Professional Documents
Culture Documents
Hospira Plum A Plus 43095226001 Rev05 01 Muen
Hospira Plum A Plus 43095226001 Rev05 01 Muen
Operating
Manuaf
For use with List 11971-04
Abbott Laboratories
;g;h Chicago, IL 60064
‘.
SECTION 1
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.1 P RODUCT O VERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...*............................................................... l-l
1.2 C ONVENTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . l-2
Warnings, Cautions, and Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . l-2
1.3 D EFINITIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ....... l-3
SECTION 2
DEVICE DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . 2-l
2.1 EATLJRES ................................................................................................................................... 2-l
Therapies ............................................................................................................................... 2-l
Line Programming Options .................................................................................................. 2-l
PlumSet Capabilities .............................................................................................................. 2-l
Air Management ................................................................................................................... 2-l
Battery ................................................................................................................................... 2-2
Biomedical ............................................................................................................................ 2-2
Options .................................................................................................................................. 2-2
Other Features ....................................................................................................................... 2-2
2.2 U SER Q UALIFICATION ................................................................................................................ 2-3
2.3 SYSTEM SELF-TE STS .................................................................................................................. 2-3
SECTION 3
INSTRUCTIONS FOR USE ........................................................................................................................ 3- 1
3.1 SETUP ......................................................................................................................................... 3-1
Unpacking ............................................................................................................................. 3-l
Tandem Carrier Instructions ................................................................................................. 3-2
3.2 A DMINISTRATION S ETS .............................................................................................................. 3-2
Preparing the Administration Set .......................................................................................... 3-3
Priming the Administration Set ............................................................................................ 3-4
Loading the Cassette ............................................................................................................. 3-5
Preparing the Secondary Line ............................................................................................... 3-6
3.3 D ISCONTINUING E LECTRONIC F LOW C ONTROL & SE?TING G RAVITY F LOW ........................... .3-8
3.4 D ISCONTINUING F LUID ADMINISTRATION ................................................................................. 3-8
...
111 430-95226-001 (Rev. 901)
Contents
SECTION 4
BASIC OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-l
4.1 GETTING STARTED ...................................................................................................................... 4-l
System Self-Tests ................................................................................................................... 4-l
Data Retention ................................................................................ ..‘~ .................................. 4-3
4.2 OPERATING KEYS ........................... ............................................................................................ 4-3
4.3 INDICATORS ................................................................................................................................ 4-5
4.4 REAR CASECONTROLS .............................................................................................................. 4-5
4.5 POWER ON .................................................................................................................................. 4-7
4.6 SIMPLE DELIVERY ....................................................................................................................... 4-8
4.7 EXAMPLES OF AUTOMATIC CALCULATION ................................................................................ 4-9
4.8 RTFL4TION ................................................................................................................................ 4-10
4.9 SIMPLE DELIVERY USING D RUG L IST ..................................................................................... .4-12
4.10 SIMPLE D ELIVERY USING D ELAYED START ............................................................................ 4-14
4.11 PIGGYBACK D ELIVERY ........................................................................................................... 4-15
4.12 P IGGYBACK WITH NURSE C ALLBACK .................................................................................... 4-17
4.13 CONCURRENT DELIVERY ....................................................................................................... .4-18
4.14 S ELECT O PTION - VOLUMES INFUSED , PRESSURE / POST INFUSION R ATE , AND L IGHTING /C• ....... .4-20
4.15 S TOP AND S TART WITH ONLY 1 L INE P UMPING ..................................................................... 4-23
4.16 S TOP AND S TART WITH B OTH L INES P UMPING ...................................................................... 4-23
4.17 BACKPRIMING ......................................................................................................................... 4-25
4.18 C LEARING P ROGRAM SETI-INGS ............................................................................................. 4-25
4.19 POSSIBLE NON -DELIVERY PROGRAMMED ............................................................................. 4-26
SECTION 5
THERAPIES (FOR ADVANCED USERS*) ............................................................................................... 5-l
5.1 D OSE C ALC (MCG/KC~MIN ON A) ................................................................................................ 5-l
5.2 D OSE C ALC (MG/HR OS B) ........................... .............................................................................. 5-5
5.3 L OADING D OSE .......................................................................................................................... 5-9
5.4 MULTISTEP PROG~WX~MING .................................................................................................... 5-15
SECTION 6
ALARMS AND TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6. -l
6.1 W ARNING ME.SSAGE~ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6. -l
6.2 R ESPONSE TO ALARMS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
General Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2 .
Line A Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Line B Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4
Malfunctions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
SECTION 7
MAINTENANCE AND STORAGE ............................................................................................................ 7-l
7.1 C LEANING , M AINTENANCE , AND STORAGE ................................................................................ 7-l
Cleaning and Sanitizing ......................................................................................................... 7-l
Battery Maintenance .............................................................................................................. 7-2
Storage .................................................................................................................................. 7-2
Service .................................................................................................................................. 7-2
SECTION 8
PRECAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ................... 8-l
8.1 ARTIFACTS ................................................................................................................................. 8-1
8.2 HEALTHCARE PROFESSIONALS AND P ATIENT R ELATED ............................................................ 8-1
8.3 C ONCURRENT F LOW ................................................................................................................... 8-2
8.4 E PIDURAL A DMINISTRATION ...................................................................................................... 8-3
8.5 BATTERY O PERATION ................................................................................................................ 8-4
8.6 S ETS A N D ACCESSORIES ............................................................................................................. 8-4
8.7 BACKPRIMING ............................................................................................................................. 8-5
8.8 G ENERAL .................................................................................................................................... 8-5
SECTION 9
SPECIFICATIONS ....................................................................................................................................... 9-l
PHYSICAL .......................................................................................................................................... 9-l
ELECTRICAL ...................................................................................................................................... 9-l
VTBI RANGE ..................................................................................................................................... 9-l
ENVIRONMENT ................................................................................................................................... 9-2
D ELIVERY RATE R ANGE ................................................................................................................... 9-2
AIR-IN-L INE ALARM ......................................................................................................................... 9-2
OCCLUSION ALARM AND LIMITS ...................................................................................................... 9-3
T IME TO D ETECT DOWNSTREAM O CCLUSIONS ................................................................................. 9-3
BOLUS V OLUME R ELEASED AFTER D OWNSTREAM O CCLUSIONS ARE C ORRECTED ......................... 9-3
D ELIVERY ACCURACY ....................................................................................................................... 9-4
~~LJM~ET CURVES ............................................................................................................................. 9-4
SECTION 10
SUPPLIES AND ACCESSORIES ............................................................................................................. 10-l
10.1 ADMINISTRATION F~um ........................................................................................................ 10-l
Administration Fluids ......................................................................................................... 10-l
Enteral and High Viscosity Fluids ...................................................................................... 10-l
Containers ............................................................................................................................ 10-l
10.2 ACCESSORIES ........................................................................................................................... 10-l
1 .l Product Overview
The Plum A+TM Volumetric Infusion System is designed to meet the fluid delivery requirements of today’s
evolving healthcare environments. The Plum A+ is a cassette based, two lines in and one line out, multi-
function infusion pump. The pump can be used for standard, piggyback. 05 concurrent delivery. Therapy
modes include:
. Standard Infusions
l Multistep Programming
l Loading Dose
l Dose Calculation
The Plum A+ is designed to deliver parenteral and enteral or epidural infusions over a broad range of
infusion rates from multiple fluid container types.
The system includes a pumping module (hereafter called the pump) and an assortment of disposable IV sets
(hereafter called a set), optional accessories, and this operator’s manual.
The optional Barcode Reader allows drug name identification by scanning a provided list of drug names
located on the Flash Tips & Barcode Directory.
The Plum A+ is designed to be used in most areas of patient care, including, but not limited to:
1.2 Conventions
This section describes the conventions used throughout this manual, as follows:
Alert messages used throughout this manual are described below. Pay particular attention to these
messages.
WARNING
A WARNING MESSAGE CONTAINS SPECIAL SAFETY EMPHASIS AND MUST BE OBSERVED AT ALL
TIMES. FAILURE TO OBSERVE A WARNING MESSAGE IS POTENTIALLY LIFE THREATENING.
CAUTION: A CAUTION USUALLY APPEARS IN FRONT OF A PROCEDURE OR STATEMENT. IT CONTAINS
INF-ORMATION TRAT COULD PRBVENT IRREVERSIBLE PRODUCT DAMAGE OR RARDWARE FAILURE. FAXLXJRE
TO OBSERVE A CAUTION COULD RESULT M SERIOUS PATIENT OR USER INJURY.
a 1.
This symbol directs the user to consult accompanying documents.
Note: Figures are rendered as graphic representations to approximate the actual product. Therefore,
figures may not exactly reflect the product.
Alarms Any condition of the pump that requires operator.attentfon. These condf-
tions will cause both audible and visible alarm indicators. Confirmation of
the condition is usuallu reauired.
Backprime Removal of air orfluid from the proximal line and cassette air trap. Nofluid
is delivered distal to the cassette during backpriming.
Bolus A single uninterrupted discrete volume of jlufd delivered in a discrete
period of time.
Cassette A recognized cassette type installed with the door closed.
Installed
Concurrent Simultaneous delivery of twofluids at fndependentjlow rates.
Delivery
Delfvery Delivery option which allows lines A and B to deliver concurrently.
Together
Delivery Rate The Delivery rate set by the operator.
Duration The time period used to deliver a VTBI at a delivery rate set by the operator.
Error A condition resulting when an operator attempts a nonacceptable input,
such as an invalid key press.
Flow Rate The resulting rate of fluidflow.
KVO (Keep Vein Delivery of fluid at a slow rate in or&r to prevent occluding the patient
Open) catheter.
Line A Cassette Inlet line on the top center of the cassette when looking at the cas-
sette from the front of the Device.
Line B Cassette Inlet line on the top right of the cassette when looking at the cas-
sette from the front of the Device.
Malfunction Failure of the device to pass internal integrity or performance tests.
Pfgg ybac k The mode where line B delivers first, followed by line A.
PressurelPost Selected using A [OptfonslVol Inflt this mode is used to set KVO Rate and Dfs-
Infusion Rate tal Pressure Limit. It is also used to view Distal Pressure in real-time.
Units A predetermined delivery quantity adopted as a standard of measure.
VTBI Volume To Be Infused - The amount offlufd remaining to be delivered to the
vatfent.
Warning An indication to advise the operator:
a) Of a possibly hazardous condition such as a Low-Battery warning.
b) That an attempt has been made to use a function in the wrong sequence,
with the wrong values, or at the wrong time such as an Invalid Key Warn-
ing.
c) That a function has been initiated in the proper manner such as a Valid
Keu Warnfnq.
2.1 Features
The Plum A+ is a dual-line volumetric infusion system designed to meet the growing demand for hospital-
wide, as well as alternate site and home healthcare. standardization. With its primary line, secondary line,
and piggyback fluid delivery capability, the Plum A+ is suited for a wide range of medical/surgical and
critical care applications. Full compatibility with LifeCare Plum Series administration sets and accessories
and the LifeShield@ needleless protection systems makes the Plum A+ a convenient and cost-effective
infusion system.
THERAPIES
l Dose Calculation
l Loading Dose
l Multistep Programming
. Concurrent delivery
l Titration
l Micro 0.1-99.9 mL/hr (in 0.1 mL increments) flow rate range for both lines
l Macro loo-999 mL/hr (in 1 mL increments) flow rate range for both lines
l Automated Secondary drug delivery (Piggyback)
PLUMSET CAPABILITIES
l Anti Free-Flow Protection
AIR MANAGEMENT
l Air Trap
l Air Removal/Backpriming
l Air Detection-Proximal
l Air Detection-Distal
BATTERY
l Battery Gauge
l Six Hour Battery Recharge Time
l Long battery life (6 hours) for emergency backup and tempprary portable operation
BIOMEDICAL
l Serial Communication
l Upgradability (Field)
OPTIONS
l Volumes Infused (A, B, Total Volume)
l KVO at dose end (1.0 ml/hr or less depending on delivery rate) or Continue Rate (CR) to
continue at the current rate
l Variable Distal Pressure Settings
OTHER FEATURES
Nonpulsatile volumetric accuracy
Microprocessor control
Large liquid crystal display (LCD) screen
Panel back illumination on mains power
Lockout switch
Standard fullfill, partfill, syringe and vial use
Parenteral and nonparenteral (enteral) fluid delivery
Blood and blood products delivery
Wide range of Standard and Specialty administration sets
WARNING
ADMINISTER ONLY ANESTHETICS/ANALGESICS APPROVED FOR EPIDURAL ADMINISTRATION
(AS INDICATED OR ALLOWED BY THE DRUGS’ FDA APPROVEiD LABELING). EPIDURAL
ADMINISTRATION OF DRUGS OTHER THAN THOSE INDICATED FOR EPIDURAL USE COULD
RESULT IN SERIOUS INJURY TO THE PATIENT.
‘.
This section explains the Plum A+ setup and administration set use.
3.1 Setup
WARNING
ARRANGE TUBING, CORDS, AND CABLES TO MINIMIZE THE RISK OF PATIENT STRANGULATION
OR ENTANGLEMENT.
UNPACKING
CAUTION: PR ODUCT D A MAGE MAY OCCUR UNLESS PROPER CARE IS EXBRCISED DURING UNPACXING
AND INSTALLATION. Do NOT USE TRE PLJJM A+ IF IT APPEARS DAMAGED IN ANY WAY. TRE BATTERY MAY
NOT BE CRARGED UPON RECEIPT.
Inspect the Plum A+ packaging for possible shipping damage. If damage is found, contact the delivery
company immediately.
Use care when unpacking the Plum A+. Retain the packing slip and save all packing material in case the
Plum A+ is damaged or fails the pump self-test and has to be returned to Abbott Laboratories.
To set up the Plum A+, connect the mains cord to a properly earthed mains receptacle, unless temporary
battery operation is desired.
Note: Use mains power whenever possible. Store the Plum A+ connected to mains to confirm a fully
charged battery for emergencies.
Note: If the quality of earth grounding source is in doubt, use battery power.
CAUTION: THE PLUM A+ SYSTEM IS DESIGNED TO OPERATE NORMALLY JN TRB PRESENCE OF MOST
ENCOUNTERED ELECTROMAGNETIC INTERFERENCE (EMI) CONDITIONS. m TRE EVENT OF EXTREME
LEVELS OF INTERFERENCE, SUCH AS ENCOUNTERED NEXT To AN ELECTROSURGICAL GENERATOR.
CELLULAR TELEPHONES, OR 2-WAY RADIOS, IT IS POSSIBLE THAT THE NORMAL OPERATION OF A SENSOR
OR MICROCOMPUTER MIGHT BE D I SR UPT ED . O PE RA TION OF THE PUMP UND ER THESE CON D ITIONS SROULD
BE AVOIDED.
The Plum A+ may be safely and conveniently mounted on an IV stand or on a tandem carrier mounted on an
IV stand.
Always mount the tandem carrier (List# 12276-O 1) to the pole before
attaching the pump(s) to the tandem carrier.
=Ei
Pnrnaty Line-
87
IK!B Finger Grip
Flow Regulator
4
Pumping Chamber
-~
430-95226-001 (Rev. 5/01) 3- 2 Plum A+ Infusion Pump
Administration Sets
Use the aseptic technique to prepare the administration set for priming, th$n proceed as follows:
Note: To avoid getting Distal Air Alarms, confirm that the fluid bag is positioned higher than the
pump. This should be set up prior to priming the cassette.
To load the primed cassette into the Plum A+, proceed as follows:
k.
Open the cassette door by lifting up the handle.
.T,.
Insert cassette into door guides. Close the cassette door. Confirm
there is no flow and no kinks appear in tubing.
Note: If flow is observed, close the tubing clamps and replace the administration set.
Note: The administration set should be changed per CDC guidelines or healthcare provider policy.
Discard after use.
CAUTION: C ONS UL T THE DRUG LABELING TO CO N F IR M DRUG COEXPATIBILITP, CONCE NTRA TION ,
DELIVERY RATES, AND VOLUMES ARE ALL SUITABLE FOR CONCURRENT DELIVERY OR PIGGYBACK DELIVERY
(SECONDARY FOLLOWED AUTOMATICALLY BY PRIMARY) MODES. ,
The Plum A+ features concurrent or piggyback delivery modes when therapy requires administering more
than one fluid.
In addition to standard containers, the Plum A+ can use syringes or vials on the secondary port. The
secondary line can be prepared without removing or repriming the cassette.
To prepare the secondary line, use aseptic technique and proceed as follows:
With cassette door closed: loosen and remove white cap, then
discard. Add secondary tubing or syringe.
Vial or Syringe: Secure the container to the cassette door using the
optional container support arm (List 12095-04-01).
E
Vial: Attach the vial adapter to the secondary port. Backprime the air from the vial
adapter into the vial, if necessary.
If electronic flow needs to be stopped, the precision flow regulator on the cassette can be used to manually
set and maintain fluid flow in a gravity mode (see the following graphics).
@ PRESS STOP
@ CLOSE DOOR
xkIrlr
Softkeys Label Region j
I Softkeys
AA
- Numeric Keys
Start and Stop Keys - moom UP/LEFT
e Selection Key
Charging/Line
Indicator DOWN/RIGHT
k
ON/OFF Key ___) + Silence Key
SYSTEM SELF-TESTS
CAUTION: Do NOT P L ACE TRE PLUM A+ IN SER VI CE IF IT FAILS THE S YSTEM S ELF -TES TS .
Connect the mains cord to mains power, then confirm that the Charging/Line indicator illuminates (above
the [ON/OFF] key on the front panel).
A systematic self-testing of the processing, delivery, and safeiy systems is performed whenever the Plum A+
is turned on, to verify readiness for operation.
Note: If the quality of earth grounding source is in doubt, use battery power.
Press the [ON/OFF] key to turn the power on. Check screen display and listen for a beep to indicate the
audio is working. Wait for the self-tests to complete. If successful, put a fully primed macro cassette into the
cassette door and close the door. See L.oaciing Cassette in Section 3 for details of correct procedure.
When the Charging/Line indicator is off and the battery symbol on the display is flashing, this indicates
that the Plum A+ is operating on low battery power and should be recharged.
CAUTION: Do NOT OPERATE THE PLUM A+ ON PATIENTS WITH THE -BATTERY REMOVED. USE OF A
q
A PROPERLY MAINTAINBD AND CHARGED BATTERY HELPS CONFIRM PROPER OPERATION.
To confirm battery is fully charged, connect the Plum A+ to AC power for a minimum of six hours while in
the OFF mode.
If an alarm occurs during the power on self-test, identify the alarm message, then take corrective action
(see Section 6, Alarms and Troubleshooting).
Start pump. If the self-test alarm recurs, remove the Plum A+ from service and contact the local Abbott
Laboratories representative.
DATA RETENTION
Delivery program settings and programming option selections are retained in non-volatile memory. If the
Plum A+ has been turned off for longer than four hours, all delivery settings are cleared and programming
option selections are restored to their default selections for next use.
4.3 Indicators
The Charge/Line Indicator is illuminated to indicate
CHARGE the battery is currently charging.
4.5 Power On
1. Press [ON/OFF] to turn on the Plum A+ and you will see
the System Self-Tests screen. c
2. To set up the Plum A+, insert selected Plum Set and close
cassette holder door lever.
L
A
4.8 Titration
r 1. Select Line A [A] (or A [BJ at the Main Delivery Screen.
I
I A I
R
A ,AlAlAlA]
0.0 Jp,& 0.0 Note: To select a Drug Name using a Therapy (i.e., Dose Calc) you
must first select which line you want (A or B), next select A [Therapy],
0.0 ““,:“’ 0.0 and then A [Program Options]. If you select a Drug Name first, it will
be cleared upon entry in Therapy mode.
Duration 0O:OO hrzmin 3. Use [SELECT] to highlight the VTBl field and enter a
value. The Duration will be automatically calculated.
VTBI WmL
Duration 50:00 hrmin
mt%r
Note: Delayed line drip indicator is solid, not flashing. It will begin
5.0 O-O flashing when actually pumping.
O-O Vd’nf
mL
0.0
L
‘1
A
Note: Upon entry to this screen, and to all lower screens from here,
the Barcode Reader (BCR) is enabled as indicated by “Wand Active”
when attached to the Plum A+. To use the BCR, swipe tip of BCRwand
across entire barcode, from dot to dot, on the Abbott supplied drug
list card. The drug name also can be selected using A IPrwrcun
Options].
4. To enter delivery rate, use numeric keypad.
Mode Piggyback Note: While the VTBI is being entered, the duration is automatically
computed. If the computed value is greater than 99:59, dashes (- -)
appear in place of hours. To understand the Automatic Calculation
feature, see Section 4.7 Examples of Automatic Calculation,
Duration 00:30 hcmin
7. Use [SELECT] key to move down to the DURATION field.
Mode Piggyback When all parameters are displayed, changing any one
Rate 133 mL/hr parameter will result in other parameters automatically
ml 100 mL being computed.
Duration OOa hnmin
CAUTION: V E RI FY RATE. VTBI. AND DURATION BEFO RE
PRESSING [START].
1
Program Options B The default screen for Program Options appears.
wuld MIU
2. Use [SELECT] to highlight “Callback” field and press A
Delay Start for 0O:oO htzmin
[Yes/No], if necessary to obtain “Yes”, then press A [Enter].
CallbaCk I
I
m
\rIAlAlAj
S EE S ECTION 8, FOR INF O R M A TION ON C ONC URRENT
SECTION 9 FOR DELJVER Y RATE RANGES).
Note:
FLOW (SEE
Note: Upon entry to this screen, and to alI lower screens from here,
the Barcode Reader (BCR) is enabled as indicated by “Wand Active”
when attached to the Plum A+. To use the BCR. swipe tip of BCR wand
across entire barcode, from dot to dot, on the Abbott supplied drug
list card. The drug name also can be selected using A fProgram
OptionsJ
Verify change to “Concurrent” mode.
-
Mode B 3. Use [SELECT] to highlight Rate field.
Rate 0.0 mUhr
Vi-B1 0.0 mL
Duration 00: 00 hrrmin
LineA 393 mL
4. Use [SELECT] to scroll to Line A or B.
tine B 0.0 mL
Total Vol m mL
100 ,“,a:;, 0.0 Note: If A [A] (or A [BJ is pressed prior to lSTOP], the line is stopped,
the display will show the progr amming screen. Press A [CanceYBack]
8.5 v;:“’ 0.0 to return to the Main Delivery screen.
A B Options
Vol Ini
<A A A A,
Example of STOPPED delivery.
4. Open door, remove set and close door. Or, press [START]
to resume infusion.
Note: If A [A] (or A [BJJ is pressed prior to [STOP], the display will
show the programming screen. Press A [Cancel/Back] to return to the
Main Delivery screen.
3.0 “y 20.4
4.17 Backpriming
A) STOPPED 1 STOPPED 1 B BACKPRIMING Removal of air or fluid from the proximal line and
cassette airtrap.
5.5 :;tr 0.0
1. To initiate a backprime, press A [Back Prime].
8.9 “Z:” 0.0
Note: Backpriming is available only when all delivery is stopped.
Secondary container is necessary to receive the air.
Bsck A
prime R%z
AA A A,
2. Press and hold A [Back Prime] until fluid pumped from
line A to line B has expelled the air from cassette air trap
into Secondary container.
4. Select A [Yes].
Loading Dose- Allows programming of an initial infusion rate for a specific volume,
followed automatically by a maintenance rate from the same container
(e.g., a fluid challenge). If Dose Calc is used, the Loading Dose and
Maintenance Dose are in the same unit of measure, over the same period
of time (mcg/min), from the same container.
Multistep- Allows a sequential program to deliver up to 10 steps; fluid volumes and
delivery rates may be programmed for each step. The program may be
entered based on Rate and Volume or Volume and Time. If Dose Calc is
used, the delivery steps are in the same unit of measure, over the same
period of time, from the same container.
IAI Proaram Dose Calc I The Program Dose Calc screen ap’pears.
Note: Screen defaults to the mcg/kg/min option or the previously
selected option.
1Al Program Dose Calc The Container Concentration Selection screen appears
whenever the selected Dose Units are gram-based (e.g.,
Drug Cone in Container
grams, mg, mcg, or ng) otherwise this screen is skipped.
Note: Default choice for possible gram-based Container
Concentration Unit is mg or the previously selected option.
Weight -kg 3. Use keypad to enter desired values. Once each value is
o.cra mr)*pmrkl
entered, use [SELECT] to highlight the next programming
‘ield.
Note: The Concentration elements and weight value must be
entered before the (SELECT] key allows access to remaining fields.
A ( Program Dose Calc 9. Continue to enter values using the keypad and advancing
- wvKIAdi4#
to the next field using [SELECT].
400 mg 250 mL
70.0 kg Note: As the Dose value is entered, the Rate is automatically
m 0.0 lr#g!nlin ml_ calculated. To understand the Autom’atic Calculation feature, see
mm kc*
7.9 ml.&, Section 4.7 Examples of Automatic Calculation,
rA I Program Dose Calc 10. Use [SELECT] to scroll down to the VTBI field.
- -MN0
E
I
A I Program Dose Calc 11. Enter a value for VTBI using the keypad.
- WMd-
Note: As the VTBI value is entered or changes, Duration is
ccnc mmg 250 rnL
automatically updated.
Weight 70.0 kg
KY
0watkel
Note: The A [Program Options] path to setting Delayed Start or
bs assignin a drug name from the internal list functions exactly as
describe 2 in Basic Operation for the Simple Delivery in Section 4.
12. After entering the desired VTBI value and any desired A
[Program Options],,press [START] to begin infusion.
O”
Vol
mL
Inf 0.0 Clear ALL settings when changing patients or multiple
programming parameters.
2. Press A [Therapy].
Program Dose Calc IB 10. Continue entering desired values using the keypad. Use
v&nd pdive
[SELECT] to advance to the next field.
cone 200 w fo0 mL
Mode Concurrent
Note: When entering the RATE, with-a VTBI value entered, the
DOSE and DURATION fields are automatically updated. To
understand the Automatic Calculation feature, see Section 4.7
Examples of Automatic Calculation,
Note: The Duration field temporarily than es to dashes when the
computed value is larger than what can be /?isplayed.
Program Dose Calc IB CAUTION: VERIFY DOSE, RATE, VTBI. AND DURATION BEFORE
wAc6u [START].
PRESSING
PUMPING 1 PUMPING Upon [START], the Main Delivery Screen shows delivery
status for both A and B lines.
1.0 n&e
mLI hr
Note: See Pelivery Rate Range in Section 9. The ratio of concurrent
5.0 Vol Inf rates has recommendations specified in Section 8.3.
mL
Program Dose Calc IB 14. To update and identify the infusing drug, enter the drug
Wmd_ -
name by using either the BCR or via A [Program Options]
then A [Drug List] with an entry for a selection.
4 1 PUMPING 1 PUMPING 1 E CAUTION: VERIFY DOSE, BATE, VTBI, AND DURATION BEFORE
PRESSING [START].
D%o c%c%z
1.0 Rate 0.5
mL1 hr 15. Press [START] to confirm entries and return to the Main
5.0 Vd Inl 0.2 Delivery screen. I.
mL
Dose s,
2. Press A [Therapy].
2.k -
mL
“pgp
1 mm 0.0 0o.m
2 0.0 0.0 WE00
A ( Program Loading Dose 9. Enter Volume To Be Infused (VTBI) from 0.1 to 9999 mL
Wsnd A&w
using the keypad.
Rate vrsl “;FmT Note: After a Rate and VTBI value are entered, the Duration is
mUhr
1 MO & w:c6 automatically computed. Manually changing the Duration field will
2 0.0 0.0 lYJ:W
automatically recompute the Rate. To understand the Automatic
Calculation feature, see Section 4.7 Exaniples of Automatic
Calculation,
10. After the values for the loading dose Step-l are entered,
use [SELECT] to highlight the maintenance dose Step-2.
Note: The loading dose therapy has two steps. The first is the
loading dose followed by the maintenance dose. These words will
appear on the Main Screen as each step is delivered.
6
I
A I Program Loading Dose II. Enter the Rate using the keypad.
WUdAC4iV8
Al Program Loadzdse 1 12. Use [SELECT] to highlight the next programming field.
Rate v-ml
miAw mL “h%
1 C&X
2 w:w
A I Program Loading Dose 13. Enter the desired VTBI using the keypad.
WandMive
Note: After the desired Rate and VTBI values are selected, the
Duration is automatically updated.
I A I PUMPING STOPPED I B I After the programmed delay period has expired, the
-Do-
Plum A+ starts “Pumping”. This is indicated where
500 Iwe 0.0 “Delayed” previously was displayed.
mL/hr
Loading Dose
Multistep
A) Program Multistep 10. Use [SELECT] to highlight the next programming field
WMd*aMI
(Diluent-Amount).
cow 50.0 mg m rnL
6.
Note: The Rate value is automatically computed. To understand the
Automatic Calculation feature, see Section 4.7 Exa~npZes ofAutomatic
Calculatfon,
AI Program Multistep 14. Use [SELECT] to highlight the Duration or VTBI field.
v&a MM
ceix 50.0 mg so0 rnL Note: The VTBI value is automatically computed when the
Duration value is entered. If VTBI is entered, the Duration value is
automatically computed.
16. Use [SELECT] to scroll past the Rate. The cursor will
drop down to Step-2 Dose field.
17. Enter the set of desired values for Step-2 of the program.
Continue process for each additionaf.Step.
Note: The **” in a Step number field indicates the step is delivering
AI Program ME2 1 or “d” indicates each completed step.
AI Program Multistep
If Dose value is changed, the Duration and Rate will
WMd*mve automatically change.
cone 50.0 mg 5X mL
Note: Only the current delivering step and future steps can be
changed.
Note: Other error and warning messages will occur when appropriate for the current
selection context. For instance, when a change to a required parameter would result in
clearing some delivery settings, the warning message will appear when that field is
selected. When an input error has just been made, the three quick beeps of an Invalid Key
Warning will sound and an appropriate warning messa e will be shown to explain the
problem. All such messages will be cleared or replaced %y another message upon the next
valid input key press.
Note: Alarm messages begin with an alphanumeric code for tracking purposes only. If
troubleshooting does not correct the problem, record the code number and contact
technical support.
GENERAL ALARMS
LINE A ALARMS
El 84,
El 90,
El91
a Proximal line obstructed, lineA
Program new VTBI.
LINE B ALARMS
E23 1 h-ox Air B, When proximal air detected Check for clamps, or empty
Backprime exceeds the limit for the line. container. Correct as necessary.
Backprime into line B tubing or
1 syringe.
13a
L
MALFUNCTIONS
‘.
7.1 Cleaning, Maintenance, and Storage
The cleaning, maintenance, and storage of the Plum A+ are described in this section.
On a routine basis, clean all of the elements behind the cassette door using cotton-tipped swabs saturated
with cleaning solution. The cassette door may be unlatched from the door handle to facilitate cleaning.
To unlatch the cassette door from its handle, tilt the device back, open the cassette door, then push the
door release tab to open the door fully.
CAUTION: To AVOID MECRANICXL OR ELECTRONIC DAMAGE, DO NOT IMMERSE THE PLUM A+ IN ANY
FLUIDS OR CLEANING SOLUTIONS.
CAUTION: CERTAIN CLEANING AND MTIZING SOLUTIONS M AY SLOWLY DEGRADE COM PONENTS MADE
FROM SOME PLASTIC MATERIALS. USING ABRASIVE CLEANERS OR CLEANING SOLUTIONS NOT
RECOMMENDED BY ABBOTT LABORATORIES MAY RESULT IN PRODUCT DAMAGE. Do NOT USE COMPOUNDS
CONTAINING COMBINATIONS OF ISOPROPYL ALCOHOL AND DIMETWL BEti? AMMONIUM CHLORIDE.
CAUTION: N EVER USE SHARP OBJECTS SUCH AS FINGER N A I L S, PAPER CL&S, OR NE EDL ES TO CLEAN
ANY PART OF THE PLUM A+.
BATTERY MAINTENANCE
CAUTION: Do NOT OPERATE THE PLUM A+ ON PATIENTS WITa THE BATTERY REMOVED. USE OF A
PROPERLY MAINTAINED AND CHARGED BATTERY HELPS CONFIRM PROPER OPERATION.
CAUTION: IF THE LOW-BATTERY ALARM SOUNDS. CONNECT TRE PLUM A+ TO b%AINS POWER
IMMEDIATELY.
The Plum A+ is battery powered for emergency backup and temporary portable operation. A fully charged,
new battery provides six hours of operation at 125 ml/hr or 500 ml total volume delivered, whichever
occurs first.
Note: For optimum battery life, the Plum A+ should be operated on battery power for six continuous
hours at least once every six months and then charged for a minimum of six hours.
The battery charges whenever the Plum A+ is connected to mains power. If the Plum A+ is switched off,
recharge takes approximately six hours. Recharge takes longer if the Plum A+ is turned on.
As a general rule, the more often the battery is partially discharged and recharged, the sooner it will need to
be replaced. Consult a qualified biomedical technician for battery replacement if necessary.
To maintain maxImum battery charge and to prolong battery life, connect the Plum A+ to mains power
whenever possible.
STORAGE
To prolong the life of the Plum A+, observe the following storage precautions:
l Store the Plum A+ away from excessive heat, cold, and humidity
l Store the Plum A+ connected to mains power
l Switch the Plum A+ off using the [ON/OFF] key
SERVICE
All servicing or adjustments to the Plum A+ should be referred to qualified technical personnel. A technical
service manual may be ordered from the local Abbott Laboratories sales office.
The Plum A+ has been designed and manufactured to be safe, reliable, and easy to use. This section details
precautions and possible hazards.
For safe operation of the Plum A+, observe the following precautions and hazards.
8.1 Artifacts
Nonhazardous, low-level electrical potentials are commonly observed when fluids are administered using
infusion devices. These potentials are well within accepted safety standards, but may create artifacts on
voltage-sensing equipment such as ECG, EMG. and EEG machines. These artifacts vary at a rate that is
associated with the infusion rate. If the monitoring machine is not operating correctly or has loose or
defective connections to its sensing electrodes, these artifacts may be accentuated so as to simulate actual
physiological signals. To determine if the abnormality in the monitoring equipment is caused by the
infusion device instead of some other source in the environment, set the Infusion device so that it is
temporarily not delivering fluid. Disappearance of the abnormality indicates that it was probably caused by
the electronic noise generated by the infusion device. Proper setup and maintenance of the monitoring
equipment should eliminate the artifact. Refer to the appropriate monitoring equipment system
documentation for setup and maintenance instructions.
The Plum A+ system is designed to operate normally in the presence of most encountered electromagnetic
interference (EMI) conditions. In the event of extreme levels of Interference, such as encountered next to an
electrosurgical generator, it is possible that the normal operation of a sensor or microcomputer might be
disrupted. Even in this event, the outcome would likely be a false alarm or detected system malfunction and
would not result in a hazard to patient or operator.
Use of radio frequency emitting devices such as cellular telephones and a-way radios in close proximity of
E:
this device may affect its operation.
Setting the primary rate greater than the secondary rate will result in a more rapid infusion of any residual
secondary drug remaining in the line and the cassette.
Consult drug labeling to co&ii drug compatibility, concentration, delivery rates, and volumes are all
suitable for secondary, concurrent and piggyback delivery modes.
Connecting more than one pump Into one patient line may significantly affect the infusion rate of at least
one of the pumps.
Arrange tubing, cords, and cables to minimize the risk of patient strangulation or entanglement.
Before opening the door, close clamp on the primary line or remove the sebondary container from the
secondary port to prevent mixing of primary and secondary fluids.
Although unlikely, failure of certain robust mechanical components such as the anti-free flow mechanism
or valve control springs could cause fluid delivery limited to the contents of the fluid container. Single fault
failure of certain electronic/motor control components would result in no more than 5 mL of unexpected
fluid delivery.
A small amount of fluid is expelled from the set (less than 0.05 ml) each time the door is opened or closed
with a set installed. If potent drugs are being used, take appropriate action to guard against
overmedication of the patient.
Before disconnecting a syringe from the cassette, pull up the plunger slightly to avoid spilling the fluid. For
rigid containers (for example, vials), close the upper slide clamp, open the cassette door, then remove and
invert the cassette (ports down).
Air bubbles may form distal to the cassette as result of normal out-gassing of dissolved air in the fluid.
This may occur if chilled solution is in use, if the pump is mounted significantly above the patient, or when
using certain fluids known to routinely outgas. In these cases, an air eliminating filter may be used.
Repeated opening and closing of the door may defeat the proximal air-in-line alarm and’may cause a distal
air-in-line alarm, requiring repriming.
The screen displays the VTBI (volume to be infused) in integers when value is above 99.9. Any fraction of a
milliliter delivered is not displayed, but is retained in memory.
When delivering short half-life critical drugs (See Critical Drugs, this section) using the Plum A+ in the
Concurrent mode, the following delivery rate guidelines should be observed:
l If the critical drug (with half-life less than 6 minutes) is to be infused at less than 2.0 mL/hr, the
other infusion should be no faster than 5 times the critical drug’s rate. Dopamine, for example,
delivered at 1.5 mL/br should not be accompanied by an infusion programmed any faster than
7.5 mL/hr.
l If the critical drug (with half-life less than 6 minutes) is to be infused at 2.0 to 5.0 mL/hr the
other infusion should be no faster than ten times the critical drug’s rate. Dopamine, for example,
delivered at 3.5 ml/hr should not be accompanied by an infusion programmed any faster than 35
mL/hr.
l If the critical drug (with half-life less than 6 minutes) is to be infused at 5.1 mL/hr or greater, the
other infusion can be programmed at any desired rate.
Note:The total of the primary rate plus the secondary rate cannot exceed 500 mL/hr.
NoteThese guidelines apply only when Infusing short half-1iJe critical drugs in Concurrent mode. Individual
patient responses may vary requiring adjustment of delivery rates.
Delivery Rate Guidelines
Short Half-life (less than 6 minutes) Maximum Rate of Accompanying Infusion
Critical Drug Infusion Rate ‘, - I ‘.
0.5 - 1.9 mL/hr 5 Times the Critical Drug Rate
2.0 - 5.0 mL/hr 10 Times the Critical Drug Rate
5.1 or Greater Any Desired Ratio
CRITICAL D RUGS
Examples of drugs with a short half-life (approximately 6 minutes or less when given IV) Include:
l Dobutamine l Nitroglycerin
l Dopamine l Nitroprusside
l Epinephrine l Norepinephrine
l Esmolol l oxytocin
l Isoproterenol l Procainamide
l Lidocaine l Trimethaphan
For these drugs, the concurrent flow guidelines should be followed when the infusion rate of the drug will be
5 mL/hr or less.
Note: This list of critical drugs is not intended to be all-inclusive of critical drugs or drugs with a short
half-life.
The clinician should become familiar with the pharmacodynamics of any critical drug before
administration.
This information is presented to inform clinicians of a rare situation that could be misinterpreted if they are
unfamiliar with this phenomenon.
E:
8.4 Epidural Administration
Recommended use of the epidural route is to provide anesthesia or analgesia for periods up to 96 hours.
This device can be used to administer only those anesthetics/analgesics approved for epidural
administration (as indicated or allowed by the drugs’ FDA approved labeling). Epidural administration of
drugs other than those Indicated for epidural use could result in serious Injury to the patient.
For epidural administration, the use of Abbott Laboratories catheter, List No. 1193 or List No. 6947, pump
sets without Y-sites, and “epidural” stickers indicating ongoing epidural administration are recommended.
Administration of drugs via the epidural route should be limited to personnel familiar with associated
techniques and patient management problems. Proper epidural placement of the catheter is essential since
catheter migration could result in intravascular or intrathecal administration. Facilities practicing epidural
administration must be equipped with resuscitative equipment, oxygen, naloxone, and other resuscitative
drugs. Adequate monitoring equipment (e.g., Orrimetry) is recommended for continuous monitoring of the
patient during epidural administration. Patients must be observed frequently for side effects in a fully-
equipped and staffed environment for at least 24 hours following completion of drug administration by the
epidural route. DELAYED RESPIRATORY DEPRESSION FOLLOWING CONTINUOUS EPIDURAL
ADMINISTRATION OF PRESERVATIVE-FREE MORPHINE SULFATE HAS BEEN REPORTED.
The epidural space has 58 openings through which fluid can exit. Pressure buildup during administration
is transient. However, if a large volume of fluid is administered over a short time period, the pressure will
take longer to return to normal. If overdelivery occurs during administration, observe the patient closely for
signs of spinal cord compression (disorientation, headache, transient neuralgias) and drug overdose.
The battery may not be fully charged upon receipt. Connect the Plum A+ to AC power for at least six hours.
Use AC power whenever possible. Connect the Plum A+ to AC power during storage to ensure a fully
charged battery for emergencies. If quality earth grounding source is in doubt, use battery power.
Administration sets should be changed per CDC guidelines or healthcare provider policy. Discard after use.
LifeCare IV Infusion sets with integral nonblood filters are not for use in the administration of blood, blood
products, emulsions, suspensions, or any medications not totally soluble in the solution being
administered. These medications may be administered through the lower Y-injection site, below the filter.
When infusing at low delivery rates (5.0 mL/hr or less) the use of thick-walled microbore Plum sets is
recommended. This will reduce the amount of the fluid bolus that may be delivered when a distal line
occlusion is released.
Use List 11986-48, syringe adapter when using syringes 10 cc or smaller on the secondary line. (Syringes
must be larger than 3 cc.)
Use a 19-gauge or larger needle or catheter at the venipuncture site for viscous fluids if operating at rates
greater than 500 ml/hr.
8.7 Backpriming
Backpriming is not recommended for reconstituting secondary containers containing dry powders.
To avoid pressurization when backpriming Into a syringe or a vial, the user must confirm that these
containers have sufficient empty space to accept the backprimed fluid.
8.8 General
Possible explosion hazard exists if used in the presence of flammable anesthetics.
Keep the cassette door securely closed while the Plum A+ is not in use, to avoid cassette door damage.
Use the following procedure to avoid the administration of a bolus following a distal occlusion:
If a secondary container is in use. clamp proximal tubing before opening cassette door
Open cassette door and remove the cassette
Open the flow regulator briefly to dissipate the pressure and then close it
Eliminate the source of occlusion
Reinsert the cassette and close the cassette door
Open all clamps and resume infusion
PHYSICAL
Dimensions: Approximately 8” X 8” X 6”, excluding pole clamp
protrusion and power cord storage.
Weight: Approximately 9.5 Ibs. with battery.
Casing: High-impact plastic
ELECTRICAL
Power Requirements: 120 V-, 50-60 Hz, 35 VA.
Fuses: Fl, F2,25OV-, 0.5 A. (internal)
Power Cord: Hospital-grade AC cord. 10 ft long, with transparent plug
and retainer plate.
Battery: One sealed, lead-acid, rechargeable 6 V battery, internal to
device.
Battery Life: With a new fully charged battery, the infuser shall operate
for a minimum of six hours at 125 mIJhr or less, or deliver
500 mL at 126 mL/hr or greater on one line.
Recharge: The battery charges whenever the Plum A+ is connected to
AC power. The recharge time is approximately six hours
with the device operating at 125 mIJhr on one line.
Electrical Leakage: Risk current limits meet AAMI/ANSI SCL (ungrounded)
12/78 standard.
NURSE-CALL System: NURSE-CALL alarm is factory set for Normally- Open
(NO)
Note: Contact the Technical Services Center to make an
internal adjustment to change the device from Norrnally-
Open (NO) to Normally Closed (NC) system.
Circuitry Ratings: Voltage-30 VDC Max
Current- 0.25 Amps Max
Contact Rating- 3 Watts Max
VTBI RANGE
VTBI Range: 0.1 to 99.9 mL (in 0.1 mL. increments)
100 to 9999 mL (in 1 mL increments)
ENVIRONMENT
Operating Temperature: 5” to 40” C
Storage Temperature: -20” to 60” C
Atmospheric Pressure: 0 - 10,000 feet (0 - 3,000m) or equivalent pressure
Relative Humidity: 10 - 90% (40” C Max)
AIR-IN-LINE ALARM
PlumSet (Distal): Bolus 0.1 mL or larger
Cumulative 0.26 rnL out of 2.5 mL
PlumSet (Proximal): Bolus at 0.5 mL, Total 1.0 mL
TRUMPET CURVES
The Trumpet Curve Graphs following the Example show representative maximum and minimum percent
flow rate deviation from the programmed rate over time. This information was developed in accordance with
IEC 60601-2-24: 1998, Sub-Clause 50.102. Refer to this standard for detailed information.
Hog
p& W T wi : The graphs following the
Example plot flow rates at 30 second intervals for the first 2 hours and for the 96th hour of delivery. The
graph plots mean delivery rate error (Average of 3 pumps) for the 2nd hour and the 96th hour as a straight
line. The graph also presents maximum and minimum average delivery rate error for this interval plotted by
averaging delivery errors over intervals of 2, 5, 11, 19 and 3 1 minutes (“Trumpet Curve”).
l Whole Blood
l Blood fractions
l Isomil Powder
l Similac Powder
l Ensure plus HN
l TwoCal HN
CONTAINERS
Dual Chamber Parenteral flexible container (Nutrimix)
Large Volume Parenteral flexible plastic containers, including premixed
Large Volume Parenteral glass containers, including premixed and nutritional
Part Fill Parenteral flexible plastic containers, including ADD-Vantage
Part Fill Parenteral glass containers
Small Volume Parenterals
Syringes, vials. includes part fills (may require special adapters)
Top Filled Enteral bags
Ready to hang Enteral solution containers
10.2 Accessories
Syringe Adaptor- List# 11986-48
Vial Adaptor- List# 1188448
Mini Pole- List# 12096-04-91
Secondary Container Support- List# 12095-04-61
Bar Code Reader (Available as a Field-Installed Infuser Option)- List# 11972-04-01
Tandem Carrier- List# 12270-O 1