You are on page 1of 6

Designation: G 109 – 07

Standard Test Method for


Determining Effects of Chemical Admixtures on Corrosion
of Embedded Steel Reinforcement in Concrete Exposed to
Chloride Environments1
This standard is issued under the fixed designation G 109; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.

1. Scope C 876 Test Method for Half-Cell Potentials of Uncoated


1.1 This test method covers a procedure for determining the Reinforcing Steel in Concrete
effects of chemical admixtures on the corrosion of metals in C 881/C 881M Specification for Epoxy-Resin-Base Bond-
concrete. This test method can be used to evaluate materials ing Systems for Concrete
intended to inhibit chloride-induced corrosion of steel in C 1152/C 1152M Test Method for Acid-Soluble Chloride in
concrete. It can also be used to evaluate the corrosivity of Mortar and Concrete
admixtures in a chloride environment. D 448 Classification for Sizes of Aggregate for Road and
1.2 The values stated in SI units are to be regarded as the Bridge Construction
standard. The values given in parentheses are for information D 632 Specification for Sodium Chloride
only. E 177 Practice for Use of the Terms Precision and Bias in
1.3 This standard does not purport to address all of the ASTM Test Methods
safety concerns, if any, associated with its use. It is the E 691 Practice for Conducting an Interlaboratory Study to
responsibility of the user of this standard to establish appro- Determine the Precision of a Test Method
priate safety and health practices and determine the applica- G 3 Practice for Conventions Applicable to Electrochemical
bility of regulatory limitations prior to use. Measurements in Corrosion Testing
G 15 Terminology Relating to Corrosion and Corrosion
2. Referenced Documents Testing
2.1 ASTM Standards: 2 G 33 Practice for Recording Data from Atmospheric Cor-
A 615/A 615M Specification for Deformed and Plain rosion Tests of Metallic-Coated Steel Specimens
Carbon-Steel Bars for Concrete Reinforcement G 46 Guide for Examination and Evaluation of Pitting
C 33 Specification for Concrete Aggregates Corrosion
C 143/C 143M Test Method for Slump of Hydraulic- 2.2 NACE Standards:3
Cement Concrete SSPC-SP 5/NACE No. 1 White Metal Blast Cleaning
C 150 Specification for Portland Cement
3. Significance and Use
C 173/C 173M Test Method for Air Content of Freshly
Mixed Concrete by the Volumetric Method 3.1 This test method provides a reliable means for predict-
C 192/C 192M Practice for Making and Curing Concrete ing the inhibiting or corrosive properties of admixtures to be
Test Specimens in the Laboratory used in concrete.
C 231 Test Method for Air Content of Freshly Mixed 3.2 This test method is useful for development studies of
Concrete by the Pressure Method corrosion inhibitors to be used in concrete.
C 511 Specification for Mixing Rooms, Moist Cabinets, 3.3 This test method has been used elsewhere with good
Moist Rooms, and Water Storage Tanks Used in the agreement between corrosion as measured by this test method
Testing of Hydraulic Cements and Concretes and corrosion damage on the embedded steel (1-4).4 This test
method might not properly rank the performance of different
corrosion inhibitors, especially at concrete covers over the steel
1
This test method is under the jurisdiction of ASTM Committee G01 on less than 40 mm (1.5 in.) or water-to-cement ratios above 0.45.
Corrosion of Metals and is the direct responsibility of Subcommittee G01.14 on The concrete mixture proportions and cover over the steel are
Corrosion of Metals in Construction Materials.
Current edition approved Nov. 1, 2007. Published December 2007. Originally
approved in 1992. Last previous edition approved in 2005 as G 109–99a(2005).
2 3
For referenced ASTM standards, visit the ASTM website, www.astm.org, or Available from The Society for Protective Coatings (SSPC), 40 24th St., 6th
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Floor, Pittsburgh, PA 15222-4656.
4
Standards volume information, refer to the standard’s Document Summary page on The boldface numbers in parentheses refer to a list of references at the end of
the ASTM website. this standard.

Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.

--```,``,``,``,,,,`,,,``,,,`,`,,-`-`,,`,,`,`,,`---

Copyright ASTM International 1


Provided by IHS under license with ASTM Licensee=MHI - NAGOYA related to 3944000/3944000013
No reproduction or networking permitted without license from IHS Not for Resale, 01/09/2008 23:23:13 MST
G 109 – 07
chosen to accelerate chloride ingress. Some inhibitors might
have an effect on this process, which could lead to results that
would differ from what would be expected in actual use (5).
4. Apparatus
4.1 The apparatus required for the evaluation of corrosion
inhibitors includes a high impedance voltmeter (at least one
Mohm) capable of measuring to 0.01 mV, a 100 V (65 %)
resistor.
5. Reagents and Materials
5.1 Cement, that conforms to Type I or Type II of Specifi-
cation C 150. Coarse aggregate shall conform to Specification
C 33 and Classification D 448, with nominal maximum size
between 9.5 and 19 mm (3⁄8 and 3⁄4 in.).
NOTE 1—Preferred maximum size aggregate is 12.5 mm (0.5 in.).
5.2 Steel Reinforcement Bars, deformed, meeting the re-
quirement of Specification A 615/A 615M; with a diameter NOTE—All measurements in inches (25.4 mm = 1 in.).
FIG. 1 Concrete Beam
between 10 mm (0.4 in.) and 16 mm (0.6 in.), and a length of
360 mm (14 in.), drilled and tapped at one end to be fitted with
coarse-thread stainless steel and nuts, as described in 5.3 and

--```,``,``,``,,,,`,,,``,,,`,`,,-`-`,,`,,`,`,,`---
5.4. These bars shall be used to manufacture the test specimens, 5.13 Silicone Caulk, for sealing the outside of the plastic
as described in Section 6. dam to the top of the concrete specimen.6,9
5.14 Reference Electrode, such as a saturated calomel or
NOTE 2—Interlaboratory test program and statistical data in Section 11 silver/silver chloride electrode for measuring the corrosion
are based upon 13-mm (0.5-in.) steel bars, 12.5-mm maximum size
potential of the bars, as defined in Terminology G 15.
aggregate, and 19-mm (0.75-in.) and 25-mm (1 in.) cover.
5.15 Hexane.
5.3 316 Stainless Steel Screws, with diameter smaller than
bar diameter (coarse thread < 5 mm (0.2 in.)), 25 to 35-mm (1 6. Preparation of Test Specimens
to 1.5-in.) long (one per bar).
6.1 Power wire brush or sand blast the bars to near white
5.4 316 Stainless Steel Nuts, two per bar to fit stainless steel
metal (see SSPC-SP 5/NACE No. 1), clean by soaking in
screws, as described in 5.3.
hexane, and allow to air dry.
5.5 Two-part Waterproof Epoxy5,6—This epoxy shall meet
the chemical resistance requirements of a Type IV, Grade 3, NOTE 3—Pickling the bars with 10 % sulfuric acid for 10 to 15 min and
Class E of Specification C 881/C 881M. rinsing with potable water prior to wire brushing is recommended when
5.6 Sulfuric Acid, 10 % by mass, for pickling (optional). the bars have an excessive amount of rust.
5.7 Electroplater’s Tape.6,7 6.2 Use the same method to clean all bars in the test
5.8 Neoprene Tubing, with 3-mm (1⁄8-in.) wall thickness and program.
the same ID as the diameter of the bar used. 6.3 Drill and tap one end of each bar, attach a stainless steel
5.9 Sodium Chloride, complying with Specification D 632. screw and two nuts, as described in 5.3 and 5.4, and tape each
5.10 Salt Solution, prepared by dissolving 3 parts of sodium end of the bar with electroplater’s tape so that a 200-mm (8-in.)
chloride (as described in 5.9) in 97 parts of water mass. portion in the middle of the bar is bare. Place a 90-mm (3.5 in.)
5.11 Epoxy Sealer, for application to the concrete specimens length of neoprene tubing, as described in 5.8, over the
after manufacture. This sealer shall be of Type III, Grade 1, electroplater’s tape at each end of the bar, and fill the length of
Class C in accordance with Specification C 881/C 881M.6,8 tubing protruding from the bar ends with the two-part epoxy, as
5.12 Plastic Dams, 75-mm (3-in.) wide and 150-mm (6-in.) described in 5.5.
long with a minimum height of 75 mm (3 in.) for placement on 6.4 Specimen size is 280 3 150 3 115 mm (11 3 6 3
the test specimens. The wall thickness shall be 61 mm (1⁄8 6 4.5 in.). Place two bars, as described in 5.2, 25 mm (1 in.) from
1⁄32 in. the bottom, and one bar at the top such that the distance from
its top to the top surface of the specimen is twice the maximum
5
aggregate size, as shown in Fig. 1.
The sole source of supply of the apparatus known to the committee at this time
is PC-Epoxy, made by Protective Coating Co., Allentown, PA. NOTE 4—For example, for a 12.5-mm (0.5 in.) aggregate, place the top
6
If you are aware of alternative suppliers, please provide this information to bar 25 mm (1 in.) from the surface. For a 9.5-mm (0.375-in.) aggregate,
ASTM International Headquarters. Your comments will receive careful consider- place the bar 19 mm (0.75 in.) from the top surface.
ation at a meeting of the responsible technical committee,1 which you may attend.
7
The sole source of supply of the apparatus known to the committee at this time
is Minnesota Mining and Manufacturing Company (3M), 1999 Mt. Read Boulevard,
Rochester, NY 14615.
8 9
The sole source of supply of the apparatus known to the committee at this time The sole source of supply of the apparatus known to the committee at this time
is Epoxy Concrete Scaler # 12560, made by Devcon. is 3M Marine Adhesive 5200.

Copyright ASTM International 2


Provided by IHS under license with ASTM Licensee=MHI - NAGOYA related to 3944000/3944000013
No reproduction or networking permitted without license from IHS Not for Resale, 01/09/2008 23:23:13 MST
G 109 – 07
6.5 Place the bars in the molds so that 40 mm (approxi-
mately 1.5 in.) of the bars are protected within each exit end
from the concrete (minimizes edge effects). This will expose
200 mm (8 in.) of steel. Place the bars with the longitudinal ribs
so that they are nearer the side of the beam, that is, both ridges
are equidistant from the top or bottom of the specimen.
6.6 Make the concrete specimens (controls and those with
admixtures to be tested) in accordance with Practice C 192/
C 192M, using the same source of materials. Determine the air
content, using either Test Method C 231 or C 173/C 173M. The
water-to-cement ratio (w/c) shall not exceed 0.5. The minimum
slump is 50 mm (2 in.) (See Test Method C 143/C 143M).
Place and consolidate the concrete in the molds containing the
bars in accordance with Practice C 192/C 192M.
NOTE 5—The concrete parameters used in the interlaboratory test were
as follows: cement content of 355 6 3 kg/m3 (600 6 5 lb/yd3), 0.50 6
0.01 w/c (ssd aggregates), and 6 6 1 % air.
6.7 Add the admixture to be tested at the manufacturer’s
recommended dosages. A water reducer is allowed, if needed, NOTE—All measurements in inches (not to scale) (25.4 mm = 1 in.).
to achieve the desired slump. Record the admixtures used. FIG. 2 Concrete Beam (Side View)
Except for the test admixtures, use the same admixtures in all
mixtures. a depth of 40 mm (1.5 in.). Use a plastic loose fitting cover to
6.8 A minimum of three replicates shall be made. Make the minimize evaporation. Maintain a relative humidity around the
same number of replicates per admixture tested and control specimens of 50 6 5 %. After two weeks, vacuum off the
(see Note 6). An addition cylinder 100 3 200 mm (4 3 8 in.) solution and allow the samples to dry for two weeks. Repeat
in diameter shall be produced for background chloride analy- this cycle.
sis. 7.2 Measure the voltage across the resistor at the beginning
of the second week of ponding using the voltmeter defined in
NOTE 6—A larger number of replicates is preferred. 4.1. Calculate the current, Ij, from the measured voltage across
6.9 Apply a wood float finish after consolidation. After the 100 V resistor, Vj, measured in volts (see Note 8) as:
removal from the forms, cure the specimens for 28 days in a Ij 5 Vj /100
moist room in accordance with Test Method C 192/C 192M
NOTE 8—With the common terminal on the bottom bar, negative
and Specification C 511.
voltages correspond to positive galvanic current (that is, the top bar is the
6.10 Upon removal from the moist room, hand wire brush anode).

--```,``,``,``,,,,`,,,``,,,`,`,,-`-`,,`,,`,`,,`---
the specimens on the concrete top surface (wood floated
surface). Allow the specimens to dry for two weeks in a 50 % 7.3 At the same time, measure the corrosion potential of the
relative humidity (RH) environment before sealing the four bars against a reference electrode that is placed in the dam
vertical sides with an epoxy sealer, as described in 5.11, in containing the salt solution (see Practice G 3 and Test Method
accordance with the manufacturer’s recommendation. Place a C 876). Connect the voltmeter between the reference electrode
plastic dam with dimensions, as described in 5.12, on the (ground or common terminal) and the bars.
specimen, as shown in Fig. 1, and about 13 mm (0.5 in.) from 8. Period of Testing
each side so that it does not extend over the taped sections of
8.1 Monitor the current as a function of time once every
the bars (see Fig. 2). Use a silicone caulk to seal the dam from
four weeks, as described in 7.2, until the average integrated
the outside, and apply epoxy sealer to the top surface outside of
macrocell current of the control specimens is 150 C or greater,
the dam.
as determined in 10.1.8, and at least half the samples show
NOTE 7—Allowing the specimens to dry before applying the concrete integrated macrocell currents equal to or greater than 150 C
epoxy will make the initial exposure to chloride more severe, and more (see Note 9).
closely follow the interlaboratory test program conditions.
NOTE 9—The value of 150 C is consistent with a macrocell current of
6.11 Attach wires and resistors. 10 µA over six months. The value of 10 µA was measured by all
laboratories on all specimens showing corrosion (controls and samples
7. Procedure with calcium chloride at 19-mm (3⁄4-in.) cover). This degree of integrated
7.1 Support each test specimen on two nonelectrically macrocell current is sufficient to ensure the presence of sufficient
conducting supports at least 13-mm (0.5-in.) thick, thus allow- corrosion for visual evaluation.
ing air flow under most of the specimen. Start the test one 8.2 In those cases where the admixtures being tested are
month after the samples are removed from the 100 % RH corrosive, end the test three full cycles after an average
atmosphere (moist room). Pond the specimens for two weeks at integrated macrocell current of 75 C is observed and the
23 6 3°C (73 6 5°F) with the salt solution, as described in integrated macrocell current of at least half the specimens
5.10. The volume of this solution is approximately 400 mL at being tested is equal or greater than 75 C.

Copyright ASTM International 3


Provided by IHS under license with ASTM Licensee=MHI - NAGOYA related to 3944000/3944000013
No reproduction or networking permitted without license from IHS Not for Resale, 01/09/2008 23:23:13 MST
G 109 – 07
9. Examination of Embedded Bars
9.1 At the conclusion of testing, break the specimens and
examine the reinforcement bars for extent of corrosion, mea-
sure the corroded area, and record the percentage of corroded
area recorded, as described in Practice G 33.
NOTE 10—Photograph the bars at the end of the test to provide a record
of the corrosion damage.
9.2 Determine the acid soluble chloride content at the depth
corresponding to the cover over the top-reinforcing bar, using
Test Method C 1152/C 1152M.
9.3 Determine the acid soluble chloride content in the
specimen produced for background chloride analysis, using
Test Method C 1152/C 1152M. This value is to be subtracted
from the acid soluble chloride, as determined in 9.2, to provide
a corrected acid soluble chloride content reflecting ingressed
FIG. 3 Standard Deviation of Repeatability
chloride.

10. Report participated in the study. The repeatability and reproducibility


10.1 Report the following information: of the test results were dependent on the magnitude of the mean
10.1.1 Full details of the concrete proportions, air content, macrocell current.
and slump of the concrete used in the control and test 11.2 Precision is as follows:
specimens, 11.2.1 95 % Repeatability Limit (Within Laboratory)—The
10.1.2 A plot of the corrosion current and potential for each within-laboratory precision of the average macrocell current
concrete specimen versus time, (for each laboratory), as expressed by the repeatability limit, r,
10.1.3 A plot of the average integrated current for each is given by the following relation:
condition of concrete versus time, log r 5 0.931 log Iavg 1 0.441 (1)
10.1.4 Time to failure, as considered to be the time for the
11.2.2 95 % Reproducibility Limit (Between
average macrocell current to reach 10 µA and at least half the
Laboratories)—The between-laboratory precision of the aver-
samples showing a current greater than 10 µA,
age macrocell current (for all laboratories), as expressed by the
10.1.5 Results of the visual inspection of each bar. The
reproducibility, R, is given by the following relation:
report shall include the percentage of original exposed steel
surface corroded and optionally the number and depths of log R 5 0.833 log Iavg 1 0.624 (2)
corrosion pits where present, as described in Practice G 46, 11.2.3 The repeatability and reproducibility limits of the
10.1.6 Photographs of the bars at the end of the test average macrocell current were calculated in accordance with
(optional), and Practice E 177. The respective standard deviations of the
10.1.7 Chloride content at the top reinforcing bar depth variation among test results can be obtained by dividing by 2.8
from the surface. This value is the corrected total chloride the values of r and R calculated using (Eq 1) and (Eq 2). The
content, as corrected 9.3. following equations were then obtained:
--```,``,``,``,,,,`,,,``,,,`,`,,-`-`,,`,,`,`,,`---

10.1.8 The ratio of total integrated current of the test log Sr 5 0.931 log Iavg 2 0.006 (3)
specimen to that of the control and time the test ended. The
log S R 5 0.833 log Iavg 1 0.177 (4)
total integrated current is:
TCj 5 TC j21 1 [~tj 2 t j21! 3 ~ij 1 i j21!/2]
11.2.4 The data used for compiling the test method preci-
sion, together with the statistical parameters as defined in
where: Practice E 691, are given in the research report.10 The graphi-
TC = total corrosion (coulombs), cal representations of the repeatability and reproducibility
tj = time (seconds) at which measurement of the macro- limits are given in Figs. 3 and 4.
cell current is carried out, and 11.2.5 The time to failure has been analyzed using Practice
ij = macrocell current (amps) at time, tj. E 691. This analysis is given in the research report.10
A sample calculation is given in Appendix X1. 11.2.6 The maximum end of the 95 % confidence interval
for time to failure for control specimens with 19-mm (0.75-in.)
11. Precision and Bias 10 concrete cover is six months for both intralaboratory and
11.1 Information on the precision of the results obtained by interlaboratory tests. The maximum ends of the 95 % confi-
this test method was derived from an interlaboratory test with dence intervals are two and six months for intra- and interlabo-
two to three specimens per laboratory. Eleven laboratories ratory tests respectively for specimens containing calcium
chloride.
11.2.7 The complete data for percent area corroded is given
10
Supporting data have been filed at ASTM International Headquarters and may in the research report.10 In all cases where there was corrosion,
be obtained by requesting Research Report RR: G01–1009. the macrocell current was greater than 9 µA. However, not

Copyright ASTM International 4


Provided by IHS under license with ASTM Licensee=MHI - NAGOYA related to 3944000/3944000013
No reproduction or networking permitted without license from IHS Not for Resale, 01/09/2008 23:23:13 MST
G 109 – 07
enough laboratories reported percent area corroded to carry out
a statistical analysis following Practice E 691.
11.3 Bias—The procedure given in this test method has no
bias because the effects of chemical admixtures on the corro-
sion of embedded steel of reinforcement are defined only in
terms of this test method.

12. Keywords
12.1 admixtures; concrete; corrosion; corrosivity; inhibitor;
reinforcing steel

FIG. 4 Standard Deviation of Reproducibility

APPENDIX

(Nonmandatory Information)

X1. TOTAL CORROSION CALCULATION

X1.1 Total Corrosion Calculation: X1.1.3 At the end of the 60 day period:
TC j 5 TCj21 1 [~t j 2 tj21! * ~i j 1 ij 21!/2] (X1.1) TC2 5 25.92 1 [60 2 30! * 86400 * ~20 1 27!/2*1026# 5 86.83 C
X1.1.1 Assume the following readings were obtained over a (X1.3)
90 day period of time: X1.1.4 At the end of the 90 day period:
Days 0 30 60 90
imac (µA) 0 20 27 35 TC3 5 86.83 1 [~90 2 60! * 86400 * ~27 1 35!/2*1026# 5 167.18 C
(X1.4)
X1.1.2 At the end of the first 30 day period the total
--```,``,``,``,,,,`,,,``,,,`,`,,-`-`,,`,,`,`,,`---

corrosion is: NOTE X1.1—Conversion factor from days to seconds = 24 3 60 3 60


26
TC1 5 0 1 [~30 2 0! * 86400 * ~20 1 0 !/2 * 10 # 5 25.92 C = 86 400.
(X1.2)

REFERENCES

(1) Berke, N. S., Shen, D. F., and Sundberg, K. M., “Comparison of the Admixture or Epoxy-Coated Reinforcing Bars as Corrosion Protection
Polarization Resistance Technique to the Macrocell Corrosion Tech- Systems,” Report No. FHWA/RD-83/-12, Federal Highway Adminis-
nique,” Corrosion Rates of Steel in Concrete, ASTM STP 1065, N. S. tration, Washington DC, 1983, pp. 71.
Berke, V. Chaker, and D. Whitney, editors, ASTM, August 1990, pp. (4) Berke, N. S., Pfeifer, D. W., and Weil, T. G., “Protection Against
38–51. Chloride-Induced Corrosion,” Concrete International, December
(2) Berke, N. S. and Hicks, M. C., “Electrochemical Methods of
1988, pp. 45–55.
Determining the Corrosivity of Steel in Concrete,” Corrosion Testing
and Evaluation: Silver Anniversary Volume, Babraiam/Dean editors, (5) Berke, N. S., Hicks, M. C., Hoopes, R. J., and Tourney, P. J., “Use of
ASTM STP 1000, ASTM, November 1990, pp. 425–440. Laboratory Techniques to Evaluate Long-Term Durability of Steel
(3) Virmani, Y. P., Clear, K. C., and Pasko, T. J., “Time-to Corrosion of Reinforced Concrete Exposed to Chloride Ingress,” ACI SP 145-16,
Reinforcing Steel in Concrete Slabs, Volume 5: Calcium Nitrite 1994, pp. 299-328.

Copyright ASTM International 5


Provided by IHS under license with ASTM Licensee=MHI - NAGOYA related to 3944000/3944000013
No reproduction or networking permitted without license from IHS Not for Resale, 01/09/2008 23:23:13 MST
G 109 – 07

ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).

Copyright ASTM International


--```,``,``,``,,,,`,,,``,,,`,`,,-`-`,,`,,`,`,,`---
6
Provided by IHS under license with ASTM Licensee=MHI - NAGOYA related to 3944000/3944000013
No reproduction or networking permitted without license from IHS Not for Resale, 01/09/2008 23:23:13 MST

You might also like