Purpose: To ensure that instruments and systems are correctly installed and functioning as designed, in the
customer's facility. Correct installation is the first step in ensuring that instruments and systems operate reliably
over an extended lifetime.
Customer Responsibilities
The customer should ensure that the installation site is prepared in accordance with the specifications contained in
the relevant site preparation document and that the necessary operating supplies, consumables and usage dependent
items such as vials, paper, solvents, etc. are available. A customer representative should be available during the
installation.
Note: Typical installation time is 1.5 hours for a system and 30 minutes for a module.
If installation time was not sold for any section, check that section as not applicable.
Issued: Oct 2008 – Rev. 6.5 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
LC System Information
Installation Checklist
Note:
If one section contains multiple choices for installed Hardware please encircle all modules that are part of the
system and fill in the corresponding Serial Number.
Note: Shipping containers should not be opened until an Agilent Technologies representative is present
Unpack all boxes and place equipment on bench. To prevent injury, get lifting assistance if needed.
Retain shipping containers and material until installation is complete and performance is verified
Notify order fulfillment of any discrepancies
Unpack and verify condition and completeness of shipment
Check serial numbers and fill in installation documentation S/N fields if necessary
Verify that power cords are correct (must mate with power sockets) and are plugged into power receptacle
When applicable, install computer system and connect all power interconnect cables
Check for the Agilent diagnostic utility (LMD/LA) CD-ROM and install this software component
Verify that the instrument is the correct module for the regional voltage (100 V/115 V/230V). Do NOT use
instruments which have an incompatible voltage for ELSD.
Remote Advisor
Ask the Customer is remote advisor is installed
If yes in install the Data Source Software
Issued: Oct 2008 – Rev. 6.5 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
G1322A/79A/B– Degasser
Section NOT Applicable
G1310A/11A/12A – Pump
Issued: Oct 2008 – Rev. 6.5 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
G1322A/79A/B– Degasser
Issued: Oct 2008 – Rev. 6.5 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
G1322A/79A/B– Degasser
Issued: Oct 2008 – Rev. 6.5 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Issued: Oct 2008 – Rev. 6.5 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Issued: Oct 2008 – Rev. 6.5 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
G1329A/B – Autosampler
1. Record Firmware
Issued: Oct 2008 – Rev. 6.5 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
1. Record Firmware
Issued: Oct 2008 – Rev. 6.5 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
1. Record Firmware
Issued: Oct 2008 – Rev. 6.5 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
The following section applies if a G1330B - ALS Thermostat is installed together with any of the well-plate-,
autosamplers or fraction collectors to generate a: G1329A/B + G1330B - Thermostat. ALS G1364B/C/D +
G1330B Therm. Fraction Collector G1367B + G1330B - Thermostat. High Performance Autosampler G1377A
+ G1330B - Thermostat. Micro WPS G2260A + G1330B - Thermostat. Preparative ALS
Setup the ALS Thermostat(s) underneath the module(s) that has (have) to be thermostatted.
WARNING: Do not connect the ALS thermostat cable while the line power is connected to the any
of the modules. This would damage the sampler or fraction collector and the thermostat electronics.
air channel adapter and plate
ALS thermostat cable
ALS thermostat waste drain
Issued: Oct 2008 – Rev. 6.5 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
1. Record Firmware
Issued: Oct 2008 – Rev. 6.5 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Issued: Oct 2008 – Rev. 6.5 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Issued: Oct 2008 – Rev. 6.5 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
G1157A/58A/59A/60A/62A/63A Valves
Issued: Oct 2008 – Rev. 6.5 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
G1314A/B/C/D/E – VWD
1.Record Firmware
Lamp
Intensity Pass/Fail
Zero-order
WL Deviation
calibration 656nm
Deviation
*
Please note that ‘D/E’ detector requires ChemStation B.04.01 and diagnostic utility
Issued: Oct 2008 – Rev. 6.5 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
G1315C/D-DAD / G1365C/D-MWD
Lamp 1.
Intensity Pass/Fail
1. Alpha
WL Deviation
calibration Beta
Deviation
Lamp 2.
Intensity Pass/Fail
2. Alpha
WL Deviation
calibration Beta
Deviation
Issued: Oct 2008 – Rev. 6.5 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
G1321A – FLD
1.Record Firmware
Lamp
Intensity Pass/Fail
Excitation
WL Deviation
verification Emission
Deviation
Issued: Oct 2008 – Rev. 6.5 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
G1362A – RID
1.Record Firmware
Diode
Balance Value
Issued: Oct 2008 – Rev. 6.5 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
G4218A – ELSD
1.Record Firmware
G4218A Checkout
Switch on the instrument
Prepare the instrument:
o If a new column is used, flush the system for 15 minutes AND
o set the detector temperature to 35 °C and equilibrate for 15 minutes
Set up a method for the checkout sample (see user manual for test conditions)
Run a test measurement
Start a run
Attach results to installation documentation
Issued: Oct 2008 – Rev. 6.5 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Issued: Oct 2008 – Rev. 6.5 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
1.Record Firmware
Note: Agilent LAN connection is made between one (1) of the Agilent 1200 Series System modules and the
Agilent ChemStation PC. If a detector is installed connect the LAN to this module
1.Record Firmware
Issued: Oct 2008 – Rev. 6.5 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Software Product:_________________________
Software Revision:________________________
Start Windows
Start ChemStations -> Add Licenses and add the license numbers. You find these numbers in the 'Software
Certificate and Registration Packet' envelopes.
Configure TCP/IP if the instrument will be connected using LAN.
If necessary install and Configure IO Libraries (LC/MS)
If necessary install and configure Agilent Bootp Service for LAN connection.
Use the Configuration Editor and configure your instruments.
Install and Configure Software Add-Ons (if applicable)
If present, install the PC Image software and create a backup using the Backup Solution User Guide as
reference.
If present, install the Agilent LMD/LA (diagnostic utility) software.
Software Product:_________________________
Software Revision:_____________________
Start Windows
Install ChemStation using the ChemStation Installation guide for reference.
Check if a printer driver is installed on the system
Verify that the regional settings are set to English-US
Install and Configure TCP/IP if the instrument will be connected using LAN.
If necessary install and Configure IO Libraries (LC/MS).
If necessary install and configure AGILENT Bootp Service if the instrument will be connected using LAN Use
the Configuration Editor
Install the Agilent LMD/LA (diagnostic utility) software
Remote Advisor
Issued: Oct 2008 – Rev. 6.5 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
1200 LC Checkout
Prepare:
1. Isocratic sample 6. detector(s)
2. pump(s) 7. fraction collector(s) (if present)
3. online degasser (s) (if present) 8. Valve(s) (if present)
4. injection device(s) 9. recording device
5. column compartment
Start a run
Attach results to installation documentation
Fill out instrument logbook
Verify connection to the Agilent diagnostic utility.
Issued: Oct 2008 – Rev. 6.5 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Service Review
Library - http://www.agilent.com/chem/library
Education - www.agilent.com/chem/education
eSeminars - www.agilent.com/chem/eseminars
Supplies - www.agilent.com/chem/supplies
Date completed
__________________________
Customer Signature
__________________________
*01200-90606*
01200-90606 Edition October / 2008 Agilent Technologies
Printed in Germany Hewlett-Packard-Strasse 8
*01200-90606*
Issued: Oct 2008 – Rev. 6.5 Copyright Part
2008Number: 01200-90606 WAD Product Support © Select
Agilent Technologies Agilent Technologies,
76337Pages
Waldbronn, Inc.
Germany
for Required 2008
Modules
*01200-90606*
01200-90606 Edition 02 / 2008 Agilent Technologies
Printed in Germany Hewlett-Packard-Strasse 8