Agilent 1100/1200 Series

Liquid Chromatograph

Hardware and Software Installation Checklist

Hardware / Software Installation

Purpose: To ensure that instruments and systems are correctly installed and functioning as designed, in the
customer's facility. Correct installation is the first step in ensuring that instruments and systems operate reliably
over an extended lifetime.

Customer Responsibilities
The customer should ensure that the installation site is prepared in accordance with the specifications contained in
the relevant site preparation document and that the necessary operating supplies, consumables and usage dependent
items such as vials, paper, solvents, etc. are available. A customer representative should be available during the
installation.

Note: Typical installation time is 1.5 hours for a system and 30 minutes for a module.

If installation time was not sold for any section, check that section as not applicable.

Service Engineers Responsibilities


Please only complete/printout pages that relate to the system module(s) that have been installed along with this
page, LC System information and Service Review page and fill the total number of pages that will be included at
the bottom of this page

Additional 1200 SL Modules


If SL modules have been included in this installation please insert these module pages into this document (Please
note that the revision number will be different to that of the main document).

Total Number of pages for this Document:____________

Issued: Oct 2008 – Rev. 6.5 Copyright  2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph

Hardware and Software Installation Checklist

LC System Information

System Name: ____________________________

System Location: _________________________

Contact Name: ___________________________

Installation Checklist

 Section NOT Applicable

Note:
If one section contains multiple choices for installed Hardware please encircle all modules that are part of the
system and fill in the corresponding Serial Number.

Unpacking and inspection of shipped materials

Note: Shipping containers should not be opened until an Agilent Technologies representative is present

 Unpack all boxes and place equipment on bench. To prevent injury, get lifting assistance if needed.
 Retain shipping containers and material until installation is complete and performance is verified
 Notify order fulfillment of any discrepancies
 Unpack and verify condition and completeness of shipment
 Check serial numbers and fill in installation documentation S/N fields if necessary
 Verify that power cords are correct (must mate with power sockets) and are plugged into power receptacle
 When applicable, install computer system and connect all power interconnect cables
 Check for the Agilent diagnostic utility (LMD/LA) CD-ROM and install this software component
 Verify that the instrument is the correct module for the regional voltage (100 V/115 V/230V). Do NOT use
instruments which have an incompatible voltage for ELSD.

Remote Advisor
 Ask the Customer is remote advisor is installed
 If yes in install the Data Source Software

Issued: Oct 2008 – Rev. 6.5 Copyright  2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph

Hardware and Software Installation Checklist

G1322A/79A/B– Degasser
 Section NOT Applicable

1. Record model number

1. Record serial number

 Set up the degasser(s)


 solvent inlet/outlet
 Priming each channel with Isopropanol from solvent bottle to pump inlet

G1310A/11A/12A – Pump

1. Record model number 2. Record model number 3. Record model number

1. Record serial number 2. Record serial number 3. Record serial number

1. Record Firmware 2. Record Firmware 3. Record Firmware

 Set up the pump(s)


 solvent inlet/outlet
 waste drain
 remote control (if used)
 relay contacts (if installed)
 seal wash tubing (if installed)
 CAN
 Agilent LAN (see note*)

 Priming the pump


 prepare solvents
 turn on
 purge
 purge mobile phase
 monitor the pressure

Issued: Oct 2008 – Rev. 6.5 Copyright  2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph

Hardware and Software Installation Checklist

G1322A/79A/B– Degasser

1. Record model number

1. Record serial number

 Set up the degasser(s)


 solvent inlet/outlet
 Priming each channel with Isopropanol from solvent bottle to pump inlet

Issued: Oct 2008 – Rev. 6.5 Copyright  2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph

Hardware and Software Installation Checklist

G1322A/79A/B– Degasser

 Section NOT Applicable

1. Record model number

1. Record serial number

 Set up the degasser(s)


 solvent inlet/outlet
 Priming each channel with Isopropanol from solvent bottle to pump inlet

G1382A/G1376A Capillary Pump with G1379B Micro Degasser


&
G2225A/G2226A Nano pump with G1379B Micro Degasser

1. Record model number 2. Record model number

1. Record serial number 2. Record serial number

1. Record Firmware 2. Record Firmware

 Set up the pump


 solvent inlet/outlet
 waste drain
 remote control (if used)
 relay contacts (if installed)
 CAN
 Agilent LAN (see note*)
 set up the degasser
 solvent inlet/outlet
 priming each channel with Isopropanol from solvent bottle to pump inlet
 Priming the pump
 prepare solvents
 turn on
 purge
 purge mobile phase
 monitor the pressure

Issued: Oct 2008 – Rev. 6.5 Copyright  2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph

Hardware and Software Installation Checklist

G1361A – Prep. Pump

1. Record model number 2. Record model number

1. Record serial number 2. Record serial number

1. Record Firmware 2. Record Firmware

 Set up the pump(s)


 solvent inlet/outlet
 waste drain
 remote control (if used)
 relay contacts (if installed)
 seal wash tubing
 CAN
 Agilent LAN (see note*)

 Priming a prep pump


 Prime each pump individually
 Use the syringe + adapter + tubing and connect to the waste outlet of the pump`s EMPV
 Turn ON the pump and electronically switch the EMPV to waste
 Manually suck the solvent into and trough the pump
 Pump at least 0.5 liter of IPA through the pump

Issued: Oct 2008 – Rev. 6.5 Copyright  2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph

Hardware and Software Installation Checklist

G1328B – Manual Injector or 5065-9922 Prep. Manual Injector

 Section NOT Applicable

1. Record model number

1. Record serial number

 Setup the injector


 start/stop cable to pump
 base plate & mounting pole
 position valve
 inlet/outlet connected
 sample loop
 vent tubes

Issued: Oct 2008 – Rev. 6.5 Copyright  2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph

Hardware and Software Installation Checklist

G1329A/B – Autosampler

1. Record model number

1. Record serial number

1. Record Firmware

 Setup the autosampler(s)


 solvent inlet/outlet
 waste drain (must be in a straight line down, no bends, kinks or curves
 large volume tube
(if installed)
 BCD output (if installed)
 remote control (if used)
 relay contacts (if installed)
 CAN
 Agilent LAN (see note*)

Issued: Oct 2008 – Rev. 6.5 Copyright  2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph

Hardware and Software Installation Checklist

G1367A/B/C/D – High Performance Autosampler G1377A - Micro WPS

1. Record model number

1. Record serial number

1. Record Firmware

 Setup the autosampler(s)


 solvent inlet/outlet
 waste drain (must be in a straight line down, no bends, kinks or curves
 large volume tube
(if installed)
 BCD output (if installed)
 remote control (if used)
 relay contacts (if installed)
 CAN
 Agilent LAN (see note*)

Issued: Oct 2008 – Rev. 6.5 Copyright  2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph

Hardware and Software Installation Checklist

G2258A – Dual Loop ALS /G2260A - Preparative ALS

1. Record model number

1. Record serial number

1. Record Firmware

 Setup the autosampler(s)


 solvent inlet/outlet
 waste drain (must be in a straight line down, no bends, kinks or curves
 large volume tube
(if installed)
 BCD output (if installed)
 remote control (if used)
 relay contacts (if installed)
 CAN
 Agilent LAN (see note*)

Issued: Oct 2008 – Rev. 6.5 Copyright  2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph

Hardware and Software Installation Checklist

G1330B - ALS Thermostat

The following section applies if a G1330B - ALS Thermostat is installed together with any of the well-plate-,
autosamplers or fraction collectors to generate a: G1329A/B + G1330B - Thermostat. ALS G1364B/C/D +
G1330B Therm. Fraction Collector G1367B + G1330B - Thermostat. High Performance Autosampler G1377A
+ G1330B - Thermostat. Micro WPS G2260A + G1330B - Thermostat. Preparative ALS

1. Record model number

1. Record serial number

 Setup the ALS Thermostat(s) underneath the module(s) that has (have) to be thermostatted.

WARNING: Do not connect the ALS thermostat cable while the line power is connected to the any
of the modules. This would damage the sampler or fraction collector and the thermostat electronics.
 air channel adapter and plate
 ALS thermostat cable
 ALS thermostat waste drain

Issued: Oct 2008 – Rev. 6.5 Copyright  2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph

Hardware and Software Installation Checklist

G1316A/B - Column Compartment

1. Record model number

1. Record serial number

1. Record Firmware

 Setup the TCC


 solvent inlet/outlet
 column
 waste drain
 remote control (if used)
 CAN

Issued: Oct 2008 – Rev. 6.5 Copyright  2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph

Hardware and Software Installation Checklist

G1316C - Column Compartment


(2 or 3 Column Compartment configurations)

1. Record model number 2. Record model number 3. Record model number

1. Record serial number 2. Record serial number 3. Record serial number

1. Record Firmware 2. Record Firmware 3. Record Firmware

TCC # TCC # TCC #

 Setup the TCC


 Remove transportation locks
 Install the valve heads (inlet valve & outlet valve)
 Install the 4 heat exchanger carriers and fitting holders
 Install all “non-column capillaries (ALS->TCC (inlet]; TCC -> Detector [outlet], waste and bypass)
 Install all capillaries of the inlet valve and outlet valve (use color codes for the fittings)
 Install the checkout column (1 column only!)
 waste drain
 remote control (if used)
 CAN
 Cluster the column compartments
 Define inlet and outlet valve, bypass and waste positions

Issued: Oct 2008 – Rev. 6.5 Copyright  2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph

Hardware and Software Installation Checklist

Valve-Pos ___ Color_______ TCC-position_________


TCC-# ___
Ser-#____________ Valve-Pos ___ Color_______ TCC-position_________
Valve-Pos ___ Color_______ TCC-position_________
 Inlet –valve
 Outlet-valve Valve-Pos ___ Color_______ TCC-position_________

TCC-# ___ Valve-Pos ___ Color_______ TCC-position_________


Ser-#____________ Valve-Pos ___ Color_______ TCC-position_________
 Inlet –valve Valve-Pos ___ Color_______ TCC-position_________
 Outlet-valve Valve-Pos ___ Color_______ TCC-position_________

TCC-# ___ Valve-Pos ___ Color_______ TCC-position_________


Ser-#____________ Valve-Pos ___ Color_______ TCC-position_________
 Inlet –valve Valve-Pos ___ Color_______ TCC-position_________
 Outlet-valve
Valve-Pos ___ Color_______ TCC-position_________

 Bypass Valve-Pos ___  Waste Valve-Pos ___

 Post column cooler TCC-# ___ TCC-position  left


 right

Issued: Oct 2008 – Rev. 6.5 Copyright  2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph

Hardware and Software Installation Checklist

G1157A/58A/59A/60A/62A/63A Valves

 Section NOT Applicable

1. Record model number 2. Record model number 3. Record model number

1. Record serial number 2. Record serial number 3. Record serial number

1. Record Firmware 2. Record Firmware 3. Record Firmware

 Setup the Valves


 Setup the 1200 Series Valve
 solvent inlet/outlet
 CAN
 CAN-DC-In from other 1200 Module (e.g. G1367B/C/D Autosampler, G1364B/C/D Fraction Collector or
G1361A Prep. Pump)

G1383A Column Organizer


 Section NOT Applicable

1. Record model number

1. Record serial number

 Setup the column organizer


 base plate and organizer plate
 rods
 traverses
 additional holders (if used)

Issued: Oct 2008 – Rev. 6.5 Copyright  2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph

Hardware and Software Installation Checklist

G1314A/B/C/D/E – VWD

1. Record model number

1. Record serial number

1.Record Firmware

 Set up the detector


 flow cell
 solvent inlet/outlet
 waste drain
 remote control (if used)
 analog output (if used)
 relay contacts (if installed)
 CAN
 Agilent LAN (see note*)

 Verify the operation of the detector


 Perform the Lamp Intensity Test
 Perform the Wavelength Calibration Test

Lamp
Intensity Pass/Fail
Zero-order
WL Deviation
calibration 656nm
Deviation
*
Please note that ‘D/E’ detector requires ChemStation B.04.01 and diagnostic utility

Issued: Oct 2008 – Rev. 6.5 Copyright  2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph

Hardware and Software Installation Checklist

G1315C/D-DAD / G1365C/D-MWD

1. Record model number 2. Record model number

1. Record serial number 2. Record serial number

1. Record Firmware 2. Record Firmware

 Set up the detector(s)


 deuterium lamp (if not installed)
 solvent inlet/outlet
 waste drain
 remote control (if used)
 analog output (if used)
 relay contacts (if installed)
 CAN
 Agilent LAN (see note*)
 Verify the operation of the detector
 Perform the Lamp Intensity Test
 Perform the Wavelength Calibration Test

Lamp 1.
Intensity Pass/Fail
1. Alpha
WL Deviation
calibration Beta
Deviation

Lamp 2.
Intensity Pass/Fail
2. Alpha
WL Deviation
calibration Beta
Deviation

Issued: Oct 2008 – Rev. 6.5 Copyright  2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph

Hardware and Software Installation Checklist

G1321A – FLD

1. Record model number

1. Record serial number

1.Record Firmware

 Set up the detector


 solvent inlet/outlet
 waste drain
 remote control (if used)
 analog output (if used)
 CAN
 Agilent LAN (see note*)

 Verify the operation of the detector


 Perform the Lamp Intensity Test & compare profile
 Perform the Wavelength
 Verification Test using the Raman band.

Lamp
Intensity Pass/Fail
Excitation
WL Deviation
verification Emission
Deviation

 If test fails perform Wavelength Calibration.

Issued: Oct 2008 – Rev. 6.5 Copyright  2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph

Hardware and Software Installation Checklist

G1362A – RID

1. Record model number

1. Record serial number

1.Record Firmware

 Set up the detector


WARNING: Make sure that all plugs are removed from the ports and the appropriate capillaries
(inlet/outlet/recycling) are connected (maximum backpressure of 5 bar for flow cell).
 solvent inlet/outlet
 solvent recycle
 waste drain
 remote control (if used)
 analog output (if used)
 CAN
 Agilent LAN (see note*)

 Verify the operation of the detector

Diode
Balance Value

Issued: Oct 2008 – Rev. 6.5 Copyright  2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph

Hardware and Software Installation Checklist

G4218A – ELSD

1. Record model number

1. Record serial number

1.Record Firmware

 Set up the detector


 Verify that the instrument is the correct module for the regional voltage (100 V/115 V/230V). Do NOT use
instruments which have an incompatible voltage
 connect the exhaust line to a fume hood or lab exhaust line
 set the correct pressure at the regulator for the pressurized nitrogen line and connect it to the detector
 install the nebulizer
 install the nebulization chamber
 install the capillary connection from detector to HPLC
 install the AGP remote connection from detector to HPLC
 install the RS-232 connection from detector to Chemstation control PC (if applicable)
 connect the siphon overflow to the waste container
 optionally connect the analog signal cable
 optionally configure the detector in Chemstation

G4218A Checkout
 Switch on the instrument
 Prepare the instrument:
o If a new column is used, flush the system for 15 minutes AND
o set the detector temperature to 35 °C and equilibrate for 15 minutes
 Set up a method for the checkout sample (see user manual for test conditions)
 Run a test measurement
 Start a run
 Attach results to installation documentation

Issued: Oct 2008 – Rev. 6.5 Copyright  2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph

Hardware and Software Installation Checklist

G1364B/C/D – Fraction Collector

1. Record model number 2. Record model number 3. Record model number

1. Record serial number 2. Record serial number 3. Record serial number

1. Record Firmware 2. Record Firmware 3. Record Firmware

 Setup the fraction collector(s)


 solvent inlet/outlet
 waste drain
 remote control (if used)
 relay contacts (if installed)
 CAN
 Agilent LAN (see note*)

Issued: Oct 2008 – Rev. 6.5 Copyright  2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph

Hardware and Software Installation Checklist

G1390A – Universal Interface Box (UIB)


 Section NOT Applicable

1. Record model number

1. Record serial number

1.Record Firmware

 Setup the universal interface box


 CAN
 Analog In (if used)
 GPIO / Analog In (if used)
 Relay Contacts (if used)

Note: Agilent LAN connection is made between one (1) of the Agilent 1200 Series System modules and the
Agilent ChemStation PC. If a detector is installed connect the LAN to this module

G4208A Instant Pilot


 Section NOT Applicable

1. Record model number

1. Record serial number

1.Record Firmware

 Set up the control module


 Connect via CAN to available instrument module

G1323B Control Module


 Section NOT Applicable

1. Record model number 1. Record model number 1. Firmware

 Set up the Control


 Module
 Connect via CAN to available instrument module

Issued: Oct 2008 – Rev. 6.5 Copyright  2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph

Hardware and Software Installation Checklist

Software Installation (Bundled)


 Section NOT Applicable

Software Product:_________________________

Software Revision:________________________

 Start Windows
 Start ChemStations -> Add Licenses and add the license numbers. You find these numbers in the 'Software
Certificate and Registration Packet' envelopes.
 Configure TCP/IP if the instrument will be connected using LAN.
 If necessary install and Configure IO Libraries (LC/MS)
 If necessary install and configure Agilent Bootp Service for LAN connection.
 Use the Configuration Editor and configure your instruments.
 Install and Configure Software Add-Ons (if applicable)
 If present, install the PC Image software and create a backup using the Backup Solution User Guide as
reference.
 If present, install the Agilent LMD/LA (diagnostic utility) software.

Software Installation (Un-Bundled)

 Section NOT Applicable

Software Product:_________________________

Software Revision:_____________________

 Start Windows
 Install ChemStation using the ChemStation Installation guide for reference.
 Check if a printer driver is installed on the system
 Verify that the regional settings are set to English-US
 Install and Configure TCP/IP if the instrument will be connected using LAN.
 If necessary install and Configure IO Libraries (LC/MS).
 If necessary install and configure AGILENT Bootp Service if the instrument will be connected using LAN Use
the Configuration Editor
 Install the Agilent LMD/LA (diagnostic utility) software

Remote Advisor

 Ask the Customer is remote advisor is installed


 If yes in install the Data Source Software

Issued: Oct 2008 – Rev. 6.5 Copyright  2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph

Hardware and Software Installation Checklist

1200 LC Checkout
 Prepare:
1. Isocratic sample 6. detector(s)
2. pump(s) 7. fraction collector(s) (if present)
3. online degasser (s) (if present) 8. Valve(s) (if present)
4. injection device(s) 9. recording device
5. column compartment
 Start a run
 Attach results to installation documentation
 Fill out instrument logbook
 Verify connection to the Agilent diagnostic utility.

Issued: Oct 2008 – Rev. 6.5 Copyright  2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph

Hardware and Software Installation Checklist

Service Review

 Affix any reports generated to this Checklist


 Record in the instrument logbook date and time install was completed
 Explain Agilent warranty for Hardware, Software and return policy for the Instrument service Center
 Explain how to log an instrument service call
 Explain support services
 Explain how to use manuals
 Advise customer of additional training options
 Advise the customer of the following useful Agilent web sites:
e.g. 1200 Series Firmware (Downloads/patches)

Exclusive offers: www.agilent.com/chem/exclusiveoffers 

Library - http://www.agilent.com/chem/library

Education - www.agilent.com/chem/education

eSeminars - www.agilent.com/chem/eseminars

Software Status Bulletins and Patches - www.agilent.com/chem/techsupport

Supplies - www.agilent.com/chem/supplies

Tech Support - www.agilent.com/chem/techsupp


Note: Password is the Registration number for the software

 Sales Order Number


__________________________

 Service Order (SO) Number


__________________________

 Date completed
__________________________

 Customer Signature
__________________________

 Support Provider Signature


__________________________

*01200-90606* Part Number: 01200-90606 © Agilent Technologies, Inc. 2008

*01200-90606*
01200-90606 Edition October / 2008 Agilent Technologies
Printed in Germany Hewlett-Packard-Strasse 8

*01200-90606*
Issued: Oct 2008 – Rev. 6.5 Copyright Part
2008Number: 01200-90606 WAD Product Support © Select
Agilent Technologies Agilent Technologies,
76337Pages
Waldbronn, Inc.
Germany
for Required 2008
Modules

*01200-90606*
01200-90606 Edition 02 / 2008 Agilent Technologies
Printed in Germany Hewlett-Packard-Strasse 8

You might also like