You are on page 1of 21

113-

Blood gas control during cardiopulmonary bypass


DT Pearson Freeman Hospital, Newcastle-upon-Tyne

Introduction Why control blood gases during


hypothermic CPB?
The functional replacement of the heart and lungs
by blood pump and artificial oxygenator during
a The maintenance of adequate perfusion to the
open heart surgery is now commonplace. To in- brain and other organs is essential in all patients
crease the tolerance of the heart to ischaemia supported by a pump oxygenator. Reduced cere-
during the surgery, systemic hypothermia is em- bral perfusion may result from systemic arterial
ployed ; augmenting cardioplegia and topical hypotension,~ though the level and degree of au-
myocardial cooling. toregulation of cerebral blood flow (CBF), in the
The conduct of cardiopulmonary bypass (CPB) absence of haemodynamically significant
is not only concerned with the preservation of atherosclerotic lesions, is dependent upon acid-
organ viability but also with maintenance of organ base status during hypothermia. The use of a
function. An important safety feature of hypo- pH-stat regime uncouples cerebral flow and
thermic CPB concerns the following important metabolism with resultant pressure passive CBF
considerations related to blood gas control: and loss of autoregulation below a mean arterial
pressure (MAP) of 55 mmHg.2,3 Alpha-stat
1) Why control blood gases during hypothermic management maintains autoregulation of CBF at
CPB?
a lower MAP with a more physiological relation-
2) Alpha-stat and pH-stat acid base management ship between CBF and cerebral metabolic rate. 4,5
during hypothermic CPB. High arterial P02 (Pa02) is vasoconstricting, nar-
3) Significance of the term ’temperature correc- rowing the CBF autoregulatory plateau and low
tion’ as it applies to blood gas measurement
Pa02 shifts the lower limit of autoregulation to
during hypothermic CPB. a higher level.6 6 Whilst the optimal levels of arterial
4) Measurement technique and frequency of esti- PC02 (PaC02), blood flow and MAP during
mation of blood gases.
Do current hypothermic cardiopulmonary bypass are yet to
commercially available artificial be
5) established,’ the potential exists that failure
oxygenators allow the perfusionist to control of autoregulation of CBF when Pa02, PaC02 and
blood gases to prescribed levels?
levels are abnormal may precipitate cerebral
6) Which bubble oxygenator design features in- pH damage.
fluence blood gas control?
The level of Pa02 directly influences the gaseous
microemboli emission from bubble
(GME)
Address for correspondence: Dr DT Pearson, Regional The life
Cardiothoracic Unit, Freeman Hospital, Freeman Road, High oxygenators.8~9 span and stability of these
Heaton, Newcastle-upon-Tyne NE7 7DN, UK. 100% 02 bubbles present in the arterial blood is

Downloaded from prf.sagepub.com at UCSF LIBRARY & CKM on April 20, 2015
114

inversely related to the partial pressure Pa02 blood total C02 content constant, maintaining
gradient between the bubble and blood.10 temperature uncorrected PaC02 at 5.3 kPa. 21
There is improved myocardial electrical stability
in patients undergoing hypothermic CPB with
alpha-stat management as compared with pH-stat of the term ’temperature
Significance
management. 11,12 correction’ it applies to blood gas
as
Systemic arterial microembolism has been measurement during hypothermic CPB
suggested as an important cause of cerebral
damage.’3,’4 It has been reported in the literature Blood gas analyser: the estimation of blood gas
that oxygenators: (a) release gaseous microem- values from samples taken from the cardio-
boli8,14,15; (b) damage formed blood elementsl6-18; pulmonary bypass circuit during hypothermia in-
I

and (c) do not allow the perfusionist to maintain volves the use of blood gas analysers (BGA) in
physiological levels of Pa02 and PaC02 during which the blood sample is heated to 37°C. The
cardiopulmonary bypass,9,19,20 all of which can values obtained may be ’temperature corrected’
theoretically produce multi-organ damage. using the BGA in-built formulae to allow for the
The current balance of clinical opinion favours difference in temperature between the patient and
alpha-stat rather than pH-stat acid-base manage- the analyser’s electrodes.
ment to optimize organ function during hypo-
pH-stat: Values corrected for actual body
thermic CPB.5,7,21 The need to control PaC02,
temperature
Pa02 and pH values during hypothermic CPB is Alpha-stat: Values not corrected for actual
therefore mandatory, based on clinical and
body temperature
physiological evidence relating to organ function.
In-line electrodes: blood gas values measured
at patient temperature during hypothermic CPB.
Alpha-stat and pH-stat acid-base The use of the term ’temperature correction’ in
management during hypothermic CPB this context is different to that applied to blood
gas values when BGAs are used involving the use
The principles associated with the use of the terms of in-built formulae to correct the values to 37°C.
alpha-stat and pH-stat in relation to acid/base pH-stat: Values not corrected to 37°C
status, are effected by alteration in C02 solubility Alpha-stat: Values corrected to 37°C
during hypothermia.
During hypothermic CPB, the perfusionist at-
Henry’s Law: Gases in solution tempts to maintain PaC02 5.3 kPa and pH 7.40.
Content =
partial pressure x solubility It may be necessary to apply ’temperature correc-
During hypothermia C02 solubility [up] tion’ of the blood gas values dependent upon the
use of either alpha-stat or pH-stat acid-base
pH-stat:
Content [up] =
partial pressure [normal] x management techniques and measurement by
solubility [up] BGA or inline electrodes. Representative values
pH [normal] . for PaC02 and pH during hypothermic CPB at
25°C are shown.
Alpha-stat:
Content [normal] =
Partial pressure [down] x
solubility [up] Blood gas machine
pH [up]
During hypothermic CPB, acid-base management Measured at 37°C
may be directed towards maintenance of tempera- pH-stat pH 7.22 PaC02 8.3 kPa
ture corrected arterial PC02 (PaC02) at 5.3 kPa Alpha-stat: pH 7.40 PaC02 5.3 kPa
and pH 7.4 (pH-stat). To compensate for the
increased solubility of C02 during hypothermia,
blood total C02 content must be raised by addition Corrected to 25°C
of C02 and/or hypoventilation. The alternative pH-stat pH 7.40 PaC02 5.3 kPa
acid-base management (alpha-stat) is to keep Alpha-stat: pH 7.57 PaC02 3. 3 kPa

Downloaded from prf.sagepub.com at UCSF LIBRARY & CKM on April 20, 2015
115

In-line electrodes Do current commercially available


Measured at 25°C artificial oxygenators allow the
pH-stat: pH 7.40 PaC02 5.3 kPa perfusionist to control blood gases to
Alpha-stat:
.

pH 7.57 PaC02 3.3 kPa prescribed levels?

Corrected to 37°C Introduction


pH-stat: pH 7.22 PaC02 8.3 kPa Currently available bubble oxygenators effect
Alpha-stat: pH 7.40 PaC02 5.3 kPa gas transfer by dispersing gas (100% 02) into the
venous blood via small holes in a gas sparger.
Gas exchange occurs in the resultant foam due
to the intimate mixing of blood and oxygen (direct
Measurement technique and frequency of blood gas interface). Once this has been achieved,
estimation of blood gases coalescence of the foam and removal of remaining
bubbles is accomplished by the use of a combin-
Significantly better control of Pa02 can be ation of the action of surface active substances in
achieved using data from an inline electrode rather the oxygenator defoamer, settling and filtration.
than intermittent blood gas sample values.9 The Total gas flow, size of bubbles generated in the
use of data from an inline P02 electrode22,z3 to gas sparger, turbulence and gas transfer occurring
control Pa02 levels has many advantages over over the heat exchange area and in the defoamer
conventional intermittent blood sampling. The are all critical factors in bubble oxygenator per-
time constants of changes in Pa02 during formance. Small bubbles produce excellent oxy-
hypothermic cardiopulmonary bypass do not genation but are inefficient at C02 removal and
permit intermittent sampling to determine Pa02 difficult to remove by the defoamer. Large bubbles
adequately and hence allow adjustment of the are essential for C02 removal and easily removed

gas/blood flow ratios needed to maintain by the defoamer, but less efficient as far as oxy-
physiological levels of Pa02. Adequate manage- genation is concerned. There always exists a com-
ment of C02 clearance during CPB can be promise between oxygenation and C02 elimin-
achieved by estimation of PaC02 at 15-minute ation in bubble oxygenators (Galletti and Brecher,
intervals. During a clinical evaluation of the gas 1966) .
transfer characteristics of 10 Bentley BIO-10 The use of low gas/blood flow ratios and
bubble oxygenators, inline Pa02 values were cor- ’efficient’ oxygenation may result in inadequate
related with BGA PaC02 values from blood sam- C02 clearance. Typically during hypothermic car-
ples taken at five-minute intervals. The scatter- diopulmonary bypass, as metabolic rate and there-
grams of the five-minute and 15-minute data are fore 02 demand falls, the perfusionist controls
shown in Figure 1. There was no significant differ- the Pa02 by reducing the gas to blood flow ratio.
ence between the group Pa02 and PaC02 values However, if this results in inadequate C02 clear-
when the five-minute and 15-minute data were ance, he or she is faced with the dilemma of
compared. raising the gas to blood flow ratio producing high
A high degree of temperature-dependent error Pa02 and normal PaC02 values or accepting
with underestimation of P02 can result from use normal Pa02 values with C02 retention. Control
of blood gas analyser results.24-26 Correlation of of Pa02 to physiological levels may be associated
the Pa02 values obtained from an inline P02 with C02 retention as shown in Figure 3.9 Whilst
electrode (Cardiomet 1000: Biomedical Sensors) this may be acceptable during hypothermic per-
and samples analysed by BGA (Instrumentation fusion using a pH-stat technique of blood gas
Laboratories IL1302) is shown in Figure 2. The management, the need to maintain C02 clearance
temperature-compensated BGA tended to under- using an alpha-stat technique can present prob-
estimate the Pa02 (mean ± sd difference between lems for the perfusionist.
pairs of values 7.0 ± 8.6). The independent control of Pa02 and PaC02
and the ability to adjust the flow rate and 02
concentration of the ventilating gas in membrane

Downloaded from prf.sagepub.com at UCSF LIBRARY & CKM on April 20, 2015
116

Figure 1 Scattergrams of blood gas data for Pa02 and PaC02 in Bentley BIO-1 0 bubble oxygenator group: (a) 15-minute

samples (Pa02 34.0 -t 5.1 kPa; PaC02 5.2 -L 0.9 kPa; n 193); (b) 5-minute samples (Pa02 30.3 -t 14.9 kPa; Pa02 5.2 ±
=
1.1
kPa; n =
67) mean ± SD. Shaded area indicates blood gas target range

Downloaded from prf.sagepub.com at UCSF LIBRARY & CKM on April 20, 2015
117

Figure 2 Comparison of Pao2 measurement techniques: (a) scattergram of Pa02 data from IL1302 blood gas analyser and
Cardiomet 1000 in-line electrode with line of identity (correlation coefficient 0.79); (b) scattergram of difference between IL1302
and Cardiomet 1000 data

Downloaded from prf.sagepub.com at UCSF LIBRARY & CKM on April 20, 2015
118

Figure 3 Scattergrams of blood gas data for Pa02 and PaC02 in Harvey H1500 bubble oxygenator groups: (a) intermittent P02
data (Pa02 30.9 ± 7.1 kPa; PaC02 5.9 ± 0.7 kPa); (b) continuous POz data (PaO~ 14.0 ± 2.0 kPa; PaC02 6.3 ± 0.7 kPa) mean ±
SD. Shaded area indicates blood gas target range

Downloaded from prf.sagepub.com at UCSF LIBRARY & CKM on April 20, 2015
119

oxygenators (which do not have a direct blood Table 2 Clinical profile of oxygenator groups
gas interface) enables the perfusionist to manage
blood gas values more easily.2’ The use of nitrogen
to dilute the 02 concentration in the ventilating
gas of bubble oxygenators has been proposed as
a method of controlling the PaOb 28 but it cannot
be recommended for clinical use.29-31

Clinical oxygenator evaluation


During clinical hypothermic cardiopulmonary
bypass (CPB) in adult patients, the ability of the
perfusionist to control arterial P02 and PC02 has
been evaluated in: (a) five groups of bubble
oxygenators (Polystan venotherm, Harvey H1700,
Bentley BIO-10, Gambro 10 and Shiley S100A
HED); (b) two groups of membrane oxygenators
(Harvey HF4000 and Cobe CML) using a pH-stat
blood gas management technique; and (c) three
groups of bubble oxygenators (Bentley-10B,
Bentley-l0Plus and Harvey H1700) using alpha- was mounted on a Cobe Stockert Pump Console.
stat blood gas management technique (Table 1). No prebypass or arterial line filters were included
’Ten oxygenators from each manufacturer were in the circuit. The oxygenator and tubing circuit
used in each of the 10 study groups. were flushed with 100% C02 prior to priming
with Plasmalyte 148 (Travenol Laboratories).
Table 1 Blood gas control: oxygenators evaluated
Perfusion technique: to conform with current
clinical practice, the blood flow rate (QB) was
maintained at 2.4 l.min-1.m2 body surface area
during cardiopulmonary bypass and reduced by
25% during hypothermia commensurate with ac-
ceptable oxygenator reservoir levels and surgical
technique and preference. Gas was delivered to
the oxygenator from back pressure compensated
anaesthetic rotameters.
In the bubble oxygenators (pH-stat), the flow
rates of 100% 02, with or without the addition
Patients: the study included 100 patients under- of small amounts of 100% C02, were manually
going open-heart surgery for acquired valvular or adjusted in an attempt to maintain predetermined
ischaemic heart disease randomly assigned to one target levels of Pa02 and PaC02 (Table 4). No
of the 10 oxygenator groups according to the C02 was added during the rewarming phase of
oxygenator used for CPB. The clinical profile of CPB. In the bubble oxygenators (alpha-stat), only
the patients allocated to each oxygenator group 100% 02 was used, without the addition of any
is shown in Tables 2 and 3. All patients were C02. Two important restraints were placed upon
anaesthetized by the same anaesthetist (DTP), the perfusionist controlling the blood gases:
using a modified form of neuroleptanalgesia sup- a) Total gas flow rate was not adjusted below
plemented by nitrous oxide and muscle relaxants. 0.51 per minute even if the Pa02 was above
Cardiopulmonary bypass: oxygenator evalu- the target range.
ation was undertaken during hypothermic CPB
(24-28°C) with the addition of cold cardioplegia. b) Inadequate C02 clearance with associated
elevation of PaC02 was corrected by increas-
The bypass circuit, consisting of the oxygenator
to be evaluated, cardiotomy reservoir (Harvey ing gas flow rate and took precedence over
maintenance of target Pa02 levels.
H705F or H4700) and customized tubing circuit,

Downloaded from prf.sagepub.com at UCSF LIBRARY & CKM on April 20, 2015
120

Table 3 Characteristics of oxygenator groups (patient and bypass data)

Table 4 Target blood gases


hypothermia. Pa02 was controlled by alteration
of the setting of the gas controller valve on the
top of the oxygenator.
In the membrane oxygenators (pH-stat), Pa02
and PaC02 levels were adjusted by alteration of
Continuous Pao2 measurement using Cardiomet 1000 the percentage 02 and C02 in the gas to the
Intermittent PaC02 measurement using IL 1302 device; a total gas flow rate of 10 l.min-1 (Cobe
CML) or 4.0 l.min-1 (Harvey HF4000) was main-
The Bentley-lOB is a typical bubble oxygenator tained throughout the period of bypass.
in which all of the gas (100% 02) is dispersed Gaseous microemboli (GME): GME were
into small bubbles at the sparger to mix with the monitored in the arterial line leading from the
incoming venous blood and effect gas transfer. oxygenator using a TM8 Microbubble Activity
This oxygenator has been modified in the Bentley- Monitor. Signals from the TM8 were computer
lOPlus to incorporate an integral gas proportioning processed to generate GME.1-1 during the CPB. 32
valve (gas controller). When this proportioning From this data an oxygenator group mean ± SD
valve is set at maximum, all of the gas entering GME level was calculated.
the device is directed to the sparger. As the valve Blood gas analysis: in all oxygenator groups, a
is progressively adjusted from maximum to the Cardiomet 1000 inline P02 electrode (Biomedical
minimum setting, an increasingly greater propor- Sensors) was used to provide continuous data for
tion of the total gas flow is directed, via the blood temperature and to control the Pa02-
original quick prime port, to the heat exchange Arterial blood samples were analysed for PaC02
area of the oxygenator (Figure 5). At minimum using an IL1302 blood gas analyser. PaC02 and
setting of the valve, 10% of the total gas flow Pa02 values were blood temperature uncorrected
passes through the sparger and 90% is directed in the alpha-stat oxygenator groups (five-minute
to the heat exchange area. The technique of con- samples) and blood temperature corrected in the
trol of blood gases in the Bentley-l0Plus pH-stat oxygenator groups (15-minute samples).
oxygenator (alpha-stat) differed from the other PaC02 values were corrected with the inline Pa02
bubble oxygenators. A gas (Qa) to blood (QB) values at the time of blood sampling.
flow ratio of 1:1 with a QB of 2.4 l.min-l.m-2 Blood samples: whole blood samples (25ml)
body surface area was used at the start of CPB. were withdrawn for haematology studies before
The initial QG was maintained throughout the and after CPB. Values for plasma haemoglobin
bypass even when QB was reduced during and platelet count were corrected for haemo-

Downloaded from prf.sagepub.com at UCSF LIBRARY & CKM on April 20, 2015
121

Figure 4 Schematic diagram of Bentley 10PIus showing effect of alteration of gas flow
pattern engendered by manipulation of the gas controller

dilution. Nopatient was taking antiplatelet drugs Bubble oxygenators (pH-stat): the number of
(aspirin dipyridamole) prior to surgery and no
or blood gas values in each group is similar, but the
donor blood was transfused before or during the distribution of blood gas values and the percentage
time of blood sampling. Anticoagulation was of values when both PaC02 and Pa02 were within
maintained using heparin; the accelerated clotting the target range is markedly different for some
time being monitored using a Haemochron and of the oxygenator groups (Figure 5 and Table 5).
maintained >400 seconds throughout bypass. Only 3% of the values obtained in the Bentley
BIO-10 group fell within the defined normal
Results range, compared with 17% for the Gambro 10
Scattergrams of the blood gas data for all 10 group, 20% for the Shiley S-100A HED group,
oxygenators in each evaluation group are dis- 31% for the Polystan venotherm group and 36%
played in Figures 5, 6 and 7, together with the for the Harvey H1700 group. The Bentley BIO-10
target range for PaC02 and Pa02. The oxygenator bubble oxygenator group mean Pa02 values were
group blood gas values (mean ± SD) and the significantly higher (p<0.05) than the other
percentage of values when both PaC02 and Pa02 oxygenators evaluated, though mean Pa02 values
are in the target range are shown in Table 5. In in all groups were above the normal range (Table
all oxygenator groups, it was possible to control 5).
the PaC02 to within the target range, confirming When the bubble oxygenator groups are
the ability of the perfusionist to adhere to the compared, the Polystan venotherm releases sig-
prescribed protocol of giving priority to C02 nificantly less GME than the other oxygenators
clearance. The degree of Pa02 control is (p<0.02) and the Shiley S100A HED releases
demonstrated by the spread of data points in the significantly greater numbers of GME than the
scattergrams and percentage values in the target other oxygenators (p<0.02) except for the
range. Gambro 10. No significant differences could be

Downloaded from prf.sagepub.com at UCSF LIBRARY & CKM on April 20, 2015
122

Harvey H1700

Downloaded from prf.sagepub.com at UCSF LIBRARY & CKM on April 20, 2015
123

Bentley BIO-10

Gambro 10

Downloaded from prf.sagepub.com at UCSF LIBRARY & CKM on April 20, 2015
124

Figure 5 Scattergrams of blood gas data for Pa02 and PaC02 in five bubble oxygenator groups using a pH-stat acid base
management technique: (a) Polystan venotherm; (b) Harvey H1700; (c) Bentley BIO-1 0; (d) Gambro 10; (e) Shiley S1 OOA HED.
Shaded area indicates blood gas target range

demonstrated between the Bentley BIO-10, pore size of the bubble disperser (using data
Harvey H1700 and Gambro 10 (Table 6, Figure 8). provided by the manufacturers). The results are
In the bubble oxygenator groups (Table 6), shown in Figure 10. A statistically significant in-
there were statistically significant differences verse relationship has been established (p<0.05).

between the gas/blood flow ratios used by the The larger bubble disperser pore size in the
perfusionist, within the confines of the evaluation Polystan Venotherm necessitated the use of com-
protocol, to control the PaC02 and Pa02 within paratively higher gas/blood flow ratios to attain
the target range. Correlation of Pa02, GME levels adequate gas exchange with significantly lower
and gas/blood flow ratios in the bubble oxygenator GME production. The lower gas/blood flow ratio
groups (Figures 8 and 9, Tables 5 and 6) indicates used in the Shiley S100A HED oxygenator group
that the use of low gas/blood flow ratios is not was associated with a small pore size and sig-

necessarily associated with low GME levels and nificantly higher GME levels.
inadequate oxygenation as evidenced by the Bubble oxygenators (alpha-stat): the Bentley-
Bentley BIO-10 data. The Bentley BIO-10, lOPlus allowed significantly better control of Pao2
Gambro 10 and Shiley S100A used significantly (p<0.05) and a smaller scatter of values than the
lower gas/blood flow ratios than the two other Bentley-lOB. Both oxygenator group PaC02
bubble oxygenators evaluated, but had high GME values were within the target range. The gas
levels in the arterial blood. The lowest levels of controller valve permitted a considerable degree
GME, from the Polystan venotherm bubble of independent PaC02 and Pa02 control in the
oxygenator group, were associated with a high Bentley-10Plus, though there was no significant
gas/blood flow ratio. difference in GME emission between the two
The mean GME levels in the five bubble oxygenators. The Bentley-10 series of oxygenators
oxygenator groups have been correlated with the offers an improvement in blood gas control and

Downloaded from prf.sagepub.com at UCSF LIBRARY & CKM on April 20, 2015
125

Figure 6 Scattergrams of blood gas data for Pa02 and PaC02 in two membrane oxygenator groups using a pH-stat acid base
management technique: (a) Cobe CML; (b) Harvey HF 4000. Shaded area indicates blood gas target range

Downloaded from prf.sagepub.com at UCSF LIBRARY & CKM on April 20, 2015
126

Bentley 10PLUS

Downloaded from prf.sagepub.com at UCSF LIBRARY & CKM on April 20, 2015
127

Figure 7 Scattergrams of blood gas data for Pa02 and PaC02 in three bubble oxygenator groups using an alpha-stat acid base
management technique: (a) Bentley 10B; (b) Bentley 10PIus; (c) Harvey H1700. Shaded area indicates blood gas target range

Table 5 Blood gas values for 10 oxygenator groups

Downloaded from prf.sagepub.com at UCSF LIBRARY & CKM on April 20, 2015
128

GME over Bentley BIO-10 (Figure 7). compares favourably with the Harvey H1700 (pH-
The Harvey H1700 bubble oxygenator was stat) and Bentley-10Plus.
evaluated using an alpha-stat acid-base manage- Membrane oxygenators (pH-stat): in both the
ment technique (Figure 7). Although only 15% membrane oxygenators evaluated, the in-
of values for Pa02 and PaC02 were within the dependence of Pa02 and PaC02 control allowed
target range, the scatter of values is small and a high percentage of blood gas values to fall in
the target area with minimal scatter. GME
Table 6 Gaseous microemboli and gas/blood flow ratio in emission was minimal and significantly lower than
10 oxygenator groups in any of the bubble oxygenators evaluated.
Haematology: although all oxygenator groups
showed a rise in plasma haemoglobin when the
pre- and post-CPB values were compared, no
individual value exceeded 120 mg.dl-1 (Table 7).
Over the period of CPB, there was a marked
difference in the mean percentage reduction in
platelet count in the different oxygenator groups
(Table 7). The post-CPB platelet count was sig-
nificantly lower than the pre-CPB (p<0.05) in all
oxygenator groups except the Cobe CML and
Gambro 10. The platelet count fall could not be
correlated with gas/blood flow ratio or GME
emission in the bubble oxygenator groups. The
use of the gas controller in the Bentley-10Plus

oxygenator group did not significantly reduce


platelet depletion as compared with the Bentley-
10B.

Figure 8 Oxygenator group gaseous microemboli for 10 oxygenators (mean ± SD)

Downloaded from prf.sagepub.com at UCSF LIBRARY & CKM on April 20, 2015
129

Figure 9 Oxygenator group gas/blood flow ratio for ten oxygenators (mean ± SD)

Table 7 Oxygenator group plasma haemoglobin (mg.dl-’) and platelet count (x 109.1-1): haematocrit corrected

Downloaded from prf.sagepub.com at UCSF LIBRARY & CKM on April 20, 2015
130

size (correlation
Figure 10 Comparison of gaseous microemboli production of bubble oxygenators and bubble disperser pore
coefficient -0.91 ): (a) oxygenator group gaseous microemboli (mean ± SD) in five bubble oxygenators (pH-stat acid-base
management); (b) bubble disperser pore size

Downloaded from prf.sagepub.com at UCSF LIBRARY & CKM on April 20, 2015
131

Discussion added to the oxygenator gas supply in bubble


Choice of oxygenator for clinical CPB must oxygenators.
take into account the gas transfer characteristics Diversity of design characteristics in the bubble
of the device and how these relate to the acid-base oxygenators evaluated may account for the differ-
management technique used during hypothermia. ences in gas/blood flow ratios used during CPB.
Clinically significant differences have been It was noted that the devices achieving adequate
demonstrated in the 10 oxygenators studied. gas exchange at the lower gas/blood flow ratios
When compared with the Bentley-lOB, the (Bentley BIP-10, Gambro 10 and Shiley S100A
Bentley-10Plus is a significantly improved design, HED) generated relatively high gaseous micro-
though no reduction in GME or improvement in emboli (GME) levels in the arterial blood. The
platelet depletion have been associated with its size of bubbles generated in the bubble disperser
use in the current study. It is possible that by of bubble oxygenators is a function of the pore
adjustment of the total QG in conjunction with size of the gas disperser.35 The inverse correlation
use of the gas controller, a further improvement of mean GME levels in the five bubble oxygenator
in control of blood gas values could have been groups (pH-stat) with the pore size of the bubble
attained in the Bentley-10PIus oxygenator. disperser supports this concept. The larger bubble
The bubble oxygenator using the highest gas/ disperser pore size in the Polystan Venotherm
blood flow ratio in the pH-stat study (Harvey necessitated the use of comparatively higher gas/
H1700) was evaluated to assess its performance blood flow ratios to attain adequate gas exchange
when an alpha-stat regime was used. This with significantly lower GME production; theor-
oxygenator allowed superior control of blood etically consistent with the production of larger,
gases in the bubble oxygenator (pH-stat) evalu- ’more easily defoamed’ bubbles in the oxygenation
ation. Whilst there was interdependence of PaC02 column of the device and relatively ’less efficient’
and Pa02, satisfactory control of blood gases, gas transfer. The lower gas/blood flow ratio used
comparable to the Bentley-10Plus, was achieved in the Shiley S100A HED oxygenator group was
with no increase in GME or platelet depletion. associated with a small pore size and significantly
In membrane oxygenators there is considerable higher GME levels.
flexibility in the control of blood gases. The in- During clinical perfusion with any individual
dependent control of Pa02 and PaC02 and the bubble oxygenator, there is a direct relationship
ability to adjust the flow rate and 02 and C02 between gas/blood flow rate, arterial P02 and
concentration of the ventilating gas enables the GME levels in the arterial blood. GME levels are
perfusionist more easily to manage blood gas also inversely related to arterial reservoir level.
values using either a pH-stat or alpha-stat regime. The efficiency of defoaming in bubble oxygenators
is related to many factors, including design, size
and shape of the defoamer, flow pattern through
the defoamer and dynamic holdup (resident time).
Which bubble oxyenator design features It is directly comparable to those established
influence blood gas control? factors influencing the defoaming capacity of
cardiotomy reservoirs.36
Improvement in ’efficiency’ of oxygenation by the The Bentley-10Plus bubble oxygenator allows
use of low gas/blood flow ratios in bubble the level of Pa02 to be adjusted with the gas
oxygenators may result from an increase in gas/ controller valve. Maintenance of a fixed gas flow
blood surface area interface consequent upon of sufficient magnitude to ensure adequate C02
generation of smaller oxygen bubbles in the oxy- elimination can be presented from inducing over-
genation column of the device.33,34 However, the oxygenation and high Pa02 values during
defoamer is not as efficient at removing smaller hypothermic CPB by manipulation of the gas
bubbles which remain after gas transfer and controller valve. This valve proportions the gas
systemic arterial gaseous microembolism may away from the oxygenation efficient gas sparger
ensue. The lower gas/blood flow ratios used, whilst to the less efficient heat exchange area. Gas
allowing adequate 02 transfer, may also result in diverted to the heat exchange area of the device
inadequate C02 clearance. C02 should not be no longer contributes significantly to further

Downloaded from prf.sagepub.com at UCSF LIBRARY & CKM on April 20, 2015
132

oxygenation, but is able to continue to effect C02 is gratefully acknowledged. The technical assist-
elimination. A unique degree of independence of ance of the perfusionists in the Cardiothoracic
control of Pa02 and PaC02 has been confirmed Unit at Freeman Hospital, Newcastle-upon-Tyne,
using the oxygenator in the clinical environment. is appreciated.

Summary
References
The need to control PaC02, Pa02 and pH values
during hypothermic CPB is mandatory, based on 1 Stockard JJ, Bickford RG, Myers RR, Aung
clinical and physiological evidence relating to MH, Dilley RB, Schauble JF. Hypotension
organ function. Control of blood gases contributes induced changes in cerebral function during
to the overall safety of CPB. cardiac surgery. Stroke 1974; 5: 730-46.
Since the current balance of clinical opinion 2 Henriksen L, Hjelms E, Lindeburgh T. Brain
favours alpha-stat rather than pH-stat acid-base hyperperfusion during cardiac operations. J
management during hypothermic cardio- Thorac Cardiovasc Surg 1983; 86: 202-208.
pulmonary bypass, a comprehensive understand- 3 Lundar T, Lindegaard KF, Froysaker T, Aasid
ing of the principles involved, associated blood R, Grip A, Nornes H. Dissociation between
gas measurement techniques and implications of cerebral autoregulation and carbon dioxide
temperature correction are essential. reactivity during nonpulsatile
Adequate and accurate control of Pa02 during cardiopulmonary bypass. Ann Thorac Surg
CPB necessitates the use of inline P02 electrodes. 1985; 40: 582-87.
Commercially available bubble oxygenators do 4 Govier AV, Reves JG, McKay RD
et al.
not allow the perfusionist adequately to control Factors and their influence on regional
blood gases. Control of PaC02 and Pa02 are cerebral blood flow during nonpulsatile
interdependent. cardiopulmonary bypass. Ann Thorac Surg
The of bubble oxygenators results in an
use 1984; 38: 592-600.
overall to hyperoxygenation and an in-
tendency 5 Murkin JM. CBF, CMRO
2 and PaCO 2 during
ability to clear C02. This tendency can cause CPB. The influence of acid-base management
on cerebral blood flow and metabolism during
problems in blood gas control when alpha-stat
blood gas management techniques are used during cardiopulmonary bypass. Proc Amer Acad
hypothermic CPB. Cardiovasc Perf 1986; 7: 121-23.
Bubble oxygenators using relatively high gas/ 6 Henriksen L. Brain luxury perfusion during
blood flow ratios allow better control of blood cardiopulmonary bypass in humans. A study
gases and lower GME emission. of the cerebral blood flow response to changes
The Bentley-10Plus bubble oxygenator offers a in CO
, O
2 2 and blood pressure. J Cerebral
degree of independent control of Pa02 and Blood Flow and Metabolism 1986; 6: 366-78
PaC02- 7 Gravlee GP. Cerebral blood flow during
The useof membrane oxygenators enables cardiopulmonary bypass. Proc Amer Acad
superior control of blood gases with independent Cardiovasc Perf 1986; 7: 126-30.
control of Pa02 and PaC02 and GME free per- 8 Pearson DT, Holden MP, Waterhouse PS,
fusion to be achieved by the perfusionist. Clark JI. Gaseous miocroemboli during open
heart surgery: detection and prevention. Proc
Acknowledgements AmerAcad Cardiovasc Perf 1983; 4:103-109.
A grant from the Department of Health and 9 Pearson DT, Holden MP, Poslad SJ, Murray
Social Security enabled this work to be under- A, Waterhouse PS. A clinical evaluation of
taken. Dr A Murray and Mr R Clayton, Regional the gas transfer characteristics and gaseous
Department of Medical Physics, Freeman Hos- microemboli production of two bubble
pital, Newcastle-upon-Tyne, assisted in the pro- oxygenators. Life Support Systems 1984; 2:
duction of the data. The provision of the 252-66.
oxygenators for evaluation by the manufacturers 10 Butler BD. Biophysical aspects of gas bubbles

Downloaded from prf.sagepub.com at UCSF LIBRARY & CKM on April 20, 2015
133

in blood. Med Instrument 1985; 18: 59-62. a reusable oxygen electrode. Anaesthesia 1983;
11 Swain JA, White FN, Peters RM. The effect 39: 362-69.
of pH on the hypothermic ventricular 24 Reeder GD, Hood AG. Accuracy of oxygen
fibrillation threshold. J Thorac Cardiovasc partial pressure measurements: an in vitro
Surg 1984; 87: 445-51. study. J Extracorp Tech 1983; 15: 89-95.
12 Kroncke GM, Nichols RD, Mendenhall JT, 25 Hill AG, Vinasky RP, Todd RW, Groom RC,
Myerowitz PD, Starling JR. Ectothermic Lefrak EA. Continuous measurement of
philosophy of acid-base balance to prevent oxygen tension during cardiopulmonary
fibrillation during hypothermia. Arch Surg bypass. Proc Amer Acad Cardiovasc Perf 1984;
1986; 121: 303-304. 5: 39-42.
13 Brennan RW, Patterson RH, Kessler J. 26 Poslad SJ, Pearson DT, Murray A. Clinical
Cerebral blood flow and metabolism during comparison of one in vitro and three inline
cardiopulmonary bypass: evidence of blood gas PO
2 sensors during cardiopulmonary
microembolic encephalopathy. Neurology bypass. Clin Phys Physiol Meas 1987; 8: 249-
1971; 21: 665-66. 59.
14 Slogoff S, Girgis KZ, Keats AS. Etiologic 27 Pearson DT, Holden MP, Poslad SJ, Murray
factors in neuropsychiatric complications A, Waterhouse PS. A clinical evaluation of
associated with cardiopulmonary bypass. the performance characteristics of one
Anaesth Analg 1982; 61: 903-11. membrane and five bubble oxygenators: gas
15 Gallager EG, Pearson DT. Ultrasonic transfer and gaseous microemboli production.
identification of sources of gaseous Perfusion 1986; 1: 15-26.
microemboli during open heart surgery. 28 Riley JB, Hurdle MB, Winn BA, Jallad MS.
Thorax 1973; 28: 295-305. Use of nitrogen in ventilation of artificial blood
16 van den Dungen JJAM, Karliczek GF, oxygenators. J Extracorp Tech 1984;
16: 45-46.
Brenken U, Wildevuur CRH. Clinical study 29 Kurusz M, Butler BD. Problems relating to
of blood trauma during perfusion with gaseous microemboli and the use of nitrogen
membrane and bubble oxygenators. J Thorac with bubble oxygenators. J Extracorp Tech
Cardiovasc Surg 1982; 83: 108-16. 1984; 16: 114-15.
17 Wright G, Sanderson JM. Cellular aggregation 30 Yost G. The bubble oxygenator as a source
and destruction during blood circulation and of gaseous microemboli. Med Instrument 1985;
oxygenation. Thorax 1976; 31: 405-409. 19: 67-69.
18 van Oeveren W, Kazatchkine MD, Descamps- 31 Bastianen GW. The use of air with bubble
Latscha B et al. Deleterious effects of oxygenators and its influence on the formation
cardiopulmonary bypass: prospective study of of microgas emboli. J NeSECC 1984; 9: 25-27.
bubble versus membrane oxygenation. J 32 Pearson DT, Poslad SJ, Murray A, Clayton
Thorac Cardiovasc Surg 1985; 89: 888-99. R. Extracorporeal circulation material
19 de Wall R. What constitutes adequate evaluation: microemboli. Life Support Systems
perfusion? J Cardiovasc Surg 161975 : ; 458-64. 1987; 5: 53-67.
20 Newland PE, Pastoriza-Pinol JV, McMillan J, 33 Ferguson TB, Burbank A, Burford TH. The
Stirling GR. A report on the use of five bubble disposable bubble oxygenator. Evaluation of
an improved model. Surgery 1967; 61: 260-64.
oxygenators for cardiopulmonary bypass
10: 45-49.
surgery. Anaesth Intens Care 1982; 34 Hammond GL, Bowley WW. Bubble
21 Swain JA. Acid-base, hypothermia and cardiac mechanics and oxygen transfer. J Thorac
surgery. Perfusion 1986; 1: 231-38. Cardiovasc Surg 1976; 71: 422-28.
22 Parker D, Delpy DT, Halsall DN. A new 35 Galletti PM, Brecher GA. Heart lung bypass:
approach to inline gas monitoring: principles. New York: Grune and Stratton,
development of an oxygen sensor. Medical and 1966.
Biological Engineering and Computing 1983; 36 Pearson DT, Carter RF, Hammo MB,
21: 134-37. Waterhouse PS. Gaseous microemboli during
23 Claremont DJ, Pagdin TM, Walton N. open-heart surgery. In: Longmore DB ed.,
Continuous monitoring of blood PO 2 in Towards safer cardiac surgery, Lancaster: MTP
extracorporeal circuits: an
in vitro analysis of Press, 1981: 325-54.

Downloaded from prf.sagepub.com at UCSF LIBRARY & CKM on April 20, 2015

You might also like