Professional Documents
Culture Documents
HBM
' spectris group
The Application
·and Installation of
Load Cells
Gert Gommola
ta
HBM
(spectris group
The Application
and Installation of
Load Cells
Gert Gommola
Gommola, Gert:
The application and installation of load cells : the ins and outs of weighing technology ;
a planning manual for the implementation of weighing technology systems and
equipment/Gert Gommola. Spectris Messtechnik GmbH. -- Frankfurt/Main : Zarbock, 2000
ISBN 3-9805544-4-9
Introduction .. . . . . . . . . . . . . . . . . . . . . . . . ·· 1-1
Introduction
Meanwhile the same units are now used internationally for measurement.
The basic unit is the kilogram, abbreviated to kg. The law also permits
the use of the gram (g) and the tonne (t).
The growing demand for ever higher quality has combined with
increased product responsibility to drive this development forward in
recent years.
Ever since the invention of the strain gauge (=») (S.G.) this application
of a physical effect has also been applied to electromechanical balances.
effect holds good for both positive and negative strain (elongation or
compression). Within defined limits (Hooke's Law (=») in the
stress/strain diagram) the procedures described are close to linear and
easily reproducible.
(b)
(c)
-
€)
(a)
Mechanical structure of
weighing equipment using
strain gauge load cells
._ I LI <hl
RTN (1 .. .470!) Z7/ HLC (0.5 ... 1 0t) PW(3...660kg)
Figure 3-1 Selected types of HBM load cells and their typical load
capacity ranges
Installing load cells 3--7
Il
In the case of tubular spring elements there can be problems with the
load introduction, and a relatively high hysteresis (=»). For this reason
the design shown in Figure 3.4 is often used. In this case the tube is
divided into segments with the aim of reducing the hysteresis. The load
is introduced through fixed platforms.
Installing load cells 3--9
-. - - - - - - -
'
ll
- - - - -
-.-
- .-
= (T]
For even lighter loads a beam spring element can be used. However, in
this case the signal is dependent on the bending moment. If the load
introduction point is changed under identical loading, the output signals
are different. For this reason a single-beam spring element is rarely used
in load cells, since special measures are required in order to maintain a
constant load introduction point (the lever arm effect perpendicular
distance of fulcrum from line of action of force).
0
D D 0
05 0
Figure 3-8 Variants of the dual-beam spring element
Figure 3-9 Load cell with pressure compensation
In contrast to bending strains, shear strains are constant from the load
introduction point to the clamping point. Shear strains of less than ± 45
degrees arise in the unbending web and are not superimposed by bending
strains in this area, Figure 3 .11.
Installing ioad cells 3-15
5%
IA
7
t
Shear
application they can for the most part be compensated for in the strain
gauge bridge circuitry.
2F
Load cells with diaphragm spring elements, Figure 3.13, can have a very
low profile and are often used where space is restricted. Their spring
deflection at maximum capacity is very small, resulting in a very high
natural frequency (=») which in some load cells may even exceed 20
kHz. They are relatively uncomplicated to produce and can be
hermetically sealed with ease. The diaphragm is specially shaped in order
to obtain a strain field favourable to measurement.
Installing load cells 3--17
Figure 3.14 shows the comparative sizes of a load cell with dual-beam
spring element, maximum capacity 10 kg, and a diaphragm load cell,
maximum capacity 1000 kg.
3-18 Mechanical structure of weighing equipment using strain gauge load cells
Apart from the financial aspect, the chief attraction of such solutions is
when existing plant or equipment cannot be rebuilt.
Installing load cells 3-19
For instance, in most cases container legs are made of untempered steel.
Temperature changes set up additional material strains. Inadequate
compensation for these effects will give false readings which the
downstream electronics can correct by mathematical means to only a
limited extent.
The required long-term stability (=») of the measuring device and its
protection against environmental influences, where fitted, should
similarly be taken into consideration at the design stage. The conditions
for carrying out the calibration (=) and recalibration ( ⇒) of such
weighing equipment are also an important part of the initial
considerations. Consider for example that if just one measuring point has
to be re-applied due to damage, the entire system has to be recalibrated.
Installing load cells 3--21
Load cells should be used in the prescribed direction of loading only. The
direction of loading is indicated by an arrow on many HBM load cells.
(a)
F (b)
V X
(c) ()
F
F
X X
(a) Central load introduction
(b) Of--centre load introduction
(c) Non-axial load introduction producing lateral forces
(d) Torque-loading (torsional) load introduction
(a)
F (b)
F
X
(a) Bearing surface too small, mechanical
deformation of the load cell
(b) Uneven bearing surface
Maximum
capacity
jl
Service load
Measuring range
The specified error limits are not
di
exceeded.
Breakin g
Operating range Sate load limit
'
lo ad
The specified error limits may be
exceeded. }
Maximum loading range Destructive range
Specified error limits are exceeded without damaging The load cell undergoes
the load cell. permanent changes.
In the case of load cells designed for both tensile and compressive
loading, two different minimum zero signals can occur as a consequence
of the most recently applied direction of loading.
and negative maximum capacity. The whole span between positive and
negative maximum capacity amounts to 200 %. For certain load cells the
maximum maintained vibration bandwidth can definitely be said to
exceed 100 %.
•
Signal S
Characteristic curve
for decreasing
loading
Characteristic curve
for increasing
loading
lo Ly
oadL J
Loadings which take these matters into account should as far as possible
be less than or equal to the maximum capacity of the load cell. They
must on no account exceed the safe load limit (=») for the load cell
concerned. If it is likely that the safe load limit will be exceeded, suitable
measures to protect the load cell from damage will need to be built in
(see also Chapter 6 -"Overload protection").
3-28 Mechanical structure of weighing equipment using strain gauge load cells
Installing load cells 4--29
Although there is no need for conventional stay rod restraint, care should
be taken to ensure the container is held firmly in position, especially if
the centre of gravity is not in a favourable position . In the container
weighing field, elastomer bearing platforms constitute cost-effective and
easily implemented structural elements that are satisfactory for low to
mid-range accuracy requirements.
«.t
0
Figure 4-1 ZEL elastomer bearing element, showing its degree of
freedom to absorb a load
C16 self centring load cells from HBM are available as pendulum load
cells with maximum capacities from 20 t to 200 t. They are suitable for
medium to high accuracy requirements.
4i
'T
«
Figure 4-2 C16 load cell (diagram) showing its degrees of freedom
Using pendle bearings and pendle supports, standard "beam form" and
C-type load cells achieve similar properties to self-centring load cells.
This makes it possible to construct container weighing equipment to high
standards of accuracy. The design of the mounting accessories on the
pendle bearing allows a displacement of some 3° without significant
measurement errors. Horizontal displacement of the points of support can
also be tolerated within limits. Whereas ZPL pendle bearings mostly
Installing load cells 4--33
consist of two support pieces and one cylindrical support, on the other
hand ZPS pendle supports fulfil the required function with just one
cylindrical support and one EP03 upper support each.
41}
D
.-.
7
)
1.
I I «fl
When using a multiple ball bearing, note that before putting the
equipment into service the locking screws on the upper plate of the
multiple ball bearing must be removed. They must be replaced with
shorter fastening screws.
Installing load cells 4--35
A multiple ball bearing with a locking screw still in place cannot move
laterally and behaves like a solid bearing. The play which the multiple
ball bearing makes possible must not impede the bearing from reaching
as far as its stop when in situ.
Figure 4.5 shows the structure of a movable support fitted with a load
cell. The load cell stands on a multiple ball bearing and the load is
transmitted to it through an upper support (reference EPO).
Figure 4-5 Load cell for compressive loading with multiple ball bearing
and EPO (upper support) mounting accessories
4-36 Mounting accessories for load cells
\
I
\
I
I
Force feedbacks are used in combination with load cells in the form of
dual-beam spring elements, and enable moment-free loading with tensile
or compressive forces along their line of effect. Their use tends to be
restricted to applications having only one load cell per platform, or they
can have knuckle eyes attached at both ends for use in situations where
the load is suspended.
Knuckle eyes are suitable for use in the event of quasi-static tensile
loading (load reversal = lOHz). The onward connection commonly takes
the form of an actuating fork. In the event of high-frequency dynamic
loading, flexible tension bars should be used.
4-38 Mounting accessories for load cells
8
a) and b) Saddle bearings
c) Ball-and-socket joint with concave
spherical shape
+ + d) Ball-and-socket joint with rounded
corners
a) c)
e) Ball-and-socket joint with groove
b)
LEI
Figure 4-10 Examples of tilting bearings
4.7.1 Stops
4 5 = Bearer
6 5 6 = Foundation plate
Stretched cords:
These transmit no forces in the vertical direction and are therefore very
good at preventing unwanted force bypasses.
4-42 Mounting accessories for load cells
<===lrl==lt---�20--===== =====�es
---=---
)
I I
.�-.-.-- �-
\ -�� \
Bolted stay rods ensure minimal force bypasses in the vertical direction.
On the other hand any slight tilting already present in the stay rods can
cause tightening and set up frictional forces that give rise to vertical force
bypasses. They must therefore be fitted with great care. The container
must also be restrained in such a way that any displacement which occurs
cannot cause the bolted stay rods to tilt.
•
Figure 4-16 Bolted stay rods
Tubular stay rods with knuckle eyes behave in much the same way as
bolted stay rods. The main difference is that the omnidirectional jointed
mounting prevents tilting. Apart from the need for horizontal alignment
of the stay rods during fitting, these tubular stay rods with knuckle eyes
4-44 Mounting accessories for load cells
The knuckle eyes should be protected to prevent their joints from seizing
up, especially when used in the open.
Weigh modules
i
J
·-
Even in these cases it is still necessary to consider how the solid bearings
and modules can best be aligned. Solid bearings exhibit maximum
rigidity along the line of the support rod and maximum flexibility
5-50 Mounting accessories for load cells
perpendicular to the rod. Wind forces and their effects on the weighing
result must also be considered in this connection.
3x12A
•
1 + ¢
+7 +
2a
$
l a
• d
l Plane parallelism
------� 30
f If the bottom plate and head plate are not parallel,
this non-parallelism imposes an additional lateral
0 0 force on the force introduction parts. This lateral
force is absorbed by the linking bar in the
longitudinal direction. The pendle bearing is
deflected in the transverse direction until the
equilibrium of forces is restored. However, in order
to keep the effect of lateral forces to a minimum, the
angle of inclination must never exceed 3°.
Axial misalignment
Axial misalignment between the top plate and the
built-in force introduction gives no cause for
concern, since the plates are screwed together.
Overload protection
Note Some types of HBM load cells are already titted with a built-in
mechanical overload protection.
Experience shows that load cells with a rather low maximum capacity
often become overloaded more quickly than load cells in the higher load
6-54 Overload protection
Design engineers can also set up and use their own preferred overload
protection for a range of load cell configurations. Figure 6.1 shows a
suggested design for an overload protection in the HBM type Z6 load
cell. In this arrangement, any distortion or spring deflection in the load
cell is limited by a mechanical stop.
Spring deflection
limitation (stop)
\!!I
Fine-threaded stop screw
'7LJ
Mechanical stop
Figure 6-1 Suggested overload protection for the HBM type Z6 load
cell
Feeler gauges are used to adjust the stop. Since the typical displacement
by distortion is very small in practice only a few 1/10mm
finethreaded adjustment screws have proved extremely reliable in this
connection. The gap between the load cell and the stop should also be
Installing load cells 6--55
protected from dust and dirt, since contamination of this kind is likely to
cause false measurements due to incorrect spring deflection limitation.
F
Spring device
Mechanical stop
Load cell C2
For instance containers are quite often exposed to the elements or subject
to production-specific influences.
When erecting new containers in the open (e.g. silos, coal bunkers, etc.)
these constitute structures for which the relevant building regulations
have to be observed.
For example, in the case of wind loading this is DIN 1055 Part 4
"Design loads for buildings".
7-58 Design and construction of container weighing systems
Centre of gravity
A s
I I
]«
B]
I I
Height of fill 1
Centre of gravity 1
Height of fill 2
Load cells
Figure 7-2 How the centre of gravity in a container with a sloping floor
changes position in relation to the height of fill
For the sake of stability it is best if the centre of gravity is below the
points of support. The position of the centre of gravity as a function of
the height of fill has a decisive influence on the number of load cells to
be used. If the contents are filled symmetrically to the load cell
Installing load cells 7--61
The horizontal pipe is springy in the vertical direction and its effect
diminishes with increasing length. The tensile or compressive
mechanical force which the pipe artificially exerts on the load cells
diminishes accordingly and ceases to threaten the accuracy of
measurements.
7-62 Design and construction of container weighing systems
----r---
'
I
----r---
'
----r---
'
Several flexible couplings can also be used instead of one long pipe
connection (Figure 7.4). Hose connections made of slightly deformable
elastic material are effective in preventing force bypasses. It is essential
to check that the elastic material is compatible with the material used to
fill or clean the container (for instance if foodstuffs or pharmaceuticals
are involved).
Note Metal bellows are not allowed in certain areas of the food industry where
scrupulous cleanliness is required.
The open connecting branch shown in Figure 7.7 is the best solution for
keeping force bypasses to a minimum. An open connecting branch
prevents contact between the pipe and the container. This type of
7-64 Design and construction of container weighing systems
---,--- I
I
Note Care must always be taken to ensure that the proportion of material in
the connecting pipes is weighed as well. The amount of fill in the inlet
and outlet lines directly connected to the container should therefore be
reproducible at the time of weighing. This means that the lines should be
either always full or always empty during data acquisition.
In the case of liquids and loose materials with centrally filled contents,
possible configurations include two fixed supports and one load cell.
This will work for containers that are constructed symmetrically, so that
regardless of the height of the contents the centre of gravity moves in a
line which is close enough to vertical for the degree of accuracy required.
In all other cases, particularly where even higher accuracy is needed, at
least three load cells should be fitted and in some circumstances even
more.
7-66 Design and construction of container weighing systems
Figure 7-9 Upright tank with two solid bearings and one load cell with a
levelling piece
This design for measuring height of fill uses one load cell in a pendle
configuration and two solid bearings that also restrain the container in
the horizontal direction. This cost-effective design protects load cells
from unacceptable influences.
7--68 Design and construction of container weighing systems
---------------�
4
Eek ....==.::::::::� acnes.saw
platform (diagram)
I
'
-----�-----
e i
'
e
I
I
Weigh Weigh
module module
Weigh module Z6 for Weigh module C2A for Weigh module C16 for
light loads medium loads heavy loads
Three weigh modules with integral stay rods attached tangentially to the
circumference of the structure make sure the container remains
horizontally stable without the need for further measures. Each weigh
Installing load cells 7--71
Weigh modules
(diagram)
Riog�
Ee,
Bracing struts are fitted or welded to the inner wall of the container. The
load is rigidly transmitted to the load cell through a foot . Load cell
weigh modules should preferably be used in this case as well, since for
one thing they already include a lift-off preventer (not shown in Figure
7 .12 for the sake of simplicity).
The system usually needs a sealant, and this is provided by a circular ring
gasket which is flexible enough not to act as a force bypass.
Installing load cells 7--73
Occasionally also the weigh vessel is held firm when not taking part in
the weighing procedure and is only released for weighing.
Simple, flexible tensile tubular stay rods are often used on suspended
containers to overcome or ease the problems of centring, height
adjustment and so forth. As well as the ever-present need for protection
against toppling, stay rods are needed to prevent swaying and twisting.
Installing load cells 7--75
Stay rod
Stay rod
Shift in centre
of gravity
•
- -
S1
Figure 7-16 Horizontal liquid container with one C16 load cell
(diagram)
Using only one load cell limits the options for carrying out compensating
corrections on account of errors caused by off centre weight
introduction.
Ideally this load cell is located centrally beneath the platform. However,
the mechanical design of the load cell (in most cases dual-beam spring
elements, see also Section 3.2.2) also permits off-centre configurations
such as a cantilevered platform.
Balances using only one load cell clearly have a price advantage.
f • • • • •
Spacer
Overload
protection
Ever more frequently these inexpensive designs are also being applied to
packaging machines and check balances as well as tasks for weighing
technology applications in the industrial sector.
In practice, an even more reliable way of protecting the load cell than
placing uncomplicated overload trips directly under the load cell to
restrict spring deflection has been to place stops under each comer of the
platform. These prevent harmful torsional loads from reaching or
damaging the load cell.
8-82 Design and construction of platform balances
Since the platform rests on several load cells it is almost impossible for
torsion loading to affect them.
One disadvantage found in practice is that for many types of load cell,
the weighing configuration requires corner correction. This is done by
moving a calibration weight from place to place on the platform and
8--84 Design and construction of platform balances
adjusting the sensitivity of one or more load cells in the weighing device
accordingly. The aim of comer conection is to obtain the same weight
information regardless of where the object being weighed is placed on
the platform.
The load cells are connected in parallel. This increases the cost of
installation. Also the actual number of load cells incurs more expense
than is the case for single-point solutions.
Electrical engineering
structure of weighing
equipment using strain
gauge load cells
t Us
Us
1 «( t
The arms of the bridge, also known as branches, are formed from
resistors RI to R4. The widest possible diversity of names for these will
be found in the literature. However, the important thing is always to view
them in conjunction with the associated bridge equations.
Zero balance
Linearity correction
Calibration
Resistance balance
The sensitivity is the output signal delivered by a load cell when acted
upon by the nominal input quantity (i.e. maximum capacity). The output
signal results from the effect of the input quantity (in this case the
measured quantity is weight). It therefore excludes the zero signal.
The sensitivity and the signal have the same unit, usually mV /V, without
any reference to the measured quantity. Consequently load cells can have
the same sensitivity for different sizes of input quantities and/or loads.
where Lp is a dead load and S% the corresponding output signal for that
dead load. S is the output signal at maximum capacity. For IBM
loadcells the lower measuring range limit is determined by the load cell
9--90 Electrical engineering structure of weighing equipment using strain gauge load cells
under no load condition, and the equation can therefore be simplified for
L,=0:
C=S, S,
Sn
S,-S,
Sp C = L II -Lp ·L
Lp Load L Ln
► Example:
A C 16/60t load cell has a nominal sensitivity of 2mV /V, that is to say, the
load cell delivers 2 m V of output signal per volt of excitation voltage at a
loading of 60t. Knowing the excitation voltage makes it possible to
calculate the loading acting upon the load cell.
• Nominal sensitivity: 2 mV/V (at 60t)
• = 5V (10V at 60t)
Bridge excitation: Up
• Measured output voltage: U A = 4m V
• 10mV/V: 60t = AmV/V: L,
p Calculated load: L, = 24t
Maximum
capacity
60t
Load cells are passive transducers that convert a mechanical quantity into
an electrical signal. To obtain usable signals requires downstream
electronics. The measuring amplifier, as the second link in the
measurement chain, has two main tasks. The first is to generate a highly
stable excitation voltage; the second is to amplify the millivolt-range
output voltage from load cells up to a standardised volt-range level with
as few errors as possible. The amplified measurement signal commonly
undergoes analogue / digital conversion. In the simplest case this is
merely used for a digital display, but frequently it forms the basis for the
widest diversity of weighing processor configurations and formulation
controls. The digitised value can be forwarded over appropriate
interfaces to a PC, a PLC or a control. Analogue electrical unit signals
(0 ... lOV, 0/4 ... 20mA) are also standard output signals from such
electronics.
,--------- -----7
1 I
I I
I Power pack I
Mechanical : I
quantity Preamplifier
I
D
I
I
,--y __
I
7
I
I
I
I Signal voltage I I I
I
I I I I
I I I Final stage I
I I
I Excitation voltage I I I
I I
L I
I
L J
R compensation
,-------- -----7
1 I
I Input stage I
I Power pack I
Mechanical I Demodulator I
quantity I
I
I
I I
I I
I
I I
Si nal volta e
I
Final stage I
I
I
Excitation voltage I
I
L _J
I
J
I
L
Figures 9.7. and 9.8. below are diagrammatically enlarged so that the
error fractions normally occurring only as parts per thousand of the
output quantity can be clearly seen.
9--96 Electrical engineering structure of weighing equipment using strain gauge load cells
p Constant misalignment
of measurement data
2
z t1l
2
o
s
5
o
0-1-------+------+--,,<-------
0
Input quantity J
t
2
za
2
c
� 01---------------------1
5
0
0
Input quantity J
Linearity error
Ohmic resistance in the supply leads causes a voltage drop between the
measuring amplifier and the load cell. The load cell therefore receives a
lower voltage than that supplied by the amplifier's internal voltage
generator. On the other hand the voltage drop in the measuring leads is so
small as to be negligible, since the amplifier input is high-impedance.
Consequently it is not always possible to make use of the convincing
advantages of a six-wire circuit (=). The possible ways of combining the
connection between load cells and amplifiers are described below.
9--98 Electrical engineering structure of weighing equipment using strain gauge load cells
t
Rn.
,neause
12. 0,036
350 •
3.5 10 20 30 40 50
Load cells using four-wire technology are calibrated in the factory so that
when excitation voltage is applied and they are loaded at maximum
capacity, the nominal output signal (nominal sensitivity) is available at
the cable termination. The cable is therefore included in the load cell
calibration, Figure 9.10.
Installing load cells 9--99
----
f
I
( A 1
I Signal voltage I I D 4
I
I I I
I I I
I I B 2
I
I Excitation voltage
I I
I I C 3
I
L ______ _J
U
Amplifier
Note In order to avoid the possible errors outlined in the previous section and
to simplify load cell operation, measuring amplifiers and load cells have
for many years been produced almost exclusively in sixwire circuit form.
- - - - - - - - - -7
r-------r7-r- - - -
Measuring leads
_j-u11,�
r-----J-t--------------c=J-------t-t-----+-
I
'
I
'
Excitation voltage
Generator
I
'
R, ->0
«. I
I
n- - '
Comparator Reference
I R, -> oo voltage I
''
'
'--- J Sense leads ----1y TTFii/1/er
�----------------
_J
Figure 9-11 Load cell and measuring amplifier using six-wire
technology
----
"I 0 1
I Signal voltage I I D 4
I
I I I
I I I
I I B 2
I I
I Excitation voltage
I
I I C 3
I
Lt
u F 3
G 2'
Amplifier
-----
The feedback leads for the six-wire amplifier must be connected without
fail. This can be done directly on the amplifier in the form of a bridge, as
shown in Figure 9 .12, or if cables have to be lengthened using six-wire
9-102 Electrical engineering structure of weighing equipment using strain gauge load cells
When connecting to the amplifier, the transducer wires for the excitation
voltage and the associated sense lead (in HBM devices typically
blue-green and black-grey) should be joined together. The load cell then
takes on four-wire characteristics. In the main, the aspects covered under
Point 9.5.2 are also applicable here.
Several load cells of the same kind connected in parallel then have a
standardised sensitivity of 2mV/V at total cumulative load.
Caution In the event of uneven load distribution, an individual load cell must not
be loaded above maximum capacity. Overloading of individual load cells
cannot be detected in the output signal.
Note Further information can be found in the Operating Manual for the VKK 2
and VKE.
Electrical engineering
protection measures
The cable shielding is formed in such a way that the measuring system
itself is completely enclosed in a Faraday cage. The cable shield is fitted
smoothly onto the load cell enclosure and extends over the connector
housing to the amplifier enclosure. These measures reduce the effects of
electromagnetic interference to an absolute minimum.
GRD GRD
For effective EMC protection, Figure 9.15, the following points are
therefore important:
■ All parts of the measurement chain must be surrounded by a closed,
EMC-compliant shielding
■ All shield transitions must have an extended, closed connection
■ The shield must not at any point go inside a device or instrument
■ Signal ground, earth and shield must be completely separate
■ When there are differences of potential in the measurement system, a
potential equalisation line must be fitted. Standard value: highly
flexible stranded wire, line cross-section 10mm
■ Make sure the neutral conductor for the operating voltage is not
earthed in downstream devices
■ Strong network interference must be reduced with the aid of filters
connected on line side; circuits for measuring instruments and
energised equipment such as contactors, thyristor controls, etc.
should be as widely separated as possible
■ Never lay measurement cables alongside power cables. Minimum
separation should be 50 cm. Even with this gap it is still advisable to
ensure that measurement cables are shielded against magnetic fields,
for instance by laying them in steel conduits
10-108 Electrical engineering protection measures
Note By using HBM load cells that comply with the HBM shieiding design,
together with distribution boxes VKK2-6 or VKK1 -4, EMC-compliant
applications can be produced that extend all the way to the measuring
amplifier.
Load cell
Greenline connector
GRD
a
GRD
PE
To prevent adverse effects, the signal ground and earth or shield in HBM
devices are kept separate. If there is still a 50-Hz hum despite this, it is
advisable to ensure that devices or load cells are isolated as far as
possible and that the measurement chain is earthed at one point only.
Installing load cells 10-109
10.1.4 Summary
Group II: electrical equipment for all areas where there is a risk of
explosion, other than mine workings at risk from fire-damp
Division by temperature:
Most of the standard load cells in the product range available from HBM
W~getechnik GmbI meet the conditions for construction in Area 2 and
Area 11. A manufacturer's certificate to this effect can be requested.
In Area 1 it is only permissible to use devices that have been tested and
certified by a European testing centre. They must be constructed in
accordance with one of the harmonised Ignition Protection Systems
described in standards EN 50015 to EN 50028, which shall be deemed to
be of equal standard.
Zener barriers limit the voltage present in the circuit with the aid of
redundant Zener diodes. If the Zener voltage is exceeded in the event of a
fault, a safety fuse connected on line side blows and cuts off the circuit.
With this Ignition Protection System, the parts that could ignite an
explosive atmosphere are housed in an enclosure. This enclosure is so
solidly constructed that it can even withstand the pressure of an internal
explosion and prevent spreading it to the surrounding atmosphere. If the
enclosure is not gas-tight, any gaps that are present must be dimensioned
to standard EN 50018. They must not be altered, since in the event of an
internal explosion their surfaces are required to cool the hot gas as it
leaves. The gaps must without fail be periodically checked for deposits
or rust at regular intervals.
HBM supplies a wide range of load cells for this degree of protection
also. An advantage compared to Intrinsic Safety "i" is that there is no
reduction in accuracy, which is especially important for verifiable
applications.
Installing load cells 10-115
Thunder clouds store huge charges which generate electrical fields above
the ground and even one above the other. When this power is suddenly
discharged to ground we call it a lightning stroke. Throughout the world,
lightning strikes the earth about 100 times every second. Lightning is one
of the most powerful sources of interference. Discharges are in the
megavolt range accompanied by currents of over 100 kiloamps.
In the main there are two pathways by which lightning can affect
weighing technology systems:
■ A direct hit on the installation concerned, together with extremely
high currents accompanied by severe overvoltage
■ Strongly fluctuating magnetic fields in the neighbourhood of the
lightning stroke, caused by the great and rapidly changing strength of
the discharge current.
These changing magnetic fields can induce strong currents and voltages
in the conducting loops of any nearby weighing equipment. In
unfavourable conditions severe overvoltage can occur, even despite the
presence of built-in overvoltage protection equipment. Most of the
damage occasioned to weighing technology systems can be traced back
to this form of lightning effect.
The sort of effects that can occur range from one-off disruption to a
particular function up to and including the destruction of weighing
technology components such as strain gauge load cells, balancing
resistors and electronic switching circuits.
Calibrating weighing
devices
For the sake of clarity here is a brief explanation of the terms involved:
Setting (=)
Verification (=)
11.2 Balancing
In most case, the first step is to carry out a zero balance. Zero balancing
arises from the requirement that when the transducer (i.e. load cell) is not
under load, that is to say when it is static and unladen, the display should
also read zero. This is achieved by adjusting zero balance or using the
"Zero balance" function to bring about a parallel shift in the
characteristic curve of the amplifier.
;
2
ccU
2
U
5
0
5 0
0
Zero point
adjustment
�
Original
characteristic
curve
Input quantity J
After zero balancing the characteristic curve passes through the zero
point. The display required at maximum capacity can be set up by
changing the slope of this line.
t
Display
5000
Original
characteristic
curve
0 5kg
Input quantity J
There are three main ways in which balancing can be carried out for the
various combinations of load cells and devices:
• By simulating a signal on a load cell
• By loading one, several or all load cells with a "real" load
• By making use of special circuits and calibration signals
Not every one of these methods is possible or usable in practice for every
combination of equipment. It is advisable to balance load cells and other
Installing load cells 11-123
devices with real loads (weights, water, etc.) whenever possible. It is also
advisable to check the balancing once completed, in order be sure of
correctly installed load cells.
Sequence:
1. Zero balance
2. Introduction of a known load in/on the weighing device following
zero balance
3. Set up the measuring amplifier/weighing electronics on the
appropriate display
11-124 Calibrating weighing devices
However, if balancing can in fact be carried out using direct loading, this
represents a decisive advantage for the user: all interference effects and
errors due to the force introduction or adjustment of the load cell in
question are ascertained.
Amplifier
5kg
Example: Carry out zero balance, introduce 5 kg load, set display to 5000
Installing load cells 11--125
Practical tips:
Note This form of "field calibration" in which known loads are introduced not
only offers significant advantages over other methods, but in many cases
is actually the method prescribed.
A special case arises when calibrating silos or other containers with high
storage capacity, if insufficient calibration weights are available or the
load-carrying surface is too small for the maximum calibration load. In
cases such as these, staggered loading should be carried out:
A calibration instrument used in place of a load cell can generate and set
up defined signals. These signals simulate the strain gauge
misalignments that occur in a load cell or in several load cell circuits as a
result of a defined loading. The circuitry of a simple calibration
instrument uses the principle of the shunt resistor to generate various
whole-number bridge misalignments.
On the other hand, since there may be a difference between the zero
signal from the calibration instrument and that produced by the load cell,
the zero point of the calibration instrument must first be balanced. Then a
defined signal/ S.G. bridge misalignment is generated and the measuring
amplifier is adjusted.
At its simplest, balancing can be carried out for load cells with
standardised sensitivity (Example 1), but there are no problems
associated with arbitrary characteristic values either (Example 2). With
all load cells, nominal sensitivity has a tolerance that is dependent on the
quality of the load cell itself. If a higher level of calibration accuracy is
required, the precise sensitivity must be found from the test report for the
load cell concerned.
► Example 1:
Load cell Z6FC3 / 100 kg = 2 m V / V, calibration signal 2 m V / V
-» Adjust indicator to 100.00
► Example 2:
Load cell HLCAC3 / 550 kg= 1.94 mV/V, calibration signal 2 mV/V
- Conversion 550kg:
1.94 mV / V = X kg: 2mV / V ➔ X = 567.01 kg
-» Adjust indicator to 567.01
The following points are very closely connected with Sections 9.5.2 and
9.5.3, and are restricted mainly to load cells in four-wire technology.
Resistance in the supply lead causes a voltage drop. The extent of the
loss depends on the cable resistance which is itself a product of the
Installing load cells 11--129
► Example 3:
► Example 4:
AR, =2,50
11--130 Calibrating weighing devices
z
Ql
E
2
%%
u .£
o
c
5
(/)
cU
Ql
E
w ] (rd t
}«i)
· i [) 'lb
•
] ·I :l ! w- l
l.2'.ansducer
cable Connection box
_I
Extension cable _ _J _]
Figure 11-4 Connecting load cells in parallel
By the laws of physics, strain gauge load cells are treated as electrical
resistors (R).
► Example 5:
This is due to the fact that the "high" 350 O resistance according to
I= ub
R,
produces a small current flow and therefore a small loss. If the load cells
connected in parallel are then reconnected following calibration, the
intrinsic resistance is lower. Whereas the current strength and signal loss
increase, the display in fact operates incorrectly. As a result the indicator
is no longer in agreement because the values it displays are too low. In
such cases it is advisable to use a calibration instrument with 120 Q.
Installing load cells 11--133
The voltage drop in the supply lead gives rise to a loss, the extent of
which is dependent not only on the cable resistance/ length and
cross-section, but also on the size of the current flowing through the
cable. The strength of the current is a function of load resistance and
depends whether just one load cell is used or several.
The difference between the end value and the measuring range is
enormously important. Whether balancing can or indeed must be carried
out is dependent on which type of adjustment is possible.
If the amplifier expects two points to be input, the zero point must be set
up first followed by the end value or one other value on the characteristic
curve. However, if a measuring range can be set up on the digital display
of the measuring instrument, the set-up sequence is unimportant and
independent of the zero point. For instance the measuring range of 2
m VN can be adjusted first, and the zero point second. In this case,
changing the zero point has no effect on the either the characteristic
curve or the output signal.
Conventional analogue amplifiers do not work this way, unless the zero
signal from the load cell is suppressed for calibration purposes.
Installing load cells 11--135
11.5 Summary
Depending on the instrument and the load cell, different methods can be
used for balancing a measurement chain.
Caution Comply with the maximum permissible value for the excitation voltage as
shown on the data sheet; too high an excitation voltage leads to
increased local heating and possibly to measurement errors or damage
to the strain gauge application
• Finally switch the display adjustment so that for a load cell with say
5 kg, the value 5000 will appear in the display at maximum capacity.
When strain gauge load cells are used for the purpose of finding the mass
of an object, the determining factor is the weighing force. Since the local
acceleration due to gravity forms an intrinsic part of a measurement, very
high levels of accuracy may even require deviations from standard
gravitational acceleration to be taken into account. Weight can be
measured with very great accuracy using strain gauge load cells, the
uncertainty being expressed not as a percentage but in ppm (parts per
million) in extreme cases. For the majority of applications, though, this is
not necessary. The inaccuracy of a standard load cell is significantly less
than 1%.
12--138 Measurement accuracy and signal resolution
The selection criterion for load cells known as the accuracy class (=») is
nowadays only used for load cells incorporated into applications not
requiring verification against official specifications. On the other hand,
load cells approved for use with applications that require verification
against official specifications are increasingly being used for the most
widely differing reasons, even in equipment that does not require such
verification. Of particular importance is the expression Vmin, which
represents the smallest permissible division tick value (minimum
division tick value (=»)) expressing accuracy as a function of maximum
capacity.
The entire measurement chain, consisting of load cells together with the
technical structure such as the load introduction point and all lines in and
out, is subjected to critical testing under load, at which time any errors
present show up when the characteristic curve is displayed. A specialist
can assess the test specimen on the basis of this characteristic curve
display and take corrective measures in good time.
Load cells complete with cable and analysis equipment normally deliver
linear, reproducible measurement results, and any deviations from these
are determined by the tolerances - see HBM data sheets. If any
calibration characteristic exhibits unacceptable deviations, the causes
must be investigated and corrected. These causes almost always turn out
to be mechanical in nature. Friction, force bypasses, tilting,
self-contained stress, distortion, thermal strain and vibration are initial
disturbance variables which must on no account be regarded as constants.
The attentive reader will notice that some of the points mentioned here
have also been dealt with in previous chapters. However, in view of the
fundamental importance of the subject they will be unhesitatingly
repeated wherever it is appropriate to do so.
13--146 Interference effects on load cells
Temperature at Chemical
effects l effects
Chemical effects can impair load cells or strain gauges, perhaps due to
moisture or an attack by corrosive substances. One possible remedy is to
use hermetically-sealed load cells. Additional measures may be needed
to protect a load cell, such as applying a coat of appropriate paint.
Thermal effects that act unevenly upon a load cell can alter the
measurement result. If uneven ambient temperatures cannot be
prevented, load cells must be protected with an insulating jacket.
Gradual, even temperature changes have little or no effect on the
measurement result, since most load cells include a means of
compensating for this kind of temperature variation. In the case of load
cells calibrated in a four-wire circuit, the original cable for the load cell
must have the same temperature as the load cell itself, otherwise
measurement errors will occur. For load cells in a six-wire circuit, the
cable temperature effect is corrected by the electronics in the measuring
instrument.
Installing load cells 13-147
Figure 13.3 shows the components and the formulae for calculating
them. From the specified values it can be seen that for an angle of
5 degrees to the measurement direction, some 99.6 % of the load is still
applied, but lateral force constitutes as much as 8.7 % of this total. A
lateral force of this size applied close to maximum capacity is enough to
destroy certain types of load cell. For instance the static lateral load limit
for load cells with column-shaped spring elements can be as low as 10 %
of maximum capacity. Other types may be much less sensitive. For
example most dual-beam spring elements have a relative, static lateral
load limit of between 50 and 200 %.
13--148 Interference effects on load cells
F
f4 = F ·cos 0
Fg =F·sin o
F F
+
'
d I d
« JR
Maximum
capacity
'
} Maximum
'\/+capacity
It is very common for the effect of a falling mass on one or more load
cells to be underestimated.
A steel ball (m = 135 g) falls from height h = 30 cm onto the load button
of a load cell (maximum capacity 1000 kg).
Where:
mp =0,135kg
Ball 135g
>
Load cell 1t DCamplifier Storage oscilloscope
The example makes it impressively clear that a load cell can be very
rapidly overloaded and broken by a falling mass. In the wrong conditions
a load cell can be destroyed simply by falling to the floor. Also to avoid
the possibility of damage in transit, load cells must be suitably packed
before dispatch. For normal operation it may be useful to place a
compressible washer between the weigh platform and the load button of
the load cell. This function is carried out for example by the elastomer
bearing platforms described in Section 4.2.
The whole of the load intended for weighing must be received by defined
supports and no part of the weighing force may be diverted elsewhere by
force bypasses, since this would give rise to measurement errors.
13--152 Interference effects on load cells
Regarding the importance of this aspect for the accuracy of the weighing
system, it would be beneficial to refer again to the topic of the layout of
supply connections for containers as described in Section 7.3.
container. In this way external forces and moments that would otherwise
have an effect on the container are avoided. If it is intended to carry out
weighing whilst this equipment is operating, all points of support must be
fitted with load cells. The resulting summation compensates for the
uneven load distribution on the points of support due to fluctuating fluid
levels. A low-pass filter (=») in the weighing electronics completes the
configuration.
13.3.1 Temperature
The specifications in HBM data sheets for load cells state their nominal
and service temperature ranges (=»). Within the nominal temperature
range the individual and cumulative errors specified in the data sheet are
never exceeded. The service temperature range defines the temperature
limits between which the load cell is in principle capable of operating
without suffering damage. Certain errors may however exceed the data
sheet specification. The storage temperature range (=») is also shown in
the data sheets.
Most load cells and their associated accessories are either galvanised or
made of stainless-steel alloy, so the transducer resists these particular
weather effects.
In the case of equipment that does not permit cyclical taring as part of the
weighing process, accumulated snow or water on top of the container
will lead straight to a measurement error. Taring (=») is not possible in
Installing load cells 13-155
Lateral forces
With regard to the lateral force acting on the stay rods, the total wind
resistance is a decisive factor. It consists of frictional resistance and drag:
2
F. =c·0 '5·4
c = Drag constant (consult relevant tables)
p =Atmospheric density (kg/m?)
vo = Wind speed (m/s)
A, = Projected area
Container diameter: Sm
=»c =0.71
Lifting forces
This lateral force can also cause a container to lift off from its points of
support.
The weigh modules supplied by HBM already incorporate stay rods and
lift-off preventers that are appropriately dimensioned to accept lateral
and lifting forces. These either consist of one or two threaded rods (Cl6
module) or they are already built in (C2A module).
Installing load cells 13-157
Additional
loading on h
the load cell Lifting force
t b=h/2
Three-point support a
Four-point support
a = 0.75d,
a1 = 0.707 d,
a2 = d
► Example calculation:
Three-point support
Four-point support
r _Fis:b93kN·8,5m
A2 2 5m
_ 158,1kN
" (simplified, F~alone is
adequate)
As can easily be seen from Figure 13.7 and the calculation, the lateral
force due to wind loading causes not only a lifting force, but also an
additional loading in the measurement direction on the load cell
positioned on the side away from the wind. This additional loading is the
same size as the lifting force.
Note This additional loading is added to the tare and net load of the container.
It is therefore important to check whether the permissible limit load of the
load cell(s) will be exceeded.
13--160 Interference effects on load cells
Installing load cells 14--161
Opening remarks
This chapter explains some of the terms and expressions frequently used
in this brochure.
14--162 Terms, expressions and characteristic quantities related to load cells
■ Accuracy class
The accuracy class specified for HBM load cells means that the
largest individual error in the load cell output signal, expressed as a
percentage, is below the value specified for the accuracy class at
which the characteristic tolerance will not be taken into
consideration. Accuracy classes for load cells are divided into
Classes A, B, C and D with appropriate error limits. (= Figure
14-1)
These accuracy classes are assigned to balance accuracy classes
G GDand (
■ Adjustment
Adjustment means setting up a measurement chain (i.e. the weighing
device as a whole) in such a way that measuring deviation is kept to
a minimum. That is, the procedure permanently changes the
behaviour of the measurement chain. In the main, such adjustments
are made to the measuring amplifier or the weighing electronics.
This procedure is often referred to somewhat inaccurately as
calibration.
■ Amplifier
An amplifier (also called a measuring amplifier) is a component in
which the input quantity and output quantity are physically the same
and the value of the input quantity is increased (i.e. amplified); in
terms of the electronics it is a power amplifier which delivers a
signal corresponding to the measured quantity with higher energy.
■ Balancing
(=») see Adjustment
■ Bending stress
Bending is a change of shape (deflection or sag) in a rod or bar
(spring element) that is stressed on one side or supported at both
ends due to a loading. For a given cross-section, length and force,
the deflection is inversely proportional to the modulus of elasticity
of the material and the areal moment of inertia of the cross-section.
Bending exerts a positive strain on one side of the bent rod (spring
element), and a negative strain on the opposite side. In the case of
exactly symmetrical cross-sections the positive and negative strains
have the same value.
Installing load cells 14--163
Creep
Spring materials develop spontaneous strain in reaction to
incremental loading. Under constant loading the material slowly
continues developing strain in the direction of loading, and begins to
creep. This property is accurately described as an elastic material
after-effect for which the design of the strain gauge geometry can
compensate almost completely, so that any remaining creep error
(=») is very small.
■ Creep error
In load cells, the error resulting from creep is described by the
expression d_, for creep under load, relative to nominal sensitivity in
percentage terms over a period of 30 minutes.
The creep error is the maximum change in the output signal from the
load cell within this specified period, compared with the value that
appears when the loading changes rapidly by an amount equal to
maximum load capacity (rapidly means within 5s max.).
a Dead load D,a(=» Figure 14-2)
The dead load is the partial loading acting on a load cell in the
specified measurement direction ignoring weigh goods. The dead
load is merely the result of the dead weight, mounting accessories,
etc.
■ Dead weight
Dead weight designates the intrinsic weight of the components in a
load cell, excluding mounting accessories and so forth; it acts in the
specified measurement direction upon the measuring element of the
load cell (spring element).
Deflection at maximum load capacity S~
Deflection at maximum load capacity is the spring deflection of the
external load introduction points on load cells in the measurement
direction relative to one other, under the effects of a change of load
by the amount of the maximum capacity.
■ Degree of protection IP
The degree of protection specifies the extent to which a load cell is
protected against penetration by dust and water (tested to DIN VDE
0470-1 or EN 60529).
Installing load cells 14-165
Q)
5
6d --------------------�
e 1.5
G
1.0
~
£
2 0
E=
5.g
0.5
0.0
c >
8 -0.5
I
1.0
-1.5 ---------'-------------�
■ Linking bars
The mobile supports for preventing horizontal forces during
container weighing allow the container (weigh vessel) to move
horizontally; for this reason the container has to be restrained in the
horizontal direction. The basic aim is for the linking bars to transmit
virtually no forces to the weigh vessel in the vertical direction (force
bypasses (=»)), but to be strong enough in the horizontal direction to
absorb peak horizontal forces (such as wind forces, agitation forces).
■ Load cell characteristic curve (=» Figure 14-2)
A load cell characteristic curve is a graphical display of the
functional connection between the input quantity Load E and the
output quantity S (analogue electrical signal) on a load cell.
Meanwhile all other parameters in the system are assumed to remain
constant.
■ Load cells
Load cells are measured quantity converters which convert the load
acting upon them into an analogue or digital electrical signal. This
conversion makes use of a spring element which distorts under the
influence of the load being measured, and this distortion is then
detected and captured with the aid of strain gauges (=).
■ Load E(=» 2)
The loading of a load cell refers to the mass m that acts on this load
cell in its measurement direction as a dependent weighing force
taking into account the local acceleration due to gravity and its
associated impetus.
■ Loading, permissible dynamic
The permissible dynamic loading of a transducer is specified as the
maximum permissible vibration bandwidth of the loading in the
measurement direction. The vibration bandwidth is the difference
between the largest overload and the smallest "underload" occurring
during 107 vibration cycles without adversely affecting the
measurement capability of the load cell (to DIN 50100).
Installing load cells 14-169
■ Long-term stability
Long-term stability is a characteristic of the long-term behaviour of
a transducer (i.e. load cell), such as ageing or drift, when all other
boundary conditions and operating conditions are constant. Nor is it
possible to influence or correct the long-term stability of transducers
by means of an electronic device connected on load side. Among the
quality criteria for long-term stability are humidity resistance tests
regulated by standards; these tests include both static and dynamic
humid and hot cycles.
■ Low-pass filter
A low-pass filter is an electrical circuit that freely passes alternating
current in a particular (i.e. low) frequency range (pass band) and
blocks current at other frequencies (stop band). A low-pass filter
blocks all current with a frequency greater than the specified cut-off
frequency.
■ Maximum capacity (also known as maximum load capacity)
E, (=» Figure 14-2)
The maximum capacity represents the upper limit of the measuring
range. Depending on the load cell type, the maximum capacity may
be a tensile or compressive loading. The guaranteed error limits are
not exceeded within the measuring range. For purposes of
calibration, maximum capacity is expressed in units of mass (e.g.
kg).
a Maximum load capacity signal S, (=» Figure 14-2)
,~.
The maximum load capacity signal is the output signal which the
load cell delivers when at maximum capacity
■ Measurement direction
The manufacturer defines the measurement direction of the load cell
by specifying the axis of measurement and direction of action
relative to the design and construction of the load cell.
14-170 Terms, expressions and characteristic quantities related to load cells
■ Measuring deviation
The aim of every measurement is to determine the true value of a
measured quantity. However, every measurement result is falsified:
by imperfections in the measuring object or measuring instruments,
by environmental effects and by changes over time in relation to all
such sources of error. Measuring deviations always have to be taken
into account. A distinction is made between systematic deviations
and chance deviations.
Measuring range B(=» Figure 14-2)
The measuring range is the load range for which the measuring
deviations of a load cell are required to stay within the defined error
limits. The measuring range is specified in terms of a starting and
final value. The difference between these two values is the nominal
measuring range expressed as a measuring span B~a=Ema Limn.
Minimum dead load E,~a (=» Figure 14-2)
The minimum dead load means the lowest loading of a load cell in
the specified measurement direction (e.g. with dead load and
mounting accessories) above which the error limits are met.
w Minimum division tick value van
The minimum division tick value is the smallest division tick value
within the load cell at which the defined error limits are met.
Multi-range balance
A multi-range balance is a balance with various weighing ranges,
each starting at zero. Each weighing range has its own division tick
value with defined minimum and maximum load specifications and
in some cases different levels of accuracy.
Natural frequency fa
The natural frequency is the frequency f which characterises the
trend of a periodic natural vibration over time. It is equal to the
reciprocal of the period T of the vibration phenomenon (f = 1/T) and
is usually specified in Hertz (lHz = 1/s = one cycle of vibration per
second). The natural frequency of a load cell refers to the frequency
with which a load cell having no mounting accessories vibrates or
oscillates along the line of its axis of measurement following a short,
sharp excitation.
Installing load cells 14-171
■ Net load
The net load (in effect the net weight) is the weight of the weigh
goods less the weight of any packaging or container or means of
transportation (tare weight) with which it has been weighed.
■ Nominal range of the excitation voltage
The nominal range of the excitation voltage is the range within
which the load cell can be operated during practical application and
within which all technical specifications and error specifications
continue to be met.
Nominal sensitivity C,
In a load cell, the characteristic nominal value for sensitivity C.
Nominal temperature range B
The nominal temperature range is the range within which load cells
can be operated during practical application and within which they
continue to meet their technical specifications and error
specifications.
Non--linearity du
Non-linearity is the largest deviation of the nominal characteristic
from the specified reference curve with increasing load, by reference
to sensitivity C. The difference is normally expressed in terms of the
greatest actual amount.
• OIML
Abbreviation for Organisation Internationale de M~trologie L~gale.
This international organisation for legal metrology pursues the
standardisation of legal metrology, including verification
regulations, in individual countries. To this end it publishes
international recommendations for individual measuring instruments
(OIML Recommendations).
■ Output resistance at reference temperature []
The output resistance is the electrical resistance measured at
reference temperature at the output terminals of the load cell
including the standard electricity supply cable.
14--172 Terms, expressions and characteristic quantities related to load cells
Strain e [ m/m]
Strain e is defined as the quotient of a change in length Dl and a
reference length la. The technical term strain is used as a generic
term for both the elongation (tensile) and contraction (compressive)
process; elongation is called "positive strain" and contraction
(compression) is known as "negative strain".
■ Strain gauge (S.G.)
A strain gauge is a measuring element, usually in the form of a
looped electrical resistor in the shape of a strip; when incorporated
into a transducer it is mounted on flexible backing material. When a
force (such as a weighing force) is exerted on the flexible backing
material in contact with the S.G., the resulting distortion of the
backing material creates a proportional change in the electrical
resistance of the S.G. Within the range of elasticity, the size of the
mechanical distortion remains proportional to the applied
mechanical quantity, so that here again the change of resistance in
the S.G. is in electrically measurable proportion to the input
quantity.
■ Sustained vibration amplitude
The sustained vibration amplitude is the amplitude at which a
system that is capable of vibrating responds periodically to an
externally originating periodic excitation once the transient
phenomenon has decayed.
■ Tare weight
The tare weight refers to the mass of any packaging, container or
transportation used with the weigh goods.
■ Taring
Taring means compensating for tare weight (=») with or without
adjustment of the tare load. In colloquial language, taring is often
incorrectly used to mean "zeroing".
■ Temperature effect on sensitivity TKe
The temperature effect on sensitivity is the change in actual
sensitivity relative to nominal sensitivity following a temperature
change of 10 K. It is expressed as the maximum temperature effect
per 10 K in the nominal temperature range.
Installing load cells 14--175
1
s
Overload protection
cuts in
Sma
C ,A
S%%
B,
s%
B,ma
E = Load C = Sensitivity
s = Output signal By Payload
Ek= Minimum dead load S% Zero signal
Ema»= Maximum capacity S%%= Minimum dead load output
E Service load Sma= Maximum load capacity signal
E = Safe load limit B,= Minimum measuring range
E, = Breaking load B,ma= Nominal measuring range
D%= Dead load Dama= Maximum !CG loading
Nominal charge: 24 OM
©2000
Spectris Messtechnik GmbH, Langen
Hottinger Baldwin Messtechnik, Darmstadt
ISBN 3-9805544-4-9