2002) 1995–2002

Table 4
Densities of corrosion currents values (106 A/cm2) of the samples in different media determined by the Stern–Geary method

icorr (106 A/cm2) Ti grade 2 TiAl6V4 NiTi NiTiCo

Fusayama Meyer saliva 170.2 1.570.2 2072 1572


Acidified saliva 370.3 3.570.3 4574 3073
Fluoridated saliva 10.572 970.2 5075 6075
Fluoridated–acidified saliva 500755 480750 250730 300725

Table 5
Polarization resistance values (kO cm2) of the samples in different media determined by the Stern–Geary method

Rp (O cm2) Ti grade 2 TiAl6V4 NiTi NiTiCo

Fusayama Meyer saliva 500745 430745 6076 7078


Acidified saliva 300735 350730 4575 4075
Fluoridated saliva 6075 4574 2073 1572
Fluoridated–acidified saliva 170.2 270.3 570.5 370.3

Fig. 9. Polarization curves for NiTiCo Sentalloy in different


media.
Fig. 11. SEM photomicrograph  1000 of NiTiCo alloy after
chronoamperometry in Fusayama Meyer saliva.

in the much more aggressive fluoridated–acidified saliva


medium the metal is characterized by localized pitting.
These results confirm those of Grimsdottir [16] who also
observed surface degradation of NiTi alloys in the form
of localized pitting corrosion.

4. Conclusion

The results of the present study have made it possible


to classify the different titanium alloys currently used in
Odontology according to their corrosion resistance, with
pure titanium grade 2 used as the reference.
The electrochemical tests performed using poten-
Fig. 10. Polarization curves for NiTiCo alloy in different media. tiometric and voltametric techniques revealed an
N. Schiff et al. / Biomaterials 23 (2002) 1995–2002 2001

Fig. 12. SEM photomicrograph  1000 of NiTi Sentalloy after Fig. 14. SEM photomicrograph  1000 of NiTi Sentalloy after
chronoamperometry in Fusayama Meyer saliva. chronoamperometry in fluoridated–acidified saliva.

In the meantime, it would be advisable to limit the use


of fluoridated gels or mouth rinces for young people
wearing orthodontic braces with NiTi wires. Thus,
nickel is known to cause toxic and allergic reactions in
the body, and concerns have been expressed regarding
the biocompatibility of material.
In all cases, fluoride was found to have a negative
effect on pure titanium and the titanium-based alloys
and this information should be widely diffused to all
dental practitioners.

References

[1] Zitter H, Plenk H. The electrochemical behavior of metallic


implant materials as an indicator of their biocompatibility.
Fig. 13. SEM photomicrograph  1000 of NiTiCo alloy after
J Biomed Mater Res 1987;21:881–96.
chronoamperometry in fluoridated–acidified saliva.
[2] Assad M, Lemieux N, Rivard CH, Yahia LH. Comparative
in vitro biocompatibility of nickel-titanium, pure nickel, pure
titanium, and stainless steel: genotoxicity and atomic absorption
evaluation. Biomed Mater Eng 1999;9:1–12.
interaction between the fluoride ions and titanium, [3] Strietzel R, Hosch A, Kalbfleisch H, Buch D. In vitro corrosion of
which caused changes to the protective passive layer of titanium. Biomaterials 1998:;19:1495–9.
the metal. [4] Gil FJ, Planell JA. Shape memory alloys for medical applications.
Patients wearing dental prosthetic reconstructions or Proc Inst Mech Eng 1998:;212:473–88.
who have titanium implants should therefore be advised [5] Torrisi L. The NiTi superelastic alloy application to the dentistry
field. Biomed Mater Eng 1999;9:39–47.
not to use fluoridated odontological gels. [6] Wever DJ, Veldhuizen AG, Sanders MM, Schakenraad JM, Van
The corrosion rates of the TiAl6V4 alloy were found Horn JR. Cytotoxic, allergic and genotoxic activity of a nickel-
to be similar to those of the reference sample. titanium alloy. Biomaterials 1997;18:1115–20.
In the case of NiTi and NiTiCo alloys, corrosion [7] Widu F, Drescher D, Junker R, Bourauel C. Corrosion and
biocompatibility of orthodontic wires. J Mater Sci: Mater Med
resistance was lower than that of the other alloys,
1999;10:275–81.
especially in neutral and acid saliva media. Further [8] Boere G. Influence of fluoride on titanium in an acidic
studies are now needed to investigate these phenomena environment measured by polarization resistance technique.
in more detail. J Appl Biomater 1995;6:283–8.
2002 N. Schiff et al. / Biomaterials 23 (2002) 1995–2002

[9] Johansson BI, Bergman B. Corrosion of titanium and [22] Gluszek J, Masalski J, Furman P, Nitsch K. Structural and
amalgam couples: effect of fluoride, area size, surface preparation electrochemical examinations of PACVD TiO2 films in ringer
and fabrication procedures. Dent Mater 1995;11:41–6. solution. Biomaterials 1997;18:789–94.
[10] Nakagawa M, Matsuya S, Shiraishi T, Ohta M. Effect of fluoride [23] Kedici SP, Aksut AA, Kilicarslan MA, Bayramoglu G, Gokdemir
concentration and pH on corrosion behavior of titanium for K. Corrosion behaviour of dental metals and alloys in different
dental use. J Dent Res 1999;78:1568–72. media. J Oral Rehabil 1998;25:800–8.
.
[11] Probster .
L, Lin W, Huttermann H. Effect of fluoride prophylactic [24] Bayramoglu G, Alemdaroglu T, Kedici S, Aksut AA. The effect
agents on titanium surfaces. Int J Oral Maxillofac Implants of pH on the corrosion of dental metal alloys. J Oral Rehabil
1992;7:390–4. 2000;27:563–75.
[12] Reclaru L, Meyer JM. Effects of fluorides on titanium and other [25] Zavanelli RA, Pessanha Henriques GE, Ferreira I, Almeida Rollo
dental alloys in dentistry. Biomaterials 1998;19:85–92. JMD. Corrosion fatigue life of commercially pure titanium and
[13] Siirila HS, Kononen M. The effect of oral topical fluorides on the Ti-6Al-4V alloys in different storage environments. J Prosthet
surface of commercially pure titanium. Int J Oral Maxillofac Dent 2000;84:274–9.
Implants 1991;6:50–4. [26] Khan MA, Williams RL, Williams DF. In-vitro corrosion and
[14] Thomas DE, Bomberger HB. The effect of chlorides and fluorides wear of titanium alloys in the biological environment. Biomate-
on titanium alloys in simulated scrubber environments. Mater rials 1996;17:2117–26.
Perform 1983;22(11):29–36. [27] Oshida Y, Miyazaki S. Corrosion and biocompatibility of shape
[15] Toumelin-Chemla F, Rouelle F, Burdairon G. Corrosive proper- memory alloys. Corros Eng 1991;40:1009–25.
ties of fluoride-containing odontologic gels against titanium. [28] Ryhanen J, Niemi E, Serlo W, Niemela E, Sandvik P, Pernu H,
J Dent 1996;24:109–15. Salo T. Biocompatibility of nickel-titanium shape memory metal
[16] Grimsdottir MR, Hensten-Pettersen A. Cytotoxic and antibacter- and its corrosion behavior in human cell cultures. J Biomed Mater
ial effects of orthodontic appliances. Scand J Dent Res 1993; Res 1997;35:451–7.
101:229–31. [29] Wever DJ, Veldhuizen AG, Sanders MM, Schakenraad JM, Van
[17] Leung VW, Darvell BW. Artificial salivas for in vitro studies of Horn JR. Cytotoxic, allergic and genotoxic activity of a nickel-
dental materials. J Dent 1997;25:475–84. titanium alloy. Biomaterials 1997;18:1115–20.
[18] Dickinson WH, Lewandowski Z. Electrochemical concepts and [30] Rondelli G. Corrosion resistance tests on NiTi shape memory
techniques in the study of stainless steel ennoblement. Biode- alloy. Biomaterials 1996;17:2003–8.
gradation 1998;9:11–21. [31] Rondelli G, Vicentini B. Localized corrosion behaviour in
[19] Grosgogeat B, Reclaru L, Lissac M, Dalard F. Measurement simulated human body fluids of commercial Ni-Ti orthodontic
and evaluation of galvanic corrosion between titanium/ wires. Biomaterials 1999;20:785–92.
Ti6A14V implants and dental alloys by electrochemical [32] Kim H, Johnson JW. Corrosion of stainless steel, nickel-titanium,
techniques and auger spectrometry. Biomaterials 1999;20: coated nickel-titanium, and titanium orthodontic wires. Angle
933–41. Orthod 1999;69:39–44.
[20] Canay S, Hersek N, Culha A, Bilgic S. Evaluation of titanium in [33] Chern Lin JH, Moser JB, Taira M, Greener EH. Cu–Ti, Co–Ti
oral conditions and its electrochemical corrosion behaviour. and Ni–Ti systems: corrosion and microhardness. J Oral Rehabil
J Oral Rehabil 1998;25:759–64. 1990;17:383–93.
[21] Eliades T. Passive film growth on titanium alloys: physicochem- [34] Kononen MH, Lavonius ET, Kivilahti JK. SEM observations on
ical and biologic considerations. Int J Oral Maxillofac Implants stress corrosion cracking of commercially pure titanium in a
1997;12:621–7. topical fluoride solution. Dent Mater 1995;11:269–72.

You might also like