You are on page 1of 8

Journal of Physics: Conference Series

PAPER • OPEN ACCESS

Detection of sodium aerosol using laser induced breakdown


spectroscopy
To cite this article: Z Alhamid and A Khumaeni 2019 J. Phys.: Conf. Ser. 1217 012020

View the article online for updates and enhancements.

This content was downloaded from IP address 36.78.248.210 on 09/09/2019 at 07:12


ISNPINSA 2018 IOP Publishing
IOP Conf. Series: Journal of Physics: Conf. Series 1217 (2019) 012020 doi:10.1088/1742-6596/1217/1/012020

Detection of sodium aerosol using laser induced breakdown


spectroscopy

Z Alhamid and A Khumaeni


1
Master Program of Physics, Department of Physics, Faculty of Science and
Mathematics, Diponegoro University, Tembalang, Semarang 50275, Indonesia
2
Department of Physics, Faculty of Science and Mathematics, Diponegoro University
Jl. Prof. Soedharto, SH, Tembalang, Semarang 50275, Indonesia
E-mail: khumaeni@fisika.undip.ac.id

Abstract. Analysis of sodium aerosol is very important in various field including nuclear
power station, electronic industry, and medical field. In medical field, aerosol detection can be
employed to earlier detection of disease based on liquid in human being. In this present work,
aerosol sodium detection has been performed using laser-induced breakdown spectroscopy
(LIBS) utilizing pulse Nd:YAG laser. In the study, sodium aerosol made of sodium chloride
(NaCl) liquid was deposited on a copper (Cu) metal plate. The Cu plate functions as a metal
subtarget to initiate a plasma emission and to improve the emission intensity of sodium line.
Experimentally, a pulse laser Nd YAG (1064 nm, 7 ns, dan 45 mJ) was focused on a sodium
aerosol to induce a luminous plasma. The plasma was then sent into the spectrometer to obtain
the emission spectrum of sodium. The observed sodium aerosol samples were made in five
variations based on their various molarities (concentrations). Variations are made, ranging
from 1 molar to 3 molar with the difference of each sample of 0.5 molar. Sharp Na lines at
588.9 nm and 589.3 nm, which overlap due to low resolution of spectrometer were clearly
detected. The other lines coming from Cu subtarget occurs at 324.7 nm and 327.4 nm. Detail
study of sodium analysis using laser induced breakdown spectroscopy will be presented.

1. Introduction
Sodium is one of the most useful ingredients in everyday life. In aerosol form, sodium is applied to
nuclear, electronic industries, and medical fields. In the medical field, aerosol sodium detection can be
used for early detection of diseases based on fluids in humans. Because there are many ingredients in
sodium aerosols found in the human body. Several methods have been used to regulate health in
humans including LIBS [2]. LIBS is a direct and fast method of analysis in detecting an element. The
popularity of this method is shown by a large number of papers published in this method in many
journals (more than 1500 papers in the last 5 years) [6-10]. The elements detected by LIBS are solids,
liquids or gases. In medical detection that is widely used is liquid. When LIBS is used to analyze
fluids, there are technical problems that interfere with the accuracy of the analysis performance.
Among them are, shock waves that accompany plasma formation cause sparks, waves, bubbles, and
aerosols. Various approaches are described in the literature that address this issue. So that aerosols are
chosen as a substitute for fluids to increase the accuracy of the target to be analyzed. To shoot a metal-
targeted aerosol is needed as a container before being shot. By depositing aerosols on the sub target
metal detection can be performed [1].

Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution
of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.
Published under licence by IOP Publishing Ltd 1
ISNPINSA 2018 IOP Publishing
IOP Conf. Series: Journal of Physics: Conf. Series 1217 (2019) 012020 doi:10.1088/1742-6596/1217/1/012020

2. Material and Methods


In this research, we used a pulse Nd: YAG laser as the radiation source. The laser wavelenght, energy,
repetition rate and pulse width were 1064 nm, 45 mJ, 10 Hz, and 7 ns respectively. Experimentally, a
pulse laser Nd YAG was focused on a sodium aerosol to induce a luminous plasma as illustrated in
Fig. 1. The plasma was then sent into the spectrometer to obtain the emission spectrum of sodium. In
the process of plasma induction of sodium aerosol there is warming caused by the focused laser
energy. Because of the heat then there is evaporation. The observed sodium aerosol samples were
made in five variations based on their various molarities (concentrations). Variations are made,
ranging from 1 molar to 3 molar with the difference of each sample of 0.5 molar.

1000
900
800
700
600
Intensity

(Na I, 501)
500 (Cu, 429)
400
300
200
100
0
300 400 500 600 700 800
Wave Length

(a) (b)

Figure 1 (a) Experimental setup used in this work and emission spectrum of ionic Ca obtained from
the human blood using the present setup. (b) shows emission spectrum of Na line obtained from the
sodium aerosol. Sharp Na lines at 588.9 nm and 589.3 nm, which overlap due to low resolution of
spectrometer were clearly detected. The other lines coming from Cu subtarget occurs at 324.7 nm and
327.4 nm.

3. Result and Discussion


A high-power narrow pulse laser beam on a solid target surface will cause the target to experience
ablation, evaporating atoms emitting out experiencing excitation and ionization and forming plasma,
which is characterized by continuous emissions as well as emission of atomic and ionic lines [4].

2
ISNPINSA 2018 IOP Publishing
IOP Conf. Series: Journal of Physics: Conf. Series 1217 (2019) 012020 doi:10.1088/1742-6596/1217/1/012020

Figure 2 Formed Plasma in high-pressure atmospheric of the order of 1 atmosphere from


target natrium and sub target Cu (cooper)

When plasma generation is carried out in a high-pressure atmospheric or gas atmosphere of the
order of 1 atmosphere, the formed plasma is spherical-shaped with a millimeter-order radius and has a
high density due to high surrounding gas pressure and emits a strong continous emission spectrum
characterized by a very white color bright with atomic and ionic emissions. Continuous emissions take
place in a very short period of time. Continus emission is caused by recombination process and inverse
bremsstrahlung. [5]

Figure 3 Formed Plasma in low-pressure atmospheric of about 1 plasma torr from target natrium
and Sub Target Cu (Cooper)

At a low pressure of about 1 plasma torr that occurs consisting of two different regions [3, 5]
namely the primary plasma area and the secondary plasma. The first area called primary plasam is a
narrow area located right in the bright white laser-target interaction area because it emits a continuous
emission spectrum in a short time. The second area is called secondary plasma located outside the
primary plasma due to the primary plasma membrane to the surrounding area. A secondary spherical
plasma that emits sharp atomic and ionic emission spectral lines. The secondary plasma radius is a
function of gas buffer pressure, wavelength, and pulse width of laser and laser pulse energy. Figure 3
showed Plasma in low-pressure atmospheric of about 1 plasma torr from target natrium and sub

3
ISNPINSA 2018 IOP Publishing
IOP Conf. Series: Journal of Physics: Conf. Series 1217 (2019) 012020 doi:10.1088/1742-6596/1217/1/012020

target Cu (Cooper). Colour green showed material cooper. Whereas for the detection of sodium is
shown by the graph of the wavelength and intensity obtained from the detection by OMA of the light
that appears. The following are the results of detection by OMA from the light that appears. For 1
molar sodium aerosols, an intensity of 501 was detected.

Figure 4 shows emission spectrum of Na line obtained from the sodium aerosol. Sharp Na lines at
1000

900

800

700

600 (Na I, 501)


Intensity

500 (Cu,
400

300

200

100

0
300 350 400 450 500 550 600 650 700 750 800
Wave Length

Figure 4 588.9 nm and 589.3 nm, which overlap due to low resolution of spectrometer were clearly
detected. The other lines coming from Cu subtarget occurs at 324.7 nm and 327.4 nm.

For sodium aerosol 1. 5 molar the intensity of 1978 was detected

5000

4500

4000

3500

3000
Intensity

2500
(Na, 1978)
(Cu, 1900)
2000
(Cu, 1532)
1500

1000

500

0
300 400 500 600 700 800
Wave Length

Figure 5 shows emission spectrum of Na line obtained from the sodium aerosol. Sharp Na lines at
588.9 nm and 589.3 nm, which overlap due to low resolution of spectrometer were clearly detected.
The other lines coming from Cu subtarget occurs at 324.7 nm and 327.4 nm.

For sodium aerosol 2 molar. The intensity of 2608 was detected

4
ISNPINSA 2018 IOP Publishing
IOP Conf. Series: Journal of Physics: Conf. Series 1217 (2019) 012020 doi:10.1088/1742-6596/1217/1/012020

5000

4500

4000

3500

3000 (Na, 2608)


Intensity

2500
(Cu, 1828) (Cu, 1954)
2000

1500

1000

500

0
250 350 450 550 650 750 850
Wave Length

Figure 6 shows emission spectrum of Na line obtained from the sodium aerosol. Sharp Na lines at
588.9 nm and 589.3 nm, which overlap due to low resolution of spectrometer were clearly detected.
The other lines coming from Cu subtarget occurs at 324.7 nm and 327.4 nm.

For sodium aerosol 2. 5 molar the intensity of 2832 was detected

5000

4500

4000

3500

3000 (Na, 2823)


Intensity

2500

2000 (Cu , 1763) (Cu, 1850)

1500

1000

500

0
250 350 450 550 650 750 850
Wave Length

Figure 7 shows emission spectrum of Na line obtained from the sodium aerosol. Sharp Na lines at
588.9 nm and 589.3 nm, which overlap due to low resolution of spectrometer were clearly detected.
The other lines coming from Cu subtarget occurs at 324.7 nm and 327.4 nm.

For sodium aerosol 3 molar the intensity of 2880 was detected

5
ISNPINSA 2018 IOP Publishing
IOP Conf. Series: Journal of Physics: Conf. Series 1217 (2019) 012020 doi:10.1088/1742-6596/1217/1/012020

5000

4500

4000

3500

3000 (Na, 2880)


Intensity

2500

2000
(Cu, 1610)
1500 (Cu, 1306)

1000

500

0
250 350 450 550 650 750 850
Wave Length

Figure 8 shows emission spectrum of Na line obtained from the sodium aerosol. Sharp Na lines at
588.9 nm and 589.3 nm, which overlap due to low resolution of spectrometer were clearly detected.
The other lines coming from Cu subtarget occurs at 324.7 nm and 327.4 nm.

From the detection results of sodium aerosols obtained can be made the ratio as follows
1,9

1,8

1,7

1,6

1,5
Ratio

1,4

1,3

1,2

1,1

1
1 1,5 2 2,5 3 3,5
Molarity

Figure 9 shows ratio from emission spectrum of Na line obtained from the sodium aerosol.

From the ratio graph shown can be concluded the greater the molarity of sodium aerosols, the
likelihood of the intensity detection results will be even greater.

4. Conclusion
Detection of sodium aerosol has been successfully performed using LIBS method with laser Nd:
YAG. This detection is indicated by the detected wavelength and the emission color emitted. Thus, the
LIBS method can be used to detect aerosol-shaped elements by using Cu-shaped solids as sub-targets.

Acknowledgment
This study was financially supported by Ministry of Research and Technology and Higher Education,
Indonesia under the project of INSINAS RISET PRATAMA scheme 2018.

6
ISNPINSA 2018 IOP Publishing
IOP Conf. Series: Journal of Physics: Conf. Series 1217 (2019) 012020 doi:10.1088/1742-6596/1217/1/012020

References
[1] Khumaeni A, Kurihara K, Lie Z S, Kagawa K, Lee Y I. 2014. Current Applied Physics. 14 451
[2] St-Onge L, Kwong E, Sabsabi M, Vadas E B. 2004. Journal of Pharmaceutical and Biomedical
Analysis. 36 277
[3] K. Kagawa, M. Ohtani, S. Yokoi and S. Nakajima. 1984. Spectrochim. Acta. 39 525-536
[4] W. S, W. T. Baskoro, M. Pardede, H. Kurniawan. 1999. Appl. Spectrosc. 53 (11) 1347-1351
[5] Kagawa, K and S. Yokoi. 1982. Acta B 37 789-795
[6] D. A. Rusak, B. C. Castle, B. W. Smith, J. D. Winefordner. 1998. trends in analyt. Chemist 458-
461
[7] D. A. Cremers, L. J. Radziemski. 2006. Handbook of laser-induced breakdown spectroscopy, John
Wiley & Sons Ltd, New York
[8] W. Miziolek, V. Palleschi, I. Schechter: Laser-induced Breakdown Spectroscopy (LIBS):
Fundamentals and Applications (Cambridge University Press, New York, 2006) p. 17
[9] N. Omenetto. 2004. J. Anal At. Spectrochim. Acta. 59 (B) 147-161
[10] J. M. Vadillo, J. 2004. J. Laserna, Spectrochim. Acta. 59 (B) 147-161

You might also like